IEC TS 62903:2023 is available as IEC TS 62903:2023 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC TS 62903:2023:
a ) establishes the free-field convergent spherical wave self-reciprocity method for ultrasonic transducer calibration,
b) establishes the measurement conditions and experimental procedure required to determine the transducer's electroacoustic parameters and acoustic output power using the self-reciprocity method,
c) establishes the criteria for checking the reciprocity of these transducers and the linear range of the focused field, and
d) provides guiding information for the assessment of the overall measurement uncertainties for radiation conductance.
This document is applicable to:
1) circular spherically curved concave focusing transducers without a centric hole working in the linear amplitude range,
2) measurements in the frequency range 0,5 MHz to 15 MHz, and
3) acoustic pressure amplitudes in the focused field within the linear amplitude range.
Characterization and sensitivity calibration of hydrophones using the reciprocity method are not addressed in this document but covered in IEC 62127-2 and IEC 60565-1.
IEC TS 62903:2023 cancels and replaces the first edition published in 2018. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) Several quantities are recognized as complex-valued quantities in the definitions and in the main text.
b) Annex I was added to provide typical measurement ranges and to provide example calibration results.

  • Technical specification
    53 pages
    English language
    sale 15% off

IEC TS 62736:2023 is available as IEC TS 62736:2023 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC TS 62736:2023 specifies requirements and methods for periodic testing of the quality of diagnostic medical ultrasound systems using reflection-mode (pulse-echo) imaging. Image measurement and interpretation workstations are included. This document includes minimum sets of such tests intended for frequent users of medical ultrasound systems, for quality assurance professionals in their organizations, or those hired from other quality-control and/or service-provider organizations. The procedures are for a wide range of more common diagnostic ultrasound systems, currently operating from 1 MHz to 40 MHz, although available phantoms meet the specifications only from 1 MHz to 23 MHz.
The tests are defined in three levels, with the simplest and most cost-effective performed most frequently:
- Level 1 comprises five quick tests/observations to be performed daily to monthly by those normally operating the systems.
- Level 2 includes one necessary test for all systems in addition to those of Level 1, two Level 1 tests performed more rigorously, two tests that are for special situations or equipment, and one that is just optional, included because it is highly developed. Level 2 tests are performed annually by those with meaningful quality assurance training.
- Level 3 extends the two special situations tests to all systems, adds one optional test and includes a periodic review of the QA programme.
Frequent distance-measurement accuracy tests are recommended in this document only for certain classes of position encoding that are not now known to be highly stable and without bias. QA in all dimensions is recommended in this document as the first test for such systems.
The test methodology is applicable for transducers operating in the 1 MHz to 23 MHz frequency range. The types of transducers used with these scanners include
a) electronic phased arrays,
b) linear arrays,
c) convex arrays,
d) mechanical transducers,
e) two-dimensional arrays operated in a 2D imaging mode,
f) transducers operating in 3D imaging mode for a limited number of sets of reconstructed 2D images, and
g) three-dimensional scanning transducers based on a combination of the above types.
All tests on scanners considered here evaluate basic pulse-echo techniques and might detect most failures in other modes. Dedicated Doppler systems, or other systems for detection of blood motion, are excluded from this scope as specialized equipment is required to test them. Such test equipment can be specific to the intended application of the Doppler system. This document includes definition of terms and specifies methods for measuring the maximum relative depth of penetration of real-time ultrasound B MODE scanners, though this penetration measure is listed as less frequently applied.
IEC TS 62736:2023 cancels and replaces the first edition published in 2016. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) expansion of the applicable types of transducers and the frequency range of application;
b) extension of test protocols and image assessments, including for very-low-echo spheres;
c) revision of phantom designs and their acoustic properties, consistent with the second edition of IEC TS 62791;
d) inclusion of luminance tests for system-image display consistency at scanner and remote monitors;
e) addition of special considerations for 3D-imaging transducers (Annex D) and workbook examples (Annex E).

  • Technical specification
    66 pages
    English language
    sale 15% off

IEC 62127-3:2022 is available as IEC 62127-3:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 62127-3:2022 specifies relevant hydrophone characteristics. This document is applicable to:
- hydrophones employing piezoelectric sensor elements, designed to measure the pulsed and continuous wave ultrasonic fields generated by ultrasonic equipment;
- hydrophones used for measurements made in water;
- hydrophones with or without an associated pre-amplifier.
IEC 62127-3:2022 cancels and replaces the first edition published in 2007 and Amendment 1:2013. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition.
a) The upper frequency limit of 40 MHz has been removed.
b) Hydrophone sensitivity definitions have been changed to recognize sensitivities as complex-valued quantities.
c) Procedures to determine the effective hydrophone size have been changed according to the rationale outlined in Annex B.
d) Requirements on the frequencies for which the effective hydrophone size shall be provided have been changed to achieve practicality for increased frequency bands.
e) The new Annex B and Annex C have been added.
f) Annex A has been updated to reflect the changes of the normative parts.

  • Standard
    62 pages
    English and French language
    sale 15% off

IEC TR 61390:2022 describes representative methods of measuring the performance of complete real-time medical ultrasonic imaging equipment in the frequency range 0,5 MHz to 23 MHz.
This document is relevant for real-time ultrasonic scanners based on the pulse-echo principle, for the types listed below:
- mechanical sector scanner;
- electronic phased array sector scanner;
- electronic linear array scanner;
-electronic curved array sector scanner;
- water-bath scanner based on any of the above four scanning mechanisms;
- plane-wave/fast imaging scanners;
- combination of several of the above methods (e.g. a linear array phased at the edge to produce a sector there to enlarge the field of view.
The methods described are based on evaluation of:
- sonograms obtained by scanning of tissue mimicking objects (phantoms);
- sonograms obtained by scanning of artificial, low- or highly reflective targets in suitable environments;
- parameters of the ultrasound field transmitted by the measured scanner.
This document does not relate to methods for measuring electrical parameters of the scanner’s electronic systems.
IEC TR 61390:2022 cancels and replaces the first edition published in 1996. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) Several additional phantom designs are included in the main body of the document;
b) Several additional transducer types are included in the Scope;
c) Methods of analysis are presented in new Annex B.

  • Technical report
    55 pages
    English language
    sale 15% off

IEC TS 62791:2022 defines terms and specifies methods for quantifying detailed imaging performance of real-time, ultrasound B-mode scanners. Detail is assessed by imaging phantoms containing small, low-echo spherical targets in a tissue-mimicking background and analysing sphere detectability. Specifications are given for phantom properties. In addition, procedures are described for acquiring images, conducting qualitative analysis of sphere detectability, and carrying out quantitative analysis by detecting sphere locations and computing their contrast-to-noise ratios. With appropriate choices in design, results can be applied, for example:
• to assess the relative ability of scanner configurations (scanner make and model, scan head and console settings) to delineate the boundary of a tumour or identify specific features of tumours;
• to choose scanner control settings, such as frequency or the number and location of transmit foci, which maximize spatial resolution;
• to detect defects in probes causing enhanced sidelobes and spurious echoes.
The types of transducers used with these scanners include:
a) phased arrays,
b) linear arrays,
c) convex arrays,
d) mechanical sector scanners,
e) 3-D probes operating in 2-D imaging mode, and
f) 3-D probes operating in 3-D imaging mode for a limited number of sets of reconstructed 2 D images.
The test methodology is applicable for transducers operating in the 1 MHz to 23 MHz frequency range.
IEC TS 62791:2022 cancels and replaces the first edition published in 2015. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition.
a) It introduces necessary corrections to the analysis methods; these have been published in the literature.
b) It increases the range of contrast levels of low-echo spheres in phantoms that meet this Technical Specification. Previous specification was -20 dB, but two additional levels, -6 dB and either -30 dB or, if possible, -40 dB, are now specified.
c) It includes a wider range of uses of the methodology, including testing the effectiveness of scanner pre-sets for specific clinical tasks and detecting flaws in transducers and in beamforming.
d) It decreases the manufacturing cost by decreasing phantoms' dimensions and numbers of low-echo, backscattering spheres embedded in each phantom.

  • Technical specification
    73 pages
    English language
    sale 15% off

IEC 62127-1:2022 is available as IEC 62127-1:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 62127-1:2022 specifies methods of use of calibrated hydrophones for the measurement in liquids of acoustic fields generated by ultrasonic medical equipment including bandwidth criteria and calibration frequency range requirements in dependence on the spectral content of the fields to be characterized.
This document:
- defines a group of acoustic parameters that can be measured on a physically sound basis;
- defines a second group of parameters that can be derived under certain assumptions from these measurements, and called derived intensity parameters;
- defines a measurement procedure that can be used for the determination of acoustic pressure parameters;
- defines the conditions under which the measurements of acoustic parameters can be made using calibrated hydrophones;
- defines procedures for correcting for limitations caused by the use of hydrophones with finite bandwidth and finite active element size, and for estimating the corresponding uncertainties.
IEC 62127-1:2022 cancels and replaces the first edition published in 2007 and Amendment 1:2013. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition.
a) The upper frequency limit of 40 MHz has been removed.
b) Hydrophone sensitivity definitions have been changed to recognize sensitivities as complex-valued quantities.
c) Procedures and requirements for narrow-band approximation and broadband measurements have been modified; details on waveform deconvolution have been added.
d) Procedures for spatial averaging correction have been amended.
e) Annex D, Annex E and bibliography have been updated to support the changes of the normative parts.

  • Standard
    211 pages
    English and French language
    sale 15% off

IEC 61689:2022 is available as IEC 61689:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61689:2022 is applicable to ultrasonic equipment designed for physiotherapy containing an ultrasonic transducer generating continuous or quasi-continuous (e.g. tone burst) wave ultrasound in the frequency range 0,5 MHz to 5 MHz. This document only relates to ultrasonic physiotherapy equipment employing a single plane non-focusing circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head. This document specifies:
- methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on reference testing methods;
- characteristics to be specified by manufacturers of ultrasonic physiotherapy equipment based on reference testing methods;
- guidelines for safety of the ultrasonic field generated by ultrasonic physiotherapy equipment;
- methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on routine testing methods;
- acceptance criteria for aspects of the output of ultrasonic physiotherapy equipment based on routine testing methods.
Therapeutic value and methods of use of ultrasonic physiotherapy equipment are not within the scope of this document.
Ultrasonic physiotherapy equipment using ultrasound in the frequency range from 20 kHz to 500 kHz is dealt with in IEC 63009.
IEC 61689:2022 cancels and replaces the third edition published in 2013. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition.
a) The requirement on water oxygen content is specified in 6.1.
b) Former recommendations in 6.2 have been changed to become requirements.
c) Several definitions in Clause 3 have been updated in line with other TC 87 documents.
d) The formerly informative Annex A has been changed to become normative, and now contains details on how conformance with IEC 60601-2-5 requirements is checked.
e) Annex D has been considerably shortened and reference to a now withdrawn regulatory document has been removed.

  • Standard
    129 pages
    English and French language
    sale 15% off

IEC 61828:2020
- provides definitions for the transmitted field characteristics of focusing and nonfocusing transducers for applications in medical ultrasound;
- relates these definitions to theoretical descriptions, design, and measurement of the transmitted fields of focusing transducers;
- gives measurement methods for obtaining defined field characteristics of focusing and nonfocusing transducers;
- specifies beam axis alignment methods appropriate for focusing and nonfocusing transducers.
IEC 61828:2021 relates to focusing ultrasonic transducers operating in the frequency range appropriate to medical ultrasound (0,5 MHz to 40 MHz) for both therapeutic and diagnostic applications. It shows how the characteristics of the transmitted field of transducers can be described from the point of view of design, as well as measured by someone with no prior knowledge of the construction details of a particular device. The transmitted ultrasound field for a specified excitation is measured by a hydrophone in either a standard test medium (for example, water) or in a given medium. This document applies only to media where the field behaviour is essentially like that in a fluid (i.e. where the influence of shear waves and elastic anisotropy is small), including soft tissues and tissue-mimicking gels. Any aspects of the field that affect their theoretical description or are important in design are also included. These definitions would have use in scientific communications, system design and description of the performance and safety of systems using these devices.
IEC 61828:2021 incorporates definitions from other related standards where possible, and supplies more specific terminology, both for defining focusing characteristics and for providing a basis for measurement of these characteristics.
IEC 61828:2021 cancels and replaces the first edition published in 2001. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) Clause 6 on Measurement procedures has been replaced by Clause 6: "Acoustic field measurement: equipment" and Clause 7: "Measurement procedure" and related definitions.
b) Reorganization of definitions and measurement section to accommodate specific sets of measurements for focusing, nonlinearity, beam axis alignment, beam area, beam maximum, numerical projection, plane wave, high intensity therapeutic ultrasound, multiple sources, spatial impulse response and compound plane waves. Clause 3 has been moved to Annex B.
c) The normative references have been updated and the Bibliography has been expanded from 8 to 40 references.
d) Twelve figures have been updated and seven new figures (B.1, B.3, B.7, B.10, B.11, B.12, B.13, B.14) have been added to facilitate measurements and be consistent with measurement terminology.
e) New measurements have been added for time delays, arrays, plane waves and spatial impulse response.
f) Annex A has been expanded to provide general guidance on pulsed waves, system responses, focusing gains and minimum beamwidth estimation.
g) New annexes have been added:
• Annex B (informative) Rationale for focusing and nonfocusing definitions
• Annex E (informative) Uncertainties;
• Annex F (informative) Transducer and hydrophone positioning systems;
• Annex G (informative) Planar scanning of a hydrophone to determine acoustic output power;
• Annex H (informative) Properties of water;
In addition, Annex A was reorganized and new Clauses A.1, A.5 and A.6 were added.
h) Guidelines for remaining within the manufacturer’s pressure and intensity hydrophone limits and the determination of the extent of nonlinearity in the field have been added.

  • Standard
    227 pages
    English and French language
    sale 15% off

IEC 63045:2020 is applicable to
– therapy equipment using extracorporeally induced non-focused or weakly focused pressure pulses;
– therapy equipment producing extracorporeally induced non-focused or weakly focused mechanical energy,
where the pressure pulses are released as single events of duration up to 25 µs.
This document does not apply to
– therapy equipment using focusing pressure pulse sources such as extracorporeal lithotripsy equipment;
– therapy equipment using other acoustic waveforms like physiotherapy equipment, low intensity ultrasound equipment and HIFU/HITU equipment.
This document specifies
– measurable parameters which are used in the declaration of the acoustic output of extracorporeal equipment producing a non-focused or weakly focused pressure pulse field,
– methods of measurement and characterization of non-focused or weakly focused pressure pulse fields.
This document has been developed for equipment intended for use in pressure pulse therapy, for example therapy of orthopaedic pain like shoulder pain, tennis elbow pain, heel spur pain, muscular trigger point therapy, lower back pain, etc. It is not intended to be used for extracorporeal lithotripsy equipment (as described in IEC 61846), physiotherapy equipment using other waveforms (as described in IEC 61689) and HIFU/HITU equipment (see IEC 60601 2-62 and IEC TR 62649).

  • Standard
    61 pages
    English language
    sale 15% off
  • Standard
    124 pages
    English and French language
    sale 15% off

IEC 60565-1:2020 specifies methods and procedures for free-field calibration of hydrophones, as well as individual electroacoustic transducers that can be used as hydrophones (receivers) and/or projectors (source transducers). Two general types of calibration are covered within this document: absolute calibration using the method of three-transducer spherical-wave reciprocity, and relative calibration by comparison with a reference device which has already been the subject of an absolute calibration.
The maximum frequency range of the methods specified in this document is from 200 Hz to 1 MHz. The lowest acoustic frequency of application will depend on a number of factors, and will typically be in the range 200 Hz to 5 kHz depending mainly on the dimensions of the chosen test facility, The highest frequency of application for the methods described here is 1 MHz.
Procedures for pressure hydrophone calibration at low frequencies can be found in IEC 60565 2 [1] . Procedures for hydrophone calibration at acoustic frequencies greater than 1 MHz are covered by IEC 62127-2 [2].
Excluded from the scope of this document are low-frequency pressure calibrations of hydrophones, which are described in IEC 60565-2 [1]. Also excluded are calibrations of digital hydrophones and systems, calibration of marine autonomous acoustic recorders, calibration of acoustic vector sensors such as particle velocity sensors and pressure gradient hydrophones, calibration of passive sonar arrays consisting of multiple hydrophones, and calibration of active sonar arrays consisting of projectors and hydrophones.
This document presents a description of the requirements for free-field calibration in terms of test facility, equipment and instrumentation, signal processing, and frequency limitations. A description of achievable uncertainty and rules for the presentation of the calibration data are provided. Also included are informative annexes that provide additional guidance on
• measurement of directional response of a hydrophone or projector,
• measurement of electrical impedance of hydrophones and projectors,
• electrical loading corrections,
• acoustic far-field criteria in underwater acoustic calibration,
• pulsed techniques in free-field calibrations,
• assessment of uncertainty in the free-field calibration of hydrophones and projectors,
• derivation of the formulae for three-transducer spherical-wave reciprocity calibrations,
• calibration using travelling-wave tubes,
• calibration of hydrophones using optical interferometry, and
• calibrations in reverberant water tanks using continuous signals.
IEC 60565-1:2020 together with IEC 60565-2:2019, cancels and replaces the second edition of IEC 60565 published in 2006. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
1) removal of all descriptions of methods for pressure calibrations of hydrophones – these are now included in Part 2;
2) removal of the derivations of formulae for free-field reciprocity calibration (both amplitude sensitivity and phase sensitivity) and placement of these into an informative annex;
3) inclusion within the scope of the calibration of the transmitting response of individual source transducers and hydrophones (but not sonar arrays);
4) re-ordering of the sections within the document such that the more general procedures for calibration such as guidance on obtaining conditions of acoustic free-field, far-field, and steady-state, appear before the descriptions of procedures for absolute or relative calibrations;
5) revision of informative Annex A to include guidance on measurement of directional response of a hydrophone or projector;
6) addition of a new informative Annex B on measurement of electrical impedance of hydrophones and projectors;
7) revision of the previous informative annex on electrical loading corrections to include corrections to account for e

  • Standard
    183 pages
    English and French language
    sale 15% off

IEC TS 63081:2019:
• defines key quantities relevant to ultrasonic materials characterization;
• specifies methods for direct measurement of many key ultrasonic materials parameters.
This document is applicable to all measurements of properties of passive acoustic materials under drive conditions that are not subject to nonlinear acoustic propagation. Whilst there are materials properties that may be of interest in a nonlinear drive regime, these are currently outside the scope of this document.

  • Technical specification
    33 pages
    English language
    sale 15% off

IEC 60565-2:2019 specifies the methods for low frequency pressure calibration of hydrophones at frequencies from 0,01 Hz to several kilohertz depending on calibration method.
IEC 60565-2:2019, together with IEC 60565-1, replaces the second edition of IEC 60565 published in 2006. This edition constitutes a technical revision.
IEC 60565-2:2019 includes the following significant technical changes with respect to the previous edition.
1) IEC 60565 has been divided into two parts:
• Part 1: Procedures for free-field calibration;
• Part 2: Procedures for low frequency pressure calibration (this document).
2) A relative calibration method has been added to Clause 8: Calibration by piezoelectric compensation.
3) A relative calibration method has been added to Clause 11: Calibration by vibrating column.
4) Clause 12: Calibration by static pressure transducer, has been added.
5) Annex A: Equivalent circuit of the excitation system for calibration with a vibrating column, has been deleted.
6) Subclauses 9.6, 9.7 and 9.8 have been moved to form a new Annex A: Advanced acoustic coupler calibration methods.

  • Standard
    108 pages
    English and French language
    sale 15% off

IEC 63009:2019 is applicable to ultrasonic equipment designed for physiotherapy containing an ultrasonic transducer generating ultrasound in the frequency range 20 kHz to 500 kHz.
This document only relates to ultrasonic physiotherapy equipment employing a single plane non-focusing circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head.
This document specifies:
• methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on reference testing methods;
• characteristics to be specified by manufacturers of ultrasonic physiotherapy equipment;
• methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on routine testing methods;
• acceptance criteria for aspects of the output of ultrasonic physiotherapy equipment.
The therapeutic value and methods of use of ultrasonic physiotherapy equipment are not within the scope of this document.
Excluded equipment includes, but is not limited to:
• equipment in which ultrasound waves are intended to destroy conglomerates (for example stones in the kidneys or the bladder) or tissue of any type;
• equipment in which a tool is driven by ultrasound (for example surgical scalpels, phacoemulsifiers, dental scalers or intracorporeal lithotripters);
• equipment in which ultrasound waves are intended to sensitize tissue to further therapies (for example radiation or chemotherapy);
• equipment in which ultrasound waves are intended to treat cancerous (i.e., malignant) or pre-cancerous tissue, or benign masses, such as High Intensity Focused Ultrasound (HIFU) or High Intensity Therapeutic Ultrasound (HITU).

  • Standard
    73 pages
    English and French language
    sale 15% off

IEC TS 63070:2019 is applicable to ultrasonic equipment designed for the medical field of application. It covers both diagnostic and therapeutic (physiotherapy and HITU) equipment.
This document describes transducer evaluation by the infrared imaging technique using a split TMM-phantom for qualitative and quantitative estimation of temperature distributions in tissue-mimicking material, resulting from absorption of ultrasound and from heating of the transducer itself.
This document also describes a method to measure transducer-surface temperature, while the transducer is driven under the still-air condition.

  • Technical specification
    26 pages
    English language
    sale 15% off

IEC TS 63001:2019 provides a technique of measurement and evaluation of ultrasound in liquids for use in cleaning devices and equipment. It specifies
• the cavitation measurement at 2,25 f0 in the frequency range 20 kHz to 150 kHz, and
• the cavitation measurement by extraction of broadband spectral components in the frequency range 10 kHz to 5 MHz.
IEC TS 63001:2019 covers the measurement and evaluation of the cavitation, but not its secondary effects (cleaning results, sonochemical effects, etc.).

  • Technical specification
    29 pages
    English language
    sale 15% off

IEC TS 62903:2018, which is a Technical Specification,
a) establishes the free-field convergent spherical wave self-reciprocity method for ultrasonic transducer calibration,
b) establishes the measurement conditions and experimental procedure required to determine the transducer's electroacoustic parameters and acoustic output power using the self-reciprocity method,
c) establishes the criteria for checking the reciprocity of these transducers and the linear range of the focused field, and
d) provides guiding information for the assessment of the overall measurement uncertainties for radiation conductance.
This document is applicable to:
i) circular spherically curved concave focusing transducers without a centric hole working in the linear amplitude range,
ii) measurements in the frequency range 0,5 MHz to 15 MHz, and
iii) acoustic pressure amplitudes in the focused field within the linear amplitude range.

  • Technical specification
    47 pages
    English language
    sale 15% off
  • Technical report
    31 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC TS 62462:2017(E) describes methods meant to assist users of ultrasound physiotherapy systems in checking the performance of such systems. It is applicable primarily to physiotherapists, general medical practitioners, chiropractors, osteopaths, beauty therapists, sports professionals, biomedical engineers, medical physicists, medical device service agents, commercial testers, test houses or manufacturers. This new edition includes the following significant technical change with respect to the previous edition: it includes a novel method for periodic testing regarding possible changes of the effective radiating area using thermochromic absorbers in a new Annex E.

  • Technical specification
    30 pages
    English language
    sale 15% off

IEC 60500:2017 specifies the relevant characteristics and properties of hydrophones in the frequency range 1 Hz to 500 kHz, and specifies how to report these characteristics. Provides guidance on the choice of a hydrophone with appropriate performance for a specific application. This standard is applicable to:
- hydrophones employing piezoelectric sensor elements, designed to respond to sound pressure in water and measure underwater acoustical signals;
- hydrophones with or without an integral pre-amplifier.
This new edition includes the following significant technical changes with respect to the previous edition:
- the format and scope of IEC 60500 have been changed to be compatible with other IEC standards;
- the upper limit of the frequency range of hydrophones has been expanded.

  • Standard
    47 pages
    English and French language
    sale 15% off

IEC 62127-2:2007 specifies absolute hydrophone calibration methods and relative (comparative) hydrophone calibration methods. This standard is applicable to:
- hydrophones used for measurements made in water and in the ultrasonic frequency range up to 40 MHz;
- hydrophones employing circular piezoelectric sensor elements, designed to measure the pulsed wave and continuous wave ultrasonic fields generated by ultrasonic equipment;
- and hydrophones with or without a hydrophone pre-amplifier. IEC 62127-1, IEC 62127-2 and IEC 62127-3 are being published simultaneously. Together these cancel and replace IEC 60866:1987, IEC 61101:1991, IEC 61102:1991, IEC 61220:1993 and IEC 62092:2001. The contents of the corrigendum of August 2008 have been included in this copy. The French version of this standard has not been voted upon.

  • Standard
    172 pages
    English and French language
    sale 15% off

IEC TS 62736:2016(E) specifies requirements and methods for periodic testing of the quality of diagnostic medical ultrasound systems with linear array, curved linear array, single element, annular array, phased array, matrix linear array transducers and two-dimensional arrays.

  • Technical specification
    38 pages
    English language
    sale 15% off

IEC TS 62791:2015(E) defines terms and specifies methods for quantifying the imaging performance of real-time, ultrasound B-mode scanners. The test methodology is applicable for transducers operating in the 2 MHz to 15 MHz frequency range.

  • Technical specification
    60 pages
    English language
    sale 15% off

IEC TS 62556:2014 is a technical specification applicable to high intensity therapeutic ultrasound (HITU) devices. It specifies the relevant parameters for quantifying the field; the measurement methods at relatively low output levels and methodology for extrapolating these to higher therapeutic level fields; some considerations of sidelobes and pre-focal maxima; and parameters relevant to HITU transducers of different construction and geometry, including non-focusing, focusing with or without lenses, collimated, diverging and convergent transducers, multi-element transducers, scanning transducers and multiple sources. This technical specification is intended to support the ultrasonic measurement requirements given in IEC 60601-2-62. These specifications would have use in quality assurance, safety testing, and the standardization of communications regarding the clinical performance of HITU systems. Where possible, this technical specification incorporates specifications from other related standards. This technical specification does not apply to the following types of devices, which are covered by other standards: lithotripters (see IEC 61846); surgical equipment (see IEC 61847); and physiotherapy devices (see IEC 61689).

  • Technical specification
    93 pages
    English language
    sale 15% off

IEC 62555:2013 establishes general principles relevant to HITU fields for the use of radiation force balances in which an obstacle (target) intercepts the sound field to be measured; specifies a calorimetric method of determining the total emitted acoustic power of ultrasonic transducers based on the measurement of thermal expansion of a fluid-filled target; specifies requirements related to the statement of electrical power characteristics of ultrasonic transducers; provides guidance related to the avoidance of acoustic cavitation during measurement; provides guidance related to the measurement of HITU transducers of different construction and geometry, including collimated, diverging and convergent transducers, and multi-element transducers; provides guidance on the choice of the most appropriate measurement method; and provides information on assessment of overall measurement uncertainties. This International Standard is applicable to the measurement of ultrasonic power generated by HITU equipment up to 500 W in the frequency range from 0,5 MHz to 5 MHz. HITU equipment may generate convergent, collimated or divergent fields. For frequencies less than 500 kHz, no validations exist and the user should assess the uncertainties of the power measurement and measurement system at the frequencies of operation. This International Standard does not apply to ultrasound equipment used for physiotherapy, for lithotripsy for general pain relief.

  • Standard
    113 pages
    English and French language
    sale 15% off

IEC/TR 62799:2013(E), which is a technical report, provides background information for users of IEC 62359 to understand the relative merits of several of the potential replacements for the thermal index as described in IEC 60601-2-37 and IEC 62359. The report discusses parameters related to thermal aspects of diagnostic ultrasonic fields; and methods for the determination of an exposure parameter relating to temperature rise in theoretical tissue-equivalent models, resulting from absorption of ultrasound. The report is intended to be used:
- by those involved in the development and maintenance of IEC 62359;
- manufacturers of medical electrical equipment for risk assessment;
- and health care regulatory authorities, test houses and other organizations responsible for implementing standards for medical electrical equipment.

  • Technical report
    44 pages
    English language
    sale 15% off

IEC 61689:2013 is applicable to ultrasonic equipment designed for physiotherapy containing an ultrasonic transducer generating continuous or quasi-continuous wave ultrasound in the frequency range 0,5 MHz to 5 MHz. This standard only relates to ultrasonic physiotherapy equipment employing a single plane non-focusing circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head. This standard specifies:
- methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on reference testing methods;
- characteristics to be specified by manufacturers of ultrasonic physiotherapy equipment based on reference testing methods;
- guidelines for safety of the ultrasonic field generated by ultrasonic physiotherapy equipment;
- methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on routine testing methods;
- and acceptance criteria for aspects of the output of ultrasonic physiotherapy equipment based on routine testing methods. Therapeutic value and methods of use of ultrasonic physiotherapy equipment are not covered by the scope of this standard. This third edition cancels and replaces the second edition published in 2007. It constitutes a technical revision which includes the following significant technical changes with respect to the previous edition:
- restriction introduced of 0,2 W/cm2 effective intensity during hydrophone measurements for treatment heads with ka≤20, to limit the likelihood of cavitation;
- change in the factor Fac, to determine the effective radiating area, from 1,354 to 1,333;
- and change to SI units for terms and definitions.
This publication is to be read in conjunction with IEC 60601-2-5:2009.

  • Standard
    124 pages
    English and French language
    sale 15% off

IEC 61161:2013 specifies a method of determining the total emitted acoustic power of ultrasonic transducers based on the use of a radiation force balance; establishes general principles for the use of radiation force balances in which an obstacle (target) intercepts the sound field to be measured; establishes limitations of the radiation force method related to cavitation and temperature rise; establishes quantitative limitations of the radiation force method in relation to diverging and focused beams; provides information on estimating the acoustic power for diverging and focused beams using the radiation force method; and provides information on assessment of overall measurement uncertainties. This International Standard is applicable to the measurement of ultrasonic power up to 1 W based on the use of a radiation force balance in the frequency range from 0,5 MHz to 25 MHz; the measurement of ultrasonic power up to 20 W based on the use of a radiation force balance in the frequency range 0,75 MHz to 5 MHz; the measurement of total ultrasonic power in well-collimated, diverging and focused ultrasonic fields; and to the use of radiation force balances of the gravimetric type or force feedback type. This third edition cancels and replaces the second edition published in 2006. It constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- whereas the second edition tacitly dealt with circular transducers only, the present edition as far as possible deals with both circular and rectangular transducers, including a number of symbols for rectangular transducers;
- attention is paid to focused cases and the influence of scanning has been added;
- the method of calibrating the radiation force balance now depends on whether the set-up is used as a primary or as secondary measurement tool;
- and Annex B (basic formulae) has been updated and in Annex C the buoyancy change method is mentioned (see also future IEC 62555).

  • Standard
    100 pages
    English and French language
    sale 15% off

IEC/TR 62781:2012(E), which is a technical report, is applicable to all measurements of ultrasonic fields where water is the transmission medium. Water provides an inexpensive and readily available medium with characteristic acoustic impedance comparable to biological tissue. However, basic tap water is far from optimum for ultrasonic measurement as it contains many dissolved, absorbed and suspended contaminants. Measurements can be affected in many ways by these impurities. This technical aims to provide a unified source to establish a water treatment process for ultrasonic measurements. The quality and treatment methods for water used within a radiation force balance (RFB) may be different from that required for hydrophone based acoustic measurements. Chemical based methods of water treatment (e.g. algaecides) may be appropriate for these applications. However, in this document, chemical means are noted but appropriately discouraged for acoustic pressure/intensity measurements. This Technical Report describes methods for:
- degassing water to be used in ultrasonic measurements;
- to decrease the ionic content of water to be used in ultrasonic measurements;
- to decrease the biological content of water to be used in ultrasonic measurements; and
- to reduce the suspended particulate content of water to be used in ultrasonic measurements.

  • Technical report
    21 pages
    English language
    sale 15% off

IEC/TS 62558:2011(E) specifies essential characteristics of a phantom and method for the measurement of void-detectability ratio for medical ultrasound systems and related transducers. It is restricted to the aspect of long-term reproducibility of testing results. Medical diagnostic ultrasound systems and related transducers need periodic testing as the quality of medical decisions based on ultrasonic images may decrease over time due to progressive degradation of essential systems characteristics. The TMM (Tissue Mimicking Material) phantom is intended to be used to measure and to enable documentation of changes in void-detectability ratio in periodic tests over years of use. This technical specification establishes:
- Important characteristics and requirements for a TMM 3D artificial cyst phantom using anechoic voids
- A design example of a 3D artificial cyst phantom, the necessary test equipment and use of relevant computer software algorithms.
This technical specification is currently applicable for linear array transducers. A uniformity test prior to void-detectability ratio (VDR) measurement is recommended.

  • Technical specification
    49 pages
    English language
    sale 15% off

IEC 62359:2010 is applicable to medical diagnostic ultrasound fields. It establishes:
- parameters related to thermal and non-thermal exposure aspects of diagnostic ultrasonic fields;
- methods for the determination of an exposure parameter relating to temperature rise in theoretical tissue-equivalent models, resulting from absorption of ultrasound;
- methods for the determination of an exposure parameter appropriate to certain non-thermal effects.
This second edition cancels and replaces the first edition, published in 2005. It constitutes a technical revision. Major changes with respect to the previous edition include the following:
- The methods of determination set out in the first edition of this standard were based on those contained in the American standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment (ODS) and were intended to yield identical results. While this second edition also follows the ODS in principal and uses the same basic formulae and assumptions (see Annex A), it contains a few significant modifications which deviate from the ODS.
- One of the primary issues dealt with in preparing this second edition of IEC 62359 was "missing" TI equations. In Edition 1 there were not enough equations to make complete "at-surface" and "below-surface" summations for TIS and TIB in combined-operating modes. Thus major changes with respect to the previous edition are related to the introduction of new calculations of thermal indices to take into account both "at-surface" and "below-surface" thermal effects. The contents of the corrigendum of March 2011 have been included in this copy.

  • Standard
    136 pages
    English and French language
    sale 15% off

IEC/TR 62649:2010(E) is relevant to the measurement and specification of ultrasound fields intended for medical therapeutic purposes. Lithotripsy and physiotherapy are excluded, since there are existing International Standards for these applications. IEC/TR 62649:2010 establishes:
- topics where there is a consensus that the development of International Standards would benefit the industries and/or patients;
- topics where the writing of standards should start immediately;
- topics where the writing of technical specifications should start immediately in order to gain practical experience and establish consensus prior to standardisation;
- topics which require future standardisation but where further research is required before initiating the writing of standards or technical specifications. IEC/TR 62649:2010 addresses primarily the requirements for measurement standards related to high intensity therapeutic ultrasound (HITU) [also known as high intensity focused ultrasound (HIFU)] fields which are both high intensity and focused and where the main mechanism for action is thermal. However, aspects of the discussion, conclusions and any resulting standards or technical specifications may also be relevant to therapeutic applications which are either focused or high intensity or where the main mechanism is not thermal. Scientific literature has been reviewed and responses to a questionnaire which was sent to experts around the world are reported.

  • Technical report
    93 pages
    English language
    sale 15% off

IEC 61391-2:2010 defines terms and specifies methods for measuring the maximum depth of penetration and the local dynamic range of real time ultrasound B MODE scanners. The types of transducers used with these scanners include:
- mechanical probes;
- electronic phased arrays;
- linear arrays;
- curved arrays;
- two-dimensional arrays;
- three-dimensional scanning probes based on a combination of the above types.
All scanners considered are based on pulse-echo techniques. The test methodology is applicable for transducers operating in the 1 MHz to 15 MHz frequency range operating both in fundamental mode and in harmonic modes that extend to 15 MHz. However, testing of harmonic modes above 15 MHz is not covered by this standard.

  • Standard
    74 pages
    English and French language
    sale 15% off

This Technical Specification describes means to allow "attenuated" acoustic quantities to be calculated under conditions where the associated acoustic measurements, made in water using standard procedures, may be accompanied by significant finite-amplitude effects. This Technical Specification establishes: - the general concept of the limits of applicability of acoustic measurements in water resulting from finite-amplitude acoustic effects; - a method to ensure that measurements are made under quasi-linear conditions in order to minimise finite-amplitude effects; - the definition of an acoustic quantity appropriate for establishing quasi-linear conditions; - a threshold value for the acoustic quantity as an upper limit for quasi-linear conditions; - a method for the estimation of attenuated acoustic quantities under conditions of nonlinear propagation in water.

  • Technical specification
    31 pages
    English language
    sale 15% off

IEC 62127-1:2007 specifies methods of use of calibrated hydrophones for the measurement in liquids of acoustic fields generated by ultrasonic medical equipment operating in the frequency range up to 40 MHz. The objectives of this standard are:
- to define a group of acoustic parameters that can be measured on a physically sound basis;
- to define a second group of parameters that can be derived under certain assumptions from these measurements, and called derived intensity parameters;
- to define a measurement procedure that may be used for the determination of acoustic pressure parameters;
- to define the conditions under which the measurements of acoustic parameters can be made in the frequency range up to 40 MHz using calibrated hydrophones;
- and to define procedures for correcting, for limitations caused by the use of hydrophones with finite bandwidth and finite active element size. IEC 62127-1, IEC 62127-2 and IEC 62127-3 are being published simultaneously. Together these cancel and replace IEC 60866:1987, IEC 61101:1991, IEC 61102:1991, IEC 61220:1993 and IEC 62092:2001. The contents of the corrigendum of August 2008 have been included in this copy. The French version of this standard has not been voted upon.

  • Standard
    166 pages
    English and French language
    sale 15% off

IEC 62127-3:2007 specifies relevant hydrophone characteristics. This standard is applicable to hydrophones employing piezoelectric sensor elements, designed to measure the pulsed and continuous wave ultrasonic fields generated by ultrasonic equipment; hydrophones used for measurements made in water; and hydrophones with or without an associated pre-amplifier. IEC 62127-1, IEC 62127-2 and IEC 62127-3 are being published simultaneously. Together these cancel and replace IEC 60866:1987, IEC 61101:1991, IEC 61102:1991, IEC 61220:1993 and IEC 62092:2001. The French version of this standard has not been voted upon.

  • Standard
    41 pages
    English and French language
    sale 15% off

IEC 61157:2007 specifies a standard means and format for the reporting of the acoustic output of medical diagnostic ultrasonic equipment. The numerical values for reporting purposes represent the average values for the maximum output conditions for a given discrete- or combined-operating mode and are derived from measurements made in water. Intensity parameters are specified in this standard, but these are regarded as derived quantities that are meaningful only under certain assumptions related to the ultrasonic field being measured. This International Standard is applicable to medical diagnostic ultrasonic equipment. It provides a set of traceable acoustic parameters describing the acoustic fields. It defines a standard means and format for the reporting of the acoustic output information. It also describes a reduced dataset recommended for equipment generating low acoustic output levels. This second edition cancels and replaces the first edition published in 1992. This edition constitutes a minor revision. The changes with respect to the previous edition are listed hereafter:
- maintenance on this standard and the referenced standards IEC 61161 and IEC 62127-1;
- and a clause on compliance has been added. The contents of the corrigendum of August 2008 have been included in this copy. The French version of this standard has not been voted upon.

  • Standard
    66 pages
    English and French language
    sale 15% off

Describes methods of calibrating the spatial measurement facilities and point-spread function of ultrasonic imaging equipment in the ultrasonic frequency range 0,5 MHz to 15 MHz. This standard is relevant for ultrasonic scanners based on the pulse echo principle of the types listed below: - mechanical sector scanners; - electronic phased-array sector scanners; - electronic linear-array scanners; - electronic curved-array sector scanners; - water bath scanners based on any of the above four scanning mechanisms; - 3D-volume reconstruction systems.

  • Standard
    91 pages
    English and French language
    sale 15% off

This Technical Specification is applicable to medical diagnostic ultrasonic fields. This Technical Specification establishes: - the important characteristics and terminology for test objects containing an embedded thermal sensor to determine ultrasonic heating; - requirements for defining particular types of test object for specific applications; - examples of design features which will ensure that the performance specifications are achieved; - validation procedures; - general test methods for using test objects to determine standardised temperature rises.

  • Technical specification
    75 pages
    English and French language
    sale 15% off

Specifies parameters for a flow Doppler test object representing a blood vessel of known diameter at a certain depth in human tissue, carrying a steady flow. Establishes a flow Doppler test object which can be used to assess various aspects of the performance of Doppler diagnostic equipment.

  • Standard
    36 pages
    English language
    sale 15% off

This International Standard - provides definitions for the transmitted field characteristics of focusing transducers for applications in medical ultrasound; - relates these definitions to theoretical descriptions, design, and measurement of the transmitted fields of focusing transducers; - gives measurement methods for obtaining defined characteristics of focusing transducers; - specifies beam axis alignment methods appropriate for focusing transducers. This International Standard relates to focusing ultrasonic transducers operating in the frequency range appropriate to medical ultrasound (0,5 MHz to 40 MHz) for both therapeutic and diagnostic applications. It shows how the characteristics of the transmitted field of transducers may be described from the point of view of design, as well as measured by someone with no prior knowledge of the construction details of a particular device. The radiated ultrasound field for a specified excitation is measured by a hydrophone in either a standard test medium (for example, water) or in a given medium. The standard applies only to media where the field behaviour is essentially like that in a fluid (i.e. where the influence of shear waves and elastic anisotropy is small), including soft tissues and tissue-mimicking gels. Any aspects of the field that affect their theoretical description or are important in design are also included. These definitions would have use in scientific communications, system design and description of the performance and safety of systems using these devices.

  • Standard
    103 pages
    English and French language
    sale 15% off

This technical specification describes: - test methods for measuring the performance of pulsed Doppler ultrasound systems; - Doppler test objects for carrying out these tests; and applies to - tests made on an overall pulsed Doppler ultrasound system, a system which is not disassembled or disconnected; - tests made on pulsed Doppler ultrasound systems whether they are stand-alone or as part of another ultrasound equipment.

  • Technical specification
    32 pages
    English language
    sale 15% off