11.040.50 - Radiographic equipment
ICS 11.040.50 Details
Radiographic equipment
Radiologische und sonstige Diagnosegerate
Appareils de radiographie
Radiografska oprema
General Information
e-Library Subscription
Create subscription and get permanent access to documents within 11.040.50 - Radiographic equipment
Currently subscription includes documents marked with .We are working on making all documents available within the subscription.
IEC 61267:2025 applies to test procedures which, for the determination of characteristics of systems or components of medical diagnostic X-ray equipment, require well-defined X-ray radiation conditions. This document deals with methods for generating X-ray radiation conditions which can be used under test conditions typically found in test laboratories or in manufacturing facilities for the determination of characteristics of medical diagnostic X-ray equipment.
IEC 61267:2025 cancels and replaces the second edition published 2005. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) removing former Annex C “Measurement of the practical peak voltage”;
b) inserting informative “Tabulated values for the squared signal-to-noise ratio per air kerma (SNR2in)” and normative “Additional X-ray radiation conditions as used in mammography and determination of the corresponding nominal aluminium half-value layers”;
c) revision of X-ray radiation conditions;
d) new method for verification of X-ray radiation conditions;
e) change of term definitions.
- Standard44 pagesEnglish languagesale 15% off
- Standard47 pagesFrench languagesale 15% off
- Standard91 pagesEnglish and French languagesale 15% off
IEC 62570:2025 applies to medical devices and other items that are anticipated to enter the magnetic resonance (MR) environment.
This document specifies the marking of items anticipated to enter the MR environment by means of terms and icons, and recommends information that should be included in the labeling.
MR image artifacts are not in the scope of the mandatory portions of this practice because they do not present a direct safety issue resulting from specific characteristics of the MR examination.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62570:2025 applies to medical devices and other items that are anticipated to enter the magnetic resonance (MR) environment. This document specifies the marking of items anticipated to enter the MR environment by means of terms and icons, and recommends information that should be included in the labeling. MR image artifacts are not in the scope of the mandatory portions of this practice because they do not present a direct safety issue resulting from specific characteristics of the MR examination.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 63322:2025 establishes security requirements of ME EQUIPMENT using high-activity SEALED RADIOACTIVE SOURCES, directly or indirectly, for medical treatment and other clinical procedures. ME EQUIPMENT containing SEALED RADIOACTIVE SOURCES that are defined as Category 1, 2 and 3 RADIOACTIVE SOURCES by IAEA are subject to this document. The object of this document is to specify requirements for the security of ME EQUIPMENT containing high-activity SEALED RADIOACTIVE SOURCES with the aim to minimize the risk of unauthorized access to the contained SEALED RADIOACTIVE SOURCES, and to serve as the basis for other standards. This document contains requirements for the MANUFACTURER of the ME EQUIPMENT and, separately, for the RESPONSIBLE ORGANIZATION regarding security at the location during use and storage. The requirements of this document apply when the SEALED RADIOACTIVE SOURCES are contained in the ME EQUIPMENT, i.e. from the time when the SEALED RADIOACTIVE SOURCES are inserted into the ME EQUIPMENT, during the INTENDED USE and when the ME EQUIPMENT is not being used for its INTENDED USE or taken out of regular use, until the equipment is being decommissioned, i.e. until all SEALED RADIOACTIVE SOURCES are permanently removed from the equipment.
- Standard68 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62570:2025 applies to medical devices and other items that are anticipated to enter the magnetic resonance (MR) environment.
This document specifies the marking of items anticipated to enter the MR environment by means of terms and icons, and recommends information that should be included in the labeling.
MR image artifacts are not in the scope of the mandatory portions of this practice because they do not present a direct safety issue resulting from specific characteristics of the MR examination.
- Standard18 pagesEnglish languagesale 15% off
- Standard27 pagesFrench languagesale 15% off
- Standard45 pagesEnglish and French languagesale 15% off
IEC 63322:2025 establishes security requirements of ME EQUIPMENT using high-activity SEALED RADIOACTIVE SOURCES, directly or indirectly, for medical treatment and other clinical procedures. ME EQUIPMENT containing SEALED RADIOACTIVE SOURCES that are defined as Category 1, 2 and 3 RADIOACTIVE SOURCES by IAEA are subject to this document. The object of this document is to specify requirements for the security of ME EQUIPMENT containing high-activity SEALED RADIOACTIVE SOURCES with the aim to minimize the risk of unauthorized access to the contained SEALED RADIOACTIVE SOURCES, and to serve as the basis for other standards. This document contains requirements for the MANUFACTURER of the ME EQUIPMENT and, separately, for the RESPONSIBLE ORGANIZATION regarding security at the location during use and storage. The requirements of this document apply when the SEALED RADIOACTIVE SOURCES are contained in the ME EQUIPMENT, i.e. from the time when the SEALED RADIOACTIVE SOURCES are inserted into the ME EQUIPMENT, during the INTENDED USE and when the ME EQUIPMENT is not being used for its INTENDED USE or taken out of regular use, until the equipment is being decommissioned, i.e. until all SEALED RADIOACTIVE SOURCES are permanently removed from the equipment.
- Standard68 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 63322:2025 establishes security requirements of ME EQUIPMENT using high-activity SEALED RADIOACTIVE SOURCES, directly or indirectly, for medical treatment and other clinical procedures. ME EQUIPMENT containing SEALED RADIOACTIVE SOURCES that are defined as Category 1, 2 and 3 RADIOACTIVE SOURCES by IAEA are subject to this document. The object of this document is to specify requirements for the security of ME EQUIPMENT containing high-activity SEALED RADIOACTIVE SOURCES with the aim to minimize the risk of unauthorized access to the contained SEALED RADIOACTIVE SOURCES, and to serve as the basis for other standards. This document contains requirements for the MANUFACTURER of the ME EQUIPMENT and, separately, for the RESPONSIBLE ORGANIZATION regarding security at the location during use and storage. The requirements of this document apply when the SEALED RADIOACTIVE SOURCES are contained in the ME EQUIPMENT, i.e. from the time when the SEALED RADIOACTIVE SOURCES are inserted into the ME EQUIPMENT, during the INTENDED USE and when the ME EQUIPMENT is not being used for its INTENDED USE or taken out of regular use, until the equipment is being decommissioned, i.e. until all SEALED RADIOACTIVE SOURCES are permanently removed from the equipment.
- Standard66 pagesEnglish languagesale 15% off
- Standard70 pagesFrench languagesale 15% off
- Standard136 pagesEnglish and French languagesale 15% off
IEC 60601-2-68:2025 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of X-ray based IMAGE-GUIDED RADIOTHERAPY equipment for use with EXTERNAL BEAM EQUIPMENT (EBE). This document covers safety aspects of kilovoltage (kV) and megavoltage (MV) X-ray imaging devices integrated in a specified geometrical relationship with EBE for the purpose of IGRT. It covers aspects of communication and relationships between the EXTERNAL BEAM EQUIPMENT and X-ray imaging devices, attached or not directly attached to, but in the same RADIATION shielded area as, and dedicated for use only with, the EXTERNAL BEAM EQUIPMENT. This document deals with equipment for OFFLINE X-IGRT, ONLINE X-IGRT and REAL-TIME X-IGRT. It covers procedures to reduce the risk of over-reliance on the X-IGRT EBE SYSTEM. For example, in the case of ONLINE X-IGRT, the MANUFACTURER will provide an interactive interface for user interaction with the correction suggested by the system. This document does not apply to CT SCANNERS, X-RAY EQUIPMENT for RADIOGRAPHY, and X-RAY EQUIPMENT for RADIOSCOPY, which are not intended for use for IGRT. Requirements that are being tested according to another standard can be identified by the manufacturer and if equivalent do not require retesting, instead evidence can refer to the CT SCANNER, X-RAY EQUIPMENT for RADIOGRAPHY, or X-RAY EQUIPMENT for RADIOSCOPY EQUIPMENT manufacturer's providing compliance statements or test reports. If the X-IGRT EQUIPMENT is combined with an MEE, any requirement that is the same for the X-IGRT EQUIPMENT and the MEE, such as a PATIENT POSITIONER, is not required to be tested twice, but can be accepted as tested by the MEE. This document applies for X-ray equipment for radiography, radioscopy, and COMPUTER tomography used for IGRT. If a clause or subclause is specifically intended to be applicable to X-IGRT EBE SYSTEMS, the content of that clause or subclause will say so. Where that is not the case, the clause or subclause applies only to X-IGRT EQUIPMENT.
This document, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the MANUFACTURER and some installation aspects of X-IGRT EBE SYSTEMS intended to be:
• for NORMAL USE, operated under the authority of the RESPONSIBLE ORGANIZATION by QUALIFIED PERSONS having the required skills for a particular medical application, for particular specified clinical purposes, e.g., STATIONARY RADIOTHERAPY or MOVING BEAM RADIOTHERAPY,
• maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE, and
• subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON.
IEC 80601-2-68:2024 cancels and replaces the first edition published in 2014. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) alignment with the new editions of the relevant standards:
– IEC 60601-2-1:2020;
– IEC 60601-2-44:2009, IEC 60601-2-44:2009/AMD1:2012 and IEC 60601-2-44:2009/AMD2:2016;
– IEC 60601-2-64:2014;
b) clarification of the use of IEC 60601-2-68 for CT SCANNERS, X-RAY EQUIPMENT for RADIOGRAPHY and RADIOSCOPY used in the same room with an EXTERNAL BEAM EQUIPMENT (EBE);
c) introduction of updated requirements related to MECHANICAL HAZARDS, RADIATION HAZARDS, PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS), ACCOMPANYING DOCUMENTATION of an ME SYSTEM, and REMOTE OPERATION.
- Standard79 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61846:2025 is applicable to:
– therapy equipment using extracorporeally induced focused pressure pulse waves;
– therapy equipment producing focused mechanical energy excluding thermal energy.
This document does not apply to percutaneous and laser lithotripsy equipment.
This document does not apply to:
– histotripsy or other therapeutic ultrasound bursts of longer time duration than that of the pressure pulse;
– non-focused pressure pulse equipment.
This document specifies:
– measurable parameters which could be used in the declaration of the acoustic output of extracorporeal focused pressure pulse equipment;
– methods of measurement and characterization of the pressure field generated by focused pressure pulse equipment.
While this document has been developed for equipment intended for use in lithotripsy, it has been developed such that, as long as no other specific standards are available to be used for other medical applications of therapeutic extracorporeal focused pressure pulse equipment, this document can be used as a guideline.
IEC 61846:2025 cancels and replaces the first edition published in 1998. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) Change of title: "pressure pulse lithotripters" in the previous edition is changed to "therapeutic focused short pressure pulse sources" in order to take into account the development in the relevant technical and biomedical applications of such sources, which were originally used only for (kidney) lithotripsy, while recent applications include a wide range for the treatment of, for example, stone diseases, orthopaedic pain, tissue, cardiac and brain diseases.
The term "focused" was added to differentiate IEC 61846 from IEC 63045.
The term "short" was added to align the nomenclature to IEC 63045 and to differentiate IEC 61846 from standards in the HIFU and HITU fields.
b) Clause 1 and elsewhere in the document: The term "lithotripsy" is changed to "therapy" in order to account for the wide range of applications beyond stone diseases.
c) Clause 3: The "−6 dB" parameter definitions are replaced by "−n dB" to avoid misconceptions in the significance and use of these parameters and to account for newer findings in literature.
Additional "n MPa" parameters are introduced for the same reasons.
The definitions of "derived" parameters are aligned to those in recently published standards, for example IEC 62127-1.
New definitions were added which describe parameters appearing in newer relevant literature, for example "momentum", "average positive acoustic pressure", "cavitation induction index", "pulse to pulse variability", "total pressure pulse energy dose".
d) Clause 6: The terms "focus hydrophone" and "field hydrophone" were removed to account for newer technical developments. New terms distinguish between "hydrophones for pressure pulse measurements" and "hydrophones for quality assurance".
e) Annexes: Descriptions, tables and figures were edited to account for newer literature and standards as well as technical developments.
- Standard41 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61846:2025 is applicable to: – therapy equipment using extracorporeally induced focused pressure pulse waves; – therapy equipment producing focused mechanical energy excluding thermal energy. This document does not apply to percutaneous and laser lithotripsy equipment. This document does not apply to: – histotripsy or other therapeutic ultrasound bursts of longer time duration than that of the pressure pulse; – non-focused pressure pulse equipment. This document specifies: – measurable parameters which could be used in the declaration of the acoustic output of extracorporeal focused pressure pulse equipment; – methods of measurement and characterization of the pressure field generated by focused pressure pulse equipment. While this document has been developed for equipment intended for use in lithotripsy, it has been developed such that, as long as no other specific standards are available to be used for other medical applications of therapeutic extracorporeal focused pressure pulse equipment, this document can be used as a guideline. IEC 61846:2025 cancels and replaces the first edition published in 1998. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) Change of title: "pressure pulse lithotripters" in the previous edition is changed to "therapeutic focused short pressure pulse sources" in order to take into account the development in the relevant technical and biomedical applications of such sources, which were originally used only for (kidney) lithotripsy, while recent applications include a wide range for the treatment of, for example, stone diseases, orthopaedic pain, tissue, cardiac and brain diseases. The term "focused" was added to differentiate IEC 61846 from IEC 63045. The term "short" was added to align the nomenclature to IEC 63045 and to differentiate IEC 61846 from standards in the HIFU and HITU fields. b) Clause 1 and elsewhere in the document: The term "lithotripsy" is changed to "therapy" in order to account for the wide range of applications beyond stone diseases. c) Clause 3: The "−6 dB" parameter definitions are replaced by "−n dB" to avoid misconceptions in the significance and use of these parameters and to account for newer findings in literature. Additional "n MPa" parameters are introduced for the same reasons. The definitions of "derived" parameters are aligned to those in recently published standards, for example IEC 62127-1. New definitions were added which describe parameters appearing in newer relevant literature, for example "momentum", "average positive acoustic pressure", "cavitation induction index", "pulse to pulse variability", "total pressure pulse energy dose". d) Clause 6: The terms "focus hydrophone" and "field hydrophone" were removed to account for newer technical developments. New terms distinguish between "hydrophones for pressure pulse measurements" and "hydrophones for quality assurance". e) Annexes: Descriptions, tables and figures were edited to account for newer literature and standards as well as technical developments.
- Standard41 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-40:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ELECTROMYOGRAPHS and EVOKED RESPONSE EQUIPMENT, hereafter referred to as ME EQUIPMENT.
The following ME EQUIPMENT are excluded:
- ME EQUIPMENT intended for therapeutic application;
- ME EQUIPMENT intended for transcutaneous electrical nerve stimulators and electrical muscle stimulators (ME EQUIPMENT covered by IEC 60601-2-10).
IEC 60601-2-40:2024 cancels and replaces the second edition published in 2016. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) added requirements for constant voltage stimulators;
b) clarified requirements for VISUAL STIMULATORS.
- Standard33 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61846:2025 is applicable to:
– therapy equipment using extracorporeally induced focused pressure pulse waves;
– therapy equipment producing focused mechanical energy excluding thermal energy.
This document does not apply to percutaneous and laser lithotripsy equipment.
This document does not apply to:
– histotripsy or other therapeutic ultrasound bursts of longer time duration than that of the pressure pulse;
– non-focused pressure pulse equipment.
This document specifies:
– measurable parameters which could be used in the declaration of the acoustic output of extracorporeal focused pressure pulse equipment;
– methods of measurement and characterization of the pressure field generated by focused pressure pulse equipment.
While this document has been developed for equipment intended for use in lithotripsy, it has been developed such that, as long as no other specific standards are available to be used for other medical applications of therapeutic extracorporeal focused pressure pulse equipment, this document can be used as a guideline.
IEC 61846:2025 cancels and replaces the first edition published in 1998. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) Change of title: "pressure pulse lithotripters" in the previous edition is changed to "therapeutic focused short pressure pulse sources" in order to take into account the development in the relevant technical and biomedical applications of such sources, which were originally used only for (kidney) lithotripsy, while recent applications include a wide range for the treatment of, for example, stone diseases, orthopaedic pain, tissue, cardiac and brain diseases.
The term "focused" was added to differentiate IEC 61846 from IEC 63045.
The term "short" was added to align the nomenclature to IEC 63045 and to differentiate IEC 61846 from standards in the HIFU and HITU fields.
b) Clause 1 and elsewhere in the document: The term "lithotripsy" is changed to "therapy" in order to account for the wide range of applications beyond stone diseases.
c) Clause 3: The "−6 dB" parameter definitions are replaced by "−n dB" to avoid misconceptions in the significance and use of these parameters and to account for newer findings in literature.
Additional "n MPa" parameters are introduced for the same reasons.
The definitions of "derived" parameters are aligned to those in recently published standards, for example IEC 62127-1.
New definitions were added which describe parameters appearing in newer relevant literature, for example "momentum", "average positive acoustic pressure", "cavitation induction index", "pulse to pulse variability", "total pressure pulse energy dose".
d) Clause 6: The terms "focus hydrophone" and "field hydrophone" were removed to account for newer technical developments. New terms distinguish between "hydrophones for pressure pulse measurements" and "hydrophones for quality assurance".
e) Annexes: Descriptions, tables and figures were edited to account for newer literature and standards as well as technical developments.
- Standard80 pagesEnglish and French languagesale 15% off
IEC 60601-2-54:2022 is available as IEC 60601-2-54:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 60601-2-54:2022 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ME EQUIPMENT and ME SYSTEMS intended to be used for projection RADIOGRAPHY and INDIRECT RADIOSCOPY. IEC 60601-2-43 applies to ME EQUIPMENT and ME SYSTEMS intended to be used for interventional applications and refers to applicable requirements in this document. ME EQUIPMENT and ME SYSTEMS intended to be used for bone or tissue absorption densitometry, computed tomography, mammography or dental or radiotherapy applications are excluded from the scope of this document. The scope of this document also excludes radiotherapy simulators. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. IEC 60601-2-54:2022 edition cancels and replaces the first edition published in 2009, Amendment 1:2015 and Amendment 2:2018. This edition constitutes a technical revision. This edition includes editorial and technical changes to reflect the IEC 60601 1:2005/AMD2:2020. It also contains corrections and technical improvements. Significant technical changes with respect to the previous edition are as follows:
a) a new specific term DOSIMETER is introduced to replace the general term DOSEMETER;
b) terms and definitions taken exclusively from IEC TR 60788:2004 and which are specifically applicable in this document have been moved to 201.3;
c) the collateral standards IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020, IEC 60601-1-12:2014 and IEC 60601-1-12:2014/AMD1:2020 are applicable if MANUFACTURER so declares;
d) the subclause 201.11.101 “Protection against excessive temperatures of X-ray tube assemblies” has been removed from this document as its requirements are sufficiently and clearly covered by IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-28:2017;
e) to adopt changes which are introduced with respect to indicator lights in 7.8.1 of the IEC 60601-1:2005/AMD2:2020 clarification of requirements is provided to avoid conflicts with requirements of indicator lights stipulated for X-RAY EQUIPMENT;
f) explanation of the term ESSENTIAL PERFORMANCE is provided in Annex AA to emphasize the performance of the clinical function under NORMAL and SINGLE FAULT CONDITIONS. CONDITIONS.
- Standard83 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment13 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-54:2022 is available as IEC 60601-2-54:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-54:2022 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ME EQUIPMENT and ME SYSTEMS intended to be used for projection RADIOGRAPHY and INDIRECT RADIOSCOPY. IEC 60601-2-43 applies to ME EQUIPMENT and ME SYSTEMS intended to be used for interventional applications and refers to applicable requirements in this document. ME EQUIPMENT and ME SYSTEMS intended to be used for bone or tissue absorption densitometry, computed tomography, mammography or dental or radiotherapy applications are excluded from the scope of this document. The scope of this document also excludes radiotherapy simulators. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. IEC 60601-2-54:2022 edition cancels and replaces the first edition published in 2009, Amendment 1:2015 and Amendment 2:2018. This edition constitutes a technical revision. This edition includes editorial and technical changes to reflect the IEC 60601 1:2005/AMD2:2020. It also contains corrections and technical improvements. Significant technical changes with respect to the previous edition are as follows: a) a new specific term DOSIMETER is introduced to replace the general term DOSEMETER; b) terms and definitions taken exclusively from IEC TR 60788:2004 and which are specifically applicable in this document have been moved to 201.3; c) the collateral standards IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020, IEC 60601-1-12:2014 and IEC 60601-1-12:2014/AMD1:2020 are applicable if MANUFACTURER so declares; d) the subclause 201.11.101 “Protection against excessive temperatures of X-ray tube assemblies” has been removed from this document as its requirements are sufficiently and clearly covered by IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-28:2017; e) to adopt changes which are introduced with respect to indicator lights in 7.8.1 of the IEC 60601-1:2005/AMD2:2020 clarification of requirements is provided to avoid conflicts with requirements of indicator lights stipulated for X-RAY EQUIPMENT; f) explanation of the term ESSENTIAL PERFORMANCE is provided in Annex AA to emphasize the performance of the clinical function under NORMAL and SINGLE FAULT CONDITIONS. CONDITIONS.
- Standard83 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment13 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61674:2024 specifies the performance and some related constructional requirements of DIAGNOSTIC DOSIMETERS intended for the measurement of AIR KERMA, AIR KERMA LENGTH PRODUCT or AIR KERMA RATE, in photon radiation fields used in medical X-ray imaging, such as RADIOGRAPHY, RADIOSCOPY and COMPUTED TOMOGRAPHY (CT), for X-RADIATION with generating potentials in the range of 20 kV to 150 kV. This document is applicable to the performance of DOSIMETERS with VENTED IONIZATION CHAMBERS and/or SEMICONDUCTOR DETECTORS as used in X-ray diagnostic imaging.
IEC 61674:2024 cancels and replaces the second edition published in 2012. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) for mammography, the manufacturer specifies the REFERENCE VALUE for the RADIATION QUALITY;
b) for mammography, the manufacturer provides the MINIMUM RATED RANGE of RADIATION QUALITIES for the compliance test on energy dependence of response;
c) the compliance test for analogue displays was removed;
d) the compliance tests for range reset, the effect of leakage and recombination losses were removed. These tests are already covered by the test on linearity and cannot be conducted for modern devices. The estimation of COMBINED STANDARD UNCERTAINTY was changed accordingly;
e) the compliance test for mains rechargeable and battery-operated dosimeters were updated for modern devices
- Standard39 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61674:2024 specifies the performance and some related constructional requirements of DIAGNOSTIC DOSIMETERS intended for the measurement of AIR KERMA, AIR KERMA LENGTH PRODUCT or AIR KERMA RATE, in photon radiation fields used in medical X-ray imaging, such as RADIOGRAPHY, RADIOSCOPY and COMPUTED TOMOGRAPHY (CT), for X-RADIATION with generating potentials in the range of 20 kV to 150 kV. This document is applicable to the performance of DOSIMETERS with VENTED IONIZATION CHAMBERS and/or SEMICONDUCTOR DETECTORS as used in X-ray diagnostic imaging. IEC 61674:2024 cancels and replaces the second edition published in 2012. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) for mammography, the manufacturer specifies the REFERENCE VALUE for the RADIATION QUALITY; b) for mammography, the manufacturer provides the MINIMUM RATED RANGE of RADIATION QUALITIES for the compliance test on energy dependence of response; c) the compliance test for analogue displays was removed; d) the compliance tests for range reset, the effect of leakage and recombination losses were removed. These tests are already covered by the test on linearity and cannot be conducted for modern devices. The estimation of COMBINED STANDARD UNCERTAINTY was changed accordingly; e) the compliance test for mains rechargeable and battery-operated dosimeters were updated for modern devices
- Standard39 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61674:2024 specifies the performance and some related constructional requirements of DIAGNOSTIC DOSIMETERS intended for the measurement of AIR KERMA, AIR KERMA LENGTH PRODUCT or AIR KERMA RATE, in photon radiation fields used in medical X-ray imaging, such as RADIOGRAPHY, RADIOSCOPY and COMPUTED TOMOGRAPHY (CT), for X-RADIATION with generating potentials in the range of 20 kV to 150 kV. This document is applicable to the performance of DOSIMETERS with VENTED IONIZATION CHAMBERS and/or SEMICONDUCTOR DETECTORS as used in X-ray diagnostic imaging.
IEC 61674:2024 cancels and replaces the second edition published in 2012. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) for mammography, the manufacturer specifies the REFERENCE VALUE for the RADIATION QUALITY;
b) for mammography, the manufacturer provides the MINIMUM RATED RANGE of RADIATION QUALITIES for the compliance test on energy dependence of response;
c) the compliance test for analogue displays was removed;
d) the compliance tests for range reset, the effect of leakage and recombination losses were removed. These tests are already covered by the test on linearity and cannot be conducted for modern devices. The estimation of COMBINED STANDARD UNCERTAINTY was changed accordingly;
e) the compliance test for mains rechargeable and battery-operated dosimeters were updated for modern devices
- Standard116 pagesEnglish languagesale 15% off
- Standard75 pagesEnglish and French languagesale 15% off
IEC 61223-3-8:2024 applies to evaluation of the imaging performance and related quality control parameters of X‑ray equipment for radiography and radioscopy that conform to IEC 60601-2-54:2022 or IEC 60601-2-43:2022.
This document applies to the evaluation of the imaging performance of the entire imaging chain from image acquisition, image processing and image display.
This document applies to acceptance tests and constancy tests, which are part of the quality assurance program in medical imaging departments and are intended to be performed by or under the responsibility of the responsible organization. A detailed discussion of the position of these tests within the medical radiological equipment lifecycle is provided in Clause A.2. The methods included rely mainly on non-invasive measurements that use appropriate test equipment and are performed after the installation is completed in accordance with the manufacturer’s installation instructions.
IEC 60601-2-54:2022 and IEC 60601-2-43:2022 require information to be provided to the responsible organization with respect to quality control. This document provides guidance to manufacturers regarding the acceptance and constancy tests for the X‑ray equipment in a manufacturer supplied quality control manual. Annex G provides guidance for such a manual.
- Standard100 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61223-3-8:2024 applies to evaluation of the imaging performance and related quality control parameters of X‑ray equipment for radiography and radioscopy that conform to IEC 60601-2-54:2022 or IEC 60601-2-43:2022. This document applies to the evaluation of the imaging performance of the entire imaging chain from image acquisition, image processing and image display. This document applies to acceptance tests and constancy tests, which are part of the quality assurance program in medical imaging departments and are intended to be performed by or under the responsibility of the responsible organization. A detailed discussion of the position of these tests within the medical radiological equipment lifecycle is provided in Clause A.2. The methods included rely mainly on non-invasive measurements that use appropriate test equipment and are performed after the installation is completed in accordance with the manufacturer’s installation instructions. IEC 60601-2-54:2022 and IEC 60601-2-43:2022 require information to be provided to the responsible organization with respect to quality control. This document provides guidance to manufacturers regarding the acceptance and constancy tests for the X‑ray equipment in a manufacturer supplied quality control manual. Annex G provides guidance for such a manual.
- Standard100 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61223-3-8:2024 applies to evaluation of the imaging performance and related quality control parameters of X‑ray equipment for radiography and radioscopy that conform to IEC 60601-2-54:2022 or IEC 60601-2-43:2022.
This document applies to the evaluation of the imaging performance of the entire imaging chain from image acquisition, image processing and image display.
This document applies to acceptance tests and constancy tests, which are part of the quality assurance program in medical imaging departments and are intended to be performed by or under the responsibility of the responsible organization. A detailed discussion of the position of these tests within the medical radiological equipment lifecycle is provided in Clause A.2. The methods included rely mainly on non-invasive measurements that use appropriate test equipment and are performed after the installation is completed in accordance with the manufacturer’s installation instructions.
IEC 60601-2-54:2022 and IEC 60601-2-43:2022 require information to be provided to the responsible organization with respect to quality control. This document provides guidance to manufacturers regarding the acceptance and constancy tests for the X‑ray equipment in a manufacturer supplied quality control manual. Annex G provides guidance for such a manual.
- Standard199 pagesEnglish and French languagesale 15% off
- Corrigendum3 pagesEnglish and French languagesale 10% offe-Library read for1 day
- Corrigendum3 pagesEnglish and French languagesale 10% offe-Library read for1 day
- Standard1 pageEnglish and French languagesale 15% off
IEC 62220-2-1:2023 describes the performance metrics associated with DUAL-ENERGY IMAGING capable DIGITAL X-RAY IMAGING DEVICES meant for medical applications and specifies the methods for their determination. These metrics can be used to analyse TISSUE-SUBTRACTED IMAGES and to evaluate dose performance, noise characteristics, and tissue-subtraction efficacy of DIGITAL X-RAY IMAGING DEVICES. The described methods indicate the procedures to obtain MULTI-SPECTRAL PRIMARY DATA and to compute their derived TISSUE-SUBTRACTED IMAGES.
The intended users of this document are MANUFACTURERS and well-equipped test laboratories. This document is restricted to DIGITAL X-RAY IMAGING DEVICES that are used for single or multiple exposure dual-energy radiographic imaging based on, for example, CR systems, direct and indirect flat panel-detector based systems.
This document excludes and is not applicable to:
– DIGITAL X-RAY IMAGING DEVICES intended to be used in mammography or in dental RADIOGRAPHY;
– slot scanning DIGITAL X-RAY IMAGING DEVICES;
– COMPUTED TOMOGRAPHY or CONE-BEAM COMPUTED TOMOGRAPHY;
– photon-energy discriminating devices such as photon counting X-RAY IMAGING DEVICES;
– devices for dynamic imaging (where series of images are acquired, as in fluoroscopy or cardiac imaging).
– DIGITAL X-RAY IMAGING DEVICES intended to be used with RADIOTHERAPY beams.
- Standard24 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62220-2-1:2023 describes the performance metrics associated with DUAL-ENERGY IMAGING capable DIGITAL X-RAY IMAGING DEVICES meant for medical applications and specifies the methods for their determination. These metrics can be used to analyse TISSUE-SUBTRACTED IMAGES and to evaluate dose performance, noise characteristics, and tissue-subtraction efficacy of DIGITAL X-RAY IMAGING DEVICES. The described methods indicate the procedures to obtain MULTI-SPECTRAL PRIMARY DATA and to compute their derived TISSUE-SUBTRACTED IMAGES. The intended users of this document are MANUFACTURERS and well-equipped test laboratories. This document is restricted to DIGITAL X-RAY IMAGING DEVICES that are used for single or multiple exposure dual-energy radiographic imaging based on, for example, CR systems, direct and indirect flat panel-detector based systems. This document excludes and is not applicable to: – DIGITAL X-RAY IMAGING DEVICES intended to be used in mammography or in dental RADIOGRAPHY; – slot scanning DIGITAL X-RAY IMAGING DEVICES; – COMPUTED TOMOGRAPHY or CONE-BEAM COMPUTED TOMOGRAPHY; – photon-energy discriminating devices such as photon counting X-RAY IMAGING DEVICES; – devices for dynamic imaging (where series of images are acquired, as in fluoroscopy or cardiac imaging). – DIGITAL X-RAY IMAGING DEVICES intended to be used with RADIOTHERAPY beams.
- Standard24 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62220-2-1:2023 describes the performance metrics associated with DUAL-ENERGY IMAGING capable DIGITAL X-RAY IMAGING DEVICES meant for medical applications and specifies the methods for their determination. These metrics can be used to analyse TISSUE-SUBTRACTED IMAGES and to evaluate dose performance, noise characteristics, and tissue-subtraction efficacy of DIGITAL X-RAY IMAGING DEVICES. The described methods indicate the procedures to obtain MULTI-SPECTRAL PRIMARY DATA and to compute their derived TISSUE-SUBTRACTED IMAGES.
The intended users of this document are MANUFACTURERS and well-equipped test laboratories. This document is restricted to DIGITAL X-RAY IMAGING DEVICES that are used for single or multiple exposure dual-energy radiographic imaging based on, for example, CR systems, direct and indirect flat panel-detector based systems.
This document excludes and is not applicable to:
– DIGITAL X-RAY IMAGING DEVICES intended to be used in mammography or in dental RADIOGRAPHY;
– slot scanning DIGITAL X-RAY IMAGING DEVICES;
– COMPUTED TOMOGRAPHY or CONE-BEAM COMPUTED TOMOGRAPHY;
– photon-energy discriminating devices such as photon counting X-RAY IMAGING DEVICES;
– devices for dynamic imaging (where series of images are acquired, as in fluoroscopy or cardiac imaging).
– DIGITAL X-RAY IMAGING DEVICES intended to be used with RADIOTHERAPY beams.
- Standard43 pagesEnglish and French languagesale 15% off
IEC 60601-2-57:2023 applies to basic safety and essential performance of equipment incorporating one or more sources of optical radiation in the wavelength range 200 nm to 3 000 nm, with the exception of laser radiation, and intended to create photobiological effects in humans for therapeutic, diagnostic, monitoring, and cosmetic or aesthetic applications; hereafter referred to as light source equipment (ls equipment).
- Standard59 pagesEnglish and French languagesale 15% off
IEC 61676:2023 is available as IEC 61676:2023 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61676:2023 specifies the performance requirements of instruments as used in the non-invasive measurement of X-RAY tube voltage up to 150 kV and the relevant compliance tests. This document also describes the method for calibration and gives guidance for estimating the uncertainty in measurements performed under conditions different from those during calibration. Applications for such measurement are found in diagnostic radiology including mammography, computed tomography (CT), dental radiology and radioscopy. This document is not concerned with the safety aspect of such instruments. The requirements for electrical safety applying to them are contained in IEC 61010-1. IEC 61676:2023 cancels and replaces first edition published in 2002, Amendment 1:2008. This edition constitutes a technical revision. It includes an assessment of the combined standard uncertainty for the performance of a hypothetical instrument for the non-invasive measurement of the tube high voltage (in Annex A) which replaces Annex A of the edition 1.1 titled "Recommended performance criteria for the invasive divider".
- Standard37 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61676:2023 is available as IEC 61676:2023 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61676:2023 specifies the performance requirements of instruments as used in the non-invasive measurement of X-RAY tube voltage up to 150 kV and the relevant compliance tests. This document also describes the method for calibration and gives guidance for estimating the uncertainty in measurements performed under conditions different from those during calibration. Applications for such measurement are found in diagnostic radiology including mammography, computed tomography (CT), dental radiology and radioscopy. This document is not concerned with the safety aspect of such instruments. The requirements for electrical safety applying to them are contained in IEC 61010-1. IEC 61676:2023 cancels and replaces first edition published in 2002, Amendment 1:2008. This edition constitutes a technical revision. It includes an assessment of the combined standard uncertainty for the performance of a hypothetical instrument for the non-invasive measurement of the tube high voltage (in Annex A) which replaces Annex A of the edition 1.1 titled "Recommended performance criteria for the invasive divider".
- Standard37 pagesEnglish languagesale 10% offe-Library read for1 day
Within the context of the ISO/IEEE 11073 family of standards for device communication, this document establishes a normative definition of communication between personal telehealth weighing scale devices and compute engines (e.g., cell phones, personal computers, personal health appliances, and set top boxes) in a manner that enables plug-and-play interoperability. It leverages appropriate portions of existing standards including ISO/IEEE 11073 terminology, information models, application profile standards, and transport standards. It specifies the use of specific term codes, formats, and behaviors in telehealth environments restricting optionality in base frameworks in favor of interoperability. This document defines a common core of communication functionality for personal telehealth weighing scales.
- Standard64 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61676:2023 specifies the performance requirements of instruments as used in the non-invasive measurement of X-RAY tube voltage up to 150 kV and the relevant compliance tests. This document also describes the method for calibration and gives guidance for estimating the uncertainty in measurements performed under conditions different from those during calibration. Applications for such measurement are found in diagnostic radiology including mammography, computed tomography (CT), dental radiology and radioscopy. This document is not concerned with the safety aspect of such instruments. The requirements for electrical safety applying to them are contained in IEC 61010-1. IEC 61676:2023 cancels and replaces first edition published in 2002, Amendment 1:2008. This edition constitutes a technical revision. It includes an assessment of the combined standard uncertainty for the performance of a hypothetical instrument for the non-invasive measurement of the tube high voltage (in Annex A) which replaces Annex A of the edition 1.1 titled "Recommended performance criteria for the invasive divider".
The contents of the corrigendum of January 2024 have been included in this copy.
- Standard104 pagesEnglish languagesale 15% off
- Standard67 pagesEnglish and French languagesale 15% off
20211012-JO-Link to 93/42/EEC and M/295 removed
- Amendment9 pagesEnglish languagesale 10% offe-Library read for1 day
20211012-JO-Link to 93/42/EEC and M/295 removed
- Amendment9 pagesEnglish languagesale 10% offe-Library read for1 day
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of both FIXED and MOBILE X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, hereafter referred to as INTERVENTIONAL X-RAY EQUIPMENT. Its scope excludes, in particular:
- equipment for RADIOTHERAPY;
- equipment for COMPUTED TOMOGRAPHY;
- ACCESSORIES intended to be introduced into the PATIENT;
- mammographic X-RAY EQUIPMENT;
- dental X-RAY EQUIPMENT.
NOTE 1 Examples of RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, for which the use of INTERVENTIONAL X-RAY EQUIPMENT complying with this standard is recommended, are given in Annex AA.
NOTE 2 Specific requirements for magnetic navigation devices, and for the use of INTERVENTIONAL X-RAY EQUIPMENT in an operating room environment were not considered in this particular standard; therefore no specific requirements have been developed for these devices or uses. In any case, such devices or uses remain under the general clause requirements.
NOTE 3 INTERVENTIONAL X-RAY EQUIPMENT, when used for cone-beam CT mode, is covered by this standard and not by IEC 60601-2-44 [2]2. No additional requirements for operation in cone-beam CT mode were identified for this standard (see also Note 4 in 203.6.4.5).
INTERVENTIONAL X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, which does not include a PATIENT SUPPORT as part of the system, is exempt from the PATIENT SUPPORT provisions of this standard.
If a clause or subclause is specifically intended to be applicable to INTERVENTIONAL X-RAY EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to INTERVENTIONAL X-RAY EQUIPMENT and to ME SYSTEMS, as relevant.
NOTE 4 See also 4.2 of the general standard.
The subclauses of this standard supersede IEC 60601-2-54 subclauses. IEC 60601-2-54 applies only with regards to the cited subclauses; non-cited subclauses of IEC 60601-2-54 do not apply.
- Standard65 pagesEnglish languagesale 10% offe-Library read for1 day
This International Standard is applicable to X-RAY SOURCE ASSEMBLIES and X-RAY TUBE ASSEMBLIES, for use in MEDICAL DIAGNOSTIC RADIOLOGY for techniques in which the X-RAY PATTERN will be received simultaneously in all points of the IMAGE RECEPTION AREA.
This standard specifies a method for the determination of the greatest geometrically symmetrical RADIATION FIELD at a specified distance from the FOCAL SPOT for which the percentage AIR KERMA RATE along the major axes of the RADIATION FIELD does not fall below a permitted value.
NOTE 1 AIR KERMA or AIR KERMA RATE are the only practical verifiable physical quantities for X-RAY SOURCES. X-RAY SOURCES must be tested independently from MEDICAL ELECTRICAL SYSTEMS. Conversion to the characteristics of the X-RAY IMAGE RECEPTOR used in a MEDICAL ELECTRICAL SYSTEM may be done in addition.
In case multiple FOCAL SPOTS are not super-imposed, each focal spot has its own REFERENCE AXIS. Then the maximum RADIATION FIELD may be given for each FOCAL SPOT separately
NOTE 2 The maximum symmetrical RADIATION FIELD may change from its initial value as the X-RAY TUBE ages through use.
NOTE 3 If, for certain MEDICAL ELECTRICAL SYSTEMs the scope of IEC 60806 does not fit, then the special RADIATION FIELD requirements could be incorporated in the MEDICAL ELECTRICAL SYSTEM particular standard. However, a statement on the RADIATION FIELD while referring IEC 60806 is then no longer possible.
- Standard14 pagesEnglish languagesale 10% offe-Library read for1 day
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of both FIXED and MOBILE X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, hereafter referred to as INTERVENTIONAL X-RAY EQUIPMENT. Its scope excludes, in particular: - equipment for RADIOTHERAPY; - equipment for COMPUTED TOMOGRAPHY; - ACCESSORIES intended to be introduced into the PATIENT; - mammographic X-RAY EQUIPMENT; - dental X-RAY EQUIPMENT. NOTE 1 Examples of RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, for which the use of INTERVENTIONAL X-RAY EQUIPMENT complying with this standard is recommended, are given in Annex AA. NOTE 2 Specific requirements for magnetic navigation devices, and for the use of INTERVENTIONAL X-RAY EQUIPMENT in an operating room environment were not considered in this particular standard; therefore no specific requirements have been developed for these devices or uses. In any case, such devices or uses remain under the general clause requirements. NOTE 3 INTERVENTIONAL X-RAY EQUIPMENT, when used for cone-beam CT mode, is covered by this standard and not by IEC 60601-2-44 [2]2. No additional requirements for operation in cone-beam CT mode were identified for this standard (see also Note 4 in 203.6.4.5). INTERVENTIONAL X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, which does not include a PATIENT SUPPORT as part of the system, is exempt from the PATIENT SUPPORT provisions of this standard. If a clause or subclause is specifically intended to be applicable to INTERVENTIONAL X-RAY EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to INTERVENTIONAL X-RAY EQUIPMENT and to ME SYSTEMS, as relevant. NOTE 4 See also 4.2 of the general standard. The subclauses of this standard supersede IEC 60601-2-54 subclauses. IEC 60601-2-54 applies only with regards to the cited subclauses; non-cited subclauses of IEC 60601-2-54 do not apply.
- Standard65 pagesEnglish languagesale 10% offe-Library read for1 day
This International Standard is applicable to X-RAY SOURCE ASSEMBLIES and X-RAY TUBE ASSEMBLIES, for use in MEDICAL DIAGNOSTIC RADIOLOGY for techniques in which the X-RAY PATTERN will be received simultaneously in all points of the IMAGE RECEPTION AREA. This standard specifies a method for the determination of the greatest geometrically symmetrical RADIATION FIELD at a specified distance from the FOCAL SPOT for which the percentage AIR KERMA RATE along the major axes of the RADIATION FIELD does not fall below a permitted value. NOTE 1 AIR KERMA or AIR KERMA RATE are the only practical verifiable physical quantities for X-RAY SOURCES. X-RAY SOURCES must be tested independently from MEDICAL ELECTRICAL SYSTEMS. Conversion to the characteristics of the X-RAY IMAGE RECEPTOR used in a MEDICAL ELECTRICAL SYSTEM may be done in addition. In case multiple FOCAL SPOTS are not super-imposed, each focal spot has its own REFERENCE AXIS. Then the maximum RADIATION FIELD may be given for each FOCAL SPOT separately NOTE 2 The maximum symmetrical RADIATION FIELD may change from its initial value as the X-RAY TUBE ages through use. NOTE 3 If, for certain MEDICAL ELECTRICAL SYSTEMs the scope of IEC 60806 does not fit, then the special RADIATION FIELD requirements could be incorporated in the MEDICAL ELECTRICAL SYSTEM particular standard. However, a statement on the RADIATION FIELD while referring IEC 60806 is then no longer possible.
- Standard14 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-43:2022 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of both FIXED and MOBILE X RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, hereafter referred to as INTERVENTIONAL X-RAY EQUIPMENT. Its scope excludes, in particular:
- equipment for RADIOTHERAPY;
- equipment for COMPUTED TOMOGRAPHY;
- ACCESSORIES intended to be introduced into the PATIENT;
- mammographic X RAY EQUIPMENT;
- dental X RAY EQUIPMENT.
INTERVENTIONAL X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, which does not include a PATIENT SUPPORT as part of the system, is exempt from the PATIENT SUPPORT provisions of this document.
If a clause or subclause is specifically intended to be applicable to INTERVENTIONAL X-RAY EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to INTERVENTIONAL X-RAY EQUIPMENT and to ME SYSTEMS, as relevant.
IEC 60601-2-54 applies only with regards to the cited subclauses; non-cited subclauses of IEC 60601-2-54 do not apply.
IEC 60601-2-43:2022 cancels and replaces the second edition published in 2010, Amendment 1:2017 and Amendment 2:2019. This edition constitutes a technical revision.
This edition includes editorial and technical changes to reflect the IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-54:2022. It also contains corrections and technical improvements. Significant technical changes with respect to the previous edition are as follows:
a) a new specific term DOSIMETER is introduced to replace the general term DOSEMETER as in IEC 60601-2-54:2022;
b) several terms and definitions that are moved from IEC TR 60788:2004 to 201.3 of IEC 60601-2-54:2022 are also referenced from IEC 60601-2-54:2022.
c) the collateral standards IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020, IEC 60601-1-12:2014 and IEC 60601-1-12:2014/AMD1:2020 are applicable if MANUFACTURER so declares;
d) the former subclause 201.11.101 “Protection against excessive temperature of X-RAY TUBE ASSEMBLIES” is removed since covered by IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-28:2017, and the former subclause 201.11.102 is renumbered as 201.11.101, as in IEC 60601-2-54:2022;
e) to adopt changes in subclause 7.8.1 “Colours of indicator lights” in IEC 60601 1:2005/AMD2:2020, clarification of requirements is provided in 201.7.8.1 to avoid conflicts with requirements of indicator lights stipulated for X-RAY EQUIPMENT, as in IEC 60601-2-54:2022;
f) explanation of the term ESSENTIAL PERFORMANCE is provided in Annex AA to emphasize the performance of the clinical function under NORMAL CONDITIONS and SINGLE FAULT CONDITIONS.
- Standard192 pagesEnglish languagesale 15% off
- Standard126 pagesEnglish and French languagesale 15% off
IEC 60806:2022 is applicable to X-RAY SOURCE ASSEMBLIES and X-RAY TUBE ASSEMBLIES.
This document specifies a method for the determination of the greatest geometrically symmetrical RADIATION FIELD at a specified distance from the FOCAL SPOT for which the percentage AIR KERMA RATE along the major axes of the RADIATION FIELD does not fall below a permitted value.
In case multiple FOCAL SPOTS are not super-imposed, each FOCAL SPOT has its own REFERENCE AXIS. Then the maximum RADIATION FIELD can be given for each FOCAL SPOT separately.
IEC 60806:2022 cancels and replaces the first edition published in 1984. This edition constitutes a technical revision.
This edition includes the following significant technical change with respect to the previous edition:
a) addition of solid state detectors as they have become more common since the first edition of 1984.
- Standard21 pagesEnglish and French languagesale 15% off
IEC 61223-3-5:2019 applies to CT SCANNERS that conform to IEC 60601-2-44:2009, IEC 60601-2-44:2009/AMD1:2012 and IEC 60601-2-44:2009/AMD1:2012AMD2:2016. IEC 60601-2-44 and this document • defines the essential parameters which describe the performance of CT SCANNERS with regard to image quality, RADIATION OUTPUT and PATIENT positioning; the list of parameters to be tested can be found in section 4.3, • defines the methods of testing the essential parameters, and • evaluates compliance with the tolerances of the parameters SPECIFIED by the ACCOMPANYING DOCUMENTS. The methods defined in IEC 60601-2-44 and this document rely on non-invasive measurements, using appropriate test equipment, performed during or after installation. Signed statements covering steps in the installation procedure can be used as part of the ACCEPTANCE TEST report. This document applies to ACCEPTANCE TESTS and CONSTANCY TESTS on a CT SCANNER. The aim of the ACCEPTANCE TESTS is to verify compliance of the installation or MAJOR SERVICE ACTION with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning. The CONSTANCY TESTS are performed to ensure that the functional performance of EQUIPMENT meets ESTABLISHED CRITERIA and to enable the early recognition of changes in the properties of components of the EQUIPMENT, and to verify compliance with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning. This document also contains requirements associated with ACCEPTANCE and CONSTANCY TESTING for the ACCOMPANYING DOCUMENTS of the CT SCANNER. This document does not apply to • aspects of mechanical and electrical safety, and • aspects of mechanical, electrical and software performance, unless they are essential for performing the ACCEPTANCE TESTS and CONSTANCY TESTS, and are directly affecting image quality, RADIATION OUTPUT and PATIENT positioning. IEC 61223-3-5:2019 cancels and replaces the first edition published in 2004, and the second edition of IEC 61223-2-6 published in 2006. This edition constitutes a technical revision. This edition includes the following significant technical change with respect to the previous edition and to IEC 61223-2-6: 1) modification of the RADIATION protection and control; 2) modification of the acceptance testing; 3) introduction of constancy testing.
- Standard64 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62985:2019 applies to – CT SCANNERS that are able to display and report CTDIVOL in accordance with IEC 60601-2-44, and – RADIATION dose index monitoring software (RDIMS) for the purpose of calculating, displaying and recording the SIZE SPECIFIC DOSE ESTIMATE (SSDE) and its associated components. Specifically, this document provides standardized methods and requirements for calculating, displaying, or recording of SSDE, SSDE(z), WATER EQUIVALENT DIAMETER (DW), and DW(z), where z represents a specific longitudinal position of the scanned object. This document provides a method of determining a reference WATER EQUIVALENT DIAMETER, DW,REF(z), using CT scans of two cylindrical water PHANTOMS and one or more anthropomorphic PHANTOM(S), which conform to the specifications defined in this document. The method of calculating the WATER EQUIVALENT DIAMETER that is implemented by the MANUFACTURER, DW,IMP(z), is tested and validated against DW,REF(z) using the TEST OBJECTS and methods defined within this document. This document also describes the methods for calculating SSDE and DW, which represent the average values of SSDE(z) and DW(z) over the RECONSTRUCTION LENGTH.
- Standard24 pagesEnglish languagesale 10% offe-Library read for1 day
NEXT ACTION: PUBLICATION EXPECTED BY 2019-10-31
2019-07-04 IP: BT decision D163/C033 - remove link to MDD
2017-10-09: Publication on hold due to negative assessment
- Amendment8 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 60601-2-54:2022 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ME EQUIPMENT and ME SYSTEMS intended to be used for projection RADIOGRAPHY and INDIRECT RADIOSCOPY. IEC 60601-2-43 applies to ME EQUIPMENT and ME SYSTEMS intended to be used for interventional applications and refers to applicable requirements in this document. ME EQUIPMENT and ME SYSTEMS intended to be used for bone or tissue absorption densitometry, computed tomography, mammography or dental or radiotherapy applications are excluded from the scope of this document. The scope of this document also excludes radiotherapy simulators. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. IEC 60601-2-54:2022 edition cancels and replaces the first edition published in 2009, Amendment 1:2015 and Amendment 2:2018. This edition constitutes a technical revision. This edition includes editorial and technical changes to reflect the IEC 60601 1:2005/AMD2:2020. It also contains corrections and technical improvements. Significant technical changes with respect to the previous edition are as follows:
a) a new specific term DOSIMETER is introduced to replace the general term DOSEMETER;
b) terms and definitions taken exclusively from IEC TR 60788:2004 and which are specifically applicable in this document have been moved to 201.3;
c) the collateral standards IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020, IEC 60601-1-12:2014 and IEC 60601-1-12:2014/AMD1:2020 are applicable if MANUFACTURER so declares;
d) the subclause 201.11.101 “Protection against excessive temperatures of X-ray tube assemblies” has been removed from this document as its requirements are sufficiently and clearly covered by IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-28:2017;
e) to adopt changes which are introduced with respect to indicator lights in 7.8.1 of the IEC 60601-1:2005/AMD2:2020 clarification of requirements is provided to avoid conflicts with requirements of indicator lights stipulated for X-RAY EQUIPMENT;
f) explanation of the term ESSENTIAL PERFORMANCE is provided in Annex AA to emphasize the performance of the clinical function under NORMAL and SINGLE FAULT CONDITIONS. CONDITIONS.
- Standard243 pagesEnglish languagesale 15% off
- Standard161 pagesEnglish and French languagesale 15% off
IEC TR 61390:2022 describes representative methods of measuring the performance of complete real-time medical ultrasonic imaging equipment in the frequency range 0,5 MHz to 23 MHz.
This document is relevant for real-time ultrasonic scanners based on the pulse-echo principle, for the types listed below:
- mechanical sector scanner;
- electronic phased array sector scanner;
- electronic linear array scanner;
-electronic curved array sector scanner;
- water-bath scanner based on any of the above four scanning mechanisms;
- plane-wave/fast imaging scanners;
- combination of several of the above methods (e.g. a linear array phased at the edge to produce a sector there to enlarge the field of view.
The methods described are based on evaluation of:
- sonograms obtained by scanning of tissue mimicking objects (phantoms);
- sonograms obtained by scanning of artificial, low- or highly reflective targets in suitable environments;
- parameters of the ultrasound field transmitted by the measured scanner.
This document does not relate to methods for measuring electrical parameters of the scanner’s electronic systems.
IEC TR 61390:2022 cancels and replaces the first edition published in 1996. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) Several additional phantom designs are included in the main body of the document;
b) Several additional transducer types are included in the Scope;
c) Methods of analysis are presented in new Annex B.
- Technical report55 pagesEnglish languagesale 15% off
IEC Corrected version
- Corrigendum3 pagesEnglish and French languagesale 10% offe-Library read for1 day
IEC Corrected version
- Corrigendum3 pagesEnglish and French languagesale 10% offe-Library read for1 day
IEC Corrected version
- Corrigendum4 pagesEnglish and French languagesale 10% offe-Library read for1 day
NEXT ACTION: PUBLICATION EXPECTED BY 2019-10-31
2019-07-04 IP: BT decision D163/C033 - remove link to MDD
NEXT ACTION: TB Action by 21-22 May 2019 i.e. Next CLC TC 62 Plenary Meeting
2019-01-24: JO - In contact with CLC TC 62 secretary to know if an annex ZZ is expected or if a CIB will be launched to decouple this standard from the harmonization process.
- Amendment5 pagesEnglish languagesale 10% offe-Library read for1 day





