Abstract

This document specifies a method for the determination of the time-weighted average mass concentration of sulfuric acid and phosphoric acid in workplace air by ion chromatography. The anions are detected by conductivity.
The method is applicable to the personal sampling of airborne particles, as defined in ISO 7708, and to static (area) sampling.
The method does not apply to the determination of sulfur trioxide.
The procedure does not differentiate between the acids and their corresponding salts if both are present in the workplace air.
The procedure does not differentiate between phosphoric acid and diphosphorus pentoxide (phosphoric anhydride) if both are present in the workplace air.

Status
Published
Public Enquiry End Date
02-Aug-2026
Publication Date
13-Sep-2026
Technical Committee
KAZ - Air quality
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
06-Aug-2026
Due Date
11-Oct-2026
Completion Date
14-Sep-2026

Buy Documents

Standard

SIST ISO 21438-1:2026

English language (28 pages)
Preview
Preview
e-Library read for
×1 day
Standard

ISO 21438-1:2022 - Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 1: Non-volatile acids (sulfuric acid and phosphoric acid) Released:5/24/2022

English language (22 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 21438-1:2022 - Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 1: Non-volatile acids (sulfuric acid and phosphoric acid) Released:5/24/2022

French language (23 pages)
sale 15% off
Preview
sale 15% off
Preview

SIST ISO 21438-1:2026 is the Slovenian standard for workplace atmospheres that adopts ISO 21438-1. It specifies how to determine the time-weighted average mass concentration of sulfuric acid and phosphoric acid in workplace air by ion chromatography, with conductivity detection, for personal sampling of airborne particles and for static area sampling. In practice, it is used by occupational hygienists, laboratories, and workplace safety teams that need to measure exposure to these non-volatile acids.

What does SIST ISO 21438-1:2026 specify?

SIST ISO 21438-1:2026 specifies a method for measuring sulfuric acid and phosphoric acid in workplace air and for reporting the result as a time-weighted average mass concentration. The standard covers how to collect the sample, extract the acids from the filter, separate them by ion chromatography, and detect them by conductivity.

The document is organized into clauses on scope, references, terms, principle, reagents, apparatus, occupational exposure assessment, sampling, analysis, expression of results, method performance, and test report. It also includes one informative annex.

AnnexWhat it covers
Annex ATemperature and pressure correction for the indicated volumetric flow rate and for recalculating concentrations to reference conditions

The method applies to personal sampling and static sampling, but not to sulfur trioxide. It also treats sulfuric acid and phosphoric acid together with their corresponding salts when both are present in workplace air, and it does not distinguish phosphoric acid from diphosphorus pentoxide when both are present.

What are the key requirements of SIST ISO 21438-1:2026?

SIST ISO 21438-1:2026 is built around a filter-based sampling and ion-chromatographic analysis workflow. The practical focus is on choosing the right sampler, avoiding contamination, extracting the collected material consistently, and controlling the instrument with blanks, calibration, and quality checks.

Sampling and collection

Clause 4 states that a known volume of air is drawn through a filter to collect acid mist, using a sampler suited to the relevant particle fraction. That matters because the measured result must match the exposure metric used for the workplace limit value, such as inhalable or thoracic sampling.

Clause 9 requires the sampler to be selected for the applicable limit value and particle fraction, and for size-selective samplers to meet the particle-fraction definitions in ISO 7708 and the performance expectations of EN 13205. In practice, the sampler choice is not just a convenience - it controls what part of the airborne aerosol is actually measured.

Clause 9.1.2 says the sampling period must be long enough to support a usable result but not so long that the filter is overloaded. The clause gives a method for estimating minimum sampling time, which helps the sampler and flow rate be matched to the expected concentration.

Clause 9.1.3 and Annex A deal with temperature and pressure effects on flow measurement and on reporting concentrations. This matters when the flowmeter calibration conditions and workplace conditions differ enough to affect the sample volume or when the result needs to be compared with a reference condition.

Sample preparation and analysis

Clause 9.5 and Clause 10 require careful transport, extraction, and filtration of the collected sample before ion chromatography. The standard gives specific handling for quartz fibre, PVC, and PTFE filters, and it emphasizes protecting the sample from contact with container walls and from contamination.

Clause 10.1 describes extraction with the eluent, use of ultrasonic treatment without heating, and filtration of the extract before analysis. In practice, this is the step that transfers the sampled acids from the filter into a form the chromatograph can measure reproducibly.

Clause 10.2 requires multi-point calibration within the linear range, analysis of blanks, and continuing calibration checks during the run. If calibration drift is too large, the instrument must be corrected or recalibrated, and the affected samples must be reanalysed or reprocessed.

Quality control and performance

Clause 10.3 requires estimation of instrumental detection limits, method detection limits, and quantification limits under the working conditions. That gives the laboratory a realistic view of how low it can measure with confidence, using the actual extraction and analysis process rather than the instrument alone.

Clause 10.4 requires reagent blanks, laboratory blanks, spiked samples, spiked duplicate samples, certified reference materials, and, where available, participation in external quality assessment. This is what keeps routine results credible, because sulfate and phosphate are common environmental contaminants and can elevate blanks.

Clause 10.5 recommends estimating measurement uncertainty in line with ISO/IEC Guide 98-3:2008. For users, that means the reported concentration should be accompanied by an uncertainty estimate that reflects sampling, extraction, calibration, and analytical variability.

Expression of results and interferences

Clause 11 sets out how to calculate the air concentration from the measured sulfate or phosphate concentrations, after blank correction and any dilution correction. Clause 12.6 adds that if sulfate or phosphate from an emission source may be present, the air result may need correction to free-acid content using a parallel source sample.

The method does not separate acid from salt when both are present, and it does not separate phosphoric acid from diphosphorus pentoxide. In practice, the user must decide whether a workplace result is adequate as reported or whether a source sample and corrective interpretation are needed.

What terms does SIST ISO 21438-1:2026 define?

  • Blank solution - a solution made by taking a reagent blank, laboratory blank, or field blank through the same dissolution procedure as a sample.
  • Calibration blank solution - a calibration solution prepared without the working standard solution, treated as zero sulfate and zero phosphate.
  • Calibration solution - a diluted solution of the working standard used to build the instrument calibration.
  • Extraction solution - the solvent or solution used to dissolve the analytes from the collected sample.
  • Sample dissolution - the process of obtaining a solution containing sulfate and phosphate from a sample, whether or not the sample fully dissolves.
  • Sample solution - the solution produced from a sample by sample dissolution.
  • Stock standard solution - a certified sulfate and/or phosphate solution traceable to national standards and used to prepare calibration solutions.
  • Working standard solution - a diluted stock standard solution prepared at a concentration more suitable for calibration work.

Who uses SIST ISO 21438-1:2026?

SIST ISO 21438-1:2026 is used by agencies concerned with health and safety at work, industrial hygienists, public health professionals, analytical laboratories, industrial users of sulfuric and phosphoric acids, and workers whose exposure may need to be measured. It is relevant where exposure monitoring, method validation, or compliance support is needed for workplace air measurements.

Quality managers and buyers also use it to judge whether a sampling and analysis service, or the associated equipment, is appropriate for a specific workplace exposure assessment task. The document is especially relevant where respirable, inhalable, or other size-selective sampling decisions affect the reported result.

What changed in SIST ISO 21438-1:2026 from the previous edition?

SIST ISO 21438-1:2026 replaces ISO 21438-1:2007. The second edition states that it has been technically revised throughout.

Which standards are used with SIST ISO 21438-1:2026?

  • ISO 1042 - supports the use of one-mark volumetric flasks for preparing solutions.
  • ISO 3585 - defines borosilicate glass 3.3 for laboratory glassware.
  • ISO 3696 - specifies the water quality used for reagent preparation and dilution.
  • ISO 7708 - defines the airborne particle size fractions used for health-related sampling.
  • ISO 8655-1, ISO 8655-2, ISO 8655-6 - support the use and verification of piston-operated volumetric apparatus.
  • ISO 18158 - provides workplace-air terminology used with this document.
  • ISO 21832:2018 - gives requirements for evaluating measuring procedures and is used for working range, limits of quantification, and method performance concepts.
  • EN 13205 - is used for assessing sampler performance for airborne particle concentrations.
  • ISO 13137 - is used for the pumps recommended for personal sampling.
  • ISO 20581 - provides general performance requirements for procedures for measuring chemical agents in workplace air.

What does the SIST ISO 21438-1:2026 document contain?

SIST ISO 21438-1:2026 contains practical instructions for workplace-air sampling, sample handling, and ion chromatography analysis of sulfate and phosphate from sulfuric and phosphoric acid. It includes lists of reagents and apparatus, including samplers, filters, pumps, flowmeters, laboratory glassware, and chromatograph components.

It also contains methods for estimating detection and quantification limits, checking calibration and continuing calibration, and evaluating blanks, recovery, precision, and uncertainty. For reporting, it gives the information that should be kept in the test record, including sample identity, sampling conditions, analytical variables, and deviations. Annex A provides the temperature and pressure corrections used when flow measurement or reporting conditions need adjustment.

Buy Documents

Standard

SIST ISO 21438-1:2026

English language (28 pages)
Preview
Preview
e-Library read for
×1 day
Standard

ISO 21438-1:2022 - Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 1: Non-volatile acids (sulfuric acid and phosphoric acid) Released:5/24/2022

English language (22 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 21438-1:2022 - Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 1: Non-volatile acids (sulfuric acid and phosphoric acid) Released:5/24/2022

French language (23 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

ECOCERT

Organic and sustainability certification.

COFRAC France Verified

CIS Institut d.o.o.

Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

SA Slovenia Verified

Eurofins Food Testing Global

Global leader in food, environment, and pharmaceutical product testing.

COFRAC Luxembourg Verified

Sponsored listings

Frequently Asked Questions

SIST ISO 21438-1:2026 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Workplace atmospheres - Determination of inorganic acids by ion chromatography - Part 1: Non-volatile acids (sulfuric acid and phosphoric acid)". This standard covers: This document specifies a method for the determination of the time-weighted average mass concentration of sulfuric acid and phosphoric acid in workplace air by ion chromatography. The anions are detected by conductivity. The method is applicable to the personal sampling of airborne particles, as defined in ISO 7708, and to static (area) sampling. The method does not apply to the determination of sulfur trioxide. The procedure does not differentiate between the acids and their corresponding salts if both are present in the workplace air. The procedure does not differentiate between phosphoric acid and diphosphorus pentoxide (phosphoric anhydride) if both are present in the workplace air.

This document specifies a method for the determination of the time-weighted average mass concentration of sulfuric acid and phosphoric acid in workplace air by ion chromatography. The anions are detected by conductivity. The method is applicable to the personal sampling of airborne particles, as defined in ISO 7708, and to static (area) sampling. The method does not apply to the determination of sulfur trioxide. The procedure does not differentiate between the acids and their corresponding salts if both are present in the workplace air. The procedure does not differentiate between phosphoric acid and diphosphorus pentoxide (phosphoric anhydride) if both are present in the workplace air.

SIST ISO 21438-1:2026 is classified under the following ICS (International Classification for Standards) categories: 13.040.30 - Workplace atmospheres; 71.040.50 - Physicochemical methods of analysis; 71.060.30 - Acids. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST ISO 21438-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 blank solution
  • 3.2 calibration blank solution
  • 3.3 calibration solution
  • 3.4 extraction solution
  • 3.5 sample dissolution
  • 3.6 sample solution
  • 3.7 stock standard solution
  • 3.8 test solution
  • 3.9 working standard solution
  • 4 Principle
  • 5 Requirement
  • 6 Reagents
  • 7 Apparatus
  • 7.1 Sampling equipment
  • 7.2 Laboratory apparatus
  • 8 Occupational exposure assessment
  • 9 Sampling
  • 9.1 Preliminary considerations
  • 9.2 Preparation for sampling
  • 9.3 Sampling position
  • 9.4 Collection of samples
  • 9.5 Transportation
  • 10 Analysis
  • 10.1 Preparation of test, calibration solutions and filter samples
  • 10.2 Instrumental analysis
  • 10.3 Estimation of detection and quantification limits
  • 10.4 Quality control
  • 10.5 Measurement uncertainty
  • 11 Expression of results
  • 12 Method performance
  • 12.1 Sample collection and stability
  • 12.2 Quantification limits
  • 12.3 Upper limits of the analytical range
  • 12.4 Bias and precision
  • 12.5 Uncertainty of sampling and analysis method
  • 12.6 Interferences
  • 13 Test report
  • 13.1 Test record
  • 13.2 Laboratory report
  • Annex A
  • Bibliography

Introduction

The health of workers in many industries is at risk through exposure by inhalation of particulate acids such as sulfuric acid or phosphoric acid compounds. Industrial hygienists and other public health professionals need to determine the effectiveness of measures taken to control workers’ exposure, and this is generally achieved by making workplace air measurements. This document has been published in order to make available a method for making valid exposure measurements for particulate acids in use in industry. It will be of benefit to: agencies concerned with health and safety at work; industrial hygienists and other public health professionals; analytical laboratories; and industrial users of sulfuric and phosphoric acids, and their workers.

It has been assumed in the drafting of the ISO 21438 series that the execution of its provisions and the interpretation of the results obtained are entrusted to appropriately qualified and experienced people.

SIST ISO 21438-1:2026 INTERNATIONAL STANDARD ISO 21438-1:2022(E) Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 1:

Non-volatile acids (sulfuric acid and phosphoric acid)

1 Scope

This document specifies a method for the determination of the time-weighted average mass concentration of sulfuric acid and phosphoric acid in workplace air by ion chromatography. The anions are detected by conductivity.

The method is applicable to the personal sampling of airborne particles, as defined in ISO 7708, and to static (area) sampling.

The method does not apply to the determination of sulfur trioxide. The procedure does not differentiate between the acids and their corresponding salts if both are present in the workplace air.

The procedure does not differentiate between phosphoric acid and diphosphorus pentoxide (phosphoric anhydride) if both are present in the workplace air.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 1042, Laboratory glassware — One-mark volumetric flasks ISO 3585, Borosilicate glass 3.3 — Properties ISO 3696, Water for analytical laboratory use — Specification and test methods ISO 7708, Air quality — Particle size fraction definitions for health-related sampling ISO 8655-1, Piston-operated volumetric apparatus — Part 1: Terminology, general requirements and user recommendations ISO 8655-2, Piston-operated volumetric apparatus — Part 2: Pipettes ISO 8655-6, Piston-operated volumetric apparatus — Part 6: Gravimetric reference measurement procedure for the determination of volume ISO 18158, Workplace air — Terminology ISO 21832:2018, Workplace air — Metals and metalloids in airborne particles — Requirements for evaluation of measuring procedures EN 13205, Workplace exposure — Assessment of sampler performance for measurement of airborne particle concentrations

SIST ISO 21438-1:2026 ISO 21438-1:2022(E)

3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 18158 and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

3.1 blank solution

solution prepared by taking a reagent blank, laboratory blank or field blank through the same procedure used for sample dissolution (3.5)

3.2 calibration blank solution

calibration solution (3.3) prepared without the addition of any working standard solution (3.9) Note 1 to entry: The concentration of sulfate and phosphate in the calibration blank solution is taken to be zero.

3.3 calibration solution

solution prepared by dilution of the working standard solution (3.9), containing sulfate and phosphate at concentrations that are suitable for use in calibration of the analytical instrument

3.4 extraction solution

solvent or solution used to solubilise the analyte(s) of interest

3.5 sample dissolution

process of obtaining a solution containing sulfate and phosphate from a sample, which might or might not involve complete dissolution of the sample

3.6 sample solution

solution prepared from a sample by the process of sample dissolution (3.5) Note 1 to entry: A sample solution can sometimes need to be subjected to further operations, e.g. dilution, in order to produce a test solution (3.8) that is ready for analysis.

3.7 stock standard solution

solution, used for preparation of the calibration solutions (3.3), containing sulfate and/or phosphate at a certified concentration that is traceable to national standards

3.8 test solution

blank solution (3.1) or sample solution (3.6) that has been subjected to all operations required to bring it into a state in which it is ready for analysis, e.g. dilution, filtered Note 1 to entry: The blank test solution is the blank solution and the sample test solution is the sample solution if these solutions are not subjected to any further operations before analysis.

3.9 working standard solution

solution, prepared by dilution of the stock standard solution(s) (3.7), that contains sulfate and phosphate at concentrations that are better suited to preparation of calibration solutions (3.3) than the concentration of sulfate and phosphate in the stock standard solutions

SIST ISO 21438-1:2026 ISO 21438-1:2022(E)

4 Principle

A known volume of air is drawn through a filter to collect acid mist. The filter is mounted in a sampler designed to collect the required particle fraction of airborne particles (see 7.1.1). Currently, limit values in different countries exist for sulfuric acid either as “total particulate mass”, defined by the performance of a sampler, or in the inhalable or the thoracic size selective fraction. A suitable sampling device for the applicable particle fraction shall be used considering the existing limit value/particle fraction.

The collected sample is then treated with eluent and ultrasonic application, without heating, to extract sulfuric and phosphoric acids.

Aliquots of the sample solution are subjected to ion chromatography in order to separate the extracted sulfate and phosphate from other anions. Following this separation, the anions are measured using a conductivity detector.

Analytical results are obtained by plotting the measured conductivity as a function of concentration. They can be used for assessment of occupational exposure to sulfuric acid and phosphoric acid (and diphosphorus pentoxide) in air.

5 Requirement

The measuring procedure shall conform to any relevant international, European or national standard which specifies performance requirements for procedures for measuring chemical agents in workplace [3] air (e.g. ISO 20581 ).

6 Reagents

Use only reagents of recognized analytical grade and only water as specified in 6.1. It is advisable to check the blank values of all chemicals before use.

NOTE 1 Sulfates and phosphates are found ubiquitously in the environment and the presence of sulfates and phosphates in reagents can lead to high blank values.

6.1 Water, from a purification system that delivers ultrapure water of grade 1 (in accordance with ISO 3696).

NOTE 2 State of art water purification systems deliver water of grade 1 with higher quality than specified in ISO 3696:1987 (e.g. resistivity greater than 18 MΩ·cm).

6.2 Sodium carbonate (Na CO ), anhydrous, mass fraction > 99,9 %. 2 3 6.3 Sodium hydrogencarbonate (NaHCO ), anhydrous, mass fraction > 99,9 %.

NOTE 3 An alkaline solution (potassium hydroxide or sodium hydroxide) can be used in place of the buffer solution according to modifications to the method, but the resulting solution will require adjustment before analysis.

6.4 Sodium carbonate/sodium hydrogen carbonate extraction and eluent stock solution. Create a stock solution for the extraction and eluent solution. The stock solution has to be appropriate to the used application.

6.5 Sodium carbonate/sodium hydrogen carbonate extraction and eluent solution. The concentration of the extraction and eluent solution are typically matched, but this can be modified according to the laboratory procedure.

SIST ISO 21438-1:2026 ISO 21438-1:2022(E) 6.6 Sulfate and phosphate stock solutions.

Use a commercial standard solution with a certificate of its concentration, e.g. 1 000 mg/l, traceable to national standards. Observe the manufacturer’s expiry date or recommended shelf-life. If necessary, dilute the standard solution (e.g. from 1 000 mg/l to 100 mg/l) to prepare the stock solution. 6.7 Sulfate and phosphate calibration solutions.

Pipette appropriate volumes of the stock solution to create the calibration. The calibration range shall be in accordance with the calculated working range (from ISO 21832:2018, 5.2.1) and the limit of quantification (from ISO 21832:2018, 8.1.2). A multi-point (the number of points to be consistent with the quality system of the laboratory) calibration shall be constructed within the linear range of the analytical method.

6.8 Sulfate and phosphate check-standard.

Use a certified reference material (CRM) as described in 10.4.3 with a certificate of its concentration, e.g. 400 mg/l for sulfate and 600 mg/l for phosphate, traceable to national standards. Observe the manufacturer’s expiry date or recommended shelf-life.

6.9 Sulfate and phosphate check-standard working solution.

The check-standard working solution should be in the middle of the calibration range. Pipette a proper volume of the sulfate and phosphate check-standard (6.8) to create a check-standard with the eluent (6.4). Prepare this solution fresh for every calibration. If necessary, create a check-standard working solution for sulfate and phosphate separately.

7 Apparatus

7.1 Sampling equipment

7.1.1 Samplers as required to collect samples consistent with applicable limit values. The

performance of samplers shall be in accordance with EN 13205 where specific size-selective fractions are stipulated.

The operating instructions supplied by the manufacturer should be consulted to find out whether particulate matter deposited on the internal surfaces of the sampler forms part of the sample. NOTE 1 In general, personal samplers for collection of the inhalable fraction of airborne particles do not exhibit the same size-selective characteristics if used for static sampling. NOTE 2 Some inhalable samplers are designed to collect the inhalable fraction of airborne particles on the filter, and any particulate matter deposited on the internal surfaces of the sampler is not of interest. Other inhalable samplers are designed such that airborne particles which pass through the entry orifice(s) match the inhalable convention, in which case particulate matter deposited on the internal surfaces of the sampler does form part of the sample. (Samplers of this second type generally incorporate an internal filter cassette or cartridge that can be removed from the sampler to enable this material to be easily recovered.) [9] NOTE 3 PD CEN/TR 15230 gives examples of inhalable samplers with the potential to meet the requirements of EN 13205, including published reports on their performance. NOTE 4 Reference [13] discusses the relationship between inhalable and thoracic fractions of sulfuric acid in workplaces.

SIST ISO 21438-1:2026 ISO 21438-1:2022(E) 7.1.2 Filters, of a diameter suitable for use with the samplers, with a collection efficiency ≥ 99 % for particles with a 0,3 µm diffusion diameter and manufactured from a material that is compatible with the sample preparation and analysis method.

Sulfuric acid and phosphoric acid are strong acids. They react (e.g. by dehydration) with many organic and polymeric materials, and destroy the filter material. Therefore, correct selection of the filter used for sample collection is of paramount importance. In particular, the filter has to be manufactured from a material that does not react with the acids. Also, some filters (e.g. glass fibre filters) can contain metals (e.g. barium) that react with sulfuric and/or phosphoric acids to produce insoluble salts. The following filter types are generally suitable for use:

  • binder-free quartz fibre filters;
  • polyvinyl chloride (PVC) membrane filters, of pore size 5 µm or less;
  • polytetrafluoroethylene (PTFE) membrane filters, of pore size 5 µm or less; Sulfates and phosphates are found ubiquitously in the environment and the presence of sulfates and phosphates in filter materials can lead to high blank values. It is therefore essential to check the blank values of each batch of filters used.

[2] 7.1.3 Sampling pumps, in accordance with ISO 13137 , should be used. [2] If the sampling pump is used outside the range of conditions specified in ISO 13137 appropriate action should be taken to ensure that the performance requirements are met. For instance, at sub-zero temperatures it can be necessary to keep the pump warm.

7.1.4 Flowmeter, portable, with an accuracy that is sufficient to enable the volumetric flow rate to be measured to within ±5 %.

The calibration of the flowmeter shall be checked against a primary standard, i.e. a flowmeter whose accuracy is traceable to national standards. If appropriate (see 9.1.3), record the atmospheric temperature and pressure at which the calibration of the flowmeter was checked. It is advisable that the flowmeter used is capable of measuring the volumetric flow rate to within ±2 % or better.

7.1.5 Ancillary equipment:

  • Flexible tubing, e.g. silicone of an outer diameter of 10 mm and an inner diameter of 6 mm for making a leakproof connection from the samplers to the sampling pumps without compressing as a result of the pressure difference between ambient atmosphere and the in-line vacuum at the used −1 flow rate of 3,5 l min .
  • Belts or harnesses, which the sampling pumps can conveniently fix for personal sampling.
  • Tweezers, manufactured from or tipped with PTFE, for handling filters.
  • Filter transport cassettes, or similar, if required (see 9.5.1), in which to transport samples to the laboratory.
  • Thermometer, of range 0 °C to 50 °C, graduated in divisions of 1 °C or less, for measurement of atmospheric temperature. For applications at temperatures below freezing, the range of the thermometer shall extend to the appropriate desired range.
  • Barometer, suitable for measurement of atmospheric pressure.

SIST ISO 21438-1:2026 ISO 21438-1:2022(E)

7.2 Laboratory apparatus

CAUTION — Both sulfates and phosphates are found ubiquitously in the environment. This can lead to elevated blanks so it is especially important to take great care that all disposable plastic labware is checked for sulfate and phosphate contamination and that all reusable laboratory apparatus is thoroughly clean before use.

7.2.1 Disposable gloves, impermeable, to avoid the possibility of contamination from the hands and to protect them from contact with toxic and corrosive substances. PVC gloves are suitable. 7.2.2 Glassware, made of borosilicate glass 3.3, conforming to the requirements of ISO 3585, cleaned before use with water (6.1).

Alternatively, the glassware may be cleaned with a suitable phosphate-free laboratory detergent using a laboratory washing machine and afterwards rinsed thoroughly with water (6.1).

7.2.3 One-mark volumetric flasks, of capacities up to 2 l, conforming to the requirements of

ISO 1042.

7.2.4 Plastic labware:

  • Screw-cap polyethylene vessels proper for the extraction of collected sample filters.
  • Beakers, of appropriate capacity.
  • Disposable membrane filters, of PTFE, of pore size 0,45 µm, for use in ion chromatography.
  • Disposable syringes, of appropriate capacity, appropriate needles.
  • Autosampler vials, of appropriate capacity.
7.2.5 Piston-operated volumetric instruments, of capacities of 50 µl to 10 ml, conforming to the

requirements of ISO 8655-1, and tested in accordance with ISO 8655-6; pipettors, conforming to the requirements of ISO 8655-2, as an alternative to one-mark pipettes for the preparation of standard solutions, calibration solutions and dilution of samples.

7.2.6 Ultrasonic bath, preferably with a timer, suitable for use in the ultrasonic extraction method for sulfuric acid or phosphoric acid.

7.2.7 Ion chromatograph, having the following listed components inclusive. Components and tubing

that come into contact with the sample solution or eluent shall, as far as possible, be comprised of inert materials, e.g. polyetheretherketone (PEEK).

  • Pump, capable of delivering a constant flow within the range 0,1 ml/min to 5 ml at a pressure of 15 MPa to 150 MPa.
  • Eluent generation system, for producing an eluent suitable for use with the selected separator column, as an alternative to use of a manually prepared eluent (see, for example, Reference [23]).
  • Sample injection system, comprising a low dead-volume, non-metallic valve fitted with a sample loop, for injecting the sample solution into the eluent stream.
  • Guard and separator column, packed with high capacity pellicular anion exchange resin, suitable for resolving sulfates and phosphates from other inorganic anions.
  • Suppressor module (chemically suppressed or electronically suppressed) for ion chromatography, suitable for use with the separator column.
  • Conductivity detector, flow through, low volume, with a non-metallic flow path.

SIST ISO 21438-1:2026 ISO 21438-1:2022(E)

8 Occupational exposure assessment

[3] [10] Refer to relevant international, European or national standards (e.g. ISO 20581 , EN 689 , [8] ASTM E1370 ) for guidance on how to develop an appropriate assessment strategy and for general guidance on measurement strategy.

9 Sampling

9.1 Preliminary considerations

9.1.1 Selection and use of samplers

Select samplers (7.1.1) according to the requirements of the applicable limit value. Size selective samplers shall be designed to collect the appropriate fraction of airborne particles as defined in ISO 7708 and tested in accordance with EN 13205.

If possible, the samplers selected should be manufactured from conducting material, since samplers manufactured in non-conducting material have electrostatic properties that can influence representative sampling.

If the samplers selected have an internal filter cassette or cartridge that has to be rinsed during sample preparation (see 9.5.2 and 10.1.3), this cassette shall be manufactured from a material that does not react with acids.

Use the samplers at their design flow rate and in accordance with the instructions provided by the [9] manufacturer. See PD CEN/TR 15230 for further guidance.

9.1.2 Sampling period

Select a sampling period that is appropriate for the measurement task (see Clause 8), but ensure that it is long enough to enable sulfuric acid or phosphoric acid to be determined with acceptable uncertainty at levels of industrial hygiene significance. When high concentrations of airborne particles are anticipated, select a sampling period that is not so long as to risk overloading the filter with particulate matter. For example, estimate the minimum sampling time, t , in minutes, required to ensure that the min amount collected is above the lower limit of the working range of the analytical method when sulfuric acid or phosphoric acid is present in the test atmosphere at a concentration of 0,1 times its limit value, using Formula (1):

m lower t = (1) min q ××01, ρ V LV where m is the lower limit, in micrograms, of the analytical range; lower q is design flow rate, in litres per minute, of the sampler; V ρ is the limit value, in milligrams per cubic metre.

...

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 blank solution
  • 3.2 calibration blank solution
  • 3.3 calibration solution
  • 3.4 extraction solution
  • 3.5 sample dissolution
  • 3.6 sample solution
  • 3.7 stock standard solution
  • 3.8 test solution
  • 3.9 working standard solution
  • 4 Principle
  • 5 Requirement
  • 6 Reagents
  • 7 Apparatus
  • 7.1 Sampling equipment
  • 7.2 Laboratory apparatus
  • 8 Occupational exposure assessment
  • 9 Sampling
  • 9.1 Preliminary considerations
  • 9.2 Preparation for sampling
  • 9.3 Sampling position
  • 9.4 Collection of samples
  • 9.5 Transportation
  • 10 Analysis
  • 10.1 Preparation of test, calibration solutions and filter samples
  • 10.2 Instrumental analysis
  • 10.3 Estimation of detection and quantification limits
  • 10.4 Quality control
  • 10.5 Measurement uncertainty
  • 11 Expression of results
  • 12 Method performance
  • 12.1 Sample collection and stability
  • 12.2 Quantification limits
  • 12.3 Upper limits of the analytical range
  • 12.4 Bias and precision
  • 12.5 Uncertainty of sampling and analysis method
  • 12.6 Interferences
  • 13 Test report
  • 13.1 Test record
  • 13.2 Laboratory report
  • Annex A
  • Bibliography

Introduction

The health of workers in many industries is at risk through exposure by inhalation of particulate acids such as sulfuric acid or phosphoric acid compounds. Industrial hygienists and other public health professionals need to determine the effectiveness of measures taken to control workers’ exposure, and this is generally achieved by making workplace air measurements. This document has been published in order to make available a method for making valid exposure measurements for particulate acids in use in industry. It will be of benefit to: agencies concerned with health and safety at work; industrial hygienists and other public health professionals; analytical laboratories; and industrial users of sulfuric and phosphoric acids, and their workers.

It has been assumed in the drafting of the ISO 21438 series that the execution of its provisions and the interpretation of the results obtained are entrusted to appropriately qualified and experienced people.

INTERNATIONAL STANDARD ISO 21438-1:2022(E) Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 1:

Non-volatile acids (sulfuric acid and phosphoric acid)

1 Scope

This document specifies a method for the determination of the time-weighted average mass concentration of sulfuric acid and phosphoric acid in workplace air by ion chromatography. The anions are detected by conductivity.

The method is applicable to the personal sampling of airborne particles, as defined in ISO 7708, and to static (area) sampling.

The method does not apply to the determination of sulfur trioxide. The procedure does not differentiate between the acids and their corresponding salts if both are present in the workplace air.

The procedure does not differentiate between phosphoric acid and diphosphorus pentoxide (phosphoric anhydride) if both are present in the workplace air.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 1042, Laboratory glassware — One-mark volumetric flasks ISO 3585, Borosilicate glass 3.3 — Properties ISO 3696, Water for analytical laboratory use — Specification and test methods ISO 7708, Air quality — Particle size fraction definitions for health-related sampling ISO 8655-1, Piston-operated volumetric apparatus — Part 1: Terminology, general requirements and user recommendations ISO 8655-2, Piston-operated volumetric apparatus — Part 2: Pipettes ISO 8655-6, Piston-operated volumetric apparatus — Part 6: Gravimetric reference measurement procedure for the determination of volume ISO 18158, Workplace air — Terminology ISO 21832:2018, Workplace air — Metals and metalloids in airborne particles — Requirements for evaluation of measuring procedures EN 13205, Workplace exposure — Assessment of sampler performance for measurement of airborne particle concentrations

ISO 21438-1:2022(E)

3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 18158 and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

3.1 blank solution

solution prepared by taking a reagent blank, laboratory blank or field blank through the same procedure used for sample dissolution (3.5)

3.2 calibration blank solution

calibration solution (3.3) prepared without the addition of any working standard solution (3.9) Note 1 to entry: The concentration of sulfate and phosphate in the calibration blank solution is taken to be zero.

3.3 calibration solution

solution prepared by dilution of the working standard solution (3.9), containing sulfate and phosphate at concentrations that are suitable for use in calibration of the analytical instrument

3.4 extraction solution

solvent or solution used to solubilise the analyte(s) of interest

3.5 sample dissolution

process of obtaining a solution containing sulfate and phosphate from a sample, which might or might not involve complete dissolution of the sample

3.6 sample solution

solution prepared from a sample by the process of sample dissolution (3.5) Note 1 to entry: A sample solution can sometimes need to be subjected to further operations, e.g. dilution, in order to produce a test solution (3.8) that is ready for analysis.

3.7 stock standard solution

solution, used for preparation of the calibration solutions (3.3), containing sulfate and/or phosphate at a certified concentration that is traceable to national standards

3.8 test solution

blank solution (3.1) or sample solution (3.6) that has been subjected to all operations required to bring it into a state in which it is ready for analysis, e.g. dilution, filtered Note 1 to entry: The blank test solution is the blank solution and the sample test solution is the sample solution if these solutions are not subjected to any further operations before analysis.

3.9 working standard solution

solution, prepared by dilution of the stock standard solution(s) (3.7), that contains sulfate and phosphate at concentrations that are better suited to preparation of calibration solutions (3.3) than the concentration of sulfate and phosphate in the stock standard solutions

ISO 21438-1:2022(E)

4 Principle

A known volume of air is drawn through a filter to collect acid mist. The filter is mounted in a sampler designed to collect the required particle fraction of airborne particles (see 7.1.1). Currently, limit values in different countries exist for sulfuric acid either as “total particulate mass”, defined by the performance of a sampler, or in the inhalable or the thoracic size selective fraction. A suitable sampling device for the applicable particle fraction shall be used considering the existing limit value/particle fraction.

The collected sample is then treated with eluent and ultrasonic application, without heating, to extract sulfuric and phosphoric acids.

Aliquots of the sample solution are subjected to ion chromatography in order to separate the extracted sulfate and phosphate from other anions. Following this separation, the anions are measured using a conductivity detector.

Analytical results are obtained by plotting the measured conductivity as a function of concentration. They can be used for assessment of occupational exposure to sulfuric acid and phosphoric acid (and diphosphorus pentoxide) in air.

5 Requirement

The measuring procedure shall conform to any relevant international, European or national standard which specifies performance requirements for procedures for measuring chemical agents in workplace [3] air (e.g. ISO 20581 ).

6 Reagents

Use only reagents of recognized analytical grade and only water as specified in 6.1. It is advisable to check the blank values of all chemicals before use.

NOTE 1 Sulfates and phosphates are found ubiquitously in the environment and the presence of sulfates and phosphates in reagents can lead to high blank values.

6.1 Water, from a purification system that delivers ultrapure water of grade 1 (in accordance with ISO 3696).

NOTE 2 State of art water purification systems deliver water of grade 1 with higher quality than specified in ISO 3696:1987 (e.g. resistivity greater than 18 MΩ·cm).

6.2 Sodium carbonate (Na CO ), anhydrous, mass fraction > 99,9 %. 2 3 6.3 Sodium hydrogencarbonate (NaHCO ), anhydrous, mass fraction > 99,9 %.

NOTE 3 An alkaline solution (potassium hydroxide or sodium hydroxide) can be used in place of the buffer solution according to modifications to the method, but the resulting solution will require adjustment before analysis.

6.4 Sodium carbonate/sodium hydrogen carbonate extraction and eluent stock solution. Create a stock solution for the extraction and eluent solution. The stock solution has to be appropriate to the used application.

6.5 Sodium carbonate/sodium hydrogen carbonate extraction and eluent solution. The concentration of the extraction and eluent solution are typically matched, but this can be modified according to the laboratory procedure.

ISO 21438-1:2022(E) 6.6 Sulfate and phosphate stock solutions.

Use a commercial standard solution with a certificate of its concentration, e.g. 1 000 mg/l, traceable to national standards. Observe the manufacturer’s expiry date or recommended shelf-life. If necessary, dilute the standard solution (e.g. from 1 000 mg/l to 100 mg/l) to prepare the stock solution. 6.7 Sulfate and phosphate calibration solutions.

Pipette appropriate volumes of the stock solution to create the calibration. The calibration range shall be in accordance with the calculated working range (from ISO 21832:2018, 5.2.1) and the limit of quantification (from ISO 21832:2018, 8.1.2). A multi-point (the number of points to be consistent with the quality system of the laboratory) calibration shall be constructed within the linear range of the analytical method.

6.8 Sulfate and phosphate check-standard.

Use a certified reference material (CRM) as described in 10.4.3 with a certificate of its concentration, e.g. 400 mg/l for sulfate and 600 mg/l for phosphate, traceable to national standards. Observe the manufacturer’s expiry date or recommended shelf-life.

6.9 Sulfate and phosphate check-standard working solution.

The check-standard working solution should be in the middle of the calibration range. Pipette a proper volume of the sulfate and phosphate check-standard (6.8) to create a check-standard with the eluent (6.4). Prepare this solution fresh for every calibration. If necessary, create a check-standard working solution for sulfate and phosphate separately.

7 Apparatus

7.1 Sampling equipment

7.1.1 Samplers as required to collect samples consistent with applicable limit values. The

performance of samplers shall be in accordance with EN 13205 where specific size-selective fractions are stipulated.

The operating instructions supplied by the manufacturer should be consulted to find out whether particulate matter deposited on the internal surfaces of the sampler forms part of the sample. NOTE 1 In general, personal samplers for collection of the inhalable fraction of airborne particles do not exhibit the same size-selective characteristics if used for static sampling. NOTE 2 Some inhalable samplers are designed to collect the inhalable fraction of airborne particles on the filter, and any particulate matter deposited on the internal surfaces of the sampler is not of interest. Other inhalable samplers are designed such that airborne particles which pass through the entry orifice(s) match the inhalable convention, in which case particulate matter deposited on the internal surfaces of the sampler does form part of the sample. (Samplers of this second type generally incorporate an internal filter cassette or cartridge that can be removed from the sampler to enable this material to be easily recovered.) [9] NOTE 3 PD CEN/TR 15230 gives examples of inhalable samplers with the potential to meet the requirements of EN 13205, including published reports on their performance. NOTE 4 Reference [13] discusses the relationship between inhalable and thoracic fractions of sulfuric acid in workplaces.

ISO 21438-1:2022(E) 7.1.2 Filters, of a diameter suitable for use with the samplers, with a collection efficiency ≥ 99 % for particles with a 0,3 µm diffusion diameter and manufactured from a material that is compatible with the sample preparation and analysis method.

Sulfuric acid and phosphoric acid are strong acids. They react (e.g. by dehydration) with many organic and polymeric materials, and destroy the filter material. Therefore, correct selection of the filter used for sample collection is of paramount importance. In particular, the filter has to be manufactured from a material that does not react with the acids. Also, some filters (e.g. glass fibre filters) can contain metals (e.g. barium) that react with sulfuric and/or phosphoric acids to produce insoluble salts. The following filter types are generally suitable for use:

  • binder-free quartz fibre filters;
  • polyvinyl chloride (PVC) membrane filters, of pore size 5 µm or less;
  • polytetrafluoroethylene (PTFE) membrane filters, of pore size 5 µm or less; Sulfates and phosphates are found ubiquitously in the environment and the presence of sulfates and phosphates in filter materials can lead to high blank values. It is therefore essential to check the blank values of each batch of filters used.

[2] 7.1.3 Sampling pumps, in accordance with ISO 13137 , should be used. [2] If the sampling pump is used outside the range of conditions specified in ISO 13137 appropriate action should be taken to ensure that the performance requirements are met. For instance, at sub-zero temperatures it can be necessary to keep the pump warm.

7.1.4 Flowmeter, portable, with an accuracy that is sufficient to enable the volumetric flow rate to be measured to within ±5 %.

The calibration of the flowmeter shall be checked against a primary standard, i.e. a flowmeter whose accuracy is traceable to national standards. If appropriate (see 9.1.3), record the atmospheric temperature and pressure at which the calibration of the flowmeter was checked. It is advisable that the flowmeter used is capable of measuring the volumetric flow rate to within ±2 % or better.

7.1.5 Ancillary equipment:

  • Flexible tubing, e.g. silicone of an outer diameter of 10 mm and an inner diameter of 6 mm for making a leakproof connection from the samplers to the sampling pumps without compressing as a result of the pressure difference between ambient atmosphere and the in-line vacuum at the used −1 flow rate of 3,5 l min .
  • Belts or harnesses, which the sampling pumps can conveniently fix for personal sampling.
  • Tweezers, manufactured from or tipped with PTFE, for handling filters.
  • Filter transport cassettes, or similar, if required (see 9.5.1), in which to transport samples to the laboratory.
  • Thermometer, of range 0 °C to 50 °C, graduated in divisions of 1 °C or less, for measurement of atmospheric temperature. For applications at temperatures below freezing, the range of the thermometer shall extend to the appropriate desired range.
  • Barometer, suitable for measurement of atmospheric pressure.

ISO 21438-1:2022(E)

7.2 Laboratory apparatus

CAUTION — Both sulfates and phosphates are found ubiquitously in the environment. This can lead to elevated blanks so it is especially important to take great care that all disposable plastic labware is checked for sulfate and phosphate contamination and that all reusable laboratory apparatus is thoroughly clean before use.

7.2.1 Disposable gloves, impermeable, to avoid the possibility of contamination from the hands and to protect them from contact with toxic and corrosive substances. PVC gloves are suitable. 7.2.2 Glassware, made of borosilicate glass 3.3, conforming to the requirements of ISO 3585, cleaned before use with water (6.1).

Alternatively, the glassware may be cleaned with a suitable phosphate-free laboratory detergent using a laboratory washing machine and afterwards rinsed thoroughly with water (6.1).

7.2.3 One-mark volumetric flasks, of capacities up to 2 l, conforming to the requirements of

ISO 1042.

7.2.4 Plastic labware:

  • Screw-cap polyethylene vessels proper for the extraction of collected sample filters.
  • Beakers, of appropriate capacity.
  • Disposable membrane filters, of PTFE, of pore size 0,45 µm, for use in ion chromatography.
  • Disposable syringes, of appropriate capacity, appropriate needles.
  • Autosampler vials, of appropriate capacity.
7.2.5 Piston-operated volumetric instruments, of capacities of 50 µl to 10 ml, conforming to the

requirements of ISO 8655-1, and tested in accordance with ISO 8655-6; pipettors, conforming to the requirements of ISO 8655-2, as an alternative to one-mark pipettes for the preparation of standard solutions, calibration solutions and dilution of samples.

7.2.6 Ultrasonic bath, preferably with a timer, suitable for use in the ultrasonic extraction method for sulfuric acid or phosphoric acid.

7.2.7 Ion chromatograph, having the following listed components inclusive. Components and tubing

that come into contact with the sample solution or eluent shall, as far as possible, be comprised of inert materials, e.g. polyetheretherketone (PEEK).

  • Pump, capable of delivering a constant flow within the range 0,1 ml/min to 5 ml at a pressure of 15 MPa to 150 MPa.
  • Eluent generation system, for producing an eluent suitable for use with the selected separator column, as an alternative to use of a manually prepared eluent (see, for example, Reference [23]).
  • Sample injection system, comprising a low dead-volume, non-metallic valve fitted with a sample loop, for injecting the sample solution into the eluent stream.
  • Guard and separator column, packed with high capacity pellicular anion exchange resin, suitable for resolving sulfates and phosphates from other inorganic anions.
  • Suppressor module (chemically suppressed or electronically suppressed) for ion chromatography, suitable for use with the separator column.
  • Conductivity detector, flow through, low volume, with a non-metallic flow path.

ISO 21438-1:2022(E)

8 Occupational exposure assessment

[3] [10] Refer to relevant international, European or national standards (e.g. ISO 20581 , EN 689 , [8] ASTM E1370 ) for guidance on how to develop an appropriate assessment strategy and for general guidance on measurement strategy.

9 Sampling

9.1 Preliminary considerations

9.1.1 Selection and use of samplers

Select samplers (7.1.1) according to the requirements of the applicable limit value. Size selective samplers shall be designed to collect the appropriate fraction of airborne particles as defined in ISO 7708 and tested in accordance with EN 13205.

If possible, the samplers selected should be manufactured from conducting material, since samplers manufactured in non-conducting material have electrostatic properties that can influence representative sampling.

If the samplers selected have an internal filter cassette or cartridge that has to be rinsed during sample preparation (see 9.5.2 and 10.1.3), this cassette shall be manufactured from a material that does not react with acids.

Use the samplers at their design flow rate and in accordance with the instructions provided by the [9] manufacturer. See PD CEN/TR 15230 for further guidance.

9.1.2 Sampling period

Select a sampling period that is appropriate for the measurement task (see Clause 8), but ensure that it is long enough to enable sulfuric acid or phosphoric acid to be determined with acceptable uncertainty at levels of industrial hygiene significance. When high concentrations of airborne particles are anticipated, select a sampling period that is not so long as to risk overloading the filter with particulate matter. For example, estimate the minimum sampling time, t , in minutes, required to ensure that the min amount collected is above the lower limit of the working range of the analytical method when sulfuric acid or phosphoric acid is present in the test atmosphere at a concentration of 0,1 times its limit value, using Formula (1):

m lower t = (1) min q ××01, ρ V LV where m is the lower limit, in micrograms, of the analytical range; lower q is design flow rate, in litres per minute, of the sampler; V ρ is the limit value, in milligrams per cubic metre.

...

Contents

  • Introduction
  • 1 Domaine d’application
  • 2 Références normatives
  • 3 Termes et définitions
  • 3.1 solution de blanc
  • 3.2 solution de blanc d’étalonnage
  • 3.3 solution d’étalonnage
  • 3.4 solution d’extraction
  • 3.5 mise en solution de l’échantillon
  • 3.6 solution d’échantillon
  • 3.7 solution étalon mère
  • 3.8 solution d’essai
  • 3.9 solution étalon de travail
  • 4 Principe
  • 5 Exigence
  • 6 Réactifs
  • 6.4 Solutions mères d’extraction et d’éluant carbonate de sodium/hydrogénocarbonate de
  • 7 Appareillage
  • 7.1 Matériel de prélèvement
  • 7.2 Appareillage de laboratoire
  • 8 Évaluation de l’exposition professionnelle
  • 9 Prélèvement
  • 9.1 Facteurs préliminaires à considérer
  • 9.2 Préparation du prélèvement
  • 9.3 Point de prélèvement
  • 9.4 Prélèvement des échantillons
  • 9.5 Transport
  • 10 Analyse
  • 10.1 Préparation des solutions d’essai et d’étalonnage et des filtres échantillons
  • 10.2 Analyse instrumentale
  • 10.3 Estimation des limites de détection et de quantification
  • 10.4 Contrôle qualité
  • 10.5 Incertitude de mesure
  • 11 Expression des résultats
  • 12 Performances de la méthode
  • 12.1 Prélèvement et conservation de l’échantillon
  • 12.2 Limites de quantification
  • 12.3 Limites supérieures de la gamme d’analyse
  • 12.4 Biais et fidélité
  • 12.5 Incertitude de la méthode de prélèvement et d’analyse
  • 12.6 Interférences
  • 13 Rapport d’essai
  • 13.1 Procès-verbal de l’essai
  • 13.2 Rapport de laboratoire
  • Annexe A
  • Bibliographie

Introduction

Dans de nombreuses industries, la santé des travailleurs est menacée en raison de l’exposition par inhalation d’acides particulaires tels que des composés d’acide sulfurique ou phosphorique. Les hygiénistes industriels et autres professionnels de santé publique ont besoin de déterminer l’efficacité des mesures prises pour contrôler l’exposition des travailleurs, généralement en effectuant des mesurages de l’air des lieux de travail. Le présent document vise à fournir une méthode permettant d’effectuer des mesurages d’exposition valides pour les acides particulaires utilisés dans l’industrie. Elle s’adresse aux agences concernées par l’hygiène et la sécurité du travail, aux hygiénistes industriels et autres professionnels de santé publique, aux laboratoires d’analyse, aux industriels utilisateurs d’acides sulfurique et phosphorique ainsi qu’à leurs employés. Lors de l’élaboration de la série ISO 21438, il a été présumé que l’exécution de ses dispositions et l’interprétation des résultats obtenus sont confiées à des personnes dûment qualifiées et expérimentées.

NORME INTERNATIONALE ISO 21438-1:2022(F) Air des lieux de travail — Détermination des acides inorganiques par chromatographie ionique — Partie 1:

Acides non volatils (acide sulfurique et acide phosphorique)

1 Domaine d’application

Le présent document spécifie une méthode qui permet de déterminer, par chromatographie ionique, la concentration massique moyenne pondérée dans le temps des acides sulfurique et phosphorique dans l’air des lieux de travail. Les anions sont détectés par conductivité. La méthode s’applique au prélèvement individuel des particules en suspension dans l’air, telles que définies dans l’ISO 7708, ainsi qu’au prélèvement à point fixe. La méthode ne s’applique pas à la détermination du trioxyde de soufre. La procédure ne permet pas de distinguer les acides et leurs sels correspondants, si tous sont présents dans l’air des lieux de travail.

La procédure ne permet pas de distinguer l’acide phosphorique et le pentoxyde de diphosphore (anhydride phosphorique), s’ils sont tous deux présents dans l’air des lieux de travail.

2 Références normatives

Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique. Pour les références non datées, la dernière édition du document de référence s’applique (y compris les éventuels amendements).

ISO 1042, Verrerie de laboratoire — Fioles jaugées à un trait ISO 3585, Verre borosilicaté 3.3 — Propriétés ISO 3696, Eau pour laboratoire à usage analytique — Spécification et méthodes d’essai ISO 7708, Qualité de l’air — Définitions des fractions de taille des particules pour l’échantillonnage lié aux problèmes de santé ISO 8655-1, Appareils volumétriques à piston — Partie 1: Définitions, exigences générales et recommandations pour l’utilisateur ISO 8655-2, Appareils volumétriques à piston — Partie 2: Pipettes à piston ISO 8655-6, Appareils volumétriques à piston — Partie 6: Méthodes gravimétriques pour la détermination de l’erreur de mesure ISO 18158, Qualité de l’air — Terminologie ISO 21832:2018, Air des lieux de travail — Métaux et métalloïdes dans les particules en suspension dans l’air — Exigences relatives à l’évaluation des procédures de mesure

ISO 21438-1:2022(F) EN 13205, Exposition sur les lieux de travail — Évaluation des performances des dispositifs de prélèvement pour le mesurage des concentrations de particules en suspension dans l’air

3 Termes et définitions

Pour les besoins du présent document, les termes et les définitions de l’ISO 18158 ainsi que les suivants s’appliquent.

L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation, consultables aux adresses suivantes:

  • ISO Online browsing platform: disponible à l’adresse https:// www .iso .org/ obp
  • IEC Electropedia: disponible à l’adresse https:// www .electropedia .org/

3.1 solution de blanc

solution préparée à partir d’un blanc de réactifs, d’un blanc de laboratoire ou d’un blanc de terrain en utilisant la même procédure que pour la mise en solution de l’échantillon (3.5)

3.2 solution de blanc d’étalonnage

solution d’étalonnage (3.3) préparée sans ajout d’une solution étalon de travail (3.9) Note 1 à l'article: La concentration de sulfate et de phosphate dans la solution de blanc d’étalonnage est considérée comme nulle.

3.3 solution d’étalonnage

solution obtenue par dilution de la solution étalon de travail (3.9), contenant du sulfate et du phosphate à des concentrations appropriées à l’étalonnage de l’instrument d’analyse

3.4 solution d’extraction

solvant ou solution utilisé pour solubiliser l’analyte ou les analytes considérés

3.5 mise en solution de l’échantillon

processus permettant d’obtenir une solution contenant du sulfate et du phosphate à partir d’un échantillon, processus pouvant impliquer ou non la mise en solution complète de l’échantillon

3.6 solution d’échantillon

solution préparée au moyen du processus de mise en solution de l’échantillon (3.5) Note 1 à l'article: Une solution d’échantillon peut parfois nécessiter des traitements ultérieurs, par exemple une dilution, afin d’obtenir une solution d’essai (3.8) prête à l’analyse.

3.7 solution étalon mère

solution, utilisée dans la préparation des solutions d’étalonnage (3.3), contenant du sulfate et/ou du phosphate à une concentration certifiée dont la traçabilité aux étalons nationaux est établie

3.8 solution d’essai

solution de blanc (3.1) ou solution d’échantillon (3.6) ayant été soumise à toutes les opérations permettant de la préparer à l’analyse, par exemple une dilution ou une filtration Note 1 à l'article: Si ces solutions ne sont pas soumises à des traitements ultérieurs avant l’analyse, la solution d’essai à blanc est la solution de blanc et la solution d’essai d’échantillon est la solution d’échantillon.

ISO 21438-1:2022(F)

3.9 solution étalon de travail

solution préparée par dilution d’une ou de plusieurs solutions étalons mères (3.7) contenant du sulfate et du phosphate à des concentrations mieux adaptées à la préparation des solutions d’étalonnage (3.3) que la concentration en sulfate et en phosphate des solutions étalons mères

4 Principe

Un volume défini d’air est prélevé à travers un filtre afin de recueillir le brouillard d’acide. Le filtre est placé dans un dispositif de prélèvement conçu pour recueillir la fraction requise des particules en suspension dans l’air (voir 7.1.1).

À l’heure actuelle, différents pays ont fixé des valeurs limites pour l’acide sulfurique qui s’appliquent soit à la «masse particulaire totale», définie par les performances d’un dispositif de prélèvement, soit à la fraction inhalable ou la fraction thoracique. Dans ce cas, un dispositif de prélèvement adapté doit être utilisé en tenant compte de la fraction conventionnelle pour laquelle la valeur limite est définie. L’échantillon recueilli est ensuite traité avec l’éluant et soumis aux ultrasons, sans chauffage, de façon à extraire les acides sulfurique et phosphorique.

Des aliquotes de la solution d’échantillon sont traitées par chromatographie ionique pour séparer le sulfate et le phosphate extraits des autres anions. À la suite de cette séparation, les anions sont mesurés à l’aide d’un détecteur conductimétrique.

Les résultats d’analyse sont obtenus en reportant la conductivité mesurée en fonction de la concentration. Ils peuvent être utilisés pour évaluer l’exposition professionnelle aux acides sulfurique et phosphorique (et au pentoxyde de diphosphore) dans l’air.

5 Exigence

La procédure de mesure doit être conforme à toute Norme internationale, européenne ou nationale pertinente qui spécifie des exigences de performances pour les procédures de mesure des agents [3] chimiques présents dans l’air des lieux de travail (ISO 20581 , par exemple).

6 Réactifs

Utiliser uniquement des réactifs de qualité analytique reconnue et de l’eau telle que spécifiée en 6.1. Il est recommandé de vérifier les valeurs de blanc de tous les produits chimiques avant utilisation. NOTE 1 Les sulfates et les phosphates sont omniprésents dans l’environnement et leur présence dans les réactifs peut se traduire par des valeurs de blanc élevées.

6.1 Eau, obtenue au moyen d’un système de purification fournissant de l’eau ultra-pure de qualité 1 (conformément à l’ISO 3696).

NOTE 2 Les systèmes de purification d’eau actuels du marché fournissent de l’eau de qualité 1 dont la qualité est supérieure à celle spécifiée dans l’ISO 3696:1987 (résistivité supérieure à 18 MΩ·cm, par exemple). 6.2 Carbonate de sodium (Na CO ), anhydre, fraction massique > 99,9 %. 2 3 6.3 Hydrogénocarbonate de sodium (NaHCO ), anhydre, fraction massique > 99,9 %.

NOTE 3 Une solution alcaline (hydroxyde de potassium ou de sodium) peut être utilisée à la place de la solution tampon après modification de la méthode, mais la solution résultante nécessitera un ajustement avant analyse.

ISO 21438-1:2022(F)

6.4 Solutions mères d’extraction et d’éluant carbonate de sodium/hydrogénocarbonate de

sodium.

Préparer une solution mère pour les solutions d’extraction et d’éluant. La solution mère doit être adaptée à l’application utilisée.

6.5 Solutions d’extraction et d’éluant carbonate de sodium/hydrogénocarbonate de sodium. Les concentrations des solutions d’extraction et d’éluant sont généralement identiques, mais elles peuvent être modifiées selon la procédure du laboratoire.

6.6 Solutions mères de sulfate et de phosphate.

Utiliser une solution étalon du commerce de concentration certifiée (1 000 mg/l, par exemple) et dont la traçabilité aux étalons nationaux est établie. Respecter la date d’expiration fixée par le fabricant ou la durée de conservation recommandée. Si nécessaire, diluer la solution étalon (par exemple, de 1 000 mg/l à 100 mg/l) pour préparer la solution mère.

6.7 Solutions d’étalonnage de sulfate et de phosphate.

Transférer à la pipette des volumes appropriés de la solution mère pour obtenir la solution d’étalonnage. Le domaine d’étalonnage doit être conforme au domaine de travail calculé (à partir de l’ISO 21832:2018, 5.2.1) et à la limite de quantification (déterminée à partir de l’ISO 21832:2018, 8.1.2). Un étalonnage multipoint (le nombre de points doit être en cohérence avec le système qualité du laboratoire) doit être réalisé dans les limites du domaine de linéarité de la méthode d’analyse. 6.8 Étalon de contrôle de sulfate et de phosphate.

Utiliser un matériau de référence certifié (MRC) tel que défini en 10.4.3, de concentration certifiée (par exemple 400 mg/l pour le sulfate et 600 mg/l pour le phosphate) et dont la traçabilité aux étalons nationaux est établie. Respecter la date d’expiration fixée par le fabricant ou la durée de conservation recommandée.

6.9 Solution de travail pour l’étalon de contrôle de sulfate et de phosphate. Il convient que la solution de travail pour l’étalon de contrôle soit située au milieu de la gamme d’étalonnage. Transférer à la pipette un volume approprié des étalons de contrôle de sulfate et de phosphate 6.8 afin d’obtenir un étalon de contrôle dans l’éluant (6.4). Préparer une solution fraîche pour chaque étalonnage. Si nécessaire, préparer une solution de travail pour étalon de contrôle séparément pour le sulfate et le phosphate.

7 Appareillage

7.1 Matériel de prélèvement

7.1.1 Dispositifs de prélèvement selon les exigences de collecte des échantillons et conformément aux valeurs limites applicables. Les performances des dispositifs de prélèvement doivent être conformes à l’EN 13205 lorsque des fractions sélectives en taille spécifiques sont stipulées. Il convient de consulter la notice d’utilisation fournie par le fabricant pour déterminer si la matière particulaire déposée sur les surfaces internes du dispositif de prélèvement fait partie de l’échantillon. NOTE 1 En général, les dispositifs de prélèvement individuels utilisés pour recueillir la fraction inhalable des particules en suspension dans l’air n’ont pas les mêmes caractéristiques de sélection de taille s’ils sont utilisés pour un prélèvement à point fixe.

ISO 21438-1:2022(F) NOTE 2 Certains dispositifs de prélèvement sont conçus pour recueillir la fraction inhalable des particules en suspension dans l’air sur le filtre et la matière particulaire déposée sur les surfaces internes du dispositif de prélèvement ne présente alors aucun intérêt. D’autres dispositifs de prélèvement de la fraction inhalable sont conçus de sorte que les particules en suspension dans l’air qui passent dans les orifices d’entrée correspondent à la convention inhalable, auquel cas la matière particulaire déposée sur les surfaces internes du dispositif de prélèvement fait partie de l’échantillon (les dispositifs de prélèvement de ce type comportent généralement une cassette porte-filtre interne ou une cartouche amovible afin de faciliter la récupération de la matière déposée). [9] NOTE 3 Le PD CEN/TR 15230 donne des exemples de dispositifs de prélèvement de la fraction inhalable susceptibles de satisfaire aux exigences de l’EN 13205, y compris des rapports publiés sur leurs performances. NOTE 4 La référence [13] décrit la relation entre les fractions inhalable et thoracique de l’acide sulfurique sur les lieux de travail.

7.1.2 Filtres, d’un diamètre compatible avec les dispositifs de prélèvement, présentant une efficacité de prélèvement ≥ 99 % pour les particules dont le diamètre de diffusion est égal à 0,3 µm et fabriqués à partir d’un matériau compatible avec la méthode de préparation et d’analyse de l’échantillon. Les acides sulfurique et phosphorique sont des acides forts. Ils réagissent (par exemple par déshydratation) avec la plupart des matières organiques et polymères et détruisent le matériau filtrant. Par conséquent, il est essentiel de choisir le filtre adapté au prélèvement de l’échantillon. En particulier, le filtre doit être fabriqué à partir d’un matériau ne réagissant pas aux acides. De même, certains filtres (par exemple les filtres en fibres de verre) peuvent contenir des métaux (du baryum par exemple) qui réagissent avec les acides sulfurique et/ou phosphorique pour produire des sels insolubles. Les types de filtres cités ci-dessous conviennent généralement:

  • les filtres en fibres de quartz sans liant;
  • les membranes filtrantes en polychlorure de vinyle (PVC), de porosité inférieure ou égale à 5 µm;
  • les membranes filtrantes en polytétrafluoroéthylène (PTFE), de porosité inférieure ou égale à 5 µm. Les sulfates et les phosphates sont omniprésents dans l’environnement et leur présence dans les matériaux du filtre peut se traduire par des valeurs de blanc élevées. Il est donc essentiel de vérifier les valeurs de blanc de chaque lot de filtres utilisé.

[2] 7.1.3 Pompes de prélèvement, conformes aux exigences de l’ISO 13137 . [2] Si la pompe de prélèvement est utilisée en dehors de la plage de conditions définies dans l’ISO 13137 , il convient de prendre des mesures appropriées pour s’assurer que les exigences de performances sont satisfaites. Par exemple, à des températures inférieures à zéro, il peut s’avérer nécessaire de maintenir la pompe au chaud.

7.1.4 Débitmètre, portable, offrant une précision suffisante pour permettre de mesurer le débit

volumétrique à ±5 %.

L’étalonnage du débitmètre doit être vérifié par rapport à un étalon primaire, c’est-à-dire un débitmètre dont la traçabilité de la précision aux étalons nationaux est établie. S’il y a lieu (voir 9.1.3), noter la température et la pression atmosphériques auxquelles l’étalonnage du débitmètre a été vérifié. Il est recommandé que le débitmètre utilisé soit capable de mesurer le débit volumétrique à ±2 % au minimum.

7.1.5 Matériel auxiliaire:

  • Tubes flexibles, par exemple en silicone, ayant un diamètre extérieur de 10 mm et un diamètre intérieur de 6 mm pour raccorder les dispositifs de prélèvement aux pompes de prélèvement de manière étanche et sans compression résultant de la différence de pression entre l’atmosphère −1 ambiante et la dépression créée dans le circuit au débit utilisé de 3,5 l min .

ISO 21438-1:2022(F) - Ceintures ou harnais, auxquels les pompes de prélèvement peuvent être facilement fixées pour le prélèvement individuel.

  • Pinces en PTFE ou à embouts en PTFE, pour la manipulation des filtres.
  • Cassettes de transport de filtre, ou matériel similaire, si nécessaire (voir 9.5.1), permettant de transporter les échantillons jusqu’au laboratoire.
  • Thermomètre, de 0 °C à 50 °C, gradué en échelons de 1 °C ou moins, pour mesurer la température atmosphérique. Pour des applications à des températures inférieures au point de congélation, la plage du thermomètre doit couvrir la plage de mesure correspondante souhaitée.
  • Baromètre, adapté au mesurage de la pression atmosphérique.

7.2 Appareillage de laboratoire

AVERTISSEMENT — Les sulfates et les phosphates sont omniprésents dans l’environnement, ce qui peut se traduire par des niveaux de blanc élevés. Il est donc particulièrement important de contrôler soigneusement la verrerie de laboratoire en plastique à usage unique afin de déceler toute contamination aux sulfates et aux phosphates, et de nettoyer complètement tous les appareillages de laboratoire réutilisables avant utilisation.

7.2.1 Gants à usage unique, imperméables, pour éviter toute possibilité de contamination par les

mains et pour les protéger de tout contact avec des substances toxiques et corrosives. Les gants en PVC conviennent.

7.2.2 Verrerie, en verre borosilicaté 3.3, conforme aux exigences de l’ISO 3585, nettoyée à l’eau (6.1) avant utilisation.

Il est également permis de nettoyer la verrerie avec un détergent de laboratoire adapté, sans phosphates, en utilisant un lave-vaisselle de laboratoire, puis en la rinçant soigneusement à l’eau (6.1).

7.2.3 Fioles jaugées à un trait, d’une capacité allant jusqu’à 2 l, conformes aux exigences

de l’ISO 1042.

7.2.4 Matériel de laboratoire en plastique:

  • Récipients en polyéthylène à bouchon à vis, adaptés pour l’extraction des échantillons collectés sur les filtres.
  • Béchers, de capacité appropriée.
  • Membranes filtrantes à usage unique, en PTFE, d’une porosité de 0,45 µm, destinées à la chromatographie ionique.
  • Seringues jetables, de capacité appropriée, avec aiguilles adaptées.
  • Flacons de passeur automatique, de capacité appropriée.
7.2.5 Appareils volumétriques à piston, d’une capacité de 50 µl à 10 ml, conformes aux exigences

de l’ISO 8655-1 et soumis à essai conformément à l’ISO 8655-6; possibilité d’utiliser des pipettes conformes aux exigences de l’ISO 8655-2, à la place des pipettes à un trait pour la préparation des solutions étalons, des solutions d’étalonnage et la dilution des échantillons.

7.2.6 Cuve à ultrasons, de préférence équipée d’une minuterie, permettant d’appliquer la méthode

d’extraction par ultrasons de l’acide sulfurique ou phosphorique.

ISO 21438-1:2022(F)

7.2.7 Système de chromatographie ionique, comprenant les composants énumérés ci-après. Les

composants et les tuyaux qui entrent en contact avec la solution d’échantillon ou l’éluant doivent, dans la mesure du possible, être constitués de matériaux inertes (polyétheréthercétone (PEEK) par exemple). - Pompe, capable de délivrer un débit constant dans la plage comprise entre 0,1 ml/min et 5 ml/min à une pression de 15 MPa à 150 MPa.

  • Système de génération d’éluant permettant de produire un éluant compatible avec la colonne de séparation sélectionnée, comme alternative à l’utilisation d’un éluant préparé manuellement (voir, par exemple, la référence [23]).

...