Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-3: Particular requirements - Test configuration, operational conditions and performance criteria for transducers with integrated or remote signal conditioning

IEC 61326-2-3:2020 is available as IEC 61326-2-3:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61326-2-3:2020 specifies more detailed test configurations, operational conditions and performance criteria for transducers with integrated or remote signal conditioning. This document applies only to transducers characterized by their ability to transform, with the aid of an auxiliary energy source, a non-electric quantity to a process-relevant electrical signal, and to output the signal at one or more PORTS. This document includes transducers for electrochemical and biological measured quantities. The transducers covered by this document can be powered by AC or DC voltage and/or by battery or with internal power supply.

Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen - Teil 2-3: Besondere Anforderungen - Prüfanordnung, Betriebsbedingungen und Leistungsmerkmale für Messgrößenumformer mit integrierter oder abgesetzter Signalaufbereitung

Matériel électrique de mesure, de commande et de laboratoire - Exigences relatives à la CEM - Partie 2-3: Exigences particulières - Configurations d’essai, conditions de fonctionnement et critères de performance des transducteurs avec un système de conditionnement du signal intégré ou à distance

IEC 61326-2-3:2020 est disponible sous forme de IEC 61326-2-3:2020 RLV qui contient la Norme internationale et sa version Redline, illustrant les modifications du contenu technique depuis l'édition précédente.L’IEC 61326-2-3:2020 spécifie de façon plus détaillée les configurations d’essai, les conditions de fonctionnement et les critères de performance des transducteurs avec un système de conditionnement du signal intégré ou à distance. Le présent document s’applique uniquement aux transducteurs caractérisés par leur capacité à transformer, avec l’aide d’une source d’énergie auxiliaire, une grandeur non électrique en un signal électrique approprié pour un processus, et à fournir un signal sur un ou plusieurs ACCES. Le présent document inclut les transducteurs pour le mesurage de grandeurs électrochimiques et biologiques. Les transducteurs couverts par ce document peuvent être alimentés par une tension alternative ou continue et/ou par batterie ou par une alimentation interne.

Električna oprema za merjenje, nadzor in laboratorijsko uporabo - Zahteve za elektromagnetno združljivost (EMC) - 2-3. del: Posebne zahteve - Preskusna konfiguracija, obratovalni pogoji in merila učinkovitosti za delovanje pretvornikov z vgrajenim ali daljinskim kondicioniranjem signalov (IEC 61326-2-3:2020)

General Information

Status
Published
Publication Date
03-Jun-2021
Current Stage
6060 - Document made available - Publishing
Start Date
04-Jun-2021
Completion Date
04-Jun-2021

Relations

Buy Standard

Standard
EN IEC 61326-2-3:2021
English language
25 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-oktober-2021
Nadomešča:
SIST EN 61326-2-3:2013
Električna oprema za merjenje, nadzor in laboratorijsko uporabo - Zahteve za
elektromagnetno združljivost (EMC) - 2-3. del: Posebne zahteve - Preskusna
konfiguracija, obratovalni pogoji in merila učinkovitosti za delovanje pretvornikov
z vgrajenim ali daljinskim kondicioniranjem signalov (IEC 61326-2-3:2020)
Electrical equipment for measurement, control and laboratory use - EMC requirements -
Part 2-3: Particular requirements - Test configuration, operational conditions and
performance criteria for transducers with integrated or remote signal conditioning (IEC
61326-2-3:2020)
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen - Teil 2-3:
Besondere Anforderungen - Prüfanordnung, Betriebsbedingungen und
Leistungsmerkmale für Messgrößenumformer mit integrierter oder abgesetzter
Signalaufbereitung (IEC 61326-2-3:2020)
Matériel électrique de mesure, de commande et de laboratoire - Exigences relatives à la
CEM - Partie 2-3: Exigences particulières - Configurations d'essai, conditions de
fonctionnement et critères de performance des transducteurs avec un système de
conditionnement du signal intégré ou à distance (IEC 61326-2-3:2020)
Ta slovenski standard je istoveten z: EN IEC 61326-2-3:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
33.100.01 Elektromagnetna združljivost Electromagnetic compatibility
na splošno in general
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD EN IEC 61326-2-3

NORME EUROPÉENNE
EUROPÄISCHE NORM
June 2021
ICS 25.040.40; 17.220.20; 33.100.20 Supersedes EN 61326-2-3:2013 and all of its
amendments and corrigenda (if any)
English Version
Electrical equipment for measurement, control and laboratory
use - EMC requirements - Part 2-3: Particular requirements -
Test configuration, operational conditions and performance
criteria for transducers with integrated or remote signal
conditioning
(IEC 61326-2-3:2020)
Matériel électrique de mesure, de commande et de Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-
laboratoire - Exigences relatives à la CEM - Partie 2-3: Anforderungen - Teil 2-3: Besondere Anforderungen -
Exigences particulières - Configurations d'essai, conditions Prüfanordnung, Betriebsbedingungen und
de fonctionnement et critères de performance des Leistungsmerkmale für Messgrößenumformer mit
transducteurs avec un système de conditionnement du integrierter oder abgesetzter Signalaufbereitung
signal intégré ou à distance (IEC 61326-2-3:2020)
(IEC 61326-2-3:2020)
This European Standard was approved by CENELEC on 2020-11-30. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.

European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 61326-2-3:2021 E

European foreword
The text of document 65A/980/FDIS, future edition 3 of IEC 61326-2-3, prepared by SC 65A "System
aspects" of IEC/TC 65 "Industrial-process measurement, control and automation" was submitted to the
IEC-CENELEC parallel vote and approved by CENELEC as EN IEC 61326-2-3:2021.
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2021-12-04
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2024-06-04
document have to be withdrawn
This document supersedes EN 61326-2-3:2013 and all of its amendments and corrigenda (if any).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association.
Endorsement notice
The text of the International Standard IEC 61326-2-3:2020 was approved by CENELEC as a
European Standard without any modification.

Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1 Where an International Publication has been modified by common modifications, indicated by (mod), the
relevant EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is available
here: www.cenelec.eu.
The Annex ZA of EN IEC 61326-1:2021 applies with the following additions:
Publication Year Title EN/HD Year
IEC 61326-1 2020 Electrical equipment for measurement, EN IEC 61326-1 2021
control and laboratory use - EMC
requirements - Part 1: General
requirements
IEC 61326-2-3 ®
Edition 3.0 2020-10
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Electrical equipment for measurement, control and laboratory use –

EMC requirements –
Part 2-3: Particular requirements – Test configuration, operational conditions

and performance criteria for transducers with integrated or remote signal

conditioning
Matériel électrique de mesure, de commande et de laboratoire –

Exigences relatives à la CEM –

Partie 2-3: Exigences particulières – Configurations d’essai, conditions de

fonctionnement et critères de performance des transducteurs avec un système

de conditionnement du signal intégré ou à distance

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 17.220.20; 25.040.40; 33.100.20 ISBN 978-2-8322-8980-8

– 2 – IEC 61326-2-3:2020 © IEC 2020
CONTENTS
FOREWORD . 4
1 Scope . 6
2 Normative references . 8
3 Terms and definitions . 8
4 General . 9
5 EMC test plan . 9
5.1 General . 9
5.2 Configuration of EUT during testing . 9
5.3 Operation conditions of EUT during testing . 10
5.4 Specification of FUNCTIONAL PERFORMANCE . 10
5.5 Test description . 10
6 Immunity requirements . 10
6.1 Conditions during the tests. 10
6.2 Immunity test requirements . 11
6.3 Random aspects . 11
6.4 Performance criteria . 11
7 Emission requirements . 12
7.1 Conditions during measurements . 12
7.2 Emission limits . 12
8 Test results and test report . 12
9 Instructions for use . 12
Annex A (normative) Immunity test requirements for PORTABLE TEST AND MEASUREMENT
EQUIPMENT powered by battery or from the circuit being measured . 13
Annex AA (normative) Additional requirements and exceptions for specific types of
transducers – Transducers for measurement of tension and compressive forces (force
transducers) . 14
AA.1 General considerations . 14
AA.2 Test configuration . 14
AA.3 Operation conditions . 15
Annex BB (normative) Additional requirements and exceptions for specific types of
transducers – Transducers for measurement of pressure (pressure transducers) . 17
BB.1 General considerations . 17
BB.2 Test configuration . 17
BB.3 Operation conditions . 18
Annex CC (normative) Additional requirements and exceptions for specific types of
transducers – Transducers for measurement of temperature (temperature transducer) . 19
CC.1 General considerations . 19
CC.2 Test configuration . 19
CC.3 Operation conditions . 20
Bibliography . 22

Figure 101 – Example of a TRANSDUCER WITH INTEGRATED SIGNAL CONDITIONING . 7
Figure 102 – Example of a TRANSDUCER WITH REMOTE SIGNAL CONDITIONING . 8
Figure AA.1 – Example of the configuration of a force transducer with remote signal
conditioning . 15

IEC 61326-2-3:2020 © IEC 2020 – 3 –
Figure BB.1 – Example of the configuration of a pressure transducer . 18
Figure CC.1 – Example of the configuration of a temperature transducer with sensor
and signal conditioning in the same housing . 20
Figure CC.2 – Example of the configuration of a temperature transducer with remote
signal conditioning . 20

Table 101 – Performance criteria for the different functions . 12
Table AA.1 – Circuitry actions for generating an output signal for simulation of a
mechanical load on the transducer . 16

– 4 – IEC 61326-2-3:2020 © IEC 2020
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
ELECTRICAL EQUIPMENT FOR MEASUREMENT,
CONTROL AND LABORATORY USE –
EMC REQUIREMENTS –
Part 2-3: Particular requirements –
Test configuration, operational conditions and pe
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.