Standard Test Method for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer

SIGNIFICANCE AND USE
5.1 The complex shear modulus is an indicator of the stiffness or resistance of asphalt binder to deformation under load. The phase angle is a measure of the relative portion of the response to an applied load that is elastic (recoverable) or viscous (nonrecoverable).  
5.2 The test procedure is applicable to measurements in the linear region where the measured modulus and phase angle are independent of the amplitude of the strain.  
5.3 The complex modulus and the phase angle are used to calculate performance-related criteria in accordance with Specification D6373 or D8239.
Note 1: The quality of the results produced by this standard are dependent on the competence of the personnel performing the procedure and the capability, calibration, and maintenance of the equipment used. Agencies that meet the criteria of Specification D3666 are generally considered capable of competent and objective testing, sampling, inspection, etc. Users of this standard are cautioned that compliance with Specification D3666 alone does not completely ensure reliable results. Reliable results depend on many factors; following the suggestions of Specification D3666 or some similar acceptable guideline provides a means of evaluating and controlling some of those factors.
SCOPE
1.1 This test method covers the determination of the complex shear modulus and phase angle of asphalt binders when tested in dynamic (oscillatory) shear using parallel plate geometry.  
1.2 This test method is intended for determining the linear viscoelastic properties of asphalt binders as required for specification testing and is not intended as a comprehensive procedure for the full characterization of the viscoelastic properties of asphalt binder.  
1.3 This standard is appropriate for unaged asphalt binder, conditioned asphalt binder, and asphalt binder recovered from either asphalt mixtures or asphalt emulsions. To keep the language in this standard precise, the term “asphalt binder” is used to refer to the material being tested.  
1.4 This procedure is limited to asphalt binders that contain particles with largest dimension less than 250 μm.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 Warning—Mercury has been designated by the United States Environmental Protection Agency (EPA) and many state agencies as a hazardous material that can cause central nervous system, kidney, and liver damage. Mercury, or its vapor, may be hazardous to health and corrosive to materials. Caution should be taken when handling mercury and mercury-containing products. See the applicable product Safety Data Sheet (SDS) for details and EPA’s website— www.epa.gov/mercury/faq.htm—for additional information. Users should be aware that selling mercury, mercury-containing products, or both, into your state may be prohibited by state law.  
1.7 This standard may involve hazardous materials, operations, and equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
14-Dec-2023
Drafting Committee
D04.44 - Rheological Tests

Relations

Effective Date
15-Dec-2023
Effective Date
01-Feb-2024
Effective Date
01-Jan-2024
Effective Date
15-Jun-2015
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023
Effective Date
15-Dec-2023

Overview

ASTM D7175-23 is the Standard Test Method for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer (DSR). Developed by ASTM International, this test method is essential for measuring the complex shear modulus and phase angle of asphalt binders. These rheological properties are fundamental in assessing an asphalt binder's stiffness and its resistance to deformation, critical for ensuring road durability under varying traffic and temperature conditions.

This standard targets both unaged and conditioned asphalt binders, including those recovered from mixtures or emulsions, provided the material contains no particles larger than 250 μm. The DSR method outlined here forms a core part of performance-related specifications for paving materials and is widely used in both research and quality assurance testing in the asphalt industry.

Key Topics

  • Complex Shear Modulus (G)*
    Indicates the binder's stiffness or resistance to deformation under load. A higher modulus value means increased rigidity.

  • Phase Angle (δ)
    Measures the balance between elastic (recoverable) and viscous (nonrecoverable) responses of the asphalt binder. It helps identify how much deformation will be permanent.

  • Linear Viscoelastic Region
    The test is valid for measurements where modulus and phase angle are independent of strain amplitude, ensuring material behavior reflects actual binder performance.

  • Test Coverage
    Applies to unaged, aged (in accordance with D2872 and D6521), and recovered asphalts, using dynamic (oscillatory) shear under parallel plate geometry.

  • Equipment and Conditions
    Requires use of a dynamic shear rheometer, precise temperature control between 4°C and 88°C, plate diameters (8 mm and 25 mm), and standard calibration practices.

  • Repeatability and Calibration
    Emphasizes the importance of standardized apparatus, regular verification, and qualified personnel to ensure reliable results.

Applications

ASTM D7175-23 plays a pivotal role in the asphalt and paving industry by:

  • Quality Control
    Enabling manufacturers and agencies to verify asphalt binder performance before use in pavement construction.

  • Performance Grading
    Supplying key measurements (G* and δ) for performance-graded (PG) binder specifications, supporting standards such as ASTM D6373 and AASHTO M320.

  • Research and Development
    Being widely adopted in laboratory research for developing advanced asphalt formulations and evaluating aging or modification effects.

  • Forensic Analysis
    Assisting in diagnosing pavement failures by testing recovered asphalt binders from field samples.

  • Compliance and Certification
    Ensuring binders meet regulatory and client requirements through standardized, reproducible testing.

Related Standards

ASTM D7175-23 references and integrates with several other critical standards and practices:

  • ASTM D6373 - Specification for Performance-Graded Asphalt Binder
  • ASTM D3666 - Specification for Agencies Testing and Inspecting Road and Paving Materials
  • ASTM D6521 - Practice for Accelerated Aging of Asphalt Binder Using a Pressurized Aging Vessel (PAV)
  • AASHTO M320 & T315 - Performance Grading and DSR Testing Methods
  • ASTM D2872 - Rolling Thin-Film Oven Test (RTFOT) for binder aging
  • ISO/IEC 17025 - General requirements for the competence of testing and calibration laboratories

Practical Value

By standardizing the method for determining the rheological properties of asphalt binders, ASTM D7175-23 supports the delivery of reliable, high-performance paving materials. The measurements obtained influence pavement design, construction quality, and longer-lasting roads, enhancing safety and reducing maintenance costs for agencies and contractors.

Monitoring rheological properties using dynamic shear rheometry as prescribed by ASTM D7175 ensures consistent binder quality, meets specification criteria, and fosters innovation in road infrastructure materials.

Buy Documents

Standard

ASTM D7175-23 - Standard Test Method for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer

English language (13 pages)
sale 15% off
sale 15% off
Standard

REDLINE ASTM D7175-23 - Standard Test Method for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer

English language (13 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

Zavod za gradbeništvo Slovenije (ZAG) - Inšpekcija

ZAG inspection body for construction products, structures, and materials.

SA Slovenia Verified

Sponsored listings

Frequently Asked Questions

ASTM D7175-23 is a standard published by ASTM International. Its full title is "Standard Test Method for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer". This standard covers: SIGNIFICANCE AND USE 5.1 The complex shear modulus is an indicator of the stiffness or resistance of asphalt binder to deformation under load. The phase angle is a measure of the relative portion of the response to an applied load that is elastic (recoverable) or viscous (nonrecoverable). 5.2 The test procedure is applicable to measurements in the linear region where the measured modulus and phase angle are independent of the amplitude of the strain. 5.3 The complex modulus and the phase angle are used to calculate performance-related criteria in accordance with Specification D6373 or D8239. Note 1: The quality of the results produced by this standard are dependent on the competence of the personnel performing the procedure and the capability, calibration, and maintenance of the equipment used. Agencies that meet the criteria of Specification D3666 are generally considered capable of competent and objective testing, sampling, inspection, etc. Users of this standard are cautioned that compliance with Specification D3666 alone does not completely ensure reliable results. Reliable results depend on many factors; following the suggestions of Specification D3666 or some similar acceptable guideline provides a means of evaluating and controlling some of those factors. SCOPE 1.1 This test method covers the determination of the complex shear modulus and phase angle of asphalt binders when tested in dynamic (oscillatory) shear using parallel plate geometry. 1.2 This test method is intended for determining the linear viscoelastic properties of asphalt binders as required for specification testing and is not intended as a comprehensive procedure for the full characterization of the viscoelastic properties of asphalt binder. 1.3 This standard is appropriate for unaged asphalt binder, conditioned asphalt binder, and asphalt binder recovered from either asphalt mixtures or asphalt emulsions. To keep the language in this standard precise, the term “asphalt binder” is used to refer to the material being tested. 1.4 This procedure is limited to asphalt binders that contain particles with largest dimension less than 250 μm. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 Warning—Mercury has been designated by the United States Environmental Protection Agency (EPA) and many state agencies as a hazardous material that can cause central nervous system, kidney, and liver damage. Mercury, or its vapor, may be hazardous to health and corrosive to materials. Caution should be taken when handling mercury and mercury-containing products. See the applicable product Safety Data Sheet (SDS) for details and EPA’s website— www.epa.gov/mercury/faq.htm—for additional information. Users should be aware that selling mercury, mercury-containing products, or both, into your state may be prohibited by state law. 1.7 This standard may involve hazardous materials, operations, and equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

SIGNIFICANCE AND USE 5.1 The complex shear modulus is an indicator of the stiffness or resistance of asphalt binder to deformation under load. The phase angle is a measure of the relative portion of the response to an applied load that is elastic (recoverable) or viscous (nonrecoverable). 5.2 The test procedure is applicable to measurements in the linear region where the measured modulus and phase angle are independent of the amplitude of the strain. 5.3 The complex modulus and the phase angle are used to calculate performance-related criteria in accordance with Specification D6373 or D8239. Note 1: The quality of the results produced by this standard are dependent on the competence of the personnel performing the procedure and the capability, calibration, and maintenance of the equipment used. Agencies that meet the criteria of Specification D3666 are generally considered capable of competent and objective testing, sampling, inspection, etc. Users of this standard are cautioned that compliance with Specification D3666 alone does not completely ensure reliable results. Reliable results depend on many factors; following the suggestions of Specification D3666 or some similar acceptable guideline provides a means of evaluating and controlling some of those factors. SCOPE 1.1 This test method covers the determination of the complex shear modulus and phase angle of asphalt binders when tested in dynamic (oscillatory) shear using parallel plate geometry. 1.2 This test method is intended for determining the linear viscoelastic properties of asphalt binders as required for specification testing and is not intended as a comprehensive procedure for the full characterization of the viscoelastic properties of asphalt binder. 1.3 This standard is appropriate for unaged asphalt binder, conditioned asphalt binder, and asphalt binder recovered from either asphalt mixtures or asphalt emulsions. To keep the language in this standard precise, the term “asphalt binder” is used to refer to the material being tested. 1.4 This procedure is limited to asphalt binders that contain particles with largest dimension less than 250 μm. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 Warning—Mercury has been designated by the United States Environmental Protection Agency (EPA) and many state agencies as a hazardous material that can cause central nervous system, kidney, and liver damage. Mercury, or its vapor, may be hazardous to health and corrosive to materials. Caution should be taken when handling mercury and mercury-containing products. See the applicable product Safety Data Sheet (SDS) for details and EPA’s website— www.epa.gov/mercury/faq.htm—for additional information. Users should be aware that selling mercury, mercury-containing products, or both, into your state may be prohibited by state law. 1.7 This standard may involve hazardous materials, operations, and equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM D7175-23 is classified under the following ICS (International Classification for Standards) categories: 93.080.20 - Road construction materials. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM D7175-23 has the following relationships with other standards: It is inter standard links to ASTM D7175-15, ASTM C670-24a, ASTM C670-24, ASTM C670-15, ASTM D6373-23, ASTM D8189-19, ASTM D6626-23, ASTM D7552-22, ASTM D7944-22, ASTM D8239-23, ASTM D4887/D4887M-11(2021)e1, ASTM D7654/D7654M-10(2018), ASTM D7643-22, ASTM D7173-20, ASTM D7405-20. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM D7175-23 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D7175 − 23
Standard Test Method for
Determining the Rheological Properties of Asphalt Binder
Using a Dynamic Shear Rheometer
This standard is issued under the fixed designation D7175; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope use. It is the responsibility of the user of this standard to
establish appropriate safety, health, and environmental prac-
1.1 This test method covers the determination of the com-
tices and determine the applicability of regulatory limitations
plex shear modulus and phase angle of asphalt binders when
prior to use.
tested in dynamic (oscillatory) shear using parallel plate
1.8 This international standard was developed in accor-
geometry.
dance with internationally recognized principles on standard-
1.2 This test method is intended for determining the linear
ization established in the Decision on Principles for the
viscoelastic properties of asphalt binders as required for
Development of International Standards, Guides and Recom-
specification testing and is not intended as a comprehensive
mendations issued by the World Trade Organization Technical
procedure for the full characterization of the viscoelastic
Barriers to Trade (TBT) Committee.
properties of asphalt binder.
1.3 This standard is appropriate for unaged asphalt binder, 2. Referenced Documents
conditioned asphalt binder, and asphalt binder recovered from
2.1 ASTM Standards:
either asphalt mixtures or asphalt emulsions. To keep the
C670 Practice for Preparing Precision and Bias Statements
language in this standard precise, the term “asphalt binder” is
for Test Methods for Construction Materials
used to refer to the material being tested.
D8 Terminology Relating to Materials for Roads and Pave-
1.4 This procedure is limited to asphalt binders that contain
ments
particles with largest dimension less than 250 μm.
D3666 Specification for Minimum Requirements for Agen-
cies Testing and Inspecting Road and Paving Materials
1.5 The values stated in SI units are to be regarded as
D6373 Specification for Performance-Graded Asphalt
standard. No other units of measurement are included in this
Binder
standard.
D8239 Specification for Performance-Graded Asphalt
1.6 Warning—Mercury has been designated by the United
Binder Using the Multiple Stress Creep and Recovery
States Environmental Protection Agency (EPA) and many state
(MSCR) Test
agencies as a hazardous material that can cause central nervous
E1 Specification for ASTM Liquid-in-Glass Thermometers
system, kidney, and liver damage. Mercury, or its vapor, may
E77 Test Method for Inspection and Verification of Ther-
be hazardous to health and corrosive to materials. Caution
mometers
should be taken when handling mercury and mercury-
E644 Test Methods for Testing Industrial Resistance Ther-
containing products. See the applicable product Safety Data
mometers
Sheet (SDS) for details and EPA’s website— www.epa.gov/
E882 Guide for Accountability and Quality Control in the
mercury/faq.htm—for additional information. Users should be
Chemical Analysis Laboratory
aware that selling mercury, mercury-containing products, or
2.2 AASHTO Standards:
both, into your state may be prohibited by state law.
R 29 Practice for Grading or Verifying the Performance
1.7 This standard may involve hazardous materials,
Grade of an Asphalt Binder
operations, and equipment. This standard does not purport to
address all of the safety concerns, if any, associated with its
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
This test method is under the jurisdiction of ASTM Committee D04 on Road contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
and Paving Materials and is the direct responsibility of Subcommittee D04.44 on Standards volume information, refer to the standard’s Document Summary page on
Rheological Tests. the ASTM website.
Current edition approved Dec. 15, 2023. Published January 2024. Originally Available from American Association of State Highway and Transportation
approved in 2005. Last previous edition approved in 2015 as D7175 – 15. DOI: Officials (AASHTO), 444 N. Capitol St., NW, Suite 249, Washington, DC 20001,
10.1520/D7175-23. http://www.transportation.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7175 − 23
T 315 Standard Test Method for Determining the Rheologi- 3.2.11 thermal equilibrium, n—condition where the tem-
cal Properties of Asphalt Binder Using a Dynamic Shear perature of the test specimen mounted between the test plates
Rheometer is constant with time.
2.3 ISO Standard:
4. Summary of Test Method
ISO/IEC 17025 General Requirements for the Competence
4.1 Test specimens are formed between parallel metal plates
of Testing and Calibration Laboratories
and one of the parallel plates is oscillated with respect to the
other at pre-selected temperatures, frequencies, and rotational
3. Terminology
strain amplitudes. The required amplitudes are specified to
3.1 Definitions—Definitions for many terms common to
ensure that the measurements are within the region of linear
asphalt binder are found in Terminology D8.
behavior.
3.2 Definitions of Terms Specific to This Standard:
4.2 The test specimen is maintained at the test temperature
3.2.1 annealing, n—the process of removing the effects of
by positive heating and cooling of the upper and lower plates
steric hardening by heating the binder until it is sufficiently
or by enclosing the upper and lower plates in a thermally
fluid so that it can be easily poured.
controlled environment or test chamber.
3.2.2 complex shear modulus (G*), n—ratio calculated by
4.3 Specification testing is performed at a test frequency of
dividing the absolute value of the peak-to-peak shear stress, τ,
10 rad/s. The complex modulus (G*) and phase angle (δ) are
by the absolute value of the peak-to-peak shear strain, γ.
calculated automatically as part of the operation of the rheom-
3.2.3 linear viscoelastic, adj—within context of this test
eter using proprietary computer software supplied by the
method, refers to a region of behavior in which the complex
instrument manufacturer.
shear modulus is independent of the amplitude of the shear
stress or strain. 5. Significance and Use
3.2.4 oscillatory shear, n—refers to a type of loading in
5.1 The complex shear modulus is an indicator of the
which a shear stress or shear strain is applied to a test sample
stiffness or resistance of asphalt binder to deformation under
in an oscillatory manner such that the shear stress or strain
load. The phase angle is a measure of the relative portion of the
varies in amplitude about zero in a sinusoidal manner.
response to an applied load that is elastic (recoverable) or
viscous (nonrecoverable).
3.2.5 parallel plate geometry, n—refers to a testing geom-
etry in which the test specimen is sandwiched between two
5.2 The test procedure is applicable to measurements in the
rigid parallel plates.
linear region where the measured modulus and phase angle are
independent of the amplitude of the strain.
3.2.6 phase angle (δ), n—the angle in degrees between the
peak of a sinusoidally applied strain and the resultant peak of
5.3 The complex modulus and the phase angle are used to
respective sinusoidal stress.
calculate performance-related criteria in accordance with
Specification D6373 or D8239.
3.2.7 portable thermometer, n—refers to an electronic de-
vice that is separate from the dynamic shear rheometer and that
NOTE 1—The quality of the results produced by this standard are
consists of a detector (probe containing a thermocouple or
dependent on the competence of the personnel performing the procedure
resistive element), associated electronic circuitry, and readout and the capability, calibration, and maintenance of the equipment used.
Agencies that meet the criteria of Specification D3666 are generally
system.
considered capable of competent and objective testing, sampling,
3.2.8 reference thermometer, n—refers to a liquid-in-glass
inspection, etc. Users of this standard are cautioned that compliance with
or electronic thermometer that is used as a laboratory reference Specification D3666 alone does not completely ensure reliable results.
Reliable results depend on many factors; following the suggestions of
standard and is calibrated by an agency accredited to ISO/IEC
Specification D3666 or some similar acceptable guideline provides a
17025.
means of evaluating and controlling some of those factors.
3.2.9 steric hardening, n—a process where molecular asso-
ciations occur between asphalt binder molecules during storage 6. Apparatus
at ambient temperature.
6.1 Dynamic Shear Rheometer (DSR) Test System—A dy-
3.2.9.1 Discussion—Steric hardening decreases the phase
namic shear rheometer test system consisting of test plates, an
angle and increases the complex shear modulus. Steric hard-
environmental chamber, an internal thermometer, a loading
ening is time and temperature dependent and is asphalt binder
device, and a data acquisition system as described in 6.1.1 –
specific. It may significantly affect the test results after just a
6.1.5. The manufacturer of the device shall provide a certificate
few hours of storage at ambient temperature.
certifying that the frequency, deflection angle, and torque are
controlled and measured with an accuracy of 1 % or less in the
3.2.10 strain sweep—an isothermal test procedure in which
the strain is increased stepwise at a linear rate of increase. range of this measurement.
6.1.1 Test Plates—Metal plates cylindrical in shape, formed
from steel or aluminum, with smooth ground surfaces are
required. Specification testing to D6373 or D8239 requires two
Available from International Organization for Standardization (ISO), ISO
plates (upper and lower) 8.00 6 0.02 mm in diameter and two
Central Secretariat, Chemin de Blandonnet 8, CP 401, 1214 Vernier, Geneva,
Switzerland, https://www.iso.org. plates (upper and lower) 25.00 6 0.05 mm in diameter (Item A
D7175 − 23
in Fig. 1). In some cases, a laboratory may perform testing 6.1.4 Loading Device—The loading device shall be capable
using only one plate size. In this case, only that plate size is of applying a sinusoidal oscillatory load to the specimen at a
required. The lower test plate shall have a minimum thickness frequency of 10.0 6 0.1 rad/s. If frequencies other than 10
or raised portion of 1.5 mm to adequately trim the specimen rad/s are used, the frequency shall be accurate to 1 %.
(Item B in Fig. 1). The plates shall be formed as an integral part
6.1.5 Data Acquisition System—The data acquisition system
of the test fixtures that are used to mount the plates in the DSR
shall provide a record of temperature, frequency, deflection
as shown in Fig. 1.
angle, and torque. The manufacturer of the rheometer shall
provide a certificate certifying that the frequency, deflection
angle, and torque are controlled and measured with an accu-
racy of at least 1 %.
6.2 Specimen Mold (optional)—The overall dimensions of
the silicone rubber mold for forming asphalt binder test
specimens may vary, but the overall thickness shall be at least
5 mm thick.
NOTE 4—The following dimensions have been found suitable: For a
25 mm test plate with a 1 mm gap, a mold cavity with a concave bottom
with an approximate diameter of 18 mm and a depth of at least 2.0 mm;
FIG. 1 Plate Diagram
and for an 8 mm test plate with a 2 mm gap, a mold cavity with a concave
bottom with an approximate diameter of 8 mm and a depth of at least
NOTE 2—The upper and lower plates should be concentric with each
2.5 mm.
other. At the present there is no suitable procedure for the user to check the
6.3 Trimming Tool—A tool with a straightedge at least 4 mm
concentricity except to visually observe whether or not the upper and
lower plates are centered with respect to each other. The moveable plate wide suitable for trimming excess binder from the periphery of
should rotate without any observable horizontal or vertical wobble. This
the test specimen to produce a smooth face on the test
may be checked visually or with a dial gage held in contact with the edge
specimen that is parallel and coincident with the outer diameter
of the moveable plate while it is being rotated.
of the upper and lower plates.
6.1.2 Environmental Chamber—A chamber for controlling
6.4 Reference Thermometer—A standard reference ther-
the temperature of the test specimen. The medium for heating
mometer for standardizing the portable thermometer. The
and cooling the specimen in the environmental chamber shall
reference thermometer must be calibrated annually by an
not affect asphalt binder properties. The temperature in the
agency accredited to ISO/IEC 17025. Alternatively, if the
chamber may be controlled by the circulation of fluid—
portable thermometer described in 6.5 is calibrated by an
conditioned gas, nitrogen or water is acceptable—or by a
outside agency accredited to ISO/IEC 17025, the reference
suitable arrangement of actively temperature controlled heating
thermometer is not needed. The thermometer shall have a
elements (for example, solid-state Peltier elements) surround-
measurement temperature range covering the DSR’s range of
ing the sample. When laboratory air is used in a forced air
test temperatures and an accuracy of 60.05 °C and shall be one
oven, a suitable dryer must be included to prevent condensa-
of the following:
tion of moisture on the test plates. The environmental chamber
6.4.1 A partial immersion liquid-in-glass thermometer read-
and the temperature controller shall control the temperature of
able to the nearest 0.01 °C meeting the requirements of
the test specimen mounted between the test plates, including
Specification E1 readable to the nearest 0.05 °C. The thermom-
any thermal gradients within the test specimen, at the test
eter shall be calibrated annually in accordance with Test
temperature 60.1 °C. The chamber or the water in the chamber
Method E77.
shall completely enclose the top and the bottom plates to
minimize thermal gradients within the fixtures and test speci- 6.4.2 A platinum resistance thermometer (PRT) readable to
the nearest 0.01 °C, with a Pt 100 Class AA tolerance rating
men.
and either a three- or four-wire configuration and an overall
NOTE 3—A circulating bath unit separate from the DSR that pumps the
sheath length at least 50 mm greater than the immersion depth.
water through the test chamber may be required if a fluid medium is used.
The thermometer shall be calibrated annually in accordance
6.1.2.1 Temperature Controller—A temperature controller
with Test Method E644. Corrections shall be applied to ensure
capable of maintaining the temperature of the test specimen at
accurate measurements within 0.05 °C.
the test temperature for the entire range of test temperatures.
6.4.3 A thermistor readable to the nearest 0.01 °C, cali-
6.1.3 Internal Thermometer—A platinum resistance ther-
brated annually in accordance with Test Method E644. Correc-
mometer (PRT) readable to the nearest 0.1 °C. The PRT shall
tions shall be applied to ensure accurate measurements within
be mounted within the environmental chamber as an integral
0.05 °C.
part of the DSR and in close proximity to the fixed plate. The
internal thermometer shall be used to control the temperature 6.5 Portable Thermometer—A calibrated portable thermom-
of the test specimen between the plates and shall provide a eter consisting of a resistive detector, associated electronic
continuous readout of temperature during the mounting, circuitry, and digital readout, fitted with a specimen-shaped
conditioning, and testing of the specimen. Standardize the wafer embedded with temperature sensor. The thickness of the
internal thermometer and apply corrections in accordance with sensor shall be no greater than 2.0 mm such that it can be
8.4.1 of this standard. inserted between the test plates. The portable thermometer
D7175 − 23
shall be a PRT or thermistor constructed in the same fashion as between the test plates) at intervals of six months or less using
described in 6.4.2 or 6.4.3, standardized internally using the the reference thermometer. If the reference thermometer (see
reference thermometer as described in 6.4 or externally by an 6.4) is also used as a portable thermometer to measure the
outside agency accredited to ISO/IEC 17025. temperature between the test plates, it shall be standardized as
6.5.1 The reference thermometer (see 6.4) may be used as per 8.3. The portable thermometer shall be standardized using
the portable thermometer if its detector is fitted with a wafer the same meters and circuitry (wiring) that are used when
embedded with a temperature sensor. temperature measurements are made between the plates.
8.3.1 Recommended Standardization Procedure—Bring the
NOTE 5—Guide E882 may be used for evaluating test data.
reference thermometer into intimate contact with the probe of
6.6 Micrometer—A micrometer readable to 0.01 mm for
the portable thermometer and place them in a thermostatically
verifying plate diameter.
controlled and stirred liquid bath. Obtain measurements with
both thermometers at test temperatures that will be used when
7. Materials
conducting tests with the DSR. Allow the bath to come to
7.1 Wiping Material—Clean cloth, paper towels, cotton thermal equilibrium at each temperature. For this purpose,
thermal equilibrium is defined for each thermometer as the
swabs, or other suitable material as required for wiping the
point when three successive readings read at 1 min intervals do
plates.
not change by more than 0.02 °C. Record the temperature on
7.2 Cleaning Solvents—Solvents for cleaning the plates
each thermometer when thermal equilibrium is reached. The
shall leave no residue on the surface of the plates.
difference between the two readings is the temperature correc-
NOTE 6—Mineral oil, citrus-based solvents, mineral spirits, toluene, or
tion that shall be applied to the portable thermometer.
similar solvents have been found to be suitable for cleaning the plates.
8.4 Standardization of Internal Thermometer—The internal
Organic solvent that does not leave a residue such as heptane, acetone, or
ethyl alcohol can also be used for removing solvent residue from the thermometer shall be verified at an interval of no greater than
surfaces of the plates.
six months. When the differences between the DSR thermom-
eter and the portable thermometer are 0.02 °C or greater, apply
7.3 Anti-Seize Compound (optional)—Used to ensure full
a temperature correction to correct to the temperature displayed
contact and heat transfer between the faces of the test plates
by the portable thermometer.
and the faces of the portable thermometer probe.
7.4 Reference Fluid—An organic polymer reference fluid NOTE 8—Some DSRs are programmed to automatically determine the
temperature correction and software within the DSR automatically applies
with a known viscosity traceable to the international system of
the correction to the reported test temperature.
units through a national metrology institute (such as NIST) that
8.4.1 Internal Thermometer Standardization Procedure—
is approximately 270 Pa-s at 64 °C, as determined through
For the entire range of test temperatures used by the DSR and
capillary viscosity measurements. The known viscosity for the
at intervals of 6 °C, place the portable thermometer between
production run (lot) of the reference standard shall be printed
the 25 mm test plates and close the gap to bring the wafer into
on the label of the bottle.
contact with the upper and lower plate so that the wafer makes
NOTE 7—A suitable reference fluid is available from Cannon Instrument
complete contact with the surfaces of the upper and lower
Company as viscosity standard number N2700000SP or DSR64C. The
plates. If needed, apply a thin layer of petroleum grease or
reference fluid is reported in cP which is numerically equal to mPa-s.
anti-seize compound to completely fill any void space between
the wafer and the plates. Complete contact is needed to ensure
8. Standardizations, Checks, and Verifications
proper heat transfer across the plates and wafer.
8.1 Standardize, check, and verify the DSR and its compo-
8.4.2 At each temperature verified, ensure thermal equilib-
nents as described in this section when the DSR is newly
rium has been reached and record the temperature indicated by
installed, when it is moved to a new location, and/or whenever
the portable thermometer and the DSR thermometer to the
the accuracy of the DSR and/or any of its components is
nearest 0.01 °C. Thermal equilibrium is reached when the
suspect. Four items require standardization, check, or verifica-
temperature indicated by both the portable thermometer and
tion: test plate diameter, portable thermometer, internal
the DSR thermometer do not vary by more than 0.03 °C over
thermometer, and overall operation of the DSR. Standardize
a 3 min period.
the internal thermometer before verifying the overall operation
8.4.3 Application of Temperature Correction—If an adjust-
of the DSR.
ment to the DSR thermometer is required, it is typically
8.2 Check of Plate Diameter—Before first using an upper or
incorporated automatically within the software on with newer
lower test plate and every six months thereafter, measure its
rheometers using the temperature verification data. If the DSR
diameter (average of three different locations ~120° apart) to does not have this capability, perform a manual temperature
the nearest 0.01 mm using the micrometer described in 6.6.
correction as described in Appendix X1 and adjust the target
Verify that the plates are in compliance with the requirements test temperature in accordance with the correction such that the
specified in 6.1.1. Maintain a log of the measured diameters so
temperature reported by the DSR includes the correction.
that the measurements are clearly identified with specific
8.5 Verification of Overall Operation of DSR—Verify the
plates.
overall operation of the DSR using a reference fluid once every
8.3 Standardization of Portable Thermometer—Standardize six months or when the accuracy of the test measurements are
the portable thermometer (used to measure the temperature suspect.
D7175 − 23
8.5.1 The reference fluid shall be used only at 64 °C. Below acetone. If necessary, use a dry cotton swab or soft cloth to
that temperature the phase angle differs sufficiently from 90° ensure that no moisture condenses on the plates.
such that the values calculated from G* and ω are no longer
9.4 Mount Test Plates and Fixtures—If the fixtures or plates
accurate. Above 64 °C the fluid tends to flow from the gap
are removed for cleaning, attach them as directed by the DSR
between the plates invalidating the measured value for G*.
manufacturer.
NOTE 9—A newly installed or reconditioned instrument should be
9.5 Zero the Test Gap—Zero the test gap in accordance with
verified on a weekly basis using the procedures in 8.5 until acceptable
one of the three options described in 9.5.1 – 9.5.3. If the test
verification has been demonstrated.
temperature differs by more than 612 °C from the temperature
8.5.2 Measurement of the Viscosity of the Reference Fluid—
at which the gap is set, re-zero the gap.
Measure the complex modulus, G*, of the reference fluid at
NOTE 11—The frame, detectors, and fixtures in the DSR change
64 °C and 10 rad/s. Additional information regarding use of the
dimension with temperature, causing the zero gap to change with changes
reference fluid is given in Annex A1. Precaution should be
in temperature. Adjustments in the gap are not necessary when measure-
taken to ensure that there are no air bubbles in the test
ments are made over a limited range of temperatures. The gap should be
set at the test temperature or, when tests are to be conducted over a range
specimen. Newer rheometers are typically programmed to
of temperatures, at the middle of the expected range of test temperatures.
automatically calculate and report the measured viscosity, η ,
M
If the instrument has thermal gap compensation, the gap may be set at the
in units of mP-s. When making the calculation directly, convert
first test temperature instead of the middle of the range of test tempera-
the numerical value of G* in kPa at 10 rad/s to viscosity in
tures.
mPa-s by multiplying the numerical value of G* by 10 . For
9.5.1 Manual Gap Setting—Manually spin the moveable
example, if G* = 2.66 kPa, η = 266 000 mPa-s. If the units for
M
plate, and while the moveable plate is spinning, slowly close
G* are in Pa, multiply by 10 .
the gap. The zero gap is reached when the plate just stops
spinning completely.
NOTE 10—The reference fluid may be reported in cP or mPa-s, which
are numerically equivalent; for example, 1 mPa = 1cP. The simplified
9.5.2 Normal Force Transducer—For rheometers with nor-
conversion given above is provided to make the calculation of η
M
mal force transducers, set the zero gap by closing the gap and
5 2
straightforward for the typical user. The multipliers 10 and 10 are
observing the normal force. Select the zero gap as the position
necessary for the proper conversion of units.
where the normal force is approximately zero.
8.5.3 The viscosity, η , measured with the DSR shall be
M
9.5.3 Automatic Gap Setting—Zero the gap automatically
within 3 % of the viscosity, η , reported by the manufacturer of
R
according to operating procedures specified by the instrument
the reference fluid. Otherwise, the overall operation of the DSR
manufacturer.
shall be considered suspect. Calculate the estimated percent
9.6 Preheating Test Plates—Once the zero gap is estab-
difference as:
lished as per 9.5.1, move the plates apart to approximately set
~η 2 η !
R M
the test gap and preheat the plates. Preheating the plates
percent difference 5 100 % × (1)
F G
η
R
promotes adhesion between the asphalt binder and the plates,
especially at the intermediate grading temperatures.
where:
9.6.1 Preheating 25 mm Plate—Bring the test plates to the
η = the viscosity as reported by the s
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D7175 − 15 D7175 − 23
Standard Test Method for
Determining the Rheological Properties of Asphalt Binder
Using a Dynamic Shear Rheometer
This standard is issued under the fixed designation D7175; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers the determination of the complex shear modulus and phase angle of asphalt binders when tested in
dynamic (oscillatory) shear using parallel plate geometry.
1.2 This test method covers the determination of the dynamic shear modulus and phase angle of asphalt binders when tested in
dynamic (oscillatory) shear using parallel plate geometry. It is applicable to asphalt binders having dynamic shear modulus values
in the range from 100 Pa to 10 MPa. This range in modulus is typically obtained between 4 and 88°C at 10 rad/s. This test method
is is intended for determining the linear viscoelastic properties of asphalt binders as required for specification testing and is not
intended as a comprehensive procedure for the full characterization of the viscoelastic properties of asphalt binders.binder.
1.3 This standard is appropriate for unaged asphalt binder, conditioned asphalt binder, and asphalt binder recovered from either
asphalt mixtures or asphalt emulsions. To keep the language in this standard precise, the term “asphalt binder” is used to refer to
the material being tested.
1.4 This standard is appropriate for unaged materials, material aged in accordance with Test Method D2872, material aged in
accordance with Practice D6521, or material aged in accordance with both Test Method D2872 and Practice D6521. This procedure
is limited to asphalt binders that contain particles with largest dimension less than 250 μm.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 Warning—Mercury has been designated by the United States Environmental Protection Agency (EPA) and many state
agencies as a hazardous material that can cause central nervous system, kidney, and liver damage. Mercury, or its vapor, may be
hazardous to health and corrosive to materials. Caution should be taken when handling mercury and mercury-containing products.
See the applicable product Safety Data Sheet (SDS) for details and EPA’s website— www.epa.gov/mercury/faq.htm—for
additional information. Users should be aware that selling mercury, mercury-containing products, or both, into your state may be
prohibited by state law.
1.7 This standard may involve hazardous materials, operations, and equipment. This standard does not purport to address all
of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate
safety and healthsafety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
This test method is under the jurisdiction of ASTM Committee D04 on Road and Paving Materials and is the direct responsibility of Subcommittee D04.44 on Rheological
Tests.
Current edition approved July 1, 2015Dec. 15, 2023. Published August 2015January 2024. Originally approved in 2005. Last previous edition approved in 20082015 as
D7175 – 08.D7175 – 15. DOI: 10.1520/D7175-15.10.1520/D7175-23.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7175 − 23
1.8 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
C670 Practice for Preparing Precision and Bias Statements for Test Methods for Construction Materials
D140D8 Practice for Sampling Asphalt MaterialsTerminology Relating to Materials for Roads and Pavements
D2170 Test Method for Kinematic Viscosity of Asphalts
D2171 Test Method for Viscosity of Asphalts by Vacuum Capillary Viscometer
D2872D3666 Test Method for Effect of Heat and Air on a Moving Film of Asphalt Binder (Rolling Thin-Film Oven
Test)Specification for Minimum Requirements for Agencies Testing and Inspecting Road and Paving Materials
D6373 Specification for Performance-Graded Asphalt Binder
D6521D8239 Practice for Accelerated Aging of Specification for Performance-Graded Asphalt Binder Using a Pressurized
Aging Vessel (PAV)the Multiple Stress Creep and Recovery (MSCR) Test
E1 Specification for ASTM Liquid-in-Glass Thermometers
E77 Test Method for Inspection and Verification of Thermometers
E563 Practice for Preparation and Use of an Ice-Point Bath as a Reference Temperature
E644 Test Methods for Testing Industrial Resistance Thermometers
E882 Guide for Accountability and Quality Control in the Chemical Analysis Laboratory
2.2 AASHTO Standards:
M320 Standard Specification for Performance-Graded Asphalt Binder
R29R 29 Practice for Grading or Verifying the Performance Grade of an Asphalt Binder
T315T 315 Standard Test Method for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear
Rheometer
2.3 Deutsche Industrie Norm (DIN) ISO Standard:
43760ISO/IEC 17025 Standard for Calibration of ThermocouplesGeneral Requirements for the Competence of Testing and
Calibration Laboratories
3. Terminology
3.1 Definitions—Definitions for many terms common to asphalt binder are found in Terminology D8.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 annealing, n—the process of removing the effects of steric hardening by heating the binder until it is sufficiently fluid so that
it can be easily poured.
3.1.2 asphalt binder, n—an asphalt-based cement that is produced from petroleum residue either with or without the addition of
non-particulate modifiers.
3.2.2 complex shear modulus (G*), n—ratio calculated by dividing the absolute value of the peak-to-peak shear stress, τ, by the
absolute value of the peak-to-peak shear strain, γ.
3.1.4 dummy test specimen, n—a specimen formed between the DSR test plates from asphalt binder or other polymer for the
purpose of determining the temperature in the asphalt binder between the plates.
3.1.4.1 Discussion—
The dummy test specimen is not used to measure the rheological properties of asphalt binder but is used solely to determine
temperature corrections.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from American Association of State Highway and Transportation Officials (AASHTO), 444 N. Capitol St., NW, Suite 249, Washington, DC 20001,
http://www.transportation.org.
Available from Beuth Verlag GmbH (DIN-- DIN Deutsches Institut fur Normung e.V.), Burggrafenstrasse 6, 10787, Berlin, Germany, http://www.en.din.de. International
Organization for Standardization (ISO), ISO Central Secretariat, Chemin de Blandonnet 8, CP 401, 1214 Vernier, Geneva, Switzerland, https://www.iso.org.
D7175 − 23
3.2.3 linear viscoelastic, adj—within context of this test method, refers to a region of behavior in which the dynamiccomplex shear
modulus is independent of the amplitude of the shear stress or strain.
3.1.6 steric hardening, n—refers to time-dependent associations that occur between asphalt binder molecules during storage at
ambient temperature. The effect of molecular association or steric hardening on the dynamic shear modulus is asphalt specific and
may be apparent even after a few hours of storage.
3.2.4 oscillatory shear, n—refers to a type of loading in which a shear stress or shear strain is applied to a test sample in an
oscillatory manner such that the shear stress or strain varies in amplitude about zero in a sinusoidal manner.
3.2.5 parallel plate geometry, n—refers to a testing geometry in which the test specimen is sandwiched between two rigid parallel
plates and subjected to shear.plates.
3.2.6 phase angle (δ), n—the angle in degrees between the peak of a sinusoidally applied strain and the resultant sinusoidal stress
in a controlled-strain testing mode, or between the applied stress and the resultant strain in a controlled-stress testing mode.peak
of respective sinusoidal stress.
3.2.7 portable temperature measuring device, thermometer, n—refers to an electronic device that is separate from the dynamic
shear rheometer and that consists of a detector (probe containing a thermocouple or resistive element), associated electronic
circuitry, and readout system.
3.2.8 reference temperature measuring device, thermometer, n—refers to a NIST-traceable liquid-in-glass or electronic thermom-
eter that is used as a laboratory standard.reference standard and is calibrated by an agency accredited to ISO/IEC 17025.
3.2.9 steric hardening, n—a process where molecular associations occur between asphalt binder molecules during storage at
ambient temperature.
3.2.9.1 Discussion—
Steric hardening decreases the phase angle and increases the complex shear modulus. Steric hardening is time and temperature
dependent and is asphalt binder specific. It may significantly affect the test results after just a few hours of storage at ambient
temperature.
3.2.10 strain sweep—an isothermal test procedure in which the strain is increased stepwise at a linear rate of increase.
3.2.11 thermal equilibrium, n—condition where the temperature of the test specimen mounted between the test plates is constant
with time.
4. Summary of Test Method
4.1 This standard contains the procedure used to measure the complex shear modulus (G*) and phase angle (δ) of asphalt binders
using dynamic shear rheometer and parallel plate geometry.
4.2 The standard is suitable for use when the dynamic shear modulus varies between 100 Pa and 10 MPa. This range in modulus
is typically obtained between 4 and 88°C, depending upon the grade, test temperature, and conditioning (aging) of the asphalt
binder.
4.3 Test specimens, nominally 25 mm in diameter by 1 mm thick or 8 mm in diameter by 2 mm thick, are formed between parallel
metal plates.
4.1 During testing, Test specimens are formed between parallel metal plates and one of the parallel plates is oscillated with respect
to the other at pre-selected frequencies and angular deflection (or torque) temperatures, frequencies, and rotational strain
amplitudes. The required amplitude depends upon the value of the complex shear modulus of the asphalt binder being tested. The
required amplitudes have been selected so that, for most asphalt binders, the testing specified in this standard is amplitudes are
specified to ensure that the measurements are within the region of linear behavior.
D7175 − 23
4.2 The test specimen is maintained at the test temperature 6 0.1°C by positive heating and cooling of the upper and lower plates
or by enclosing the upper and lower plates in a thermally controlled environment or test chamber.
4.3 Oscillatory loading frequencies using this standard can range from 1 to 160 rad/s. Specification testing is performed at a test
frequency of 10 rad/s. The complex modulus (G*) and phase angle (δ) are calculated automatically as part of the operation of the
rheometer using proprietary computer software supplied by the instrument manufacturer.
5. Significance and Use
5.1 The test temperature for this test is related to the temperature experienced by the pavement in the geographical area for which
the asphalt binder is intended to be used.
5.1 The complex shear modulus is an indicator of the stiffness or resistance of asphalt binder to deformation under load. The
complex shear modulus and the phase angle define the resistance to shear deformation of the asphalt binder in the linear
viscoelastic region. The complex modulus and the phase angle are used to calculate performance-related criteria in accordance with
Specificationphase angle is a measure of the relative portion of the response to an applied load that is elastic (recoverable) D6373or
AASHTO Standard M320.viscous (nonrecoverable).
5.2 The test procedure is applicable to measurements in the linear region where the measured modulus and phase angle are
independent of the amplitude of the strain.
5.3 The complex modulus and the phase angle are used to calculate performance-related criteria in accordance with Specification
D6373 or D8239.
NOTE 1—The quality of the results produced by this standard are dependent on the competence of the personnel performing the procedure and the
capability, calibration, and maintenance of the equipment used. Agencies that meet the criteria of Specification D3666 are generally considered capable
of competent and objective testing, sampling, inspection, etc. Users of this standard are cautioned that compliance with Specification D3666 alone does
not completely ensure reliable results. Reliable results depend on many factors; following the suggestions of Specification D3666 or some similar
acceptable guideline provides a means of evaluating and controlling some of those factors.
6. Interferences
6.1 Particulate material in the asphalt binder is limited to particles with longest dimensions less than 250 μm. Particles with
dimensions greater than 250 μm approach the dimensions of the gap (1000 μm). In order to accurately characterize a two-phase
material containing particulate material it is well accepted that the thickness of the test specimen must be at least four times the
maximum particle size.
6.1.1 The calculation of the complex modulus from the data obtained from the DSR is highly dependent upon an accurate
measurement of the diameter of the test specimen. In the procedure, the diameter of the test specimen is assumed equal to the
diameter of the test plates. This assumption is valid only if the test sample is properly trimmed.
6.1.2 The physical properties of asphalt binders are very sensitive to test temperature and thermal history. Thermal history is the
number of times asphalt binder sample has been heated prior to testing. Controlling the test temperature to 6 0.1°C and limiting
the number of times the asphalt sample is heated prior to testing (only one heating is recommended) is essential in order to obtain
repeatable test results within a laboratory as well as to reproduce results between laboratories.
6. Apparatus
6.1 Dynamic Shear Rheometer (DSR) Test System—A dynamic shear rheometer test system consisting of parallel metal plates, a
means for controlling the temperature of the test specimen, test plates, an environmental chamber, an internal thermometer, a
loading device, and a control and data acquisition system. system as described in 6.1.1 – 6.1.5. The manufacturer of the device
shall provide a certificate certifying that the frequency, deflection angle, and torque are controlled, measured, or both, with
controlled and measured with an accuracy of 1 % or less in the range of this measurement.
6.1.1 Test Plates—Metal plates cylindrical in shape, formed from steel or aluminum, with smooth ground surfaces are required.
Two Specification testing to D6373 or D8239plates requires two plates (upper and lower) 8.00 6 0.02 mm in diameter and two
D7175 − 23
plates (upper and lower) 25.00 6 0.05 mm in diameter are required. The test plates(Item A in Fig. 1). In some cases, a laboratory
may perform testing using only one plate size. In this case, only that plate size is required. The lower test plate shall have a
minimum thickness or raised portion of 1.5 mm to allow sufficient clearance for trimming the specimen. adequately trim the
specimen (Item B in Fig. 1). The plates shall be formed as an integral part of the test fixtures that are used to mount the plates
in the DSR as shown in Fig. 1.
FIG. 1 Plate DimensionsDiagram
NOTE 2—The upper and lower plates should be concentric with each other. At the present there is no suitable procedure for the user to check the
concentricity except to visually observe whether or not the upper and lower plates are centered with respect to each other. The moveable plate should
rotate without any observable horizontal or vertical wobble. This may be checked visually or with a dial gage held in contact with the edge of the
moveable plate while it is being rotated.
6.1.2 Environmental Chamber—A chamber for controlling the temperature of the test specimen. The medium for heating and
cooling the specimen in the environmental chamber shall not affect asphalt binder properties. The temperature in the chamber may
be controlled by the circulation of fluid—conditioned gas, nitrogen or water is acceptable—or by a suitable arrangement of actively
temperature controlled heating elements (for example, solid state solid-state Peltier elements) surrounding the sample. When
laboratory air is used in a forced air oven, a suitable dryer must be included to prevent condensation of moisture on the test plates.
The environmental chamber and the temperature controller shall control the temperature of the test specimen mounted between
the test plates, including any thermal gradients within the test specimen, at the test temperature 60.1°C.60.1 °C. The chamber or
the water in the chamber shall completely enclose the top and the bottom plates to minimize thermal gradients within the fixtures
and test specimen.
NOTE 3—A circulating bath unit,unit separate from the DSR that pumps the water through the test chamber may be required if a fluid medium is used.
6.1.2.1 Temperature Controller—A temperature controller capable of maintaining the temperature of the test specimen at the test
temperature for the entire range of test temperatures.
6.1.3 Internal DSR Temperature Measurement Device—Thermometer—A platinum resistance temperature measurement device
(PRT) thermometer (PRT) readable to the nearest 0.1 °C. The PRT shall be mounted within the environmental chamber as an
integral part of the DSR and in close proximity to the fixed plate, with a range of 4 to 88°C, and with a resolution of 0.1°C. This
temperature measurement device plate. The internal thermometer shall be used to control the temperature of the test specimen
between the plates and shall provide a continuous readout of temperature during the mounting, conditioning, and testing of the
specimen. Standardize the internal thermometer and apply corrections in accordance with 8.4.1 of this standard.
6.1.4 Loading Device—The loading device shall be capable of applying a sinusoidal oscillatory load to the specimen at a
frequency of 10.0 6 0.1 rad/s. If frequencies other than 10 rad/s are used, the frequency shall be accurate to 1 %. The loading
device shall be capable of providing either a stress controlled or strain controlled load within a range of stress or strain necessary
to make the measurements described in this test method.
6.1.5 Data Acquisition System—The data acquisition system shall provide a record of temperature, frequency, deflection angle, and
torque. The manufacturer of the rheometer shall provide a certificate certifying that the frequency, deflection angle, and torque are
reported controlled and measured with an accuracy of at least 1 %.
6.2 Specimen Mold (optional)—The overall dimensions of the silicone rubber mold for forming asphalt binder test specimens may
vary, but the overall thickness shall be at least 5 mm thick.
D7175 − 23
NOTE 4—The following dimensions have been found suitable: For a 25-mm25 mm test plate with a 1-mm gap 1 mm gap, a mold cavity with a concave
bottom with an approximate diameter of 18 mm and a depth of at least 2.0 mmmm; and for an 8-mm 8 mm test plate with a 2-mm gap 2 mm gap, a
mold cavity with a concave bottom with an approximate diameter of 8 mm and a depth of at least 2.5 mm.2.5 mm.
6.3 Trimming Tool—A tool with a straightedge at least 4 mm wide suitable for trimming excess binder from the periphery of the
test specimen to produce a smooth face on the test specimen that is parallel and coincident with the outer diameter of the upper
and lower plates.
6.4 Reference Temperature Measurement Device—Thermometer—Either a NIST-traceable liquid-in-glass thermometer(s) (see A
standard reference thermometer for standardizing the portable thermometer. The reference thermometer must be calibrated
annually by an agency accredited to ISO/IEC 17025. Alternatively, if the portable thermometer described in 7.4.16.5) or
NIST-traceable digital electronic thermometer (see is calibrated by an outside agency accredited to ISO/IEC 17025, the reference
thermometer is 7.4.2) shall be maintained in the laboratory as a temperature standard. not needed. The thermometer shall have a
measurement temperature range covering the DSR’s range of test temperatures and an accuracy of 60.05 °C and shall be one of
the following:
6.4.1 Liquid-in-Glass Thermometer—NIST-traceable liquid-in-glass thermometer(s) with a range between 0 to 88°C and with
subdivisions of 0.1°C. The thermometer(s) shall be partial immersion thermometers with an ice point. The liquid-in-glass
thermometers A partial immersion liquid-in-glass thermometer readable to the nearest 0.01 °C meeting the requirements of
Specification E1shall be verified at least once a year readable to the nearest 0.05 °C. The thermometer shall be calibrated annually
in accordance with test method Test Method E77 and Practice E563.
NOTE 4—An Optical Viewing Device is recommended as an optional viewing device for use with liquid-in-glass thermometers because it enhances
readability and minimizes parallax when reading the liquid-in-glass reference thermometer.
6.4.2 Digital Temperature Measurement Device—An electronic thermometer that incorporates a thermometric device or resistive
detector with an accuracy of 60.05°C and a resolution of 0.01°C. The electronic A platinum resistance thermometer (PRT)
readable to the nearest 0.01 °C, with a Pt 100 Class AA tolerance rating and either a three- or four-wire configuration and an overall
sheath length at least 50 mm greater than the immersion depth. The thermometer shall be calibrated at least once per year
byannually in accordance with Test Method E644a commercial calibrating service using a NIST-traceable reference standard in
accordance with Test Methods . Corrections shall be applied to ensure accurate measurements within 0.05 °C.E644.
6.4.3 A thermistor readable to the nearest 0.01 °C, calibrated annually in accordance with Test Method E644. Corrections shall
be applied to ensure accurate measurements within 0.05 °C.
6.5 Portable Temperature Measurement Device—Thermometer—A calibrated portable thermometer consisting of a thermometric
device or resistive detector, associated electronic circuitry, and digital readout. readout, fitted with a specimen-shaped wafer
embedded with temperature sensor. The thickness of the detectorsensor shall be no greater than 2.0 mm such that it can be inserted
between the test plates. The portable thermometer shall be a PRT or thermistor constructed in the same fashion as described in 6.4.2
or 6.4.3, standardized internally using the reference thermometer as described in 6.4 or externally by an outside agency accredited
to ISO/IEC 17025.
6.5.1 The reference temperature measurement device thermometer (see 7.46.4) may be used for this purpose as the portable
thermometer if its detector fits within the dummy specimen as required byis fitted with a wafer embedded with a 9.4.2.temperature
sensor.
NOTE 5—Guide E882 may be used for evaluating test data.
6.6 Micrometer—A micrometer readable to 0.01 mm for verifying plate diameter.
7. Materials
7.1 Wiping Material—Clean cloth, paper towels, cotton swabs, or other suitable material as required for wiping the plates.
7.2 Cleaning Solvents: Solvents—
Solvents for cleaning the plates shall leave no residue on the surface of the plates.
D7175 − 23
NOTE 6—Mineral oil, citrus-based solvents, mineral spirits, toluene, or similar solvents have been found to be suitable for cleaning the plates. Organic
solvent that does not leave a residue such as heptane, acetone, or ethyl alcohol can also be used for removing solvent residue from the surfaces of the
plates.
7.3 Anti-Seize Compound (optional)—Used to ensure full contact and heat transfer between the faces of the test plates and the
faces of the portable thermometer probe.
7.4 Reference Fluid—An organic polymer reference fluid with a known viscosity traceable to the international system of units
through a national metrology institute (such as NIST) that is approximately 270 Pa-s at 64 °C, as determined through capillary
viscosity measurements. The known viscosity for the production run (lot) of the reference standard shall be printed on the label
of the bottle.
NOTE 7—A suitable reference fluid is available from Cannon Instrument Company as viscosity standard number N2700000SP or DSR64C. The reference
fluid is reported in cP which is numerically equal to mPa-s.
8. Verification Standardizations, Checks, and Verifications
8.1 Verify Standardize, check, and verify the DSR and its components as described in this section when the DSR is newly
installed, when it is moved to a new location, and/or whenever the accuracy of the DSR and/or any of its components is suspect.
Four items require standardization, check, or verification: test plate diameter, DSR torque transducer, portable thermometer, and
DSR test specimen temperature. Verify the DSR temperatureportable thermometer, internal thermometer, and overall operation of
the DSR. Standardize the internal thermometer before verifying the torque transducer.overall operation of the DSR.
8.2 VerificationCheck of Plate Diameter—Before first using an upper or lower test plate, plate and every six months thereafter,
measure its diameter (average of 3three different locations ~120° apart) to the nearest 0.02 mm. 0.01 mm using the micrometer
described in 6.6. Verify that the plates are in compliance with the requirements specified in 7.1.16.1.1. Maintain a log of the
measured diameters so that the measurements are clearly identified with specific plates.
8.3 VerificationStandardization of Portable Thermometer—VerifyStandardize the portable thermometer (used to measure the
temperature between the test plates) at intervals of six months or less using the laboratory reference thermometer. If the reference
thermometer (see 7.46.4) is also used as a portable thermometer to measure the temperature between the test plates, it shall be
verifiedstandardized as per 9.3.18.3. Electronic thermometers The portable thermometer shall be verifiedstandardized using the
same meters and circuitry (wiring) that are used when temperature measurements are made between the plates.
8.3.1 Recommended VerificationStandardization Procedure—Bring the reference thermometer into intimate contact with the
detector fromprobe of the portable thermometer and place them in a thermostatically controlled and stirred water bath (see liquid
Note 7). Ensure that de-ionized water is used to prevent electrical conduction from occurring between electrodes of the resistive
temperature sensitive element. If this is not available, encase the reference thermometer and the detector of the portable
thermometer into a water proof plastic bag prior to placement into the bath. Obtain measurements at intervals of 6°C over the range
of test temperatures allowing the with both thermometers at test temperatures that will be used when conducting tests with the
DSR. Allow the bath to come to thermal equilibrium at each temperature. If the readings of the portable thermometer and the
reference thermometer differ by 0.1°C or more, record the difference at each temperature as a temperature correction and maintain
the corrections in a log.For this purpose, thermal equilibrium is defined for each thermometer as the point when three successive
readings read at 1 min intervals do not change by more than 0.02 °C. Record the temperature on each thermometer when thermal
equilibrium is reached. The difference between the two readings is the temperature correction that shall be applied to the portable
thermometer.
NOTE 7—A recommended procedure is to use a stirred water bath that is controlled to 6 0.1°C such as the viscosity bath used for Test Method D2170
or Test Method D2171. Bring the probe from the portable thermometer into contact with the reference thermometer and hold the assembly in intimate
contact. A rubber band works well for this purpose. Immerse the assembly in the water bath and bring the water bath to thermal equilibrium. Record the
temperature on each device when thermal equilibrium is reached.
8.4 Temperature Offset—Standardization of Internal Thermometer—Thermal gradients within the rheometer can cause differences
between the temperature of the test specimen and the temperature indicated by The internal thermometer shall be verified at an
interval of no greater than six months. When the differences between the DSR thermometer (also used to control the temperature
of the DSR). When these differences are 0.1°C or greater, determineand the portable thermometer are 0.02 °C or greater, apply a
temperature correction by using a thermal detector mounted in a silicone rubber wafer (see to correct to 9.4.1) or by placing asphalt
D7175 − 23
binder (dummy sample) between the plates and inserting the detector of the portable thermometer into the asphalt binder (see the
9.4.2). The temperature offset shall be verified each time the portable thermometer is verified.temperature displayed by the portable
thermometer.
NOTE 8—Some DSRs are programmed to automatically determine the temperature correction and software within the DSR automatically applies the
correction to the reported test temperature.
8.4.1 Method Using Silicone Rubber Wafer—Internal Thermometer Standardization Procedure—Place the wafer For the entire
range of test temperatures used by the DSR and at intervals of 6 °C, place the portable thermometer between the 25 mm test plates
and close the gap to bring the wafer into contact with the upper and lower plate so that the silicone rubber wafer makes complete
contact with the surfaces of the upper and lower plates. If needed, apply a thin layer of petroleum grease or anti-seize compound
(see Note 8) to completely fill any void space between the silicone rubber wafer and the plates. Complete contact is needed to
ensure proper heat transfer across the plates and silicone rubber wafer (see wafer.Note 9) Determine any needed temperature
correction as per 9.4.3.
NOTE 8—Anti-seize compound available by that name at hardware and the auto supply stores is much less apt to contaminate the circulating water than
petroleum jelly.
NOTE 9—The currently available silicone wafer is 2 mm thick and slightly greater than 25 mm in diameter.
9.4.2 Method Using Dummy Test Specimen—The dummy test specimen shall be formed from asphalt binder, or other polymer that
can be readily formed between the plates. Mount the dummy test specimen between the test plates and insert the detector (probe)
of the portable thermometer into the dummy test specimen. Close the gap to the test gap (1 mm for 25-mm plates and 2 mm for
8-mm plates) keeping the detector centered vertically and radially in the dummy test specimen. Heat the plates as needed to allow
the dummy test specimen to completely fill the gap between the test plates. It is not necessary to trim the dummy test specimen
but avoid excessive material around the edges of the plates. Develop control charts using Guide E882. Determine any needed
temperature correction as per 9.4.3.
NOTE 10—Silly putty can leave a residue of silicone oil on the surfaces of the plates and for this reason its use as a dummy specimen is not recommended.
8.4.2 Determination of Temperature Offset—Obtain simultaneous temperature measurements with the DSR thermometer and the
portable thermometer at 6°C increments to cover the range of test temperatures. At each temperature increment, afterverified,
ensure thermal equilibrium has been reached, reached and record the temperature indicated by the portable thermometer and the
DSR thermometer to the nearest 0.1°C. Temperature0.01 °C. Thermal equilibrium is reached when the temperature indicated by
both the DSRportable thermometer and the portableDSR thermometer do not vary by more than 0.1°C0.03 °C over a five minute
time period. Obtain additional measurements to include the entire temperature range that will be used for measuring the dynamic
shear modulus.3 min period.
8.4.3 Plot Offset versus Specimen Temperature (optional)—Application of Temperature Correction—Using the data obtained in If
an adjustment 9.4.3, prepare a plot of the difference between the two temperature measurements versus the temperature measured
with the portable thermometer, Fig. 2. This difference is the temperature correction that must be applied to the DSR temperature
controller to obtain the desired temperature in the test specimen between the test plates. Report the temperature correction at the
respective testthermometer is required, it is typically incorporated automatically within the software on with newer rheometers
using the temperature verification data. If the DSR does not have this capability, perform a manual temperature correction as
described in Appendix X1 temperature from the plot and report the corrected test temperature between the plates as the test
temperature. Alternatively, the instrument software may be written to incorporate these temperature corrections.and adjust the
target test temperature in accordance with the correction such that the temperature reported by the DSR includes the correction.
NOTE 11—The difference between the two temperature measurements may not be a constant for a given rheometer but may vary with differences between
the test temperature and the ambient laboratory temperature as well as with fluctuations in ambient temperature. The difference between the two
temperature measurements is caused in part by thermal gradients in the test specimen and fixtures.
8.5 Verification of Overall Operation of DSR—Verify the accuracy of the torque transducer and angular displacement transducer
whenever the DSR is newly installed, when it is moved, every six months, each time temperature offset is verified, and/or whenever
overall operation of the DSR using a reference fluid once every six months or when the accuracy of measurements with the DSR
is the test measurements are suspect.
D7175 − 23
NOTE 12—A newly installed or reconditioned instrument should be verified on a weekly basis using the procedures in 9.5 until acceptable verification
has been demonstrated. Maintaining the data in the form of a control chart where the verification measurements are plotted versus calendar date is
recommended (see Appendix X2).
8.5.1 The reference fluid shall be used only at 64 °C. Below that temperature the phase angle differs sufficiently from 90° such
that the values calculated from G* and ω are no longer accurate. Above 64 °C the fluid tends to flow from the gap between the
plates invalidating the measured value for G*.
NOTE 9—A newly installed or reconditioned instrument should be verified on a weekly basis using the procedures in 8.5 until acceptable verification has
been demonstrated.
8.5.2 Measurement of the Viscosity of the Reference Fluid—Measure the complex modulus, G*, of the reference fluid at 64 °C and
10 rad/s. Additional information regarding use of the reference fluid is given in Annex A1. Precaution should be taken to ensure
that there are no air bubbles in the test specimen. Newer rheometers are typically programmed to automatically calculate and report
the measured viscosity, η , in units of mP-s. When making the calculation directly, convert the numerical value of G* in kPa at
M
10 rad/s to viscosity in mPa-s by multiplying the numerical value of G* by 10 . For example, if G* = 2.66 kPa, η = 266 000
M
mPa-s. If the units for G* are in Pa, multiply by 10 .
NOTE 10—The reference fluid may be reported in cP or mPa-s, which are numerically equivalent; for example, 1 mPa = 1cP. The simplified conversion
5 2
given above is provided to make the calculation of η straightforward for the typical user. The multipliers 10 and 10 are necessary for the proper
M
conversion of units.
8.5.3 Verification of Torque Transducer—Verify The viscosity, η the calibration of the torque transducer, measured with the DSR
M
shall be within 3 % of the viscosity, η using a reference fluid or manufacturer-supplied fixtures whenever the calibration of the
R
torque transducer is suspect and/or when the dynamic viscosity, as measured for the reference fluid, indicates that the torque
transducer is not in calibration., reported by the manufacturer of the reference fluid. Otherwise, the overall operation of the DSR
shall be considered suspect. Calculate the estimated percent difference as:
9.5.1.1 Verification of Torque Transducer with Reference Fluid (see Annex A1)—The complex viscosity measured with the DSR
shall be within 3 % of the capillary viscosity as reported by the manufacturer of the reference fluid, otherwise the calibration of
the torque transducer shall be considered suspect. Calculate the complex viscosity as the complex modulus, G* divided by the
angular frequency in rad/s. If the requirements of this section can not be met, discontinue use of the device and consult the
manufacturer or other qualified service personnel.
η 2 η
~ !
R M
percent difference 5 100% × (1)
F G
η
R
NOTE 13—A suitable reference fluid is available from Cannon Instrument Company as viscosity standard number N2700000SP.
where:
η = the viscosity as reported by the supplier of the reference fluid, mP-s (cP), and
R
η = the measured viscosity, converted to mPa-s (cP).
M
9.5.1.2 Verification of Torque Transducer with Fixtures—Verify the calibration of the torque transducer using the manufacturer-
supplied fixtures in accordance with the instructions supplied by the manufacturer. Suitable manufacturer-supplied fixtures are not
widely available. If suitable fixtures are not available, this requirement shall be waived.
8.5.4 Verification of Angular Displacement Transducer—If manufacturer-supplied fixtures are available, verify the calibration
every six months and/or whenever the calibration of the DSR is suspect. If suitable fixtures are the overall DSR operation cannot
be successfully verified according to 8.5not available, this requirement shall be waived., it shall not be used for testing in
accordance with this standard until it has been successfully calibrated by the manufacturer or other qualified service personnel.
9. Preparation of Apparatus
9.1 Prepare the apparatus for testing in accordance with the manufacturer’s recommendations. Specific requirements will vary for
different DSR models and manufacturers.
D7175 − 23
9.2 Inspect Test Plates—Inspect the surfaces of the test plates and discard any plates with jagged or rounded edges or deep
scratches.
9.3 Preparation of Test Plates—Clean any asphalt binder residue from the plates with an organic solvent such as mineral oil,
mineral spirits, a citrus-based solvent, or toluene. Remove any remaining solvent residue by wiping the surface of the plates with
a cotton swab or a soft cloth dampened with reagent grade organic solvent such as heptane, ethyl alcohol, or acetone. If necessary,
use a dry cotton swab or soft cloth to ensure that no moisture condenses on the plates.
9.4 Mount Test Plates and Fixtures—Mount the test plates and fixtures in the DSR taking care to visually ensure that the plates
are parallel to each other and tighten the plates and fixtures to firmly seat them into the DSR, taking care to ensure that the plates
are parallel to each other.If the fixtures or plates are removed for cleaning, attach them as directed by the DSR manufacturer.
9.5 Zero the Test Gap—Select the testing temperature according to the expected grade of the asphalt binder or according to the
pre selected testing schedule. When multiple test temperatures are used, zero the gap at the middle of the expected range of testZero
the test gap in accordance with one of the three options described in 9.5.1 – 9.5.3temperatures. Allow the DSR to reach a stabilized
temperature within 6 0.1°C of test temperature. . If the test temperature differs by more than 6 12°C 612 °C from the temperature
at which the gap is set, re-zero the gap. Zero the gap prior to each time a new specimen is formed between the plates.
NOTE 11—If the instrument has thermal gap compensation, the gap may The frame, detectors, and fixtures in the DSR change dimension with temperature,
causing the zero gap to change with changes in temperature. Adjustments in the gap are not necessary when measurements are made over a limited range
of temperatures. The gap should be set at the first test temperature instead of in test temperature or, when tests are to be conducted over a range of
temperatures, at the middle of the expected range of test temperatures. It is very important to set and zero the gap correctly. Incorrect gap setting produces
significant errors as shown inIf the instrument has thermal gap compensation, the gap may be set at the first test temperature instead of Fig. 3.the middle
of the range of test temperatures.
NOTE 15—Specification D6373 and AASHTO Practice R29 provide guidance on the selection of test temperatures.
9.5.1 Manual Gap Setting—Manually spin the moveable plate, and while the moveable plate is spinning, slowly close the gap. The
zero gap is reached when the plate just stops spinning completely.
9.5.2 Determining Zero Gap—Normal Force Transducer—Establish For rheometers with normal force transducers, set the zero
gap by one of three methods.closing the gap and observing the normal force. Select the zero gap as the position where the normal
force is approximately zero.
10.5.1.1 Manual Gap Setting—Spin the moveable plate, and while the moveable plate is spinning, close the gap until the movable
plate touches the fixed plate. The zero gap is reached when the plate just stops spinning completely.
10.5.1.2 Normal Force Transducer—For rheometers with normal force transducers, set the zero gap by closing the gap and
observing the normal force. After establishing contact between the plates, set the zero gap as the position where the normal force
is approximately zero.
10.5.1.3 Automatic Gap Setting—Zero the gap automatically according to operating procedures specified by the instrument
manufacturer.
9.5.3 Automatic Gap Setting—Zero the gap automatically according to operating procedures specified by the instrument
manufacturer.
9.6 Preheating Test Plates—Once the zero gap is established as per 10.5.19.5.1, move the plates apart to approximately set the
test gap and preheat the plates. Preheating the plates promotes adhesion between the asphalt binder and the plates, especially at
the intermediate grading temperatures.
9.6.1 Preheating 25-mm 25 mm Plate—Bring the test plates to the test temperature or the lowest test temperature if testing is to
be conducted at more than one temperature.
9.6.2 Preheating 8-mm 8 mm Plate—Bring the plates to between 34 and 46°C46 °C to preheat the upper and lower plates.
D7175 − 23
NOTE 12—Specifications D6373 and D8239 and AASHTO R 29 provide guidance on the selection of test temperatures.
NOTE 13—Preheating the test plates is important in order to obtain adequate adhesion between the asphalt binder and the test plates. Preheating is
especially critical when the silicone mold is used to prepare the asphalt binder for transfer to the test plates and when the testing is conducted with the
8 mm plates. When the direct placement method is used, as long as the test plates are immediately brought in contact with
...