General Information

Abstract

IEC 60627:2013 is applicable to anti-scatter grids used in medical diagnostic X-ray imaging equipment. Anti-Scatter Grids are used to reduce the incidence of scattered radiation, produced particularly in the body of the patient, upon the image reception area and thus to improve the contrast of the X-ray pattern. This International Standard specifies the definitions, determination and indication of characteristics of anti-scatter grids. In this standard only linear grids are considered. Since at present only focused grids are used in mammography, this standard is restricted to focused grids where mammographic anti-scatter grids are concerned. This standard is not intended to be applied for acceptance tests. This standard does not cover the homogeneity of performance over the area of a grid. This standard is intended to be applied for the determination of the characteristics of anti-scatter grids under test conditions. These conditions are not usually available at the site of the responsible organization. This third edition cancels and replaces the second edition published in 2001, and constitutes a technical revision. In this revision calcium tungstate phosphor fluorescent screens, which are no longer available, have been replaced by gadolinium oxysulphide (GOS) fluorescent screens. Further, a new quality parameter is introduced: the Image Improvement Factor or Q-factor, which better describes the properties of the anti-scatter grid, especially for digital detector applications. Further differences between this third edition and the previous second edition are:
- new instrumentation is prescribed for measurements of the transmission of primary radiation, the transmission of scattered radiation and the transmission of total radiation, because fluorescent screens made of calcium tungstate phosphors are outdated and are no longer available;
- the definition of the phantom used for measurements of the transmission of primary radiation, the transmission of scattered radiation and the transmission of total radiation is modified and references to IEC 61267 are omitted;
- and the radiation conditions used for the measurements have been adapted and are now the RQR and RQR-M conditions specified in IEC 61267:2005.

Status
Published
Publication Date
08-Jul-2013
Drafting Committee
MT 50 - TC 62/SC 62B/MT 50
Current Stage
PPUB - Publication issued
Start Date
09-Jul-2013
Completion Date
31-Jul-2013

Buy Documents

Standard

IEC 60627:2013 - Diagnostic X-ray imaging equipment - Characteristics of general purpose and mammographic anti-scatter grids Released:7/9/2013

ISBN:978-2-8322-0917-2
English and French language (69 pages)
sale 15% off
Preview
sale 15% off
Preview

Overview

IEC 60627:2013 is an international standard developed by the International Electrotechnical Commission (IEC) that specifies the characteristics of anti-scatter grids used in diagnostic X-ray imaging equipment. Anti-scatter grids are essential components in medical radiography, employed to minimize scattered radiation that occurs within the patient’s body, ultimately enhancing image contrast and diagnostic quality. This standard focuses on linear anti-scatter grids for general diagnostic purposes and mammography, particularly addressing focused grids used in mammographic applications.

The 2013 version is the third edition, updating prior editions by replacing outdated calcium tungstate fluorescent screens with gadolinium oxysulphide (GOS) screens. It also introduces a new quality parameter, the Image Improvement Factor (Q-factor), better suited for digital detector performance evaluation.

Key Topics

  • Scope and Application
    IEC 60627:2013 applies to determination and specification of anti-scatter grid characteristics under controlled test conditions, not for acceptance testing or homogeneity evaluation over the grid surface.

  • Grid Types Covered

    • General purpose linear anti-scatter grids
    • Focused mammographic anti-scatter grids
  • Measurement Methodology
    The standard prescribes new instrumentation and measurement arrangements for:

    • Transmission of primary radiation
    • Transmission of scattered radiation
    • Transmission of total radiation

    These use narrow-beam and broad-beam conditions consistent with updated X-ray radiation quality reference conditions (RQR and RQR-M) defined in IEC 61267:2005.

  • Physical Characteristics Definitions
    Important parameters defined and calculated include:

    • Grid Selectivity (Σ)
    • Contrast Improvement Ratio (K)
    • Grid Exposure Factor (B)
    • Image Improvement Factor (Q) – a novel parameter emphasizing grid performance with digital detectors
  • Standard Requirements
    The standard outlines manufacturing tolerances, application limits, accuracy requirements for physical characteristics, and mandatory markings and documentation to accompany anti-scatter grids.

Applications

IEC 60627:2013 supports the design, testing, and quality control of anti-scatter grids in diagnostic radiology and mammography units. Its application helps medical equipment manufacturers and testing laboratories to:

  • Improve image contrast by optimizing anti-scatter grid characteristics
  • Ensure consistent grid performance in removing scatter radiation while maintaining primary beam transmission
  • Facilitate the integration of anti-scatter grids with digital X-ray detectors using the Image Improvement Factor (Q-factor) for better imaging outcomes
  • Standardize test conditions across different laboratories and manufacturers to support international harmonization
  • Provide clear and standardized documentation and markings for end-users and maintenance personnel

By adhering to IEC 60627, healthcare providers can expect improved diagnostic image quality, leading to more accurate patient diagnoses in X-ray imaging, including sensitive mammographic examinations.

Related Standards

  • IEC 61267:2005 – Medical diagnostic X-ray equipment - Radiation conditions for use in the determination of characteristics
    Provides the radiation quality reference conditions (RQR and RQR-M) used for anti-scatter grid measurements in IEC 60627.

  • IEC 60601-2-54 – Particular requirements for the safety of mammographic X-ray equipment
    Complements IEC 60627 by addressing safety aspects of mammographic imaging systems.

  • ISO 4090:2020 – Medical electrical equipment - Determination of the image quality characteristics of X-ray imaging devices
    Provides broader guidelines on image quality and testing of diagnostic X-ray equipment, including scatter reduction components.


Keywords: IEC 60627, anti-scatter grid, diagnostic X-ray imaging, mammographic grids, scattered radiation reduction, image contrast, Image Improvement Factor, Q-factor, gadolinium oxysulphide screens, X-ray equipment standards, IEC standards, medical radiography, digital detectors.

Relations

Effective Date
05-Sep-2023

Buy Documents

Standard

IEC 60627:2013 - Diagnostic X-ray imaging equipment - Characteristics of general purpose and mammographic anti-scatter grids Released:7/9/2013

ISBN:978-2-8322-0917-2
English and French language (69 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

IEC 60627:2013 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Diagnostic X-ray imaging equipment - Characteristics of general purpose and mammographic anti-scatter grids". This standard covers: IEC 60627:2013 is applicable to anti-scatter grids used in medical diagnostic X-ray imaging equipment. Anti-Scatter Grids are used to reduce the incidence of scattered radiation, produced particularly in the body of the patient, upon the image reception area and thus to improve the contrast of the X-ray pattern. This International Standard specifies the definitions, determination and indication of characteristics of anti-scatter grids. In this standard only linear grids are considered. Since at present only focused grids are used in mammography, this standard is restricted to focused grids where mammographic anti-scatter grids are concerned. This standard is not intended to be applied for acceptance tests. This standard does not cover the homogeneity of performance over the area of a grid. This standard is intended to be applied for the determination of the characteristics of anti-scatter grids under test conditions. These conditions are not usually available at the site of the responsible organization. This third edition cancels and replaces the second edition published in 2001, and constitutes a technical revision. In this revision calcium tungstate phosphor fluorescent screens, which are no longer available, have been replaced by gadolinium oxysulphide (GOS) fluorescent screens. Further, a new quality parameter is introduced: the Image Improvement Factor or Q-factor, which better describes the properties of the anti-scatter grid, especially for digital detector applications. Further differences between this third edition and the previous second edition are: - new instrumentation is prescribed for measurements of the transmission of primary radiation, the transmission of scattered radiation and the transmission of total radiation, because fluorescent screens made of calcium tungstate phosphors are outdated and are no longer available; - the definition of the phantom used for measurements of the transmission of primary radiation, the transmission of scattered radiation and the transmission of total radiation is modified and references to IEC 61267 are omitted; - and the radiation conditions used for the measurements have been adapted and are now the RQR and RQR-M conditions specified in IEC 61267:2005.

IEC 60627:2013 is applicable to anti-scatter grids used in medical diagnostic X-ray imaging equipment. Anti-Scatter Grids are used to reduce the incidence of scattered radiation, produced particularly in the body of the patient, upon the image reception area and thus to improve the contrast of the X-ray pattern. This International Standard specifies the definitions, determination and indication of characteristics of anti-scatter grids. In this standard only linear grids are considered. Since at present only focused grids are used in mammography, this standard is restricted to focused grids where mammographic anti-scatter grids are concerned. This standard is not intended to be applied for acceptance tests. This standard does not cover the homogeneity of performance over the area of a grid. This standard is intended to be applied for the determination of the characteristics of anti-scatter grids under test conditions. These conditions are not usually available at the site of the responsible organization. This third edition cancels and replaces the second edition published in 2001, and constitutes a technical revision. In this revision calcium tungstate phosphor fluorescent screens, which are no longer available, have been replaced by gadolinium oxysulphide (GOS) fluorescent screens. Further, a new quality parameter is introduced: the Image Improvement Factor or Q-factor, which better describes the properties of the anti-scatter grid, especially for digital detector applications. Further differences between this third edition and the previous second edition are: - new instrumentation is prescribed for measurements of the transmission of primary radiation, the transmission of scattered radiation and the transmission of total radiation, because fluorescent screens made of calcium tungstate phosphors are outdated and are no longer available; - the definition of the phantom used for measurements of the transmission of primary radiation, the transmission of scattered radiation and the transmission of total radiation is modified and references to IEC 61267 are omitted; - and the radiation conditions used for the measurements have been adapted and are now the RQR and RQR-M conditions specified in IEC 61267:2005.

IEC 60627:2013 is classified under the following ICS (International Classification for Standards) categories: 11.040.50 - Radiographic equipment; 33.060.30 - Radio relay and fixed satellite communications systems. The ICS classification helps identify the subject area and facilitates finding related standards.

IEC 60627:2013 has the following relationships with other standards: It is inter standard links to IEC 60627:2001. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

IEC 60627:2013 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


IEC 60627 ®
Edition 3.0 2013-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Diagnostic X-ray imaging equipment –
Characteristics of general purpose and mammographic anti-scatter grids

Équipements de diagnostic par imagerie à rayonnement X –
Caractéristiques des grilles antidiffusantes d'usage général et de
mammographie
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.

Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les
microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.

IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.

Useful links:
IEC publications search - www.iec.ch/searchpub Electropedia - www.electropedia.org
The advanced search enables you to find IEC publications The world's leading online dictionary of electronic and
by a variety of criteria (reference number, text, technical electrical terms containing more than 30 000 terms and
committee,…). definitions in English and French, with equivalent terms in
It also gives information on projects, replaced and additional languages. Also known as the International
withdrawn publications. Electrotechnical Vocabulary (IEV) on-line.

IEC Just Published - webstore.iec.ch/justpublished Customer Service Centre - webstore.iec.ch/csc
Stay up to date on all new IEC publications. Just Published If you wish to give us your feedback on this publication
details all new publications released. Available on-line and or need further assistance, please contact the
also once a month by email. Customer Service Centre: csc@iec.ch.

A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.

A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.

Liens utiles:
Recherche de publications CEI - www.iec.ch/searchpub Electropedia - www.electropedia.org
La recherche avancée vous permet de trouver des Le premier dictionnaire en ligne au monde de termes
publications CEI en utilisant différents critères (numéro de électroniques et électriques. Il contient plus de 30 000
référence, texte, comité d’études,…). termes et définitions en anglais et en français, ainsi que
Elle donne aussi des informations sur les projets et les les termes équivalents dans les langues additionnelles.
publications remplacées ou retirées. Egalement appelé Vocabulaire Electrotechnique
International (VEI) en ligne.
Just Published CEI - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications de la CEI.
Just Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur
Disponible en ligne et aussi une fois par mois par email. cette publication ou si vous avez des questions
contactez-nous: csc@iec.ch.
IEC 60627 ®
Edition 3.0 2013-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Diagnostic X-ray imaging equipment –

Characteristics of general purpose and mammographic anti-scatter grids

Équipements de diagnostic par imagerie à rayonnement X –

Caractéristiques des grilles antidiffusantes d'usage général et de

mammographie
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX V
ICS 11.040.50 ISBN 978-2-8322-0917-2

– 2 – 60627 © IEC:2013
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
1 Scope . 8
2 Normative references . 8
3 Terms and definitions . 8
3.1 Grid definitions . 9
3.2 Geometric characteristics . 10
3.3 Physical characteristics . 10
3.4 Other terms . 11
4 Structure of ANTI-SCATTER GRIDS . 12
5 Measurement and determination of physical characteristics . 12
5.1 Method and arrangement for measurement . 12
5.1.1 Determination of physical characteristics . 12
5.1.2 Instrumentation . 13
5.1.3 Phantoms . 14
5.1.4 Arrangements . 14
5.1.5 Radiation conditions . 16
5.1.6 Constancy of source . 16
5.2 Physical characteristics . 16
5.2.1 Measurements for the TRANSMISSION OF PRIMARY RADIATION (T ). 16
p
5.2.2 Measurements for the TRANSMISSION OF SCATTERED RADIATION (T ) . 16
s
5.2.3 Measurements for the TRANSMISSION OF TOTAL RADIATION (T ) . 17
t
5.2.4 Calculation of the GRID SELECTIVITY (Σ) . 17
5.2.5 Calculation of the CONTRAST IMPROVEMENT RATIO (K) . 17
5.2.6 Calculation of the GRID EXPOSURE FACTOR (B) . 17
5.2.7 Calculation of the IMAGE IMPROVEMENT FACTOR (Q) . 17
5.2.8 Accuracy of measurements . 17
6 Requirements for ANTI-SCATTER GRIDS . 18
6.1 Manufacturing tolerances . 18
6.2 Determination of the APPLICATION LIMITS . 18
6.3 Accuracy of characteristics . 18
6.3.1 GRID SELECTIVITY . 18
6.3.2 CONTRAST IMPROVEMENT RATIO . 18
6.3.3 GRID EXPOSURE FACTOR . 18
6.3.4 IMAGE IMPROVEMENT FACTOR . 19
6.4 Markings and ACCOMPANYING DOCUMENTS . 19
6.4.1 Data in ACCOMPANYING DOCUMENTS . 19
6.4.2 Mandatory markings and indications for LINEAR GRIDS . 19
6.4.3 Mandatory markings and indications for FOCUSED GRIDS . 19
6.4.4 Additional mandatory markings and indications . 19
6.4.5 Further requirements . 20
Annex A (normative) Calculation of the APPLICATION LIMITS . 27
Annex B (informative) Influence of scatter fraction on the physical characteristics . 30
Bibliography . 32
Index of defined terms used in this standard . 33

60627 © IEC:2013 – 3 –
Figure 1 – Structure of ANTI-SCATTER GRIDS . 21
Figure 2 – RADIATION DETECTOR . 22
Figure 3 – Measurement arrangement for general purpose ANTI-SCATTER GRIDS with
NARROW-BEAM CONDITION (determination of the TRANSMISSION OF PRIMARY RADIATION) . 23
Figure 4 – Measurement arrangement for general-purpose ANTI-SCATTER GRIDS with
BROAD BEAM CONDITION (determination of the TRANSMISSION OF SCATTERED RADIATION). 24
Figure 5 – Measurement arrangement for MAMMOGRAPHIC ANTI-SCATTER GRIDS with
NARROW-BEAM CONDITION (determination of the TRANSMISSION OF PRIMARY RADIATION) . 25
Figure 6 – Measurement arrangement for MAMMOGRAPHIC ANTI-SCATTER GRIDS with
BROAD BEAM CONDITION (determination of the TRANSMISSION OF SCATTERED RADIATION). 26
Figure B.1 – Physical characteristics as function of scatter fraction: SELECTIVITY Σ
(dash-dotted line), CONTRAST IMPROVEMENT RATIO K (dotted line), GRID EXPOSURE
B (dashed line), and IMAGE IMPROVEMENT FACTOR Q (solid line) . 31
FACTOR
– 4 – 60627 © IEC:2013
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
DIAGNOSTIC X-RAY IMAGING EQUIPMENT –

Characteristics of general purpose and
mammographic anti-scatter grids

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60627 has been prepared by subcommittee 62B: Diagnostic
imaging equipment, of IEC technical committee 62: Electrical equipment in medical practice.
This third edition cancels and replaces the second edition published in 2001, and constitutes
a technical revision.
In this revision calcium tungstate phosphor FLUORESCENT SCREENS, which are no longer
available, have been replaced by gadolinium oxysulphide (GOS) FLUORESCENT SCREENS.
IMAGE IMPROVEMENT FACTOR or Q-factor,
Further, a new quality parameter is introduced: the
which better describes the properties of the ANTI-SCATTER GRID, especially for digital detector
applications.
Further differences between this third edition and the previous second edition are:
– some definitions have been modified and others added to improve clarity, harmonization
or generality;
60627 © IEC:2013 – 5 –
– new instrumentation is prescribed for measurements of the TRANSMISSION OF PRIMARY
RADIATION, the TRANSMISSION OF SCATTERED RADIATION and the TRANSMISSION OF TOTAL
RADIATION, because FLUORESCENT SCREENS made of calcium tungstate phosphors are
outdated and are no longer available;
– the definition of the PHANTOM used for measurements of the TRANSMISSION OF PRIMARY
RADIATION, the TRANSMISSION OF SCATTERED RADIATION and the TRANSMISSION OF TOTAL
RADIATION is modified and references to IEC 61267 are omitted;
– the RADIATION CONDITIONS used for the measurements have been adapted and are now the
RQR and RQR-M conditions specified in IEC 61267:2005;
– tolerances are specified for the dimensions in the arrangements for the measurements of
the TRANSMISSION OF PRIMARY RADIATION, the TRANSMISSION OF SCATTERED RADIATION and
the TRANSMISSION OF TOTAL RADIATION.
The text of this standard is based on the following documents:
FDIS Report on voting
62B/914/FDIS 62B/922/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
Informative material appearing outside of tables, such as notes, examples and references: smaller type. Normative text of
–
tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THIS STANDARD OR IN OTHER IEC PUBLICATIONS REFERENCED IN
THIS STANDARD: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the numbered divisions within the table of contents, inclusive of all
subdivisions (e.g., Clause 5 includes subclauses 5.1, 5.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g., 5.1, 5.2 and 5.2.1 are all
subclauses of Clause 5).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or”, so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
– 6 – 60627 © IEC:2013
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
NOTE The attention of National Committees is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC publication in
which to make products in accordance with the new requirements and to equip themselves for conducting new or
revised tests.
It is the recommendation of the committee that the content of this publication be adopted for implementation
nationally not earlier than 36 months from the date of publication.

IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
60627 © IEC:2013 – 7 –
INTRODUCTION
The first edition of IEC 60627 was intended for ANTI-SCATTER GRIDS used in general
radiography and is not appropriate for ANTI-SCATTER GRIDS used in mammography. As a
consequence, a complementary standard IEC 61953 was published. Later, it was decided to
revise and merge together the two standards covering ANTI-SCATTER GRIDS. Wherever
possible, a harmonized approach has been used. This constituted the second edition of
IEC 60627 published in 2001.
This third edition is a revision of the second edition. This revision was initiated by the fact that
calcium tungstate phosphors have become obsolete, and are no longer available.
Instrumentation with FLUORESCENT SCREENS made of gadolinium oxysulphide (GOS) is the
present state of the art.
Further, a new quality parameter is introduced: the IMAGE IMPROVEMENT FACTOR Q. This factor
better describes the properties of ANTI-SCATTER GRIDS than the GRID EXPOSURE FACTOR B and
the CONTRAST IMPROVEMENT FACTOR K, especially for digital detector applications. Namely, the
signal-to-noise ratio (SNR) for digital X-ray detectors is increased proportionally with the
square root of the factor Q when an ANTI-SCATTER GRID is applied. This effect is due to the
efficient reduction of SCATTERED RADIATION and overcompensates the loss of PRIMARY
RADIATION when using an ANTI-SCATTER GRID in situations where a considerable amount of
SCATTERED RADIATION is present. The name IMAGE IMPROVEMENT FACTOR is chosen to reflect
the improved image quality (characterized by SNR and other parameters) under equal
RADIATION dose conditions.
Special laboratory provisions and carefully controlled test conditions are needed for the
measurements described here.
– 8 – 60627 © IEC:2013
DIAGNOSTIC X-RAY IMAGING EQUIPMENT –

Characteristics of general purpose and
mammographic anti-scatter grids

1 Scope
This International Standard is applicable to ANTI-SCATTER GRIDS used in medical diagnostic X-
ray imaging equipment. ANTI-SCATTER GRIDS are used to reduce the incidence of SCATTERED
RADIATION, produced particularly in the body of the PATIENT, upon the IMAGE RECEPTION AREA
and thus to improve the contrast of the X-RAY PATTERN. This International Standard specifies
the definitions, determination and indication of characteristics of ANTI-SCATTER GRIDS.
In this standard only LINEAR GRIDS are considered.
Since at present only FOCUSED GRIDS are used in mammography, this standard is restricted to
FOCUSED GRIDS where MAMMOGRAPHIC ANTI-SCATTER GRIDS are concerned.
This standard is not intended to be applied for ACCEPTANCE TESTS.
This standard does not cover the homogeneity of performance over the area of a grid.
This standard is intended to be applied for the determination of the characteristics of ANTI-
SCATTER GRIDS under test conditions. These conditions are not usually available at the site of
the RESPONSIBLE ORGANIZATION.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
Amendment 1:2012
IEC 60601-1-3:2008, Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic
X-ray equipment
Amendment 1:2013
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC 61267:2005, Medical diagnostic X-ray equipment – Radiation conditions for use in the
determination of characteristics
3 Terms and definitions
For the purposes of this document, the terms and definitions given in
IEC 60601-1:2005+A1:2012, IEC 60601-1-3:2008+A1:2013, IEC/TR 60788:2004 and the
following apply.
60627 © IEC:2013 – 9 –
3.1 Grid definitions
3.1.1
ANTI-SCATTER GRID
device to be placed before the IMAGE RECEPTION AREA in order to reduce the incidence of
SCATTERED RADIATION upon that area and thus increase the contrast in the X-RAY PATTERN
3.1.2
LINEAR GRID
ANTI-SCATTER GRID composed of highly absorbing strips and highly transmitting interspaces
which are parallel in their longitudinal direction
3.1.3
PARALLEL GRID
LINEAR GRID in which the planes of the absorbing strips are parallel to each other and
perpendicular to the incident face
3.1.4
FOCUSED GRID
LINEAR GRID in which the planes of the absorbing strips converge to a straight line at the
FOCUSING DISTANCE
3.1.5
TAPERED GRID
LINEAR GRID in which the height of the absorbing strips decreases as the distance between the
absorbing strips and the TRUE CENTRAL LINE increases. This decrease is symmetrical about
the TRUE CENTRAL LINE
3.1.6
CROSS-GRID
ANTI-SCATTER GRID composed of two LINEAR GRIDS built together in such a way that the
directions of their absorbing strips form an angle
3.1.7
ORTHOGONAL CROSS-GRID
CROSS-GRID in which the directions of the absorbing strips form an angle of 90°
3.1.8
OBLIQUE CROSS-GRID
CROSS-GRID in which the directions of the absorbing strips form an angle other than 90°
3.1.9
STATIONARY GRID
ANTI-SCATTER GRID used in such a way that it does not move in relation to the RADIATION BEAM
3.1.10
MOVING GRID
ANTI-SCATTER GRID used in an ACCESSORY that enables the ANTI-SCATTER GRID to be moved,
when irradiated by a RADIATION BEAM, in order to avoid the imaging of the absorbing strips and
the consequent loss of information
3.1.11
MAMMOGRAPHIC ANTI-SCATTER GRID
FOCUSED GRID specially designed for mammography
Note 1 to entry: In this standard the term "general purpose ANTI-SCATTER GRID" is used to describe any ANTI-
SCATTER GRID not specially designed for mammography.

– 10 – 60627 © IEC:2013
3.2 Geometric characteristics
3.2.1
STRIP FREQUENCY
N
–1
number of absorbing strips per unit length of a LINEAR GRID (unit: cm )
3.2.2
GRID RATIO
r
ratio between the height of the absorbing strips and the distance between the absorbing strips
in the centre of a LINEAR GRID
3.2.3
FOCUSING DISTANCE
f
distance between the incident face of a FOCUSED GRID and the line into which the planes of the
absorbing strips of the grid converge (unit: cm)
Note 1 to entry: Attention is drawn to the differences between "FOCUSING DISTANCE", "FOCAL SPOT to grid distance"
and "FOCAL SPOT to film distance".
3.2.4
APPLICATION LIMITS
f , f
1 2
lower, f , and upper, f , limits of the distance from the FOCAL SPOT to the incident face of a
1 2
FOCUSED GRID or a PARALLEL GRID between which the obtained radiological information can be
considered acceptable for many purposes (unit: cm)
Note 1 to entry: See Annex A for details on the calculation of the APPLICATION LIMITS.
3.2.5
TRUE CENTRAL LINE
– for a PARALLEL GRID: line on the incident face in the direction of the absorbing strips
and passing through the centre of the grid area
– for a FOCUSED GRID: perpendicular projection onto the incident face of the grid, of the
line into which the planes of the absorbing strips converge
– for a TAPERED GRID: line on the incident face in the direction of the absorbing strips
and lying within a symmetry plane of the grid structure
Note 1 to entry: A CROSS-GRID has two TRUE CENTRAL LINES.
3.2.6
CENTRAL-LINE INDICATION
marking on the incident face of a LINEAR GRID, which is intended to indicate the position and
direction of the TRUE CENTRAL LINE
Note 1 to entry: In most cases, this marking coincides with the geometric centre line of the grid's incident face.
3.3 Physical characteristics
3.3.1
TRANSMISSION OF PRIMARY RADIATION
T
p
characteristic of an object, evaluated as the ratio of the MEASURED VALUE of the quantity or
rate of PRIMARY RADIATION with the object placed in a RADIATION BEAM to that with the object
removed from the beam, under specific measuring conditions

60627 © IEC:2013 – 11 –
3.3.2
TRANSMISSION OF SCATTERED RADIATION
T
s
characteristic of an object, evaluated as the ratio of the MEASURED VALUE of the quantity or
rate of SCATTERED RADIATION with the object placed in a RADIATION BEAM to that with the object
removed from the beam, under specific measuring conditions
3.3.3
TRANSMISSION OF TOTAL RADIATION
T
t
characteristic of an object, evaluated as the ratio of the MEASURED VALUE of the quantity or
rate of total RADIATION with the object placed in a RADIATION BEAM to that with the object
removed from the beam, under specific measuring conditions
3.3.4
GRID SELECTIVITY
Σ
characteristic of an ANTI-SCATTER GRID, evaluated as the ratio of the TRANSMISSION OF PRIMARY
RADIATION to the TRANSMISSION OF SCATTERED RADIATION
3.3.5
CONTRAST IMPROVEMENT RATIO
K
characteristic of an ANTI-SCATTER GRID, evaluated as the ratio of the TRANSMISSION OF PRIMARY
RADIATION to the TRANSMISSION OF TOTAL RADIATION
3.3.6
GRID EXPOSURE FACTOR
B
characteristic of an ANTI-SCATTER GRID, evaluated as the reciprocal value of the TRANSMISSION
OF TOTAL RADIATION
3.3.7
IMAGE IMPROVEMENT FACTOR
Q
characteristic of an ANTI-SCATTER GRID, evaluated as the ratio of the square of the
TRANSMISSION OF PRIMARY RADIATION to the TRANSMISSION OF TOTAL RADIATION
3.4 Other terms
3.4.1
DECENTRING
distance between the TRUE CENTRAL LINE of a FOCUSED GRID and the perpendicular projection
of the FOCAL SPOT of an X-RAY TUBE onto the incident face of the grid
3.4.2
DEFOCUSING
difference between the distance from the FOCAL SPOT of an X-RAY TUBE to the incident face of
a FOCUSED GRID and the FOCUSING DISTANCE of that grid
Note 1 to entry: See Clause A.1 for an explanation of DECENTRING and DEFOCUSING.
3.4.3
SERIAL NUMBER
number and/or other designation to identify an individual unit of a certain model of equipment
or ACCESSORY
– 12 – 60627 © IEC:2013
4 Structure of ANTI-SCATTER GRIDS
ANTI-SCATTER GRIDS usually consist of strips of highly absorbent material, of thickness d and
height h, arranged at regular intervals D from each other; see Figure 1.
The height h of the strips is either constant over the area of the ANTI-SCATTER GRID or
decreases in TAPERED GRIDS from the highest strip – with height h – towards two edges.
NOTE D and d are measured at the incident face of the grid.
The interspaces between the strips are usually filled with highly transmitting material. The
ANTI-SCATTER GRID may have a frame and covers to protect against mechanical damage and to
ensure the rigidity of the grid.
The STRIP FREQUENCY shall be determined according to the formula
N =
(d + D)
The GRID RATIO shall be determined according to one of the following formulae
– for a PARALLEL GRID and for a FOCUSED GRID:
h
r =
D
– for a TAPERED GRID:
h
r =
D
– for a CROSS-GRID:
h
r =
D
h
r =
D
Letter symbols without index designate general properties of a LINEAR GRID. The index "0"
designates quantities at the TRUE CENTRAL LINE. Indices "1" or "2" designate quantities for the
LINEAR GRIDS forming a CROSS-GRID.
5 Measurement and determination of physical characteristics
5.1 Method and arrangement for measurement
5.1.1 Determination of physical characteristics
For the purpose of this standard, the values of the TRANSMISSION OF PRIMARY RADIATION, the
TRANSMISSION OF SCATTERED RADIATION and the TRANSMISSION OF TOTAL RADIATION shall be
determined as the ratio of the two MEASURED VALUES obtained with the instrumentation
described in 5.1.2 and the PHANTOM described in 5.1.3, in the arrangements described in
5.1.4, and with the RADIATION CONDITIONS described in 5.1.5.

60627 © IEC:2013 – 13 –
5.1.2 Instrumentation
5.1.2.1 General
A RADIATION DETECTOR shall be used which incorporates a FLUORESCENT SCREEN and a photo-
detector; see Figure 2. The FLUORESCENT SCREEN shall be made from a terbium-activated
gadolinium oxysulphide (GOS, Gd O S:Tb) scintillator, preferably without dye.
2 2
The area density of the scintillator shall be
–2 –2
a) for general purpose ANTI-SCATTER GRIDS 75 mg ⋅ cm ± 10 mg ⋅ cm ;
–2 –2
b) for MAMMOGRAPHIC ANTI-SCATTER GRIDS 30 mg ⋅ cm ± 3 mg ⋅ cm .
NOTE The previously specified scintillator material, calcium tungstate, is no longer commercially available.
Current state-of-the-art is GOS, which is non-toxic and non-hygroscopic as opposed to thallium-doped cesium
iodide (CsI:Tl). Experimental results show equivalent outcome for calcium tungstate and GOS scintillators.
The diameter of the measuring field shall be 6,0 mm ± 0,5 mm.
The luminance produced at the FLUORESCENT SCREEN shall be measured with the photo-
detector, which shall be sensitive in the energy range of the light photons produced.
The ADDITIONAL FILTRATION between the supporting plane of the ANTI-SCATTER GRID and the
active layer of the FLUORESCENT SCREEN shall be not more than
a) for general purpose ANTI-SCATTER GRIDS 0,5 mm Al;
b) for MAMMOGRAPHIC ANTI-SCATTER GRIDS 0,1 mm Al
for the RADIATION CONDITION applied.
The dark current and the direct IRRADIATION of the photo-detector shall not significantly affect
the result of the measurements.
The response of the photo-detector shall be linearly proportional to the RADIATION intensity.
5.1.2.2 Test for dark current and direct IRRADIATION
Use the following test PROCEDURE to check the effects of the dark current and the direct
IRRADIATION of the photo-detector:
a) use an arrangement as described in 5.2.3, except that the ANTI-SCATTER GRID is removed,
and apply a RADIATION CONDITION as specified in 5.1.5;
b) measure the detector signal at the maximum X-RAY TUBE CURRENT used for the grid
measurements with the photo-detector shielded by X-ray transparent material against the
visible light excited in the FLUORESCENT SCREEN and with the photo-detector unshielded;
c) measure the detector signal without IRRADIATION (this is the dark-current value of the
RADIATION DETECTOR);
d) calculate the ratio of the MEASURED VALUES for the shielded and unshielded measurements
after subtraction of the dark-current value;
e) this ratio shall not exceed 0,002.
5.1.2.3 Test for linearity
Use the following test PROCEDURE to check the linearity of the photo-detector:
a) use an arrangement as described in 5.2.3, except that the ANTI-SCATTER GRID is removed,
and apply a RADIATION CONDITION as specified in 5.1.5;

– 14 – 60627 © IEC:2013
b) measure the detector signal at the maximum X-RAY TUBE CURRENT used for the grid
measurements, at one half of that current, and without IRRADIATION, while using the same
X-RAY TUBE VOLTAGE for these three measurements;
c) the MEASURED VALUE at one half of the maximum X-RAY TUBE CURRENT shall be within ± 5 %
MEASURED VALUES.
of the average of the other two
5.1.3 Phantoms
a) For general purpose ANTI-SCATTER GRIDS, the PHANTOM used for the determination of
TRANSMISSION OF PRIMARY RADIATION and for the determination of TRANSMISSION OF
SCATTERED RADIATION shall be a water-filled container. The container shall have
– sides of outside dimensions of 300 mm ± 1 mm and height of 200 mm ± 1 mm;
– top and bottom surfaces and side walls made of polymethyl-methacrylate (PMMA),
each having 10 mm ± 2 mm thickness; and
– the interior filled with water.
When used under NARROW-BEAM CONDITIONS, the above PHANTOM may be replaced by a
PHANTOM identical except for reduced outside dimensions. This equivalence should be
checked.
Alternatively to the above PHANTOM, a PHANTOM made up of water-equivalent solid material
having the same overall dimensions as the container, may be used. This equivalence
should be checked.
b) b) For MAMMOGRAPHIC ANTI-SCATTER GRIDS, the PHANTOM used for the determination of
TRANSMISSION OF PRIMARY RADIATION and for the determination of TRANSMISSION OF
SCATTERED RADIATION shall be a PMMA block of square cross-section with sides of
150 mm ± 1 mm and a thickness of 50 mm ± 1 mm.
5.1.4 Arrangements
a) For general purpose ANTI-SCATTER GRIDS, the arrangements for the measurements shall be
according to the arrangements shown in Figure 3 (PHANTOM in the upper position,
NARROW-BEAM CONDITION) and Figure 4 (PHANTOM in the lower position, BROAD BEAM
CONDITION).
The positions of the FOCAL SPOT, general purpose ANTI-SCATTER GRID and measuring field
are the same for both configurations. All distances in the figures are given with a tolerance
of maximum ± 10 mm, unless specified otherwise.
The distance from the FOCAL SPOT to the supporting plane of the general purpose ANTI-
SCATTER GRID shall be 100 cm (1 000 mm), even if the FOCUSING DISTANCE of the FOCUSED
GRID under consideration is not 100 cm. For the geometry described, the results of the
measurements are insensitive to the FOCUSING DISTANCE f .
The general purpose ANTI-SCATTER GRID shall be so fixed that the central line as defined by
the CENTRAL-LINE INDICATION is above the centre of the measuring field. The supporting
plane of the grid shall be perpendicular within ± 0,2° to the plane containing the FOCAL
SPOT and the central line as defined by the CENTRAL-LINE INDICATION. The distance
between the supporting plane of the grid and the output plane of the FLUORESCENT SCREEN
of the RADIATION DETECTOR shall be 20 mm.
The DIAPHRAGMS shown in Figures 3 and 4 and the PRIMARY RADIATION blocker shown in
Figure 4 shall be made of lead of a thickness of 5 mm ± 1 mm. The upper DIAPHRAGM shall
be positioned at a distance between 150 mm and 300 mm from the FOCAL SPOT. The lower
DIAPHRAGM shall be positioned at a distance of 220 mm from the supporting plane of the
grid. For the measurements in NARROW-BEAM CONDITION (see Figure 3), the PHANTOM shall
be positioned with its top surface against the upper DIAPHRAGM with an extra DIAPHRAGM
positioned against its bottom surface. For the measurements in BROAD BEAM CONDITION
(see Figure 4), the PHANTOM shall be positioned with its top surface against the lower
DIAPHRAGM so that its bottom surface is at a distance of 20 mm from the supporting plane
of the grid.
60627 © IEC:2013 – 15 –
b) For MAMMOGRAPHIC ANTI-SCATTER GRIDS, the arrangements for the measurements shall be
according to the arrangements shown in Figure 5 (PHANTOM in the upper position, NARROW-
BEAM CONDITION) and Figure 6 (PHANTOM in the lower position, BROAD BEAM CONDITION).
The positions of the FOCAL SPOT, MAMMOGRAPHIC ANTI-SCATTER GRID and measuring field
are the same for both configurations. They are described in terms of the arrangements
with the PHANTOM in the lower position (see Figure 6). All distances in the figures are given
with a tolerance of maximum ± 5 mm, unless specified otherwise.
The distance from the FOCAL SPOT to the supporting plane of the MAMMOGRAPHIC ANTI-
SCATTER GRID shall be 60 cm (600 mm), even if the FOCUSING DISTANCE of the grid under
consideration is not 60 cm. For the geometry described, the results of the measurements
are insensitive to the FOCUSING DISTANCE f .
For the measurement with the MAMMOGRAPHIC ANTI-SCATTER GRID in place and the
PHANTOM in the lower position (see Figure 6), the FOCAL SPOT, the centre of the bottom
surface of the PHANTOM, and the centre of the measuring field shall be co-linear. The
normal from the FOCAL SPOT to the bottom surface of the PHANTOM shall bisect one of the
side faces of the PHANTOM.
The incident face of the MAMMOGRAPHIC ANTI-SCATTER GRID shall be parallel to the bottom
surface of the PHANTOM. The central line of the grid as defined by the CENTRAL-LINE
INDICATION shall be parallel to a side of the PHANTOM. The chest-wall side of the grid, if
applicable, shall be oriented to, and aligned with, that side of the PHANTOM which is
bisected by the normal from the FOCAL SPOT to the bottom surface of the PHANTOM.
The MAMMOGRAPHIC ANTI-SCATTER GRID shall be so aligned that the normal from the FOCAL
SPOT to the bottom surface of the PHANTOM intersects the incident face of the grid at the
central line as defined by the CENTRAL-LINE INDICATION. The supporting plane of the grid
shall be perpendicular within ± 0,2° to the plane containing the FOCAL SPOT and the central
line as defined by the CENTRAL-LINE INDICATION. The distance between the supporting plane
of the grid and the output plane of the FLUORESCENT SCREEN of the RADIATION DETECTOR
shall be 10 mm.
The measurement configuration described above may need to be modified for either or
both of the following special situations:
– the grid is smaller than the PHANTOM:
the grid shall then be positioned so that the FOCAL SPOT, the centre of the grid and the
centre of the measuring field are co-linear;
– the grid lines run parallel to the chest-wall side:
the grid shall then be tilted to compensate for DECENTRING, which shall be achieved by
raising or lowering the side of the grid distant from the CENTRAL-LINE INDICATION, such
that the absorbing strips above the measuring field are directed towards the FOCAL
SPOT and the TRANSMISSION OF PRIMARY RADIATION is maximized.
NOTE In practice, a tilting angle of approximately 7° is sufficient.
The DIAPHRAGMS shown in Figures 5 and 6 and the PRIMARY RADIATION blocker shown in
Figure 6 shall be made of lead of a thickness between 1 mm and 2 mm. The upper
DIAPHRAGM shall be positioned at a distance of 200 mm or less from the FOCAL SPOT. The
lower DIAPHRAGM shall be positioned at a distance of 60 mm from the supporting plane of
MAMMOGRAPHIC ANTI-SCATTER GRID. For the measurements in BROAD BEAM CONDITION
the
(see Figure 6), the PHANTOM shall be positioned with its top surface against the lower
DIAPHRAGM so that its bottom surface is at a distance of 10 mm from the supporting plane
of the grid.
NARROW-BEAM CONDITION (see Figure 5) shall be
The arrangement for the measurements in
the same as described above, except that the PHANTOM is positioned with its top surface
against the upper DIAPHRAGM, close to the X-RAY TUBE with the RADIATION BEAM passing the
PHANTOM near its centre, and with an extra DIAPHRAGM positioned against the bottom
surface of the PHANTOM.
– 16 – 60627 © IEC:2013
5.1.5 Radiation conditions
a) For general-purpose ANTI-SCATTER GRIDS, the measurements shall be performed with
RADIATION CONDITION RQR 6, as defined in IEC 61267:2005, applied in NARROW-BEAM or
BROAD BEAM CONDITIONS as specified in 5.1.4 a).
Where a general-purpose ANTI-SCATTER GRID is specified for low-energy use, additional
measurements may be performed with RADIATION CONDITION RQR 4.
Where a general-purpose ANTI-SCATTER GRID is specified for high-energy use, additional
measurements may be performed with RADIATION CONDITION RQR 9.
b) For MAMMOGRAPHIC ANTI-SCATTER GRIDS, the measurements shall be performed with
RADIATION CONDITION RQR-M 2, as defined in IEC 61267:2005, applied in NARROW-BEAM or
BROAD BEAM CONDITIONS as specified in 5.1.4 b).
An X-RAY TUBE producing little EXTRA-FOCAL RADIATION should be selected.
NOTE The use of an X-RAY TUBE with a large amount of EXTRA-FOCAL RADIATION can slightly increase the value of
the TRANSMISSION OF SCATTERED RADIATION.
5.1.6 Constancy of source
The LOADING FACTORS of the X-RAY TUBE shall be so controlled that the effect of energy
fluence rate variations on the accuracy of each individual measurement is less than ± 0,5 %.
5.2 Physical characteristics
5.2.1 Measurements for the TRANSMISSION OF PRIMARY RADIATION (T )
p
The two measurements required for the determination of the TRANSMISSION OF PRIMARY
RADIATION shall be performed under NARROW-BEAM CONDITIONS:
– with the PHANTOM and the ANTI-SCATTER GR
...