Abstract

This document establishes a vocabulary of terms and semantics for all fields of respiratory care involving mechanical ventilation, such as intensive-care ventilation, anaesthesia ventilation, emergency and transport ventilation and home-care ventilation, including sleep-apnoea breathing-therapy equipment. It is applicable
—     in lung ventilator and breathing-therapy device standards,
—     in health informatics standards,
—     for labelling on medical electrical equipment and medical electrical systems,
—     in medical electrical equipment and medical electrical system instructions for use and accompanying documents,
—     for medical electrical equipment and medical electrical systems interoperability, and
—     in electronic health records.
This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator and ventilator breathing system.
NOTE    This document can also be used for other applications relating to lung ventilation, including non-electrical devices and equipment, research, description of critical events, forensic analysis and adverse event (vigilance) reporting systems.
This document does not specify terms specific to breathing-therapy equipment, or to physiologic closed-loop ventilation, high-frequency ventilation or negative-pressure ventilation; nor to respiratory support using liquid ventilation or extra-corporeal gas exchange, or oxygen, except where it has been considered necessary to establish boundaries between bordering concepts.

Status
Not Published
Public Enquiry End Date
26-Oct-2026
Technical Committee
VAZ - Healthcare
Current Stage
4020 - Public enquire (PE) (Adopted Project)
Start Date
14-Aug-2026
Due Date
01-Jan-2027
Completion Date
02-Nov-2026

Buy Documents

Draft

oSIST prEN ISO 19223-1:2026 - BARVE

English language (141 pages)
Preview
Preview
e-Library read for
×1 day

oSIST prEN ISO 19223-1:2026 is the Slovenian draft adoption of ISO/DIS 19223-1:2026, published by SIST. It specifies vocabulary and semantics for lung ventilators in respiratory care, including intensive-care, anaesthesia, emergency and transport, home-care, and sleep-apnoea applications. The document is written for standards work, device labelling, instructions for use, interoperability, health informatics, and electronic health records.

What does oSIST prEN ISO 19223-1:2026 specify?

oSIST prEN ISO 19223-1:2026 defines the terms used for artificial ventilation and the relationships between those terms. The scope is terminology, not ventilator design, performance limits, or test methods.

The document applies to:

  • lung ventilator and breathing-therapy device standards
  • health informatics standards
  • labelling on medical electrical equipment and medical electrical systems
  • instructions for use and accompanying information
  • interoperability
  • electronic health records

It also applies to accessories intended by the manufacturer to be connected to a ventilator breathing system or a ventilator when those accessories can affect basic safety or essential performance.

The document does not cover terms specific to high-frequency ventilation or jet ventilation, breathing-therapy equipment, physiologic closed-loop ventilation, negative-pressure ventilation, liquid ventilation, extra-corporeal gas exchange, or oxygen, except where needed to separate bordering concepts.

Clause 1 gives the scope, clause 2 addresses references, and clause 3 contains the terms, definitions, symbols, and abbreviated terms. The main body of clause 3 is organised by topic, from general artificial-ventilation terminology through breath, flow, pressure, volume, initiation and termination, baseline and PEEP, mode terminology, bi-level terminology, and safety limits and alarm terminology.

AnnexWhat it covers
Annex AIllustrations of ventilation terms
Annex BConcepts relating to baseline airway pressures and PEEP
Annex CSystematic Coding for Health Informatics and Medical Device Communications
Annex DImplementation of this terminology in ventilator labelling and instructions
Annex EAlphabetized index of defined terms

What are the key requirements of oSIST prEN ISO 19223-1:2026?

oSIST prEN ISO 19223-1:2026 requires consistent use of defined vocabulary so that ventilator settings, displays, records, labels, and communications mean the same thing across devices and documents. Because this is a terminology standard, the practical requirement is to use the document’s exact concepts and relationships, not local or proprietary wording.

Use the defined value types correctly

Clause 3.1.2 distinguishes set, measured, preset, and actual value. In practice, that separates what the user commands, what a device measures, what a stored configuration changes, and what exists physically.

This matters when a ventilator displays a parameter, when a clinician documents a setting, or when software exchanges data. It prevents confusion between a target, a sensor reading, and the underlying condition.

Use the ventilator and circuit terms precisely

Clause 3.1.3 defines ventilator, airway, airway device, ventilator breathing system (VBS), patient-connection port, gas output port, gas return port, and exhaust port. These terms tell users what part of the system they are referring to.

That distinction matters in labelling, design reviews, troubleshooting, and interoperability work because the standard separates the machine, the breathing system, and the patient interface. It also helps identify accessories that can affect basic safety or essential performance.

Describe ventilation mode as a combination of pattern and inflation-type

The document’s central model is in clauses 3.3 and 3.13. A ventilation-mode is built from a ventilation-pattern and an inflation-type. The pattern determines when inflations happen and how the ventilator responds to patient-trigger events, while the inflation-type determines how pressure or flow is regulated during an inflation.

In practice, this reduces ambiguity in mode names and makes modes easier to teach, compare, and document across ventilators. The document also explains that some terms are mainly for technical documents and informatics, while others are more useful for labelling and instructions.

Use the baseline pressure and PEEP terms in a controlled way

Clause 3.12 defines baseline airway pressure (BAP), PEEP, total PEEP, and auto-PEEP, and Annex B explains how they relate. The document uses these terms to separate the intended baseline pressure from the pressure remaining in the lungs at the end of expiration.

This matters in practice because a difference between BAP and measured end-expiratory pressure can point to too short an expiratory time, airway obstruction, or leakage. Annex B shows how expiratory control, flow limitation, and incomplete emptying change what the user sees and measures.

Keep initiation and termination terms separate

Clause 3.1.2.6 uses initiate as the general term for starting a process, while clause 3.11 covers trigger, termination, and synchronisation terms. The document deliberately avoids using trigger as a catch-all word.

That matters because a ventilator can start an inflation by a patient-trigger event, a timed signal, a manual input, a remote device, or another defined condition. Clear wording is important when setting modes such as CMV, IMV, SIMV, A/C, CSV, or bi-level AV.

Treat alarms and safety limits as separate concepts from control settings

Clause 3.1.2.5 defines limit as a point beyond which a parameter cannot pass without action, and clause 3.15 covers safety limits and alarm terminology. In the document, these terms are used for patient protection that is independent of normal control settings.

For users, this means alarm limits and high-pressure or pressure-relief terms must be interpreted separately from the selected ventilation strategy. That separation matters in safety documentation, risk management, and service procedures.

Use the respiratory mechanics terms with the measurement method

Clause 3.1.4 defines airway resistance, respiratory system compliance, pulmonary compliance, driving pressure, static compliance, dynamic compliance, airway occlusion pressure, NIF/MIP, and leak terms. The document also notes that some measured values need the method of measurement attached to them.

This is important because the same patient can produce different compliance values depending on whether the value is static or dynamic, and whether the reference is the respiratory system or the lungs alone. The terminology helps users interpret measurements correctly in testing, monitoring, and research.

What terms does oSIST prEN ISO 19223-1:2026 define?

  • Ventilator (lung ventilator) - a medical device or medical electrical equipment intended to provide artificial ventilation.
  • Ventilator breathing system (VBS) - the gas pathway to and from the patient at respiratory pressures, bounded by the gas inlet, patient-connection port, and gas exhaust port.
  • Baseline airway pressure (BAP) - the baseline pressure level used as the reference for pressure terms in this document.
  • Positive end-expiratory pressure (PEEP) - the pressure remaining at end expiration, which is discussed with total PEEP and auto-PEEP in Annex B.
  • Ventilation-pattern - the part of a ventilation-mode that determines when inflations occur and how the ventilator responds to patient-trigger events.
  • Inflation-type - the part of a ventilation-mode that determines how pressure or flow is regulated during an inflation.
  • Assured inflation - an inflation selected for programmed delivery in a predetermined pattern.
  • Auto-PEEP - the portion of end-expiratory pressure above the intended baseline that remains because the lungs have not fully emptied.

Who uses oSIST prEN ISO 19223-1:2026?

The document is used by ventilator manufacturers, standards writers, clinicians, health informatics teams, and quality or regulatory staff. It is also relevant for buyers and technical evaluators who need to understand what a ventilator mode, label, or data field actually means.

In practice, users apply it when:

  • naming ventilator modes and inflation types
  • writing labels, instructions for use, and accompanying information
  • mapping ventilator data into health informatics systems and electronic health records
  • designing interoperability and communication profiles
  • reporting incidents, adverse events, or forensic findings
  • training staff to move between ventilators without learning conflicting local terms

Which standards are used with oSIST prEN ISO 19223-1:2026?

Clause 2 has no normative references, but the document names several related standards and source works that support its terminology and presentation.

StandardWhat it contributes
ISO 19223-2:2025Terminology for high-frequency and jet ventilation
ISO 19223-3:2025Terminology for breathing-therapy equipment
ISO 704:2022Principles and methods for terminology work, used for the structure of the document
ISO 10241-1:2011General requirements and examples for terminological entries in standards
IEC 60050-880Source for several general medical-device and healthcare terminology entries
ISO/IEEE 11073-10101:2004+Amd1:2017Health informatics nomenclature for point-of-care medical device communication
EN ISO 11073-10101:2020/prA1Additional nomenclature definitions for device interoperability
EN ISO 80601-2-12:2023Critical care ventilator safety and essential performance context

What does the oSIST prEN ISO 19223-1:2026 document contain?

The document contains a large, structured set of definitions arranged by topic, plus an alphabetized index in Annex E. Many definitions include notes that explain how the terms are used in practice and how they relate to other terms.

Annex A provides figures of ventilation terms and waveform examples, including mode and phase illustrations. Annex B explains baseline airway pressure, PEEP, total PEEP, auto-PEEP, and expiratory control concepts using explanatory waveforms.

Annex C provides systematic coding for health informatics and medical device communications, including classifications of inflation-types and ventilation-modes. Annex D shows how the terminology can be implemented in ventilator labelling and instructions, including equivalence tables for manufacturer wording and standardized designation.

The document also includes a formula for static compliance and several tables and figures that support mode classification, coded naming, and interpretation of breath and pressure patterns.

Relations

Effective Date
04-Jun-2025

Buy Documents

Draft

oSIST prEN ISO 19223-1:2026 - BARVE

English language (141 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

oSIST prEN ISO 19223-1:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Lung ventilators and related equipment - Vocabulary and semantics - Part 1: Lung ventilators (ISO/DIS 19223-1:2026)". This standard covers: This document establishes a vocabulary of terms and semantics for all fields of respiratory care involving mechanical ventilation, such as intensive-care ventilation, anaesthesia ventilation, emergency and transport ventilation and home-care ventilation, including sleep-apnoea breathing-therapy equipment. It is applicable — in lung ventilator and breathing-therapy device standards, — in health informatics standards, — for labelling on medical electrical equipment and medical electrical systems, — in medical electrical equipment and medical electrical system instructions for use and accompanying documents, — for medical electrical equipment and medical electrical systems interoperability, and — in electronic health records. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator and ventilator breathing system. NOTE This document can also be used for other applications relating to lung ventilation, including non-electrical devices and equipment, research, description of critical events, forensic analysis and adverse event (vigilance) reporting systems. This document does not specify terms specific to breathing-therapy equipment, or to physiologic closed-loop ventilation, high-frequency ventilation or negative-pressure ventilation; nor to respiratory support using liquid ventilation or extra-corporeal gas exchange, or oxygen, except where it has been considered necessary to establish boundaries between bordering concepts.

This document establishes a vocabulary of terms and semantics for all fields of respiratory care involving mechanical ventilation, such as intensive-care ventilation, anaesthesia ventilation, emergency and transport ventilation and home-care ventilation, including sleep-apnoea breathing-therapy equipment. It is applicable — in lung ventilator and breathing-therapy device standards, — in health informatics standards, — for labelling on medical electrical equipment and medical electrical systems, — in medical electrical equipment and medical electrical system instructions for use and accompanying documents, — for medical electrical equipment and medical electrical systems interoperability, and — in electronic health records. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator and ventilator breathing system. NOTE This document can also be used for other applications relating to lung ventilation, including non-electrical devices and equipment, research, description of critical events, forensic analysis and adverse event (vigilance) reporting systems. This document does not specify terms specific to breathing-therapy equipment, or to physiologic closed-loop ventilation, high-frequency ventilation or negative-pressure ventilation; nor to respiratory support using liquid ventilation or extra-corporeal gas exchange, or oxygen, except where it has been considered necessary to establish boundaries between bordering concepts.

oSIST prEN ISO 19223-1:2026 is classified under the following ICS (International Classification for Standards) categories: 01.040.11 - Health care technology (Vocabularies); 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

oSIST prEN ISO 19223-1:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 19223:2021. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

oSIST prEN ISO 19223-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms,definitions, symbols, and abbreviated terms
  • 3.1 General artificial-ventilation terminology
  • 3.1.1.1 normal use
  • 3.1.1.2 intended use
  • 3.1.1.3 normal condition
  • 3.1.1.4 single fault condition
  • 3.1.1.5 protection device
  • 3.1.1.6 accompanying information
  • 3.1.1.7 port
  • 3.1.2.1 set
  • 3.1.2.2 measured
  • 3.1.2.3 preset
  • 3.1.2.4 actual value
  • 3.1.2.5 limit
  • 3.1.2.6 initiate
  • 3.1.2.7 mandatory
  • 3.1.2.8 rise time
  • 3.1.3.1 ventilator
  • 3.1.3.2 airway
  • 3.1.3.3 airway device
  • 3.1.3.4 VBS
  • 3.1.3.5 sleep-apnoea breathing-therapy equipment
  • 3.1.3.6 exhaust port
  • 3.1.3.7 gas output port
  • 3.1.3.8 gas return port
  • 3.1.3.9 patient-connection port
  • 3.1.3.10 fraction of inspired oxygen
  • 3.1.3.11 heliox
  • 3.1.4.1 lung
  • 3.1.4.2 respiratory system
  • 3.1.4.3 airway resistance
  • 3.1.4.4 respiratory compliance
  • 3.1.4.5 pulmonary compliance
  • 3.1.4.6 driving pressure
  • 3.1.4.7 static compliance
  • 3.1.4.8 dynamic compliance
  • 3.1.4.9 airway occlusion pressure
  • 0.1 DEPRECATED: P
  • 3.1.4.10 negative inspiratory force
  • 3.1.4.11 airway leak
  • 3.1.4.12 ventilator breathing system leak
  • 3.1.4.13 elastance
  • 3.1.5.1 ventilation
  • 3.1.5.2 artificial ventilation
  • 3.1.5.3 mechanical ventilation
  • 3.1.5.4 automatic ventilation
  • 3.1.5.5 PPV
  • 3.1.5.6 negative-pressure ventilation
  • 3.1.5.7 non-invasive ventilation
  • 3.1.5.8 high-frequency ventilation
  • 3.1.5.9 jet ventilation
  • 3.2 Breath terminology
  • 3.2.1 breath
  • 3.2.2 breathe
  • 3.2.3 spontaneous breath
  • 3.2.4 natural breathing
  • 3.2.5 unrestricted breathing
  • 3.2.6 respiratory activity
  • 3.2.7 inspiratory effort
  • 3.2.8 expiratory effort
  • 3.2.9 inspiration
  • 3.2.10 exhalation
  • 3.2.11 unassisted breath
  • 3.2.12 supported breath
  • 3.2.13 assisted breath
  • 3.2.14 synchronized breath
  • 3.2.15 controlled breath
  • 3.2.16 additional breath
  • 3.2.17 concurrent breath
  • 3.2.18 work of breathing
  • 3.2.19 imposed work of breathing
  • 3.2.20 ventilator work of breathing
  • 3.2.21 mechanical power
  • 3.3 Ventilation control paradigms
  • 3.3.1 flow-regulation
  • 3.3.2 pressure-regulation
  • 3.3.3 demand flow

Introduction

The characteristics of ventilation-modes of current automatic lung ventilators are complex and often not well understood. Some ventilator manufacturers have created proprietary terms to describe ways of ventilating patients, and other manufacturers have used existing terms with different meanings in different situations. This has led to patient safety hazards, an example being that lung ventilator clinical orders (settings) for one model of ventilator can be quite different from those required to get the same result from a different ventilator. International standards are necessary to prevent inconsistencies that may result in risks to patients. This document provides standardized terminology and semantics for artificial ventilation. Where possible, terms already in general use have been incorporated, while clearly defining each term and limiting its potential for misuse. New terms and deprecation of previous terms have been necessary to ensure a clear understanding of the interactions between the patient and the ventilator. This document has been developed with the participation, cooperation and assistance of clinicians, members of other standards development organizations, and major international ventilator manufacturers. The applications include lung ventilators, anaesthesia and respiratory therapy equipment, health informatics systems facilitating clinical care, research, interoperability, incident reporting and equipment maintenance. Ventilators have become increasingly interactive with the patient, and terminology must take this into account.

The terminology in this document is defined to facilitate both the setting of a ventilator and how to describe and record the resultant ventilator-patient interactions, continuously and at defined points within the course of ventilation. This includes the result of the complex interactions that occur when additional breaths are taken during an assured-inflation cycle, as can occur, for example, during APRV (airway pressure release ventilation).

Ventilation-modes of modern ventilators require the selection of both the pattern that determines when and what inflations occur, and the method used to inflate the lungs; functions that can be conveniently referred to as the ventilation-pattern and the inflation-type. In most cases the choice of inflation-type is largely independent of the choice of ventilation-pattern.

In this document, the ventilation-mode is a composite of these two independent, user-selectable elements:

the ventilation-pattern determines how the ventilator will respond to patient-trigger events and what it will cause to be delivered, when, irrespective of the patient’s respiratory activity and the inflation-type which determines how the pressure or flow at the patient-connection port will be regulated during an inflation, once initiated.

This format, with its focus on the pattern to which inflations are initiated and on the type of inflation that is delivered, is independent of which ventilation-pattern and inflation-type combination can be used with which setting, and for which clinical intention. Its structure not only reduces the number of items to be remembered but also makes them much easier to teach, learn, remember and recognize, thereby improving ventilator usability. The overall objective is to encourage a consistent use of ventilator vocabulary so that users trained in the application of this document will be able to move easily from one ventilator to another and operate each one, with confidence, after a minimum amount of training. Some of the terms in this document are principally intended for technical documents, informatics and related applications (see Annex C) and might have little applicability to ventilator labelling and instructions for use (see Annex D).

The notes to some definitions refer to post-coordinated terms. These are terms formed by the combination of two defined terms, or of a defined term with a natural language word, to form a new compound term. Usually, the additional term or word qualifies the base term by reference to an alternative site, pressure level or point of occurrence within a respiratory cycle. It can place a restriction on the applicability of the base term.

Examples of the application of this document are illustrated in the figures of Annex A and Annex B but these are not intended to indicate a requirement, nor to impose any restriction on the design of artificial

oSIST prEN ISO 19223-1:2026 ISO/DIS 19223-1:2026(en) ventilation devices. Colour coding is employed in most of these figures of this document to help distinguish between some of the specific characteristics being illustrated. [1] NOTE The following figures have been adapted from Reference with permission:

  • Figures: Figure A.1 to Figure A.35 and Figure B.1 to Figure B.7.

oSIST prEN ISO 19223-1:2026 DRAFT International Standard ISO/DIS 19223-1:2026(en) Lung ventilators and related equipment — Vocabulary and semantics — Part 1:

Lung ventilators

1 Scope

This document defines terms for fields of respiratory care involving artificial ventilation, such as intensivecare ventilation, anaesthesia ventilation, emergency and transport ventilation and home-care ventilation. This document is applicable:

  • in lung ventilator and breathing-therapy equipment device standards,
  • in health informatics standards,
  • for labelling on medical electrical equipment and medical electrical systems,
  • in medical electrical equipment and medical electrical system instructions for use and accompanying information,
  • for medical electrical equipment and medical electrical systems interoperability, and
  • in electronic health records.

This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator and ventilator breathing system. NOTE This document can also be used for other applications relating to lung ventilation, including non-electrical devices and equipment, research, description of critical events, forensic analysis and adverse event (vigilance) reporting systems.

This document does not specify terms specific to high-frequency ventilation or jet ventilation, which [2] are defined in ISO 19223-2:2025 , terms specific to breathing-therapy equipment, which are defined [3] in ISO 19223-3:2025 , or to physiologic closed-loop ventilation, negative-pressure ventilation; nor to respiratory support using liquid ventilation or extra-corporeal gas exchange, or oxygen, except where it has been considered necessary to establish boundaries between bordering concepts. [4] This document is structured based on the guidance and requirements specified in ISO 704:2022 and [5] ISO 10241-1:2011 .

2 Normative references

There are no normative references in this document.

3 Terms,definitions, symbols, and abbreviated terms

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp

oSIST prEN ISO 19223-1:2026 ISO/DIS 19223-1:2026(en) - IEC Electropedia: available at http:// www .electropedia .org/ NOTE For convenience, an index and a list of sources of all defined terms used in this document are provided in Annex E.

3.1 General artificial-ventilation terminology

3.1.1 Referenced terms

3.1.1.1 normal use

operation, including standby, routine inspection and adjustments by any user, in accordance with the accompanying information (3.1.1.6) or, for those medical devices provided without accompanying information (3.1.1.6), with generally accepted practice Note 1 to entry: Normal use should not be confused with intended use (3.1.1.2). While both include the concept of use as intended by the manufacturer, intended use (3.1.1.2) focuses on the medical purpose while normal use incorporates not only the medical purpose, but also matters such as maintenance, transport. [6] [SOURCE: IEC 60050-880 , 880-21-07]

3.1.1.2 intended use

intended purpose use for which a product or process is intended according to the specifications, instructions for use and other accompanying information (3.1.1.6) [6] [SOURCE: IEC 60050-880 , 880-21-06] Note 1 to entry: Intended use should not be confused with normal use (3.1.1.1). While both include the concept of use as intended by the manufacturer, intended use focuses on the medical purpose while normal use (3.1.1.1) incorporates not only the medical purpose, but also such matters as maintenance, service and transport.

3.1.1.3 normal condition

condition of a medical device or accessory in which all risk control measures are intact [6] [SOURCE: IEC 60050-880 , 880-18-11]

3.1.1.4 single fault condition

condition of a medical device or accessory in which a single risk control measure is defective [6] [SOURCE: IEC 60050-880 , 880-18-25.]

3.1.1.5 protection device

part or function of a medical device that, without intervention by the user, protects against a hazard or a hazardous output [6] [SOURCE: IEC 60050-880 , 880-16-08.]

3.1.1.6 accompanying information

DEPRECATED: accompanying document information supplied by the manufacturer with or marked on a medical device or accessory for the user or the responsible organization, particularly regarding safe use [6] [SOURCE: IEC 60050-880 , 880-11-01.]

oSIST prEN ISO 19223-1:2026 ISO/DIS 19223-1:2026(en)

3.1.1.7 port

fluid port opening for the passage of a fluid through a specified interface Note 1 to entry: Typical interfaces where ports occur are:

  • where breathing gas enters a ventilator,
  • where water for humidification enters a humidifier,
  • where operator-detachable tubing is connected to a medical device, and
  • where a VBS (3.1.3.4) is connected to the patient or to an airway device (3.1.3.3). Note 2 to entry: A port can be in the form of a specific connector or be designed to not allow connection with any connector.

[6] [SOURCE: IEC 60050-880 , 880-05-02. Modified - added preferred term port]

3.1.2 Qualifiers defined in this document

3.1.2.1 set

setting allocated a specific value EXAMPLE 1 A set pressure limit (3.1.2.5).

EXAMPLE 2 The set Δ inspiratory pressure (3.8.6).

Note 1 to entry: Set is used in this document as a prefix to distinguish an intended value for a controlled artificial ventilation (3.1.5.2) variable from a measured (3.1.2.2) or actual value (3.1.2.4) of the same quantity, if this is not evident from the context of use.

Note 2 to entry: A set value might be determined directly by the user or indirectly by selection of an algorithm that determines the setting based on other set or measured (3.1.2.2) values. Note 3 to entry: See also the references to settings in Annex A and Annex C.

3.1.2.2 measured

determined by a measuring device or system Note 1 to entry: This term is used in this document as a prefix to distinguish the value of a quantity as determined by a measuring device or system, from an actual value (3.1.2.4) or set (3.1.2.1) value of the same quantity, if this is not evident from the context of use.

Note 2 to entry: Measured values might be displayed or recorded as discrete values or as a continuous waveform.

3.1.2.3 preset

one of a set of stored configuration parameter(s), including selection of algorithms and initial values for use by algorithms, which affects or modifies the performance of the ventilator (3.1.3.1) Note 1 to entry: Presets are commonly configured by the manufacturer or a responsible organization. Note 2 to entry: Access to a preset value is typically controlled by - a tool, - a responsible-organization password and a technical description, separate from the instructions for use, - an individual user password,

oSIST prEN ISO 19223-1:2026 ISO/DIS 19223-1:2026(en) - voice recognition, or - biometric means.

3.1.2.4 actual value

value of a quantity as it exists in fact Note 1 to entry: This is the true value of a quantity, which might, or might not, be determinable by a measuring device. Note 2 to entry: The definitions of terms denoting quantities, in this document, denote the actual value of that quantity. Set (3.1.2.1) values are the means by which an user informs the ventilator (3.1.3.1) of the intended actual value, and measured (3.1.2.2) values are displays or records of the actual value, to the accuracy and resolution of the measuring system.

3.1.2.5 limit

point or level beyond which the value of a parameter cannot pass without an action by the ventilator (3.1.3.1) Note 1 to entry: The action might be a notification or the implementation of means to prevent or mitigate a hazardous situation.

Note 2 to entry: In this document, this term is restricted to the designation of safety constraints that provide patient protection that is completely independent of the controlled ventilation (3.1.5.1) parameters. Note 3 to entry: An alarm system uses an alarm limit in determining an alarm condition. Note 4 to entry: See also Annex E and safety limits and alarm terminology (3.15).

3.1.2.6 initiate

cause a process or action to begin Note 1 to entry: This word has been adopted in this document as the general term to designate the concept of causing a process or action to begin. This is to counter a tendency to use the word trigger (3.11.1) for this purpose; a trend that removes the ability of that term to differentiate its own special meaning from that of a simple timed switching action. In this document, an inflation (3.4.1.1) can be initiated by, for example:

  • a patient-trigger event (3.11.5);
  • a timed signal;
  • a manual input;
  • a conditional termination (3.11.12) of an expiratory phase (3.5.6);
  • a signal from a remote device.

Note 2 to entry: In this document, an inflation (3.4.1.1) that is initiated by a timed signal may be referred to as being ventilator-initiation (3.11.9).

Note 3 to entry: A conditional termination (3.11.12) of an expiratory phase (3.5.6) is typically achieved by establishing an expiratory flow (3.9.5) threshold at which the next inflation (3.4.1.1) is initiated. An initiation of this type is sometimes used, for example, to optimize the expiratory time (3.6.6) in an APRV (3.13.3.9) mode. Such a threshold flow is typically expressed as either a set (3.1.2.1) expiratory flow (3.9.5) or a set (3.1.2.1) percentage of the peak expiratory flow (3.9.5).

oSIST prEN ISO 19223-1:2026 ISO/DIS 19223-1:2026(en)

3.1.2.7 mandatory

required to occur Note 1 to entry: This term is defined as used in this document, with a specific meaning of the word mandatory, which in natural language has a spectrum of meanings. It has become firmly established in the vocabulary of artificial ventilation (3.1.5.2) but, because of its ambiguity, it can denote either ‘total control’ or, as in this definition, ‘required to occur’. In early artificial ventilation (3.1.5.2) practice, it was usual for all aspects of the patient's ventilation (3.1.5.1) to be taken over by the ventilator (3.1.3.1) and so every breath (3.2.1) could be described as mandatory in its broadest sense. Since then, with the introduction of breath synchronization (3.11.6) and the concept of support for spontaneous breaths (3.2.3), only a small percentage of patients currently have their ventilation (3.1.5.1) totally controlled. However, there remains a mandatory component to all forms of artificial ventilation (3.1.5.2) but a key aspect for an user when setting a contemporary ventilator (3.1.3.1) is being assured that, when a selected inflation-type (3.4.1.2) is delivered within the selected ventilation-pattern (3.13.1.3), it will provide at least a minimum level of assistance and, in the case of apnoea, that the ventilation (3.1.5.1) will be totally controlled. These developments have led manufacturers to increasingly restrict the use of the term ‘mandatory’ to the context of ventilation (3.1.5.1) that is assured to occur by the programmed delivery of a selected inflation-type (3.4.1.2), in predetermined patterns, at a rate that is independent of the patient's respiratory activity (3.2.6). This is the only sense in which the term mandatory is used in this document; which is mainly in the explanation of the classical mode names such as continuous mandatory ventilation (3.13.3.2), intermittent mandatory ventilation (3.13.3.4) and synchronized intermittent mandatory ventilation (3.13.3.5). For other purposes, wherever possible, the word assured is used in its place.

3.1.2.8 rise time

indication of the time for the regulated parameter to rise to a set (3.1.2.1) value following the initiation (3.1.2.6) of an inflation (3.4.1.1) Note 1 to entry: The rise time is often expressed as the slope of a ramp or as the time-constant of the rise although neither of these terms depicts the actual typical trajectory of this pressure rise precisely. Note 2 to entry: For pressure-regulation (3.3.2), this is the time to reach a set (3.1.2.1) inspiratory pressure (3.8.2), for flow-regulation (3.3.1) it is the time to reach a set (3.1.2.1) inspiratory flow (3.9.2). Note 3 to entry: See also Figure A.1 and Figure A.9 to Figure A.13. 3.1.2.9 Δ delta difference between two quantities Note 1 to entry: This symbol is used in this document as a prefix to denote that the qualified parameter is referenced to another parameter. In particular, it is used in this document to qualify an inspiratory pressure (3.8.2) or expiratory pressure (3.8.8) to denote when it is referenced to a baseline airway pressure (3.12.1) level instead of to the default, ambient pressure level.

Note 2 to entry: In verbal communication, Δ, as used in this document, is expressed as either ‘delta’ or ‘differential’. Note 3 to entry: See also Figure A.1, Figure A.9, Figure A.10, Figure A.29, Figure A.30 and Figure A.33.

3.1.3 Ventilation equipment terms

3.1.3.1 ventilator

lung ventilator DEPRECATED: respirator medical device or medical electrical equipment intended to provide artificial ventilation (3.1.5.2) Note 1 to entry: In cases of possible ambiguity the full term, lung ventilator, should be used.

oSIST prEN ISO 19223-1:2026 ISO/DIS 19223-1:2026(en)

3.1.3.2 airway

patient airway connected, gas-containing cavities and passages within the respiratory system (3.1.4.2), that conduct gas between the alveoli and the oral and nasal orifices on the surface of the face, or the patient-connection port (3.1.3.9) if an airway device (3.1.3.3) is used Note 1 to entry: This is a well-established term that is commonly used in isolation in references to the airway of a patient. Depending on the context, it is sometimes more helpful to use the admitted term patient airway.

3.1.3.3 airway device

device intended for use as an interface between the patient-connection port (3.1.3.9) of a ventilator (3.1.3.1) and the patient's airway (3.1.3.2), and which has no auxiliary features on which the ventilator (3.1.3.1) is dependent for its normal operation EXAMPLE Endotracheal tube; tracheotomy tube; face mask; supralaryngeal airway. Note 1 to entry: The connection to the patient's airway (3.1.3.2) can be at the face (non-invasive) or internal to the patient (invasive).

Note 2 to entry: A face mask that intentionally vents respiratory gas to atmosphere by means of a bleed orifice is a functional part of the ventilator breathing system (3.1.3.4) and therefore not an airway device. With that arrangement, the face seal of the mask becomes the patient-connection port (3.1.3.9) and there is no patient-connection port (3.1.3.9) connector, nor an airway device.

...