Abstract

This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971[2].
This document does not apply to MLMD employing large language models (LLM) or generative AI.

Status
Published
Public Enquiry End Date
29-Apr-2026
Publication Date
13-Jul-2026
Technical Committee
VAZ - Healthcare
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
09-Jul-2026
Due Date
13-Sep-2026
Completion Date
14-Jul-2026

Buy Documents

Technical specification

SIST-TS CEN ISO/TS 24971-2:2026

English language (41 pages)
Preview
Preview
e-Library read for
×1 day

SIST-TS CEN ISO/TS 24971-2:2026 is the Slovenian adoption of CEN ISO/TS 24971-2:2026, a European Technical Specification from CEN and ISO on applying ISO 14971 to machine learning in artificial intelligence for medical devices. It gives guidance on AI and ML risks and on using the ISO 14971 risk management process for ML-enabled medical devices (MLMD). The document is meant to be used with ISO 14971 and ISO/TR 24971, and it excludes MLMD using large language models (LLM) or generative AI.

What does SIST-TS CEN ISO/TS 24971-2:2026 specify?

SIST-TS CEN ISO/TS 24971-2:2026 specifies guidance for identifying, assessing, controlling and monitoring risks that are specific to AI and ML in MLMD. The document is organized as a risk-management companion to ISO 14971, with core clauses on the risk management system, risk analysis, evaluation, control, residual risk, review, and post-production activities.

It also includes four informative annexes that focus on bias, example hazards, safety-related characteristics, and autonomy.

AnnexWhat it covers
Annex AExplanation of bias and examples of how bias can affect MLMD
Annex BExamples of hazards, hazardous situations, events, circumstances and risk control measures for MLMD
Annex CQuestions for identifying hazards and characteristics related to safety in use, design and post-production
Annex DConsiderations for MLMD that have a level of autonomy

What are the key requirements of SIST-TS CEN ISO/TS 24971-2:2026?

SIST-TS CEN ISO/TS 24971-2:2026 keeps the ISO 14971 risk-management structure, but it applies it to the specific realities of MLMD. In practice, that means the manufacturer has to look closely at data, model behavior, user reliance, update logic, transparency and post-market monitoring, not just the device output.

Clause 4 - build the risk management system around MLMD realities

Clause 4 says the ISO 14971 risk management process applies to MLMD, including identification of hazards, estimation and evaluation of risks, risk control, overall residual risk, and production and post-production monitoring. For MLMD, the risk management team should include people who understand machine learning, IT, software validation, usability engineering, clinical workflow and good data management.

That matters because MLMD risk often depends on how training data and test data were chosen, how the model behaves in practice, and how users interpret the output. The risk management plan should also address performance monitoring, updating or retraining, criteria for initiating those actions, and the possibility of returning to a previous version.

Clause 5 - treat data, misuse and bias as part of risk analysis

Clause 5 emphasizes that intended use and reasonably foreseeable misuse for MLMD include use errors related to lack of transparency, use outside the intended patient population, inadequate input data and overreliance on the output. Annex C adds questions that help the manufacturer examine how the device is used, what data it receives, how it connects to other devices or software, and whether it has a level of autonomy.

The document also points to training data and test data quality, representativeness and separation. In practice, that means the manufacturer should check whether the data reflect the intended patient population and clinical setting, because poor coverage or mismatch can create bias, underfitting, overfitting or misleading performance results.

Clause 5.5 and Clause 6 - estimate and evaluate risk in a way that fits MLMD

Clause 5.5 recognizes that, for some ML-related risks, the probability of harm may be hard to estimate. In those cases, risk estimation may need to rely on the severity of possible harm, while also considering hardware, software and IT reliability, usability and the quality of the training and test data.

Clause 6 keeps the ISO 14971 requirement to decide whether each risk is acceptable. The practical point is simple: MLMD does not get a separate risk threshold, but the evidence used to judge acceptability may have to come from data quality checks, usability work and performance testing rather than from classical design assumptions alone.

Clause 7 - control risk in the right order and verify the controls

Clause 7 follows the ISO 14971 priority order for risk control: inherently safe design, protective measures in the device or process, and information for safety. For MLMD, that includes algorithm and model choice, data quality assurance, sequestering training data from test data, realistic user interfaces, access control, human oversight, alarms, and instructions that explain the output and its limits.

Clause 7.2 is especially important because it links risk control to testing. The document says testing the ML model is needed to confirm that acceptance criteria are met, synthetic data may be used in simulated conditions, and rollback to a previous version should be verified by simulation and review.

Clauses 8 to 10 - judge overall residual risk and keep monitoring after release

Clause 8 asks the manufacturer to evaluate overall residual risk against the benefits of intended use, with extra attention to complexity, silent failures, autonomy, user interface issues, overtrust and novelty of the technology. Clause 8.2 also requires disclosure of significant residual risks, including performance limitations, recommended checks before use, and differences in performance across patient groups.

Clauses 9 and 10 move the focus into production and post-production. The manufacturer should review the risk management plan before distribution, actively collect safety-related information, monitor drift and performance changes, and decide when retraining, updating or rollback is needed. These decisions and their rationale belong in the risk management file.

What terms does SIST-TS CEN ISO/TS 24971-2:2026 define?

  • Bias - A systematic difference in treatment of certain objects, people or groups compared with others.
  • Explainability - The ability of a system to express the main factors behind its results in a way humans can understand.
  • Machine learning (ML) - A system function that learns from input data instead of following only fixed instructions.
  • Machine learning algorithm (ML algorithm) - An algorithm that determines machine learning model parameters from data according to given criteria.
  • Machine learning model (ML model) - The logical representation that produces predictions from input data after training.
  • ML-enabled medical device (MLMD) - A medical device that uses machine learning.
  • Test data - Data used to assess the performance of the ML model and kept separate from training data.
  • Training data - Data used to train the ML model.

Who uses SIST-TS CEN ISO/TS 24971-2:2026?

SIST-TS CEN ISO/TS 24971-2:2026 is used by medical device manufacturers, ML developers, risk managers, quality managers, regulatory and clinical teams, and usability and software specialists working on MLMD. It is relevant to organizations developing diagnostics, decision support, image analysis, treatment planning, monitoring systems and other medical devices that use ML.

It is also useful for buyers and reviewers who need to understand whether an MLMD risk-management file addresses data quality, bias, explainability, autonomy, update control and post-market monitoring in a way that fits ISO 14971.

Which standards are used with SIST-TS CEN ISO/TS 24971-2:2026?

  • ISO 14971:2019 - The core medical device risk management standard; this document is built on its process.
  • ISO/TR 24971:2020 - The companion guidance for applying ISO 14971; this document is intended to be used with it.
  • IEC 62366-1 - Supports usability engineering, especially where transparency, understandability and overtrust affect safe use.
  • IEC 62304 - Supports software life cycle processes for software-based MLMD controls and verification.
  • IEC/TR 60601-4-1 - Provides guidance on medical electrical equipment with a degree of autonomy.
  • IEC/TR 80002-1 - Gives guidance on applying ISO 14971 to medical device software.
  • IEC 80001-1 - Addresses risk management for IT networks incorporating medical devices.
  • ISO/IEC 22989, ISO/IEC 23894 and ISO/IEC TR 24027 - Provide AI concepts, AI risk management guidance and bias terminology that support the technical background of the document.

What does the SIST-TS CEN ISO/TS 24971-2:2026 document contain?

The document contains a structured set of clauses that walk the reader through risk management for MLMD, from planning and analysis to post-production action. It also contains figures that explain the relationship between concept development, training and testing, and how underfitting, good fit and overfitting differ.

Annex A explains bias and gives examples of how bias can arise in data, model development and device performance. Annex B gives example hazards, hazardous situations and potential controls, including data quality, data management, cybersecurity, usability and functionality. Annex C provides question sets for use, design, training, testing and post-production review. Annex D explains autonomy concepts and gives ways to think about levels of autonomy and human oversight in MLMD.

Buy Documents

Technical specification

SIST-TS CEN ISO/TS 24971-2:2026

English language (41 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

SIST-TS CEN ISO/TS 24971-2:2026 is a technical specification published by the Slovenian Institute for Standardization (SIST). Its full title is "Medical devices - Guidance on the application of ISO 14971 - Part 2: Machine learning in artificial intelligence (ISO/TS 24971-2:2026)". This standard covers: This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971[2]. This document does not apply to MLMD employing large language models (LLM) or generative AI.

This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971[2]. This document does not apply to MLMD employing large language models (LLM) or generative AI.

SIST-TS CEN ISO/TS 24971-2:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.01 - Medical equipment in general; 35.240.01 - Application of information technology in general; 35.240.80 - IT applications in health care technology. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST-TS CEN ISO/TS 24971-2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 bias
  • 3.2 explainability
  • 3.3 machine learning
  • 3.4 machine learning algorithm
  • 3.5 machine learning model
  • 3.6 ML-enabled medical device
  • 3.7 overfitting
  • 3.8 test data
  • 3.9 training
  • 3.10 training data
  • 3.11 transparency
  • 3.12 underfitting
  • 4 General requirements for risk management system
  • 4.1 Risk management process
  • 4.2 Management responsibilities
  • 4.3 Competence of personnel
  • 4.4 Risk management plan
  • 4.5 Risk management file
  • 5 Risk analysis
  • 5.1 Risk analysis process
  • 5.2 Intended use and reasonably foreseeable misuse
  • 5.3 Identification of characteristics related to safety
  • 5.4 Identification of hazards and hazardous situations
  • 5.5 Risk estimation
  • 6 Risk evaluation
  • 7 Risk control
  • 7.1 Risk control option analysis
  • 7.2 Implementation of risk control measures
  • 7.3 Residual risk evaluation and subsequent steps
  • 8 Evaluation of overall residual risk
  • 8.1 General considerations for MLMD
  • 8.2 Disclosure of significant residual risks
  • 9 Risk management review
  • 10 Production and post-production activities
  • 10.1 General
  • 10.2 Information collection
  • 10.3 Information review
  • 10.4 Actions
  • Annex A
  • Annex B
  • 1 Data selection Inadequate control of data collection process
  • 2 Data annotation Inconsistent annotation performance
  • 3 Data augmentation Poorly designed or excessive augmentation
  • 4 Data shift Data entry errors, data dropout
  • 5 Usability User unable to intervene in an autonomous process (see IEC/TR 60601-4-1 )
  • 6 Update process Insufficient control of retraining process
  • 8 ML model training Insufficient training data
  • 9 Functionality False positive results
  • 10 Installation Incompatibility of MLMD with software or hardware environment
  • Annex C
  • Annex D
  • Bibliography

Introduction

Artificial intelligence (AI) is rapidly advancing and offers transformative potential for the healthcare sector. These advantages can be related to improved benefits for the patient, increased efficiency in clinical workflows and more effective management of healthcare overall. However, the implementation of new technologies such as AI can also present new risks and can, for example, jeopardize patient safety, affect privacy and security, influence user actions, undermine trust in healthcare or adversely affect the management of healthcare.

[16] The safety and effectiveness of AI in medical devices was explored in an AAMI-BSI document , which identified three ways in which AI-based medical devices differed from “traditional” (non-AI) medical devices:

a) Training. These medical devices can process large amounts of data and learn from these data to improve their results. Thus, they can have positive effects on patient health within the scope of the intended use of the medical device.

b) Level of autonomy. These medical devices can generate different treatment options, select the best option [8] based on a trained model and execute the selected option (see for example IEC/TR 60601-4-1 ). These steps can be performed with reduced or even without direct user action. Therefore, human oversight is critical for their safe application.

c) Explainability. These medical devices often rely on complex algorithms and large datasets to generate output. However, the inherent opacity of these algorithms makes it challenging to interpret how specific conclusions or recommendations are derived. This aspect is sometimes referred to as the “black-box problem”. This can make it difficult for clinicians, other healthcare personnel and individuals without specialist knowledge to understand the rationale for the output. Many different AI-based technologies and algorithms exist today, which are differentiated by data topology and model architecture, including decision trees, genetic algorithms and deep learning-based technologies [4] [6] such as generative AI and neural networks. ISO/IEC 22989 and ISO/IEC 23894 provide general guidance on AI concepts, terminology and risk management, but they do not specifically address the application of AI to medical devices. It is noted that “risk” is defined in these documents as the effect of uncertainties on [3] objectives (see also ISO 31000 ). This definition is useful for organizational or business risk management. The term “risk” used in the healthcare sector is different and is defined in ISO 14971:2019 as the combination of the probability of occurrence of harm and the severity of that harm. Figure 1 — Concept development, training and testing stages of the ML model and its relationship with the ML algorithm, the training data and the test data This document focuses on machine learning (ML) techniques and is restricted to ML-enabled medical devices (MLMD). Machine learning is considered a subset of AI that involves an ML model and an ML algorithm. See Figure 1. The ML model and the ML algorithm are the results of the concept development for a new MLMD, together with acceptance criteria for the eventual MLMD. It is noted that the supporting infrastructure (computing framework, hardware, network and other IT components) can be an important aspect in concept development. The MLMD acceptance criteria are different from the criteria for risk acceptability. It

SIST-TS CEN ISO/TS 24971-2:2026 ISO/TS 24971-2:2026(en) is important to establish the MLMD acceptance criteria at the start, as part of concept development, and not at the end of the MLMD development. Otherwise, the results of MLMD testing could influence the decisions when establishing those criteria.

After concept development, the ML model is trained by using an ML algorithm enabling it to learn patterns from training data without being explicitly programmed. Next, the trained ML model is applied to test data to verify its performance. The training data and the test data are different (disjoint) sets. They can be actual patient data or synthetic data, i.e. data created to simulate a patient for training or testing purposes. The tested ML model can then be applied to new patient data in a clinical setting. More information on MLMD [21] [23] [17][18] can be found in IMDRF documents N67 and N88 and in guidance documents from FDA, Health Canada and MHRA.

It is recognized that the ML model can require retraining after a period of use to redefine its parameters and to ensure its continued performance. This can be achieved by planned retraining with collected patient data or on a continuous basis with each new patient data. The latter is referred to as “continuous learning” throughout this document and sometimes as “adaptive” in other documents. All medical devices come with inherent risks. Manufacturers are required to demonstrate that their medical devices do not pose unacceptable risks, and that the benefits of the intended use outweigh the overall residual risk. ISO 14971 details how manufacturers can identify, assess and control risks to protect the patients, the users and other persons as well as property (for example objects, data, other equipment) and the environment. This includes risks related to data and systems security and cybersecurity. Guidance on [2] [11] the application of ISO 14971 is provided in document ISO/TR 24971 . Additionally, IEC 80001-1 and [12] IEC/TR 80002-1 address software internal to a medical device that can support AI or ML. This document was developed to provide specific guidance on the application of ISO 14971 to MLMD. It does not provide a new risk management process, nor does it expand the requirements of ISO 14971. This document addresses risks related to machine learning and topics such as data management, feature extraction, unwanted bias, information security, training the ML model by an ML algorithm, evaluation and testing of the [14] trained ML model. See Annex A for an explanation of bias. The report AAMI TIR34971 provided valuable input for this document.

SIST-TS CEN ISO/TS 24971-2:2026 Technical Specification ISO/TS 24971-2:2026(en) Medical devices — Guidance on the application of ISO 14971 — Part 2:

Machine learning in artificial intelligence

1 Scope

This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This [2] document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971 . This document does not apply to MLMD employing large language models (LLM) or generative AI.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 14971:2019, Medical devices — Application of risk management to medical devices

3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 14971:2019 and the following apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

3.1 bias

systematic difference in treatment of certain objects, people, or groups in comparison to others Note 1 to entry: Treatment is any kind of action, including perception, observation, representation, prediction or decision.

[7] [SOURCE: ISO/IEC TR 24027:2021, 3.2.2 ]

3.2 explainability

property of a system to express important factors influencing the system’s results in a way that humans can understand Note 1 to entry: It is intended to answer the question “Why?” without actually attempting to argue that the course of action that was taken was necessarily optimal.

[4] [SOURCE: ISO/IEC 22989:2022, 3.5.7 , modified — “AI system” was changed to “system”.]

SIST-TS CEN ISO/TS 24971-2:2026 ISO/TS 24971-2:2026(en)

3.3 machine learning

ML function of a system that can learn from input data instead of strictly following a set of specific instructions Note 1 to entry: Machine learning focuses on prediction based on known properties learned from the input data. [13] [SOURCE: AAMI TIR66: 2017 , 3.19]

3.4 machine learning algorithm

ML algorithm algorithm to determine parameters of a machine learning model (3.5) from data according to given criteria [4] [SOURCE: ISO/IEC 22989:2022, 3.3.6 , modified — The example was removed.]

3.5 machine learning model

ML model logical representation of a system that generates predictions based on input data Note 1 to entry: A machine learning model results from training (3.9) based on a machine learning algorithm (3.4). [4] Note 2 to entry: Adapted from ISO/IEC 22989:2022, 3.1.23 and 3.3.7 .

3.6 ML-enabled medical device

MLMD medical device that utilizes machine learning (3.3) Note 1 to entry: MLMD can involve multiple ML models and multiple ML algorithms (3.4).

3.7 overfitting

creating a model that fits the training data (3.10) too precisely and fails to generalize on new data Note 1 to entry: Overfitting can occur because the trained model has learned from non-essential features in the training data (i.e. features that do not generalize to useful outputs), because of excessive noise in the training data (e.g. excessive number of outliers) or because the model is too complex for the training data. Note 2 to entry: Overfitting can be identified when there is a significant difference between errors measured on training data and on separate test data. The performance of overfitted models is especially impacted when there is a significant mismatch between training data and new data.

Note 3 to entry: See Figure 2 for a graphical comparison with underfitting (3.12). [5] [SOURCE: ISO/IEC 23053:2022, 3.1.4 , modified — In the definition, “which” was changed to “that”; in Note 1 to entry, “because of” was added; in Note 2 to entry, “separate test and validation data” was changed to “separate test data”, and “production data” was changed to “new data”; Note 3 to entry was added.]

3.8 test data

data used to assess the performance of the machine learning model (3.5) Note 1 to entry: Test data is disjoint from training data (3.10). [4] [SOURCE: ISO/IEC 22989:2022, 3.2.14 , modified — The preferred term “evaluation data” was removed; in the definition, “a final model” was changed to “the machine learning model”; “validation data” was deleted from Note 1 to entry.]

SIST-TS CEN ISO/TS 24971-2:2026 ISO/TS 24971-2:2026(en)

3.9 training

process to determine or to improve the parameters of a machine learning model (3.5), based on a machine learning algorithm (3.4), by using training data (3.10) [4] [SOURCE: ISO/IEC 22989:2022, 3.3.15 , modified — The admitted term “model training” was removed.]

3.10 training data

data used to train a machine learning model (3.5) [4] [SOURCE: ISO/IEC 22989:2022, 3.3.16 ]

3.11 transparency

property of a system that appropriate information about the system is made available to relevant stakeholders Note 1 to entry: Appropriate information for system transparency can include aspects such as features, performance, residual risks, limitations, components, procedures, measures, design goals, design choices and assumptions, data sources and labelling protocols.

Note 2 to entry: Inappropriate disclosure of some aspects of a system can violate security, privacy or confidentiality requirements.

[4] [SOURCE: ISO/IEC 22989:2022, 3.5.15 , modified — “residual risks” was added in Note 1 to entry.]

3.12 underfitting

creating a model that does not fit the training data (3.10) closely enough and produces incorrect predictions on new data Note 1 to entry: Underfitting can occur when features are poorly selected, when there is insufficient training time or when the model is too simple to learn from large training data due to limited model capacity (i.e. expressive power). Note 2 to entry: See Figure 2 for a graphical comparison with overfitting (3.7). [5] [SOURCE: ISO/IEC 23053:2022, 3.1.5 , modified — In Note 1 to entry, “when there is” was added; Note 2 to entry was added.] a) Underfitting b) Good fit c) Overfitting Key training data model prediction X input variable Y output variable Figure 2 — Fit of the ML model on the training data

SIST-TS CEN ISO/TS 24971-2:2026 ISO/TS 24971-2:2026(en)

4 General requirements for risk management system

4.1 Risk management process

ISO 14971 specifies a process for risk management of medical devices. This process also applies to ML-enabled medical devices. The process covers the identification of hazards and hazardous situations, the estimation and evaluation of the associated risks, the need for risk control, the evaluation of overall residual risk and the production and post-production activities for monitoring the effectiveness of the risk control measures.

4.2 Management responsibilities

ISO 14971 requires top management to:

a) provide adequate resources; b) assign competent personnel; c) define and document a policy for establishing the criteria for risk acceptability; and d) review the suitability of the risk management process to ensure its continuing effectiveness. [2] There is no specific guidance for MLMD in addition to what is provided in ISO/TR 24971 .

4.3 Competence of personnel

ISO 14971 requires that persons performing risk management tasks are competent and, where appropriate, have knowledge of and experience with the particular medical device, its use, the technologies involved, and the risk management techniques employed.

For MLMD specifically, this knowledge should cover the relevant stages of the MLMD life cycle. This can include general knowledge about the training data and test data used for the MLMD, awareness of the data quality, understanding of the context of the data and knowledge of what the data actually mean in clinical practice.

The team performing the risk management tasks for the MLMD should include:

[23] a) individuals with an understanding of good machine learning practices (see IMDRF document N88 ), who are trained on ML model and ML algorithm development to identify hazardous situations that can occur for the MLMD; b) individuals with knowledge of information technology (IT) to identify risks and risk control measures related to the computing platforms (when applicable) where the MLMD is to be installed and used, including the security risk control measures for MLMD connected to the internet or to other equipment; c) individuals trained on simulation and algorithm testing to identify risk control measures for the MLMD, [9] either in software or in hardware (see IEC 62304 ); d) individuals with knowledge of software validation to develop verification methods for the risk control measures; e) individuals trained on usability engineering (human factors engineering) in the context of the MLMD [10] use in accordance with IEC 62366-1 , including how transparency and explainability can affect the MLMD use (e.g. the potential for bias in interpreting the output); f) individuals with knowledge of the clinical workflow for the MLMD, understanding of the environment in which the MLMD is to be used, understanding of the interactions of the MLMD with medical practitioners and patients, and the capability to evaluate the relevance of the results produced by the MLMD in clinical practice; and

SIST-TS CEN ISO/TS 24971-2:2026 ISO/TS 24971-2:2026(en) g) individuals with an understanding of good data management practices to identify risks related to data management, including data integrity, quality of training data and test data (e.g. data robustness, data fitness to the ML model and free of unintended bias), and data privacy. NOTE An explanation of bias is given in Annex A.

4.4 Risk management plan

ISO 14971 requires the manufacturer to establish a risk management plan that includes:

a) the scope of the risk management activities; b) the assignment of responsibilities and authorities; c) requirements for the review of risk management activities; d) the criteria for risk acceptability; e) the method to evaluate overall residual risk and the criteria for acceptability of the overall residual risk; f) activities for verification of the implementation and effectiveness of risk control measures; and g) activities to collect and review production and post-production information. Specifically for MLMD, the manufacturer should consider whether the post-production activities need to include monitoring the MLMD performance, updating the MLMD or retraining the ML model. If monitoring the MLMD performance is considered necessary for safety, the risk management plan shall define the methods and processes for such monitoring and which data are to be collected and reviewed to support this activity. If updating the MLMD or retraining the ML model is considered necessary for the purpose of managing risk, the risk management plan should include the criteria for initiating an update or retraining, the frequency of applying those criteria (occasionally, after a set period or continuously) and the planned activities in the update or retraining process itself. These activities may further include software maintenance and provisions for returning the MLMD to a previous version. NOTE Retraining the ML model and returning the MLMD to a previous version can be subject to regulatory [1] requirements (see ISO 13485:2016, 7.3.9 and 8.2.3 ).

EXAMPLE 1 Methods for monitoring the MLMD performance, including periodic testing with new test data, to verify that the MLMD continues to perform as intended.

EXAMPLE 2 Processes for collecting specific data from reported complaints, processes for collecting necessary input data (e.g. required data input fields to be filled by users before the MLMD can operate).

4.5 Risk management file

ISO 14971 requires the manufacturer to establish and maintain a risk management file. The risk management file contains the results produced by the risk management process as well as traceability from each identified hazard to the risk analysis, the risk evaluation, the implementation and verification of risk control measures and the results of evaluation of the residual risk.

[2] There is no specific guidance for MLMD in addition to what is provided in ISO/TR 24971 .

5 Risk analysis

5.1 Risk analysis process

ISO 14971 requires the manufacturer to perform risk analysis for the particular medical device and to record the results of the planned risk analysis activities in the risk management file. The techniques that support risk [2] analysis (see ISO/TR 24971:2020, Annex B ) can also be applied to ML-enabled medical devices.

SIST-TS CEN ISO/TS 24971-2:2026 ISO/TS 24971-2:2026(en)

5.2 Intended use and reasonably foreseeable misuse

ISO 14971 requires the manufacturer to document a clear description of the medical device, its intended use and the reasonably foreseeable misuse. It is emphasized that reasonably foreseeable misuse includes use error (for example related to lack of transparency), use outside the intended patient population, use with inadequate input data, overreliance on MLMD output. Questions related to MLMD use that can assist in identifying hazards and characteristics related to safety are given in Annex C. Differences in MLMD performance can relate to the selection of training data from specific patient groups, leading to differences in performance for other patient groups. The manufacturer can decide to restrict the intended use based on such differences in performance.

...