kSIST FprEN IEC 63483:2026 is a SIST standard for methods for spectral imaging performance evaluation of computed tomography (CT). It specifies selected methods and metrics for conventional image, virtual monoenergetic image [VMI], virtual non-contrast image [VNC], iodine concentration image, and electron density image outputs from CT scanners that conform to IEC 60601-2-44. The document is written for people who need to measure, record, and compare CT spectral imaging performance with specified phantoms and test objects.

What does kSIST FprEN IEC 63483:2026 specify?

kSIST FprEN IEC 63483:2026 specifies a framework for evaluating CT spectral imaging performance, not a full clinical performance profile. It sets out the image types covered, the standard CT metrics that can still be used, and dedicated measurement methods for spectral image types.

The document applies to spectral images generated from data provided by a CT scanner conforming to IEC 60601-2-44. The images may be generated on the scanner or with separate software. Clause 1 limits the scope to these image types:

AnnexWhat it covers
ACalculation of CT number in virtual monoenergetic images
BCorrelations with DICOM Supplement 188, YY/T 1766.3, and AAPM TG291/TG299
CLimitations of phantom measurements
DWhy the document focuses on the listed spectral image types

Clause 1 also states that the methods are selected methods and metrics. They do not replace manufacturer performance specifications or testing protocols, and they do not set acceptance or constancy criteria.

What are the key requirements of kSIST FprEN IEC 63483:2026?

Test conditions and reporting

Clause 4.1 requires the manufacturer’s recommended CT conditions of operation and reconstruction settings to be used when evaluating spectral imaging performance. If those are not specified, the manufacturer’s reference clinical protocol should be used instead. The scan settings, reconstruction settings, method, measurement results, reference values, and dose information shall be recorded.

In practice, this means the evaluation must be reproducible and tied to the intended clinical use. Clause 4.1 also says the same CT conditions of operation should be used for both conventional CT metric evaluation and spectral imaging metric evaluation.

Clause 4.4 adds an important limitation: the results depend on scanner hardware, acquisition protocol, reconstruction method, and phantom design. The methods are based on measurable physical quantities, so they are not intended to evaluate every individual clinical task.

Conventional CT metrics and where they apply

Clause 4.2 separates the document into two method groups: conventional CT dose and imaging metrics from other IEC standards, and dedicated spectral-image methods. Clause 5.1 allows these conventional CT metrics to be used:

  • reconstructed section thickness
  • noise
  • accuracy of mean CT number
  • uniformity
  • high-contrast spatial resolution

Clause 5.2 limits their use to conventional image, VMI, and VNC. For VMI, three energy settings must be evaluated - one low, one midrange, and one high energy level within 50 keV to 110 keV.

In practice, this tells the reader where the usual CT quality tests still make sense. Clause 5 also excludes low-contrast resolution and low-contrast detectability from the conventional metric set used here.

Phantom and test object requirements

Clause 6.2.1 requires the phantom to be made from tissue-equivalent or water-equivalent base material, with attenuation properties suitable for diagnostic CT energies. The phantom should be cylindrical with a circular or elliptical cross-section, and two sizes should be available: a small size for head-type use and a large size for body-type use. The length must be sufficient to avoid edge effects.

Clause 6.2.2 requires cylindrical test objects for iodine concentration and electron density measurements. The iodine objects should represent clinically relevant iodine-in-water concentrations, and the electron-density objects should include equivalents for water, soft tissue, dense bone, and air. The test objects are to be aligned parallel to the phantom axis and placed away from the center and edge to reduce beam hardening artefacts.

Clause 6.2.3 requires the phantom to be centered and aligned with the scanner z-axis. The chosen phantom size should match the anatomical region and the scan range should be centered on the test object.

Image-specific measurement methods

Clause 6.3.1 defines the VNC method. It uses iodine-in-water test objects at 0, 2, 5, 10, and 15 mg/ml. A region of interest [ROI] is placed at the center of each test object and the average pixel value is measured. The blank object, or a matching base-material ROI, is used as the reference. This checks how well iodine signal is removed.

Clause 6.3.2 defines CT number accuracy in VMI. It uses iodine-in-water test objects between 5 mg/ml and 15 mg/ml, and evaluates the selected VMI images from Clause 5.2. The center ROI is measured in each keV image, then compared with the provided CT number or with a calculated reference value from Annex A. In practice, this confirms whether the VMI output matches the expected CT number at the chosen energy.

Clause 6.3.3 covers iodine concentration image accuracy. It uses 2, 5, 10, and 15 mg/ml test objects and allows measurements at different distances from the phantom center to check positional dependence. If the image is reported in Hounsfield units, it is converted to mass density using the manufacturer’s factor when available. The phantom manufacturer’s nominal iodine concentration is used as the reference.

Clause 6.3.4 covers electron density image accuracy. It uses tissue-mimicking objects equivalent to water, soft tissue, dense bone, and air. A center ROI is measured for each object, and the phantom manufacturer’s nominal electron density is the reference value.

What terms does kSIST FprEN IEC 63483:2026 define?

  • CT spectral imaging - CT imaging based on separate sets of data representing different energy spectra, so the resulting images reflect the spectral properties of the scanned materials.
  • Spectral image - A CT image generated by CT spectral imaging.
  • Conventional image - A spectral image whose pixel values represent CT number under a specified tube voltage and filtration.
  • Virtual monoenergetic image [VMI] - An image whose pixel values represent CT numbers for a monoenergetic x-ray source at a specified energy in keV.
  • Virtual non-contrast image [VNC] / virtual un-enhanced image [VUE] - A material-removed image that excludes the attenuation contribution of a contrast-agent-related material such as iodine.
  • Iodine concentration image - A material-specific image whose pixel values represent iodine-related attenuation or concentration.
  • Electron density image - An image whose pixel values correspond to electrons per unit volume, such as electron density relative to water.
  • Region of interest [ROI] - A localized part of an image with a defined shape and size used for measurement.
  • Test object - An individual object with defined properties, such as size, shape, and material, used for image quality evaluation.

Who uses kSIST FprEN IEC 63483:2026?

kSIST FprEN IEC 63483:2026 is used by CT manufacturers, application specialists, medical physicists, service engineers, quality managers, and imaging departments that evaluate spectral CT outputs. It is also relevant to test laboratories and procurement teams that need a common method for checking image performance claims.

In practice, these users apply it to choose phantoms, set scan conditions, measure VMI, VNC, iodine concentration, and electron density images, and document reproducible results for quality assurance or product evaluation.

Which standards are used with kSIST FprEN IEC 63483:2026?

  • IEC 60601-1:2005 - Provides the general requirements for basic safety and essential performance of medical electrical equipment.
  • IEC 60601-1-3:2008 - Provides the collateral standard for radiation protection in diagnostic X-ray equipment.
  • IEC 60601-2-44:2009 and IEC 60601-2-44:2009/AMD2:2016 - Provide the CT-specific safety and essential performance framework, including CT conditions of operation and CT scanner definitions used throughout the document.
  • IEC 61223-3-5:2019 - Provides the conventional CT acceptance and constancy test concepts, definitions, and methods used for the conventional metrics referenced in Clause 5.
  • IEC 62945:2018 - Provides the source definition for test object.

Annex B also relates the document to DICOM Supplement 188, YY/T 1766.3, and AAPM TG291 and TG299 as related multi-energy CT work.

What does the kSIST FprEN IEC 63483:2026 document contain?

The document contains definitions, general evaluation guidance, applicability rules for conventional CT metrics, and dedicated methods for spectral image performance testing. Table 1 shows which conventional CT metrics apply to each image type, including the special case for VMI at selected energies.

Annex A gives the calculation approach for CT number in VMI, including mass attenuation coefficient handling. Annex B maps the document’s image types to related multi-energy CT work. Annex C explains why phantom-based results can change with scanner settings, decomposition method, and phantom design. Annex D explains why the standard focuses on the listed spectral image types and excludes other clinically motivated outputs that are not suitable for this type of performance evaluation.

Buy Documents

Draft

oSIST prEN IEC 63483:2025

English language (24 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

kSIST FprEN IEC 63483:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Methods for spectral imaging performance evaluation of computed tomography". This standard covers: Methods for spectral imaging performance evaluation of computed tomography

Methods for spectral imaging performance evaluation of computed tomography

kSIST FprEN IEC 63483:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.50 - Radiographic equipment; 35.240.80 - IT applications in health care technology. The ICS classification helps identify the subject area and facilitates finding related standards.

kSIST FprEN IEC 63483:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 Terms defined in this standard
  • 3.1.1 CT spectral imaging
  • 3.1.2 Spectral image
  • 3.1.3 Conventional image
  • 3.1.4 Virtual monoenergetic image
  • 3.1.5 Virtual non-contrast image
  • 3.1.6 Iodine concentration image
  • 3.1.7 Electron density image
  • 3.2 Terms defined in other standards
  • 3.2.1 CT conditions of operation
  • 3.2.2 CT scanner
  • 3.2.3 CT number
  • 3.2.4 Mean CT number
  • 3.2.5 Noise
  • 3.2.6 Reconstructed section thickness
  • 3.2.7 Region of interest
  • 3.2.8 Spatial resolution
  • 3.2.9 Test object
  • 3.2.10 Uniformity
  • 4 General consideration of spectral imaging performance evaluation
  • 4.1 Typical conditions of operation for CT spectral imaging (3.1.1)
  • 4.2 Performance evaluation methodologies for CT spectral imaging (3.1.1)
  • 4.3 Dosimetry of spectral imaging
  • 4.4 Technical considerations for proper use for the evaluation methods
  • 5 Applicability of conventional CT imaging performance metrics for spectral
  • 5.1 Conventional CT performance evaluation metrics
  • 5.2 Applicability to different types of spectral images (3.1.2)
  • 6 CT spectral imaging (3.1.1) performance evaluation
  • 6.1 General
  • 6.2 Phantoms for spectral imaging performance evaluation
  • 6.3 Evaluation of spectral imaging performance
  • Annex A
  • Annex B
  • Annex C
  • Annex D
  • Bibliography

Introduction

The purpose of the document is to provide a set of standardized methods and metrics for the evaluation of CT spectral imaging (3.1.1).

Main content of this document includes:

  • Key definitions of CT spectral imaging (3.1.1) and also spectral image (3.1.2) types differentiated as conventional CT
  • Applicability and scenarios of conventional imaging performance evaluation metrics and methodology in CT spectral imaging (3.1.1)
  • Dedicated performance evaluation metrics, methodologies and applicable phantom specifications for corresponding image types from CT spectral imaging (3.1.1)
  • Necessary notes/informative annex about CT spectral imaging (3.1.1) technology and references

oSIST prEN IEC 63483:2025

IEC 2025

1 Scope

This document applies to spectral image (3.1.2) types based on data provided by a CT scanner (3.2.2) that confirms to IEC 60601-2-44 . The spectral images (3.1.2) can be generated either on the CT scanner (3.2.2) or with separate software.

The purpose of this standard is to provide selected methods and metrics for evaluation of imaging performance associated with the following spectral image (3.1.2) type (if available):

- Conventional image (3.1.3) - Virtual monoenergetic image (3.1.4) - Virtual non contrast image (3.1.5) - Iodine concentration image (3.1.6) - Electron density image (3.1.7) NOTE The methods presented in this document do not provide comprehensive technical performance evaluation of CT spectral imaging (3.1.1).

The results obtained using the methods of this document do not replace CT spectral imaging (3.1.1) performance specifications and testing protocols as provided by manufacturer. This standard does neither specify performance criteria for acceptance testing or constancy testing nor does it specify requirements for manufacturers to provide such criteria. However, manufacturers may provide performance data and/or quality assurance test criteria according to the methods described in this standard.

Additionally, phantom requirements are specified accordingly.

Spectral image (3.1.2) types not listed above are not covered by this document, as explained in Annex D.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance IEC 60601-1-3:2008, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X- ray equipment IEC 60601-2-44:2009, Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography IEC 60601-2-44:2009/AMD2:2016, Amendment 2 - Medical electrical equipment - Part 2-44:

Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography IEC 61223-3-5:2019, Evaluation and routine testing in medical imaging departments - Part 3-5:

Acceptance and constancy tests - Imaging performance of computed tomography X-ray equipment

oSIST prEN IEC 63483:2025

IEC 2025

IEC 62945:2018, Radiation protection instrumentation - Measuring the imaging performance of X-ray computed tomography (CT) security-screening systems

3 Terms and definitions

For the purposes of this document, the terms and defintions given in IEC 60601-2-44:2009, IEC 60601-2-44:2009/AMD2:2016 and IEC 61223-3-5:2019, and the following terms and definitions apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • IEC Electropedia: available at https://www.electropedia.org/
  • ISO Online browsing platform: available at https://www.iso.org/obp

3.1 Terms defined in this standard

3.1.1 CT spectral imaging

CT imaging based on separate sets of data representing different energy spectra, so that the resulting CT images can reflect the different spectral properties of the materials in the object scanned.

Note 1 to entry: CT spectral imaging can be realized based on different techniques.

3.1.2 Spectral image

CT images generated by CT spectral imaging (3.1.1).

Note 1 to entry: Spectral image types addressed in this document are listed in Clause 1. For those spectral image (3.1.2) types not covered in this document, refer to rationale in Annex D.

3.1.3 Conventional image

Image created by CT spectral imaging (3.1.1), where the pixel value in the image represents the CT number (3.2.3) of the object scanned under a specified tube voltage and filtration. Note 1 to entry: Manufacturers may have different naming conventions on their products. Note 2 to entry: Examples of conventional images (3.1.3) include virtual monoenergetic images (3.1.4) at a specified keV and image types named e.g., "Combined Images", "Mixed Images" and "Single kVp like Images". Note 3 to entry: The pixel value in the conventional image (3.1.3) can deviate from the pixel value obtained by nonspectral CT imaging.

3.1.4 Virtual monoenergetic image

VMI Image created by CT spectral imaging (3.1.1), where the pixel values represent the CT numbers (3.2.3) of the object scanned with a monoenergetic x-ray source at a specified energy (in keV).

Note 1 to entry: Virtual monoenergetic images (3.1.4) are also referred as virtual monochromatic images or synthetic monoenergetic images.

Note 2 to entry: Manufacturers may have different naming conventions on their products for above image type.

3.1.5 Virtual non-contrast image

VNC Virtual un-enhanced image

oSIST prEN IEC 63483:2025

IEC 2025

VUE Material-removed image created by CT spectral imaging (3.1.1), where the pixel values represent the CT numbers (3.2.3) after excluding the attenuation contribution of a specific material component related to a contrast-agent (e.g., iodine).

Note 1 to entry: Manufacturers may have different naming conventions on their products for above image type. Note 2 to entry: There may be image types which can be generated in the similar way but with different pixel values represented, e.g. the pixel values representing the density of water. In this case the methods and metrics specified for Virtual non-contrast images (3.1.5) in this document can not be directly applied before converting the pixel values to CT numbers (3.2.3).

3.1.6 Iodine concentration image

Material-specific image created by CT spectral imaging (3.1.1), where the pixel values represent the attenuation contribution or concentration of the iodine-like material component. Pixel values are given in units of mass density or, in case of representation as the corresponding iodine enhancement image, in Hounsfield units.

Note 1 to entry: Manufacturers may have different naming conventions on their products for above image type.

3.1.7 Electron density image

Image created by CT spectral imaging (3.1.1), where the pixel values correspond to the number of electrons per unit volume (e.g., given as a particle number density or electron density relative to water).

Note 1 to entry: Manufacturers may have different naming conventions on their products for above image type.

3.2 Terms defined in other standards

3.2.1 CT conditions of operation

all selectable parameters governing the operation of a CT scanner (3.2.2) Note 1 to entry: Examples of such conditions include nominal tomographic section thickness (IEC 60601-2- 44:2009, 201.3.206), CT pitch factor (IEC 60601-2-44:2009/AMD2:2016, 201.3.204), filtration (IEC 60601-1- 3:2008, 3.24), peak X-ray tube voltage and either X-ray tube current and loading time or current time product. Note 2 to entry: Some CT conditions of operation (3.2.1) may vary during the exposure. Note 3 to entry: CT conditions of operation (3.2.1) include parameters that are derived by the system from the userselectable parameters. [SOURCE: IEC 60601-2-44:2009, 201.3.202]

3.2.2 CT scanner

X-ray equipment intended to generate cross-sectional images of the body by computer reconstruction of X-ray transmission data obtained at different angles, which may include signal analysis and display equipment, patient support, support parts and accessories Note 1 to entry: The scope of this document is limited to CT scanners (3.2.2) intended to be used for both head and body characterised by an enclosure of the X-ray source(s) and imaging detector(s) in a common protective cover in the shape of a toroid.

Note 2 to entry: Secondary imaging processing is not included in the scope of this standard. [SOURCE: IEC 60601-2-44:2009, 201.3.201]

oSIST prEN IEC 63483:2025

IEC 2025

3.2.3 CT number

number used to represent the mean X-ray attenuation (IEC 60601-1-3:2008, 3.7) associated with each elemental area of the computed tomography image Note 1 to entry: The CT number (3.2.3) is normally expressed in Hounsfield units. Measured values of the linear attenuation (IEC 60601-1-3:2008, 3.7) coefficients are transformed into CT numbers (3.2.3) using the international Hounsfield scale, using the expression:

𝜇 − 𝜇 𝑚𝑎𝑡𝑒𝑟𝑖𝑎𝑙 𝑤𝑎𝑡𝑒𝑟 𝐶𝑇 𝑛𝑢𝑚𝑏𝑒𝑟 = × 1000 𝜇 𝑤𝑎𝑡𝑒𝑟 where 𝜇 isthe linear attenuation (IEC 60601-1-3:2008, 3.7) coefficient. Note 2 to entry: The CT number (3.2.3) scale is defined so that water has a value of 0 and air a value of –1 000 (𝜇 is assumed to be 0).

𝑎𝑖𝑟 [SOURCE: IEC 61223-3-5:2019, 3.6]

3.2.4 Mean CT number

mean value of the CT number (3.2.3) of all pixels within a certain defined region of interest (3.2.7) [SOURCE: IEC 61223-3-5:2019, 3.14]

3.2.5 Noise

variation of CT numbers (3.2.3) from a mean value in a defined area in the image of a uniform substance Note 1 to entry: The magnitude of noise (3.2.5) is indicated by the standard deviation of the CT numbers (3.2.3) of a uniform substance in the region of interest (3.2.7).

[SOURCE: IEC 61223-3-5:2019, 3.15]

3.2.6 Reconstructed section thickness

Full width at half maximum (IEC 61223-3-5:2019, 3.10), of the sensitivity profile (IEC 60601- 2-44:2009, 201.3.207) of a reconstructed image [SOURCE: IEC 61223-3-5:2019, 3.17]

3.2.7 Region of interest

ROI Localized part of an image, of defined shape and dimension, which is of particular interest at a given time [SOURCE: IEC 61223-3-5:2019, 3.18]

3.2.8 Spatial resolution

CT scanners (3.2.2) ability to resolve spatially adjacent objects in the image, when the difference in attenuation (IEC 60601-1-3:2008, 3.7) between the objects and the background is large compared to noise (3.2.5) [SOURCE: IEC 61223-3-5:2019, 3.20]

oSIST prEN IEC 63483:2025

IEC 2025

3.2.9 Test object

Individual object having specific properties (size, shape, materials, etc.) that when imaged by the system allows a certain image quality evaluation to be carried out. [SOURCE: ,IEC 62945:2018, 3.1.17]

3.2.10 Uniformity

consistency of the CT numbers (3.2.3) of the image of a homogeneous material across the scan field [SOURCE: IEC 61223-3-5:2019, 3.24]

4 General consideration of spectral imaging performance evaluation

4.1 Typical conditions of operation for CT spectral imaging (3.1.1)

If available, the manufacturer's recommended CT conditions of operation (3.2.1) and reconstruction settings shall be used when evaluating spectral imaging performance. The CT conditions of operation (3.2.1) should align with the clinical application of CT spectral imaging (3.1.1). If not specified, the reference CT conditions of operation (3.2.1) according to the clinical protocols provided by the manufacturer for CT spectral imaging (3.1.1) shall be used. NOTE 1 Reference protocols on the system may not be optimized for or compatible with a specific phantom. The CT conditions of operation (3.2.1), reconstruction settings, performance testing methodology, measurement results, and reference values shall be recorded. For each of the spectral image (3.1.2) types, the same CT conditions of operation (3.2.1) should be applied for both conventional CT imaging metric evaluation and spectral imaging metric evaluation, as described in sections Clause 5 and Clause 6 respectively. Dose information associated with the scan shall be recorded.

NOTE 2 The scanner settings and measurement techniques are important to record so that the tests and results can be effectively reproduced.

4.2 Performance evaluation methodologies for CT spectral imaging (3.1.1)

The performance evaluation of CT spectral imaging (3.1.1), as described in this document, includes two kinds of metrics/methodologies:

- Conventional CT dose/imaging performance metrics, defined in other IEC standards (IEC 60601-2-44 and IEC 61223-3-5) - Particular performance evaluation metrics and methodology for CT spectral imaging (3.1.1) related image types NOTE For certain image types which could be used for either routine diagnostics or spectral applications (e.g., virtual monoenergetic images), both methods may be applicable For conventional metrics, this document will specify the corresponding applicability for CT spectral imaging (3.1.1). Relevant contents are specified in Clause 5. For particular metrics for CT spectral imaging (3.1.1), standardized procedures and also dedicated phantom requirements are specified in Clause 6.

oSIST prEN IEC 63483:2025

IEC 2025

4.3 Dosimetry of spectral imaging

The dose performance metrics identified in IEC 60601-2-44 are applicable.

...