oSIST prEN IEC 80601-2-86:2026
(Main)Medical electrical equipment - Part 2-86: Particular requirements for the basic safety and essential performance of electrocardiographs, including diagnostic equipment, monitoring equipment, ambulatory equipment, electrodes, cables and leadwires
- Status
- Not Published
- Public Enquiry End Date
- 30-Sep-2026
- Technical Committee
- IEMO - Electrical equipment in medical practice
- Current Stage
- 4020 - Public enquire (PE) (Adopted Project)
- Start Date
- 23-Jul-2026
- Due Date
- 10-Dec-2026
- Completion Date
- 29-Sep-2026
oSIST prEN IEC 80601-2-86:2026 is the Slovenian standard identical to prEN IEC 80601-2-86:2026. It sets particular requirements for the basic safety and essential performance of ECG ME EQUIPMENT, including diagnostic equipment, monitoring equipment, ambulatory equipment, electrodes, cables and leadwires. The document is a Committee Draft for Vote, so it is still under study and subject to change.
What does oSIST prEN IEC 80601-2-86:2026 specify?
oSIST prEN IEC 80601-2-86:2026 specifies ECG-specific requirements that supplement the IEC 60601 general standard for medical electrical equipment. It covers ECG ME EQUIPMENT and ACCESSORIES for professional healthcare facilities, emergency medical service environments and home healthcare environments, with requirements that depend on the claimed intended use.
The standard organizes its content around the general IEC 60601 structure, then adds ECG-specific clauses for scope, definitions, testing, classification, identification and marking, electrical and mechanical hazards, EMC, alarms, and accessory performance. Annexes AA, BB, CC, DD, EE and GG provide guidance, electrode information, clinical use cases, ECG measurement rules, a CTS test atlas and alarm diagrams.
| Annex | What it covers |
|---|---|
| AA | General guidance and rationale |
| BB | Positions, identifications and colour codes of additional ECG ELECTRODEs |
| CC | ECG clinical use cases |
| DD | Definitions and rules for the measurement of electrocardiograms |
| EE | CTS test atlas |
| GG | Alarm diagrams of Clause 208/IEC 60601-1-8:2006 +A1:2012 |
The scope includes ECG devices or parts of systems that acquire a body-surface ECG signal and display, record, analyze or transmit the data for diagnosis, treatment or monitoring. It excludes direct fetal ECG equipment and Fetal & Maternal MONITORING EQUIPMENT, and it excludes MRI-specific ECG trunk cables and lead wires.
What are the key requirements of oSIST prEN IEC 80601-2-86:2026?
The standard’s main job is to make sure ECG ME EQUIPMENT remains safe and performs as claimed in normal use and under foreseeable fault or test conditions. The practical consequence is that the manufacturer’s intended use drives which tests, disclosures and accessory checks apply.
Scope, intended use and equipment categories
Clause 201.1 separates ECG ME EQUIPMENT into general, diagnostic, monitoring and ambulatory requirements. Diagnostic ECG ME EQUIPMENT, MONITORING ECG ME EQUIPMENT and AMBULATORY ECG ME EQUIPMENT are treated differently because they support different clinical decisions and use patterns.
This matters because a device’s labeling and intended use determine the performance claims it must support. If a device claims 12-lead diagnosis, ambulatory recording, or monitoring with alarms, the relevant ECG-specific clauses apply.
Defibrillation behavior and applied-part safety
Clause 201.8.5.5.2 requires ECG ME EQUIPMENT to behave safely if a defibrillator discharge occurs at the patient connection. The standard distinguishes DEFIBRILLATION PROOF equipment, which is designed to recover after defibrillation, from DEFIBRILLATION SAFE equipment, which must not create a hazard even if it is damaged by the shock.
This matters in practice because patient-connected ECG systems may stay attached during emergency defibrillation. The standard also requires testing with the manufacturer’s recommended accessories and, where needed, worst-case accessory combinations.
Protection against water ingress and power interruptions
Clause 201.11.6.5 gives transportable ECG ME EQUIPMENT an ingress protection requirement against accidental wetting. Clause 201.11.8 then sets the behavior for SUPPLY MAINS interruption, including retention of operator settings and stored patient data, continuation from an internal power source when present, and visible indication when operating from that source.
Clause 201.11.8.101 adds battery-depletion protection for monitoring equipment. In practice, the device has to warn the operator before battery exhaustion becomes a functional problem and then power down in a controlled way.
ECG signal quality, frequency response and ST measurement
Clause 201.12.4.102 defines the frequency response expectations for ECG ME EQUIPMENT, with different requirements for general ECG use, ST monitoring and diagnostic ECG use. For diagnostic devices and devices that interpret the ST segment, the standard also checks how the system handles waveform shape, baseline recovery and ringing.
This matters because ECG interpretation depends on faithful reproduction of the cardiac waveform, not just a visible trace. A device that distorts low-frequency or ST-segment content can mislead diagnosis or automated analysis.
Pacemaker pulse handling and arrhythmia performance
Clauses 201.12.4.113 and 201.12.4.114 cover pacemaker pulse detection, display and rejection capability. The standard requires the manufacturer to disclose detection limits, identify additional test patterns when needed, and document false pulse detection behavior.
Clauses 201.12.4.119.3 and Table 201.114 cover performance reporting for arrhythmia algorithms. In practice, this means the manufacturer must evaluate and report algorithm behavior for rhythm and event detection using defined databases, summary matrices and record-by-record statistics.
ECG accessories: trunk cables, lead wires and electrodes
Clauses 201.15.101.1.3 and 201.15.101.1.4 address dielectric withstand and noise for TRUNK CABLEs and LEAD WIRES. Related tables cover flex life, tensile strength, connector mating cycles and lead-wire resistance.
This matters because a good ECG unit can still fail in use if the cable assembly is noisy, mechanically weak or electrically unsafe. The standard treats the accessory chain as part of the medical device system, not as a separate afterthought.
EMC, HF surgery interference and alarms
Clause 202 applies the EMC collateral standard with ECG-specific immunity requirements, including electrosurgery interference. Clause 208 applies the alarm-system collateral standard to MONITORING ECG ME EQUIPMENT, with ECG-specific alarm priorities, inoperable-device indications, low-battery alarms, audio-off indication and alarm timing rules.
That is important because monitoring ECG devices often drive immediate clinical action. The alarm behavior has to be clear, timely and consistent with risk management.
What terms does oSIST prEN IEC 80601-2-86:2026 define?
- ECG ME EQUIPMENT - Any ECG device, system or part of a system that acquires a body-surface ECG and displays, records, analyzes or transmits it for diagnosis, treatment or monitoring.
- DIAGNOSTIC ECG ME EQUIPMENT - ECG ME EQUIPMENT intended to acquire 12 or more LEADs for diagnosis from rhythm, conduction and waveform contour.
- MONITORING ECG ME EQUIPMENT - ECG ME EQUIPMENT used to monitor cardiac activity, including alarm-related use where ECG results may drive immediate clinical action.
- AMBULATORY ECG ME EQUIPMENT - ECG ME EQUIPMENT that acquires ECG from an ambulatory patient, including Holter systems, telemetry and patch recorders.
- ECG ELECTRODE - A sensor in contact with a specified part of the body to detect electrical cardiac activity.
- DEFIBRILLATION PROOF - ECG ME EQUIPMENT and applied parts designed to withstand defibrillator discharge and recover operation afterward.
- DEFIBRILLATION SAFE - ECG ME EQUIPMENT and applied parts designed so that leaving them attached during defibrillation does not harm the patient, even if the device is damaged.
Who uses oSIST prEN IEC 80601-2-86:2026?
Manufacturers use this standard to design ECG monitors, diagnostic recorders, ambulatory recorders, electrodes, trunk cables and lead wires to the expected safety and performance profile. Test laboratories use it to plan electrical, EMC, defibrillation, accessory and algorithm verification.
Quality, regulatory and clinical engineering teams use it to check labeling, intended use, test evidence, alarm behavior and technical documentation. Buyers and hospital specifiers use it to compare ECG systems on the basis of claimed use, accessory compatibility and alarm/monitoring behavior.
What changed in oSIST prEN IEC 80601-2-86:2026 from the previous edition?
This first edition cancels and replaces IEC 60601-2-25:2011, IEC 60601-2-27:2011 and IEC 60601-2-47:2012. It is described as a technical revision aligned with IEC 60601-1:2005 with its amendments and newer collateral standards.
The document also combines the former ECG particular standards into one text for general ECG requirements plus monitoring, diagnostic and ambulatory ECG systems. It adds requirements for ECG TRUNK CABLEs, PATIENT LEAD WIREs, DISPOSABLE ECG ELECTRODEs and cardiac rhythm and ST segment measurement algorithms, and it brings in requirements previously omitted from IEC 60601-2-51.
Which standards are used with oSIST prEN IEC 80601-2-86:2026?
- IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 - the general standard for basic safety and essential performance that this part amends and supplements.
- IEC 60601-1-2:2014 and IEC 60601-1-2:2014/AMD1:2020 - the EMC collateral standard, used here with ECG-specific modifications in Clause 202.
- IEC 60601-1-8:2006, IEC 60601-1-8:2006/AMD1:2012 and IEC 60601-1-8:2006/AMD2:2020 - the alarm-system collateral standard, used here with ECG-specific modifications in Clause 208.
- IEC 60601-2-2:2017 - the HF surgical equipment standard, referenced for electrosurgery-interference testing.
- IEC 60529:1989 + AMD1:1999 + AMD2:2013 - used for the wetting and ingress-protection test applied to transportable equipment.
What does the oSIST prEN IEC 80601-2-86:2026 document contain?
The document contains many ECG-specific test figures, including defibrillation test circuits, frequency-response setups, common-mode rejection tests, baseline recovery, pacemaker pulse simulation, cable noise measurement, dielectric withstand and HF surgery protection measurement. It also contains tables for electrode colour codes and positions, accessory mechanical and electrical tests, and the data structures used for algorithm performance reporting.
Annexes and tables support both device design and verification work. Annex DD explains ECG measurement terms and rules, Annex EE provides a CTS test atlas, and Annex GG gives alarm diagrams linked to Clause 208. The algorithm clauses use database-based reporting tables for arrhythmia, ST and HRV evaluation, so readers can see how performance claims are to be documented.
Relations
- Effective Date
- 30-Jun-2026
- Effective Date
- 26-May-2026
- Effective Date
- 26-May-2026
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
oSIST prEN IEC 80601-2-86:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Medical electrical equipment - Part 2-86: Particular requirements for the basic safety and essential performance of electrocardiographs, including diagnostic equipment, monitoring equipment, ambulatory equipment, electrodes, cables and leadwires". This standard covers: Medical electrical equipment - Part 2-86: Particular requirements for the basic safety and essential performance of electrocardiographs, including diagnostic equipment, monitoring equipment, ambulatory equipment, electrodes, cables and leadwires
Medical electrical equipment - Part 2-86: Particular requirements for the basic safety and essential performance of electrocardiographs, including diagnostic equipment, monitoring equipment, ambulatory equipment, electrodes, cables and leadwires
oSIST prEN IEC 80601-2-86:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.55 - Diagnostic equipment; 11.040.99 - Other medical equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
oSIST prEN IEC 80601-2-86:2026 has the following relationships with other standards: It is inter standard links to SIST EN 60601-2-25:2015, SIST EN 60601-2-27:2016, SIST EN 60601-2-47:2015. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
oSIST prEN IEC 80601-2-86:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 53 Figure 201.112 – Simulated QRS complex
- 54 Figure 201.113 – Pacemaker test circuit
- 55 Figure 201.114: dielectric withstand test
- 56 Figure 201.115 — Test setup for cable noise measurement
- 57 Figure 201.116 - Flex life test setup
- 58 Figure 201.117 – Test circuit for offset instability/internal noise determination
- 59 Figure 201.118 – Defibrillation overload test circuit (all capacitor and resistor values have a 60 tolerance of 10%)
- 61 Figure 201.119 – Defibrillation overload test circuit for cables containing resistors
- 62 Figure 202.101 – Test layout for radiated and conducted emission tests
- 63 Figure 202.102 – Set-up for immunity tests
- 64 Figure 202.103 – Test circuit for HF surgery protection measurement
- 65 Figure 202.104 – Test setup for HF surgery protection measurement
- 66 Figure AA.1 – Applied part with multiple PATIENT connections
- 67 Figure AA.2 – D-sub connector
- 68 Figure AA.3 Simulation of a rectangular pulse filtered by 0,05 Hz causal highpass filter (right) 69 and a 0,5 Hz non-causal highpass filter (left)
- 70 Figure AA.4 - Example of scatter plot of ST amplitude measurement
- 71 Figure AA.5 - Example of scatter plot of ST amplitude error values for all measurements
- 72 Figure AA.6 - Example of scatter plot of ST amplitude error values for ST reference values 73 from -200 microvolt to + 200 microvolt
- 74 Figure AA.7 - Example of scatter plot of ST slope measurements
- 75 Figure AA.8 - Example of scatter plot of all ST slope measurement errors
- 76 Figure AA.9—Example of scatter plot of all ST slope measurement errors for ST reference 77 values from -2.0 mV/sec to + 2.0 mV/sec
- 79 Figure BB.1 – LEADs and colours for fetal ecg
- 80 Figure BB.2 – Positions of the ECG ELECTRODEs on the fetus for fetal ECG (see Table BB.2)
- 81 Figure BB.3 – LEAD positions and colours for fetal scalp ECG (see Table BB.2)
- 82 Figure DD.1 – Normal electrocardiogram
- 83 Figure DD.2 – Determination of global intervals (example)
- 84 Figure DD.3 – Waveform durations, isoelectric segments
- 85 Figure DD.4 – QRS complex with small R-wave(s) (see Figure DD.5, DD.6)
- 86 Figure DD.5 – Detail of small accepted R-wave
- 87 Figure DD.6 – Detail of small rejected R-wave
- 88 Figure FF.1 – Nomenclature of CALIBRATION ECG S
- 89 Figure FF.2 – Nomenclature of analytical ECG s
- 90 Figure FF.3 – Noise Characteristics
- 91 Figure GG.1 – NON-LATCHING ALARM SIGNALS without ALARM RESET
- 92 Figure GG.2 – NON-LATCHING ALARM SIGNALS with ALARM RESET
- 93 Figure GG.3 – LATCHING ALARM SIGNALS with ALARM RESET
- 94 Figure GG.4 – Two ALARM CONDITIONS with ALARM RESET
- 96 Table 201.101 – ESSENTIAL PERFORMANCE requirements
- 97 Table 201.102 – ECG ELECTRODE colour codes, identification codes, and position, for standard 98 12 LEAD & Frank ECG placement
- 99 Table 201.103 – Identification and colour for AMBULATORY ECG placement
- 100 Table 201.104 – Protection against the effect of defibrillation (test conditions)
- 101 Table 201.105 – Frequency response by ECG category
- 102 Table 201.106 – Electrode connection configurations for test of multichannel crosstalk
- 103 Table 201.107 – Acceptable mean differences and standard deviations for global intervals and 104 Q-, R-, S-durations on calibration and analytical ECGs
- 105 Table 201.108 – Acceptable mean differences and standard deviations for global durations 106 and intervals for biological ECGs
- 107 Table 201.109 – Disclosed changes of measurements caused by NOISE on ECGs
- 108 Table 201.110 – Example for the documentation of accuracy measures for diagnostic 109 interpretative statements
- 110 Table 201.111 – Example for the documentation of accuracy measures for rhythm 111 interpretative statements
263 INTRODUCTION 264 This particular standard concerns the BASIC SAFETY and ESSENTIAL PERFORMANCE of 265 electrocardiographic equipment, including diagnostic equipment, monitoring equipment, 266 ambulatory equipment, electrodes, cables and LEAD WIREs. It amends and supplements 267 IEC 60601-1:2005 (third edition), IEC 60601-1:2005/AMD1:2012 and IEC 60601- 268 1:2005/AMD2:2020: Medical electrical equipment – Part 1: General requirements for basic 269 safety and ESSENTIAL PERFORMANCE hereinafter referred to as the general standard. 270 The aim of this first edition is to bring this particular standard up to date with reference to the 271 third edition of the general standard through reformatting and technical changes. Furthermore, 272 the purpose of this is to also accomplish the following additional goals:
273 1) To combine the three ECG particular standards into one particular standard, which will 274 cover general ECG requirements as well as specific requirements for monitoring, 275 diagnostic, and ambulatory ECG systems.
276 2) Inclusion of previously omitted requirements from 60601-2-51.
277 3) Addition of requirements for ECG TRUNK CABLES and patient LEAD WIREs (ANSI/AAMI 278 EC53), DISPOSABLE ECG ELECTRODES (ANSI/AAMI EC12), and cardiac rhythm and ST 279 segment measurement algorithms (ANSI/AAMI EC57).
280 The requirements of this particular standard take priority over those of the general standard. 281 A “General guidance and rationale” for the more important requirements of this particular 282 standard is included in Annex AA. It is considered that knowledge of the reasons for these 283 requirements will not only facilitate the proper application of the standard but will, in due course, 284 expedite any revision necessitated by changes in clinical practice or as a result of developments 285 in technology. However, Annex AA does not form part of the requirements of this standard.
287 MEDICAL ELECTRICAL EQUIPMENT – 288 289 Part 2-86: Particular requirements for the basic safety and essential 290 performance of electrocardiographic equipment, including diagnostic 291 equipment, monitoring equipment, ambulatory equipment, electrodes, 292 cables and lead wires 293 Scope, objective and related standards 1) 294 Clause 1 of the general standard applies, except as follows:
295 201.1.1 * Scope 296 Replacement:
297 This particular standard applies to BASIC SAFETY and ESSENTIAL PERFORMANCE of ECG ME 298 EQUIPMENT and ACCESSORIES (including ECG ELECTRODES, TRUNK CABLES, and LEAD WIRES) as 299 defined in 201.3.211, 201.3.240, and 201.3.222, hereinafter also referred to as ECG ME 300 EQUIPMENT. Additional specific requirements apply to the ECG ME EQUIPMENT based on the 301 INTENDED USE claimed by the MANUFACTURER for equipment intended for ECG PATIENT monitoring, 302 diagnostic, or ambulatory use as defined in 201.3.226, 201.3.207 and 201.3.202 respectively. 303 This particular standard applies to ECG ME EQUIPMENT for use in professional healthcare facilities 304 as well as in EMERGENCY MEDICAL SERVICE ENVIRONMENTS and HOME HEALTHCARE ENVIRONMENTS. 305 In addition, clauses may be applied to ACCESSORIES (including ECG ELECTRODES, TRUNK CABLES, 306 and LEAD WIRES) or a part or parts of an ECG system (e.g. AMBULATORY ECG RECORDER). 307 The scope for each of these types of ECG ME EQUIPMENT are described as follows:
308 ECG ME EQUIPMENT Any ECG device or system or part of a system, including ECG ELECTRODEs, 309 TRUNK CABLEs, and LEAD WIREs and interconnecting means, that acquires an ECG signal from the 310 body surface of one PATIENT and/or displays, records, analyses, or transmits the resultant data 311 for the purpose of diagnosing, treating, or monitoring that PATIENT. This includes any device or 312 system or part of a system that meets the definition of MONITORING, DIAGNOSTIC, or AMBULATORY 313 ECG ME EQUIPMENT as well as any other device that acquires and/or processes an ECG signal for 314 the purpose of diagnosing, treating or monitoring a PATIENT based on the acquired body surface 315 ECG signal. This does not include devices that are used for diagnosing, treating or monitoring 316 PATIENTS based only on other types of cardiac signals such as magnetocardiographic signals 317 or derived from other cardiac signals such as heart rate monitors derived from a photo 318 plethysmograph or phonocardiograph. In addition, ECG ME EQUIPMENT does not include devices 319 that are solely used for health and wellness (e.g. fitness trackers measuring heart rate solely 320 to improve general wellness or athletic performance). However, devices that are part of 321 consumer devices (such as watches, patches, body-worn devices and similar devices) that 322 acquired an ECG and displays, records, analyses, and/or transmits the ECG for the purpose of 323 diagnosing, treating or monitoring the PATIENT are included in the scope of an ECG ME 324 EQUIPMENT. This does not include external defibrillators unless they have an INTENDED USE for 325 making a diagnosis, monitoring or treatment from the ECG waveform in addition to the function 326 of delivering a defibrillation shock based on the ECG.
327 This does not include direct fetal ECG equipment and Fetal & Maternal MONITORING EQUIPMENT 329 Note: Fetal & Maternal Monitor: A Fetal-Maternal monitor that non-invasively measures and displays Fetal heart rate 330 (FHR), Maternal heart rate (MHR) and uterine activity (UA). The equipment acquires and displays the FHR & MHR 331 tracing from abdominal surface ELECTRODES that pick up the Fetal ECG (FECG) and Maternal ECG (MECG) signal. 332 Note: while this standard applies to certain EQUIPMENT, manufacturers may consider relevant requirements of this 333 standards for testing similar equipment not in scope.
334 DIAGNOSTIC ECG ME EQUIPMENT 2 335 Any ECG ME EQUIPMENT that acquires and/or displays, records, analyzes, or transmits a 12 LEAD 336 ECG signal for the purpose of diagnosis from the interpretation of the cardiac rhythm, conduction, 337 and waveform contour. This includes any device that acquires diagnostic resting ECG as well as 338 any other type of ECG ME EQUIPMENT that claims to have the capability of acquiring a diagnostic 339 12 LEAD ECG. Devices which have multiple purposes and include the DIAGNOSTIC ECG ME EQUIPMENT 340 definition in the INTENDED USE are included within this scope.
341 AMBULATORY ECG ME EQUIPMENT 342 Any ECG ME EQUIPMENT that acquires and/or displays, records, analyzes, or transmits an ECG 343 signal from an ambulatory PATIENT. Examples of AMBULATORY ECG ME EQUIPMENT include Holter 344 systems and continuous long-term ambulatory full disclosure recording systems; Holter 345 recorders, mobile cardiac telemetry and long-term ambulatory patch recorders, event recorders, 346 and non-continuous loop recorders. Devices which have multiple purposes and include the 347 AMBULATORY ECG ME EQUIPMENT definition in the INTENDED USE are included within this scope. 348 MONITORING ECG ME EQUIPMENT 349 Any ECG ME EQUIPMENT that acquires and/or displays, records, analyzes, or transmits an ECG for the 350 purpose of monitoring the cardiac activity of one PATIENT and devices used as part of an ALARM SYSTEM. 351 This includes displaying ECG, displaying measurements such as heart rate, and generating alarms based 352 on analysis of the ECG. It also includes devices where ECG or ECG-derived information device is 353 intended to be relied upon in deciding to take immediate clinical action. Examples of MONITORING ECG 354 EQUIPMENT include ECG and hardwired bedside patient monitors, telemetry systems, and other system 355 used to identify of patient conditions requiring intervention based on manual and/or automated analysis 356 of the ECG signal.
357 ECG TRUNK CABLES and PATIENT LEAD WIRES 358 ECG TRUNK CABLES and PATIENT LEAD WIRES that are used in applications that require special 359 characteristics, such as a magnetic resonance imaging (MRI) suite, are excluded from this standard. 360 However, non-MRI ECG TRUNK CABLES and PATIENT LEAD WIRES must meet all of the requirements of this 361 standard, unless a requirement is specifically excluded.
362 ECG ELECTRODES 363 This standard establishes minimum labeling, safety, and performance requirements for disposable ECG 364 ELECTRODES.
365 Included within the scope of this standard is any disposable ECG ELECTRODE SYSTEM 366 ELECTRODES excluded from the scope of this standard are active ELECTRODES, needle ELECTRODES, 367 reusable (non-disposable) ELECTRODES, ELECTRODES intended to deliver therapeutic energy, and 368 ELECTRODES intended solely for the measurement of physiologic signals other than the 369 ELECTROCARDIOGRAM (e.g., ELECTRODES used with apnea monitors, if the ELECTRODE is solely used for 2
Unless otherwise specified, the term 12 LEAD indicates 12 or more LEADs, including at least the limb LEADs and precordial LEADs, as defined elsewhere in this document.
370 non-ECG purposes, e.g., impedance plethysmography). Also, requirements for electrolyte composition 371 are not covered by this standard.
372 201.1.2 Object 373 Replacement:
374 The object of this particular standard is to establish BASIC SAFETY and ESSENTIAL PERFORMANCE 375 requirements for ECG ME EQUIPMENT.
376 201.1.2.101 * Organization of this particular standard 377 The requirements in this particular standard are organized into the following types of 378 requirements:
379 1) General requirements – all requirements are considered general requirements and apply 380 to all types of ECG ME EQUIPMENT unless explicitly noted.
381 2) Diagnostic requirements specific to DIAGNOSTIC ECG ME EQUIPMENT will be explicitly 382 identified as such.
383 3) Monitoring requirements specific to MONITORING ECG ME EQUIPMENT will be explicitly 384 identified as such.
385 4) Ambulatory requirements specific to AMBULATORY ECG ME EQUIPMENT will be explicitly 386 identified as such.
387 Additionally, the following requirements have been added to this particular standard:
388 • ECG TRUNK CABLE and LEAD WIREs, as in ANSI/AAMI EC53. These requirements apply 389 to any ECG ME EQUIPMENT, including MONITORING, DIAGNOSTIC, AND AMBULATORY ECG 390 ME EQUIPMENT.
391 • DISPOSABLE ECG ELECTRODE, as in ANSI/AAMI EC12. These requirements apply to 392 any ECG ME EQUIPMENT that includes DISPOSABLE ECG ELECTRODES as an APPLIED PART. 393 • Requirements from the last edition of 60601-2-51 that were omitted when it was 394 combined with 60601-2-25.
395 A summary of requirements for compliance to this standard for ECG ME EQUIPMENT are listed in 396 Table AA.1 – AA.4 in Annex AA.
397 201.1.3 Collateral standards 398 Addition:
399 This particular standard refers to those applicable collateral standards that are listed in Clause 400 2 of the general standard and Clause 201.2 of this particular standard.
401 IEC 60601-1-2:2014 and IEC 60601-1-2:2014/AMD1:2020, and IEC 60601-1-8:2006, IEC 402 60601-1-8:2006/AMD1:2012 and IEC 60601-1-8:2006/AMD2:2020 apply as modified in Clauses 403 202 and 208 respectively. All other published collateral standards in the IEC 60601-1 series 404 apply as published.
405 201.1.4 Particular standards 406 Replacement:
407 In the IEC 60601 series, particular standards may modify, replace or delete requirements 408 contained in the general standard and collateral standards as appropriate for the particular ECG 409 ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL 410 PERFORMANCE requirements.
411 A requirement of a particular standard takes priority over the general standard. 412 For brevity, IEC 60601-1 is referred to in this particular standard as the general standard. 413 Collateral standards are referred to by their document number.
414 The numbering of clauses and subclauses of this particular standard corresponds to that of the 415 general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content of 416 Clause 1 of the general standard) or applicable collateral standard with the prefix “20x” where 417 x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this particular 418 standard addresses the content of Clause 4 of the IEC 60601-1-2 collateral standard, 203.4 in 419 this particular standard addresses the content of Clause 4 of the IEC 60601-1-3 collateral 420 standard, etc.). The changes to the text of the general standard are specified by the use of the 421 following words:
422 “Replacement” means that the clause or subclause of the general standard or applicable 423 collateral standard is replaced completely by the text of this particular standard. 424 “Addition” means that the text of this particular standard is additional to the requirements of the 425 general standard or applicable collateral standard.
426 “Amendment” means that the clause or subclause of the general standard or applicable 427 collateral standard is amended as indicated by the text of this particular standard. 428 Subclauses, Figures or Tables which are additional to those of the general standard are 429 numbered starting from 201.101. However, because definitions in the general standard are 430 numbered 3.1 through 3.139, additional definitions in this standard are numbered beginning 431 from 201.3.201. Additional Annexes are lettered AA, BB, etc., and additional items aa), bb), etc. 432 Subclauses, Figures or Tables which are additional to those of a collateral standard are 433 numbered starting from 20x, where “x” is the final digit(s) of the collateral standard, e.g. 202 for 434 IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
435 The term “this standard” is used to reference to the general standard, any applicable collateral 436 standards and this particular standard taken together.
437 Where there is no corresponding clause or subclause in this particular standard, the clause or 438 subclause of the general standard or applicable collateral standard, although possibly not 439 relevant, applies without modification; where it is intended that any part of the general standard 440 or applicable collateral standard, although possibly relevant, is not to be applied, a statement 441 to that effect is given in this particular standard.
442 Normative references 443 Clause 2 of the general standard applies, except as follows:
444 Replacement:
445 IEC 60601-1-2:2014 and IEC 60601-1-2:2014/AMD1:2020, Medical electrical equipment – Part 446 1-2: General requirements for basic safety and ESSENTIAL PERFORMANCE – Collateral standard:
447 Electromagnetic disturbances – Requirements and tests 448 IEC 60601-1-8:2006, IEC 60601-1-8:2006/AMD1:2012 and IEC 60601-1-8:2006/AMD2:2020, 449 Medical electrical equipment – Part 1-8: General requirements for basic s
...