IEC 60601-2-54:2009
(Main)Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
IEC 60601-2-54:2009 applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems intended to be used for projection radiography and radioscopy. The minimum safety requirements specified in this particular standard are considered to provide for a practical degree of safety in the operation of ME equipment for radiography and radioscopy. Requirements for additional provisions for ME equipment for interventional applications are covered by IEC 60601-2-43. IEC 60601-2-54:2009 has been developed for use with the third edition of IEC 60601-1 (2005) and its collateral standards. IEC 60601-2-54:2009 is about type testing of X-ray systems. For the type test of a given product only one particular standard applies. Thus, IEC 60601-2-28 cannot be used in the type test according to IEC 60601-2-54:2009. The system aspects of the X-ray source integrated into X-ray systems are covered by IEC 60601-2-54:2009. IEC 60601-2-32 and IEC 60601-2-7, which had been written for use with the second edition of IEC 60601-1 (1988), cannot be used with the third edition of IEC 60601-1 (2005) and IEC 60601-2-54. The contents of the corrigenda of March 2010 and June 2011 have been included in this copy.
Appareils électromédicaux - Partie 2-54: Exigences particulières pour la sécurité de base et les performances essentielles des appareils à rayonnement X utilisés pour la radiographie et la radioscopie
La CEI 60601-2-54:2009 s'applique à la sécurité de base et aux performances essentielles des appareils electromédicaux et des systèmes electromédicaux destinés à la radiographie de projection et à la radioscopie. Les exigences de sécurité minimales spécifiées dans la présente norme particulière sont estimées assurer un degré de sécurité réalisable dans le cadre du fonctionnement des appareils EM utilisés pour la radiographie et la radioscopie. Des dispositions supplémentaires concernant les procédures d'intervention applicables aux appareils EM sont spécifiées dans la CEI 60601-2-43. La CEI 60601-2-54:2009 a été élaborée en vue d'une utilisation avec la troisième édition de la CEI 60601-1 (2005) et ses normes collatérales. Le contenu des corrigendum de mars 2010 et juin 2011 a été pris en considération dans cet exemplaire.
General Information
Relations
Standards Content (Sample)
IEC 60601-2-54 ®
Edition 1.0 2009-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X utilisés pour la radiographie et la
radioscopie
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by
any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either IEC or
IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.
IEC Central Office
3, rue de Varembé
CH-1211 Geneva 20
Switzerland
Email: inmail@iec.ch
Web: www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
ƒ Catalogue of IEC publications: www.iec.ch/searchpub
The IEC on-line Catalogue enables you to search by a variety of criteria (reference number, text, technical committee,…).
It also gives information on projects, withdrawn and replaced publications.
ƒ IEC Just Published: www.iec.ch/online_news/justpub
Stay up to date on all new IEC publications. Just Published details twice a month all new publications released. Available
on-line and also by email.
ƒ Electropedia: www.electropedia.org
The world's leading online dictionary of electronic and electrical terms containing more than 20 000 terms and definitions
in English and French, with equivalent terms in additional languages. Also known as the International Electrotechnical
Vocabulary online.
ƒ Customer Service Centre: www.iec.ch/webstore/custserv
If you wish to give us your feedback on this publication or need further assistance, please visit the Customer Service
Centre FAQ or contact us:
Email: csc@iec.ch
Tel.: +41 22 919 02 11
Fax: +41 22 919 03 00
A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.
ƒ Catalogue des publications de la CEI: www.iec.ch/searchpub/cur_fut-f.htm
Le Catalogue en-ligne de la CEI vous permet d’effectuer des recherches en utilisant différents critères (numéro de référence,
texte, comité d’études,…). Il donne aussi des informations sur les projets et les publications retirées ou remplacées.
ƒ Just Published CEI: www.iec.ch/online_news/justpub
Restez informé sur les nouvelles publications de la CEI. Just Published détaille deux fois par mois les nouvelles
publications parues. Disponible en-ligne et aussi par email.
ƒ Electropedia: www.electropedia.org
Le premier dictionnaire en ligne au monde de termes électroniques et électriques. Il contient plus de 20 000 termes et
définitions en anglais et en français, ainsi que les termes équivalents dans les langues additionnelles. Egalement appelé
Vocabulaire Electrotechnique International en ligne.
ƒ Service Clients: www.iec.ch/webstore/custserv/custserv_entry-f.htm
Si vous désirez nous donner des commentaires sur cette publication ou si vous avez des questions, visitez le FAQ du
Service clients ou contactez-nous:
Email: csc@iec.ch
Tél.: +41 22 919 02 11
Fax: +41 22 919 03 00
IEC 60601-2-54 ®
Edition 1.0 2009-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X utilisés pour la radiographie et la
radioscopie
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
XB
CODE PRIX
ICS 11.040.50 ISBN 978-2-88910-227-3
– 2 – 60601-2-54 © IEC:2009
CONTENTS
FOREWORD.4
INTRODUCTION.6
201.1 Scope, object and related standards .7
201.2 Normative references.8
201.3 Terms and definitions.9
201.4 General requirements .10
201.5 General requirements for testing of ME EQUIPMENT .11
201.6 Classification of ME EQUIPMENT and ME SYSTEMS.11
201.7 ME EQUIPMENT identification, marking and documents .11
201.8 Protection against electrical HAZARDS from ME EQUIPMENT .14
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS .17
201.10 Protection against unwanted and excessive radiation HAZARDS .21
201.11 Protection against excessive temperatures and other HAZARDS .21
201.12 Accuracy of controls and instruments and protection against hazardous
outputs .22
201.13 HAZARDOUS SITUATIONS and fault conditions .22
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS).22
201.15 Construction of ME EQUIPMENT .22
201.16 ME SYSTEMS .22
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS .22
202 Electromagnetic compatibility – Requirements and tests .22
203 Radiation protection in diagnostic X-ray equipment.23
Annexes .58
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS.59
Annex AA (informative) Particular guidance and rationale.61
Bibliography.64
Index of defined terms used in this particular standard.65
Figure 203.101 – Zone of EXTRA-FOCAL RADIATION .41
Figure 203.102 – Discrepancies in covering the IMAGE RECEPTION AREA.43
Figure 203.103 – Discrepancies in visual indication of the X-RAY FIELD .47
Figure 203.104 – Testing for STRAY RADIATION (X-RAY BEAM horizontal with X-RAY
SOURCE ASSEMBLY below the PATIENT SUPPORT) .55
Figure 203.105 – Testing for STRAY RADIATION (X-RAY BEAM vertical with X-RAY SOURCE
ASSEMBLY below the PATIENT SUPPORT) .55
Figure 203.106 – Testing for STRAY RADIATION (X-RAY BEAM horizontal with X-RAY
SOURCE ASSEMBLY above the PATIENT SUPPORT) .56
Figure 203.107 – Testing for STRAY RADIATION (X-RAY BEAM vertical with X-RAY SOURCE
ASSEMBLY above the PATIENT SUPPORT).57
Table 201.101 – Distributed ESSENTIAL PERFORMANCE requirements .10
Table 203.101 – Tests for verifying reproducibility and linearity .29
60601-2-54 © IEC:2009 – 3 –
Table 203.102 – Loadings for testing AUTOMATIC EXPOSURE CONTROLS.31
Table 203.103 – ATTENUATION for the measurement of AIR KERMA .33
Table 203.104 – ATTENUATION EQUIVALENT of items in the X-RAY BEAM.49
Table 203.105 – Application categories .51
Table 203.106 – Requirements for PRIMARY PROTECTIVE SHIELDING .51
Table 203.107 – STRAY RADIATION in SIGNIFICANT ZONES OF OCCUPANCY .53
Table 201.C.101 – Marking on the outside of ME EQUIPMENT or its parts .59
Table 201.C.102 – Subclauses requiring statements in ACCOMPANYING DOCUMENTS.59
– 4 – 60601-2-54 © IEC:2009
INTERNATIONAL ELECTROTECHNICAL COMMISSION
_____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-2-54 has been prepared by IEC subcommittee 62B:
Diagnostic imaging equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
IEC 60601-2-54 has been developed for use with the third edition of IEC 60601-1 (2005). It
replaces and supersedes IEC 60601-2-7 and IEC 60601-2-32, as well as IEC 60601-2-
28:1993 (currently under revision), all of which were developed to amend earlier editions of
IEC 60601-1 and consequently no longer apply to this particular standard.
The text of this particular standard is based on the following documents:
FDIS Report on voting
62B/735/FDIS 62B/750/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.
60601-2-54 © IEC:2009 – 5 –
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
ERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
– T
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
The contents of the corrigenda of March 2010 and June 2011 have been included in this
copy.
– 6 – 60601-2-54 © IEC:2009
INTRODUCTION
This particular standard has been prepared to provide, based on IEC 60601-1:2005 (third
edition) and its collaterals, a complete set of safety requirements for ME EQUIPMENT for
RADIOGRAPHY and RADIOSCOPY. While the previously existing standards for such equipment
were dedicated to components and subsystems, this particular standard addresses the
system level of X-RAY EQUIPMENT, which consists of a combination of an X-RAY GENERATOR,
ASSOCIATED EQUIPMENT and ACCESSORIES. Component functions are addressed as far as
necessary.
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of ME EQUIPMENT for RADIOGRAPHY and
RADIOSCOPY. Requirements for additional provisions for ME EQUIPMENT for interventional
applications are covered by IEC 60601-2-43.
60601-2-54 © IEC:2009 – 7 –
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
201.1 Scope, object and related standards
1)
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
ME EQUIPMENT and ME SYSTEMS intended to be used for projection RADIOGRAPHY and
RADIOSCOPY. IEC 60601-2-43 applies to ME EQUIPMENT and ME SYSTEMS intended to be used
for interventional applications and refers to applicable requirements in this particular
standard.
ME EQUIPMENT and ME SYSTEMS intended to be used for bone or tissue absorption
densitometry, computed tomography, mammography or dental applications are excluded from
the scope of this International Standard. The scope of this International Standard also
excludes radiotherapy simulators.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
NOTE Taking into account economic and social factors, the scope of this particular standard includes
ME EQUIPMENT intended to be used for DIRECT RADIOSCOPY. In some countries examinations performed with
DIRECT RADIOSCOPY are prohibited.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for ME EQUIPMENT and ME SYSTEMS for RADIOGRAPHY and
RADIOSCOPY.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-3 apply as modified in Clauses 202 and 203 respectively.
IEC 60601-1-8 and IEC 60601-1-10 do not apply. All other published collateral standards in
the IEC 60601-1 series apply as published.
—————————
1)
The general standard is IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance
– 8 – 60601-2-54 © IEC:2009
NOTE OPERATORS of X-RAY EQUIPMENT are used to audible signals as required in this particular standard rather
than to the concepts of IEC 60601-1-8. Therefore IEC 60601-1-8 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography on page 64.
60601-2-54 © IEC:2009 – 9 –
Clause 2 of the general standard applies, except as follows:
Addition:
IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis –
Characteristics of focal spots
IEC 60580:2000, Medical electrical equipment – Dose area product meters
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC 60806, Determination of the maximum symmetrical radiation field from a rotating anode
X-ray tube for medical diagnosis
IEC 62220-1:2003, Medical electrical equipment – Characteristics of digital X-ray imaging
devices – Part 1: Determination of the detective quantum efficiency
Amendment:
IEC 60601-1-2:2007 Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance– Collateral standard: Electromagnetic compatibility –
Requirements and tests
IEC 60601-1-3:2008 Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic X-
ray equipment
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005,
applicable collateral standards and IEC 60788:2004 apply, except as follows:
NOTE An index of defined terms is found beginning on page 65.
Addition:
201.3.201
DIRECT RADIOGRAPHY
RADIOGRAPHY in which the permanent recording is effected at an IMAGE RECEPTION AREA
Example: film-screen or film radiography.
201.3.202
DIRECT RADIOSCOPY
RADIOSCOPY in which the visible images are presented at the IMAGE RECEPTION AREA, or close
to it, in the RADIATION BEAM
201.3.203
DOSE AREA PRODUCT
product of the area of the cross-section of an X-RAY BEAM and the averaged AIR KERMA over
that cross-section. The unit is the gray square metre (Gy⋅m )
201.3.204
ENTRANCE FIELD SIZE
dimensions of the field in the entrance plane of an X-RAY IMAGE RECEPTOR that can be used for
the transmission of an X-RAY PATTERN under specific conditions
– 10 – 60601-2-54 © IEC:2009
201.3.205
INDIRECT RADIOGRAPHY
RADIOGRAPHY in which the permanent recording is effected after TRANSFER of the information
obtained at an IMAGE RECEPTION AREA
Examples: CR systems, digital detector systems, image intensifier systems.
201.3.206
INDIRECT RADIOSCOPY
RADIOSCOPY in which the images are presented at a location outside the RADIATION BEAM after
TRANSFER of the information
201.3.207
INTERLOCK
means preventing the start or the continued operation of ME EQUIPMENT unless certain
predetermined conditions prevail
201.3.208
NOMINAL SHORTEST IRRADIATION TIME
shortest LOADING TIME for which a required constancy of the controlled RADIATION QUANTIY is
maintained
NOTE The IRRADIATION TIME is controlled by a HIGH-VOLTAGE GENERATOR with AUTOMATIC CONTROL SYSTEMS.
201.3.209
SERIAL RADIOGRAPHY
RADIOGRAPHY in which the information is obtained and recorded in a regular or irregular series
of LOADINGS with equal or unequal LOADING FACTORS
201.4 General requirements
Clause 4 of the general standard applies, except as follows
201.4.3 ESSENTIAL PERFORMANCE
Additional subclause:
201.4.3.101 * Additional ESSENTIAL PERFORMANCE requirements
Additional ESSENTIAL PERFORMANCE requirements are found in the subclauses listed in Table
201.101.
Table 201.101 – Distributed ESSENTIAL PERFORMANCE requirements
Requirement Subclause
Accuracy of LOADING FACTORS 203.6.4.3.104
Reproducibility of the RADIATION output 203.6.3.2
AUTOMATIC CONTROL SYSTEM 203.6.5
Imaging performance 203.6.7
201.4.10.2 Supply mains for ME EQUIPMENT and ME SYSTEMS
Addition:
60601-2-54 © IEC:2009 – 11 –
The internal impedance of a SUPPLY MAINS is to be considered sufficiently low for the operation
of X-RAY EQUIPMENT for RADIOGRAPHY and RADIOSCOPY if the value of the APPARENT RESISTANCE
OF SUPPLY MAINS does not exceed the value specified in the ACCOMPANYING DOCUMENTS.
Either the APPARENT RESISTANCE OF SUPPLY MAINS or other appropriate SUPPLY MAINS
specifications used in a facility shall be specified in the ACCOMPANYING DOCUMENTS.
NOTE If a NOMINAL voltage is claimed for a mains power supply system, it is assumed that there is no voltage of a
higher value between any of the conductors of the system or between any of these conductors and earth.
An alternating voltage is considered in practice to be sinusoidal if any instantaneous value of the waveform
concerned differs from the instantaneous value of the ideal waveform at the same moment by no more than ± 2 %
of the peak value of the ideal waveform.
A three-phase SUPPLY MAINS is considered to have a practical symmetry if it delivers symmetrical voltages and
produces, when loaded symmetrically, symmetrical currents.
The requirements of this standard are based upon the assumption that three-phase systems have a symmetrical
configuration of the MAINS VOLTAGE with respect to earth. Single-phase systems may be derived from such
three-phase systems. Where the supply system is not earthed at the source it is assumed that adequate measures
have been provided to detect, limit and remedy any disturbance of symmetry within a reasonably short time.
X-RAY EQUIPMENT is considered to comply with the requirements of this standard only if its specified NOMINAL
ELECTRIC POWER can be demonstrated at an APPARENT RESISTANCE OF SUPPLY MAINS having a value not less than
the APPARENT RESISTANCE OF SUPPLY MAINS specified by the MANUFACTURER in the ACCOMPANYING DOCUMENTS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
201.7.2.7 Electrical input power from the SUPPLY MAINS
Addition:
For ME EQUIPMENT that is specified to be PERMANENTLY INSTALLED, the information may be
stated in the ACCOMPANYING DOCUMENTS only.
The information on the input power shall be specified in terms of combinations of
a) the RATED MAINS VOLTAGE of the ME EQUIPMENT in volts; see 7.2.1 and 7.2.6 of the general
standard,
b) the number of phases; see 7.2.1 and 7.2.6 of the general standard,
c) the frequency, in hertz; see 7.2.1 and 7.2.6 of the general standard,
d) the maximum permissible value for APPARENT RESISTANCE OF SUPPLY MAINS, in ohms;
e) the characteristics of OVER-CURRENT RELEASES required in the SUPPLY MAINS.
NOTE These requirements are adapted from 6.1j) of IEC 60601-2-7:1998.
– 12 – 60601-2-54 © IEC:2009
201.7.2.15 Cooling conditions
Addition:
If cooling is necessary for safe operation of ME EQUIPMENT, or a subassembly thereof, the
cooling requirements shall be indicated in the ACCOMPANYING DOCUMENT, including as
appropriate:
– the maximum heat dissipation into the surrounding air, given separately for each
subassembly that dissipates more than 100 W and might be separately located on
installation;
– the maximum heat dissipation into forced air cooling devices, and the corresponding flow
rate and temperature rise of the forced air stream;
– the maximum heat dissipation into a cooling medium utility and the permissible input
temperature range, minimum flow rate and pressure requirements for the utility.
NOTE These requirements are adapted from 6.1t) of IEC 60601-2-7:1998.
Additional subclause:
201.7.2.101 Beam limiting device
BEAM LIMITING DEVICES shall be provided with the following markings:
– those required in subclause 7.2.2 of the general standard;
– serial designation or individual identification;
TOTAL FILTRATION in terms of QUALITY EQUIVALENT FILTRATION.
–
NOTE These requirements are adapted from 6.1 of IEC 60601-2-28:1993.
201.7.8.1 Colours of indicator lights
Addition:
The indication of X-RAY related states shall be excluded from subclause 7.8 in the general
standard. Subclauses 203.6.4.2 and 203.6.4.101 shall apply instead.
201.7.9 ACCOMPANYING DOCUMENTS
201.7.9.1 General
Addition:
The ACCOMPANYING DOCUMENTS shall contain quality control procedures to be performed on
the X-RAY EQUIPMENT by the RESPONSIBLE ORGANISATION. These shall include acceptance
criteria and frequency for the tests.
Additionally for X-RAY EQUIPMENT provided with an integrated digital X-RAY IMAGE RECEPTOR,
the ACCOMPANYING DOCUMENTS shall contain:
– a description of image processing applied to ORIGINAL DATA including the revision number
or how to determine it and identification of the version if applicable;
– a description of the file transfer format of the images acquired with this unit and of any
data associated with these images;
The performance of means required to present the images for diagnostic purpose shall be
stated according to the INTENDED USE.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
60601-2-54 © IEC:2009 – 13 –
201.7.9.2 Instructions for use
201.7.9.2.1 General
Addition:
201.7.9.2.1.101 LOADING FACTORS
In the instructions for use the LOADING FACTORS shall be stated as described below. The
following combinations and data shall be stated:
a) The corresponding NOMINAL X-RAY TUBE VOLTAGE for RADIOSCOPY and RADIOGRAPHY
together with the highest X-RAY TUBE CURRENT obtainable from the ME EQUIPMENT when
operated at that X-RAY TUBE VOLTAGE.
b) The corresponding highest X-RAY TUBE CURRENT for RADIOSCOPY and RADIOGRAPHY
together with the highest X-RAY TUBE VOLTAGE obtainable from the ME EQUIPMENT when
operating at that X-RAY TUBE CURRENT.
c) The corresponding combination of X-RAY TUBE VOLTAGE for RADIOSCOPY and RADIOGRAPHY,
and X-RAY TUBE CURRENT which results in the highest electric power in the high-voltage
circuit (see 203.4.101).
d) The NOMINAL ELECTRIC POWER given as the highest constant electric power in kilowatts
which the ME EQUIPMENT can produce/generate, for a LOADING TIME of 0,1 s at an X-RAY
TUBE VOLTAGE of 100 kV or, if these values are not selectable, with nearest parameters
(see 203.4.101).
The NOMINAL ELECTRIC POWER shall be given together with the combination of X-RAY TUBE
VOLTAGE and X-RAY TUBE CURRENT and the LOADING TIME.
e) For ME EQUIPMENT indicating precalculated or measured CURRENT TIME PRODUCT, the
lowest CURRENT TIME PRODUCT or the combinations of LOADING FACTORS resulting in the
lowest CURRENT TIME PRODUCT.
If the value of the lowest CURRENT TIME PRODUCT depends upon the X-RAY TUBE VOLTAGE or
upon certain combinations of values of LOADING FACTORS, the lowest CURRENT TIME
PRODUCT may be given as a table or curve showing the dependence.
f) The NOMINAL SHORTEST IRRADIATION TIME used in AUTOMATIC EXPOSURE CONTROL systems of
ME EQUIPMENT.
If the NOMINAL SHORTEST IRRADIATION TIME depends upon LOADING FACTORS such as X-RAY
TUBE VOLTAGE and X-RAY TUBE CURRENT, the ranges of these LOADING FACTORS for which
the NOMINAL SHORTEST IRRADIATION TIME is valid shall be stated.
The maximum possible range of the X-RAY TUBE VOLTAGE and/or the X-RAY TUBE CURRENT
during IRRADIATIONS, controlled with the AUTOMATIC EXPOSURE CONTROL SYSTEMS, shall be
stated in the instructions for use.
NOTE These requirements are adapted from 6.8.2 a) of IEC 60601-2-7:1998.
201.7.9.2.1.102 X-RAY SOURCE ASSEMBLY
The instructions for use shall state the maximum symmetrical RADIATION FIELD of the
integrated X-RAY SOURCE ASSEMBLY determined according to IEC 60806.
NOTE This requirement is adapted from 6.8.2 (dd) of IEC 60601-2-28:1993.
201.7.9.2.1.103 Integrated X-RAY IMAGE RECEPTOR
For X-RAY EQUIPMENT provided with an integrated X-RAY IMAGE RECEPTOR, the instructions for
use shall contain a description of the particular handling and maintenance of the X-RAY IMAGE
RECEPTOR.
Compliance is checked by inspection of the instructions for use.
– 14 – 60601-2-54 © IEC:2009
201.7.9.3 Technical description
Additional subclause:
201.7.9.3.101 X-RAY SOURCE ASSEMBLY
The technical description of the integrated X-RAY SOURCE ASSEMBLIES shall specify the
following, in addition to the data required to be marked according to subclause 7.2 of the
general standard:
a) specification of the REFERENCE AXIS to which the TARGET ANGLE(s) and the FOCAL SPOT
characteristics of the X-RAY SOURCE ASSEMBLY refer:
b) TARGET ANGLE(s) with respect to the specified REFERENCE AXIS;
c) position of the FOCAL SPOT and its tolerances on the REFERENCE AXIS;
d) NOMINAL FOCAL SPOT VALUE(s) determined according to IEC 60336 for the specified
REFERENCE AXIS.
NOTE These requirements are adapted from 6.8.3 dd) of IEC 60601-2-28:1993.
Additional subclause:
201.7.9.101 Additional statements in ACCOMPANYING DOCUMENTS
Additional requirements for statements in ACCOMPANYING DOCUMENTS (which include
instructions for use and technical description) are found in the subclauses listed in Table
201.C.102 of Annex C.
201.8 Protection against electrical HAZARDS from ME EQUIPMENT
Clause 8 of the general standard applies, except as follows:
201.8.4 Limitation of voltage, current or energy
201.8.4.3 ME EQUIPMENT intended to be connected to a power source by a plug
Addition:
201.8.4.3.101 HIGH-VOLTAGE CABLE CONNECTIONS
Detachable HIGH-VOLTAGE CABLE CONNECTIONS shall either be designed so that the use of
TOOLS is required to disconnect them or they shall be provided with INTERLOCKS so that at all
times when protective covers or HIGH-VOLTAGE CONNECTIONS are removed:
– the ME EQUIPMENT is disconnected from its power supply, and
– capacitances in the high-voltage circuit are discharged within the minimum time necessary
to gain access to the high-voltage circuit, and
– the discharged state is maintained.
Compliance is checked by inspection and by measurement.
NOTE These requirements are adapted from Clause 15, item aa) of IEC 60601-2-7:1998.
Additional subclause:
201.8.4.101 Limitation of X-RAY TUBE VOLTAGE
ME EQUIPMENT shall be designed so as not to deliver in INTENDED USE, to any connected X-RAY
TUBE ASSEMBLY, a voltage greater than the NOMINAL X-RAY TUBE VOLTAGE for the X-RAY TUBE
60601-2-54 © IEC:2009 – 15 –
concerned or greater than the NOMINAL X-RAY TUBE VOLTAGE the X-RAY TUBE ASSEMBLY is
designed for, whichever is the lower voltage.
NOTE This requirement is adapted from 3.1 of IEC 60601-2-7:1998.
201.8.5 Separation of parts
201.8.5.1 MEANS OF PROTECTION (MOP)
Additional subclause:
201.8.5.1.101 Additional limitation of voltage, current or energy
Provision shall be made to prevent the appearance of an unacceptably high voltage in the
MAINS PART or in any other low-voltage circuit.
NOTE This may be achieved for example
– by provision of a winding layer or a conductive screen connected to the PROTECTIVE EARTH TERMINAL between
high-voltage and low-voltage circuits;
– by provision of a voltage limiting device across terminals to which external devices are connected and between
which an excessive voltage might arise if the external path becomes discontinuous.
Compliance is checked by inspection of design data and construction.
NOTE These requirements are adapted from 15bb) of IEC 60601-2-7:1998.
201.8.5.4 WORKING VOLTAGE
Addition:
201.8.5.4.101 Stator and stator circuit dielectric strength testing
The test voltage for the dielectric strength testing of stator and stator circuit used for the
operation of the rotating anode of the X-RAY TUBE is to be referred to the voltage existing after
reduction of the stator supply voltage to its steady state operating value.
NOTE This requirement is adapted from 20.4 l) of IEC 60601-2-7:1998.
201.8.6 Protective earthing, functional earthing and potential equalization of ME
EQUIPMENT
201.8.6.4 Impedance and current-carrying capability
Addition:
The flexible conductive screen is not to be recognized as satisfying a requirement for a
PROTECTIVE EARTH CONNECTION between the devices connected by the cable.
Additional subclause:
201.8.6.101 X-RAY TUBE ASSEMBLY
a) Accessible high-voltage cables connecting x-RAY TUBE ASSEMBLIES to their associated
HIGH-VOLTAGE GENERATOR shall incorporate a flexible conductive screen, having a
-1
resistance per unit length not exceeding 1 Ω m , and covered with a non-conductive
material capable of protecting the screen against mechanical damage. The screen shall
be connected to the conductive ENCLOSURE of the HIGH-VOLTAGE GENERATOR.
Compliance is checked by visual inspection and by measurement.
– 16 – 60601-2-54 © IEC:2009
NOTE These requirements are adopted from Clause 16, item aa) of IEC 60601-2-28:1993.
b) In all cases, there shall be electrical continuity between the screen of a fitted high-voltage
cable and the ACCESSIBLE METAL PARTS of its receptacle on the X-RAY TUBE ASSEMBLY.
Compliance is checked by visual inspection and by measurement.
NOTE These requirements are adopted from Clause 16, item cc) of IEC 60601-2-28:1993.
201.8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS
201.8.7.3 *Allowable values
Item c) is amended as follows:
For MOBILE X-RAY EQUIPMENT and TRANSPORTABLE X-RAY EQUIPMENT, the TOUCH CURRENT under
SINGLE FAULT CONDITION shall not exceed 2 mA.
Item d) is replaced by:
For MOBILE X-RAY EQUIPMENT and TRANSPORTABLE X-RAY EQUIPMENT, the allowable values of the
EARTH LEAKAGE CURRENT are 2,5 mA in NORMAL CONDITION and 5 mA in SINGLE FAULT
CONDITIONS. For PERMANENTLY INSTALLED ME EQUIPMENT, the allowable value of EARTH LEAKAGE
CURRENT is 10 mA in NORMAL CONDITION and in SINGLE FAULT CONDITIONS.
Item e) is amended as follows:
For PERMANENTLY INSTALLED ME EQUIPMENT including HIGH-VOLTAGE GENERATORS, the allowable
value of EARTH LEAKAGE CURRENT is 20 mA in NORMAL CONDITION and SINGLE FAULT CONDITION.
NOTE These requirements are adapted from 19.3 of IEC 60601-2-7:1998.
201.8.8.3 Dielectric strength
Amendment to the compliance test for high-voltage circuit:
The high-voltage circuit of the ME EQUIPMENT is tested by applying no more than half the test
voltage, and then the test voltage is gradually raised over a period of 10 s to the full value,
which is maintained for 3 min in RADIOGRAPHY and 15 min in RADIOSCOPY.
Addition to the test conditions for high-voltage circuit:
The test for the high-voltage circuit shall be made without an X-RAY TUBE ASSEMBLY connected
and with a test voltage of 1,2 times the NOMINAL X-RAY TUBE VOLTAGE of the ME EQUIPMENT.
If the ME EQUIPMENT can be tested only with the X-RAY TUBE ASSEMBLY connected and if the X-
RAY TUBE does not allow the ME EQUIPMENT to be tested with a test voltage of 1,2 times the
NOMINAL X- RAY TUBE VOLTAGE, the test voltage may be lower but not less than 1,1 times that
voltage.
For ME EQUIPMENT in which the NOMINAL X-RAY TUBE VOLTAGE for RADIOSCOPY does not exceed
80 % of that for RADIOGRAPHY, the test voltage for the high-voltage circuit shall be referred to
the value for RADIOGRAPHY, and the test shall be carried out in that mode only.
If during the dielectric strength test there is a RISK of overheating a transformer under test, it
is permitted to carry out the test at a higher supply frequency.
60601-2-54 © IEC:2009 – 17 –
During the dielectric strength test, the test voltage in the high-voltage circuit should be kept
as close as possible to 100 %, and is not to be outside the range of 100 % and 105 % of the
value required.
During the dielectric strength test, slight corona discharges in the high-voltage circuit are to
be disregarded if they cease when the test voltage is lowered to
...
IEC 60601-2-54 ®
Edition 1.1 2015-04
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X utilisés pour la radiographie et la
radioscopie
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing more than 30 000 terms and
Technical Specifications, Technical Reports and other definitions in English and French, with equivalent terms in 15
documents. Available for PC, Mac OS, Android Tablets and additional languages. Also known as the International
iPad. Electrotechnical Vocabulary (IEV) online.
IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a More than 60 000 electrotechnical terminology entries in
variety of criteria (reference number, text, technical English and French extracted from the Terms and Definitions
committee,…). It also gives information on projects, replaced clause of IEC publications issued since 2002. Some entries
and withdrawn publications. have been collected from earlier publications of IEC TC 37,
77, 86 and CISPR.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements
Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales,
électriques. Il contient plus de 30 000 termes et définitions en
Spécifications techniques, Rapports techniques et autres
anglais et en français, ainsi que les termes équivalents dans
documents de l'IEC. Disponible pour PC, Mac OS, tablettes
15 langues additionnelles. Egalement appelé Vocabulaire
Android et iPad.
Electrotechnique International (IEV) en ligne.
Recherche de publications IEC - www.iec.ch/searchpub
Glossaire IEC - std.iec.ch/glossary
La recherche avancée permet de trouver des publications IEC Plus de 60 000 entrées terminologiques électrotechniques, en
en utilisant différents critères (numéro de référence, texte, anglais et en français, extraites des articles Termes et
comité d’études,…). Elle donne aussi des informations sur les Définitions des publications IEC parues depuis 2002. Plus
projets et les publications remplacées ou retirées. certaines entrées antérieures extraites des publications des
CE 37, 77, 86 et CISPR de l'IEC.
IEC Just Published - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications IEC. Just
Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur cette
Disponible en ligne et aussi une fois par mois par email. publication ou si vous avez des questions contactez-nous:
csc@iec.ch.
IEC 60601-2-54 ®
Edition 1.1 2015-04
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les
performances essentielles des appareils à rayonnement X utilisés pour la
radiographie et la radioscopie
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.50 ISBN 978-2-8322-2619-3
IEC 60601-2-54 ®
Edition 1.1 2015-04
CONSOLIDATED VERSION
REDLINE VERSION
VERSION REDLINE
colour
inside
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X utilisés pour la radiographie et la
radioscopie
– 2 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
INTRODUCTION TO AMENDMENT 1 . 7
201.1 Scope, object and related standards. 8
201.2 Normative references . 9
201.3 Terms and definitions . 10
201.4 General requirements . 11
201.5 General requirements for testing of ME EQUIPMENT . 12
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 12
201.7 ME EQUIPMENT identification, marking and documents . 12
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 15
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 18
201.10 Protection against unwanted and excessive radiation HAZARDS . 22
201.11 Protection against excessive temperatures and other HAZARDS . 22
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 23
201.13 HAZARDOUS SITUATIONS and fault conditions . 23
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 23
201.15 Construction of ME EQUIPMENT . 23
201.16 ME SYSTEMS. 23
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 24
202 Electromagnetic compatibility – Requirements and tests . 24
203 Radiation protection in diagnostic X-ray equipment . 24
Annexes . 60
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS . 61
Annex AA (informative) Particular guidance and rationale . 63
Bibliography . 66
Index of defined terms used in this particular standard. 67
Figure 203.101 – Zone of EXTRA-FOCAL RADIATION . 43
Figure 203.102 – Discrepancies in covering the IMAGE RECEPTION AREA . 45
Figure 203.103 – Discrepancies in visual indication of the X-RAY FIELD . 49
Figure 203.104 – Testing for STRAY RADIATION (X-RAY BEAM horizontal with X-RAY
SOURCE ASSEMBLY below the PATIENT SUPPORT) . 57
Figure 203.105 – Testing for STRAY RADIATION (X-RAY BEAM vertical with X-RAY SOURCE
ASSEMBLY below the PATIENT SUPPORT) . 57
Figure 203.106 – Testing for STRAY RADIATION (X-RAY BEAM horizontal with X-RAY
SOURCE ASSEMBLY above the PATIENT SUPPORT) . 58
Figure 203.107 – Testing for STRAY RADIATION (X-RAY BEAM vertical with X-RAY SOURCE
ASSEMBLY above the PATIENT SUPPORT) . 59
Table 201.101 – Distributed potential ESSENTIAL PERFORMANCE requirements . 11
+AMD1:2015 CSV IEC 2015
Table 203.101 – Tests for verifying reproducibility and linearity . 31
Table 203.102 – Loadings for testing AUTOMATIC EXPOSURE CONTROLS . 33
Table 203.103 – ATTENUATION for the measurement of AIR KERMA . 35
Table 203.104 – ATTENUATION EQUIVALENT of items in the X-RAY BEAM . 51
Table 203.105 – Application categories . 53
Table 203.106 – Requirements for PRIMARY PROTECTIVE SHIELDING . 53
Table 203.107 – STRAY RADIATION in SIGNIFICANT ZONES OF OCCUPANCY . 55
Table 201.C.101 – Marking on the outside of ME EQUIPMENT or its parts . 61
Table 201.C.102 – Subclauses requiring statements in ACCOMPANYING DOCUMENTS . 61
– 4 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
INTERNATIONAL ELECTROTECHNICAL COMMISSION
_____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This consolidated version of the official IEC Standard and its amendment has been prepared
for user convenience.
IEC 60601-2-54 edition 1.1 contains the first edition (2009-06) [documents 62B/735/FDIS and
62B/750/RVD] and its amendment 1 (2015-04) [documents 62B/929/CDV and 62B/956/RVC].
In this Redline version, a vertical line in the margin shows where the technical content is
modified by amendment 1. Additions and deletions are displayed in red, with deletions being
struck through. A separate Final version with all changes accepted is available in this
publication.
+AMD1:2015 CSV IEC 2015
International standard IEC 60601-2-54 has been prepared by IEC subcommittee 62B:
Diagnostic imaging equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
IEC 60601-2-54 has been developed for use with the third edition of IEC 60601-1 (2005). It
replaces and supersedes IEC 60601-2-7 and IEC 60601-2-32, as well as parts of IEC 60601-
2-28:1993 (currently under revision), all of which were developed to amend earlier editions of
IEC 60601-1 and consequently no longer apply to this particular standard.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
The committee has decided that the contents of the base publication and its amendment will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
The contents of the corrigenda of March 2010 and June 2011 have been included in this
copy.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
+AMD1:2015 CSV IEC 2015
INTRODUCTION
This particular standard has been prepared to provide, based on IEC 60601-1:2005 (third
edition) and its collaterals, a complete set of safety requirements for ME EQUIPMENT for
RADIOGRAPHY and RADIOSCOPY. While the previously existing standards for such equipment
were dedicated to components and subsystems, this particular standard addresses the
system level of X-RAY EQUIPMENT, which consists of a combination of an X-RAY GENERATOR,
ASSOCIATED EQUIPMENT and ACCESSORIES. Component functions are addressed as far as
necessary.
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of ME EQUIPMENT for RADIOGRAPHY and
RADIOSCOPY. Requirements for additional provisions for ME EQUIPMENT for interventional
applications are covered by IEC 60601-2-43.
INTRODUCTION TO AMENDMENT 1
The purpose of this first amendment to IEC 60601-2-54:2009 is to introduce changes to
reference the first amendment (2012) to IEC 60601-1:2005. As neither IEC 60601-2-54:2009
nor this amendment refers to specific elements of IEC 60601-1-2, the introduction of a dated
reference to the latter document has been removed. In addition, a number of technical errors
have been corrected.
– 8 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
201.1 Scope, object and related standards
1)
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
ME EQUIPMENT and ME SYSTEMS intended to be used for projection RADIOGRAPHY and
RADIOSCOPY. IEC 60601-2-43 applies to ME EQUIPMENT and ME SYSTEMS intended to be used
for interventional applications and refers to applicable requirements in this particular
standard.
ME EQUIPMENT and ME SYSTEMS intended to be used for bone or tissue absorption
densitometry, computed tomography, mammography or dental applications are excluded from
the scope of this International Standard. The scope of this International Standard also
excludes radiotherapy simulators.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
NOTE Taking into account economic and social factors, the scope of this particular standard includes
ME EQUIPMENT intended to be used for DIRECT RADIOSCOPY. In some countries examinations performed with
DIRECT RADIOSCOPY are prohibited.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for ME EQUIPMENT and ME SYSTEMS for RADIOGRAPHY and
RADIOSCOPY.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-3 apply as modified in Clauses 202 and 203 respectively.
IEC 60601-1-8, IEC 60601-1-10, IEC 60601-1-11 and IEC 60601-1-12 do not apply. All other
published collateral standards in the IEC 60601-1 series apply as published.
—————————
1)
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance
+AMD1:2015 CSV IEC 2015
NOTE OPERATORS of X-RAY EQUIPMENT are used to audible signals as required in this particular standard rather
than to the concepts of IEC 60601-1-8. Therefore IEC 60601-1-8 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography on page 66.
– 10 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
Clause 2 of the general standard applies, except as follows:
Addition:
IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis –
Characteristics of focal spots
IEC 60580:2000, Medical electrical equipment – Dose area product meters
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC 60806, Determination of the maximum symmetrical radiation field from a rotating anode
X-ray tube for medical diagnosis
IEC 62220-1:2003, Medical electrical equipment – Characteristics of digital X-ray imaging
devices – Part 1: Determination of the detective quantum efficiency
Amendment:
IEC 60601-1-2:2007 Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance– Collateral standard: Electromagnetic compatibility –
Requirements and tests
IEC 60601-1-3:2008 Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic X-
ray equipment
IEC 60601-1-3:2008/AMD1:2013
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1:2005/AMD1:2012, applicable collateral standards and IEC 60788:2004 apply,
except as follows:
NOTE An index of defined terms is found beginning on page 66.
Addition:
201.3.201
DIRECT RADIOGRAPHY
RADIOGRAPHY in which the permanent recording is effected at an IMAGE RECEPTION AREA
Example: film-screen or film radiography.
201.3.202
DIRECT RADIOSCOPY
RADIOSCOPY in which the visible images are presented at the IMAGE RECEPTION AREA, or close
to it, in the RADIATION BEAM
201.3.203
DOSE AREA PRODUCT
product of the area of the cross-section of an X-RAY BEAM and the averaged AIR KERMA over
)
that cross-section. The unit is the gray square metre (Gy⋅m
+AMD1:2015 CSV IEC 2015
201.3.204
ENTRANCE FIELD SIZE
dimensions of the field in the entrance plane of an X-RAY IMAGE RECEPTOR that can be used for
the transmission of an X-RAY PATTERN under specific conditions
201.3.205
INDIRECT RADIOGRAPHY
RADIOGRAPHY in which the permanent recording is effected after TRANSFER of the information
obtained at an IMAGE RECEPTION AREA
Examples: CR systems, digital detector systems, image intensifier systems.
201.3.206
INDIRECT RADIOSCOPY
RADIOSCOPY in which the images are presented at a location outside the RADIATION BEAM after
TRANSFER of the information
201.3.207
INTERLOCK
means preventing the start or the continued operation of ME EQUIPMENT unless certain
predetermined conditions prevail
201.3.208
NOMINAL SHORTEST IRRADIATION TIME
shortest LOADING TIME for which a required constancy of the controlled RADIATION QUANTIY is
maintained
NOTE The IRRADIATION TIME is controlled by a HIGH-VOLTAGE GENERATOR with AUTOMATIC CONTROL SYSTEMS.
201.3.209
SERIAL RADIOGRAPHY
RADIOGRAPHY in which the information is obtained and recorded in a regular or irregular series
of LOADINGS with equal or unequal LOADING FACTORS
201.4 General requirements
Clause 4 of the general standard applies, except as follows
201.4.3 ESSENTIAL PERFORMANCE
Additional subclause:
201.4.3.101 * Additional potential ESSENTIAL PERFORMANCE requirements
Additional potential ESSENTIAL PERFORMANCE requirements are found in the subclauses listed
in Table 201.101.
Table 201.101 – Distributed potential ESSENTIAL PERFORMANCE requirements
Requirement Subclause
Accuracy of LOADING FACTORS 203.6.4.3.104
Reproducibility of the RADIATION output 203.6.3.2
AUTOMATIC CONTROL SYSTEM 203.6.5
Imaging performance 203.6.7
– 12 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
201.4.10.2 Supply mains for ME EQUIPMENT and ME SYSTEMS
Addition:
The internal impedance of a SUPPLY MAINS is to be considered sufficiently low for the operation
of X-RAY EQUIPMENT for RADIOGRAPHY and RADIOSCOPY if the value of the APPARENT RESISTANCE
OF SUPPLY MAINS does not exceed the value specified in the ACCOMPANYING DOCUMENTS.
Either the APPARENT RESISTANCE OF SUPPLY MAINS or other appropriate SUPPLY MAINS
specifications used in a facility shall be specified in the ACCOMPANYING DOCUMENTS.
NOTE If a NOMINAL voltage is claimed for a mains power supply system, it is assumed that there is no voltage of a
higher value between any of the conductors of the system or between any of these conductors and earth.
An alternating voltage is considered in practice to be sinusoidal if any instantaneous value of the waveform
concerned differs from the instantaneous value of the ideal waveform at the same moment by no more than ± 2 %
of the peak value of the ideal waveform.
A three-phase SUPPLY MAINS is considered to have a practical symmetry if it delivers symmetrical voltages and
produces, when loaded symmetrically, symmetrical currents.
The requirements of this standard are based upon the assumption that three-phase systems have a symmetrical
configuration of the MAINS VOLTAGE with respect to earth. Single-phase systems may be derived from such
three-phase systems. Where the supply system is not earthed at the source it is assumed that adequate measures
have been provided to detect, limit and remedy any disturbance of symmetry within a reasonably short time.
X-RAY EQUIPMENT is considered to comply with the requirements of this standard only if its specified NOMINAL
ELECTRIC POWER can be demonstrated at an APPARENT RESISTANCE OF SUPPLY MAINS having a value not less than
the APPARENT RESISTANCE OF SUPPLY MAINS specified by the MANUFACTURER in the ACCOMPANYING DOCUMENTS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
201.7.2.7 Electrical input power from the SUPPLY MAINS
Addition:
For ME EQUIPMENT that is specified to be PERMANENTLY INSTALLED, the information may be
stated in the ACCOMPANYING DOCUMENTS only.
The information on the input power shall be specified in terms of combinations of
a) the RATED MAINS VOLTAGE of the ME EQUIPMENT in volts; see 7.2.1 and 7.2.6 of the general
standard,
b) the number of phases; see 7.2.1 and 7.2.6 of the general standard,
c) the frequency, in hertz; see 7.2.1 and 7.2.6 of the general standard,
+AMD1:2015 CSV IEC 2015
d) the maximum permissible value for APPARENT RESISTANCE OF SUPPLY MAINS, in ohms;
e) the characteristics of OVER-CURRENT RELEASES required in the SUPPLY MAINS.
NOTE These requirements are adapted from 6.1j) of IEC 60601-2-7:1998.
201.7.2.15 Cooling conditions
Addition:
If cooling is necessary for safe operation of ME EQUIPMENT, or a subassembly thereof, the
cooling requirements shall be indicated in the ACCOMPANYING DOCUMENT, including as
appropriate:
– the maximum heat dissipation into the surrounding air, given separately for each
subassembly that dissipates more than 100 W and might be separately located on
installation;
– the maximum heat dissipation into forced air cooling devices, and the corresponding flow
rate and temperature rise of the forced air stream;
– the maximum heat dissipation into a cooling medium utility and the permissible input
temperature range, minimum flow rate and pressure requirements for the utility.
NOTE These requirements are adapted from 6.1t) of IEC 60601-2-7:1998.
Additional subclause:
201.7.2.101 Beam limiting device
BEAM LIMITING DEVICES shall be provided with the following markings:
– those required in subclause 7.2.2 of the general standard;
– serial designation or individual identification;
– TOTAL FILTRATION in terms of QUALITY EQUIVALENT FILTRATION of all materials together that
are permanently fixed and intercept the X-RAY BEAM.
NOTE These requirements are adapted from 6.1 of IEC 60601-2-28:1993.
201.7.8.1 Colours of indicator lights
Addition:
The indication of X-RAY related states shall be excluded from subclause 7.8 in the general
standard. Subclauses 203.6.4.2 and 203.6.4.101 shall apply instead.
201.7.9 ACCOMPANYING DOCUMENTS
201.7.9.1 General
Addition:
The ACCOMPANYING DOCUMENTS shall contain quality control procedures to be performed on
the X-RAY EQUIPMENT by the RESPONSIBLE ORGANISATION. These shall include acceptance
criteria and frequency for the tests.
Additionally for X-RAY EQUIPMENT provided with an integrated digital X-RAY IMAGE RECEPTOR,
ACCOMPANYING DOCUMENTS shall contain:
the
– a description of image processing applied to ORIGINAL DATA including the revision number
or how to determine it and identification of the version if applicable;
– a description of the file transfer format of the images acquired with this unit and of any
data associated with these images;
– 14 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
The performance of means required to present the images for diagnostic purpose shall be
stated according to the INTENDED USE.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.7.9.2 Instructions for use
201.7.9.2.1 General
Addition:
201.7.9.2.1.101 LOADING FACTORS
In the instructions for use the LOADING FACTORS shall be stated as described below. The
following combinations and data shall be stated:
a) The corresponding NOMINAL X-RAY TUBE VOLTAGE for RADIOSCOPY and RADIOGRAPHY
together with the highest X-RAY TUBE CURRENT obtainable from the ME EQUIPMENT when
operated at that X-RAY TUBE VOLTAGE.
b) The corresponding highest X-RAY TUBE CURRENT for RADIOSCOPY and RADIOGRAPHY
together with the highest X-RAY TUBE VOLTAGE obtainable from the ME EQUIPMENT when
operating at that X-RAY TUBE CURRENT.
c) The corresponding combination of X-RAY TUBE VOLTAGE for RADIOSCOPY and RADIOGRAPHY,
and X-RAY TUBE CURRENT which results in the highest electric power in the high-voltage
circuit (see 203.4.101).
d) The NOMINAL ELECTRIC POWER given as the highest constant electric power in kilowatts
which the ME EQUIPMENT can produce/generate, for a LOADING TIME of 0,1 s at an X-RAY
TUBE VOLTAGE of 100 kV or, if these values are not selectable, with nearest parameters
(see 203.4.101).
The NOMINAL ELECTRIC POWER shall be given together with the combination of X-RAY TUBE
VOLTAGE and X-RAY TUBE CURRENT and the LOADING TIME.
e) For ME EQUIPMENT indicating precalculated or measured CURRENT TIME PRODUCT, the
lowest CURRENT TIME PRODUCT or the combinations of LOADING FACTORS resulting in the
lowest CURRENT TIME PRODUCT.
If the value of the lowest CURRENT TIME PRODUCT depends upon the X-RAY TUBE VOLTAGE or
upon certain combinations of values of LOADING FACTORS, the lowest CURRENT TIME
PRODUCT may be given as a table or curve showing the dependence.
f) The NOMINAL SHORTEST IRRADIATION TIME used in AUTOMATIC EXPOSURE CONTROL systems of
ME EQUIPMENT.
If the NOMINAL SHORTEST IRRADIATION TIME depends upon LOADING FACTORS such as X-RAY
TUBE VOLTAGE and X-RAY TUBE CURRENT, the ranges of these LOADING FACTORS for which
the NOMINAL SHORTEST IRRADIATION TIME is valid shall be stated.
The maximum possible range of the X-RAY TUBE VOLTAGE and/or the X-RAY TUBE CURRENT
during IRRADIATIONS, controlled with the AUTOMATIC EXPOSURE CONTROL SYSTEMS, shall be
stated in the instructions for use.
NOTE These requirements are adapted from 6.8.2 a) of IEC 60601-2-7:1998.
201.7.9.2.1.102 X-RAY SOURCE ASSEMBLY
The instructions for use shall state the maximum symmetrical RADIATION FIELD of the
integrated X-RAY SOURCE ASSEMBLY determined according to IEC 60806.
NOTE This requirement is adapted from 6.8.2 (dd) of IEC 60601-2-28:1993.
+AMD1:2015 CSV IEC 2015
201.7.9.2.1.103 Integrated X-RAY IMAGE RECEPTOR
For X-RAY EQUIPMENT provided with an integrated X-RAY IMAGE RECEPTOR, the instructions for
use shall contain a description of the particular handling and maintenance of the X-RAY IMAGE
RECEPTOR.
Compliance is checked by inspection of the instructions for use.
201.7.9.2.17 ME EQUIPMENT emitting radiation
Replacement:
For X-RAY EQUIPMENT the instructions for use shall provide information as required in 203.5.
201.7.9.3 Technical description
Additional subclause:
201.7.9.3.101 X-RAY SOURCE ASSEMBLY
The technical description of the integrated X-RAY SOURCE ASSEMBLIES shall specify the
following, in addition to the data required to be marked according to subclause 7.2 of the
general standard:
a) specification of the REFERENCE AXIS to which the TARGET ANGLE(s) and the FOCAL SPOT
characteristics of the X-RAY SOURCE ASSEMBLY refer:
b) TARGET ANGLE(s) with respect to the specified REFERENCE AXIS;
c) position of the FOCAL SPOT and its tolerances on the REFERENCE AXIS;
d) NOMINAL FOCAL SPOT VALUE(s) determined according to IEC 60336 for the specified
REFERENCE AXIS.
NOTE These requirements are adapted from 6.8.3 dd) of IEC 60601-2-28:1993.
Additional subclause:
201.7.9.101 Additional statements in ACCOMPANYING DOCUMENTS
Additional requirements for statements in ACCOMPANYING DOCUMENTS (which include
instructions for use and technical description) are found in the subclauses listed in Table
201.C.102 of Annex C.
201.8 Protection against electrical HAZARDS from ME EQUIPMENT
Clause 8 of the general standard applies, except as follows:
201.8.4 Limitation of voltage, current or energy
201.8.4.3 ME EQUIPMENT intended to be connected to a power source by a plug
Addition:
201.8.4.3.101 HIGH-VOLTAGE CABLE CONNECTIONS
Detachable HIGH-VOLTAGE CABLE CONNECTIONS shall either be designed so that the use of
TOOLS is required to disconnect them or they shall be provided with INTERLOCKS so that at all
times when protective covers or HIGH-VOLTAGE CONNECTIONS are removed:
– the ME EQUIPMENT is disconnected from its power supply, and
– 16 – IEC 60601-2-54:2009
+AMD1:2015 CSV IEC 2015
– capacitances in the high-voltage circuit are discharged within the minimum time necessary
to gain access to the high-voltage circuit, and
– the discharged state is maintained.
Compliance is checked by inspection and by measurement.
NOTE These requirements are adapted from Clause 15, item aa) of IEC 60601-2-7:1998.
Additional subclause:
201.8.4.101 Limitation of X-RAY TUBE VOLTAGE
ME EQUIPMENT shall be designed so as not to deliver in INTENDED USE, to any connected X-RAY
TUBE ASSEMBLY, a voltage greater than the NOMINAL X-RAY TUBE VOLTAGE for the X-RAY TUBE
concerned or greater than the NOMINAL X-RAY TUBE VOLTAGE the X-RAY TUBE ASSEMBLY is
designed for, whichever is the lower voltage.
NOTE This requirement is adapted from 3.1 of IEC 60601-2-7:1998.
201.8.5 Separation of parts
201.8.5.1 MEANS OF PROTECTION (MOP)
Additional subclause:
201.8.5.1.101 Additional limitation of voltage, current or energy
Provision shall be made to prevent the appearance of an unacceptably high voltage in the
MAINS PART or in any other low-voltage circuit.
NOTE This may be achieved for example
– by provision of a winding layer or a conductive screen connected to the PROTECTIVE EARTH TERMINAL between
high-voltage and low-voltage circuits;
– by provision of a voltage limiting dev
...
IEC 60601-2-54 ®
Edition 1.2 2018-06
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X utilisés pour la radiographie et la
radioscopie
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing 21 000 terms and definitions in
Technical Specifications, Technical Reports and other English and French, with equivalent terms in 16 additional
documents. Available for PC, Mac OS, Android Tablets and languages. Also known as the International Electrotechnical
iPad. Vocabulary (IEV) online.
IEC publications search - webstore.iec.ch/advsearchform IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a 67 000 electrotechnical terminology entries in English and
variety of criteria (reference number, text, technical French extracted from the Terms and Definitions clause of
committee,…). It also gives information on projects, replaced IEC publications issued since 2002. Some entries have been
and withdrawn publications. collected from earlier publications of IEC TC 37, 77, 86 and
CISPR.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: sales@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements
Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales,
électriques. Il contient 21 000 termes et définitions en anglais
Spécifications techniques, Rapports techniques et autres
et en français, ainsi que les termes équivalents dans 16
documents de l'IEC. Disponible pour PC, Mac OS, tablettes
langues additionnelles. Egalement appelé Vocabulaire
Android et iPad.
Electrotechnique International (IEV) en ligne.
Recherche de publications IEC -
Glossaire IEC - std.iec.ch/glossary
webstore.iec.ch/advsearchform
67 000 entrées terminologiques électrotechniques, en anglais
La recherche avancée permet de trouver des publications IEC et en français, extraites des articles Termes et Définitions des
en utilisant différents critères (numéro de référence, texte, publications IEC parues depuis 2002. Plus certaines entrées
comité d’études,…). Elle donne aussi des informations sur les antérieures extraites des publications des CE 37, 77, 86 et
projets et les publications remplacées ou retirées. CISPR de l'IEC.
IEC Just Published - webstore.iec.ch/justpublished Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications IEC. Just Si vous désirez nous donner des commentaires sur cette
Published détaille les nouvelles publications parues. publication ou si vous avez des questions contactez-nous:
Disponible en ligne et aussi une fois par mois par email. sales@iec.ch.
IEC 60601-2-54 ®
Edition 1.2 2018-06
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X utilisés pour la radiographie et la
radioscopie
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.50 ISBN 978-2-8322-5862-0
IEC 60601-2-54 ®
Edition 1.2 2018-06
CONSOLIDATED VERSION
REDLINE VERSION
VERSION REDLINE
colour
inside
Medical electrical equipment –
Part 2-54: Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
Appareils électromédicaux –
Partie 2-54: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X utilisés pour la radiographie et la
radioscopie
– 2 – IEC 60601-2-54:2009+AMD1:2015
+AMD2:2018 CSV IEC 2018
CONTENTS
FOREWORD. 4
INTRODUCTION . 7
INTRODUCTION TO AMENDMENT 1 . 7
INTRODUCTION TO AMENDMENT 2 . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 9
201.3 Terms and definitions . 10
201.4 General requirements . 12
201.5 General requirements for testing of ME EQUIPMENT . 13
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 13
201.7 ME EQUIPMENT identification, marking and documents . 13
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 16
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 19
201.10 Protection against unwanted and excessive radiation HAZARDS . 23
201.11 Protection against excessive temperatures and other HAZARDS . 23
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 24
201.13 HAZARDOUS SITUATIONS and fault conditions . 24
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 24
201.15 Construction of ME EQUIPMENT . 24
201.16 ME SYSTEMS . 24
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 25
202 Electromagnetic compatibility – Requirements and tests . 25
203 Radiation protection in diagnostic X-ray equipment . 25
Annexes . 65
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS . 66
Annex AA (informative) Particular guidance and rationale . 68
Bibliography . 73
Index of defined terms used in this particular standard . 75
Figure 203.101 – Zone of EXTRA-FOCAL RADIATION . 47
Figure 203.102 – Discrepancies in covering the IMAGE RECEPTION AREA . 49
Figure 203.103 – Discrepancies in visual indication of the X-RAY FIELD . 53
Figure 203.104 – Testing for STRAY RADIATION (X-RAY BEAM horizontal with X-RAY
SOURCE ASSEMBLY below the PATIENT SUPPORT) . 62
Figure 203.105 – Testing for STRAY RADIATION (X-RAY BEAM vertical with X-RAY SOURCE
ASSEMBLY below the PATIENT SUPPORT) . 62
Figure 203.106 – Testing for STRAY RADIATION (X-RAY BEAM horizontal with X-RAY
SOURCE ASSEMBLY above the PATIENT SUPPORT) . 63
Figure 203.107 – Testing for STRAY RADIATION (X-RAY BEAM vertical with X-RAY SOURCE
ASSEMBLY above the PATIENT SUPPORT) . 64
+AMD2:2018 CSV IEC 2018
Table 201.101 – Distributed potential ESSENTIAL PERFORMANCE requirements . 12
Table 203.101 – Tests for verifying reproducibility and linearity. 33
Table 203.102 – Loadings for testing AUTOMATIC EXPOSURE CONTROLS . 35
Table 203.103 – ATTENUATION for the measurement of AIR KERMA . 37
Table 203.104 – ATTENUATION EQUIVALENT of items in the X-RAY BEAM . 55
Table 203.105 – Application categories . 57
Table 203.106 – Requirements for PRIMARY PROTECTIVE SHIELDING . 58
Table 203.107 – STRAY RADIATION in SIGNIFICANT ZONES OF OCCUPANCY . 60
Table 201.C.101 – Marking on the outside of ME EQUIPMENT or its parts . 66
Table 201.C.102 – Subclauses requiring statements in ACCOMPANYING DOCUMENTS . 66
– 4 – IEC 60601-2-54:2009+AMD1:2015
+AMD2:2018 CSV IEC 2018
INTERNATIONAL ELECTROTECHNICAL COMMISSION
_____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This consolidated version of the official IEC Standard and its amendments has been prepared
for user convenience.
IEC 60601-2-54 edition 1.2 contains the first edition (2009-06) [documents 62B/735/FDIS and
62B/750/RVD], its amendment 1 (2015-04) [documents 62B/929/CDV and 62B/956/RVC],
its amendment 2 (2018-06) [documents 62B/1089/FDIS and 62B/1097/RVD] and its
corrigenda 1 (March 2010) and 2 (June 2011).
In this Redline version, a vertical line in the margin shows where the technical content is
modified by amendments 1 and 2. Additions and deletions are displayed in red, with deletions
being struck through. A separate Final version with all changes accepted is available in this
publication.
+AMD2:2018 CSV IEC 2018
In this Redline version, a vertical line in the margin shows where the technical content
is modified by amendments 1 and 2. Additions and deletions are displayed in red, with
deletions being struck through. A separate Final version with all changes accepted is
available in this publication.
International standard IEC 60601-2-54 has been prepared by IEC subcommittee 62B:
Diagnostic imaging equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
IEC 60601-2-54 has been developed for use with the third edition of IEC 60601-1 (2005). It
replaces and supersedes IEC 60601-2-7 and IEC 60601-2-32, as well as parts of IEC 60601-
2-28:1993 (currently under revision), all of which were developed to amend earlier editions of
IEC 60601-1 and consequently no longer apply to this particular standard.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-2-54:2009+AMD1:2015
+AMD2:2018 CSV IEC 2018
The committee has decided that the contents of the base publication and its amendments will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
+AMD2:2018 CSV IEC 2018
INTRODUCTION
This particular standard has been prepared to provide, based on IEC 60601-1:2005 (third
edition) and its collaterals, a complete set of safety requirements for ME EQUIPMENT for
RADIOGRAPHY and RADIOSCOPY. While the previously existing standards for such equipment
were dedicated to components and subsystems, this particular standard addresses the
system level of X-RAY EQUIPMENT, which consists of a combination of an X-RAY GENERATOR,
ASSOCIATED EQUIPMENT and ACCESSORIES. Component functions are addressed as far as
necessary.
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of ME EQUIPMENT for RADIOGRAPHY and
RADIOSCOPY. Requirements for additional provisions for ME EQUIPMENT for interventional
applications are covered by IEC 60601-2-43.
INTRODUCTION TO AMENDMENT 1
The purpose of this first amendment to IEC 60601-2-54:2009 is to introduce changes to
reference the first amendment (2012) to IEC 60601-1:2005. As neither IEC 60601-2-54:2009
nor this amendment refers to specific elements of IEC 60601-1-2, the introduction of a dated
reference to the latter document has been removed. In addition, a number of technical errors
have been corrected.
INTRODUCTION TO AMENDMENT 2
The purpose of this second amendment to IEC 60601-2-54:2009 is to introduce changes
which take the current state of the art into account. Therefore, X-RAY EQUIPMENT specified for
DIRECT RADIOSCOPY is no longer in the scope of this document. The normative references were
also updated in this amendment, and editorial clarifications and new terms and definitions
were added. Provisions for QUALITY CONTROL PROCEDURES to be recommended by the
MANUFACTURER are emphasized. Specific attention is paid to EXAMINATION PROTOCOLS in a new
subclause which differentiate between adult and paediatric applications, in particular for
X-RAY EQUIPMENT without an AUTOMATIC CONTROL SYSTEM. In addition, fixed periods for
termination of LOADING after release of the RADIATION control by the OPERATOR are stipulated
for RADIOSCOPY.
A new subclause on electronic documentation of EXAMINATION PROTOCOLS is introduced. It
recommends providing access to electronic documentation containing relevant parameters of
the PRE-PROGRAMMED EXAMINATION PROTOCOL. In another new subclause, the creation of basic
documentation of the RADIATION DOSE STRUCTURED REPORT (RDSR) according to IEC 61910-1 is
recommended. Furthermore, the subclause describing the LAST IMAGE HOLD RADIOGRAM has
been revised and requires that the last image in RADIOSCOPY be displayed rather than provide
just a means to display it.
This amendment recommends providing a graphical DISPLAY of the position of the BEAM
LIMITING DEVICE blades on the IMAGE DISPLAY DEVICE in the subclause "Indication on the X-RAY
EQUIPMENT".
Finally, the requirement for providing means to limit the FOCAL SPOT TO SKIN DISTANCES for
radioscopic X-RAY EQUIPMENT differentiates between MOBILE and FIXED EQUIPMENT and
extends, in the latter case, the minimum distance in possible clinical applications.
– 8 – IEC 60601-2-54:2009+AMD1:2015
+AMD2:2018 CSV IEC 2018
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
201.1 Scope, object and related standards
1)
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
ME EQUIPMENT and ME SYSTEMS intended to be used for projection RADIOGRAPHY and INDIRECT
RADIOSCOPY. IEC 60601-2-43 applies to ME EQUIPMENT and ME SYSTEMS intended to be used
for interventional applications and refers to applicable requirements in this particular
standard.
ME EQUIPMENT and ME SYSTEMS intended to be used for bone or tissue absorption
densitometry, computed tomography, mammography or dental or radiotherapy applications
are excluded from the scope of this International Standard. The scope of this International
Standard also excludes radiotherapy simulators.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
NOTE Taking into account economic and social factors, the scope of this particular standard includes
ME EQUIPMENT intended to be used for DIRECT RADIOSCOPY. In some countries examinations performed with
DIRECT RADIOSCOPY are prohibited.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for ME EQUIPMENT and ME SYSTEMS for RADIOGRAPHY and
RADIOSCOPY.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-3 apply as modified in Clauses 202 and 203 respectively.
IEC 60601-1-8, IEC 60601-1-9, IEC 60601-1-10, IEC 60601-1-11 and IEC 60601-1-12 do not
apply. All other published collateral standards in the IEC 60601-1 series apply as published.
—————————
1)
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance
+AMD2:2018 CSV IEC 2018
NOTE OPERATORS of X-RAY EQUIPMENT are used to audible signals as required in this particular standard rather
than to the concepts of IEC 60601-1-8. Therefore IEC 60601-1-8 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography on page 73.
– 10 – IEC 60601-2-54:2009+AMD1:2015
+AMD2:2018 CSV IEC 2018
Clause 2 of the general standard applies, except as follows:
Addition:
IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis –
Characteristics of focal spots
IEC 60580:2000, Medical electrical equipment – Dose area product meters
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC 60806, Determination of the maximum symmetrical radiation field from a rotating anode
X-ray tube for medical diagnosis
IEC 62220-1-1:2003 2015, Medical electrical equipment – Characteristics of digital X-ray
imaging devices – Part 1-1: Determination of the detective quantum efficiency – Detectors
used in radiographic imaging
Amendment:
IEC 60601-1-2:2007 Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance– Collateral standard: Electromagnetic compatibility –
Requirements and tests
IEC 60601-1-3:2008 Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic X-
ray equipment
IEC 60601-1-3:2008/AMD1:2013
IEC 61910-1:2014, Medical electrical equipment – Radiation dose documentation – Part 1:
Radiation dose structured reports for radiography and radioscopy
IEC 62494-1:2008, Medical electrical equipment – Exposure index of digital X-ray imaging
systems – Part 1: Definitions and requirements for general radiography
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1:2005/AMD1:2012, applicable collateral standards and IEC 60788:2004 apply,
except as follows:
NOTE An index of defined terms is found beginning on page 73.
Addition:
201.3.201
DIRECT RADIOGRAPHY
RADIOGRAPHY in which the permanent recording is effected at an IMAGE RECEPTION AREA
Example: film-screen or film radiography.
+AMD2:2018 CSV IEC 2018
201.3.202
DIRECT RADIOSCOPY
RADIOSCOPY in which the visible images are presented at the IMAGE RECEPTION AREA, or close
to it, in the RADIATION BEAM
201.3.203
DOSE AREA PRODUCT
product of the area of the cross-section of an X-RAY BEAM and the averaged AIR KERMA over
that cross-section. The unit is the gray square metre (Gy⋅m )
201.3.204
ENTRANCE FIELD SIZE
dimensions of the field in the entrance plane of an X-RAY IMAGE RECEPTOR that can be used for
the transmission of an X-RAY PATTERN under specific conditions
201.3.205
INDIRECT RADIOGRAPHY
RADIOGRAPHY in which the permanent recording is effected after TRANSFER of the information
obtained at an IMAGE RECEPTION AREA
Examples: CR systems, digital detector systems, image intensifier systems.
201.3.206
INDIRECT RADIOSCOPY
RADIOSCOPY in which the images are presented at a location outside the RADIATION BEAM after
TRANSFER of the information
201.3.207
INTERLOCK
means preventing the start or the continued operation of ME EQUIPMENT unless certain
predetermined conditions prevail
201.3.208
NOMINAL SHORTEST IRRADIATION TIME
shortest LOADING TIME for which a required constancy of the controlled RADIATION QUANTIY is
maintained
NOTE The IRRADIATION TIME is controlled by a HIGH-VOLTAGE GENERATOR with AUTOMATIC CONTROL SYSTEMS.
201.3.209
SERIAL RADIOGRAPHY
RADIOGRAPHY in which the information is obtained and recorded in a regular or irregular series
of LOADINGS with equal or unequal LOADING FACTORS
201.3.210
EXAMINATION PROTOCOL
full set of any programmed technical factors, control functions and settings, including image
processing settings, designed to optimize the image acquisition and DISPLAY
201.3.211
EXAMINATION PROTOCOL SELECTION CONTROL
control to select a PRE-PROGRAMMED EXAMINATION PROTOCOL
201.3.212
LAST-IMAGE HOLD RADIOGRAM
LIH RADIOGRAM
single image obtained by sampling or temporal processing of one or more images from the
end of a radioscopic IRRADIATION
Note 1 to entry: This note applies to the French language only.
– 12 – IEC 60601-2-54:2009+AMD1:2015
+AMD2:2018 CSV IEC 2018
201.3.213
PRE-PROGRAMMED EXAMINATION PROTOCOL
single hardware or software setting, or both, which is associated with an EXAMINATION
PROTOCOL
201.3.214
RADIOSCOPY REPLAY IMAGE SEQUENCE
series of the most recent images of the most recent RADIOSCOPY IRRADIATION-EVENT
201.4 General requirements
Clause 4 of the general standard applies, except as follows
201.4.3 ESSENTIAL PERFORMANCE
Additional subclause:
201.4.3.101 * Additional potential ESSENTIAL PERFORMANCE requirements
Additional potential ESSENTIAL PERFORMANCE requirements are found in the subclauses listed
in Table 201.101.
Table 201.101 – Distributed potential ESSENTIAL PERFORMANCE requirements
Requirement Subclause
Accuracy of LOADING FACTORS 203.6.4.3.104
Reproducibility of the RADIATION output 203.6.3.2
AUTOMATIC CONTROL SYSTEM 203.6.5
Imaging performance 203.6.7
201.4.10.2 Supply mains for ME EQUIPMENT and ME SYSTEMS
Addition:
The internal impedance of a SUPPLY MAINS is to be considered sufficiently low for the operation
of X-RAY EQUIPMENT for RADIOGRAPHY and RADIOSCOPY if the value of the APPARENT RESISTANCE
OF SUPPLY MAINS does not exceed the value specified in the ACCOMPANYING DOCUMENTS.
Either the APPARENT RESISTANCE OF SUPPLY MAINS or other appropriate SUPPLY MAINS
specifications used in a facility shall be specified in the ACCOMPANYING DOCUMENTS.
NOTE If a NOMINAL voltage is claimed for a mains power supply system, it is assumed that there is no voltage of a
higher value between any of the conductors of the system or between any of these conductors and earth.
An alternating voltage is considered in practice to be sinusoidal if any instantaneous value of the waveform
concerned differs from the instantaneous value of the ideal waveform at the same moment by no more than ± 2 %
of the peak value of the ideal waveform.
A three-phase SUPPLY MAINS is considered to have a practical symmetry if it delivers symmetrical voltages and
produces, when loaded symmetrically, symmetrical currents.
The requirements of this standard are based upon the assumption that three-phase systems have a symmetrical
configuration of the MAINS VOLTAGE with respect to earth. Single-phase systems may be derived from such
three-phase systems. Where the supply system is not earthed at the source it is assumed that adequate measures
have been provided to detect, limit and remedy any disturbance of symmetry within a reasonably short time.
+AMD2:2018 CSV IEC 2018
X-RAY EQUIPMENT is considered to comply with the requirements of this standard only if its specified NOMINAL
ELECTRIC POWER can be demonstrated at an APPARENT RESISTANCE OF SUPPLY MAINS having a value not less than
the APPARENT RESISTANCE OF SUPPLY MAINS specified by the MANUFACTURER in the ACCOMPANYING DOCUMENTS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
201.7.2.7 Electrical input power from the SUPPLY MAINS
Addition:
For ME EQUIPMENT that is specified to be PERMANENTLY INSTALLED, the information may be
stated in the ACCOMPANYING DOCUMENTS only.
The information on the input power shall be specified in terms of combinations of
a) the RATED MAINS VOLTAGE of the ME EQUIPMENT in volts; see 7.2.1 and 7.2.6 of the general
standard,
b) the number of phases; see 7.2.1 and 7.2.6 of the general standard,
c) the frequency, in hertz; see 7.2.1 and 7.2.6 of the general standard,
d) the maximum permissible value for APPARENT RESISTANCE OF SUPPLY MAINS, in ohms;
e) the characteristics of OVER-CURRENT RELEASES required in the SUPPLY MAINS.
NOTE These requirements are adapted from 6.1j) of IEC 60601-2-7:1998.
201.7.2.15 Cooling conditions
Addition:
If cooling is necessary for safe operation of ME EQUIPMENT, or a subassembly thereof, the
cooling requirements shall be indicated in the ACCOMPANYING DOCUMENT, including as
appropriate:
– the maximum heat dissipation into the surrounding air, given separately for each
subassembly that dissipates more than 100 W and might be separately located on
installation;
– the maximum heat dissipation into forced air cooling devices, and the corresponding flow
rate and temperature rise of the forced air stream;
– the maximum heat dissipation into a cooling medium utility and the permissible input
temperature range, minimum flow rate and pressure requirements for the utility.
NOTE These requirements are adapted from 6.1t) of IEC 60601-2-7:1998.
Additional subclause:
– 14 – IEC 60601-2-54:2009+AMD1:2015
+AMD2:2018 CSV IEC 2018
201.7.2.101 Beam limiting device
EAM LIMITING DEVICES shall be provided with the following markings:
B
– those required in subclause 7.2.2 of the general standard;
– serial designation or individual identification;
– TOTAL FILTRATION in terms of QUALITY EQUIVALENT FILTRATION of all materials together that
are permanently fixed and intercept the X-RAY BEAM.
NOTE These requirements are adapted from 6.1 of IEC 60601-2-28:1993.
201.7.8.1 Colours of indicator lights
Addition:
The indication of X-RAY related states shall be excluded from subclause 7.8 in the general
standard. Subclauses 203.6.4.2 and 203.6.4.101 shall apply instead.
201.7.9 ACCOMPANYING DOCUMENTS
201.7.9.1 General
Addition:
The ACCOMPANYING DOCUMENTS shall contain instructions for MANUFACTURER-recommended
QUALITY CONTROL PROCEDURES and tests to be performed on the X-RAY EQUIPMENT by the
RESPONSIBLE ORGANIZATION. These shall include acceptance criteria for each test and
frequency for the each tests.
NOTE The intention is to perform these QUALITY CONTROL PROCEDURES and tests using only the supplied
information.
Additionally for X-RAY EQUIPMENT provided with an integrated digital X-RAY IMAGE RECEPTOR,
the ACCOMPANYING DOCUMENTS shall contain:
– a description an identification of adjustable or selectable image processing applied to
ORIGINAL DATA including the revision version number or how to determine it and
identification of the version if applicable;
– a description of the file transfer format of the images acquired with this unit and of any
data associated with these images;
The performance of means required to present the images for diagnostic purpose shall be
stated according to the INTENDED USE.
If the test or PROCEDURE requires a device-specific TOOL that is only available from the
MANUFACTURER, the MANUFACTURER shall make this TOOL available to the RESPONSIBLE
ORGANIZATION.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.7.9.2 Instructions for use
201.7.9.2.1 General
Addition:
201.7.9.2.1.101 LOADING FACTORS
In the instructions for use the LOADING FACTORS shall be stated as described below. The
following combinations and data shall be stated:
+AMD2:2018 CSV IEC 2018
a) The corresponding NOMINAL X-RAY TUBE VOLTAGE for RADIOSCOPY and RADIOGRAPHY
together with the highest X-RAY TUBE CURRENT obtainable from the ME EQUIPMENT when
operated at that X-RAY TUBE VOLTAGE.
RAY TUBE CURRENT for RADIOSCOPY and RADIOGRAPHY
b) The corresponding highest X-
together with the highest X-RAY TUBE VOLTAGE obtainable from the ME EQUIPMENT when
operating at that X-RAY TUBE CURRENT.
c) The corresponding combination of X-RAY TUBE VOLTAGE for RADIOSCOPY and RADIOGRAPHY,
RAY TUBE CURRENT which results in the highest electric power in the high-voltage
and X-
circuit (see 203.4.101).
d) The NOMINAL ELECTRIC POWER given as the highest constant electric power in kilowatts
which the ME EQUIPMENT can produce/generate, for a LOADING TIME of 0,1 s at an X-RAY
TUBE VOLTAGE of 100 kV or, if these values are not selectable, with nearest parameters
(see 203.4.101).
The NOMINAL ELECTRIC POWER shall be given together with the combination of X-RAY TUBE
VOLTAGE and X-RAY TUBE CURRENT and the LOADING TIME.
e) For ME EQUIPMENT indicating precalculated or measured CURRENT TIME PRODUCT, the
lowest CURRENT TIME PRODUCT or the combinations of LOADING FACTORS resulting in the
lowest CURRENT TIME PRODUCT.
If the value of the lowest CURRENT TIME PRODUCT depends upon the X-RAY TUBE VOLTAGE or
upon certain combinations of values of LOADING FACTORS, the lowest CURRENT TIME
PRODUCT may be given as a table or curve showing the dependence.
NOMINAL SHORTEST IRRADIATION TIME used in AUTOMATIC EXPOSURE CONTROL systems of
f) The
ME EQUIPMENT.
If the NOMINAL SHORTEST IRRADIATION TIME depends upon LOADING FACTORS such as X-RAY
TUBE VOLTAGE and X-RAY TUBE CURRENT, the ranges of these LOADING FACTORS for which
NOMINAL SHORTEST IRRADIATION TIME is valid shall be stated.
the
The maximum possible range of the X-RAY TUBE VOLTAGE and/or the X-RAY TUB
...












Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...