EN 61326-2-4:2013
(Main)Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-4: Particular requirements - Test configurations, operational conditions and performance criteria for insulation monitoring devices according to IEC 61557-8 and for equipment for insulation fault location according to IEC 61557-9
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-4: Particular requirements - Test configurations, operational conditions and performance criteria for insulation monitoring devices according to IEC 61557-8 and for equipment for insulation fault location according to IEC 61557-9
IEC 61236-2-4:2012 specifies detailed test configurations, operational conditions and performance criteria for equipment for insulation monitoring according to IEC 61557-8 and insulation fault location according to IEC 61557-9. This applies to insulation monitoring devices and insulation fault location systems permanently or semi-permanently connected to the distribution system. This second edition cancels and replaces the first edition published in 2006 and constitutes a technical revision. It includes the following significant technical change: update of the document with respect to IEC 61326-1:2012.
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen - Teil 2-4: Besondere Anforderungen - Prüfanordnung, Betriebsbedingungen und Leistungsmerkmale für Isolationsüberwachungsgeräte gemäß IEC 61557-8 und Geräte zur Isolationsfehlerortung gemäß IEC 61557-9
Matériel électrique de mesure, de commande et de laboratoire - Exigences relatives à la CEM - Partie 2-4: Exigences particulières - Configurations d'essai, conditions de fonctionnement et critères de performance pour les contrôleurs d'isolement conformes à la CEI 61557-8 et pour les dispositifs de localisation de défaut d'isolement conformes à la CEI 61557-9
La CEI 61326-2-4:2012 spécifie en détail, les configurations d'essai, les conditions de fonctionnement et les critères de performance pour les équipements pour le contrôle de l'isolement conforme à la CEI 61557-8 et la localisation des défauts d'isolement conformes à la CEI 61557-9. Elle s'applique aux contrôleurs d'isolement et aux systèmes de localisation des défauts d'isolement connectés de manière permanente ou semi-permanente au réseau de distribution. Cette deuxième édition annule et remplace la première édition parue en 2006 et elle constitue une révision technique. Elle inclut la modification technique majeure suivante: mise à jour du document par rapport à la CEI 61326-1:2012.
Električna oprema za merjenje, kontrolo in laboratorijsko uporabo - Zahteve za elektromagnetno združljivost (EMC) - 2-4. del: Posebne zahteve - Preskusne konfiguracije, obratovalni pogoji in merila za lastnosti naprav za stalno preverjanje izolacije po IEC 61557-8 in opreme za ugotavljanje mesta okvare izolacije po IEC 61557-9 (IEC 61326-2-4:2012)
Poleg standarda IEC 61236-1 ta del standarda IEC 61326 določa podrobnejše preskusne konfiguracije, obratovalne pogoje in merila za lastnosti kot standard IEC 61326-1 za opremo za stalno preverjanje izolacije po IEC 61557-8 in ugotavljanje mesta okvare izolacije po IEC 61557-9. Uporablja se za naprave za stalno preverjanje izolacije in sisteme za ugotavljanje mesta okvare izolacije, ki so trajno ali poltrajno priključeni na razdelilni sistem.
General Information
Relations
Frequently Asked Questions
EN 61326-2-4:2013 is a standard published by CLC. Its full title is "Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-4: Particular requirements - Test configurations, operational conditions and performance criteria for insulation monitoring devices according to IEC 61557-8 and for equipment for insulation fault location according to IEC 61557-9". This standard covers: IEC 61236-2-4:2012 specifies detailed test configurations, operational conditions and performance criteria for equipment for insulation monitoring according to IEC 61557-8 and insulation fault location according to IEC 61557-9. This applies to insulation monitoring devices and insulation fault location systems permanently or semi-permanently connected to the distribution system. This second edition cancels and replaces the first edition published in 2006 and constitutes a technical revision. It includes the following significant technical change: update of the document with respect to IEC 61326-1:2012.
IEC 61236-2-4:2012 specifies detailed test configurations, operational conditions and performance criteria for equipment for insulation monitoring according to IEC 61557-8 and insulation fault location according to IEC 61557-9. This applies to insulation monitoring devices and insulation fault location systems permanently or semi-permanently connected to the distribution system. This second edition cancels and replaces the first edition published in 2006 and constitutes a technical revision. It includes the following significant technical change: update of the document with respect to IEC 61326-1:2012.
EN 61326-2-4:2013 is classified under the following ICS (International Classification for Standards) categories: 17.220 - Electricity. Magnetism. Electrical and magnetic measurements; 19.080 - Electrical and electronic testing; 25.040.40 - Industrial process measurement and control; 33.100 - Electromagnetic compatibility (EMC). The ICS classification helps identify the subject area and facilitates finding related standards.
EN 61326-2-4:2013 has the following relationships with other standards: It is inter standard links to EN 61326-2-4:2006, EN IEC 61326-2-4:2021. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN 61326-2-4:2013 is associated with the following European legislation: EU Directives/Regulations: 2004/108/EC, 2004/108/EU, 2014/30/EU. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
You can purchase EN 61326-2-4:2013 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of CLC standards.
Standards Content (Sample)
SLOVENSKI STANDARD
01-marec-2013
1DGRPHãþD
SIST EN 61326-2-4:2007
(OHNWULþQDRSUHPD]DPHUMHQMHNRQWURORLQODERUDWRULMVNRXSRUDER=DKWHYH]D
HOHNWURPDJQHWQR]GUXåOMLYRVW(0&GHO3RVHEQH]DKWHYH3UHVNXVQH
NRQILJXUDFLMHREUDWRYDOQLSRJRMLLQPHULOD]DODVWQRVWLQDSUDY]DVWDOQRSUHYHUMDQMH
L]RODFLMHSR,(&LQRSUHPH]DXJRWDYOMDQMHPHVWDRNYDUHL]RODFLMHSR,(&
,(&
Electrical equipment for measurement, control and laboratory use - EMC requirements -
Part 2-4: Particular requirements - Test configurations, operational conditions and
performance criteria for insulation monitoring devices according to IEC 61557-8 and for
equipment for insulation fault location according to IEC 61557-9 (IEC 61326-2-4:2012)
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen - Teil 2-4:
Besondere Anforderungen - Prüfanordnung, Betriebsbedingungen und
Leistungsmerkmale für Isolationsüberwachungsgeräte gemäß IEC 61557-8 und Geräte
zur Isolationsfehlerortung gemäß IEC 61557-9 (IEC 61326-2-4:2012)
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
Matériel électrique de mesure, de commande et de laboratoire - Exigences relatives à la
CEM - Partie 2-4: Exigences particulières - Configurations d'essai, conditions de
fonctionnement et critères d'aptitude à la fonction pour les contrôleurs d'isolement
conformes à la CEI 61557-8 et pour les dispositifs de localisation de défaut d'isolement
conformes à la CEI 61557-9 (CEI 61326-2-4:2012)
Ta slovenski standard je istoveten z: EN 61326-2-4:2013
ICS:
19.080 (OHNWULþQRLQHOHNWURQVNR Electrical and electronic
SUHVNXãDQMH testing
33.100.01 Elektromagnetna združljivost Electromagnetic compatibility
na splošno in general
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN 61326-2-4
NORME EUROPÉENNE
January 2013
EUROPÄISCHE NORM
ICS 17.220; 19.080; 25.040.40; 33.100 Supersedes EN 61326-2-4:2006
English version
Electrical equipment for measurement, control and laboratory use -
EMC requirements -
Part 2-4: Particular requirements -
Test configurations, operational conditions and performance criteria for
insulation monitoring devices according to IEC 61557-8 and for equipment
for insulation fault location according to IEC 61557-9
(IEC 61326-2-4:2012)
Matériel électrique de mesure, de Elektrische Mess-, Steuer-, Regel- und
commande et de laboratoire - Laborgeräte -
Exigences relatives à la CEM - EMV-Anforderungen -
Partie 2-4: Exigences particulières - Teil 2-4: Besondere Anforderungen -
Configurations d'essai, conditions de Prüfanordnung, Betriebsbedingungen und
fonctionnement et critères de performance Leistungsmerkmale für
pour les contrôleurs d'isolement Isolationsüberwachungsgeräte gemäß
conformes à la CEI 61557-8 et pour les IEC 61557-8 und Geräte zur
dispositifs de localisation de défaut Isolationsfehlerortung gemäß IEC 61557-9
d'isolement conformes à la CEI 61557-9 (IEC 61326-2-4:2012)
(CEI 61326-2-4:2012)
This European Standard was approved by CENELEC on 2012-08-14. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the CEN-CENELEC Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the CEN-CENELEC Management Centre has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus,
the Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany,
Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland,
Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.
CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
Management Centre: Avenue Marnix 17, B - 1000 Brussels
© 2013 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 61326-2-4:2013 E
Foreword
The text of document 65A/630/FDIS, future edition 2 of IEC 61326-2-4, prepared by SC 65A, "System
aspects", of IEC TC 65, "Industrial-process measurement, control and automation" was submitted to the
IEC-CENELEC parallel vote and approved by CENELEC as EN 61326-2-4:2013.
The following dates are fixed:
• latest date by which the document has (dop) 2013-07-10
to be implemented at national level by
publication of an identical national
standard or by endorsement
(dow) 2015-08-14
• latest date by which the national
standards conflicting with the
document have to be withdrawn
This document supersedes EN 61326-2-4:2006.
4:2006:
– update of the document with respect to EN 61326-1:2013.
of clauses, subclauses, tables and figures.
When a particular subclause of EN 61326-1 is not mentioned in this part, that subclause applies as far as
is reasonable. When this standard states “addition”, “modification” or “replacement”, the relevant text in
EN 61326-1 is to be adapted accordingly.
NOTE The following numbering system is used:
– subclauses, tables and figures that are numbered starting from 101 are additional to those in EN 61326-1;
– unless notes are in a new subclause or involve notes in EN 61326-1, they are numbered starting from 101 including
those in a replaced clause or subclause;
– additional annexes are lettered AA, BB, etc.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such patent
rights.
This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).
For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this
document.
Endorsement notice
The text of the International Standard IEC 61326-2-4:2012 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following note has to be added for the standard indicated:
IEC 61557-1:2007 NOTE Harmonized as EN 61557-1:2007 (not modified).
- 3 - EN 61326-2-4:2013
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
NOTE When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.
Annex ZA of EN 61326-1:2013 applies, except as follows:
Publication Year Title EN/HD Year
Addition:
IEC 61326-1 2012 Electrical equipment for measurement, EN 61326-1 2013
control and laboratory use -
EMC requirements -
Part 1: General requirements
IEC 61557-8 2007 Electrical safety in low voltage distribution EN 61557-8 2007
+ corr. May 2007 systems up to 1 000 V a.c. and 1 500 V d.c. -
Equipment for testing, measuring or
monitoring of protective measures -
Part 8: Insulation monitoring devices for IT
systems
IEC 61557-9 2009 Electrical safety in low voltage distribution EN 61557-9 2009
systems up to 1 000 V a.c. and 1 500 V d.c. -
Equipment for testing, measuring or
monitoring of protective measures -
Part 9: Equipment for insulation fault location
in IT systems
Annex ZZ
(informative)
Coverage of Essential Requirements of EU Directives
This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association and within its scope the standard covers
protection requirements of Annex I, Article 1 of the EC Directive 2004/108/EC.
Compliance with this standard provides one means of conformity with the specified essential
requirements of the Directive[s] concerned.
NOTE Other requirements and other EU Directives may be applicable to the products falling within the scope of
this standard.
IEC 61326-2-4 ®
Edition 2.0 2012-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Electrical equipment for measurement, control and laboratory use –
EMC requirements –
Part 2-4: Particular requirements – Test configurations, operational conditions and
performance criteria for insulation monitoring devices according to IEC 61557-8 and
for equipment for insulation fault location according to IEC 61557-9
Matériel électrique de mesure, de commande et de laboratoire – Exigences relatives
à la CEM –
Partie 2-4: Exigences particulières – Configurations d'essai, conditions de
fonctionnement et critères de performance pour les contrôleurs d'isolement conformes
à la CEI 61557-8 et pour les dispositifs de localisation de défaut d’isolement conformes
à la CEI 61557-9
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX N
ICS 17.220; 19.080; 25.040.40; 33.100 ISBN 978-2-83220-209-8
– 2 – 61326-2-4 © IEC:2012
CONTENTS
FOREWORD . 3
1 Scope . 5
2 Normative references . 5
3 Terms and definitions . 5
4 General . 6
5 EMC test plan . 6
5.1 General . 6
5.2 Configuration of EUT during testing . 7
5.2.1 General . 7
5.2.2 Composition of EUT . 7
5.2.3 Assembly of EUT . 7
5.2.4 I/O ports . 7
5.2.5 Auxiliary equipment . 7
5.2.6 Cabling and earthing (grounding) . 7
5.3 Operation conditions of EUT during testing . 7
5.3.101 Operational conditions . 7
5.4 Specification of functional performance . 8
5.5 Test description . 8
6 Immunity requirements . 8
6.1 Conditions during the tests . 8
6.1.101 Electrostatic discharge immunity tests . 8
6.1.102 Electromagnetic field tests . 9
6.1.103 Burst tests . 9
6.1.104 Surge immunity tests. 9
6.1.105 Conducted RF tests . 9
6.1.106 Power frequency magnetic field tests . 9
6.2 Immunity test requirements . 9
6.3 Random aspects. 10
6.4 Performance criteria . 10
7 Emission requirements . 12
7.1 Conditions during measurements. . 12
7.2 Emission limits . 12
8 Test results and test report . 12
9 Instructions for use . 13
Annex A (normative) Immunity test requirements for portable test and measurement
equipment powered by battery or from the circuit being measured . 14
Bibliography . 15
Table 101 – Immunity tests . 10
Table 102 – Performance criteria definition . 11
Table 103 – Test conditions for quiescent and operate modes . 12
61326-2-4 © IEC:2012 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
ELECTRICAL EQUIPMENT FOR MEASUREMENT,
CONTROL AND LABORATORY USE –
EMC REQUIREMENTS –
Part 2-4: Particular requirements –
Test configurations, operational conditions and performance
criteria for insulation monitoring devices according to IEC 61557-8
and for equipment for insulation fault location according to IEC 61557-9
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, u
...
La norme EN 61326-2-4:2013 précise les exigences EMC pour les équipements électriques utilisés pour la mesure, le contrôle et les laboratoires, avec un accent particulier sur les dispositifs de surveillance d'isolation selon IEC 61557-8 et les équipements de localisation de défauts d'isolation selon IEC 61557-9. Son champ d'application est essentiel pour garantir des performances fiables dans des configurations de test spécifiques, des conditions opérationnelles et des critères de performance des appareils connectés de manière permanente ou semi-permanente au système de distribution. Parmi les points forts de cette norme, on note son approche détaillée des configurations de test, qui assure que les équipements respectent des critères stricts pour leur bon fonctionnement dans divers environnements. Cette norme constitue également une mise à jour technique significative par rapport à la première édition de 2006, intégrant les évolutions apportées par la norme IEC 61326-1:2012, ce qui renforce sa pertinence dans le secteur des équipements pour la mesure et le contrôle. La norme EN 61326-2-4:2013 est particulièrement pertinente pour les fabricants et utilisateurs d'appareils de surveillance d'isolation, car elle garantit que ces équipements répondent à des exigences de qualité et de sécurité élevées. En standardisant les exigences et les performances, elle contribue à une meilleure protection contre les risques d'isolation, favorisant ainsi la sécurité des installations électriques. Grâce à ses spécifications techniques précises, cette norme joue un rôle crucial dans l'amélioration de la fiabilité et de l'efficacité des systèmes concernés.
Die Norm EN 61326-2-4:2013 behandelt essentielle Aspekte der elektrischen Ausrüstung zur Messung, Steuerung und Laboranwendung, insbesondere die elektromagnetische Verträglichkeit (EMV) von Geräten zur Isolationsüberwachung und zur Isolationsfehlersuche. Sie definiert detaillierte Prüfkonfigurationen, Betriebsbedingungen und Leistungskriterien für Isolationsüberwachungsgeräte gemäß IEC 61557-8 und Systeme zur Isolationsfehlersuche gemäß IEC 61557-9. Ein herausragendes Merkmal dieser Norm ist ihre umfassende Herangehensweise, die Geräte abdeckt, die dauerhaft oder semipermanente mit dem Verteilungssystem verbunden sind. Dies gewährleistet, dass die Anforderungen an die Sicherheit und Funktionalität dieser Geräte in realen Einsatzszenarien gründlich berücksichtigt werden. Zudem stellt die Norm sicher, dass alle relevanten Aspekte der EMV-Einhaltung in Bezug auf die getesteten Geräte abgedeckt sind, was für die Sicherheit im Betrieb von entscheidender Bedeutung ist. Die Initiative zur Aktualisierung der Norm, die die erste Ausgabe aus dem Jahr 2006 ersetzt, bringt bedeutende technische Änderungen mit sich, unter anderem die Anpassung an IEC 61326-1:2012. Diese Anpassungen sind besonders relevant, da sie die Norm auf den neuesten Stand der Technik bringen und sicherstellen, dass die Anforderungen den gegenwärtigen technischen Gegebenheiten und Standards entsprechen. Zusammengefasst bietet die EN 61326-2-4:2013 eine robuste Grundlage für die Evaluierung und den Einsatz von Isolationsüberwachungs- und Isolationsfehlersuchsystemen. Ihre Relevanz für Hersteller, Prüfinstitutionen und Anwender ist unbestritten, da sie die Einhaltung eines hohen Sicherheitsstandards gewährleistet, der im Rahmen moderner industrieller Anwendungen unerlässlich ist.
The EN 61326-2-4:2013 standard provides a crucial framework for the operational and performance criteria regarding insulation monitoring devices and insulation fault location systems, addressing the specifics outlined in IEC 61557-8 and IEC 61557-9. This document stands out for its comprehensive scope, detailing the necessary test configurations and operational conditions that ensure the reliability and safety of electrical equipment for measurement, control, and laboratory use. One of the primary strengths of this standard is its focus on insulation monitoring devices that are either permanently or semi-permanently connected to distribution systems. By delineating clear operational criteria, the standard facilitates compliance for manufacturers and ensures that end-users can trust in the performance and functionality of insulation fault location equipment. The technical revisions incorporated in this second edition significantly enhance its relevance. The update concerning IEC 61326-1:2012 is particularly noteworthy, as it ensures that the standard aligns with the latest EMC requirements, thus reinforcing the integrity of equipment operating within potentially challenging electrical environments. This alignment not only promotes safety but also aids manufacturers in achieving compliance with broader regulatory frameworks. Furthermore, the clarity provided in the definitions of test configurations and performance criteria enhances usability for industry professionals. By providing detailed guidance, EN 61326-2-4:2013 serves as an essential reference that aids in the proper assessment and assurance of insulation monitoring technologies, fostering innovation while maintaining high standards of quality and safety within the field. Ultimately, the EN 61326-2-4:2013 standard represents a pivotal development in the oversight and regulation of insulation monitoring devices and fault location techniques, ensuring that both manufacturers and users benefit from well-defined performance standards tailored to the complexities of electrical measurement and control environments. Its relevance in the current landscape of electrical safety and compliance continues to be paramount.
EN 61326-2-4:2013は、測定、制御、および実験室用の電気機器に関する標準であり、特にIEC 61557-8およびIEC 61557-9に基づく絶縁監視装置と絶縁故障位置特定装置のためのテスト構成、運用条件、性能基準を定めています。この標準の範囲は、絶縁監視装置および絶縁故障位置特定システムが分配システムに恒久的または半恒久的に接続されている場合に適用されます。 この文書の強みは、絶縁監視のための詳細なテスト構成と明確な運用条件を提供し、機器とシステムの性能基準を明確に定義している点です。特に、IEC 61557-8およびIEC 61557-9に基づく要求事項を満たすことに重点を置いているため、業界標準に従った高い信頼性を確保することができます。また、2012年版で更新された内容は、最新の技術基準に適合していることを保証し、技術的な改訂が行われている点も重要です。 EN 61326-2-4:2013は、特に絶縁監視と故障位置特定に関わる機器の選定や使用に対して、利用者にとって非常に大きな意義を持つ標準です。これにより、適切な安全性を確保し、運用効率を向上させることが可能となります。全体として、この標準は、関連する業界において重要なガイドラインを提供し、技術的な進歩と顧客の要求に応えるために必要不可欠な要素となっています。
SIST EN 61326-2-4:2013 문서는 측정, 제어 및 실험실 용 전기 장비를 위한 EMC 요구 사항의 특정 요구 사항을 다루고 있습니다. 이 표준의 주요 초점은 IEC 61557-8과 IEC 61557-9에 따른 절연 모니터링 장치 및 절연 결함 위치 장비의 테스트 구성, 운영 조건 및 성능 기준을 명확히 규명하는 것입니다. 이 표준의 범위는 IEC 61557-8에 따른 절연 모니터링 장치와 IEC 61557-9에 따른 절연 결함 위치 장치에 적용됩니다. 이는 배전 시스템에 영구적 혹은 반영구적으로 연결된 장치에 적합합니다. 최신판인 이 문서는 2006년에 발행된 첫 번째 판을 대체하며, 기술적 개정 사항을 포함하고 있습니다. 특히 IEC 61326-1:2012에 대한 업데이트가 중요한 기술적 변화로 설명됩니다. 이 표준의 강점은 복잡한 장비의 안전성과 신뢰성을 보장하는 데 필수적인 명확한 테스트 구성 및 운영 조건을 제공한다는 점입니다. 사용자는 이러한 세부 사항을 통해 절연 모니터링 장치와 절연 결함 위치 시스템의 성능을 효과적으로 평가하고, 필요한 경우 개선할 수 있습니다. 또한, 이 표준은 사용자가 신뢰할 수 있는 절연 모니터링 장치를 선택하는 데 있어 중요한 가이드라인을 제공합니다. SIST EN 61326-2-4:2013은 최근 기술 동향과 산업 표준에 부합하는 포괄적인 정보를 제공하며, 전기 장비의 신뢰성, 안전성 및 성능을 최적화하는 데 기여하는 중요한 문서로 자리 잡고 있습니다. 이러한 이유로 이 표준은 현재 및 미래의 전기 장비 설계 및 운영에 있어 매우 높은 적합성과 중요성을 지니고 있습니다.








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...