Evaluation of absorbed power density related to human exposure to radio frequency fields from wireless communication devices operating between 6 GHz and 300 GHz

IEC/IEEE TR 63572:2026 describes the computation and measurement techniques and test approaches for evaluating the local peak absorbed power density (pAPD) and peak spatial average absorbed (epithelial) power density (psAPD) induced in a human body from a wireless device transmitting in close proximity to the user at frequencies between 6 GHz and 300 GHz.
This document provides information on the testing of portable devices transmitting at distances close to the human body, such as mobile phones, tablets, wearable devices, etc. The information in this document is also relevant to exposure in the close proximity of base stations.

General Information

Status
Published
Publication Date
23-Mar-2026
Drafting Committee
JWG 12 - TC 106/JWG 12
Current Stage
PPUB - Publication issued
Start Date
24-Mar-2026
Completion Date
30-Jan-2026

Overview

IEC/IEEE TR 63572:2026 provides comprehensive guidance on evaluating the absorbed power density (APD) related to human exposure from wireless communication devices operating in the frequency range of 6 GHz to 300 GHz. Developed jointly by the International Electrotechnical Commission (IEC) and IEEE, this technical report addresses the measurement and computational techniques required to assess human exposure to radio frequency (RF) fields, especially from devices used in close proximity to the body such as mobile phones, tablets, and wearables. The document also provides methodologies applicable to exposure scenarios near base stations.

Complying with global health and safety guidelines, this standardization document highlights practical test approaches for quantifying both local peak absorbed power density (pAPD) and peak spatial average absorbed (epithelial) power density (psAPD), supporting manufacturers, test laboratories, and regulatory bodies tasked with ensuring the safety of emerging wireless technologies, including 5G and future generations.

Key Topics

  • Frequency Range Coverage: Techniques and considerations specifically for devices and scenarios operating within 6 GHz to 300 GHz, including millimeter-wave spectrum use cases.
  • Absorbed Power Density (APD): Definitions, significance, and relevance of APD as a basic restriction metric for electromagnetic field (EMF) exposure above 6 GHz.
  • Measurement and Computational Methods:
    • Use of measurement phantoms and computational body models.
    • Procedures for APD and psAPD evaluation.
    • Uncertainty evaluation and reporting of results.
  • Device Under Test (DUT) Configurations: Guidance on positioning, necessary test configurations for portable and wearable devices during measurements.
  • System Check and Validation: Steps for verifying the accuracy and reliability of measurement and computational methods.
  • Uncertainty Budgets: Techniques for calculating and considering uncertainty in both measurement and computational assessments.
  • Exposure Metrics: Discussion of physical metrics and averaging metrics including spatial and temporal averaging relevant to APD.

Applications

  • Portable Wireless Device Compliance: Ensures mobile phones, tablets, laptops, and wearable devices meet international RF exposure safety requirements, particularly for products operating above 6 GHz.
  • Base Station Proximity Assessment: Provides methodologies for evaluating absorbed power density for individuals in close proximity to base stations, supporting equipment manufacturers and network operators in regulatory compliance.
  • Regulatory and Certification Bodies: Supplies actionable testing frameworks necessary for product approval and market access in accordance with regional and global EMF safety standards.
  • Research and Product Development: Assists researchers and engineers in advancing new wireless technologies by offering standardized approaches to address human EMF exposure.
  • Consumer Safety: Supports development of safer wireless products, enhancing consumer confidence in rapidly evolving high-frequency wireless communications such as 5G, 6G, and beyond.

Related Standards

  • IEC/IEEE 62209-1528:2020: Specifies the assessment of specific absorption rate (SAR) for wireless devices up to 10 GHz, forming the base for APD assessment standards above this frequency.
  • IEC/IEEE 63195-1 and 63195-2: Define methods for assessment of incident power density (IPD) from 6 GHz to 300 GHz for portable and body-worn devices.
  • IEC 62232:2022: Provides procedures for evaluating EMF exposure from base stations in the 110 MHz to 300 GHz range.
  • ICNIRP Guidelines (2020): International exposure guidelines referencing APD as a basic restriction above 6 GHz.
  • IEEE Std C95.1™-2019: Contains requirements for equivalent assessment of epithelial power density above 6 GHz.
  • IEC PAS 63446: Outlines methods for converting SAR results into APD for the 6 GHz to 10 GHz range.

By following IEC/IEEE TR 63572:2026, organizations and professionals can ensure scientifically robust, internationally recognized assessment of human EMF exposure linked to the latest wireless communication technologies. This enhances product safety, regulatory compliance, and public trust in high-frequency wireless devices.

Buy Documents

Technical report

IEC/IEEE TR 63572:2026 - Evaluation of absorbed power density related to human exposure to radio frequency fields from wireless communication devices operating between 6 GHz and 300 GHz

ISBN:978-2-8327-1099-9
Release Date:24-Mar-2026
English language (109 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

BSMI (Bureau of Standards, Metrology and Inspection)

Taiwan's standards and inspection authority.

TAF Taiwan Verified

Sponsored listings

Frequently Asked Questions

IEC/IEEE TR 63572:2026 is a technical report published by the International Electrotechnical Commission (IEC). Its full title is "Evaluation of absorbed power density related to human exposure to radio frequency fields from wireless communication devices operating between 6 GHz and 300 GHz". This standard covers: IEC/IEEE TR 63572:2026 describes the computation and measurement techniques and test approaches for evaluating the local peak absorbed power density (pAPD) and peak spatial average absorbed (epithelial) power density (psAPD) induced in a human body from a wireless device transmitting in close proximity to the user at frequencies between 6 GHz and 300 GHz. This document provides information on the testing of portable devices transmitting at distances close to the human body, such as mobile phones, tablets, wearable devices, etc. The information in this document is also relevant to exposure in the close proximity of base stations.

IEC/IEEE TR 63572:2026 describes the computation and measurement techniques and test approaches for evaluating the local peak absorbed power density (pAPD) and peak spatial average absorbed (epithelial) power density (psAPD) induced in a human body from a wireless device transmitting in close proximity to the user at frequencies between 6 GHz and 300 GHz. This document provides information on the testing of portable devices transmitting at distances close to the human body, such as mobile phones, tablets, wearable devices, etc. The information in this document is also relevant to exposure in the close proximity of base stations.

IEC/IEEE TR 63572:2026 is classified under the following ICS (International Classification for Standards) categories: 17.220.20 - Measurement of electrical and magnetic quantities. The ICS classification helps identify the subject area and facilitates finding related standards.

IEC/IEEE TR 63572:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


IEC/IEEE TR 63572 ®
Edition 1.0 2026-03
TECHNICAL
REPORT
Evaluation of absorbed power density related to human exposure to radio
frequency fields from wireless communication devices operating between 6 GHz
and 300 GHz
ICS 17.220.20  ISBN 978-2-8327-1099-9

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or
by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing being
secured. Requests for permission to reproduce should be addressed to either IEC at the address below or IEC’s
member Natio nal Committee in the country of the requester or from IEEE.

IEC Secretariat Institute of Electrical and Electronics Engineers, Inc.
3, rue de Varembé 3 Park Avenue
CH-1211 Geneva 20 New York, NY 10016-5997
Switzerland United States of America
Tel.: +41 22 919 02 11 stds.ipr@ieee.org
info@iec.ch www.ieee.org
www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About the IEEE
IEEE is the world’s largest professional association dedicated to advancing technological innovation and excellence for the
benefit of humanity. IEEE and its members inspire a global community through its highly cited publications, conferences,
technology standards, and professional and educational activities.

About IEC/IEEE publications
The technical content of IEC/IEEE publications is kept under constant review by the IEC and IEEE. Please make sure that
you have the latest edition, a corrigendum or an amendment might have been published.

IEC publications search - webstore.iec.ch/advsearchform IEC Products & Services Portal - products.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews, graphical symbols and the glossary. With a
committee, …). It also gives information on projects, replaced subscription you will always have access to up to date content
and withdrawn publications. tailored to your needs.

IEC Just Published - webstore.iec.ch/justpublished
Electropedia - www.electropedia.org
Stay up to date on all new IEC publications. Just Published
The world's leading online dictionary on electrotechnology,
details all new publications released. Available online and
containing more than 22 500 terminological entries in English and
once a month by email.
French, with equivalent terms in 25 additional languages. Also

known as the International Electrotechnical Vocabulary (IEV)
IEC Customer Service Centre - webstore.iec.ch/csc
online.
If you wish to give us your feedback on this publication or need

further assistance, please contact the Customer Service
Centre: sales@iec.ch.
CONTENTS
FOREWORD . 7
INTRODUCTION . 9
1 Scope . 10
2 Normative references . 10
3 Terms and definitions . 10
4 Symbols and abbreviated terms . 14
4.1 Physical quantities . 14
4.2 Constants . 14
4.3 Abbreviated terms . 14
5 Absorbed power density (epithelial power density). 16
5.1 General . 16
5.2 Specification of the spatial-average absorbed power density . 16
5.3 Alternative definitions . 18
6 Fundamentals of evaluating APD . 19
6.1 General . 19
6.2 Relevant characteristics of the human body . 19
6.3 Evaluation of the APD . 20
7 Measurement and computational phantoms and body models . 20
7.1 Measurement phantom . 20
7.2 Computational phantom models . 22
7.3 Anatomical models of the human body . 22
8 Measurement methods . 22
8.1 General . 22
8.2 Measurement evaluation procedures . 23
8.3 DUT position and testing configurations . 23
8.4 System check and validation . 23
8.4.1 System check . 23
8.4.2 System validation . 23
9 Computational methods . 23
9.1 General . 23
9.2 Model validation . 24
9.3 Evaluation of the averaged APD . 25
9.3.1 Planar surfaces . 25
9.3.2 Non-planar surfaces . 26
10 Uncertainty evaluation . 26
10.1 Measurement uncertainty . 26
10.1.1 General . 26
10.1.2 Calibration uncertainty . 27
10.1.3 Probe/scanning uncertainty . 28
10.1.4 Phantom uncertainty . 28
10.1.5 Postprocessing uncertainty . 28
10.1.6 DUT uncertainty. 29
10.2 Computational uncertainty . 29
10.2.1 General . 29
10.2.2 Uncertainty contributions due to the computational parameters . 30
10.2.3 Uncertainty contribution of the computational representation of the DUT
model . 31
10.2.4 Uncertainty contribution of the computational representation of the
phantom model . 32
10.2.5 Uncertainty of the maximum exposure evaluation . 33
10.2.6 Uncertainty budget . 33
10.3 Combined measurement and computational uncertainty . 34
11 Reporting . 35
11.1 General . 35
11.2 Items to be recorded in measurement exposure evaluation reports . 35
11.3 Items recorded in computational exposure evaluation reports . 36
Annex A (informative) Human skin from EMF exposure perspective . 37
A.1 Structure of the human skin from the perspective of EMF exposure . 37
A.2 Dielectric properties of the skin tissues . 37
A.3 Tissue-equivalent and reflectivity-based human models . 37
Annex B (informative) Exposure in terms of the absorbed power density. 40
B.1 Previous work . 40
B.1.1 Incident and transmitted power density in one-dimensional skin models . 40
B.1.2 Three-dimensional modelling of the absorbed power density . 40
B.1.3 Heating factor . 41
B.1.4 Measurement studies. 42
B.2 Relationship with incident power density . 42
Annex C (informative) Review of the metrics relevant to the absorbed power density . 43
C.1 General . 43
C.2 Physical metrics . 43
C.2.1 Absorbed power . 43
C.2.2 Transmitted power . 44
C.2.3 Reflected power. 44
C.2.4 Temperature . 47
C.3 Averaging metrics . 48
C.3.1 Evaluation surface . 48
C.3.2 Spatial averaging . 48
C.3.3 Temporal averaging . 50
Annex D (informative) Absorbed power density for modulated signals . 51
Annex E (informative) APD assessment using conventional SAR measurement
systems . 54
Annex F (informative) APD assessment using E-field probes and band-limited
phantoms . 55
F.1 General . 55
F.2 Phantom design . 55
F.3 Probe design . 58
F.4 Probe calibration . 59
F.5 Induced E-field scanning and APD reconstruction . 60
F.6 Uncertainty . 60
F.7 Validation. 62
Annex G (informative) APD evaluation based on free-space E-field measurement and
reflectivity-based phantom . 64
G.1 General . 64
G.2 Reflectivity-based phantoms with enhanced transmission . 65
G.3 Reconstruction of APD from the measured E-field . 66
G.4 Measurement system . 68
G.5 Validation using reference antennas . 68
G.6 Uncertainty . 72
Annex H (informative) APD evaluation from infrared-based temperature
measurements using reflectivity-based phantoms . 74
H.1 General . 74
H.2 Infrared imaging . 75
H.3 Reflectivity-based phantoms optimised for infrared measurements . 75
H.4 Reconstruction of APD from thermal measurements . 76
H.5 Measurement system . 79
H.6 Application examples . 80
H.7 Validation using reference antennas . 81
H.8 Uncertainty . 84
Annex I (informative) APD assessment by backward transformation from the
measured E-field behind the low-loss phantom . 86
I.1 General . 86
I.2 Low-loss phantom . 86
I.3 Reconstruction of the APD by backward transformation . 86
Annex J (informative) APD assessment using hybrid electromagnetic and thermal
measurements . 88
Annex K (informative) Probe technologies . 89
K.1 General . 89
K.2 Temperature-based probes . 89
K.3 Radio-frequency electromagnetic probes . 89
Annex L (informative) APD assessment using measurements of incident electric field
with source and phantom modelling . 90
L.1 Methodology . 90
L.2 Validation of OTAA for incident power density measurements . 91
L.3 Validation of the OTAA for APD measurements . 93
L.4 Practical use case application . 94
L.5 Measurement of devices using digitally modulated signals . 98
Annex M (informative) Validation of the measurement methods . 99
M.1 General . 99
M.2 Validation sources and target values . 99
M.2.1 Cavity-fed dipole arrays . 99
M.2.2 Pyramidal horns with slot arrays . 100
Annex N (informative) Computational algorithms for the calculation of the APD . 101
N.1 FDTD . 101
N.2 FEM . 101
Annex O (informative) Code verification . 102
Bibliography . 103

Figure 1 – Minimum size of the planar APD evaluation surface . 22
Figure A.1 – Reflection and transmission of an electromagnetic wave from the human
body and phantoms above 6 GHz . 38
Figure C.1 – Incident power density as a function of separation distance d and an
antenna input power of 23 dBm. 45
Figure C.2 – Absorbed power density as a function of separation distance d and an
antenna input power of 23 dBm. 46
Figure C.3 – Distribution of the psAPD at a separation distance d = 10 mm . 46
Figure C.4 – Power transmission coefficients as a function of antenna to skin
separation distance d . 47
Figure D.1 – Measurement results in time domain and converted to frequency domain . 51
Figure D.2 – Sine wave associated with a waveform CW vs modulated signal . 52
Figure D.3 – Waveform power density integrated over the bandwidth . 52
Figure D.4 – Signal-to-noise ratio for same waveform with different measurement
settings: 40 dB (top), 20 dB (centre), and 10 dB (bottom) . 53
Figure F.1 – Cross-section of the phantom for assessing APD in the skin and skin

model . 56
Figure F.2 – Comparison of the APD in the phantom and the layered skin surface using
plane-waves versus normalized x component of wave vector (k /k where k is the free
x 0 0
space wave vector) . 57
Figure F.3 – psAPD in dB (0 dB [1,0 times] corresponding to 1 W/m ) averaged over
1 cm of the skin surface as well as in the phantom as a function of the separation
between the dipole antenna and the phantom surface . 58
Figure F.4 – Measurement validation of the APD probe axial receiving pattern . 59
Figure F.5 – Calibration setup using an open-ended waveguide radiating into the TSL . 59
Figure F.6 – Decay of the APD probe signal inside the calibration setup matched to the
computational and analytical target by the probe sensitivity factor; the sensitivity limit
(noise floor) of the APD probe is also shown . 60
Figure F.7 – Cosine similarity of the reflection coefficient of a standard gain horn
computed and measured at varied distances in front of the APD phantom . 63
Figure F.8 – Comparison of the measured and computationally determined APD of a
standard gain horn at 30 GHz in touch against the APD phantom . 63
Figure G.1 – Schematic representation of the APD assessment method based on free-
space E-field vector measurement, using a reflectivity-based phantom . 64
Figure G.2 – Reflectivity-based phantom characteristics vs incidence angle at 60 GHz
for TM and TE polarizations . 65
Figure G.3 – Schematic representation of the measurement setup . 68
Figure G.4 – Reference antennas (see IEC/IEEE 63195-1) used for validation . 69
Figure G.5 – Dielectric loading of DUT by reflectivity-based phantom and skin . 69
Figure G.6 – APD distribution for the reference slotted horn antenna at 60 GHz . 70
Figure G.7 – APD as a function of the evaluation distance d for the horn antenna . 71
Figure H.1 – Schematic representation of the APD assessment method based on
IR-based heat dynamics measurement of a phantom of thickness w at distance d . 75
Figure H.2 – The power reflection coefficient of the phantom for TM and TE
polarizations at 60 GHz. 76
Figure H.3 – Schematic representation of the measurement setup . 80
Figure H.4 – Example of APD spatial distributions at z = 0 plane retrieved from IR
measurements for a 2 × 2 patch antenna array . 81
Figure H.5 – Example of spatial distributions of APD in the z = 0 mm plane retrieved
from IR measurements for a conical horn antenna . 81
Figure H.6 – APD distribution for the horn antenna at 60 GHz; the accepted power is
20 mW . 83
Figure H.7 – APD as a function of d for the reference horn antenna: The results are
normalized to the antenna input power of 10 mW . 84
Figure L.1 – Flowchart of the procedure for measurements of the absorbed power
density by means of an OTA augmented approach . 91
Figure L.2 – OTA spherical scanning measurement setups used in [85] . 92
Figure L.3 – Comparison of the psAPD for the cavity-fed dipole array
(IEC/IEEE 63195-1) at 30 GHz obtained by means of the OTAA approach compared
with full-wave computations . 94
Figure L.4 – Comparison of the psAPD for the slotted-horn array (IEC/IEEE 63195-1)
at 30 GHz obtained by means of the OTAA approach compared with full-wave
computations . 94
Figure L.5 – Model of a portable millimetre-wave device and example of far-field
directivity . 95
Figure L.6 – Comparison of the psAPD for a mmW portable device obtained by means
of the OTAA approach compared to full-wave computations . 96
Figure L.7 – Comparison of the psAPD for a mmW portable device obtained by means
of the OTAA approach compared to full-wave computations . 97
Figure L.8 – Comparison of the spatial averaged absorbed power density distribution
(1 cm ) on the phantom surface for the portable device at 1,25 mm distance . 97

Table 1 – An example of budget of the uncertainty contributions of measurement
evaluations . 27
Table 2 – Budget of the uncertainty contributions of the computational algorithm for the
validation setup or testing setup . 30
Table 3 – Budget of the uncertainty of the developed model of the setup . 33
Table 4 – Computational uncertainty budget . 34
Table 5 – Example of uncertainty budget of the exposure assessment combining
measurements and computational methods . 35
Table D.1 – 5G bands . 52
Table F.1 – Thickness and dielectric properties (30 GHz) of the skin tissue layer model
used to estimate APD in the skin . 56
Table F.2 – Properties of the optimized dosimetric phantom emulating the APD in the
human skin from 24 GHz to 33 GHz . 57
Table F.3 – Changes in the mean value of the ratio of APD in the phantom and the
layered skin model for all angles of incidence . 58
Table F.4 – APD bandlimited phantom uncertainty budget . 61
Table G.1 – ∆ (i) [%] for free space E-field method . 72
j
Table G.2 – APD uncertainty budget for free-space E-field method . 73
Table H.1 – ∆ (i) [%] for infrared-based temperature method . 84
j
Table H.2 – APD uncertainty budget for infrared-based temperature method . 85
Table L.1 – psIPD values for the 30 GHz HSA as defined in IEC/IEEE 63195-1 . 92
Table L.2 – psIPD values for the 60 GHz HSA as defined in IEC/IEEE 63195-1 . 93
Table L.3 – psIPD values for the 30 GHz CDA as defined in IEC/IEEE 63195-1 . 93
Table L.4 – psIPD values for the 60 GHz CDA as defined in IEC/IEEE 63195-1 . 93
Table L.5 – Dielectric phantom parameters for the evaluation of the absorbed power
density with the OTA augmented approach . 93
Table L.6 – Comparison psIPD values obtained for the portable device mock-up based
on full-wave computations and OTA augmented approach . 95
Table M.1 – Target values for the cavity-fed dipole array at 10 GHz, 30 GHz, 60 GHz,
and 90 GHz computed at a distance of 2 mm in front of the reference skin model
normalized to a TRP of 0 dBm . 99
Table M.2 – Target values for the pyramidal horns with slot arrays at 10 GHz, 30 GHz,
60 GHz and 90 GHz computed at a distance of 2 mm in front of the reference skin
model normalized to a TRP of 0 dBm . 100

Evaluation of absorbed power density related to human exposure
to radio frequency fields from wireless communication
devices operating between 6 GHz and 300 GHz

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC document(s)"). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation.
IEEE Standards documents are developed within IEEE Societies and subcommittees of IEEE Standards
Association (IEEE SA) Board of Governors. IEEE develops its standards through an accredited consensus
development process, which brings together volunteers representing varied viewpoints and interests to achieve
the final product. IEEE standards are documents developed by volunteers with scientific, academic, and industry-
based expertise in technical working groups. Volunteers are not necessarily members of IEEE or IEEE SA and
participate without compensation from IEEE. While IEEE administers the process and establishes rules to promote
fairness in the consensus development process, IEEE does not independently evaluate, test, or verify the
accuracy of any of the information or the soundness of any judgments contained in its standards. IEEE documents
are made available for use subject to important notices and legal disclaimers (see
https://standards.ieee.org/ipr/disclaimers.html for more information).
IEC collaborates closely with IEEE in accordance with conditions determined by agreement between the two
organizations. This Dual Logo International Standard was jointly developed by the IEC and IEEE under the terms
of that agreement.
2) The formal decisions of IEC on technical matters express, as nearly as possible, an international consensus of
opinion on the relevant subjects since each technical committee has representation from all interested IEC
National Committees. The formal decisions of IEEE on technical matters, once consensus within IEEE Societies
and Standards Coordinating Committees has been reached, is determined by a balanced ballot of materially
interested parties who indicate interest in reviewing the proposed standard. Final approval of the IEEE standards
document is given by the IEEE Standards Association (IEEE SA) Standards Board.
3) IEC/IEEE Publications have the form of recommendations for international use and are accepted by IEC National
Committees/IEEE Societies in that sense. While all reasonable efforts are made to ensure that the technical
content of IEC/IEEE Publications is accurate, IEC or IEEE cannot be held responsible for the way in which they
are used or for any misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
(including IEC/IEEE Publications) transparently to the maximum extent possible in their national and regional
publications. Any divergence between any IEC/IEEE Publication and the corresponding national or regional
publication shall be clearly indicated in the latter.
5) IEC and IEEE do not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC and IEEE are not responsible
for any services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or IEEE or their directors, employees, servants or agents including individual
experts and members of technical committees and IEC National Committees, or volunteers of IEEE Societies and
the Standards Coordinating Committees of the IEEE Standards Association (IEEE SA) Standards Board, for any
personal injury, property damage or other damage of any nature whatsoever, whether direct or indirect, or for
costs (including legal fees) and expenses arising out of the publication, use of, or reliance upon, this IEC/IEEE
Publication or any other IEC or IEEE Publications.
8) Attention is drawn to the normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that implementation of this IEC/IEEE Publication may require use of material
covered by patent rights. By publication of this standard, no position is taken with respect to the existence or
validity of any patent rights in connection therewith. IEC or IEEE shall not be held responsible for identifying
Essential Patent Claims for which a license may be required, for conducting inquiries into the legal validity or
scope of Patent Claims or determining whether any licensing terms or conditions provided in connection with
submission of a Letter of Assurance, if any, or in any licensing agreements are reasonable or non-discriminatory.
Users of this standard are expressly advised that determination of the validity of any patent rights, and the risk
of infringement of such rights, is entirely their own responsibility.
IEC/IEEE TR 63572 was prepared by IEC technical committee 106: Methods for the assessment
of electric, magnetic and electromagnetic fields associated with human exposure, in
cooperation with the International Committee on Electromagnetic Safety of the IEEE Standards
Association, under the IEC/IEEE Dual Logo Agreement between IEC and IEEE. It is a Technical
Report.
This document is published as an IEC/IEEE Dual Logo standard.
The text of this Technical Report is based on the following IEC documents:
Draft Report on voting
106/705/DTR 106/720/RVDTR
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
The language used for the development of this Technical Report is English.
This document was drafted in accordance with the rules given in the ISO/IEC Directives, Part 2,
available at www.iec.ch/members_experts/refdocs. The main document types developed by IEC
are described in greater detail at www.iec.ch/publications/.
The IEC Technical Committee and IEEE Technical Committee have decided that the contents
of this document will remain unchanged until the stability date indicated on the IEC website
under webstore.iec.ch in the data related to the specific document. At this date, the document
will be
– reconfirmed,
– withdrawn, or
– revised.
INTRODUCTION
This document describes computational and measurement methods for the evaluation of
absorbed (epithelial) power density related to human exposures due to electromagnetic field
(EMF) transmitting devices operating in close proximity to the user at frequencies between
6 GHz and 300 GHz. The types of devices include but are not limited to mobile telephones,
tablets, laptops, etc.
For portable devices, the specific absorption rate (SAR) assessment standard for wireless
devices used in close proximity to the user IEC/IEEE 62209-1528:2020 [1] is specified up to
10 GHz. The IEC/IEEE 63195-1 and IEC/IEEE 63195-2 standards on the assessment of the
incident power density (IPD) for wireless devices used in close proximity to the user are valid
from 6 GHz to 300 GHz. For exposure to EMF emitted from base stations, IEC 62232:2022 [2]
specifies methods to assess the compliance boundaries based on reference levels and basic
restrictions for a frequency range from 110 MHz to 300 GHz.
The absorbed power density (APD) is considered as the relevant local basic restriction and
exposure metric above 6 GHz in the ICNIRP 2020 guidelines [3] and in the Health Canada
Notice [4]. Similarly, IEEE Std C95.1™-2019 [5] requires equivalent assessment of epithelial
power density above 6 GHz. IEC PAS 63446 [6] describes methods to convert SAR results into
APD in frequency range of 6 GHz to 10 GHz.
IEC TC 106 and IEEE ICES TC 34 (IEC/IEEE) have previously noted the necessity to extend
compliance assessment standards for portable devices to cover the basic restrictions on APD.
To ensure timely publication of the available knowledge on the computational and measurement
technologies on APD assessment, IEC TC 106 and IEEE ICES TC 34 decided on a two-step
strategy to ensure that the fundamental approaches are available.
In 2023 and 2024, the focus was on the development of a Technical Report (this document),
specifying the state of the art of computational and measurement techniques and test
approaches for evaluating portable devices based on absorbed power density measurements
from 6 GHz to 300 GHz.
Upon drafting this document, a Technical Report, a new work item proposal has been initiated
to develop a Dual Logo International Standard (IS) jointly among IEEE and IEC on the
computational and measurement procedures based on leveraging the content of this document.
This document is an informative document that serves as the starting point for the International
Standards on computational and measurement assessment procedures of the APD. The
methodologies and approaches described in this document can be useful for the assessment
of APD in the early phase of computational and measurement technology development. It also
contains recommendations for future standardization work and highlights areas that require
additional investigation or consideration.

1 Scope
This document describes the computation and measurement techniques and test approaches
for evaluating the local peak absorbed power density (pAPD) and peak spatial average
absorbed (epithelial) power density (psAPD) induced in a human body from a wireless device
transmitting in close proximity to the user at frequencies between 6 GHz and 300 GHz.
This document provides information on the testing of portable devices transmitting at distances
close to the human body, such as mobile phones, tablets, wearable devices, etc. The
information in this document is also relevant to exposure in the close proximity of base stations.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies.
For undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC/IEEE 63195-1, Assessment of power density of human exposure to radio frequency fields
from wireless devices in close proximity to the head and body (frequency range of 6 GHz to
300 GHz) - Part 1: Measurement procedure
IEC/IEEE 63195-2, Assessment of power density of human exposure to radio frequency fields
from wireless devices in close proximity to the head and body (frequency range of 6 GHz to
300 GHz) - Part 2: Computational procedure
3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC/IEEE 63195-1,
IEC/IEEE 63195-2, and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following
addresses:
– IEC Electropedia: available at https://www.electropedia.org/
– ISO Online browsing platform: available at https://www.iso.org/obp
– IEEE Dictionary Online: available at http://dictionary.ieee.org
3.1
incident power density
IPD
function of the complex Poynting vector S at the location r
Note 1 to entry: The formula to compute IPD from S depends on the applicable exposure guidelines or national
regulations.
Note 2 to entry: IEC/IEEE 63195-1:2022 and IEC/IEEE 63195-2:2022 refer to the incident power density and all
derived quantities as power density.
3.2
absorbed power density
APD
epithelial power density
function of the Poynting vector (Re{S}) that is transmitted into a phantom or a human body and
integrated over a surface to compute the sAPD (3.6)
Note 1 to entry: See Clause 5 for details and formulae for the APD.
3.3
body model
phantom
mock-up that represents the exposed human body or a part of it in a measurement set-up or in
a simulation
3.4
spatial-average incident power density
sIPD
IPD (3.1) averaged over an averaging area
Note 1 to entry: sIPD is a function of the location. It is determined on the evaluation surface (3.13).
2 2
Note 2 to entry: Example averaging area sizes specified in exposure guidelines are 1 cm and/or 4 cm .
Note 3 to entry: IEC/IEEE 63195-1:2022 and IEC/IEEE 63195-2:2022 refer to the spatial-average incident power
density as spatial-average power density.
3.5
peak absorbed power density
pAPD
local maximum of the APD (3.2)
3.6
spatial-average absorbed power density
sAPD
APD (3.2) averaged over an averaging area (3.12)
Note 1 to entry: sAPD is a function of the location. It is determined on the evaluation surface (3.13).
2 2
Note 2 to entry: Example averaging area sizes specified in exposure guidelines are 1 cm and/or 4 cm .
3.7
peak spatial-average incident power density
psIPD
global maximum value of the sIPD (3.4) on the evaluation surface (3.13)
Note 1 to entry: Other local maxima (i.e. secondary peak spatial-average power density values) can exist.
Note 2 to entry: IEC/IEEE 63195-1:2022 and IEC/IEEE 63195-2:2022 refer to the spatial-average incident power
density as spatial-average power density.
3.8
peak spatial-average absorbed power density
psAPD
global maximum value of the sAPD (3.6) on the evaluation surface (3.13)
Note 1 to entry: Other local maxima (i.e. se
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...