ASTM G220-20
(Practice)Standard Practice for Replacing Saturated Calomel Reference Electrode (SCE) for Measuring Electrode Potentials
Standard Practice for Replacing Saturated Calomel Reference Electrode (SCE) for Measuring Electrode Potentials
SIGNIFICANCE AND USE
5.1 Electrode potential measurement is an essential step in many electrochemical corrosion test methods and practices.
5.2 The SCE has been widely specified for laboratory tests because it is reliable, accurate, simple to use, and it has been readily available. However, because this device contains mercury, and mercury has recently been banned by some governmental agencies, it may not be available in locations where mercury is banned. As a result, test methods using the SCE may not be possible in locations where mercury is banned. Therefore, it is necessary to establish an alternative reference electrode for these standards.
5.3 The KCl saturated, silver/silver chloride reference electrode, Ag/AgCl (sat KCl), is a satisfactory replacement for the SCE in laboratory test methods and practices. This reference electrode should provide comparable performance and accuracy to the SCE.
5.4 It will be necessary to carry out interlaboratory test programs for each test method where the Ag/AgCl (sat KCl) electrode is specified to replace the SCE, in order to develop the precision of the method using the new reference electrode.
Note 1: In cases where a test method specifies that either the SCE or the Ag/AgCl (sat KCl) reference electrode may be used, and the ILS program used to develop the precision of the method included significant numbers of participants using each of these reference electrodes, no additional testing is required.
SCOPE
1.1 This practice provides the steps necessary to prepare revisions to standards that specify the saturated calomel reference electrode (SCE) for measuring or controlling electrode potentials.
1.2 The SCE contains mercury and, as a result, it may not be available in locations where mercury has been banned. This practice covers the selection and use of an alternative reference electrode.
1.3 In test methods where the SCE is specified, it will be necessary to develop a new precision statement using the alternative reference electrode.
1.4 This practice will not apply to electrometric pH determinations where the SCE has been used. (Electrometric pH measurement is an analytical method that is covered elsewhere.)
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 Warning—Mercury has been designated by many regulatory agencies as a hazardous substance that can cause serious medical issues. Mercury, or its vapor, has been demonstrated to be hazardous to health and corrosive to materials. Use caution when handling mercury and mercury-containing products. See the applicable product Safety Data Sheet (SDS) for additional information. The potential exists that selling mercury or mercury-containing products, or both, is prohibited by local or national law. Users must determine legality of sales in their location.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
- Status
- Published
- Publication Date
- 31-Oct-2020
- Technical Committee
- G01 - Corrosion of Metals
- Drafting Committee
- G01.11 - Electrochemical Measurements in Corrosion Testing
Relations
- Effective Date
- 01-May-2019
- Effective Date
- 01-Jan-2017
- Effective Date
- 01-May-2016
- Effective Date
- 15-Dec-2014
- Effective Date
- 01-Dec-2013
- Effective Date
- 01-May-2013
- Effective Date
- 01-Nov-2011
- Effective Date
- 01-May-2010
- Effective Date
- 01-Oct-2008
- Effective Date
- 01-Nov-2005
- Effective Date
- 01-Nov-2004
- Effective Date
- 10-Aug-1999
- Effective Date
- 10-May-1999
Overview
ASTM G220-20 - Standard Practice for Replacing Saturated Calomel Reference Electrode (SCE) for Measuring Electrode Potentials provides essential guidance for laboratories and organizations that utilize electrochemical test methods for corrosion testing. Traditionally, the saturated calomel reference electrode (SCE) has been the standard for measuring or controlling electrode potentials due to its reliability and precision. However, the use of mercury in SCEs has led to restrictions and bans in several regions due to environmental and safety concerns. ASTM G220-20 addresses this challenge by establishing a rigorous process for replacing the SCE with a mercury-free alternative: the KCl-saturated silver/silver chloride (Ag/AgCl, sat KCl) reference electrode.
This standard is relevant for laboratories seeking compliance with local regulations, environmental safety policies, and international quality practices within electrochemical testing.
Key Topics
Mercury Restrictions and SCE Limitations:
Due to global initiatives to minimize mercury usage, laboratories may no longer be able to use SCEs for measuring electrode potentials. ASTM G220-20 details the implications of such restrictions for corrosion test methods.Alternative Reference Electrode Selection:
The standard identifies the Ag/AgCl (sat KCl) reference electrode as a suitable and comparable alternative for most laboratory applications, offering similar accuracy and reliability as the SCE.Conversion of Electrode Potential Values:
When transferring existing test methods to the new reference electrode, all electrode potential readings previously referenced to the SCE must be adjusted by adding 0.045 V to align with the Ag/AgCl (sat KCl) scale.Verification, Precision, and Maintenance:
ASTM G220-20 outlines procedures for:- Verifying performance of the Ag/AgCl (sat KCl) electrode using ASTM Guide G215 and comparing to SCE master electrodes.
- Requiring interlaboratory test programs (per ASTM E691) to establish new precision statements for methods that incorporate the replacement reference electrode.
- Following established maintenance and validation protocols for reference electrodes to uphold testing integrity.
Health, Safety, and Compliance:
The standard emphasizes user responsibility for safe handling of mercury-containing materials and strict adherence to relevant laws and environmental safety standards.
Applications
Corrosion Testing Laboratories:
Laboratories engaged in electrochemical corrosion tests now have a clear, standardized path to phase out SCEs in favor of Ag/AgCl (sat KCl) electrodes, ensuring continued compliance with regulatory and safety requirements.Standards Development:
Organizations updating or revising their internal or published standards pertaining to electrode potentials can use ASTM G220-20 to amend references from SCE to Ag/AgCl (sat KCl), ensuring test methods remain accurate and legally compliant.Quality Assurance and Regulatory Compliance:
By following the practice set out in ASTM G220-20, testing facilities can verify the validity, accuracy, and consistency of electrode potential measurements while adhering to evolving environmental legislation and global trade policies.
Related Standards
- ASTM G3: Practice for Conventions Applicable to Electrochemical Measurements in Corrosion Testing
- ASTM G215: Guide for Electrode Potential Measurement
- ASTM E691: Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
- NACE/ASTM G193: Terminology and Acronyms Relating to Corrosion
Keywords: Ag/AgCl (sat KCl), saturated calomel reference electrode, SCE replacement, reference electrode standard, electrode potentials, electrochemical corrosion testing, mercury-free electrode, ASTM G220, laboratory compliance, corrosion standards.
By implementing ASTM G220-20, laboratories and organizations ensure robust, compliant, and environmentally responsible electrochemical measurement practices for years to come.
Buy Documents
ASTM G220-20 - Standard Practice for Replacing Saturated Calomel Reference Electrode (SCE) for Measuring Electrode Potentials
Get Certified
Connect with accredited certification bodies for this standard

ECOCERT
Organic and sustainability certification.

Element Materials Technology
Materials testing and product certification.
National Aerospace and Defense Contractors Accreditation Program (NADCAP)
Global cooperative program for special process quality in aerospace.
Sponsored listings
Frequently Asked Questions
ASTM G220-20 is a standard published by ASTM International. Its full title is "Standard Practice for Replacing Saturated Calomel Reference Electrode (SCE) for Measuring Electrode Potentials". This standard covers: SIGNIFICANCE AND USE 5.1 Electrode potential measurement is an essential step in many electrochemical corrosion test methods and practices. 5.2 The SCE has been widely specified for laboratory tests because it is reliable, accurate, simple to use, and it has been readily available. However, because this device contains mercury, and mercury has recently been banned by some governmental agencies, it may not be available in locations where mercury is banned. As a result, test methods using the SCE may not be possible in locations where mercury is banned. Therefore, it is necessary to establish an alternative reference electrode for these standards. 5.3 The KCl saturated, silver/silver chloride reference electrode, Ag/AgCl (sat KCl), is a satisfactory replacement for the SCE in laboratory test methods and practices. This reference electrode should provide comparable performance and accuracy to the SCE. 5.4 It will be necessary to carry out interlaboratory test programs for each test method where the Ag/AgCl (sat KCl) electrode is specified to replace the SCE, in order to develop the precision of the method using the new reference electrode. Note 1: In cases where a test method specifies that either the SCE or the Ag/AgCl (sat KCl) reference electrode may be used, and the ILS program used to develop the precision of the method included significant numbers of participants using each of these reference electrodes, no additional testing is required. SCOPE 1.1 This practice provides the steps necessary to prepare revisions to standards that specify the saturated calomel reference electrode (SCE) for measuring or controlling electrode potentials. 1.2 The SCE contains mercury and, as a result, it may not be available in locations where mercury has been banned. This practice covers the selection and use of an alternative reference electrode. 1.3 In test methods where the SCE is specified, it will be necessary to develop a new precision statement using the alternative reference electrode. 1.4 This practice will not apply to electrometric pH determinations where the SCE has been used. (Electrometric pH measurement is an analytical method that is covered elsewhere.) 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 Warning—Mercury has been designated by many regulatory agencies as a hazardous substance that can cause serious medical issues. Mercury, or its vapor, has been demonstrated to be hazardous to health and corrosive to materials. Use caution when handling mercury and mercury-containing products. See the applicable product Safety Data Sheet (SDS) for additional information. The potential exists that selling mercury or mercury-containing products, or both, is prohibited by local or national law. Users must determine legality of sales in their location. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
SIGNIFICANCE AND USE 5.1 Electrode potential measurement is an essential step in many electrochemical corrosion test methods and practices. 5.2 The SCE has been widely specified for laboratory tests because it is reliable, accurate, simple to use, and it has been readily available. However, because this device contains mercury, and mercury has recently been banned by some governmental agencies, it may not be available in locations where mercury is banned. As a result, test methods using the SCE may not be possible in locations where mercury is banned. Therefore, it is necessary to establish an alternative reference electrode for these standards. 5.3 The KCl saturated, silver/silver chloride reference electrode, Ag/AgCl (sat KCl), is a satisfactory replacement for the SCE in laboratory test methods and practices. This reference electrode should provide comparable performance and accuracy to the SCE. 5.4 It will be necessary to carry out interlaboratory test programs for each test method where the Ag/AgCl (sat KCl) electrode is specified to replace the SCE, in order to develop the precision of the method using the new reference electrode. Note 1: In cases where a test method specifies that either the SCE or the Ag/AgCl (sat KCl) reference electrode may be used, and the ILS program used to develop the precision of the method included significant numbers of participants using each of these reference electrodes, no additional testing is required. SCOPE 1.1 This practice provides the steps necessary to prepare revisions to standards that specify the saturated calomel reference electrode (SCE) for measuring or controlling electrode potentials. 1.2 The SCE contains mercury and, as a result, it may not be available in locations where mercury has been banned. This practice covers the selection and use of an alternative reference electrode. 1.3 In test methods where the SCE is specified, it will be necessary to develop a new precision statement using the alternative reference electrode. 1.4 This practice will not apply to electrometric pH determinations where the SCE has been used. (Electrometric pH measurement is an analytical method that is covered elsewhere.) 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 Warning—Mercury has been designated by many regulatory agencies as a hazardous substance that can cause serious medical issues. Mercury, or its vapor, has been demonstrated to be hazardous to health and corrosive to materials. Use caution when handling mercury and mercury-containing products. See the applicable product Safety Data Sheet (SDS) for additional information. The potential exists that selling mercury or mercury-containing products, or both, is prohibited by local or national law. Users must determine legality of sales in their location. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM G220-20 is classified under the following ICS (International Classification for Standards) categories: 25.220.99 - Other treatments and coatings; 71.040.50 - Physicochemical methods of analysis; 77.060 - Corrosion of metals. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM G220-20 has the following relationships with other standards: It is inter standard links to ASTM G3-14(2019), ASTM G215-17, ASTM G215-16, ASTM G3-14, ASTM G3-13, ASTM E691-13, ASTM E691-11, ASTM G3-89(2010), ASTM E691-08, ASTM E691-05, ASTM G3-89(2004), ASTM G3-89(1999), ASTM E691-99. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM G220-20 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: G220 − 20
Standard Practice for
Replacing Saturated Calomel Reference Electrode (SCE) for
Measuring Electrode Potentials
This standard is issued under the fixed designation G220; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.8 This international standard was developed in accor-
dance with internationally recognized principles on standard-
1.1 This practice provides the steps necessary to prepare
ization established in the Decision on Principles for the
revisions to standards that specify the saturated calomel
Development of International Standards, Guides and Recom-
reference electrode (SCE) for measuring or controlling elec-
mendations issued by the World Trade Organization Technical
trode potentials.
Barriers to Trade (TBT) Committee.
1.2 TheSCEcontainsmercuryand,asaresult,itmaynotbe
available in locations where mercury has been banned. This
2. Referenced Documents
practicecoverstheselectionanduseofanalternativereference
2.1 ASTM Standards:
electrode.
E691Practice for Conducting an Interlaboratory Study to
1.3 In test methods where the SCE is specified, it will be
Determine the Precision of a Test Method
necessary to develop a new precision statement using the
G3Practice for Conventions Applicable to Electrochemical
alternative reference electrode.
Measurements in Corrosion Testing
G215Guide for Electrode Potential Measurement
1.4 This practice will not apply to electrometric pH deter-
minations where the SCE has been used. (Electrometric pH 2.2 NACE/ASTM Standard:
measurement is an analytical method that is covered else- G193Terminology and Acronyms Relating to Corrosion
where.)
3. Terminology
1.5 The values stated in SI units are to be regarded as
3.1 Definitions—For definitions of terms used in this
standard. No other units of measurement are included in this
practice, see NACE/ASTM Terminology G193.
standard.
1.6 Warning—Mercury has been designated by many regu-
4. Summary of Practice
latory agencies as a hazardous substance that can cause
4.1 IntestmethodsandpracticeswheretheSCEisspecified
serious medical issues. Mercury, or its vapor, has been
for electrode potential measurement or control, the KCl satu-
demonstrated to be hazardous to health and corrosive to
rated silver/silver-chloride reference electrode, Ag/AgCl (sat
materials. Use caution when handling mercury and mercury-
KCl), shall be used as an acceptable alternative.
containing products. See the applicable product Safety Data
Sheet (SDS) for additional information. The potential exists
4.2 If the Ag/AgCl (sat KCl) reference electrode is used in
that selling mercury or mercury-containing products, or both,
atestmethod,allpreviouselectrodepotentialvaluesreferredto
is prohibited by local or national law. Users must determine
the SCE can be converted to the Ag/AgCl (sat KCl) scale by
legality of sales in their location.
adding 0.045 V to the SCE values.
1.7 This standard does not purport to address all of the
4.3 New precision statements are required for test methods
safety concerns, if any, associated with its use. It is the
where the alternative reference electrode, Ag/AgCl (sat KCl),
responsibility of the user of this standard to establish appro-
is now specified.The interlaboratory test program and analysis
priate safety, health, and environmental practices and deter-
of results will be required within 5 years of completing the
mine the applicability of regulatory limitations prior to use.
standard incorporating the new Ag/AgCl (sat KCl) reference
electrode requirement.
This practice is under the jurisdiction ofASTM Committee G01 on Corrosion
ofMetalsandisthedirectresponsibilityofSubcommitteeG01.11onElectrochemi- For referenced ASTM standards, visit the ASTM website, www.astm.org, or
cal Measurements in Corrosion Testing. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Current edition approved Nov. 1, 2020. Published December 2020. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/G0220-20. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
G220 − 20
5. Significance and Use 6.4 Theissueofpolarizationofthereferenceelectrodefrom
the measuring instrument is seldom a concern with modern
5.1 Electrode potential measurement is an essential step in
electronics.
many electrochemical corrosion test methods and practices.
6.5 ThefactthattestmethodsandpracticesspecifytheSCE,
5.2 The SCE has been widely specified for laboratory tests
and these standards have acceptable precision and stability
because it is reliable, accurate, simple to use, and it has been
shows that the saturated KCl fill solution is not an issue for
readily available. However, because this device contains
thesestandards.Therefore,thissolutionshouldbeadequatefor
mercury, and mercury has recently been banned by some
the silver/silver chloride reference electrode.
governmental agencies, it may not be available in locations
where mercury is banned. As a result, test methods using the
6.6 Although silver ions are problematic in many
SCE may not be possible in locations where mercury is
environments, the concentration of silver ions in a saturated
banned. Therefore, it is necessary to establish an alternative
–10
KCl solution is less than 10 M at ambient temperature.
reference electrode for these standards.
Contamination of the test environment from leakage from an
5.3 The KCl saturated, silver/silver chloride reference
Ag/AgCl (sat KCl) liquid junction with silver ions is unlikely
electrode,Ag/AgCl (sat KCl), is a satisfactory replacement for
to ever be a problem. The concentration of mercurous ions
++ –18
the SCE in laboratory test methods and practices. This refer-
(Hg ) in a saturated KCl solution is less than 10 Mat
ence electrode should provide comparable performance and
ambient temperature, also indicating that the SCE does not
accuracy to the SCE.
cause mercury contamination of test environments. See Ap-
pendix X1 for the details on the calculations of these concen-
5.4 It will be necessary to carry out interlaboratory test
trations.
programs for each test method where the Ag/AgCl (sat KCl)
electrode is specified to replace the SCE, in order to develop
6.7 Although the silver/silver chloride electrodes have been
the precision of the method using the new reference el
...




Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...