Specification for radio disturbance and immunity measuring apparatus and methods - Part 4-2: Uncertainties, statistics and limit modelling - Uncertainty in EMC measurements

This part of CISPR 16 is designated a basic standard, which specifies the manner in which measurement uncertainty is to be taken in to account in determining compliance with CISPR limits. The material is also relevant to any EMC test when interpretation of the results and conclusions reached will be impacted by the uncertainty of the instrumentation used during the testing. Annex A contains the background material used in providing the amount of measurement uncertainty found in generating the CISPR values shown in Clause 4 and hence provides valuable background material for those needing both initial and further information on measurement uncertainty and how to take into account individual uncertainties in the measurement chain. The annex however is not intended to be a tutorial of user manual or to be copied when making uncertainty calculations. CISPR 16-1, CISPR 16-2, CISPR 16-3 and CISPR 16-4 have been reorganised into 14 parts, to accommodate growth and easier maintenance. This first edition of CISPR 16-4-2 cancels and replaces the first edition of CISPR 16-4 published in 2002. It contains the clauses of CISPR 16-4 without technical changes.

Spécifications des méthodes et des appareils de mesure des perturbations radioélectriques et de l'immunité aux perturbations radioélectriques - Partie 4-2: Incertitudes, statistiques et modélisation des limites - Incertitudes de mesure CEM

La présente partie de la CISPR 16 est une norme fondamentale qui spécifie la manière dont il faut prendre en compte les incertitudes de mesure lors de la détermination de la conformité aux limites du CISPR. Cela s'applique également à tous les essais CEM lorsque les interprétations des résultats et les conclusions peuvent être influencées par l'incertitude de l'instrumentation utilisée pendant les essais. L'annexe A contient les informations utilisées pour fournir la valeur de l'incertitude de mesure CISPR donnée à l'article 4 et, en conséquence, fournit des informations importantes pour ceux qui ont besoin à la fois de l'information initiale et de l'information complémentaire sur les incertitudes de mesure et sur la façon de prendre en compte les incertitudes individuelles dans la chaîne de mesure. Toutefois, l'annexe A n'est pas destinée à être un modèle de manuel d'utilisateur ou à être recopiée lors des calculs d'incertitude. Les publications CISPR 16-1, CISPR 16-2, CISPR 16-3 and CISPR 16-4 ont été réorganisées en 14 parties, dans le but de pouvoir gérer plus facilement leur évolution et maintenance. Cette première édition de la CISPR 16-4-2 annule et remplace la CISPR 16-4 publiée en 2002. Elle contient les articles de la CISPR 16-4 sans modifications de leur contenu technique.

General Information

Status
Published
Publication Date
25-Nov-2003
Current Stage
DELPUB - Deleted Publication
Start Date
08-Jun-2011
Completion Date
26-Oct-2025
Ref Project

Relations

Standard
CISPR 16-4-2:2003 - Specification for radio disturbance and immunity measuring apparatus and methods - Part 4-2: Uncertainties, statistics and limit modelling - Uncertainty in EMC measurements Released:11/26/2003 Isbn:2831873010
English and French language
43 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


COMMISSION
CISPR
ELECTROTECHNIQUE
16-4-2
INTERNATIONALE
INTERNATIONAL
Première édition
ELECTROTECHNICAL
First edition
2003-11
COMMISSION
COMITÉ INTERNATIONAL SPÉCIAL DES PERTURBATIONS RADIOÉLECTRIQUES
INTERNATIONAL SPECIAL COMMITTEE ON RADIO INTERFERENCE
Spécifications des méthodes et des appareils
de mesure des perturbations radioélectriques et
de l'immunité aux perturbations radioélectriques –
Partie 4-2:
Incertitudes, statistiques et modélisation des
limites – Incertitudes de mesure CEM

Specification for radio disturbance and immunity
measuring apparatus and methods –
Part 4-2:
Uncertainties, statistics and limit modelling –
Uncertainty in EMC measurements
Numéro de référence
Reference number
Editions consolidées Consolidated editions
Les versions consolidées de certaines publications de la The IEC is now publishing consolidated versions of its
CEI incorporant les amendements sont disponibles. Par publications. For example, edition numbers 1.0, 1.1
exemple, les numéros d’édition 1.0, 1.1 et 1.2 indiquent and 1.2 refer, respectively, to the base publication,
respectivement la publication de base, la publication de the base publication incorporating amendment 1 and
base incorporant l’amendement 1, et la publication de the base publication incorporating amendments 1
base incorporant les amendements 1 et 2 and 2.
Informations supplémentaires Further information on IEC publications
sur les publications de la CEI
Le contenu technique des publications de la CEI est The technical content of IEC publications is kept
constamment revu par la CEI afin qu'il reflète l'état under constant review by the IEC, thus ensuring that
actuel de la technique. Des renseignements relatifs à the content reflects current technology. Information
cette publication, y compris sa validité, sont dispo- relating to this publication, including its validity, is
nibles dans le Catalogue des publications de la CEI available in the IEC Catalogue of publications
(voir ci-dessous) en plus des nouvelles éditions, (see below) in addition to new editions, amendments
amendements et corrigenda. Des informations sur les and corrigenda. Information on the subjects under
sujets à l’étude et l’avancement des travaux entrepris consideration and work in progress undertaken by the
par le comité d’études qui a élaboré cette publication, technical committee which has prepared this
ainsi que la liste des publications parues, sont publication, as well as the list of publications issued,
également disponibles par l’intermédiaire de: is also available from the following:
• Site web de la CEI (www.iec.ch) • IEC Web Site (www.iec.ch)
• Catalogue des publications de la CEI • Catalogue of IEC publications
Le catalogue en ligne sur le site web de la CEI The on-line catalogue on the IEC web site
(www.iec.ch/searchpub) vous permet de faire des (www.iec.ch/searchpub) enables you to search by a
recherches en utilisant de nombreux critères, variety of criteria including text searches,
comprenant des recherches textuelles, par comité technical committees and date of publication. On-
d’études ou date de publication. Des informations en line information is also available on recently
ligne sont également disponibles sur les nouvelles issued publications, withdrawn and replaced
publications, les publications remplacées ou publications, as well as corrigenda.
retirées, ainsi que sur les corrigenda.
• IEC Just Published • IEC Just Published
Ce résumé des dernières publications parues This summary of recently issued publications
(www.iec.ch/online_news/justpub) est aussi (www.iec.ch/online_news/justpub) is also available
disponible par courrier électronique. Veuillez by email. Please contact the Customer Service
prendre contact avec le Service client (voir ci- Centre (see below) for further information.
dessous) pour plus d’informations.
• Service clients • Customer Service Centre
Si vous avez des questions au sujet de cette If you have any questions regarding this
publication ou avez besoin de renseignements publication or need further assistance, please
supplémentaires, prenez contact avec le Service contact the Customer Service Centre:
clients:
Email: custserv@iec.ch Email: custserv@iec.ch
Tél: +41 22 919 02 11 Tel: +41 22 919 02 11
Fax: +41 22 919 03 00 Fax: +41 22 919 03 00
.
COMMISSION
CISPR
ELECTROTECHNIQUE
16-4-2
INTERNATIONALE
INTERNATIONAL
Première édition
ELECTROTECHNICAL
First edition
2003-11
COMMISSION
COMITÉ INTERNATIONAL SPÉCIAL DES PERTURBATIONS RADIOÉLECTRIQUES
INTERNATIONAL SPECIAL COMMITTEE ON RADIO INTERFERENCE
Spécifications des méthodes et des appareils
de mesure des perturbations radioélectriques et
de l'immunité aux perturbations radioélectriques –
Partie 4-2:
Incertitudes, statistiques et modélisation des
limites – Incertitudes de mesure CEM

Specification for radio disturbance and immunity
measuring apparatus and methods –
Part 4-2:
Uncertainties, statistics and limit modelling –
Uncertainty in EMC measurements
© IEC 2003 Droits de reproduction réservés ⎯ Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in any
utilisée sous quelque forme que ce soit et par aucun procédé, form or by any means, electronic or mechanical, including
électronique ou mécanique, y compris la photocopie et les photocopying and microfilm, without permission in writing from
microfilms, sans l'accord écrit de l'éditeur. the publisher.
International Electrotechnical Commission, 3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, Switzerland
Telephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch Web: www.iec.ch
CODE PRIX
Commission Electrotechnique Internationale S
PRICE CODE
International Electrotechnical Commission
Международная Электротехническая Комиссия
Pour prix, voir catalogue en vigueur
For price, see current catalogue

– 2 – CISPR 16-4-2  CEI:2003
SOMMAIRE
AVANT-PROPOS .4
INTRODUCTION.6
TABLEAU RÉCAPITULATIF DES RÉFÉRENCES CROISÉES.8

1 Domaine d'application.10
2 Références normatives.10
3 Définitions et symboles .12
3.1 Symboles généraux .12
3.2 Valeurs mesurées (mesurandes) .12
3.3 Quantités en entrée .12
4 Incertitude de mesure de l’instrumentation .14
4.1 Vue d'ensemble.14
4.2 Valeurs à considérer pour les mesures des perturbations conduites à un
accès secteur .16
4.3 Valeurs à considérer pour les mesures de puissance perturbatrice.16
4.4 Valeurs à considérer pour les mesures des perturbations rayonnées du
champ électrique sur un emplacement de mesure en espace libre ou un autre
emplacement d’essai possible .16

Annexe A (informative) Bases pour les valeurs de U du tableau 1 .20
cispr
Bibliographie .42

CISPR 16-4-2  IEC:2003 – 3 –
CONTENTS
FOREWORD.5
INTRODUCTION.7
TABLE RECAPITULATING CROSS-REFERENCES .9

1 Scope.11
2 Normative references .11
3 Definitions and symbols.13
3.1 General symbols .13
3.2 Measurands .13
3.3 Input quantities.13
4 Measurement instrumentation uncertainty .15
4.1 Overview .15
4.2 Quantities to be considered for conducted disturbance measurements at a
mains port .17
4.3 Quantities to be considered for disturbance power measurements.17
4.4 Quantities to be considered for radiated disturbance measurements of
electric field strength on an open area test site or alternative test site .17

Annex A (informative) Basis for U values in Table 1.21
cispr
Bibliography.43

– 4 – CISPR 16-4-2 © CEI:2003
COMMISSION ÉLECTROTECHNIQUE INTERNATIONALE
COMITÉ INTERNATIONAL SPÉCIAL DES PERTURBATIONS RADIOÉLECTRIQUES
___________
SPÉCIFICATIONS DES MÉTHODES ET DES APPAREILS
DE MESURE DES PERTURBATIONS RADIOÉLECTRIQUES ET
DE L'IMMUNITÉ AUX PERTURBATIONS RADIOÉLECTRIQUES –

Partie 4-2: Incertitudes, statistiques et modélisation des limites –
Incertitudes de mesure CEM
AVANT-PROPOS
1) La Commission Electrotechnique Internationale (CEI) est une organisation mondiale de normalisation composée
de l'ensemble des comités électrotechniques nationaux (Comités nationaux de la CEI). La CEI a pour objet de
favoriser la coopération internationale pour toutes les questions de normalisation dans les domaines de
l'électricité et de l'électronique. A cet effet, la CEI – entre autres activités – publie des Normes internationales,
des Spécifications techniques, des Rapports techniques, des Spécifications accessibles au public (PAS) et des
Guides (ci-après dénommés "Publication(s) de la CEI"). Leur élaboration est confiée à des comités d'études,
aux travaux desquels tout Comité national intéressé par le sujet traité peut participer. Les organisations
internationales, gouvernementales et non gouvernementales, en liaison avec la CEI, participent également aux
travaux. La CEI collabore étroitement avec l'Organisation Internationale de Normalisation (ISO), selon des
conditions fixées par accord entre les deux organisations.
2) Les décisions ou accords officiels de la CEI concernant les questions techniques représentent, dans la mesure
du possible, un accord international sur les sujets étudiés, étant donné que les Comités nationaux de la CEI
intéressés sont représentés dans chaque comité d’études.
3) Les Publications de la CEI se présentent sous la forme de recommandations internationales et sont agréées
comme telles par les Comités nationaux de la CEI. Tous les efforts raisonnables sont entrepris afin que la CEI
s'assure de l'exactitude du contenu technique de ses publications; la CEI ne peut pas être tenue responsable de
l'éventuelle mauvaise utilisation ou interprétation qui en est faite par un quelconque utilisateur final.
4) Dans le but d'encourager l'uniformité internationale, les Comités nationaux de la CEI s'engagent, dans toute la
mesure possible, à appliquer de façon transparente les Publications de la CEI dans leurs publications
nationales et régionales. Toutes divergences entre toutes Publications de la CEI et toutes publications
nationales ou régionales correspondantes doivent être indiquées en termes clairs dans ces dernières.
5) La CEI n’a prévu aucune procédure de marquage valant indication d’approbation et n'engage pas sa
responsabilité pour les équipements déclarés conformes à une de ses Publications.
6) Tous les utilisateurs doivent s'assurer qu'ils sont en possession de la dernière édition de cette publication.
7) Aucune responsabilité ne doit être imputée à la CEI, à ses administrateurs, employés, auxiliaires ou
mandataires, y compris ses experts particuliers et les membres de ses comités d'études et des Comités
nationaux de la CEI, pour tout préjudice causé en cas de dommages corporels et matériels, ou de tout autre
dommage de quelque nature que ce soit, directe ou indirecte, ou pour supporter les coûts (y compris les frais
de justice) et les dépenses découlant de la publication ou de l'utilisation de cette Publication de la CEI ou de
toute autre Publication de la CEI, ou au crédit qui lui est accordé.
8) L'attention est attirée sur les références normatives citées dans cette publication. L'utilisation de publications
référencées est obligatoire pour une application correcte de la présente publication.
9) L’attention est attirée sur le fait que certains des éléments de la présente Publication de la CEI peuvent faire
l’objet de droits de propriété intellectuelle ou de droits analogues. La CEI ne saurait être tenue pour
responsable de ne pas avoir identifié de tels droits de propriété et de ne pas avoir signalé leur existence.
La Norme internationale CISPR 16-4-2 a été établie par le sous-comité A du CISPR : Mesures
des perturbations radioélectriques et méthodes statistiques.
Cette première édition de la CISPR 16-4-2 annule et remplace la CISPR 16-4 publiée en 2002.
Elle contient les articles de la CISPR 16-4 sans modifications de leur contenu technique.
Cette publication a été rédigée selon les Directives ISO/CEI, Partie 2.
Le comité a décidé que le contenu de cette publication ne sera pas modifié avant 2006. A cette
date, la publication sera
• reconduite;
• supprimée;
• remplacée par une édition révisée, ou
• amendée.
CISPR 16-4-2 © IEC:2003 – 5 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
INTERNATIONAL SPECIAL COMMITTEE ON RADIO INTERFERENCE
____________
SPECIFICATION FOR RADIO DISTURBANCE AND IMMUNITY
MEASURING APPARATUS AND METHODS –

Part 4-2: Uncertainties, statistics and limit modelling –
Uncertainty in EMC measurements

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard CISPR 16-4-2 has been prepared by CISPR subcommittee A: Radio
interference measurements and statistical methods.
This first edition of CISPR 16-4-2 cancels and replaces the first edition of CISPR 16-4
published in 2002. It contains the clauses of CISPR 16-4 without technical changes.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
The committee has decided that the contents of this publication will remain unchanged until
2006. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.
– 6 – CISPR 16-4-2  CEI:2003
INTRODUCTION
Les publications CISPR 16-1, CISPR 16-2, CISPR 16-3 et CISPR 16-4 ont été réorganisées en
14 parties, dans le but de pouvoir gérer plus facilement leur évolution et maintenance. Les
nouvelles parties portent de nouveaux numéros. Voir la liste donnée ci-dessous.
Anciennes publications CISPR 16
Nouvelles publications CISPR 16
Appareils de mesure
CISPR 16-1-1
Appareils de mesure
CISPR 16-1-2 Matériels auxiliaires – Perturbations conduites
des perturbations
radioélectriques
CISPR 16-1-3 Matériels auxiliaires – Puissance perturbatrice
CISPR 16-1
et de l'immunité
aux perturbations
Matériels auxiliaires – Perturbations rayonnées
CISPR 16-1-4
radioélectriques
Emplacements d'essai pour l'étalonnage des
CISPR 16-1-5
antennes de 30 MHz et 1 000 MHz
CISPR 16-2-1 Mesures des perturbations conduites
Méthodes de mesure
Mesure de la puissance perturbatrice
CISPR 16-2-2
des perturbations et
CISPR 16-2
de l'immunité
CISPR 16-2-3 Mesures des perturbations rayonnées
CISPR 16-2-4
Mesures de l'immunité
CISPR 16-3 Rapports techniques du CISPR
Incertitudes dans les essais normalisés en CEM
CISPR 16-4-1
Rapports et
Incertitudes de l'instrumentation de mesure
CISPR 16-3 recommandations CISPR 16-4-2
du CISPR
Considérations statistiques dans la détermination
CISPR 16-4-3
de la conformité CEM des produits fabriqués en
grand nombre
Statistiques des plaintes pour le calcul
Incertitudes dans
CISPR 16-4 CISPR 16-4-4
des limites
les mesures CEM
Des informations plus spécifiques concernant la relation entre l' "ancienne" CISPR 16-4 et la
"nouvelle" CISPR 16-4-2 sont données dans le tableau qui suit cette introduction (TABLEAU
RÉCAPITULATIF DES RÉFÉRENCES CROISÉES).
Les spécifications des appareils de mesure sont données dans les cinq nouvelles parties de la
CISPR 16-1, alors que les méthodes de mesure des perturbations radioélectriques sont
désormais couvertes par les quatre nouvelles parties de la CISPR 16-2. Différents rapports
avec des informations sur le contexte du CISPR et sur les perturbations radioélectriques en
général sont donnés dans la CISPR 16-3. La CISPR 16-4 contient des informations relatives
aux incertitudes, aux statistiques et à la modélisation des limites.
La CISPR 16-4 est constituée des quatre parties suivantes, sous le titre général Spécifications
des méthodes et des appareils de mesure des perturbations radioélectriques et de l'immunité
aux perturbations radioélectriques – Incertitudes, statistiques et modélisation des limites:
• Partie 4-1: Incertitudes dans les essais normalisés en CEM,
• Partie 4-2: Incertitudes de mesure CEM,
• Partie 4-3: Considérations statistiques dans la détermination de la conformité CEM des
produits fabriqués en grand nombre,
• Partie 4-4: Statistiques des plaintes pour le calcul des limites.

CISPR 16-4-2  IEC:2003 – 7 –
INTRODUCTION
CISPR 16-1, CISPR 16-2, CISPR 16-3 and CISPR 16-4 have been reorganised into 14 parts,
to accommodate growth and easier maintenance. The new parts have also been renumbered.
See the list given below.
Old CISPR 16 publications New CISPR 16 publications
CISPR 16-1-1 Measuring apparatus
CISPR 16-1-2 Ancillary equipment – Conducted disturbances
Radio disturbance
and immunity
CISPR 16-1-3 Ancillary equipment – Disturbance power
CISPR 16-1
measuring
apparatus
Ancillary equipment – Radiated disturbances
CISPR 16-1-4
Antenna calibration test sites for 30 MHz to
CISPR 16-1-5
1 000 MHz
CISPR 16-2-1 Conducted disturbance measurements
Methods of
Measurement of disturbance power
CISPR 16-2-2
measurement of
CISPR 16-2
disturbances and
CISPR 16-2-3 Radiated disturbance measurements
immunity
CISPR 16-2-4
Immunity measurements
CISPR 16-3 CISPR technical reports
Uncertainties in standardised EMC tests
CISPR 16-4-1
Reports and
Measurement instrumentation uncertainty
CISPR 16-3 recommendations CISPR 16-4-2
of CISPR
Statistical considerations in the
CISPR 16-4-3
determination of EMC compliance of mass-
produced products
Statistics of complaints and a model for the
Uncertainty in EMC
CISPR 16-4 CISPR 16-4-4
calculation of limits
measurements
More specific information on the relation between the ‘old’ CISPR 16-4 and the present ‘new’
CISPR 16-4-2 is given in the table after this introduction (TABLE RECAPITULATING CROSS
REFERENCES).
Measurement instrumentation specifications are given in five new parts of CISPR 16-1, while
the methods of measurement are covered now in four new parts of CISPR 16-2. Various
reports with further information and background on CISPR and radio disturbances in general
are given in CISPR 16-3. CISPR 16-4 contains information related to uncertainties, statistics
and limit modelling.
CISPR 16-4 consists of the following parts, under the general title Specification for radio
disturbance and immunity measuring apparatus and methods - Uncertainties, statistics and
limit modelling:
• Part 4-1: Uncertainties in standardised EMC tests,
• Part 4-2: Uncertainty in EMC measurements,
• Part 4-3: Statistical considerations in the determination of EMC compliance of mass-
produced products,
• Part 4-4: Statistics of complaints and a model for the calculation of limits.

– 8 – CISPR 16-4-2  CEI:2003
TABLEAU RÉCAPITULATIF DES RÉFÉRENCES CROISÉES

Première édition de la CISPR 16-4 Première édition de la CISPR 16-4-2
Articles Articles
1 1
2 2
3 3
4 4
Annexe Annexe
A A
CISPR 16-4-2  IEC:2003 – 9 –
TABLE RECAPITULATING CROSS-REFERENCES

First edition of CISPR 16-4 First edition of CISPR 16-4-2
Clauses Clauses
1 1
2 2
3 3
4 4
Annex Annex
A A
– 10 – CISPR 16-4-2  CEI:2003
SPÉCIFICATIONS DES MÉTHODES ET DES APPAREILS
DE MESURE DES PERTURBATIONS RADIOÉLECTRIQUES ET
DE L'IMMUNITÉ AUX PERTURBATIONS RADIOÉLECTRIQUES –

Partie 4-2: Incertitudes, statistiques et modélisation des limites –
Incertitudes de mesure CEM
1 Domaine d'application
La présente partie de la CISPR 16 est une norme fondamentale qui spécifie la manière dont il
faut prendre en compte les incertitudes de mesure lors de la détermination de la conformité
aux limites du CISPR. Cela s'applique également à tous les essais CEM lorsque les
interprétations des résultats et les conclusions peuvent être influencées par l'incertitude de
l'instrumentation utilisée pendant les essais. L'annexe A contient les informations utilisées pour
fournir la valeur de l'incertitude de mesure CISPR donnée à l'article 4 et, en conséquence,
fournit des informations importantes pour ceux qui ont besoin à la fois de l'information initiale
et de l'information complémentaire sur les incertitudes de mesure et sur la façon de prendre en
compte les incertitudes individuelles dans la chaîne de mesure. Toutefois, l'annexe A n'est pas
destinée à être un modèle de manuel d'utilisateur ou à être recopiée lors des calculs
d'incertitude. Pour cela, il convient d’utiliser les références données dans la bibliographie.
Les spécifications de l'instrumentation de mesure sont données dans la CISPR 16-1 et les
méthodes de mesure sont couvertes par la CISPR 16-2. Des informations complémentaires sur
les perturbations radioélectriques et sur l’origine du CISPR sont données dans la CISPR 16-3.
Les autres parties de la CISPR 16-4 contiennent des informations complémentaires sur
l’incertitude en général, les statistiques et la modélisation des limites.
2 Références normatives
Les documents de référence suivants sont indispensables pour l'application du présent
document. Pour les références datées, seule l'édition citée s'applique. Pour les références non
datées, la dernière édition du document de référence s'applique (y compris les éventuels
amendements).
CISPR 16-1 (toutes les parties), Spécifications des méthodes et des appareils de mesure des
perturbations radioélectriques et de l'immunité aux perturbations radioélectriques – Appareils
de mesure des perturbations radioélectriques et de l'immunité aux perturbations
radioélectriques
CISPR 16-2 (toutes les parties), Spécifications pour les appareils et méthodes de mesure des
perturbations radioélectriques et de l'immunité aux perturbations radioélectriques – Méthodes
de mesure des perturbations et de l'immunité
CISPR 16-3, Spécifications des méthodes et des appareils de mesure des perturbations
radioélectriques et de l'immunité aux perturbations radioélectriques – Partie 3: Rapports
techniques du CISPR
CISPR 16-4-1, Spécifications des méthodes et des appareils de mesure des perturbations
radioélectriques et de l'immunité aux perturbations radioélectriques – Partie 4-1: Incertitudes,
statistiques et modélisation des limites – Incertitudes dans les essais normalisés en CEM
CISPR 16-4-3, Spécifications des méthodes et des appareils de mesure des perturbations
radioélectriques et de l'immunité aux perturbations radioélectriques – Partie 4-3: Incertitudes,
statistiques et modélisation des limites – Considérations statistiques dans la détermination de
la conformité CEM des produits fabriqués en grand nombre

CISPR 16-4-2  IEC:2003 – 11 –
SPECIFICATION FOR RADIO DISTURBANCE AND IMMUNITY
MEASURING APPARATUS AND METHODS –

Part 4-2: Uncertainties, statistics and limit modelling –
Uncertainty in EMC measurements

1 Scope
This part of CISPR 16 is designated a basic standard, which specifies the manner in which
measurement uncertainty is to be taken in to account in determining compliance with CISPR
limits. The material is also relevant to any EMC test when interpretation of the results and
conclusions reached will be impacted by the uncertainty of the instrumentation used during
the testing. Annex A contains the background material used in providing the amount of
measurement uncertainty found in generating the CISPR values shown in Clause 4 and hence
provides valuable background material for those needing both initial and further information
on measurement uncertainty and how to take into account individual uncertainties in the
measurement chain. The annex however is not intended to be a tutorial of user manual or to
be copied when making uncertainty calculations. For that, the references shown in the
bibliography should be used.
Measurement instrumentation specifications are given in CISPR 16-1, while the methods of
measurement are covered in CISPR 16-2. Further information and background on CISPR and
radio disturbances is given in CISPR 16-3. The other parts of CISPR 16-4 contain further
information on uncertainties in general, statistics and limit modelling.
2 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
CISPR 16-1 (all parts), Specification for radio disturbance and immunity measuring apparatus
and methods – Radio disturbance and immunity measuring apparatus
CISPR 16-2 (all parts), Specification for radio disturbance and immunity measuring apparatus
and methods – Methods of measurement of disturbances and immunity
CISPR 16-3, Specification for radio disturbance and immunity measuring apparatus and
methods – Part 3: CISPR technical reports
CISPR 16-4-1, Specification for radio disturbance and immunity measuring apparatus and
methods – Part 4-1: Uncertainties, statistics and limit modelling - Uncertainties in
standardised EMC tests
CISPR 16-4-3, Specification for radio disturbance and immunity measuring apparatus and
methods – Part 4-3: Uncertainties, statistics and limit modelling - Statistical considerations in
the determination of EMC compliance of mass-produced products

– 12 – CISPR 16-4-2  CEI:2003
CISPR 16-4-4, Spécifications des méthodes et des appareils de mesure des perturbations
radioélectriques et de l'immunité aux perturbations radioélectriques – Partie 4-4: Incertitudes,
statistiques et modélisation des limites – Statistiques des plaintes pour le calcul des limites
3 Définitions et symboles
Pour les besoins de la présente partie de la CISPR 16, les symboles suivants s’appliquent.
NOTE Les termes généraux et définitions concernant l'incertitude sont donnés dans la référence bibliographique
[2]. Les définitions générales sont données dans la référence bibliographique [1]. Ces définitions ne seront pas
répétées ici.
3.1 Symboles généraux
X Quantité en entrée
i
x Estimation de X
i i
u(x ) Incertitude-type de x
i i
c Coefficient de sensibilité
i
y Résultat d’une mesure, (l’estimation d’une valeur mesurée), corrigée de tous les
effets systématiques significatifs reconnus
u (y) Incertitude-type (composée) de y
c
k Facteur d'élargissement
U Incertitude élargie de y
3.2 Valeurs mesurées (mesurandes)
V Tension, en dB(µV)
P Puissance perturbatrice, en dB(pW)
E Niveau de champ électrique, en dB(µV/m)
3.3 Quantités en entrée
V Lecture en tension sur le récepteur, en dB(µV)
r
L Atténuation de la connexion entre le récepteur et le réseau fictif secteur, la pince
c
absorbante ou l’antenne, en dB
L Facteur de division en tension du réseau fictif, en dB
amn
L Perte d’insertion de la pince absorbante, en dB
ac
AF Facteur d’antenne, en dB(/m)
δV Correction de l’imprécision en tension sinusoïdale du récepteur, en dB
sw
δV Correction d’une réponse imparfaite en amplitude du récepteur pour les impulsions,
pa
en dB
δV Correction d’une réponse en impulsion imparfaite du récepteur avec la fréquence de
pr
répétition, en dB
δV Correction de l’effet du plancher de bruit du récepteur, en dB
nf
δM Correction de l’erreur due à la désadaptation, en dB
δMD Correction de l’erreur due aux perturbations secteur, en dB
δZ Correction d’une impédance imparfaite du réseau fictif, en dB
δE Correction de l’effet de l’environnement, en dB
δAF Correction de l’erreur d’interpolation du facteur d’antenne, en dB
f
CISPR 16-4-2  IEC:2003 – 13 –
CISPR 16-4-4, Specification for radio disturbance and immunity measuring apparatus and
methods – Part 4-4: Uncertainties, statistics and limit modelling – A model for the calculation
of limits
3 Definitions and symbols
For the purpose of this part of CISPR 16, the following symbols apply.
NOTE General uncertainty terms and definitions used in uncertainty are contained in bibliography reference [2].
General definitions are contained in bibliography reference [1]. These definitions will not be repeated here.
3.1 General symbols
X input quantity
i
x estimate of X
i i
u(x ) standard uncertainty of x
i i
c sensitivity coefficient
i
y result of a measurement, (the estimate of the measurand), corrected for all
recognised significant systematic effects
u (y) (combined) standard uncertainty of y
c
k coverage factor
U expanded uncertainty of y
3.2 Measurands
V Voltage, in dB(µV)
P Disturbance power, in dB(pW)
E Electric field strength, in dB(µV/m)
3.3 Input quantities
V Receiver voltage reading, in dB(µV)
r
L Attenuation of the connection between the receiver and the artificial mains network,
c
absorbing clamp or antenna, in dB
L Voltage division factor of the AMN, in dB
amn
Insertion loss of the absorbing clamp, in dB
L
ac
AF Antenna factor, in dB(/m)
δV Correction for receiver sine wave voltage inaccuracy, in dB
sw
δV Correction for imperfect receiver pulse amplitude response, in dB
pa
δV Correction for imperfect receiver pulse repetition rate response, in dB
pr
δV Correction for the effect of the receiver noise floor, in dB
nf
δM Correction for the error caused by mismatch, in dB
δMD Correction for the error caused by mains disturbances, in dB
δZ Correction for imperfect AMN impedance, in dB
δE Correction for the effect of the environment, in dB
δAF Correction for antenna factor interpolation error, in dB
f
– 14 – CISPR 16-4-2  CEI:2003
δAF Correction de la différence entre la variation du facteur d’antenne avec la hauteur, et
h
la variation avec la hauteur du facteur d’antenne d’un dipôle de référence, en dB
δA Correction de la directivité de l’antenne, en dB
dir
δA Correction de la localisation du centre de phase de l’antenne, en dB
ph
δA Correction de la réponse en polarisation croisée de l’antenne, en dB
cp
δA Correction de la dissymétrie de l’antenne, en dB
bal
δSA Correction d’une atténuation imparfaite de l'emplacement, en dB
δd Correction de la distance imparfaite de l’antenne, en dB
δh Correction d’une hauteur imparfaite de la table au-dessus du plan de sol, en dB

4 Incertitude de mesure de l’instrumentation
4.1 Vue d'ensemble
On doit tenir compte de l’incertitude de mesure de l’instrumentation lors de la détermination de
la conformité ou de la non-conformité avec une limite de perturbation.
L’incertitude de mesure de l’instrumentation pour un laboratoire d’essai doit être évaluée pour
les mesures indiquées dans les paragraphes suivants, en tenant compte de chacune des
quantités énumérées ici. L’incertitude-type u(x ) en décibels et le coefficient de sensibilité c
i i
doivent être évalués pour l’estimation x de chaque quantité. L’incertitude-type composée u (y)
i c
de l'estimation y doit être calculée comme suit:
2 2
u (y)= c u (x )
c ∑ i i
i
L’incertitude élargie de l’instrumentation de mesure U pour un laboratoire d’essai doit être
lab
calculée comme suit:
U = 2 u (y)
lab c
et doit être indiquée dans le rapport d’essai.
NOTE 1 Le facteur d'élargissement k = 2 conduit approximativement à un niveau de confiance de 95 % pour la
distribution quasi normale typique de la plupart des résultats de mesure.
La conformité ou la non-conformité à une limite de perturbation doit être déterminée de la
manière suivante.
Si U est inférieure ou égale à U du tableau 1, alors:
lab cispr
– la conformité est estimée atteinte si aucune perturbation mesurée ne dépasse la limite de
perturbation;
– la non-conformité est estimée atteinte si une perturbation mesurée quelconque dépasse la
limite de perturbation.
NOTE 2 U ressemble à une valeur d’incertitude de mesure pour un essai spécifique , qui a été déterminée en
cispr
considérant les incertitudes associées aux valeurs listées en 4.2.
Si U est supérieure à U du tableau 1, alors:
lab cispr
– la conformité est estimée atteinte si aucune perturbation mesurée, augmentée de
(U – U ), ne dépasse la limite de perturbation;
lab cispr
– la non-conformité est estimée atteinte si une perturbation mesurée quelconque, augmentée
de (U – U ), dépasse la limite de perturbation.
lab cispr
CISPR 16-4-2  IEC:2003 – 15 –
δAF Correction for the difference between the antenna factor variation with height, and
h
the variation with height of the antenna factor of a reference dipole, in dB
δA Correction for antenna directivity, in dB
dir
δA Correction for antenna phase centre location, in dB
ph
δA Correction for antenna cross-polarisation response, in dB
cp
δA Correction for antenna unbalance, in dB
bal
δSA Correction for imperfect site attenuation, in dB
δd Correction for imperfect antenna distance, in dB
δh Correction for imperfect table height above ground plane, in dB
4 Measurement instrumentation uncertainty
4.1 Overview
Measurement instrumentation uncertainty shall be taken into account when determining
compliance or non-compliance with a disturbance limit.
The measurement instrumentation uncertainty for a test laboratory shall be evaluated for
those measurements addressed in the following subclauses, taking into consideration each of
the quantities listed there. The standard uncertainty u(x ) in decibels and the sensitivity
i
coefficient c shall be evaluated for the estimate x of each quantity. The combined standard
i i
uncertainty u (y) of the estimate y of the measurand shall be calculated as
c
2 2
u (y)= c u (x )
c i i

i
The expanded measurement instrumentation uncertainty U for a test laboratory shall be
lab
calculated as
U = 2 u (y)
lab c
and shall be stated in the test report.
NOTE 1 The coverage factor k = 2 yields approximately a 95 % level of confidence for the near-normal
distribution typical of most measurement results.
Compliance or non-compliance with a disturbance limit shall be determined in the following
manner.
If U is less than or equal to U in Table 1, then:
lab cispr
− compliance is deemed to occur if no measured disturbance exceeds the disturbance limit;
− non-compliance is deemed to occur if any measured disturbance exceeds the disturbance
limit.
NOTE U resembles a value of measurement uncertainty for a specific test, which was determined by
cispr
considering uncertainties associated with the quantities listed in 4.2.
If U is greater than U in Table 1, then:
lab cispr
− compliance is deemed to occur if no measured disturbance, increased by (U – U ),
lab cispr
exceeds the disturbance limit;
− non-compliance is deemed to occur if any measured disturbance, increased by
(U – U ), exceeds the disturbance limit.
lab cispr
– 16 – CISPR 16-4-2  CEI:2003
Tableau 1 – Valeurs de U
cispr
Mesure U
cispr
(9 kHz – 150 kHz) 4,0 dB
Perturbation conduite (accès secteur)
(150 kHz – 30 MHz) 3,6 dB
Puissance perturbatrice (30 MHz – 300 MHz) 4,5 dB
Perturbation rayonnée (niveau du champ électrique sur un emplacement
(30 MHz – 1 000 MHz) 5,2 dB
d’essai en espace libre ou un autre emplacement d’essai possible)
Autre A l’étude
NOTE 3 Les valeurs de U du tableau 1 sont basées sur les incertitudes élargies de l’annexe A, qui ont été
cispr
évaluées en considérant les incertitudes associées aux valeurs citées ci-dessous.
Rien dans cet article ne dispense de la nécessité pour l’appareil de mesure d'être conforme
aux spécifications de la CISPR 16-1.
4.2 Valeurs à considérer pour les mesures des perturbations conduites à un accès
secteur
– Lecture sur le récepteur
– Atténuation de la connexion entre le réseau fictif secteur et le récepteur
– Facteur de division en tension du réseau fictif secteur
– Précision du récepteur en tension sinusoïdale
– Réponse du récepteur en amplitude pour les impulsions
– Variation de la réponse du récepteur en impulsion avec la fréquence de répétition
– Plancher de bruit du récepteur
– Effets de désadaptation entre l’accès récepteur du réseau fictif secteur et le récepteur
– Impédance du réseau fictif secteur
4.3 Valeurs à considérer pour les mesures de puissance perturbatrice
– Lecture sur le récepteur
– Atténuation de la connexion entre la pince absorbante et le récepteur
– Perte d’insertion de la pince absorbante
– Précision du récepteur en tension sinusoïdale
– Réponse du récepteur en amplitude pour les impulsions
– Variation de la réponse en impulsion du récepteur avec la fréquence de répétition
– Plancher de bruit du récepteur
– Effets de désadaptation entre l’accès récepteur de la pince absorbante et le récepteur
– Effet des perturbations secteur
– Effet de l’environnement
4.4 Valeurs à considérer pour les mesures des perturbations rayonnées du champ
électrique sur un emplacement de mesure en espace libre ou un autre
emplacement d’essai possible
– Lecture sur le récepteur
– Atténuation de la connexion entre l’antenne et le récepteur
– Facteur d’antenne
– Précision en tension sinusoïdale du récepteur

CISPR 16-4-2  IEC:2003 – 17 –
Table 1 – Values of U
cispr
Measurement U
cispr
(9 kHz – 150 kHz) 4,0 dB
Conducted disturbance (mains port)
(150 kHz – 30 MHz) 3,6 dB
Disturbance power (30 MHz – 300 MHz) 4,5 dB
Radiated disturbance (electric field strength on an open area test
(30 MHz – 1 000 MHz) 5,2 dB
site or alternative test site)
Under
Other
consideration
NOTE 2 The values of U in Table 1 are based on the expanded uncertainties in Annex A, which were
cispr
evaluated by considering uncertainties associated with the quantities listed below.
Nothing in this clause removes the requirement for measuring apparatus to comply with
specifications in CISPR 16-1.
4.2 Quantities to be considered for conducted disturbance measurements
at a mains port
– Receiver reading
– Attenuation of the connection between artificial mains network and receiver
– Artificial mains network voltage division factor
– Receiver sine-wave voltage accuracy
– Receiver pulse amplitude response
– Receiver pulse response variation with repetition frequency
– Receiver noise floor
– Mismatch effects between artificial mains network receiver port and receiver
– Artificial mains network impedance
4.3 Quantities to be considered for disturbance power measurements
– Receiver reading
– Attenuation of the connection between absorbing clamp and receiver
– Absorbing clamp insertion loss
– Receiver sine-wave voltage accuracy
– Receiver pulse amplitude response
– Receiver pulse response variation with repetition f
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...