ASTM E2926-17
(Test Method)Standard Test Method for Forensic Comparison of Glass Using Micro X-ray Fluorescence (µ-XRF) Spectrometry
Standard Test Method for Forensic Comparison of Glass Using Micro X-ray Fluorescence (µ-XRF) Spectrometry
SIGNIFICANCE AND USE
4.1 µ-XRF provides a means of simultaneously detecting major, minor, and trace elemental constituents in small glass fragments such as those frequently examined in forensic case work. It can be used at any point in the analytical scheme without concern for changing sample shape or sample properties, such as RI, due to its totally nondestructive nature.
4.2 Limits of detection (LOD) are dependent on several factors, including instrument configuration and operating parameters, sample thickness, and atomic number of the individual elements. Typical LODs range from parts per million (µgg-1) to percent (%).
4.3 µ-XRF provides simultaneous qualitative analysis for elements having an atomic number of eleven or greater. This multi-element capability permits detection of elements typically present in glass such as magnesium (Mg), silicon (Si), aluminum (Al), calcium (Ca), potassium (K), iron (Fe), titanium (Ti), strontium (Sr), and zirconium (Zr), as well as other elements that may be detectable in some glass by µ-XRF (for example, molybdenum (Mo), selenium (Se), or erbium (Er)) without the need for a predetermined elemental menu.
4.4 µ-XRF comparison of glass fragments provides additional discrimination power beyond that of RI or density comparisons, or both, alone.
4.5 The method precision should be established in each laboratory for the specific conditions and instrumentation in that laboratory.
4.6 When using small fragments having varying surface geometries and thicknesses, precision deteriorates due to take-off-angle and critical depth effects. Flat fragments with thickness greater than 1.5 mm do not suffer from these constraints, but are not always available as questioned specimens received in casework. As a consequence of the deterioration in precision for small fragments and the lack of appropriate calibration standards, quantitative analysis by µ-XRF is not typically used.
4.7 Appropriate sampling techniques should be used to account for natural h...
SCOPE
1.1 This test method is for the determination of major, minor, and trace elements present in glass fragments. The elemental composition of a glass fragment can be measured through the use of µ-XRF analysis for comparisons of glass.
1.2 This test method covers the application of µ-XRF using mono- and poly- capillary optics, and an energy dispersive X-ray detector (EDS).
1.3 This test method does not replace knowledge, skill, ability, experience, education, or training and should be used in conjunction with professional judgment.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
- Status
- Published
- Publication Date
- 31-Jan-2017
- Technical Committee
- E30 - Forensic Sciences
- Drafting Committee
- E30.01 - Criminalistics
Relations
- Effective Date
- 01-May-2014
- Effective Date
- 01-May-2013
- Effective Date
- 15-Jun-2012
- Effective Date
- 01-Oct-2010
- Effective Date
- 01-Oct-2008
- Refers
ASTM E177-06b - Standard Practice for Use of the Terms Precision and Bias in ASTM Test Methods - Effective Date
- 15-Nov-2006
- Refers
ASTM E177-06a - Standard Practice for Use of the Terms Precision and Bias in ASTM Test Methods - Effective Date
- 01-Nov-2006
- Effective Date
- 01-Nov-2004
- Effective Date
- 01-Nov-2004
- Refers
ASTM E177-04e1 - Standard Practice for Use of the Terms Precision and Bias in ASTM Test Methods - Effective Date
- 01-Nov-2004
- Effective Date
- 01-Aug-2004
- Refers
ASTM E177-90a(2002) - Standard Practice for Use of the Terms Precision and Bias in ASTM Test Methods - Effective Date
- 10-Jan-2002
Overview
ASTM E2926-17 is the Standard Test Method for Forensic Comparison of Glass Using Micro X-ray Fluorescence (μ-XRF) Spectrometry, established by ASTM International. This standard outlines a nondestructive analytical approach for comparing glass fragments through the simultaneous detection of major, minor, and trace elements using μ-XRF. Commonly applied in forensic science, this method enhances discrimination power in glass fragment comparison, supporting investigations and legal proceedings.
Key Topics
- μ-XRF Analysis: Employs capillary optics and energy dispersive X-ray detectors to analyze elemental composition in glass fragments without altering sample properties or shape.
- Elemental Detection: Capable of detecting elements with atomic number 11 or higher, including magnesium (Mg), silicon (Si), aluminum (Al), calcium (Ca), potassium (K), iron (Fe), titanium (Ti), strontium (Sr), and zirconium (Zr).
- Non-destructive Testing: Preserves the integrity and physical properties of glass evidence, allowing repeated and multi-step analyses.
- Detection Limits: Typical μ-XRF limits of detection range from parts per million (ppm) to percent, depending on instrument settings, sample thickness, and element type.
- Qualitative and Semi-Quantitative Results: Provides qualitative profiles and semi-quantitative comparisons (using elemental peak intensity ratios) between questioned and known glass samples.
- Precision and Best Practices: Emphasizes the need for laboratory-specific precision validation, appropriate sample preparation, and regular system calibration to maintain data integrity.
Applications
- Forensic Glass Comparison: Vital in crime scene investigations involving glass fragments (e.g., burglary, vandalism, traffic accidents), supporting source attribution or exclusion.
- Legal Evidence: μ-XRF data is used as robust supplementary evidence, providing higher discrimination than refractive index (RI) or density measurement alone.
- Material Science: Assists in glass quality control and source verification in manufacturing and research contexts.
- Non-invasive Sampling: Especially valuable when only small or precious specimen samples are available, as μ-XRF is entirely nondestructive.
- Database Collection: Facilitates creation and use of spectral databases for known glass types, streamlining casework comparisons and interlaboratory studies.
Related Standards
- ASTM E2330: Standard Test Method for Determination of Concentrations of Elements in Glass Samples Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) for Forensic Comparisons. Provides destructive, quantitative trace analysis as a complement to μ-XRF.
- ASTM E177: Practice for Use of the Terms Precision and Bias in ASTM Test Methods. Cited for establishing test precision and repeatability in μ-XRF applications.
- ICP-OES and LA-ICP-MS Methods: Offer alternative techniques, including destructive and non-destructive options, depending on analytical requirements.
- Standard Reference Materials (SRM): Use of NIST SRM 1831 and other certified reference materials ensures calibration reliability and data comparability.
Practical Value
The adoption of ASTM E2926-17 improves forensic laboratories’ capability to perform objective, sensitive, and repeatable comparisons of glass evidence. By following standardized procedures, forensic practitioners ensure consistent generation of high-quality analytical data that withstands scrutiny in legal and scientific forums. This standard is critical for enhancing evidential value in glass analysis, upholding best practices in forensic science, and supporting reliable outcomes in both investigative and quality control settings.
Keywords: forensic glass comparison, micro X-ray fluorescence, μ-XRF spectrometry, elemental analysis, ASTM E2926-17, forensic science, nondestructive testing.
Buy Documents
ASTM E2926-17 - Standard Test Method for Forensic Comparison of Glass Using Micro X-ray Fluorescence (µ-XRF) Spectrometry
REDLINE ASTM E2926-17 - Standard Test Method for Forensic Comparison of Glass Using Micro X-ray Fluorescence (µ-XRF) Spectrometry
Frequently Asked Questions
ASTM E2926-17 is a standard published by ASTM International. Its full title is "Standard Test Method for Forensic Comparison of Glass Using Micro X-ray Fluorescence (µ-XRF) Spectrometry". This standard covers: SIGNIFICANCE AND USE 4.1 µ-XRF provides a means of simultaneously detecting major, minor, and trace elemental constituents in small glass fragments such as those frequently examined in forensic case work. It can be used at any point in the analytical scheme without concern for changing sample shape or sample properties, such as RI, due to its totally nondestructive nature. 4.2 Limits of detection (LOD) are dependent on several factors, including instrument configuration and operating parameters, sample thickness, and atomic number of the individual elements. Typical LODs range from parts per million (µgg-1) to percent (%). 4.3 µ-XRF provides simultaneous qualitative analysis for elements having an atomic number of eleven or greater. This multi-element capability permits detection of elements typically present in glass such as magnesium (Mg), silicon (Si), aluminum (Al), calcium (Ca), potassium (K), iron (Fe), titanium (Ti), strontium (Sr), and zirconium (Zr), as well as other elements that may be detectable in some glass by µ-XRF (for example, molybdenum (Mo), selenium (Se), or erbium (Er)) without the need for a predetermined elemental menu. 4.4 µ-XRF comparison of glass fragments provides additional discrimination power beyond that of RI or density comparisons, or both, alone. 4.5 The method precision should be established in each laboratory for the specific conditions and instrumentation in that laboratory. 4.6 When using small fragments having varying surface geometries and thicknesses, precision deteriorates due to take-off-angle and critical depth effects. Flat fragments with thickness greater than 1.5 mm do not suffer from these constraints, but are not always available as questioned specimens received in casework. As a consequence of the deterioration in precision for small fragments and the lack of appropriate calibration standards, quantitative analysis by µ-XRF is not typically used. 4.7 Appropriate sampling techniques should be used to account for natural h... SCOPE 1.1 This test method is for the determination of major, minor, and trace elements present in glass fragments. The elemental composition of a glass fragment can be measured through the use of µ-XRF analysis for comparisons of glass. 1.2 This test method covers the application of µ-XRF using mono- and poly- capillary optics, and an energy dispersive X-ray detector (EDS). 1.3 This test method does not replace knowledge, skill, ability, experience, education, or training and should be used in conjunction with professional judgment. 1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
SIGNIFICANCE AND USE 4.1 µ-XRF provides a means of simultaneously detecting major, minor, and trace elemental constituents in small glass fragments such as those frequently examined in forensic case work. It can be used at any point in the analytical scheme without concern for changing sample shape or sample properties, such as RI, due to its totally nondestructive nature. 4.2 Limits of detection (LOD) are dependent on several factors, including instrument configuration and operating parameters, sample thickness, and atomic number of the individual elements. Typical LODs range from parts per million (µgg-1) to percent (%). 4.3 µ-XRF provides simultaneous qualitative analysis for elements having an atomic number of eleven or greater. This multi-element capability permits detection of elements typically present in glass such as magnesium (Mg), silicon (Si), aluminum (Al), calcium (Ca), potassium (K), iron (Fe), titanium (Ti), strontium (Sr), and zirconium (Zr), as well as other elements that may be detectable in some glass by µ-XRF (for example, molybdenum (Mo), selenium (Se), or erbium (Er)) without the need for a predetermined elemental menu. 4.4 µ-XRF comparison of glass fragments provides additional discrimination power beyond that of RI or density comparisons, or both, alone. 4.5 The method precision should be established in each laboratory for the specific conditions and instrumentation in that laboratory. 4.6 When using small fragments having varying surface geometries and thicknesses, precision deteriorates due to take-off-angle and critical depth effects. Flat fragments with thickness greater than 1.5 mm do not suffer from these constraints, but are not always available as questioned specimens received in casework. As a consequence of the deterioration in precision for small fragments and the lack of appropriate calibration standards, quantitative analysis by µ-XRF is not typically used. 4.7 Appropriate sampling techniques should be used to account for natural h... SCOPE 1.1 This test method is for the determination of major, minor, and trace elements present in glass fragments. The elemental composition of a glass fragment can be measured through the use of µ-XRF analysis for comparisons of glass. 1.2 This test method covers the application of µ-XRF using mono- and poly- capillary optics, and an energy dispersive X-ray detector (EDS). 1.3 This test method does not replace knowledge, skill, ability, experience, education, or training and should be used in conjunction with professional judgment. 1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
ASTM E2926-17 is classified under the following ICS (International Classification for Standards) categories: 07.140 - Forensic science; 81.040.10 - Raw materials and raw glass. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM E2926-17 has the following relationships with other standards: It is inter standard links to ASTM E177-14, ASTM E177-13, ASTM E2330-12, ASTM E177-10, ASTM E177-08, ASTM E177-06b, ASTM E177-06a, ASTM E177-06, ASTM E177-04, ASTM E177-04e1, ASTM E2330-04, ASTM E177-90a(2002). Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM E2926-17 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E2926 − 17 An American National Standard
Standard Test Method for
Forensic Comparison of Glass Using Micro X-ray
Fluorescence (µ-XRF) Spectrometry
This standard is issued under the fixed designation E2926; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
One objective of a forensic glass examination is to compare glass specimens to determine if they
canbediscriminatedusingtheirphysical,opticalorchemicalproperties(forexample,color,refractive
index (RI), density, elemental composition). If the specimens are distinguishable, except for
acceptable and explainable variations, in any of these observed and measured properties, it may be
concluded that they did not originate from the same source of broken glass. If the specimens are
indistinguishable in all of these observed and measured properties, the possibility that they originated
from the same source of glass cannot be eliminated. The use of an elemental analysis method such as
micro X-ray fluorescence spectrometry (µ-XRF) yields high discrimination among sources of glass.
1. Scope 2. Referenced Documents
2.1 ASTM Standards:
1.1 This test method is for the determination of major,
E177 Practice for Use of the Terms Precision and Bias in
minor, and trace elements present in glass fragments. The
ASTM Test Methods
elemental composition of a glass fragment can be measured
E2330 Test Method for Determination of Concentrations of
through the use of µ-XRF analysis for comparisons of glass.
Elements in Glass Samples Using Inductively Coupled
1.2 This test method covers the application of µ-XRF using Plasma Mass Spectrometry (ICP-MS) for Forensic Com-
mono- and poly- capillary optics, and an energy dispersive parisons
X-ray detector (EDS).
3. Summary of Test Method
1.3 This test method does not replace knowledge, skill,
3.1 µ-XRF is a nondestructive elemental analysis technique
ability,experience,education,ortrainingandshouldbeusedin
based on the emission of characteristic X-rays following the
conjunction with professional judgment.
excitation of the specimen by an X-ray source using capillary
optics. Simultaneous multi-elemental analysis is typically
1.4 The values stated in SI units are to be regarded as
achieved for elements of atomic number eleven or greater.
standard. No other units of measurement are included in this
standard. 3.2 Glass fragments usually do not require sample prepara-
tion prior to analysis by µ-XRF. Cleaning of specimens may be
1.5 This standard does not purport to address all of the
performed to remove any surface debris.
safety concerns, if any, associated with its use. It is the
3.3 Specimens are mounted and placed into the instrument
responsibility of the user of this standard to establish appro-
chamber and subjected to an X-ray beam. The characteristic
priate safety and health practices and determine the applica-
X-rays emitted by the specimen are detected using an energy
bility of regulatory limitations prior to use.
dispersive X-ray detector and displayed as a spectrum of
energy versus intensity.
3.4 Qualitative analysis is accomplished by identifying
elements present in the specimen based on their characteristic
X-ray energies.
ThistestmethodisunderthejurisdictionofASTMCommitteeE30onForensic
Sciences and is the direct responsibility of Subcommittee E30.01 on Criminalistics. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Feb. 1, 2017. Published February 2017. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2013. Last previous edition approved in 2013 as E2926 – 13. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E2926-17. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E2926 − 17
3.5 Semi-quantitative analysis is accomplished by compar- capability and less analysis time. LA-ICP-MS drawbacks are
ing the relative area under the peaks of characteristic X-rays of greater instrument cost and complexity of operation.
certain elements.
4.10 Scanning Electron Microscopy with EDS (SEM-EDS)
3.6 Spectral and elemental ratio comparisons of the glass is also available for elemental analysis, but it is of limited use
for forensic glass source discrimination due to poor detection
specimens are conducted for source discrimination or associa-
tion. limits for higher atomic number elements present in glass at
trace concentration levels. However, discrimination of sources
4. Significance and Use that have indistinguishable RIs and densities may be possible.
4.1 µ-XRF provides a means of simultaneously detecting
5. Interferences
major, minor, and trace elemental constituents in small glass
5.1 Peak overlaps occur in various regions of the EDS
fragments such as those frequently examined in forensic case
spectrum (1). In glass, such interferences include the overlap
work. It can be used at any point in the analytical scheme
of characteristic X-ray lines (for example, Ti K-series and Ba
without concern for changing sample shape or sample
L-series), sum peaks, primary X-ray source excitation peaks
properties, such as RI, due to its totally nondestructive nature.
(for example, Rh), and escape peaks. In general, automated
4.2 Limits of detection (LOD) are dependent on several
deconvolution algorithms are included in data processing
factors, including instrument configuration and operating
software that adequately address such overlaps. EDS spectra
parameters, sample thickness, and atomic number of the
shall be manually inspected to ensure that potential peak
individual elements. Typical LODs range from parts per
overlaps are considered and addressed.
-1
million (µgg ) to percent (%).
4.3 µ-XRF provides simultaneous qualitative analysis for 6. Apparatus
elements having an atomic number of eleven or greater. This
6.1 A µ-XRF spectrometer with an EDS detector is em-
multi-element capability permits detection of elements typi-
ployed. Most commercial-grade µ-XRF systems with EDS
cally present in glass such as magnesium (Mg), silicon (Si),
detectors should be adequate for forensic analysis of glass.The
aluminum (Al), calcium (Ca), potassium (K), iron (Fe), tita-
µ-XRFsystemmust,however,meetthefollowingperformance
nium (Ti), strontium (Sr), and zirconium (Zr), as well as other
specifications:
elements that may be detectable in some glass by µ-XRF (for
6.1.1 The spot size(s) must be within the range(s) of
example, molybdenum (Mo), selenium (Se), or erbium (Er))
approximately 10 µm to 2 mm; the spot size used may be
without the need for a predetermined elemental menu.
adjustable to different sizing for instruments with appropriate
4.4 µ-XRF comparison of glass fragments provides addi- optics.
tional discrimination power beyond that of RI or density 6.1.2 The instrument must be capable of operating at an
comparisons, or both, alone. accelerating voltage of 35 kV or greater.
6.1.3 The EDS detector must be capable of a resolution that
4.5 The method precision should be established in each
is typically less than 180 eV, measured as the full width at half
laboratory for the specific conditions and instrumentation in
themaximumheightoftheMnKαpeak;betterresolutionswill
that laboratory.
provide improved discrimination of adjacent or overlapping
4.6 When using small fragments having varying surface
peaks, or both.
geometries and thicknesses, precision deteriorates due to take-
6.1.4 Acalibrated, scaled display of energy units (keV) and
off-angle and critical depth effects. Flat fragments with thick-
the ability to identify and label X-ray lines is required for the
ness greater than 1.5 mm do not suffer from these constraints,
EDS system.
but are not always available as questioned specimens received
6.2 Energy Calibration Material—Capable of calibrating
incasework.Asaconsequenceofthedeteriorationinprecision
the EDS detector at both the low (<2 keV) and high (>6 keV)
for small fragments and the lack of appropriate calibration
X-ray spectral regions.
standards,quantitativeanalysisbyµ-XRFisnottypicallyused.
6.3 An X-ray source that does not yield significant spectral
4.7 Appropriate sampling techniques should be used to
interferences with the characteristic X-ray lines for the ele-
account for natural heterogeneity of the material, varying
ments typically found in glass is required. Several X-ray
surface geometries, and potential critical depth effects.
sources are available; a rhodium X-ray source is preferred for
4.8 Inductively Coupled Plasma-Optical Emission Spec-
appropriate excitation energy and minimal spectral interfer-
trometry (ICP-OES) and Inductively Coupled Plasma-Mass
ences for elements in glass. Other X-ray sources such as Mo
Spectrometry (ICP-MS) may also be used for trace elemental
X-ray tubes cause interferences with discriminating elements
analysis of glass and offer lower minimum detection levels and
such as Zr.
the ability for quantitative analysis. However, these methods
6.4 A vacuum sample chamber, sample stage, and visual-
are destructive, and require larger sample sizes and much
ization system are required.
longer sample preparation times (Test Method E2330).
4.9 LaserAblation-InductivelyCoupledPlasma-MassSpec-
trometry (LA-ICP-MS) uses comparable specimen sizes to
The boldface numbers in parentheses refer to a list of references at the end of
those used for µ-XRF but offers better LODs, quantitative this standard.
E2926 − 17
6.5 The sample holder, sample support film, and mounting established by the laboratory or analyst for this procedure (10
material (for example, adhesive with low trace elements) must % tolerance is recommended).
prevent background interferences.
8.2.2 Demonstrate that Ti and Sr have LOD in a soda-lime
glass matrix of 75 ppm or less (as described in 11.1) for the
7. Hazards instrumental parameters used for collection of spectra from the
glass specimens. NIST SRM 1831 is a suitable sample for this
7.1 The X-ray sources emit radiation when energized. For
purpose.
operator safety, appropriate shielding and safety interlocks
must be in place and operational.
9. Procedure
8. Calibration and Standardization
9.1 Specimen Preparation:
9.1.1 Examine glass fragments using stereomicroscopy to
8.1 Apparatus—The instrument must be optimized as in
determine an appropriate preparation method for the specimen.
accordance with manufacturer’s instruction.
9.1.2 Ifnecessary,cleanthespecimentoremoveanysurface
8.1.1 Energy Calibration—Calibrate the X-ray energy scale
contamination. Cleaning may include washing specimens with
to characteristic X-ray emission lines by either measuring the
soap and water, with or without ultrasonication, and rinsing in
centroid energy of a low- (<2 keV) and high- (>6 keV) energy
deionized water, followed by rinsing in acetone, methanol, or
peak or by using software provided by the instrument manu-
ethanol,anddrying.Soakinginvariousconcentrationsofnitric
facturer. For example, the aluminum (1.486 keV) and copper
acid for 30 minutes or longer, rinsing with deionized water and
(Cu) (8.040 keV) Kα-X-ray energy lines may be used.
ethanol, and drying prior to analysis removes most surface
8.1.2 Stage Calibration—For automated or multiple point
contamination without affecting the measured concentrations
analysis, initialize the stage position to assure that the stage
of elements inherent in the glass. However, the use of nitric
coordinates accurately reflect the stage position.
acid may remove any surface coating that may be present.
8.1.3 Optical Alignment:
9.1.3 Mount the specimen for analysis.
8.1.3.1 Align X-ray optics to obtain the maximum count
9.1.3.1 The specimen mounting technique depends on the
rate.
sample size and shape, beam size, X-ray fluorescence spec-
8.1.3.2 Align visualization optics to ensure that the visual
trometer chamber design and purpose of the examination.
target area coincides with the X-ray beam position.
9.1.3.2 Raise specimens off the surface of the stage for
8.1.4 Spot Size Measurement—Determine spot size of the
analysis using an X-ray transparent sample holder or support-
X-ray beam at the focal point of the visualization optics. For
ive X-ray film, or both. This positioning reduces X-ray scatter
instruments with continuous variable spot size options, deter-
off of the surface of the stage and, hence, improves sample
mine the spot size at multiple settings and interpolate the
signal-to-noise. Because analysis is performed under vacuum,
others.
ensurethatspecimensretaintheirpositiononthesampleholder
8.1.5 Reference Materials—Analyze a glass certified refer-
by securing with adhesive. Prior to analysis, analyze a small
ence material (CRM) (for example, NISTSRM 1831) to verify
amount of the adhesive to determine the presence of any
the calibration of X-ray energy lines for elements present in
elements that could interfere with those in the specimen.When
glass and determine if the instrument response is within
small amounts of adhesive are used and beam overspill (X-ray
acceptable limits. Measure this glass CRM using the same
beam extending beyond the perimeter of the specimen) is
analysis parameters as the glass specimens. Use this reference
avoided, little to no interference from the adhesive will be
glass sample to normalize element ratios for interlaboratory
observed.
comparisons, intralaboratory data collection from different
9.1.3.3 Position specimens to present as flat a surface as
analytical runs, and databasing applications to improve preci-
possible to the impinging excitation X-ray beam. If necessary,
sion.
use a small amount of adhesive to facilitate this positioning.
8.1.6 Blanks—Collect a spectrum of a specimen devoid of
9.1.3.4 For comparisons, glass specimen should be of simi-
elements having an atomic number of 11 or greater, such as the
lar size, shape, and thickness to each other. For full thickness
plastic stage plate or an area of the support material having no
fragments of float glass, comparisons should be made between
glass present. Record any system peaks present for future
similar surface types (for example, non-float surface to non-
reference.
float surface).
8.2 Quality Assurance:
9.1.4 Place sample(s) in the instrument’s analysis chamber.
8.2.1 The performance of the instrument must be monitored
For automated multiple point analyses, it may be necessary to
routinely and the frequency and tolerances should be set by
secure the sample/sample holder to the instrument stage.
each laboratory.
9.1.5 Evacuate the chamber; samples should be run under
8.2.1.1 Check the system calibration prior to the perfor-
vacuum.
mance of an analysis.
9.1.6 Target specimen areas that are relatively flat in topog-
8.2.1.2 Check the performance of the X-ray source using a
raphy and focus imaging optics. Avoid excitation beam over-
known element standard (for example, Cu). Maximum counts
spill when possible.
for the system should be obtained utilizing system operating
parameters established by the laboratory. Maximum counts 9.2 Operating Conditions—The following are suggested as
should not show appreciable drift from acceptable parameters a general guide for instrument operating conditions:
E2926 − 17
9.2.1 Turn on the X-ray source. Set the excitation voltage to 10.6 Compare the spectra using peak identification, spectral
at least 35 kV in order to provide sufficient overvoltage comparisons, and peak intensity ratio comparisons.
necessary for efficient X-ray excitation of the K-lines of higher
10.6.1 Peak Identification—Compare detected elements of
atomic number elements, such as As, Rb, Sr, Zr, and Mo. the questioned and known glass spectra.
Higherbeamenergieswilltypicallyimprovedetectionlimitsof
10.6.2 Spectral Comparisons—Visually compare the spec-
those elements. Once the beam excitation voltage is
tral shapes and relative peak heights of the questioned and
established, it should not be changed between specimens in a
known glass specimen spectra.
givencomparisonset.MostoftheX-raylinesproducedmaybe
10.6.3 Peak Intensity Ratio Comparisons—When the area
displayed with an energy range of 0 to 20 keV.
of a characteristic energy peak of an element has a signal-to-
9.2.2 Select an appropriate X-ray optic size for the analysis
noise ratio of ten or more, that element may be used in
of the specimen. For instruments with variable spot size
semi-quantitativeanalysissuchaspeakintensityratiocompari-
options, this size is determined from the setpoint calibrated as
sons. This criterion may not apply when spectral artifacts,
described in 8.1.4. Once the spot size is established, it should
overlapping energies of other elements, or shoulder peaks, or a
not be changed between specimens in a given comparison set.
combination thereof, are present within the characteristic
9.2.3 Set the pulse processor time constant a
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E2926 − 13 E2926 − 17
Standard Test Method for
Forensic Comparison of Glass Using Micro X-ray
Fluorescence (μ-XRF) Spectrometry
This standard is issued under the fixed designation E2926; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
One objective of a forensic glass examination is to compare glass specimens to determine if they
can be discriminated using their physical, optical or chemical properties (for example, color, refractive
index (RI), density, elemental composition). If the specimens are distinguishable, except for
acceptable and explainable variations, in any of these observed and measured properties, it may be
concluded that they did not originate from the same source of broken glass. If the specimens are
indistinguishable in all of these observed and measured properties, the possibility that they originated
from the same source of glass cannot be eliminated. The use of an elemental analysis method such as
micro X-ray fluorescence spectrometry (μ-XRF) yields high discrimination among sources of glass.
1. Scope
1.1 This test method is for the determination of major, minor, and trace elements present in glass fragments. The elemental
composition of a glass fragment can be measured through the use of μ-XRF analysis for comparisons of glass.
1.2 This test method covers the application of μ-XRF using mono- and poly- capillary optics, and an energy dispersive X-ray
detector (EDS).
1.3 This test method does not replace knowledge, skill, ability, experience, education, or training and should be used in
conjunction with professional judgment.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
E177 Practice for Use of the Terms Precision and Bias in ASTM Test Methods
E2330 Test Method for Determination of Concentrations of Elements in Glass Samples Using Inductively Coupled Plasma Mass
Spectrometry (ICP-MS) for Forensic Comparisons
3. Summary of Test Method
3.1 μ-XRF is a nondestructive elemental analysis technique based on the emission of characteristic X-rays following the
excitation of the specimen by an X-ray source using capillary optics. Simultaneous multi-elemental analysis is typically achieved
for elements of atomic number eleven or greater.
3.2 Glass fragments usually do not require sample preparation prior to analysis by μ-XRF. Cleaning of specimens may be
performed to remove any surface debris.
This test method is under the jurisdiction of ASTM Committee E30 on Forensic Sciences and is the direct responsibility of Subcommittee E30.01 on Criminalistics.
Current edition approved June 15, 2013Feb. 1, 2017. Published July 2013February 2017. Originally approved in 2013. Last previous edition approved in 2013 as E2926
– 13. DOI: 10.1520/E2926-13.10.1520/E2926-17.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E2926 − 17
3.3 Specimens are mounted and placed into the instrument chamber and subjected to an X-ray beam. The characteristic X-rays
emitted by the specimen are detected using an energy dispersive X-ray detector and displayed as a spectrum of energy versus
intensity.
3.4 Qualitative analysis is accomplished by identifying elements present in the specimen based on their characteristic X-ray
energies.
3.5 Semi-quantitative analysis is accomplished by comparing the relative area under the peaks of characteristic X-rays of certain
elements.
3.6 Spectral and elemental ratio comparisons of the glass specimens are conducted for source discrimination or association.
4. Significance and Use
4.1 μ-XRF provides a means of simultaneously detecting major, minor, and trace elemental constituents in small glass fragments
such as those frequently examined in forensic case work. It can be used at any point in the analytical scheme without concern for
changing sample shape or sample properties, such as RI, due to its totally nondestructive nature.
4.2 Limits of detection (LOD) are dependent on several factors, including instrument configuration and operating parameters,
-1
sample thickness, and atomic number of the individual elements. Typical LODs range from parts per million (μgg ) to percent (%).
4.3 μ-XRF provides simultaneous qualitative analysis for elements having an atomic number of eleven or greater. This
multi-element capability permits detection of elements typically present in glass such as magnesium (Mg), silicon (Si), aluminum
(Al), calcium (Ca), potassium (K), iron (Fe), titanium (Ti), strontium (Sr), and zirconium (Zr), as well as other elements that may
be detectable in some glass by μ-XRF (for example, molybdenum (Mo), selenium (Se), or erbium (Er)) without the need for a
predetermined elemental menu.
4.4 μ-XRF comparison of glass fragments provides additional discrimination power beyond that of RI or density comparisons,
or both, alone.
4.5 The method precision should be established in each laboratory for the specific conditions and instrumentation in that
laboratory.
4.6 When using small fragments having varying surface geometries and thicknesses, precision deteriorates due to take-off-angle
and critical depth effects. Flat fragments with thickness greater than 1.5 mm do not suffer from these constraints, but are not always
available as questioned specimens received in casework. As a consequence of the deterioration in precision for small fragments
and the lack of appropriate calibration standards, quantitative analysis by μ-XRF is not typically used.
4.7 Appropriate sampling techniques should be used to account for natural heterogeneity of the material, varying surface
geometries, and potential critical depth effects.
4.8 Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES) and Inductively Coupled Plasma-Mass Spectrom-
etry (ICP-MS) may also be used for trace elemental analysis of glass and offer lower minimum detection levels and the ability for
quantitative analysis. However, these methods are destructive, and require larger sample sizes and much longer sample preparation
times (Test Method E2330).
4.9 Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry (LA-ICP-MS) uses comparable specimen sizes to those
used for μ-XRF but offers better LODs, quantitative capability and less analysis time. LA-ICP-MS drawbacks are greater
instrument cost and complexity of operation.
4.10 Scanning Electron Microscopy with EDS (SEM-EDS) is also available for elemental analysis, but it is of limited use for
forensic glass source discrimination due to poor detection limits for higher atomic number elements present in glass at trace
concentration levels. However, discrimination of sources that have indistinguishable RIs and densities may be possible.
5. Interferences
5.1 Peak overlaps occur in various regions of the EDS spectrum.spectrum (1). In glass, such interferences include the overlap
of characteristic X-ray lines (for example, Ti K-series and Ba L-series), sum peaks, primary X-ray source excitation peaks (for
example, Rh), and escape peaks. In general, automated deconvolution algorithms are included in data processing software that
adequately address such overlaps. EDS spectra shall be manually inspected to ensure that potential peak overlaps are considered
and addressed.
Available from X-ray Transition Energies Database, National Institute of Standards and Technology (NIST), 100 Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070,
http://physics.nist.gov/PhysRefData/XrayTrans/Html/search.html.
Latkoczy, C., Becker, S. Ducking, M., Gunther, D., Hoogewerff, J.A., Almirall, J., Buscaglia, J., Dobney, A., Koons, R., Montero, S., van der Peijl, G., Stoecklein, W.,
Trejos, T., Watling, J.R., and Zdanowicz, V., “Development and evaluation of a standard method for the quantitative determination of elements in float glass samples by
LA-ICP-MS,” Journal of Forensic Sciences, Vol 50, No. 6, 2005, pp. 1327–1341.The boldface numbers in parentheses refer to a list of references at the end of this standard.
E2926 − 17
6. Apparatus
6.1 A μ-XRF spectrometer with an EDS detector is employed. Most commercial-grade μ-XRF systems with EDS detectors
should be adequate for forensic analysis of glass. The μ-XRF system must, however, meet the following performance
specifications:
6.1.1 The spot size(s) must be within the range(s) of approximately 10 μm to 2 mm; the spot size used may be adjustable to
different sizing for instruments with appropriate optics.
6.1.2 The instrument must be capable of operating at an accelerating voltage of 35 kV or greater.
6.1.3 The EDS detector must be capable of a resolution that is typically less than 180 eV, measured as the full width at half the
maximum height of the Mn Kα peak; better resolutions will provide improved discrimination of adjacent or overlapping peaks,
or both.
6.1.4 A calibrated, scaled display of energy units (keV) and the ability to identify and label X-ray lines is required for the EDS
system.
6.2 Energy Calibration Material—Capable of calibrating the EDS detector at both the low (<2 keV) and high (>6 keV) X-ray
spectral regions.
6.3 An X-ray source that does not yield significant spectral interferences with the characteristic X-ray lines for the elements
typically found in glass is required. Several X-ray sources are available; a rhodium X-ray source is preferred for appropriate
excitation energy and minimal spectral interferences for elements in glass. Other X-ray sources such as Mo X-ray tubes cause
interferences with discriminating elements such as Zr.
6.4 A vacuum sample chamber, sample stage, and visualization system are required.
6.5 The sample holder, sample support film, and mounting material (for example, adhesive with low trace elements) must
prevent background interferences.
7. Hazards
7.1 The X-ray sources emit radiation when energized. For operator safety, appropriate shielding and safety interlocks must be
in place and operational.
8. Calibration and Standardization
8.1 Apparatus—The instrument must be optimized as in accordance with manufacturer’s instruction.
8.1.1 Energy Calibration—calibrateCalibrate the X-ray energy scale to characteristic X-ray emission lines by either measuring
the centroid energy of a low- (<2 keV) and high- (>6 keV) energy peak or by using software provided by the instrument
manufacturer. For example, the aluminum (1.486 keV) and copper (Cu) (8.040 keV) Kα-X-ray energy lines may be used.
8.1.2 Stage Calibration—For automated or multiple point analysis, initialize the stage position to assure that the stage
coordinates accurately reflect the stage position.
8.1.3 Optical Alignment:
8.1.3.1 Align X-ray optics to obtain the maximum count rate.
8.1.3.2 Align visualization optics to ensure that the visual target area coincides with the X-ray beam position.
8.1.4 Spot Size Measurement—Determine spot size of the X-ray beam at the focal point of the visualization optics. For
instruments with continuous variable spot size options, determine the spot size at multiple settings and interpolate the others.
8.1.5 Reference Materials—Analyze a glass certified reference material (CRM) (for example, NIST SRM 1831) to verify the
calibration of X-ray energy lines for elements present in glass and determine if the instrument response is within acceptable limits.
Measure this glass CRM using the same analysis parameters as the glass specimens. Use this reference glass sample to normalize
element ratios for interlaboratory comparisons, intralaboratory data collection from different analytical runs, and databasing
applications to improve precision.
8.1.6 Blanks—Collect a spectrum of a specimen devoid of elements having an atomic number of 11 or greater, such as the plastic
stage plate or an area of the support material having no glass present. Record any system peaks present for future reference.
8.2 Quality Assurance:
8.2.1 The performance of the instrument must be monitored routinely and the frequency and tolerances should be set by each
laboratory.
8.2.1.1 Check the system calibration prior to the performance of an analysis.
8.2.1.2 Check the performance of the X-ray source using a known element standard (for example, Cu). Maximum counts for
the system should be obtained utilizing system operating parameters established by the laboratory. Maximum counts should not
show appreciable drift from acceptable parameters established by the laboratory or analyst for this procedure (10 % tolerance is
recommended).
8.2.2 Demonstrate that Ti and Sr have LOD in a soda-lime glass matrix of 75 ppm or less (as described in 11.1) for the
instrumental parameters used for collection of spectra from the glass specimens. NIST SRM 1831 is a suitable sample for this
purpose.
E2926 − 17
9. Procedure
9.1 Specimen Preparation:
9.1.1 Examine glass fragments using stereomicroscopy to determine an appropriate preparation method for the specimen.
9.1.2 If necessary, clean the specimen to remove any surface contamination. Cleaning may include washing specimens with
soap and water, with or without ultrasonication, and rinsing in deionized water, followed by rinsing in acetone, methanol, or
ethanol, and drying. Soaking in various concentrations of nitric acid for 30 minutes or longer, rinsing with deionized water and
ethanol, and drying prior to analysis removes most surface contamination without affecting the measured concentrations of
elements inherent in the glass. However, the use of nitric acid may remove any surface coating that may be present.
9.1.3 Mount the specimen for analysis.
9.1.3.1 The specimen mounting technique depends on the sample size and shape, beam size, X-ray fluorescence spectrometer
chamber design and purpose of the examination.
9.1.3.2 Raise specimens off the surface of the stage for analysis using an X-ray transparent sample holder or supportive X-ray
film, or both. This positioning reduces X-ray scatter off of the surface of the stage and, hence, improves sample signal-to-noise.
Because analysis is performed under vacuum, ensure that specimens retain their position on the sample holder by securing with
adhesive. Prior to analysis, analyze a small amount of the adhesive to determine the presence of any elements that could interfere
with those in the specimen. When small amounts of adhesive are used and beam overspill (X-ray beam extending beyond the
perimeter of the specimen) is avoided, little to no interference from the adhesive will be observed.
9.1.3.3 Position specimens to present as flat a surface as possible to the impinging excitation X-ray beam. If necessary, use a
small amount of adhesive to facilitate this positioning.
9.1.3.4 For comparisons, glass specimen should be of similar size, shape, and thickness to each other. For full thickness
fragments of float glass, comparisons should be made between similar surface types (for example, non-float surface to non-float
surface).
9.1.4 Place sample(s) in the instrument’s analysis chamber. For automated multiple point analyses, it may be necessary to secure
the sample/sample holder to the instrument stage.
9.1.5 Evacuate the chamber; samples should be run under vacuum.
9.1.6 Target specimen areas that are relatively flat in topography and focus imaging optics. Avoid excitation beam overspill
when possible.
9.2 Operating Conditions—The following are suggested as a general guide for instrument operating conditions:
9.2.1 Turn on the X-ray source. Set the excitation voltage to at least 35 kV in order to provide sufficient overvoltage necessary
for efficient X-ray excitation of the K-lines of higher atomic number elements, such as As, Rb, Sr, Zr, and Mo. Higher beam
energies will typically improve detection limits of those elements. Once the beam excitation voltage is established, it should not
be changed between specimens in a given comparison set. Most of the X-ray lines produced may be displayed with an energy range
of 0 to 20 keV.
9.2.2 Select an appropriate X-ray optic size for the analysis of the specimen. For instruments with variable spot size options,
this size is determined from the setpoint calibrated as described in 8.1.4. Once the spot size is established, it should not be changed
between specimens in a given comparison set.
9.2.3 Set the pulse processor time constant at a midrange value; this is a compromise between maximum count rate and
maximum spectral resolution. The optimal count rate is generally provided by the instrument manufacturer. Once established, do
not change the pulse processor time between specimens in a given comparison set.
9.2.4 Adjust the beam current for each specimen as needed to yield a maximum X-ray detector dead time not to exceed 50
percent.
9.3 Specimen Analysis:
9.3.1 For each specimen, collect a spectrum for a live time that provides reasonable counting statistics for trace element peaks.
For a μ-XRF system with a 100 μm monocapillary and a Si(Li) detector, 1200 live seconds is generally sufficient.
9.3.2 Collect replicate spectra to ensure that the questioned glass fragments and known glass source(s) are adequately
characterized. When practical, analyze a minimum of three replicates on each questioned specimen examined and nine replicates
on known glass sources.
10. Calculation and Interpretation of Results
10.1 Examine the spectrum, and identify and label the peaks.
10.2 Characteristic X-ray energies may be obtained from automatic element identification software; however, peak identifica-
tions shall be manually verified. Energy slide rules, computer generated theoretical fit curves that can be superimposed on the
spectrum, and published tables can be used for this purpose.
10.3 Spectral artifacts, such as sum peaks and escape peaks associated with the major peaks, should be considered and
corrected.
----------------------
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...