(Main)Radiation protection instrumentation - In vivo counters - Classification, general requirements and test procedures for portable, transportable and installed equipment
Radiation protection instrumentation - In vivo counters - Classification, general requirements and test procedures for portable, transportable and installed equipment
Specifies the classification, general design requirements, performance characteristics and test procedures for in vivo counting systems for detecting trace amounts of radionuclides in the bodies of persons working in nuclear power plants, laboratories and facilities handling radionuclides, and inhabitants living on territory which may be contaminated. The purpose is to determine the dose equivalent to organs and the effective dose of internal radiation for the whole body.
Strahlenschutz-Messgeräte - Einrichtungen für die in-vivo-Überwachung - Ganz- und Teilkörperzähler - Klassifizierung, allgemeine Anforderungen und Prüfverfahren für tragbare, transportable und festinstallierte Einrichtungen
Instrumentation pour la radioprotection - Systèmes de mesure in vivo - Classification, exigences générales et procédures d'essai pour les appareils portables, mobiles ou à poste fixe
Spécifie la classification, les exigences générales de conception, les caractéristiques de performance et les procédures d'essai des systèmes de mesure in vivo utilisés pour détecter les traces de radionucléides dans le corps des personnes travaillant dans des centrales nucléaires, des laboratoires et des installations où l'on manipule des radionucléides, ou des personnes vivant dans des zones susceptibles d'être contaminées. Il s'agit de déterminer la dose équivalente aux organes et la dose efficace du rayonnement interne pour l'ensemble du corps.
Oprema za varstvo pred sevanjem – Merilniki notranje kontaminacije – Razvrstitev, splošne zahteve in postopki preskušanja za prenosno, premično in inštalirano opremo
General Information
Standards Content (Sample)
SLOVENSKI SIST EN 61582:2006
STANDARD
september 2006
Oprema za varstvo pred sevanjem – Merilniki notranje kontaminacije –
Razvrstitev, splošne zahteve in postopki preskušanja za prenosno, premično
in inštalirano opremo
Radiation protection instrumentation - In vivo counters - Classification, general
requirements and test procedures for portable, transportable and installed
equipment
ICS 13.280; 17.240 Referenčna številka
SIST EN 61582:2006(en)
© Standard je založil in izdal Slovenski inštitut za standardizacijo. Razmnoževanje ali kopiranje celote ali delov tega dokumenta ni dovoljeno
---------------------- Page: 1 ----------------------
EUROPEAN STANDARD
EN 61582
NORME EUROPÉENNE
May 2006
EUROPÄISCHE NORM
ICS 13.280
English version
Radiation protection instrumentation -
In vivo counters -
Classification, general requirements and test procedures
for portable, transportable and installed equipment
(IEC 61582:2004, modified)
Instrumentation pour la radioprotection - Strahlenschutz-Messgeräte -
Systèmes de mesure in vivo - Einrichtungen für die
Classification, exigences générales in-vivo-Überwachung -
et procédures d'essai pour les appareils Ganz- und Teilkörperzähler -
portables, mobiles ou à poste fixe Klassifizierung, allgemeine Anforderungen
(CEI 61582:2004, modifiée) und Prüfverfahren für tragbare,
transportable und festinstallierte
Einrichtungen
(IEC 61582:2004, modifiziert)
This European Standard was approved by CENELEC on 2006-02-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Cyprus, the Czech
Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.
CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
Central Secretariat: rue de Stassart 35, B - 1050 Brussels
© 2006 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 61582:2006 E
---------------------- Page: 2 ----------------------
EN 61582:2006 - 2 -
Foreword
The text of the International Standard IEC 61582:2004, prepared by SC 45B, Radiation protection
instrumentation, of IEC TC 45, Nuclear instrumentation, together with common modifications prepared by
the CENELEC BTTF 111-3, Nuclear instrumentation and radiation protection instrumentation, was
submitted to the formal vote and was approved by CENELEC as EN 61582 on 2006-02-01.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
(dop) 2007-02-01
national standard or by endorsement
– latest date by which the national standards conflicting
(dow) 2009-02-01
with the EN have to be withdrawn
Annex ZA has been added by CENELEC.
__________
---------------------- Page: 3 ----------------------
- 3 - EN 61582:2006
Endorsement notice
The text of the International Standard IEC 61582:2004 was approved by CENELEC as a European
Standard with agreed common modifications as given below.
COMMON MODIFICATIONS
1 Scope and object
th
Delete in the 10 paragraph, second line: “of the whole body monitor”.
3 Terms and definitions
3.10 relative intrinsic error
Delete the last word of the last paragraph: “activity”.
3.13 energy resolution
Extend the term in order to read “relative energy resolution”.
Delete “full width at half maximum (FWHM)” below the indication of the term.
Shift the first paragraph behind the formula and start the shifted paragraph with
“Where the full width is measured …”.
Replace in the text in front of the formula “R ” with “ε ”
e R’
Replace in the formula “FWHM” with “ε ”
R’
Delete the “%” at the end of the formula.
4 Classification
4.1 General classification
In the first line, replace “type of elements” with “radionuclides”.
7 Characteristics of equipment for high-energy emitter measurements
100 keV to 3 MeV
7.3 Energy resolution
In the second paragraph, replace “9 %” with “10 %”.
7.4 Integral non-linearity
In the first and second paragraph, replace “conversion response (integral non-
linearity)” with “integral non-linearity (conversion response)”.
8 Performance requirements and test procedures for low-energy emitter
measurements
8.1 General test procedures
8.1.2 Tests performed under standard test conditions
Replace reference to “Tables 2, 3 and 4” with reference to “Tables 2 and 3”.
---------------------- Page: 4 ----------------------
EN 61582:2006 - 4 -
8.4 Radiation characteristics
8.4.2 Linearity (in relation to activity)
8.4.2.3 Expression of results
Correct the explanation of F in order to read “is the relative uncertainty of the
SA
conventionally true activity of the test phantom in per cent (95 % confidence)”,
extend the explanation of F in order to read “… of the test phantom relative to the
SR
other sources …”
8.4.4 Determination of the integral non-linearity (INL) error
8.4.4.2 INL determination
Insert in the first paragraph behind “conversion characteristics”: “(channel number N
versus energy E )”,
γ
replace the first explanation below the formula with “the channel number for energy
ZERO”
and in the second explanation, replace “k” with “K”
8.4.8 Determination of the maximum count rate
8.4.8.2 Method of test
1 2
Replace in the formula “C ” with “C ”
2 1
8.4.9 Determination of stability
8.4.9.2 Method of test
Replace in the last formula and the paragraph i
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.