ISO/TR 22572:2026
(Main)Ophthalmic optics — Spectacle lenses — Measurement of refractive index
- Abstract
This document describes the possible methods of determining the refractive index of spectacle lenses, whether in uncut, cut or final mounted form. It lists several destructive and non-destructive test methods having different accuracies.
- Status
- Published
- Publication Date
- 15-Nov-2026
- Technical Committee
- ISO/TC 172/SC 7 - Ophthalmic optics and instruments
- Drafting Committee
- ISO/TC 172/SC 7/WG 3 - Spectacle lenses
- Current Stage
- 6060 - International Standard published - 60 Publication
- Start Date
- 16-Nov-2026
- Due Date
- 16-Nov-2026
- Completion Date
- 16-Nov-2026
ISO/TR 22572:2026 is an ISO Technical Report on measuring the refractive index of spectacle lenses. It describes destructive and non-destructive methods for uncut, cut, and final mounted lenses, and it is written for people who need to compare lens materials or verify a manufacturer’s claimed index.
This report is especially useful where a finished spectacle lens should be checked without being ruined. It gives a practical overview of several measurement approaches with different accuracies, from traditional prism methods to newer interferometric and prism-coupled instruments.
What does ISO/TR 22572:2026 specify?
ISO/TR 22572:2026 describes possible ways to determine the refractive index of spectacle lenses in uncut, cut, or final mounted form. The report compares destructive and non-destructive methods and sets the context for choosing a method based on available equipment, required accuracy, and whether the lens may be damaged.
The document is organized into eight main clauses and five informative annexes.
| Annex | What it covers |
|---|---|
| Annex A | Comparison of the different destructive methods |
| Annex B | Expected uncertainty when used with spectacle lens samples |
| Annex C | Group refractive index and phase refractive index |
| Annex D | Discussion of the surface power method |
| Annex E | Details of refractive index matching liquids |
Clause 4 addresses laboratory temperature. Clauses 5, 6, 7, and 8 cover traditional destructive methods, classical non-destructive methods, newer non-destructive methods, and a prism coupled refractometer.
What are the key points of ISO/TR 22572:2026?
ISO/TR 22572:2026 is not a test standard with one single method. Instead, it shows several methods that can be used on spectacle lenses, explains how they work, and points out the practical limits of each method.
Laboratory temperature matters
Clause 4 says measurement can be carried out at ambient laboratory temperature, usually either 23 °C or 27 °C, according to national practice. In practice, this means the work can be done under normal lab conditions, but the temperature should be controlled enough to support the accuracy needed for refractive index measurement.
Destructive methods require sample preparation
Clause 5 covers the minimum deviation method, the V-block method, and the Abbe refractometer. These methods generally require a prism or slab to be cut from the lens, with polished edges or a flat surface, so they are suitable for stock lenses but not for a wearer’s finished lens when damage must be avoided.
The minimum deviation method uses a goniometer or spectrometer and relies on prism geometry and Snell’s law. The V-block method uses a 90° prism in a V-block refractometer and may use a refractive index matching liquid to reduce scattered light. The Abbe refractometer uses the critical angle with a contacting liquid and is limited by the refractive index of the instrument prism and the sample.
Accuracy depends strongly on preparation and geometry
Clause 5.4 states that, for spectacle lenses, a minimum accuracy of 0.01 is expected. The report also shows that destructive methods are affected by how well the sample is cut, polished, and aligned, so the practical result may be less precise than ideal theory suggests.
This matters because a lab or buyer should not assume that the theoretical capability of the instrument is the same as the result on an actual spectacle lens sample.
Classical non-destructive methods use lens shape and power
Clause 6 explains methods based on sagittal depth, radius of curvature, vertex power, and immersion. These approaches can avoid destroying the lens, which is their main advantage for finished eyewear and for verifying claims on the final product.
The report shows that the sag formula can be used to estimate radius of curvature, and then the surface power equations can be used to derive refractive index. It also explains why errors in sag measurement, surface asphericity, and lens power can affect the result, especially for progressive or otherwise non-spherical surfaces.
Immersion methods need careful orientation and the right meridian
Clause 6.2 explains that immersing the lens in a known medium changes the measured vertex power, allowing the refractive index to be derived. For astigmatic lenses, the same meridian should be used in air and in the medium, and the more steeply curved surface should be oriented toward the support when possible.
In practice, this means the method is sensitive to how the lens is held and measured. The report also notes that power-variation lenses can be difficult because the measured area may not represent the whole lens.
New methods include low-coherence interferometry
Clause 7 describes low-coherence interferometry, which measures optical path differences and can derive the group refractive index [GRI]. The report then shows how the phase refractive index [PRI] can be calculated from GRI, with Annex C explaining the relationship in more detail.
This is important because some modern devices can measure finished spectacle lenses quickly and without cutting them. The report also notes that some instruments can measure multiple wavelengths and estimate the Abbe number.
Prism coupled refractometry can measure finished lenses without sample preparation
Clause 8 describes a prism-coupled refractometer based on the critical angle at the interface between a reference prism and the lens. No sample preparation is needed, and the prism rests on the convex surface of the lens.
The practical caution is that coated lenses may be measured as a coating-plus-substrate system, so the result may reflect the effective index of the coating rather than the substrate material alone.
What terms does ISO/TR 22572:2026 define?
- V-block prism - A prism with a 90° channel made from optical glass with a known refractive index.
- refractive index matching liquid - A transparent liquid whose refractive index is close to that of the specimen.
- contacting liquid - A transparent liquid whose refractive index is higher than that of the specimen.
- group refractive index (GRI) - The refractive index used for pulse or wave-packet propagation in the low-coherence interferometry section.
- phase refractive index (PRI) - The refractive index at a specified wavelength, derived from GRI in Annex C.
- vertex power - The lens power used in the non-destructive methods to relate measured power changes to refractive index.
- Abbe refractometer - An instrument that measures refractive index by the critical-angle principle with a prism and a contacting liquid.
Who uses ISO/TR 22572:2026?
ISO/TR 22572:2026 is relevant to spectacle lens manufacturers, test laboratories, optical instrument users, and trading standards authorities. The introduction says it was prepared to help identify methods that can determine the refractive index of finished spectacle lenses and to support verification of manufacturers’ claims.
It is also useful for quality control staff, buyers, and optical technicians who need to choose between destructive and non-destructive checking. Because the report compares methods with different accuracies and sample-preparation demands, it helps users decide whether they need a laboratory method, a production check, or a finished-lens verification method.
Which standards are used with ISO/TR 22572:2026?
- ISO 9802 - Provides vocabulary for raw optical glass, supporting the terminology used in the report.
- ISO 13666 - Provides vocabulary for spectacle lenses, which helps keep the optical terms consistent.
- ISO 80000-1 - Supports the general use of quantities and units.
- ISO 80000-3 - Supports the use of space and time quantities and units.
- ISO 21395-1 - Gives detailed instructions for the minimum deviation method used in clause 5.1.
- ISO 21395-2 - Gives detailed instructions for the V-block refractometer method used in clause 5.2.
- ISO 489 - Is cited for the critical-angle refractometer principle and for handling contacting liquids.
What does the ISO/TR 22572:2026 document contain?
The document contains schematic figures for the minimum deviation method, the V-block refractometer, the Abbe refractometer, immersion methods, low-coherence interferometry, and the prism coupled refractometer. It also contains equations for sag, surface power, immersion measurement, interferometry, and critical-angle calculation.
Annex A compares destructive methods, Annex B reports experimental uncertainty on spectacle lens samples, Annex C explains how GRI and PRI are related, Annex D discusses surface power errors, and Annex E lists refractive index matching liquids. The report also includes worked experimental results and practical notes on sample preparation, measurement orientation, coating effects, and liquid selection.
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ISO/TR 22572:2026 is a technical report published by the International Organization for Standardization (ISO). Its full title is "Ophthalmic optics — Spectacle lenses — Measurement of refractive index". This standard covers: This document describes the possible methods of determining the refractive index of spectacle lenses, whether in uncut, cut or final mounted form. It lists several destructive and non-destructive test methods having different accuracies.
This document describes the possible methods of determining the refractive index of spectacle lenses, whether in uncut, cut or final mounted form. It lists several destructive and non-destructive test methods having different accuracies.
ISO/TR 22572:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.70 - Ophthalmic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO/TR 22572:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 V-block prism
- 3.2 refractive index matching liquid
- 3.3 contacting liquid
- 4 Laboratory temperature
- 5 Traditional destructive methods
- 5.1 Minimum deviation method
- 5.2 V-block method
- 1 V-block
- 5.3 Abbe refractometer
- 5.4 Practical results
- 6.1 Sagittal depth gauge and radius of curvature methods
- 6.2 Immersion methods
- 7.1 Low-coherence interferometry
- 7.2 Measuring apparatus
- 8 Prism coupled refractometer
- 8.1 Theory
- Annex A
- Annex B
- Annex C
- Annex D
- Annex E
- Bibliography
Introduction
This document results from work in ISO/TC 172/SC 7/WG 3 to identify methods that could determine the refractive index of spectacle lenses with the aim of being able to differentiate between the refractive index of finished spectacle lenses. This would enable trading standards authorities to verify manufacturers' claims on their products.
Three methods of measurement of the refractive index of optical materials have been standardized, the minimum deviation method in ISO 21395-1, the V-block refractometer method in ISO 21395-2, as well as the Abbe refractometer method. These methods use specialised instruments and require either a prism to be cut out of the plane of the lens and the cut edges polished or a zone on the front surface to be surfaced flat and polished. Hence, although they can be used to measure the properties of stock spectacle lenses, they cannot be applied to a wearer’s spectacle lenses since they are ruined. For optimum results, two of the methods require specialised contacting liquids of similar index to that of the lens under test. Care is needed with the contacting liquid if it presents an environmental and health hazard during handling, storage, and disposal – see the warning in ISO 489.
In general, there is no restriction on the value of the refractive index of finished spectacle lenses, but it can be necessary to determine the refractive index of finished spectacle lenses and whether the index is what the manufacturer claims it to be. Both the industry and spectacle wearers prefer a non-destructive method to be used to check the lenses, since these methods do not harm the products. For example, the methods that combine vertex power measurements and either the sagittal depth gauge and radius of curvature or the immersion methods have no specific instruments for determining the refractive index. In addition, there are some recent devices in the market that provide refractive index measurement for finished spectacle lenses, such as critical angle and interferometric based measurements. This document is intended to aid in measuring the refractive index of spectacle lenses and improve the communications between spectacle manufacturers and customers. For the purposes of spectacle lenses, an accuracy of 0,01 is expected as the minimum requirement.
Technical Report ISO/TR 22572:2026(en) Ophthalmic optics — Spectacle lenses — Measurement of refractive index
1 Scope
This document describes the possible methods of determining the refractive index of spectacle lenses, whether in uncut, cut or final mounted form.
It lists several destructive and non-destructive test methods having different accuracies.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 9802, Raw optical glass — Vocabulary ISO 13666, Ophthalmic optics — Spectacle lenses — Vocabulary ISO 80000-1, Quantities and units — Part 1: General ISO 80000-3, Quantities and units — Part 3: Space and time
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 9802, ISO 13666, ISO 80000-1, ISO 80000-3 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 V-block prism
prism with a 90° channel manufactured from optical glass with a known refractive index [SOURCE: ISO 21395-2:2022, 3.1, modified — The words “with a 90° channel” were added.]
3.2 refractive index matching liquid
transparent liquid having the refractive index close to that of the specimen [SOURCE: ISO 21395-2:2022, 3.2]
3.3 contacting liquid
transparent liquid having the refractive index higher than the specimen
ISO/TR 22572:2026(en)
4 Laboratory temperature
For the precision required for this purpose, measurement can be at ambient laboratory temperature, which is usually either (23 ± 5) °C or (27 ± 5) °C in accordance with national practice. NOTE The temperature of 23 °C is normally the standard laboratory temperature in temperate areas while 27 °C is the standard laboratory temperature in tropical and subtropical areas.
5 Traditional destructive methods
5.1 Minimum deviation method
This method uses a goniometer or spectrometer and requires a prism with a (55 ± 10)° apical angle to be cut out of the plane of the lens, and the surfaces of the two cut edges polished – see Figure 1. For a comparison of the different destructive methods see Annex A.
Key 1 laser or light source, slit and collimator 2 turntable, with angular scale 3 prism cut from spectacle lens 4 detector or viewing telescope Figure 1 — Schematic diagram of the minimum deviation method The apical angle of the prism is measured by reflecting light from the collimator part of the instrument from the two adjacent surfaces and viewing the reflected beam in the telescope. The relative positions of the collimator, telescope and orientation of the prism are then adjusted to allow light from the collimator to pass through the prism and be viewed in the telescope. The angle of incidence on the first face of the prism is then adjusted to find the minimum angle of deviation. At this angle, the ray path through the prism is symmetrical, so that the final angle of refraction equals the angle of incidence. The refractive index is calculated using Snell’s law and the geometry of the apical angle and angle of incidence. Detailed instructions for applying this method are specified in ISO 21395-1.
5.2 V-block method
This method uses a V-block refractometer and requires a 90° prism to be cut out of the plane of the lens, and the two cut edges polished. The prism is then mounted in the V-block, which is effectively a parallel sided block of glass with a V-shaped trough into which the prism is placed – see Figure 2. Although shown in Figure 2, the refractive index matching liquid is not required if the cut edges are precisely prepared and well polished. The liquid, which has a similar refractive index to that of the lens under test, reduces the scattered light when entering and leaving the lens prism if the surfaces are not well polished, thus improving precision. The liquid can be chosen initially on the basis of the claimed refractive index and by making a first
ISO/TR 22572:2026(en) measurement and then, if necessary, using that result to make an improved choice of liquid and measuring a second time. Details of some refractive index matching liquids are given in Annex E. The prism samples used in the experiments reported in Annex B were prepared using a spectacle lens edging machine. The final angle of refraction/deviation through the prism assembly enables the refractive index of the lens to be calculated. Detailed instructions for applying this method are specified in ISO 21395-2. Key
1 V-block
2 prism cut from spectacle lens 3 refractive index matching liquid Figure 2 — Schematic diagram of the V-block refractometer The introduction to ISO 21395-2 states:
“The minimum deviation method is most accurate in refractive index measurement but requires an advanced technical skill to prepare a specimen with a precise shape and to measure the refractive index.” “In contrast the V-block refractometer method is easier and faster when preparing a specimen and requires less technical skill for measurement.” The standard describes the potential trade-offs between the tolerance on the 90° apical angle of the prism and the refractive index difference between the refractive index matching liquid and the (lens) prism. The preferred tolerance for the apical angle of the prism is ±2’ arc.
5.3 Abbe refractometer
This method uses an Abbe refractometer and requires a slab of material to be cut out of the plane of the lens with dimensions suitable to fit on the fixed prism of the instrument. This is then surfaced flat. A contacting liquid of similar (i.e. a refractive index matching liquid) or higher refractive index than that of the lens material is used to couple the sample to the refractometer. The method is described in ISO 489 but basically relies on the measurement of the critical angle, that is with, preferably monochromatic, light from a diffuse source entering the sample and passing at grazing incidence into the instrument's prism – see Figure 3. The ray shown identifies the direction of these rays, while rays that had
ISO/TR 22572:2026(en) a smaller angle of incidence in the sample all lie above this direction. A telescope is usually used to enhance the observation.
Key 1 high refractive index prism 2 block with plane surface cut from spectacle lens 3 contacting liquid 4 telescope Figure 3 — Schematic diagram of the Abbe refractometer Instruments are generally calibrated directly to give the refractive index. NOTE Optically, an Abbe refractometer can only measure samples with refractive index lower than that of its prism. This often gives an upper limit for measurement of the refractive index of 1,700. Similarly, the contacting liquid has to have a refractive index at least that of the sample.
5.4 Practical results
Because of the difficulties in sample preparation, the results are not as precise as those summarised in Table A.1. For the purposes of spectacle lenses, a minimum accuracy of 0,01 is required. The results from experiments on spectacle lens samples are presented in Annex B. 1) If the instrument is not directly calibrated for refractive index, it can be calculated from the formula:
nnsin()sin () cos()s in() where n is the refractive index of the prism and α and δ are defined in Figure 3. p p
ISO/TR 22572:2026(en) 6 “Classical” non-destructive methods
6.1 Sagittal depth gauge and radius of curvature methods
6.1.1 Accurate theory and considerations
The radii of curvature of the surfaces of a lens can be measured directly using reflection instruments or by using a sagittal depth gauge to measure the depth (sag) across a known chord. The radii of curvature, in metres, can then be calculated from the sag formula, Formula 1, where s and c are in millimetres:
sc
r or sr1000 ()1000 rc (1) 2000s where r radius of curvature of a surface of a lens, in metres; s is the sagittal depth, in millimetres; c is the semi-chord, i.e. half the distance between the two outer pins for a 3-pin instrument, or the radius of the annulus for a ring-based instrument, in millimetres. If the radii of curvature of the surfaces of a lens are known, application of the surface power equations allows a calculation of the refractive index. The theory is as follows:
Formula 2 gives the back vertex power of a lens:
F
F' F (2) v 2 t 1F
n where F’ is the back vertex power of a lens;
F and F are the front and back surface powers, in D (m ); 1 2 t is the centre thickness, in metres; n is the refractive index of the material of the lens.
The surface powers can be expressed in terms of their radii of curvature in metres, r and r respectively, 1 2 using Formula 3:
n 1 1n F and F (3) 1 2 r r 1 2 Formula 2 can then be rewritten as Formula 4:
n1 r 1n
F (4)
t n1 r
1 n r
Formula 4 can be simplified and solved as a quadratic equation for n in terms of r , r , t and the back vertex 1 2
power. In the usual quadratic equation, an + bn + c = 0, the coefficients and solution are given in Formulae 5 and 6 respectively:
11 tF '
a
rr r 12 1 tF ' 11 2t
b F' (5)
rr rr r 12 12 1
ISO/TR 22572:2026(en) t 1 c F'
rr 12 and
bb 4ac n (6) 2a where the positive root is taken for positive powered lenses, the negative for negative powered lenses.
The larger the chord, 2c, the larger the sag, s, (approximately proportional to c ). Hence, an error in the measurement of the sag will cause a smaller error or uncertainty in the calculated radius for larger
chords than shorter ones (again, approximately proportional to c ). For example, for a surface of power +6,00 D on a lens with a refractive index of 1,500, an error of 0,001 in the index is generated by an error or uncertainty of 0,001 2 mm (1,2 μm) for a chord of 20 mm diameter as opposed to an error or uncertainty of 0,005 mm (5,0 μm) for a chord of 40 mm diameter.
Conversely, if the surface is aspherical by design or inadvertently from the surfacing process, the larger the chord, the greater the probable deviation from an assumed spherical surface. For the same surface as above, it requires a p-value of 0,448 for the 20 mm chord as opposed to 0,869 for the 40 mm chord to give an error of +0,001 in the index, a smaller amount of elliptical flattening from the assumed spherical shape, for which p = 1,000. Negative errors of 0,001 in the index are produced by p-values of 1,560 and 1,137 respectively, which correspond to steepening ellipses.
Instruments working by measuring the reflected beam provide the curve along the 2 principal meridians; mapping devices often use an aperture in the region of 8 mm diameter at every point while single point measurement devices often use an aperture of 4 mm diameter, similar to that used by focimeters (lensmeters). Depth gauges can be supplied in various diameters. The British firm, OTI (Optical Tools for Industry, https:// www .opticaltools .co .uk states that a digital clock gauge to measure the sag can be accurate to 10 μm, while a Heidenhain linear gauge can be accurate to 0,1 μm. Obviously, a ring-based depth gauge cannot be used on a toroidal surface, and only a measuring system that uses a small area could be used on a power-variation lens since the surface power will be relatively uniform in only the primary reference point’s measurement area.
6.1.2 Approximate or simplified theory and considerations
By expanding the fraction and ignoring terms higher than F , Formula 2 can be simplified to become
Formula 7:
t
FF FF (7) v 11 2 n
If the term in F in Formula 7 is also ignored, i.e. the lens is regarded as a “thin lens”, and the powers
expressed in terms of the refractive index of the lens and the surface radii, Formula 7 can be simplified to Formula 8:
Fn()1 (8)
rr 12 where F the vertex power, as it now no longer matters which orientation the lens is in – but see below;
r and r are the radii of the two surfaces and are regarded as being positive if convex, negative if concave; 1 2 n is the refractive index of the material of the lens.
ISO/TR 22572:2026(en) Rearranging Formula 8 to determine n gives Formula 9:
F
n1 (9)
rr
Some simple instruments, e.g. the “lens measure”, are frequently calibrated in surface power for a specific index, often 1,530, rather than radius of curvature. In this case, the actual refractive index can be calculated from Formula 10:
F
n11()n (10)
FF 12 where F and F are the indicated front and back surface powers; 1 2 n is the refractive index used for the calibration of the instrument.
tF⋅
The omission of the term in Formula 8 onwards can lead to significant errors, particularly for positive n powered lenses where both the thickness and front (convex) surface power are greater than for negative [7] powered lenses. Burek recommended that for these positive-powered lenses, the calculation in Formulae 8 or 9 will be more accurate if based on the front vertex power, i.e. with the lens reversed from its usual orientation in the focimeter(lensmeter). As an example, consider a lens of BVP +5,00 DS with a front surface power of +8,00 D, back surface power of -3,21 D, centre thickness of 5 mm and index 1,600. Based on the back vertex power, the calculated index according to Formula 9 is 1,626 whereas based on the calculated front vertex power, the index is 1,604.
Annex D describes the effects of some errors in sag measurement.
6.2 Immersion methods
6.2.1 Theory
For a lens that is immersed in a medium of refractive index, n , Formula 8 needs rewriting as Formula 11:
M
Fn()n (11)
v M rr 12 This shows that the power of a lens depends on both its refractive index and that of the surrounding medium. By immersing the lens in a fluid of known refractive index, a new vertex power can be determined. From Formulae 8 and 11, the values of the VP in air and in the medium can be shown to give the refractive index of the lens as shown in Formula 12:
nF F Mivvn airin medium n (12) FF vvin air in medium where F is the vertex power measured in air v in air F is the vertex power measured in the medium with the lens in the same orientation as when v in medium measured in air t
This method relies on approximating the real lens as a thin lens by omitting the term ⋅F in Formula 7. As
n described in 6.1.2, the value of this is greater for a positive powered than a negative-powered lens, because of both the greater centre thickness and the higher convex/front surface power. The effect of this can be minimised to give better estimates for the refractive index if the lens is turned over, and the front vertex 2) Using the sign convention adopted for Formula 8.
ISO/TR 22572:2026(en) powers in air and with the immersing device are used, i.e. always do these measurements with the more steeply curved surface oriented towards the focimeter's or lensmeter’s lens support. With astigmatic-power lenses, care is taken to use the power in the same meridian both in air and in the medium. Using the principal meridian of larger absolute power will probably give better results. Measurement of power-variation lenses could lead to uncertain results since the powers in the two situations are measured in an area of stable power – perhaps well above the fitting cross for progressive-power lenses or in the near portion for degressive-power lenses.
...