IEC 60601-2-63:2012
(Main)Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment
Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment
IEC 60601-2-63:2012 applies to the basic safety and essential performance of dental extra-oral X-ray equipment, hereafter also called ME equipment. The scope includes ME systems containing such ME equipment. This particular standard has been prepared to provide, based on IEC 60601-1:2005 and its collaterals, a complete set of basic safety and essential performance requirements for dental extra-oral x-ray equipment. While the previously existing standards for such equipment were dedicated to components and subsystems, this particular standard addresses the system level of dental extra-oral x-ray equipment. Components and their functions are addressed as far as necessary. The scope of this standard is restricted to X-ray equipment where:
- the x-ray tube assembly contains the high-voltage transformer assembly; and
- the geometrical relations between the X-ray source, the anatomical object being imaged in the patient, and the X-ray image receptor, are preset in the design and cannot be arbitrarily altered by the operator during intended use.
Appareils électromédicaux - Partie 2-63: Exigences particulières pour la sécurité de base et les performances essentielles des appareils à rayonnement X dentaires extra-oraux
La CEI 60601-2-63:2012 s'applique à la sécurité de base et aux performances essentielles des appareils à rayonnement X dentaires extra-oraux, également désignés ci-après par appareils EM. Le domaine d'application inclut les systèmes EM contenant de tels appareils EM. La présente norme particulière a été préparée pour fournir, sur la base de la CEI 60601-1:2005 et de ses normes collatérales, un ensemble complet d'exigences de sécurité de base et de performances essentielles pour les appareils à rayonnement x dentaires extra-oraux. Alors que les normes antérieures pour de tels appareils étaient consacrées aux composants et aux sous-systèmes, la présente norme particulière concerne le niveau système des appareils à rayonnement x dentaires extra-oraux. Les composants et leurs fonctions sont mentionnés dans la mesure où cela est nécessaire. Le domaine d'application de la présente norme est limité aux appareils à rayonnement X dans lesquels:
- la gaine équipée contient l'ensemble transformateur haute tension; et
- les relations géométriques entre la source de rayonnement X, l'objet anatomique soumis à l'imagerie chez le patient et le récepteur radiologique sont préréglées au niveau de la conception et ne peuvent pas être modifiées de façon arbitraire par l'opérateur pendant l'utilisation prévue.
General Information
Relations
Standards Content (Sample)
IEC 60601-2-63 ®
Edition 1.2 2021-05
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.
IEC publications search - webstore.iec.ch/advsearchform IEC online collection - oc.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews. With a subscription you will always
committee, …). It also gives information on projects, replaced have access to up to date content tailored to your needs.
and withdrawn publications.
Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
The world's leading online dictionary on electrotechnology,
Stay up to date on all new IEC publications. Just Published
containing more than 22 000 terminological entries in English
details all new publications released. Available online and
and French, with equivalent terms in 18 additional languages.
once a month by email.
Also known as the International Electrotechnical Vocabulary
(IEV) online.
IEC Customer Service Centre - webstore.iec.ch/csc
If you wish to give us your feedback on this publication or
need further assistance, please contact the Customer Service
Centre: sales@iec.ch.
IEC 60601-2-63 ®
Edition 1.2 2021-05
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential
performance of dental extra-oral X-ray equipment
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 11.040.50 ISBN 978-2-8322-9806-0
IEC 60601-2-63 ®
Edition 1.2 2021-05
CONSOLIDATED VERSION
REDLINE VERSION
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
– 2 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
INTRODUCTION TO AMENDMENT 1 . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terminology and definitions . 11
201.4 General requirements . 13
201.5 General requirements for testing of ME EQUIPMENT . 14
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 14
201.7 ME EQUIPMENT identification, marking and documents . 14
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 17
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 18
201.10 Protection against unwanted and excessive radiation HAZARDS . 19
201.11 Protection against excessive temperatures and other HAZARDS . 19
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 19
201.13 HAZARDOUS SITUATIONS and fault conditions . 19
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 19
201.15 Construction of ME EQUIPMENT . 19
201.16 ME SYSTEMS . 19
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 19
202 Electromagnetic compatibility – Requirements and tests . 20
203 Radiation protection in diagnostic X-ray equipment . 20
Annexes . 33
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS . 34
Annex AA (informative) Particular guidance and rationale . 35
Bibliography . 42
Index of defined terms used in this particular standard . 45
Figure 203.101 – Zone of EXTRA-FOCAL RADIATION . 29
Figure AA.1 – PANORAMIC X-RAY EQUIPMENT . 35
Figure AA.2 – AIR KERMA during IRRADIATION with direct current X-RAY GENERATOR . 37
Figure AA.3 – AIR KERMA during IRRADIATION with ONE-PEAK X-RAY GENERATOR . 38
Figure AA.4 – Example – series of (numerous) pulsed IRRADIATIONS for a CBCT (cone
beam computed tomography) IRRADIATION event, with CONSTANT POTENTIAL HIGH-
VOLTAGE GENERATOR and time-width modulation . 39
Figure AA.5 – Example – series of two irradiations for PANORAMIC-like views of right
and left TMJ (temporo-mandibolar joint) in the same image, with ONE-PEAK HIGH-
VOLTAGE GENERATOR . 40
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
MANUFACTURER in the RISK MANAGEMENT PROCESS . 13
+AMD2:2021 CSV © IEC 2021
Table 201.C.101 – Marking on the outside of ME EQUIPMENT or its parts . 34
Table 201.C.102 – Subclauses requiring statements in ACCOMPANYING DOCUMENTS . 34
– 4 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This consolidated version of the official IEC Standard and its amendments has been
prepared for user convenience.
IEC 60601-2-63 edition 1.2 contains the first edition (2012-09) [documents 62B/888/FDIS
and 62B/898/RVD], its amendment 1 (2017-07) [documents 62B/1049/FDIS and
62B/1058/RVD] and its amendment 2 (2021-05) [documents 62B/1232/FDIS and
62B/1237/RVD].
In this Redline version, a vertical line in the margin shows where the technical content
is modified by amendments 1 and 2. Additions are in green text, deletions are in
strikethrough red text. A separate Final version with all changes accepted is available
in this publication.
+AMD2:2021 CSV © IEC 2021
International Standard IEC 60601-2-63 has been prepared by IEC subcommittee 62B:
Diagnostic imaging equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: in roman type.
– Test specifications: in italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
The committee has decided that the contents of the base publication and its amendments will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
+AMD2:2021 CSV © IEC 2021
INTRODUCTION
This particular standard has been prepared to provide, based on IEC 60601-1:2005 (third
edition), and its collaterals, a complete set of BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for DENTAL EXTRA-ORAL X-RAY EQUIPMENT. While the previously existing standards
for such equipment were dedicated to components and subsystems, this particular standard
addresses the system level of DENTAL EXTRA-ORAL X-RAY EQUIPMENT. Components and their
functions are addressed as far as necessary.
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of DENTAL EXTRA-ORAL
X-RAY EQUIPMENT
The minimum safety requirements for DENTAL INTRA-ORAL X-RAY EQUIPMENT are specified in a
separate particular standard IEC 60601-2-65 to simplify and improve the readability
Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions
of the collateral standard IEC 60601-1-3, the particular standards IEC 60601-2-28
IEC 60601-2-7, or IEC 60601-2-32 have been extracted and moved into this particular
standard.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard.
INTRODUCTION TO AMENDMENT 1
The purpose of this first amendment to IEC 60601-2-63:2012 is to introduce changes to
reference the Amendment 1 (2012) to IEC 60601-1:2005. As neither IEC 60601-2-63:2012 nor
this amendment refers to specific elements of IEC 60601-1-2, the introduction of a dated
reference to the latter document has been removed.
– 8 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT, hereafter also called ME EQUIPMENT. The scope includes
ME SYSTEMS containing such ME EQUIPMENT.
NOTE 1 This includes PANORAMIC equipment, CEPHALOMETRIC equipment, and equipment for dental volumetric
reconstruction (hereafter DVR) as defined in 201.3.203 below. An example of such equipment is an equipment
designed to perform PANORAMIC, CEPHALOMETRIC and DENTAL VOLUMETRIC RECONSTRUCTION (hereafter DVR) as
defined in 201.3.203.
NOTE 2 DVR includes dental CBCT (cone beam computed tomography), which is also known with other names in
certain parts of the world, e.g. DVT (digital volumetric tomography); DVR also includes tomosynthesis.
NOTE 3 This may include the imaging of other anatomical parts (e.g. the hand) as long as required for dental
treatment (e.g. orthodontic treatment).
NOTE 4 This may include anatomical objects of interest to the ENT (ear, nose, and throat) specialist.
The scope of this standard is restricted to X-RAY EQUIPMENT where:
• the X-RAY TUBE ASSEMBLY contains the HIGH-VOLTAGE TRANSFORMER ASSEMBLY and
• the geometrical relations between the X-RAY SOURCE, the anatomical object being imaged
in the PATIENT, and the X-RAY IMAGE RECEPTOR, are preset in the design and cannot be
OPERATOR during INTENDED USE.
arbitrarily altered by the
DENTAL EXTRA-ORAL X-RAY EQUIPMENT are X-RAY EQUIPMENT designed for EXTRA-ORAL
in which the geometrical relations between the X-RAY SOURCE, the anatomical
RADIOGRAPHY
object being imaged in the PATIENT, and the X-RAY IMAGE RECEPTOR, are preset in the design
and cannot be arbitrarily altered by the OPERATOR during INTENDED USE. In such equipment, the
X-RAY TUBE ASSEMBLY contains the HIGH-VOLTAGE TRANSFORMER ASSEMBLY.
NOTE 5 DENTAL INTRA-ORAL X-RAY EQUIPMENT is excluded from the scope of this standard.
NOTE 6 FOCAL SPOT TO IMAGE RECEPTOR DISTANCE and FOCAL SPOT to object distance are preset in the design of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
NOTE 7 For DENTAL X-RAY EQUIPMENT not in the scope of this document because of the restriction above,
applicable clauses of IEC 60601-2-54 may be used with this document.
ME EQUIPMENT and ME SYSTEMS in the scope of IEC 60601-2-44, IEC 60601-2-54,
IEC 60601-2-45, IEC 60601-2-65 or IEC 60601-2-43 are excluded from the scope of this
particular standard. The scope of this International Standard also excludes RADIOTHERAPY
SIMULATORS and equipment for bone or tissue absorption densitometry. Excluded from the
ME EQUIPMENT intended to be used for DENTAL RADIOSCOPY.
scope is also
—————————
1)
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance
+AMD2:2021 CSV © IEC 2021
Within its specific scope, the clauses of this particular standard supersede and replace those
of IEC 60601-2-7, Medical electrical equipment – Particular requirements for the safety of
high-voltage generators of diagnostic X-ray generators and of IEC 60601-2-32, Medical
electrical equipment – Particular requirements for the safety of associated equipment of X-ray
equipment.
NOTE 8 Requirements for X-RAY GENERATORS and for ASSOCIATED EQUIPMENT, which were previously specified in
IEC 60601-2-7 and IEC 60601-2-32, have been included in either IEC 60601-1:2005 (Ed3) or this particular
rd
standard. Therefore IEC 60601-2-7 and IEC 60601-2-32 are not part of the IEC 60601-1 3 edition scheme for
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard. Therefore IEC 60601-2-28 does not apply to ME EQUIPMENT in the scope of this
International Standard with the exception of X-RAY TUBE ASSEMBLIES that are replaceable in
the field by entities other than the manufacturer.
NOTE 9 Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions of the
collateral standard IEC 60601-1-3 or the particular standard IEC 60601-2-28 have been extracted and moved into
this particular standard.
NOTE 10 For X-RAY EQUIPMENT in the scope of this particular standard X-RAY TUBE ASSEMBLIES are X-RAY
MONOBLOCK ASSEMBLIES.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for ME EQUIPMENT for EXTRA-ORAL DENTAL RADIOGRAPHY.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-3 apply as modified in Clause 202 and 203 respectively.
2) ) )
3 4
IEC 60601-1-8, IEC 60601-1-10 , IEC 60601-1-11 and IEC 60601-1-12 do not apply. All
other published collateral standards in the IEC 60601-1 series apply as published.
NOTE OPERATORS of DENTAL EXTRA-ORAL X-RAY EQUIPMENT are used to audible signals as required in this
particular standard rather than to the concepts of IEC 60601-1-8. Therefore IEC 60601-1-8 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard or collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
—————————
)
IEC 60601-1-10, Medical electrical equipment – Part 1-10: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for the development of physiologic closed-loop controllers
)
IEC 60601-1-11, Medical electrical equipment – Part 1-11: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment
4)
IEC 60601-1-12, Medical electrical equipment – Part 1-12: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems intended for use in the emergency medical services environment
– 10 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 42.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic compatibility –
Requirements and tests
IEC 60601-1-3:2008, Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic X-
+AMD2:2021 CSV © IEC 2021
ray equipment
IEC 60601-1-3:2008/AMD1:2013
Addition:
IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis –
Characteristics of focal spots
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-2-29:2008, Medical electrical equipment – Part 2-29: Particular requirements for
the basic safety and essential performance of radiotherapy simulators
IEC 60601-2-54:2009, Medical electrical equipment – Part 2-54: Particular requirements for
the basic safety and essential performance of X-ray equipment for radiography and
radioscopy
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC/PAS 61910-1:20072014, Medical electrical equipment – Radiation dose documentation –
Part 1: Equipment Radiation dose structured reports for radiography and radioscopy
201.3 Terminology and definitions
Amendment:
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1:2005/AMD1:2012, its applicable collateral standards, IEC/TR 60788:2004 and
the following apply:
NOTE An index of defined terms is found beginning on page 45.
Addition:
201.3.201
CEPHALOMETRIC
related to PROJECTION RADIOGRAPHY of the whole dento-maxillo-facial anatomy, whereas the
projection geometry is such to minimize geometrical image distortions
Note 1 to entry: This is usually achieved by setting a sufficiently large source-to-object-distance and source-to-
detector-distance.
Note 2 to entry: Another term often used for CEPHALOMETRIC RADIOGRAPHY is teleradiography.
201.3.202
DENTAL
related to structures in the dento-maxillo-facial district of the PATIENT, including dentition
201.3.203
*DENTAL VOLUMETRIC RECONSTRUCTION
DVR
reconstruction of the 3-dimensional attenuation distribution of the whole or part of the
irradiated volume from a series of 2-dimensional projections produced by an X-RAY BEAM on
an X-RAY IMAGE RECEPTOR moving around the head of the PATIENT
– 12 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
201.3.204
DOSE AREA PRODUCT
product of the area of the cross-section of an X-RAY BEAM and the averaged AIR KERMA over
that cross-section. The unit is the gray square metre (Gy⋅m ).
[SOURCE: IEC 60601-2-54:2009, 201.3.203]
201.3.205
ELECTRONIC X-RAY IMAGE RECEPTOR
X-RAY IMAGE RECEPTOR comprising an electrically-powered conversion method
201.3.206
EXTRA-ORAL
related to DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located outside the oral
cavity
201.3.207
INTERLOCK
means preventing the start or the continued operation of ME EQUIPMENT unless certain
predetermined conditions prevail
[SOURCE:IEC 60601-2-54:2009, 201.3.207]
201.3.208
INTRA-ORAL
related to DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located, wholly or
partially, inside the oral cavity
201.3.209
NOMINAL SHORTEST IRRADIATION TIME
shortest LOADING TIME for which a required constancy of the controlled RADIATION QUANTIY is
maintained
Note 1 to entry: The IRRADIATION TIME is controlled by a HIGH-VOLTAGE GENERATOR with AUTOMATIC CONTROL
SYSTEMS.
[SOURCE: IEC 60601-2-54:2009, 201.3.208]
201.3.210
ONE-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers an unrectified
output voltage, or rectified output voltage with one peak during each cycle of the supply
[SOURCE: IEC 60601-2-65:2012, 201.3.208
201.3.211
* PANORAMIC
related to DENTAL RADIOGRAPHY, produced by the coordinated motion of a scanning fan-shaped
X-RAY BEAM, oriented parallel to the cranio-caudal axis of the PATIENT, and an X-RAY IMAGE
RECEPTOR, both rotating around the head of the PATIENT
Note 1 to entry: A tomographic layer is produced with respect to the plane perpendicular to the rotational axis.
The resulting image is a focused projection on a surface parallel to the rotational axis.
Note 2 to entry: The scanning axis is usually vertical.
201.3.212
TWO-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers a rectified
output voltage with two peaks during each cycle of the supply
+AMD2:2021 CSV © IEC 2021
[SOURCE: IEC 60601-2-65:2012, 201.3.209]
201.3.213
X-RAY MONOBLOCK ASSEMBLY
X-RAY TUBE ASSEMBLY containing the HIGH-VOLTAGE TRANSFORMER ASSEMBLY
Note 1 to entry: The term X-RAY MONOBLOCK ASSEMBLY excludes the BEAM LIMITING DEVICE.
201.3.214
DENTAL CONE BEAM COMPUTED TOMOGRAPHY
DENTAL CBCT
3-dimensional imaging of DENTAL anatomical structures, performed by reconstruction of a
volume from a series of 2-dimensional projections produced by circular or rectangular
collimated X-RAY BEAM on an X-RAY IMAGE RECEPTOR rotating around the head of the PATIENT
201.3.215
EXAMINATION PROTOCOL
full set of programmed LOADING FACTORS, control functions and settings, including image
processing settings, designed to the image acquisition and DISPLAY
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.3 ESSENTIAL PERFORMANCE
Addition:
201.4.3.101 Additional potential ESSENTIAL PERFORMANCE requirements
The list in Table 201.101 is a list of potential ESSENTIAL PERFORMANCE to be considered by the
MANUFACTURER in the RISK MANAGEMENT PROCESS.
NOTE Subclause 203.6.4.3.102 (Accuracy of LOADING FACTORS) specifies a limitation in applying subclause
203.6.4.3.102.3 (Accuracy of X-RAY TUBE VOLTAGE) and 203.6.4.3.102.4 (Accuracy of X-RAY TUBE CURRENT). This
limitation is also valid for the ESSENTIAL PERFORMANCE list.
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
MANUFACTURER in the RISK MANAGEMENT PROCESS
Requirement Subclause
Accuracy of LOADING FACTORS 203.6.4.3.102
Reproducibility of the RADIATION output 203.6.3.2
201.4.10.1 Source of power for ME EQUIPMENT
Addition:
201.4.10.1.101 Connection to SUPPLY MAINS
ME EQUIPMENT shall be PERMANENTLY INSTALLED unless the INTENDED USE requires it to be
MOBILE.
201.4.10.2 SUPPLY MAINS for ME EQUIPMENT and ME SYSTEMS
Addition:
– 14 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
The internal impedance of a SUPPLY MAINS is to be considered sufficiently low for the operation
of ME EQUIPMENT if the value of the APPARENT RESISTANCE OF SUPPLY MAINS does not exceed
the value specified in the ACCOMPANYING DOCUMENTS.
For this purpose, the APPARENT RESISTANCE of SUPPLY MAINs R is determined according to the
formula:
UU−
R =
I
where
U is the no-load MAINS VOLTAGE;
U is the MAINS VOLTAGE under load;
I is the mains current under load.
ME EQUIPMENT is considered to comply with the requirements of this standard only if its
specified NOMINAL ELECTRIC POWER can be demonstrated at a resistance of supply mains
having a value not less than the APPARENT RESISTANCE OF SUPPLY MAINS specified by the
MANUFACTURER in the ACCOMPANYING DOCUMENTS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and by functional test.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
201.7.2.7 Electrical input power from the SUPPLY MAINS
Addition:
Except for items a) to c) below, for ME EQUIPMENT that is specified to be PERMANENTLY
INSTALLED, the information may be stated in the ACCOMPANYING DOCUMENTS only.
The information on the input power shall be specified in terms of combinations of
a) the rated MAINS VOLTAGE of the ME EQUIPMENT in volts; see 7.2.1 and 7.2.6 of the general
standard;
b) the number of phases; see 7.2.1 and 7.2.6 of the general standard;
c) the frequency, in hertz; see 7.2.1 and 7.2.6 of the general standard;
d) the maximum permissible value for APPARENT RESISTANCE OF SUPPLY MAINS, in ohms;
e) the characteristics of OVER-CURRENT RELEASES required in the SUPPLY MAINS.
+AMD2:2021 CSV © IEC 2021
NOTE These requirements are adapted from IEC 60601-2-7 subclause 6.1j).
201.7.2.15 Cooling conditions
Addition:
If cooling is necessary for safe operation of ME EQUIPMENT, or a subassembly thereof, the
cooling requirements shall be indicated in the ACCOMPANYING DOCUMENTS, as identified in the
RISK MANAGEMENT PROCESS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and the RISK
MANAGEMENT FILE.
Additional subclause:
201.7.2.101 BEAM LIMITING DEVICE
Where detachable in NORMAL USE, BEAM LIMITING DEVICES shall be provided with the following
markings:
– those required in subclause 7.2.2 of the general standard;
– serial designation or individual identification;
– ADDITIONAL FILTRATION, if the additional value is more than the equivalent of 0,2 mm Al.
Compliance is checked by inspection.
201.7.8.1 Colours of indicator lights
Addition:
The indication of X-ray related states shall be excluded from subclause 7.8 in the general
standard. 203.6.4.2 and 203.6.4.101 shall apply instead.
201.7.9 ACCOMPANYING DOCUMENTs
201.7.9.1 General
Addition:
NOTE 101 Annex C Table 201.C.102 lists the requirements of this particular standard that are additional to those
of the general standard for statements in the ACCOMPANYING DOCUMENTS.
The ACCOMPANYING DOCUMENTS shall contain quality control PROCEDURES to be performed on
the ME EQUIPMENT by the RESPONSIBLE ORGANISATION. These shall include acceptance criteria
and frequency for the tests.
Additionally for ME EQUIPMENT provided with one or more integrated ELECTRONIC X-RAY IMAGE
RECEPTORs, the ACCOMPANYING DOCUMENTS shall contain a description of the minimum
performance required of the means, used to display the images for diagnostic purpose
according to the INTENDED USE for each ELECTRONIC X-RAY IMAGE RECEPTOR used.
If a test or a QUALITY CONTROL PROCEDURE recommended by the MANUFACTURER requires a
device-specific arrangement (including a TOOL, a PHANTOM, a special software or a software
setting); that is only available from the MANUFACTURER, the MANUFACTURER shall provide this
arrangement for the RESPONSIBLE ORGANIZATION.
NOTE 103 The intention is to perform these QUALITY CONTROL PROCEDURES and tests using only the
ACCOMPANYING DOCUMENTS.
NOTE 102 For instance, the minimum required number of pixels and number of discernible grey levels at the
DISPLAY.
– 16 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
If the test or PROCEDURE requires a device-specific TOOL that is only available from the
MANUFACTURER, the MANUFACTURER shall make this TOOL available to the RESPONSIBLE
ORGANIZATION.
NOTE 104 The MANUFACTURER can provide PHANTOM with the equipment, if specified in the local regulations.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.7.9.2 Instructions for use
201.7.9.2.1 General
Additional subclause:
201.7.9.2.1.101 LOADING FACTORS
In the instructions for use the LOADING FACTORS shall be stated as described below. The
following combinations and data shall be stated:
a) value(s) or range of X-RAY TUBE VOLTAGE settings;
b) value(s) or range of X-RAY TUBE CURRENT settings;
c) values or range of IRRADIATION TIME settings;
d) maximum X-RAY TUBE CURRENT at each X-RAY TUBE VOLTAGE setting, if different from (b);
e) maximum and minimum IRRADIATION TIME at each X-RAY TUBE VOLTAGE and X-RAY TUBE
CURRENT setting, if different from (c);
f) for ME EQUIPMENT indicating precalculated or measured CURRENT TIME PRODUCT, the lowest
CURRENT TIME PRODUCT or the combinations of LOADING FACTORS resulting in the lowest
CURRENT TIME PRODUCT;
g) if the value of the lowest CURRENT TIME PRODUCT depends upon the X-RAY TUBE VOLTAGE or
upon certain combinations of values of LOADING FACTORS, the lowest CURRENT TIME
PRODUCT may be given as a table or curve showing the dependence;
h) the NOMINAL SHORTEST IRRADIATION TIME used in AUTOMATIC EXPOSURE CONTROL systems of
ME EQUIPMENT;
i) if the NOMINAL SHORTEST IRRADIATION TIME depends upon LOADING FACTORS such as X-RAY
TUBE VOLTAGE and X-RAY TUBE CURRENT, the ranges of these LOADING FACTORS for which
the NOMINAL SHORTEST IRRADIATION TIME is valid;
j) the maximum possible range of the X-RAY TUBE VOLTAGE and/or the X-RAY TUBE CURRENT
during IRRADIATION, controlled with the AUTOMATIC CONTROL SYSTEM
NOTE these requirements are adapted from IEC 60601-2-7 subclause 6.8.2 a).
Compliance is checked by inspection of the instructions for use.
201.7.9.3 Technical description
Additional subclauses:
201.7.9.3.101 Technical description for DENTAL EXTRA-ORAL X-RAY EQUIPMENT
In addition to the data required to be marked according to subclause 7.2 of the general
standard, the technical description shall specify a description of the geometric relationship of
the FOCAL SPOT, X-RAY BEAM dimensions, PATIENT position and IMAGE RECEPTION AREA, as well
as a method used to estimate the DOSE AREA PRODUCT.
Additional subclause:
+AMD2:2021 CSV © IEC 2021
201.7.9.3.102 X-RAY SOURCE ASSEMBLY
The technical description of the integrated X-RAY SOURCE ASSEMBLY shall specify the following,
in addition to the data required to be marked according to subclause 7.2 of the general
standard:
a) specification of the REFERENCE AXIS to which the TARGET ANGLE(s) and the FOCAL SPOT
characteristics of the X-RAY SOURCE ASSEMBLY refer:
b) TARGET ANGLE(s) with respect to the specified REFERENCE AXIS;
c) NOMINAL FOCAL SPOT VALUE(s) determined according to IEC 60336 for the specified
REFERENCE AXIS.
Compliance is checked by inspection of the technical description.
Additional subclause:
201.7.9.101 Requirements to the SUPPLY MAINS
The information on the RATED electrical input power for DENTAL EXTRA-ORAL X-RAY GENERATORS
shall also include:
• either the maximum permissible value for the APPARENT RESISTANCE OF SUPPLY MAINS or
other appropriate SUPPLY MAINS specifications used in a facility, and
• the characteristics of OVER-CURRENT RELEASES eventually required in the SUPPLY MAINS.
201.8 Protection against electrical HAZARDS from ME EQUIPMENT
Clause 8 of the general standard applies, except as follows:
201.8.5 Separation of parts
201.8.5.1 MEANS OF PROTECTION (MOP)
Additional subclause:
201.8.5.1.101 Additional limitation of voltage, current or energy
Provisions shall be made to prevent the appearance of an unacceptably HIGH VOLTAGE in the
MAINS PART or in any other low-voltage circuit.
NOTE This may be achieved for example by
– provision of a winding layer or a conductive screen connected to the PROTECTIVE EARTH TERMINAL between
HIGH VOLTAGE and low-voltage circuits, or
– provision of a voltage limiting device across terminals to which external devices are connected and between
which an excessive voltage might arise if the external path becomes open-circuited.
Compliance is checked by inspection of design data and construction.
NOTE These requirements are adapted from IEC 60601-2-7:1998, subclause 15bb).
201.8.5.4 Working voltage
Addition:
201.8.5.4.101 Stator and stator circuit dielectric strength testing
The test voltage for the dielectric strength testing of stator and stator circuit used for the
operation of the rotating anode of the X-RAY TUBE is to be referred to the voltage existing after
reduction of the stator supply voltage to its steady state operating value.
– 18 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
NOTE These requirements are adapted from IEC 60601-2-7:1998, subclause 20.4 l).
201.8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS
201.8.7.3 * Allowable values
Item c) is amended as follows:
For non-PERMANENTLY INSTALLED X-RAY GENERATORS the allowable value of TOUCH CURRENT in
SINGLE FAULT CONDITION is 2 mA.
NOTE This relaxation from the requirement of the general standard does not apply to PATIENT LEAKAGE CURRENT.
Item e) is amended as follows:
For PERMANENTLY INSTALLED ME EQUIPMENT, the allowable value of EARTH LEAKAGE CURRENT is
20 mA in NORMAL CONDITION and SINGLE FAULT CONDITION.
201.8.8.3 * Dielectric strength
Addition:
Instead of subclause 8.8.3 of the ge
...
IEC 60601-2-63 ®
Edition 1.0 2012-09
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X dentaires extra-oraux
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les
microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
Useful links:
IEC publications search - www.iec.ch/searchpub Electropedia - www.electropedia.org
The advanced search enables you to find IEC publications The world's leading online dictionary of electronic and
by a variety of criteria (reference number, text, technical electrical terms containing more than 30 000 terms and
committee,…). definitions in English and French, with equivalent terms in
It also gives information on projects, replaced and additional languages. Also known as the International
withdrawn publications. Electrotechnical Vocabulary (IEV) on-line.
IEC Just Published - webstore.iec.ch/justpublished Customer Service Centre - webstore.iec.ch/csc
Stay up to date on all new IEC publications. Just Published If you wish to give us your feedback on this publication
details all new publications released. Available on-line and or need further assistance, please contact the
also once a month by email. Customer Service Centre: csc@iec.ch.
A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.
Liens utiles:
Recherche de publications CEI - www.iec.ch/searchpub Electropedia - www.electropedia.org
La recherche avancée vous permet de trouver des Le premier dictionnaire en ligne au monde de termes
publications CEI en utilisant différents critères (numéro de électroniques et électriques. Il contient plus de 30 000
référence, texte, comité d’études,…). termes et définitions en anglais et en français, ainsi que
Elle donne aussi des informations sur les projets et les les termes équivalents dans les langues additionnelles.
publications remplacées ou retirées. Egalement appelé Vocabulaire Electrotechnique
International (VEI) en ligne.
Just Published CEI - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications de la CEI.
Just Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur
Disponible en ligne et aussi une fois par mois par email. cette publication ou si vous avez des questions
contactez-nous: csc@iec.ch.
IEC 60601-2-63 ®
Edition 1.0 2012-09
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X dentaires extra-oraux
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
X
CODE PRIX
ICS 11.040.50 ISBN 978-2-83220-382-8
– 2 – 60601-2-63 © IEC:2012
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terms and definitions . 11
201.4 General requirements . 12
201.5 General requirements for testing of ME EQUIPMENT . 13
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 13
201.7 ME EQUIPMENT identification, marking and documents . 13
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 16
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 17
201.10 Protection against unwanted and excessive radiation HAZARDS . 18
201.11 Protection against excessive temperatures and other HAZARDS . 18
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 18
201.13 HAZARDOUS SITUATIONS and fault conditions . 18
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 18
201.15 Construction of ME EQUIPMENT . 18
201.16 ME SYSTEMS . 18
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 19
202 Electromagnetic compatibility – Requirements and tests . 19
203 Radiation protection in diagnostic X-ray equipment . 19
Annexes . 31
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS . 32
Annex AA (informative) Particular guidance and rationale . 33
Bibliography . 39
Index of defined terms used in this particular standard . 42
Figure 203.101 – Zone of EXTRA-FOCAL RADIATION . 28
Figure AA.1 – PANORAMIC X-RAY EQUIPMENT . 33
Figure AA.2 – AIR KERMA during IRRADIATION with direct current X-RAY GENERATOR . 35
Figure AA.3 – AIR KERMA during IRRADIATION with ONE-PEAK X-RAY GENERATOR . 36
Figure AA.4 – Example – series of (numerous) pulsed IRRADIATIONS for a CBCT (cone
beam computed tomography) IRRADIATION event, with CONSTANT POTENTIAL HIGH-
VOLTAGE GENERATOR and time-width modulation . 37
Figure AA.5 – Example – series of two irradiations for PANORAMIC-like views of right
and left TMJ (temporo-mandibolar joint) in the same image, with ONE-PEAK HIGH-
VOLTAGE GENERATOR . 37
60601-2-63 © IEC:2012 – 3 –
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
MANUFACTURER in the RISK MANAGEMENT PROCESS . 13
Table 201.C.101 – Marking on the outside of ME EQUIPMENT or its parts . 32
Table 201.C.102 – Subclauses requiring statements in ACCOMPANYING DOCUMENTS . 32
– 4 – 60601-2-63 © IEC:2012
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60601-2-63 has been prepared by IEC subcommittee 62B:
Diagnostic imaging equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
The text of this particular standard is based on the following documents:
FDIS Report on voting
62B/888/FDIS 62B/898/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
60601-2-63 © IEC:2012 – 5 –
– Requirements and definitions: in roman type.
– Test specifications: in italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – 60601-2-63 © IEC:2012
The committee has decided that the contents of this publication will remain unchange-d until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
60601-2-63 © IEC:2012 – 7 –
INTRODUCTION
This particular standard has been prepared to provide, based on IEC 60601-1:2005 (third
edition), and its collaterals, a complete set of BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for DENTAL EXTRA-ORAL X-RAY EQUIPMENT. While the previously existing standards
for such equipment were dedicated to components and subsystems, this particular standard
addresses the system level of DENTAL EXTRA-ORAL X-RAY EQUIPMENT. Components and their
functions are addressed as far as necessary.
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of DENTAL EXTRA-ORAL
X-RAY EQUIPMENT
The minimum safety requirements for DENTAL INTRA-ORAL X-RAY EQUIPMENT are specified in a
separate particular standard IEC 60601-2-65 to simplify and improve the readability
Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions
of the collateral standard IEC 60601-1-3, the particular standards IEC 60601-2-28
IEC 60601-2-7, or IEC 60601-2-32 have been extracted and moved into this particular
standard.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard.
– 8 – 60601-2-63 © IEC:2012
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT, hereafter also called ME EQUIPMENT. The scope includes
ME SYSTEMS containing such ME EQUIPMENT.
NOTE 1 This includes PANORAMIC equipment, CEPHALOMETRIC equipment, and equipment for dental volumetric
reconstruction (hereafter DVR) as defined in 201.3.203 below.
NOTE 2 DVR includes dental CBCT (cone beam computed tomography), which is also known with other names in
certain parts of the world, e.g. DVT (digital volumetric tomography); DVR also includes tomosynthesis.
NOTE 3 This may include the imaging of other anatomical parts (e.g. the hand) as long as required for dental
treatment (e.g. orthodontic treatment).
NOTE 4 This may include anatomical objects of interest to the ENT (ear, nose, and throat) specialist.
The scope of this standard is restricted to X-RAY EQUIPMENT where:
• the X-RAY TUBE ASSEMBLY contains the HIGH-VOLTAGE TRANSFORMER ASSEMBLY and
• the geometrical relations between the X-RAY SOURCE, the anatomical object being imaged
PATIENT, and the X-RAY IMAGE RECEPTOR, are preset in the design and cannot be
in the
arbitrarily altered by the OPERATOR during INTENDED USE.
NOTE 5 DENTAL INTRA-ORAL X-RAY EQUIPMENT is excluded from the scope of this standard.
NOTE 6 FOCAL SPOT TO IMAGE RECEPTOR DISTANCE and FOCAL SPOT to object distance are preset in the design of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
NOTE 7 For DENTAL X-RAY EQUIPMENT not in the scope of this document because of the restriction above,
applicable clauses of IEC 60601-2-54 may be used with this document.
ME EQUIPMENT and ME SYSTEMS in the scope of IEC 60601-2-44, IEC 60601-2-54,
IEC 60601-2-45, IEC 60601-2-65 or IEC 60601-2-43 are excluded from the scope of this
RADIOTHERAPY
particular standard. The scope of this International Standard also excludes
SIMULATORS and equipment for bone or tissue absorption densitometry. Excluded from the
scope is also ME EQUIPMENT intended to be used for DENTAL RADIOSCOPY.
Within its specific scope, the clauses of this particular standard supersede and replace those
of IEC 60601-2-7, Medical electrical equipment – Particular requirements for the safety of
high-voltage generators of diagnostic X-ray generators and of IEC 60601-2-32, Medical
electrical equipment – Particular requirements for the safety of associated equipment of X-ray
equipment.
—————————
1)
The general standard is IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance
60601-2-63 © IEC:2012 – 9 –
NOTE 8 Requirements for X-RAY GENERATORS and for ASSOCIATED EQUIPMENT, which were previously specified in
IEC 60601-2-7 and IEC 60601-2-32, have been included in either IEC 60601-1:2005 (Ed3) or this particular
rd
standard. Therefore IEC 60601-2-7 and IEC 60601-2-32 are not part of the IEC 60601-1 3 edition scheme for
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard. Therefore IEC 60601-2-28 does not apply to ME EQUIPMENT in the scope of this
International Standard with the exception of X-RAY TUBE ASSEMBLIES that are replaceable in
the field.
NOTE 9 Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions of the
collateral standard IEC 60601-1-3 or the particular standard IEC 60601-2-28 have been extracted and moved into
this particular standard.
NOTE 10 For X-RAY EQUIPMENT in the scope of this particular standard X-RAY TUBE ASSEMBLIES are X-RAY
MONOBLOCK ASSEMBLIES.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for ME EQUIPMENT for EXTRA-ORAL DENTAL RADIOGRAPHY.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-3 apply as modified in Clause 202 and 203 respectively.
2) )
IEC 60601-1-8, IEC 60601-1-10 and IEC 60601-1-11 do not apply. All other published
collateral standards in the IEC 60601-1 series apply as published.
NOTE OPERATORS of DENTAL EXTRA-ORAL X-RAY EQUIPMENT are used to audible signals as required in this
particular standard rather than to the concepts of IEC 60601-1-8. Therefore IEC 60601-1-8 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard or collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
—————————
)
IEC 60601-1-10, Medical electrical equipment – Part 1-10: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for the development of physiologic closed-loop controllers
)
Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance –
Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the
home healthcare environment
– 10 – 60601-2-63 © IEC:2012
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 39.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic compatibility –
Requirements and tests
IEC 60601-1-3:2008, Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic X-
ray equipment
Addition:
IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis –
Characteristics of focal spots
IEC 60601-2-29:2008, Medical electrical equipment – Part 2-29: Particular requirements for
the basic safety and essential performance of radiotherapy simulators
60601-2-63 © IEC:2012 – 11 –
IEC 60601-2-54:2009, Medical electrical equipment – Part 2-54: Particular requirements for
the basic safety and essential performance of X-ray equipment for radiography and
radioscopy
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC/PAS 61910-1:2007, Medical electrical equipment – Radiation dose documentation –
Part 1: Equipment for radiography and radioscopy
201.3 Terminology and definitions
Amendment:
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005, its
applicable collateral standards, IEC/TR 60788:2004 and the following apply:
NOTE An index of defined terms is found beginning on page 42.
Addition:
201.3.201
CEPHALOMETRIC
related to PROJECTION RADIOGRAPHY of the whole dento-maxillo-facial anatomy, whereas the
projection geometry is such to minimize geometrical image distortions
Note 1 to entry: This is usually achieved by setting a sufficiently large source-to-object-distance and source-to-
detector-distance.
Note 2 to entry: Another term often used for CEPHALOMETRIC RADIOGRAPHY is teleradiography.
201.3.202
DENTAL
related to structures in the dento-maxillo-facial district of the PATIENT, including dentition
201.3.203
*DENTAL VOLUMETRIC RECONSTRUCTION
DVR
reconstruction of the 3-dimensional attenuation distribution of the whole or part of the
irradiated volume from a series of 2-dimensional projections produced by an X-RAY BEAM on
an X-RAY IMAGE RECEPTOR moving around the head of the PATIENT
201.3.204
DOSE AREA PRODUCT
product of the area of the cross-section of an X-RAY BEAM and the averaged AIR KERMA over
that cross-section. The unit is the gray square metre (Gy⋅m ).
[SOURCE: IEC 60601-2-54:2009, 201.3.203]
201.3.205
ELECTRONIC X-RAY IMAGE RECEPTOR
X-RAY IMAGE RECEPTOR comprising an electrically-powered conversion method
201.3.206
EXTRA-ORAL
DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located outside the oral
related to
cavity
– 12 – 60601-2-63 © IEC:2012
201.3.207
INTERLOCK
means preventing the start or the continued operation of ME EQUIPMENT unless certain
predetermined conditions prevail
[SOURCE:IEC 60601-2-54:2009, 201.3.207]
201.3.208
INTRA-ORAL
related to DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located, wholly or
partially, inside the oral cavity
201.3.209
NOMINAL SHORTEST IRRADIATION TIME
shortest LOADING TIME for which a required constancy of the controlled RADIATION QUANTIY is
maintained
Note 1 to entry: The IRRADIATION TIME is controlled by a HIGH-VOLTAGE GENERATOR with AUTOMATIC CONTROL
SYSTEMS.
[SOURCE: IEC 60601-2-54:2009, 201.3.208]
201.3.210
ONE-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers an unrectified
output voltage, or rectified output voltage with one peak during each cycle of the supply
[SOURCE: IEC 60601-2-65:2012, 201.3.208
201.3.211
* PANORAMIC
related to DENTAL RADIOGRAPHY, produced by the coordinated motion of a scanning fan-shaped
X-RAY BEAM, oriented parallel to the cranio-caudal axis of the PATIENT, and an X-RAY IMAGE
RECEPTOR, both rotating around the head of the PATIENT
Note 1 to entry: A tomographic layer is produced with respect to the plane perpendicular to the rotational axis.
The resulting image is a focused projection on a surface parallel to the rotational axis.
Note 2 to entry: The scanning axis is usually vertical.
201.3.212
TWO-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers a rectified
output voltage with two peaks during each cycle of the supply
[SOURCE: IEC 60601-2-65:2012, 201.3.209]
201.3.213
X-RAY MONOBLOCK ASSEMBLY
X-RAY TUBE ASSEMBLY containing the HIGH-VOLTAGE TRANSFORMER ASSEMBLY
Note 1 to entry: The term X-RAY MONOBLOCK ASSEMBLY excludes the BEAM LIMITING DEVICE.
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.3 ESSENTIAL PERFORMANCE
Addition:
60601-2-63 © IEC:2012 – 13 –
201.4.3.101 Additional ESSENTIAL PERFORMANCE requirements
The list in Table 201.101 is a list of potential ESSENTIAL PERFORMANCE to be considered by the
MANUFACTURER in the RISK MANAGEMENT PROCESS.
NOTE Subclause 203.6.4.3.102 (Accuracy of LOADING FACTORS) specifies a limitation in applying subclause
203.6.4.3.102.3 (Accuracy of X-RAY TUBE VOLTAGE) and 203.6.4.3.102.4 (Accuracy of X-RAY TUBE CURRENT). This
limitation is also valid for the ESSENTIAL PERFORMANCE list.
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
in the RISK MANAGEMENT PROCESS
MANUFACTURER
Requirement Subclause
Accuracy of LOADING FACTORS 203.6.4.3.102
Reproducibility of the RADIATION output 203.6.3.2
201.4.10.1 Source of power for ME EQUIPMENT
Addition:
201.4.10.1.101 Connection to SUPPLY MAINS
ME EQUIPMENT shall be PERMANENTLY INSTALLED unless the INTENDED USE requires it to be
MOBILE.
201.4.10.2 SUPPLY MAINS for ME EQUIPMENT and ME SYSTEMS
Addition:
The internal impedance of a SUPPLY MAINS is to be considered sufficiently low for the operation
of ME EQUIPMENT if the value of the APPARENT RESISTANCE OF SUPPLY MAINS does not exceed
the value specified in the ACCOMPANYING DOCUMENTS.
ME EQUIPMENT is considered to comply with the requirements of this standard only if its
specified NOMINAL ELECTRIC POWER can be demonstrated at a resistance of supply mains
having a value not less than the APPARENT RESISTANCE OF SUPPLY MAINS specified by the
MANUFACTURER in the ACCOMPANYING DOCUMENTS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and by functional test.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
– 14 – 60601-2-63 © IEC:2012
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
201.7.2.7 Electrical input power from the SUPPLY MAINS
Addition:
Except for items a) to c) below, for ME EQUIPMENT that is specified to be PERMANENTLY
INSTALLED, the information may be stated in the ACCOMPANYING DOCUMENTS only.
The information on the input power shall be specified in terms of combinations of
a) the rated MAINS VOLTAGE of the ME EQUIPMENT in volts; see 7.2.1 and 7.2.6 of the general
standard;
b) the number of phases; see 7.2.1 and 7.2.6 of the general standard;
c) the frequency, in hertz; see 7.2.1 and 7.2.6 of the general standard;
d) the maximum permissible value for APPARENT RESISTANCE OF SUPPLY MAINS, in ohms;
e) the characteristics of OVER-CURRENT RELEASES required in the SUPPLY MAINS.
NOTE These requirements are adapted from IEC 60601-2-7 subclause 6.1j).
201.7.2.15 Cooling conditions
Addition:
If cooling is necessary for safe operation of ME EQUIPMENT, or a subassembly thereof, the
ACCOMPANYING DOCUMENTS, as identified in the
cooling requirements shall be indicated in the
RISK MANAGEMENT PROCESS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and the RISK
MANAGEMENT FILE.
Additional subclause:
201.7.2.101 BEAM LIMITING DEVICE
Where detachable in NORMAL USE, BEAM LIMITING DEVICES shall be provided with the following
markings:
– those required in subclause 7.2.2 of the general standard;
– serial designation or individual identification;
– ADDITIONAL FILTRATION, if the additional value is more than the equivalent of 0,2 mm Al.
Compliance is checked by inspection.
201.7.8.1 Colours of indicator lights
Addition:
The indication of X-ray related states shall be excluded from subclause 7.8 in the general
standard. 203.6.4.2 and 203.6.4.101 shall apply instead.
201.7.9 ACCOMPANYING DOCUMENTs
201.7.9.1 General
Addition:
NOTE 101 Annex C Table 201.C.102 lists the requirements of this particular standard that are additional to those
of the general standard for statements in the ACCOMPANYING DOCUMENTS.
60601-2-63 © IEC:2012 – 15 –
The ACCOMPANYING DOCUMENTS shall contain quality control PROCEDURES to be performed on
the ME EQUIPMENT by the RESPONSIBLE ORGANISATION. These shall include acceptance criteria
and frequency for the tests.
Additionally for ME EQUIPMENT provided with one or more integrated ELECTRONIC X-RAY IMAGE
RECEPTORs, the ACCOMPANYING DOCUMENTS shall contain a description of the minimum
performance required of the means, used to display the images for diagnostic purpose
according to the INTENDED USE for each ELECTRONIC X-RAY IMAGE RECEPTOR used.
NOTE 102 For instance, the minimum required number of pixels and number of discernible grey levels at the
DISPLAY.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.7.9.2 Instructions for use
201.7.9.2.1 General
Additional subclause:
201.7.9.2.1.101 LOADING FACTORS
In the instructions for use the LOADING FACTORS shall be stated as described below. The
following combinations and data shall be stated:
a) value(s) or range of X-RAY TUBE VOLTAGE settings;
b) value(s) or range of X-RAY TUBE CURRENT settings;
c) values or range of IRRADIATION TIME settings;
d) maximum X-RAY TUBE CURRENT at each X-RAY TUBE VOLTAGE setting, if different from (b);
e) maximum and minimum IRRADIATION TIME at each X-RAY TUBE VOLTAGE and X-RAY TUBE
CURRENT setting, if different from (c);
f) for ME EQUIPMENT indicating precalculated or measured CURRENT TIME PRODUCT, the lowest
CURRENT TIME PRODUCT or the combinations of LOADING FACTORS resulting in the lowest
CURRENT TIME PRODUCT;
g) if the value of the lowest CURRENT TIME PRODUCT depends upon the X-RAY TUBE VOLTAGE or
upon certain combinations of values of LOADING FACTORS, the lowest CURRENT TIME
PRODUCT may be given as a table or curve showing the dependence;
h) the NOMINAL SHORTEST IRRADIATION TIME used in AUTOMATIC EXPOSURE CONTROL systems of
ME EQUIPMENT;
i) if the NOMINAL SHORTEST IRRADIATION TIME depends upon LOADING FACTORS such as X-RAY
TUBE VOLTAGE and X-RAY TUBE CURRENT, the ranges of these LOADING FACTORS for which
the NOMINAL SHORTEST IRRADIATION TIME is valid;
j) the maximum possible range of the X-RAY TUBE VOLTAGE and/or the X-RAY TUBE CURRENT
during IRRADIATION, controlled with the AUTOMATIC CONTROL SYSTEM
NOTE these requirements are adapted from IEC 60601-2-7 subclause 6.8.2 a).
Compliance is checked by inspection of the instructions for use.
201.7.9.3 Technical description
Additional subclauses:
201.7.9.3.101 Technical description for DENTAL EXTRA-ORAL X-RAY EQUIPMENT
In addition to the data required to be marked according to subclause 7.2 of the general
standard, the technical description shall specify a description of the geometric relationship of
– 16 – 60601-2-63 © IEC:2012
the FOCAL SPOT, X-RAY BEAM dimensions, PATIENT position and IMAGE RECEPTION AREA, as well
as a method used to estimate the DOSE AREA PRODUCT.
Additional subclause:
201.7.9.3.102 X-RAY SOURCE ASSEMBLY
The technical description of the integrated X-RAY SOURCE ASSEMBLY shall specify the following,
in addition to the data required to be marked according to subclause 7.2 of the general
standard:
a) specification of the REFERENCE AXIS to which the TARGET ANGLE(s) and the FOCAL SPOT
characteristics of the X-RAY SOURCE ASSEMBLY refer:
b) TARGET ANGLE(s) with respect to the specified REFERENCE AXIS;
c) NOMINAL FOCAL SPOT VALUE(s) determined according to IEC 60336 for the specified
REFERENCE AXIS.
Compliance is checked by inspection of the technical description.
Additional subclause:
201.7.9.101 Requirements to the SUPPLY MAINS
The information on the RATED electrical input power for DENTAL EXTRA-ORAL X-RAY GENERATORS
shall also include:
• either the maximum permissible value for the APPARENT RESISTANCE OF SUPPLY MAINS or
SUPPLY MAINS specifications used in a facility, and
other appropriate
• the characteristics of OVER-CURRENT RELEASES eventually required in the SUPPLY MAINS.
201.8 Protection against electrical HAZARDS from ME EQUIPMENT
Clause 8 of the general standard applies, except as follows:
201.8.5 Separation of parts
201.8.5.1 MEANS OF PROTECTION (MOP)
Additional subclause:
201.8.5.1.101 Additional limitation of voltage, current or energy
Provisions shall be made to prevent the appearance of an unacceptably HIGH VOLTAGE in the
MAINS PART or in any other low-voltage circuit.
NOTE This may be achieved for example by
– provision of a winding layer or a conductive screen connected to the PROTECTIVE EARTH TERMINAL between
HIGH VOLTAGE and low-voltage circuits, or
– provision of a voltage limiting device across terminals to which external devices are connected and between
which an excessive voltage might arise if the external path becomes open-circuited.
Compliance is checked by inspection of design data and construction.
NOTE These requirements are adapted from IEC 60601-2-7:1998, subclause 15bb).
201.8.5.4 Working voltage
Addition:
60601-2-63 © IEC:2012 – 17 –
201.8.5.4.101 Stator and stator circuit dielectric strength testing
The test voltage for the dielectric strength testing of stator and stator circuit used for the
operation of the rotating anode of the X-RAY TUBE is to be referred to the voltage existing after
reduction of the stator supply voltage to its steady state operating value.
NOTE These requirements are adapted from IEC 60601-2-7:1998, subclause 20.4 l).
201.8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS
201.8.7.3 * Allowable values
Item c) is amended as follows:
PERMANENTLY INSTALLED X-RAY GENERATORS the allowable value of TOUCH CURRENT in
For non-
SINGLE FAULT CONDITION is 2 mA.
NOTE This relaxation from the requirement of the general standard does not apply to PATIENT LEAKAGE CURRENT.
Item e) is amended as follows:
For PERMANENTLY INSTALLED ME EQUIPMENT, the allowable value of EARTH LEAKAGE CURRENT is
20 mA in NORMAL CONDITION and SINGLE FAULT CONDITION.
201.8.8.3 * Dielectric strength
Addition:
Instead of subclause 8.8.3 of the general standard the HIGH VOLTAGE circuit of X-RAY
MONOBLOCK ASSEMBLIES shall be tested as follows:
The test for the HIGH VOLTAGE circuit shall be made with a test voltage between 1,1 and 1,15
NOMINAL X-RAY TUBE VOLTAGE of the X-RAY MONOBLOCK ASSEMBLY. If the
times the maximum
HIGH VOLTAGE circuit is not accessible, the voltage measurement may be indirect.
The HIGH VOLTAGE circuit of X-RAY MONOBLOCK ASSEMBLIES is tested by applying the test
voltage for a time equal to two times the maximum permissible IRRADIATION TIME for NORMAL
USE, as specified in the ACCOMPANYING DOCUMENTS. The test is repeated three times with a
minimum interval of two minutes between each test.
ONE-PEAK HIGH-VOLTAGE GENERATORS, the test voltage for the HIGH VOLTAGE circuit shall be
For
referred to the no-load half cycle if the X-RAY TUBE VOLTAGE for the no-load half cycle is higher
than in the on-load half cycle.
If during the dielectric strength test there is a RISK of overheating a transformer under test, it
is permitted to carry out the test at a higher supply frequency.
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and
ME SYSTEMS
Clause 9 of the general standard applies, except as follows:
201.9.8 HAZARDS associated with support systems
201.9.8.4 Systems with MECHANICAL PROTECTIVE DEVICES
Additional subclause:
– 18 – 60601-2-63 © IEC:2012
201.9.8.4.101 MECHANICAL PROTECTIVE DEVICE
Ropes, chains or bands running parallel to other rope, chains or bands may be regarded as a
MECHANICAL PROTECTIVE DEVICE if they are not loaded during NORMAL USE.
Ropes, chains or bands used as a MECHANICAL PROTECTIVE DEVICE shall be accessible for
inspection and the ACCOMPANYING DOCUMENTS shall give appropriate instructions for
inspection.
Compliance is checked by functional test and inspection of ACCOMPANYING DOCUMENTS.
201.10 Protection against unwanted and excessive radiation HAZARDS
Clause 10 of the general standard applies.
NOTE The collateral standard IEC 60601-1-3 is referenced in the general standard and is covered under clause
203 of this document.
201.11 Protection against excessive temperatures and othe
...
IEC 60601-2-63 ®
Edition 1.2 2021-05
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X dentaires extra-oraux
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.
IEC publications search - webstore.iec.ch/advsearchform IEC online collection - oc.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews. With a subscription you will always
committee, …). It also gives information on projects, replaced have access to up to date content tailored to your needs.
and withdrawn publications.
Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
The world's leading online dictionary on electrotechnology,
Stay up to date on all new IEC publications. Just Published
containing more than 22 000 terminological entries in English
details all new publications released. Available online and
and French, with equivalent terms in 18 additional languages.
once a month by email.
Also known as the International Electrotechnical Vocabulary
(IEV) online.
IEC Customer Service Centre - webstore.iec.ch/csc
If you wish to give us your feedback on this publication or
need further assistance, please contact the Customer Service
Centre: sales@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Recherche de publications IEC - IEC online collection - oc.iec.ch
webstore.iec.ch/advsearchform Découvrez notre puissant moteur de recherche et consultez
La recherche avancée permet de trouver des publications IEC gratuitement tous les aperçus des publications. Avec un
en utilisant différents critères (numéro de référence, texte, abonnement, vous aurez toujours accès à un contenu à jour
comité d’études, …). Elle donne aussi des informations sur adapté à vos besoins.
les projets et les publications remplacées ou retirées.
Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
Le premier dictionnaire d'électrotechnologie en ligne au
Restez informé sur les nouvelles publications IEC. Just
monde, avec plus de 22 000 articles terminologiques en
Published détaille les nouvelles publications parues.
anglais et en français, ainsi que les termes équivalents dans
Disponible en ligne et une fois par mois par email.
16 langues additionnelles. Egalement appelé Vocabulaire
Electrotechnique International (IEV) en ligne.
Service Clients - webstore.iec.ch/csc
Si vous désirez nous donner des commentaires sur cette
publication ou si vous avez des questions contactez-nous:
sales@iec.ch.
IEC 60601-2-63 ®
Edition 1.2 2021-05
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential
performance of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les
performances essentielles des appareils à rayonnement X dentaires extra-oraux
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.50 ISBN 978-2-8322-4309-1
IEC 60601-2-63 ®
Edition 1.2 2021-05
CONSOLIDATED VERSION
REDLINE VERSION
VERSION REDLINE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X dentaires extra-oraux
– 2 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
INTRODUCTION TO AMENDMENT 1 . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terminology and definitions . 11
201.4 General requirements . 13
201.5 General requirements for testing of ME EQUIPMENT . 14
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 14
201.7 ME EQUIPMENT identification, marking and documents . 14
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 17
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 18
201.10 Protection against unwanted and excessive radiation HAZARDS . 19
201.11 Protection against excessive temperatures and other HAZARDS . 19
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 19
201.13 HAZARDOUS SITUATIONS and fault conditions . 19
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 19
201.15 Construction of ME EQUIPMENT . 19
201.16 ME SYSTEMS . 19
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 19
202 Electromagnetic compatibility – Requirements and tests . 20
203 Radiation protection in diagnostic X-ray equipment . 20
Annexes . 33
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS . 34
Annex AA (informative) Particular guidance and rationale . 35
Bibliography . 42
Index of defined terms used in this particular standard . 45
Figure 203.101 – Zone of EXTRA-FOCAL RADIATION . 29
Figure AA.1 – PANORAMIC X-RAY EQUIPMENT . 35
Figure AA.2 – AIR KERMA during IRRADIATION with direct current X-RAY GENERATOR . 37
Figure AA.3 – AIR KERMA during IRRADIATION with ONE-PEAK X-RAY GENERATOR . 38
Figure AA.4 – Example – series of (numerous) pulsed IRRADIATIONS for a CBCT (cone
beam computed tomography) IRRADIATION event, with CONSTANT POTENTIAL HIGH-
VOLTAGE GENERATOR and time-width modulation . 39
Figure AA.5 – Example – series of two irradiations for PANORAMIC-like views of right
and left TMJ (temporo-mandibolar joint) in the same image, with ONE-PEAK HIGH-
VOLTAGE GENERATOR . 40
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
MANUFACTURER in the RISK MANAGEMENT PROCESS . 13
+AMD2:2021 CSV © IEC 2021
Table 201.C.101 – Marking on the outside of ME EQUIPMENT or its parts . 34
Table 201.C.102 – Subclauses requiring statements in ACCOMPANYING DOCUMENTS . 34
– 4 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This consolidated version of the official IEC Standard and its amendments has been
prepared for user convenience.
IEC 60601-2-63 edition 1.2 contains the first edition (2012-09) [documents 62B/888/FDIS
and 62B/898/RVD], its amendment 1 (2017-07) [documents 62B/1049/FDIS and
62B/1058/RVD] and its amendment 2 (2021-05) [documents 62B/1232/FDIS and
62B/1237/RVD].
In this Redline version, a vertical line in the margin shows where the technical content
is modified by amendments 1 and 2. Additions are in green text, deletions are in
strikethrough red text. A separate Final version with all changes accepted is available
in this publication.
+AMD2:2021 CSV © IEC 2021
International Standard IEC 60601-2-63 has been prepared by IEC subcommittee 62B:
Diagnostic imaging equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: in roman type.
– Test specifications: in italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
The committee has decided that the contents of the base publication and its amendments will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
+AMD2:2021 CSV © IEC 2021
INTRODUCTION
This particular standard has been prepared to provide, based on IEC 60601-1:2005 (third
edition), and its collaterals, a complete set of BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for DENTAL EXTRA-ORAL X-RAY EQUIPMENT. While the previously existing standards
for such equipment were dedicated to components and subsystems, this particular standard
addresses the system level of DENTAL EXTRA-ORAL X-RAY EQUIPMENT. Components and their
functions are addressed as far as necessary.
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of DENTAL EXTRA-ORAL
X-RAY EQUIPMENT
The minimum safety requirements for DENTAL INTRA-ORAL X-RAY EQUIPMENT are specified in a
separate particular standard IEC 60601-2-65 to simplify and improve the readability
Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions
of the collateral standard IEC 60601-1-3, the particular standards IEC 60601-2-28
IEC 60601-2-7, or IEC 60601-2-32 have been extracted and moved into this particular
standard.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard.
INTRODUCTION TO AMENDMENT 1
The purpose of this first amendment to IEC 60601-2-63:2012 is to introduce changes to
reference the Amendment 1 (2012) to IEC 60601-1:2005. As neither IEC 60601-2-63:2012 nor
this amendment refers to specific elements of IEC 60601-1-2, the introduction of a dated
reference to the latter document has been removed.
– 8 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT, hereafter also called ME EQUIPMENT. The scope includes
ME SYSTEMS containing such ME EQUIPMENT.
NOTE 1 This includes PANORAMIC equipment, CEPHALOMETRIC equipment, and equipment for dental volumetric
reconstruction (hereafter DVR) as defined in 201.3.203 below. An example of such equipment is an equipment
designed to perform PANORAMIC, CEPHALOMETRIC and DENTAL VOLUMETRIC RECONSTRUCTION (hereafter DVR) as
defined in 201.3.203.
NOTE 2 DVR includes dental CBCT (cone beam computed tomography), which is also known with other names in
certain parts of the world, e.g. DVT (digital volumetric tomography); DVR also includes tomosynthesis.
NOTE 3 This may include the imaging of other anatomical parts (e.g. the hand) as long as required for dental
treatment (e.g. orthodontic treatment).
NOTE 4 This may include anatomical objects of interest to the ENT (ear, nose, and throat) specialist.
The scope of this standard is restricted to X-RAY EQUIPMENT where:
• the X-RAY TUBE ASSEMBLY contains the HIGH-VOLTAGE TRANSFORMER ASSEMBLY and
• the geometrical relations between the X-RAY SOURCE, the anatomical object being imaged
in the PATIENT, and the X-RAY IMAGE RECEPTOR, are preset in the design and cannot be
arbitrarily altered by the OPERATOR during INTENDED USE.
DENTAL EXTRA-ORAL X-RAY EQUIPMENT are X-RAY EQUIPMENT designed for EXTRA-ORAL
RADIOGRAPHY in which the geometrical relations between the X-RAY SOURCE, the anatomical
object being imaged in the PATIENT, and the X-RAY IMAGE RECEPTOR, are preset in the design
and cannot be arbitrarily altered by the OPERATOR during INTENDED USE. In such equipment, the
X-RAY TUBE ASSEMBLY contains the HIGH-VOLTAGE TRANSFORMER ASSEMBLY.
NOTE 5 DENTAL INTRA-ORAL X-RAY EQUIPMENT is excluded from the scope of this standard.
NOTE 6 FOCAL SPOT TO IMAGE RECEPTOR DISTANCE and FOCAL SPOT to object distance are preset in the design of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
NOTE 7 For DENTAL X-RAY EQUIPMENT not in the scope of this document because of the restriction above,
applicable clauses of IEC 60601-2-54 may be used with this document.
ME EQUIPMENT and ME SYSTEMS in the scope of IEC 60601-2-44, IEC 60601-2-54,
IEC 60601-2-45, IEC 60601-2-65 or IEC 60601-2-43 are excluded from the scope of this
particular standard. The scope of this International Standard also excludes RADIOTHERAPY
SIMULATORS and equipment for bone or tissue absorption densitometry. Excluded from the
ME EQUIPMENT intended to be used for DENTAL RADIOSCOPY.
scope is also
—————————
1)
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance
+AMD2:2021 CSV © IEC 2021
Within its specific scope, the clauses of this particular standard supersede and replace those
of IEC 60601-2-7, Medical electrical equipment – Particular requirements for the safety of
high-voltage generators of diagnostic X-ray generators and of IEC 60601-2-32, Medical
electrical equipment – Particular requirements for the safety of associated equipment of X-ray
equipment.
NOTE 8 Requirements for X-RAY GENERATORS and for ASSOCIATED EQUIPMENT, which were previously specified in
IEC 60601-2-7 and IEC 60601-2-32, have been included in either IEC 60601-1:2005 (Ed3) or this particular
rd
standard. Therefore IEC 60601-2-7 and IEC 60601-2-32 are not part of the IEC 60601-1 3 edition scheme for
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard. Therefore IEC 60601-2-28 does not apply to ME EQUIPMENT in the scope of this
International Standard with the exception of X-RAY TUBE ASSEMBLIES that are replaceable in
the field by entities other than the manufacturer.
NOTE 9 Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions of the
collateral standard IEC 60601-1-3 or the particular standard IEC 60601-2-28 have been extracted and moved into
this particular standard.
NOTE 10 For X-RAY EQUIPMENT in the scope of this particular standard X-RAY TUBE ASSEMBLIES are X-RAY
MONOBLOCK ASSEMBLIES.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for ME EQUIPMENT for EXTRA-ORAL DENTAL RADIOGRAPHY.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-3 apply as modified in Clause 202 and 203 respectively.
2) ) )
3 4
IEC 60601-1-8, IEC 60601-1-10 , IEC 60601-1-11 and IEC 60601-1-12 do not apply. All
other published collateral standards in the IEC 60601-1 series apply as published.
NOTE OPERATORS of DENTAL EXTRA-ORAL X-RAY EQUIPMENT are used to audible signals as required in this
particular standard rather than to the concepts of IEC 60601-1-8. Therefore IEC 60601-1-8 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard or collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
—————————
)
IEC 60601-1-10, Medical electrical equipment – Part 1-10: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for the development of physiologic closed-loop controllers
)
IEC 60601-1-11, Medical electrical equipment – Part 1-11: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment
4)
IEC 60601-1-12, Medical electrical equipment – Part 1-12: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems intended for use in the emergency medical services environment
– 10 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 42.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic compatibility –
Requirements and tests
IEC 60601-1-3:2008, Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic X-
+AMD2:2021 CSV © IEC 2021
ray equipment
IEC 60601-1-3:2008/AMD1:2013
Addition:
IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis –
Characteristics of focal spots
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-2-29:2008, Medical electrical equipment – Part 2-29: Particular requirements for
the basic safety and essential performance of radiotherapy simulators
IEC 60601-2-54:2009, Medical electrical equipment – Part 2-54: Particular requirements for
the basic safety and essential performance of X-ray equipment for radiography and
radioscopy
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC/PAS 61910-1:20072014, Medical electrical equipment – Radiation dose documentation –
Part 1: Equipment Radiation dose structured reports for radiography and radioscopy
201.3 Terminology and definitions
Amendment:
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1:2005/AMD1:2012, its applicable collateral standards, IEC/TR 60788:2004 and
the following apply:
NOTE An index of defined terms is found beginning on page 45.
Addition:
201.3.201
CEPHALOMETRIC
related to PROJECTION RADIOGRAPHY of the whole dento-maxillo-facial anatomy, whereas the
projection geometry is such to minimize geometrical image distortions
Note 1 to entry: This is usually achieved by setting a sufficiently large source-to-object-distance and source-to-
detector-distance.
Note 2 to entry: Another term often used for CEPHALOMETRIC RADIOGRAPHY is teleradiography.
201.3.202
DENTAL
related to structures in the dento-maxillo-facial district of the PATIENT, including dentition
201.3.203
*DENTAL VOLUMETRIC RECONSTRUCTION
DVR
reconstruction of the 3-dimensional attenuation distribution of the whole or part of the
irradiated volume from a series of 2-dimensional projections produced by an X-RAY BEAM on
an X-RAY IMAGE RECEPTOR moving around the head of the PATIENT
– 12 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
201.3.204
DOSE AREA PRODUCT
product of the area of the cross-section of an X-RAY BEAM and the averaged AIR KERMA over
that cross-section. The unit is the gray square metre (Gy⋅m ).
[SOURCE: IEC 60601-2-54:2009, 201.3.203]
201.3.205
ELECTRONIC X-RAY IMAGE RECEPTOR
X-RAY IMAGE RECEPTOR comprising an electrically-powered conversion method
201.3.206
EXTRA-ORAL
related to DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located outside the oral
cavity
201.3.207
INTERLOCK
means preventing the start or the continued operation of ME EQUIPMENT unless certain
predetermined conditions prevail
[SOURCE:IEC 60601-2-54:2009, 201.3.207]
201.3.208
INTRA-ORAL
related to DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located, wholly or
partially, inside the oral cavity
201.3.209
NOMINAL SHORTEST IRRADIATION TIME
shortest LOADING TIME for which a required constancy of the controlled RADIATION QUANTIY is
maintained
Note 1 to entry: The IRRADIATION TIME is controlled by a HIGH-VOLTAGE GENERATOR with AUTOMATIC CONTROL
SYSTEMS.
[SOURCE: IEC 60601-2-54:2009, 201.3.208]
201.3.210
ONE-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers an unrectified
output voltage, or rectified output voltage with one peak during each cycle of the supply
[SOURCE: IEC 60601-2-65:2012, 201.3.208
201.3.211
* PANORAMIC
related to DENTAL RADIOGRAPHY, produced by the coordinated motion of a scanning fan-shaped
X-RAY BEAM, oriented parallel to the cranio-caudal axis of the PATIENT, and an X-RAY IMAGE
RECEPTOR, both rotating around the head of the PATIENT
Note 1 to entry: A tomographic layer is produced with respect to the plane perpendicular to the rotational axis.
The resulting image is a focused projection on a surface parallel to the rotational axis.
Note 2 to entry: The scanning axis is usually vertical.
201.3.212
TWO-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers a rectified
output voltage with two peaks during each cycle of the supply
+AMD2:2021 CSV © IEC 2021
[SOURCE: IEC 60601-2-65:2012, 201.3.209]
201.3.213
X-RAY MONOBLOCK ASSEMBLY
X-RAY TUBE ASSEMBLY containing the HIGH-VOLTAGE TRANSFORMER ASSEMBLY
Note 1 to entry: The term X-RAY MONOBLOCK ASSEMBLY excludes the BEAM LIMITING DEVICE.
201.3.214
DENTAL CONE BEAM COMPUTED TOMOGRAPHY
DENTAL CBCT
3-dimensional imaging of DENTAL anatomical structures, performed by reconstruction of a
volume from a series of 2-dimensional projections produced by circular or rectangular
collimated X-RAY BEAM on an X-RAY IMAGE RECEPTOR rotating around the head of the PATIENT
201.3.215
EXAMINATION PROTOCOL
full set of programmed LOADING FACTORS, control functions and settings, including image
processing settings, designed to the image acquisition and DISPLAY
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.3 ESSENTIAL PERFORMANCE
Addition:
201.4.3.101 Additional potential ESSENTIAL PERFORMANCE requirements
The list in Table 201.101 is a list of potential ESSENTIAL PERFORMANCE to be considered by the
MANUFACTURER in the RISK MANAGEMENT PROCESS.
NOTE Subclause 203.6.4.3.102 (Accuracy of LOADING FACTORS) specifies a limitation in applying subclause
203.6.4.3.102.3 (Accuracy of X-RAY TUBE VOLTAGE) and 203.6.4.3.102.4 (Accuracy of X-RAY TUBE CURRENT). This
limitation is also valid for the ESSENTIAL PERFORMANCE list.
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
MANUFACTURER in the RISK MANAGEMENT PROCESS
Requirement Subclause
Accuracy of LOADING FACTORS 203.6.4.3.102
Reproducibility of the RADIATION output 203.6.3.2
201.4.10.1 Source of power for ME EQUIPMENT
Addition:
201.4.10.1.101 Connection to SUPPLY MAINS
ME EQUIPMENT shall be PERMANENTLY INSTALLED unless the INTENDED USE requires it to be
MOBILE.
201.4.10.2 SUPPLY MAINS for ME EQUIPMENT and ME SYSTEMS
Addition:
– 14 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
The internal impedance of a SUPPLY MAINS is to be considered sufficiently low for the operation
of ME EQUIPMENT if the value of the APPARENT RESISTANCE OF SUPPLY MAINS does not exceed
the value specified in the ACCOMPANYING DOCUMENTS.
For this purpose, the APPARENT RESISTANCE of SUPPLY MAINs R is determined according to the
formula:
UU−
R =
I
where
U is the no-load MAINS VOLTAGE;
U is the MAINS VOLTAGE under load;
I is the mains current under load.
ME EQUIPMENT is considered to comply with the requirements of this standard only if its
specified NOMINAL ELECTRIC POWER can be demonstrated at a resistance of supply mains
having a value not less than the APPARENT RESISTANCE OF SUPPLY MAINS specified by the
MANUFACTURER in the ACCOMPANYING DOCUMENTS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and by functional test.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
201.7.2.7 Electrical input power from the SUPPLY MAINS
Addition:
Except for items a) to c) below, for ME EQUIPMENT that is specified to be PERMANENTLY
INSTALLED, the information may be stated in the ACCOMPANYING DOCUMENTS only.
The information on the input power shall be specified in terms of combinations of
a) the rated MAINS VOLTAGE of the ME EQUIPMENT in volts; see 7.2.1 and 7.2.6 of the general
standard;
b) the number of phases; see 7.2.1 and 7.2.6 of the general standard;
c) the frequency, in hertz; see 7.2.1 and 7.2.6 of the general standard;
d) the maximum permissible value for APPARENT RESISTANCE OF SUPPLY MAINS, in ohms;
e) the characteristics of OVER-CURRENT RELEASES required in the SUPPLY MAINS.
+AMD2:2021 CSV © IEC 2021
NOTE These requirements are adapted from IEC 60601-2-7 subclause 6.1j).
201.7.2.15 Cooling conditions
Addition:
If cooling is necessary for safe operation of ME EQUIPMENT, or a subassembly thereof, the
cooling requirements shall be indicated in the ACCOMPANYING DOCUMENTS, as identified in the
RISK MANAGEMENT PROCESS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and the RISK
MANAGEMENT FILE.
Additional subclause:
201.7.2.101 BEAM LIMITING DEVICE
Where detachable in NORMAL USE, BEAM LIMITING DEVICES shall be provided with the following
markings:
– those required in subclause 7.2.2 of the general standard;
– serial designation or individual identification;
– ADDITIONAL FILTRATION, if the additional value is more than the equivalent of 0,2 mm Al.
Compliance is checked by inspection.
201.7.8.1 Colours of indicator lights
Addition:
The indication of X-ray related states shall be excluded from subclause 7.8 in the general
standard. 203.6.4.2 and 203.6.4.101 shall apply instead.
201.7.9 ACCOMPANYING DOCUMENTs
201.7.9.1 General
Addition:
NOTE 101 Annex C Table 201.C.102 lists the requirements of this particular standard that are additional to those
of the general standard for statements in the ACCOMPANYING DOCUMENTS.
The ACCOMPANYING DOCUMENTS shall contain quality control PROCEDURES to be performed on
the ME EQUIPMENT by the RESPONSIBLE ORGANISATION. These shall include acceptance criteria
and frequency for the tests.
Additionally for ME EQUIPMENT provided with one or more integrated ELECTRONIC X-RAY IMAGE
RECEPTORs, the ACCOMPANYING DOCUMENTS shall contain a description of the minimum
performance required of the means, used to display the images for diagnostic purpose
according to the INTENDED USE for each ELECTRONIC X-RAY IMAGE RECEPTOR used.
If a test or a QUALITY CONTROL PROCEDURE recommended by the MANUFACTURER requires a
device-specific arrangement (including a TOOL, a PHANTOM, a special software or a software
setting); that is only available from the MANUFACTURER, the MANUFACTURER shall provide this
arrangement for the RESPONSIBLE ORGANIZATION.
NOTE 103 The intention is to perform these QUALITY CONTROL PROCEDURES and tests using only the
ACCOMPANYING DOCUMENTS.
NOTE 102 For instance, the minimum required number of pixels and number of discernible grey levels at the
DISPLAY.
– 16 – IEC 60601-2-63:2012+AMD1:2017
+AMD2:2021 CSV © IEC 2021
If the test or PROCEDURE requires a device-specific TOOL that is only available from the
MANUFACTURER, the MANUFACTURER shall make this TOOL available to the RESPONSIBLE
ORGANIZATION.
NOTE 104 The MANUFACTURER can provide PHANTOM with the equipment, if specified in the local regulations.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.7.9.2 Instructions for use
201.7.9.2.1 General
Additional subclause:
201.7.9.2.1.101 LOADING FACTORS
In the instructions for use the LOADING FACTORS shall be stated as described below. The
following combinations and data shall be stated:
a) value(s) or range of X-RAY TUBE VOLTAGE settings;
b) value(s) or range of X-RAY TUBE CURRENT settings;
c) values or range of IRRADIATION TIME settings;
d) maximum X-RAY TUBE CURRENT at each X-RAY TUBE VOLTAGE setting, if different from (b);
e) maximum and minimum IRRADIATION TIME at each X-RAY TUBE VOLTAGE and X-RAY TUBE
CURRENT setting, if different from (c);
f) for ME EQUIPMENT indicating precalculated or measured CURRENT TIME PRODUCT, the lowest
CURRENT TIME PRODUCT or the combinations of LOADING FACTORS resulting in the lowest
CURRENT TIME PRODUCT;
g) if the value of the lowest CURRENT TIME PRODUCT depends upon the X-RAY TUBE VOLTAGE or
upon certain combinations of values of LOADING FACTORS, the lowest CURRENT TIME
PRODUCT may be given as a table or curve showing the dependence;
h) the NOMINAL SHORTEST IRRADIATION TIME used in AUTOMATIC EXPOSURE CONTROL systems of
ME EQUIPMENT;
i) if the NOMINAL SHORTEST IRRADIATION TIME depends upon LOADING FACTORS such as X-RAY
TUBE VOLTAGE and X-RAY TUBE CURRENT, the ranges of these LOADING FACTORS for which
the NOMINAL SHORTEST IRRADIATION TIME is valid;
j) the maximum possible range of the X-RAY TUBE VOLTAGE and/or the X-RAY TUBE CURRENT
during IRRADIATION, controlled with the AUTOMATIC CONTROL SYSTEM
NOTE these requirements are adapted from IEC 60601-2-7 subclause 6.8.2 a).
Compliance is checked by inspection of the instructions for use.
201.7.9.3 Technical description
Additional subclauses:
201.7.9.3.101 Technical description for DENTAL EXTRA-ORAL X-RAY EQUIPMENT
In addition to the data required to be marked according to subclause 7.2 of the general
standard, the technical description shall specify a description of the geometric relationship of
the FOCAL SPOT, X-RAY BEAM dimensions, PATIENT position and IMAGE RECEPTION AREA, as well
as a method used to estimate the DOSE AREA PRODUCT.
Additional subclause:
+AMD2:2021 CSV © IEC 2021
201.7.9.3.102 X-RAY SOURCE ASSEMBLY
RAY SOURCE ASSEMBLY shall specify the following,
The technical description of the integrated X-
in addition to the data required to be marked according to subclause 7.2 of the general
standard:
a) specification of the REFERENCE AXIS to which the TARGET ANGLE(s) and t
...
IEC 60601-2-63 ®
Edition 1.1 2017-07
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X dentaires extra-oraux
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing 20 000 terms and definitions in
Technical Specifications, Technical Reports and other English and French, with equivalent terms in 16 additional
documents. Available for PC, Mac OS, Android Tablets and languages. Also known as the International Electrotechnical
iPad. Vocabulary (IEV) online.
IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a 65 000 electrotechnical terminology entries in English and
variety of criteria (reference number, text, technical French extracted from the Terms and Definitions clause of
committee,…). It also gives information on projects, replaced IEC publications issued since 2002. Some entries have been
and withdrawn publications. collected from earlier publications of IEC TC 37, 77, 86 and
CISPR.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements
Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales,
électriques. Il contient 20 000 termes et définitions en anglais
Spécifications techniques, Rapports techniques et autres
et en français, ainsi que les termes équivalents dans 16
documents de l'IEC. Disponible pour PC, Mac OS, tablettes
langues additionnelles. Egalement appelé Vocabulaire
Android et iPad.
Electrotechnique International (IEV) en ligne.
Recherche de publications IEC - www.iec.ch/searchpub
Glossaire IEC - std.iec.ch/glossary
La recherche avancée permet de trouver des publications IEC
65 000 entrées terminologiques électrotechniques, en anglais
en utilisant différents critères (numéro de référence, texte,
et en français, extraites des articles Termes et Définitions des
comité d’études,…). Elle donne aussi des informations sur les
publications IEC parues depuis 2002. Plus certaines entrées
projets et les publications remplacées ou retirées.
antérieures extraites des publications des CE 37, 77, 86 et
CISPR de l'IEC.
IEC Just Published - webstore.iec.ch/justpublished
Restez informé sur les nouvelles publications IEC. Just Service Clients - webstore.iec.ch/csc
Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur cette
Disponible en ligne et aussi une fois par mois par email. publication ou si vous avez des questions contactez-nous:
csc@iec.ch.
IEC 60601-2-63 ®
Edition 1.1 2017-07
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential
performance of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les
performances essentielles des appareils à rayonnement X dentaires extra-oraux
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.50 ISBN 978-2-8322-4584-2
IEC 60601-2-63 ®
Edition 1.1 2017-07
CONSOLIDATED VERSION
REDLINE VERSION
VERSION REDLINE
colour
inside
Medical electrical equipment –
Part 2-63: Particular requirements for the basic safety and essential performance
of dental extra-oral X-ray equipment
Appareils électromédicaux –
Partie 2-63: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X dentaires extra-oraux
– 2 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
INTRODUCTION TO AMENDMENT 1 . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terminology and definitions . 11
201.4 General requirements . 13
201.5 General requirements for testing of ME EQUIPMENT . 13
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 13
201.7 ME EQUIPMENT identification, marking and documents . 14
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 16
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 18
201.10 Protection against unwanted and excessive radiation HAZARDS . 18
201.11 Protection against excessive temperatures and other HAZARDS . 18
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 18
201.13 HAZARDOUS SITUATIONS and fault conditions . 18
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 19
201.15 Construction of ME EQUIPMENT . 19
201.16 ME SYSTEMS . 19
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 19
202 Electromagnetic compatibility – Requirements and tests . 19
203 Radiation protection in diagnostic X-ray equipment . 19
Annexes . 31
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS . 32
Annex AA (informative) Particular guidance and rationale . 33
Bibliography . 39
Index of defined terms used in this particular standard . 42
Figure 203.101 – Zone of EXTRA-FOCAL RADIATION . 28
Figure AA.1 – PANORAMIC X-RAY EQUIPMENT . 33
Figure AA.2 – AIR KERMA during IRRADIATION with direct current X-RAY GENERATOR . 35
Figure AA.3 – AIR KERMA during IRRADIATION with ONE-PEAK X-RAY GENERATOR . 36
Figure AA.4 – Example – series of (numerous) pulsed IRRADIATIONS for a CBCT (cone
beam computed tomography) IRRADIATION event, with CONSTANT POTENTIAL HIGH-
VOLTAGE GENERATOR and time-width modulation . 37
Figure AA.5 – Example – series of two irradiations for PANORAMIC-like views of right
and left TMJ (temporo-mandibolar joint) in the same image, with ONE-PEAK HIGH-
VOLTAGE GENERATOR . 37
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
MANUFACTURER in the RISK MANAGEMENT PROCESS . 13
© IEC 2017
Table 201.C.101 – Marking on the outside of ME EQUIPMENT or its parts . 32
Table 201.C.102 – Subclauses requiring statements in ACCOMPANYING DOCUMENTS . 32
– 4 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This consolidated version of the official IEC Standard and its amendment has been prepared
for user convenience.
IEC 60601-2-63 edition 1.1 contains the first edition (2012-09) [documents 62B/888/FDIS and
62B/898/RVD] and its amendment 1 (2017-07) [documents 62B/1049/FDIS and 62B/1058/RVD].
In this Redline version, a vertical line in the margin shows where the technical content is
modified by amendment 1. Additions are in green text, deletions are in strikethrough red text.
A separate Final version with all changes accepted is available in this publication.
© IEC 2017
International Standard IEC 60601-2-63 has been prepared by IEC subcommittee 62B:
Diagnostic imaging equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: in roman type.
– Test specifications: in italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
The committee has decided that the contents of the base publication and its amendment will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
© IEC 2017
INTRODUCTION
This particular standard has been prepared to provide, based on IEC 60601-1:2005 (third
edition), and its collaterals, a complete set of BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for DENTAL EXTRA-ORAL X-RAY EQUIPMENT. While the previously existing standards
for such equipment were dedicated to components and subsystems, this particular standard
addresses the system level of DENTAL EXTRA-ORAL X-RAY EQUIPMENT. Components and their
functions are addressed as far as necessary.
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of DENTAL EXTRA-ORAL
X-RAY EQUIPMENT
The minimum safety requirements for DENTAL INTRA-ORAL X-RAY EQUIPMENT are specified in a
separate particular standard IEC 60601-2-65 to simplify and improve the readability
Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions
of the collateral standard IEC 60601-1-3, the particular standards IEC 60601-2-28
IEC 60601-2-7, or IEC 60601-2-32 have been extracted and moved into this particular
standard.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard.
INTRODUCTION TO AMENDMENT 1
The purpose of this first amendment to IEC 60601-2-63:2012 is to introduce changes to
reference the Amendment 1 (2012) to IEC 60601-1:2005. As neither IEC 60601-2-63:2012 nor
this amendment refers to specific elements of IEC 60601-1-2, the introduction of a dated
reference to the latter document has been removed.
– 8 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-63: Particular requirements for the basic safety
and essential performance of dental extra-oral X-ray equipment
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT, hereafter also called ME EQUIPMENT. The scope includes
ME SYSTEMS containing such ME EQUIPMENT.
NOTE 1 This includes PANORAMIC equipment, CEPHALOMETRIC equipment, and equipment for dental volumetric
reconstruction (hereafter DVR) as defined in 201.3.203 below.
NOTE 2 DVR includes dental CBCT (cone beam computed tomography), which is also known with other names in
certain parts of the world, e.g. DVT (digital volumetric tomography); DVR also includes tomosynthesis.
NOTE 3 This may include the imaging of other anatomical parts (e.g. the hand) as long as required for dental
treatment (e.g. orthodontic treatment).
NOTE 4 This may include anatomical objects of interest to the ENT (ear, nose, and throat) specialist.
The scope of this standard is restricted to X-RAY EQUIPMENT where:
• the X-RAY TUBE ASSEMBLY contains the HIGH-VOLTAGE TRANSFORMER ASSEMBLY and
• the geometrical relations between the X-RAY SOURCE, the anatomical object being imaged
PATIENT, and the X-RAY IMAGE RECEPTOR, are preset in the design and cannot be
in the
arbitrarily altered by the OPERATOR during INTENDED USE.
NOTE 5 DENTAL INTRA-ORAL X-RAY EQUIPMENT is excluded from the scope of this standard.
NOTE 6 FOCAL SPOT TO IMAGE RECEPTOR DISTANCE and FOCAL SPOT to object distance are preset in the design of
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
NOTE 7 For DENTAL X-RAY EQUIPMENT not in the scope of this document because of the restriction above,
applicable clauses of IEC 60601-2-54 may be used with this document.
ME EQUIPMENT and ME SYSTEMS in the scope of IEC 60601-2-44, IEC 60601-2-54,
IEC 60601-2-45, IEC 60601-2-65 or IEC 60601-2-43 are excluded from the scope of this
RADIOTHERAPY
particular standard. The scope of this International Standard also excludes
SIMULATORS and equipment for bone or tissue absorption densitometry. Excluded from the
scope is also ME EQUIPMENT intended to be used for DENTAL RADIOSCOPY.
Within its specific scope, the clauses of this particular standard supersede and replace those
of IEC 60601-2-7, Medical electrical equipment – Particular requirements for the safety of
high-voltage generators of diagnostic X-ray generators and of IEC 60601-2-32, Medical
electrical equipment – Particular requirements for the safety of associated equipment of X-ray
equipment.
—————————
1)
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance
© IEC 2017
NOTE 8 Requirements for X-RAY GENERATORS and for ASSOCIATED EQUIPMENT, which were previously specified in
IEC 60601-2-7 and IEC 60601-2-32, have been included in either IEC 60601-1:2005 (Ed3) or this particular
rd
standard. Therefore IEC 60601-2-7 and IEC 60601-2-32 are not part of the IEC 60601-1 3 edition scheme for
DENTAL EXTRA-ORAL X-RAY EQUIPMENT.
All requirements addressing integrated X-RAY TUBE ASSEMBLIES are covered by this particular
standard. Therefore IEC 60601-2-28 does not apply to ME EQUIPMENT in the scope of this
International Standard with the exception of X-RAY TUBE ASSEMBLIES that are replaceable in
the field.
NOTE 9 Requirements particular to DENTAL X-RAY-EQUIPMENT which were included in previous editions of the
collateral standard IEC 60601-1-3 or the particular standard IEC 60601-2-28 have been extracted and moved into
this particular standard.
NOTE 10 For X-RAY EQUIPMENT in the scope of this particular standard X-RAY TUBE ASSEMBLIES are X-RAY
MONOBLOCK ASSEMBLIES.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for ME EQUIPMENT for EXTRA-ORAL DENTAL RADIOGRAPHY.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-3 apply as modified in Clause 202 and 203 respectively.
2) ) )
3 4
IEC 60601-1-8, IEC 60601-1-10 , IEC 60601-1-11 and IEC 60601-1-12 do not apply. All
other published collateral standards in the IEC 60601-1 series apply as published.
NOTE OPERATORS of DENTAL EXTRA-ORAL X-RAY EQUIPMENT are used to audible signals as required in this
particular standard rather than to the concepts of IEC 60601-1-8. Therefore IEC 60601-1-8 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard or collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
—————————
)
IEC 60601-1-10, Medical electrical equipment – Part 1-10: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for the development of physiologic closed-loop controllers
)
IEC 60601-1-11, Medical electrical equipment – Part 1-11: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment
4)
IEC 60601-1-12, Medical electrical equipment – Part 1-12: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems intended for use in the emergency medical services environment
– 10 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 39.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic compatibility –
Requirements and tests
IEC 60601-1-3:2008, Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral standard: Radiation protection in diagnostic X-
ray equipment
IEC 60601-1-3:2008/AMD1:2013
Addition:
© IEC 2017
IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis –
Characteristics of focal spots
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-2-29:2008, Medical electrical equipment – Part 2-29: Particular requirements for
the basic safety and essential performance of radiotherapy simulators
IEC 60601-2-54:2009, Medical electrical equipment – Part 2-54: Particular requirements for
the basic safety and essential performance of X-ray equipment for radiography and
radioscopy
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC/PAS 61910-1:2007, Medical electrical equipment – Radiation dose documentation –
Part 1: Equipment for radiography and radioscopy
201.3 Terminology and definitions
Amendment:
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1:2005/AMD1:2012, its applicable collateral standards, IEC/TR 60788:2004 and
the following apply:
NOTE An index of defined terms is found beginning on page 42.
Addition:
201.3.201
CEPHALOMETRIC
related to PROJECTION RADIOGRAPHY of the whole dento-maxillo-facial anatomy, whereas the
projection geometry is such to minimize geometrical image distortions
Note 1 to entry: This is usually achieved by setting a sufficiently large source-to-object-distance and source-to-
detector-distance.
Note 2 to entry: Another term often used for CEPHALOMETRIC RADIOGRAPHY is teleradiography.
201.3.202
DENTAL
related to structures in the dento-maxillo-facial district of the PATIENT, including dentition
201.3.203
*DENTAL VOLUMETRIC RECONSTRUCTION
DVR
reconstruction of the 3-dimensional attenuation distribution of the whole or part of the
irradiated volume from a series of 2-dimensional projections produced by an X-RAY BEAM on
an X-RAY IMAGE RECEPTOR moving around the head of the PATIENT
201.3.204
DOSE AREA PRODUCT
product of the area of the cross-section of an X-RAY BEAM and the averaged AIR KERMA over
that cross-section. The unit is the gray square metre (Gy⋅m ).
[SOURCE: IEC 60601-2-54:2009, 201.3.203]
– 12 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
201.3.205
ELECTRONIC X-RAY IMAGE RECEPTOR
X-RAY IMAGE RECEPTOR comprising an electrically-powered conversion method
201.3.206
EXTRA-ORAL
related to DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located outside the oral
cavity
201.3.207
INTERLOCK
means preventing the start or the continued operation of ME EQUIPMENT unless certain
predetermined conditions prevail
[SOURCE:IEC 60601-2-54:2009, 201.3.207]
201.3.208
INTRA-ORAL
related to DENTAL RADIOGRAPHY where the X-RAY IMAGE RECEPTOR is located, wholly or
partially, inside the oral cavity
201.3.209
NOMINAL SHORTEST IRRADIATION TIME
shortest LOADING TIME for which a required constancy of the controlled RADIATION QUANTIY is
maintained
Note 1 to entry: The IRRADIATION TIME is controlled by a HIGH-VOLTAGE GENERATOR with AUTOMATIC CONTROL
SYSTEMS.
[SOURCE: IEC 60601-2-54:2009, 201.3.208]
201.3.210
ONE-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers an unrectified
output voltage, or rectified output voltage with one peak during each cycle of the supply
[SOURCE: IEC 60601-2-65:2012, 201.3.208
201.3.211
* PANORAMIC
related to DENTAL RADIOGRAPHY, produced by the coordinated motion of a scanning fan-shaped
X-RAY BEAM, oriented parallel to the cranio-caudal axis of the PATIENT, and an X-RAY IMAGE
RECEPTOR, both rotating around the head of the PATIENT
Note 1 to entry: A tomographic layer is produced with respect to the plane perpendicular to the rotational axis.
The resulting image is a focused projection on a surface parallel to the rotational axis.
Note 2 to entry: The scanning axis is usually vertical.
201.3.212
TWO-PEAK HIGH-VOLTAGE GENERATOR
HIGH-VOLTAGE GENERATOR for operation on a single-phase supply that delivers a rectified
output voltage with two peaks during each cycle of the supply
[SOURCE: IEC 60601-2-65:2012, 201.3.209]
201.3.213
X-RAY MONOBLOCK ASSEMBLY
X-RAY TUBE ASSEMBLY containing the HIGH-VOLTAGE TRANSFORMER ASSEMBLY
Note 1 to entry: The term X-RAY MONOBLOCK ASSEMBLY excludes the BEAM LIMITING DEVICE.
© IEC 2017
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.3 ESSENTIAL PERFORMANCE
Addition:
201.4.3.101 Additional potential ESSENTIAL PERFORMANCE requirements
The list in Table 201.101 is a list of potential ESSENTIAL PERFORMANCE to be considered by the
MANUFACTURER in the RISK MANAGEMENT PROCESS.
NOTE Subclause 203.6.4.3.102 (Accuracy of LOADING FACTORS) specifies a limitation in applying subclause
203.6.4.3.102.3 (Accuracy of X-RAY TUBE VOLTAGE) and 203.6.4.3.102.4 (Accuracy of X-RAY TUBE CURRENT). This
limitation is also valid for the ESSENTIAL PERFORMANCE list.
Table 201.101 – List of potential ESSENTIAL PERFORMANCE to be considered by
MANUFACTURER in the RISK MANAGEMENT PROCESS
Requirement Subclause
Accuracy of LOADING FACTORS 203.6.4.3.102
Reproducibility of the RADIATION output 203.6.3.2
201.4.10.1 Source of power for ME EQUIPMENT
Addition:
201.4.10.1.101 Connection to SUPPLY MAINS
ME EQUIPMENT shall be PERMANENTLY INSTALLED unless the INTENDED USE requires it to be
MOBILE.
201.4.10.2 SUPPLY MAINS for ME EQUIPMENT and ME SYSTEMS
Addition:
The internal impedance of a SUPPLY MAINS is to be considered sufficiently low for the operation
of ME EQUIPMENT if the value of the APPARENT RESISTANCE OF SUPPLY MAINS does not exceed
the value specified in the ACCOMPANYING DOCUMENTS.
ME EQUIPMENT is considered to comply with the requirements of this standard only if its
specified NOMINAL ELECTRIC POWER can be demonstrated at a resistance of supply mains
having a value not less than the APPARENT RESISTANCE OF SUPPLY MAINS specified by the
MANUFACTURER in the ACCOMPANYING DOCUMENTS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and by functional test.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
– 14 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
201.7.2.7 Electrical input power from the SUPPLY MAINS
Addition:
ME EQUIPMENT that is specified to be PERMANENTLY
Except for items a) to c) below, for
INSTALLED, the information may be stated in the ACCOMPANYING DOCUMENTS only.
The information on the input power shall be specified in terms of combinations of
a) the rated MAINS VOLTAGE of the ME EQUIPMENT in volts; see 7.2.1 and 7.2.6 of the general
standard;
b) the number of phases; see 7.2.1 and 7.2.6 of the general standard;
c) the frequency, in hertz; see 7.2.1 and 7.2.6 of the general standard;
d) the maximum permissible value for APPARENT RESISTANCE OF SUPPLY MAINS, in ohms;
e) the characteristics of OVER-CURRENT RELEASES required in the SUPPLY MAINS.
NOTE These requirements are adapted from IEC 60601-2-7 subclause 6.1j).
201.7.2.15 Cooling conditions
Addition:
If cooling is necessary for safe operation of ME EQUIPMENT, or a subassembly thereof, the
cooling requirements shall be indicated in the ACCOMPANYING DOCUMENTS, as identified in the
RISK MANAGEMENT PROCESS.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS and the RISK
MANAGEMENT FILE.
Additional subclause:
201.7.2.101 BEAM LIMITING DEVICE
Where detachable in NORMAL USE, BEAM LIMITING DEVICES shall be provided with the following
markings:
– those required in subclause 7.2.2 of the general standard;
– serial designation or individual identification;
– ADDITIONAL FILTRATION, if the additional value is more than the equivalent of 0,2 mm Al.
Compliance is checked by inspection.
201.7.8.1 Colours of indicator lights
Addition:
The indication of X-ray related states shall be excluded from subclause 7.8 in the general
standard. 203.6.4.2 and 203.6.4.101 shall apply instead.
© IEC 2017
201.7.9 ACCOMPANYING DOCUMENTs
201.7.9.1 General
Addition:
NOTE 101 Annex C Table 201.C.102 lists the requirements of this particular standard that are additional to those
of the general standard for statements in the ACCOMPANYING DOCUMENTS.
The ACCOMPANYING DOCUMENTS shall contain quality control PROCEDURES to be performed on
the ME EQUIPMENT by the RESPONSIBLE ORGANISATION. These shall include acceptance criteria
and frequency for the tests.
Additionally for ME EQUIPMENT provided with one or more integrated ELECTRONIC X-RAY IMAGE
RECEPTORs, the ACCOMPANYING DOCUMENTS shall contain a description of the minimum
performance required of the means, used to display the images for diagnostic purpose
according to the INTENDED USE for each ELECTRONIC X-RAY IMAGE RECEPTOR used.
NOTE 102 For instance, the minimum required number of pixels and number of discernible grey levels at the
DISPLAY.
Compliance is checked by inspection of the ACCOMPANYING DOCUMENTS.
201.7.9.2 Instructions for use
201.7.9.2.1 General
Additional subclause:
201.7.9.2.1.101 LOADING FACTORS
In the instructions for use the LOADING FACTORS shall be stated as described below. The
following combinations and data shall be stated:
a) value(s) or range of X-RAY TUBE VOLTAGE settings;
b) value(s) or range of X-RAY TUBE CURRENT settings;
c) values or range of IRRADIATION TIME settings;
d) maximum X-RAY TUBE CURRENT at each X-RAY TUBE VOLTAGE setting, if different from (b);
e) maximum and minimum IRRADIATION TIME at each X-RAY TUBE VOLTAGE and X-RAY TUBE
CURRENT setting, if different from (c);
f) for ME EQUIPMENT indicating precalculated or measured CURRENT TIME PRODUCT, the lowest
CURRENT TIME PRODUCT or the combinations of LOADING FACTORS resulting in the lowest
CURRENT TIME PRODUCT;
g) if the value of the lowest CURRENT TIME PRODUCT depends upon the X-RAY TUBE VOLTAGE or
upon certain combinations of values of LOADING FACTORS, the lowest CURRENT TIME
PRODUCT may be given as a table or curve showing the dependence;
h) the NOMINAL SHORTEST IRRADIATION TIME used in AUTOMATIC EXPOSURE CONTROL systems of
ME EQUIPMENT;
i) if the NOMINAL SHORTEST IRRADIATION TIME depends upon LOADING FACTORS such as X-RAY
TUBE VOLTAGE and X-RAY TUBE CURRENT, the ranges of these LOADING FACTORS for which
the NOMINAL SHORTEST IRRADIATION TIME is valid;
j) the maximum possible range of the X-RAY TUBE VOLTAGE and/or the X-RAY TUBE CURRENT
during IRRADIATION, controlled with the AUTOMATIC CONTROL SYSTEM
NOTE these requirements are adapted from IEC 60601-2-7 subclause 6.8.2 a).
Compliance is checked by inspection of the instructions for use.
– 16 – IEC 60601-2-63:2012+AMD1:2017 CSV
© IEC 2017
201.7.9.3 Technical description
Additional subclauses:
201.7.9.3.101 Technical description for DENTAL EXTRA-ORAL X-RAY EQUIPMENT
In addition to the data required to be marked according to subclause 7.2 of the general
standard, the technical description shall specify a description of the geometric relationship of
the FOCAL SPOT, X-RAY BEAM dimensions, PATIENT position and IMAGE RECEPTION AREA, as well
as a method used to estimate the DOSE AREA PRODUCT.
Additional subclause:
201.7.9.3.102 X-RAY SOURCE ASSEMBLY
The technical description of the integrated X-RAY SOURCE ASSEMBLY shall specify the following,
in addition to the data required to be marked according to subclause 7.2 of the general
standard:
a) specification of the REFERENCE AXIS to which the TARGET ANGLE(s) and the FOCAL SPOT
characteristics of the X-RAY SOURCE ASSEMBLY refer:
b) TARGET ANGLE(s) with respect to the specified REFERENCE AXIS;
c) NOMINAL FOCAL SPOT VALUE(s) determined according to IEC 60336 for the specified
REFERENCE AXIS.
Compliance is checked by inspection of the technical description.
Additional subclause:
201.7.9.101 Requirements to the SUPPLY MAINS
The information on the RATED electrical input power for DENTAL EXTRA-ORAL X-RAY GENERATORS
shall also include:
• either the maximum permissible value for the APPARENT RESISTANCE OF SUPPLY MAINS or
other appropriate SUPPLY MAINS specifications used in a facility, and
• the characteristics of OVER-CURRENT RELEASES eventually required in the SUPPLY MAINS.
201.8 Protection against electrical HAZARDS from ME EQUIPMENT
Clause 8 of the general standard applies, except as follows:
201.8.5 Separation of parts
201.8.5.1 MEANS OF PROTECTION (MOP)
Additional subclause:
201.8.5.1.101 Additional limitation of voltage, current or energy
Provisions shall be made to prevent the appearance of an unacceptably HIGH VOLTAGE in the
MAINS PART or in any other low-voltage circuit.
NOTE This may be achieved for example by
– provision of a winding layer or a conductive screen connected to the PROTECTIVE EARTH TERMINAL between
HIGH VOLTAGE and low-voltage circuits, or
– provision of a voltage limiting device across terminals to which external devices are connected and between
which an excessive voltage might arise if the external path becomes open-circuited.
© IEC 2017
...














Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...