Standard Practice for Cone Beam Computed Tomographic (CT) Examination

SIGNIFICANCE AND USE
4.1 The purpose of this practice is to establish instruction, procedures, and equipment qualification to inspect a component for specified discontinuities. This practice is written so it can be specified on the engineering drawing, specification, or contract. This practice requires a detailed procedure delineating the technique or procedure requirements and shall be approved by the Cognizant Engineering Organization (CEO).  
4.2 The requirements in this practice shall be used when placing a CT system into NDT service and establishing a baseline of system performance measures. Monitoring the system performance over time shall be performed, including detector correction procedures, performance measurements, and system maintenance.  
4.3 For CT examinations where specified discontinuities are not known/identified/requested/available (that is, engineering analysis, engineering information only), portions of Section 7 – 9 requirements do not necessarily apply. For these types of examinations, the purchase order or contract shall specify the development of a specific procedure, approved by purchaser and supplier, to specify the requirements of this practice to be incorporated into the examination.
SCOPE
1.1 This practice establishes the minimum requirements for computed tomography (CT) examination of components using cone beam systems. The purpose of this practice is to establish instruction, procedures, and equipment qualification to inspect a component for specified discontinuities.  
1.2 This practice applies to systems with a Digital Detector Array (DDA) and an X-ray source. It does not cover the use of isotope sources.  
1.3 There are sections in this practice that may require agreement between the purchaser and the supplier, or specific direction from the cognizant engineering organization (CEO). These items should be addressed in the purchase order or the contract. Generally, the items are application-specific, performance related, or both.  
1.4 Applications for cone-beam CT are diverse. This practice is not intended to be limiting or restrictive. Refer to Guide E1441 for guidance on CT fundamentals and Guide E1672 for tradeoffs in selection of a CT system.  
1.5 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. NCRP 144 or NIST Handbook 114, or both, may be used as guides to ensure that radiographic procedures are performed so that personnel shall not receive a radiation dosage exceeding the maximum permitted by the city, state, or national codes.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Jan-2023
Current Stage
Ref Project

Buy Standard

Standard
ASTM E3375-23 - Standard Practice for Cone Beam Computed Tomographic (CT) Examination
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E3375 − 23
Standard Practice for
1
Cone Beam Computed Tomographic (CT) Examination
This standard is issued under the fixed designation E3375; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2
1.1 This practice establishes the minimum requirements for 2.1 ASTM Standards:
computed tomography (CT) examination of components using E543 Specification for Agencies Performing Nondestructive
cone beam systems. The purpose of this practice is to establish Testing
instruction, procedures, and equipment qualification to inspect E1316 Terminology for Nondestructive Examinations
a component for specified discontinuities. E1441 Guide for Computed Tomography (CT)
E1672 Guide for Computed Tomography (CT) System Se-
1.2 This practice applies to systems with a Digital Detector
lection
Array (DDA) and an X-ray source. It does not cover the use of
E1695 Test Method for Measurement of Computed Tomog-
isotope sources.
raphy (CT) System Performance
1.3 There are sections in this practice that may require
E1817 Practice for Controlling Quality of Radiological Ex-
agreement between the purchaser and the supplier, or specific
amination by Using Representative Quality Indicators
direction from the cognizant engineering organization (CEO).
(RQIs)
These items should be addressed in the purchase order or the
E2002 Practice for Determining Image Unsharpness and
contract. Generally, the items are application-specific, perfor-
Basic Spatial Resolution in Radiography and Radioscopy
mance related, or both.
E2339 Practice for Digital Imaging and Communication in
1.4 Applications for cone-beam CT are diverse. This prac- Nondestructive Evaluation (DICONDE)
E2597/E2597M Practice for Manufacturing Characterization
tice is not intended to be limiting or restrictive. Refer to Guide
E1441 for guidance on CT fundamentals and Guide E1672 for of Digital Detector Arrays
E2698 Practice for Radiographic Examination Using Digital
tradeoffs in selection of a CT system.
Detector Arrays
1.5 Units—The values stated in SI units are to be regarded
E2767 Practice for Digital Imaging and Communication in
as standard. No other units of measurement are included in this
Nondestructive Evaluation (DICONDE) for X-ray Com-
standard.
puted Tomography (CT) Test Methods
1.6 This standard does not purport to address all of the
E3147 Practice for Evaluating DICONDE Interoperability of
safety concerns associated with its use. It is the responsibility
Nondestructive Testing and Inspection Systems
of the user of this standard to establish appropriate safety,
3
2.2 ANSI Standards:
health, and environmental practices and determine the appli-
ANSI/ASNT CP 189 Standard for Qualification and Certifi-
cability of regulatory limitations prior to use. NCRP 144 or
cation of Nondestructive Testing Personnel
NIST Handbook 114, or both, may be used as guides to ensure
SNT-TC-1A Recommended Practice - Personnel Qualifica-
that radiographic procedures are performed so that personnel
tion and Certification in Nondestructive Testing
shall not receive a radiation dosage exceeding the maximum
4
2.3 ISO Standards:
permitted by the city, state, or national codes.
ISO 9712 Non-destructive Testing - Qualification and Cer-
1.7 This international standard was developed in accor-
tification of NDT Personnel
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Development of International Standards, Guides and Recom-
2
mendations issued by the World Trade Organization Technical For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Barriers to Trade (TBT) Committee.
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
1 3
This practice is under the jurisdiction of ASTM Committee E07 on Nonde- Available from American National Standards Institute (ANSI), 25 W. 43rd St.,
structive Testing and is the direct responsibility of Subcommittee E07.01 on 4th Floor, New York, NY 10036, http://www.ansi.org.
4
Radiology (X and Gamma) Method. Available from International Organization for Standardization (ISO), ISO
Current edition approved Feb. 1, 2023. Published February 2023. DOI: 10.1520/ Central Secretariat, Chemin de Blandonnet 8, CP 401, 1214 Vernier, Geneva,
E3375-23. Switzerland, https://www.iso.org.