2021-02.09 -JO - positive assessment for MDD (EN IEC 60601-2-83:2020 - PR= 64477) and the comments of the HAS consultant for MDR was resolved by the TC for EN IEC 60601-2-83:2020/prA11 - PR= 72969
This A11 introduces a revised Annex ZA & Annex ZZ  to EN IEC 60601-2-83:2020 in view of offering it for citation in OJEU under MDD (93/42/EEC) and MDR (2017/745)

  • Amendment
    10 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2021-02.09 -JO - positive assessment for MDD (EN IEC 60601-2-83:2020 - PR= 64477) and the comments of the HAS consultant for MDR was resolved by the TC for EN IEC 60601-2-83:2020/prA11 - PR= 72969
This A11 introduces a revised Annex ZA & Annex ZZ  to EN IEC 60601-2-83:2020 in view of offering it for citation in OJEU under MDD (93/42/EEC) and MDR (2017/745)

  • Amendment
    10 pages
    English language
    sale 10% off
    e-Library read for
    1 day

EN-IEC 60601-2-66 applies to the BASIC SAFETY of HEARING AIDS and HEARING AID SYSTEMS, hereafter also referred to as ME EQUIPMENT or ME SYSTEM. If a clause or subclause is specifically intended to be applicable to HEARING AIDS only, or to HEARING AID SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to HEARING AIDS and to HEARING AID SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of HEARING AIDS or HEARING AID SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 201.7.9.2 and 201.9.6. ACCESSORIES to HEARING AIDS in the HOME HEALTHCARE ENVIRONMENT (e.g. remote control units, audio streamers, battery chargers, power supplies) can be tested according to the applicable standard, IEC 60065, IEC 60950-1, IEC 62368-1 or other applicable IEC safety standards. Alternatively, the general standard may be applied. HEARING AIDS do not have a MAINS PART intended for connection to AC SUPPLY MAINS. The connection to the SUPPLY MAINS of a HEARING AID SYSTEM is covered by power supply, charger or other types of ACCESSORIES. ACCESSORIES with FUNCTIONAL CONNECTION to a HEARING AID may form a HEARING AID SYSTEM. HEARING AID related ACCESSORIES that are not physically connected to the HEARING AID during NORMAL USE are not considered to be APPLIED PART, because they do not directly contribute to the INTENDED USE of the HEARING AID. Wireless programming interfaces are covered by the applicable standard IEC 60065, IEC 60950-1, IEC 62368-1 or other applicable IEC safety standards. Alternatively, the general standard may be applied. Programming interfaces with wired connection to the HEARING AID are covered by the general standard.

  • Standard
    64 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60601-2-66:2019 is available as IEC 60601-2-66:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-66:2019 applies to the BASIC SAFETY of HEARING AIDS and HEARING AID SYSTEMS, hereafter also referred to as ME EQUIPMENT or ME SYSTEM. IEC 60601-2-66:2019 cancels and replaces the second edition published in 2015. It constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) revision of the definition about ESSENTIAL PERFORMANCE; b) revision of the application of IEC 60601-1-2:2014 for electromagnetic disturbances; c) correction of the used voltage for HEARING AIDS from 1,6 V to 4,5 V; d) correction of the drop test level from 1,5 m to 1,0 m; e) correction of the wording of IEC 60601-2-66:2015.

  • Standard
    64 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60601-2-83:2019 is applicable to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HOME LIGHT THERAPY EQUIPMENT, intended for use in the HOME HEALTHCARE ENVIRONMENT. HOME LIGHT THERAPY EQUIPMENT is typically used by a LAY OPERATOR. The scope of this document includes all light sources except laser.

  • Standard
    34 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60118-13:2019 covers the relevant EMC phenomena for hearing aids. Hearing aid immunity to high frequency fields originating from digital wireless devices such as mobile phones was identified as one of the most relevant EMC phenomena impacting hearing aids. IEC 60118-13:2019 cancels and replaces the fourth edition published in 2016 and constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) it introduces a new measurement method and set of EMC requirements for hearing aids immunity to mobile digital wireless devices; b) generic EMC requirements for hearing aids are no longer included – should be covered by other standards as appropriate.

  • Standard
    25 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2020-03-17-JO- BT decision BT164/DG11529/DV to remove the link to the MDD (93/42/EEC)
Publication on hold due to negative assessment

  • Amendment
    8 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This particular standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
ELECTROMYOGRAPHS and EVOKED RESPONSE EQUIPMENT, hereafter referred to as ME EQUIPMENT.
NOTE Myofeedback equipment, where the capturing of muscle contraction is based on electromyography, is
within the scope of this particular standard.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.

  • Standard
    33 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 82304-1:2016 applies to the safety and security of health software products designed to operate on general computing platforms and intended to be placed on the market without dedicated hardware, and its primary focus is on the requirements for manufacturers. It covers the entire lifecycle including design, development, validation, installation, maintenance, and disposal of health software products.

  • Standard
    30 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This part of ISO 80369 specifies the functional requirements for SMALL-BORE CONNECTORS intended to be used for CONNECTIONS of MEDICAL DEVICES and related ACCESSORIES. This part of ISO 80369 specifies functional requirements for the essential performance of SMALL-BORE CONNECTORS. This part of ISO 80369 does not specify the functional requirements for the MEDICAL DEVICES or  ACCESSORIES that use these CONNECTORS. Such requirements are given in particular International Standards for specific MEDICAL DEVICES or ACCESSORIES.

  • Standard
    38 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of LIGHT
ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of PATIENTS.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively
to the manufacture and some installation aspects of LIGHT ION BEAM ME EQUIPMENT
– intended for RADIOTHERAPY in human medical practice, including those in which the
selection and DISPLAY of operating parameters can be controlled automatically by
PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS),
– that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in
the range 10 MeV/n to 500 MeV/n,
and
– intended to be
– for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED
PERSONS by OPERATORS having the required skills for a particular medical application,
for particular SPECIFIED clinical purposes maintained in accordance with the
recommendations given in the INSTRUCTIONS FOR USE,
– subject to regular quality assurance performance and calibration checks by a QUALIFIED
PERSON.
NOTE 1 In this particular standard, all references to installation refer to installation in the RESPONSIBLE
ORGANIZATION’S premises.
NOTE 2 In this particular standard, all references to ABSORBED DOSE refer to ABSORBED DOSE in water.
NOTE 3 Information regarding x-ray image guidance can be found in IEC 60601-2-68 (under development).
NOTE 4 IEC 61217 gives guidance on the designation of ME EQUIPMENT movements, the marking of scales, their
zero positions and the direction of movement with increasing value (see 201.7.4.101).

  • Standard
    58 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This particular standard applies to BASIC SAFETY and ESSENTIAL PERFORMANCE of equipment for
EXTRACORPOREALLY INDUCED LITHOTRIPSY as defined in 201.3.206 including equipment for
other medical applications of therapeutic extracorporeally induced focused PRESSURE PULSES,
hereafter referred to as ME EQUIPMENT. The applicability of this particular standard is limited to
components directly involved in the LITHOTRIPSY treatment, such as, but not limited to, the
generator of the PRESSURE PULSE, PATIENT support device, and their interactions with imaging
and monitoring devices. Other devices, such as PATIENT treatment planning computers, X-ray
and ultrasonic devices, are excluded from this standard, because they are treated in other
applicable IEC standards.
This particular standard does not apply to:
– ULTRASOUND PHYSIOTHERAPY EQUIPMENT intended to be used for physiotherapy;
– ULTRASOUND equipment intended to be used for high intensity therapeutic ULTRASOUND
(HITU) and other therapy equipment as described in Annex AA;

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60601-2-57:2011 applies to basic safety and essential performance of equipment incorporating one or more sources of optical radiation in the wavelength range 200 nm to 3 000 nm, with the exception of laser radiation, and intended to create non-visual photo-biological effects in humans or animals for therapeutic, diagnostic, monitoring, cosmetic/aesthetic or veterinary applications; hereafter referred to as light source equipment. IEC 60601-2-57:2011 does not apply to equipment for sun tanning, for ophthalmic instruments or for infant phototherapy. Light source equipment may consist of a single or multiple sources of optical radiation, with or without power supply, or may be incorporated into a complex system that includes optical, electricalor mechanical systems or sources of other radiation.

  • Standard
    40 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Defines the concept of permanent filtration in X-ray tube assemblies for medical diagnosis and radiotherapy and describes a method for its determination. It contains requirements for statements of compliance for accompanying documents and for markings on X-ray tube assemblies. Methods are given to determine the permanent filtration in an X-ray tube assembly with an accuracy that is sufficient to enable the appropriate additional filtration to be provided in order to attain the desired total filtration.

  • Standard
    12 pages
    English language
    sale 10% off
    e-Library read for
    1 day

D127/107: Project under Benchemark 6 * Extension of target date approved until 2007-09-12 (no more extension allowed) * TC 62 decision on 2006-10-24: prEN 50447 deleted from work programme due to lack of sustained interest from stakeholders * D129/C066: Cancellation of project confirmed * Standstill released

  • Draft
    15 pages
    English language
    sale 10% off
    e-Library read for
    1 day

D127/107: Project under Benchemark 6 * Extension of target date approved until 2007-09-12 (no more extension allowed) * TC 62 decision on 2006-10-24: prEN 50447 deleted from work programme due to lack of sustained interest from stakeholders

  • Draft
    15 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60601-2-75:2017 applies to the basic safety and essential performance of photodynamic therapy and photodynamic diagnosis equipment. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. This document applies to photodynamic therapy and photodynamic diagnosis equipment used for compensation or alleviation of disease, injury or disability. In the case of combined equipment (e.g. equipment additionally provided with a function or an applied part for the target area), such equipment also comply with any particular standard specifying safety requirements for the additional function. This particular standard does not apply to: - light therapy equipment intended for use in photothermal ablation, coagulation, and hyperthermia; - low-level laser therapy equipment not intended for use with a photosensitizer; - illumination equipment intended for use in observation, monitoring, and diagnosis, not intended for use with a photosensitizer. This document is to be used with IEC 60601-1:2005 and its amendment 1.

  • Standard
    32 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Establishes particular requirements for the safety of electromyographs and evoked response equipment as defined in clause 2 of this standard.

  • Standard
    33 pages
    English language
    sale 10% off
    e-Library read for
    1 day

NEW!IEC 60601-2-2:2017 is available as IEC 60601-2-2:2017 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-2:2017 applies to the basic safety and essential performance of HF surgical equipment and HF surgical accessories. HF surgical equipment having a rated output power not exceeding 50 W (for example for micro-coagulation, or for use in dentistry or ophthalmology) is exempt from certain of the requirements of this particular standard. These exemptions are indicated in the relevant requirements. The object of this particular standard is to establish particular basic safety and essential performance requirements for HF surgical equipment and HF surgical accessories. This sixth edition cancels and replaces the fifth edition published in 2009. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: - refinement and additions to the defined terms; - additional separation of the requirements for HF surgical equipment and HF surgical accessories; - a new requirement for adult neutral electrodes to be contact quality monitoring neutral electrodes; - new requirements for devices that have or use a high current mode.

  • Standard
    93 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies requirements for the process by which a programmable electrical medical system is designed. Serves as the basis of requirements of Particular Standards, including serving as a guide to safety requirements for the purpose of reducing and managing risk.This standard covers requirement specification, architecture, detailed design and implementation software development, modification, verification and validation, marking and accompanying documents.

  • Standard
    34 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies requirements and tests for the safety of ultrasonic physiotherapy equipment. The aim of this second edition is to bring this particular standard up to date with reference to publications EN 60601-1 (1988) including the amendments 1 (1991) and 2 (1995), EN 60601-1-2 (1993) and EN 61689 (1996).

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Establishes particular requirements for the safety of endoscopic equipment and enables parts of endoscopic equipment to be tested together or individually.

  • Standard
    29 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies the particular safety requirements for electrocardiographs, intended for the production of detachable electrocardiograms for diagnostic purposes. Also applies to vectorcardiographs and equipment for stress testing.

  • Standard
    33 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Establishes requirements to be complied with by manufacturers in the design and construction of electron accelerators for use in radiotherapy and defines type tests and site tests. Places limits on the degradation of equipment performance beyond which it can be presumed that a fault condition exists and where an interlock then operates to prevent continued operation of the equipment.

  • Standard
    69 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 62467-1:2009 specifies the performance and some related constructional requirements of well-type ionization chambers and associated measurement apparatus, intended for the determination of a quantity, such as air kerma strength or reference air kerma rate in photon radiation fields or absorbed dose to water at a depth, in photon and beta radiation fields used in brachytherapy, after appropriate calibration for a given type of source. IEC 62467-1:2009 covers the techniques for the quantification of the quantity appropriate for the brachytherapy source under consideration. This quantity may be air kerma strength or reference air kerma rate at 1 m, or absorbed dose to water at a depth (e.g. 2 mm or 5 mm). Measurement of these quantities may be accomplished by a variety of well-type ionization chambers or systems currently available for this purpose. This standard applies to products intended for low dose rate, high dose rate, intravascular, both photon and beta, brachytherapy measurements. It does not apply to instruments for nuclear medicine applications. The application of the standard is limited to instruments that incorporate well-type ionization chambers as detectors. The intended use is the measurement of the output of radioactive, encapsulated sources for intracavitary (insertion into body cavities) or interstitial (insertion into body tissue) applications. The object of IEC 62467-1:2009 is a) to establish requirements for a satisfactory level of performance for well-type chamber systems, and b) to standardize the methods for the determination of compliance with this level of performance. IEC 62467-1:2009 is not concerned with the safety aspects of well-type chamber systems. The well-type chamber systems covered by this standard are not intended for use in patient environment. The electrical safety of well-type chamber systems is covered in IEC 61010-1. The operation of the electrometer measuring system is covered in IEC 60731.

  • Standard
    28 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Applies to the safety of equipment for extracorporeally induced lithotripsy, i.e. device for treatment with extracorporally generated pressure pulses. The applicability of this standard is limited to components directly involved in the lithotripsy treatment, such as, but not limited to, the generator of the pressure pulse, patient support device, and their interactions with imaging and monitoring devices.

  • Standard
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60601-2-66:2015 applies to the basic safety of hearing instruments and hearing instrument systems, hereafter also referred to as ME equipment or ME system. This second edition cancels and replaces the first edition published in 2012. It constitutes a technical revision to adapt IEC 60601-2-66:2012 to the technical corrections introduced by Amendment 1 (2012) to IEC 60601-1:2005, as well as to clarify and correct the wording of this particular standard and to implement minor changes requested by interested parties.

  • Standard
    61 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60601-2-64:2014 applies to the basic safety and essential performance of Light Ion Beam Medical Electrical equipment, hereafter referred to as ME equipment, used for treatment of patients. This particular standard, with the inclusion of type tests and site tests, applies respectively to the manufacture and some installation aspects of ME equipment intended for radiotherapy in human medical practice, including those in which the selection and display of operating parameters can be controlled automatically by programmable electronic subsystems, that, in normal use, deliver a radiation beam of light ions having energy per nucleon in the range 10 MeV/n to 500 MeV/n. Key words: nuclear medicine, radiation dosimetry.

  • Standard
    58 pages
    English language
    sale 10% off
    e-Library read for
    1 day

IEC 60601-2-36:2014 applies to basic safety and essential performance of equipment for extracorporeally induced lithotripsy including equipment for other medical applications of therapeutic extracorporeally induced focused pressure pulses. The applicability of this particular standard is limited to components directly involved in the lithotripsy treatment, such as, but not limited to, the generator of the pressure pulse, patient support device, and their interactions with imaging and monitoring devices. Other devices, such as patient treatment planning computers, X-ray and ultrasonic devices, are excluded from this standard, because they are treated in other applicable IEC standards. This particular standard does not apply to ultrasound physiotherapy equipment intended to be used for physiotherapy; abd to ultrasound equipment intended to be used for high intensity therapeutic ultrasound (HITU) and other therapy equipment as described in Annex AA. This second edition cancels and replaces the first edition of IEC 60601-2-36 published in 1997. This edition constitutes a technical revision to align structurally with IEC 60601-1:2005 and its Amendment1:2012.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies the particular safety requirements for ambulatory electrocardiographic systems. Within the scope of this standard are systems of the following types: a) systems that provide continuous recording and continuous analysis of the ECG allowing full re-analysis giving essentially similar results. The systems may first record and store the ECG and analyse it later on a separate unit, or record and analyse the ECG simultaneously. The type of storage media used is irrelevant with regard to this standard; b) systems that provide continuous analysis and only partial or limited recording not allowing a full re-analysis of the ECG. The safety aspects of this standard apply to all types of systems falling in one of the above- mentioned categories.

  • Standard
    47 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies the minimum safety requirements for single patient haemodialysis, haemodiafiltration and haemofiltration equipment. These devices are intended for use either by medical staff or under the supervision of medical expertise, including haemodialysis, haemodiafiltration and haemofiltration equipment operated by the patient.

  • Standard
    29 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies the requirements for infusion pumps, infusion controllers, syringe pumps and pumps for ambulatory use. These devices are intended for use by medical staff and home patients as prescribed and medically indicated. It is the responsibility of the manufacturer to ensure that the requirements of this standard are reliably implemented.

  • Standard
    67 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This standard specifies requirements for the safety of gamma beam therapy equipment intended for radiotherapy in human medical practice and includes equipment in which the selection and display of operating parameters can be controlled by a programmable electronic system. Applies to equipment which is intended to deliver a gamma radiation beam(s) at normal treatment distances greater than 5 cm using a sealed radioactive source(s). Establishes requirements to ensure the radiation safety and enhance the electrical and mechanical safety of gamma beam therapy equipment used in human medical practice and specifies tests for demonstrating compliance with those requirements.

  • Standard
    51 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This standard concerns the safety of short-wave therapy equipment having a rated output power not exceeding 500 W. This second edition revises earlier shortcomings and deals in greater detail with inductive applicators.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Establishes particular requirements for the safety of electromyographs and evoked response equipment as defined in clause 2 of this standard.

  • Standard
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Establishes safety requirements for the design and manufacture of X-ray equipment for prolonged radioscopically guided interventional procedures. Specifies information which is to be provided with such equipment for the assistance of the user and operator in managing the radiation risk arising from these procedures which could affect patients and staff.

  • Standard
    39 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment, with special attention being paid to the avoidance of hazards due to the inflation process.

  • Standard
    49 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Specifies requirements for safety of electrically operated hospital beds. The object of this standard is to keep the safety hazards to patients, operators and the environment as low as possible, and to describe tests to verify that these requirements are attained.

  • Standard
    38 pages
    English language
    sale 10% off
    e-Library read for
    1 day

Establishes general requirements for protection against ionizing radiation in medical diagnostic X-ray equipment, in order that the dose equivalent to the patient, the operator and other staff can be kept as low as reasonably achievable.

  • Standard
    58 pages
    English language
    sale 10% off
    e-Library read for
    1 day