SIST EN IEC 61010-2-061:2021
(Main)Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-061: Particular requirements for laboratory atomic spectrometers with thermal atomization and ionization (IEC 61010-2-061:2018)
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-061: Particular requirements for laboratory atomic spectrometers with thermal atomization and ionization (IEC 61010-2-061:2018)
NEW!IEC 61010-2-061:2018 is available as IEC 61010-2-061:2018 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61010-2-061:2018 applies to electrically powered laboratory atomic spectrometers with thermal atomization. It has the status of a group safety publication in accordance with IEC Guide 104. This fourth edition cancels and replaces the third edition published in 2015. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- adaptation of changes introduced by Amendment 1 of IEC 61010-1;
- added tolerance for stability of AC voltage test equipment to Clause 6;
- added requirement for interlock systems containing electric/electronic or programmable components to Clause 15.
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte -Teil 2-061: Besondere Anforderungen an Labor-Atomspektrometer mit thermischer Atomisierung und Ionisation (IEC 61010-2-061:2018)
Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-061: Exigences particulières pour spectromètres atomiques de laboratoire avec vaporisation et ionisation thermiques (IEC 61010-2-061:2018)
NEW!IEC 61010-2-061:2018 est disponible sous forme de IEC 61010-2-061:2018 RLV qui contient la Norme internationale et sa version Redline, illustrant les modifications du contenu technique depuis l'édition précédente.
L’IEC 61010-2-061:2018 s’applique aux spectromètres atomiques de laboratoire équipés d’une alimentation électrique et d’une vaporisation thermique. Elle a le statut d'une publication groupée de sécurité conformément au Guide IEC 104. Cette quatrième édition annule et remplace la troisième édition parue en 2015. Cette édition constitue une révision technique. Cette édition inclut les modifications techniques majeures suivantes par rapport à l'édition précédente:
- adaptation des modifications introduites par l’Amendement 1 de l’IEC 61010-1;
- ajout à l’Article 6 de la tolérance pour la stabilité du matériel d’essai en tension alternative;
- ajout à l’Article 15 d’une exigence relative aux systèmes de verrouillage contenant des composants électriques/électroniques ou programmables.
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 2-061. del: Posebne zahteve za laboratorijske atomske spektrometre s termično atomizacijo in ionizacijo (IEC 61010-2-061:2018)
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-december-2021
Nadomešča:
SIST EN 61010-2-061:2015
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-061. del: Posebne zahteve za laboratorijske atomske spektrometre s
termično atomizacijo in ionizacijo (IEC 61010-2-061:2018)
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 2-061: Particular requirements for laboratory atomic spectrometers with
thermal atomization and ionization (IEC 61010-2-061:2018)
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte -Teil 2-
061: Besondere Anforderungen an Labor-Atomspektrometer mit thermischer
Atomisierung und Ionisation (IEC 61010-2-061:2018)
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-061: Exigences particulières pour spectromètres atomiques de
laboratoire avec vaporisation et ionisation thermiques (IEC 61010-2-061:2018)
Ta slovenski standard je istoveten z: EN IEC 61010-2-061:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN IEC 61010-2-061
NORME EUROPÉENNE
EUROPÄISCHE NORM
September 2021
ICS 19.080; 71.040.20 Supersedes EN 61010-2-061:2015 and all of its
amendments and corrigenda (if any)
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-061: Particular requirements
for laboratory atomic spectrometers with thermal atomization and
ionization
(IEC 61010-2-061:2018)
Règles de sécurité pour appareils électriques de mesurage, Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
de régulation et de laboratoire - Partie 2-061: Exigences Regel- und Laborgeräte -Teil 2-061: Besondere
particulières pour spectromètres atomiques de laboratoire Anforderungen an Labor-Atomspektrometer mit thermischer
avec vaporisation et ionisation thermiques Atomisierung und Ionisation
(IEC 61010-2-061:2018) (IEC 61010-2-061:2018)
This European Standard was approved by CENELEC on 2018-10-26. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 61010-2-061:2021 E
European foreword
The text of document 66/643/CDV, future edition 4 of IEC 61010-2-061, prepared by IEC/TC 66
"Safety of measuring, control and laboratory equipment" was submitted to the IEC-CENELEC parallel
vote and approved by CENELEC as EN IEC 61010-2-061:2021.
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2022-08-02
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2024-08-02
document have to be withdrawn
This document supersedes EN 61010-2-061:2015 and all of its amendments and corrigenda (if any).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a Standardization Request given to CENELEC by the
European Commission and the European Free Trade Association, and supports essential
requirements of EU Directive(s) / Regulation(s).
For the relationship with EU Directive(s) / Regulation(s), see informative Annex ZZ, which is an
integral part of EN IEC 61010-2-061:2021/A11:2021.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
Endorsement notice
The text of the International Standard IEC 61010-2-061:2018 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards
indicated:
IEC 62061 NOTE Harmonized as EN 62061
ISO 13849 (series) NOTE Harmonized as EN ISO 13849 (series)
IEC 61010-2-061 ®
Edition 1.0 2018-09
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
GROUP SAFETY PUBLICATION
PUBLICATION GROUPÉE DE SÉCURITÉ
Safety requirements for electrical equipment for measurement, control, and
laboratory use –
Part 2-061: Particular requirements for laboratory atomic spectrometers with
thermal atomization and ionization
Exigences de sécurité pour appareils électriques de mesurage, de régulation et
de laboratoire –
Partie 2-061: Exigences particulières pour spectromètres atomiques de
laboratoire avec vaporisation et ionisation thermiques
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 19.080; 71.040.20 ISBN 978-2-8322-6040-1
– 2 – IEC 61010-2-061:2018 © IEC 2018
CONTENTS
FOREWORD . 3
1 Scope and object . 5
2 Normative references. 5
3 Terms and definitions . 5
4 Tests . 6
5 Marking and documentation . 6
6 Protection against electric shock . 9
7 Protection against mechanical HAZARDS . 9
8 Resistance to mechanical stresses . 9
9 Protection against the spread of fire . 10
10 Equipment temperature limits and resistance to heat . 10
11 Protection against HAZARDS from fluids and solid foreign objects . 10
12 Protection against radiation, including laser sources, and against sonic and
ultrasonic pressure . 11
13 Protection against liberated gases and substances, explosion and implosion . 12
14 Components and subassemblies . 13
15 Protection by interlocks . 14
16 HAZARDS resulting from application. 14
17 RISK assessment . 14
Annexes . 15
Annex A (normative) Routine tests . 16
Annex L (informative) Index of defined terms . 17
Bibliography . 18
Figure 101 – Example of a GAS LOCK . 12
Table 101 – Time-temperature conditions . 9
IEC 61010-2-061:2018 © IEC 2018 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR
MEASUREMENT, CONTROL, AND LABORATORY USE –
Part 2-061: Particular requirements for laboratory atomic
spectrometers with thermal atomization and ionization
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61010-2-061 has been prepared by IEC technical committee 66:
Safety of measuring, control and laboratory equipment.
It has the status of a group safety publication in accordance with IEC Guide 104.
This fourth edition cancels and replaces the third edition published in 2015. This edition
constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous
edition:
a) adaptation of changes introduced by Amendment 1 of IEC 61010-1;
b) added tolerance for stability of AC voltage test equipment to Clause 6;
– 4 – IEC 61010-2-061:2018 © IEC 2018
c) added requirement for interlock systems containing electric/electronic or programmable
components to Clause 15.
The text of this International Standard is based on the following documents:
CDV Report on voting
66/643/CDV 66/668/RVC
Full information on the voting for the approval of this International Standard can be found in
the report on voting indicated in the above table.
This document has been drafted in accordance with the ISO/IEC Directives, Part 2.
A list of all parts of the IEC 61010 series, under the general title, Safety requirements for
electrical equipment for measurement, control, and laboratory use, may be found on the IEC
website.
This Part 2-061 is intended to be used in conjunction with IEC 61010-1. It was established on
the basis of the third edition (2010) and its Amendment 1 (2016).
This Part 2-061 supplements or modifies the corresponding clauses in IEC 61010-1 so as to
convert that publication into the IEC standard: Particular requirements for laboratory atomic
spectrometers with thermal atomization and ionization.
Where a particular subclause of Part 1 is not mentioned in this Part 2, that subclause applies
as far as is reasonable. Where this part states “addition”, “modification” or “replacement”, or
“deletion”, the relevant requirement, test specification or note in Part 1 should be adapted
accordingly.
In this standard:
1) the following print types are used:
– requirements: in roman type;
– NOTES: in small roman type;
– conformity and test: in italic type;
– terms used throughout this standard which have been defined in Clause 3: SMALL
ROMAN CAPITALS;
2) subclauses, figures, tables and notes which are additional to those in Part 1 are numbered
starting from 101. The additional annexes are lettered starting from AA and additional list
items are lettered from aa).
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under "http://webstore.iec.ch" in the data related to
the specific document. At this date, the document will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IEC 61010-2-061:2018 © IEC 2018 – 5 –
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR
MEASUREMENT, CONTROL, AND LABORATORY USE –
Part 2-061: Particular requirements for laboratory atomic
spectrometers with thermal atomization and ionization
1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1 Scope
1.1.1 Equipment included in scope
Replacement:
Replace the text, except the first paragraph, with the following new text:
This part of IEC 61010 applies to electrically powered laboratory atomic spectrometers with
thermal atomization.
NOTE 1 Examples include atomic absorption spectrometers, emission flame photometers, atomic fluorescence
spectrophotometers, inductively coupled plasma spectrometers, microwave coupled plasma spectrometers and
mass spectrometers, all with thermal atomization and ionization (including tubing and connectors which are
provided by the manufacturer for connection to external supplies).
NOTE 2 If all or part of the equipment falls within the scope of one or more other Part 2 documents of IEC 61010
as well as within the scope of this document, consideration is given to those other Part 2 documents.
1.1.2 Equipment excluded from scope
Addition:
Add, before the first paragraph, the following new text:
This document does not apply to thermal atomization detectors (flame ionization detectors)
used in gas chromatography.
2 Normative references
This clause of Part 1 is applicable.
3 Terms and definitions
This clause of Part 1 is applicable except as follows:
Addition:
Add the following new terms:
3.2.101
SPRAY CHAMBER
chamber in which droplets of sample in aerosol are allowed to separate so that the droplets of
necessary size can be passed onward to the burner, with the remainder draining to waste
– 6 – IEC 61010-2-061:2018 © IEC 2018
3.2.102
GAS LOCK
device to allow drainage of waste sample liquid, and to prevent unintentional escape of gas
from the SPRAY CHAMBER through its drain outlet
Note 1 to entry: See for example Figure 101.
3.5.101
FLASH-BACK
event during which the flame travels back through the burner with the result that the gas in
the mixing chamber is caused to ignite
4 Tests
This clause of Part 1 is applicable except as follows:
Addition:
Add the following new subclauses:
4.4.2.101 Sampling probe tip
Any system designed to withdraw a sampling probe tip after sampling has been completed
shall be overridden, so as to leave the tip in its most exposed position when a sample vessel
is removed.
Exceptions:
The withdrawal system need not be overridden if the sampling probe:
a) cannot cause a HAZARD to the OPERATOR when it is exposed;
b) is designed in such a manner that no SINGLE FAULT CONDITION can cause the tip to remain
exposed after sampling has been completed.
4.4.2.102 Failure, or partial failure, of the MAINS supply
The voltage of the power supply to the equipment from the MAINS supply shall first be reduced
to just less than 90 % of the RATED voltage, and shall then be switched off.
5 Marking and documentation
This clause of Part 1 is applicable except as follows:
Addition:
Add the following new subclause:
5.1.5.101 Gas and liquid connections
The following shall be unambiguously marked adjacent to the connector on the equipment
(see 5.2):
a) the identity of the gas or liquid;
b) the maximum permitted pr
...