Abstract

This document specifies various calibration strategies for physicochemical and chemical analytical methods and specifies the calculation of analytical results.
It defines the general context for linear calibration so that individual standards dealing with analytical methods for the examination of water quality can make reference to it.

Status
Published
Public Enquiry End Date
19-Sep-2026
Publication Date
12-Oct-2026
Technical Committee
KAV - Water quality
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
08-Oct-2026
Due Date
13-Dec-2026
Completion Date
13-Oct-2026

Buy Documents

Standard

SIST ISO 8466-1:2026 - BARVE

English language (47 pages)
Preview
Preview
e-Library read for
×1 day
Standard

ISO 8466-1:2021 - Water quality — Calibration and evaluation of analytical methods — Part 1: Linear calibration function/17/2021

Release Date:17-Nov-2021
English language (41 pages)
sale 15% off
Preview
sale 15% off
Preview

SIST ISO 8466-1:2026 is the Slovenian standard based on ISO 8466-1:2021, published by SIST. It sets out linear calibration for physicochemical and chemical analytical methods used in water quality work, including how to establish the working range, calculate analytical results, and check whether a calibration remains valid. It is intended for laboratories, analysts, and quality staff who need a common calibration framework for routine water analysis and related method validation.

What does SIST ISO 8466-1:2026 specify?

SIST ISO 8466-1:2026 specifies calibration strategies for physicochemical and chemical analytical methods and the calculation of analytical results. It also defines the general context for linear calibration so that other standards for water quality analysis can refer to it.

The document is organized into scope, terms and symbols, determination of the linear working range, calibration strategies, validity checks, and four annexes.

AnnexWhat it covers
Annex AGoodness-of-fit test according to Mandel, method standard deviation, variation coefficient, and confidence interval
Annex BExamples of linearity testing
Annex CNormative examination of the linear working range using the empirical test of curvature
Annex DWeighted regression using 1/x weighting

What are the key requirements of SIST ISO 8466-1:2026?

SIST ISO 8466-1:2026 treats calibration as a two-step process: first determine the linear working range, then calculate the calibration function. In practice, that means a laboratory has to prove where the method behaves linearly before it uses the line for reporting results.

Establish the working range before fitting the line

Clause 5 requires a preliminary working range based on the practical application, technical feasibility, and the expected linear relationship. The lower end of the range should be at least the limit of quantification, and the laboratory must be able to carry out dilution or concentration steps accurately. This matters because a range that is too wide can make results unreliable at the low end.

Test linearity with measured calibration solutions

Clause 5 calls for at least five calibration solutions with different contents and the corresponding indicated values. The contents may be equally spaced, geometric, or based on powers of ten for wide ranges. The standard allows two practical ways to judge linearity: visual inspection of the content/indicated-value diagram and calculation of point-to-point slopes. In practice, this helps the analyst see where the curve starts to bend instead of assuming the whole range is usable.

Use the slope pattern to locate the linear part

Clause 5.3.3 says to calculate slopes between consecutive points and compare them with the median slope. Random scatter around zero supports linearity; systematic drift signals a non-linear section at the beginning or end of the range. Annex C gives an alternative empirical test of curvature, and both methods are intended to lead to the same range decision. This is useful when a spreadsheet or instrument software handles the slope-based check more easily than a full curve test.

Choose one of five calibration strategies

Clause 6 describes five calibration modes: external standard for the measuring method, internal standard for the measuring method, external standard for the total procedure, internal standard for the total procedure, and standard addition. The choice depends on whether matrix effects, sample preparation effects, or analyte conversion need to be corrected.

  • External standard, measuring method - used with matrix-free standards; matrix or preparation effects are handled separately by recovery.
  • Internal standard, measuring method - used when sample preparation or matrix effects vary and can be compensated by an internal standard.
  • External standard, total procedure - used when standards must go through the same preparation as the samples, for example after derivatization.
  • Internal standard, total procedure - used when both preparation effects and matrix effects need correction.
  • Standard addition - used when each sample matrix affects the signal in its own way and the sample itself must be calibrated.

Calculate results from the regression model

Clause 6.2 states that the calibration function is obtained by simple linear regression from the calibration pairs. The document gives separate result calculations for each calibration mode, including formulas for external standard, internal standard, total procedure, and standard addition. The practical effect is that the final result is not just read from the instrument signal - it is converted through the selected calibration model and any recovery or internal-standard correction.

Check recovery and validity during routine work

Clause 6.3 and 6.4 include recovery-rate determination for analyte or internal standard, using spiked samples. Clause 7 requires the calibration to be checked each working day before measurement, and more often for large series if needed. The document gives two validity checks: a control solution or control sample within the working range, and a slope check using at least two control levels near the upper and lower range limits. This keeps the calibration tied to actual instrument performance, not just the original calibration run.

What terms does SIST ISO 8466-1:2026 define?

  • Working range - the interval between the highest and lowest content covered by calibration, with the lower limit at least at the limit of quantification.
  • One-point calibration - calibration between the origin and the indicated value for one calibration sample.
  • Two-point calibration - calibration using two calibration samples at the upper and lower limits of the working range.
  • Indicated value - the quantity value provided by a measuring instrument or measuring system.
  • Content - the quantitative expression of the concentration or fraction of a substance.
  • Total procedure - the full analytical method from sample pre-treatment to reporting of results.
  • Calibration sample - a standard solution with a matrix.
  • Responsivity - the indicated value divided by the corresponding content in the calibration sample.

Who uses SIST ISO 8466-1:2026?

SIST ISO 8466-1:2026 is used by laboratories that measure water quality using physicochemical or chemical methods. It is relevant for analysts who build calibration curves, quality managers who set validation and control rules, and buyers or technical managers who compare calibration approaches for routine analysis.

It is especially useful where matrix effects, sample preparation losses, or analyte conversion can influence results. That includes method development, method validation, recovery studies, and day-to-day verification of calibration for instrument-based water analysis.

What changed in SIST ISO 8466-1:2026 from the previous edition?

SIST ISO 8466-1:2026 cancels and replaces ISO 8466-1:1990 and is technically revised. The document states these main changes:

  • the title was modified;
  • the scope was updated to calibration for routine analysis;
  • calculation of performance characteristics was moved to informative Annex A;
  • the calibration range was extended to several orders of magnitude;
  • verification of homogeneity of variances was deleted;
  • the linearity test was modified;
  • several calibration strategies were added or described more clearly;
  • the document was editorially revised.

Which standards are used with SIST ISO 8466-1:2026?

SIST ISO 8466-1:2026 cites several related standards and documents that support parts of its calibration approach:

  • ISO 8466-2:2001 - covers non-linear second-order calibration functions and supports the quadratic comparison used in Annex A.
  • ISO 11095:1996 - provides linear calibration using reference materials.
  • ISO 11352 - gives estimation of measurement uncertainty from validation and quality control data.
  • ISO/TS 13530 - gives guidance on analytical quality control for chemical and physicochemical water analysis.
  • ISO/TS 21231:2019 - helps select a representative matrix.
  • ISO/TS 28037:2010 - covers straight-line calibration functions and uncertainty treatment.
  • ISO/IEC Guide 99:2007 - provides the metrology vocabulary used for terms such as indication.

What does the SIST ISO 8466-1:2026 document contain?

SIST ISO 8466-1:2026 contains the main linear calibration workflow, the regression equations, recovery calculations, and validity checks needed for routine water analysis. Table 1 summarizes the available calibration strategies and their intended use, which helps a laboratory choose a method that matches its matrix behavior and preparation steps.

Annex A collects statistical tools that support calibration assessment, including Mandel goodness-of-fit, method standard deviation, variation coefficient, and confidence interval. Annex B gives worked examples of linearity testing for nitrite, ammonium, copper, lead, desisopropylatrazine, and carbamazepine, showing how slope patterns are interpreted in practice. Annex C provides the empirical curvature test with examples for lead, chloride, and iron. Annex D gives weighted regression with 1/x weighting, which is useful when low-concentration performance needs better emphasis over a wide calibration range.

Relations

Effective Date
01-Nov-2026
Effective Date
21-Oct-2026
Effective Date
25-Aug-2026
Effective Date
25-Aug-2026
Effective Date
25-Aug-2026
Effective Date
25-Aug-2026
Effective Date
25-Aug-2026
Effective Date
25-Aug-2026
Effective Date
25-Aug-2026
Effective Date
25-Aug-2026
Effective Date
01-Jul-2026
Effective Date
10-Jun-2026
Effective Date
20-May-2026
Effective Date
06-May-2026
Effective Date
29-Apr-2026

Buy Documents

Standard

SIST ISO 8466-1:2026 - BARVE

English language (47 pages)
Preview
Preview
e-Library read for
×1 day
Standard

ISO 8466-1:2021 - Water quality — Calibration and evaluation of analytical methods — Part 1: Linear calibration function/17/2021

Release Date:17-Nov-2021
English language (41 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

CIS Institut d.o.o.

Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

SA Slovenia Verified

Kiwa BDA Testing

Building and construction product certification.

RVA Netherlands Verified

Kmetijski inštitut Slovenije

Agricultural Institute of Slovenia. Soil testing, plant health, agricultural product analysis.

SA Slovenia Verified

Sponsored listings

Frequently Asked Questions

SIST ISO 8466-1:2026 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Water quality - Calibration and evaluation of analytical methods - Part 1: Linear calibration function". This standard covers: This document specifies various calibration strategies for physicochemical and chemical analytical methods and specifies the calculation of analytical results. It defines the general context for linear calibration so that individual standards dealing with analytical methods for the examination of water quality can make reference to it.

This document specifies various calibration strategies for physicochemical and chemical analytical methods and specifies the calculation of analytical results. It defines the general context for linear calibration so that individual standards dealing with analytical methods for the examination of water quality can make reference to it.

SIST ISO 8466-1:2026 is classified under the following ICS (International Classification for Standards) categories: 13.060.01 - Water quality in general. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST ISO 8466-1:2026 has the following relationships with other standards: It is inter standard links to SIST ISO 8466-1:1996, SIST EN ISO 22032:2026, SIST EN 17503:2023, SIST EN ISO 17294-2:2024, SIST EN ISO 10304-4:2022, SIST EN ISO 20596-2:2023, SIST EN ISO 20236:2025, SIST EN ISO 18475:2025, SIST EN ISO 15923-1:2025, SIST EN 17892:2024, SIST EN ISO 13646:2026, SIST EN ISO 15192:2026, SIST EN ISO 18127:2026, SIST EN ISO 18724:2025, SIST EN ISO 23697-1:2026. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

SIST ISO 8466-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 working range
  • 3.2 one-point calibration
  • 3.3 two-point calibration
  • 3.4 indicated value
  • 3.5 content
  • 3.6 total procedure
  • 3.7 calibration solution
  • 3.8 calibration sample
  • 3.9 measurement procedure
  • 3.10 responsivity
  • 4 Symbols
  • 5 Determination of the linear working range and establishment of the
  • 5.1 General
  • 5.2 Preliminary choice of working range
  • 5.3 Estimation of the linear working range
  • 6 Calibration strategies
  • 6.1 General
  • 6.2 Calculation of the calibration function
  • 6.3 Calibration of the measuring method using an external standard, including
  • 6.4 Calibration of the measuring method using an internal standard, including
  • 6.5 Calibration of the total procedure using an external standard
  • 6.6 Calibration of the total procedure using an internal standard
  • 6.7 Standard addition
  • 7 Strategies for testing the validity of calibration
  • 7.1 General
  • 7.2 Testing by means of a control solution or control sample
  • 7.3 Testing the slope of the calibration line
  • Annex A
  • Annex B
  • Annex C
  • Annex D
  • Bibliography

Introduction

Calibration is a prerequisite for the quantification of analytes by means of physicochemical and chemical methods. In most cases, simple linear regression is applied because many measuring methods show a linear relationship between the indicated value and the sample content. Since the publication of ISO 8466-1 in 1990, a huge progress has been made in the field of instrumental analysis, a consequence of which is that various calibration strategies have been developed in order to make best use of the equipment. The calibration range of many analytical methods was constrained to a maximum of one order of magnitude by the theoretical statistical requirement to only apply simple linear regression if homogeneity of variances exists across the selected working range. Due to the estimation of measurement uncertainty by calculation of the confidence interval in ISO 8466-1:1990, it had been necessary to conform to the required homogeneity of variances. Meanwhile, other methods for the estimation of measurement uncertainty that are independent of calibration have been established (e.g. ISO 11352).

Calibration is always done in two steps. The first step comprises the determination of the linear range, the second step is the calculation of the calibration function. The calibration strategies that are described in this document enable the analyst to individually define the calibration effort according to specified requirements. The method that is described in ISO 8466-1:1990 remains part of the informative annex since it can still be useful for specific purposes (e.g. method validation).

SIST ISO 8466-1:2026 INTERNATIONAL STANDARD ISO 8466-1:2021(E) Water quality — Calibration and evaluation of analytical methods — Part 1:

Linear calibration function

1 Scope

This document specifies various calibration strategies for physicochemical and chemical analytical methods and specifies the calculation of analytical results.

It defines the general context for linear calibration so that individual standards dealing with analytical methods for the examination of water quality can make reference to it.

2 Normative references

There are no normative references in this document.

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminological databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

3.1 working range

interval, determined by calibration, between the highest and the lowest content, where the lowest possible limit of the working range is the limit of quantification of the analytical method

3.2 one-point calibration

calibration between the origin and the indicated value corresponding to the content in the calibration sample (3.8)

3.3 two-point calibration

calibration using two calibration samples (3.8) with different contents at the upper and at the lower working range limit

3.4 indicated value

y quantity value provided by a measuring instrument or a measuring system [7] Note 1 to entry: In accordance with definition 4.1 “indication” of ISO/IEC Guide 99 .

SIST ISO 8466-1:2026 ISO 8466-1:2021(E)

3.5 content

general term for the quantitative expression of the concentration or fraction of a substance in or of a substance mixture Note 1 to entry: For example, it is the generic term for mass concentration, molar concentration, mass fraction.

3.6 total procedure

analytical method comprising all steps from sample pre-treatment to the report of results

3.7 calibration solution

standard solution without a matrix

3.8 calibration sample

standard solution with a matrix

3.9 measurement procedure

comprises all details of a measurement including all calculations for obtaining the measurement result

3.10 responsivity

R input–output gain of a detector system, i.e. indicated value divided by the corresponding content in the calibration sample (3.8) Note 1 to entry: For a system that responds linearly to its input, there is a unique responsivity. For nonlinear systems, the responsivity is the local slope.

4 Symbols

A recovery rate for the sample s A recovery rate of the internal standard Is a intercept of the calibration line or the regression line for standard addition b slope of the calibration line or the regression line for standard addition or the coefficient of the linear term of the second-order calibration function b point-to-point slope i

b median of the slopes b m i c regression coefficient of the quadratic term of the second-order calibration function F factor (taking into account sample preparation, e.g. enrichment factor) f conversion factor (e.g. 100 for expression in %) f or f degree of freedom 1 2 i index of the calibration solutions (1, 2, …, N) m number of replicate measurements per spiking level for standard addition N number of calibration solutions

SIST ISO 8466-1:2026 ISO 8466-1:2021(E) n number of spiking levels for the method of standard addition (including unspiked sample) P confidence level R responsivity R responsivity threshold

R responsivity median m V volume of the spiking solution i ssi V volume of the measurement solution M V volume of the sample s V volume of the sub-samples su x analyte content

mean value of the contents x

x analyte content in the calibration solution e x analyte content in the spiked calibration sample eg x analyte content in the sample after calibration of the total procedure g x content of the internal standard I in the calibration solution Ie x content of the internal standard I in the calibration sample Ieg x content of the internal standard I in the sample Ig x measured content of the internal standard I in the measurement solution IM x analyte content in the calibration solution i

x analyte content in the measurement solution M x analyte content in the original sample o x analyte content in the sample s x content of the spiked sample sp x content of the spiking solution ss x spiked content in the sample z x spiked content in the sub-samples i zi y indicated value y mean value of the indicated values y

y indicated value for external calibration e y indicated value for external calibration of the total procedure eg y indicated value of the analyte in the sample g

SIST ISO 8466-1:2026 ISO 8466-1:2021(E) y indicated value of the internal standard I for calibration of the measuring method Ie y indicated value of the internal standard I for calibration of the total procedure Ieg y indicated value of the internal standard I in the sample Ig y indicated value of the calibration solution i or indicated value of the sub-sample i for standard

addition y indicated value of the (unspiked) sample s Δb difference between the slope b and the median of the slopes b im− i m

5 Determination of the linear working range and establishment of the

calibration range

5.1 General

Linearity of the analytical measuring method is tested within the practice-oriented working range in accordance with the following pattern:

a) First step: Establishment of the preliminary working range. Prepare and analyse calibration solutions (with or without internal standards, analyte-dependent but matrix-independent) over one or more orders of magnitude.

b) Second step: Test linearity and establishment of the linear working range. Calculate the linear calibration function.

5.2 Preliminary choice of working range

Each calibration starts with the selection of a preliminary working range. This is subject to:

  • the objective with respect to the practical application;
  • the technically feasible possibilities;
  • the linearity of the functional relationship between the indicated values and the contents. The working range should largely cover a wide application range for the purpose in hand. The available measuring instruments often allow the choice of a wide linear working range (sometimes over several orders of magnitude). On the other hand, it is required that the indicated values obtained near the lower limit of the working range can be at least distinguished from the indicated values of the procedure blank. A lower limit of the working range is only reasonable if it is greater than or equal to the limit of quantification of this method. Moreover, dilution and concentration steps are required to be feasible and accurate.

5.3 Estimation of the linear working range

5.3.1 General

For testing the linearity of a working range covering one or more orders of magnitude, the following procedure which is based on the first-order calibration function has proved to be effective. Prepare a minimum of five calibration solutions with different contents (x) and determine the

corresponding indicated values (y ). Subsequently, plot both the measuring points and the point-to-

point slopes in a diagram and visually estimate the linear working range (see 5.3.2).

SIST ISO 8466-1:2026 ISO 8466-1:2021(E) Here, the contents may be equidistant or be based on geometric series (e.g. 1; 2; 4; 8; …) or, when testing linear working ranges that are expected to be large, such as for ICP-MS, be based on the power of ten. Multiple measurements increase the reliability of the statement. For many measuring methods showing a linear behaviour over two to five orders of magnitude, a series of contents (x ) in the form of … 0,1; 0,2; 0,5; 1; 2; 5; 10; 20; 50; 100; … is suitable.

NOTE 1 For the recommended series of contents, the measuring points for the contents are nearly equidistant on a logarithmic scale. Instead of a series of contents of e.g. [… 0,2; 2; 20; …], the series of contents of [… 0,25; 2,5; 25; …] can also be applied.

NOTE 2 Multiple calibration solutions increase the reliability more than multiple determination.

5.3.2 Visual testing of measurement data — Testing using the x/y-diagram

The measuring points are plotted in a content/indicated value-diagram and the developing of the measuring points is visually assessed.

NOTE 1 When using a spreadsheet programme, the additional presentation of a linear regression function and a comparing second-order regression function (stating the correlation coefficient or the coefficient of determination) can be helpful.

NOTE 2 Analysis of residuals using a preliminary linear regression can give additional information.

5.3.3 Estimation of the linear range by calculating the point-to-point slope

For estimation of the linear working range calculate section-wise the slope b from the data sets of each

two consecutive measuring points.

In case of a linear calibration function, a constant slope can be expected. Since there might be a nonlinear sub-section, the median of the slope values should be used instead of the arithmetic mean value. In case of a linear working range, the section-wise calculated slopes should scatter around the median. A non-linear range at the beginning or at the end will be recognised by systematically increasing or decreasing deviations from the median of the slopes b .

m For evaluation, plot the differences (Δb = b − b ) between the slopes b and the median of the slopes i−m i m i b in a diagram.

m If the procedure is linear across the selected working range, the differences will be normally distributed around zero (vary unsystematically around zero). Emerging trends indicate a non-linear relationship between content and indicated value.

If, in addition, a tolerance range with limiting lines above and below the zero-line is entered, then the working range that is accepted to be linear can be easily estimated. The tolerance range is defined in accordance with the accuracy requirements for analysis. The linear working range ends (or starts) at the measuring point (characterized as continuous index) that is just located within the accepted scatter around the zero-line or from which a systematic trend is recognized. Annex B describes this procedure and gives examples.

NOTE 1 The tolerance range can affect measurement uncertainty. NOTE 2 Especially in the case of measuring points not being equidistantly distributed, single points can visually lead to a false estimation of linearity. The limit of the linear working range can be specified more precisely by adding further measuring points.

SIST ISO 8466-1:2026 ISO 8466-1:2021(E) As an alternative to the estimation of the linear range by determining the point-to-point slope, Annex C describes the testing of the linear working range by means of the empirical test of curvature. Both methods lead to the same result.

NOTE 3 An Excel spreadsheet for the testing of the linear working range can be downloaded from the website [10] of the Water Chemistry Society . It includes several sample data sets.

6 Calibration strategies

6.1 General

The calibration function that has been determined for a substance is only valid for the covered working range. Moreover, it depends on the operating condition of the measuring instrument and shall be tested within each series of measurements (see Clause 7). The calibration function is valid as long as the requirements for the measurement uncertainty are fulfilled.

The measurement uncertainty of analytical results comprises contributions from random and systematic deviations. The determination of the measurement uncertainty using quality assurance data is described in ISO 11352. Calibration itself contributes to the random and systematic deviations depending on the number of calibration levels applied and the number of parallel determinations, or to what extent a correction of matrix-related systematic deviations is done during calibration. Therefore, more or less complex calibration strategies can be selected according to the required measurement uncertainty for analysis. In this context, this document does not clearly specify the number of calibration levels, the number of standard additions or the number of parallel determinations. For establishing the calibration functions, five modes of operation are described:

a) calibration of the measuring method using an external standard; b) calibration of the measuring method using an internal standard; c) calibration of the total procedure using an external standard; d) calibration of the total procedure using an internal standard; e) calibration according to the method of standard addition.

The following options for working ranges and for the number of calibration levels can be applied:

- for one-point calibration, selection of a content level at the upper working range limit. Blanks shall be negligible (below the limit of detection); - if blanks are not negligible or are varying, measurement of at least one more content level near the lower working range limit and exclusion of the origin for the calculation of the calibration function; - for multi-point calibration, distribution of content levels across the calibration range; - for one-point and two-point calibration, a minimum of three measurements for each content level. The lowest content level of the calibration range shall be greater than or equal to the limit of quantification. The determination of the limit detection and of the limit of quantification is carried out in accordance with documented procedures, for example according to ISO/TS 13530. Table 1 summarizes the basic conditions and application ranges of various calibration strategies. For large working ranges, some software packages offer the possibility of weighting by means of 1/x in order to improve trueness in the lower part of the working range, if necessary. This weighting procedure is presented in Annex D, including formulae for calculation.

SIST ISO 8466-1:2026 ISO 8466-1:2021(E) Table 1 — Summary of various calibration strategies Calibration method Subclause Characteristics/basic conditions Preferred application range a) calibration of 6.3 — matrix-free calibration solution; Analytical methods for which the measuring it is known due to experience - abscissa: content values; method using from the standardization an external process or intralaboratory - ordinate: indicated values; standard, validation that the recovincluding ery rate is predominantly - regardless of the sample matrix, determination of constant.

the content value of the sample is the recovery rate determined from the indicated value of the analyte of the sample and the calibration function; - matrix effects or process-related effects are corrected by the recovery rate of spiked real or synthetic samples, if necessary b) calibration of 6.4 — matrix-free calibration solution with Analytical methods for which the measuring internal standard; the matrices influence the method using measurement results during - abscissa: ratio of the analyte content an internal preparation of different samin the calibration solution to the standard, ples to an extent that is not content of the internal standard in the including negligible and not predictacalibration solution; determination of ble the recovery rate - ordinate: ratio of the corresponding of the internal indicated value of the standard standard substance to the indicated value of the internal standard; - the content value of the sample is determined from the ratio of the indicated values of the analyte and the internal standard and the calibration function; - matrix effects are corrected by addition of the internal standard to the sample prior to sample preparation c) calibration of the 6.5 — synthetic samples or analyte-free real Analytical methods for which total procedure samples are spiked with an analyte it is known due to experience using an external in order to prepare calibration from the standardization standard samples and are subjected to the total process or intralaboratory procedure; validation that matrix-related interferences are present - abscissa: content values; which, however, have a similar impact on the indicated - ordinate: indicated values; values regarding amount and direction.

  • regardless of the sample matrix, the content value of the sample is determined from the indicated value of the sample and the calibration function;
  • the matrix effect is corrected by the recovery rate of spiked samples

SIST ISO 8466-1:2026 ISO 8466-1:2021(E) Table 1 (continued) Calibration method Subclause Characteristics/basic conditions Preferred application range d) calibration of the 6.6 — for the preparation of calibration Analytical methods for which total procedure samples, synthetic or analyte-free the matrices influence the using an internal real samples are spiked with an measurement results during standard analyte and with an internal standard preparation of different samand finally subjected to the total ples to an extent that is not procedure; negligible and not predictable.

  • abscissa: ratio of the analyte content Analytical methods for which in the calibration sample to the the analyte is converted content of the internal standard in the during sample preparation calibration sample; to a different form that is not available as a standard
  • ordinate: ratio of the corresponding substance.

indicated value of the standard substance to the indicated value of the internal standard; - matrix effects will be corrected due to the addition of the internal standard to the sample prior to sample preparation e) calibration 6.7 — the sample itself is used for the Analytical methods for which according to preparation of the calibration samples it is known that each matrix the method of by spiking it with an analyte; significantly affects the standard addition indicated value in a non-pre- - abscissa: spiked contents; dictable manner.

  • ordinate: corresponding indicated values;
  • the content value of the sample is determined by dividing the indicated value of the unspiked sample by the slope of the calibration function

...

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 working range
  • 3.2 one-point calibration
  • 3.3 two-point calibration
  • 3.4 indicated value
  • 3.5 content
  • 3.6 total procedure
  • 3.7 calibration solution
  • 3.8 calibration sample
  • 3.9 measurement procedure
  • 3.10 responsivity
  • 4 Symbols
  • 5 Determination of the linear working range and establishment of the
  • 5.1 General
  • 5.2 Preliminary choice of working range
  • 5.3 Estimation of the linear working range
  • 6 Calibration strategies
  • 6.1 General
  • 6.2 Calculation of the calibration function
  • 6.3 Calibration of the measuring method using an external standard, including
  • 6.4 Calibration of the measuring method using an internal standard, including
  • 6.5 Calibration of the total procedure using an external standard
  • 6.6 Calibration of the total procedure using an internal standard
  • 6.7 Standard addition
  • 7 Strategies for testing the validity of calibration
  • 7.1 General
  • 7.2 Testing by means of a control solution or control sample
  • 7.3 Testing the slope of the calibration line
  • Annex A
  • Annex B
  • Annex C
  • Annex D
  • Bibliography

Introduction

Calibration is a prerequisite for the quantification of analytes by means of physicochemical and chemical methods. In most cases, simple linear regression is applied because many measuring methods show a linear relationship between the indicated value and the sample content. Since the publication of ISO 8466-1 in 1990, a huge progress has been made in the field of instrumental analysis, a consequence of which is that various calibration strategies have been developed in order to make best use of the equipment. The calibration range of many analytical methods was constrained to a maximum of one order of magnitude by the theoretical statistical requirement to only apply simple linear regression if homogeneity of variances exists across the selected working range. Due to the estimation of measurement uncertainty by calculation of the confidence interval in ISO 8466-1:1990, it had been necessary to conform to the required homogeneity of variances. Meanwhile, other methods for the estimation of measurement uncertainty that are independent of calibration have been established (e.g. ISO 11352).

Calibration is always done in two steps. The first step comprises the determination of the linear range, the second step is the calculation of the calibration function. The calibration strategies that are described in this document enable the analyst to individually define the calibration effort according to specified requirements. The method that is described in ISO 8466-1:1990 remains part of the informative annex since it can still be useful for specific purposes (e.g. method validation).

INTERNATIONAL STANDARD ISO 8466-1:2021(E) Water quality — Calibration and evaluation of analytical methods — Part 1:

Linear calibration function

1 Scope

This document specifies various calibration strategies for physicochemical and chemical analytical methods and specifies the calculation of analytical results.

It defines the general context for linear calibration so that individual standards dealing with analytical methods for the examination of water quality can make reference to it.

2 Normative references

There are no normative references in this document.

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminological databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

3.1 working range

interval, determined by calibration, between the highest and the lowest content, where the lowest possible limit of the working range is the limit of quantification of the analytical method

3.2 one-point calibration

calibration between the origin and the indicated value corresponding to the content in the calibration sample (3.8)

3.3 two-point calibration

calibration using two calibration samples (3.8) with different contents at the upper and at the lower working range limit

3.4 indicated value

y quantity value provided by a measuring instrument or a measuring system [7] Note 1 to entry: In accordance with definition 4.1 “indication” of ISO/IEC Guide 99 .

ISO 8466-1:2021(E)

3.5 content

general term for the quantitative expression of the concentration or fraction of a substance in or of a substance mixture Note 1 to entry: For example, it is the generic term for mass concentration, molar concentration, mass fraction.

3.6 total procedure

analytical method comprising all steps from sample pre-treatment to the report of results

3.7 calibration solution

standard solution without a matrix

3.8 calibration sample

standard solution with a matrix

3.9 measurement procedure

comprises all details of a measurement including all calculations for obtaining the measurement result

3.10 responsivity

R input–output gain of a detector system, i.e. indicated value divided by the corresponding content in the calibration sample (3.8) Note 1 to entry: For a system that responds linearly to its input, there is a unique responsivity. For nonlinear systems, the responsivity is the local slope.

4 Symbols

A recovery rate for the sample s A recovery rate of the internal standard Is a intercept of the calibration line or the regression line for standard addition b slope of the calibration line or the regression line for standard addition or the coefficient of the linear term of the second-order calibration function b point-to-point slope i

b median of the slopes b m i c regression coefficient of the quadratic term of the second-order calibration function F factor (taking into account sample preparation, e.g. enrichment factor) f conversion factor (e.g. 100 for expression in %) f or f degree of freedom 1 2 i index of the calibration solutions (1, 2, …, N) m number of replicate measurements per spiking level for standard addition N number of calibration solutions

ISO 8466-1:2021(E) n number of spiking levels for the method of standard addition (including unspiked sample) P confidence level R responsivity R responsivity threshold

R responsivity median m V volume of the spiking solution i ssi V volume of the measurement solution M V volume of the sample s V volume of the sub-samples su x analyte content

mean value of the contents x

x analyte content in the calibration solution e x analyte content in the spiked calibration sample eg x analyte content in the sample after calibration of the total procedure g x content of the internal standard I in the calibration solution Ie x content of the internal standard I in the calibration sample Ieg x content of the internal standard I in the sample Ig x measured content of the internal standard I in the measurement solution IM x analyte content in the calibration solution i

x analyte content in the measurement solution M x analyte content in the original sample o x analyte content in the sample s x content of the spiked sample sp x content of the spiking solution ss x spiked content in the sample z x spiked content in the sub-samples i zi y indicated value y mean value of the indicated values y

y indicated value for external calibration e y indicated value for external calibration of the total procedure eg y indicated value of the analyte in the sample g

ISO 8466-1:2021(E) y indicated value of the internal standard I for calibration of the measuring method Ie y indicated value of the internal standard I for calibration of the total procedure Ieg y indicated value of the internal standard I in the sample Ig y indicated value of the calibration solution i or indicated value of the sub-sample i for standard

addition y indicated value of the (unspiked) sample s Δb difference between the slope b and the median of the slopes b im− i m

5 Determination of the linear working range and establishment of the

calibration range

5.1 General

Linearity of the analytical measuring method is tested within the practice-oriented working range in accordance with the following pattern:

a) First step: Establishment of the preliminary working range. Prepare and analyse calibration solutions (with or without internal standards, analyte-dependent but matrix-independent) over one or more orders of magnitude.

b) Second step: Test linearity and establishment of the linear working range. Calculate the linear calibration function.

5.2 Preliminary choice of working range

Each calibration starts with the selection of a preliminary working range. This is subject to:

  • the objective with respect to the practical application;
  • the technically feasible possibilities;
  • the linearity of the functional relationship between the indicated values and the contents. The working range should largely cover a wide application range for the purpose in hand. The available measuring instruments often allow the choice of a wide linear working range (sometimes over several orders of magnitude). On the other hand, it is required that the indicated values obtained near the lower limit of the working range can be at least distinguished from the indicated values of the procedure blank. A lower limit of the working range is only reasonable if it is greater than or equal to the limit of quantification of this method. Moreover, dilution and concentration steps are required to be feasible and accurate.

5.3 Estimation of the linear working range

5.3.1 General

For testing the linearity of a working range covering one or more orders of magnitude, the following procedure which is based on the first-order calibration function has proved to be effective. Prepare a minimum of five calibration solutions with different contents (x) and determine the

corresponding indicated values (y ). Subsequently, plot both the measuring points and the point-to-

point slopes in a diagram and visually estimate the linear working range (see 5.3.2).

ISO 8466-1:2021(E) Here, the contents may be equidistant or be based on geometric series (e.g. 1; 2; 4; 8; …) or, when testing linear working ranges that are expected to be large, such as for ICP-MS, be based on the power of ten. Multiple measurements increase the reliability of the statement. For many measuring methods showing a linear behaviour over two to five orders of magnitude, a series of contents (x ) in the form of … 0,1; 0,2; 0,5; 1; 2; 5; 10; 20; 50; 100; … is suitable.

NOTE 1 For the recommended series of contents, the measuring points for the contents are nearly equidistant on a logarithmic scale. Instead of a series of contents of e.g. [… 0,2; 2; 20; …], the series of contents of [… 0,25; 2,5; 25; …] can also be applied.

NOTE 2 Multiple calibration solutions increase the reliability more than multiple determination.

5.3.2 Visual testing of measurement data — Testing using the x/y-diagram

The measuring points are plotted in a content/indicated value-diagram and the developing of the measuring points is visually assessed.

NOTE 1 When using a spreadsheet programme, the additional presentation of a linear regression function and a comparing second-order regression function (stating the correlation coefficient or the coefficient of determination) can be helpful.

NOTE 2 Analysis of residuals using a preliminary linear regression can give additional information.

5.3.3 Estimation of the linear range by calculating the point-to-point slope

For estimation of the linear working range calculate section-wise the slope b from the data sets of each

two consecutive measuring points.

In case of a linear calibration function, a constant slope can be expected. Since there might be a nonlinear sub-section, the median of the slope values should be used instead of the arithmetic mean value. In case of a linear working range, the section-wise calculated slopes should scatter around the median. A non-linear range at the beginning or at the end will be recognised by systematically increasing or decreasing deviations from the median of the slopes b .

m For evaluation, plot the differences (Δb = b − b ) between the slopes b and the median of the slopes i−m i m i b in a diagram.

m If the procedure is linear across the selected working range, the differences will be normally distributed around zero (vary unsystematically around zero). Emerging trends indicate a non-linear relationship between content and indicated value.

If, in addition, a tolerance range with limiting lines above and below the zero-line is entered, then the working range that is accepted to be linear can be easily estimated. The tolerance range is defined in accordance with the accuracy requirements for analysis. The linear working range ends (or starts) at the measuring point (characterized as continuous index) that is just located within the accepted scatter around the zero-line or from which a systematic trend is recognized. Annex B describes this procedure and gives examples.

NOTE 1 The tolerance range can affect measurement uncertainty. NOTE 2 Especially in the case of measuring points not being equidistantly distributed, single points can visually lead to a false estimation of linearity. The limit of the linear working range can be specified more precisely by adding further measuring points.

ISO 8466-1:2021(E) As an alternative to the estimation of the linear range by determining the point-to-point slope, Annex C describes the testing of the linear working range by means of the empirical test of curvature. Both methods lead to the same result.

NOTE 3 An Excel spreadsheet for the testing of the linear working range can be downloaded from the website [10] of the Water Chemistry Society . It includes several sample data sets.

6 Calibration strategies

6.1 General

The calibration function that has been determined for a substance is only valid for the covered working range. Moreover, it depends on the operating condition of the measuring instrument and shall be tested within each series of measurements (see Clause 7). The calibration function is valid as long as the requirements for the measurement uncertainty are fulfilled.

The measurement uncertainty of analytical results comprises contributions from random and systematic deviations. The determination of the measurement uncertainty using quality assurance data is described in ISO 11352. Calibration itself contributes to the random and systematic deviations depending on the number of calibration levels applied and the number of parallel determinations, or to what extent a correction of matrix-related systematic deviations is done during calibration. Therefore, more or less complex calibration strategies can be selected according to the required measurement uncertainty for analysis. In this context, this document does not clearly specify the number of calibration levels, the number of standard additions or the number of parallel determinations. For establishing the calibration functions, five modes of operation are described:

a) calibration of the measuring method using an external standard; b) calibration of the measuring method using an internal standard; c) calibration of the total procedure using an external standard; d) calibration of the total procedure using an internal standard; e) calibration according to the method of standard addition.

The following options for working ranges and for the number of calibration levels can be applied:

- for one-point calibration, selection of a content level at the upper working range limit. Blanks shall be negligible (below the limit of detection); - if blanks are not negligible or are varying, measurement of at least one more content level near the lower working range limit and exclusion of the origin for the calculation of the calibration function; - for multi-point calibration, distribution of content levels across the calibration range; - for one-point and two-point calibration, a minimum of three measurements for each content level. The lowest content level of the calibration range shall be greater than or equal to the limit of quantification. The determination of the limit detection and of the limit of quantification is carried out in accordance with documented procedures, for example according to ISO/TS 13530. Table 1 summarizes the basic conditions and application ranges of various calibration strategies. For large working ranges, some software packages offer the possibility of weighting by means of 1/x in order to improve trueness in the lower part of the working range, if necessary. This weighting procedure is presented in Annex D, including formulae for calculation.

ISO 8466-1:2021(E) Table 1 — Summary of various calibration strategies Calibration method Subclause Characteristics/basic conditions Preferred application range a) calibration of 6.3 — matrix-free calibration solution; Analytical methods for which the measuring it is known due to experience - abscissa: content values; method using from the standardization an external process or intralaboratory - ordinate: indicated values; standard, validation that the recovincluding ery rate is predominantly - regardless of the sample matrix, determination of constant.

the content value of the sample is the recovery rate determined from the indicated value of the analyte of the sample and the calibration function; - matrix effects or process-related effects are corrected by the recovery rate of spiked real or synthetic samples, if necessary b) calibration of 6.4 — matrix-free calibration solution with Analytical methods for which the measuring internal standard; the matrices influence the method using measurement results during - abscissa: ratio of the analyte content an internal preparation of different samin the calibration solution to the standard, ples to an extent that is not content of the internal standard in the including negligible and not predictacalibration solution; determination of ble the recovery rate - ordinate: ratio of the corresponding of the internal indicated value of the standard standard substance to the indicated value of the internal standard; - the content value of the sample is determined from the ratio of the indicated values of the analyte and the internal standard and the calibration function; - matrix effects are corrected by addition of the internal standard to the sample prior to sample preparation c) calibration of the 6.5 — synthetic samples or analyte-free real Analytical methods for which total procedure samples are spiked with an analyte it is known due to experience using an external in order to prepare calibration from the standardization standard samples and are subjected to the total process or intralaboratory procedure; validation that matrix-related interferences are present - abscissa: content values; which, however, have a similar impact on the indicated - ordinate: indicated values; values regarding amount and direction.

  • regardless of the sample matrix, the content value of the sample is determined from the indicated value of the sample and the calibration function;
  • the matrix effect is corrected by the recovery rate of spiked samples

ISO 8466-1:2021(E) Table 1 (continued) Calibration method Subclause Characteristics/basic conditions Preferred application range d) calibration of the 6.6 — for the preparation of calibration Analytical methods for which total procedure samples, synthetic or analyte-free the matrices influence the using an internal real samples are spiked with an measurement results during standard analyte and with an internal standard preparation of different samand finally subjected to the total ples to an extent that is not procedure; negligible and not predictable.

  • abscissa: ratio of the analyte content Analytical methods for which in the calibration sample to the the analyte is converted content of the internal standard in the during sample preparation calibration sample; to a different form that is not available as a standard
  • ordinate: ratio of the corresponding substance.

indicated value of the standard substance to the indicated value of the internal standard; - matrix effects will be corrected due to the addition of the internal standard to the sample prior to sample preparation e) calibration 6.7 — the sample itself is used for the Analytical methods for which according to preparation of the calibration samples it is known that each matrix the method of by spiking it with an analyte; significantly affects the standard addition indicated value in a non-pre- - abscissa: spiked contents; dictable manner.

  • ordinate: corresponding indicated values;
  • the content value of the sample is determined by dividing the indicated value of the unspiked sample by the slope of the calibration function

...