oSIST prEN ISO 6980-1:2025
(Main)Nuclear energy - Reference beta-particle radiation - Part 1: Methods of production (ISO 6980-1:2023)
Nuclear energy - Reference beta-particle radiation - Part 1: Methods of production (ISO 6980-1:2023)
This document specifies the requirements for reference beta radiation fields produced by radioactive sources to be used for the calibration of personal and area dosemeters and dose-rate meters to be used for the determination of the quantities Hp(0,07), H'(0,07;Ω), Hp(3) and H'(3;Ω), and for the determination of their response as a function of beta particle energy and angle of incidence. The basic quantity in beta dosimetry is the absorbed-dose rate in a tissue-equivalent slab phantom. This document gives the characteristics of radionuclides that have been used to produce reference beta radiation fields, gives examples of suitable source constructions and describes methods for the measurement of the residual maximum beta particle energy and the dose equivalent rate at a depth of 0,07 mm in the International Commission on Radiation Units and Measurements (ICRU) sphere. The energy range involved lies between 0,22 MeV and 3,6 MeV maximum beta energy corresponding to 0,07 MeV to 1,2 MeV mean beta energy and the dose equivalent rates are in the range from about 10 µSv·h-1 to at least 10 Sv·h-1.. In addition, for some sources, variations of the dose equivalent rate as a function of the angle of incidence are given. However, as noted in ICRU 56[5], the ambient dose equivalent, H*(10), used for area monitoring, and the personal dose equivalent, Hp(10), as used for individual monitoring, of strongly penetrating radiation, are not appropriate quantities for any beta radiation, even that which penetrates 10 mm of tissue (Emax > 2 MeV).
This document is applicable to two series of reference beta radiation fields, from which the radiation necessary for determining the characteristics (calibration and energy and angular dependence of response) of an instrument can be selected.
Series 1 reference radiation fields are produced by radioactive sources used with beam-flattening filters designed to give uniform dose equivalent rates over a large area at a specified distance. The proposed sources of 106Ru/106Rh, 90Sr/90Y, 85Kr, 204Tl and 147Pm produce maximum dose equivalent rates of approximately 200 mSv·h–1.
Series 2 reference radiation fields are produced without the use of beam-flattening filters, which allows large area planar sources and a range of source-to-calibration plane distances to be used. Close to the sources, only relatively small areas of uniform dose rate are produced, but this series has the advantage of extending the energy and dose rate ranges beyond those of series 1. The series also include radiation fields using polymethylmethacrylate (PMMA) absorbers to reduce the maximum beta particle energy. The radionuclides used are those of series 1; these sources produce dose equivalent rates of up to 10 Sv·h–1.
Énergie nucléaire - Rayonnement bêta de référence - Partie 1: Méthodes de production (ISO 6980-1:2023)
Le présent document spécifie les exigences relatives aux champs de rayonnement bêta de référence produits par les sources radioactives à utiliser pour l'étalonnage des dosimètres de zone, des dosimètres individuels et des débitmètres de dose destinés à être utilisés pour le mesurage des grandeurs Hp(0,07), H'(0,07;Ω), Hp(3) et H'(3;Ω), et pour la détermination de leur réponse en fonction de l'énergie des particules bêta et de l'angle d'incidence. En dosimétrie bêta, la grandeur fondamentale est le débit de dose absorbée dans un fantôme-plaque équivalent tissu. Le présent document donne les caractéristiques des radionucléides qui ont été utilisés pour produire des champs de rayonnement bêta de référence; il donne également des exemples de construction de sources appropriées et décrit des méthodes de mesure de l'énergie maximale résiduelle des particules bêta et du débit d'équivalent de dose à une profondeur de 0,07 mm dans la sphère ICRU (de l’anglais «International Commission on Radiation Units and Measurements» - Commission internationale des unités et mesures radiologiques). La plage d'énergie concernée se situe entre 0,22 MeV et 3,6 MeV en énergie bêta maximale, soit 0,07 MeV à 1,2 MeV en énergie bêta moyenne, et les débits d'équivalent de dose sont compris entre 10 µSv·h–1 et au moins 10 Sv·h–1. Pour certaines sources, les variations du débit d'équivalent de dose en fonction de l'angle d'incidence sont également fournies. Comme le souligne toutefois le rapport ICRU 56[5], l'équivalent de dose ambiant, H*(10), et l'équivalent de dose individuel, Hp(10), utilisés pour la surveillance de zone et pour la surveillance individuelle, respectivement, dans le cas de rayonnements fortement pénétrants ne sont pas des grandeurs appropriées pour un rayonnement bêta, même pour un rayonnement capable de traverser une épaisseur de tissu de 10 mm (Emax > 2 MeV).
Le présent document s'applique à deux séries de champs de rayonnement bêta de référence parmi lesquels le rayonnement nécessaire pour déterminer les caractéristiques (étalonnage, dépendance énergétique et angulaire de la réponse) d'un instrument peut être sélectionné.
Les champs de rayonnement de référence de la série 1 sont produits par des sources radioactives qui sont utilisées avec des filtres égalisateurs de faisceau conçus de façon à obtenir des débits d'équivalent de dose uniformes sur une surface étendue à une distance spécifiée. Les sources de 106Ru/106Rh, 90Sr/90Y, 85Kr, 204Tl et 147Pm qui sont proposées produisent des débits d'équivalent de dose maximaux d'environ 200 mSv·h–1.
Les champs de rayonnement de référence de la série 2 sont produits sans interposition de filtres égalisateurs de faisceau, ce qui permet d'utiliser des sources planes de surface étendue et une plage de distances entre la source et le plan d'étalonnage. À proximité des sources ne sont produites que des zones de débit de dose uniforme relativement peu étendues, mais cette série présente l'avantage d'étendre les plages d'énergie et de débits de dose au-delà de celles de la série 1. Cette série inclut également des champs de rayonnement utilisant des absorbeurs en polyméthacrylate de méthyle (PMMA) pour réduire l'énergie maximale des particules bêta. Les radionucléides utilisés sont ceux de la série 1; ces sources produisent des débits d'équivalent de dose pouvant atteindre 10 Sv·h–1.
Jedrska energija - Referenčno sevanje delcev beta - 1. del: Metode izdelave (ISO 6980-1:2023)
General Information
Standards Content (Sample)
SLOVENSKI STANDARD
01-julij-2025
Jedrska energija - Referenčno sevanje delcev beta - 1. del: Metode izdelave (ISO
6980-1:2023)
Nuclear energy - Reference beta-particle radiation - Part 1: Methods of production (ISO
6980-1:2023)
Énergie nucléaire - Rayonnement bêta de référence - Partie 1: Méthodes de production
(ISO 6980-1:2023)
Ta slovenski standard je istoveten z: prEN ISO 6980-1
ICS:
17.240 Merjenje sevanja Radiation measurements
27.120.01 Jedrska energija na splošno Nuclear energy in general
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
INTERNATIONAL ISO
STANDARD 6980-1
Third edition
2023-11
Nuclear energy — Reference beta-
particle radiation —
Part 1:
Methods of production
Énergie nucléaire — Rayonnement bêta de référence —
Partie 1: Méthodes de production
Reference number
ISO 6980-1:2023(E)
ISO 6980-1:2023(E)
© ISO 2023
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
ISO 6980-1:2023(E)
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 Requirements for reference beta-particle radiation fields at the calibration distance .2
4.1 Standard test conditions . 2
4.2 Energy of the reference radiation fields . 2
4.3 Shape of the beta-particle spectrum . 2
4.4 Uniformity of the dose rate . 3
4.5 Photon contamination . 3
4.6 Variation of the beta-particle emission with time . 3
5 Radionuclides suitable for reference beta-particle radiation fields .3
6 Source characteristics and their measurement . 4
6.1 Fundamental characteristics of reference sources . 4
6.1.1 Construction of reference sources . 4
6.1.2 Measurement and/or simulation of characteristics of the reference
radiation fields . 4
6.1.3 Beta particle contamination . 13
6.1.4 Photon contamination . 13
6.2 Characteristics of the two series of reference beta-particle radiation fields .13
6.2.1 General .13
6.2.2 Series 1 reference beta-particle radiation fields .13
6.2.3 Series 2 reference beta-particle radiation fields . 14
7 Source calibration .15
Annex A (normative) Tissue substitutes .17
Annex B (normative) Reference conditions and standard test conditions .18
Annex C (informative) Characteristics of the recommended sources —Examples of source
construction .20
Bibliography .21
iii
ISO 6980-1:2023(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO document should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use
of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed
patent rights in respect thereof. As of the date of publication of this document, ISO had not received
notice of (a) patent(s) which may be required to implement this document. However, implementers are
cautioned that this may not represent the latest information, which may be obtained from the patent
database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all
such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to
the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 85, Nuclear energy, nuclear technologies,
and radiological protection, Subcommittee SC 2, Radiological protection.
This third edition of ISO 6980-1 cancels and replaces ISO 6980-1:2022, of which it constitutes a minor
revision.
The main changes are as follows:
— editorial changes throughout the document.
A list of all the parts in the ISO 6980 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
ISO 6980-1:2023(E)
Introduction
ISO 6980 series covers the production, calibration, and use of reference beta-particle radiation fields for
the calibration of dosemeters and dose-rate meters for protection purposes. This document describes
the methods of production and characterization of the reference radiation. ISO 6980-2 describes
procedures for the determination of absorbed dose rate to a reference depth of tissue from beta particle
reference radiation fields. ISO 6980-3 describes procedures for the calibration of dosemeters and dose-
rate meters and the determination of their response as a function of beta-particle energy and angle of
beta-particle incidence.
For beta particles, the calibration and the determination of the response of dosemeters and dose-rate
meters is essentially a three-step process. First, the basic field quantity, absorbed dose to tissue at
a depth of 0,07 mm (and optionally also at a depth of 3 mm) in a tissue-equivalent slab geometry is
measured at the point of test, using methods described in ISO 6980-2. Then, the appropriate operational
quantity is derived by the application of a conversion coefficient that relates the quantity measured
(reference absorbed dose) to the selected operational quantity for the selected irradiation geometry.
Finally, the reference point of the device under test is placed at the point of test for the calibration
and determination of the response of the dosemeter. Depending on the type of dosemeter under test,
the irradiation is either carried out on a phantom or free-in-air for personal and area dosemeters,
respectively. For individual and area monitoring, this document describes the methods and the
conversion coefficients to be used for the determination of the response of dosemeters and dose-rate
meters in terms of the ICRU operational quantities, i.e., directional dose equivalent, H′(0,07;Ω) and
H′(3;Ω), as well as personal dose equivalent, H (0,07) and H (3).
p p
v
INTERNATIONAL STANDARD ISO 6980-1:2023(E)
Nuclear energy — Reference beta-particle radiation —
Part 1:
Methods of production
1 Scope
This document specifies the requirements for reference beta radiation fields produced by radioactive
sources to be used for the calibration of personal and area dosemeters and dose-rate meters to be used
for the determination of the quantities H (0,07), H'(0,07;Ω), H (3) and H'(3;Ω), and for the determination
p p
of their response as a function of beta particle energy and angle of incidence. The basic quantity in beta
dosimetry is the absorbed-dose rate in a tissue-equivalent slab phantom. This document gives the
characteristics of radionuclides that have been used to produce reference beta radiation fields, gives
examples of suitable source constructions and describes methods for the measurement of the residual
maximum beta particle energy and the dose equivalent rate at a depth of 0,07 mm in the International
Commission on Radiation Units and Measurements (ICRU) sphere. The energy range involved lies
between 0,22 MeV and 3,6 MeV maximum beta energy corresponding to 0,07 MeV to 1,2 MeV mean beta
−1
-1
energy and the dose equivalent rates are in the range from about 10 µSv·h to at least 10 Svh·. In
addition, for some sources, variations of the dose equivalent rate as a function of the angle of incidence
[5]
are given. However, as noted in ICRU 56 , the ambient dose equivalent, H*(10), used for area
monitoring, and the personal dose equivalent, H (10), as used for individual monitoring, of strongly
p
penetrating radiation, are not appropriate quantities for any beta radiation, even that which penetrates
10 mm of tissue (E > 2 MeV).
max
This document is applicable to two series of reference beta radiation fields, from which the radiation
necessary for determining the characteristics (calibration and energy and angular dependence of
response) of an instrument can be selected.
Series 1 reference radiation fields are produced by radioactive sources used with beam-flattening
filters designed to give uniform dose equivalent rates over a large area at a specified distance. The
106 106 90 90 85 204 147
proposed sources of Ru/ Rh, Sr/ Y, Kr, Tl and Pm produce maximum dose equivalent
–1
rates of approximately 200 mSv·h .
Series 2 reference radiation fields are produced without the use of beam-flattening filters, which allows
large area planar sources and a range of source-to-calibration plane distances to be used. Close to the
sources, only relatively small areas of uniform dose rate are produced, but thi
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.