SIST EN ISO 13199:2013
(Main)Stationary source emissions - Determination of total volatile organic compounds (TVOCs) in waste gases from non-combustion processes - Non-dispersive infrared analyser equipped with catalytic converter (ISO 13199:2012)
Stationary source emissions - Determination of total volatile organic compounds (TVOCs) in waste gases from non-combustion processes - Non-dispersive infrared analyser equipped with catalytic converter (ISO 13199:2012)
ISO 13199:2012 specifies the principle, the essential performance criteria and quality assurance/quality control (QA/QC) procedures of an automatic method for measuring total volatile organic compound (TVOC) content in waste gases of stationary sources, using a non-dispersive infrared absorption (NDIR) analyser equipped with a catalytic converter which oxidizes VOC to carbon dioxide.
This method is suitable for the measurement of TVOC emissions from non-combustion processes. This method allows continuous monitoring with permanently installed measuring systems, as well as intermittent measurements of TVOC emissions.
The method has been tested on field operation for painting and printing processes, where TVOC concentrations in the waste gases were from about 70 mg/m3 to 600 mg/m3.
Emissionen aus stationären Quellen - Bestimmung der Summe der flüchtigen organischen Verbindungen (TVOC) in Abgasen, die nicht aus Verbrennungsprozessen stammen - Nicht-dispersives Infrarot-Messgerät mit Konverter (ISO 13199:2012)
Diese Internationale Norm legt das Prinzip, die wesentlichen Leistungskriterien und die QA/QC-Maßnahmen (Qualitätssicherung/Qualitätslenkung) für ein automatisches Verfahren zur Messung des Gehalts der Summe der flüchtigen organischen Verbindungen (TVOC) in Abgasen aus stationären Quellen fest, wobei ein Messgerät mit nicht-dispersiver Infrarotabsorption (NDIR), das mit einem Konverter zur Oxidation der VOC zu Kohlendioxid ausgestattet ist, verwendet wird.
Dieses Verfahren ist zur Messung von TVOC-Emissionen aus Prozessen geeignet, bei denen keine Ver-brennung stattfindet. Es erlaubt sowohl kontinuierliche Messungen mit fest installierten Messeinrichtungen als auch diskontinuierliche Messungen.
Das Verfahren wurde an Lackier- und Druckanlagen getestet, deren Abgase TVOC-Konzentrationen von 70 mg/m3 bis 600 mg/m3 aufwiesen.
Émissions de sources fixes - Détermination des composés organiques volatils totaux (COVT) dans les effluents gazeux des processus sans combustion - Analyseur à infrarouge non dispersif équipé d'un convertisseur catalytique (ISO 13199:2012)
L'ISO 13199:2012 spécifie le principe, les critères de performance essentiels et les procédures d'assurance qualité/contrôle qualité (AQ/CQ) d'une méthode automatique pour mesurer la teneur en composés organiques volatils totaux (COVT) dans les effluents gazeux de sources fixes, en utilisant un analyseur par rayonnement infrarouge non dispersif (NDIR) équipé d'un convertisseur catalytique qui oxyde les COV en dioxyde de carbone.
Cette méthode est appropriée au mesurage des émissions de COVT résultant de processus autres que les processus de combustion. Cette méthode permet une surveillance en continu par des systèmes de mesurage installés en permanence ainsi que des mesurages ponctuels des émissions de COVT.
Cette méthode a été testée sur site pour des processus de peinture et d'impression dans lesquels les concentrations en COVT dans les effluents gazeux pouvaient atteindre environ 70 mg/m3 à 600 mg/m3.
Emisije nepremičnih virov - Določevanje celotnih hlapnih organskih spojin (TVOC) v odpadnih plinih nezgorevalnih procesov - Nedisperzivni infrardeči analizator, opremljen s katalitskim pretvornikom (ISO 13199:2012)
Ta mednarodni standard določa načelo, osnovna merila učinkovitosti in postopke za zagotavljanje/nadzor kakovosti (QA/QC) samodejne metode za merjenje celotne hlapne organske spojine (TVOC) v odpadnih plinih nepremičnih virov, in sicer z uporabo analizatorja nedisperzne infrardeče absorpcije (NDIR), opremljenega s katalizatorjem, ki hlapno organsko spojino oksidira v ogljikov dioksid. Metoda je primerna za merjenje emisij celotne hlapne organske spojine iz nezgorevalnih procesov. Metoda omogoča neprekinjen nadzor s trajno vgrajenimi merilnimi sistemi in občasne meritve emisij celotne hlapne organske spojine. Metoda je bila preskušena pri terenskem delovanju v procesih barvanja in tiskanja, pri čemer so koncentracije celotne hlapne organske spojine v odpadnih plinih merile od približno 70 mg/m3 do 600 mg/m3.
General Information
- Status
- Published
- Publication Date
- 13-Mar-2013
- Technical Committee
- KAZ - Air quality
- Current Stage
- 6060 - National Implementation/Publication (Adopted Project)
- Start Date
- 04-Mar-2013
- Due Date
- 09-May-2013
- Completion Date
- 14-Mar-2013
Overview
EN ISO 13199:2012 / ISO 13199:2012 defines an automatic method for measuring total volatile organic compounds (TVOCs) in waste gases from non‑combustion stationary sources. The method uses a non‑dispersive infrared (NDIR) analyser equipped with a catalytic converter that oxidizes VOCs to carbon dioxide (CO2) and measures the resulting CO2 signal as a proxy for TVOC mass concentration. EN ISO 13199:2012 covers the measurement principle, essential performance criteria and the required quality assurance/quality control (QA/QC) procedures for continuous and intermittent monitoring. The standard has been field‑tested on painting and printing processes with TVOC levels roughly 70–600 mg/m3.
Key topics and requirements
- Measurement principle: Extractive NDIR detection after catalytic oxidation of VOC to CO2.
- Scope: Applicable to waste gases from non‑combustion processes (not for combustion sources).
- Performance criteria: Specifies essential performance characteristics and how to determine measurement uncertainty (see clauses on performance characteristics).
- Automatic Measuring Systems (AMS): Requirements for permanently installed continuous systems and for intermittent measurements.
- Sampling & measurement procedure: Guidance on probe installation, sampling, data collection and calculation of TVOC mass concentration.
- QA/QC procedures: Calibration, frequency of checks, operational gases and procedures for ensuring reliable results.
- Annexes: Include schematic diagrams of NDIR analysers, operational gas specifications, procedures for performance testing and example compliance assessments.
- Standards referenced: Normative references include ISO 9169:2006 (AMS performance), ISO 14956 (suitability evaluation) and ISO 20988 (measurement uncertainty).
Applications and users
EN ISO 13199:2012 is intended for:
- Environmental regulators setting monitoring requirements for solvent‑using industries.
- Plant operators and environmental managers in painting, printing, coating, degreasing and similar non‑combustion processes who need continuous emissions data or intermittent compliance checks.
- Air quality laboratories and field engineers responsible for installation, calibration and QA/QC of NDIR‑based TVOC monitors.
- Instrument manufacturers developing NDIR analysers with catalytic converters and demonstrating conformity with performance criteria.
Practical uses include continuous emissions monitoring, process control to reduce VOC losses, verification of abatement systems and compliance reporting.
Related standards
- ISO 9169:2006 - Definition and determination of AMS performance characteristics
- ISO 14956 - Evaluation of suitability by comparison with required measurement uncertainty
- ISO 20988 - Guidelines for estimating measurement uncertainty
Keywords: EN ISO 13199:2012, ISO 13199:2012, TVOC, total volatile organic compounds, NDIR analyser, catalytic converter, stationary source emissions, waste gases, air quality, QA/QC, automatic measuring system.
Frequently Asked Questions
SIST EN ISO 13199:2013 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Stationary source emissions - Determination of total volatile organic compounds (TVOCs) in waste gases from non-combustion processes - Non-dispersive infrared analyser equipped with catalytic converter (ISO 13199:2012)". This standard covers: ISO 13199:2012 specifies the principle, the essential performance criteria and quality assurance/quality control (QA/QC) procedures of an automatic method for measuring total volatile organic compound (TVOC) content in waste gases of stationary sources, using a non-dispersive infrared absorption (NDIR) analyser equipped with a catalytic converter which oxidizes VOC to carbon dioxide. This method is suitable for the measurement of TVOC emissions from non-combustion processes. This method allows continuous monitoring with permanently installed measuring systems, as well as intermittent measurements of TVOC emissions. The method has been tested on field operation for painting and printing processes, where TVOC concentrations in the waste gases were from about 70 mg/m3 to 600 mg/m3.
ISO 13199:2012 specifies the principle, the essential performance criteria and quality assurance/quality control (QA/QC) procedures of an automatic method for measuring total volatile organic compound (TVOC) content in waste gases of stationary sources, using a non-dispersive infrared absorption (NDIR) analyser equipped with a catalytic converter which oxidizes VOC to carbon dioxide. This method is suitable for the measurement of TVOC emissions from non-combustion processes. This method allows continuous monitoring with permanently installed measuring systems, as well as intermittent measurements of TVOC emissions. The method has been tested on field operation for painting and printing processes, where TVOC concentrations in the waste gases were from about 70 mg/m3 to 600 mg/m3.
SIST EN ISO 13199:2013 is classified under the following ICS (International Classification for Standards) categories: 13.040.40 - Stationary source emissions. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN ISO 13199:2013 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-april-2013
(PLVLMHQHSUHPLþQLKYLURY'RORþHYDQMHFHORWQLKKODSQLKRUJDQVNLKVSRMLQ792&
YRGSDGQLKSOLQLKQH]JRUHYDOQLKSURFHVRY1HGLVSHU]LYQLLQIUDUGHþLDQDOL]DWRU
RSUHPOMHQVNDWDOLWVNLPSUHWYRUQLNRP,62
Stationary source emissions - Determination of total volatile organic compounds
(TVOCs) in waste gases from non-combustion processes - Non-dispersive infrared
analyser equipped with catalytic converter (ISO 13199:2012)
Emissionen aus stationären Quellen - Bestimmung der Summe der flüchtigen
organischen Verbindungen (TVOC) in Abgasen, die nicht aus Verbrennungsprozessen
stammen - Nicht-dispersives Infrarot-Messgerät mit Konverter (ISO 13199:2012)
Émissions de sources fixes - Détermination des composés organiques volatils totaux
(COVT) dans les effluents gazeux des processus sans combustion - Analyseur à
infrarouge non dispersif équipé d'un convertisseur catalytique (ISO 13199:2012)
Ta slovenski standard je istoveten z: EN ISO 13199:2012
ICS:
13.040.40 (PLVLMHQHSUHPLþQLKYLURY Stationary source emissions
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN ISO 13199
NORME EUROPÉENNE
EUROPÄISCHE NORM
October 2012
ICS 13.040.40
English Version
Stationary source emissions - Determination of total volatile
organic compounds (TVOCs) in waste gases from non-
combustion processes - Non-dispersive infrared analyser
equipped with catalytic converter (ISO 13199:2012)
Émissions de sources fixes - Détermination des composés
Emissionen aus stationären Quellen - Bestimmung der
organiques volatils totaux (COVT) dans les effluents Summe der flüchtigen organischen Verbindungen (TVOCs)
gazeux des processus sans combustion - Analyseur à in Abgasen, die nicht aus Verbrennungsprozessen
infrarouge non dispersif équipé d'un convertisseur stammen - Nicht-dispersives Infrarot-Messgerät mit
catalytique (ISO 13199:2012) Konverter (ISO 13199:2012)
This European Standard was approved by CEN on 14 October 2012.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same
status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United
Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2012 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 13199:2012: E
worldwide for CEN national Members.
Contents Page
Foreword .3
Foreword
This document (EN ISO 13199:2012) has been prepared by Technical Committee ISO/TC 146 "Air quality" in
collaboration with Technical Committee CEN/TC 264 “Air quality” the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an identical
text or by endorsement, at the latest by April 2013, and conflicting national standards shall be withdrawn at the
latest by April 2013.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights.
According to the CEN/CENELEC Internal Regulations, the national standards organisations of the following
countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech
Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece,
Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.
Endorsement notice
The text of ISO 13199:2012 has been approved by CEN as a EN ISO 13199:2012 without any modification.
INTERNATIONAL ISO
STANDARD 13199
First edition
2012-10-15
Stationary source emissions —
Determination of total volatile organic
compounds (TVOCs) in waste gases
from non-combustion processes — Non-
dispersive infrared analyser equipped
with catalytic converter
Émissions de sources fixes — Détermination des composés organiques
volatils totaux (COVTs) dans les effluents gazeux des processus
sans combustion — Analyseur à infrarouge non dispersif équipé d’un
convertisseur catalytique
Reference number
ISO 13199:2012(E)
©
ISO 2012
ISO 13199:2012(E)
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or ISO’s
member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2012 – All rights reserved
ISO 13199:2012(E)
Contents Page
Foreword .iv
Introduction . v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Symbols and abbreviated terms . 3
5 Principle . 4
5.1 Method of measurement . 4
5.2 Analyser equipment . 6
5.3 Performance criteria fulfilment . 6
6 Performance criteria and determination of the performance characteristics . 6
6.1 Performance criteria . 6
6.2 Determination of the performance characteristics and measurement uncertainty . 6
7 Measurement procedure . 8
7.1 General . 8
7.2 Choice of the measuring system . 8
7.3 Sampling . 8
7.4 Data collection . 8
7.5 Calculation . 9
8 Quality assurance and quality control procedures . 9
8.1 General . 9
8.2 Frequency of checks . 9
8.3 AMS for intermittent measurements . 9
8.4 Permanently installed AMS . 11
9 Test report .13
Annex A (informative) Schematic diagrams of NDIR analysers .14
Annex B (normative) Operational gases .17
Annex C (normative) Procedures for determination of the performance characteristics during the
general performance test .19
Annex D (informative) Example of assessment of compliance of the NDIR method .23
Annex E (informative) Results of comparison tests.26
Bibliography .30
ISO 13199:2012(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International
Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 13199 was prepared by Technical Committee ISO/TC 146, Air quality, Subcommittee SC 1, Stationary
source emissions.
iv © ISO 2012 – All rights reserved
ISO 13199:2012(E)
Introduction
Volatile organic compounds (VOCs) play significant roles in atmospheric chemistry, especially the formation
of photochemical oxidants and/or ozone (O ) and suspended particulate matter (SPM), which are known to
have negative impacts on human health and biological systems. There are many areas of the world where their
atmospheric concentrations are close to or above the level of the WHO guidelines for environmental standards
in each nation. It is therefore strongly required in many nations to reduce the emission of VOC from various
anthropogenic sources.
In order to manage VOC emission from stationary emission sources, it is essential for enterprises to evaluate
the quantity of VOCs emitted from their facilities where organic solvents as well as other chemicals are used
for industrial processes like painting, printing, cleaning, and degreasing.
A technique for continuously and precisely measuring the concentration of total VOCs (TVOCs) in waste
gases emitted from ducts to atmosphere, which is easy to operate and to maintain, is very helpful for both
governments and enterprises to control and reduce VOC emissions. This measurement method, based on
the use of an NDIR analyser equipped with a catalytic converter for oxidation of TVOCs to CO , has some
advantages compared to measurement methods using flame ionization detection (FID) and FID–GC (flame
ionization detection–gas chromatography), namely:
a) high-safety operation is possible, since no flame and no hydrogen are used;
b) response factors of individual VOCs are not different from each other;
c) no interference due to oxygen is observed.
Note, however, that this method is not applicable to waste gas from combustion processes.
INTERNATIONAL STANDARD ISO 13199:2012(E)
Stationary source emissions — Determination of total volatile
organic compounds (TVOCs) in waste gases from non-
combustion processes — Non-dispersive infrared analyser
equipped with catalytic converter
1 Scope
This International Standard specifies the principle, the essential performance criteria and quality
assurance/quality control (QA/QC) procedures of an automatic method for measuring total volatile organic
compound (TVOC) content in waste gases of stationary sources, using a non-dispersive infrared absorption
(NDIR) analyser equipped with a catalytic converter which oxidizes VOC to carbon dioxide.
This method is suitable for the measurement of TVOC emissions from non-combustion processes. This
method allows continuous monitoring with permanently installed measuring systems, as well as intermittent
measurements of TVOC emissions.
The method has been tested on field operation for painting and printing processes, where TVOC concentrations
3 3
in the waste gases were from about 70 mg/m to 600 mg/m .
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced document
(including any amendments) applies.
ISO 9169:2006, Air quality — Definition and determination of performance characteristics of an automatic
measuring system
ISO 14956, Air quality — Evaluation of the suitability of a measurement procedure by comparison with a
required measurement uncertainty
ISO 20988, Air quality — Guidelines for estimating measurement uncertainty
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
automatic measuring system
AMS
measuring system interacting with the waste gas under investigation, returning an output signal proportional to
the physical unit of the measurand in unattended operation
NOTE 1 Adapted from ISO 9169:2006, 2.1.2.
NOTE 2 In the sense of this document, an AMS is a system that can be attached to a duct to continuously or intermittently
measure and record TVOC mass concentrations passing through the duct.
3.2
analyser
analytical part in an extractive or in situ AMS
[3]
[ISO 12039:2001, 3.3]
ISO 13199:2012(E)
3.3
calibration of an automatic measuring system
procedure for establishing the statistical relationship between values of the measurand indicated by the
automatic measuring system and the corresponding values given by an independent method of measurement
implemented simultaneously at the same measuring point
3.4
interference
negative or positive effect upon the response of the measuring system, due to a component of the sample that
is not the measurand
3.5
interferent
interfering substance
substance present in the air mass under investigation, other than the measurand, that affects the response
[ISO 9169:2006, 2.1.12]
3.6
lack of fit
systematic deviation, within the range of application, between the accepted value of a reference material applied
to the measuring system and the corresponding result of measurement produced by the measuring system
[ISO 9169:2006, 2.2.9]
3.7
mass concentration
concentration of a substance in a waste gas expressed as mass per volume
[3]
NOTE 1 Adapted from ISO 12039:2001, 3.10.
NOTE 2 Mass concentration is often expressed in milligrams per cubic metre (mg/m ).
3.8
measurand
particular quantity subject to measurement
[4]
[ISO/IEC Guide 98-3:2008, B.2.9]
EXAMPLE The TVOC mass concentration (mg/m ) in waste gas.
3.9
performance characteristic
one of the quantities assigned to equipment in order to define its performance
NOTE Performance characteristics can be described by values, tolerances or ranges.
3.10
period of unattended operation
maximum interval of time for which the performance characteristics remain within a predefined range without
external servicing, e.g. refill, adjustment
[ISO 9169:2006, 2.2.11]
NOTE The period of unattended operation is often called maintenance interval.
3.11
residence time
time period for the sampled gas to be transported from the inlet of the probe to the inlet of the measurement cell
2 © ISO 2012 – All rights reserved
ISO 13199:2012(E)
3.12
response time
time interval between the instant when a stimulus is subjected to a specified abrupt change and the instant when
the response reaches and remains within specified limits around its final stable value, determined as the sum of
the lag time and the rise time in the rising mode, and the sum of the lag time and the fall time in the falling mode
[ISO 9169:2006, 2.2.4]
3.13
span gas
gas or gas mixture used to adjust and check a specific point on a calibration curve
[3]
NOTE Adapted from ISO 12039:2001, 3.4.1.
EXAMPLE Normally a mixture of propane and air is used.
3.14
span point
value of the output quantity (measured signal) of the automatic measuring ssytem for the purpose of calibration,
adjustment, etc. that represents a correct measured value generated by reference material
3.15
standard uncertainty
uncertainty of the result of measurement expressed as a standard deviation
[4]
[ISO/IEC Guide 98-3:2008, 2.3.1]
NOTE The standard uncertainty of a result of measurement is an estimate of the standard deviation of the population
of all possible results of measurement which can be obtained by means of the same method of measurement for the
measurand exhibiting a unique value.
3.16
total volatile organic compounds
TVOCs
by convention, total organic compounds present with a partial pressure below their saturated vapour pressure
at ambient air pressure and temperature
NOTE Measured TVOC values (mass concentration or volume concentration) are usually referred to carbon.
3.17
uncertainty (of measurement)
parameter, associated with the result of a measurement, that characterizes the dispersion of the values that
could reasonably be attributed to the measurand
[4]
[ISO/IEC Guide 98-3:2008, 2.2.3]
3.18
zero gas
gas or gas mixture used to establish the zero point on a calibration curve within a given concentration range
[3]
[ISO 12039:2001, 3.4.2]
3.19
zero point
specified value of the output quantity (measured signal) of the AMS and which, in the absence of the measured
component, represents the zero crossing of the calibration line
4 Symbols and abbreviated terms
g
TVOC mass concentration
ISO 13199:2012(E)
γ
grand mean of measured TVOC mass concentration
−6
ϕ TVOC volume fraction ( = 10 )
e residual (lack of fit) at level i
i
k coverage factor
M molar mass of carbon ( = 12 g/mol)
C
n number of measurements
s standard deviation of level j
j
s standard deviation of repeatability
r,j
s standard deviation of reproducibility
R,j
u standard uncertainty of TVOC mass concentration
combined uncertainty of TVOC mass concentration
u γ
()
TVOC
expanded uncertainty of mass concentration
U γ
()
TVOC
V molar volume (22,4 l/mol)
m
C coefficient of variation of repeatability
V,r
C coefficient of variation of reproducibility
V,R
C coefficient of variation of the standard uncertainty
V,u
x
average of the measured values x
i
x ith measured value
i
average of the measured value at level i
x
i
value estimated by the regression line at level i
xˆ
i
AMS automatic measuring system
NDIR non-dispersive infrared absorption
QA quality assurance
QC quality control
5 Principle
5.1 Method of measurement
The measuring system consists of a sample conditioning system and the NDIR analyser for measuring CO
equipped with the converter for oxidation of TVOCs to CO as shown in Figure 1. A portion of sample gas (gas
A) passes through the converter (and the moisture removal system), and goes into the NDIR analyser, while
another portion of sample gas (gas B) passes through (the moisture removal system), and goes into the NDIR
analyser. The difference in the CO concentration between gas A and gas B is equal to the concentration of
CO which comes from TVOCs.
4 © ISO 2012 – All rights reserved
ISO 13199:2012(E)
5.2 Analyser equipment
The gas analysers use, as the measurement principle, the absorption of infrared radiation (IR) by the component
measured in characteristic wavelength ranges. The analysers operate according to the non-dispersive IR
(NDIR) method, while the selectivity of measurement is achieved by the radiation detector which is filled with
the component to be measured. Schematic diagrams of a typical NDIR analyser for measuring TVOC mass
concentration are given in Figures A.1, A.2, and A.3.
5.3 Performance criteria fulfilment
5.3.1 General. The automatic measuring system (AMS) based on the NDIR method shall comply with
the performance criteria specified in Table 1. The associated performance characteristics are determined as
specified in 5.3.2 to 5.3.4.
5.3.2 General performance test. The manufacturer of the measuring system shall demonstrate in a general
performance test that the relevant performance criteria listed in Table 1 are fulfilled by the instrument type. The
procedure of this general performance test shall comply with the relevant standards.
5.3.3 Ongoing quality assurance (QA) and quality control (QC) in the laboratory. The user of AMS shall
demonstrate during regular laboratory tests conducted within the ongoing QC programme that the relevant
performance criteria listed in Table 1 are fulfilled for the specific AMS.
5.3.4 Quality assurance during operation in the field. The user of AMS shall check during field operation
that the relevant performance criteria listed in Table 1 are fulfilled.
6 Performance criteria and determination of the performance characteristics
6.1 Performance criteria
Table 1 gives an overview of the relevant performance characteristics and performance criteria of the analyser
and measurement system to be evaluated at three levels, during a general performance test, by means of
ongoing QA/QC in the laboratory and during field operation. In the rightmost column, values included in the
calculation of the expanded uncertainty are indicated.
6.2 Determination of the performance characteristics and measurement uncertainty
6.2.1 Performance test
The performance characteristics of the AMS shall be determined during the general performance test in
accordance with applicable international or national standards. The values of the performance characteristics
determined shall meet the performance criteria specified in Table 1. The procedures for the determination of
these performance characteristics are described in Annex C.
The ambient conditions applied during the general performance test shall be documented.
The expanded uncertainty of the AMS measured values shall be calculated in accordance with ISO 14956 on
the basis of the performance characteristics determined during the general performance test and shall meet
the uncertainty specified for the measurement objective.
6.2.2 Ongoing quality control
The user shall check specific performance characteristics during ongoing operation of the measuring system
with a periodicity specified in Table 2.
The measurement uncertainty during field application shall be determined by the user of the measuring system
in accordance with applicable international or national standards. It can be determined by a direct or an indirect
6 © ISO 2012 – All rights reserved
ISO 13199:2012(E)
approach for uncertainty estimation as specified in ISO 20988. The uncertainty of the measured values under
field operation is not only influenced by the performance characteristics of the analyser itself, but also by
uncertainty contributions due to:
a) the sampling line and conditioning system;
b) the site-specific conditions;
c) the calibration gases used.
Table 1 — Relevant performance criteria of the analyser and the measuring system to be evaluated
during the general performance test and by means of ongoing QA/QC (laboratory) and field operation
General Term for
QA/QC
Performance Field
Performance criterion performance evaluating
characteristic operation
(lab)
test uncertainty
Response time ≤120 s —
P P P
Standard deviation of ≤1 % of the upper limit
a
repeatability in laboratory of the lowest measuring —
P P P
at zero point range used
Standard deviation of ≤2 % of the upper limit
a
repeatability in laboratory of the lowest measuring P P — P
at span point range used
≤2 % of the upper limit
Lack of fit of the lowest measuring P P — P
range used
≤2 % of the upper limit
Zero drift within 24 h of the lowest measuring —
P P P
range used
≤2 % of the upper limit
Span drift within 24 h of the lowest measuring P — P P
range used
Influence of atmospheric ≤2 % of the upper limit
pressure, for a pressure of the lowest measuring P — — P
b
change of 2 kPa range used
Influence of ambient ≤2 % of the upper limit
temperature, for a change of the lowest measuring — —
P P
of 10 K range used
Influence of electric
≤2 % of the upper limit of
voltage for the variation P — — P
the range per 10 V
per 10 V
Influence of CO
≤4 % of the upper limit of
and other interfering P P — P
the range used
c
components
Converter efficiency,
≥95 % P P — P
tested with CH
Losses and leakage in
≤2 % of the measured
the sampling line and — — P —
value
conditioning system
The upper limit of the lowest measuring range used should be set suitable to the application so that the measurement
values lie within 20 %–80 % of the analyser range.
a 3
Each zero gas and span gas containing 1 000 mg/m of CO shall be used for ongoing QA/QC test in the laboratory.
b
The tested pressure is defined in the manufacturer’s recommendations.
c
See C.5.
ISO 13199:2012(E)
6.2.3 Establishment of the uncertainty budget
An uncertainty budget shall be established to determine whether the analyser and its associated sampling
system fulfil the requirements for a maximum allowable expanded uncertainty. This uncertainty budget shall be
drawn up according to the procedures specified in ISO 14956 or ISO 20988, taking into account all the relevant
characteristics included in calculation of expanded uncertainty given in Table 1. An example of the evaluation
of an uncertainty budget is given in Annex D.
7 Measurement procedure
7.1 General
The AMS shall be operated according to the manufacturer’s instructions. The QA/QC procedures specified
in Clause 8 shall be strictly observed. During the measurement, the ambient conditions should be applied in
ranges during the general performance test.
7.2 Choice of the measuring system
To choose an appropriate analyser, sampling line, and conditioning unit, the following characteristics of waste
gases should be known before the field operation:
a) temperature of the waste gas;
b) water vapour content of the waste gas;
c) dust load of the waste gas;
d) expected concentration range of TVOCs;
e) expected concentration of potentially interfering substances.
To avoid long response times and memory effects, the sampling line should be as short as possible. If
necessary, a bypass pump should be used. If there is a high dust loading in the sample gas, an appropriate
heated filter shall be used.
Before conducting field measurements, the user shall verify that the necessary QA/QC procedures have
been performed.
7.3 Sampling
7.3.1 Sampling location
The sampling location chosen for the measurement devices and sampling shall be of sufficient size and
construction in order to obtain a representative emission measurement suitable for the measurement task.
In addition, the sampling location shall be chosen with regard to safety of the personnel, accessibility and
availability of electrical power.
7.3.2 Sampling point(s)
Ensure that the gas concentrations measured are representative of the average conditions inside the waste
gas duct. Therefore, the sampling points shall be selected to allow for a representative sampling.
[2] [6]
NOTE The selection of sampling points for representative sampling is described e.g. in ISO 9096 and EN 15259.
7.4 Data collection
The values measured with the calibrated NDIR analyser under operating conditions of the waste gas shall be
recorded by an internal or external data-logging system and averaged in accordance with the measurement task.
8 © ISO 2012 – All rights reserved
ISO 13199:2012(E)
7.5 Calculation
The TVOC mass concentration at standard conditions is the quotient of the mass of TVOCs to the volume of
the dry gas under specified reference conditions of temperature and pressure (273 K, 1 013 hPa), normally
expressed in milligrams per cubic metre (mg/m ) which is calculated as carbon. Results of the measurement
shall be expressed as mass concentrations at reference conditions of water vapour (dry gas).
If the TVOC concentration is provided as a volume concentration, Equation (1) shall be used to calculate the
mass concentration at standard conditions of temperature and pressure (273 K, 1 013 hPa):
M
C
γϕ= (1)
V
m
where
−6
ϕ is the TVOC volume fraction (= 10 );
M is the molar mass of carbon (= 12 g/mol);
C
V is the molar volume (= 22,4 l/mol).
m
8 Quality assurance and quality control procedures
8.1 General
QA/QC are important in order to ensure that the uncertainty of the measured TVOC values is kept within the
limits specified for the measurement task.
Two different applications of the automatic measuring system require distinction:
— AMS for intermittent measurements;
— permanently installed AMS for continuous monitoring.
8.2 Frequency of checks
Table 2 shows the minimum required frequency of checks. The user shall implement the relevant standards for
determination of performance characteristics or procedures described in Annex C.
The user shall implement a procedure to guarantee that the zero gases and span gases used meet the
uncertainty requirement specified in Annex B, e.g. by comparison with a reference gas of higher quality.
8.3 AMS for intermittent measurements
8.3.1 General
AMS for intermittent measurements shall be adjusted and checked in accordance with 8.3.2 at the frequencies
specified in Table 2.
The results of the QA/QC procedures shall be documented.
8.3.2 Adjustments and functional tests
8.3.2.1 Instrument adjustment
Instrument adjustments with zero and span gases shall be carried out at least at the beginning of each
measurement series. Appropriate safety procedures shall be followed.
ISO 13199:2012(E)
The zero and span gas shall be introduced under the same flow and pressure conditions using the sample port
of the instrument or according to the manufacturer’s instructions when using individual zero and span ports.
The adjustment procedure shall be carried out as follows:
a) feed zero gas into the NDIR analyser and set the zero;
b) feed span gas and adjust the instrument accordingly;
c) feed zero gas into the NDIR analyser once more and check the reading returns to zero.
Steps a) to c) shall be repeated if the reading does not return to zero.
Table 2 — Minimum frequency of checks for QA/QC during the operation
Minimum frequency
Check
AMS for intermittent measurements Permanently installed AMS
Response time once a year once a year
Standard deviation of repeatability once a year once a year
at zero point
Standard deviation of repeatability once a year once a year
at span point
Lack of fit once a year and after repair of the once a year and after repair of the
AMS AMS
Calibration — at regular time intervals specified e.g.
in legislation or applicable standards
by comparison with an independent
method of measurement
Check of CO influence once a year once a year
Converter check once for each measurement series once a year
Sampling system and leakage once for each measurement series once a year
check
Cleaning or changing of particulate once for each measurement series, if once in the period of unattended
a
filters at the sampling inlet and at needed operation
the monitor inlet
Zero drift every 3 h and at the end of measuring once in the period of unattended
period operation
Span drift every 3 h and at the end of measuring once in the period of unattended
period operation
Regular maintenance of the as required by the manufacturer once in the period of unattended
analyser operation
a
The particulate filter shall be changed periodically depending on the dust load at the sampling site. During this filter change the
filter housing shall be cleaned.
8.3.2.2 Response time
The response time of the AMS response shall be checked in accordance with C.2 at least once a year.
8.3.2.3 Standard deviation of repeatability at zero point
The standard deviation of repeatability at zero point shall be checked in accordance with C.3.2 at least once a year.
8.3.2.4 Standard deviation of repeatability at span point
The standard deviation of repeatability at span point shall be checked in accordance with C.3.3 at least once a year.
10 © ISO 2012 – All rights reserved
ISO 13199:2012(E)
8.3.2.5 Check of lack of fit (linearity check)
The linearity of the AMS response shall be checked in accordance with C.4 at least once a year.
8.3.2.6 Check of CO influence
The CO influence shall be checked in accordance with C.5 at least once a year.
8.3.2.7 Check of the converter efficiency
The converter efficiency shall be checked in accordance with C.6 at least once for each measurement series.
8.3.2.8 Sampling system and leakage check
The sampling system of the AMS shall be checked in accordance with C.7 at least once for each
measurement series.
8.3.2.9 Cleaning or changing of particulate filters
The particulate filter shall be checked at least once for each measurement series and changed if needed.
During the filter change the filter housing shall be cleaned.
8.3.2.10 Zero and span drift
The zero and span drift shall be checked in accordance with C.8 at least every 3 h and at the end of the
measuring period.
8.3.2.11 Regular maintenance of the analyser
The regular maintenance of the analyser shall be performed as required by the manufacturer.
8.3.2.12 Measurement uncertainty
The uncertainty of measured values obtained by AMS for intermittent monitoring shall be determined in
accordance with the principles laid down in ISO 20988. The measurement uncertainty shall be representative
of the intended application of the AMS. It shall take into account all relevant sources of uncertainty.
NOTE The uncertainty of measured values obtained by AMS for intermittent monitoring can be determined by a direct
or by an indirect approach described in ISO 20988. The direct approach can be based on comparison measurements with
an independent method of measurement under conditions of the intended operation of the AMS. ISO 20988 describes
procedures to evaluate such comparison measurements. A detailed description of the indirect approach is given in ISO 14956.
The uncertainty of the measured values shall meet the uncertainty criterion specified for the measurement objective.
8.4 Permanently installed AMS
8.4.1 General
Permanently installed AMS for continuous monitoring shall meet the performance criteria specified in Table 1.
General QA/QC procedures for permanently installed AMS specified in the relevant standards shall be observed.
The results of the QA/QC procedures shall be documented.
ISO 13199:2012(E)
8.4.2 Adjustments and functional tests
8.4.2.1 Instrument adjustment
Permanently installed AMS shall be adjusted in accordance with 8.3.2.1 at least once in the period of
unattended operation.
8.4.2.2 Response time
The response time of the AMS response shall be checked in accordance with C.2 at least once a year.
8.4.2.3 Standard deviation of repeatability at zero point
The standard deviation of repeatability at zero point shall be checked in accordance with C.3.2 at least once a year.
8.4.2.4 Standard deviation of repeatability at span point
The standard deviation of repeatability at span point shall be checked in accordance with C.3.3 at least once a year.
8.4.2.5 Check of lack of fit (Linearity check)
The linearity of the AMS response shall be checked in accordance with C.4 at least once a year.
8.4.2.6 Check of CO influence
The CO influence shall be checked in accordance with C.5 at least once a year.
8.4.2.7 Check of the converter efficiency
The converter efficiency shall be checked in accordance with C.6 at least once a year.
8.4.2.8 Sampling system and leakage check
The sampling system of the AMS shall be checked in accordance with C.7 at least once a year.
8.4.2.9 Zero and span drift
The zero and span drift shall be checked in accordance with C.8 at least once in the period of unattended
operation. This manual check is also needed for AMS with internal automatic zero and span checks.
8.4.2.10 Regular maintenance of the analyser
The regular maintenance of the analyser shall be performed once in the period of unattended operation
according to the manufacturer’s instructions and documentation.
8.4.3 Calibration, validation, and measurement uncertainty
Permanently installed AMS for continuous monitoring shall be calibrated and validated by comparison with an
independent method of measurement. The validation shall include the determination of the uncertainty of the
measured values obtained by the calibrated AMS.
NOTE The continuous flame ionization detector method for the determination of TVOC mass concentration in waste
[5]
gases specified in EN 13526 may be used as an independent method of measurement.
The AMS shall be subject to adjustments and functional tests according to 8.4.2 before each calibration and validation.
The calibration and validation of the AMS shall be performed at regular intervals and after repair of the analyser
in accordance with the relevant standards.
12 © ISO 2012 – All rights reserved
ISO 13199:2012(E)
The uncertainty of measured values obtained by permanently installed AMS for continuous monitoring shall
be determined by comparison measurements with an independent method of measurement as part of the
calibration and validation of the AMS. This ensures that the measurement uncertainty is representative of the
application at the specific plant.
NOTE The determination of the uncertainty of measured values obtained by permanently installed AMS for continuous
monitoring on the basis of a comparison with an independent method of measurement is described, for example, in ISO 20988.
The uncertainty of the measured values shall meet the uncertainty criterion specified for the measurement objective.
9 Test report
The test report shall be in accordance with international or national regulations. If not specified otherwise, it
shall include at least the following information:
a) reference to this International Standard (ISO 13199:2012);
b) description of the measurement objective;
c) principle of gas sampling;
d) information about the analyser and description of the sampling and conditioning line;
e) identification of the analyser used, and the performance characteristics of the analyser as listed in Table 1;
f) operating range;
g) details of the quality and the concentration of the span gases used;
h) description of plant and process;
i) identification of the sampling plane;
j) actions taken to achieve representative samples;
k) description of the location of the sampling point(s) in the sampling plane;
l) description of the operating conditions of the plant process;
m) changes in the plant operations during sampling;
n) sampling date, time, and duration;
o) time averaging on relevant periods;
p) measured values;
q) measurement uncertainty;
r) results of any checks;
s) any deviations from this International Standard.
ISO 13199:2012(E)
Annex A
(informative)
Schematic diagrams of NDIR analysers
A.1 General
Schematic diagrams of NDIR analysers for measuring TVOCs are shown in Figures A.1, A.2, and A.3,
respectively. Type 1 has a flowing reference cell through which the TVOC sample gas (no converter, gas B)
passes. Type 2 has a reference cell which is filled with a non-IR absorbent gas and has a switch valve. Type 3
has a modulation valve with a single cell.
A.2 Schematic diagram for NDIR analyser — Type 1
A schematic diagram of a flowing reference cell type NDIR analyser (dual beam) is shown in Figure A.1. This
analyser consists of the reference cell and samp
...




Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...