ISO 16840-10:2026
(Main)Wheelchair seating — Part 10: Requirements and test method for the assessment of resistance to ignition of postural support devices
- Abstract
This document specifies requirements and a test method to assess the resistance to ignition by a smouldering cigarette equivalent of integrated or non-integrated components of a wheelchair intended either to protect tissue integrity or to provide postural support, or both. The electrical ignition source is also a simulation of other potential sources of environmental ignition hazards. The test measures only the resistance to ignition of the items tested, and not the ignitability of the complete wheelchair. It gives an indication of, but cannot guarantee, the ignition behaviour of the assembled devices of a complete wheelchair. This document does not apply to resistance to ignition of structural parts of a wheelchair. This document does not cover changes in resistance to ignition as a result of regular washing or use of the postural support devices. This document allows for the separate testing of components of a wheelchair that are normally used in the horizontal plane (e.g. a seat cushion) from those normally used in the vertical plane (e.g. a back support). This document describes testing an assembly of the composite of materials as used in the component. The results of the test in this document do not give any indication of the resistance to ignition of any of the separate individual materials of the test sample. NOTE 1 The intent of this document is primarily to address components that interface with the human body, such as cushions for positioning, or whose described purpose is that of protecting skin tissue against pressure, shear, and maceration related damage, as well as textile, foam, and plastic-based postural support devices. NOTE 2 The pass criteria within this document have been set at a basic minimal level and can differ from current requirements in some countries. NOTE 3 Requirements for the control of risks from sources of fire created by electrical and electronic components are included in ISO 7176-14.
- Status
- Published
- Publication Date
- 08-Nov-2026
- Technical Committee
- ISO/TC 173/SC 1 - Wheelchairs
- Drafting Committee
- ISO/TC 173/SC 1/WG 11 - Wheelchair seating
- Current Stage
- 6060 - International Standard published
- Start Date
- 09-Nov-2026
- Completion Date
- 09-Nov-2026
ISO 16840-10:2026 is an International Standard from ISO for wheelchair seating. It specifies a test method and minimum requirements to assess the resistance to ignition of postural support devices, using a smouldering cigarette equivalent heat source and a reproducible electrical hot-wire source.
The standard is written for manufacturers, test laboratories, purchasers, and anyone selecting wheelchair seating components that are intended to protect tissue integrity or provide postural support. In practice, it helps them compare materials and assemblies for ignition resistance without treating the result as a test of the full wheelchair.
What does ISO 16840-10:2026 specify?
ISO 16840-10:2026 covers integrated and non-integrated wheelchair components that protect tissue integrity, provide postural support, or do both. It sets out how to test those components for resistance to ignition and what result they must achieve.
The standard is limited to the items tested as assemblies of the materials used in the component. It does not assess the ignitability of the complete wheelchair, structural parts of the wheelchair, or changes caused by regular washing or use.
It is organized into nine clauses and three informative annexes.
| Annex | What it covers |
|---|---|
| Annex A | Guidance on fabrication of the test coil apparatus, including parts, winding, and a sample holder |
| Annex B | Guidance on selecting the surface or surfaces at risk for testing |
| Annex C | Example test report |
What are the key requirements of ISO 16840-10:2026?
ISO 16840-10:2026 requires testing the composite of materials as used in the final component, not the individual materials on their own. That matters because a cover, foam, webbing, interliner, or other assembled layer can behave differently from each material tested separately.
Test items and sample condition
Clause 6.5 requires test samples to be new or unused and assembled as delivered for use, including cover, filling, and any other relevant components. If a device is intended to be used with specific covers, each cover must be tested. If it may be used with or without a cover, both configurations must be tested.
Clause 7.1 and 7.2 require the sample to be identified, selected on the surfaces judged to be at risk, and conditioned for at least 12 h in an environment between 10 °C and 30 °C and 15 % to 80 % relative humidity. In practice, the lab must test the product in a controlled state, not an arbitrary workshop condition.
Orientation and surface selection
Clause 7.3 separates the test into horizontal and vertical orientations. If a component is intended only for the horizontal plane within ±30°, use the horizontal test. For all other supports, use the vertical test. If the item is intended for use in either plane, only the vertical test is necessary.
Annex B explains how to choose the surface or surfaces most likely to be exposed to a fire source, which may be different from the normal body-contact surface. For a buyer or designer, this means the test should focus on the real exposure risk, not just the most obvious contact face.
Heat source and calibration
Clause 6.3 specifies a hot-wire coil made from NiCr alloy resistance wire, heated by a DC power supply in constant-current mode. Clause 7.4 requires calibration with a folded paper strip so that visible ignition occurs within 10 s to 14 s. The coil is also conditioned before use and cleaned between tests. This is important because the standard aims for repeatable ignition behavior from a controlled source rather than a variable real cigarette.
Test execution
Clauses 7.6 and 7.7 require the heat source to be applied for 60 s ± 1 s, then removed and the sample observed for a further 120 s. The test is repeated twice on a new or unaffected area of the sample.
For smaller postural support devices, clause 7.8 says the heat source is applied to the centre of the largest area that could be exposed to an external ignition source, in the normal orientation of use. That matters for belts, harnesses, head supports, and similar parts where only a limited area may be exposed in service.
Pass criteria
Clause 8 sets two performance requirements for both horizontal and vertical testing:
- no flaming during or after the test
- no smouldering 120 s after removal of the heat source
The standard therefore treats any flaming ignition as a failure and also requires the material to stop smouldering promptly.
Test reporting
Clause 9 requires the report to identify the standard used, the test lab, sample construction, calibration status, environmental conditions, photos before and after testing, deviations, and unusual features. It also includes a statement that the result applies only to the ignitability of the tested material combination under the test conditions.
What terms does ISO 16840-10:2026 define?
Flaming is combustion in the gas phase with visible light. In this standard, flaming is the most serious outcome because it means the sample has ignited.
Non-integrated means detachable without tools. The standard treats some cushions and supports as non-integrated even when they are held in place by hook and loop fastenings.
Postural support device (PSD) means a structure attached to a wheelchair with a surface that contacts the occupant’s body and supports, corrects, or stabilizes sitting posture. Examples include seat supports, back supports, lateral supports, head supports, and positioning belts.
Smouldering means self-propagating exothermic oxidation without flaming. The test checks for smouldering after the heat source is removed because continued burning can still injure the user.
Who uses ISO 16840-10:2026?
ISO 16840-10:2026 is used by wheelchair manufacturers, seating and cushion designers, and suppliers of postural support devices when selecting materials and validating product designs. Test laboratories use it to run ignition-resistance tests in a controlled and repeatable way.
It is also relevant to quality managers, risk managers, and purchasers who need a documented basis for product selection or product verification. In practice, it supports specification, procurement, product development, and conformity testing for wheelchair seating components.
What changed in ISO 16840-10:2026 from the previous edition?
ISO 16840-10:2026 is the third edition. It cancels and replaces ISO 16840-10:2021 and incorporates ISO 16840-10:2021/Amd. 1:2024.
The main change stated in the foreword is the removal of the original test method and the incorporation of the test method from the amendment. The document is also described as technically revised.
Which standards are used with ISO 16840-10:2026?
ISO 7176-26, Wheelchairs - Part 26: Vocabulary, supplies the terms and definitions used in this standard.
ISO 7176-14, Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods, is identified in the introduction as the standard for risks from electrical and electronic fire sources.
ISO 10993-1, Biological evaluation of medical devices - Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process, is referenced for material biocompatibility and risk management considerations.
ISO 14971, Medical devices - Application of risk management to medical devices, is referenced for risk analysis and design decisions where ignition resistance has to be balanced against other harms and functional needs.
What does the ISO 16840-10:2026 document contain?
ISO 16840-10:2026 contains a defined hot-wire test method, environmental and conditioning requirements, sample preparation steps, and pass/fail criteria for ignition resistance. It also includes health and safety guidance for test personnel and a detailed description of the apparatus, including the wire coil, power supply, paper calibration strip, and airflow-controlled test enclosure.
Annex A provides practical fabrication guidance for the test coil apparatus, with figures and a parts list. Annex B gives examples of surfaces that may be at risk and therefore suitable for test selection. Annex C provides a sample test report format.
The document uses figures to show coil dimensions, sample placement, calibration setup, and example positions for horizontal, vertical, and small PSD testing. It also includes a bibliography linking the method to wheelchair power-system fire-risk control, biological safety, and risk management.
Relations
- Effective Date
- 07-Dec-2025
- Effective Date
- 07-Dec-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ISO 16840-10:2026 is a standard published by the International Organization for Standardization (ISO). Its full title is "Wheelchair seating — Part 10: Requirements and test method for the assessment of resistance to ignition of postural support devices". This standard covers: This document specifies requirements and a test method to assess the resistance to ignition by a smouldering cigarette equivalent of integrated or non-integrated components of a wheelchair intended either to protect tissue integrity or to provide postural support, or both. The electrical ignition source is also a simulation of other potential sources of environmental ignition hazards. The test measures only the resistance to ignition of the items tested, and not the ignitability of the complete wheelchair. It gives an indication of, but cannot guarantee, the ignition behaviour of the assembled devices of a complete wheelchair. This document does not apply to resistance to ignition of structural parts of a wheelchair. This document does not cover changes in resistance to ignition as a result of regular washing or use of the postural support devices. This document allows for the separate testing of components of a wheelchair that are normally used in the horizontal plane (e.g. a seat cushion) from those normally used in the vertical plane (e.g. a back support). This document describes testing an assembly of the composite of materials as used in the component. The results of the test in this document do not give any indication of the resistance to ignition of any of the separate individual materials of the test sample. NOTE 1 The intent of this document is primarily to address components that interface with the human body, such as cushions for positioning, or whose described purpose is that of protecting skin tissue against pressure, shear, and maceration related damage, as well as textile, foam, and plastic-based postural support devices. NOTE 2 The pass criteria within this document have been set at a basic minimal level and can differ from current requirements in some countries. NOTE 3 Requirements for the control of risks from sources of fire created by electrical and electronic components are included in ISO 7176-14.
This document specifies requirements and a test method to assess the resistance to ignition by a smouldering cigarette equivalent of integrated or non-integrated components of a wheelchair intended either to protect tissue integrity or to provide postural support, or both. The electrical ignition source is also a simulation of other potential sources of environmental ignition hazards. The test measures only the resistance to ignition of the items tested, and not the ignitability of the complete wheelchair. It gives an indication of, but cannot guarantee, the ignition behaviour of the assembled devices of a complete wheelchair. This document does not apply to resistance to ignition of structural parts of a wheelchair. This document does not cover changes in resistance to ignition as a result of regular washing or use of the postural support devices. This document allows for the separate testing of components of a wheelchair that are normally used in the horizontal plane (e.g. a seat cushion) from those normally used in the vertical plane (e.g. a back support). This document describes testing an assembly of the composite of materials as used in the component. The results of the test in this document do not give any indication of the resistance to ignition of any of the separate individual materials of the test sample. NOTE 1 The intent of this document is primarily to address components that interface with the human body, such as cushions for positioning, or whose described purpose is that of protecting skin tissue against pressure, shear, and maceration related damage, as well as textile, foam, and plastic-based postural support devices. NOTE 2 The pass criteria within this document have been set at a basic minimal level and can differ from current requirements in some countries. NOTE 3 Requirements for the control of risks from sources of fire created by electrical and electronic components are included in ISO 7176-14.
ISO 16840-10:2026 is classified under the following ICS (International Classification for Standards) categories: 11.180.10 - Aids and adaptation for moving. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 16840-10:2026 has the following relationships with other standards: It is inter standard links to ISO 16840-10:2021/Amd 1:2024, ISO 16840-10:2021. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO 16840-10:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
International
Standard
ISO 16840-10
Third edition
Wheelchair seating —
2026-11
Part 10:
Requirements and test method for
the assessment of resistance to
ignition of postural support devices
Sièges de fauteuils roulants —
Partie 10: Exigences et méthode d'essai pour l'évaluation de la
résistance à l'inflammation des dispositifs de soutien postural
Reference number
© ISO 2026
All rights reserved.
This ISO publication is protected by copyright and is owned by ISO and/or its licensors.
The content of this ISO publication is provided under licence, not sold. Use is subject to the applicable licence terms issued by ISO,
an ISO member body, or an authorized third-party distributor.
Except as required for implementation or expressly permitted by a separate licence, no part of this ISO publication may be
reproduced, distributed, modified, used, or made available in any form or by any means – electronic or mechanical, including
photocopying, scanning, recording, or posting on internal or external digital platforms.
Any use beyond the scope of the granted rights is prohibited and may result in legal action.
ISO copyright office
CP 401 • CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Principle . 2
5 Health and safety of test personnel . 2
5.1 General .2
5.2 Enclosure .2
5.3 Extinguishers .3
6 Apparatus . 3
6.1 Test enclosure .3
6.2 Timer .3
6.3 Heat source .3
6.3.1 General .3
6.3.2 Wire coil .3
6.3.3 Power supply and connection to wire coil.4
6.3.4 Conditioning the wire coil .4
6.3.5 Paper calibration strip .4
6.4 Test environment .5
6.5 Test sample .5
7 Test procedure . 5
7.1 Sample preparation .5
7.2 Sample conditioning .5
7.3 Selection of testing orientation .5
7.4 Heat source calibration .5
7.5 Wire coil cleaning between tests .6
7.6 Horizontal test .6
7.7 Vertical test.7
7.8 Smaller PSDs .8
8 Requirements . 9
9 Test report . 9
Annex A (informative) Guidance on fabrication of the test coil apparatus .11
Annex B (informative) Selection of surface(s) at risk, to be used for testing .18
Annex C (informative) Test report example . 19
Bibliography .22
iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 173, Assistive products, Subcommittee SC 1,
Wheelchairs.
This third edition cancels and replaces the second edition (ISO 16840-10:2021) which has been technically
revised. It also incorporates the Amendment ISO 16840-10:2021/Amd. 1:2024.
The main changes are:
— removal of the original test method and incorporation of the test method from the amendment.
A list of all parts in the ISO 16840 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
Introduction
The ignition and subsequent burning of wheelchairs is very rare, but this can occur as a result of
— being close to a burning object such as a fire beside the wheelchair,
— overheating of any electrical or electronic device on the wheelchair, and
— contact from sparks or flames (such as welding sparks, cigarettes, or matches).
Wheelchair occupants are at particular risk of injury or death from these fires and resulting fumes because
it is very possible that they do not have the ability to move away from the wheelchair.
NOTE In the United States, data collected in the 1990s showed that only a small number of individuals per
[5] st
million wheelchair users had died due to fire . Public FDA records indicate most 21 century flammability incidents
[6]
involving wheelchairs are from electrical faults .
Wheelchairs can be considered to comprise the following components:
a) structural components such as the frame, wheels, etc., that are essential to the mechanical integrity of
the wheelchair;
b) power-related components such as motors, energy sources, controllers etc., that are required for the
functioning of powered devices on wheelchairs.
c) integrated or non-integrated devices to manage tissue integrity, such as seat and back support cushions,
that are intended to have primarily a clinical function to minimize the risks of skin damage (these can
also be intended to control or accommodate posture);
d) postural support devices, including, but not limited to, sling seats, sling back supports, arm supports,
foot supports, pelvic positioning supports (hip belts), anterior trunk supports (harnesses and chest
belts), lateral pelvic/trunk supports (lateral pads) etc., that are attached to the wheelchair and are
primarily intended to give positioning and postural support to the wheelchair occupant (these can also
be intended to aid in pressure redistribution);
Each of the above components has a different severity of risk associated with its likelihood of igniting and its
resulting harm to the wheelchair occupant. This document specifies requirements and a test method to help
manufacturers and purchasers of wheelchairs to design and procure wheelchairs and their components that
are appropriate for the risk of ignition balanced against the functional needs of the wheelchair occupant.
The aim of this document is to provide appropriate alternatives to using furniture-based flammability
standards, to reflect the uses and purposes of wheelchairs and their accessories.
There is yet no International Standard on the resistance to ignition of structural components [see list item
a) above].
The power related components [see list item b) above] are the subject of ISO 7176-14 which specifies
requirements to prevent overheating in electrical components that could lead to a fire.
The tissue integrity devices [see list item c) above] and postural support devices [see list item d) above] are
the subject of this document, regardless of the seating application (whether used on or off a wheelchair),
for the same benefits of the support device to the individual. For this purpose, a smouldering cigarette as
a standardised heat source is not necessarily a reproducible heat source, since the heat output between
cigarettes from within one pack is variable, and between different packs even more so. Since the early
furnishing flammability standards were introduced, filter tipped and fire safe cigarettes have evolved,
and thus today, testing is not being carried out with the equivalent heat source as when the furnishing
flammability standards were first drafted. The aim of this document is to describe a reproducible
standardised heat source output which reproduces the burn pattern and potential for ignition of a cigarette.
Cross-lab validation testing has shown that the previously recommended cartridge heater did not reproduce
the burn pattern or potential for ignition of a cigarette, and it did not reliably produce the same test results
as the hot wire coil.
v
Various existing flammability standards allow for active flaming for two-minutes after removal of the heat
source. By this time a wheelchair occupant could suffer severe injuries directly or by secondary ignition of
other materials in the environment. This current document differs in that it tests for resistance to ignition
‒ no flaming ignition is allowed. Given the full consideration of the wheelchair user’s safety and long-term
health, this document has been drafted to be used as the flammability standard for all seating textiles/soft
components of a wheelchair that interface with the human body.
Global efforts are under way to eliminate the use of chemicals which present a serious risk to humans and
the environment. Good practice is also to use biocompatible materials that minimize the risk of exposure to
the wheelchair occupant to toxic substances. Materials chosen are required to conform to biocompatibility
requirements and risk management guidance of ISO 10993-1. The change of emphasis around the materials
to be used minimises the use of toxic and hazardous flame-retardant chemicals, which is preferable due to
the minimal contribution of seating system materials to fire severity.
Materials chosen should comply with biocompatibility requirements and risk management guidance of
ISO 10993-1, which includes consideration of cytotoxicity, sensitization, and irritation evaluations for
devices making contact with intact skin.
The pass criteria within this document have been set at a basic minimal level and can differ from current
requirements in some countries. Where practical, manufacturers should use materials with superior
resistance to ignition, when the overall risk analysis indicates the benefits outweigh any unintended harm,
in accordance with ISO 14971.
The day to day usage of a wheelchair can affect its materials’ resistance to ignition through cyclic loading,
movement of materials, washing, cleaning, etc. Manufacturers will often take this effect into account as part
of their risk assessment when selecting materials for their products to minimize the effects of this normal
use. However, although this document can be used on parts that have been used, etc., the test samples
specify new or unused parts. Different environments commonly encountered by some wheelchair occupants
can also affect the flammability of materials. For example, home oxygen systems, delivery systems for drugs
carried in an inflammable medium, etc., can turn an inert material into a flammable one. Dust and other
materials accumulated within the chair have also been found to be a source of readily ignitable material.
Wheelchair manufacturers and occupants should be aware of these risks, and should design and use
wheelchairs accordingly, as covered by ISO 14971.
vi
International Standard ISO 16840-10:2026(en)
Wheelchair seating —
Part 10:
Requirements and test method for the assessment of
resistance to ignition of postural support devices
1 Scope
This document specifies requirements and a test method to assess the resistance to ignition by a smouldering
cigarette equivalent of integrated or non-integrated components of a wheelchair intended either to protect
tissue integrity or to provide postural support, or both. The electrical ignition source is also a simulation of
other potential sources of environmental ignition hazards.
The test measures only the resistance to ignition of the items tested, and not the ignitability of the complete
wheelchair. It gives an indication of, but cannot guarantee, the ignition behaviour of the assembled devices
of a complete wheelchair.
This document does not apply to resistance to ignition of structural parts of a wheelchair. This document
does not cover changes in resistance to ignition as a result of regular washing or use of the postural support
devices.
This document allows for the separate testing of components of a wheelchair that are normally used in the
horizontal plane (e.g. a seat cushion) from those normally used in the vertical plane (e.g. a back support).
This document describes testing an assembly of the composite of materials as used in the component. The
results of the test in this document do not give any indication of the resistance to ignition of any of the
separate individual materials of the test sample.
NOTE 1 The intent of this document is primarily to address components that interface with the human body, such
as cushions for positioning, or whose described purpose is that of protecting skin tissue against pressure, shear, and
maceration related damage, as well as textile, foam, and plastic-based postural support devices.
NOTE 2 The pass criteria within this document have been set at a basic minimal level and can differ from current
requirements in some countries.
NOTE 3 Requirements for the control of risks from sources of fire created by electrical and electronic components
are included in ISO 7176-14.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 7176-26, Wheelchairs — Part 26: Vocabulary
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 7176-26 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
flaming
undergoing combustion in the gaseous phase with the emission of light
3.2
non-integrated
detachable without the use of tools
Note 1 to entry: Seat or back support cushions held in place by hook and loop fastenings are also considered non-
integrated.
3.3
postural support device
PSD
structure, attached to a wheelchair, that has a surface that contacts the occupant’s body and is used either to
support, correct or stabilise the occupant’s sitting posture
EXAMPLE Seat support, back support, lateral support, head support, pelvic positioning belt, anterior trunk
support.
[SOURCE: ISO 7176-26:2007, 4.7.3 modified — The deprecated term “postural support” has been removed; in
the definition, “modify or accommodate” has been changed to “support, correct or stabilise”.]
3.4
smouldering
exothermic oxidation, not accompanied by flaming (3.1), that is self-propagating, i.e. without an external
ignition source
Note 1 to entry: It is possible that smouldering is accompanied by incandescence.
4 Principle
Test materials used in integrated and non-integrated seat and back supports, and other postural support
devices as used in wheelchairs are assembled in either horizontal or vertical samples and subjected to a heat
source that is equivalent to a smouldering cigarette in its thermal output and capacity to ignite materials
used in wheelchair seating. The resulting effects on the test materials are observed and measured.
NOTE The tests are repeated a second time to ensure that they are repeatable, but without incurring the costs of
more tests than needed as a minimum.
5 Health and safety of test personnel
5.1 General
WARNING — The following test requirements call for the use of procedures that can be hazardous to
health if adequate precautions are not taken.
Ensure that there is no hidden smouldering of the sample before disposal.
5.2 Enclosure
For safety, the tests should be conducted in a non-combustible fume cupboard. If such a cupboard is not
available, a test enclosure shall be constructed (see 6.1) so that the operator is protected from the fumes.
...