ISO 20342-1:2019
(Main)Assistive products for tissue integrity when lying down — Part 1: General requirements
Assistive products for tissue integrity when lying down — Part 1: General requirements
This document specifies general requirements and related test methods that are relevant to assistive products for tissue integrity (APTI) in the lying position in different application environments such as hospitals, home care and institutions. This document applies to the safety of APTI, which are intended to remain in situ during periods of lying, and to prevent and/or treat pressure injuries. This document covers a range of different lying support surfaces intended to be used in combination with the appropriate support platform or as a whole integrated system. This document also covers assistive products primarily intended for tissue integrity for changing a lying position and assistive products for maintaining a lying position. This document does not apply to lying support surfaces used in combination with incubators. This document addresses the combination of a full body support surface and an adjustable mattress support platform. It also covers safety and performance test methods to ensure protection against injuries to the user. This document specifies requirements and test methods for APTI within the following classifications of ISO 9999:2016: 04 33 06 Assistive products for tissue integrity when lying down such as but not limited to: — Mattresses and mattress overlays for pressure injury prevention; — Mattress coverings for pressure injury prevention mattresses. 12 31 03 Assistive products for sliding and turning such as but not limited to: Devices for changing position or direction of a person using sliding or turning techniques. The only products included are those intended to be used in a lying position and remain in situ as part of the lying support surface. They are the following: — sliding products that glide one way and lock the other way; — sheets and underlays in flexible materials with low friction; — fabric sold by the metre, cut as required for repositioning use; — powered turning product; This excludes sliding boards unless the product is intended to be left in situ. 09 07 06 Positioning pillows, positioning cushions and positioning systems such as but not limited to: — leg positioners, — arm positioners, and — multipurpose body positioners. 18 12 15 Bedding such as but not limited to: — draw sheets.
Produits d'assistance pour l'intégrité des tissus en position allongée — Partie 1: Exigences générales
Le présent document spécifie des exigences générales et des méthodes d'essai associées applicables aux produits d'assistance pour l'intégrité des tissus (PAIT) en position allongée dans différents environnements d'application tels que les hôpitaux, les soins à domicile et les institutions. Le présent document s'applique à la sécurité des PAIT, qui sont destinés à rester in situ pendant les périodes d'alitement et à aider à prévenir et/ou à traiter les escarres. Le présent document couvre une gamme de différentes surfaces d'alitement destinées à être utilisées en combinaison avec la structure porteuse appropriée ou en tant que système complet intégré. Le présent document traite également de produits d'assistance principalement destinés à l'intégrité des tissus pour changer de position allongée et de produits d'assistance pour maintenir une position allongée. Le présent document ne s'applique pas aux surfaces d'alitement utilisées en combinaison avec des incubateurs. Le présent document traite de la combinaison d'un support de redistribution de la pression du corps entier et d'un sommier réglable. Il intègre également des méthodes d'essai de sécurité et de performance visant à assurer la protection de l'utilisateur contre toute blessure. Le présent document spécifie des exigences et des méthodes d'essai pour les PAIT entrant dans le cadre des classifications suivantes de l'ISO 9999:2016: 04 33 06 Produits d'assistance pour l'intégrité des tissus en position allongée, tels que, sans toutefois s'y limiter: — les matelas et surmatelas pour la prévention des escarres; — les protège-matelas pour la prévention des escarres. 12 31 03 Produits d'assistance au glissement et produits d'assistance au retournement, tels que, sans toutefois s'y limiter: les dispositifs permettant de modifier la position ou la direction d'une personne en utilisant des techniques de glissement ou des techniques de retournement. Les seuls produits inclus sont les produits destinés à être utilisés en position allongée et à rester in situ en tant que partie intégrante de la surface d'alitement. Ces produits sont les suivants: — les produits de glissement qui coulissent dans un sens et se verrouillent dans l'autre sens; — les planches et les supports en matériaux flexibles à faible frottement; — le tissu vendu au mètre, découpé tel que requis pour une utilisation de repositionnement; — les produits électriques de repositionnement. Les planches de glissements sont exclues, à moins que le produit soit destiné à rester in situ. 09 07 06 Oreillers de positionnement, coussins de positionnement et systèmes de positionnement, tels que, sans toutefois s'y limiter: — les dispositifs de positionnement des jambes; — les dispositifs de positionnement des bras; — les dispositifs de positionnement du corps polyvalents. 18 12 15 Literie, telle que, sans toutefois s'y limiter: — les alèses.
General Information
Relations
Buy Standard
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 20342-1
First edition
2019-06
Assistive products for tissue integrity
when lying down —
Part 1:
General requirements
Produits d'assistance pour l'intégrité des tissus en position
allongée —
Partie 1: Exigences générales
Reference number
ISO 20342-1:2019(E)
©
ISO 2019
---------------------- Page: 1 ----------------------
ISO 20342-1:2019(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2019
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2019 – All rights reserved
---------------------- Page: 2 ----------------------
ISO 20342-1:2019(E)
Contents Page
Foreword .v
Introduction .vi
1 Scope . 1
2 Normative references . 2
3 Terms and definitions . 3
4 General requirements and safety . 7
4.1 General requirements . 7
4.2 Intended use . 7
4.2.1 General requirements . 7
4.2.2 Consideration regarding intended use . 7
4.2.3 Intended use statement . 8
4.3 APTI risk management. 8
4.4 APTI usability . 8
4.4.1 General. 8
4.4.2 Design requirements in relation to persons with cognitive impairment . 9
4.5 Design controls . 9
4.6 Clinical evaluation . 9
4.7 Foreseeable misuse . 9
4.8 Test conditions . 9
4.9 Lifting and carrying means.10
5 Safety requirements .10
5.1 Requirements for information supplied by the manufacturer .10
5.1.1 General.10
5.1.2 APTI traceability .10
5.1.3 Education and training .11
5.1.4 Pre-sale information .11
5.1.5 User information .11
5.1.6 Service information and inspection .12
5.1.7 Labelling .13
5.1.8 Marking of user weight and maximum load .13
5.1.9 Packaging.13
5.2 APTI which can be dismantled .13
5.2.1 General requirements .13
5.2.2 Small parts.13
5.2.3 Fasteners and connections .14
5.3 Resistance to corrosion .14
5.4 Noise and vibration .14
5.5 Sound audible acoustic energy .14
5.6 Default indicators .15
5.7 Feedback .15
6 Flammability .16
6.1 General .16
6.2 Flammability .16
6.3 Moulded parts used as enclosures for electrical equipment .16
7 Mechanical safety .17
7.1 Prevention of traps for the human body .17
7.2 Safety of moving and folding parts .17
7.3 V-shaped openings .19
7.4 Surfaces, corners, edges and protruding parts .19
7.5 Folding and adjusting mechanisms .19
7.6 Instability hazard .19
7.7 Temperature of parts that come into contact with human skin .20
© ISO 2019 – All rights reserved iii
---------------------- Page: 3 ----------------------
ISO 20342-1:2019(E)
7.8 Ergonomic principles .20
7.9 Additional consideration .21
8 Safety of electrical equipment .21
8.1 General electrical requirements .21
8.2 Electromagnetic compatibility .21
8.2.1 General.21
8.2.2 Emissions .21
8.2.3 Immunity .21
8.2.4 Power frequency magnetic field immunity .21
8.3 Liquid ingress .22
8.4 Interruption of power supply/supply mains to an APTI .22
8.5 Hold to run activation .22
8.6 Emergency stop functions .22
9 Biocompatibility .23
9.1 Biocompatibility and toxicity .23
9.2 Animal tissue .23
10 Contamination .23
10.1 Liquid ingress .23
10.2 Cleaning and disinfection . .24
10.3 Cross infection and microbial contamination .24
Annex A (informative) General information.25
Annex B (informative) Environmental and consumer related guidance .29
Annex C (informative) Periodic inspection .33
Bibliography .34
iv © ISO 2019 – All rights reserved
---------------------- Page: 4 ----------------------
ISO 20342-1:2019(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see www .iso
.org/iso/foreword .html.
This document was prepared by Technical Committee ISO/TC 173, Assistive products.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www .iso .org/members .html.
© ISO 2019 – All rights reserved v
---------------------- Page: 5 ----------------------
ISO 20342-1:2019(E)
Introduction
This document specifies general requirements that are relevant to assistive products for tissue
integrity (APTI) in the lying position in different application environments such as hospitals, home care,
and institutions. Some of the devices can be used/reused in more than one application environment.
This means that different requirements and test methods can apply to the same Assistive Products
for Tissue Integrity (APTI), depending on the application environment. For an APTI to conform with
this document, all relevant clauses need to be fulfilled, depending on the type of APTI. For example,
some APTI do not include electrical components; therefore, the clauses related to electrical components
might not be relevant.
APTI play a very important role in the prevention and treatment of pressure injuries. Another important
role in the prevention and treatment of pressure injury is the clinical practice and the clinical evaluation.
Guidance can be found in the NPUAP/EPUAP/PPPIA Guidelines, "Prevention and Treatment of Pressure
Ulcers: Clinical Practice Guideline,” from 2014.
Surfaces applied on operating theatre tables can also impact in the process of patient management
and may need to be taken into consideration. It should be recognized however, patient stability and
specialist equipment used during an operation often create conflicting priorities to those of an APTI.
Using this document, clinicians and manufacturers should consider the impact of other items (including
additional APTI) used in conjunction with an APTI on tissue integrity and safety.
This document only covers general requirements to ensure safety of users. However, the intention is to
develop a series of standards to cover the broad range of issues related to the APTI.
vi © ISO 2019 – All rights reserved
---------------------- Page: 6 ----------------------
INTERNATIONAL STANDARD ISO 20342-1:2019(E)
Assistive products for tissue integrity when lying down —
Part 1:
General requirements
1 Scope
This document specifies general requirements and related test methods that are relevant to assistive
products for tissue integrity (APTI) in the lying position in different application environments such as
hospitals, home care and institutions. This document applies to the safety of APTI, which are intended
to remain in situ during periods of lying, and to prevent and/or treat pressure injuries.
This document covers a range of different lying support surfaces intended to be used in combination
with the appropriate support platform or as a whole integrated system.
This document also covers assistive products primarily intended for tissue integrity for changing a
lying position and assistive products for maintaining a lying position.
This document does not apply to lying support surfaces used in combination with incubators.
This document addresses the combination of a full body support surface and an adjustable mattress
support platform. It also covers safety and performance test methods to ensure protection against
injuries to the user.
This document specifies requirements and test methods for APTI within the following classifications of
ISO 9999:2016:
04 33 06 Assistive products for tissue integrity when lying down such as but not limited to:
— Mattresses and mattress overlays for pressure injury prevention;
— Mattress coverings for pressure injury prevention mattresses.
12 31 03 Assistive products for sliding and turning such as but not limited to:
Devices for changing position or direction of a person using sliding or turning techniques. The only
products included are those intended to be used in a lying position and remain in situ as part of the
lying support surface. They are the following:
— sliding products that glide one way and lock the other way;
— sheets and underlays in flexible materials with low friction;
— fabric sold by the metre, cut as required for repositioning use;
— powered turning product;
This excludes sliding boards unless the product is intended to be left in situ.
09 07 06 Positioning pillows, positioning cushions and positioning systems such as but not limited to:
— leg positioners,
— arm positioners, and
— multipurpose body positioners.
© ISO 2019 – All rights reserved 1
---------------------- Page: 7 ----------------------
ISO 20342-1:2019(E)
18 12 15 Bedding such as but not limited to:
— draw sheets.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 554, Standard atmospheres for conditioning and/or testing — Specifications
ISO 9614-1, Acoustics — Determination of sound power levels of noise sources using sound intensity —
Part 1: Measurement at discrete points
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
ISO 13732-1, Ergonomics of the thermal environment — Methods for the assessment of human responses to
contact with surfaces — Part 1: Hot surfaces
ISO 14155, Clinical investigation of medical devices for human subjects — Good clinical practice
ISO 15223-1, Medical devices — Symbols to be used with medical device labels, labelling and information to
be supplied — Part 1: General requirements
ISO 22442-1, Medical devices utilizing animal tissues and their derivatives — Part 1: Application of risk
management
IEC 60601-1:2006, Medical electrical equipment — Part 1: General requirements for basic safety and
essential performance
IEC 60601-1-2, Medical electrical equipment — Part 1-2: General requirements for basic safety and essential
performance — Collateral standard: Electromagnetic compatibility — Requirements and tests
IEC 60601-1-6, Medical electrical equipment — Part 1-6: General requirements for basic safety and essential
performance — Collateral standard: Usability
IEC 60601-1-8, Medical electrical equipment — Part 1-8: General requirements for basic safety and
essential performance — Collateral standard: General requirements, tests and guidance for alarm systems
in medical electrical equipment and medical electrical systems
IEC 60601-1-11, Medical electrical equipment — Part 1-11: General requirements for basic safety and
essential performance — Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
IEC 60529, Degrees of protection provided by enclosures (IP Code)
IEC 60695-11-10, Fire hazard testing — Part 11-10: Test flames — 50 W horizontal and vertical flame
test methods
IEC 61000-3-2, Electromagnetic compatibility (EMC) — Part 3-2: Limits — Limits for harmonic current
emissions (equipment input current ≤16 A per phase)
IEC 61000-3-3, Electromagnetic compatibility (EMC) — Part 3-3: Limits — Limitation of voltage changes,
voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤16
A per phase and not subject to conditional connection
IEC 61000-4-3, Electromagnetic compatibility (EMC) — Part 4-3: Testing and measurement techniques —
Radiated, radio-frequency, electromagnetic field immunity test
2 © ISO 2019 – All rights reserved
---------------------- Page: 8 ----------------------
ISO 20342-1:2019(E)
IEC 61000-4-8, Electromagnetic compatibility (EMC) — Part 4-8: Testing and measurement techniques —
Power frequency magnetic field immunity test
IEC 61672-1, Electroacoustics — Sound level meters — Part 1: Specifications
IEC 61672-2, Electroacoustics — Sound level meters — Part 2: Pattern evaluation tests
IEC 80601-2-35, Medical electrical equipment — Part 2-35: Particular requirements for the basic safety
and essential performance of heating devices using blankets, pads or mattresses and intended for heating
in medical use
EN 716-2:2017, Furniture — Children’s cots and folding cots for domestic use — Part 2: Test methods
EN 1041, Information supplied by the manufacturer of medical devices
CISPR 11, Industrial, scientific and medical (ISM) radio-frequency equipment — Electromagnetic
disturbance characteristics — Limits and methods of measurement
European Commission, MEDDEV 2.7/1 CLINICAL EVALUATION: A GUIDE FOR MANUFACTURERS AND
NOTIFIED BODIES UNDER DIRECTIVES 93/42/EEC and 90/385/EEC
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https: //www .iso .org/obp
— IEC Electropedia: available at http: //www .electropedia .org/
3.1
application environment 4
care provided in a domestic area where the APTI is used to alleviate or compensate for an injury,
disability or disease
Note 1 to entry: This excludes use in all other application environments (e.g. nursing homes, rehabilitation and
geriatric facilities) when an APTI is purely designed for application environment 4.
[SOURCE: IEC 60601-2-52, 201.3.204, modified — "APTI" replaced "ME equipment"]
3.2
applied part
part of the APTI (3.5) that in normal use comes into physical contact with the user of the APTI (3.5) or a
medical system to perform its function
[SOURCE: IEC 60601-1:2006, 3.8, modified — "APTI" replaced ME Equipment", "of the" replaced "for",
"user" replaced "patient" and "necessary" and notes not included]
3.3
assistant
person who is helping a user (3.30) of the APTI (3.5)
EXAMPLE The ways assistants help persons with a disability (3.11) can be reposition in bed, bed ingress and
egress, operating hoists and assisting with transferring in/out of seats.
Note 1 to entry: An assistant can be a health care professional or a non-professional e.g. a relative.
3.4
assistive product
instrument, equipment or technical system intended by the manufacturer to be used for the prevention,
treatment or alleviation of or compensation for impairment (3.13)
© ISO 2019 – All rights reserved 3
---------------------- Page: 9 ----------------------
ISO 20342-1:2019(E)
3.5
assistive product for tissue integrity
APTI
surface intended to protect body tissue, designed to interface with the body when lying down (3.15) or
in adjusted position
3.6
bedding
items normally placed on a mattress (3.16)
EXAMPLE Mattress covers, underlays, sheets, blankets, quilts (duvets) and their covers, cushions, pillows,
bolsters and pillow cases.
3.7
body mass index
BMI
value derived from the mass (weight in kilograms) and height (in metres) of an individual, defined as
2
the body mass divided by the square of the body length, expressed in units of kg/m , calculated by the
following formula:
2
BMI = m/l
where
m is the mass in kg;
l is the length in metres.
3.8
clinical evaluation
assessment and analysis of clinical data pertaining to a medical device (3.21) to verify the clinical safety
and performance of the device when used as intended by the manufacturer
Note 1 to entry: Can include a compilation of clinical data, any scientific literature and the results of any clinical
investigations (3.9), taking into account any relevant harmonized standards.
Note 2 to entry: Guidance for clinical data evaluation is given in MEDDEV 2.7/1.
[SOURCE: ISO 13485:2016, 3.3, modified — notes added]
3.9
clinical investigation
systematic study into human subjects, undertaken to verify the safety and performance of a specific
medical device (3.21), under the manufacturer’s intended conditions of use
[SOURCE: ISO 22523:2006, 3.7]
3.10
detachable part
part designed to be unfastened or disconnected without damage to the part or the whole
3.11
disability
impairments (3.13), activity limitations, and participation restrictions denoting the negative aspects of
the interaction between an individual (with a health condition) and that individual's contextual factors
(environmental and personal factors)
[SOURCE: ICF 2001
...
NORME ISO
INTERNATIONALE 20342-1
Première édition
2019-06
Produits d'assistance pour l'intégrité
des tissus en position allongée —
Partie 1:
Exigences générales
Assistive products for tissue integrity when lying down —
Part 1: General requirements
Numéro de référence
ISO 20342-1:2019(F)
©
ISO 2019
---------------------- Page: 1 ----------------------
ISO 20342-1:2019(F)
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2019
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
Fax: +41 22 749 09 47
E-mail: copyright@iso.org
Web: www.iso.org
Publié en Suisse
ii © ISO 2019 – Tous droits réservés
---------------------- Page: 2 ----------------------
ISO 20342-1:2019(F)
Sommaire Page
Avant-propos .v
Introduction .vi
1 Domaine d’application . 1
2 Références normatives . 2
3 Termes et définitions . 3
4 Exigences générales et sécurité . 8
4.1 Exigences générales . 8
4.2 Utilisation prévue . 8
4.2.1 Exigences générales . 8
4.2.2 Considération concernant l’utilisation prévue . 8
4.2.3 Déclaration d’utilisation prévue . 8
4.3 Gestion des risques découlant d’un PAIT . 9
4.4 Aptitude à l’utilisation du PAIT . 9
4.4.1 Généralités . 9
4.4.2 Exigences de conception pour les personnes présentant une déficience
cognitive . 9
4.5 Contrôle de la conception .10
4.6 Évaluation clinique .10
4.7 Mauvaise utilisation prévisible .10
4.8 Conditions d’essai .10
4.9 Moyens de levage et de portage .11
5 Exigences de sécurité .11
5.1 Exigences relatives aux informations fournies par le fabricant .11
5.1.1 Généralités .11
5.1.2 Traçabilité des PAIT .11
5.1.3 Sensibilisation et formation .12
5.1.4 Informations de prévente .12
5.1.5 Informations relatives à l’utilisation .13
5.1.6 Informations relatives à l’entretien et inspection .14
5.1.7 Étiquetage .14
5.1.8 Marquage du poids de l’utilisateur et de la charge maximale .14
5.1.9 Conditionnement .15
5.2 PAIT pouvant être démonté .15
5.2.1 Exigences générales .15
5.2.2 Petits éléments .15
5.2.3 Fixations et raccords .15
5.3 Résistance à la corrosion .15
5.4 Bruit et vibration .15
5.5 Énergie acoustique des sons audibles .15
5.6 Indicateurs de défaut .16
5.7 Retour d’information .17
6 Inflammabilité .17
6.1 Généralités .17
6.2 Inflammabilité .18
6.3 Parties moulées utilisées comme boîtiers d’équipement électrique .18
7 Sécurité mécanique .18
7.1 Prévention des piégeages pour le corps humain .18
7.2 Sécurité des parties mobiles et articulées .19
7.3 Ouvertures en V .20
7.4 Surfaces, coins, bords et parties saillantes .21
7.5 Mécanismes de pliage et de réglage .21
7.6 Risque d’instabilité .21
© ISO 2019 – Tous droits réservés iii
---------------------- Page: 3 ----------------------
ISO 20342-1:2019(F)
7.7 Température des parties entrant en contact avec la peau humaine .21
7.8 Principes ergonomiques .22
7.9 Autre considération .22
8 Équipements électriques de sécurité.23
8.1 Exigences électriques générales .23
8.2 Compatibilité électromagnétique .23
8.2.1 Généralités .23
8.2.2 Émissions .23
8.2.3 Immunité .23
8.2.4 Immunité au champ magnétique à la fréquence du réseau .23
8.3 Pénétration de liquides .24
8.4 Coupure de l’alimentation/du réseau d’alimentation d’un PAIT . .24
8.5 Activation à action maintenue .24
8.6 Fonctions d’arrêt d’urgence .25
9 Biocompatibilité .25
9.1 Biocompatibilité et toxicité .25
9.2 Tissu d’origine animale .25
10 Contamination .25
10.1 Pénétration de liquides .25
10.2 Nettoyage et désinfection .26
10.3 Contamination croisée infectieuse et microbienne .26
Annexe A (informative) Informations générales .28
Annexe B (informative) Recommandations environnementales et relatives au consommateur .32
Annexe C (informative) Inspection périodique .37
Bibliographie .38
iv © ISO 2019 – Tous droits réservés
---------------------- Page: 4 ----------------------
ISO 20342-1:2019(F)
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui
concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2 (voir www
.iso .org/directives).
L’attention est attirée sur le fait que certains des éléments du présent document peuvent faire l’objet de
droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable
de ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant
les références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de
l’élaboration du document sont indiqués dans l’Introduction et/ou dans la liste des déclarations de
brevets reçues par l’ISO (voir www .iso .org/brevets).
Les appellations commerciales éventuellement mentionnées dans le présent document sont données
pour information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l’ISO liés à l’évaluation de la conformité, ou pour toute information au sujet de l’adhésion
de l’ISO aux principes de l’Organisation mondiale du commerce (OMC) concernant les obstacles
techniques au commerce (OTC), voir le lien suivant: www .iso .org/iso/fr/avant -propos.
Le présent document a été élaboré par le comité technique ISO/TC 173, Appareils et accessoires
fonctionnels pour les personnes handicapées.
Il convient que l’utilisateur adresse tout retour d’information ou toute question concernant le présent
document à l’organisme national de normalisation de son pays. Une liste exhaustive desdits organismes
se trouve à l’adresse www .iso .org/fr/members .html.
© ISO 2019 – Tous droits réservés v
---------------------- Page: 5 ----------------------
ISO 20342-1:2019(F)
Introduction
Le présent document spécifie des exigences générales applicables aux produits d’assistance pour
l’intégrité des tissus (PAIT) en position allongée dans différents environnements d’application tels que
les hôpitaux, les soins à domicile et les institutions. Certains de ces dispositifs peuvent être utilisés/
réutilisés dans plusieurs environnements d’application. Cela signifie que différentes exigences et
méthodes d’essai peuvent s’appliquer au même produit d’assistance pour l’intégrité des tissus (PAIT)
en fonction de l’environnement d’application. Pour qu’un PAIT se conforme au présent document, il est
nécessaire que toutes les dispositions applicables soient satisfaites, selon le type de PAIT considéré. Par
exemple, certains PAIT ne contiennent pas de composants électriques; par conséquent, les dispositions
relatives aux composants électriques peuvent ne pas s’appliquer.
Les PAIT jouent un rôle très important dans la prévention et le traitement des escarres. La pratique et
l’évaluation cliniques jouent également un rôle majeur dans la prévention et le traitement des escarres.
Des recommandations sont disponibles dans les lignes directrices NPUAP/EPUAP/PPPIA «Prévention
et Traitement des Escarres: Recommandations de pratiques cliniques» de 2014.
Les supports utilisés sur les tables d’opérations peuvent également avoir un impact sur le processus
de prise en charge des patients et il peut être nécessaire d’en tenir compte. Cependant, il convient de
reconnaître que la stabilité du patient et l'équipement du spécialiste utilisé durant une opération créent
souvent des priorités en conflit avec celles d'un PAIT.
Dans le cadre de l’utilisation du présent document, il convient que les cliniciens et les fabricants tiennent
compte de l’impact des autres éléments (incluant les autres PAIT) utilisés conjointement à un PAIT sur
l’intégrité des tissus et la sécurité.
Le présent document couvre uniquement les exigences générales permettant de garantir la sécurité
des utilisateurs. Cependant, l’objectif est de développer une série de normes visant à couvrir le large
éventail de préoccupations liées aux PAIT.
vi © ISO 2019 – Tous droits réservés
---------------------- Page: 6 ----------------------
NORME INTERNATIONALE ISO 20342-1:2019(F)
Produits d'assistance pour l'intégrité des tissus en position
allongée —
Partie 1:
Exigences générales
1 Domaine d’application
Le présent document spécifie des exigences générales et des méthodes d’essai associées applicables
aux produits d’assistance pour l’intégrité des tissus (PAIT) en position allongée dans différents
environnements d’application tels que les hôpitaux, les soins à domicile et les institutions. Le présent
document s’applique à la sécurité des PAIT, qui sont destinés à rester in situ pendant les périodes
d’alitement et à aider à prévenir et/ou à traiter les escarres.
Le présent document couvre une gamme de différentes surfaces d’alitement destinées à être utilisées
en combinaison avec la structure porteuse appropriée ou en tant que système complet intégré.
Le présent document traite également de produits d’assistance principalement destinés à l’intégrité
des tissus pour changer de position allongée et de produits d’assistance pour maintenir une position
allongée.
Le présent document ne s’applique pas aux surfaces d’alitement utilisées en combinaison avec des
incubateurs.
Le présent document traite de la combinaison d’un support de redistribution de la pression du corps
entier et d’un sommier réglable. Il intègre également des méthodes d’essai de sécurité et de performance
visant à assurer la protection de l’utilisateur contre toute blessure.
Le présent document spécifie des exigences et des méthodes d’essai pour les PAIT entrant dans le cadre
des classifications suivantes de l’ISO 9999:2016:
04 33 06 Produits d’assistance pour l’intégrité des tissus en position allongée, tels que, sans toutefois
s’y limiter:
— les matelas et surmatelas pour la prévention des escarres;
— les protège-matelas pour la prévention des escarres.
12 31 03 Produits d’assistance au glissement et produits d’assistance au retournement, tels que, sans
toutefois s’y limiter:
les dispositifs permettant de modifier la position ou la direction d’une personne en utilisant des
techniques de glissement ou des techniques de retournement. Les seuls produits inclus sont les produits
destinés à être utilisés en position allongée et à rester in situ en tant que partie intégrante de la surface
d’alitement. Ces produits sont les suivants:
— les produits de glissement qui coulissent dans un sens et se verrouillent dans l’autre sens;
— les planches et les supports en matériaux flexibles à faible frottement;
— le tissu vendu au mètre, découpé tel que requis pour une utilisation de repositionnement;
— les produits électriques de repositionnement.
Les planches de glissements sont exclues, à moins que le produit soit destiné à rester in situ.
© ISO 2019 – Tous droits réservés 1
---------------------- Page: 7 ----------------------
ISO 20342-1:2019(F)
09 07 06 Oreillers de positionnement, coussins de positionnement et systèmes de positionnement, tels
que, sans toutefois s’y limiter:
— les dispositifs de positionnement des jambes;
— les dispositifs de positionnement des bras;
— les dispositifs de positionnement du corps polyvalents.
18 12 15 Literie, telle que, sans toutefois s’y limiter:
— les alèses.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur
contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique.
Pour les références non datées, la dernière édition du document de référence s’applique (y compris les
éventuels amendements).
ISO 554, Atmosphères normales de conditionnement et/ou d'essai — Spécifications
ISO 9614-1, Acoustique — Détermination par intensimétrie des niveaux de puissance acoustique émis par
les sources de bruit — Partie 1: Mesurages par points
ISO 10993-1, Évaluation biologique des dispositifs médicaux — Partie 1: Évaluation et essais au sein d'un
processus de gestion du risque
ISO 13732-1, Ergonomie des ambiances thermiques — Méthodes d'évaluation de la réponse humaine au
contact avec des surfaces — Partie 1: Surfaces chaudes
ISO 14155, Investigation clinique des dispositifs médicaux pour sujets humains — Bonnes pratiques cliniques
ISO 15223-1, Dispositifs médicaux — Symboles à utiliser avec les étiquettes, l'étiquetage et les informations
à fournir relatifs aux dispositifs médicaux — Partie 1: Exigences générales
ISO 22442-1, Dispositifs médicaux utilisant des tissus animaux et leurs dérivés — Partie 1: Application de la
gestion des risques
IEC 60601-1:2006, Appareils électromédicaux — Partie 1: Exigences générales pour la sécurité de base et
les performances essentielles
IEC 60601-1-2, Appareils électromédicaux — Partie 1-2: Exigences générales pour la sécurité de base et les
performances essentielles — Norme collatérale: Perturbations électromagnétiques — Exigences et essais
IEC 60601-1-6, Appareils électromédicaux — Partie 1-6: Exigences générales pour la sécurité de base et les
performances essentielles — Norme collatérale: Aptitude à l’utilisation
IEC 60601-1-8, Appareils électromédicaux — Partie 1-8: Exigences générales pour la sécurité de base et
les performances essentielles — Norme collatérale: Exigences générales, essais et guides pour les systèmes
d'alarme des appareils et des systèmes électromédicaux
IEC 60601-1-11, Appareils électromédicaux — Partie 1-11: Exigences générales pour la sécurité de base
et les performances essentielles — Norme collatérale: Exigences pour les appareils électromédicaux et les
systèmes électromédicaux utilisés dans l'environnement des soins à domicile
IEC 60529, Degrés de protection procurés par les enveloppes (Code IP)
IEC 60695-11-10, Essais relatifs aux risques du feu — Partie 11-10: Flammes d’essai — Méthodes d’essai
horizontale et verticale à la flamme de 50 W
2 © ISO 2019 – Tous droits réservés
---------------------- Page: 8 ----------------------
ISO 20342-1:2019(F)
IEC 61000-3-2, Compatibilité électromagnétique (CEM) — Partie 3-2: Limites — Limites pour les émissions
de courant harmonique (courant appelé par les appareils ≤ 16 A par phase)
IEC 61000-3-3, Compatibilité électromagnétique (CEM) — Partie 3-3: Limites — Limitation des variations
de tension, des fluctuations de tension et du papillotement dans les réseaux publics d’alimentation basse
tension, pour les matériels ayant un courant assigné ≤ 16 A par phase et non soumis à un raccordement
conditionnel
IEC 61000-4-3, Compatibilité électromagnétique (CEM) — Partie 4-3: Techniques d’essai et de mesure —
Essai d’immunité aux champs électromagnétiques rayonnés aux fréquences radioélectriques
IEC 61000-4-8, Compatibilité électromagnétique (CEM) — Partie 4-8: Techniques d’essai et de mesure —
Essai d’immunité au champ magnétique à la fréquence du réseau
IEC 61672-1, Électroacoustique — Sonomètres — Partie 1: Spécifications
IEC 61672-2, Électroacoustique — Sonomètres — Partie 2: Essais d’évaluation d’un modèle
IEC 80601-2-35, Appareils électromédicaux — Partie 2-35: Exigences particulières pour la sécurité de base
et les performances essentielles des dispositifs de réchauffage utilisant des couvertures, des coussins ou des
matelas chauffants et destinés au réchauffage des patients en usage médical
EN 716-2:2017, Mobilier — Lits à nacelle fixes et pliants à usage domestique pour enfants — Partie 2:
Méthodes d’essai
EN 1041, Informations fournies par le fabricant de dispositifs médicaux
CISPR 11, Appareils industriels, scientifiques et médicaux — Caractéristiques de perturbations
radioélectriques — Limites et méthodes de mesure.
Commission européenne, MEDDEV 2.7/1 CLINICAL EVALUATION: A GUIDE FOR MANUFACTURERS
AND NOTIFIED BODIES UNDER DIRECTIVES 93/42/EEC and 90/385/EEC.
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s’appliquent.
L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en
normalisation, consultables aux adresses suivantes:
— ISO Online browsing platform: disponible à l’adresse https: //www .iso .org/obp;
— IEC Electropedia: disponible à l’adresse http: //www .electropedia .org/.
3.1
environnement d’application 4
soins dispensés à domicile et pour lesquels les PAIT sont utilisés pour atténuer ou compenser les effets
d’une blessure, d’une incapacité ou d’une maladie
Note 1 à l'article: Ceci exclut une utilisation dans tous les autres environnements d’application (par exemple
maisons de soins, centres de réadaptation et centres de gériatrie), lorsqu’un PAIT est purement conçu pour
l’environnement d’application 4.
[SOURCE: IEC 60601-2-52, 201.3.204, modifiée — «appareils EM» a été remplacé par «PAIT»]
© ISO 2019 – Tous droits réservés 3
---------------------- Page: 9 ----------------------
ISO 20342-1:2019(F)
3.2
partie appliquée
partie du PAIT (3.5) qui, en utilisation normale, vient en contact physique avec l’utilisateur du PAIT
(3.5) ou d’un système médical pour assurer sa fonction
[SOURCE: IEC 60601-1:2006, 3.8, modifiée— «appareil EM» a été remplacé par «PAIT», «pour que» a
été remplacé par «du» et la fin remaniée en conséquence, «patient» a été remplacé par «utilisateur» et
«nécessairement» a été supprimé]
3.3
accompagnant
personne aidant un utilisateur (3.30) du PAIT (3.5)
EXEMPLE Les manières dont les accompagnants aident les personnes en situation d’incapacité (3.11)
peuvent notamment inclure le repositionnement dans le lit, l’installation dans le lit et la sortie du lit, l’utilisation
de lève-personnes et l’aide au transfert dans et hors des fauteuils et des lits.
N
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.