ISO TS 24971-2:2026
(Main)Medical devices - Guidance on the application of ISO 14971 - Part 2: Machine learning in artificial intelligence
- Abstract
ISO TS 24971-2:2026 This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971[2].
This document does not apply to MLMD employing large language models (LLM) or generative AI.
- Status
- Published
- Publication Date
- 16-Jun-2026
- Technical Committee
- SC 62A - Common aspects of medical equipment, software, and systems
- Drafting Committee
- JWG 1 - Application of risk management to medical devices
- Current Stage
- PPUB - Publication issued - 60 Publication
- Start Date
- 22-Jun-2026
- Due Date
- 31-Jul-2026
- Completion Date
- 31-Jul-2026
ISO/TS 24971-2:2026 is a Technical Specification from ISO, prepared in collaboration with IEC, for medical devices. It gives guidance on applying the ISO 14971 risk management process to machine learning in artificial intelligence for ML-enabled medical devices (MLMD). Manufacturers, risk managers, and related technical teams use it to address ML-specific issues such as bias, data quality, explainability, autonomy, monitoring, updating, and retraining.
What does ISO/TS 24971-2:2026 specify?
ISO/TS 24971-2:2026 specifies guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and on how to apply ISO 14971 to ML-enabled medical devices. It is intended to be used with ISO 14971 and ISO/TR 24971, so it adds ML-focused guidance to an existing medical-device risk management framework rather than creating a separate process.
The document excludes ML-enabled medical devices that employ large language models (LLM) or generative AI.
Its structure is straightforward: the early clauses cover scope, references, and definitions, then the main body follows the ISO 14971 risk management flow, and the annexes give supporting explanations, examples, questions, and autonomy considerations.
| Annex | What it covers |
|---|---|
| Annex A | Explanation of bias |
| Annex B | Examples of hazards and hazardous situations |
| Annex C | Identification of hazards and characteristics related to safety |
| Annex D | Considerations for MLMD having a level of autonomy |
What are the key requirements of ISO/TS 24971-2:2026?
The document keeps ISO 14971 as the base process
Clause 4.1 says the ISO 14971 risk management process also applies to MLMD. In practice, a manufacturer still has to identify hazards, estimate and evaluate risk, choose controls, judge overall residual risk, and follow up after release. The standard’s value is in showing where ML-based products need extra attention, not in replacing the usual medical-device risk process.
Competence has to cover ML, data, IT, usability, and clinical context
Clause 4.3 expands competence for MLMD teams. The people doing risk management should understand ML model and algorithm development, IT and security, software validation, usability engineering, clinical workflow, and data management. That matters because many ML-related hazards come from poor training data, poor interpretation of output, or weak understanding of how the device is actually used.
The risk management plan needs ML-specific post-production decisions
Clause 4.4 requires the risk management plan to define the scope, responsibilities, review criteria, risk acceptability criteria, residual-risk evaluation method, verification activities, and production and post-production review. For MLMD, the plan should also consider performance monitoring, updating, retraining, and possible return to a previous version. In practice, this means the manufacturer must decide in advance how drift, updates, and retraining will be handled.
Risk analysis must include intended use, misuse, and ML-specific safety characteristics
Clauses 5.2 to 5.5 emphasize intended use, reasonably foreseeable misuse, and characteristics related to safety. The document highlights misuse such as use outside the intended patient population, inadequate input data, overreliance on output, and problems linked to lack of transparency. It also points readers to Annex C for questions that help identify MLMD hazards and safety-related characteristics. This matters because some ML risks only appear when the device is used in real clinical settings, with real users and imperfect data.
Risk control should start with design, then protective measures, then information for safety
Clause 7.1 follows the ISO 14971 priority order. For MLMD, examples include selecting and designing the ML algorithm and model, assuring data quality, separating training data from test data, adding human oversight, using alarms or intervention mechanisms, and giving users clear information and training. The document also links explainability to safety information, so users can understand important factors affecting output.
Verification can use testing, synthetic data, and usability evaluation
Clause 7.2 explains that implementation and effectiveness of risk controls have to be verified. For MLMD, this can include testing the ML model against acceptance criteria, using synthetic data in simulated use, checking for demographic bias, verifying return-to-previous-version procedures, and using usability evaluation to see whether an intervention or explanation really works. In practice, the standard expects evidence that the controls are not just designed well but actually function as intended.
Overall residual risk must consider autonomy, bias, silence of failures, and novelty
Clause 8.1 adds MLMD-specific considerations to the overall residual risk evaluation. The evaluation should reflect the complexity of use, possible silent failures, the level of autonomy, use-related residual risks such as overtrust, and the degree of novelty or innovation. Clause 8.2 then requires significant residual risks to be disclosed, including performance limits, relevant checks before use, transparency, explainability, and known differences in performance across patient groups.
Production and post-production activities must watch for drift and trigger action
Clause 10 requires active collection and review of information related to safety. For MLMD, this includes monitoring performance over time, tracking drift, reviewing changes in clinical practice or patient populations, and deciding when to retrain, update, or return to a previous version. Actions taken after release must be documented in the risk management file, with the rationale and evidence used for the decision.
What terms does ISO/TS 24971-2:2026 define?
- Bias - A systematic difference in treatment of certain objects, people, or groups compared with others.
- Explainability - The property of a system to express the important factors behind its results in a way humans can understand.
- Machine learning (ML) - A function of a system that learns from input data instead of following only fixed instructions.
- Machine learning algorithm (ML algorithm) - An algorithm that determines the parameters of a machine learning model from data according to given criteria.
- Machine learning model (ML model) - A logical representation of a system that generates predictions based on input data.
- ML-enabled medical device (MLMD) - A medical device that uses machine learning.
- Training data - Data used to train a machine learning model.
- Test data - Data used to assess the performance of a machine learning model and kept separate from training data.
Who uses ISO/TS 24971-2:2026?
ISO/TS 24971-2:2026 is used by medical device manufacturers, risk management teams, quality and regulatory staff, software and ML engineers, clinical specialists, usability engineers, IT and cybersecurity specialists, and data management personnel. It is particularly relevant when developing, verifying, validating, monitoring, updating, or retraining ML-enabled medical devices.
It also helps buyers, assessors, and auditors understand what evidence to look for when an MLMD claims safe use in a clinical workflow, especially where output affects diagnosis, treatment planning, or other clinical actions.
Which standards are used with ISO/TS 24971-2:2026?
| Standard | What it contributes |
|---|---|
| ISO 14971:2019 | The base medical-device risk management process used throughout the document |
| ISO/TR 24971:2020 | Additional guidance on applying ISO 14971, which this document uses as its companion guidance |
| IEC 62304 | Guidance for software life cycle processes, relevant to MLMD software development and verification |
| IEC 62366-1 | Usability engineering guidance, relevant to transparency, explainability, and use-related risk |
| IEC 80001-1:2021 | Risk management for IT networks with medical devices, relevant to connected MLMD and security |
| IEC/TR 80002-1:2009 | Guidance on applying ISO 14971 to medical device software |
| IEC/TR 60601-4-1:2017 | Guidance on medical electrical equipment with a degree of autonomy |
| ISO/IEC 22989:2022 | AI concepts and terminology used in the background and definitions |
| ISO/IEC 23894:2023 | General AI risk management guidance |
| ISO/IEC/TR 24027:2021 | Guidance on bias in AI systems, used for the bias discussion |
What does the ISO/TS 24971-2:2026 document contain?
The document includes figures that show the ML concept development, training, and testing stages, the fit of an ML model to training data, and options for concept improvement and retraining. Annex B includes example tables of events, circumstances, hazardous situations, harm, and potential risk control measures.
Annex C provides question sets for identifying hazards and characteristics related to safety across use, design and development, and post-production. Annex D describes autonomy concepts and ways to categorize MLMD by how much the device monitors, generates options, selects options, executes actions, and allows human intervention.
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ISO TS 24971-2:2026 is a technical specification published by the International Electrotechnical Commission (IEC). Its full title is "Medical devices - Guidance on the application of ISO 14971 - Part 2: Machine learning in artificial intelligence". This standard covers: ISO TS 24971-2:2026 This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971[2]. This document does not apply to MLMD employing large language models (LLM) or generative AI.
ISO TS 24971-2:2026 This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971[2]. This document does not apply to MLMD employing large language models (LLM) or generative AI.
ISO TS 24971-2:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.01 - Medical equipment in general; 35.240.01 - Application of information technology in general; 35.240.80 - IT applications in health care technology. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO TS 24971-2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 bias
- 3.2 explainability
- 3.3 machine learning
- 3.4 machine learning algorithm
- 3.5 machine learning model
- 3.6 ML-enabled medical device
- 3.7 overfitting
- 3.8 test data
- 3.9 training
- 3.10 training data
- 3.11 transparency
- 3.12 underfitting
- 4 General requirements for risk management system
- 4.1 Risk management process
- 4.2 Management responsibilities
- 4.3 Competence of personnel
- 4.4 Risk management plan
- 4.5 Risk management file
- 5 Risk analysis
- 5.1 Risk analysis process
- 5.2 Intended use and reasonably foreseeable misuse
- 5.3 Identification of characteristics related to safety
- 5.4 Identification of hazards and hazardous situations
- 5.5 Risk estimation
- 6 Risk evaluation
- 7 Risk control
- 7.1 Risk control option analysis
- 7.2 Implementation of risk control measures
- 7.3 Residual risk evaluation and subsequent steps
- 8 Evaluation of overall residual risk
- 8.1 General considerations for MLMD
- 8.2 Disclosure of significant residual risks
- 9 Risk management review
- 10 Production and post-production activities
- 10.1 General
- 10.2 Information collection
- 10.3 Information review
- 10.4 Actions
- Annex A
- Annex B
- 1 Data selection Inadequate control of data collection process
- 2 Data annotation Inconsistent annotation performance
- 3 Data augmentation Poorly designed or excessive augmentation
- 4 Data shift Data entry errors, data dropout
- 5 Usability User unable to intervene in an autonomous process (see IEC/TR 60601-4-1 )
- 6 Update process Insufficient control of retraining process
- 8 ML model training Insufficient training data
- 9 Functionality False positive results
- 10 Installation Incompatibility of MLMD with software or hardware environment
- Annex C
- Annex D
- Bibliography
Introduction
Artificial intelligence (AI) is rapidly advancing and offers transformative potential for the healthcare sector. These advantages can be related to improved benefits for the patient, increased efficiency in clinical workflows and more effective management of healthcare overall. However, the implementation of new technologies such as AI can also present new risks and can, for example, jeopardize patient safety, affect privacy and security, influence user actions, undermine trust in healthcare or adversely affect the management of healthcare.
[16] The safety and effectiveness of AI in medical devices was explored in an AAMI-BSI document , which identified three ways in which AI-based medical devices differed from “traditional” (non-AI) medical devices:
a) Training. These medical devices can process large amounts of data and learn from these data to improve their results. Thus, they can have positive effects on patient health within the scope of the intended use of the medical device.
b) Level of autonomy. These medical devices can generate different treatment options, select the best option [8] based on a trained model and execute the selected option (see for example IEC/TR 60601-4-1 ). These steps can be performed with reduced or even without direct user action. Therefore, human oversight is critical for their safe application.
c) Explainability. These medical devices often rely on complex algorithms and large datasets to generate output. However, the inherent opacity of these algorithms makes it challenging to interpret how specific conclusions or recommendations are derived. This aspect is sometimes referred to as the “black-box problem”. This can make it difficult for clinicians, other healthcare personnel and individuals without specialist knowledge to understand the rationale for the output. Many different AI-based technologies and algorithms exist today, which are differentiated by data topology and model architecture, including decision trees, genetic algorithms and deep learning-based technologies [4] [6] such as generative AI and neural networks. ISO/IEC 22989 and ISO/IEC 23894 provide general guidance on AI concepts, terminology and risk management, but they do not specifically address the application of AI to medical devices. It is noted that “risk” is defined in these documents as the effect of uncertainties on [3] objectives (see also ISO 31000 ). This definition is useful for organizational or business risk management. The term “risk” used in the healthcare sector is different and is defined in ISO 14971:2019 as the combination of the probability of occurrence of harm and the severity of that harm. Figure 1 — Concept development, training and testing stages of the ML model and its relationship with the ML algorithm, the training data and the test data This document focuses on machine learning (ML) techniques and is restricted to ML-enabled medical devices (MLMD). Machine learning is considered a subset of AI that involves an ML model and an ML algorithm. See Figure 1. The ML model and the ML algorithm are the results of the concept development for a new MLMD, together with acceptance criteria for the eventual MLMD. It is noted that the supporting infrastructure (computing framework, hardware, network and other IT components) can be an important aspect in concept development. The MLMD acceptance criteria are different from the criteria for risk acceptability. It
ISO/TS 24971-2:2026(en) is important to establish the MLMD acceptance criteria at the start, as part of concept development, and not at the end of the MLMD development. Otherwise, the results of MLMD testing could influence the decisions when establishing those criteria.
After concept development, the ML model is trained by using an ML algorithm enabling it to learn patterns from training data without being explicitly programmed. Next, the trained ML model is applied to test data to verify its performance. The training data and the test data are different (disjoint) sets. They can be actual patient data or synthetic data, i.e. data created to simulate a patient for training or testing purposes. The tested ML model can then be applied to new patient data in a clinical setting. More information on MLMD [21] [23] [17][18] can be found in IMDRF documents N67 and N88 and in guidance documents from FDA, Health Canada and MHRA.
It is recognized that the ML model can require retraining after a period of use to redefine its parameters and to ensure its continued performance. This can be achieved by planned retraining with collected patient data or on a continuous basis with each new patient data. The latter is referred to as “continuous learning” throughout this document and sometimes as “adaptive” in other documents. All medical devices come with inherent risks. Manufacturers are required to demonstrate that their medical devices do not pose unacceptable risks, and that the benefits of the intended use outweigh the overall residual risk. ISO 14971 details how manufacturers can identify, assess and control risks to protect the patients, the users and other persons as well as property (for example objects, data, other equipment) and the environment. This includes risks related to data and systems security and cybersecurity. Guidance on [2] [11] the application of ISO 14971 is provided in document ISO/TR 24971 . Additionally, IEC 80001-1 and [12] IEC/TR 80002-1 address software internal to a medical device that can support AI or ML. This document was developed to provide specific guidance on the application of ISO 14971 to MLMD. It does not provide a new risk management process, nor does it expand the requirements of ISO 14971. This document addresses risks related to machine learning and topics such as data management, feature extraction, unwanted bias, information security, training the ML model by an ML algorithm, evaluation and testing of the [14] trained ML model. See Annex A for an explanation of bias. The report AAMI TIR34971 provided valuable input for this document.
Technical Specification ISO/TS 24971-2:2026(en) Medical devices — Guidance on the application of ISO 14971 — Part 2:
Machine learning in artificial intelligence
1 Scope
This document provides guidance on risks specific to artificial intelligence (AI) and machine learning (ML) and how to apply the risk management process of ISO 14971 to ML-enabled medical devices (MLMD). This [2] document is intended to be used in conjunction with ISO 14971 and ISO/TR 24971 . This document does not apply to MLMD employing large language models (LLM) or generative AI.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 14971:2019, Medical devices — Application of risk management to medical devices
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 14971:2019 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 bias
systematic difference in treatment of certain objects, people, or groups in comparison to others Note 1 to entry: Treatment is any kind of action, including perception, observation, representation, prediction or decision.
[7] [SOURCE: ISO/IEC TR 24027:2021, 3.2.2 ]
3.2 explainability
property of a system to express important factors influencing the system’s results in a way that humans can understand Note 1 to entry: It is intended to answer the question “Why?” without actually attempting to argue that the course of action that was taken was necessarily optimal.
[4] [SOURCE: ISO/IEC 22989:2022, 3.5.7 , modified — “AI system” was changed to “system”.]
ISO/TS 24971-2:2026(en)
3.3 machine learning
ML function of a system that can learn from input data instead of strictly following a set of specific instructions Note 1 to entry: Machine learning focuses on prediction based on known properties learned from the input data. [13] [SOURCE: AAMI TIR66: 2017 , 3.19]
3.4 machine learning algorithm
ML algorithm algorithm to determine parameters of a machine learning model (3.5) from data according to given criteria [4] [SOURCE: ISO/IEC 22989:2022, 3.3.6 , modified — The example was removed.]
3.5 machine learning model
ML model logical representation of a system that generates predictions based on input data Note 1 to entry: A machine learning model results from training (3.9) based on a machine learning algorithm (3.4). [4] Note 2 to entry: Adapted from ISO/IEC 22989:2022, 3.1.23 and 3.3.7 .
3.6 ML-enabled medical device
MLMD medical device that utilizes machine learning (3.3) Note 1 to entry: MLMD can involve multiple ML models and multiple ML algorithms (3.4).
3.7 overfitting
creating a model that fits the training data (3.10) too precisely and fails to generalize on new data Note 1 to entry: Overfitting can occur because the trained model has learned from non-essential features in the training data (i.e. features that do not generalize to useful outputs), because of excessive noise in the training data (e.g. excessive number of outliers) or because the model is too complex for the training data. Note 2 to entry: Overfitting can be identified when there is a significant difference between errors measured on training data and on separate test data. The performance of overfitted models is especially impacted when there is a significant mismatch between training data and new data.
Note 3 to entry: See Figure 2 for a graphical comparison with underfitting (3.12). [5] [SOURCE: ISO/IEC 23053:2022, 3.1.4 , modified — In the definition, “which” was changed to “that”; in Note 1 to entry, “because of” was added; in Note 2 to entry, “separate test and validation data” was changed to “separate test data”, and “production data” was changed to “new data”; Note 3 to entry was added.]
3.8 test data
data used to assess the performance of the machine learning model (3.5) Note 1 to entry: Test data is disjoint from training data (3.10). [4] [SOURCE: ISO/IEC 22989:2022, 3.2.14 , modified — The preferred term “evaluation data” was removed; in the definition, “a final model” was changed to “the machine learning model”; “validation data” was deleted from Note 1 to entry.]
ISO/TS 24971-2:2026(en)
3.9 training
process to determine or to improve the parameters of a machine learning model (3.5), based on a machine learning algorithm (3.4), by using training data (3.10) [4] [SOURCE: ISO/IEC 22989:2022, 3.3.15 , modified — The admitted term “model training” was removed.]
3.10 training data
data used to train a machine learning model (3.5) [4] [SOURCE: ISO/IEC 22989:2022, 3.3.16 ]
3.11 transparency
property of a system that appropriate information about the system is made available to relevant stakeholders Note 1 to entry: Appropriate information for system transparency can include aspects such as features, performance, residual risks, limitations, components, procedures, measures, design goals, design choices and assumptions, data sources and labelling protocols.
Note 2 to entry: Inappropriate disclosure of some aspects of a system can violate security, privacy or confidentiality requirements.
[4] [SOURCE: ISO/IEC 22989:2022, 3.5.15 , modified — “residual risks” was added in Note 1 to entry.]
3.12 underfitting
creating a model that does not fit the training data (3.10) closely enough and produces incorrect predictions on new data Note 1 to entry: Underfitting can occur when features are poorly selected, when there is insufficient training time or when the model is too simple to learn from large training data due to limited model capacity (i.e. expressive power). Note 2 to entry: See Figure 2 for a graphical comparison with overfitting (3.7). [5] [SOURCE: ISO/IEC 23053:2022, 3.1.5 , modified — In Note 1 to entry, “when there is” was added; Note 2 to entry was added.] a) Underfitting b) Good fit c) Overfitting Key training data model prediction X input variable Y output variable Figure 2 — Fit of the ML model on the training data
ISO/TS 24971-2:2026(en)
4 General requirements for risk management system
4.1 Risk management process
ISO 14971 specifies a process for risk management of medical devices. This process also applies to ML-enabled medical devices. The process covers the identification of hazards and hazardous situations, the estimation and evaluation of the associated risks, the need for risk control, the evaluation of overall residual risk and the production and post-production activities for monitoring the effectiveness of the risk control measures.
4.2 Management responsibilities
ISO 14971 requires top management to:
a) provide adequate resources; b) assign competent personnel; c) define and document a policy for establishing the criteria for risk acceptability; and d) review the suitability of the risk management process to ensure its continuing effectiveness. [2] There is no specific guidance for MLMD in addition to what is provided in ISO/TR 24971 .
4.3 Competence of personnel
ISO 14971 requires that persons performing risk management tasks are competent and, where appropriate, have knowledge of and experience with the particular medical device, its use, the technologies involved, and the risk management techniques employed.
For MLMD specifically, this knowledge should cover the relevant stages of the MLMD life cycle. This can include general knowledge about the training data and test data used for the MLMD, awareness of the data quality, understanding of the context of the data and knowledge of what the data actually mean in clinical practice.
The team performing the risk management tasks for the MLMD should include:
[23] a) individuals with an understanding of good machine learning practices (see IMDRF document N88 ), who are trained on ML model and ML algorithm development to identify hazardous situations that can occur for the MLMD; b) individuals with knowledge of information technology (IT) to identify risks and risk control measures related to the computing platforms (when applicable) where the MLMD is to be installed and used, including the security risk control measures for MLMD connected to the internet or to other equipment; c) individuals trained on simulation and algorithm testing to identify risk control measures for the MLMD, [9] either in software or in hardware (see IEC 62304 ); d) individuals with knowledge of software validation to develop verification methods for the risk control measures; e) individuals trained on usability engineering (human factors engineering) in the context of the MLMD [10] use in accordance with IEC 62366-1 , including how transparency and explainability can affect the MLMD use (e.g. the potential for bias in interpreting the output); f) individuals with knowledge of the clinical workflow for the MLMD, understanding of the environment in which the MLMD is to be used, understanding of the interactions of the MLMD with medical practitioners and patients, and the capability to evaluate the relevance of the results produced by the MLMD in clinical practice; and
ISO/TS 24971-2:2026(en) g) individuals with an understanding of good data management practices to identify risks related to data management, including data integrity, quality of training data and test data (e.g. data robustness, data fitness to the ML model and free of unintended bias), and data privacy. NOTE An explanation of bias is given in Annex A.
4.4 Risk management plan
ISO 14971 requires the manufacturer to establish a risk management plan that includes:
a) the scope of the risk management activities; b) the assignment of responsibilities and authorities; c) requirements for the review of risk management activities; d) the criteria for risk acceptability; e) the method to evaluate overall residual risk and the criteria for acceptability of the overall residual risk; f) activities for verification of the implementation and effectiveness of risk control measures; and g) activities to collect and review production and post-production information. Specifically for MLMD, the manufacturer should consider whether the post-production activities need to include monitoring the MLMD performance, updating the MLMD or retraining the ML model. If monitoring the MLMD performance is considered necessary for safety, the risk management plan shall define the methods and processes for such monitoring and which data are to be collected and reviewed to support this activity. If updating the MLMD or retraining the ML model is considered necessary for the purpose of managing risk, the risk management plan should include the criteria for initiating an update or retraining, the frequency of applying those criteria (occasionally, after a set period or continuously) and the planned activities in the update or retraining process itself. These activities may further include software maintenance and provisions for returning the MLMD to a previous version. NOTE Retraining the ML model and returning the MLMD to a previous version can be subject to regulatory [1] requirements (see ISO 13485:2016, 7.3.9 and 8.2.3 ).
EXAMPLE 1 Methods for monitoring the MLMD performance, including periodic testing with new test data, to verify that the MLMD continues to perform as intended.
EXAMPLE 2 Processes for collecting specific data from reported complaints, processes for collecting necessary input data (e.g. required data input fields to be filled by users before the MLMD can operate).
4.5 Risk management file
ISO 14971 requires the manufacturer to establish and maintain a risk management file. The risk management file contains the results produced by the risk management process as well as traceability from each identified hazard to the risk analysis, the risk evaluation, the implementation and verification of risk control measures and the results of evaluation of the residual risk.
[2] There is no specific guidance for MLMD in addition to what is provided in ISO/TR 24971 .
5 Risk analysis
5.1 Risk analysis process
ISO 14971 requires the manufacturer to perform risk analysis for the particular medical device and to record the results of the planned risk analysis activities in the risk management file. The techniques that support risk [2] analysis (see ISO/TR 24971:2020, Annex B ) can also be applied to ML-enabled medical devices.
ISO/TS 24971-2:2026(en)
5.2 Intended use and reasonably foreseeable misuse
ISO 14971 requires the manufacturer to document a clear description of the medical device, its intended use and the reasonably foreseeable misuse. It is emphasized that reasonably foreseeable misuse includes use error (for example related to lack of transparency), use outside the intended patient population, use with inadequate input data, overreliance on MLMD output. Questions related to MLMD use that can assist in identifying hazards and characteristics related to safety are given in Annex C. Differences in MLMD performance can relate to the selection of training data from specific patient groups, leading to differences in performance for other patient groups. The manufacturer can decide to restrict the intended use based on such differences in performance.
...