IEC 62021-3:2014
(Main)Insulating liquids - Determination of acidity - Part 3: Test methods for non-mineral insulating oils
Insulating liquids - Determination of acidity - Part 3: Test methods for non-mineral insulating oils
IEC 62021-3:2014 describes two procedures for the determination of the acidity of unused and used electrical non-mineral insulating oils. Method A is potentiometric titration and Method B is colourimetric titration. The method may be used to indicate relative changes that occur in non-mineral insulating oil during use under oxidizing conditions regardless of the colour or other properties of the resulting non-mineral oil. The acidity can be used in the quality control of unused non-mineral insulating oil. As a variety of oxidation products present in used non-mineral insulating oil contribute to acidity and these products vary widely in their corrosion properties, the test cannot be used to predict corrosiveness of non-mineral insulating oil under service conditions.
Liquides isolants - Détermination de l'acidité - Partie 3: Méthodes d'essai pour les huiles non minérales isolantes
L'IEC 62021-3:2014 décrit deux procédures pour la détermination de l'acidité des huiles non minérales isolantes électriques neuves et usagées. La Méthode A est le titrage potentiométrique et la Méthode B est le titrage colorimétrique. La méthode peut être utilisée pour indiquer des modifications relatives qui se produisent dans une huile non minérale isolante pendant son utilisation par oxydation, indépendamment de la couleur ou d'autres propriétés de l'huile non minérale résultante. L'acidité peut être utilisée dans le contrôle de la qualité de l'huile non minérale isolante neuve. Comme divers produits d'oxydation présents dans l'huile non minérale isolante usagée contribuent à l'acidité et comme ces produits varient largement du point de vue de leurs propriétés de corrosion, l'essai ne peut pas être utilisé pour prévoir la corrosivité d'une huile non minérale isolante dans des conditions de service.
General Information
Standards Content (Sample)
IEC 62021-3 ®
Edition 1.0 2014-03
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Insulating liquids – Determination of acidity –
Part 3: Test methods for non-mineral insulating oils
Liquides isolants – Détermination de l’acidité –
Partie 3: Méthodes d’essai pour les huiles non minérales isolantes
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing more than 30 000 terms and
Technical Specifications, Technical Reports and other definitions in English and French, with equivalent terms in 14
documents. Available for PC, Mac OS, Android Tablets and additional languages. Also known as the International
iPad. Electrotechnical Vocabulary (IEV) online.
IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a More than 55 000 electrotechnical terminology entries in
variety of criteria (reference number, text, technical English and French extracted from the Terms and Definitions
committee,…). It also gives information on projects, replaced clause of IEC publications issued since 2002. Some entries
and withdrawn publications. have been collected from earlier publications of IEC TC 37,
77, 86 and CISPR.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements
Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales,
électriques. Il contient plus de 30 000 termes et définitions en
Spécifications techniques, Rapports techniques et autres
anglais et en français, ainsi que les termes équivalents dans
documents de l'IEC. Disponible pour PC, Mac OS, tablettes
14 langues additionnelles. Egalement appelé Vocabulaire
Android et iPad.
Electrotechnique International (IEV) en ligne.
Recherche de publications IEC - www.iec.ch/searchpub
Glossaire IEC - std.iec.ch/glossary
Plus de 55 000 entrées terminologiques électrotechniques, en
La recherche avancée permet de trouver des publications IEC
en utilisant différents critères (numéro de référence, texte, anglais et en français, extraites des articles Termes et
comité d’études,…). Elle donne aussi des informations sur les Définitions des publications IEC parues depuis 2002. Plus
projets et les publications remplacées ou retirées. certaines entrées antérieures extraites des publications des
CE 37, 77, 86 et CISPR de l'IEC.
IEC Just Published - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications IEC. Just
Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur cette
Disponible en ligne et aussi une fois par mois par email. publication ou si vous avez des questions contactez-nous:
csc@iec.ch.
IEC 62021-3 ®
Edition 1.0 2014-03
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Insulating liquids – Determination of acidity –
Part 3: Test methods for non-mineral insulating oils
Liquides isolants – Détermination de l’acidité –
Partie 3: Méthodes d’essai pour les huiles non minérales isolantes
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX T
ICS 29.040.10 ISBN 978-2-8322-1404-6
– 2 – IEC 62021-3:2014 © IEC 2014
CONTENTS
FOREWORD . 4
INTRODUCTION . 6
1 Scope . 7
2 Normative references . 7
3 Terms and definitions . 7
4 Method A: Automatic potentiometric titration . 8
4.1 Principle . 8
4.2 Reagents and auxiliary products . 8
4.2.1 Reagents . 8
4.2.2 Titration reagent . 8
4.2.3 Titration solvent . 8
4.2.4 Potassium hydrogen phthalate, primary standard . 9
4.2.5 Reference electrode electrolyte . 9
4.2.6 Aqueous buffer solutions . 9
4.2.7 Glass electrode cleaning solution . 9
4.3 Apparatus . 9
4.3.1 Potentiometric titration apparatus . 9
4.3.2 Glass indicator electrode . 9
4.3.3 Reference electrode . 10
4.3.4 Stirrer . 10
4.3.5 Titration vessel . 10
4.3.6 Titration stand . 10
4.4 Sampling. 10
4.5 Preparation and maintenance of electrode system . 10
4.5.1 Preparation . 10
4.5.2 Maintenance . 10
4.6 Calibration . 11
4.6.1 Calibration of pH titrimeter . 11
4.6.2 Settings for the potentiometric instrument . 11
4.7 Procedure . 12
4.7.1 General . 12
4.7.2 Standardization of alcoholic potassium hydroxide solution . 12
4.7.3 Blank titration . 13
4.7.4 Sample titration . 13
4.8 Calculation of result . 14
4.9 Precision . 14
4.9.1 Repeatability . 14
4.9.2 Reproducibility . 14
4.10 Report. 15
5 Method B: Colourimetric titration. 15
5.1 Principle . 15
5.2 Reagents . 15
5.2.1 General . 15
5.2.2 Titration reagent . 15
5.2.3 Titration solvent . 16
5.2.4 Potassium hydrogen phthalate, primary standard . 16
5.2.5 Standard hydrochloric acid solution . 16
5.2.6 Alkali blue 6B indicator solution . 16
5.2.7 Cobalt nitrate solution . 16
5.3 Apparatus . 16
5.3.1 Titration vessel . 16
5.3.2 Stirrer . 16
5.3.3 Burette . 16
5.4 Sampling. 16
5.5 Procedure . 17
5.5.1 General . 17
5.5.2 Standardization of alcoholic potassium hydroxide solution . 17
5.5.3 Blank titration . 17
5.5.4 Sample titration . 18
5.6 Calculation of result . 18
5.7 Precision . 18
5.7.1 Repeatability . 18
5.7.2 Reproducibility . 19
5.8 Report. 19
Annex A (informative) Determination of acidity in non-mineral electrical insulating oils
by photometric titration . 20
A.1 Principle . 20
A.2 Reagents and solvents . 20
A.3 Preparation of titration solutions and solvents . 20
A.3.1 Potassium hydroxide alcoholic solution (0,01 mol/l) . 20
A.3.2 Potassium hydrogen phthalate solution (0,01 mol/l) . 20
A.3.3 Titration solvent . 21
A.4 Apparatus . 21
A.4.1 Volumetric titrator . 21
A.4.2 Titration vessel . 21
A.4.3 Titration stand . 21
A.4.4 Stirrer . 21
A.4.5 Recorder/printer . 21
A.4.6 Photometric sensor . 21
A.5 Sampling. 22
A.6 Procedure . 22
A.6.1 Preparation and maintenance of the titration system . 22
A.6.2 Determination of acidity of the titration solvent (blank titration) . 22
A.6.3 Determination of molarity of the potassium hydroxide alcoholic
solution (0,01 mol/l) . 22
A.6.4 Titration of soluble acidity in the oil sample. 23
A.7 Calculation of result . 23
A.8 Report. 23
Figure 1 – Potentiometric titration curve . 12
Figure A.1 – Molecular structure of para-naphtholbenzein indicator in a) acidic media
and b) basic media . 24
Figure A.2 – UV spectra of para-naphtholbenzein indicator in toluene/2-propanol/water
solution in acidic media (curve a) and basic media (curve b) . 24
– 4 – IEC 62021-3:2014 © IEC 2014
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
INSULATING LIQUIDS – DETERMINATION OF ACIDITY –
Part 3: Test methods for non-mineral insulating oils
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 62021-3 has been prepared by IEC technical committee 10: Fluids
for electrotechnical applications.
The text of this standard is based on the following documents:
FDIS Report on voting
10/936/FDIS 10/942/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
A list of all parts in the IEC 62021 series, published under the general title Insulating liquids –
Determination of acidity, can be found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
· reconfirmed,
· withdrawn,
· replaced by a revised edition, or
· amended.
IMPORTANT – The “colour inside” logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct understanding
of its contents. Users should therefore print this publication using a colour printer.
– 6 – IEC 62021-3:2014 © IEC 2014
INTRODUCTION
Health and safety
This International Standard does not purport to address all the safety problems associated
with its use. It is the responsibility of the user of the standard to establish appropriate health
and safety practices and determine the applicability of regulatory limitations prior to use.
The insulating liquids which are the subject of this standard should be handled with due
regard to personal hygiene. Direct contact with the eyes may cause slight irritation. In the
case of eye contact, irrigation with copious quantities of clean running water should be carried
out and medical advice sought.
Some of the procedures referenced in this standard involve the use of processes that could
lead to a hazardous situation. Attention is drawn to the relevant standard for guidance.
Environment
This standard involves non-mineral insulating oils, chemicals, used sample containers and
fluid-contaminated solids. The disposal of these items should be carried out according to local
regulations with regard to their impact on the environment. Every precaution should be taken
to prevent the release into the environment of these oils.
INSULATING LIQUIDS – DETERMINATION OF ACIDITY –
Part 3: Test methods for non-mineral insulating oils
1 Scope
This part of IEC 62021 describes two procedures for the determination of the acidity of
unused and used electrical non-mineral insulating oils. Method A is potentiometric titration
and Method B is colourimetric titration.
NOTE 1 In unused and used non-mineral insulating oils, the constituents that may be considered to have acidic
characteristics include organic acids, phenolic compounds, some oxidation products, resins, organometallic salts
and additives.
The method may be used to indicate relative changes that occur in non-mineral insulating oil
during use under oxidizing conditions regardless of the colour or other properties of the
resulting non-mineral oil.
The acidity can be used in the quality control of unused non-mineral insulating oil.
As a variety of oxidation products present in used non-mineral insulating oil contribute to
acidity and these products vary widely in their corrosion properties, the test cannot be used to
predict corrosiveness of non-mineral insulating oil under service conditions.
NOTE 2 The acidity results obtained by potentiometric test method may or may not be numerically the same as
those obtained by colourimetric methods, but they are generally of the same magnitude.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC 60475, Method of sampling insulating liquids
ISO 5725 (all parts), Accuracy (trueness and precision) of measurement methods and results
ISO 6619, Petroleum products and lubricants – Neutralization number – Potentiometric
titration method
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
acidity
quantity of base, expressed in milligrams of potassium hydroxide per gram of sample,
required to titrate potentiometrically or colourimetrically a test portion in a specified solvent to
the end point
3.2
non-mineral insulating oil
insulating liquid, not derived from petroleum crudes
– 8 – IEC 62021-3:2014 © IEC 2014
3.3
unused oil
non-mineral insulating oil that has not been used in, or been in contact with, electrical
equipment
4 Method A: Automatic potentiometric titration
4.1 Principle
Any acid-base titration may be conducted potentiometrically. The test portion of the insulating
fluid is dissolved in solvent and titrated potentiometrically with alcoholic potassium hydroxide
using a glass-indicating electrode and a reference electrode. The potential difference (which
can be expressed as pH after calibration) is measured after the successive addition of known
increments of alcoholic potassium hydroxide.
Where a strong point of inflection is detected from the first derivative of the titration curve, this
should be used as the end point. If only a weak inflection point is present, the potential
difference corresponding to pH of 11,5 has been found more reproducible and less
instrument-dependent.
4.2 Reagents and auxiliary products
4.2.1 Reagents
Only reagents of recognized analytical grade and de-ionized water or water of equivalent
purity shall be used.
4.2.2 Titration reagent
Standard alcoholic solution between 0,01 mol/l and 0,05 mol/l potassium hydroxide (KOH).
EXAMPLE Preparation of 0,01 mol/l potassium hydroxide in 2-propanol.
Add 0,6 g of potassium hydroxide to 1 000 ml ± 10 ml of 2-propanol. Boil gently for 10 min to
effect solution. Cool and stopper the flask.
Allow the solution to stand in the dark for 2 days and then filter the supernatant liquid through
a 5 µm membrane filter. Store in a suitable amber glass bottle.
The concentration of this solution is approximately 0,01 mol/l and shall be standardized as
described in 4.7.2.
Store in such a manner that the solution is protected from atmospheric carbon dioxide by
means of a guard tube containing soda-lime absorbent and in such a way that it does not
come into contact with cork, rubber or saponifiable stopcock grease.
Commercial alcoholic potassium hydroxide solution may be used, if necessary diluting to
0,01 mol/l with 2-propanol. This shall be standardized as described in 4.7.2.
NOTE 1 For oils with high acidity, which may give an extended titration time, it may be helpful to carry out a pre-
test using 0,1 mol/l potassium hydroxide titrant to determine a suitable titrant concentration.
NOTE 2 For periodic tests on equipment in service, faster titration may be achieved by the use of 0,05 or
0,1 mol/l potassium hydroxide by agreement between the laboratory and the equipment owner, although this may
result in poorer precision and detection limit.
4.2.3 Titration solvent
The titration solvent is as follows:
– 2-propanol (isopropanol, IPA), pure.
2-propanol is the preferred solvent. It should be noted that the use of other solvents might
change the dissociation potential and thus the neutralisation point.
4.2.4 Potassium hydrogen phthalate, primary standard
This should be dried before use for 2 h at 105 °C.
A 0,1 mol/l solution of hydrochloric acid in de-ionized water, prepared as in ISO 6619, may be
used. Other acids may be used, e.g. benzoic acid, provided they are certified against a
primary standard.
4.2.5 Reference electrode electrolyte
Prepare a solution of potassium chloride in de-ionized water, or lithium chloride in ethanol, at
the concentration recommended by the electrode manufacturer. Commercially available
solutions may be used where available.
4.2.6 Aqueous buffer solutions
Buffer solutions of suitable pH for calibration of electrodes, for example, pH 4, pH 7 and
pH 11 or close to pH 12.
4.2.7 Glass electrode cleaning solution
Weigh 8 g of ammonium peroxydisulfate into a glass beaker. Carefully add 100 ml of 98 %
sulphuric acid and gently stir. Before use, the solution should be left overnight for the solid to
dissolve completely.
WARNING Ammonium peroxydisulfate is a strong oxidizing agent. Sulphuric acid is a strong
corrosive agent. Handle carefully.
Commercially available cleaning solutions as recommended by the electrode manufacturer
may be used.
4.3 Apparatus
4.3.1 Potentiometric titration apparatus
An automatic pH titrimeter or an instrument for a potentiometric titration capable of titrating to
a fixed end-point using either variable or fixed titrant increments.
The instrument shall be protected from stray electrical fields so that no change of the reading
is produced by touching any part of the system with a grounded lead.
An automatic burette with a dispensing accuracy of ±0,005 ml or better is required.
A reservoir for the titrating solution. It should be fitted with a guard tube containing soda lime
or other carbon dioxide absorbing material.
4.3.2 Glass indicator electrode
A glass electrode specifically designed for non-aqueous titrations is recommended.
The electrode shall be connected to the potentiometer by means of a suitably screened cable
such that the resistance between the screening and the entire length of the electrical
connection is greater than 50 000 MΩ.
– 10 – IEC 62021-3:2014 © IEC 2014
4.3.3 Reference electrode
The electrode shall be made of glass and shall be reserved for non-aqueous titrations.
Certain alternative electrode-electrolyte combinations have been found to give satisfactory
results, although the precision using these alternatives has not been determined. Combined
electrodes may be used provided they otherwise conform to this standard and have at least a
similar speed of response.
4.3.4 Stirrer
The stirrer should have a variable speed and be fitted with a propeller, paddle or magnetic bar
of chemically inert surface material. It shall be electrically grounded to avoid any change in
the meter reading during the course of the titration.
4.3.5 Titration vessel
This should be as small as possible, sufficient to contain the solvent, sample, stirrer and
electrodes and be inert to the reagents. Glass vessels are preferred to prevent build-up of
electrostatic charge.
4.3.6 Titration stand
This should comprise a suitable stand to support the beaker, electrodes, stirrer and burette.
4.4 Sampling
Samples shall be taken following the procedure given in IEC 60475.
Ensure that the test portion is representative by thoroughly mixing, as any sediment present
may be acidic or have adsorbed acidic material from the liquid phase.
4.5 Preparation and maintenance of electrode system
4.5.1 Preparation
Although electrodes are not particularly fragile, they should be handled carefully at all times.
Rinse the electrodes with 2-propanol and finally with de-ionized water.
Following each titration immerse the electrodes in de-ionized water to remove any surplus
electrolyte adhering to the outside of the electrode and allow excess water to drain off. The
immersion time should be sufficient to prevent any memory effects on subsequent titrations.
When in use, any plug that is present on the reference electrode should be removed and the
electrolyte level in the electrode kept above that of liquid in the titration vessel to prevent
entry of contaminants into the electrode.
4.5.2 Maintenance
4.5.2.1 Glass electrode
Clean the electrode weekly by immersing the tip in 0,1 mol/l hydrochloric acid for 12 h
followed by washing with de-ionized water. If more aggressive cleaning is required, immerse
the electrode tip in cleaning solution (see 4.2.7) for 5 min and follow this by thorough washing
with de-ionized water. This treatment should be carried out on a monthly basis when the
electrode is in regular use.
When not in use, immerse the lower half of the electrode in de-ionized water. Do not allow the
electrode to dry out. If this occurs it may be possible to reactivate by immersing in cleaning
solution (see 4.2.7) as detailed above.
4.5.2.2 Reference electrode
Drain and fill the electrode with electrolyte solution (see 4.2.5) according to the
manufacturer’s recommendations. When using the sleeve-type electrode, carefully remove the
ground-glass sleeve and thoroughly wipe both ground-glass sleeve surfaces. Replace the
sleeve loosely and allow a few drops of electrolyte to drain through to flush the ground-glass
joint and to wet the ground surfaces thoroughly with electrolyte. Set the sleeve in place and
refill with electrolyte (see 4.2.5).
When not in use, immerse the electrode in electrolyte (see 4.2.5) keeping the level of the
electrolyte in the electrode above that of the immersion fluid level. The filling apertures should
be covered during storage.
The electrode should be cleaned as necessary (at least weekly) by flushing with de-ionized
water.
4.6 Calibration
4.6.1 Calibration of pH titrimeter
Determine the pH reading for the buffer solutions (see 4.2.6) on a daily basis. The value of
the titration end-point of pH 11,5 is then extrapolated and shall be entered into the
instrumental programme.
The linearity and slope of the potentiometric titrator over the pH range 4 to 11 should comply
with the electrode manufacturer’s tolerances.
Temperature correction shall be applied.
Owing to the significant effect of temperature on the pH of the buffer solutions (see 4.2.6), it
is desirable to keep the temperature as close to the buffer manufacturer's calibration
temperature as possible.
4.6.2 Settings for the potentiometric instrument
Set a potential for an end point titration (usually between –50 mV and –100 mV), which allows
the recording of the whole titration curve. For this purpose, use the titration procedure as
described in 4.7.3 with the addition of 100 µl of 0,1 mol/l hydrochloric acid (see 4.2.4). See
Figure 1.
– 12 – IEC 62021-3:2014 © IEC 2014
DET U001
400,000
350,000
300,000
350,250,000000
200,000
150,000
100,000
50,000
0,000
–50,000
0,000 0,050 0,100 0,150 0,200 0,250 0,300
V/ml
IEC 043714
Key
Black dotted line first derivative
Red dotted line exact volume of titrant added (ml).
Figure 1 – Potentiometric titration curve
Most instruments calculate automatically the first derivative of the potential titration curve and
the exact volume of titrant added.
4.7 Procedure
4.7.1 General
Set up the apparatus in accordance with the manufacturer’s instructions.
Rinse and fill the burette with an alcoholic solution of potassium hydroxide between 0,01 mol/l
and 0,05 mol/l (see 4.2.2).
Standardize the 0,01 mol/l or 0,05 mol/l alcoholic potassium hydroxide solution at least every
two weeks against potassium hydrogen phthalate (see 4.7.2).
Carry out a blank titration on the solvent (see 4.7.3) each day and after changing to a fresh
batch of solvent.
Prepare and titrate a sample of the non-mineral insulating oil against alcoholic potassium
hydroxide (see 4.7.4).
4.7.2 Standardization of alcoholic potassium hydroxide solution
Standardize the alcoholic potassium hydroxide solution potentiometrically against 0,1 g to
0,16 g of the potassium hydrogen phthalate, weighed to an accuracy of 0,0002 g and
dissolved in approximately 100 ml of carbon dioxide free water.
Depending on the capacity of the titration vessel, the amount of potassium hydrogen
phthalate may need to be less than 0,1 g, with a smaller volume of water used to dissolve it.
The volume of water shall be enough to dissolve the phthalate and to ensure the complete
immersion of the electrode bulb.
U/mV
Calculate the molarity to the nearest 0,0005, expressed as mol/l, using the following formula.
1000 × m × p
Molarity =
204,23 ×V
(1)
where
m is the mass of potassium hydrogen phthalate in g;
p is the purity of potassium hydrogen phthalate;
204,23 is the molecular weight of potassium hydrogen phthalate, in g/mol;
V is the volume of alcoholic KOH solution (see 4.2.2) used to titrate the solution, in ml.
Alternatively, standard 0,1 mol/l acid may be used to standardize the alcoholic KOH (see
4.2.4).
V × M
A A
Molarity =
V
B
(2)
where
V is the volume of 0,1 mol/l standard hydrochloric acid used to titrate the solution, in ml;
A
M is the molarity of the standard hydrochloric acid in mol/l;
A
V is the volume of potassium hydroxide solution, in ml.
B
4.7.3 Blank titration
Perform a blank titration in duplicate as in 4.7.4, on 20 ml ± 0,1 ml of the solvent (see 4.2.3)
daily and after changing to a fresh batch of solvent.
Blank titrations shall be continued until two consecutive titrations differ by no more than
0,005 ml, based on 20 ml of solvent and the mean of these is calculated as V (see 4.8).
Where a higher solvent volume than 20 ml is required because of apparatus constraints, the
same volume of solvent shall be used for the sample titration.
High values may arise from carbon dioxide absorption or inherent 2-propanol acidity. If the
blank value is greater than 0,06 ml (based on 20 ml of solvent), steps shall be taken to
remove the cause of the high values.
4.7.4 Sample titration
Prepare the sample for titration as described in 4.4 and weigh 5 g ± 0,1 g of the non-mineral
insulating oil to the nearest 0,01 g into the titration vessel. Add 20 ml ± 0,1 ml of titration
solvent (see 4.2.3).
The amount of solvent added may depend on the testing device used, the volume and shape
of vessel, etc. Add an amount of titration solvent sufficient to ensure the complete immersion
of electrode’s bulb.
Place the titration vessel on the titration stand and stir the solution until the sample has
dissolved and the pH reading is constant, taking care to limit the speed of stirring to avoid
spattering and/or stirring air into the solution.
– 14 – IEC 62021-3:2014 © IEC 2014
Carry out the titration with a solution between 0,01 mol/l and 0,05 mol/l potassium hydroxide,
following the instrument manufacturer’s recommendations, to an end-point of pH 11,5 or to
the determined end-point potential (see 4.6.2).
NOTE Dynamic titrant addition is preferred to reduce the overall analysis time.
If the titration time exceeds 15 min, it may be necessary to prevent carbon dioxide absorption
by blanketing the solution with nitrogen.
On completion of the titration, record the burette reading V (see 4.8) at the pH reading of 10
or determine the value of the first derivative of the potentiometric titration curve (see Figure
1).
Rinse the electrodes and burette tip with titration solvent (see 4.2.3). Re-hydrate the glass
electrode by immersing the bulb in de-ionized water (see 4.5.1) and allow excess water to
drain off. Where oxidized oil is analysed, the electrode should be immersed in de-ionized
water containing a few drops of hydrochloric acid, followed by rinsing in de-ionized water.
If further titrations are not to be carried out immediately, the electrodes shall be stored in
the de-ionized water.
4.8 Calculation of result
Calculate the acidity to the nearest 0,01, expressed as mg KOH/g of oil using the following
formula:
(V −V )× M ×56,1
1 0
Acidity =
m
(3)
where
V is the volume of alcoholic KOH solution (see 4.2.2) used to titrate the test portion, in ml;
V is the volume of alcoholic KOH solution (see 4.2.2) used for blank titration, in ml;
M is the molarity of alcoholic KOH solution (see 4.2.2), in mol/l;
56,1 is the molecular weight of potassium hydroxide, in g/mol;
m is the mass of the test portion used in g.
4.9 Precision
4.9.1 Repeatability
The difference between successive test results obtained by the same operator with the same
apparatus under constant operating conditions on identical test material would, in the long
run, in the normal and correct operation of the test method, exceed the values shown below
only in one case in 20:
– unused synthetic esters and silicones: 6 % of the mean value;
– unused natural esters: 9 % of the mean value;
– used oils: 12 % of the mean value.
NOTE The repeatability values for unused oils only apply where the result is significantly above the quantification
limit, which has been established as 0,014 mg KOH/g oil.
4.9.2 Reproducibility
The difference between two single and independent results obtained by different operators
working in different laboratories on identical test material would, in the long run, in the normal
and correct operation of the test method, exceed the values shown below only in one case in
20:
– unused oils: 28 % of the mean value;
– used oils: 35 % of the mean value.
NOTE Repeatability and reproducibility limits were established in accordance with ISO 5725 for used oil. Those
for unused oil have been taken from ISO 6619.
4.10 Report
The test report shall contain at least the following information:
– the type and identification of the product tested;
– a reference to this standard;
– the result of the test (see 4.8) expressed to the nearest 0,01 mg KOH/g of oil;
– any deviation, by agreement or otherwise, from the procedure specified;
– the date of the test.
5 Method B: Colourimetric titration
5.1 Principle
The test portion is dissolved in a specified solvent and titrated colourimetrically with alcoholic
potassium hydroxide to a specified colour using Alkali Blue 6B indicator.
NOTE Colourimetric titration may not be suitable to highly coloured oils.
5.2 Reagents
5.2.1 General
Only reagents of recognized analytical grade and de-ionized water or water of equivalent
purity shall be used.
5.2.2 Titration reagent
Standard alcoholic solution between 0,01 mol/l and 0,05 mol/l potassium hydroxide (KOH).
EXAMPLE Preparation of 0,01 mol/l potassium hydroxide in 2-propanol.
Add 0,6 g of potassium hydroxide to 1 000 ml ± 10 ml of 2-propanol. Boil gently for 10 min to
effect solution. Cool and stopper the flask.
Allow the solution to stand in the dark for 2 days and then filter the supernatant liquid through
a 5 µm membrane filter. Store in a suitable amber glass bottle.
The concentration of this solution is approximately 0,01 mol/l and shall be standardized as
described in 5.5.2.
Commercial alcoholic potassium hydroxide solution may be used, if necessary diluting to
0,01 mol/l with 2-propanol. This shall be standardized as described in 5.5.2.
Store and use in such a manner that the solution is protected from atmospheric carbon
dioxide and in such a way that it does not come into contact with cork, rubber or saponifiable
stopcock grease. The solution may be protected by inert gas or by means of a guard tube
containing soda-lime absorbent.
NOTE 1 For oils with high acidity, which may give an extended titration time, it may be helpful to carry out a pre-
test using 0,1 mol/l potassium hydroxide titrant to determine a suitable titrant concentration.
– 16 – IEC 62021-3:2014 ©
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...