EN IEC 62209-3:2019
(Main)Measurement procedure for the assessment of specific absorption rate of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Part 3: Vector measurement-based systems (Frequency range of 600 MHz to 6 GHz)
Measurement procedure for the assessment of specific absorption rate of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Part 3: Vector measurement-based systems (Frequency range of 600 MHz to 6 GHz)
IEC 62209-3: 2019 specifies measurement protocols and test procedures for the reproducible measurement of peak spatial-average specific absorption rate (psSAR) induced inside a simplified model of a human head or body by radio-frequency (RF) transmitting devices, with a specified measurement uncertainty. Requirements are provided for psSAR assessment using vector measurement-based systems. Such systems determine the psSAR by three-dimensional (3D) field reconstruction within the volume of interest in accordance with the requirements herein for the measurement system, calibration, uncertainty assessment and validation methods. The protocols and procedures apply for the psSAR assessments covering a significant majority of people including children during use of wireless communication devices operated in close proximity to the head or body. This document is applicable to wireless communication devices intended to be used at a position near the human head or body at distances up to and including 200 mm. This document may be employed to evaluate SAR compliance of different types of wireless communication devices used next to the ear, in front of the face, mounted on the body, combined with other RF-transmitting or non-transmitting devices or accessories (e.g. belt-clip), or embedded in garments. The overall applicable frequency range is from 600 MHz to 6 GHz. The system validation procedures provided within this document cover frequencies from 600 MHz to 6 GHz. With a vector measurement-based system this document can be employed to evaluate SAR compliance of different types of wireless communication devices. The wireless communication device categories covered include but are not limited to mobile telephones, cordless microphones, auxiliary broadcast devices and radio transmitters in personal computers, desktop and laptop devices, multi-band, multi-antenna, and push-to-talk devices. Key Words: Human Exposure, Hand-Held and Body Mounted Wireless Communication Devices.
Messverfahren für die Beurteilung der spezifischen Absorptionsrate bei der Exposition von Personen gegenüber hochfrequenten Feldern von handgehaltenen und am Körper getragenen schnurlosen Kommunikationsgeräten - Teil 3: Auf Vektormessungen basierende Systeme (Frequenzbereich von 600 MHz bis 6 GHz)
Procédure de mesure pour l'évaluation du débit d'absorption spécifique de l'exposition humaine aux champs radiofréquence produits par les dispositifs de communications sans fil tenus à la main ou portes près du corps - Partie 3: Systèmes basés sur la mesure vectorielle (plage de fréquences comprise entre 600 MHz et 6 GHz)
IEC 62209-3: 2019 spécifie les protocoles et procédures d'essai pour le mesurage reproductible du DAS maximal moyenné dans l'espace (psSAR – peak spatial-average specific absorption rate) induit à l'intérieur d'un modèle simplifié de tête ou de corps humain par des dispositifs d'émission de radiofréquence (RF), avec une incertitude de mesure spécifiée. Des exigences sont fournies concernant l’évaluation du psSAR utilisant des systèmes basés sur la mesure vectorielle. Ces systèmes déterminent le psSAR par reconstruction de champ tridimensionnel (3D) à l'intérieur du volume à l'étude conformément aux exigences du présent document relatives au système de mesure, à l'étalonnage, à l'évaluation de l'incertitude et aux méthodes de validation. Les protocoles et procédures s'appliquent pour les évaluations du psSAR couvrant une grande majorité de personnes, y compris les enfants, lors de l'utilisation de dispositifs de communications à proximité de la tête ou du corps. Le présent document s'applique aux dispositifs de communications sans fil destinés à être utilisés proches de la tête ou du corps humain, à des distances allant jusqu'à 200 mm inclus. Le présent document peut être utilisé pour évaluer la conformité du DAS de différents types de dispositifs de communications sans fil utilisés proches de l'oreille, devant le visage, sur le corps, en combinaison avec d'autres dispositifs ou accessoires de transmission RF ou pas (une attache de ceinture, par exemple) ou intégrés aux vêtements. La plage de fréquences globale applicable est comprise entre 600 MHz et 6 GHz. Les procédures de validation du système indiquées dans le présent document couvrent les fréquences comprises entre 600 MHz et 6 GHz. Avec un système basé sur la mesure vectorielle, le présent document peut être utilisé pour évaluer la conformité du DAS de différents types de dispositifs de communications sans fil. Les catégories de dispositifs de communications sans fil incluent, entre autres, les téléphones mobiles, les micros sans fil, les dispositifs de diffusion auxiliaires et les émetteurs radio dans les ordinateurs personnels, les ordinateurs de bureau et les ordinateurs portables, ainsi que les dispositifs à plusieurs bandes, à plusieurs antennes et à boutons-poussoirs. Mots clés: Exposition Humaine, Dispositifs de communications sans fil tenus à la main ou portés près du corps
Merilni postopki za ocenjevanje stopnje specifične absorpcije pri izpostavljenosti ljudi elektromagnetnim sevanjem brezžičnih komunikacijskih naprav, ki se držijo v roki ali pritrdijo na telo - 3. del: Sistemi vektorskega merjenja (frekvenčno območje od 600 MHz do 6 GHz)
General Information
Standards Content (Sample)
SLOVENSKI STANDARD
01-junij-2021
Merilni postopki za ocenjevanje stopnje specifične absorpcije pri izpostavljenosti
ljudi elektromagnetnim sevanjem brezžičnih komunikacijskih naprav, ki se držijo v
roki ali pritrdijo na telo - 3. del: Sistemi vektorskega merjenja (frekvenčno območje
od 600 MHz do 6 GHz)
Measurement procedure for the assessment of specific absorption rate of human
exposure to radio frequency fields from hand-held and body-mounted wireless
communication devices - Part 3: Vector measurement-based systems (Frequency range
of 600 MHz to 6 GHz)
Ta slovenski standard je istoveten z: EN IEC 62209-3:2019
ICS:
13.280 Varstvo pred sevanjem Radiation protection
33.050.10 Telefonska oprema Telephone equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN IEC 62209-3
NORME EUROPÉENNE
EUROPÄISCHE NORM
November 2019
ICS 33.060.20
English Version
Measurement procedure for the assessment of specific
absorption rate of human exposure to radio frequency fields from
hand-held and body-mounted wireless communication devices -
Part 3: Vector measurement-based systems (Frequency range
of 600 MHz to 6 GHz)
(IEC 62209-3:2019)
Procédure de mesure pour l'évaluation du débit Messverfahren für die Beurteilung der spezifischen
d'absorption spécifique de l'exposition humaine aux champs Absorptionsrate bei der Exposition von Personen
radiofréquence produits par les dispositifs de gegenüber hochfrequenten Feldern von handgehaltenen
communications sans fil tenus à la main ou portes près du und am Körper getragenen schnurlosen
corps - Partie 3: Systèmes basés sur la mesure vectorielle Kommunikationsgeräten - Teil 3: Auf Vektormessungen
(plage de fréquences comprise entre 600 MHz et 6 GHz) basierende Systeme (Frequenzbereich von 600 MHz bis 6
(IEC 62209-3:2019) GHz)
(IEC 62209-3:2019)
This European Standard was approved by CENELEC on 2019-10-29. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2019 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 62209-3:2019 E
European foreword
The text of document 106/494/FDIS, future edition 1 of IEC 62209-3, prepared by IEC/TC 106
"Methods for the assessment of electric, magnetic and electromagnetic fields associated with human
exposure" was submitted to the IEC-CENELEC parallel vote and approved by CENELEC as
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2020-07-29
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2022-10-29
document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
Endorsement notice
The text of the International Standard IEC 62209-3:2019 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards
indicated:
ISO/IEC 17025 NOTE Harmonized as EN ISO/IEC 17025
ISO 3611:2010 NOTE Harmonized as EN ISO 3611:2010 (not modified)
ISO/IEC 17043 NOTE Harmonized as EN ISO/IEC 17043
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1 Where an International Publication has been modified by common modifications, indicated by (mod), the relevant
EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is available here:
www.cenelec.eu.
Publication Year Title EN/HD Year
IEC 62209-1 2016 Measurement procedure for the EN 62209-1 2016
assessment of specific absorption rate of
human exposure to radio frequency fields
from hand-held and body-mounted
wireless communication devices - Part 1:
Devices used next to the ear (Frequency
range of 300 MHz to 6 GHz)
IEC 62209-2 2010 Human exposure to radio frequency fields EN 62209-2 2010
from hand-held and body-mounted
wireless communication devices - Human
models, instrumentation, and procedures -
Part 2: Procedure to determine the specific
absorption rate (SAR) for wireless
communication devices used in close
proximity to the human body (frequency
range of 30 MHz to 6 GHz)
IEC 62479 - Assessment of the compliance of low- EN 62479 -
power electronic and electrical equipment
with the basic restrictions related to human
exposure to electromagnetic fields (10
MHz to 300 GHz)
IEC TR 62630 2010 Guidance for evaluating exposure from - -
multiple electromagnetic sources
ISO/IEC Guide 98-1 2009 Uncertainty of measurement – Part 1: - -
Introduction to the expression of
uncertainty in measurement
ISO/IEC Guide 98-3 - Uncertainty of measurement - Part 3: - -
Guide to the expression of uncertainty in
measurement (GUM:1995)
IEC/IEEE 62704-1 - Determining the peak spatial-average - -
specific absorption rate (SAR) in the
human body from wireless communications
devices, 30 MHz to 6 GHz - Part 1:
General requirements for using the finite
difference time-domain (FDTD) method for
SAR calculations
IEC 62209-3 ®
Edition 1.0 2019-09
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
HORIZONTAL STANDARD
NORME HORIZONTALE
Measurement procedure for the assessment of specific absorption rate of
human exposure to radio frequency fields from hand-held and body-mounted
wireless communication devices –
Part 3: Vector measurement-based systems (Frequency range of 600 MHz
to 6 GHz)
Procédure de mesure pour l'évaluation du débit d'absorption spécifique
de l'exposition humaine aux champs radiofréquence produits par les dispositifs
de communications sans fil tenus à la main ou portes près du corps –
Partie 3: Systèmes basés sur la mesure vectorielle (plage de fréquences
comprise entre 600 MHz et 6 GHz)
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 33.060.20 ISBN 978-2-8322-7355-5
– 2 – IEC 62209-3:2019 © IEC 2019
CONTENTS
FOREWORD . 9
INTRODUCTION . 11
1 Scope . 12
2 Normative references . 12
3 Terms and definitions . 13
4 Symbols and abbreviated terms . 14
5 Overview of the measurement procedure . 14
6 Measurement system specifications . 17
6.1 General requirements . 17
6.2 Phantom specifications . 19
6.2.1 Head phantom specifications – shell . 19
6.2.2 Body phantom specifications – shell . 19
6.2.3 Tissue-equivalent medium properties . 19
6.3 Measurement system requirements . 19
6.3.1 General . 19
6.3.2 Scanning measurement system specifications . 19
6.3.3 Array measurement system specifications . 20
6.4 Device holder specification . 21
6.5 Reconstruction algorithm and peak spatial-averaging specifications . 22
7 Protocol for SAR assessments . 22
7.1 Measurement preparation . 22
7.1.1 General . 22
7.1.2 Preparation of tissue-equivalent medium . 22
7.1.3 System check . 23
7.1.4 Preparation of the device under test (DUT) . 23
7.1.5 Operating modes . 23
7.1.6 Position of the DUT in relation to the phantom . 23
7.1.7 Positions of the DUT in relation to the flat phantom for large DUT . 23
7.1.8 Test frequencies for DUT . 24
7.2 Tests to be performed . 24
7.3 General measurement procedure . 25
7.3.1 General . 25
7.3.2 Measurement procedure for scanning systems . 25
7.3.3 Measurement procedure for array systems . 26
7.4 SAR measurements for simultaneous transmission . 26
7.4.1 General . 26
7.4.2 SAR measurements for uncorrelated signals . 27
7.4.3 SAR measurements for correlated signals . 31
8 Measurement uncertainty estimation. 32
8.1 General . 32
8.2 Requirements on the measurement uncertainty evaluation . 32
8.3 Description of measurement uncertainty models . 33
8.3.1 General . 33
8.3.2 Uncertainty models for array measurement system and scanning
measurement systems . 34
8.3.3 Example uncertainty budget templates . 35
IEC 62209-3:2019 © IEC 2019 – 3 –
9 Measurement report . 39
Annex A (normative) Phantom specifications . 40
A.1 SAM phantom specifications . 40
A.1.1 Justification . 40
A.1.2 SAM phantom geometry.
...
Questions, Comments and Discussion