ASTM D5996-16
(Test Method)Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography
Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography
SIGNIFICANCE AND USE
5.1 In the power-generation industry, high-purity water is used to reduce corrosion from anions, such as sulfate, chloride, and fluoride. These anions are known to be detrimental to materials of construction used in steam generators, reactor vessel internals and recirculation piping, heat exchangers, connective piping, and turbines. Most electric generating plants try to control these anions to
5.2 These anions and others cause low product yields in semiconductor manufacturing. They are also monitored and controlled at similarly low levels as in the electric power industry.
5.3 Low molecular weight organic acids (acetate, formate, propionate) have been detected in steam generator feed water. These low molecular weight organic materials are believed to be high-temperature degradation products of chemicals used to control cycle water pH and organic contaminants in cycle makeup water.
5.4 In the semiconductor industry, anion contaminants may come from the breakdown of low molecular weight organic materials by ultraviolet light radiation, which is frequently used to produce bacteria-free water. These organic compounds may also contribute to low product yield.
5.5 The production of high-purity water for process makeup and use frequently employs the use of demineralizers to remove unwanted anion contaminants. Also in the electric power industry, demineralizers are used in the process stream to maintain low levels of these contaminants. As such, it is important to monitor this process to ensure that water quality standards are being met. These processes can be monitored for the above-mentioned anions.
5.6 On-line measurements of these contaminants provide a greater degree of protection of the processes by allowing for frequent on-line measurement of these species. Early detection of contaminant ingress allows for quicker corrective action to locate, reduce, or eliminate, or combination thereof, the source. Grab samples will not provide the same level of pro...
SCOPE
1.1 This test method covers on-line analysis of high-purity water by the ion chromatography technique. This test method is applicable for measuring various anionic contaminants in high-purity water, typically in the range of 0.01 to 100 μg/L. This test method is used to determine the concentration of acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in a continuously flowing sample. The range of the test method is only as good as the reagent water available for preparing standards. At extremely low concentrations,
1.2 Online sample analysis of flowing streams does not lend itself to collaborative studies due to the nature of the sample and the possibility of contamination that may result from handling the sample as part of the collaborative study. Therefore this standard test method is not based on the results of a collaborative study but is intended to provide the best possible guidance for doing this type of analysis.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
- Status
- Published
- Publication Date
- 14-Feb-2016
- Technical Committee
- D19 - Water
Relations
- Effective Date
- 15-Feb-2016
- Effective Date
- 01-May-2020
- Effective Date
- 01-Aug-2018
- Effective Date
- 01-Aug-2018
- Effective Date
- 01-Feb-2017
- Effective Date
- 15-Feb-2016
- Effective Date
- 15-Jun-2012
- Effective Date
- 15-Jun-2011
- Effective Date
- 01-May-2011
- Effective Date
- 01-Feb-2011
- Effective Date
- 01-Dec-2010
- Effective Date
- 01-Mar-2010
- Effective Date
- 01-Oct-2008
- Effective Date
- 15-Jan-2008
- Effective Date
- 01-Dec-2007
Overview
ASTM D5996-16: Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography is a critical standard developed by ASTM International. This standard provides a comprehensive methodology for the on-line measurement of anionic contaminants in high-purity water using ion chromatography. The method is essential for industries where trace anionic impurities can significantly affect product quality and process performance, including power generation and semiconductor manufacturing.
ASTM D5996-16 specifies procedures for the on-line analysis of contaminants such as acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in continuously flowing high-purity water samples, typically within the range of 0.01 to 100 µg/L.
Key Topics
On-Line Ion Chromatography: The standard outlines procedures for continuous, real-time monitoring of anionic species using automated ion chromatography instruments. This minimizes the risk of sample contamination and provides early warning of contaminant ingress.
Control of Anions: Reducing and monitoring anionic contaminants such as sulfate, chloride, and fluoride is crucial for minimizing corrosion and material degradation in power plant infrastructure and ensuring high yields in semiconductor production.
Water Quality and Demineralization: The test method is used to assess the performance of demineralizers and other water purification systems, ensuring compliance with stringent water quality requirements.
Calibration and Quality Control: Guidance is provided for instrument calibration, validation, use of laboratory control samples, and matrix spike procedures to assure accuracy and reliability of analytical results.
Precision and Interferences: The document discusses the challenges of measuring very low concentrations of anions, potential interferences, and the importance of maintaining sample integrity and using high-purity reagents and materials.
Applications
Power Generation: The method is extensively utilized in power plants to continuously monitor and control corrosive anionic species in boiler feedwater, steam generator systems, reactor internals, and process piping. Quick detection allows for immediate corrective actions, preventing costly damage and outages.
Semiconductor Manufacturing: In the semiconductor industry, even trace levels of anionic contaminants can reduce product yields. On-line analysis ensures ultrapure water used in critical cleaning and manufacturing is kept free from performance-limiting impurities.
Process Water Quality Assurance: Industries relying on high-purity water systems employ this standard to monitor water quality in real time, maintain system integrity, and avoid manual sampling errors or delays.
Environmental and Laboratory Settings: This method is referenced in quality assurance programs for laboratory water and environmental monitoring applications, ensuring compliance with industry and regulatory requirements.
Related Standards
To achieve robust water quality monitoring and ensure compatibility with ASTM D5996-16, consideration should be given to the following related standards:
- ASTM D1066: Practice for Sampling Steam
- ASTM D1129: Terminology Relating to Water
- ASTM D1192: Guide for Equipment for Sampling Water and Steam in Closed Conduits
- ASTM D1193: Specification for Reagent Water
- ASTM D2777: Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
- ASTM D3370: Practices for Sampling Water from Closed Conduits
- ASTM D3864: Guide for On-Line Monitoring Systems for Water Analysis
- ASTM D5542: Test Methods for Trace Anions in High Purity Water by Ion Chromatography
- ASTM D4453: Practice for Handling of High Purity Water Samples
Practical Value
ASTM D5996-16 ensures continuous and reliable monitoring of anionic contaminants in high-purity water, supporting high operational standards in power generation, electronics, and other sectors where water purity is critical. By enabling rapid detection and correction of contamination events, the test method helps maintain system integrity, extend equipment life, and increase overall process efficiency. The on-line ion chromatography approach outlined in this standard also aids in reducing sample handling risks, minimizing contamination, and providing consistent, accurate data for regulatory compliance and quality assurance.
Keywords: ASTM D5996-16, anionic contaminants, high-purity water, on-line ion chromatography, power industry, semiconductor industry, water quality monitoring, demineralization, trace anions, water analysis.
Buy Documents
ASTM D5996-16 - Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography
REDLINE ASTM D5996-16 - Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography
Get Certified
Connect with accredited certification bodies for this standard

ECOCERT
Organic and sustainability certification.

Eurofins Food Testing Global
Global leader in food, environment, and pharmaceutical product testing.

Intertek Bangladesh
Intertek certification and testing services in Bangladesh.
Sponsored listings
Frequently Asked Questions
ASTM D5996-16 is a standard published by ASTM International. Its full title is "Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography". This standard covers: SIGNIFICANCE AND USE 5.1 In the power-generation industry, high-purity water is used to reduce corrosion from anions, such as sulfate, chloride, and fluoride. These anions are known to be detrimental to materials of construction used in steam generators, reactor vessel internals and recirculation piping, heat exchangers, connective piping, and turbines. Most electric generating plants try to control these anions to 5.2 These anions and others cause low product yields in semiconductor manufacturing. They are also monitored and controlled at similarly low levels as in the electric power industry. 5.3 Low molecular weight organic acids (acetate, formate, propionate) have been detected in steam generator feed water. These low molecular weight organic materials are believed to be high-temperature degradation products of chemicals used to control cycle water pH and organic contaminants in cycle makeup water. 5.4 In the semiconductor industry, anion contaminants may come from the breakdown of low molecular weight organic materials by ultraviolet light radiation, which is frequently used to produce bacteria-free water. These organic compounds may also contribute to low product yield. 5.5 The production of high-purity water for process makeup and use frequently employs the use of demineralizers to remove unwanted anion contaminants. Also in the electric power industry, demineralizers are used in the process stream to maintain low levels of these contaminants. As such, it is important to monitor this process to ensure that water quality standards are being met. These processes can be monitored for the above-mentioned anions. 5.6 On-line measurements of these contaminants provide a greater degree of protection of the processes by allowing for frequent on-line measurement of these species. Early detection of contaminant ingress allows for quicker corrective action to locate, reduce, or eliminate, or combination thereof, the source. Grab samples will not provide the same level of pro... SCOPE 1.1 This test method covers on-line analysis of high-purity water by the ion chromatography technique. This test method is applicable for measuring various anionic contaminants in high-purity water, typically in the range of 0.01 to 100 μg/L. This test method is used to determine the concentration of acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in a continuously flowing sample. The range of the test method is only as good as the reagent water available for preparing standards. At extremely low concentrations, 1.2 Online sample analysis of flowing streams does not lend itself to collaborative studies due to the nature of the sample and the possibility of contamination that may result from handling the sample as part of the collaborative study. Therefore this standard test method is not based on the results of a collaborative study but is intended to provide the best possible guidance for doing this type of analysis. 1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
SIGNIFICANCE AND USE 5.1 In the power-generation industry, high-purity water is used to reduce corrosion from anions, such as sulfate, chloride, and fluoride. These anions are known to be detrimental to materials of construction used in steam generators, reactor vessel internals and recirculation piping, heat exchangers, connective piping, and turbines. Most electric generating plants try to control these anions to 5.2 These anions and others cause low product yields in semiconductor manufacturing. They are also monitored and controlled at similarly low levels as in the electric power industry. 5.3 Low molecular weight organic acids (acetate, formate, propionate) have been detected in steam generator feed water. These low molecular weight organic materials are believed to be high-temperature degradation products of chemicals used to control cycle water pH and organic contaminants in cycle makeup water. 5.4 In the semiconductor industry, anion contaminants may come from the breakdown of low molecular weight organic materials by ultraviolet light radiation, which is frequently used to produce bacteria-free water. These organic compounds may also contribute to low product yield. 5.5 The production of high-purity water for process makeup and use frequently employs the use of demineralizers to remove unwanted anion contaminants. Also in the electric power industry, demineralizers are used in the process stream to maintain low levels of these contaminants. As such, it is important to monitor this process to ensure that water quality standards are being met. These processes can be monitored for the above-mentioned anions. 5.6 On-line measurements of these contaminants provide a greater degree of protection of the processes by allowing for frequent on-line measurement of these species. Early detection of contaminant ingress allows for quicker corrective action to locate, reduce, or eliminate, or combination thereof, the source. Grab samples will not provide the same level of pro... SCOPE 1.1 This test method covers on-line analysis of high-purity water by the ion chromatography technique. This test method is applicable for measuring various anionic contaminants in high-purity water, typically in the range of 0.01 to 100 μg/L. This test method is used to determine the concentration of acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in a continuously flowing sample. The range of the test method is only as good as the reagent water available for preparing standards. At extremely low concentrations, 1.2 Online sample analysis of flowing streams does not lend itself to collaborative studies due to the nature of the sample and the possibility of contamination that may result from handling the sample as part of the collaborative study. Therefore this standard test method is not based on the results of a collaborative study but is intended to provide the best possible guidance for doing this type of analysis. 1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
ASTM D5996-16 is classified under the following ICS (International Classification for Standards) categories: 71.040.50 - Physicochemical methods of analysis. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM D5996-16 has the following relationships with other standards: It is inter standard links to ASTM D5996-05(2009), ASTM D1129-13(2020)e2, ASTM D1066-18, ASTM D1066-18e1, ASTM D4453-17, ASTM D4453-16, ASTM D2777-12, ASTM D1066-11, ASTM D5810-96(2011), ASTM D4453-11, ASTM D3370-10, ASTM D1129-10, ASTM D3370-08, ASTM D2777-08, ASTM D3370-07. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM D5996-16 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D5996 − 16
Standard Test Method for
Measuring Anionic Contaminants in High-Purity Water by
On-Line Ion Chromatography
This standard is issued under the fixed designation D5996; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2.1 ASTM Standards:
1.1 This test method covers on-line analysis of high-purity
D1066Practice for Sampling Steam
water by the ion chromatography technique. This test method
D1129Terminology Relating to Water
is applicable for measuring various anionic contaminants in
D1192Guide for Equipment for Sampling Water and Steam
high-purity water, typically in the range of 0.01 to 100 µg/L.
in Closed Conduits (Withdrawn 2003)
This test method is used to determine the concentration of
D1193Specification for Reagent Water
acetate, formate, chloride, fluoride, phosphate, nitrate, and
D2777Practice for Determination of Precision and Bias of
sulfate in a continuously flowing sample.The range of the test
Applicable Test Methods of Committee D19 on Water
method is only as good as the reagent water available for
D3370Practices for Sampling Water from Closed Conduits
preparing standards. At extremely low concentrations, <1.0
D3864Guide for On-Line Monitoring Systems for Water
µg/L, preparing standards is difficult, and extra care must be
Analysis
taken in their preparation. The sample may have to be
D4453Practice for Handling of High Purity Water Samples
conditioned from higher pressures and temperatures to condi-
D5542Test Methods for TraceAnions in High Purity Water
tions that are suitable for use by on-line instruments.
by Ion Chromatography
D5810Guide for Spiking into Aqueous Samples
1.2 Onlinesampleanalysisofflowingstreamsdoesnotlend
itself to collaborative studies due to the nature of the sample D5847Practice for Writing Quality Control Specifications
for Standard Test Methods for Water Analysis
and the possibility of contamination that may result from
handling the sample as part of the collaborative study. There-
3. Terminology
fore this standard test method is not based on the results of a
3.1 Definitions:
collaborative study but is intended to provide the best possible
3.1.1 For definitions of terms used in this standard, refer to
guidance for doing this type of analysis.
Terminology D1129.
1.3 The values stated in SI units are to be regarded as
3.2 Definitions of Terms Specific to This Standard:
standard. No other units of measurement are included in this
3.2.1 analytical column, n—a column used to separate the
standard.
anions of interest.
3.2.2 analytical column set, n—a combination of one or
1.4 This standard does not purport to address all of the
more guard columns followed by one or more analytical
safety concerns, if any, associated with its use. It is the
columns.
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
3.2.3 anion suppressor device, n—a device that is placed
bility of regulatory limitations prior to use.
between the analytical columns and the detector.
3.2.3.1 Discussion—Its purpose is to inhibit detector re-
sponsetotheionicconstituentsintheeluant,soastolowerthe
This test method is under the jurisdiction ofASTM Committee D19 on Water
and is the direct responsibility of Subcommittee D19.03 on Sampling Water and For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Water-Formed Deposits,Analysis of Water for Power Generation and Process Use, contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
On-Line Water Analysis, and Surveillance of Water. Standards volume information, refer to the standard’s Document Summary page on
Current edition approved Feb. 15, 2016. Published June 2016. Originally the ASTM website.
approved in 1996. Last previous edition approved in 2009 as D5996–05 (2009). The last approved version of this historical standard is referenced on
DOI: 10.1520/D5996-16. www.astm.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D5996 − 16
detector background and at the same time enhance detector 5.2 These anions and others cause low product yields in
response to the ions of interest. semiconductor manufacturing. They are also monitored and
controlled at similarly low levels as in the electric power
3.2.4 breakthrough volume, n—the maximum sample vol-
industry.
ume that can be passed through a concentrator column before
the least tightly bound ion of interest is eluted. 5.3 Low molecular weight organic acids (acetate, formate,
propionate) have been detected in steam generator feed water.
3.2.4.1 Discussion—All of the columns in series contribute
These low molecular weight organic materials are believed to
to the overall capacity of the analytical column set.
behigh-temperaturedegradationproductsofchemicalsusedto
3.2.5 concentrator column, n—an ion exchange column
control cycle water pH and organic contaminants in cycle
used to concentrate the ions of interest and thereby increase
makeup water.
method sensitivity.
5.4 In the semiconductor industry, anion contaminants may
3.2.6 eluant, n—theionicmobilephaseusedtotransportthe
come from the breakdown of low molecular weight organic
sample through the analytical column.
materials by ultraviolet light radiation, which is frequently
3.2.7 guard column, n—acolumnusedbeforetheanalytical
used to produce bacteria-free water.These organic compounds
column to protect it from contaminants, such as particulate
may also contribute to low product yield.
matter or ionic species that may chemically foul the resins and
5.5 Theproductionofhigh-puritywaterforprocessmakeup
degrade their performance.
and use frequently employs the use of demineralizers to
3.2.8 ion chromatography, n—a form of liquid chromatog-
remove unwanted anion contaminants. Also in the electric
raphy in which ionic constituents are separated by ion ex-
power industry, demineralizers are used in the process stream
change followed by a suitable detection means.
to maintain low levels of these contaminants. As such, it is
important to monitor this process to ensure that water quality
3.2.9 resolution, n—the ability of an analytical column to
standards are being met.These processes can be monitored for
separate constituents under specific test conditions.
the above-mentioned anions.
4. Summary of Test Method
5.6 On-line measurements of these contaminants provide a
greater degree of protection of the processes by allowing for
4.1 A continuously flowing sample is injected into the
frequenton-linemeasurementofthesespecies.Earlydetection
instrument through a sample injection valve. The sample is
of contaminant ingress allows for quicker corrective action to
pumped through a concentrator column where the anions of
locate,reduce,oreliminate,orcombinationthereof,thesource.
interest are collected on ion-exchange resin. After a suitable
Grab samples will not provide the same level of protection
volume of sample has been passed through the concentrator
because of their intermittent nature and the longer time
column, sample flow is diverted and an eluant is pumped
required to obtain and then analyze the sample.
throughtheconcentratorcolumntoremovethetrappedanions.
This eluant then flows through an analytical column set where
5.7 Additionally, on-line monitoring significantly reduces
theanionsareseparatedbasedontheretentioncharacteristicof
the potential for contamination of high-purity water samples, a
each anion relative to the eluant used. The eluant stream
significant problem when sampling and testing high-purity
containing the anions of interest passes through a suppressor
water.
device where the cations from the eluant are exchanged for
hydrogen ions, converting the anions to their acid form. After
6. Interferences
the suppressor device, the eluant solution passes through a
6.1 When working with low concentration samples, blanks,
conductivity detector where the separated anions are detected.
and standards, contamination can be a serious problem. Ex-
Detectionlimitsfortheanionsareenhancedbecausetheanions
treme care must be exercised in all phases of this test method.
are in the acid form rather than the salt.
6.2 Improper sample line material or sample lines that have
4.2 The anions are identified based on the retention time as
not been properly conditioned can give results that may not be
compared to known standards. By measuring peak height or
truly representative of the process stream. Absorption/
area and comparing the detector response to known standards,
desorption of anions on sample line wall deposits can change
the anions can be quantified.
analytical results. Maintaining a minimum sample flow of 1.8
m/s(6ft/s)willminimizedepositbuilduponsamplelinewalls,
5. Significance and Use
reducing the potential for absorption/desorption of anions.
5.1 In the power-generation industry, high-purity water is
6.3 A single anion present at a concentration significantly
usedtoreducecorrosionfromanions,suchassulfate,chloride,
higher than other anions could mask closely adjacent peaks on
and fluoride. These anions are known to be detrimental to
the chromatogram.
materials of construction used in steam generators, reactor
vessel internals and recirculation piping, heat exchangers, 6.4 Low breakthrough volumes may be experienced when
connectivepiping,andturbines.Mostelectricgeneratingplants continuouslymonitoringforanionsinwaterthathashaditspH
try to control these anions to <1.0 µg/Lin the steam generator raised by ammonia, morpholine, or other additives. This
feed water. Some nuclear power plants have been able to interference can be eliminated by taking the sample from the
control anion contaminants at less than 0.02 µg/L. effluent of a cation resin column.
D5996 − 16
6.5 Identification of the anion is based on retention time of 8.3 Prepare eluant for the specific columns used and for the
the anion of interest. An interfering anion having the same anions of interest in accordance with manufacturer’s direc-
retention time as one of the anions of interest will result in tions.
erroneously high values for that anion.
8.4 Prepare regenerant for the specific suppressor used in
6.6 When loading a concentrator column, high concentra-
accordance with the manufacturer’s directions if required.
NOTE 1—There are numerous combinations of analytical columns,
tions of interfering anions may cause low breakthrough vol-
suppressors, eluants, and regenerants that may be used with this test
umes of other anions. These interfering anions may act as an
method. It is not practicable to list all the combinations. Users should use
eluantanddisplaceotheranionsfromtheconcentratorcolumn.
the appropriate combination of concentrator column, analytical column,
SeeAnnexA1todeterminebreakthroughvolume.Donotload
suppressor, eluant, and regenerant to achieve the desired resolution and
a sample volume greater than 80% of the breakthrough
detection.
volume.
8.5 Fluoride Solution, Stock (1.00 mL=1.00 mg F)—Dry
sodiumfluorideat110°Cfor2 60.5handcoolinadesiccator.
7. Apparatus
Dissolve 2.210 g of dried salt in water and dilute to 1 L.
7.1 Ion chromatograph with the following components:
8.6 Acetate Solution, Stock (1.00 mL=1.00 mg acetate)—
7.1.1 Eluant Introduction System—The wetted portion of
Dissolve 1.389 g of sodium acetate in water and dilute to 1 L
the eluant pump should be nonmetallic or of a corrosion-
with water. Store in a brown glass bottle with a TFE-
resistantmetaltopreventcontaminationofthechromatography
fluorocarbon lined cap in a refrigerator.
columns.
8.7 Formate Solution, Stock (1.00 mL=1 mg formate)—
7.1.2 Sample Injection System—The wetted portion of the
Dissolve 1.511 g sodium formate in water and dilute to 1 L
sample pump should be nonmetallic or of a corrosion-resistant
with water. Store in a brown glass bottle with a TFE-
metal to prevent metal contamination of the chromatography
fluorocarbon lined cap in a refrigerator.
columns.
7.1.3 Anion Suppressor Device.
8.8 Chloride Solution, Stock (1.00 mL=1.00 mg Cl)—Dry
7.1.4 Conductivity Cell, low dead volume (1 µL). Tempera-
sodium chloride (NaCl) for 2 6 0.5 h at 110°C and cool in a
ture compensated or corrected flow through conductivity
desiccator. Dissolve 1.648 g of the dry salt in water and dilute
detectorshouldbecapableofmeasuringconductivityfrom0to
to1L.
1000 µS/cm. If temperature controlled conductivity detector is
8.9 Phosphate Solution, Stock (1.00 mL=1.00 mg PO )—
used, temperature control should be at 60.5°C or better.
Dissolve 1.433 g of potassium dihydrogen phosphate
7.1.5 Suppressor Device Regenerant System—Some manu-
(KH PO ) in water and dilute to 1 L with water.
2 4
facturers provide integrated regenerant systems that reduce the
consumption of eluant. Electrochemical suppressor regenerant 8.10 Sulfate Solution, Stock(1.00mL=1.00mgSO )—Dry
systems can be used, eliminating the need to prepare regener-
sodium sulfate for 2 6 0.5 h at 110°C and cool in a desiccator.
ant solutions. Dissolve 1.479 g of the dried salt in water and dilute to 1 L.
8.11 Nitrate Solution, Stock (1.00 mL=1.00 mg NO )—
8. Reagents
Dryapproximately2gofsodiumnitrate(NaNO )at105°Cfor
8.1 Purity of Reagents—Reagent grade chemicals shall be
48 h. Dissolve exactly 1.371 g of the dried salt in water and
used in all tests. Unless otherwise indicated, it is intended that
dilute to 1 L with water.
all reagents conform to the specifications of the Committee on
8.12 Anion Intermediate Solutions—Prepare a 1000 µg/L
Analytical Reagents of theAmerican Chemical Society where
4 standard of each anion by diluting 1.00 mL of each stock
such specifications are available. Other grades may be used,
solution to 1 L. If acetate, formate, or phosphate are included
provided it is first ascertained that the reagent is of sufficiently
in the standard, the solution must be prepared daily. It is
high purity to permit its use without lessening the accuracy of
recommendedthatthesestandardsbepreparedseparatelyfrom
the determination.
the rest of the anions.
8.2 Purity of Water—Unless otherwise indicated, reference
8.13 Anion Working Solutions—Prepare a blank and at least
to water shall be understood to mean reagent water as defined
three different working solutions from the anion intermediate
by Specification D1193 Type 1 and shall contain less than 0.2
solution,containingtheanionsofinterest.Prepareindedicated
µg/Loftheanionsofinterest.Freshlypreparedwatershouldbe
volumetric flasks and transfer to sample containers in accor-
used for making the low-level standards used for calibration.
dance with Practice D4453. Prepare fresh daily. The range of
Detection limits will be limited by the purity of the water and
the working solutions prepared should bracket the analytical
reagents used to make standards. The purity of the water used
range of interest.Atypical range would be 5, 10, and 25 µg/L
shall be checked by the use of Test Methods D5542.
for each anion or consistent with analytical range of interest.
Systems equipped with sample preparation modules are ca-
pable of automatic standard preparation at significantly lower
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For suggestions on the testing of reagents not concentrations.
listed by the American Chemical Society, see Analar Standards for Laboratory
NOTE 2—When working with very low concentration standards, it is
Chemicals, BDH Ltd., Poole, Dorset, U.K., a
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D5996 − 05 (Reapproved 2009) D5996 − 16
Standard Test Method for
Measuring Anionic Contaminants in High-Purity Water by
On-Line Ion Chromatography
This standard is issued under the fixed designation D5996; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers on-line analysis of high-purity water by the ion chromatography technique. This test method is
applicable for measuring various anionic contaminants in high-purity water, typically in the range of 0.01 to 100 μg/L. This test
method is used to determine the concentration of acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in a
continuously flowing sample. The range of the test method is only as good as the reagent water available for preparing standards.
At extremely low concentrations, <1.0 μg/L, preparing standards is difficult, and extra care must be taken in their preparation. The
sample may have to be conditioned from higher pressures and temperatures to conditions that are suitable for use by on-line
instruments.
1.2 Online sample analysis of flowing streams does not lend itself to collaborative studies due to the nature of the sample and
the possibility of contamination that may result from handling the sample as part of the collaborative study. Therefore this standard
test method is not based on the results of a collaborative study but is intended to provide the best possible guidance for doing this
type of analysis.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
D1066 Practice for Sampling Steam
D1129 Terminology Relating to Water
D1192 Guide for Equipment for Sampling Water and Steam in Closed Conduits (Withdrawn 2003)
D1193 Specification for Reagent Water
D2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
D3370 Practices for Sampling Water from Closed Conduits
D3864 Guide for On-Line Monitoring Systems for Water Analysis
D4453 Practice for Handling of High Purity Water Samples
D5542 Test Methods for Trace Anions in High Purity Water by Ion Chromatography
D5810 Guide for Spiking into Aqueous Samples
D5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis
3. Terminology
3.1 For definitions of terms used in this test method, refer to Terminology D1129.
3.1 Definitions:
3.1.1 For definitions of terms used in this standard, refer to Terminology D1129.
This test method is under the jurisdiction of ASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.03 on Sampling Water and
Water-Formed Deposits, Analysis of Water for Power Generation and Process Use, On-Line Water Analysis, and Surveillance of Water.
Current edition approved Oct. 1, 2009Feb. 15, 2016. Published November 2009June 2016. Originally approved in 1996. Last previous edition approved in 20052009 as
D5996 – 05.D5996 – 05 (2009). DOI: 10.1520/D5996-05R09.10.1520/D5996-16.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
The last approved version of this historical standard is referenced on www.astm.org.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D5996 − 16
3.2 Definitions of Terms Specific to This Standard:
3.2.1 analytical column, n—a column used to separate the anions of interest.
3.2.2 analytical column set, n—a combination of one or more guard columns followed by one or more analytical columns.
3.2.3 anion suppressor device, n—a device that is placed between the analytical columns and the detector. Its purpose is to
inhibit detector response to the ionic constituents in the eluant, so as to lower the detector background and at the same time enhance
detector response to the ions of interest.
3.2.3.1 Discussion—
Its purpose is to inhibit detector response to the ionic constituents in the eluant, so as to lower the detector background and at the
same time enhance detector response to the ions of interest.
3.2.4 breakthrough volume, n—the maximum sample volume that can be passed through a concentrator column before the least
tightly bound ion of interest is eluted. All of the columns in series contribute to the overall capacity of the analytical column set.
3.2.4.1 Discussion—
All of the columns in series contribute to the overall capacity of the analytical column set.
3.2.5 concentrator column, n—an ion exchange column used to concentrate the ions of interest and thereby increase method
sensitivity.
3.2.6 eluant, n—the ionic mobile phase used to transport the sample through the analytical column.
3.2.7 guard column, n—a column used before the analytical column to protect it from contaminants, such as particulate matter
or ionic species that may chemically foul the resins and degrade their performance.
3.2.8 ion chromatography, n—a form of liquid chromatography in which ionic constituents are separated by ion exchange
followed by a suitable detection means.
3.2.9 resolution, n—the ability of an analytical column to separate constituents under specific test conditions.
4. Summary of Test Method
4.1 A continuously flowing sample is injected into the instrument through a sample injection valve. The sample is pumped
through a concentrator column where the anions of interest are collected on ion-exchange resin. After a suitable volume of sample
has been passed through the concentrator column, sample flow is diverted and an eluant is pumped through the concentrator
column to remove the trapped anions. This eluant then flows through an analytical column set where the anions are separated based
on the retention characteristic of each anion relative to the eluant used. The eluant stream containing the anions of interest passes
through a suppressor device where the cations from the eluant are exchanged for hydrogen ions, converting the anions to their acid
form. After the suppressor device, the eluant solution passes through a conductivity detector where the separated anions are
detected. Detection limits for the anions are enhanced because the anions are in the acid form rather than the salt.
4.2 The anions are identified based on the retention time as compared to known standards. By measuring peak height or area
and comparing the detector response to known standards, the anions can be quantified.
5. Significance and Use
5.1 In the power-generation industry, high-purity water is used to reduce corrosion from anions, such as sulfate, chloride, and
fluoride. These anions are known to be detrimental to materials of construction used in steam generators, reactor vessel internals
and recirculation piping, heat exchangers, connective piping, and turbines. Most electric generating plants try to control these
anions to <1.0 μg/L in the steam generator feed water. Some nuclear power plants have been able to control anion contaminants
at less than 0.02 μg/L.
5.2 These anions and others cause low product yields in semiconductor manufacturing. They are also monitored and controlled
at similarly low levels as in the electric power industry.
5.3 Low molecular weight organic acids (acetate, formate, propionate) have been detected in steam generator feed water. These
low molecular weight organic materials are believed to be high-temperature degradation products of chemicals used to control
cycle water pH and organic contaminants in cycle makeup water.
5.4 In the semiconductor industry, anion contaminants may come from the breakdown of low molecular weight organic
materials by ultraviolet light radiation, which is frequently used to produce bacteria-free water. These organic compounds may also
contribute to low product yield.
5.5 The production of high-purity water for process makeup and use frequently employs the use of demineralizers to remove
unwanted anion contaminants. Also in the electric power industry, demineralizers are used in the process stream to maintain low
D5996 − 16
levels of these contaminants. As such, it is important to monitor this process to ensure that water quality standards are being met.
These processes can be monitored for the above-mentioned anions.
5.6 On-line measurements of these contaminants provide a greater degree of protection of the processes by allowing for frequent
on-line measurement of these species. Early detection of contaminant ingress allows for quicker corrective action to locate, reduce,
or eliminate, or combination thereof, the source. Grab samples will not provide the same level of protection because of their
intermittent nature and the longer time required to obtain and then analyze the sample.
5.7 Additionally, on-line monitoring significantly reduces the potential for contamination of high-purity water samples, a
significant problem when sampling and testing high-purity water.
6. Interferences
6.1 When working with low concentration samples, blanks, and standards, contamination can be a serious problem. Extreme
care must be exercised in all phases of this test method.
6.2 Improper sample line material or sample lines that have not been properly conditioned can give results that may not be truly
representative of the process stream. Absorption/desorption of anions on sample line wall deposits can change analytical results.
Maintaining a minimum sample flow of 1.8 m/s (6 ft/s) will minimize deposit buildup on sample line walls, reducing the potential
for absorption/desorption of anions.
6.3 A single anion present at a concentration significantly higher than other anions could mask closely adjacent peaks on the
chromatogram.
6.4 Low breakthrough volumes may be experienced when continuously monitoring for anions in water that has had its pH raised
by ammonia, morpholine, or other additives. This interference can be eliminated by taking the sample from the effluent of a cation
resin column.
6.5 Identification of the anion is based on retention time of the anion of interest. An interfering anion having the same retention
time as one of the anions of interest will result in erroneously high values for that anion.
6.6 When loading a concentrator column, high concentrations of interfering anions may cause low breakthrough volumes of
other anions. These interfering anions may act as an eluant and displace other anions from the concentrator column. See Annex
A1 to determine breakthrough volume. Do not load a sample volume greater than 80 % of the breakthrough volume.
7. Apparatus
7.1 Ion chromatograph with the following components:
7.1.1 Eluant Introduction System—The wetted portion of the eluant pump should be nonmetallic or of a corrosion-resistant
metal to prevent contamination of the chromatography columns.
7.1.2 Sample Injection System—The wetted portion of the sample pump should be nonmetallic or of a corrosion-resistant metal
to prevent metal contamination of the chromatography columns.
7.1.3 Anion Suppressor Device.
7.1.4 Conductivity Cell, low dead volume (1 μL). Temperature compensated or corrected flow through conductivity detector
should be capable of measuring conductivity from 0 to 1000 μS/cm. If temperature controlled conductivity detector is used,
temperature control should be at 60.5°C or better.
7.1.5 Suppressor Device Regenerant System—Some manufacturers provide integrated regenerant systems that reduce the
consumption of eluant. Electrochemical suppressor regenerant systems can be used, eliminating the need to prepare regenerant
solutions.
8. Reagents
8.1 Purity of Reagents—Reagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that all
reagents conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society where such
specifications are available. Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.
8.2 Purity of Water—Unless otherwise indicated, reference to water shall be understood to mean reagent water as defined by
Specification D1193 Type 1 and shall contain less than 0.2 μg/L of the anions of interest. Freshly prepared water should be used
for making the low-level standards used for calibration. Detection limits will be limited by the purity of the water and reagents
used to make standards. The purity of the water used shall be checked by the use of Test Methods D5542.
8.3 Prepare eluant for the specific columns used and for the anions of interest in accordance with manufacturer’s directions.
Reagent Chemicals, American Chemical Society Specifications, American Chemical Society, Washington, DC. For suggestions on the testing of reagents not listed by
the American Chemical Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia and National
Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD.
D5996 − 16
8.4 Prepare regenerant for the specific suppressor used in accordance with the manufacturer’s directions if required.
NOTE 1—There are numerous combinations of analytical columns, suppressors, eluants, and regenerants that may be used with this test method. It is
not practicable to list all the combinations. Users should use the appropriate combination of concentrator column, analytical column, suppressor, eluant,
and regenerant to achieve the desired resolution and detection.
8.5 Fluoride Solution, Stock (1.00 mL = 1.00 mg F)—Dry sodium fluoride at 110°C for 2 6 0.5 h and cool in a desiccator.
Dissolve 2.210 g of dried salt in water and dilute to 1 L.
8.6 Acetate Solution, Stock (1.00 mL = 1.00 mg acetate)—Dissolve 1.389 g of sodium acetate in water and dilute to 1 L with
water. Store in a brown glass bottle with a TFE-fluorocarbon lined cap in a refrigerator.
8.7 Formate Solution, Stock (1.00 mL = 1 mg formate)—Dissolve 1.511 g sodium formate in water and dilute to 1 L with water.
Store in a brown glass bottle with a TFE-fluorocarbon lined cap in a refrigerator.
8.8 Chloride Solution, Stock (1.00 mL = 1.00 mg Cl)—Dry sodium chloride (NaCl) for 2 6 0.5 h at 110°C and cool in a
desiccator. Dissolve 1.648 g of the dry salt in water and dilute to 1 L.
8.9 Phosphate Solution, Stock (1.00 mL = 1.00 mg PO )—Dissolve 1.433 g of potassium dihydrogen phosphate (KH PO ) in
4 2 4
water and dilute to 1 L with water.
8.10 Sulfate Solution, Stock (1.00 mL = 1.00 mg SO )—Dry sodium sulfate for 2 6 0.5 h at 110°C and cool in a desiccator.
Dissolve 1.479 g of the dried salt in water and dilute to 1 L.
8.11 Nitrate Solution, Stock (1.00 mL = 1.00 mg NO )—Dry approximately 2 g of sodium nitrate (NaNO ) at 105°C for 48 h.
3 3
Dissolve exactly 1.371 g of the dried salt in water and dilute to 1 L with water.
8.12 Anion Intermediate Solutions—Prepare a 1000 μg/L standard of each anion by diluting 1.00 mL of each stock solution to
1 L. If acetate, formate, or phosphate are included in the standard, the solution must be prepared daily. It is recommended that these
standards be prepared separately from the rest of the anions.
8.13 Anion Working Solutions—Prepare a blank and at least three different working solutions from the anion intermediate
solution, containing the anions of interest. Prepare in dedicated volumetric flasks and transfer to sample containers in accordance
with Practice D4453. Prepare fresh daily. The range of the working solutions pr
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...