ASTM D7968-23
(Test Method)Standard Test Method for Determination of Polyfluorinated Compounds in Soil by Liquid Chromatography Tandem Mass Spectrometry (LC/MS/MS)
Standard Test Method for Determination of Polyfluorinated Compounds in Soil by Liquid Chromatography Tandem Mass Spectrometry (LC/MS/MS)
SIGNIFICANCE AND USE
5.1 This test method has been developed by the U.S. EPA Region 5 Chicago Regional Laboratory (CRL).
5.2 PFAS are widely used in various industrial and commercial products; they are persistent, bio-accumulative, and ubiquitous in the environment. PFAS have been reported to exhibit developmental toxicity, hepatotoxicity, immunotoxicity, and hormone disturbance. A draft Toxicological Profile for Perfluoroalkyls from the U.S. Department of Health and Human Services is available.7 PFAS have been detected in soils, sludges, and surface and drinking waters. Hence, there is a need for a quick, easy, and robust method to determine these compounds at trace levels in various soil matrices for understanding of the sources and pathways of exposure.
5.3 This method has been used to determine selected PFAS in sand (Table 4) and four ASTM reference soils (Table 5).
SCOPE
1.1 This procedure covers the determination of selected polyfluorinated alkyl substances (PFAS) in a soil matrix using solvent extraction, filtration, followed by liquid chromatography (LC) and detection with tandem mass spectrometry (MS/MS). These analytes are qualitatively and quantitatively determined by this method. This method adheres to multiple reaction monitoring (MRM) mass spectrometry. This procedure utilizes a quick extraction and is not intended to generate an exhaustive accounting of the content of PFAS in difficult soil matrices. An exhaustive extraction procedure for PFAS, such as published by Washington et al.,2 for difficult matrices should be considered when analyzing PFAS. The approach from this standard was utilized to screen laboratory coats (textiles) to identify if PFAS would be leached from the materials.
1.2 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 The method of detection limit3 and reporting range4 for the target analytes are listed in Table 1.
1.3.1 The reporting limit in this test method is the minimum value below which data are documented as non-detects. Analyte detections between the method detection limit and the reporting limit are estimated concentrations and are not reported following this test method. In most cases, the reporting limit is calculated from the concentration of the Level 1 calibration standard as shown in Table 2 for the PFAS after taking into account a 2 g sample weight and a final extract volume of 10 mL, 50 % water/50 % MeOH with 0.1 % acetic acid. The final extract volume is assumed to be 10 mL because 10 mL of 50 % water/50 % MeOH with 0.1 % acetic acid was added to each soil sample and only the liquid layer after extraction is filtered, leaving the solid and any residual solvent behind. It is raised above the Level 1 calibration concentration for PFOS, PFHxA, FHEA, and FOEA; these compounds can be identified at the Level 1 concentration but the standard deviation among replicates at this lower spike level resulted in a higher reporting limit.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
- Status
- Published
- Publication Date
- 31-Oct-2023
- Technical Committee
- D34 - Waste Management
- Drafting Committee
- D34.01.06 - Analytical Methods
Relations
- Effective Date
- 01-Nov-2023
- Effective Date
- 01-Nov-2023
- Effective Date
- 01-Nov-2023
- Effective Date
- 01-Nov-2023
Overview
ASTM D7968-23 is the internationally recognized standard test method for the determination of polyfluorinated compounds, also known as per- and polyfluoroalkyl substances (PFAS), in soil. Developed by ASTM International under the jurisdiction of Committee D34 on Waste Management, this standard outlines procedures for qualitative and quantitative analysis of selected PFAS using solvent extraction and liquid chromatography tandem mass spectrometry (LC/MS/MS), with multiple reaction monitoring (MRM).
PFAS are synthetic chemicals commonly used in industrial and commercial products due to their resistance to heat, water, and oil. However, their persistence, bioaccumulation, and mobility in the environment have raised significant health and environmental concerns. Detecting and quantifying PFAS at trace levels in soil is necessary for risk assessment, remediation, and regulatory compliance.
Key Topics
Purpose and Scope
- Provides a quick, robust, and sensitive method to detect trace-level PFAS in various soil matrices.
- Utilizes solvent extraction, filtration, and LC/MS/MS for comprehensive PFAS analysis.
Analyte Coverage
- Focuses on a targeted suite of perfluorinated carboxylic acids, sulfonates, and fluorotelomer acids (see Table 1 of the standard for details).
- Method detection limits and reporting ranges are specified for each compound.
Performance-Based Methodology
- Allows use of alternative instrumentation or procedures if data quality objectives are achieved.
- Adheres to stringent quality assurance and calibration requirements to ensure data reliability.
Sampling and Quality Control
- Emphasizes contamination avoidance by using PFAS-free sample containers and equipment.
- Involves routine use of field blanks, surrogates, and matrix spike samples.
Trace Analysis
- Specifically suitable for low-level detection of PFAS in soil and sand, providing valuable data for environmental monitoring and pathway exposure studies.
Health and Safety
- Stipulates laboratory safety practices, including the use of appropriate PPE and reference to Safety Data Sheets (SDS).
Applications
Environmental Site Assessment
- Identifying and quantifying PFAS contamination in soils at waste management facilities, industrial sites, and brownfields.
- Supports site characterization, human health risk assessment, and the development of appropriate remediation strategies.
Regulatory Compliance
- Meets data requirements for federal, regional (e.g., U.S. EPA), and international regulatory frameworks.
- Facilitates monitoring required under environmental protection statutes for persistent organic pollutants.
Research and Product Testing
- Used to screen consumer products (e.g., textiles, lab coats) for PFAS leaching potential.
- Assists in understanding the environmental fate, transport, and persistence of PFAS compounds.
Soil Quality Monitoring
- Enables regular environmental surveillance to track PFAS trends over time, aiding government agencies and private companies in managing long-term soil quality.
Related Standards
- ASTM D1193 - Specification for Reagent Water used in analytical testing.
- ASTM D2777 - Practice for Determination of Precision and Bias of Applicable Test Methods.
- ASTM D5847 - Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis.
- EPA SW-846 - Test Methods for Evaluating Solid Waste, Physical/Chemical Methods.
- 40 CFR Part 136 Appendix B - Procedures for determining method detection limits.
Summary
ASTM D7968-23 is a crucial analytical standard that enables accurate, sensitive, and reproducible measurement of PFAS in soils by employing LC/MS/MS technology. Its robust procedures and quality assurance requirements make it essential for laboratories, regulators, and stakeholders engaged in environmental monitoring, contamination assessment, and remediation planning related to polyfluorinated compounds. By following this standard, organizations can ensure reliable PFAS data to inform decisions and comply with evolving environmental regulations.
Keywords: ASTM D7968-23, PFAS in soil, LC/MS/MS, environmental analysis, polyfluorinated compounds, perfluoroalkyl substances, soil testing standard, PFAS contamination, environmental compliance.
Buy Documents
ASTM D7968-23 - Standard Test Method for Determination of Polyfluorinated Compounds in Soil by Liquid Chromatography Tandem Mass Spectrometry (LC/MS/MS)
REDLINE ASTM D7968-23 - Standard Test Method for Determination of Polyfluorinated Compounds in Soil by Liquid Chromatography Tandem Mass Spectrometry (LC/MS/MS)
Get Certified
Connect with accredited certification bodies for this standard

NSF International
Global independent organization facilitating standards development and certification.
CIS Institut d.o.o.
Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

Kiwa BDA Testing
Building and construction product certification.
Sponsored listings
Frequently Asked Questions
ASTM D7968-23 is a standard published by ASTM International. Its full title is "Standard Test Method for Determination of Polyfluorinated Compounds in Soil by Liquid Chromatography Tandem Mass Spectrometry (LC/MS/MS)". This standard covers: SIGNIFICANCE AND USE 5.1 This test method has been developed by the U.S. EPA Region 5 Chicago Regional Laboratory (CRL). 5.2 PFAS are widely used in various industrial and commercial products; they are persistent, bio-accumulative, and ubiquitous in the environment. PFAS have been reported to exhibit developmental toxicity, hepatotoxicity, immunotoxicity, and hormone disturbance. A draft Toxicological Profile for Perfluoroalkyls from the U.S. Department of Health and Human Services is available.7 PFAS have been detected in soils, sludges, and surface and drinking waters. Hence, there is a need for a quick, easy, and robust method to determine these compounds at trace levels in various soil matrices for understanding of the sources and pathways of exposure. 5.3 This method has been used to determine selected PFAS in sand (Table 4) and four ASTM reference soils (Table 5). SCOPE 1.1 This procedure covers the determination of selected polyfluorinated alkyl substances (PFAS) in a soil matrix using solvent extraction, filtration, followed by liquid chromatography (LC) and detection with tandem mass spectrometry (MS/MS). These analytes are qualitatively and quantitatively determined by this method. This method adheres to multiple reaction monitoring (MRM) mass spectrometry. This procedure utilizes a quick extraction and is not intended to generate an exhaustive accounting of the content of PFAS in difficult soil matrices. An exhaustive extraction procedure for PFAS, such as published by Washington et al.,2 for difficult matrices should be considered when analyzing PFAS. The approach from this standard was utilized to screen laboratory coats (textiles) to identify if PFAS would be leached from the materials. 1.2 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.3 The method of detection limit3 and reporting range4 for the target analytes are listed in Table 1. 1.3.1 The reporting limit in this test method is the minimum value below which data are documented as non-detects. Analyte detections between the method detection limit and the reporting limit are estimated concentrations and are not reported following this test method. In most cases, the reporting limit is calculated from the concentration of the Level 1 calibration standard as shown in Table 2 for the PFAS after taking into account a 2 g sample weight and a final extract volume of 10 mL, 50 % water/50 % MeOH with 0.1 % acetic acid. The final extract volume is assumed to be 10 mL because 10 mL of 50 % water/50 % MeOH with 0.1 % acetic acid was added to each soil sample and only the liquid layer after extraction is filtered, leaving the solid and any residual solvent behind. It is raised above the Level 1 calibration concentration for PFOS, PFHxA, FHEA, and FOEA; these compounds can be identified at the Level 1 concentration but the standard deviation among replicates at this lower spike level resulted in a higher reporting limit. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
SIGNIFICANCE AND USE 5.1 This test method has been developed by the U.S. EPA Region 5 Chicago Regional Laboratory (CRL). 5.2 PFAS are widely used in various industrial and commercial products; they are persistent, bio-accumulative, and ubiquitous in the environment. PFAS have been reported to exhibit developmental toxicity, hepatotoxicity, immunotoxicity, and hormone disturbance. A draft Toxicological Profile for Perfluoroalkyls from the U.S. Department of Health and Human Services is available.7 PFAS have been detected in soils, sludges, and surface and drinking waters. Hence, there is a need for a quick, easy, and robust method to determine these compounds at trace levels in various soil matrices for understanding of the sources and pathways of exposure. 5.3 This method has been used to determine selected PFAS in sand (Table 4) and four ASTM reference soils (Table 5). SCOPE 1.1 This procedure covers the determination of selected polyfluorinated alkyl substances (PFAS) in a soil matrix using solvent extraction, filtration, followed by liquid chromatography (LC) and detection with tandem mass spectrometry (MS/MS). These analytes are qualitatively and quantitatively determined by this method. This method adheres to multiple reaction monitoring (MRM) mass spectrometry. This procedure utilizes a quick extraction and is not intended to generate an exhaustive accounting of the content of PFAS in difficult soil matrices. An exhaustive extraction procedure for PFAS, such as published by Washington et al.,2 for difficult matrices should be considered when analyzing PFAS. The approach from this standard was utilized to screen laboratory coats (textiles) to identify if PFAS would be leached from the materials. 1.2 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.3 The method of detection limit3 and reporting range4 for the target analytes are listed in Table 1. 1.3.1 The reporting limit in this test method is the minimum value below which data are documented as non-detects. Analyte detections between the method detection limit and the reporting limit are estimated concentrations and are not reported following this test method. In most cases, the reporting limit is calculated from the concentration of the Level 1 calibration standard as shown in Table 2 for the PFAS after taking into account a 2 g sample weight and a final extract volume of 10 mL, 50 % water/50 % MeOH with 0.1 % acetic acid. The final extract volume is assumed to be 10 mL because 10 mL of 50 % water/50 % MeOH with 0.1 % acetic acid was added to each soil sample and only the liquid layer after extraction is filtered, leaving the solid and any residual solvent behind. It is raised above the Level 1 calibration concentration for PFOS, PFHxA, FHEA, and FOEA; these compounds can be identified at the Level 1 concentration but the standard deviation among replicates at this lower spike level resulted in a higher reporting limit. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM D7968-23 is classified under the following ICS (International Classification for Standards) categories: 13.080.10 - Chemical characteristics of soils. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM D7968-23 has the following relationships with other standards: It is inter standard links to ASTM D7968-17a, ASTM E3302-22, ASTM E3358-23a, ASTM F2931-19a. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM D7968-23 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D7968 − 23
Standard Test Method for
Determination of Polyfluorinated Compounds in Soil by
Liquid Chromatography Tandem Mass Spectrometry (LC/
MS/MS)
This standard is issued under the fixed designation D7968; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope reporting limit are estimated concentrations and are not re-
ported following this test method. In most cases, the reporting
1.1 This procedure covers the determination of selected
limit is calculated from the concentration of the Level 1
polyfluorinated alkyl substances (PFAS) in a soil matrix using
calibration standard as shown in Table 2 for the PFAS after
solvent extraction, filtration, followed by liquid chromatogra-
taking into account a 2 g sample weight and a final extract
phy (LC) and detection with tandem mass spectrometry (MS/
volume of 10 mL, 50 % water/50 % MeOH with 0.1 % acetic
MS). These analytes are qualitatively and quantitatively deter-
acid. The final extract volume is assumed to be 10 mL because
mined by this method. This method adheres to multiple
10 mL of 50 % water/50 % MeOH with 0.1 % acetic acid was
reaction monitoring (MRM) mass spectrometry. This proce-
added to each soil sample and only the liquid layer after
dure utilizes a quick extraction and is not intended to generate
extraction is filtered, leaving the solid and any residual solvent
an exhaustive accounting of the content of PFAS in difficult
behind. It is raised above the Level 1 calibration concentration
soil matrices. An exhaustive extraction procedure for PFAS,
2 for PFOS, PFHxA, FHEA, and FOEA; these compounds can
such as published by Washington et al., for difficult matrices
be identified at the Level 1 concentration but the standard
should be considered when analyzing PFAS. The approach
deviation among replicates at this lower spike level resulted in
from this standard was utilized to screen laboratory coats
a higher reporting limit.
(textiles) to identify if PFAS would be leached from the
1.4 This standard does not purport to address all of the
materials.
safety concerns, if any, associated with its use. It is the
1.2 Units—The values stated in SI units are to be regarded
responsibility of the user of this standard to establish appro-
as standard. No other units of measurement are included in this
priate safety, health, and environmental practices and deter-
standard.
mine the applicability of regulatory limitations prior to use.
3 4
1.3 The method of detection limit and reporting range for
1.5 This international standard was developed in accor-
the target analytes are listed in Table 1.
dance with internationally recognized principles on standard-
1.3.1 The reporting limit in this test method is the minimum
ization established in the Decision on Principles for the
value below which data are documented as non-detects. Ana-
Development of International Standards, Guides and Recom-
lyte detections between the method detection limit and the
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
2. Referenced Documents
This test method is under the jurisdiction of ASTM Committee D34 on Waste
Management and is the direct responsibility of Subcommittee D34.01.06 on
2.1 ASTM Standards:
Analytical Methods.
D1193 Specification for Reagent Water
Current edition approved Nov. 1, 2023. Published November 2023. Originally
D2777 Practice for Determination of Precision and Bias of
approved in 2014. Last previous edition approved in 2017 as D7968 – 17a. DOI:
10.1520/D7968-23. Applicable Test Methods of Committee D19 on Water
Washington, J. W., Naile, J. E., Jenkins, T. M., and Lynch, D. G., “Character-
D5847 Practice for Writing Quality Control Specifications
izing Fluorotelomer and Polyfluoroalkyl Substances in New and Aged
for Standard Test Methods for Water Analysis
Fluorotelomer-Based Polymers for Degradation Studies with GC/MS and LC/MS/
E2554 Practice for Estimating and Monitoring the Uncer-
MS,” Environmental Science and Technology, Vol 48, 2014, pp. 5762–5769.
The MDL is determined following the Code of Federal Regulations, 40 CFR
tainty of Test Results of a Test Method Using Control
Part 136, Appendix B utilizing solvent extraction of soil. A 2 g sample of Ottawa
Chart Techniques
sand was utilized. A detailed process determining the MDL is explained in the
reference and is beyond the scope of this standard to be explained here.
4 5
Reporting range concentration is calculated from Table 2 concentrations For referenced ASTM standards, visit the ASTM website, www.astm.org, or
assuming a 30 μL injection of the Level 1 calibration standard for the PFAS, and the contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
highest level calibration standard with a 10 mL final extract volume of a 2 g soil Standards volume information, refer to the standard’s Document Summary page on
sample. Volume variations will change the reporting limit and ranges. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7968 − 23
A
TABLE 1 Method Detection Limit and Reporting Range
3.2.18 PFecHS—Decaluoro-4-
MDL Reporting Limit
(pentafluoroethyl)cyclohexanesulfonate
Analyte
(ng/kg) (ng/kg)
3.2.19 PFAC—Perfluoroalkyl Carboxylic Acid
PFTreA 6.76 25–1000
PFTriA 5.26 25–1000
3.2.20 PFBA—Perfluorobutanoate
PFDoA 3.56 25–1000
PFUnA 2.45 25–1000
3.2.21 PFPeA—Perfluoropentanoate
PFDA 5.54 25–1000
PFOS 18.83 50–1000 3.2.22 PFHxA—Perfluorohexanoate
PFNA 2.82 25–1000
3.2.23 PFHpA—Perfluoroheptanoate
PFecHS 2.41 25–1000
PFOA 6.24 25–1000
3.2.24 PFOA—Perfluorooctanoate
PFHxS 7.75 25–1000
PFHpA 5.80 25–1000 3.2.25 PFNA—Perfluorononanoate
PFHxA 15.44 50–1000
3.2.26 PFDA—Perfluorodecanoate
PFBS 6.49 25–1000
PFPeA 20.93 125–5000
3.2.27 PFUnA—Perfluoroundecanoate
PFBA 22.01 125–5000
FHEA 199.04 600–20 000 3.2.28 PFTriA—Perfluorotridecanoate
FOEA 258.37 750–20 000
3.2.29 PFTreA—Perfluorotetradecanoate
FDEA 137.46 500–20 000
FOUEA 4.85 25–1000
3.2.30 FTAs and FTUAs—Fluorotelomer and Unsaturated
FhpPa 5.09 25–1000
Fluorotelomer Acids
FHUEA 3.50 25–1000
A
Abbreviations are defined in 3.2. 3.2.31 FHpPA—3-perfluoropheptyl propanoic acid
3.2.32 FOUEA—2H-perfluoro-2-decenoic acid
3.2.33 FDEA—2-perfluorodecyl ethanoic acid
2.2 Other Documents:
3.2.34 FOEA—2-perfluorooctyl ethanoic acid
EPA SW-846 Test Methods for Evaluating Solid Waste,
Physical/Chemical Methods
3.2.35 FHUEA—2H-perfluoro-2-octenoic acid
40 CFR Part 136 Appendix B Definition and Procedure for
3.2.36 FHEA—2-perfluorohexyl ethanoic acid
the Determination of the Method Detection Limit
3.2.37 MPFAS—Isotopically labeled Perfluoroal-
kylsulfonates
3. Terminology
3.2.38 MPFHxS— O -Perfluorohexylsulfonate
3.1 Definitions: 2
3.1.1 reporting limit, RL, n—the minimum concentration
3.2.39 MPFOS— C -Perfluorooctylsulfonate
below which data are documented as non-detects.
3.2.40 MPFCA—Isotopically labeled Perfluoroalkylcar-
3.2 Abbreviations:
boxylates
3.2.1 CCC—Continuing Calibration Check
3.2.41 MPFBA— C -Perfluorobutanoate
3.2.2 IC—Initial Calibration
3.2.42 MPFHxA— C -Perfluorohexanoate
3.2.3 ppt—parts per trillion, ng/kg or ng/L
3.2.43 MPFOA— C -Perfluorooctanoate
3.2.4 LC—Liquid Chromatography
3.2.44 MPFNA— C -Perfluorononanoate
3.2.5 LCS/LCSD—Laboratory Control Sample/Laboratory
3.2.45 MPFDA— C -Perfluorodecanoate
Control Sample Duplicate
3.2.46 MPFUnA— C -Perfluoroundecanoate
3.2.6 MDL—Method Detection Limit
3.2.47 MPFDoA— C -Perfluorodecanoate
3.2.7 MeOH—Methanol
3.2.48 QA—Quality Assurance
–3
3.2.8 mM—millimolar, 1 × 10 moles/L
3.2.49 QC—Quality Control
3.2.9 MRM—Multiple Reaction Monitoring
3.2.50 RL—Reporting Limit
3.2.10 MS/MSD—Matrix Spike/Matrix Spike Duplicate
3.2.51 RLCS—Reporting Limit Check Sample
3.2.11 NA—Not available
3.2.52 RSD—Relative Standard Deviation
3.2.12 ND—Non-detect
3.2.53 RT—Retention Time
3.2.13 P&A—Precision and Accuracy
3.2.54 SRM—Single Reaction Monitoring
3.2.14 PFAS—Perfluoroalkyl substances
3.2.55 SS—Surrogate Standard
3.2.15 PFBS—Perfluorobutylsulfonate
3.2.56 TC—Target Compound
3.2.16 PFHxS—Perfluorohexylsulfonate
4. Summary of Test Method
3.2.17 PFOS—Perfluorooctylsulfonate
4.1 The operating conditions presented in this test method
have been successfully used in the determination of polyfluo-
Available from National Technical Information Service (NTIS), U.S. Depart-
rinated compounds in soil; however, this test method is
ment of Commerce, 5285 Port Royal Road, Springfield, VA, 22161, http://
www.epa.gov/epawaste/hazard/testmethods/index.htm intended to be performance based and alternative operating
D7968 − 23
TABLE 2 Concentrations of Calibration Standards (ng/L)
Analyte/Surrogate LV1 LV2 LV3 LV4 LV5 LV6 LV7 LV8 LV9
PFPeA, PFBA 25 50 100 200 300 400 500 750 1000
PFTreA, PFTriA, PFDoA, PFUnA, PFDA, PFOS, PFNA, PFHxA, PFHpA, PFBS,
PFechS, PFOA, PFHxS, FOUEA, FHUEA, FHpPA, MPFBS, MPFHxA, MPFUnA, 5 10 20 40 60 80 100 150 200
MPFOA, MPFDA, MPFOS, MPFNA, MPFHxS, MPFBA
FHEA, FOEA, FDEA 100 200 400 800 1200 1600 2000 3000 4000
conditions can be used to perform this method provided data through a polypropylene filter unit. Acetic acid (~50 μL) is
quality objectives are attained. added to all the filtered samples to adjust the pH ~3 to 4 and
then analyzed by LC/MS/MS.
4.2 For PFAS analysis, samples are shipped to the lab on ice
and analyzed within 28 days of collection. A sample (2 g) is 4.3 Most of the PFAS target compounds are identified by
transferred to a polypropylene tube, spiked with surrogates (all comparing the single reaction monitoring (SRM) transition and
samples) and target PFAS (laboratory control and matrix spike its confirmatory SRM transition if correlated to the known
samples). The analytes are tumbled for an hour with 10 mL of standard SRM (Table 3) and quantitated utilizing an external
methanol:water (50:50) under basic condition (pH ~9 to 10 calibration. The surrogates and some PFAS target analytes
adjusted with ~20 μL ammonium hydroxide). The samples are (PFPeA, PFBA, FOUEA, and FHUEA) only utilize one SRM
centrifuged and the extract, leaving the solid behind, is filtered transition due to a less sensitive or non-existent secondary
TABLE 3 Retention Times, SRM Ions, and Analyte-Specific Mass Spectrometer Parameters
Primary/
Primary/ Retention Times
Chemical Cone (V) Collision (eV) MRM Transition Confirmatory SRM
Confirmatory (min)
Area Ratio
Primary 20 13 712.9→668.9
PFTreA 10.63 7.4
Confirmatory 20 30 712.9→169
Primary 25 12 662.9→618.9
PFTriA 10.17 7.4
Confirmatory 25 28 662.9→169
Primary 10 12 612.9→568.9
PFDoA 9.61 8.2
Confirmatory 10 25 612.9→169
Primary 15 10 562.9→519
PFUnA 9.05 7.2
Confirmatory 15 18 562.9→269
Primary 20 10 512.9→468.9
PFDA 8.45 6.5
Confirmatory 20 16 512.9→219
Primary 10 42 498.9→80.1
PFOS 8.78 1.3
Confirmatory 10 40 498.9→99.1
Primary 20 10 462.9→418.9
PFNA 7.78 4.9
Confirmatory 20 16 462.9→219
Primary 10 25 460.9→381
PFecHS 8.1 2.2
Confirmatory 10 25 460.9→99.1
Primary 20 10 412.9→369
PFOA 7.11 3.6
Confirmatory 20 16 412.9→169
Primary 15 32 398.9→80.1
PFHxS 7.39 1
Confirmatory 15 32 398.9→99.1
Primary 15 10 362.9→319
PFHpA 6.35 4.1
Confirmatory 15 15 362.9→169
Primary 15 8 312.9→269
PFHxA 5.54 24.1
Confirmatory 15 18 312.9→119.1
Primary 10 30 298.9→80.1
PFBS 5.66 1.6
Confirmatory 10 25 298.9→99.1
PFPeA Primary 4.68 10 8 263→219 NA
PFBA Primary 3.67 10 8 212.9→169 NA
Primary 15 20 376.9→293
FHEA 6.14 3.6
Confirmatory 15 6 376.9→313
Primary 15 18 476.9→393
FOEA 7.54 4.3
Confirmatory 15 12 476.9→413
Primary 15 8 576.8→493
FDEA 8.83 3.2
Confirmatory 15 15 576.8→513
FOUEA Primary 7.54 20 12 456.9→392.9 NA
Primary 15 12 440.9→337
FHpPA 7.54 1.1
Confirmatory 15 20 440.9→317
FHUEA Primary 6.08 10 12 357→293 NA
MPFBA Primary 3.67 10 7 217→172.1 NA
MPFHxA Primary 5.54 15 8 315→270 NA
MPFHxS Primary 7.39 15 34 402.9→84.1 NA
MPFOA Primary 7.11 15 10 417→372 NA
MPFNA Primary 7.81 15 9 467.9→423 NA
MPFOS Primary 8.78 15 40 502.9→80.1 NA
MPFDA Primary 8.45 15 10 514.9→470 NA
MPFUnA Primary 9.05 15 10 564.9→519.9 NA
MPFDoA Primary 9.61 15 12 614.9→569.9 NA
D7968 − 23
SRM transition. As an additional quality control measure, determine these compounds at trace levels in various soil
isotopically labeled PFAS surrogate (listed in 12.4) recoveries matrices for understanding of the sources and pathways of
are monitored. There is no correction to the data based upon exposure.
surrogate recoveries. The final report issued for each sample
5.3 This method has been used to determine selected PFAS
lists the concentration of PFAS, if detected, or
in sand (Table 4) and four ASTM reference soils (Table 5).
quantifiable, in ng/kg (dry weight basis) and the surrogate
recoveries.
6. Interferences
5. Significance and Use 6.1 All glassware is washed in hot water with detergent and
rinsed in hot water followed by distilled water. The glassware
5.1 This test method has been developed by the U.S. EPA
is then dried and heated in an oven at 250 °C for 15 to 30 min.
Region 5 Chicago Regional Laboratory (CRL).
All glassware is subsequently rinsed with methanol or acetoni-
5.2 PFAS are widely used in various industrial and com-
trile.
mercial products; they are persistent, bio-accumulative, and
6.2 All reagents and solvents should be pesticide residue
ubiquitous in the environment. PFAS have been reported to
purity or higher to minimize interference problems. The use of
exhibit developmental toxicity, hepatotoxicity,
PFAS-containing caps must be avoided.
immunotoxicity, and hormone disturbance. A draft Toxicologi-
cal Profile for Perfluoroalkyls from the U.S. Department of
6.3 Matrix interferences may be caused by contaminants in
Health and Human Services is available. PFAS have been
the sample. The extent of matrix interferences can vary
detected in soils, sludges, and surface and drinking waters.
considerably depending on variations in the sample matrices.
Hence, there is a need for a quick, easy, and robust method to
6.4 Contaminants have been found in reagents, glassware,
tubing, glass disposable pipettes, filters, degassers, and other
apparatus that release polyfluorinated compounds. All of these
7 materials and supplies are routinely demonstrated to be free
A draft Toxicological Profile for Perfluroalkyls can be found at http://
www.atsdr.cdc.gov/toxprofiles/tp.asp?id=1117&tid=237 (2014).
TABLE 4 Single-Laboratory Recovery Data in Ottawa Sand
Measured ng/kg from Ottawa Sand P&A Data (400 ng/kg spike for all PFAS except 2000 ng/kg for PFBA and PFPeA and 8000 ng/kg spike for FHEA,
FDEA, and FOEA)
Sample
PFTreA PFTriA PFDoA PFUnA PFDA PFNA PFOA PFHpA PFHxA PFPeA PFBA
Unspiked
Unspiked
P&A 1 389.6 394.3 384.7 376.7 362.1 347.6 345.8 232.9 222.2 1614.9 1344.5
P&A 2 462.1 424.6 397.2 379.1 378.4 376.9 365.9 247.9 229.8 1710.1 1388
P&A 3 402.7 387.7 383.1 365.9 374.7 363.3 347.1 242.4 222.9 1658.9 1376
P&A 4 403.9 397.1 395.4 381.5 379 359.4 342.7 246.8 225.8 1693.6 1401.9
P&A 5 467.2 445.8 412.6 388.5 376.8 370.3 369.7 249.3 231.4 1716.5 1433.4
P&A 6 392.1 385.3 374.2 370.9 353.2 351.7 340.3 236.7 220.5 1659 1366.4
Mean
Recovery 419.6 405.8 391.2 377.1 370.7 361.5 351.9 242.7 225.4 1675.5 1385
(ng/kg)
% Mean 104.9 101.4 97.8 94.3 92.7 90.4 88 60.7 56.4 83.8 69.3
Recovery
Standard 35.4 24.1 13.5 8 10.6 11.1 12.6 6.6 4.4 38.5 30.7
Deviation
RSD (%) 8.4 5.9 3.5 2.1 2.9 3.1 3.6 11 1.9 2.3 2.2
Sample PFBS PFHxS PFOS PFechS FOUEA FHpPA FHUEA FHEA FOEA FDEA
Unspiked
Unspiked
P&A 1 337.4 349.1 340.3 342.8 389.5 371.3 372.5 7023.5 8202.6 8564.9
P&A 2 347.3 358.3 345.9 347.2 408.7 377.2 387.1 7346.1 8542.6 9308
P&A 3 366.3 330.1 331.7 345.4 401.5 361.4 379 6844.3 7402.4 8989.2
P&A 4 348.2 343.6 338.3 347.6 404.9 377.5 388.1 7258.2 7551.9 9173.4
P&A 5 351.8 361.7 365.6 362.6 417.5 395.1 391.8 7461.3 7821.2 9287.4
P&A 6 336.7 343.4 363.7 342.5 394.5 356.9 374.5 7559.3 8002.2 8367.1
Mean
Recovery 347.9 347.7 347.7 348 402.7 373.2 382.1 7248.8 7920.5 8948.3
(ng/kg)
% Mean 87 86.9 86.9 87 100.7 93.3 95.5 90.6 99 111.9
Recovery
Standard 10.9 11.5 13.9 7.4 10 13.6 7.9 270.4 421.3 395.3
Deviation
RSD (%) 3.1 3.3 4 2.1 2.5 3.6 2.1 3.7 5.3 4.4
D7968 − 23
TABLE 5 Single-Laboratory Surrogate Recovery Data in Ottawa Sand
Measured ng/kg from Ottawa Sand – 400 ng/kg spike
Sample
MPFBA MPFHxA MPFHxS MPFOA MPFNA MPFOS MPFDA MPFUnA MPFDoA
Unspiked 1 420.0 433.5 431.8 428.0 439.4 429.2 442.6 443.3 447.7
Unspiked 2 366.5 396.8 378.5 384.9 389.8 373.6 404.9 400.8 425.8
P&A 1 361.1 364.3 356.3 377.0 376.6 354.4 384.9 391.3 409.3
P&A 2 383.6 378.4 357.3 389.4 379.7 375.7 395.7 399.2 412.2
P&A 3 374.5 378.5 375.4 390.5 378.6 372.4 382.5 386.9 402.2
P&A 4 370.1 384.4 366.1 396.3 384.4 374.2 397.8 406.2 420.5
P&A 5 370.1 386.8 372.0 395.7 381.1 372.8 394.4 399.9 421.5
P&A 6 363.6 384.8 356.1 397.9 384.9 368.6 389.5 392.3 402.9
Mean
Recovery
376.2 388.4 374.2 394.9 389.3 377.6 399.0 402.5 417.7
(ng/kg dry
weight)
% Mean 94.0 97.1 93.5 98.7 97.3 94.4 99.8 100.6 104.4
Recovery
Standard 19.0 20.4 24.9 15.0 20.7 21.9 19.0 17.6 14.9
Deviation
RSD (%) 5.1 5.3 6.7 3.8 5.3 5.8 4.8 4.4 3.6
from interferences by analyzing laboratory reagent blanks may be used. The retention times and order of elution may
under the same conditions as the samples. If found, measures change depending on the column used and need to be moni-
should be taken to remove the contamination or data should be tored.
qualified; background subtraction of blank contamination is not
7.1.3 Isolator Column—A reverse phase C18 column was
allowed.
used in this test method to separate the target analytes in the LC
system and solvents from the target analytes in the analytical
6.5 The liquid chromatography system used should consist,
sample. This column was placed between the solvent mixing
as much as practical, of sample solution or eluent contacting
chamber and the injector sample loop.
components free of PFAS target analytes of interest.
7.2 Tandem Mass Spectrometer System—An MS/MS system
6.6 Polyethylene LC vial caps or any other target analyte-
capable of multiple reaction monitoring (MRM) analysis or
free vial caps should be used.
any system that is capable of meeting the requirements in this
6.7 Polyethylene disposable pipettes or target analyte-free
test method must be used.
pipettes should be used. All disposable pipettes should be
7.3 Centrifuge—A device to centrifuge the samples.
checked for release of target analytes of interest.
6.8 Degassers are important to continuous LC operation and
7.4 Lab Rotator —A device to mix the samples by end-
most commonly are made of fluorinated polymers. To enable over-end rotation.
use, an isolator column should be placed after the degasser and
7.5 Filtration Device:
prior to the sample injection valve to separate the PFAS in the
7.5.1 Hypodermic Syringe—A luer-lock tip glass syringe
sample from the PFAS in the LC system.
capable of holding a syringe-driven filter unit.
7.5.2 A 10 mL lock tip glass syringe size is recommended
7. Apparatus
since a 10 mL sample size is used in this test method.
7.1 LC/MS/MS System:
7.5.3 Filter Unit —Polypropylene filter units were used to
7.1.1 Liquid Chromatography System—A complete LC sys-
filter the samples.
tem is required in order to analyze samples; this should include
a sample injection system, a solvent pumping system capable
8. Reagents and Materials
of mixing solvents, a sample compartment capable of main-
taining required temperature, and a temperature-controlled
8.1 Purity of Reagents—High performance liquid chroma-
column compartment. An LC system that is capable of per-
tography (HPLC) pesticide residue analysis and spectropho-
forming at the flows, pressures, controlled temperatures,
tometry grade chemicals must be used in all tests. Unless
sample volumes, and requirements of the standard must be
indicated otherwise, it is intended that all reagents must
used.
conform to the Committee on Analytical Reagents of the
7.1.2 Analytical Column —A reverse phase Charged Sur-
American Chemical Society. Other reagent grades may be
face Hybrid Phenyl-Hexyl particle column was used to develop
this test method. Any column that achieves adequate resolution
A lab rotator, or equivalent, has been found suitable to mix samples.
8 10
A Waters Acquity UPLC CSH Phenyl-Hexyl, 2.1 × 100 mm and 1.7 μm particle A 0.2 μm polypropylene membrane syringe-driven filter unit, or equivalent,
size column, or equivalent, has been found suitable for use. It was used to develop has been found suitable for use.
this test method and generate the precision and bias data presented in Section 16. If Reagent Chemicals, American Chemical Society Specifications, American
you are aware of an alternative column that meets the performance of the standard, Chemical Society, Washington, D.C. For suggestions on the testing of reagents not
please provide this information to ASTM International Headquarters. Your com- listed by the American Chemical Society, see Analar Standards for Laboratory
ments will receive careful consideration at the meeting responsible technical Chemicals, EDH Ltd., Poole, Dorset, U.K. and the United States Pharmacopeia and
committee, which you may attend. National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD.
D7968 − 23
used provided they are first determined to be of sufficiently 8.19.14 Perfluorotetradecanoate (PFTreA, CAS No. 376-06-
high purity to permit their use without affecting the accuracy of 7).
the measurements. 8.19.15 Decafluoro-4-
(pentafluoroethyl)cyclohexanesulfonate (PFechS, CAS No.
8.2 Purity of Water—Unless otherwise indicated, references
67584-42-3).
to water must be understood to mean reagent water conforming
8.19.16 3-perfluoropheptyl propanoic acid (FHpPA, CAS
to Type I of Specification D1193. It must be demonstrated that
No. 812-70-4).
this water does not contain contaminants at concentrations
8.19.17 2H-perfluoro-2-decenoic acid (FOUEA, CAS No.
sufficient to interfere with the analysis.
70887-84-2).
8.3 Gases—Ultrapure nitrogen and argon.
8.19.18 2-perfluorodecyl ethanoic acid (FDEA, CAS num-
ber not available).
8.4 Vials—2 mL amber glass or polypropylene autosampler
vials or equivalent. 8.19.19 2-perfluorooctyl ethanoic acid (FOEA, CAS No.
27854-31-5).
8.5 Polyethylene or any PFAS-free applicable autosampler
8.19.20 2H-perfluoro-2-octenoic acid (FHUEA, CAS num-
vial caps.
ber not available).
8.6 Syringe—10 or 25 mL filter-adaptable glass syringe with
8.19.21 2-perfluorohexyl ethanoic acid (FHEA, CAS No.
luer lock.
53826-12-3).
8.7 pH paper (pH range 1 to 14).
8.20 PFAS Surrogates:
8.20.1 O -Perfluorohexylsulfonate (MPFHxS).
8.8 Polypropylene Tubes—15 and 50 mL.
8.20.2 C -Perfluorooctylsulfonate (MPFOS).
8.9 Class A volumetric glassware.
8.20.3 C -Perfluorobutanoate (MPFBA).
8.10 Pipette Tips—Polypropylene pipette tips free of release
8.20.4 C -Perfluorohexanoate (MPFHxA).
agents or low retention coating of various sizes.
8.20.5 C -Perfluorooctanoate (MPFOA).
8.20.6 C -Perfluorononanoate (MPFNA).
8.11 Polyethylene disposable pipettes. 5
8.20.7 C -Perfluorodecanoate (MPFDA).
8.12 Acetonitrile (CAS No. 75-05-8).
8.20.8 C -Perfluoroundecanoate (MPFUnA).
8.13 Methanol (CAS No. 67-56-1).
8.20.9 C -Perfluorododecanoate (MPFDoA).
8.14 Ammonium acetate (CAS No. 631-61-8).
9. Hazards
8.15 Acetic acid (CAS No. 64-19-7).
9.1 Normal laboratory safety applies to this method. Ana-
8.16 2-Propanol (isopropyl alcohol, CAS No. 67-63-0).
lysts should wear safety glasses, gloves, and lab coats when
working in the lab. Analysts should review the Safety Data
8.17 Ammonium hydroxide (CAS No. 1336-21-6).
Sheets (SDS) for all reagents used in this method.
8.18 Ottawa sand (CAS No. 14808-60-7).
8.19 PFAS Standards: 10. Sampling
8.19.1 Perfluorobutylsulfonate (PFBS, CAS No. 29420-49-
10.1 Sampling and Preservation—Grab samples are col-
3).
lected in glass or polypropylene containers. Sample containers
8.19.2 Perfluorohexylsulfonate (PFHxS, CAS No. 3871-99-
and contact surfaces with PFAS must be avoided. As part of the
6).
overall quality assurance program for this test method, field
8.19.3 Perfluorooctylsulfonate (PFOS, CAS No. 1763-23-
blanks exposed to the same field conditions as samples are
1).
collected and analyzed according to this test method to assess
8.19.4 Perfluorobutanoate (PFBA, CAS No. 375-22-4).
the potential for field contamination. This test method is based
8.19.5 Perfluoropentanoate (PFPeA, CAS No. 2706-90-3).
on a 2 g sample size per analysis. If different sample sizes are
8.19.6 Perfluorohexanoate (PFHxA, CAS No. 307-24-4).
used, spiking solution amounts may need to be modified.
8.19.7 Perfluoroheptanoate (PFHpA, CAS No. 375-85-9).
Conventional sampling practices should be followed with the
8.19.8 Perfluorooctanoate (PFOA, CAS No. 335-67-1).
caution that PFAS-containing products may be present in
8.19.9 Perfluorononanoate (PFNA, CAS No. 375-95-1)
sampling equipment. All sampling equipment and supplies
8.19.10 Perfluorodecanoate (PFDA, CAS No. 335-76-2).
must be PFAS free in order to prevent contamination of the
8.19.11 Perfluoroundecanoate (PFUnA, CAS No. 2058-94-
samples. EPA publication SW-846 may be used as a sampling
8).
guide. Samples must be shipped on ice with a trip blank. Once
8.19.12 Perfluorododecanoate (PFDoA, CAS No. 307-55-
received, the sample temperature is taken and must be between
1).
freezing and 6 °C. If the receiving temperature is greater than
8.19.13 Perfluorotridecanoate (PFTriA, CAS No. 72629-94-
6 °C, the sample temperature is noted in the case narrative
8).
accompanying the data. Samples should be stored refrigerated
between 0 and 6 °C from the time of collection until analysis.
PFAS standards may be difficult to find; some sources of PFAS standards that
have been found suitable for use were from Aldrich Chemical Company,
Accustandard, Wellington Laboratories, Inc., and Wako Laboratory. Standards from PFAS surrogates from Wellington Laboratories Inc., or equivalent, have been
other vendors may be used. found suitable for use.
D7968 − 23
The sample should be analyzed within 28 days of collection. 11.3.1.6 Desolvation Gas Temperature: 450 °C.
No holding time study has been done on the various soil 11.3.1.7 Desolvation Gas Flow: 800 L/h.
matrices tested in this test method. Holding time may vary 11.3.1.8 Cone Gas Flow: 200 L/h.
depending on the matrix, and individual laboratories should 11.3.1.9 Collision Gas Flow: 0.15 mL/min.
determine the holding time in their matrix. 11.3.1.10 Low Mass Resolution 1: 2.6.
11.3.1.11 High Mass Resolution 1: 14.
11. Preparation of LC/MS/MS
11.3.1.12 Ion Energy 1: 1.
11.3.1.13 Entrance Energy: 1.
11.1 LC Chromatograph Operating Conditions:
11.3.1.14 Collision Energy: Variable depending on analyte.
11.1.1 Injections of all standards and samples are made at a
11.3.1.15 Exit Energy: 1.
30 μL volume. Other injection volumes may be used to
11.3.1.16 Low Mass Resolution 2: 2.5.
optimize conditions. Standards and sample extracts must be in
11.3.1.17 High Mass Resolution 2: 14.
a 50:50 methanol:water solution containing 0.1 % acetic acid.
In the case of extreme concentration differences amongst 11.3.1.18 Ion Energy 2: 3.
11.3.1.19 Gain: 1.0.
samples, it is wise to analyze a blank after a concentrated
sample and before a dilute sample to minimize carryover of 11.3.1.20 Multiplier: 511.1.
11.3.1.21 Inter-Scan Delay: 0.004 s.
analytes from injection to injection. However, there should not
be carryover between samples. The LC utilized to develop this
test method has a flow-through LC needle design. The gradient 12. Calibration and Standardization
conditions for liquid chromatography are shown in Table 6.
12.1 The mass spectrometer must be calibrated as per
11.2 LC Sample Manager Conditions: manufacturer’s specifications before analysis. Analytical val-
11.2.1 Needle Wash Solvent—60 % acetonitrile/40 % ues satisfying test method criteria have been achieved using the
2-propanol; Time: 5 min.
following procedures. Prepare all solutions in the lab using
Class A volumetric glassware.
11.3 Mass Spectrometer Parameters:
11.3.1 To acquire the maximum number of data points per
12.2 Calibration and Standardization—To calibrate the
SRM channel while maintaining adequate sensitivity, the tune instrument, analyze nine calibration standards of the polyfluo-
parameters may be optimized according to the instrument used.
rinated compounds prior to sample analysis as shown in Table
Each peak requires at least ten scans per peak for adequate 2. Calibration stock standard solution is prepared from the
quantitation. This test method contains nine surrogates, which
target and surrogate spike solutions directly to ensure consis-
are isotopically labeled PFAS, and 21 PFAS which are split up tency. Stock standard Solution A containing the polyfluorinated
into 18 MRM acquisition functions to optimize sensitivity.
compounds and surrogates is prepared at Level 9 concentration
Variable parameters regarding retention times, SRM and aliquots of that solution are diluted to prepare Levels 1
transitions, and cone and collision energies are shown in Table
through 8. The following steps will produce standards with the
3. Mass spectrometer parameters used in the development of concentration values shown in Table 2. The analyst is respon-
this method are listed below: sible for recording initial component weights carefully when
11.3.1.1 The instrument is set in the Electrospray negative working with pure materials and correctly carrying the weights
source setting. through the dilution calculations. At a minimum five calibra-
11.3.1.2 Capillary Voltage: 0.75 kV. tion levels are required when using a linear calibration curve
11.3.1.3 Cone: Variable depending on analyte. and six calibration levels are required when using a quadratic
11.3.1.4 Extractor: 2 Volts. calibration curve. An initial nine-point curve may be used to
11.3.1.5 Source Temperature: 150 °C. allow for the dropping of the lower level calibration points if
the individual laboratory’s instrument cannot achieve low
detection limits on certain PFAS. This should allow for at least
A guide to help and determine sample holding times can be found at
a five or six-point calibration curve to be obtained. No
http://www.epa.gov/esd/cmb/research/bs_033cmb06.pdf (2014).
problems were encountered while using the nine-point calibra-
tion curve in developing this test method.
TABLE 6 Gradient Conditions for Liquid Chromatography
12.2.1 Calibration stock standard Solution A (Level 9, Table
Percent
2) is prepared from the target and surrogate spike solutions
95 % Water:
Percent directly to ensure consistency. 500 μL of surrogate spike
5 %
Time Flow 95 % Water: Percent
Acetonitrile, (20 μg ⁄L), 500 μL Target Spike I, and 500 μL of PFAS Target
(min) (mL/min) 5 % Acetonitrile
400 mM
Spike II (refer to Table 7) is added to a 50 mL volumetric flask
Acetonitrile
Ammonium
and diluted to 50 mL with 50:50 methanol:water containing
Acetate
0 0.3 95 0 5 0.1 % acetic acid. The preparation of the Level 9 standard can
1 0.3 75 20 5
be accomplished using appropriate volumes and concentrations
6 0.3 50 45 5
of stock solutions as per a particular laboratory’s standard
13 0.3 15 80 5
14 0.4 0 95 5 procedure. It is critical to ensure that analytes are solubilized in
17 0.4 0 95 5
the Level 9 standard.
18 0.4 95 0 5
12.2.2 Aliquots of Solution A are then diluted with 50:50
21 0.4 95 0 5
methanol:water containing 0.1 % acetic acid to prepare the
D7968 − 23
TABLE 7 PFAS Target Spike Solutions (PPB)
Concentration of Analyte in PFAS Target Spike Solutions
Analyte PFAS High Target Spike Solutions
PFAS Reporting Limit Spike Solution
Target Spike I Target Spike II
PFTreA, PFTriA, PFDoA, PFUnA,
PFDA, PFOS, PFNA, PFHxA, PFHpA, 20 μg/L — 2 μg/L
PFBS, PFechS, PFOA, PFHxS
PFBA, PFPeA 100 μg/L — 10 μg/L
FOUEA, FHUEA, FHpPA — 20 μg/L 2 μg/L
FHEA, FOEA, FDEA — 400 μg/L 40 μg/L
desired calibration levels in 2 mL LC vials. The calibration 12.2.6 Linear calibration may be used if the point of origin
vials must be used within 24 h to ensure optimum results. The is excluded and a fit weighting 1/X is used in order to give
end calibration check must be prepared in a separate LC vial
more emphasis to the lower concentrations. Each calibration
near the mid-level. All calibration standards should only be point used to generate the curve must have a calculated percent
used once. The analyte concentration in the vial may change
deviation less than 25 % from the generated curve.
after the vial cap is pierced because the vial caps may not reseal
12.2.7 Quadratic calibration may be used if the point of
after puncture; if the cap reseals it may be used over again.
origin is excluded, and a fit weighting of 1/X is used in order
Changing the caps immediately after the injection should
to give more emphasis to the lower concentrations. Each
alleviate this problem. Calibration standards are not filtered.
calibration point used to generate the curve must have a
12.2.3 Inject each standard and obtain its chromatogram. An
calculated percent deviation less than 25 % from the generated
external calibration technique is used to monitor the primary
curve.
and confirmatory SRM transitions of each analyte. Calibration
12.2.8 The retention time window of the SRM transitions
software is utilized to conduct the quantitation of the target
must be within 5 % of the retention time of the analyte in a
analytes and surrogates using the primary SRM transition. The
midpoint calibration standard. If this is not the case, re-analyze
ratios of the primary/confirmatory SRM transition area counts
the calibration curve to determine if there was a shift in
are given in Table 3 and will vary depending on the individual
retention time during the analysis and the sample needs to be
tuning conditions. The primary/confirmatory SRM transition
re-injected. If the retention time is still incorrect in the sample,
area ratio must be within 35 % of the individual lab’s accepted
refer to the analyte as an unknown.
primary/confirmatory SRM transition area ratio. The primary
12.2.9 A midpoint calibration check standard must be ana-
SRM transition of each analyte is used for quantitation of and
lyzed at the end of each batch of 30 samples or within 24 h
the confirmatory SRM transition for confirmation. This gives
after the initial calibration curve was generated; the criteria in
added confirmation by isolating the parent ion, forming two
the individual lab’s quality system may be more restrictive
product ions via fragmentation, and relating it to the retention
pertaining to the number of samples. This end calibration
time in the calibration standard.
check, a new not pierced sealed vial, should come from the
12.2.4 Depending on sensitivity and matrix interference
same calibration standard solution that was used to generate the
issues dependent on sample type, the confirmatory SRM
initial curve. The results from the end calibration check
transition can be used as the primary SRM transition for
standard must have a percent deviation less than 30 % from the
quantitation during analysis. This must be explained in a
calculated concentration for the target analytes and surrogates.
narrative accompanying the data. A new primary/confirmatory
If the results are not within these criteria, corrective action
ion ratio will then be determined if switching the SRM
including reoccurrence minimization is performed and either
transitions used to quantitate and confirm. The new primary/
all samples in the batch are re-analyzed against a new
confirmatory SRM transition area ratio is required to be within
calibration curve or the affected results are qualified with an
35 % of the individual lab’s new primary/confirmatory SRM
indication that they do not fall within the performance criteria
transition area ratio.
of the test method. If the analyst inspects the vial containing
12.2.5 The calibration software manual should be consulted
the end calibration check standard and notices that the sample
to use the software correctly. The quantitation method is set as
evaporated affecting the concentration or other anomaly, a new
an external calibration using the peak areas in ppt units.
end calibration check standard may be made and analyzed. If
Concentrations may be calculated using the data system
this new end calibration check standard has a percent deviation
software to generate linear regression or quadratic calibration
less than 30 % from the calculated concentration for the target
curves. Forcing the calibration curve through the origin (X = 0,
analytes and surrogates, the results may be reported unquali-
Y = 0) is not recommended. Curves should be evaluated using
15 fied.
relative error or relative standard error.
12.3 If a laboratory has not performed the test before or if
there has been a major change in the measurement system, for
Management and Technical Requirements for Laboratories Performing Envi-
example, new analyst, new instrument, etc., an instrument
ronmental Analysis, Module 4: Quality Systems for Chemical Testing; The NELAC
Institute, 2017. qualification study including method detection limit (MDL),
D7968 − 23
calibration range determination, and precision and bias deter- method. References generating QC acceptance criteria are
mination must be performed to demonstrate laboratory capa- ASTM Practices D2777, D5847, E2554, or Method 8000 in
bility. EPA publication SW-846.
12.3.1 Analyze at least four replicates of a spiked sand
12.4 Surrogate Spiking Solution:
sample containing the PFAS and surrogates at an extract
12.4.1 A surrogate spiking solution containing nine isotopi-
concentration in the calibration range of Levels 4 through 7.
cally labeled PFAS—MPFBA, MPFHxA, MPFHxS, MPFDA,
The Level 6 concentration of the nine-point calibration curve
MPFOA, MPFOS, MPFNA, MPFUnA, and MPFDoA—is
was used to set the QC acceptance criteria in this method. The
added to all samples, method blanks, duplicates, laboratory
matrix and chemistry should be similar to the matrix used in
control samples, matrix spikes, and reporting limit checks. A
this test method. Each replicate must be taken through the
stock surrogate spiking solution is prepared at 20 μg/L in 95 %
complete analytical test method including any sample manipu-
acetonitrile: 5 % water. Spiking 40 μL of this spiking solution
lation and extraction steps.
into a 2 g soil sample results in a concentration of 400 ng/kg of
12.3.2 Calculate the mean (average) percent recovery and
the surrogate in the sample. The results obtained for the
relative standard deviation (RSD) of the four values and
surrogate recoveries must fall within the limits of Table 8. If
compare to the acceptable ranges of the QC acceptance criteria
the limits are not met, the affected results must be qualified
for the Initial Demonstration of Performance in Table 8.
with an indication that they do not fall within the performance
12.3.3 This study should be repeated until the single-
criteria of the test method.
operator precision and mean recovery are within the limits in
12.4.1.1 The surrogate spiking solution was prepared by
Table 8. If a concentration other than the recommended
adding 500 μL of a 2 mg/L PFAS surrogate mix in a 50 mL
concentration is used, refer to Practice D5847 for information
volumetric and diluted to 50 mL with 95 % acetonitrile: 5 %
on applying the F test and t test in evaluating the acceptability
water. Surrogate spiking solutions are routinely replaced every
of the mean and standard deviation.
year if not previously discarded for quality control failure.
12.3.3.1 The QC acceptance criteria for the Initial Demon-
12.5 Method Blank:
stration of Performance in Table 8 were generated from the
single-laboratory data shown in Section 16. Data from Ottawa
sand and four ASTM soil matrices are shown in Section 16. It
is recommended that each laboratory determine in-house QC 16
Surrogate mix from Wellington Laboratories, Inc. has been found suitable for
acceptance criteria which meet or exceed the criteria in this test use.
TABLE 8 QC Acceptance Criteria
NOTE 1—Table 8 data is preliminary until a multi-lab validation study is completed.
Initial Demonstration of Performance Laboratory Control Sample
Recovery (%) Precision Recovery (%)
Analyte/Surrogate Spike Conc. ng/kg
Maximum Lower Control Limit Upper Control Limit
Lower Limit Upper Limit
% RSD (LCL) % (UCL) %
PFTreA 400 70 130 30 70 130
PFTriA 400 70 130 30 70 130
PFDoA 400 70 130 30 70 130
PFUnA 400 70 130 30 70 130
PFDA 400 70 130 30 70 130
PFOS 400 70 130 30 70 130
PFNA 400 70 130 30 70 130
PFecHS 400 70 130 30 70 130
PFOA 400 70 130 30 70 130
PFHxS 400 70 130 30 70 130
PFHpA 400 50 130 30 50 130
PFHxA 400 50 130 30 50 130
PFBS 400 70 130 30 70 130
PFPeA 2000 70 130 30 70 130
PFBA 2000 50 130 30 50 130
FHEA 8000 70 130 30 70 130
FOEA 8000 70 130 30 70 130
FDEA 8000 70 130 30 70 130
FOUEA 400 70 130 30 70 130
FHpPA 400 70 130 30 70 130
FHUEA 400 70 130 30 70 130
MPFBA 400 70 130 30 70 130
MPFHxA 400 70 130 30 70 130
MPFHxS 400 70 130 30 70 130
MPFOA 400 70 130 30 70 130
MPFNA 400 70 130 30 70 130
MPFOS 400 70 130 30 70 130
MPFDA 400 70 130 30 70 130
MPFUnA 400 70 130 30 70 130
MPFDoA 400 70 130 30 70 130
D7968 − 23
12.5.1 At least two method blanks for every 30 samples are I and II in 95 % acetonitrile: 5 % water containing the 21 PFAS
prepared in 2 g of Ottawa sand to investigate for contamination at concentrations listed in Table 7. Spike 40 μL of these stock
during sample preparation and extraction. The concentration of solutions into 2 g of the site sample to yield a concentration of
target analytes in either/both blank(s) must be at less than half 2000 ng/kg (PFBA and PFPeA), 8000 ng/kg (FHEA, FDEA,
the reporting limit or the data must be qualified as having a and FOEA), and 400 ng/kg of remaining 16 PFAS (PFTreA,
blank issue and the reporting limit must be raised to at least PFTriA, PFDoA, PFUnA, PFDA, PFOS, PFNA, PFHxA,
three times above the blank contamination concentration. PFHpA, PFBS, PFechS, PFOA, PFHxS, FOUEA, FHUEA,
PFAS are common in the environment and laboratories requir- AND FHpPA) in the sample.
ing an additional method blank sample. 12.8.2 If the spiked concentration plus the background
concentration exceeds that of the Level 9 calibration standard,
12.6 Reporting Limit Check Sample (RLCS):
the sample must be diluted (using 50 % Methanol/50 % Water
12.6.1 Each batch or within the 24 h analysis window a
with 0.1 % acetic acid) to a level near the midpoint of the
reporting limit check sample must be analyzed. The reporting
calibration curve.
limit check sample is processed like a laboratory control
12.8.3 Calculate the percent recovery of the spike (P) using
sample just spiked at or near (one to two times) the reporting
Eq 1:
limit. The concentration of the RLCS may be reported below
A V 1 V 2 BV
~ !
? s s?
the reporting limit since the spike is at or near the reporting
P 5 100 (1)
CV
limit. This sample is to check if the analytes were present at the
reporting limit, they would be identified. The recovery limits
where:
for the RLCS are 35 to 150 %, if any analytes are outside of
A = concentration found in spiked sample,
these limits the QC failure is explained in a narrative accom-
B = concentration found in unspiked sample,
panying the data.
C = concentration of analyte in spiking solution,
12.6.2 Two grams of Ottawa sand is added to a 15 mL
V = volume of sample used,
s
polypropylene centrifuge tube. The sample is spiked with
V = volume of spiking solution added, and
40 μL of PFAS surrogate spiking solution and 25 μL of PFAS P = percent recovery.
reporting limit check solution (Table 7) and then taken through
12.8.4 The percent recovery of the spike must fall within the
the sample preparation and analyzed.
limits in Table 9. If the percent recovery is not wit
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D7968 − 17a D7968 − 23
Standard Test Method for
Determination of Polyfluorinated Compounds in Soil by
Liquid Chromatography Tandem Mass Spectrometry (LC/
MS/MS)
This standard is issued under the fixed designation D7968; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This procedure covers the determination of selected polyfluorinated compounds (PFCs) alkyl substances (PFAS) in a soil
matrix using solvent extraction, filtration, followed by liquid chromatography (LC) and detection with tandem mass spectrometry
(MS/MS). These analytes are qualitatively and quantitatively determined by this method. This method adheres to multiple reaction
monitoring (MRM) mass spectrometry. This procedure utilizes a quick extraction and is not intended to generate an exhaustive
accounting of the content of PFCsPFAS in difficult soil matrices. An exhaustive extraction procedure for polyfluoralkyl substances,
PFAS, such as published by Washington et al., for difficult matrices should be considered when analyzing PFCs.PFAS. The
approach from this standard was utilized to screen laboratory coats (textiles) to identify if PFAS would be leached from the
materials.
1.2 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this
standard.
3 4
1.3 The Method of Detection Limitmethod of detection limit and Reporting Rangereporting range for the target analytes are
listed in Table 1.
1.3.1 The reporting limit in this test method is the minimum value below which data are documented as non-detects. Analyte
detections between the method detection limit and the reporting limit are estimated concentrations and are not reported following
this test method. In most cases, the reporting limit is calculated from the concentration of the Level 1 calibration standard as shown
in Table 2 for the polyfluorinated compounds PFAS after taking into account a 2-g2 g sample weight and a final extract volume
of 10 mL, 50 % water/50 % MeOH with 0.1 % acetic acid. The final extract volume is assumed to be 10 mL because 10 mL of
50 % water/50 % MeOH with 0.1 % acetic acid was added to each soil sample and only the liquid layer after extraction is filtered,
leaving the solid and any residual solvent behind. It is raised above the Level 1 calibration concentration for PFOS, PFHxA, FHEA,
and FOEA; these compounds can be identified at the Level 1 concentration but the standard deviation among replicates at this
lower spike level resulted in a higher reporting limit.
This test method is under the jurisdiction of ASTM Committee D34 on Waste Management and is the direct responsibility of Subcommittee D34.01.06 on Analytical
Methods.
Current edition approved Sept. 1, 2017Nov. 1, 2023. Published September 2017November 2023. Originally approved in 2014. Last previous edition approved in 2017 as
D7968 – 17.D7968 – 17a. DOI: 10.1520/D7968-17A.10.1520/D7968-23.
Washington, J. W., Naile, J. E., Jenkins, T. M., and Lynch, D. G., “Characterizing Fluorotelomer and Polyfluoroalkyl Substances in New and Aged Fluorotelomer-Based
Polymers for Degradation Studies with GC/MS and LC/MS/MS,” Environmental Science and Technology, Vol 48, 2014, pp. 5762–5769.
The MDL is determined following the Code of Federal Regulations, 40 CFR Part 136, Appendix B utilizing solvent extraction of soil. Two-gram A 2 g sample of Ottawa
sand was utilized. A detailed process determining the MDL is explained in the reference and is beyond the scope of this standard to be explained here.
Reporting range concentration is calculated from Table 2 concentrations assuming a 30-μL 30 μL injection of the Level 1 calibration standard for the PFCs,PFAS, and
the highest level calibration standard with a 10-mL 10 mL final extract volume of a 2-g2 g soil sample. Volume variations will change the reporting limit and ranges.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7968 − 23
A
TABLE 1 Method Detection Limit and Reporting Range
MDL Reporting Limit
Analyte
(ng/kg) (ng/kg)
PFTreA 6.76 25–1000
PFTriA 5.26 25–1000
PFDoA 3.56 25–1000
PFUnA 2.45 25–1000
PFDA 5.54 25–1000
PFOS 18.83 50–1000
PFNA 2.82 25–1000
PFecHS 2.41 25–1000
PFOA 6.24 25–1000
PFHxS 7.75 25–1000
PFHpA 5.80 25–1000
PFHxA 15.44 50–1000
PFBS 6.49 25–1000
PFPeA 20.93 125–5000
PFBA 22.01 125–5000
FHEA 199.04 600–20 000
FOEA 258.37 750–20 000
FDEA 137.46 500–20 000
FOUEA 4.85 25–1000
FhpPa 5.09 25–1000
FHUEA 3.50 25–1000
A
Abbreviations are defined in 3.2.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
D1193 Specification for Reagent Water
D2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
D5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis
E2554 Practice for Estimating and Monitoring the Uncertainty of Test Results of a Test Method Using Control Chart Techniques
2.2 Other Documents:
EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods
40 CFR Part 136 Appendix B Definition and Procedure for the Determination of the Method Detection Limit
3. Terminology
3.1 Definitions:
3.1.1 reporting limit, RL, n—the minimum concentration below which data are documented as non-detects.
3.1.2 polyfluorinated compounds, PFCs, n—in this test method, eleven perfluoroalkyl carboxylic acids, three
perfluoroalkylsulfonates, Decafluoro-4-(pentafluoroethyl)cyclohexanesulfonate, and six fluorotelomer acids listed in Table 1
collectively (not including mass labeled surrogates).
3.2 Abbreviations:
3.2.1 CCC—Continuing Calibration Check
3.2.2 IC—Initial Calibration
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from National Technical Information Service (NTIS), U.S. Department of Commerce, 5285 Port Royal Road, Springfield, VA, 22161, http://www.epa.gov/
epawaste/hazard/testmethods/index.htm
D7968 − 23
TABLE 2 Concentrations of Calibration Standards (ng/L)
Analyte/Surrogate LV1 LV2 LV3 LV4 LV5 LV6 LV7 LV8 LV9
PFPeA, PFBA 25 50 100 200 300 400 500 750 1000
PFTreA, PFTriA, PFDoA, PFUnA, PFDA, PFOS, PFNA, PFHxA, PFHpA, PFBS,
PFechS, PFOA, PFHxS, FOUEA, FHUEA, FHpPA, MPFBS, MPFHxA, MPFUnA, 5 10 20 40 60 80 100 150 200
MPFOA, MPFDA, MPFOS, MPFNA, MPFHxS, MPFBA
FHEA, FOEA, FDEA 100 200 400 800 1200 1600 2000 3000 4000
3.2.3 ppt—parts per trillion, ng/kg or ng/L
3.2.4 LC—Liquid Chromatography
3.2.5 LCS/LCSD—Laboratory Control Sample/Laboratory Control Sample Duplicate
3.2.6 MDL—Method Detection Limit
3.2.7 MeOH—Methanol
-3–3
3.2.8 mM—millimolar, 1 × 10 moles/L
3.2.9 MRM—Multiple Reaction Monitoring
3.2.10 MS/MSD—Matrix Spike/Matrix Spike Duplicate
3.2.11 NA—Not available
3.2.12 ND—non-detectNon-detect
3.2.13 P&A—Precision and Accuracy
3.2.14 PFAS—PerfluoroalkylsulfonatePerfluoroalkyl substances
3.2.15 PFBS—perfluorobutylsulfonatePerfluorobutylsulfonate
3.2.16 PFHxS—perfluorohexylsulfonatePerfluorohexylsulfonate
3.2.17 PFOS—Perfluorooctylsulfonate
3.2.18 PFecHS—Decaluoro-4-(pentafluoroethyl)cyclohexanesulfonate
3.2.19 PFAC—Perfluoroalkyl Carboxylic Acid
3.2.20 PFBA—Perfluorobutanoate
3.2.21 PFPeA—Perfluoropentanoate
3.2.22 PFHxA—Perfluorohexanoate
3.2.23 PFHpA—Perfluoroheptanoate
3.2.24 PFOA—Perfluorooctanoate
D7968 − 23
3.2.25 PFNA—Perfluorononanoate
3.2.26 PFDA—Perfluorodecanoate
3.2.27 PFUnA—Perfluoroundecanoate
3.2.28 PFTriA—Perfluorotridecanoate
3.2.29 PFTreA—Perfluorotetradecanoate
3.2.30 FTAs and FTUAs—Fluorotelomer and Unsaturated Fluorotelomer Acids
3.2.31 FHpPA—3-perfluoropheptyl propanoic acid
3.2.32 FOUEA—2H-perfluoro-2-decenoic acid
3.2.33 FDEA—2-perfluorodecyl ethanoic acid
3.2.34 FOEA—2-perfluorooctyl ethanoic acid
3.2.35 FHUEA—2H-perfluoro-2-octenoic acid
3.2.36 FHEA—2-perfluorohexyl ethanoic acid
3.2.37 MPFAS—Isotopically labeled Perfluoroalkylsulfonates
3.2.38 MPFHxS— O -Perfluorohexylsulfonate
3.2.39 MPFOS— C -Perfluorooctylsulfonate
3.2.40 MPFCA—Isotopically labeled Perfluoroalkylcarboxylates
3.2.41 MPFBA— C -Perfluorobutanoate
3.2.42 MPFHxA— C -Perfluorohexanoate
3.2.43 MPFOA— C -Perfluorooctanoate
3.2.44 MPFNA— C -Perfluorononanoate
3.2.45 MPFDA— C -Perfluorodecanoate
3.2.46 MPFUnA— C -Perfluoroundecanoate
3.2.47 MPFDoA— C -Perfluorodecanoate
3.2.48 QA—Quality Assurance
3.2.49 QC—Quality Control
3.2.50 RL—Reporting Limit
D7968 − 23
3.2.51 RLCS—Reporting Limit Check Sample
3.2.52 RSD—Relative Standard Deviation
3.2.53 RT—Retention Time
3.2.54 SRM—Single Reaction Monitoring
3.2.55 SS—Surrogate Standard
3.2.56 TC—Target Compound
4. Summary of Test Method
4.1 The operating conditions presented in this test method have been successfully used in the determination of polyfluorinated
compounds in soil; however, this test method is intended to be performance based and alternative operating conditions can be used
to perform this method provided data quality objectives are attained.
4.2 For PFCPFAS analysis, samples are shipped to the lab on ice and analyzed within 28 ddays of collection. A sample (2 g) is
transferred to a polypropylene tube, spiked with surrogates (all samples) and target PFC compounds PFAS (laboratory control and
matrix spike samples). The analytes are tumbled for an hour with 10 mL of methanol:water (50:50) under basic condition (pH ~
9-10 ~9 to 10 adjusted with ~20 μL ammonium hydroxide). The samples are centrifuged and the extract, leaving the solid behind,
is filtered through a polypropylene filter unit. Acetic acid (~50 μL) is added to all the filtered samples to adjust the pH ~3-4 ~3
to 4 and then analyzed by LC/MS/MS.
4.3 Most of the PFCPFAS target compounds are identified by comparing the single reaction monitoring (SRM) transition and its
confirmatory SRM transition if correlated to the known standard SRM (Table 3) and quantitated utilizing an external calibration.
The surrogates and some PFCPFAS target analytes (PFPeA, PFBA, FOUEA, and FHUEA) only utilize one SRM transition due
to a less sensitive or non-existent secondary SRM transition. As an additional quality control measure, isotopically labeled PFC
surrogatesPFAS surrogate (listed in 12.4) recoveries are monitored. There is no correction to the data based upon surrogate
recoveries. The final report issued for each sample lists the concentration of PFCs,PFAS, if detected, or
in ng/kg (dry weight basis) and the surrogate recoveries.
5. Significance and Use
5.1 This test method has been developed by the U.S. EPA Region 5 Chicago Regional Laboratory (CRL).
5.2 PFCsPFAS are widely used in various industrial and commercial products; they are persistent, bio-accumulative, and
ubiquitous in the environment. PFCsPFAS have been reported to exhibit developmental toxicity, hepatotoxicity, immunotoxicity,
and hormone disturbance. A draft Toxicological Profile for Perfluoroalkyls from the U.S. Department of Health and Human
Services is available. PFCsPFAS have been detected in soils, sludges, and surface and drinking waters. Hence, there is a need for
a quick, easy, and robust method to determine these compounds at trace levels in various soil matrices for understanding of the
sources and pathways of exposure.
5.3 This method has been used to determine selected polyfluorinated compounds PFAS in sand (Table 4) and four ASTM reference
soils (Table 5).
6. Interferences
6.1 All glassware is washed in hot water with detergent and rinsed in hot water followed by distilled water. The glassware is then
dried and heated in an oven at 250 °C for 15 to 30 min. All glassware is subsequently rinsed with methanol or acetonitrile.
A draft Toxicological Profile for Perfluroalkyls can be found at http://www.atsdr.cdc.gov/toxprofiles/tp.asp?id=1117&tid=237 (2014).
D7968 − 23
TABLE 3 Retention Times, SRM Ions, and Analyte-Specific Mass Spectrometer Parameters
Primary/
Primary/ Retention Times
Chemical Cone (V) Collision (eV) MRM Transition Confirmatory SRM
Confirmatory (min)
Area Ratio
Primary 20 13 712.9→668.9
PFTreA 10.63 7.4
Confirmatory 20 30 712.9→169
Primary 25 12 662.9→618.9
PFTriA 10.17 7.4
Confirmatory 25 28 662.9→169
Primary 10 12 612.9→568.9
PFDoA 9.61 8.2
Confirmatory 10 25 612.9→169
Primary 15 10 562.9→519
PFUnA 9.05 7.2
Confirmatory 15 18 562.9→269
Primary 20 10 512.9→468.9
PFDA 8.45 6.5
Confirmatory 20 16 512.9→219
Primary 10 42 498.9→80.1
PFOS 8.78 1.3
Confirmatory 10 40 498.9→99.1
Primary 20 10 462.9→418.9
PFNA 7.78 4.9
Confirmatory 20 16 462.9→219
Primary 10 25 460.9→381
PFecHS 8.1 2.2
Confirmatory 10 25 460.9→99.1
Primary 20 10 412.9→369
PFOA 7.11 3.6
Confirmatory 20 16 412.9→169
Primary 15 32 398.9→80.1
PFHxS 7.39 1
Confirmatory 15 32 398.9→99.1
Primary 15 10 362.9→319
PFHpA 6.35 4.1
Confirmatory 15 15 362.9→169
Primary 15 8 312.9→269
PFHxA 5.54 24.1
Confirmatory 15 18 312.9→119.1
Primary 10 30 298.9→80.1
PFBS 5.66 1.6
Confirmatory 10 25 298.9→99.1
PFPeA Primary 4.68 10 8 263→219 NA
PFBA Primary 3.67 10 8 212.9→169 NA
Primary 15 20 376.9→293
FHEA 6.14 3.6
Confirmatory 15 6 376.9→313
Primary 15 18 476.9→393
FOEA 7.54 4.3
Confirmatory 15 12 476.9→413
Primary 15 8 576.8→493
FDEA 8.83 3.2
Confirmatory 15 15 576.8→513
FOUEA Primary 7.54 20 12 456.9→392.9 NA
Primary 15 12 440.9→337
FHpPA 7.54 1.1
Confirmatory 15 20 440.9→317
FHUEA Primary 6.08 10 12 357→293 NA
MPFBA Primary 3.67 10 7 217→172.1 NA
MPFHxA Primary 5.54 15 8 315→270 NA
MPFHxS Primary 7.39 15 34 402.9→84.1 NA
MPFOA Primary 7.11 15 10 417→372 NA
MPFNA Primary 7.81 15 9 467.9→423 NA
MPFOS Primary 8.78 15 40 502.9→80.1 NA
MPFDA Primary 8.45 15 10 514.9→470 NA
MPFUnA Primary 9.05 15 10 564.9→519.9 NA
MPFDoA Primary 9.61 15 12 614.9→569.9 NA
6.2 All reagents and solvents should be pesticide residue purity or higher to minimize interference problems. The use of
PFC-containing caps shouldPFAS-containing caps must be avoided.
6.3 Matrix interferences may be caused by contaminants in the sample. The extent of matrix interferences can vary considerably
depending on variations in the sample matrices.
6.4 Contaminants have been found in reagents, glassware, tubing, glass disposable pipettes, filters, degassers, and other apparatus
that release polyfluorinated compounds. All of these materials and supplies are routinely demonstrated to be free from interferences
by analyzing laboratory reagent blanks under the same conditions as the samples. If found, measures should be taken to remove
the contamination or data should be qualified; background subtraction of blank contamination is not allowed.
6.5 The liquid chromatography system used should consist, as much as practical, of sample solution or eluent contacting
components free of PFCPFAS target analytes of interest.
6.6 Polyethylene LC vial caps or any other target analyte-free vial caps should be used.
D7968 − 23
TABLE 4 Single-Laboratory Recovery Data in Ottawa Sand
Measured ng/kg from Ottawa Sand P&A Data (400 ng/kg spike for all PFCs except 2000 ng/kg for PFBA and PFPeA and 8000 ng/kg spike for FHEA,
Sample
FDEA, and FOEA)
Sample
PFTreA PFTriA PFDoA PFUnA PFDA PFNA PFOA PFHpA PFHxA PFPeA PFBA
PFPeA and 8000 ng/kg spike for FHEA,
FDEA, and FOEA)
Unspiked
Unspiked
P&A 1 389.6 394.3 384.7 376.7 362.1 347.6 345.8 232.9 222.2 1614.9 1344.5
P&A 2 462.1 424.6 397.2 379.1 378.4 376.9 365.9 247.9 229.8 1710.1 1388
P&A 3 402.7 387.7 383.1 365.9 374.7 363.3 347.1 242.4 222.9 1658.9 1376
P&A 4 403.9 397.1 395.4 381.5 379 359.4 342.7 246.8 225.8 1693.6 1401.9
P&A 5 467.2 445.8 412.6 388.5 376.8 370.3 369.7 249.3 231.4 1716.5 1433.4
P&A 6 392.1 385.3 374.2 370.9 353.2 351.7 340.3 236.7 220.5 1659 1366.4
Mean
Recovery 419.6 405.8 391.2 377.1 370.7 361.5 351.9 242.7 225.4 1675.5 1385
(ng/kg)
% Mean 104.9 101.4 97.8 94.3 92.7 90.4 88 60.7 56.4 83.8 69.3
Recovery
Standard 35.4 24.1 13.5 8 10.6 11.1 12.6 6.6 4.4 38.5 30.7
Deviation
RSD (%) 8.4 5.9 3.5 2.1 2.9 3.1 3.6 11 1.9 2.3 2.2
Sample PFBS PFHxS PFOS PFechS FOUEA FHpPA FHUEA FHEA FOEA FDEA
Unspiked
Unspiked
P&A 1 337.4 349.1 340.3 342.8 389.5 371.3 372.5 7023.5 8202.6 8564.9
P&A 2 347.3 358.3 345.9 347.2 408.7 377.2 387.1 7346.1 8542.6 9308
P&A 3 366.3 330.1 331.7 345.4 401.5 361.4 379 6844.3 7402.4 8989.2
P&A 4 348.2 343.6 338.3 347.6 404.9 377.5 388.1 7258.2 7551.9 9173.4
P&A 5 351.8 361.7 365.6 362.6 417.5 395.1 391.8 7461.3 7821.2 9287.4
P&A 6 336.7 343.4 363.7 342.5 394.5 356.9 374.5 7559.3 8002.2 8367.1
Mean
Recovery 347.9 347.7 347.7 348 402.7 373.2 382.1 7248.8 7920.5 8948.3
(ng/kg)
% Mean 87 86.9 86.9 87 100.7 93.3 95.5 90.6 99 111.9
Recovery
Standard 10.9 11.5 13.9 7.4 10 13.6 7.9 270.4 421.3 395.3
Deviation
RSD (%) 3.1 3.3 4 2.1 2.5 3.6 2.1 3.7 5.3 4.4
TABLE 5 Single-Laboratory Surrogate Recovery Data in Ottawa Sand
Measured ng/kg from Ottawa Sand – 400 ng/kg spike
Sample
MPFBA MPFHxA MPFHxS MPFOA MPFNA MPFOS MPFDA MPFUnA MPFDoA
Unspiked 1 420.0 433.5 431.8 428.0 439.4 429.2 442.6 443.3 447.7
Unspiked 2 366.5 396.8 378.5 384.9 389.8 373.6 404.9 400.8 425.8
P&A 1 361.1 364.3 356.3 377.0 376.6 354.4 384.9 391.3 409.3
P&A 2 383.6 378.4 357.3 389.4 379.7 375.7 395.7 399.2 412.2
P&A 3 374.5 378.5 375.4 390.5 378.6 372.4 382.5 386.9 402.2
P&A 4 370.1 384.4 366.1 396.3 384.4 374.2 397.8 406.2 420.5
P&A 5 370.1 386.8 372.0 395.7 381.1 372.8 394.4 399.9 421.5
P&A 6 363.6 384.8 356.1 397.9 384.9 368.6 389.5 392.3 402.9
Mean
Recovery
376.2 388.4 374.2 394.9 389.3 377.6 399.0 402.5 417.7
(ng/kg dry
weight)
% Mean 94.0 97.1 93.5 98.7 97.3 94.4 99.8 100.6 104.4
Recovery
Standard 19.0 20.4 24.9 15.0 20.7 21.9 19.0 17.6 14.9
Deviation
RSD (%) 5.1 5.3 6.7 3.8 5.3 5.8 4.8 4.4 3.6
6.7 Polyethylene disposable pipettes or target analyte-free pipettes should be used. All disposable pipettes should be checked for
release of target analytes of interest.
6.8 Degassers are important to continuous LC operation and most commonly are made of fluorinated polymers. To enable use, an
D7968 − 23
isolator column should be placed after the degasser and prior to the sample injection valve to separate the PFCsPFAS in the sample
from the PFCsPFAS in the LC system.
D7968 − 23
7. Apparatus
7.1 LC/MS/MS System:
7.1.1 Liquid Chromatography SystemSystem——A complete LC system is required in order to analyze samples; this should
include a sample injection system, a solvent pumping system capable of mixing solvents, a sample compartment capable of
maintaining required temperature, and a temperature-controlled column compartment. An LC system that is capable of performing
at the flows, pressures, controlled temperatures, sample volumes, and requirements of the standard shallmust be used.
7.1.2 Analytical Column —A reverse phase Charged Surface Hybrid Phenyl-Hexyl particle column was used to develop this test
method. Any column that achieves adequate resolution may be used. The retention times and order of elution may change
depending on the column used and need to be monitored.
7.1.3 Isolator ColumnColumn——A reverse phase C18 column was used in this test method to separate the target analytes in the
LC system and solvents from the target analytes in the analytical sample. This column was placed between the solvent mixing
chamber and the injector sample loop.
7.2 Tandem Mass Spectrometer SystemSystem——AAn MS/MS system capable of multiple reaction monitoring (MRM) analysis
or any system that is capable of meeting the requirements in this test method shallmust be used.
7.3 Centrifuge—A device to centrifuge the samples.
7.4 Lab Rotator —A device to mix the samples by end-over-end rotation.
7.5 Filtration Device:
7.5.1 Hypodermic Syringe—A luer-lock tip glass syringe capable of holding a syringe driven syringe-driven filter unit.
7.5.2 A 10-mL 10 mL lock tip glass syringe size is recommended since a 10-mL 10 mL sample size is used in this test method.
7.5.3 Filter Unit —Polypropylene filter units were used to filter the samples.
8. Reagents and Materials
8.1 Purity of Reagents—High performance liquid chromatography (HPLC) pesticide residue analysis and spectrophotometry
grade chemicals shallmust be used in all tests. Unless indicated otherwise, it is intended that all reagents shallmust conform to the
Committee on Analytical Reagents of the American Chemical Society. Other reagent grades may be used provided they are first
determined to be of sufficiently high purity to permit their use without affecting the accuracy of the measurements.
8.2 Purity of Water—Unless otherwise indicated, references to water shallmust be understood to mean reagent water conforming
to Type 1I of Specification D1193. It shallmust be demonstrated that this water does not contain contaminants at concentrations
sufficient to interfere with the analysis.
8.3 Gases—Ultrapure nitrogen and argon.
8.4 Vials—2-mL 2 mL amber glass or polypropylene autosampler vials or equivalent.
A Waters Acquity UPLC CSH Phenyl-Hexyl, 2.1 × 100 mm and 1.7-μm 1.7 μm particle size column, or equivalent, has been found suitable for use. It was used to develop
this test method and generate the precision and bias data presented in Section 16. If you are aware of an alternative column that meets the performance of the standard, please
provide this information to ASTM International Headquarters. Your comments will receive careful consideration at the meeting responsible technical committee, which you
may attend.
A Lab Rotator,lab rotator, or equivalent, has been found suitable to mix samples.
An Acrodisc GxF/0.2 μm GHP membrane syringe driven A 0.2 μm polypropylene membrane syringe-driven filter unit, or equivalent, has been found suitable for use.
Reagent Chemicals, American Chemical Society Specifications, American Chemical Society, Washington, D.C. For suggestions on the testing of reagents not listed by
the American Chemical Society, see AnnualAnalar Standards for Laboratory Chemicals, EDH Ltd., Poole, Dorset, U.K. and the United States Pharmacopeia and National
Formulary, U. S. U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD.
D7968 − 23
8.5 Polyethylene autosampler vial caps or equivalent.or any PFAS-free applicable autosampler vial caps.
8.6 Syringe—10 or 25 mL filter-adaptable glass syringe with luer lock.
8.7 pH paper (pH range 1-14).1 to 14).
8.8 Polypropylene Tubes—15- and 50-mL.15 and 50 mL.
8.9 Class A volumetric glassware.
8.10 Pipette Tips—Polypropylene pipette tips free of release agents or low retention coating of various sizes.
8.11 Polyethylene disposable pipettes.
8.12 Acetonitrile (CAS #No. 75-05-8).
8.13 Methanol (CAS #No. 67-56-1).
8.14 Ammonium acetate (CAS #No. 631-61-8).
8.15 Acetic acid (CAS # 64-19-7)No. 64-19-7).
8.16 2-Propanol (isopropyl alcohol, CAS #No. 67-63-0).
8.17 Ammonium hydroxide (CAS #No. 1336-21-6).
8.18 Ottawa sand (CAS #No. 14808-60-7).
8.19 PFCPFAS Standards:
8.19.1 Perfluorobutylsulfonate (PFBS, CAS #No. 29420-49-3).
8.19.2 Perfluorohexylsulfonate (PHFxS, CAS #(PFHxS, CAS No. 3871-99-6).
8.19.3 Perfluorooctylsulfonate (PFOS, CAS #No. 1763-23-1).
8.19.4 Perfluorobutanoate (PFBA, CAS #No. 375-22-4).
8.19.5 Perfluoropentanoate (PFPeA, CAS #No. 2706-90-3).
8.19.6 Perfluorohexanoate (PFHxA, CAS #No. 307-24-4).
8.19.7 Perfluoroheptanoate (PFHpA, CAS #No. 375-85-9).
8.19.8 Perfluorooctanoate (PFOA, CAS #No. 335-67-1).
8.19.9 Perfluorononanoate (PFNA, CAS #No. 375-95-1)
8.19.10 Perfluorodecanoate (PFDA, CAS #No. 335-76-2).
PFC StandardsPFAS standards may be difficult to find; some sources of PFCPFAS standards that have been found suitable for use were from Aldrich Chemical Company,
Accustandard, Wellington Laboratories, Inc., and Wako Laboratory. Standards from other vendors may be used.
D7968 − 23
8.19.11 Perfluoroundecanoate (PFUnA, CAS #No. 2058-94-8).
8.19.12 Perfluorododecanoate (PFDoA, CAS #No. 307-55-1).
8.19.13 Perfluorotridecanoate (PFTriA, CAS #No. 72629-94-8).
8.19.14 Perfluorotetradecanoate (PFTreA, CAS #No. 376-06-7).
8.19.15 Decafluoro-4-(pentafluoroethyl)cyclohexanesulfonate (PFechS, CAS #No. 67584-42-3).
8.19.16 3-perfluoropheptyl propanoic acid (FHpPA, CAS #No. 812-70-4).
8.19.17 2H-perfluoro-2-decenoic acid (FOUEA, CAS #No. 70887-84-2).
8.19.18 2-perfluorodecyl ethanoic acid (FDEA, CAS #number not available).
8.19.19 2-perfluorooctyl ethanoic acid (FOEA, CAS #No. 27854-31-5).
8.19.20 2H-perfluoro-2-octenoic acid (FHUEA, CAS #number not available).
8.19.21 2-perfluorohexyl ethanoic acid (FHEA, CAS #No. 53826-12-3).
8.20 PFCPFAS Surrogates:
8.20.1 O -Perfluorohexylsulfonate (MPFHxS).
8.20.2 C -Perfluorooctylsulfonate (MPFOS).
8.20.3 C -Perfluorobutanoate (MPFBA).
8.20.4 C -Perfluorohexanoate (MPFHxA).
8.20.5 C -Perfluorooctanoate (MPFOA).
8.20.6 C -Perfluorononanoate (MPFNA).
8.20.7 C -Perfluorodecanoate (MPFDA).
8.20.8 C -Perfluoroundecanoate (MPFUnA).
8.20.9 C -Perfluorododecanoate (MPFDoA).
9. Hazards
9.1 Normal laboratory safety applies to this method. Analysts should wear safety glasses, gloves, and lab coats when working in
the lab. Analysts should review the material safety data sheets (MSDS) Safety Data Sheets (SDS) for all reagents used in this
method.
10. Sampling
10.1 Sampling and Preservation—Grab samples are collected in glass or polypropylene containers. Sample containers and contact
surfaces with PTFE shallPFAS must be avoided. As part of the overall quality assurance program for this test method, field blanks
exposed to the same field conditions as samples are collected and analyzed according to this test method to assess the potential
for field contamination. This test method is based on a 2-g2 g sample size per analysis. If different sample sizes are used, spiking
solution amounts may need to be modified. Conventional sampling practices should be followed with the caution that
PFCPFAS surrogates from Wellington Laboratories Inc., or equivalent, have been found suitable for use.
D7968 − 23
PFC-containingPFAS-containing products may be present in sampling equipment. All sampling equipment and supplies shall be
PFC-free must be PFAS free in order to prevent contamination of the samples. EPA publicationspublication SW-846 may be used
as a sampling guide. Samples shallmust be shipped on ice with a trip blank. Once received, the sample temperature is taken and
should be less than must be between freezing and 6 °C. If the receiving temperature is greater than 6 °C, the sample temperature
is noted in the case narrative accompanying the data. Samples should be stored refrigerated between 0 and 6 °C from the time of
collection until analysis. The sample should be analyzed within 28 ddays of collection. No holding time study has been done on
the various soil matrices tested in this test method. Holding time may vary depending on the matrix, and individual laboratories
should determine the holding time in their matrix.
11. Preparation of LC/MS/MS
11.1 LC Chromatograph Operating Conditions:
11.1.1 Injections of all standards and samples are made at a 30-μL 30 μL volume. Other injection volumes may be used to optimize
conditions. Standards and sample extracts shallmust be in a 50:50 methanol:water solution containing 0.1 % acetic acid. In the case
of extreme concentration differences amongst samples, it is wise to analyze a blank after a concentrated sample and before a dilute
sample to minimize carryover of analytes from injection to injection. However, there should not be carryover between samples.
The LC utilized to develop this test method has a flow through flow-through LC needle design. The gradient conditions for liquid
chromatography are shown in Table 6.
11.2 LC Sample Manager Conditions:
11.2.1 Needle Wash Solvent—60 % acetonitrile/40 % 2-propanol; Time: 5 min.
11.3 Mass Spectrometer Parameters:
11.3.1 To acquire the maximum number of data points per SRM channel while maintaining adequate sensitivity, the tune
parameters may be optimized according to the instrument used. Each peak requires at least ten scans per peak for adequate
quantitation. This test method contains nine surrogates, which are isotopically labeled PFCs,PFAS, and 21 PFCsPFAS which are
split up into 18 MRM acquisition functions to optimize sensitivity. Variable parameters regarding retention times, SRM transitions,
and cone and collision energies are shown in Table 3. Mass spectrometer parameters used in the development of this method are
listed below:
11.3.1.1 The instrument is set in the Electrospray negative source setting.
11.3.1.2 Capillary Voltage: 0.75 kV.
11.3.1.3 Cone: Variable depending on analyte.
11.3.1.4 Extractor: 2 Volts.
TABLE 6 Gradient Conditions for Liquid Chromatography
Percent
95 % Water:
Percent
5 %
Time Flow 95 % Water: Percent
Acetonitrile,
(min) (mL/min) 5 % Acetonitrile
400 mM
Acetonitrile
Ammonium
Acetate
0 0.3 95 0 5
1 0.3 75 20 5
6 0.3 50 45 5
13 0.3 15 80 5
14 0.4 0 95 5
17 0.4 0 95 5
18 0.4 95 0 5
21 0.4 95 0 5
A guide to help and determine sample holding times can be found at http://www.epa.gov/esd/cmb/research/bs_033cmb06.pdf (2014).
D7968 − 23
11.3.1.5 Source Temperature: 150 °C.
11.3.1.6 Desolvation Gas Temperature: 450 °C.
11.3.1.7 Desolvation Gas Flow: 800 L/h.
11.3.1.8 Cone Gas Flow: 200 L/h.
11.3.1.9 Collision Gas Flow: 0.15 mL/min.
11.3.1.10 Low Mass Resolution 1: 2.6.
11.3.1.11 High Mass Resolution 1: 14.
11.3.1.12 Ion Energy 1: 1.
11.3.1.13 Entrance Energy: 1.
11.3.1.14 Collision Energy: Variable depending on analyte.
11.3.1.15 Exit Energy: 1.
11.3.1.16 Low Mass Resolution 2: 2.5.
11.3.1.17 High Mass Resolution 2: 14.
11.3.1.18 Ion Energy 2: 3.
11.3.1.19 Gain: 1.0.
11.3.1.20 Multiplier: 511.1.
11.3.1.21 Inter-Scan Delay: 0.004 s.
12. Calibration and Standardization
12.1 The mass spectrometer shallmust be calibrated as per manufacturer’s specifications before analysis. Analytical values
satisfying test method criteria have been achieved using the following procedures. Prepare all solutions in the lab using Class A
volumetric glassware.
12.2 Calibration and Standardization—To calibrate the instrument, analyze nine calibration standards of the polyfluorinated
compounds prior to sample analysis as shown in Table 2. Calibration stock standard solution is prepared from the target and
surrogate spike solutions directly to ensure consistency. Stock standard Solution A containing the polyfluorinated compounds and
surrogates is prepared at Level 9 concentration and aliquots of that solution are diluted to prepare Levels 1 through 8. The
following steps will produce standards with the concentration values shown in Table 2. The analyst is responsible for recording
initial component weights carefully when working with pure materials and correctly carrying the weights through the dilution
calculations. At a minimum five calibration levels are required when using a linear calibration curve and six calibration levels are
required when using a quadratic calibration curve. An initial nine-point curve may be used to allow for the dropping of the lower
level calibration points if the individual laboratory’s instrument cannot achieve low detection limits on certain PFCs.PFAS. This
should allow for at least a five-five or six-point calibration curve to be obtained. No problems were encountered while using the
nine-point calibration curve in developing this test method.
12.2.1 Calibration stock standard Solution A (Level 9, Table 2) is prepared from the target and surrogate spike solutions directly
to ensure consistency. 500 μL of surrogate spike (20 μg ⁄L), 500 μL Target Spike I, and 500 μL of PFCPFAS Target Spike II (refer
to Table 7) is added to a 50-mL 50 mL volumetric flask and diluted to 50 mL with 50:50 methanol:water containing 0.1 % acetic
acid. The preparation of the Level 9 standard can be accomplished using appropriate volumes and concentrations of stock solutions
as per a particular laboratory’s standard procedure. It is critical to ensure that analytes are solubilized in the Level 9 standard.
D7968 − 23
TABLE 7 PFCPFAS Target Spike Solutions (PPB)
Concentration of Analyte in PFC Target Spike Solutions
Analyte
Analyte
PFC High Target Spike Solutions
PFC Reporting Limit Spike Solution
Concentration of Analyte in PFAS Target Spike Solutions
PFAS High Target Spike Solutions
PFAS Reporting
Target Spike I Target Spike II
PFTreA, PFTriA, PFDoA, PFUnA,
PFDA, PFOS, PFNA, PFHxA, PFHpA, 20 μg/L — 2 μg/L
PFBS, PFechS, PFOA, PFHxS
PFBA, PFPeA 100 μg/L — 10 μg/L
FOUEA, FHUEA, FHpPA — 20 μg/L 2 μg/L
FHEA, FOEA, FDEA — 400 μg/L 40 μg/L
12.2.2 Aliquots of Solution A are then diluted with 50:50 methanol:water containing 0.1 % acetic acid to prepare the desired
calibration levels in 2-mL amber glass 2 mL LC vials. The calibration vials shallmust be used within 24 h to ensure optimum
results. The end calibration check shallmust be prepared in a separate LC vial near the mid-level. All calibration standards should
only be used once. The analyte concentration in the vial may change after the vial cap is pierced because the vial caps domay not
reseal after puncture. puncture; if the cap reseals it may be used over again. Changing the caps immediately after the injection
should alleviate this problem. Calibration standards are not filtered.
12.2.3 Inject each standard and obtain its chromatogram. An external calibration technique is used to monitor the primary and
confirmatory SRM transitions of each analyte. Calibration software is utilized to conduct the quantitation of the target analytes and
surrogates using the primary SRM transition. The ratios of the primary/confirmatory SRM transition area counts are given in Table
3 and will vary depending on the individual tuning conditions. The primary/confirmatory SRM transition area ratio shallmust be
within 35 % of the individual labs’lab’s accepted primary/confirmatory SRM transition area ratio. The primary SRM transition of
each analyte is used for quantitation of and the confirmatory SRM transition for confirmation. This gives added confirmation by
isolating the parent ion, forming two product ions via fragmentation, and relating it to the retention time in the calibration standard.
12.2.4 Depending on sensitivity and matrix interference issues dependent on sample type, the confirmatory SRM transition can
be used as the primary SRM transition for quantitation during analysis. This shallmust be explained in a narrative accompanying
the data. A new primary/confirmatory ion ratio will then be determined if switching the SRM transitions used to quantitate and
confirm. The new primary/confirmatory SRM transition area ratio is required to be within 35 % of the individual labs’lab’s new
primary/confirmatory SRM transition area ratio.
12.2.5 The calibration software manual should be consulted to use the software correctly. The quantitation method is set as an
external calibration using the peak areas in ppt units. Concentrations may be calculated using the data system software to generate
linear regression or quadratic calibration curves. Forcing the calibration curve through the origin (X = 0, Y = 0) is not
recommended. Curves should be evaluated using relative error or relative standard error.
12.2.6 Linear calibration may be used if the coefficient of determination, r , is ≥0.98 for the analyte. The point of origin is excluded
and a fit weighting 1/X is used in order to give more emphasis to the lower concentrations. If one of the calibration standards other
than the high or low point causes the r of the curve to be <0.98, this point shall be re-injected or a new calibration curve shall
be regenerated. Each calibration point used to generate the curve shallmust have a calculated percent deviation less than 30 %25 %
from the generated curve. If the low or high point(s), or both, are excluded, minimally a five-point curve is acceptable but the
reporting range shall be modified to reflect this change.
12.2.7 Quadratic calibration may be used if the coefficient of determination, r , is ≥0.99 for the analyte. The point of origin is
excluded, and a fit weighting of 1/X is used in order to give more emphasis to the lower concentrations. If one of the calibration
standards causes the curve to be <0.99, this point shall be re-injected or a new calibration curve shall be regenerated. If the low
or high point(s), or both, are excluded, minimally a six-point curve is acceptable but the reporting range shall be modified to reflect
this change. Each calibration point used to generate the curve shallmust have a calculated percent deviation less than 30 %25 %
from the generated curve.
Management and Technical Requirements for Laboratories Performing Environmental Analysis, Module 4: Quality Systems for Chemical Testing; The NELAC Institute,
2017.
D7968 − 23
12.2.8 The retention time window of the SRM transitions shallmust be within 5 % of the retention time of the analyte in a midpoint
calibration standard. If this is not the case, re-analyze the calibration curve to determine if there was a shift in retention time during
the analysis and the sample needs to be re-injected. If the retention time is still incorrect in the sample, refer to the analyte as an
unknown.
12.2.9 A midpoint calibration check standard shallmust be analyzed at the end of each batch of 30 samples or within 24 h after
the initial calibration curve was generated; the criteria in the individual labs’lab’s quality system may be more restrictive pertaining
to the number of samples. This end calibration check, a new not pierced sealed vial, should come from the same calibration
standard solution that was used to generate the initial curve. The results from the end calibration check standard shallmust have
a percent deviation less than 30 % from the calculated concentration for the target analytes and surrogates. If the results are not
within these criteria, corrective action including reoccurrence minimization is performed and either all samples in the batch are
re-analyzed against a new calibration curve or the affected results are qualified with an indication that they do not fall within the
performance criteria of the test method. If the analyst inspects the vial containing the end calibration check standard and notices
that the sample evaporated affecting the concentration or other anomaly, a new end calibration check standard may be made and
analyzed. If this new end calibration check standard has a percent deviation less than 30 % from the calculated concentration for
the target analytes and surrogates, the results may be reported unqualified.
12.3 If a laboratory has not performed the test before or if there has been a major change in the measurement system, for example,
new analyst, new instrument, etc., an instrument qualification study including method detection limit (MDL), calibration range
determination, and precision and bias determination shallmust be performed to demonstrate laboratory capability.
12.3.1 Analyze at least four replicates of a spiked sand sample containing the PFCsPFAS and surrogates at an extract concentration
in the calibration range of Levels 4 through 7. The Level 6 concentration of the nine-point calibration curve was used to set the
QC acceptance criteria in this method. The matrix and chemistry should be similar to the matrix used in this test method. Each
replicate shallmust be taken through the complete analytical test method including any sample manipulation and extraction steps.
12.3.2 Calculate the mean (average) percent recovery and relative standard deviation (RSD) of the four values and compare to the
acceptable ranges of the QC acceptance criteria for the Initial Demonstration of Performance in Table 8.
12.3.3 This study should be repeated until the single operator single-operator precision and mean recovery are within the limits
in Table 8. If a concentration other than the recommended concentration is used, refer to Practice D5847 for information on
applying the F test and t test in evaluating the acceptability of the mean and standard deviation.
12.3.3.1 The QC acceptance criteria for the Initial Demonstration of Performance in Table 8 were generated from the
single-laboratory data shown in Section 16. Data from Ottawa sand and four ASTM soil matrices are shown in Section 16. It is
recommended that each laboratory determine in-house QC acceptance criteria which meet or exceed the criteria in this test method.
References generating QC acceptance criteria are ASTM Practices D2777, D5847, E2554, or Method 8000 in EPA publication-
spublication SW-846.
12.4 Surrogate Spiking Solution:
12.4.1 A surrogate spiking solution containing nine isotopically labeled PFCs—MPFBA,PFAS—MPFBA, MPFHxA, MPFHxS,
MPFDA, MPFOA, MPFOS, MPFNA, MPFUnA, and MPFDoA—is added to all samples, method blanks, duplicates, laboratory
control samples, matrix spikes, and reporting limit checks. A stock surrogate spiking solution is prepared at 20 μg/L in 95 %
acetonitrile: 5 % water. Spiking 40 μL of this spiking solution into a 2-g2 g soil sample results in a concentration of 400 ng/kg of
the surrogate in the sample. The results obtained for the surrogate recoveries shallmust fall within the limits of Table 8. If the limits
are not met, the affected results shallmust be qualified with an indication that they do not fall within the performance criteria of
the test method.
12.4.1.1 The surrogate spiking solution was prepared by adding 500 μL of a 2-mg/L PFC 2 mg/L PFAS surrogate mix in a 50
mL volumetric and diluted to 50 mL with 95 % acetonitrile: 5 % water. Surrogate spiking solutions are routinely replaced every
year if not previously discarded for quality control failure.
12.5 Method Blank:
Surrogate mix from Wellington Laboratories, Inc. has been found suitable for use.
D7968 − 23
TABLE 8 QC Acceptance Criteria
NOTE 1—Table 8 data is preliminary until a multi-lab validation study is completed.
Initial Demonstration of Performance Laboratory Control Sample
Recovery (%) Precision Recovery (%)
Analyte/Surrogate Spike Conc. ng/kg
Maximum Lower Control Limit Upper Control Limit
Lower Limit Upper Limit
% RSD (LCL) % (UCL) %
PFTreA 400 70 130 30 70 130
PFTriA 400 70 130 30 70 130
PFDoA 400 70 130 30 70 130
PFUnA 400 70 130 30 70 130
PFDA 400 70 130 30 70 130
PFOS 400 70 130 30 70 130
PFNA 400 70 130 30 70 130
PFecHS 400 70 130 30 70 130
PFOA 400 70 130 30 70 130
PFHxS 400 70 130 30 70 130
PFHpA 400 50 130 30 50 130
PFHxA 400 50 130 30 50 130
PFBS 400 70 130 30 70 130
PFPeA 2000 70 130 30 70 130
PFBA 2000 50 130 30 50 130
FHEA 8000 70 130 30 70 130
FOEA 8000 70 130 30 70 130
FDEA 8000 70 130 30 70 130
FOUEA 400 70 130 30 70 130
FHpPA 400 70 130 30 70 130
FHUEA 400 70 130 30 70 130
MPFBA 400 70 130 30 70 130
MPFHxA 400 70 130 30 70 130
MPFHxS 400 70 130 30 70 130
MPFOA 400 70
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...