Standard Specification for Laboratory Weights and Precision Mass Standards

ABSTRACT
This specification covers laboratory weights and precision mass standards including their principal physical characteristics and metrological requirements. Maximum permissible error, magnetic property, density, and surface roughness for each weight shall be within the limits indicated in this specification. Physical characteristics shall be based on construction, design, surface area, materials, magnetism, density, surface finish, weight adjustment, and marking.
SCOPE
1.1 This specification covers weights and mass standards used in laboratories, specifically classes 000, 00, 0, 1, 2, 3, 4, 5, 6, and 7. This specification replaces National Bureau of Standards Circular 547, Section 1, which is out of print.  
1.2 This specification and calibration method is intended for use by weight manufacturers, national metrology institutes, weight calibration laboratories, accreditation bodies, users of weights, and regulatory bodies.  
1.3 This specification contains the principal physical characteristics and metrological requirements for weights that are used.  
1.3.1 For the verification of weighing instruments;  
1.3.2 For the calibration of weights of a lower class of accuracy; and  
1.3.3 With weighing instruments.  
1.4 Maximum Permissible Errors (formerly tolerances) and design restrictions for each class are described in order that both individual weights or sets of weights can be chosen for appropriate applications.  
1.5 Weight manufacturers must be able to provide evidence that all new weights comply with specifications in this standard (for example, material, density, magnetism, surface finish, mass values, uncertainties) to make any claim of compliance to Specification E617, Maximum Permissible Errors, weight classes, or metrological traceability.  
1.5.1 During subsequent calibrations, calibration laboratories must meet the requirements of ISO/IEC 17025:2017.  
1.5.2 Subsequent calibrations must meet all the requirements, including Sections 7, 8, and 9, Table 8 and Table 11 (environmental parameters) to make any claim of compliance to Specification E617, Maximum Permissible Errors, weight classes, or metrological traceability.
Note 1: Requirements set forth in NIST IR 6969 and NIST IR 5672 are compliant with all the requirements of Specification E617, Sections 7, 8, and 9.  
1.6 The values stated in SI units are to be regarded as standard.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
14-Aug-2023
Technical Committee
E41 - Laboratory Apparatus

Relations

Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023
Effective Date
15-Aug-2023

Overview

ASTM E617-23 is the standard specification issued by ASTM International for laboratory weights and precision mass standards. This document defines the essential physical characteristics and metrological requirements for weights used in laboratory settings. It provides a framework for maximum permissible error, as well as requirements surrounding magnetic properties, density, and surface roughness for various classes of weights. The specification is recognized internationally and serves as a reference for weight manufacturers, calibration laboratories, national metrology institutes, accreditation bodies, and users of laboratory weights who require traceability and accuracy in their measurements.

Key Topics

  • Scope and Classes: The standard covers weights and mass standards across classes 000, 00, 0, 1, 2, 3, 4, 5, 6, and 7. This broad classification ensures applicability for a wide range of laboratory and industrial applications, from high-precision scientific research to routine weighing.

  • Metrological Requirements: Sets out maximum permissible errors for each class and denomination of weight, ensuring that users can select the appropriate weights based on their accuracy needs.

  • Physical Properties: Outlines the material and construction requirements, including surface area, design, materials, density, magnetism, and surface finish. Weights must be constructed to withstand wear and corrosion, typically requiring austenitic stainless steel or similarly resistant materials.

  • Calibration and Traceability: Details the requirements for initial and subsequent calibrations, ensuring metrological traceability to national and international mass standards. Laboratories performing calibrations must adhere to ISO/IEC 17025 requirements.

  • Magnetic and Density Specifications: Specifies upper limits for magnetic properties and defines acceptable density ranges. These limits are designed to minimize the impact of environmental conditions and material variability on mass measurements.

  • Documentation and Compliance: Manufacturers and calibration laboratories must maintain evidence of compliance, including material certificates and calibration reports, to substantiate claims of adherence to ASTM E617.

Applications

ASTM E617-23 is crucial wherever laboratory weights and precision mass standards are used:

  • Calibration of Weighing Instruments: Ensures that devices such as analytical balances and industrial scales are verified using weights of known accuracy classes.

  • Metrology and Testing Laboratories: Provides the basis for producing, certifying, and periodically recalibrating reference weights, supporting traceable measurements in scientific research, pharmaceutical production, and quality control.

  • Accreditation and Inspection: Used by regulatory bodies to ensure that laboratories comply with international standards in measurement and calibration processes.

  • Manufacture and Adjustment of Weights: Guides weight manufacturers in producing weights that meet strict criteria for accuracy, material properties, and adjustability, supporting claims of compliance and traceability.

  • Routine Laboratory Operations: Laboratory personnel use ASTM E617-compliant weights for daily analytical and preparative work to assure accuracy and confidence in results.

Related Standards

Organizations that utilize ASTM E617-23 should also consider the following related documents for comprehensive compliance and best practices:

  • ISO/IEC 17025:2017 - General requirements for the competence of testing and calibration laboratories.
  • NIST Handbook 143 - State Weights and Measures Laboratories Program Handbook.
  • NIST SP 811 - Guide for the Use of the International System of Units (SI).
  • NISTIR 5672 & NISTIR 6969 - Advanced Mass Calibration and Measurement Assurance guidance.
  • OIML R111 - Technical requirements for weight classes used in legal metrology.
  • BIPM SI Brochure - The International System of Units, for understanding unit definitions and international traceability.

Conclusion

ASTM E617-23 provides a rigorous and internationally recognized framework for the specification, use, and calibration of laboratory weights. By adhering to this standard, laboratories and manufacturers ensure the accuracy, reliability, and metrological traceability of their mass measurements, supporting global consistency and quality in testing and research. This makes ASTM E617 essential for any operation reliant on precision weighing and calibration.

Buy Documents

Technical specification

ASTM E617-23 - Standard Specification for Laboratory Weights and Precision Mass Standards

English language (16 pages)
sale 15% off
sale 15% off
Technical specification

REDLINE ASTM E617-23 - Standard Specification for Laboratory Weights and Precision Mass Standards

English language (16 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

ECOCERT

Organic and sustainability certification.

COFRAC France Verified

Eurofins Food Testing Global

Global leader in food, environment, and pharmaceutical product testing.

COFRAC Luxembourg Verified

Intertek Bangladesh

Intertek certification and testing services in Bangladesh.

BAB Bangladesh Verified

Sponsored listings

Frequently Asked Questions

ASTM E617-23 is a technical specification published by ASTM International. Its full title is "Standard Specification for Laboratory Weights and Precision Mass Standards". This standard covers: ABSTRACT This specification covers laboratory weights and precision mass standards including their principal physical characteristics and metrological requirements. Maximum permissible error, magnetic property, density, and surface roughness for each weight shall be within the limits indicated in this specification. Physical characteristics shall be based on construction, design, surface area, materials, magnetism, density, surface finish, weight adjustment, and marking. SCOPE 1.1 This specification covers weights and mass standards used in laboratories, specifically classes 000, 00, 0, 1, 2, 3, 4, 5, 6, and 7. This specification replaces National Bureau of Standards Circular 547, Section 1, which is out of print. 1.2 This specification and calibration method is intended for use by weight manufacturers, national metrology institutes, weight calibration laboratories, accreditation bodies, users of weights, and regulatory bodies. 1.3 This specification contains the principal physical characteristics and metrological requirements for weights that are used. 1.3.1 For the verification of weighing instruments; 1.3.2 For the calibration of weights of a lower class of accuracy; and 1.3.3 With weighing instruments. 1.4 Maximum Permissible Errors (formerly tolerances) and design restrictions for each class are described in order that both individual weights or sets of weights can be chosen for appropriate applications. 1.5 Weight manufacturers must be able to provide evidence that all new weights comply with specifications in this standard (for example, material, density, magnetism, surface finish, mass values, uncertainties) to make any claim of compliance to Specification E617, Maximum Permissible Errors, weight classes, or metrological traceability. 1.5.1 During subsequent calibrations, calibration laboratories must meet the requirements of ISO/IEC 17025:2017. 1.5.2 Subsequent calibrations must meet all the requirements, including Sections 7, 8, and 9, Table 8 and Table 11 (environmental parameters) to make any claim of compliance to Specification E617, Maximum Permissible Errors, weight classes, or metrological traceability. Note 1: Requirements set forth in NIST IR 6969 and NIST IR 5672 are compliant with all the requirements of Specification E617, Sections 7, 8, and 9. 1.6 The values stated in SI units are to be regarded as standard. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ABSTRACT This specification covers laboratory weights and precision mass standards including their principal physical characteristics and metrological requirements. Maximum permissible error, magnetic property, density, and surface roughness for each weight shall be within the limits indicated in this specification. Physical characteristics shall be based on construction, design, surface area, materials, magnetism, density, surface finish, weight adjustment, and marking. SCOPE 1.1 This specification covers weights and mass standards used in laboratories, specifically classes 000, 00, 0, 1, 2, 3, 4, 5, 6, and 7. This specification replaces National Bureau of Standards Circular 547, Section 1, which is out of print. 1.2 This specification and calibration method is intended for use by weight manufacturers, national metrology institutes, weight calibration laboratories, accreditation bodies, users of weights, and regulatory bodies. 1.3 This specification contains the principal physical characteristics and metrological requirements for weights that are used. 1.3.1 For the verification of weighing instruments; 1.3.2 For the calibration of weights of a lower class of accuracy; and 1.3.3 With weighing instruments. 1.4 Maximum Permissible Errors (formerly tolerances) and design restrictions for each class are described in order that both individual weights or sets of weights can be chosen for appropriate applications. 1.5 Weight manufacturers must be able to provide evidence that all new weights comply with specifications in this standard (for example, material, density, magnetism, surface finish, mass values, uncertainties) to make any claim of compliance to Specification E617, Maximum Permissible Errors, weight classes, or metrological traceability. 1.5.1 During subsequent calibrations, calibration laboratories must meet the requirements of ISO/IEC 17025:2017. 1.5.2 Subsequent calibrations must meet all the requirements, including Sections 7, 8, and 9, Table 8 and Table 11 (environmental parameters) to make any claim of compliance to Specification E617, Maximum Permissible Errors, weight classes, or metrological traceability. Note 1: Requirements set forth in NIST IR 6969 and NIST IR 5672 are compliant with all the requirements of Specification E617, Sections 7, 8, and 9. 1.6 The values stated in SI units are to be regarded as standard. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM E617-23 is classified under the following ICS (International Classification for Standards) categories: 71.040.20 - Laboratory ware and related apparatus. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM E617-23 has the following relationships with other standards: It is inter standard links to ASTM E617-18, ASTM D4753-15, ASTM E2206-21, ASTM E1154-23, ASTM D7416-09(2020), ASTM E2040-19, ASTM D7551-10(2015), ASTM C1009-21, ASTM E2510-20, ASTM C114-23, ASTM E542-22, ASTM D7651-17, ASTM D8293-22, ASTM E458-08(2020), ASTM E898-20. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM E617-23 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E617 − 23
Standard Specification for
Laboratory Weights and Precision Mass Standards
This standard is issued under the fixed designation E617; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.6 The values stated in SI units are to be regarded as
standard.
1.1 This specification covers weights and mass standards
used in laboratories, specifically classes 000, 00, 0, 1, 2, 3, 4,
1.7 This international standard was developed in accor-
5, 6, and 7. This specification replaces National Bureau of
dance with internationally recognized principles on standard-
Standards Circular 547, Section 1, which is out of print.
ization established in the Decision on Principles for the
1.2 This specification and calibration method is intended for Development of International Standards, Guides and Recom-
use by weight manufacturers, national metrology institutes, mendations issued by the World Trade Organization Technical
weight calibration laboratories, accreditation bodies, users of
Barriers to Trade (TBT) Committee.
weights, and regulatory bodies.
2. Referenced Documents
1.3 This specification contains the principal physical char-
acteristics and metrological requirements for weights that are
2.1 ISO Standards:
used.
ISO/IEC 17025:2017 General Requirements for the Compe-
1.3.1 For the verification of weighing instruments;
tence of Testing and Calibration Laboratories
1.3.2 For the calibration of weights of a lower class of
2.2 NIST Standards:
accuracy; and
NIST Handbook 143 State Weights and Measures Laborato-
1.3.3 With weighing instruments.
ries Program Handbook (2019)
1.4 Maximum Permissible Errors (formerly tolerances) and
NIST SP 811 Guide for the Use of the International System
design restrictions for each class are described in order that
of Unit (SI) 2008 Edition
both individual weights or sets of weights can be chosen for
NIST SP 1038 The International System of Units (SI) –
appropriate applications.
Conversion Factors for General Use (May 2006)
1.5 Weight manufacturers must be able to provide evidence
NISTIR 5672 Advanced Mass Calibration and Measurement
that all new weights comply with specifications in this standard
Assurance Program for State Calibration Laboratories
(for example, material, density, magnetism, surface finish,
(2019)
mass values, uncertainties) to make any claim of compliance to
NISTIR 6969 Selected Laboratory and Measurement Prac-
Specification E617, Maximum Permissible Errors, weight
tices to Support Basic Mass Calibrations (2019)
classes, or metrological traceability.
NIST Technical Note 1297 Guidelines for Evaluating and
1.5.1 During subsequent calibrations, calibration laborato-
Expressing the Uncertainty of NIST Measurement Results
ries must meet the requirements of ISO/IEC 17025:2017.
(1994)
1.5.2 Subsequent calibrations must meet all the
requirements, including Sections 7, 8, and 9, Table 8 and Table
2.3 OIML Standards:
11 (environmental parameters) to make any claim of compli-
OIML D 28 Conventional Value of the Result of Weighing in
ance to Specification E617, Maximum Permissible Errors,
Air (2004)
weight classes, or metrological traceability.
OIML R 111–1 Weights of classes E1, E2, F1, F2, M1,
M1–2, M2, M2–3 and M3 Part 1: Metrological and
NOTE 1—Requirements set forth in NIST IR 6969 and NIST IR 5672
are compliant with all the requirements of Specification E617, Sections 7, Technical Requirements (Edition 2004)
8, and 9.
1 2
This specification is under the jurisdiction of ASTM Committee E41 on Available from International Organization for Standardization (ISO), 1, ch. de
Laboratory Apparatusand is the direct responsibility of Subcommittee E41.06 on la Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http://www.iso.org.
Laboratory Instruments and Equipment. Available from National Institute of Standards and Technology (NIST), 100
Current edition approved Aug. 15, 2023. Published November 2023. Originally Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http://www.nist.gov.
approved in 1978. Last previous edition approved in 2018 as E617 – 18. DOI: Available from Organisation Internationale de Metrologie Legale, 11 Rue
10.1520/E0617-23. Turgot, 75009 Paris, France, http://www.oiml.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E617 − 23
2.4 BIPM Standards: dation OIML D 28. For a weight taken at 20 °C, the conven-
VIM: JCGM 200:2012 International Vocabulary of Metrolo- tional mass is the mass of a reference weight of a density of
–3 –3
gy–Basic and General Concepts and Associated Terms
8000 kg m which it balances in air of density of 1.2 kg m .
GUM: JCGM 100:2008 Evaluation of Measurement
3.1.6 correction—mass values are traditionally expressed by
Data–Guide to the Expression of Uncertainty in Measure-
two numbers, one being the nominal mass of the weight, and
ment
the second being a correction. The mass of the weight is the
SI Brochure: 2019 The International System of Units (SI),
th assigned nominal value plus the assigned correction. Positive
9 Edition
6 corrections indicate that the weight embodies more mass than
2.5 EURAMET Standards:
is indicated by the assigned nominal value. Negative correc-
EURAMET/cg-18/V. 4.0 Guidelines on the Calibration of
tions indicate that the weight embodies less mass than is
Non-Automatic Weighing Instruments (2015)
indicated by the assigned nominal value. The correction is
equivalent to the “error.”
3. Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.7 kilogram—the SI base unit of mass. Formerly the
3.1.1 accuracy class of weights—a class of weights that international prototype kilogram (IPK), the platinum-iridium
meets certain metrological requirements intended to keep the
cylinder maintained at the International Bureau of Weights and
errors within specified limits.
Measures (BIPM) in Sèvres, France, was internationally ac-
cepted as having a defined mass of 1 kg. In 2018 the kilogram
3.1.2 balance—instrument indicating apparent mass that is
was redefined in terms of the Planck constant h (taken to be
sensitive to the following forces:
–34 2 –1
6.626 070 15 × 10 kg m s ) where the mass of primary
Force due to gravity
F 5 m·g
g
mass standards may be determined by any primary method
such as those described in the mise en pratique for the
Air buoyancy equal to the weight of
m
the displaced air. definition of the kilogram in the BIPM SI Brochure. The
F 5 v·ρ ·g 5 ρ ·g
b a a
ρ
redefinition process ensured continuity of mass measurements
before and after redefinition such that all mass values traceable
Vertical component of the magnetic
≠ H
interaction between the weight and
F 5 μ sM 1 χ Hd dV
eee to the IPK remained the same when the new definition came
z o
≠ z
v the balance or the environment, or
into effect, but the uncertainties of these mass values were
both.
increased by the relative standard uncertainty of the IPK
H and M are vectors; z is the vertical cartesian coordinate.
–8
immediately after the redefinition (1.0 × 10 ).
If magnetic effects are negligible, that is, the permanent
magnetization (M) of the weight and the magnetic suscepti-
3.1.8 magnetism—effect that generates an attractive or re-
bility (χ) are sufficiently small, and the balance is calibrated
pulsive force without the presence of charged species.
with reference weights of well-known mass, the balance can
3.1.8.1 (volume) magnetic susceptibility (χ)—measure of the
be used to indicate the conventional mass, m , of a body
c
ability of a medium to modify a magnetic field. It is related to
under conventionally chosen conditions.
the magnetic permeability (μ) by the relation: μ/μ = 1 + χ. The
3.1.3 calibration (of weights)—the acts of determining the
quantity μ/μ is sometimes referred to as the relative
mass difference between a standard of known mass value and
permeability, μ .
r
an “unknown” test weight or set of weights, establishing the
3.1.8.2 (permanent) magnetization (M)—parameter that
mass value and conventional mass value of the “unknown,”
specifies a magnetic state of material bodies such as weights, in
and of determining a quantitative estimate of the uncertainty to
the absence of an external magnetic field (most generally,
be assigned to the stated mass or conventional mass value of
the “unknown,” or both, and providing metrological traceabil- magnetization is a vector whose magnitude and direction are
ity to the “unknown.” not necessarily constant within the material). The magnetiza-
tion of a body generates an inhomogeneous magnetic field in
3.1.3.1 calibration (generally)—set of operations that
space and thus may produce magnetic forces on other materi-
establish, under specified conditions, the relationship between
als.
values of quantities indicated by a measuring instrument or
measuring system, or values represented by a material measure
3.1.9 mass—physical quantity, which can be ascribed to any
or a reference material, and the corresponding values realized
material object and which gives a measure of its quantity of
by standards.
matter. The unit of mass is the kilogram.
3.1.4 calibration certificate—certificate issued by calibra-
3.1.10 maximum permissible errors—the maximum amount
tion laboratories to document the results of a calibration.
by which the sum of the conventional mass correction of the
3.1.5 conventional mass—conventional value of the result
weight, and its associated uncertainty is allowed to deviate
of weighing in air, in accordance to International Recommen-
from the assigned nominal value.
3.1.11 metrological traceability—property of a measure-
Available from Bureau International des Poids et Mesures (BIPM), Pavilion de
ment result whereby the result can be related to a reference
Breteuil, F-92312, Sèvres Cedex, France, http://www.bipm.org.
6 through a documented unbroken chain of calibrations, each
Available from Euramet, Bundesallee 100, 38116 Braunschweig, Germany,
http://www.euramet.org. contributing to the measurement uncertainty. Metrological
E617 − 23
traceability requires an established calibration hierarchy. Ele-
Symbol Unit Definition
d m estimated distance between centers of
ments for confirming metrological traceability to be an unbro-
weights during loading
ken chain to a primary mass standard shall include a docu-
d m estimated distance from the center of
mented measurement uncertainty, a documented measurement the load receptor to one of the corners
F N air buoyancy equal to the weight of the
b
procedure, accredited technical competence, metrological
displaced air
traceability to the SI, and established calibration intervals (see
F N gravitational force
g
current VIM: JCGM 200).
F N magnetic force between a mass
z
comparator and a weight in the vertical
3.1.12 reference standard—a standard, generally of the
or z-direction
–2
highest metrological quality available at a given location, from g m s gravitational acceleration
–1
H A m magnetizing field strength
which measurements made at that location are derived.
h % relative humidity
r
3.1.13 roughness parameter or R-parameter (R or R )— I kg indication of the weighing instruments
a z
(scale division)
parameter that describes the assessed roughness profile of a
ΔI kg indication difference of the balance,
sample. The letter R is indicative of the type of assessed
where ΔI = I – I
t r
ΔI kg indication difference using an
profile, in this case R for roughness profile. The assessed 1
automatic exchange mechanism with
profile of a sample can be in terms of different profile types: a
weights in first position
roughness profile or R-parameter, primary profile or
ΔI kg indication difference using an
automatic exchange mechanism with
P-parameter, a waviness profile or W-parameter.
weights in reversed position
3.1.14 set of weights—a series of weights, usually presented
ΔI kg change in indication of balance due to
s
sensitivity weight
in a case so arranged to make possible any weighing of all
i – subscript used as an index in
loads between the mass of the weight with the smallest nominal
summations
value and the sum of the masses of all weights of the series
j – subscript for number of test weights or
number of series of measurements
with a progression in which the mass of the smallest nominal
k – coverage factor, typically 2 or 3
value weight constitutes the smallest step of the series.
–1
M A m permanent magnetization (see also
μ M)
3.1.15 temperature (t)—in degrees Celsius, is related to the
m kg mass of a rigid body (weight)
absolute thermodynamic temperature scale, called the Kelvin
Δm kg mass difference, usually between test
scale, by t = T – 273.15 K. and reference weight
δm kg maximum permissible error on the
3.1.16 test weight (m )—weight that is to be tested according
t weights
m kg mass, nominal value of the weight
to this standard.
(e.g. 1 kg)
3.1.17 tolerance test—verification that the sum of the con-
m kg conventional mass of the weight
c
Δm kg conventional mass difference between
ventional mass corrections of the weights and their correspond- c
test weight and reference weight
ing uncertainties as calibrated are correct within the maximum
¯
kg average conventional mass difference
Δm
c
permissible errors of the respective weight class.
between test weight and reference
weight
3.1.18 uncertainty—non-negative parameter characterizing
m kg conventional mass of the reference
cr
the dispersion of the quantity values being attributed to a
weight
m kg conventional mass of the test weight
ct
measurand, based on the information used.
m kg mass of the sensitivity weight
s
3.1.19 units—the units used are: (1) for mass, the milligram
m kg mass of the test weight
t
n – subscript for number of measurement
(mg), the gram (g) and the kilogram (kg); (2) for density, the
sequences
–3
kilogram per cubic meter (kg m ).
p Pa barometric pressure
R μm mean height of roughness profile (R-
a
3.1.20 weight—material measure of mass, regulated in re-
parameter)
gard to its physical and metrological characteristics: shape,
R μm maximum height of roughness profile
z
(R-parameter)
dimensions, material, surface quality, nominal value, density,
r – subscript for reference weight
magnetic properties and maximum permissible error.
s – subscript for sensitivity weight
s kg standard deviation
NOTE 2—The term “weight” is also used as the physical quantity of the
2 2
s kg variance
gravitational force of a body. From the context it is usually clear in which
T K thermodynamic temperature using the
sense the term is used. If the sense is not clear, one may use the words
International Temperature Scale of
“weight force” or “weight piece,” depending on its meaning.
1990 (ITS-90)
ΔT* °C initial difference between weight tem-
3.2 Symbols:
perature and laboratory temperature
Symbol Unit Definition t – subscript for test weight
A – represents weighing the reference t °C temperature in degrees Celsius, where
weight in a weighing cycle t = T– 273.15 K
B – represents weighing the test weight in U kg uncertainty, expanded uncertainty
a weighing cycle u kg uncertainty, standard uncertainty
C – correction factor for air buoyancy u kg uncertainty of air buoyancy correction
b
D kg difference of balance readings u kg uncertainty of the balance
ba
between minimum and maximum
u kg combined standard uncertainty
c
values from eccentricity test u kg uncertainty due to the display resolu-
d
d kg scale interval
tion of a digital balance
E617 − 23
5. Physical Characteristics
Symbol Unit Definition
u kg uncertainty due to eccentricity
E
–3
5.1 Construction:
u kg m uncertainty of the formula used to
F
calculate air density
5.1.1 Type—Weights are divided into two types based upon
u % uncertainty in relative humidity
hr
the design:
u kg uncertainty due to instability of the
inst
reference weight
5.1.1.1 Type I—These weights are of one-piece construction
u kg uncertainty due to magnetism
ma
and contain no added adjusting material. They must be
u Pa uncertainty in barometric pressure
p
specified when weights are to be used as standards for the
u kg uncertainty due to the sensitivity of the
s
balance
calibration of weights of Classes 000, 00, 0, 1, 2, and 3, and
u °C uncertainty in temperature
t
where maximum stability is required. A precise measurement
u kg uncertainty due to the weighing
w
process of density can only be made for one-piece weights.
V m volume of a solid body (weight)
5.1.1.2 Type II—Weights of this type can be of any appro-
z m vertical cartesian coordinate
–2
priate design such as screw knob, ring, or sealed plug.
μ N A magnetic permeability
–2
μ N A magnetic constant (magnetic
0 Adjusting material can be used provided it is of a material at
permeability of vacuum), μ = 4π ×
least as stable as the base material and is contained in such a
–7 –2
10 N A
way that it will not become separated from the weight.
μ M T magnetic polarization
μ – relative magnetic permeability (μ/μ )
r 0
5.1.2 Class 000, 00, and 0 shall be Type I, one piece
v – effective degrees of freedom
eff
–3 construction.
ρ kg m mass of a rigid body (weight)
–3
ρ kg m density of air as a reference value
0 5.1.3 Class 1, 2, 3, 4, 5, 6, and 7 may be either Type I or
–3
equal to 1.2 kg m
Type II depending on the application.
–3
ρ kg m density of moist air
a
–3
ρ kg m density of moist air during the last
al
5.2 Design—A weight may have any shape that does not
(previous) calibration of the reference
introduce features that reduce the reliability. Weights with
weight
–3
ρ kg m density of a reference weight with
r nominal values less than 1 g shall have unique shapes to
mass m
r
differentiate the weights from one another. See Table 2. The
–3
ρ kg m density of the weight being tested
t
shape of weights smaller than 1 mg shall be discussed and
χ – magnetic susceptibility
verified with the customer. All weights shall be free of ragged
4. Maximum Permissible Errors or sharp edges or ends. Both sheet metal and wire weights shall
be free of cracks such as may be formed from bending.
4.1 For each weight, the expanded uncertainty U at approxi-
mately 95 % confidence (see Section 9) of the conventional 5.3 Surface Area—For classes 000, 00, 0, 1, 2, 3, and 4 the
mass shall be less than or equal to one-third of the maximum surface area is not to exceed twice the area of a cylinder of
permissible error given in Table 1 as defined in Section 9. equal height and diameter for weights 1 g and above. Sheet
Subsequent calibrations must meet all the requirements (in-
metal weights or wire weights may be used below 1 g. For
cluding environmental parameters as shown in Table 11, of Classes 5, 6, and 7 the total surface areas should be minimized
Sections 7, 8, and 9; and the requirements of ISO/IEC
to the extent possible.
17025:2017, 7.8.6 to make any claim of compliance to Speci-
5.4 Material:
fication E617, Maximum Permissible Errors, weight classes, or
5.4.1 Class 000, 00, 0, 1, 2, 3, 4, and 5 Weights—Shall be
metrological traceability.
made of austenitic stainless steel or a material whose hardness
4.1.1 For each weight, the conventional mass, m (deter-
c
and resistance to corrosion is similar to or better than that of
mined with an expanded uncertainty), shall not differ by more
austenitic stainless steel.
than the difference: maximum permissible error δm minus
5.4.2 Class 6 Weights—Class 6 weights below 5 kg shall be
expanded uncertainty, from the nominal value of the weight,
made of austenitic stainless steel or a material whose hardness
m :
o
and resistance to corrosion is similar to or better than that of
m 2 ~δm 2 U! # ~m ! # m 1~δm 2 U! (1)
o c o
austenitic stainless steel. Class 6 weights of 5 kg or greater
shall be made of grey cast iron or of another material whose
4.2 Maximum permissible errors for classes 000, 00, 0, 1, 2,
brittleness and resistance to corrosion is similar to or better
3, 4, 5, 6, and 7 are given in Table 1. These maximum
than that of grey cast iron. The surface of the weights may be
permissible errors apply to conventional mass values.
treated with a suitable coating in order to improve their
4.3 Maximum Permissible Errors for weights of denomina-
corrosion resistance. This coating shall withstand shocks and
tion intermediate between those listed, the maximum permis-
outdoor weather conditions.
sible error shall be interpolated from the nearest values shown.
5.4.3 Class 7 Weights—Class 7 weights below 100 g shall
Maximum permissible errors for nominal values outside the
be made of austenitic stainless steel or a material whose
ranges listed in Table 1 shall be interpolated from the nearest
hardness and resistance to corrosion is similar to or better than
metric equivalents.
that of austenitic stainless steel. Class 7 weights of 100 g or
4.4 For class 000, 00, and 0 weights, which are always greater shall be made of grey cast iron or of another material
accompanied by certificates giving the mass values and whose brittleness and resistance to corrosion is similar to or
uncertainties, the deviation from the nominal value, m – m , better than that of grey cast iron. The surface of the weights
c 0
shall be taken into account by the user. may be treated with a suitable coating in order to improve their
E617 − 23
TABLE 1 Maximum Permissible Errors
NOTE 1—Maximum Permissible Errors are reported in SI units, typically milligrams.
NOTE 2—The “grain” is the same in avoirdupois, troy, and apothecaries units of mass.
NOTE 3—See NIST SP 811 and NIST SP 1038 for conversion and units of measure.
NOTE 4—For nominal values not listed, see 4.3.
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
Metric
5000 kg 25 g 50 g 100 g 250 g 500 g 750 g
3000 kg 15 g 30 g 60 g 150 g 300 g 450 g
2000 kg 10 g 20 g 40 g 100 g 200 g 300 g
1000 kg 5.0 g 10 g 20 g 50 g 100 g 150 g
500 kg 2.5 g 5.0 g 10 g 25 g 50 g 75 g
300 kg 1.5 g 3.0 g 6.0 g 15 g 30 g 45 g
200 kg 1.0 g 2.0 g 4.0 g 10 g 20 g 30 g
100 kg 500 mg 1.0 g 2.0 g 5.0 g 10 g 15 g
50 kg 13 mg 25 mg 63 mg 120 mg 250 500 mg 1.0 g 2.5 g 5.0 g 7.5 g
30 kg 7.5 15 38 75 150 300 600 mg 1.5 g 3.0 g 5.1 g
25 kg 6.25 12.5 31 62 125 250 500 1.3 g 2.5 g 4.5 g
20 kg 5.0 10 25 50 100 200 400 1.0 g 2.0 g 3.8 g
10 kg 2.5 5.0 13 25 50 100 200 500 mg 1.0 g 2.2 g
5 kg 1.3 2.5 6.0 12 25 50 100 250 500 mg 1.4 g
3 kg 0.75 1.5 3.8 7.5 15 30 60 150 300 1.0 g
2 kg 0.5 1.0 2.5 5.0 10 20 40 100 200 750 mg
1 kg 0.25 0.5 1.3 2.5 5.0 10 20 50 100 470
500 g 0.13 0.25 0.60 1.2 2.5 5.0 10 35 70 300
300 g 0.075 0.15 0.38 0.75 1.5 3.0 6.0 30 60 210
200 g 0.05 0.10 0.25 0.50 1.0 2.0 4.0 20 40 160
100 g 0.025 0.05 0.13 0.25 0.50 1.0 2.0 10 20 100
50 g 0.015 0.030 0.060 0.12 0.25 0.60 1.2 5.0 10 62
30 g 0.014 0.026 0.037 0.074 0.15 0.45 0.90 3.0 6.0 44
20 g 0.013 0.025 0.037 0.074 0.10 0.35 0.70 2.0 4.0 33
10 g 0.010 0.020 0.025 0.050 0.074 0.25 0.50 1.0 2.0 21
5 g 0.005 0.010 0.017 0.034 0.054 0.18 0.36 0.75 1.5 13
3 g 0.005 0.010 0.017 0.034 0.054 0.15 0.30 0.64 1.3 9.4
2 g 0.005 0.010 0.017 0.034 0.054 0.13 0.26 0.56 1.1 7.0
1 g 0.005 0.010 0.017 0.034 0.054 0.10 0.20 0.45 0.90 4.5
500 mg 0.002 0.003 0.005 0.010 0.025 0.080 0.16 0.36 0.72 3.0
300 mg 0.002 0.003 0.005 0.010 0.025 0.070 0.14 0.31 0.61 2.2
200 mg 0.002 0.003 0.005 0.010 0.025 0.060 0.12 0.27 0.54 1.8
100 mg 0.002 0.003 0.005 0.010 0.025 0.050 0.10 0.22 0.43 1.2
50 mg 0.002 0.003 0.005 0.010 0.014 0.042 0.085 0.17 0.35 0.88
30 mg 0.002 0.003 0.005 0.010 0.014 0.038 0.075 0.15 0.30 0.68
20 mg 0.002 0.003 0.005 0.010 0.014 0.035 0.070 0.13 0.26 0.56
10 mg 0.002 0.003 0.005 0.010 0.014 0.030 0.060 0.10 0.21 0.40
5 mg 0.002 0.003 0.005 0.010 0.014 0.028 0.055 0.083 0.17
3 mg 0.002 0.003 0.005 0.010 0.014 0.026 0.052 0.071 0.14
2 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.062 0.12
1 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.050 0.10
0.5 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.0540 0.080
0.3 mg 0.002 0.003 0.005 0.010 0.014 0.025
0.2 mg 0.002 0.003 0.005 0.010 0.014
0.1 mg 0.002 0.003 0.005 0.010
0.05 mg 0.002 0.003 0.005 0.010
Avoirdupois Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
Pound mg mg mg g & mg g & mg g & mg g & mg g & mg
10000 lb 45 g 91 g 230 g 450 g 680 g
5000 lb 23 g 45 g 110 g 230 g 340 g
3000 lb 14 g 27 g 68 g 140 g 200 g
2500 lb 11 g 23 g 57 g 110 g 170 g
2000 lb 9.1 g 18 g 45 g 91 g 140 g
1000 lb 4.5 g 9.1 g 23 g 45 g 68 g
500 lb 2.3 g 4.5 g 11 g 23 g 34 g
300 lb 1.4 g 2.7 g 6.8 g 14 g 20 g
200 lb 910 mg 1.8 g 4.5 g 9.1 g 14 g
100 lb 57 110 230 450 910 mg 2.3 g 4.5 g 6.8 g
50 lb 29 57 110 230 450 1.1 g 2.3 g 4.1 g
30 lb 17 32 68 140 270 680 mg 1.4 g 2.7 g
25 lb 14 28 57 110 230 570 1.1 g 2.4 g
20 lb 12 23 45 91 180 450 910 mg 2.0 g
10 lb 5.5 11 23 45 91 230 450 1.3 g
5 lb 2.7 5.4 11 23 45 110 230 780 mg
3 lb 1.7 3.4 6.8 14 27 68 140 580
E617 − 23
TABLE 1 Continued
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
2 lb 1.2 2.3 4.5 9.1 18 45 91 440
1 lb 0.55 1.1 2.3 4.5 9.1 35 70 270
0.5 lb 0.27 0.54 1.1 2.3 4.5 23 45 170
0.3 lb 0.17 0.34 0.68 1.4 2.7 14 27 120
0.2 lb 0.12 0.23 0.45 0.91 1.8 9.0 18 97
0.1 lb 0.055 0.11 0.23 0.57 1.1 4.5 9.1 59
0.05 lb 0.037 0.074 0.11 0.36 0.77 2.3 4.5 37
0.03 lb 0.029 0.059 0.083 0.32 0.59 1.4 2.7 26
0.02 lb 0.024 0.047 0.070 0.23 0.45 0.91 1.8 20
0.01 lb 0.017 0.034 0.054 0.16 0.34 0.73 1.5 12
0.005 lb 0.017 0.034 0.054 0.14 0.27 0.58 1.2 7.8
0.003 lb 0.017 0.034 0.054 0.11 0.22 0.50 0.99 5.4
0.002 lb 0.015 0.030 0.049 0.091 0.19 0.44 0.87 4.2
0.001 lb 0.0050 0.010 0.024 0.078 0.15 0.35 0.70 2.9
0.0005 lb 0.0045 0.0091 0.023 0.064 0.13 0.28 0.56 2.0
Avoirdupois Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
Ounce mg mg mg mg mg mg mg mg
10 oz 0.35 0.70 1.4 2.8 5.4 28 57 200
8 oz 0.27 0.54 1.1 2.3 4.5 23 45 170
5 oz 0.18 0.35 0.71 1.4 2.8 14 28 130
4 oz 0.14 0.28 0.57 1.1 2.3 11 23 110
3 oz 0.10 0.21 0.43 0.91 1.8 8.5 17 90
2 oz 0.070 0.14 0.28 0.64 1.3 5.7 11 70
1 oz 0.037 0.074 0.14 0.43 0.86 2.8 5.4 42
1/2 oz 0.030 0.060 0.085 0.30 0.59 1.4 2.8 27
1/4 oz 0.020 0.041 0.062 0.20 0.43 0.84 1.7 17
1/8 oz 0.017 0.034 0.054 0.16 0.31 0.67 1.3 10
1/16 oz 0.017 0.034 0.054 0.12 0.24 0.50 1.1 6.5
1/32 oz 0.014 0.029 0.047 0.095 0.19 0.43 0.87 4.2
1/64 oz 0.0050 0.010 0.023 0.077 0.15 0.35 0.69 2.8
0.5 oz 0.030 0.060 0.085 0.30 0.59 1.4 2.8 27
0.3 oz 0.068 0.23 0.45 0.89 1.8 19
0.2 oz 0.057 0.19 0.38 0.78 1.6 14
0.1 oz 0.054 0.14 0.29 0.63 1.3 9.0
0.05 oz 0.054 0.11 0.23 0.50 1.0 5.7
0.03 oz 0.045 0.095 0.19 0.43 0.85 4.1
0.02 oz 0.029 0.077 0.18 0.38 0.75 3.2
0.01 oz 0.023 0.064 0.14 0.30 0.60 2.2
Troy Ounce Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg mg g & mg
1000 oz t 310 mg 620 mg 1.6 g 3.1 g
500 oz t 160 310 780 mg 1.6 g
300 oz t 91 190 470 930 mg
200 oz t 62 120 310 620
100 oz t 31 62 160 310
50 oz t 16 31 78 160
30 oz t 9.1 19 47 93
20 oz t 6.2 12 39 77
10 oz t 3.1 6.2 30 61
5 oz t 1.6 3.1 16 31
3 oz t 0.91 1.9 9.3 19
2 oz t 0.71 1.4 6.2 12
1 oz t 0.45 0.91 3.1 6.2
0.5 oz t 0.31 0.62 1.6 3.1
0.3 oz t 0.24 0.49 0.92 1.8
0.2 oz t 0.20 0.40 0.80 1.6
0.1 oz t 0.15 0.30 0.65 1.3
0.05 oz t 0.12 0.23 0.52 1.0
0.03 oz t 0.097 0.19 0.44 0.88
0.02 oz t 0.084 0.17 0.39 0.77
0.01 oz t 0.071 0.14 0.31 0.62
0.005 oz t 0.056 0.11 0.25 0.50
Pennyweight Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg mg mg g & mg
10000 dwt 155 310 780 1.6 g
5000 dwt 78 160 390 780 mg
3000 dwt 47 91 230 470
2000 dwt 31 62 160 310
1000 dwt 16 31 78 160
500 dwt 7.8 16 43 87
300 dwt 4.7 9.1 34 68
200 dwt 3.1 6.2 30 61
100 dwt 1.6 3.1 16 31
50 dwt 0.78 1.6 7.8 16
E617 − 23
TABLE 1 Continued
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
30 dwt 0.58 1.2 4.7 9.3
20 dwt 0.46 0.91 3.1 6.2
10 dwt 0.31 0.62 1.6 3.1
5 dwt 0.22 0.44 0.89 1.7
3 dwt 0.17 0.34 0.73 1.5
2 dwt 0.15 0.3 0.65 1.3
1 dwt 0.12 0.23 0.52 1.0
Grain Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg mg mg mg
10000 gr 6.5 13 36 79
5000 gr 3.2 6.5 31 61
3000 gr 1.9 3.9 19 39
2000 gr 1.3 2.6 13 26
1000 gr 0.71 1.4 6.5 13
500 gr 0.48 0.91 3.2 6.5
300 gr 0.35 0.65 1.9 3.9
200 gr 0.28 0.57 1.3 2.6
100 gr 0.20 0.40 0.82 1.6
50 gr 0.15 0.30 0.65 1.3
30 gr 0.12 0.25 0.55 1.1
20 gr 0.11 0.21 0.48 0.91
10 gr 0.084 0.17 0.42 0.77
5 gr 0.071 0.14 0.31 0.62
3 gr 0.060 0.12 0.25 0.53
2 gr 0.054 0.11 0.22 0.45
1 gr 0.045 0.091 0.17 0.37
0.5 gr 0.039 0.078 0.14 0.31
Carat Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg mg
5000 c 10 20
3000 c 6.0 12
2000 c 4.0 8.0
1000 c 2.0 4.0
500 c 1.0 2.0
300 c 0.69 1.3
200 c 0.52 1.0
100 c 0.35 0.7
50 c 0.25 0.50
30 c 0.19 0.40
20 c 0.16 0.33
10 c 0.13 0.26
5 c 0.10 0.20
3 c 0.086 0.17
2 c 0.075 0.15
1 c 0.060 0.12
0.5 c 0.050 0.10
Apothecary Ounce mg
12 oz ap 64
10 oz ap 61
6 oz ap 37
5 oz ap 31
4 oz ap 25
3 oz ap 19
2 oz ap 12
1 oz ap 6.2
Apothecary Dram mg
6 dr ap 4.7
5 dr ap 3.9
4 dr ap 3.1
3 dr ap 2.3
2 dr ap 1.8
1 dr ap 1.4
Apothecary mg
Scruple
2 s ap 1.2
1 s ap 0.91
corrosion resistance. This coating shall withstand shocks and 5.5 Magnetism—Weights shall not exceed maximum per-
outdoor weather conditions. missible magnetic properties as listed in Tables 3 and 4 for any
E617 − 23
TABLE 2 Shape of Weights 1 g or Less
bration laboratories. For Class 000 and 00, the manufacturer
Nominal Values Polygonal Sheets Wires shall provide a measured value for the density of the weights at
5, 50, 500 mg Pentagon Pentagon time of manufacture. Use of a sample taken adjacent to the
3, 30, 300 mg Circle Circle
material from which the weight is manufactured to measure the
2, 20, 200 mg Square Square
density is permitted, however an additional uncertainty com-
1, 10, 100, 1000 mg Triangle Triangle
–5
ponent equal to 5 × 10 must be combined with the standard
uncertainty of the density determination process. For subse-
TABLE 3 Maximum Polarization, μ M, (μT) quent calibrations, the density may be verified by the calibrat-
ing laboratory if there are concerns and capability for the
000, 00,
Weight Class 1 2 and 3 4 and 5 6 and 7
and 0
measurement is available. As lower precision of measurement
Maximum
is required, so the range of density is broadened. See Table 5.
Not
polarization, 2.5 8 25 80
applicable 5.6.1 The determination of the minimum and maximum
μ M, (μT)
density limits for nominal values not listed in Table 5 shall be
converted to metric values and the limits of the metric value
next greater than the converted value used. (Example: 1
TABLE 4 Maximum Magnetic Susceptibility, χ
apothecary ounce class 4 is equal to 31.1034768 g, therefore
000, 00,
Weight Class 1 2 and 3 4 and 5 6 and 7
and 0 the density limits are equal to values listed in Table 5 for 50 g.)
m # 1 g 0.25 0.9 10 Not Not
NOTE 4—Materials used to make weights for special applications that
applicable applicable
do not fall within the density limits stated above, should have stated
2 g # m # 10 g 0.06 0.18 0.7 4 Not
densities or density determinations performed.
applicable
20 g # m 0.02 0.07 0.2 0.8 Not
5.7 Finish—The surface of the weights (including the base
applicable
and corners) shall be smooth, the edges shall be rounded, and
the weights shall not be porous.
5.7.1 The surface quality of a weight shall not exceed
portion of the weight. If the values of all local measurements of
maximum values of surface roughness, R and R through
a z
magnetization and susceptibility are less than these limits, then
visual inspection using a hand held gage. See Table 6.
it may be assumed that the uncertainty components due to the
5.7.2 The surface roughness of all weight surfaces must
magnetism of the weight are negligible. The maximum perma-
meet the limits established in Table 6 for the weight classifi-
nent magnetization and magnetic susceptibilities given in
cation.
Tables 3 and 4 are such that, at magnetic fields and magnetic
5.8 Adjustment:
field gradients possibly present on balance pans, they produce
5.8.1 Type I Weights—Weights shall be adjusted by
a change of the conventional mass of less than one tenth of the
abrasion, grinding or any appropriate method. The surface
maximum permissible error of the test weight.
requirements shall be met at the end of the adjustment process.
NOTE 3—Magnetic susceptibility may be tested in accordance with
5.8.2 Type II Weights—Weights with adjusting cavities shall
OIML R 111-1, Annex B. Cast iron cannot have a susceptibility specifi-
be adjusted with the same material from which they are made,
cation of any real value.
or with materials that are at least as stable and of similar
5.6 Density—Because of the effect of the buoyant force of
density as the base material. For weights which have sealing
air on a weight, precision measurements of mass require that
caps, the cap may be made of aluminum. The back-up spacer
the volume of the weight be known, as well as the density of
should be of a similar material as the weight. Adjusting
the air in which it is being measured, so that appropriate
material and back-up disc must meet the magnetic require-
corrections can be made. For weights of higher precision, the
ments specified for the accuracy class of the weight.
range of density is limited to values at or near the density of
well-established standards, such as are used by primary cali- 5.9 Marking:
TABLE 5 Minimum and Maximum Limits for Density
3 –3
ρ , ρ (10 kg m )
min max
Nominal
Class of Weight
Value
000 00 and 0 1 2 and 3 4 and 5 6 and 7
$ 100 g 7.967 – 8.033 7.934 – 8.067 7.81 – 8.21 7.39 – 8.73 6.4 – 10.7 $ 4.4
50 g 7.960 – 8.040 7.92 – 8.08 7.74 – 8.28 7.27 – 8.89 6.0 – 12.0 $ 4.0
20 g 7.900 – 8.090 7.84 – 8.17 7.50 – 8.57 6.6 – 10.1 4.8 – 24.0 $ 2.6
10 g 7.87 – 8.14 7.74 – 8.28 7.27 – 8.89 6.0 – 12.0 $ 4.0 $ 2.0
5 g 7.87 – 8.13 7.62 – 8.42 6.9 – 9.6 5.3 – 16.0 $ 3.0
2 g 7.62 – 8.42 7.27 – 8.89 6.0 – 12.0 $ 4.0 $ 2.0
1 g 7.28 – 8.90 6.9 – 9.6 5.3 – 16.0 $ 3.0
500 mg 7.50 – 8.60 6.3 – 10.9 $ 4.4 $ 2.2
200 mg 6.7 – 9.6 5.3 – 16.0 $ 3.0
100 mg 6.0 – 12 $ 4.4
50 mg $ 4.8 $ 3.4
20 mg $ 2.5 $ 2.3
E617 − 23
TABLE 6 Maximum Values of Surface Roughness
to those of high stability. Appendix X1 should serve as a guide
Classes 000, Classes Classes Class Classes in selecting weights for specific applications.
00, 0 1, 2 3, 4 5 6, 7
6.2 Class—Maximum permissible errors for Classes 000
R (μm) 0.1 0.2 0.4 1 25
A
through 7 are shown in Table 1. Lower numbers indicate
R (μm) 0.5 1 2 5 100
Z
smaller maximum permissible errors.
6.3 Lifters:
6.3.1 Classes 000, 00, 0, 1, 2, 3, and 4 shall be supplied with
5.9.1 Class 000, 00, and 0 weights shall not bear any
lifters when sets of weights are ordered. Individual weights
indication of nominal value and shall not be marked unless
shall be supplied with lifters when specified by the purchaser.
used to distinguish from another class 000, 00, or 0 weight,
Lifters or forceps shall securely hold the weights for which
provided that the surface quality and stability of the weights are
they are designed. Additional pressure shall not cause the
not affected by the markings or by the process used to mark it.
dropping of small weights or the forceful ejection of large
5.9.2 Numerical Value for Classes 1, 2, 3, 4, 5, 6, and
weights.
7—The nominal value of each weight shall appear on the
6.3.2 For weights 500 g or larger, the parts of the lifter that
surface of each weight. Only the numerical portion of the
come in contact with the weights shall be covered with a
weight value needs to be on the surface of weights. Weights
non-magnetic material softer than the surface of the weight,
made of wire or too small to be marked shall not be marked but
such as plastic or chamois skin from which the grease has been
should be identifiable by their shape or number of bends.
removed.
5.9.3 Units of Weight—Weights 100 g and greater may be
6.3.3 For smaller weights, the lifters may be of the same
marked with the unit name or abbreviation. In the case of sets
design where practical or may be of a non-magnetic material
of non-metric weights, at least the largest weight of a particular
softer than the weights, such as close-grained wood or plastics
set should be marked with the unit name or abbreviation. In
not affected by alcohol. When the parts of the lifters or forceps
any case the unit shall not be included where such marking
which come in contact with the weights are not covered by a
would be illegible.
soft material, they shall be smooth and polished and the edges
5.9.4 Abbreviations—The accepted abbreviation may be
on which the weight may be lifted shall be well rounded.
used in marking. Abbreviations are shown in Appendix X2.
6.3.4 If forceps are used for lifting small weights, stainless
Periods shall not be used with abbreviations in marking
steel forceps with nonmetallic tips may be used, where the tips
weights.
that come in contact with the weights are covered with a
5.9.5 Multiple Weights—Multiple weights of the same
material softer than the surface of the weight, such as plastic or
nominal value included in a set of weights shall have distin-
chamois skin from which the grease has been removed. The
guishing marks.
forceps may also be made of a material softer than the weights,
5.9.6 Depth of Markings—Markings shall be clear, shallow,
such as close-grained wood or plastics not affected by alcohol.
relatively broad, and free of burrs and sharp angles. Markings
When the parts of the forceps which come in contact with the
shall not perforate or crack sheet metal weights.
weights are not covered by a soft material, they shall be smooth
5.9.7 User Marking—It is recommended for a user to clearly
and polished and the edges on which the weight may be lifted
identify individual weights as it helps to link a weight to its
shall be well rounded.
calibration certificate or verification document. The recom-
6.4 Cases:
mended maximum values for user markings are given in Table
6.4.1 Classes 000, 00, 0, 1, 2, 3, and 4 weights, when
7.
supplied in sets, may be supplied with one or more cases or
6. Ordering Information shall meet customer specifications for cases. The case shall be
designed so that as long as the lid remains closed, the weights
6.1 Selection of type and class depends upon the application
shall be held secure, and when possible the pocket depth shall
of the weights. For reference standards, stability and informa-
be such that the shoulder of the weight does not extend above
tion about the values of the weights is more important than the
the edge of the pocket. The hinges and locks shall be adequate
closeness of the values to nominal. Weights to be used with
to hold the lid closed with any reasonable handling. There shall
balances of low precision do not require small maximum
be no discoloration of weights due to the lining of the case,
permissible errors, nor need the choice of materials be limited
such as might result from long storage in a warm or damp
location. This condition does not apply to weights not designed
to be handled manually.
TABLE 7 Recommended Maximum Number of User Markings
6.4.2 Pockets—A separate pocket shall be supplied for each
Maximum Number
Nominal Height of
Class of Signs, Numerals, weight and for each forceps and lifter, except that extremely
Value Lettering
or Letters
large lifters may not require pockets. All pockets shall be large
000 – 7 < 1 g 1 mm 2
enough so that no appreciable friction shall be encountered in
000, 00, 0 $ 1 g 2 mm 3
inserting or removing weights. If the cover is not lined, the
1 $ 1 g 3 mm 5
2 – 7 1 g to 100 g 3 mm 5 individual holes in the cover shall be smooth or lined. Pockets
2 – 7 200 g to 10 kg 5 mm 5
for weights 1 g or equivalent or larger shall be constructed of
2 - 3 $ 20 kg 7 mm 5
a smooth nonabrasive material, or lined with a smooth,
4 – 7 $ 20 kg 12 mm 5
nonabrasive material.
E617 − 23
6.5 Denominations—The customer’s purchase order or con- maintained their metrological properties. Any weights found
tract shall define the contents of the weight set. defective at the time of re-calibration shall be reviewed with
the customer.
6.6 Density Identification—Weights that are to be calibrated
7.2.2 For subsequent calibration, as a minimum, the
shall carry identification of the density of the various materials
weights shall be visually inspected for design and surface
of which the weights are manufactured. Identification of
conditions and the conventional mass or conventional mass
density shall be displayed on the certificate, or on the cover or
correction reported, with the associated measurement
interior of the box.
uncertainty, on a calibration certificate as specified in 7.1.1.
6.7 Special Requirements—If a customer has specific re-
quirements that deviate from this standard (that is, material,
8. Calibration Procedures
shape, maximum permissible errors, and so forth) the manu-
8.1 Weight manufacturers must be able to provide evidence
facturer may use this specification as a reference, not a
that all new weights comply with specifications in this standard
requirement, to provide the customer with the weights that they
(for example, material, density, magnetism, surface finish,
need.
mass values, uncertainties) to make any claim of compliance to
7. Certificates Specification E617, Maximum Permissible Errors, weight
classes, or metrological traceability.
7.1 Calibration—Laboratories issuing calibration certifi-
8.1.1 During subsequent calibrations, calibration laborato-
cates for weights and weight sets shall have evidence of
ries must meet the requirements of ISO/IEC 17025:2017.
metrological traceability to the International System of Units
8.1.2 Subsequent calibrations must meet all the
(SI). Calibration certificates shall be issued only by laborato-
requirements, including Sections 7, 8, and 9, Table 8 and Table
ries having a quality system complying with the requirements
11 (environmental parameters) to make any claim of compli-
of ISO/IEC 17025:2017, which has preferably been verified by
ance to Specification E617, Maximum Permissible Errors,
third party assessment (accreditation).
weight classes, or metrological traceability.
7.1.1 Calibration Certificates—A calibration certificate
shall state, as a minimum: the conventional mass of each
NOTE 5—Requirements set forth in NIST IR 6969 and NIST IR 5672
weight, m , an indication of whether a weight has been adjusted are compliant with all the requirements of Specification E617, Sections 7,
c
8, and 9.
prior to calibration, its expanded uncertainty, U, and the value
of the coverage factor, k, and meet the requirements of
8.2 Cleaning Weights:
ISO/IEC 17025:2017 and the GUM, as interpreted by ILAC,
8.2.1 Before calibration, dust and any foreign particles shall
for metrological traceability.
be removed by brushing with a clean soft-bristle brush or
7.1.2 Class 000, 00, and 0 weights shall be accompanied by
gentle wiping with a non-abrasive lint-free wipe. Cleaning
a calibration certificate meeting the requirements of 7.1.1.
should not remove any significant amounts of weight material.
7.1.2.1 The certificate for Class 000, 00, and 0 weights shall
Weights should be handled and stored in such a way that they
state, as a minimum, the values of conventional mass, m , the
c
stay contamination-free. Care must be taken to not change the
expanded uncertainty, U, and the coverage factor, k, and the
surface properties of the weight (that is, by scratching the
density or volume for each weight. In addition, the certificate
weight).
shall state if the density or volume was measured or estimated.
8.2.2 If a weight contains significant amounts of contami-
7.2 Calibration, Initial Verification from the Manufacturer nation that cannot be removed by brushing with a clean
and Subsequent Calibration: soft-bristle brush or gentle wiping with a non-abrasive lint-free
7.2.1 Table 8 gives the required tests for initial calibration wipe, the weight or some part of it can be washed with solvents
from the manufacturer and subsequent calibration. The catego- leaving no significant amount of resid
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E617 − 18 E617 − 23
Standard Specification for
Laboratory Weights and Precision Mass Standards
This standard is issued under the fixed designation E617; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This specification covers weights and mass standards used in laboratories, specifically classes 000, 00, 0, 1, 2, 3, 4, 5, 6, and
7. This specification replaces National Bureau of Standards Circular 547, Section 1, which is out of print.
1.2 This specification and calibration method is intended for use by weight manufacturers, national metrology institutes, weight
calibration laboratories, accreditation bodies, users of weights, and regulatory bodies.
1.3 This specification contains the principal physical characteristics and metrological requirements for weights that are used.
1.3.1 For the verification of weighing instruments;
1.3.2 For the calibration of weights of a lower class of accuracy; and
1.3.3 With weighing instruments.
1.4 Maximum Permissible Errors (formerly tolerances) and design restrictions for each class are described in order that both
individual weights or sets of weights can be chosen for appropriate applications.
1.5 Weight manufacturers must be able to provide evidence that all new weights comply with specifications in this standard (for
example, material, density, magnetism, surface finish, mass values, uncertainties). Statements of compliance by calibration
laboratories during subsequent calibrations must meet the requirements of ISO/IEC 17025, 5.10.4.2 and indicate on the calibration
report which sections have or have not been assessed. Subsequent calibrations must meet all the requirements (including
environmental parameters as shown in Table 11, of Sections uncertainties) 7, 8, and 9; and the requirements of ISO/IEC
17025:2005, 5.10.4.2 to make any claim of compliance to Specification E617,E617, Maximum Permissible Errors, weight classes,
or metrological traceability.
1.5.1 During subsequent calibrations, calibration laboratories must meet the requirements of ISO/IEC 17025:2017.
1.5.2 Subsequent calibrations must meet all the requirements, including Sections 7, 8, and 9, Table 8 and Table 11 (environmental
parameters) to make any claim of compliance to Specification E617, Maximum Permissible Errors, weight classes, or metrological
traceability.
This specification is under the jurisdiction of ASTM Committee E41 on Laboratory Apparatusand is the direct responsibility of Subcommittee E41.06 on Laboratory
Instruments and Equipment.
Current edition approved Oct. 1, 2018Aug. 15, 2023. Published November 2018November 2023. Originally approved in 1978. Last previous edition approved in 20132018
as E617 – 13.E617 – 18. DOI: 10.1520/E0617-18.10.1520/E0617-23.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E617 − 23
NOTE 1—Requirements set forth in NIST IR 6969 and NIST IR 5672 are compliant with all the requirements of Specification E617, Sections 7, 8, and
9.
1.6 The values stated in SI units are to be regarded as standard.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ISO Standards:
ISO/IEC 1702517025:2017 General Requirements for the Competence of Testing and Calibration Laboratories (2005)
2.2 NIST Standards:
NIST Handbook 143 State Weights and Measures Laboratories Program Handbook (2007)(2019)
NIST SP 811 Guide for the Use of the International System of Unit (SI) 2008 Edition
NIST SP 1038 The International System of Units (SI) – Conversion Factors for General Use (May 2006)
NISTIR 5672 Advanced Mass Calibration and Measurement Assurance Program for State Calibration Laboratories
(2014)(2019)
NISTIR 6969 Selected Laboratory and Measurement Practices to Support Basic Mass Calibrations (2017)(2019)
NIST Technical Note 1297 (1994) Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results
(1994)
2.3 OIML Standards:
OIML D 28 Conventional Value of the Result of Weighing in Air (2004)
OIML R111–1e04R 111–1 Weights of classes E1, E2, F1, F2, M1, M1–2, M2, M2–3 and M3 Part 1: Metrological and Technical
Requirements (2004)(Edition 2004)
2.4 BIPM Standards:
VIM: JCGM 200:2012 International Vocabulary of Metrology–Basic and General Concepts and Associated Terms
GUM: JCGM 100:2008 Evaluation of Measurement Data–Guide to the Expression of Uncertainty in Measurement
th
SI Brochure: 2019 The International System of Units (SI), 9 Edition
2.5 EURAMET Standards:
EURAMET/cg-18/V. 4.0 Guidelines on the Calibration of Non-Automatic Weighing Instruments (2015)
3. Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 accuracy class of weights—a class of weights that meets certain metrological requirements intended to keep the errors within
specified limits.
3.1.2 balance—instrument indicating apparent mass that is sensitive to the following forces:
Force due to gravity
F 5m·g
g
Air buoyancy equal to the weight of the displaced air.
m
F 5v·ρ ·g 5 ρ ·g
b a a
ρ
Vertical component of the magnetic interaction between the weight and the bal-
≠H
ance or the environment, or both.
F 5μ M 1 χ H dV
eees d
z o
≠z
v
H and M are vectors; z is the vertical cartesian coordinate. If magnetic effects are negligible, that is, the permanent magneti-
zation (M) of the weight and the magnetic susceptibility (χ) are sufficiently small, and the balance is calibrated with reference
weights of well-known mass, the balance can be used to indicate the conventional mass, m , of a body under conventionally
c
chosen conditions.
Available from International Organization for Standardization (ISO), 1, ch. de la Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http://www.iso.org.
Available from National Institute of Standards and Technology (NIST), 100 Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http://www.nist.gov.
Available from Organisation Internationale de Metrologie Legale, 11 Rue Turgot, 75009 Paris, France, http://www.oiml.org.
Available from Bureau International des Poids et Mesures (BIPM), Pavilion de Breteuil, F-92312, Sèvres Cedex, France, http://www.bipm.org.
Available from Euramet, Bundesallee 100, 38116 Braunschweig, Germany, http://www.euramet.org.
E617 − 23
3.1.3 calibration (of weights)—the acts of determining the mass difference between a standard of known mass value and an
“unknown” test weight or set of weights, establishing the mass value and conventional mass value of the “unknown,” and of
determining a quantitative estimate of the uncertainty to be assigned to the stated mass or conventional mass value of the
“unknown,” or both, and providing metrological traceability to the “unknown.”
3.1.3.1 calibration (generally)—set of operations that establish, under specified conditions, the relationship between values of
quantities indicated by a measuring instrument or measuring system, or values represented by a material measure or a reference
material, and the corresponding values realized by standards.
3.1.4 calibration certificate—certificate issued by calibration laboratories to document the results of a calibration.
3.1.5 conventional mass—conventional value of the result of weighing in air, in accordance to International Recommendation
OIML D 28. For a weight taken at 20°C,20 °C, the conventional mass is the mass of a reference weight of a density of 8000
3–3 3–3
kg/mkg m which it balances in air of density of 1.2 kg/mkg m .
3.1.6 correction—mass values are traditionally expressed by two numbers, one being the nominal mass of the weight, and the
second being a correction. The mass of the weight is the assigned nominal value plus the assigned correction. Positive corrections
indicate that the weight embodies more mass than is indicated by the assigned nominal value. Negative corrections indicate that
the weight embodies less mass than is indicated by the assigned nominal value. The correction is equivalent to the “error.”
3.1.7 international prototype kilogram—the SI base unit of mass. Formerly the international prototype kilogram (IPK), the
platinum-iridium cylinder maintained at the International Bureau of Weights and Measures (BIPM), at Sevres, France with an
internationally accepted (BIPM) in Sèvres, France, was internationally accepted as having a defined mass of 1 kg. In 2018 the
–34 2 –1
kilogram was redefined in terms of the Planck constant h (taken to be 6.626 070 15 × 10 kg m s ) where the mass of primary
mass standards may be determined by any primary method such as those described in the mise en pratique for the definition of
the kilogram in the BIPM SI Brochure. The redefinition process ensured continuity of mass measurements before and after
redefinition such that all mass values traceable to the IPK remained the same when the new definition came into effect, but the
uncertainties of these mass values were increased by the relative standard uncertainty of the IPK immediately after the redefinition
–8
(1.0 × 10 ).
3.1.8 magnetism—effect that generates an attractive or repulsive force.force without the presence of charged species.
3.1.8.1 (volume) magnetic susceptibility (χ)—measure of the ability of a medium to modify a magnetic field. It is related to the
magnetic permeability (μ) by the relation: μ/μ = 1 + χ. The quantity μ/μ is sometimes referred to as the relative permeability, μ .
0 0 r
3.1.8.2 (permanent) magnetization (M)—parameter that specifies a magnetic state of material bodies such as weights, in the
absence of an external magnetic field (most generally, magnetization is a vector whose magnitude and direction are not necessarily
constant within the material). The magnetization of a body generates an inhomogeneous magnetic field in space and thus may
produce magnetic forces on other materials.
3.1.9 mass—physical quantity, which can be ascribed to any material object and which gives a measure of its quantity of matter.
The unit of mass is the kilogram.
3.1.10 maximum permissible errors—the maximum amount by which the sum of the conventional mass correction of the weight,
its deviation from nominal value, and its associated uncertainty is allowed to deviate from the assigned nominal value.
3.1.11 metrological traceability—property of a measurement result whereby the result can be related to a reference through a
documented unbroken chain of calibrations, each contributing to the measurement uncertainty. Metrological traceability requires
an established calibration hierarchy. Elements for confirming metrological traceability to be an unbroken chain to an international
measurement standard or a national measurement standard (IPK or NPS), a primary mass standard shall include a documented
measurement uncertainty, a documented measurement procedure, accredited technical competence, metrological traceability to the
SI, and established calibration intervals (see current VIM: JCGM 200).
3.1.12 reference standard—a standard, generally of the highest metrological quality available at a given location, from which
measurements made at that location are derived.
3.1.13 roughness parameter or R-parameter (R or R )—parameter that describes the assessed roughness profile of a sample. The
a z
E617 − 23
letter R is indicative of the type of assessed profile, in this case R for roughness profile. The assessed profile of a sample can be
in terms of different profile types: a roughness profile or R-parameter, primary profile or P-parameter, a waviness profile or
W-parameter.
3.1.14 set of weights—a series of weights, usually presented in a case so arranged to make possible any weighing of all loads
between the mass of the weight with the smallest nominal value and the sum of the masses of all weights of the series with a
progression in which the mass of the smallest nominal value weight constitutes the smallest step of the series.
3.1.15 temperature (t)—in degrees Celsius, is related to the absolute thermodynamic temperature scale, called the Kelvin scale,
by t = T – 273.15 K.
3.1.16 test weight (m )—weight that is to be tested according to this standard.
t
3.1.17 tolerance test—verification that the sum of the conventional mass corrections of the weights and their corresponding
uncertainties as testedcalibrated are correct within the maximum permissible errors of the respective weight class.
3.1.18 uncertainty—non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand,
based on the information used.
3.1.19 units—the units used are: (1) for mass, the milligram (mg), the gram (g) and the kilogram (kg); (2) for density, the kilogram
–3
per cubic meter (kg m ).
3.1.20 U.S. National prototype standard—platinum-iridium kilogram identified as K20, maintained at the National Institute of
Standards and Technology, with value assigned relative to the International Prototype Kilogram provides the United States access
to the mass unit.
3.1.20 weight—material measure of mass, regulated in regard to its physical and metrological characteristics: shape, dimensions,
material, surface quality, nominal value, density, magnetic properties and maximum permissible error.
NOTE 2—The term “weight” is also used as the physical quantity of the gravitational force of a body. From the context it is usually clear in which sense
the term is used. If the sense is not clear, one may use the words “weight force” or “weight piece,” depending on its meaning.
3.2 Symbols:
Symbol Unit Definition
A – represents weighing the reference
weight in a weighing cycle
B – represents weighing the test weight in
a weighing cycle
C – correction factor for air buoyancy
D kg difference of balance readings
between minimum and maximum
values from eccentricity test
d kg scale interval
d m estimated distance between centers of
weights during loading
d m estimated distance from the center of
the load receptor to one of the corners
F N air buoyancy equal to the weight of the
b
displaced air
F N gravitational force
g
F N magnetic force between a mass
z
comparator and a weight in the vertical
or z-direction
–2
g m s gravitational acceleration
–1
H A m magnetizing field strength
hr % relative humidity
h % relative humidity
r
I kg indication of the weighing instruments
(scale division)
ΔI kg indication difference of the balance,
where ΔI = I – I
t r
E617 − 23
Symbol Unit Definition
ΔI kg indication difference using an
automatic exchange mechanism with
weights in first position
ΔI kg indication difference using an
automatic exchange mechanism with
weights in reversed position
ΔI kg change in indication of balance due to
s
sensitivity weight
i – subscript used as an index in
summations
j – subscript for number of test weights or
number of series of measurements
k – coverage factor, typically 2 or 3
–1
M A m permanent magnetization (see also
μ M)
m kg mass of a rigid body (weight)
Δm kg mass difference, usually between test
and reference weight
δm kg maximum permissible error on the
weights
m kg mass, nominal value of the weight
(e.g. 1 kg)
m kg conventional mass of the weight
c
Δm kg conventional mass difference between
c
test weight and reference weight
¯
kg average conventional mass difference
Δm
c
between test weight and reference
weight
m kg conventional mass of the reference
cr
weight
m kg conventional mass of the test weight
ct
m kg mass of the sensitivity weight
s
m kg mass of the test weight
t
n – subscript for number of measurement
sequences
p Pa barometric pressure
R μm mean height of roughness profile (R-
a
parameter)
R μm maximum height of roughness profile
z
(R-parameter)
r – subscript for reference weight
s – subscript for sensitivity weight
s kg standard deviation
2 2
s kg variance
T K thermodynamic temperature using the
International Temperature Scale of
1990 (ITS-90)
ΔT* °C initial difference between weight tem-
perature and laboratory temperature
t – subscript for test weight
t °C temperature in degrees Celsius, where
t = T– 273.15 K
U kg uncertainty, expanded uncertainty
u kg uncertainty, standard uncertainty
u kg uncertainty of air buoyancy correction
b
u kg uncertainty of the balance
ba
u kg combined standard uncertainty
c
u kg uncertainty due to the display resolu-
d
tion of a digital balance
u kg uncertainty due to eccentricity
E
–3
u kg m uncertainty of the formula used to cal-
F
culate air density
u % uncertainty in relative humidity
hr
u kg uncertainty due to instability of the ref-
inst
erence weight
u kg uncertainty due to magnetism
ma
u Pa uncertainty in barometric pressure
p
u kg uncertainty due to the sensitivity of the
s
balance
u °C uncertainty in temperature
t
u kg uncertainty due to the weighing pro-
w
cess
V m volume of a solid body (weight)
z m vertical cartesian coordinate
–2
μ N A magnetic permeability
E617 − 23
Symbol Unit Definition
–2
μ N A magnetic constant (magnetic
permeability of vacuum), μ = 4π ×
–7 –2
10 N A
μ M T magnetic polarization
μ – relative magnetic permeability (μ/μ )
r 0
v – effective degrees of freedom
eff
–3
ρ kg m mass of a rigid body (weight)
–3
ρ kg m density of air as a reference value
–3
equal to 1.2 kg m
–3
ρ kg m density of moist air
a
–3
ρ kg m density of moist air during the last
al
(previous) calibration of the reference
weight
–3
ρ kg m density of a reference weight with
r
mass m
r
–3
ρ kg m density of the weight being tested
t
χ – magnetic susceptibility
4. Maximum Permissible Errors
4.1 For each weight, the expanded uncertainty U at approximately 95 % confidence (See(see Section 9) of the conventional mass
shall be less than or equal to one-third of the maximum permissible error given in Table 1 as defined in Section 9. Subsequent
calibrations must meet all the requirements (including environmental parameters as shown in Table 11, of Sections 7, 8, and 9; and
the requirements of ISO/IEC 17025:2005, 5.10.4.217025:2017, 7.8.6 to make any claim of compliance to Specification E617,
Maximum Permissible Errors, weight classes, or metrological traceability.
4.1.1 For each weight, the conventional mass, m (determined with an expanded uncertainty), shall not differ by more than the
c
difference: maximum permissible error δm minus expanded uncertainty, from the nominal value of the weight, m :
o
m 2 δm 2 U # m # m 1 δm 2 U (1)
~ ! ~ ! ~ !
o c o
4.2 Maximum permissible errors for classes 000, 00, 0, 1, 2, 3, 4, 5, 6, and 7 are given in Table 1. These maximum permissible
errors apply to conventional mass values.
4.3 Maximum Permissible Errors for weights of denomination intermediate between those listed, the maximum permissible error
shall be proportional to the values shown.interpolated from the nearest values shown. Maximum permissible errors for nominal
values outside the ranges listed in Table 1 shall be interpolated from the nearest metric equivalents.
4.4 For class 000, 00, and 0 weights, which are always accompanied by certificates giving the mass values and uncertainties, the
deviation from the nominal value, m – m , shall be taken into account by the user.
c 0
5. Physical Characteristics
5.1 Construction:
5.1.1 Type—Weights are divided into two types based upon the design:
5.1.1.1 Type I—These weights are of one-piece construction and contain no added adjusting material. They must be specified when
weights are to be used as standards for the calibration of weights of Classes 000, 00, 0, 1, 2, and 3, and where maximum stability
is required. A precise measurement of density can only be made for one-piece weights.
5.1.1.2 Type II—Weights of this type can be of any appropriate design such as screw knob, ring, or sealed plug. Adjusting material
can be used provided it is of a material at least as stable as the base material and is contained in such a way that it will not become
separated from the weight.
5.1.2 Class 000, 00, and 0 shall be Type I, one piece construction. Weights with nominal values less than 1 g shall have unique
shapes to differentiate the weights from one another. See Table 2. The shape of weights smaller than 1 mg shall be discussed and
verified with the customer.
5.1.3 Class 1, 2, 3, 4, 5, 6, and 7 may be either Type I or Type II depending on the application.
E617 − 23
TABLE 1 Maximum Permissible Errors
NOTE 1—Maximum Permissible Errors are reported in SI units, typically milligrams.
NOTE 2—The “grain” is the same in avoirdupois, troy, and apothecaries units of mass.
NOTE 3—See NIST SP 811 and NIST SP 1038 for conversion and units of measure.
NOTE 4—For nominal values not listed, see 4.3.
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
Metric
5000 kg 25 g 50 g 100 g 250 g 500 g 750 g
3000 kg 15 g 30 g 60 g 150 g 300 g 450 g
2000 kg 10 g 20 g 40 g 100 g 200 g 300 g
1000 kg 5.0 g 10 g 20 g 50 g 100 g 150 g
500 kg 2.5 g 5.0 g 10 g 25 g 50 g 75 g
300 kg 1.5 g 3.0 g 6.0 g 15 g 30 g 45 g
200 kg 1.0 g 2.0 g 4.0 g 10 g 20 g 30 g
100 kg 500 mg 1.0 g 2.0 g 5.0 g 10 g 15 g
50 kg 13 mg 25 mg 63 mg 120 mg 250 500 mg 1.0 g 2.5 g 5.0 g 7.5 g
30 kg 7.5 15 38 75 150 300 600 mg 1.5 g 3.0 g 4.5 g
30 kg 7.5 15 38 75 150 300 600 mg 1.5 g 3.0 g 5.1 g
25 kg 6.25 12.5 31 62 125 250 500 1.2 g 2.5 g 3.8 g
25 kg 6.25 12.5 31 62 125 250 500 1.3 g 2.5 g 4.5 g
20 kg 5.0 10 25 50 100 200 400 1.0 g 2.0 g 3.8 g
10 kg 2.5 5.0 13 25 50 100 200 500 mg 1.0 g 2.2 g
5 kg 1.3 2.5 6.0 12 25 50 100 250 500 mg 1.4 g
3 kg 0.75 1.5 3.8 7.5 15 30 60 150 300 1.0 g
2 kg 0.5 1.0 2.5 5.0 10 20 40 100 200 750 mg
1 kg 0.25 0.5 1.3 2.5 5.0 10 20 50 100 470
500 g 0.13 0.25 0.60 1.2 2.5 5.0 10 30 50 300
500 g 0.13 0.25 0.60 1.2 2.5 5.0 10 35 70 300
300 g 0.075 0.15 0.38 0.75 1.5 3.0 6.0 20 30 210
300 g 0.075 0.15 0.38 0.75 1.5 3.0 6.0 30 60 210
200 g 0.05 0.10 0.25 0.50 1.0 2.0 4.0 15 20 160
200 g 0.05 0.10 0.25 0.50 1.0 2.0 4.0 20 40 160
100 g 0.025 0.05 0.13 0.25 0.50 1.0 2.0 9.0 10 100
100 g 0.025 0.05 0.13 0.25 0.50 1.0 2.0 10 20 100
50 g 0.015 0.030 0.060 0.12 0.25 0.60 1.2 5.6 7.0 62
50 g 0.015 0.030 0.060 0.12 0.25 0.60 1.2 5.0 10 62
30 g 0.014 0.026 0.037 0.074 0.15 0.45 0.90 4.0 5.0 44
30 g 0.014 0.026 0.037 0.074 0.15 0.45 0.90 3.0 6.0 44
20 g 0.013 0.025 0.037 0.074 0.10 0.35 0.70 3.0 3.0 33
20 g 0.013 0.025 0.037 0.074 0.10 0.35 0.70 2.0 4.0 33
10 g 0.010 0.020 0.025 0.050 0.074 0.25 0.50 2.0 2.0 21
10 g 0.010 0.020 0.025 0.050 0.074 0.25 0.50 1.0 2.0 21
5 g 0.005 0.010 0.017 0.034 0.054 0.18 0.36 1.3 2.0 13
5 g 0.005 0.010 0.017 0.034 0.054 0.18 0.36 0.75 1.5 13
3 g 0.005 0.010 0.017 0.034 0.054 0.15 0.30 0.95 2.0 9.4
3 g 0.005 0.010 0.017 0.034 0.054 0.15 0.30 0.64 1.3 9.4
2 g 0.005 0.010 0.017 0.034 0.054 0.13 0.26 0.75 2.0 7.0
2 g 0.005 0.010 0.017 0.034 0.054 0.13 0.26 0.56 1.1 7.0
1 g 0.005 0.010 0.017 0.034 0.054 0.10 0.20 0.50 2.0 4.5
1 g 0.005 0.010 0.017 0.034 0.054 0.10 0.20 0.45 0.90 4.5
500 mg 0.002 0.003 0.005 0.010 0.025 0.080 0.16 0.38 1.0 3.0
500 mg 0.002 0.003 0.005 0.010 0.025 0.080 0.16 0.36 0.72 3.0
300 mg 0.002 0.003 0.005 0.010 0.025 0.070 0.14 0.30 1.0 2.2
300 mg 0.002 0.003 0.005 0.010 0.025 0.070 0.14 0.31 0.61 2.2
200 mg 0.002 0.003 0.005 0.010 0.025 0.060 0.12 0.26 1.0 1.8
200 mg 0.002 0.003 0.005 0.010 0.025 0.060 0.12 0.27 0.54 1.8
100 mg 0.002 0.003 0.005 0.010 0.025 0.050 0.10 0.20 1.0 1.2
100 mg 0.002 0.003 0.005 0.010 0.025 0.050 0.10 0.22 0.43 1.2
50 mg 0.002 0.003 0.005 0.010 0.014 0.042 0.085 0.16 0.88
50 mg 0.002 0.003 0.005 0.010 0.014 0.042 0.085 0.17 0.35 0.88
30 mg 0.002 0.003 0.005 0.010 0.014 0.038 0.075 0.14 0.68
30 mg 0.002 0.003 0.005 0.010 0.014 0.038 0.075 0.15 0.30 0.68
20 mg 0.002 0.003 0.005 0.010 0.014 0.035 0.070 0.12 0.56
20 mg 0.002 0.003 0.005 0.010 0.014 0.035 0.070 0.13 0.26 0.56
10 mg 0.002 0.003 0.005 0.010 0.014 0.030 0.060 0.10 0.21 0.40
5 mg 0.002 0.003 0.005 0.010 0.014 0.028 0.055 0.080
5 mg 0.002 0.003 0.005 0.010 0.014 0.028 0.055 0.083 0.17
3 mg 0.002 0.003 0.005 0.010 0.014 0.026 0.052 0.070
3 mg 0.002 0.003 0.005 0.010 0.014 0.026 0.052 0.071 0.14
2 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.060
2 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.062 0.12
1 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.050 0.10
E617 − 23
TABLE 1 Continued
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
0.5 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.050
0.5 mg 0.002 0.003 0.005 0.010 0.014 0.025 0.050 0.0540 0.080
0.3 mg 0.002 0.003 0.005 0.010 0.014 0.025
0.2 mg 0.002 0.003 0.005 0.010 0.014
0.1 mg 0.002 0.003 0.005 0.010
0.05 mg 0.002 0.003 0.005 0.010
Avoirdupois Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
Pound mg mg mg g & mg g & mg g & mg g & mg g & mg
10000 lb 45 g 91 g 230 g 450 g 680 g
5000 lb 23 g 45 g 110 g 230 g 340 g
3000 lb 14 g 27 g 68 g 140 g 200 g
2500 lb 11 g 23 g 57 g 110 g 170 g
2000 lb 9.1 g 18 g 45 g 91 g 140 g
1000 lb 4.5 g 9.1 g 23 g 45 g 68 g
500 lb 2.3 g 4.5 g 11 g 23 g 34 g
300 lb 1.4 g 2.7 g 6.8 g 14 g 20 g
200 lb 910 mg 1.8 g 4.5 g 9.1 g 14 g
100 lb 57 110 230 450 910 mg 2.3 g 4.5 g 6.8 g
50 lb 29 57 110 230 450 1.1 g 2.3 g 4.1 g
30 lb 17 32 68 140 270 680 mg 1.4 g 2.7 g
25 lb 14 28 57 110 230 570 1.1 g 2.4 g
20 lb 12 23 45 91 180 450 910 mg 2.0 g
10 lb 5.5 11 23 45 91 230 450 1.3 g
5 lb 2.7 5.4 11 23 45 110 230 780 mg
3 lb 1.7 3.4 6.8 14 27 68 140 580
2 lb 1.2 2.3 4.5 9.1 18 45 91 440
1 lb 0.55 1.1 2.3 4.5 9.1 23 45 270
1 lb 0.55 1.1 2.3 4.5 9.1 35 70 270
0.5 lb 0.27 0.54 1.1 2.3 4.5 16 23 170
0.5 lb 0.27 0.54 1.1 2.3 4.5 23 45 170
0.3 lb 0.17 0.34 0.68 1.4 2.7 11 14 120
0.3 lb 0.17 0.34 0.68 1.4 2.7 14 27 120
0.2 lb 0.12 0.23 0.45 0.91 1.8 8.2 9.1 97
0.2 lb 0.12 0.23 0.45 0.91 1.8 9.0 18 97
0.1 lb 0.055 0.11 0.23 0.45 1.1 5.4 6.8 59
0.1 lb 0.055 0.11 0.23 0.57 1.1 4.5 9.1 59
0.05 lb 0.027 0.054 0.11 0.36 0.77 3.3 4.5 37
0.05 lb 0.037 0.074 0.11 0.36 0.77 2.3 4.5 37
0.03 lb 0.017 0.034 0.068 0.32 0.59 2.4 3.2 26
0.03 lb 0.029 0.059 0.083 0.32 0.59 1.4 2.7 26
0.02 lb 0.017 0.034 0.045 0.23 0.45 1.9 2.3 20
0.02 lb 0.024 0.047 0.070 0.23 0.45 0.91 1.8 20
0.01 lb 0.012 0.023 0.034 0.16 0.34 1.2 1.4 12
0.01 lb 0.017 0.034 0.054 0.16 0.34 0.73 1.5 12
0.005 lb 0.0075 0.015 0.024 0.14 0.27 0.86 0.91 7.8
0.005 lb 0.017 0.034 0.054 0.14 0.27 0.58 1.2 7.8
0.003 lb 0.0075 0.015 0.024 0.11 0.22 0.64 0.91 5.4
0.003 lb 0.017 0.034 0.054 0.11 0.22 0.50 0.99 5.4
0.002 lb 0.0075 0.015 0.024 0.091 0.19 0.50 0.91 4.2
0.002 lb 0.015 0.030 0.049 0.091 0.19 0.44 0.87 4.2
0.001 lb 0.0075 0.015 0.024 0.068 0.15 0.36 0.91 2.9
0.001 lb 0.0050 0.010 0.024 0.078 0.15 0.35 0.70 2.9
0.0005 lb 0.0045 0.0091 0.023 0.064 0.13 0.27 2.1
0.0005 lb 0.0045 0.0091 0.023 0.064 0.13 0.28 0.56 2.0
0.0003 lb 0.0045 0.0091 0.023 0.054 0.11 0.23 1.5
0.0002 lb 0.0045 0.0091 0.023 0.045 0.095 0.20 1.2
0.0001 lb 0.0045 0.0091 0.023 0.041 0.086 0.16 0.84
0.00005 lb 0.0045 0.0091 0.014 0.036 0.073 0.14
0.00003 lb 0.0045 0.0091 0.014 0.032 0.064 0.11
0.00002 lb 0.0045 0.0091 0.014 0.029 0.059 0.091
0.00001 lb 0.0045 0.0091 0.014 0.027 0.054 0.091
0.000005 lb 0.0045 0.0091 0.014 0.023 0.054 0.091
0.000003 lb 0.0045 0.0091 0.014 0.023 0.054 0.091
0.000002 lb 0.0045 0.0091 0.014 0.023 0.054 0.091
0.000001 lb 0.0045 0.0091 0.014 0.023 0.054 0.091
Avoirdupois Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
Ounce mg mg mg mg mg mg mg mg
10 oz 0.35 0.70 1.4 2.8 5.4 19 23 200
10 oz 0.35 0.70 1.4 2.8 5.4 28 57 200
8 oz 027 0.54 1.1 2.3 4.5 16 23 170
8 oz 0.27 0.54 1.1 2.3 4.5 23 45 170
5 oz 0.18 0.35 0.71 1.4 2.8 12 11 130
5 oz 0.18 0.35 0.71 1.4 2.8 14 28 130
4 oz 0.14 0.28 0.57 1.1 2.3 9.5 11 110
E617 − 23
TABLE 1 Continued
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
4 oz 0.14 0.28 0.57 1.1 2.3 11 23 110
3 oz 0.10 0.21 0.43 0.91 1.8 8.2 5.4
3 oz 0.10 0.21 0.43 0.91 1.8 8.5 17 90
2 oz 0.070 0.14 0.28 0.64 1.3 5.9 5.4 70
2 oz 0.070 0.14 0.28 0.64 1.3 5.7 11 70
1 oz 0.027 0.054 0.14 0.43 0.86 3.9 3.2 42
1 oz 0.037 0.074 0.14 0.43 0.86 2.8 5.4 42
1/2 oz 0.017 0.034 0.071 0.30 0.59 2.5 2.3 27
1/2 oz 0.030 0.060 0.085 0.30 0.59 1.4 2.8 27
1/4 oz 0.011 0.023 0.034 0.20 0.43 1.6 1.4 17
1/4 oz 0.020 0.041 0.062 0.20 0.43 0.84 1.7 17
1/8 oz 0.0075 0.015 0.024 0.16 0.31 1.1 0.91 10
1/8 oz 0.017 0.034 0.054 0.16 0.31 0.67 1.3 10
1/16 oz 0.0075 0.015 0.024 0.12 0.24 0.73 0.91 6.5
1/16 oz 0.017 0.034 0.054 0.12 0.24 0.50 1.1 6.5
1/32 oz 0.0075 0.015 0.024 0.095 0.19 0.50 4.2
1/32 oz 0.014 0.029 0.047 0.095 0.19 0.43 0.87 4.2
1/64 oz 0.0075 0.015 0.023 0.077 0.15 0.36 2.8
1/64 oz 0.0050 0.010 0.023 0.077 0.15 0.35 0.69 2.8
0.5 oz 0.071 0.30 0.59 2.5 2.3 27
0.5 oz 0.030 0.060 0.085 0.30 0.59 1.4 2.8 27
0.3 oz 0.034 0.23 0.45 1.8 1.4 19
0.3 oz 0.068 0.23 0.45 0.89 1.8 19
0.2 oz 0.034 0.19 0.38 1.4 0.91 14
0.2 oz 0.057 0.19 0.38 0.78 1.6 14
0.1 oz 0.024 0.14 0.29 0.91 0.91 9.0
0.1 oz 0.054 0.14 0.29 0.63 1.3 9.0
0.05 oz 0.024 0.11 0.23 0.64 0.91 5.7
0.05 oz 0.054 0.11 0.23 0.50 1.0 5.7
0.03 oz 0.024 0.095 0.19 0.45 0.91 4.1
0.03 oz 0.045 0.095 0.19 0.43 0.85 4.1
0.02 oz 0.024 0.077 0.18 0.40 0.91 3.2
0.02 oz 0.029 0.077 0.18 0.38 0.75 3.2
0.01 oz 0.023 0.064 0.14 0.30 0.91 2.2
0.01 oz 0.023 0.064 0.14 0.30 0.60 2.2
0.005 oz 0.023 0.054 0.11 0.23 1.5
0.003 oz 0.023 0.050 0.095 0.19 1.1
0.002 oz 0.023 0.044 0.086 0.16 0.92
0.001 oz 0.023 0.038 0.077 0.13 0.66
0.0005 oz 0.014 0.032 0.064 0.11 0.47
0.0003 oz 0.014 0.029 0.059 0.095
0.0002 oz 0.014 0.027 0.054 0.086
0.0001 oz 0.014 0.026 0.050 0.073
Troy Ounce Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg mg g & mg
1000 oz t 310 mg 620 mg 1.6 g 3.1 g
500 oz t 160 310 780 mg 1.6 g
300 oz t 91 190 470 930 mg
200 oz t 62 120 310 620
100 oz t 31 62 160 310
50 oz t 16 31 78 160
30 oz t 9.1 19 47 93
20 oz t 6.2 12 35
20 oz t 6.2 12 39 77
10 oz t 3.1 6.2 21
10 oz t 3.1 6.2 30 61
5 oz t 1.6 3.1 12
5 oz t 1.6 3.1 16 31
3 oz t 0.91 1.9 8.4
3 oz t 0.91 1.9 9.3 19
2 oz t 0.71 1.4 6.5
2 oz t 0.71 1.4 6.2 12
1 oz t 0.45 0.91 4.2
1 oz t 0.45 0.91 3.1 6.2
0.5 oz t 0.31 0.62 2.6
0.5 oz t 0.31 0.62 1.6 3.1
0.3 oz t 0.24 0.49 1.9
0.3 oz t 0.24 0.49 0.92 1.8
0.2 oz t 0.20 0.40 1.5
0.2 oz t 0.20 0.40 0.80 1.6
0.1 oz t 0.15 0.30 0.97
0.1 oz t 0.15 0.30 0.65 1.3
0.05 oz t 0.12 0.23 0.65
0.05 oz t 0.12 0.23 0.52 1.0
E617 − 23
TABLE 1 Continued
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
0.03 oz t 0.097 0.19 0.49
0.03 oz t 0.097 0.19 0.44 0.88
0.02 oz t 0.084 0.17 0.41
0.02 oz t 0.084 0.17 0.39 0.77
0.01 oz t 0.071 0.14 0.31 0.62
0.005 oz t 0.056 0.11 0.23
0.005 oz t 0.056 0.11 0.25 0.50
0.003 oz t 0.049 0.097 0.19
0.002 oz t 0.044 0.091 0.17
0.001 oz t 0.038 0.078 0.14
0.0005 oz t 0.033 0.065 0.11
0.0003 oz t 0.030 0.060 0.097
0.0002 oz t 0.028 0.056 0.084
0.0001 oz t 0.026 0.052 0.071
Pennyweight Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg g & mg g & mg
mg mg mg g & mg
10000 dwt 155 0.31 g 0.78 g
10000 dwt 155 310 780 1.6 g
5000 dwt 78 0.16 g 0.39 g
5000 dwt 78 160 390 780 mg
3000 dwt 47 91 mg 0.23 g
3000 dwt 47 91 230 470
2000 dwt 31 62 0.16 g
2000 dwt 31 62 160 310
1000 dwt 16 31 78 mg
1000 dwt 16 31 78 160
500 dwt 7.8 16 41
500 dwt 7.8 16 43 87
300 dwt 4.7 9.1 28
300 dwt 4.7 9.1 34 68
200 dwt 3.1 6.2 21
200 dwt 3.1 6.2 30 61
100 dwt 1.6 3.1 12
100 dwt 1.6 3.1 16 31
50 dwt 0.78 1.6 7.8 16
30 dwt 0.58 1.2 5.3
30 dwt 0.58 1.2 4.7 9.3
20 dwt 0.46 0.91 4.2
20 dwt 0.46 0.91 3.1 6.2
10 dwt 0.31 0.62 2.6
10 dwt 0.31 0.62 1.6 3.1
5 dwt 0.22 0.44 0.89 1.7
3 dwt 0.17 0.34 1.3
3 dwt 0.17 0.34 0.73 1.5
2 dwt 0.15 0.3 0.97
2 dwt 0.15 0.3 0.65 1.3
1 dwt 0.12 0.23 0.65
1 dwt 0.12 0.23 0.52 1.0
Grain Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg mg mg mg
10000 gr 6.5 13 36 64
10000 gr 6.5 13 36 79
5000 gr 3.2 6.5 22 32
5000 gr 3.2 6.5 31 61
3000 gr 1.9 3.9 15 20
3000 gr 1.9 3.9 19 39
2000 gr 1.3 2.6 11 14
2000 gr 1.3 2.6 13 26
1000 gr 0.71 1.4 6.5 6.4
1000 gr 0.71 1.4 6.5 13
500 gr 0.48 0.91 4.3 3.2
500 gr 0.48 0.91 3.2 6.5
300 gr 0.35 0.65 3.0 2.3
300 gr 0.35 0.65 1.9 3.9
200 gr 0.28 0.57 2.3 1.4
200 gr 0.28 0.57 1.3 2.6
100 gr 0.20 0.40 1.5 0.91
100 gr 0.20 0.40 0.82 1.6
50 gr 0.15 0.30 0.97 0.91
50 gr 0.15 0.30 0.65 1.3
30 gr 0.12 0.25 0.71 0.91
30 gr 0.12 0.25 0.55 1.1
20 gr 0.11 0.21 0.58 0.91
E617 − 23
TABLE 1 Continued
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
20 gr 0.11 0.21 0.48 0.91
10 gr 0.084 0.17 0.42 0.91
10 gr 0.084 0.17 0.42 0.77
5 gr 0.071 0.14 0.31 0.45
5 gr 0.071 0.14 0.31 0.62
3 gr 0.060 0.12 0.25 0.45
3 gr 0.060 0.12 0.25 0.53
2 gr 0.054 0.11 0.22 0.45
1 gr 0.045 0.091 0.17 0.45
1 gr 0.045 0.091 0.17 0.37
0.5 gr 0.039 0.078 0.14 0.45
0.5 gr 0.039 0.078 0.14 0.31
0.3 gr 0.035 0.071 0.12
0.2 gr 0.032 0.064 0.11
0.1 gr 0.028 0.056 0.091
0.05 gr 0.026 0.052 0.071
0.03 gr 0.025 0.051 0.071
0.02 gr 0.025 0.050 0.071
0.01 gr 0.025 0.050 0.071
0.005 gr 0.025 0.050 0.050
0.003 gr 0.025 0.050 0.050
0.002 gr 0.025 0.050 0.050
0.001 gr 0.025 0.050 0.050
Carat Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
mg mg
5000 c 10 20
3000 c 6.0 12
2000 c 4.0 8.0
1000 c 2.0 4.0
500 c 1.0 2.0
300 c 0.69 1.3
200 c 0.52 1.0
100 c 0.35 0.7
50 c 0.25 0.50
30 c 0.19 0.40
20 c 0.16 0.33
10 c 0.13 0.26
5 c 0.10 0.20
3 c 0.086 0.17
2 c 0.075 0.15
1 c 0.060 0.12
0.5 c 0.050 0.10
0.3 c 0.044 0.089
0.2 c 0.040 0.080
0.1 c 0.035 0.070
0.05 c 0.030 0.060
0.03 c 0.028 0.056
0.02 c 0.027 0.053
0.01 c 0.025 0.050
Apothecary Ounce mg
12 oz ap 45
12 oz ap 64
10 oz ap 36
10 oz ap 61
6 oz ap 23
6 oz ap 37
5 oz ap 18
5 oz ap 31
4 oz ap 16
4 oz ap 25
3 oz ap 11
3 oz ap 19
2 oz ap 9.1
2 oz ap 12
1 oz ap 4.5
1 oz ap 6.2
Apothecary Dram mg
6 dr ap 3.6
6 dr ap 4.7
5 dr ap 2.7
5 dr ap 3.9
4 dr ap 2.3
4 dr ap 3.1
3 dr ap 1.8
E617 − 23
TABLE 1 Continued
±mg except as noted
Denomination
Class 000 Class 00 Class 0 Class 1 Class 2 Class 3 Class 4 Class 5 Class 6 Class 7
3 dr ap 2.3
2 dr ap 1.8
1 dr ap 1.4
Apothecary mg
Scruple
2 s ap 1.4
2 s ap 1.2
1 s ap 0.91
5.2 Design—A weight may have any shape that does not introduce features that reduce the reliability. Weights with nominal values
less than 1 g shall have unique shapes to differentiate the weights from one another. See Table 2. The shape of weights smaller
than 1 mg shall be discussed and verified with the customer. All weights shall be free of ragged or sharp edges or ends. Both sheet
metal and wire weights shall be free of cracks such as may be formed from bending.
5.3 Surface Area—For classes 000, 00, 0, 1, 2, 3, and 4 the surface area is not to exceed twice the area of a cylinder of equal height
and diameter for weights 1 g and above. Sheet metal weights or wire weights may be used below 1 g. For Classes 5, 6, and 7 the
total surface areas should be minimized to the extent possible.
5.4 Material:
5.4.1 Class 000, 00, 0, 1, 2, 3, 4, and 5 Weights—The hardness of this material and its resistance to wear and corrosion shall be
Shall be made of austenitic stainless steel or a material whose hardness and resistance to corrosion is similar to or better than that
of austenitic stainless steel.
5.4.2 Class 6 Weights—Class 6 weights below 5 kg shall be made of austenitic stainless steel or a material whose hardness and
resistance to corrosion is similar to or better than that of austenitic stainless steel. Class 6 weights of 5 kg or greater shall be made
of grey cast iron or of another material whose brittleness and resistance to corrosion is similar to or better than that of grey cast
iron. The surface of the weights may be treated with a suitable coating in order to improve their corrosion resistance. This coating
shall withstand shocks and outdoor weather conditions.
5.4.3 Classes 6 and 7—Class 7 Weights—Cylindrical class 6 and Class 7 weights below 5 kg and class 6 and 7 weights below 100
g shall be made of austenitic stainless steel or a material whose hardness and resistance to corrosion is similar to or better than
that of steel. Other class 6 and austenitic stainless steel. Class 7 weights of 5 kg100 g or greater shall be made of grey cast iron
or of another material whose brittleness and resistance to corrosion is similar to or better than that of grey cast iron. The surface
of the weights may be treated with a suitable coating in order to improve their corrosion resistance. This coating shall withstand
shocks and outdoor weather conditions.
5.5 Magnetism—Weights shall not exceed maximum permissible magnetic properties as listed in Tables 3 and 4 for any portion
of the weight. If the values of all local measurements of magnetization and susceptibility are less than these limits, then it may
be assumed that the uncertainty components due to the magnetism of the weight are negligible. The maximum permanent
magnetization and magnetic susceptibilities given in Tables 3 and 4 are such that, at magnetic fields and magnetic field gradients
possibly present on balance pans, they produce a change of the conventional mass of less than 1/10 one tenth of the maximum
permissible error of the test weight.
NOTE 3—Magnetic susceptibility may be tested in accordance with OIML R 111-1, Annex B. Cast iron cannot have a susceptibility specification of any
real value.
TABLE 2 Shape of Weights 1 g or Less
Nominal Values Polygonal Sheets Wires
5, 50, 500 mg Pentagon Pentagon
3, 30, 300 mg Circle Circle
2, 20, 200 mg Square Square
1, 10, 100, 1000 mg Triangle Triangle
E617 − 23
TABLE 3 Maximum Polarization, μ M, (μT)
000, 00,
Weight Class 1 2 and 3 4 and 5 6 and 7
and 0
Maximum
Not
polarization, 2.5 8 25 80
applicable
μ M, (μT)
TABLE 4 Maximum Magnetic Susceptibility, χ
000, 00,
Weight Class 1 2 and 3 4 and 5 6 and 7
and 0
m # 1 g 0.25 0.9 10 Not Not
applicable applicable
2 g # m # 10 g 0.06 0.18 0.7 4 Not
applicable
20 g # m 0.02 0.07 0.2 0.8 Not
applicable
5.6 Density—Because of the effect of the buoyant force of air on a weight, precision measurements of mass require that the volume
of the weight be known, as well as the density of the air in which it is being measured, so that appropriate corrections can be made.
For weights of higher precision, the range of density is limited to values at or near the density of well-established standards, such
as are used by primary calibration laboratories. For Class 000 and 00, the manufacturer shall provide a measured value for the
density of the weights at time of manufacture. Use of a sample taken adjacent to the material from which the weight is
–5
manufactured to measure the density is permitted, however an additional uncertainty component equal to 5 × 10 must be
combined with the standard uncertainty of the density determination process. For subsequent calibrations, the density may be
verified by the calibrating laboratory if there are concerns and capability for the measurement is available. As lower precision of
measurement is required, so the range of density is broadened. See Table 5.
5.6.1 The determination of the minimum and maximum density limits for nominal values not listed in Table 5 shall be converted
to metric values and the limits of the metric value next greater than the converted value used. (Example: 1 apothecary ounce class
4 is equal to 31.1034768 g, therefore the density limits are equal to values listed in Table 5 for 50 g.)
NOTE 4—Materials used to make weights for special applications that do not fall within the density limits stated above, should have stated densities or
density determinations performed.
5.7 Finish—The surface of the weights (including the base and corners) shall be smooth, the edges shall be rounded, and the
weights shall not be porous.
5.7.1 The surface quality of a weight shall not exceed maximum values of surface roughness, R and R through visual inspection
a z
using a hand held gage. See Table 6.
TABLE 5 Minimum and Maximum Limits for Density
3 –3
ρ , ρ (10 kg m )
min max
Nominal
Class of Weight
Value
000 00 and 0 1 2 and 3 4 and 5 6 and 7
$ 100 g 7.967 – 8.033 7.934 – 8.067 7.81 – 8.21 7.39 – 8.73 6.4 – 10.7 $ 4.4
50 g 7.960 – 8.04 7.92 – 8.08 7.74 – 8.28 7.27 – 8.89 6.0 – 12.0 $ 4.0
50 g 7.960 – 8.040 7.92 – 8.08 7.74 – 8.28 7.27 – 8.89 6.0 – 12.0 $ 4.0
20 g 7.900 – 8.09 7.84 – 8.17 7.50 – 8.57 6.6 – 10.1 4.8 – 24.0 $ 2.6
20 g 7.900 – 8.090 7.84 – 8.17 7.50 – 8.57 6.6 – 10.1 4.8 – 24.0 $ 2.6
10 g 7.87 – 8.14 7.74 – 8.28 7.27 – 8.89 6.0 – 12.0 $ 4.0 $ 2.0
5 g 7.87 – 8.13 7.62 – 8.42 6.9 – 9.6 5.3 – 16.0 $ 3.0
2 g 7.62 – 8.42 7.27 – 8.89 6.0 – 12.0 $ 4.0 $ 2.0
1 g 7.28 – 8.90 6.9 – 9.6 5.3 – 16.0 $ 3.0
500 mg 7.50 – 8.60 6.3 – 10.9 $ 4.4 $ 2.2
200 mg 6.7 – 9.6 5.3 – 16.0 $ 3.0
100 mg 6.0 – 12 $ 4.4
50 mg $ 4.8 $ 3.4
20 mg $ 2.5 $ 2.3
E617 − 23
TABLE 6 Maximum Values of Surface Roughness
Classes 000, Classes Classes Class Classes
00, 0 1, 2 3, 4 5 6, 7
R (μm) 0.1 0.2 0.4 1 25
A
R (μm) 0.5 1 2 5 100
Z
5.7.2 The surface roughness of all weight surfaces must meet the limits established in Table 6 for the weight classification.
5.8 Adjustment:
5.8.1 Type I Weights—Weights shall be adjusted by abrasion, grinding or any appropriate method. The surface requirements shall
be met at the end of the adjustment process.
5.8.2 Type II Weights—Weights with adjusting cavities shall be adjusted with the same material from which they are made, or with
materials that are at least as stable and of similar density as the base material. For weights which have sealing caps, the cap may
be made of aluminum. The back-up spacer should be of a similar material as the weight. Adjusting material and back-up disc must
meet the magnetic requirements specified for the accuracy class of the weight.
5.9 Marking:
5.9.1 Class 000, 00, and 0 weights shall not bear any indication of nominal value and shall not be marked unless used to
distinguish from another class 000, 00, or 0 weight, provided that the surface quality and stability of the weights are not affected
by the markings or by the process used to mark it.
5.9.2 Numerical Value for Classes 1, 2, 3, 4, 5, 6, and 7—The nominal value of each weight shall appear on the surface of each
weight. Only the numerical portion of the weight value needs to be on the surface of weights. Weights made of wire or too small
to be marked shall not be marked but should be identifiable by their shape or number of bends.
5.9.3 Units of Weight—Weights 100 g and greater may be marked with the unit name or abbreviation. In the case of sets of
non-metric weights, at least the largest weight of a particular set should be marked with the unit name or abbreviation. In any case
the unit shall not be included where such marking would be illegible.
5.9.4 Abbreviations—The accepted abbreviation may be used in marking. Abbreviations are shown in Appendix X2. Periods shall
not be used with abbreviations in marking weights.
5.9.5 Multiple Weights—Multiple weights of the same nominal value included in a set of weights shall have distinguishing marks.
5.9.6 Depth of Markings—Markings shall be clear, shallow, relatively broad, and free of burrs and sharp angles. Markings shall
not perforate or crack sheet metal weights.
5.9.7 User Marking—It is recommended for a user to clearly identify individual weights as it helps to link a weight to its
calibration certificate or verification document. The recommended maximum values for user markings are given in Table 7.
TABLE 7 Recommended Maximum Number of User Markings
Maximum Number
Nominal Height of
Class of Signs, Num
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...