1.1 This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. NOTE Although the scope is limited to medical devices, this document can be applicable to other products and equipment. This document covers radiation processes employing irradiators using: a) the radionuclide 60Co or 137Cs; b) a beam from an electron generator; or c) a beam from an X-ray generator. 1.2 This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. NOTE For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A. 1.2.1 This document does not specify requirements for designating a medical device as sterile. NOTE Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67. 1.2.2 This document does not specify a quality management system for the control of all stages of production of medical devices. NOTE It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, REF Section_sec_4 \r \h Clause 4 08D0C9EA79F9BACE118C8200AA004BA90B02000000080000000E000000530065006300740069006F006E005F007300650063005F0034000000 ). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party. 1.2.3 This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release. 1.2.4 This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities. NOTE Regulations on safety requirements for occupational safety related to radiation can exist in some countries. 1.2.5 This document does not specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    39 pages
    English language
    sale 15% off
  • Standard
    41 pages
    French language
    sale 15% off

This document describes a method for substantiating a selected sterilization dose of 17,5 kGy, 20 kGy, 22,5 kGy, 27,5 kGy, 30 kGy, 32,5 kGy or 35 kGy that achieves a sterility assurance level (SAL) of 10−6 or less for radiation sterilization of health care products. This document also specifies a method of sterilization dose audit used to demonstrate the continued effectiveness of the substantiated sterilization dose. NOTE 1 Selection and substantiation of the sterilization dose is used to meet the requirements for establishing the sterilization dose within process definition in ISO 11137‑1. This document does not apply to other sterilization doses than the substantiation of a selected sterilization dose of 17,5 kGy, 20 kGy, 22,5 kGy, 27,5 kGy, 30 kGy, 32,5 kGy or 35 kGy. The method is not used for the substantiation of a selected sterilization dose if the average bioburden of the entire product item exceeds the limit specified for the selected sterilization dose (see Table 3). NOTE 2 The methods for substantiation of selected sterilization doses of 25 kGy and 15 kGy are not included in this document. They are described in ISO 11137‑2. If the decision is made to use this method of sterilization dose establishment, the method is intended to be followed in accordance with the requirements (shall) and guidance (should) stipulated herein.

  • Standard
    54 pages
    English language
    sale 15% off
  • Standard
    56 pages
    French language
    sale 15% off

This document provides additional guidance to that given in ISO 11137-3 on meeting the requirements specified in ISO 11137-1, ISO 11137-2 and ISO/TS 13004 for the establishment and control of a radiation sterilization process using gamma, electron beam, and X-irradiation.

  • Technical specification
    55 pages
    English language
    sale 15% off
  • Technical specification
    55 pages
    English language
    sale 15% off

ISO 11137-3:2017 gives guidance on meeting the requirements in ISO 11137-1 and ISO 11137-2 and in ISO/TS 13004 relating to dosimetry and its use in development, validation and routine control of a radiation sterilization process.

  • Standard
    41 pages
    English language
    sale 15% off
  • Standard
    45 pages
    French language
    sale 15% off

ISO 11137-2:2013 specifies methods for determining the minimum dose needed to achieve a specified requirement for sterility and methods to substantiate the use of 25 kGy or 15 kGy as the sterilization dose to achieve a sterility assurance level, SAL, of 10−6. ISO 11137-2:2013 also specifies methods of sterilization dose audit used to demonstrate the continued effectiveness of the sterilization dose. ISO 11137-2:2013 defines product families for sterilization dose establishment and sterilization dose audit.

  • Standard
    68 pages
    English language
    sale 15% off
  • Standard
    71 pages
    French language
    sale 15% off

ISO 11137-2:2011 specifies methods of determining the minimum dose needed to achieve a specified requirement for sterility and methods to substantiate the use of 25 kGy or 15 kGy as the sterilization dose to achieve a sterility assurance level, SAL, of 10-6. ISO 11137-2:2011 also specifies methods of sterilization dose audit used to demonstrate the continued effectiveness of the sterilization dose. ISO 11137-2:2011 defines product families for sterilization dose establishment and sterilization dose audit.

  • Standard
    70 pages
    English language
    sale 15% off
  • Standard
    76 pages
    French language
    sale 15% off

ISO 11137-1:2006 specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. Although the scope of ISO 11137-1:2006 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment. ISO 11137-1:2006 covers radiation processes employing irradiators using the radionuclide 60Co or 137Cs, a beam from an electron generator or a beam from an X-ray generator. ISO 11137-1:2006 does not: specify requirements for development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease; detail specified requirements for designating a medical device as sterile; specify a quality management system for the control of all stages of production of medical devices; specify requirements for occupational safety associated with the design and operation of irradiation facilities; specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    37 pages
    English language
    sale 15% off
  • Standard
    39 pages
    French language
    sale 15% off

ISO 11137-3:2006 gives guidance on the requirements in ISO 11137 parts 1 and 2 relating to dosimetry. Dosimetry procedures related to the development, validation and routine control of a radiation sterilization process are described.

  • Standard
    15 pages
    English language
    sale 15% off
  • Standard
    17 pages
    French language
    sale 15% off

ISO 11137-2:2006 specifies methods of determining the minimum dose needed to achieve a specified requirement for sterility and methods to substantiate the use of 25 kGy or 15 kGy as the sterilization dose to achieve a sterility assurance level (SAL) of 10-6. It also specifies methods of dose auditing in order to demonstrate the continued effectiveness of the sterilization dose. ISO 11137-2:2006 defines product families for dose establishment and dose auditing.

  • Standard
    60 pages
    English language
    sale 15% off
  • Standard
    62 pages
    French language
    sale 15% off

ISO/TS 13409:2002 describes a method of substantiating the suitability of 25 kGy as a sterilization dose for radiation sterilization of products with an average bioburden of less than 1 000 colony-forming units (cfu) that are manufactured in small quantities (less than 1 000 product units). This method may be used to substantiate a sterilization dose of 25 kGy for any of the following situations: a single batch of product units; initial production of a new product while the sterilization dose is being established by another method; routine production of small batches. Information collected in applying the method of dose substantiation described in this Technical Specification may be applicable in meeting the product qualification requirements for sterilization dose selection of ISO 11137:1995.

  • Technical specification
    12 pages
    English language
    sale 15% off
  • Technical specification
    12 pages
    French language
    sale 15% off
  • Standard
    1 page
    English language
    sale 15% off
  • Standard
    1 page
    French language
    sale 15% off
  • Standard
    1 page
    French language
    sale 15% off

Describes a method of substantiating the suitability of 25 kGy as a sterilization dose for radiation sterilization of health care products with an average bioburden of less than 1 000 colony forming units (cfu), that are manufactured in small quantities.

  • Technical report
    19 pages
    English language
    sale 15% off
  • Technical report
    14 pages
    French language
    sale 15% off
  • Technical report
    14 pages
    French language
    sale 15% off

Specifies requirements for validation, process control and routine monitoring in the radiation sterilization of health care products. Applies to continuous and batch type gamma irradiators using the radionuclides 60 Co and 137 Cs, and to irradiators using a beam from an electron or x-ray generator. Does not cover facility design, licensing, operator training, factors related to radiation safety, the assessment of the suitability of the product for its intended use.

  • Standard
    61 pages
    English language
    sale 15% off
  • Standard
    61 pages
    French language
    sale 15% off
  • Standard
    61 pages
    French language
    sale 15% off

Frequently Asked Questions

ISO/TC 198/WG 2 is a Working Group within the International Organization for Standardization (ISO). It is named "Radiation sterilization". This committee has published 17 standards.

ISO/TC 198/WG 2 develops ISO standards. Currently, there are 17 published standards from this working group.

The International Organization for Standardization (ISO) is an independent, non-governmental international organization that develops and publishes international standards. Founded in 1947 and headquartered in Geneva, Switzerland, ISO brings together experts from 170+ member countries to share knowledge and develop voluntary, consensus-based standards that support innovation and provide solutions to global challenges.

A Working Group in ISO is a specialized group responsible for developing standards or technical work within a defined scope. These bodies bring together international experts to create consensus-based standards that support global trade, safety, and interoperability.