SIST EN 13925-3:2005
(Main)Non destructive testing - X ray diffraction from polycrystalline and amorphous materials - Part 3: Instruments
- Abstract
This document sets out the characteristics of instruments used for X-ray powder diffraction ("powder" as defined in EN 13925-1:2003, Clause 5) as a basis for their control and hence quality assurance of the measurements made by this technique. Performance testing indicators are given for diffractometer performance testing. Different types and makes of X-ray powder diffractometer vary considerably in their design and intended fields of application. This document attempts to cover as much of this range as possible by keeping to common principles. To make the standard more readily applicable, the Bragg-Brentano configuration is addressed in most detail because of its wide use. Additional considerations and adaptations may be necessary to cover some types of instruments or configuration and some fields of application. Some of these types of instrument are described in Annex B.
- Status
- Published
- Publication Date
- 30-Jun-2005
- Technical Committee
- PKG - Testing of metallic materials
- Current Stage
- 6060 - National Implementation/Publication (Adopted Project)
- Start Date
- 01-Jul-2005
- Due Date
- 01-Jul-2005
- Completion Date
- 01-Jul-2005
SIST EN 13925-3:2005 is the Slovenian adoption of EN 13925-3:2005 from CEN for non destructive testing by X-ray powder diffraction (XRPD) on polycrystalline and amorphous materials. It specifies the characteristics of XRPD instruments, with special attention to Bragg-Brentano diffractometers, so laboratories can control equipment, align and calibrate it, and monitor measurement quality.
What does SIST EN 13925-3:2005 specify?
SIST EN 13925-3:2005 sets out the characteristics of instruments used for X-ray powder diffraction, using those characteristics as the basis for instrument control and quality assurance of measurements. The scope covers a wide range of diffractometer designs and keeps to common principles, while treating Bragg-Brentano geometry in more detail because it is widely used.
The document is organized around the equipment description, traceable characterisation data, alignment and calibration, and performance monitoring.
| Annex | What it covers |
|---|---|
| Annex A | Relationship between the XRPD standards |
| Annex B | Alignment of Bragg-Brentano diffractometers |
| Annex C | Procedures for instrument performance characterisation |
| Annex D | Sample report forms for characterisation of instruments |
Clause 4 describes the instrument parts, Clause 5 lists the characteristics to record, Clause 6 covers alignment and calibration, and Clause 7 covers performance testing and monitoring.
What are the key requirements of SIST EN 13925-3:2005?
SIST EN 13925-3:2005 requires the XRPD setup to be described and controlled as a complete measuring system, not as isolated parts. In practice, this means the source, optics, detector, goniometer, specimen stage and data collection system all have to be identified and their relevant characteristics recorded.
The instrument has to be described by its main components
Clause 4 describes a diffractometer as a combination of goniometer, X-ray source, incident beam optics, diffracted beam optics, specimen stage, detector and data collection system. The clause also notes that data processing belongs with EN 13925-2. For users, this means the measurement result depends on both the hardware and the way the data are processed, so both have to be controlled.
Source, optics and detector choices affect line position, shape and intensity
Clause 4 distinguishes conventional X-ray sources, synchrotron radiation sources, monochromators, filters, multilayer mirrors, slits, collimators and capillary optics. The document explains why these choices matter: they control wavelength purity, beam size, divergence, background, intensity and angular resolution. For example, filters can create background steps near diffraction lines, and detectors with energy discrimination need dead-time control when count rates rise.
Specimen geometry and goniometer motion have to match the analysis
Clause 4 and 4.5 describe flat, rod-shaped and irregular specimens, together with the relevant specimen motions such as omega (ω), phi (φ) and chi (χ). Bragg-Brentano geometry places the apparent source, specimen and receiving slit on the focusing circle, while Seeman-Bohlin uses a different relationship between the circles. In practice, this means specimen mounting and motion are not generic - they depend on the analysis type, and irregular or curved specimens often work better with parallel-beam geometry.
Traceable characteristics have to be recorded by manufacturer and user
Clause 5 requires nominal or measured values for important instrument characteristics to be traceable and available. The manufacturer provides values for new equipment and new components, while the user records the configuration used for each experiment. This includes source details, monochromators, filters, slits, collimators, detectors, goniometer motion, and specimen stage capabilities, so results can be linked back to the exact setup that produced them.
Alignment and calibration have to control systematic error
Clause 6 says the instrument has to be properly aligned and, where required, calibrated. Alignment is verified with an appropriate calibrant, and calibration is carried out with a suitable, preferably certified, calibrant using a written procedure. In practice, this is what limits systematic errors in 2θ position, peak width and peak shape, which are the quantities that most affect XRPD interpretation.
Performance has to be monitored on a defined schedule
Clause 7 requires performance to be defined for each type of analysis and then monitored at suitable intervals. Annex C gives progressively more detailed ways to do this, from checking line position, intensity and breadth for selected peaks to whole-pattern fitting. If monitoring shows that performance has fallen outside the accepted level, the data collected since the last accepted check have to be re-evaluated.
Annex C gives practical ways to characterise the instrument response
Annex C recommends methods for monitoring angular deviation, line breadth, intensity, profile shape, lattice parameters, the Fundamental Parameter Approach, and whole pattern fitting. These methods are used to build instrument parameter curves and to track changes over time. For the reader, this means the standard is not only about initial setup - it also gives a way to prove that the instrument stays fit for use.
What terms does SIST EN 13925-3:2005 define?
SIST EN 13925-3:2005 uses the terms and definitions of prEN 1330-11:2004. The following terms are central to the document.
- X-ray powder diffraction (XRPD) - The technique used here to study polycrystalline and amorphous materials by their diffraction pattern.
- Bragg-Brentano geometry - A parafocusing diffractometer geometry in which the apparent source, specimen and receiving slit lie on the focusing circle.
- Goniometer - The mechanical and electrical assembly that controls the positions of the diffractometer components relative to the specimen.
- Position sensitive detector (PSD) - A detector that records diffraction intensity over a range of positions at the same time.
- Full width at half-maximum (FWHM) - A line-breadth measure used to assess peak width, resolution and instrumental broadening.
- Fundamental Parameter Approach - A line-profile fitting approach that combines the specimen contribution, the instrument contribution and the spectral distribution of the beam.
- Rietveld method - A whole-pattern refinement method used in Annex C for monitoring instrument behaviour from a complete diffraction pattern.
Who uses SIST EN 13925-3:2005?
SIST EN 13925-3:2005 is used by XRPD laboratories, non destructive testing laboratories, instrument manufacturers, and quality assurance staff who need a controlled and repeatable measurement setup. It is also relevant to laboratory managers and buyers who have to specify what an instrument must provide before acceptance or routine use.
In practice, users apply it when they:
- select or compare diffractometers for a given analysis task
- document the configuration used for a measurement
- align and calibrate an instrument
- check performance over time
- compare results from different instruments or laboratories
- prepare characterisation and report forms
Which standards are used with SIST EN 13925-3:2005?
| Standard or document | What it contributes |
|---|---|
| EN 13925-1:2003 | General principles for XRPD and the definition of “powder” used in this part |
| EN 13925-2:2003 | Procedures for data processing and evaluation, including methods used in calibration and profile analysis |
| prEN 1330-11:2004 | Terminology and definitions for X-ray diffraction from polycrystalline and amorphous materials |
| WI 00138070 | Reference materials used for alignment, calibration and performance testing |
What does the SIST EN 13925-3:2005 document contain?
The document contains a structured description of XRPD instrument hardware and configuration, with separate treatment of sources, optics, detectors, goniometers and specimen stages. Clause 5 provides tables of characteristics to be supplied or recorded, and Annex D gives sample report forms based on those tables.
Annex C contains practical characterisation procedures and the kinds of plots used to follow instrument behaviour, including angular deviation curves, line breadth curves, intensity diagrams, shape analysis curves, lattice-parameter monitoring, the Fundamental Parameter Approach and whole pattern fitting. Annex B lists the main alignment actions for Bragg-Brentano diffractometers, while Annex A shows how this part fits with the rest of the XRPD standard set.
Relations
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
Get Certified
Connect with accredited certification bodies for this standard

IMP NDT d.o.o.
Non-destructive testing services. Radiography, ultrasonic, magnetic particle, penetrant, visual inspection.

Inštitut za kovinske materiale in tehnologije
Institute of Metals and Technology. Materials testing, metallurgical analysis, NDT.

Q Techna d.o.o.
NDT and quality assurance specialist. 30+ years experience. NDT personnel certification per ISO 9712, nuclear and thermal power plant inspections, QA/
Sponsored listings
Frequently Asked Questions
SIST EN 13925-3:2005 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Non destructive testing - X ray diffraction from polycrystalline and amorphous materials - Part 3: Instruments". This standard covers: This document sets out the characteristics of instruments used for X-ray powder diffraction ("powder" as defined in EN 13925-1:2003, Clause 5) as a basis for their control and hence quality assurance of the measurements made by this technique. Performance testing indicators are given for diffractometer performance testing. Different types and makes of X-ray powder diffractometer vary considerably in their design and intended fields of application. This document attempts to cover as much of this range as possible by keeping to common principles. To make the standard more readily applicable, the Bragg-Brentano configuration is addressed in most detail because of its wide use. Additional considerations and adaptations may be necessary to cover some types of instruments or configuration and some fields of application. Some of these types of instrument are described in Annex B.
This document sets out the characteristics of instruments used for X-ray powder diffraction ("powder" as defined in EN 13925-1:2003, Clause 5) as a basis for their control and hence quality assurance of the measurements made by this technique. Performance testing indicators are given for diffractometer performance testing. Different types and makes of X-ray powder diffractometer vary considerably in their design and intended fields of application. This document attempts to cover as much of this range as possible by keeping to common principles. To make the standard more readily applicable, the Bragg-Brentano configuration is addressed in most detail because of its wide use. Additional considerations and adaptations may be necessary to cover some types of instruments or configuration and some fields of application. Some of these types of instrument are described in Annex B.
SIST EN 13925-3:2005 is classified under the following ICS (International Classification for Standards) categories: 19.100 - Non-destructive testing. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN 13925-3:2005 has the following relationships with other standards: It is inter standard links to SIST EN 13925-2:2004, SIST EN 1330-11:2007, SIST-TS CEN ISO/TS 21432:2005, SIST EN 15305:2009, SIST EN ISO 21432:2020. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN 13925-3:2005 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Description of equipment
- 4.1 General
- 4.2 X-ray sources
- 4.3 Incident and diffracted X-ray beam optics
- 1 Focus of X-ray tube
- 2 Monochromator
- 3 Apparent source (on the goniometer circle)
- 1 Receiving slit
- 2 Monochromator
- 3 Detector slit
- 5 Source
- 6 Divergence slit
- 7 Knife edge
- 8 Specimen surface
- 9 Detector window
- 10 Receiving slit
- 11 Goniometer circle
- 4.4 Detectors
- 1 Source
- 2 Specimen
- 4.5 Goniometers
- 1 Focusing circle
- 2 Specimen surface
- 3 Goniometer circle
- 1 Goniometer axis
- 2 In plane of specimen surface
- 4.6 Specimen stage
- 4.7 Data collection system
- 5 Characterisation of equipment components
- 6 Equipment alignment and calibration
- 6.1 General
- 6.2 Alignment
- 6.3 Calibration
- 7 Performance testing and monitoring
- Annex A
- Annex B
- Annex C
- 2 Specimen surface displacement
- 3 Specimen transparency
- 4 Flat specimen
- 5 Axial divergence
- Annex D
- Bibliography
Introduction
X-ray powder diffraction (XRPD) is a powerful Non-Destructive Testing (NDT) method for determining a range of physical and chemical characteristics of materials. These include the type and quantities of phases present, the crystallographic unit cell and structure, crystallographic texture, macrostress, crystallite size and microstrain, and the electron radial distribution function.
This document aims to describe the general aspects of the XRPD technique and its applications but not to define a specific or detailed standard for each field of application or type of analysis. The main purposes of the standard are therefore to provide:
- practical guidance, unified concepts and terminology for use of the XRPD technique in the area of Non- Destructive Testing with general information about its capabilities and limitations of relevance to laboratories working at different levels of sophistication, from routine testing to research;
- a basis for Quality Assurance in XRPD laboratories allowing performance testing and monitoring of instruments as well as the comparison of results from different instruments;
- a general basis (without imposing specifications) for further specific NDT product standards and related Quality Assurance applications, with aspects common to most fields of application. In order to make the standard immediately usable in a wide range of laboratories and applications, diffractometers with Bragg-Brentano geometry are considered in more detail than other instruments. Radiation Protection: Exposure of any part of the human body to X-rays can be injurious to health. It is therefore essential that whenever X-ray equipment is used, adequate precautions should be taken to protect the operator and any other person in the vicinity. Recommended practice for radiation protection as well as limits for the levels of X-radiation exposure are established by national legislation in each country. If there are no official regulations or recommendations in a country, the latest recommendations of the International Commission on Radiological Protection should be applied.
SIST EN 13925-3:2005 EN 13925-3:2005 (E)
1 Scope
This document sets out the characteristics of instruments used for X-ray powder diffraction (“powder” as defined in EN 13925-1:2003, Clause 5) as a basis for their control and hence quality assurance of the measurements made by this technique. Performance testing indicators are given for diffractometer performance testing. Different types and makes of X-ray powder diffractometer vary considerably in their design and intended fields of application. This document attempts to cover as much of this range as possible by keeping to common principles. To make the standard more readily applicable, the Bragg-Brentano configuration is addressed in most detail because of its wide use. Additional considerations and adaptations may be necessary to cover some types of instruments or configuration and some fields of application. Some of these types of instrument are described in Annex B.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
EN 13925-2:2003, Non-destructive testing – X-ray diffraction from polycrystalline and amorphous materials – Part 2: Procedures prEN 1330-11:2004, Non-destructive testing – Terminology – Part 11: X-ray Diffraction from Polycrystalline and Amorphous Materials
3 Terms and definitions
For the purposes of this document, the terms and definitions of prEN 1330-11:2004 apply.
4 Description of equipment
4.1 General
This description is particularly intended for instruments dedicated to the fields of application described in EN 13925-1. For other applications, additional considerations may be required. A diffractometer generally comprises:
− goniometer; − X-ray source; − incident beam optics which may include monochromatisation or filtering, collimation and/or focusing or parallelism of the beam; − diffracted beam optics which may include monochromatisation or filtering, collimation and/or focusing or parallelism of the beam; − specimen stage; − detector; − data collection system.
These parts of the instrument are considered in more detail below. A data processing system is also required to produce measurements from the instrument. Data processing systems, whether manual or computerised, shall be included with the data processing procedures of EN 13925-2.
SIST EN 13925-3:2005 EN 13925-3:2005 (E) A well-controlled environment (temperature and pressure) is strongly recommended for analysis where reproducible measurement of line profile position, width and shape is required. Humidity is may be important because compounds in the specimen may react with water or absorb it with a consequent change in their lattice constants, e.g. clay minerals. The X-ray beam is partially scattered and attenuated by the air in the beam path with consequential effects on the detected diffraction pattern background and intensity. This effect has sometimes been minimised by use of an evacuated or helium filled beam path.
For all the items described in this clause the corresponding main characteristics to be controlled are given in Clause 5.
4.2 X-ray sources
4.2.1 General
There are several types of x-ray source that can be used for XRPD measurements ranging from conventional laboratory sources to intense and well-collimated synchrotron sources. Each source exhibits characteristics that make it more suitable for particular types of analysis. The main source types are described below.
4.2.2 Conventional X-ray sources (sealed tubes and rotating anode sources)
X-rays are obtained by bombarding a metal anode with electrons emitted by the thermoionic effect and accelerated in a strong electric field produced by a high-voltage generator. Most of the kinetic energy of the electrons is converted to heat, which limits the power of the tubes and requires efficient anode cooling. An increase of about two orders of magnitude in brilliance can be obtained using rotating anodes instead of sealed tubes. Microfocus sources operate at relatively low power settings but maintain brightness by electrostatically or magnetically steering the beam inside the X-ray tube onto the target. The spectrum emitted by a conventional X-ray source operating at sufficiently high voltage consists firstly of a continuous background of polychromatic radiation with a sharp cut off at short wavelengths determined by the maximum voltage applied. Upon this is superimposed a limited number of narrow characteristic lines whose wavelengths are characteristic of the anode material. The emitted radiation is not polarised.
The type of X-ray source and the electron emission current, and accelerating voltage applied to it by the high voltage generator have a marked effect on X-ray intensity and its energy distribution. The emission current and accelerating voltage of a conventional X-ray tube normally give reproducible adjustment of the X-ray beam intensity and energy spectrum on a time scale of days. However, experience has shown that the absolute X-ray intensity differs significantly for nominally equal sources and that it decreases with source age.
4.2.3 Synchrotron radiation sources
A beam of charged particles strongly accelerated in an electric field and deflected in a magnetic field emits a
continuous spectrum of X-rays that is as much as 10 times as brilliant as sealed X-ray tubes. It is called "synchrotron radiation". This increased brilliance relates to the total energy spectrum. Monochromatisation of the beam typically results in diffraction intensities one or two orders of magnitude greater than from conventional sources.
The main advantages of using synchrotron radiation for XRPD measurements are:
- nearly parallel-beam diffraction geometry;
- highly monochromatised and tunable radiation;
- very small and almost symmetric contribution of the instrument to the observed line shape that leads to simpler characterisation of line profiles and very good angular resolution.
SIST EN 13925-3:2005 EN 13925-3:2005 (E) This type of X-ray radiation source requires beam flux monitoring with time (the beam flux can decrease significantly during the experiment) and wavelength calibration. The emitted radiation is strongly polarised in the plane of deflection of the charged particles.
4.3 Incident and diffracted X-ray beam optics
4.3.1 General
The main characteristics of the incident and diffracted beams are their wavelength spectrum, their direction of propagation, their cross-sectional area and shape at the specimen as well as their degree of collimation (axial and equatorial divergence) and focusing. The items of equipment that determine these characteristics are called "beam optics". The effects of these items are referred to by the general term "beam conditioning". The items of equipment described below are used to obtain different degrees of radiation purity and different geometries, i.e. a so-called "focused" beam, "monochromatised" beam, "collimated" beam, or "parallel" beam.
4.3.2 Monochromators
4.3.2.1 General
The wavelengths conventionally used with laboratory sources correspond to the characteristic spectral lines from specific anode materials.
For many XRPD applications it is advantageous to eliminate all other spectral components emitted from the X-ray source or the specimen. When using the K doublet is sometimes helpful to additionally suppress the K -line, i.e. α1,2 α2 to work with highly monochromatic K -radiation.
α1 Monochromatisation devices can be used singly or in combination.
4.3.2.2 Filters and filters
ββββ Partial monochromatisation can be obtained using K filters, i.e. foils made of a metal selected as having an β absorption edge between the wavelengths of the K and K radiation emitted by the source. α β Standard K filters are designed to reduce the K intensity to about two orders of magnitude less than the β β K intensity. Such a filter also attenuates the K doublet intensity, typically by a factor of 2, and reduces the α α polychromatic radiation from the source. Two filters are sometimes combined to form so-called "balanced filters" [1].
Spectral filters shall be used with caution because the absorption edges of the filter material cause significant steps in the background close to each of the observed diffraction lines. Such steps impede accurate analysis when the complete line profile(s) are used, for example, in line-broadening analysis. In some cases it is useful to attenuate the incident or the diffracted beam. Then a foil is used that is made of a metal (often aluminium) selected as having no absorption edge near the radiation wavelength used.
4.3.2.3 Crystal Monochromators
Monochromatisation of radiation is most often accomplished by diffraction from a crystal. We may distinguish:
a) a mosaic crystal. The relative energy or wavelength resolution (∆Ε/Ε and ∆λ/λ, respectively) is typically a few parts per hundred. Most of the K radiation and of the continuous polychromatic radiation is eliminated. The β orientations of the coherently diffracting parts of the mosaic crystal (generally graphite) are spread over about 0,4 degrees for many commercially available monochromators. For such crystals more than 50 % of the incident intensity of the K doublet can be retained. Mosaic crystal monochromators are usually installed in the diffracted α1,2 beam. In this configuration, they also eliminate specimen fluorescence from all elements except the one that constitutes the anode.
SIST EN 13925-3:2005 EN 13925-3:2005 (E) b) a single crystal. The energy resolution is typically better than two parts per thousand. The two components of the K doublet can be separated and the K radiation eliminated. This type of monochromatisation causes a α1,2 α2 large loss in intensity as typically only about 15 % of the incident intensity of a spectral line is retained. Single crystal monochromators (usually almost perfect single crystals of quartz, silicon or germanium) are generally used on the primary beam often in a focusing arrangement. The intensity loss may be largely offset by the associated reduction in axial divergence of the beam, eliminating the need for one set of Soller slits (see 4.3.3.3.2). Two or more crystals (e.g. [2]) may be combined to achieve much better resolution than one, but the use of multiple crystals may yield insufficient intensity for normal powder diffraction. Care shall be taken to ensure that the harmonics of the chosen wavelength are eliminated by the positioning of suitable apertures (slits) within the monochromator, or by other means.
These devices can be used in the incident and/or diffracted beams (see Figures 1 and 2).
Key
1 Focus of X-ray tube
2 Monochromator
3 Apparent source (on the goniometer circle)
Figure 1 - Positioning the monochromator on the incident beam
SIST EN 13925-3:2005 EN 13925-3:2005 (E)
Key
1 Receiving slit
2 Monochromator
3 Detector slit
Figure 2 - Positioning the monochromator on the diffraction beam
4.3.2.4 Electronic filters
Electronic filtering uses photon-counting detectors (see 4.4) and is often referred to as pulse height discrimination or “discrimination” or “energy discrimination”. It electronically selects the pulses arising from the photons of the radiation chosen for the experiment.
An effect similar to partial monochromatisation can be achieved by the use of a solid-state detector, e.g. Peltiercooled. The relative energy resolution of these systems is of the same order as that of a mosaic crystal monochromator (about one part per 100), i.e. it is sufficient to eliminate the K component and most of the β continuous background, but not to separate the K doublet. Detectors of this type typically have high counting α1,2 efficiencies but are susceptible to dead time effects (prEN 1330-11). Such effects result in a non-linear response
above about 10 counts per second, including all wavelengths registered by the detector, whether or not used for the measurement. Care shall therefore be taken to limit the maximum count rate observed to a level at which the detector function properly. If necessary an appropriate dead time correction shall be applied.
4.3.2.5 Multilayer mirrors
Multilayer mirrors usually consist of alternating thin layers of two elements (e.g. W and C) deposited onto an appropriately curved substrate. The layers have a thickness of the order of a few times the wavelength of the X-radiation used [3].
Multilayer mirrors for X-radiation are primarily used to reconfigure a divergent beam into an intense parallel or convergent beam (see 4.3.3.3.4). The mirror unit also acts as a monochromator by suppressing the K and β polychromatic radiation from the source. The thickness of the layers deposited on the surface varies along the length of the mirror in such a way that Bragg’s law is satisfied over a range of beam divergence angles and the corresponding incidence angles at the mirror. These systems reach a reflectivity close to 100 % and thereby provide high intensity beams. A common configuration uses a parabolic graded mirror to transform the divergent incident radiation from the source into an intense nearly parallel beam. Elliptic graded mirrors may be used to transform the divergent radiation from the source into a convergent (focused) beam. For special applications several mirrors can be combined in the incident and/or diffracted beams.
SIST EN 13925-3:2005 EN 13925-3:2005 (E)
4.3.3 Beam dimensions and geometry
4.3.3.1 General
The geometric dimensions of the incident and diffracted X-ray beams can be controlled using collimators and slits. The geometric dimensions may also be altered significantly by crystal monochromators and graded multilayer mirrors.
The items described below define the beam dimensions and geometry and are illustrated schematically in Figure 3.
Key 1, 2 Possible position of divergence anti-scatter slit 3, 4 Possible position of receiving anti-scatter slit
5 Source
6 Divergence slit
7 Knife edge
8 Specimen surface
9 Detector window
10 Receiving slit
11 Goniometer circle
A, B, D Possible position of Soller slit C Centre of specimen surface and of the goniometer circle Figure 3 — Schematic arrangement of the devices for control of beam dimension and geometry with respect to the centre of the goniometer circle (C)
4.3.3.2 Slits
4.3.3.2.1 General Different types of slit can be placed in the incident and diffracted beam paths, usually to limit their equatorial and axial (see 4.5.1) divergence. Slits can also serve to reduce parasitic scattering, to define the radius of the focusing circle and to control the angular resolution. They are often made of a metal with high X-ray attenuation. Orthogonal slit pairs are used for some measurements (e.g. for crystallographic texture measurements) to control the equatorial and axial beam divergence. In 4.3.3.2.2 to 4.3.3.2.7 only slits will be dealt with, because in XRPD slits are far more widely used than other apertures.
SIST EN 13925-3:2005 EN 13925-3:2005 (E) A useful quantitative geometric treatment of slits for Bragg-Brentano diffractometers is described in [4]. 4.3.3.2.2 Variable aperture slits Mechanically or computer controlled slits can be applied as variable divergence, antiscatter and receiving slits. A variable aperture slit is most often used in diffractometers with Bragg-Brentano geometry to maintain a constant irradiated surface area of the specimen while the angle between the incident beam and the specimen surface is varied. As a consequence, the irradiated specimen volume varies during the 2θ scan. A matched variable anti-scatter slit may be used in conjunction with a variable divergence slit. This can be particularly effective in minimising parasitic scattering at low 2θ angles. 4.3.3.2.3 Divergence slits Divergence slits between the X-ray tube and the specimen. They limit the equatorial divergence of the incident beam.
4.3.3.2.4 Antiscatter slits Antiscatter slits reduce parasitic scattering caused by the interaction of the incident and diffracted beams with the air, the specimen stage and the components of the optical system. They are placed in the path of the X-ray beams directly before or after the specimen or after the receiving slit. They may be used singly or in combination. If properly aligned they do not restrict the incident or diffracted beams and they do not change the diffraction geometry.
4.3.3.2.5 Knife edge The so-called "knife edge" can be used for flat specimens. It is a flat blade or wedge with a straight edge positioned close to the specimen surface, parallel to the specimen surface and the goniometer axis. A knife edge reduces the parasitic scattering by limiting the irradiated area of the specimen. It has the same function as the divergence and the antiscatter slits. The position of the knife edge can be stationary or variable, and it can be mechanically or computer controlled.
...