SIST EN ISO 10993-6:2026
(Main)Biological evaluation of medical devices - Part 6: Tests for local effects after implantation (ISO 10993-6:2026)
- Abstract
This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required.
This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs).
These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.
- Status
- Published
- Public Enquiry End Date
- 30-Jul-2024
- Publication Date
- 10-Jun-2026
- Technical Committee
- VAZ - Healthcare
- Current Stage
- 6060 - National Implementation/Publication (Adopted Project)
- Start Date
- 28-May-2026
- Due Date
- 02-Aug-2026
- Completion Date
- 11-Jun-2026
- Directive
- 2017/745 - Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EECOJ Ref: L 10.11.2026 OJ Date: 10-Nov-2026
SIST EN ISO 10993-6:2026 is the Slovenian adoption of EN ISO 10993-6:2026, based on ISO 10993-6:2026. SIST EN ISO 10993-6:2026 specifies implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. It is used when a device or material needs biological evaluation of tissue response after implantation, including cases where skin or mucosal tissue is breached.
What does SIST EN ISO 10993-6:2026 specify?
SIST EN ISO 10993-6:2026 specifies how to carry out implantation tests to assess local tissue effects after implantation. The focus is on biological reactivity at the implant site, not on mechanical loading or functional performance.
The scope covers medical devices and materials that may be solid or non-solid, including porous materials, liquids, gels, pastes, powders and particulates. It also covers absorbable, degradable, non-absorbable materials and tissue-engineered medical products (TEMPs).
The document is organized into general clauses for planning, sample preparation, controls, animal model selection, test periods, surgery, evaluation and reporting, followed by annexes for specific implantation sites and supporting evaluation methods.
| Annex | What it covers |
|---|---|
| Annex A | Test method for implantation in subcutaneous tissue |
| Annex B | Test method for implantation in muscle |
| Annex C | Test method for implantation in bone |
| Annex D | Test method for implantation in neural tissue |
| Annex E | Test methods for devices contacting peripheral nerve tissue |
| Annex F | Examples of scoring systems used to support the evaluation of local biological effects after implantation |
| Annex G | Microscopic evaluation of tissue responses to implanted materials |
| Annex ZA | Relationship to the General Safety and Performance Requirements of Regulation (EU) 2017/745 |
What are the key requirements of SIST EN ISO 10993-6:2026?
SIST EN ISO 10993-6:2026 requires the implantation study to be planned in detail and documented before testing starts. In practice, the study plan needs to identify the test method, samples, animals, implantation site, duration, measurement frequency, assessment method and evaluation approach so the results can be interpreted later.
Study planning and controls
Clause 4.1 says animal testing should be avoided unless the needed data cannot be obtained another way. It also requires animal studies to be done in a suitably approved facility, under the applicable animal welfare rules, and under Good Laboratory Practices or another recognized quality system.
Clause 4.2 and 4.3 require test and control samples to be in the final finished state and the choice of control to be justified and documented. Comparative controls should be as similar as practical in shape, size, curvature, porosity and surface condition, because matching the control to the implant makes the tissue comparison meaningful.
For non-absorbable materials, a stable non-absorbable comparative control shall be used. For absorbable materials, a similar absorbable control should be considered when possible, because the tissue response changes as degradation proceeds.
Where and how to implant
Clause 5.1 requires the test sample to be implanted into or onto the tissue most relevant to the intended clinical use. For absorbable materials, the implantation site shall be marked so it can be identified later, and a sham surgical procedure may be used to separate procedure effects from implant effects.
Clause 5.2 ties the animal model to the principles in ISO 10993-2 and requires the test and control samples to be implanted under the same conditions. The annexes then adapt the general method to the relevant tissue type, including subcutaneous tissue, muscle, bone, neural tissue and peripheral nerve tissue.
Clause 5.3 links the test period to the expected clinical exposure or to the time needed to reach a stable tissue response. For non-absorbable materials, the document distinguishes short-term and long-term observation windows; for absorbable materials, the period is related to the estimated degradation time.
Surgery, observation and evaluation
Clause 5.4 requires surgery under general anaesthesia, regular monitoring of animal health and body mass, and euthanasia by accepted humane methods at the end of the study. This matters because surgical trauma can affect the tissue response and must be controlled and recorded.
Clause 5.5 requires both macroscopic and microscopic assessment of the implant site and the surrounding tissue. Annex D and Annex E also call for neurological and peripheral-nerve-specific observations when those tissues are the target.
Clause 6 requires the report to contain enough detail for an independent review. The report has to identify the laboratory, the samples, the animals, the implantation and retrieval methods, the histology method, the findings, and the final comparative evaluation of test and control materials.
What terms does SIST EN ISO 10993-6:2026 define?
SIST EN ISO 10993-6:2026 defines key terms used to describe implant response and study design.
- Absorb - the action of foreign material or its decomposition products passing through or being assimilated by cells and tissue over time.
- Comparative control - a medical device or material with a history of safe clinical use that serves as the comparison article for a new device.
- Coupon - smaller representative portions of a larger or more complex device that still contain the same materials, finishes and processing as the final device.
- Degradation product - an intermediate or final substance formed by physical, metabolic or chemical decomposition.
- Local effect - a tissue response visible by gross pathology and histopathology at the site where the sample is implanted or administered.
- Location marker - an inert, non-absorbable material or process used to identify the implant location in tissue.
- Reference control - a material with known properties that is considered generally inert with respect to local tissue effects.
- Steady-state - a biologically stable condition where further change in the test system is no longer detected over time.
Who uses SIST EN ISO 10993-6:2026?
SIST EN ISO 10993-6:2026 is used by medical device manufacturers, biomaterials developers, contract test laboratories, toxicologists and pathology teams. It is also used by quality, regulatory and clinical evaluation staff who need implantation data for biocompatibility files and submissions.
In practice, users apply it when they need to choose an animal model, select a control, plan implantation sites and periods, perform histopathology, or write a test report for a device or material intended to contact tissue after implantation. It is especially relevant for subcutaneous, muscle, bone, neural and peripheral nerve evaluation.
What changed in SIST EN ISO 10993-6:2026 from the previous edition?
SIST EN ISO 10993-6:2026 replaces SIST EN ISO 10993-6:2017. The European foreword also states that EN ISO 10993-6:2026 supersedes EN ISO 10993-6:2016, and that the ISO text was approved by CEN without modification.
Which standards are used with SIST EN ISO 10993-6:2026?
SIST EN ISO 10993-6:2026 depends on several other ISO 10993 parts and related standards for its full use.
| Standard | What it contributes |
|---|---|
| ISO 10993-1 | General principles for biological safety evaluation within a risk management process |
| ISO 10993-2 | Animal welfare requirements |
| ISO 10993-4 | Selection of tests for interactions with blood |
| ISO 10993-9 | Framework for identifying and quantifying potential degradation products |
| ISO 10993-11 | Tests for systemic toxicity, used when a combined study is designed |
| ISO 10993-12:2021 | Sample preparation and reference materials |
| ISO 10993-16 | Toxicokinetic study design for degradation products and leachables |
| ISO/TS 37137-1 | General requirements for absorbable medical devices, used for degradation-related control selection |
| ISO 15798, ISO 11979-5, ISO 7405, ISO 14708 series, ISO 5840 series | Vertical standards named as guidance for specific device areas |
What does the SIST EN ISO 10993-6:2026 document contain?
SIST EN ISO 10993-6:2026 contains detailed implantation methods for subcutaneous tissue, muscle, bone, neural tissue and peripheral nerve tissue, plus general rules that apply across all methods. It also contains reporting requirements that tell the user what to document about samples, animals, surgery, retrieval and histology.
Annex F gives example semiquantitative scoring systems for inflammatory and tissue-response grading, including a neural tissue example. Annex G explains how microscopic evaluation should be approached, including tissue section preparation, implant-tissue interface assessment, lymph node evaluation, and the estimation of absorption or degradation.
Annex ZA links the European standard to Regulation (EU) 2017/745 and explains how the standard can support conformity with selected General Safety and Performance Requirements.
Relations
- Effective Date
- 01-Jul-2026
- Effective Date
- 27-May-2026
- Effective Date
- 27-May-2026
- Effective Date
- 27-May-2026
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
SIST EN ISO 10993-6:2026 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Biological evaluation of medical devices - Part 6: Tests for local effects after implantation (ISO 10993-6:2026)". This standard covers: This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required. This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs). These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.
This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required. This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs). These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.
SIST EN ISO 10993-6:2026 is classified under the following ICS (International Classification for Standards) categories: 11.100.20 - Biological evaluation of medical devices. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN ISO 10993-6:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 10993-6:2017, SIST-TS CEN/TS 15277:2006, SIST EN 60601-1:2007/A2:2021, SIST EN 455-3:2007. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN ISO 10993-6:2026 is associated with the following European legislation: EU Directives/Regulations: 2017/745; Standardization Mandates: M/575, M/575 Amd 1, M/575 Amd 2. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
SIST EN ISO 10993-6:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 absorb
- 3.2 comparative control
- 3.3 coupon
- 3.4 degradation product
- 3.5 degrade
- 3.6 euthanasia
- 3.7 local effect
- 3.8 location marker
- 3.9 reference control
- 3.10 steady-state
- 4 Common provisions for implantation test methods
- 4.1 General
- 4.2 Preparation of samples for implantation
- 4.3 Selection of control samples
- 5 General aspects and requirements for implantation test
- 5.1 Tissue and implantation site
- 5.2 Animal model
- 5.3 Test periods
- 5.4 Surgery and testing conditions
- 5.5 Evaluation
- 6 Test report
- 6.1 General
- 6.2 Test laboratory
- 6.3 Implant samples
- 6.4 Animals and implantation
- 6.5 Retrieval and histological procedure
- 6.6 Macroscopic and microscopic evaluation
- 6.7 Final evaluation
- Annex A
- Annex B
- Annex C
- Annex D
- Annex E
- Annex F
- 0 Rare, 1/hpf to 5/hpf Rare, 6/hpf to 10/hpf
- Annex G
- Bibliography
Introduction
The objective of the implantation test methods is to characterize the local tissue response after implantation of a medical device or material (test sample) including integration, degradation, or absorption in an appropriate animal model.
The test sample is implanted into an anatomical site appropriate for the evaluation of the local effects of the medical device (or portion of) in an animal.
The medical device or material local effects are evaluated by a comparison of the tissue response caused by a test sample to that caused by comparative or reference control samples used in medical devices whose clinical acceptability and biocompatibility characteristics have been established. Careful study design can include other relevant biological effects to reduce the number of animals used to evaluate safety and efficacy while accomplishing all study objectives. Additionally, a long-term systemic toxicity study that is designed to incorporate the methods, biological effects, additional timepoints and outcomes of implantation testing can satisfy the requirements of this document and ISO 10993-11.
SIST EN ISO 10993-6:2026
SIST EN ISO 10993-6:2026 International Standard ISO 10993-6:2026(en) Biological evaluation of medical devices — Part 6:
Tests for local effects after implantation
1 Scope
This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required.
This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs). These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 10993-1, Biological evaluation of medical devices — Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process ISO 10993-2, Biological evaluation of medical devices — Part 2: Animal welfare requirements ISO 10993-4, Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood ISO 10993-9, Biological evaluation of medical devices — Part 9: Framework for identification and quantification of potential degradation products ISO 10993-12:2021, Biological evaluation of medical devices — Part 12: Sample preparation and reference materials ISO 10993-16, Biological evaluation of medical devices — Part 16: Toxicokinetic study design for degradation products and leachables
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 10993-1, ISO 10993-2, ISO 10993-4, ISO 10993-9, ISO 10993-12, ISO 10993-16 and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
SIST EN ISO 10993-6:2026 ISO 10993-6:2026(en)
3.1 absorb
action of a non-endogenous (foreign) material or substance, or its decomposition products passing through or being assimilated by cells and tissue over time
3.2 comparative control
medical device or material with a history of safe clinical use that is used as a point of comparison for a new medical device seeking regulatory approval
3.3 coupon
multiple smaller compositionally representative portions of a more complex or larger device that together contain all materials, surface finishes, and processing as the final, finished device
3.4 degradation product
intermediate or final substance which results from the physical, metabolic, or chemical decomposition of a material or substance [SOURCE: ISO/TS 37137-1:2021, 3.2, modified — “and/or” has been changed to “or” and “agent” has been replaced with “substance”.]
3.5 degrade
physically, metabolically or chemically decompose a material or substance [SOURCE: ISO/TS 37137-1:2021, 3.3, modified — “and/or” has been changed to “or”.]
3.6 euthanasia
humane killing of an animal by a method causing minimal physical and mental suffering [SOURCE: ISO 10993-2:2022, 3.5]
3.7 local effect
tissue response to an implanted test sample that can be seen with gross pathology and histopathology at the site of test sample administration or implantation
3.8 location marker
inert, non-absorbable material or process used to mark the location in tissue of an implanted material
3.9 reference control
material with known properties considered to be generally inert in terms of biological local tissue effects and has been established in a historical basis within a pharmacopeia or similar standard
3.10 steady-state
biologically stable condition wherein change in the test system’s cellular activity, morphology or features is no longer detected over a period of time
4 Common provisions for implantation test methods
4.1 General
The ISO 10993 series requires animal testing to be avoided unless the required data cannot be obtained by other means. Prior to choosing to pursue implantation studies, consideration to the type of data required,
SIST EN ISO 10993-6:2026 ISO 10993-6:2026(en) the appropriate method and approach to implantation studies should be given to ensure that the data collected covers the concerns for biological reactivity. It is important that the study be planned in sufficient detail such that all relevant information can be extracted from the use of each animal and each study plan or protocol according to ISO 10993-2, ISO 10993-11 and ISO 10993-16. The implantation tests shall be planned and documented in the study documentation, including the information on the selected test methods, test samples, test animals, implantation sites, test duration, measurement frequencies, assessment methods and evaluation. When relevant vertical standards exist, refer to those as guidance for biological specific evaluation (e.g. ISO 15798, ISO 11979-5, ISO 7405, ISO 14708 series, ISO 5840 series). All animal studies shall be performed in a facility approved by a nationally recognized organization and in accordance with all appropriate regulations dealing with laboratory animal welfare to conform with the requirements of ISO 10993-2. These studies shall be performed under Good Laboratory Practices or other recognized, quality assurance systems. While pilot implant studies can or cannot be conducted under Good Laboratory Practices, the test facility should have an adequate quality system. The provisions of this clause shall apply to the test methods specified in Annex A, Annex B, Annex C and Annex D.
4.2 Preparation of samples for implantation
4.2.1 Test and control samples shall be in a final finished state as described in ISO 10993-12 (unless used in a pilot study). The implant size and shape shall be documented and justified. For certain devices, clinically relevant dosing can be considered. A justification should be provided to demonstrate mimicking a clinically relevant dose in the implantation study.
Test samples for various implant sites are described in Annex A, Annex B, Annex C and Annex D. Physical properties (such as form, density, hardness, surface porosity and texture), geometrical characteristics (form, shape and size) and composition can influence the character of the tissue response to the test material and shall be recorded and taken into account when the response is characterized. Comparative control samples should be matched as closely as reasonably possible for physical properties and geometrical (form and shape) characteristics. Otherwise, a reference control (well characterized material with a well characterized response) can be used, such as high density polyethylene (HDPE). Additional control samples may be included in the study to help interpret tissue reactions [e.g. in case of combination drugs, an additional control such as a placebo device (without drug) can be considered]. 4.2.2 Each implant shall be manufactured, processed, cleaned of contaminants and sterilized by the method intended for the final product and this shall be confirmed in the study documentation. After final preparation and sterilization, the implant samples shall be handled aseptically and in such a way as to ensure that they are not damaged or contaminated in any way prior to or during implantation. The use of smaller test samples or test coupons that are consistent with the requirements as described in ISO 10993-12 can be relevant if the medical device to be tested cannot be tested as is, due to size or complex geometry. The test samples or test coupons shall contain all tissue-contacting materials and surface finishes as in the final, finished medical device. For medical devices comprising two or more different materials, implanted test samples should be of similar composition, surface finish and include each individual material to the final product. With multiple materials, it can be necessary to implant several coupons together, each containing a subset of device materials. The potential for synergies and interactions of different materials in the final product should be considered in the choice of test sample. NOTE Refer to the respective annexes for guidance on the number of test and control samples for each material and implantation period.
4.2.3 For materials used as scaffolds for tissue-engineered medical products, it can be appropriate not to use the final preparation pre-populated with cells or proteins, as the immune reaction of the animal to human cellular or protein components of such products and the reaction of the cells to the animal can interfere with the resulting local tissue response, making it difficult to interpret.
SIST EN ISO 10993-6:2026 ISO 10993-6:2026(en) NOTE 1 Implantation in an appropriate immune-deficient animal can be an option to avoid the xenograft responses, if justified.
NOTE 2 Products with living cells are not considered medical devices in all jurisdictions.
4.2.4 For multicomponent material device designed to be cured prior to placement, the components
shall be mixed before use and allowed to set before implantation. However, materials that are designed to polymerize in situ (e.g. bone cements, many dental materials, tissue sealants and glues) shall be introduced in a manner such that in situ polymerization occurs. The procedure used shall be justified and documented.
4.3 Selection of control samples
The process of implanting any material in tissue induces some degree of cellular response. In the evaluation of material local effects after implantation, the choice of control samples is important in determining the acceptability of the observed tissue reaction to the test sample. Evaluation should be performed by comparing the tissue reaction to that of a similar material (comparative control) whose clinical acceptability and biocompatibility characteristics have been established. The geometrical and physical characteristics such as shape, size, curvature and especially the surface condition (including porosity and texture) of the control article(s), should be as similar to that of the implant test samples as is practical. Any deviations shall be explained, justified and recorded. If no appropriate comparative control is available, a reference control can be used. For further guidance, see ISO 10993-12. For example, reference materials as specified in ISO 10993-12:2021, Table A.1 negative control column can provide a good control for solid, smooth, non-absorbable materials. Conversely, the reaction can likely be different if it is used to compare against a non-absorbable or degradable hernia mesh due to the differing physical form characteristics. Ideally, a commercially available material should be used as a comparative control so the tissue reaction is similar to what has been clinically proven. For non-absorbable medical devices or materials, a comparative control shall be used made of a stable non-absorbable comparative material (see Annex A, Annex B, Annex C and Annex D).
Since absorbable materials encounter changing tissue responses as degradation proceeds at rates that differ based on composition, processing and sterilization techniques, similar absorbable comparative medical device or material controls should be considered.
The absorption rate of the control material or medical device should be similar to the test material or medical device (for more information, see ISO/TS 37137-1). Alternatively, a non-absorbable comparative control can be used, if justified (e.g. there are no clinically relevant absorbable controls). If a comparative control is used, typically reference materials as specified in ISO 10993-12:2021, Annex A are also implanted into the animal for comparison and serving as a procedural control yielding a total of three separate articles included in the study. For novel materials or medical devices, if a commercially available comparative material or medical device does not exist, the choice of a control that is as close as possible is preferred. The nature of any adverse inflammatory response or high reactivity rating or similar histopathological evaluation conclusion should be discussed in the context of the control article chosen. The choice of control shall be documented and justified. If no control or a sham control is used, this also shall be documented and justified.
5 General aspects and requirements for implantation test
5.1 Tissue and implantation site
5.1.1 The test sample shall be implanted into or onto the tissues most relevant to the intended clinical use of the material. The justification for the choice of sample numbers, tissue, implantation sites and test period shall be documented. Test methods for various implantation sites are given in Annex A, Annex B, Annex C and Annex D. If other implantation sites are chosen, the general scientific principles behind the test methods described in Annex A, Annex B, Annex C and Annex D shall still be adhered to and a justification be provided for selecting an anatomical location that does not align with the use of the device. NOTE For some devices, there are vertical standards prescribing specific implant studies to evaluate local tissue responses, e.g. for intraocular lens implant (see ISO 11979-5) and dental devices (see ISO 7405). The studies described in the standards can be used to satisfy the requirements in ISO 10993-6.
SIST EN ISO 10993-6:2026 ISO 10993-6:2026(en) 5.1.2 For absorbable materials, the implantation site shall be marked in a manner suitable for identification of the site at the end of the designated time periods, taking into consideration the growth and aging of the implanted animal(s). The use of a non-invasive permanent skin marker or a template marking the placement of the sample is recommended as a single application for short-term study intervals only. This method of marking can come off as skin exfoliates and requires remarking as often as needed to maintain the marked site. Additionally, in many laboratory animals, the loose highly mobile skin limits the precision of marking sites that are deeper than skin. In most circumstances, a location marker comprising an appropriate nonabsorbable inert or biocompatible material (e.g. HDPE 1 mm by 2 mm by 5 mm, polypropylene suture, gold band, clips) may be used to identify the location of the implant site. These location markers should be far enough away from each implant site to not induce changes in the local tissue reaction to the implanted material. If this is not possible, i.e. the marker of the implanted material is adjacent to or on the edge of the implanted material, then the tissue reaction to this marker should not be included in the evaluation of the tissue reaction to the implanted material.
5.1.3 A sham surgical procedure can be used to evaluate the impact of the procedure on the tissue involved; in these cases, the specific justification shall be provided. If a sham surgical procedure is performed, the same implantation procedure that is used for the test or control should be used for the sham procedure.
5.2 Animal model
5.2.1 All aspects of animal care and accommodation shall be performed in accordance with ISO 10993-2. 5.2.2 Animals are used to evaluate local effects following implantation and are described as animal models in the context of this document. In general, small laboratory animals such as mice, rats, guinea-pigs or rabbits are preferred. The use of larger animals such as dogs, sheep, goats or pigs may be justified based upon special scientific considerations of the test sample under study, length of study in relation to animal life expectancy, or if needed to accommodate implant size, with whole device testing or applicable sham defect size.
5.2.3 Select an animal species in line with the principles set out in ISO 10993-2, giving due consideration to the size of the implant test samples, the number of implants per animal, the intended duration of the test in relation to the expected lifespan of the animals, as well as potential species' differences regarding biological responses. The number of animals and implant sites within should be the minimum to account for site to site and animal to animal variability.
5.2.4 For short-term testing, animals such as rodents or rabbits are commonly used. For long-term testing, animals such as rodents, rabbits, dogs, sheep, goats, pigs and other animals with a relatively long life expectancy are suitable.
5.2.5 Before starting an animal study with degradable materials, relevant information from in vitro degradation studies should be considered for estimating relevant retrieval timing. For absorbable materials, a pilot study in rodents can be considered to determine the expected rate of degradation and the implantation duration needed to reach the steady-state before embarking on studies in larger animals. [23] NOTE Guidance regarding in vitro degradation characterization can be found in ISO 13781 and ASTM F1635 [25] for hydrolysable polymeric constructs and in ASTM F3268 for absorbable metallic constructs. General guidance regarding the linkage of in vitro observations with the determination of relevant in vivo retrieval intervals can be found in ISO/TS 17137. Additional guidance on selection of implantation time points for absorbable materials can be found in ISO/TS 37137-1.
...