This document specifies a framework for the identification, and if necessary, quantification of constituents of a medical device, allowing the identification of biological hazards and the estimation and control of biological risks from material constituents, using a generally stepwise approach to the chemical characterization which can include one or more of the following:
—          the identification of its materials of construction (medical device configuration);
—          the characterization of the materials of construction via the identification and quantification of their chemical constituents (material composition);
—          the characterization of the medical device for chemical substances that were introduced during manufacturing (e.g. mould release agents, process contaminants, sterilization residues);
—          the estimation (using laboratory extraction conditions) of the potential of the medical device, or its materials of construction, to release chemical substances under clinical use conditions (extractables);
—          the measurement of chemical substances released from a medical device under its clinical conditions of use (leachables).
This document can also be used for chemical characterization (e.g. the identification and/or quantification) of degradation products. Information on other aspects of degradation assessment are covered in ISO 10993-9, ISO 10993-13, ISO 10993-14 and ISO 10993-15.
The ISO 10993 series is applicable when the material or medical device has direct or indirect body contact (see ISO 10993-1 for categorization by nature of body contact).
This document is intended for suppliers of materials and manufacturers of medical devices, to support a biological evaluation.

  • Standard
    81 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies general requirements for the design of tests for identifying and quantifying
degradation products from final metallic medical devices or corresponding material samples finished
as ready for clinical use.
This document is applicable only to those degradation products generated by chemical alteration of the
final metallic device in an in vitro degradation test. Because of the nature of in vitro tests, the test results
approximate the in vivo behaviour of the implant or material. The described chemical methodologies
are a means to generate degradation products for further assessments.
This document is applicable to both materials designed to degrade in the body as well as materials that
are not intended to degrade.
This document is not applicable to evaluation of degradation which occurs by purely mechanical
processes; methodologies for the production of this type of degradation product are described in
specific product standards, where available.
NOTE Purely mechanical degradation causes mostly particulate matter. Although this is excluded from the
scope of this document, such degradation products can evoke a biological response and can undergo biological
evaluation as described in other parts of ISO 10993.
Because of the wide range of metallic materials used in medical devices, no specific analytical
techniques are identified for quantifying the degradation products. The identification of trace elements
(<10–6 w/w) contained in the specific metal or alloy is not addressed in this document, nor are specific
requirements for acceptable levels of degradation products provided in this document.
This document excludes the biological activity of the degradation products. (See instead the applicable
clauses of ISO 10993-1 and ISO 10993-17).

  • Standard
    26 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to induce skin sensitization.
This document includes:
—    details of in vivo skin sensitization test procedures;
—    key factors for the interpretation of the results.
NOTE      Instructions for the preparation of materials specifically in relation to the above tests are given in Annex A.

  • Standard
    60 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to induce skin sensitization.
This document includes:
—    details of in vivo skin sensitization test procedures;
—    key factors for the interpretation of the results.
NOTE      Instructions for the preparation of materials specifically in relation to the above tests are given in Annex A.

  • Standard
    60 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2021-12-09- JO-  the EN was submitted to FV wiith the link to the legislation (90/385/EEC, 93/42/EEC)
however since these legilsations have been repealed, the link has been removed and the EN is to proceed to publication

  • Amendment
    11 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document provides general principles for the systematic evaluation of the potential and observed degradation of medical devices through the design and performance of in vitro degradation studies. Information obtained from these studies can be used in the biological evaluation described in the ISO 10993 series.
This document is applicable to both materials designed to degrade in the body as well as materials that are not intended to degrade.
This document is not applicable to:
a)   the evaluation of degradation which occurs by purely mechanical processes; methodologies for the production of this type of degradation product are described in specific product standards, where available;
NOTE    Purely mechanical degradation causes mostly particulate matter. Although this is excluded from the scope of this document, such degradation products can evoke a biological response and can undergo biological evaluation as described in other parts of ISO 10993.
b)   leachable components which are not degradation products;
c)   medical devices or components that do not contact the patient's body directly or indirectly.

  • Standard
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2021-12-09- JO-  the EN was submitted to FV wiith the link to the legislation (90/385/EEC, 93/42/EEC)
however since these legilsations have been repealed, the link has been removed and the EN is to proceed to publication

  • Amendment
    11 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and gives guidance on the procedures in the preparation of samples and the selection of reference materials for medical device testing primarily in biological test systems primarily in accordance with one or more parts of the ISO 10993 series.
Specifically, this document addresses the following:
—     test sample selection;
—     selection of representative portions from a medical device;
—     test sample preparation;
—     experimental controls;
—     selection of, and requirements for, reference materials;
—     preparation of extracts.
This document is not applicable to live cells but can be relevant to the material or medical device components of combination products containing live cells.
Extractions for chemical characterization are covered in ISO 10993-18. Clause 7, 8, 9, 10 [with the exception of 10.3.5 and 10.3.11 b)], and 11 can apply to extractions for chemical characterization. Information given in C.1 to C.4 can also be relevant.

  • Standard
    32 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document provides general principles for the systematic evaluation of the potential and observed degradation of medical devices through the design and performance of in vitro degradation studies. Information obtained from these studies can be used in the biological evaluation described in the ISO 10993 series.
This document is applicable to both materials designed to degrade in the body as well as materials that are not intended to degrade.
This document is not applicable to:
a)   the evaluation of degradation which occurs by purely mechanical processes; methodologies for the production of this type of degradation product are described in specific product standards, where available;
NOTE    Purely mechanical degradation causes mostly particulate matter. Although this is excluded from the scope of this document, such degradation products can evoke a biological response and can undergo biological evaluation as described in other parts of ISO 10993.
b)   leachable components which are not degradation products;
c)   medical devices or components that do not contact the patient's body directly or indirectly.

  • Standard
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the requirements for and provides guidance on the specification, selection, qualification, bio-decontamination, validation, operation and control of isolator systems related to aseptic processing of health care products and processing of cell-based health care products.
This document does not specify requirements for restricted access barrier systems (RABS).
This document does not supersede or replace national regulatory requirements such as Good Manufacturing Practices (GMPs) and/or compendia requirements that pertain in particular to national or regional jurisdictions.
This document does not specify requirements for isolators used for sterility testing; however, some of the principles and information in this document could be applicable to this application.
This document does not define biosafety containment requirements.

  • Standard
    39 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to produce irritation. The tests are designed to predict and classify the irritation potential of medical devices, materials or their extracts according to ISO 10993‑1 and ISO 10993‑2.
This document includes:
—     pre-test considerations for irritation, including in silico and in vitro methods for dermal exposure;
—     details of in vitro and in vivo irritation test procedures;
—     key factors for the interpretation of the results.

  • Standard
    76 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and gives guidance on the procedures in the preparation of samples and the selection of reference materials for medical device testing primarily in biological test systems primarily in accordance with one or more parts of the ISO 10993 series.
Specifically, this document addresses the following:
—     test sample selection;
—     selection of representative portions from a medical device;
—     test sample preparation;
—     experimental controls;
—     selection of, and requirements for, reference materials;
—     preparation of extracts.
This document is not applicable to live cells but can be relevant to the material or medical device components of combination products containing live cells.
Extractions for chemical characterization are covered in ISO 10993-18. Clause 7, 8, 9, 10 [with the exception of 10.3.5 and 10.3.11 b)], and 11 can apply to extractions for chemical characterization. Information given in C.1 to C.4 can also be relevant.

  • Standard
    32 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the requirements for and provides guidance on the specification, selection, qualification, bio-decontamination, validation, operation and control of isolator systems related to aseptic processing of health care products and processing of cell-based health care products.
This document does not specify requirements for restricted access barrier systems (RABS).
This document does not supersede or replace national regulatory requirements such as Good Manufacturing Practices (GMPs) and/or compendia requirements that pertain in particular to national or regional jurisdictions.
This document does not specify requirements for isolators used for sterility testing; however, some of the principles and information in this document could be applicable to this application.
This document does not define biosafety containment requirements.

  • Standard
    39 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies:
—          the general principles governing the biological evaluation of medical devices within a risk management process;
—          the general categorization of medical devices based on the nature and duration of their contact with the body;
—          the evaluation of existing relevant data from all sources;
—          the identification of gaps in the available data set on the basis of a risk analysis;
—          the identification of additional data sets necessary to analyse the biological safety of the medical device;
—          the assessment of the biological safety of the medical device.
This document applies to evaluation of materials and medical devices that are expected to have direct or indirect contact with:
—          the patient's body during intended use;
—          the user's body, if the medical device is intended for protection (e.g., surgical gloves, masks and others).
This document is applicable to biological evaluation of all types of medical devices including active, non-active, implantable and non-implantable medical devices.
This document also gives guidelines for the assessment of biological hazards arising from:
—          risks, such as changes to the medical device over time, as a part of the overall biological safety assessment;
—          breakage of a medical device or medical device component which exposes body tissue to new or novel materials.
Other parts of ISO 10993 cover specific aspects of biological assessments and related tests. Device-specific or product standards address mechanical testing.
This document excludes hazards related to bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents and other pathogens.

  • Standard
    50 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to produce irritation. The tests are designed to predict and classify the irritation potential of medical devices, materials or their extracts according to ISO 10993‑1 and ISO 10993‑2.
This document includes:
—     pre-test considerations for irritation, including in silico and in vitro methods for dermal exposure;
—     details of in vitro and in vivo irritation test procedures;
—     key factors for the interpretation of the results.

  • Standard
    76 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies:
—          the general principles governing the biological evaluation of medical devices within a risk management process;
—          the general categorization of medical devices based on the nature and duration of their contact with the body;
—          the evaluation of existing relevant data from all sources;
—          the identification of gaps in the available data set on the basis of a risk analysis;
—          the identification of additional data sets necessary to analyse the biological safety of the medical device;
—          the assessment of the biological safety of the medical device.
This document applies to evaluation of materials and medical devices that are expected to have direct or indirect contact with:
—          the patient's body during intended use;
—          the user's body, if the medical device is intended for protection (e.g., surgical gloves, masks and others).
This document is applicable to biological evaluation of all types of medical devices including active, non-active, implantable and non-implantable medical devices.
This document also gives guidelines for the assessment of biological hazards arising from:
—          risks, such as changes to the medical device over time, as a part of the overall biological safety assessment;
—          breakage of a medical device or medical device component which exposes body tissue to new or novel materials.
Other parts of ISO 10993 cover specific aspects of biological assessments and related tests. Device-specific or product standards address mechanical testing.
This document excludes hazards related to bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents and other pathogens.

  • Standard
    50 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a framework for the identification, and if necessary, quantification of constituents of a medical device, allowing the identification of biological hazards and the estimation and control of biological risks from material constituents, using a generally stepwise approach to the chemical characterization which can include one or more of the following:
—          the identification of its materials of construction (medical device configuration);
—          the characterization of the materials of construction via the identification and quantification of their chemical constituents (material composition);
—          the characterization of the medical device for chemical substances that were introduced during manufacturing (e.g. mould release agents, process contaminants, sterilization residues);
—          the estimation (using laboratory extraction conditions) of the potential of the medical device, or its materials of construction, to release chemical substances under clinical use conditions (extractables);
—          the measurement of chemical substances released from a medical device under its clinical conditions of use (leachables).
This document can also be used for chemical characterization (e.g. the identification and/or quantification) of degradation products. Information on other aspects of degradation assessment are covered in ISO 10993-9, ISO 10993-13, ISO 10993-14 and ISO 10993-15.
The ISO 10993 series is applicable when the material or medical device has direct or indirect body contact (see ISO 10993-1 for categorization by nature of body contact).
This document is intended for suppliers of materials and manufacturers of medical devices, to support a biological evaluation.

  • Standard
    81 pages
    English language
    sale 10% off
    e-Library read for
    1 day

1.1    Inclusions
1.1.1    This document specifies requirements for the development, validation and routine control of a low temperature steam and formaldehyde (LTSF) sterilization process for medical devices using a mixture of low temperature steam and formaldehyde as sterilizing agent and which operates below ambient pressure.
NOTE       Although the scope of this document is limited to medical devices, it specifies requirements and provides guidance that can be applicable to other products and equipment.
1.1.2    This document is intended to be applied by process developers, manufacturers of sterilization equipment, manufacturers of medical devices to be sterilized and the organizations with responsibility for sterilizing medical devices (see ISO 14937:2009, Table E.1).
1.2    Exclusions
1.2.1    This document does not specify requirements for the development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.
NOTE       See ISO 22442‑1, ISO 22442‑2 and ISO 22442‑3.
1.2.2    This document does not specify requirements for designating a medical device as "STERILE". Such requirements are given in EN 556‑1.
1.2.3    This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE       It is not a requirement of this document to have a complete quality management system during manufacture or reprocessing, but those elements of such a system that are required are normatively referenced at appropriate places in the text. Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production or reprocessing of medical devices including the sterilization process. Further guidance is given in E.4 of ISO 14937:2009.
1.2.4    This document does not specify requirements for occupational safety associated with the design and operation of LTSF sterilization facilities.
NOTE 1    Safety requirements for sterilizers are specified in IEC 61010‑2‑040.
NOTE 2    Attention is also drawn to the existence in some countries of regulations stipulating safety requirements.
1.2.5    This document does not cover analytical methods for determining levels or residues of formaldehyde and/or its reaction products.
NOTE 1    Attention is drawn to EN 14180.
NOTE 2    Attention is drawn to the possible existence in some countries of statutory regulations specifying limits for the level of formaldehyde residues on medical devices and products.
1.2.6    This document does not cover preparatory measures that might be necessary before sterilization such as cleaning, disinfection and packing.
NOTE       For reprocessable medical devices, the manufacturer(s) of these devices can supply information on the preparatory measures (see ISO 17664).

  • Standard
    63 pages
    English language
    sale 10% off
    e-Library read for
    1 day

20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO-  under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019. Awaiting for assessement report from E&Y.
2018-10-17 - TAN : Lack of compliance

  • Amendment
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO-  under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019- Awaiting for  the assessment report E&Y Report

  • Amendment
    19 pages
    English language
    sale 10% off
    e-Library read for
    1 day

NEXT ACTION: TC ACTION BY 2022-10-26 : TC to send a revised annex ZA for assessment at PUB
2020-02-19- JO-CEN/TC 206 to take decision either to remove the link or send a revised annex ZA for assessment at PUB

  • Standard
    26 pages
    English language
    sale 10% off
    e-Library read for
    1 day

1.1    Inclusions
1.1.1    This document specifies requirements for the development, validation and routine control of a low temperature steam and formaldehyde (LTSF) sterilization process for medical devices using a mixture of low temperature steam and formaldehyde as sterilizing agent and which operates below ambient pressure.
NOTE       Although the scope of this document is limited to medical devices, it specifies requirements and provides guidance that can be applicable to other products and equipment.
1.1.2    This document is intended to be applied by process developers, manufacturers of sterilization equipment, manufacturers of medical devices to be sterilized and the organizations with responsibility for sterilizing medical devices (see ISO 14937:2009, Table E.1).
1.2    Exclusions
1.2.1    This document does not specify requirements for the development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.
NOTE       See ISO 22442‑1, ISO 22442‑2 and ISO 22442‑3.
1.2.2    This document does not specify requirements for designating a medical device as "STERILE". Such requirements are given in EN 556‑1.
1.2.3    This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE       It is not a requirement of this document to have a complete quality management system during manufacture or reprocessing, but those elements of such a system that are required are normatively referenced at appropriate places in the text. Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production or reprocessing of medical devices including the sterilization process. Further guidance is given in E.4 of ISO 14937:2009.
1.2.4    This document does not specify requirements for occupational safety associated with the design and operation of LTSF sterilization facilities.
NOTE 1    Safety requirements for sterilizers are specified in IEC 61010‑2‑040.
NOTE 2    Attention is also drawn to the existence in some countries of regulations stipulating safety requirements.
1.2.5    This document does not cover analytical methods for determining levels or residues of formaldehyde and/or its reaction products.
NOTE 1    Attention is drawn to EN 14180.
NOTE 2    Attention is drawn to the possible existence in some countries of statutory regulations specifying limits for the level of formaldehyde residues on medical devices and products.
1.2.6    This document does not cover preparatory measures that might be necessary before sterilization such as cleaning, disinfection and packing.
NOTE       For reprocessable medical devices, the manufacturer(s) of these devices can supply information on the preparatory measures (see ISO 17664).

  • Standard
    63 pages
    English language
    sale 10% off
    e-Library read for
    1 day

20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO-  under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019- Awaiting for  the assessment report E&Y Report

  • Amendment
    19 pages
    English language
    sale 10% off
    e-Library read for
    1 day

20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO-  under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019. Awaiting for assessement report from E&Y.
2018-10-17 - TAN : Lack of compliance

  • Amendment
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 13485:2016 specifies requirements for a quality management system where an organization needs to demonstrate its ability to provide medical devices and related services that consistently meet customer and applicable regulatory requirements. Such organizations can be involved in one or more stages of the life-cycle, including design and development, production, storage and distribution, installation, or servicing of a medical device and design and development or provision of associated activities (e.g. technical support). ISO 13485:2016 can also be used by suppliers or external parties that provide product, including quality management system-related services to such organizations.
Requirements of ISO 13485:2016 are applicable to organizations regardless of their size and regardless of their type except where explicitly stated. Wherever requirements are specified as applying to medical devices, the requirements apply equally to associated services as supplied by the organization.
The processes required by ISO 13485:2016 that are applicable to the organization, but are not performed by the organization, are the responsibility of the organization and are accounted for in the organization's quality management system by monitoring, maintaining, and controlling the processes.
If applicable regulatory requirements permit exclusions of design and development controls, this can be used as a justification for their exclusion from the quality management system. These regulatory requirements can provide alternative approaches that are to be addressed in the quality management system. It is the responsibility of the organization to ensure that claims of conformity to ISO 13485:2016 reflect any exclusion of design and development controls.
If any requirement in Clauses 6, 7 or 8 of ISO 13485:2016 is not applicable due to the activities undertaken by the organization or the nature of the medical device for which the quality management system is applied, the organization does not need to include such a requirement in its quality management system. For any clause that is determined to be not applicable, the organization records the justification as described in 4.2.2.

  • Standard
    68 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    78 pages
    German language
    sale 10% off
    e-Library read for
    1 day
  • Standard – translation
    131 pages
    Slovenian and English language
    sale 10% off
    e-Library read for
    1 day

ISO 13408-2:2018 specifies requirements for sterilizing filtration as part of aseptic processing of health care products conducted in accordance with ISO 13408‑1. It also offers guidance to filter users concerning general requirements for set-up, validation and routine operation of a sterilizing filtration process.
ISO 13408-2:2018 is not applicable to removal of viruses.
Sterilizing filtration is not applicable to fluids that intentionally contain particles larger than the pore size of the filter (e.g. bacterial whole-cell vaccines).
ISO 13408-2:2018 is not applicable to high efficiency particulate air (HEPA) filters.
ISO 13408-2:2018 does not specify requirements for the development, validation and routine control of a process for removing the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.

  • Standard
    47 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11737-1:2018 specifies requirements and provides guidance on the enumeration and microbial characterization of the population of viable microorganisms on or in a health care product, component, raw material or package.
NOTE 1    The nature and extent of microbial characterization is dependent on the intended use of bioburden data.
NOTE 2    See Annex A for guidance on Clauses 1 to 9.
ISO 11737-1:2018 does not apply to the enumeration or identification of viral, prion or protozoan contaminants. This includes the removal and detection of the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE 3    Guidance on inactivating viruses and prions can be found in ISO 22442‑3, ICH Q5A(R1) and ISO 13022.
ISO 11737-1:2018 does not apply to the microbiological monitoring of the environment in which health care products are manufactured.

  • Standard
    58 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 13408-2:2018 specifies requirements for sterilizing filtration as part of aseptic processing of health care products conducted in accordance with ISO 13408‑1. It also offers guidance to filter users concerning general requirements for set-up, validation and routine operation of a sterilizing filtration process.
ISO 13408-2:2018 is not applicable to removal of viruses.
Sterilizing filtration is not applicable to fluids that intentionally contain particles larger than the pore size of the filter (e.g. bacterial whole-cell vaccines).
ISO 13408-2:2018 is not applicable to high efficiency particulate air (HEPA) filters.
ISO 13408-2:2018 does not specify requirements for the development, validation and routine control of a process for removing the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.

  • Standard
    47 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 10993-4:2017 specifies general requirements for evaluating the interactions of medical devices with blood.
It describes
a)    a classification of medical devices that are intended for use in contact with blood, based on the intended use and duration of contact as defined in ISO 10993‑1,
b)    the fundamental principles governing the evaluation of the interaction of devices with blood,
c)    the rationale for structured selection of tests according to specific categories, together with the principles and scientific basis of these tests.
Detailed requirements for testing cannot be specified because of limitations in the knowledge and precision of tests for evaluating interactions of devices with blood. This document describes biological evaluation in general terms and may not necessarily provide sufficient guidance for test methods for a specific device.
The changes in this document do not indicate that testing conducted according to prior versions of this document is invalid. For marketed devices with a history of safe clinical use, additional testing according to this revision is not recommended.

  • Standard
    82 pages
    English language
    sale 10% off
    e-Library read for
    1 day

CCMC - creation of a 2nd corrigendum as the instructions in the 1st corrigendum were incomplete and unclear

  • Corrigendum
    9 pages
    English language
    sale 10% off
    e-Library read for
    1 day

CCMC - creation of a 2nd corrigendum as the instructions in the 1st corrigendum were incomplete and unclear

  • Corrigendum
    9 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11737-1:2018 specifies requirements and provides guidance on the enumeration and microbial characterization of the population of viable microorganisms on or in a health care product, component, raw material or package.
NOTE 1    The nature and extent of microbial characterization is dependent on the intended use of bioburden data.
NOTE 2    See Annex A for guidance on Clauses 1 to 9.
ISO 11737-1:2018 does not apply to the enumeration or identification of viral, prion or protozoan contaminants. This includes the removal and detection of the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE 3    Guidance on inactivating viruses and prions can be found in ISO 22442‑3, ICH Q5A(R1) and ISO 13022.
ISO 11737-1:2018 does not apply to the microbiological monitoring of the environment in which health care products are manufactured.

  • Standard
    58 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11137-3:2017 gives guidance on meeting the requirements in ISO 11137-1 and ISO 11137-2 and in ISO/TS 13004 relating to dosimetry and its use in development, validation and routine control of a radiation sterilization process.

  • Standard
    50 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11137-3:2017 gives guidance on meeting the requirements in ISO 11137-1 and ISO 11137-2 and in ISO/TS 13004 relating to dosimetry and its use in development, validation and routine control of a radiation sterilization process.

  • Standard
    50 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11135:2014 specifies requirements for the development, validation and routine control of an ethylene oxide sterilization process for medical devices in both the industrial and health care facility settings, and it acknowledges the similarities and differences between the two applications.

  • Standard
    90 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 10993-6:2016 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices.
ISO 10993-6:2016 applies to materials that are
-      solid and non-absorbable,
-      non-solid, such as porous materials, liquids, gels, pastes, and particulates, and
-      degradable and/or absorbable, which may be solid or non-solid.
The test sample is implanted into a site and animal species appropriate for the evaluation of the biological safety of the material. These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical or functional loading. This part of ISO 10993 can also be applied to medical devices that are intended to be used topically in clinical indications where the surface or lining might have been breached, in order to evaluate local tissue responses.
The local effects are evaluated by a comparison of the tissue response caused by a test sample to that caused by control materials used in medical devices whose clinical acceptability and biocompatibility characteristics have been established. The objective of the test methods is to characterize the history and evolution of the tissue response after implantation of a medical device/biomaterial including final integration or absorption/degradation of the material. In particular for degradable/absorbable materials, the degradation characteristics of the material and the resulting tissue response should be determined.
ISO 10993-6:2016 does not deal with systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the long-term implantation studies intended for evaluation of local biological effects might provide insight into some of these properties. Systemic toxicity studies conducted by implantation might satisfy the requirements of this part of ISO 10993. When conducting combined studies for evaluating local effects and systemic effects, the requirements of both standards is to be fulfilled.

  • Standard
    42 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 10993-6:2016 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices.
ISO 10993-6:2016 applies to materials that are
-      solid and non-absorbable,
-      non-solid, such as porous materials, liquids, gels, pastes, and particulates, and
-      degradable and/or absorbable, which may be solid or non-solid.
The test sample is implanted into a site and animal species appropriate for the evaluation of the biological safety of the material. These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical or functional loading. This part of ISO 10993 can also be applied to medical devices that are intended to be used topically in clinical indications where the surface or lining might have been breached, in order to evaluate local tissue responses.
The local effects are evaluated by a comparison of the tissue response caused by a test sample to that caused by control materials used in medical devices whose clinical acceptability and biocompatibility characteristics have been established. The objective of the test methods is to characterize the history and evolution of the tissue response after implantation of a medical device/biomaterial including final integration or absorption/degradation of the material. In particular for degradable/absorbable materials, the degradation characteristics of the material and the resulting tissue response should be determined.
ISO 10993-6:2016 does not deal with systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the long-term implantation studies intended for evaluation of local biological effects might provide insight into some of these properties. Systemic toxicity studies conducted by implantation might satisfy the requirements of this part of ISO 10993. When conducting combined studies for evaluating local effects and systemic effects, the requirements of both standards is to be fulfilled.

  • Standard
    42 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 13485:2016 specifies requirements for a quality management system where an organization needs to demonstrate its ability to provide medical devices and related services that consistently meet customer and applicable regulatory requirements. Such organizations can be involved in one or more stages of the life-cycle, including design and development, production, storage and distribution, installation, or servicing of a medical device and design and development or provision of associated activities (e.g. technical support). ISO 13485:2016 can also be used by suppliers or external parties that provide product, including quality management system-related services to such organizations.
Requirements of ISO 13485:2016 are applicable to organizations regardless of their size and regardless of their type except where explicitly stated. Wherever requirements are specified as applying to medical devices, the requirements apply equally to associated services as supplied by the organization.
The processes required by ISO 13485:2016 that are applicable to the organization, but are not performed by the organization, are the responsibility of the organization and are accounted for in the organization's quality management system by monitoring, maintaining, and controlling the processes.
If applicable regulatory requirements permit exclusions of design and development controls, this can be used as a justification for their exclusion from the quality management system. These regulatory requirements can provide alternative approaches that are to be addressed in the quality management system. It is the responsibility of the organization to ensure that claims of conformity to ISO 13485:2016 reflect any exclusion of design and development controls.
If any requirement in Clauses 6, 7 or 8 of ISO 13485:2016 is not applicable due to the activities undertaken by the organization or the nature of the medical device for which the quality management system is applied, the organization does not need to include such a requirement in its quality management system. For any clause that is determined to be not applicable, the organization records the justification as described in 4.2.2.

  • Standard
    68 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    78 pages
    German language
    sale 10% off
    e-Library read for
    1 day
  • Standard – translation
    131 pages
    Slovenian and English language
    sale 10% off
    e-Library read for
    1 day

ISO 11137-2:2013 specifies methods for determining the minimum dose needed to achieve a specified requirement for sterility and methods to substantiate the use of 25 kGy or 15 kGy as the sterilization dose to achieve a sterility assurance level, SAL, of 10−6. ISO 11137-2:2013 also specifies methods of sterilization dose audit used to demonstrate the continued effectiveness of the sterilization dose.
ISO 11137-2:2013 defines product families for sterilization dose establishment and sterilization dose audit.

  • Standard
    81 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11137-1:2006 specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. Although the scope of ISO 11137-1:2006 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment.
ISO 11137-1:2006 covers radiation processes employing irradiators using the radionuclide 60Co or 137Cs, a beam from an electron generator or a beam from an X-ray generator.
ISO 11137-1:2006 does not:
specify requirements for development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease;detail specified requirements for designating a medical device as sterile;specify a quality management system for the control of all stages of production of medical devices;specify requirements for occupational safety associated with the design and operation of irradiation facilities;specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    58 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This European Standard specifies a test method and the minimum requirements for bactericidal activity of chemical disinfectant and antiseptic products that form a homogeneous, physically stable preparation when diluted with hard water, or - in the case of ready-to-use products - with water. Products can only be tested at a concentration of 80 % or less (97 % with a modified method for special cases) as some dilution is always produced by adding the test organisms and interfering substance.
This European Standard applies to products that are used in the medical area in the fields of hygienic handrub, hygienic handwash, surgical handrub, surgical handwash, instrument disinfection by immersion, and surface disinfection by wiping, spraying, flooding or other means.
This European Standard applies to areas and situations where disinfection or antisepsis is medically indicated. Such indications occur in patient care, for example:
-   in hospitals, in community medical facilities and in dental institutions;
-   in clinics of schools, of kindergartens and of nursing homes;
and may occur in the workplace and in the home. It may also include services such as laundries and kitchens supplying products directly for the patients.
NOTE 1   The method described is intended to determine the activity of commercial formulations or active substances under the conditions in which they are used.
NOTE 2   This method corresponds to a phase 2 step 1 test.
NOTE 3   This method cannot be used to evaluate the activity of products against Legionella in watersystems and against mycobacteria.
EN 14885 specifies in detail the relationship of the various tests to one another and to 'use recommendations'.

  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This European Standard specifies a test method and the minimum requirements for bactericidal activity of chemical disinfectant and antiseptic products that form a homogeneous, physically stable preparation when diluted with hard water, or - in the case of ready-to-use products - with water. Products can only be tested at a concentration of 80 % or less (97 % with a modified method for special cases) as some dilution is always produced by adding the test organisms and interfering substance.
This European Standard applies to products that are used in the medical area in the fields of hygienic handrub, hygienic handwash, surgical handrub, surgical handwash, instrument disinfection by immersion, and surface disinfection by wiping, spraying, flooding or other means.
This European Standard applies to areas and situations where disinfection or antisepsis is medically indicated. Such indications occur in patient care, for example:
-   in hospitals, in community medical facilities and in dental institutions;
-   in clinics of schools, of kindergartens and of nursing homes;
and may occur in the workplace and in the home. It may also include services such as laundries and kitchens supplying products directly for the patients.
NOTE 1   The method described is intended to determine the activity of commercial formulations or active substances under the conditions in which they are used.
NOTE 2   This method corresponds to a phase 2 step 1 test.
NOTE 3   This method cannot be used to evaluate the activity of products against Legionella in watersystems and against mycobacteria.
EN 14885 specifies in detail the relationship of the various tests to one another and to 'use recommendations'.

  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11137-2:2013 specifies methods for determining the minimum dose needed to achieve a specified requirement for sterility and methods to substantiate the use of 25 kGy or 15 kGy as the sterilization dose to achieve a sterility assurance level, SAL, of 10−6. ISO 11137-2:2013 also specifies methods of sterilization dose audit used to demonstrate the continued effectiveness of the sterilization dose.
ISO 11137-2:2013 defines product families for sterilization dose establishment and sterilization dose audit.

  • Standard
    81 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11137-1:2006 specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. Although the scope of ISO 11137-1:2006 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment.
ISO 11137-1:2006 covers radiation processes employing irradiators using the radionuclide 60Co or 137Cs, a beam from an electron generator or a beam from an X-ray generator.
ISO 11137-1:2006 does not:
specify requirements for development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease;detail specified requirements for designating a medical device as sterile;specify a quality management system for the control of all stages of production of medical devices;specify requirements for occupational safety associated with the design and operation of irradiation facilities;specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    58 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 11135:2014 specifies requirements for the development, validation and routine control of an ethylene oxide sterilization process for medical devices in both the industrial and health care facility settings, and it acknowledges the similarities and differences between the two applications.

  • Standard
    90 pages
    English language
    sale 10% off
    e-Library read for
    1 day