SIST EN ISO 10720:2007 is the SIST adoption of EN ISO 10720:2007 for steel and iron. It specifies a thermal conductimetric method after fusion in a current of inert gas for determining nitrogen content in steel and iron over the range 0,0008 % (m/m) to 0,5 % (m/m). Laboratories that analyze ferrous materials use it to generate comparable nitrogen results and report them in a standard form.
What does SIST EN ISO 10720:2007 specify?
SIST EN ISO 10720:2007 specifies how to determine nitrogen in steel and iron by fusing a test portion in helium, separating the released gases, and measuring nitrogen by thermal conductivity. The document is organized into scope, normative references, principle, reagents and materials, apparatus, sampling, procedure, expression of results, precision, and test report.
| Annex | What it covers |
|---|---|
| Annex A | Features of commercial resistive electrode furnaces and nitrogen analysers |
| Annex B | Additional information on the international cooperative tests and test samples |
| Annex C | Graphical representation of the precision data |
What are the key requirements of SIST EN ISO 10720:2007?
Principle and measurement range
SIST EN ISO 10720:2007 applies to nitrogen contents between 0,0008 % (m/m) and 0,5 % (m/m). Clause 3 states that the test portion is fused in a single-use graphite crucible under helium at high temperature, the nitrogen is extracted as molecular nitrogen, and the analyser measures it by thermal conductimetry.
In practice, this means the laboratory needs an instrument that can fuse the sample cleanly, carry the gas stream, and read the thermal conductivity signal reproducibly.
Sample preparation, blanks, and calibration
Clause 7 requires the test sample to be degreased before weighing, and it gives separate handling for lower and higher nitrogen levels. For lower nitrogen contents, the method uses a nickel capsule; for higher contents, it uses a degassed graphite crucible.
The blank test is essential. Clause 7.3 requires duplicate blanks, correction for nitrogen in the pure iron used, and investigation if the blank values are too high or too far apart. Clause 7.5 then builds a calibration graph from potassium nitrate standard solutions, with one calibration range for samples up to 0,10 % (m/m) and another for samples between 0,10 % (m/m) and 0,50 % (m/m). In practice, this is the traceability step that links the analyser reading to a nitrogen mass.
Instrument control and operating checks
Clause 7.1 requires the user to keep filters clean, verify reagent performance with a certified reference material, and stabilise the instrument after filter or reagent changes or after downtime. The clause also notes that some instruments need the sample introduction pipe cleaned to remove carbon deposits.
This matters because drift, contamination, or unstable heating power can change the reading before the operator notices.
Calculation, precision, and reporting
Clause 8 gives the calculation of nitrogen content from the analyser reading, the blank, and the test portion mass. It also provides precision data from a planned interlaboratory trial and links the repeatability and reproducibility limits to nitrogen level.
Clause 9 requires the test report to identify the sample, laboratory, and date; state the method by this standard; give the result in its reported form; and record any unusual features or any operation not specified in the standard.
What terms does SIST EN ISO 10720:2007 define?
- Thermal conductimetric method - A measurement method that determines nitrogen by comparing the thermal conductivity of the gas stream with and without the extracted nitrogen.
- Test portion - The weighed piece of sample that is actually fused and measured.
- Blank test - A control determination used to measure background nitrogen from reagents, crucibles, and handling.
- Calibration graph - The plot that links analyser signal to known amounts of nitrogen so sample readings can be converted into content.
- Repeatability limit - The closeness expected between results obtained under the same conditions in a short time.
- Reproducibility limit - The closeness expected between results from different laboratories or conditions within the precision study.
- Within-laboratory reproducibility limit (Rw) - The longer-interval precision measure used in the interlaboratory study for results obtained in the same laboratory on different days.
Who uses SIST EN ISO 10720:2007?
SIST EN ISO 10720:2007 is used by chemical-analysis laboratories that measure nitrogen in steel and iron, including laboratories serving mills, foundries, and other ferrous-product producers. It is also relevant to quality teams and buyers who need nitrogen results to compare material batches, verify specifications, or check laboratory performance.
What changed in SIST EN ISO 10720:2007 from the previous edition?
SIST EN ISO 10720:2007 takes over ISO 10720:1997 through EN ISO 10720:2007 without modification. The European approval notice states that the text was adopted as EN ISO 10720:2007 without technical changes.
Which standards are used with SIST EN ISO 10720:2007?
| Standard | What it contributes |
|---|---|
| ISO 385-1 | General requirements for burettes |
| ISO 648 | One-mark pipettes |
| ISO 1042 | One-mark volumetric flasks |
| ISO 3696 | Water for analytical laboratory use |
| ISO 5725-1 | General principles and definitions for accuracy |
| ISO 5725-2 | Repeatability and reproducibility determination |
| ISO 5725-3 | Intermediate precision measures |
| ISO 14284 | Sampling and preparation of samples for chemical composition analysis |
These references support the laboratory glassware, water quality, sampling, and precision-statistics parts of the method.
What does the SIST EN ISO 10720:2007 document contain?
The document contains the operating principle, reagent list, apparatus requirements, sampling instruction, analytical procedure, calculation, precision data, and reporting rules. It also includes calibration tables for the low- and high-range nitrogen levels, a precision table based on interlaboratory testing, and a graph showing the logarithmic relationship between nitrogen content and precision limits. Annex A describes typical furnace and analyser features, while Annexes B and C give the cooperative-test background and the precision plots used to support the method.
Relations
- Effective Date
- 25-Aug-2026
- Effective Date
- 25-Aug-2026
- Effective Date
- 25-Aug-2026
- Effective Date
- 25-Aug-2026
- Effective Date
- 25-Aug-2026
- Effective Date
- 25-Aug-2026
- Effective Date
- 25-Aug-2026
- Effective Date
- 09-Feb-2026
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
- Refers
SIST EN 1414:2000/A1:2000 - Touch and close fasteners - Cycling procedure for subsequent testing - Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
- Referred By
SIST-TP CEN/TR 10261:2008 - Iron and steel - Review of available methods of chemical analysis - Effective Date
- 28-Jan-2026
Get Certified
Connect with accredited certification bodies for this standard

Element Materials Technology
Materials testing and product certification.

Inštitut za kovinske materiale in tehnologije
Institute of Metals and Technology. Materials testing, metallurgical analysis, NDT.
Sponsored listings
Frequently Asked Questions
SIST EN ISO 10720:2007 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Steel and iron - Determination of nitrogen content - Thermal conductimetric method after fusion in a current of inert gas (ISO 10720:1997)". This standard covers: Method for determination of nitrogen content between 0,0008 % and 0,5 % in steel and iron.
Method for determination of nitrogen content between 0,0008 % and 0,5 % in steel and iron.
SIST EN ISO 10720:2007 is classified under the following ICS (International Classification for Standards) categories: 77.080.01 - Ferrous metals in general. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN ISO 10720:2007 has the following relationships with other standards: It is inter standard links to SIST ISO 1042:1998, SIST ISO 648:1995, SIST ISO 385-1:1995, SIST ISO 14284:2000, SIST ISO 5725-1:2003, SIST ISO 5725-2:2003, SIST ISO 5725-3:2003, SIST ISO 3696:1995, SIST EN 15649-3:2010, SIST EN ISO 9934-1:2002, SIST ENV 14237:2002, SIST EN 546-1:2007, SIST EN 1414:2000/A1:2000, SIST EN 602:2004, SIST-TP CEN/TR 10261:2008. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN ISO 10720:2007 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Principle
- 4 Reagents and materials
- 5 Apparatus
- 6 Sampling
- 7 Procedure
- 7.1 General instructions
- 7.2 Test portion
- 7.3 Blank test
- 7.4 Determination
- 7.5 Establishment of the calibration graph
- 7.51 Preparation of the calibration series
- 4.8.9 0,250 50 0,005 0
- 4.8.6 2,50 500 0,050
- 4.8.5 4.8.4 5,00 1000 0,100
- 4.8.2 12.5 2 500 0,500
- 4.8.1 1) Zero member
- 8 Expression of results
- 8.1 Method of calculation
- 8.2 Precision
- 9 Test report
- Annex A
- Annex B
- Annex C
1 Scope
This International Standard specifies a thermal conductimetric method after fusion under inert gas for the determination of nitrogen in steel and iron.
The method is applicable to nitrogen contents between 0,000 8 % (m/m) and 0,5% (m/m).
2 Normative references
The following standards contain provisions which, through reference in this text, constitute provisions of this International Standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this International Standard are encouraged to investigate the possibility of applying the most recent editions of the standards indicated below. Members of IEC and IS0 maintain registers of currently valid International Standards.
IS0 385-l :I 984, Laboratory glassware - Burettes - Part 1: General requirements. IS0 648:1977, Laboratory glassware - One-mark pipettes.
IS0 1042: ----I), Laboratory glassware - One-mark volumetric flasks. IS0 3696:1987, Water for analytical laboratory use - Specification and test methods. IS0 5725-l :I 994, Accuracy (trueness and precision) of measurement methods and results - Part I: General principles and definitions.
IS0 5725-2:1994, Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method. IS0 5725-3:1994, Accuracy (trueness and precision) of measurement methods and results - Part 3: Intermediate measures of the precision of a standard measurement method.
- Sampling and preparation of samples for the determination of chemical IS0 14284 :I 996, Steel and iron composition.
3 Principle
Fusion of a test portion in a single-use graphite crucible under helium gas at a high temperature (e. g. 2 200 “C) Extraction of the nitrogen in the form of molecular nitrogen in the stream of helium. Separation from the other gaseous extracts and measurement by thermal conductimetric method.
4 Reagents and materials
During the analysis, unless otherwise stated, use only reagents of recognized analytical grade and only grade 2 water as specified in IS0 3696.
1) To be published. (Revision of IS0 10427 983)
@ IS0 IS0 10720: 1997(E) 4.1 Water, prepare just before use.
4.2 Helium, high purity, total impurity content 0,000 5 % (m/m). An oxidation reagent or catalyst [copper oxide or platinum] tube heated to a temperature above 450 “C shall be used prior to a purifying unit, when the presence of organic contaminants is suspected in the helium. 4.3 Pure iron, of known low nitrogen contents less than 0,001 % (m/m). 4.4 Copper (II) oxide, on granulated support.
4.5 Anhydrous magnesium perchlorate, Mg(CIO,),, particle size from I,2 mm to 2,O mm, or anhydrous calcium sulfate, particle size from 0,60 mm to 0,85 mm.
4.6 Sodium hydroxide, on granulated support.
Particle size from 0,7 mm to I,2 mm.
Appropriate solvent, suitable for washing greasy or dirty test samples, e.g. acetone. 4.7 4.8 Potassium nitrate, standard solution.
After drying at between 100 “C and 105 OC for 2 h and allowing to cool in a desiccator, weigh, to the nearest 0,l mg, the masses of potassium nitrate [purity > 99,9 % (m/m)] indicated in table 1. Dissolve the potassium nitrate in about 50 ml of water (4.1), transfer quantitatively to a 100 ml one-mark volumetric flask, dilute to the mark with water (4.1) and mix.
1 ml of each standard solution contains the mass of nitrogen indicated in table 1. Table 1 - Standard solutions Name of the Mass of potassium Corresponding standard nitrate used concentration solution of nitrogen mg/ml
12,5 4.8.1 9,022 8 4.8.2 7,218 2 IO,0 4.8.3 5,413 8
4.8.4 3,609 1
4.8.5 1,804 6 2,50 4.8.6 0,902 3 I,25 4.8.7 0,360 9 0,50 4.8.8 [4.8.5 x l/10] ') 0,25 4.8.9 [4.8.6 x l/IO] 0,125 4.8.10 [4.8.7 x l/IO] 0,050 1) e.g. .transfer 10,O ml of the standard solution (4.8.5) into a 100 ml one-mark volumetric flask, dilute to the mark with water (4.1) and mix.
5 Apparatus
During the analysis, unless otherwise stated, use only ordinary laboratory apparatus.
@ IS0 IS0 10720: 1997(E) All volumetric glassware shall be class A, in accordance with IS0 385-l) IS0 648 or IS0 1042 as appropriate. The apparatus required for fusion of the test portion, separation and measurement of the nitrogen extracted may be obtained commercially from a number of manufacturers. Follow the manufacturer’s instructions for the operation of the instrument.
Features of commercial instruments are given in annex A.
5.1 Graphite crucible, single-use.
Use high purity crucibles suitable for use with the apparatus. 5.2 Micropipette, 100 ul and 200 ~1, limit of error shall be less than 1 ~1. 5.3 Nickel capsule.
For example, about 6 mm in diameter; 8 mm in height; 0,2 g in mass and 0,23 ml in volume, or about 6 mm in diameter; 125 mm in height; 0,5 g in mass and 0,35 ml in volume. In any case, the nitrogen content shall be less than 0,000 2 % (mlm).
5.4 Crucible tongs, for handling the crucibles used.
5.5 Glass-wool filters.
6 Sampling
Carry out sampling in accordance with IS0 14284 or appropriate national standards for steel and iron.
7 Procedure
WARNING - The risks involved when using an apparatus for fusing the test portion are mainly risks of burns. It is therefore essential to use crucible tongs (5.4) and appropriate containers for the used crucibles.
7.1 General instructions
Keep the glass-wool filters (5.5) clean. Using a certified reference material, verify the effectiveness of the installed reagents (4.4, 4.5 and 4.6) and change them if necessary.
In certain instruments, it is necessary to clean the sample introduction pipe in the furnace after each analysis in order to eliminate carbon deposits. If the electricity supply has been switched off for a long time, allow time for the instrument to stabilize as recommended by the manufacturer.
After changing the filters (5.5) and/or reagents (4.4, 4.5 and 4.6), or when the apparatus has been inoperative for a period, stabilize the instrument by carrying out trial analyses, the results of which are to be disregarded, then proceed with calibration as indicated in 7.5 before analysing the sample. If the instrument used provides a direct reading in percentage of nitrogen, adjust the instrument reading for each calibration range as follows.
Read the content of a certified reference material of high nitrogen content at various power settings. The required heating power for the determination of test samples is that at which the reading levels off. In order to determine a high alloy test sample a high alloy certified reference material shall be used to know the required heating power.
@ IS0 IS0 10720: 1997(E)
7.2 Test portion
Degrease the test sample by washing in a suitable solvent (4.7). Evaporate the last traces of the washing liquid by heating.
Weigh, to the nearest 1 mg, approximately ,0 g of the test sample for nitrogen conte nts up to 0,l % ( m/m) approximately 0,50 g for nitrogen contents reater than 0,l % (m/m) (see note . 1)
NOTE 1 The mass of the test portion may be dependent on the type of instrument used.
7.3 Blank test
Prior to the determination, carry out the following blank tests in duplicate.
...