Water meters for cold potable water and hot water - Part 3: Test report format

ISO 4064-3:2014|OIML R 49-3:2013 specifies a test report format to be used in conjunction with ISO 4064-1:2014|OIML R 49-1:2013 and ISO 4064-2:2014|OIML R 49-2:2013 for water meters for cold potable water and hot water.

Compteurs d'eau potable froide et d'eau chaude — Partie 3: Format du rapport d'essais

L'ISO 4064-3:2014|OIML R 49-3:2013 spécifie le format du rapport d'essai à utiliser dans le contexte de la mise en oeuvre de l'ISO 4064‑1:2014|OIML R 49-1:2013 et de l'ISO 4064‑2:2014|OIML R 49-2:2013 pour les compteurs d'eau potable froide et d'eau chaude.

General Information

Status
Published
Publication Date
25-May-2014
Current Stage
9599 - Withdrawal of International Standard
Start Date
19-Dec-2024
Completion Date
13-Dec-2025

Relations

Effective Date
17-Dec-2022
Effective Date
23-Aug-2014
Effective Date
17-Mar-2012

Overview

ISO 4064-3:2014 - “Water meters for cold potable water and hot water - Part 3: Test report format” specifies a standardized test report format to be used together with ISO 4064-1:2014 (metrological and technical requirements) and ISO 4064-2:2014 (test methods) for water meters. This fourth edition (identical to OIML R 49-3:2013) defines the structure, fields and checklists for type evaluation and initial verification reports for mechanical, electronic and separable water‑meter components.

Keywords: ISO 4064-3, water meter test report format, OIML R 49-3, water meters, cold potable water, hot water, type evaluation, initial verification.

Key topics and technical requirements

ISO 4064-3 focuses on the format and content of test reports rather than on new test procedures. Main elements include:

  • Scope and normative references linking to ISO 4064‑1 and ISO 4064‑2.
  • Terms, definitions and symbols used in reports (for example: + = pass, − = fail, EUT = equipment under test).
  • Type evaluation report template (Clause 4) covering:
    • Applicant, authorized representative, and testing laboratory identification.
    • Model submission checklist (mechanical, electronic, combined, separable units).
    • Detailed type information: measuring principle, accuracy class, temperature class, environmental class, electromagnetic environment.
    • Maximum admissible temperature and pressure, orientation limits, and installation details (mechanical and electrical).
    • EUT testing requirements and category/case per ISO 4064‑2.
    • Checklists and pass/fail tables for examinations and performance tests.
  • Initial verification report template (Clause 5) and example tables for separable units.
  • Annexes: normative lists of documents concerning the type (Annex A) and a listing of test equipment used (Annex B).

Practical applications

Implementing ISO 4064-3 brings practical benefits:

  • Consistent, complete test reports that ease peer review, certification and regulatory acceptance.
  • Streamlined type evaluation and initial verification across manufacturers, test laboratories and conformity assessment bodies.
  • Clear documentation for legal metrology, procurement, and installation records.
  • Interoperability for separable calculators, sensors and supplementary electronic devices through standardized report fields.

Who should use this standard

  • Water meter manufacturers preparing type approval dossiers
  • Accredited testing and calibration laboratories
  • National metrology institutes and legal metrology authorities
  • Conformity assessment bodies and certification schemes (including OIML certificate system)
  • Utilities and procurement teams requiring standardized test documentation

Related standards

  • ISO 4064‑1:2014 - Metrological and technical requirements
  • ISO 4064‑2:2014 - Test methods
  • OIML R 49 series (corresponding legal metrology guidance)

Using ISO 4064-3 helps ensure test reports are complete, auditable and aligned with international metrological and legal‑metrology requirements for water meters.

Standard

ISO 4064-3:2014 - Water meters for cold potable water and hot water

English language
71 pages
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 4064-3:2014 - Compteurs d'eau potable froide et d'eau chaude

French language
97 pages
sale 15% off
Preview
sale 15% off
Preview

Frequently Asked Questions

ISO 4064-3:2014 is a standard published by the International Organization for Standardization (ISO). Its full title is "Water meters for cold potable water and hot water - Part 3: Test report format". This standard covers: ISO 4064-3:2014|OIML R 49-3:2013 specifies a test report format to be used in conjunction with ISO 4064-1:2014|OIML R 49-1:2013 and ISO 4064-2:2014|OIML R 49-2:2013 for water meters for cold potable water and hot water.

ISO 4064-3:2014|OIML R 49-3:2013 specifies a test report format to be used in conjunction with ISO 4064-1:2014|OIML R 49-1:2013 and ISO 4064-2:2014|OIML R 49-2:2013 for water meters for cold potable water and hot water.

ISO 4064-3:2014 is classified under the following ICS (International Classification for Standards) categories: 91.140.60 - Water supply systems. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 4064-3:2014 has the following relationships with other standards: It is inter standard links to ISO 4064-3:2024, ISO 4064-1:2005, ISO 4064-3:2005. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase ISO 4064-3:2014 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.

Standards Content (Sample)


INTERNATIONAL ISO
STANDARD 4064-3
Fourth edition
2014-06-01
Water meters for cold potable water
and hot water —
Part 3:
Test report format
Compteurs d’eau potable froide et d’eau chaude —
Partie 3: Format du rapport d’essais
Reference number
©
ISO 2014
© ISO 2014
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2014 – All rights reserved

Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms, definitions, symbols, and abbreviated terms . 1
4 Type evaluation report . 1
4.1 General . 1
4.2 Information concerning the type . 2
4.3 General information concerning the test equipment .12
4.4 Check list for water meter examinations and performance tests .13
4.5 Type evaluation tests (for all water meters) .23
4.6 Type evaluation tests (for electronic water meters and mechanical water meters with
electronic components).47
5 Initial verification report.68
5.1 General .68
5.2 Information concerning the EUT verified .68
5.3 Initial verification test report (ISO 4064-2:2014|OIML R 49-2:2013, Clause 10) .68
Annex A (normative) List of documents concerning the type (ISO 4064‑1:2014|OIML R 49‑1:2013,
7.2.9) .73
Annex B (normative) Listing of test equipment used in examinations and tests .74
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2, www.iso.org/directives.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any
patent rights identified during the development of the document will be in the Introduction and/or on
the ISO list of patent declarations received, www.iso.org/patents.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
The committees responsible for this document are Technical Committee ISO/TC 30, Measurement of fluid
flow in closed conduits, Subcommittee SC 7, Volume methods including water meters and OIML Technical
Subcommittee TC 8/SC 5 Water meters.
This fourth edition of ISO 4064-3 cancels and replaces the third edition (ISO 4064-3:2005), which has
been technically revised. Provisions of the third edition are addressed in ISO 4064-2:2014|OIML R 49-
2:2013.
ISO 4064 consists of the following parts, under the general title Water meters for cold potable water and
hot water:
— Part 1: Metrological and technical requirements
— Part 2: Test methods
— Part 3: Test report format
— Part 4: Non-metrological requirements not covered in ISO 4064-1
— Part 5: Installation requirements
This edition of ISO 4064-3 is identical to the corresponding edition of OIML R 49-3, which has been
issued concurrently. OIML R 49-3 was approved for final publication by the International Committee of
Legal Metrology at its 48th meeting in Ho Chi Minh City, Vietnam in October 2013. It will be submitted
to the International Conference on Legal Metrology in 2016 for formal sanction.
iv © ISO 2014 – All rights reserved

Introduction
Implementation of this test report format is informative with regard to the implementation of
ISO 4064-1|OIML R 49-1 and ISO 4064-2|OIML R 49-2 in national regulations; however, its implementation
is required within the framework of the OIML Certificate System for Measuring Instruments
[ISO 4064-2:2014|OIML R 49-2:2013, 11.1].
Clause 4 shows the required format of a type evaluation report for a complete or combined water meter.
A type evaluation report for a separable calculator (including indicating device) or a measurement
transducer (including flow or volume sensor) requires a similar format. However, some modifications to
the tables may be required because a large number of variations in the design of these separable units
is possible.
Some examples of tables for presenting the test results for separable units are shown in Clause 5 for
initial verifications. These tables can also be adapted for type evaluation reports.
INTERNATIONAL STANDARD ISO 4064-3:2014(E)
Water meters for cold potable water and hot water —
Part 3:
Test report format
1 Scope
This part of ISO 4064|OIML R 49 specifies a test report format to be used in conjunction with
ISO 4064-1|OIML R 49-1 and ISO 4064-2|OIML R 49-2 for water meters for cold potable water and hot
water.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
ISO 4064-1:2014|OIML R 49-1:2013, Water meters for cold potable water and hot water — Part 1:
Metrological and technical requirements
ISO 4064-2:2014|OIML R 49-2:2013, Water meters for cold potable water and hot water — Part 2: Test
methods
3 Terms, definitions, symbols, and abbreviated terms
For the purposes of this document, the terms and definitions given in ISO 4064-1|OIML R 49-1 apply.
Some symbols and abbreviated terms used in the tables are as follows.
+ pass
− fail
n/a not applicable
EUT equipment under test
H horizontal
MAP maximum admissible pressure
MAT maximum admissible temperature
MPE maximum permissible error
V vertical
4 Type evaluation report
4.1 General
For each examination and test the checklist shall be completed according to this example:
+ −
× Pass
× Fail
n/a n/a Not applicable
4.2 Information concerning the type
4.2.1 General
Application _________________________________
number:
Applicant: _________________________________
Authorized _________________________________
representative:
Address: _________________________________
_________________________________
_________________________________
_________________________________
Testing _________________________________
laboratory:
Authorized _________________________________
representative:
Address: _________________________________
_________________________________
_________________________________
_________________________________
4.2.2 Model submitted
New model: _________________________________
Variant of approved model(s):
Approval number: _________________________________
Variation of approved model: _________________________________
See Table 1.
2 © ISO 2014 – All rights reserved

Table 1 — Model submitted
a a
Submitted for approval tests Yes No Remarks
Mechanical water meter (complete)
Mechanical water meter (combined)
Electronic water meter (complete)
Electronic water meter (combined)
Family of water meters
Separable calculator (including indicating device)
Separable measurement transducer (including flow
or volume sensor)
Supplementary electronic device(s) for testing (per-
manently attached to meter)
Supplementary electronic device(s) for data trans-
mission (permanently attached to meter)
Supplementary electronic device(s) for testing (tem-
porarily attached to meter)
Supplementary electronic device(s) for data trans-
mission (temporarily attached to meter)
Ancillary devices
a
Tick as appropriate.
4.2.3 Mechanical water meter (complete or combined)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Type details:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
for combination meters
Q _________ m /h
x1
Q _________ m /h
x2
Measuring principle: __________________________
Accuracy class: _____________
Temperature class: _____________
Environmental class: _____________
Electromagnetic environment: _____________
Maximum admissible _________ °C
temperature:
Maximum admissible pressure: ___________ MPa (_____ bar)
Orientation limitation: _________________________
EUT testing requirements (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Category: _________________________
Case: _________________________
Installation details:
Connection type (flange, screw _________
thread, concentric manifold):
Minimum straight length of inlet __________ mm
pipe:
Minimum straight length of out- __________ mm
let pipe:
Flow conditioner (details if _________________________
required):
Mounting: _________________________
Orientation: _________________________
Other relevant information: _________________________
NOTE If a family of meters is submitted, the details in this subclause are to be given for each size of water
meter.
4 © ISO 2014 – All rights reserved

4.2.4 Electronic water meter (complete or combined)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Type details:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
for combination meters
Q _________ m /h
x1
Q _________ m /h
x2
Measuring principle: __________________________
Accuracy class: _____________
Temperature class: _____________
Environmental class: _____________
Electromagnetic environment: _____________
Maximum admissible temperature: _________ °C
Maximum admissible pressure: ___________ MPa (_____ bar)
Orientation limitation: _________________________
EUT testing requirements (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Category: _________________________
Case: _________________________
Installation details (mechanical):
Connection type (flange, screw thread, _________
concentric manifold):
Minimum straight length of inlet pipe: __________ mm
Minimum straight length of outlet pipe: __________ mm
Flow conditioner (details if required): _________________________
Mounting: _________________________
Orientation: _________________________
Other relevant information: _________________________
Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
U : ______________________ V
max
U : ______________________ V
min
Frequency: _______________________ Hz
NOTE If a family of meters is submitted, the details in this subclause are to be given for each size of water
meter.
4.2.5 Separable calculator (including indicating device)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Type details:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
6 © ISO 2014 – All rights reserved

for combination meters
Q _________ m /h
x1
Q _________ m /h
x2
Measuring principle: __________________________
Accuracy class: _____________
Temperature class: _____________
Environmental class: _____________
Electromagnetic environment: _____________
Maximum admissible temperature: _________°C
Maximum admissible pressure: ___________MPa (_____ bar)
Orientation limitation: _________________________
EUT testing requirements (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Category: _________________________
Case: _________________________
Maximum relative error specified by the manufacturer:
Lower flow rate zone, Q ≤ Q < Q : ________________________%
1 2
Upper flow rate zone, Q ≤ Q ≤ Q : ________________________%
2 4
Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
U : ________________________ V
max
U : ________________________ V
min
Frequency: _______________________ Hz
Approval number(s) of compatible meas-
urement transducer(s) (including flow or
volume sensor): ____________________
4.2.6 Separable measurement transducer (including flow or volume sensor)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Type details:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
for combination meters
Q _________ m /h
x1
Q _________ m /h
x2
Measuring principle: __________________________
Accuracy class: _____________
Temperature class: _____________
Environmental class: _____________
Electromagnetic environment: _____________
Maximum admissible temperature: _________°C
Maximum admissible pressure: ___________MPa (_____ bar)
Orientation limitation: _________________________
EUT testing requirements (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Category: _________________________
Case: _________________________
Maximum relative error specified by the manufacturer:
Lower flow rate zone, Q ≤ Q < Q : ________________________%
1 2
Upper flow rate zone, Q ≤ Q ≤ Q : ________________________%
2 4
8 © ISO 2014 – All rights reserved

Installation details (mechanical):
Connection type (flange, screw thread, concentric mani- _________
fold):
Minimum straight length of inlet pipe: __________mm
Minimum straight length of outlet pipe: __________mm
Flow conditioner (details if required): _________________________
Mounting: _________________________
Orientation: _________________________
Other relevant information: _________________________
Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
U : ________________________ V
max
U : ________________________ V
min
Frequency: _______________________ Hz
Approval number(s) of compatible calculator(s)
(including indicating device): _________________________
4.2.7 Supplementary electronic device(s) used for testing (permanently attached to meter)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
: ________________________ V
U
max
U : ________________________ V
min
Frequency: _______________________ Hz
Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
4.2.8 Supplementary electronic device(s) used for data transmission (permanently attached to
meter)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
U : ________________________ V
max
U : ________________________ V
min
Frequency: _______________________ Hz
Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
4.2.9 Supplementary electronic device(s) used for testing (temporarily attached to meter)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
U : ________________________ V
max
U : ________________________ V
min
Frequency: _______________________ Hz
10 © ISO 2014 – All rights reserved

Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
4.2.10 Supplementary electronic device(s) used for data transmission (temporarily attached to
meter)
Manufacturer: __________________________________________________
Model number: __________________________________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
U : ________________________ V
max
U : ________________________ V
min
Frequency: _______________________ Hz
EUT testing requirements (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Category: _________________________
Case: _________________________
Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
4.2.11 Ancillary devices
Manufacturer: __________________________________________________
Model number: __________________________________________________
Power supply:
Type (battery, mains AC, mains DC): _________________________
U : ________________________ V
max
U : ________________________ V
min
Frequency: _______________________ Hz
Approval number(s) of compatible calculator(s)
(including indicating device): _________________________
EUT testing requirements (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Category: _________________________
Case: _________________________
Installation details (electrical):
Wiring instructions: _________________________
Mounting arrangement: _________________________
Orientation limitations: _________________________
Approval number(s) of compatible water
meters, calculator(s) (including indicating
device) and measurement transducer(s)
(including flow or volume sensor): _________________________
4.2.12 Documents concerning the type
A list of documents shall be submitted with the type approval application as in Annex A.
4.3 General information concerning the test equipment
Details of all items of measuring equipment and test instruments used for the type examinations, and
initial verifications shall be listed in Annex B, including:
Manufacturer
Model number
Serial number
Date of last calibration
Date of next calibration due of e.g. instruments for measuring:
— linear dimensions
— pressure gauges
— pressure transmitters
— manometers
— temperature transducers
12 © ISO 2014 – All rights reserved

— reference meters
— volume tanks
— weighing machines
— signal generators (for pulse, current or voltage)
4.4 Check list for water meter examinations and performance tests
4.4.1 Check list for water meter examinations
External examination for all water meters
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
Function of the indicating device
6.7.1.1 The indicating device shall provide an easily read, reliable
and unambiguous visual indication of the indicated volume
6.7.1.1 The indicating device shall include visual means for testing
and calibration.
6.7.1.1 The indicating device may include additional elements for
testing and calibration by other methods, e.g. for automatic
testing and calibration
Unit of measurement and its placement
6.7.1.2 The indicated volume of water shall be expressed in cubic
metres
6.7.1.2 The symbol m shall appear on the dial or immediately adja-
cent to the numbered display
Indicating range
6.7.1.3 For Q ≤ 6,3, the minimum indicating range is 0 m to
9 999 m
6.7.1.3 For 6,3 < Q ≤ 63, the minimum indicating range is 0 m to
99 999 m
6.7.1.3 For 63 < Q ≤ 630, the minimum indicating range is 0 m to
999 999 m
6.7.1.3 For 630 < Q ≤ 6 300, the minimum indicating range is 0 m
to 9 999 999 m
Colour coding for indicating device
6.7.1.4 The colour black should be used to indicate the cubic metre
and its multiples
6.7.1.4 The colour red should be used to indicate sub-multiples of a
cubic metre
6.7.1.4 The colours shall be applied to either the pointers, indexes,
numbers, wheels, discs, dials or aperture frames
6.7.1.4 Other means of indicating the cubic metre may be used
provided there is no ambiguity in distinguishing between
the primary indication and alternative displays, e.g. sub-
multiples for verification and testing
Types of indicating device: Type 1 — Analogue device
External examination for all water meters
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
6.7.2.1 The indicated volume shall be shown by continuous move-
ment of either:
a) one or more pointers moving relative to graduated
scales; or
b) one or more circular scales or drums each passing
an index
6.7.2.1 The value expressed in cubic metres for each scale division
n
shall be of the form 10 , where n is a positive or a negative
whole number or zero, thereby establishing a system of
consecutive decades.
6.7.2.1 The scale shall be graduated in values expressed in cubic
metres or accompanied by a multiplying factor
(×0,001; ×0,01; ×0,1; ×1; ×10; ×100; ×1 000 etc.)
6.7.2.1 Rotational movement of the pointers or circular scales shall
be clockwise
6.7.2.1 Linear movement of pointers or scales shall be left to right
6.7.2.1 Movement of numbered roller indicators shall be upwards
Types of indicating device: Type 2 — Digital device
6.7.2.2 The indicated volume is given by a line of digits appearing in
one or more apertures
6.7.2.2 The advance of one digit shall be completed while the digit
of the next immediately lower decade changes from 9 to 0
6.7.2.2 The apparent height of the digits shall be at least 4 mm
6.7.2.2 For non-electronic devices, movement of numbered roller
indicators (drums) shall be upwards
6.7.2.2 For non-electronic devices, the lowest value decade may
have a continuous movement, the aperture being large
enough to permit a digit to be read without ambiguity
6.7.2.2 For electronic devices with non-permanent displays the vol-
ume shall be able to be displayed at any time for at least 10 s
6.7.2.2 For electronic devices, the meter shall provide visual check-
ing of the entire display which shall have the following
sequence:
— for seven segment type displaying all the elements (e.g.
an “eights” test); and
— for seven segment type blanking all the elements (a
“blanks” test).
For graphical displays, an equivalent test is required to dem-
onstrate that display faults cannot result in any digit being
misinterpreted.
Each step of the sequence shall last at least 1 s
Types of indicating device: Type 3 — Combination of analogue and digital devices
6.7.2.3 The indicated volume is given by a combination of type 1
and type 2 devices and the respective requirements of each
shall apply
Verification devices — General requirements
14 © ISO 2014 – All rights reserved

External examination for all water meters
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
6.7.3.1 Every indicating device shall provide means for visual, non-
ambiguous verification testing and calibration
6.7.3.1 The visual verification may have either a continuous or a
discontinuous movement
6.7.3.1 In addition to the visual verification display, an indicating
device may include provisions for rapid testing by the inclu-
sion of complementary elements (e.g. star wheels or discs),
providing signals through externally attached sensors.
Verification devices — Visual verification displays
6.7.3.2.1 The value of the verification scale interval, expressed in
n n n
cubic metres, shall be of the form:1 × 10 , 2 × 10 or 5 × 10 ,
where n is a positive or negative whole number, or zero
6.7.3.2.1 The indicated volume is given by a line of digits appearing in
one or more apertures
6.7.3.2.1 For analogue or digital indicating devices with continuous
movement of the first element, the verification scale interval
may be formed from the division into 2, 5 or 10 equal parts
of the interval between two consecutive digits of the first
element. Numbering shall not be applied to these divisions
6.7.3.2.1 For digital indicating devices with discontinuous move-
ment of the first element, the verification scale interval is
the interval between two consecutive digits or incremental
movements of the first element
6.7.3.2.2 On indicating devices with continuous movement of the first
element, the apparent scale spacing shall be not less than
1 mm and not more than 5 mm
6.7.3.2.2 The scale shall consist of either:
a) lines of equal thickness not exceeding one quarter of
the scale spacing and differing only in length; or
b) contrasting bands of a constant width equal to the
scale spacing
6.7.3.2.2 The apparent width of the pointer at its tip shall not exceed
one-quarter of the scale spacing and in no case shall it be
greater than 0,5 mm
Resolution of the indicating device
6.7.3.2.3 The sub-divisions of the verification scale shall be small
enough to ensure that the resolution of the indicating device
does not exceed 0,25 % of the actual volume for accuracy
class 1 meters, and 0,5 % of the actual volume for accuracy
class 2 meters, for a 90 min test at the minimum flow rate,
Q .
NOTE 1 When the display of the first element is continuous,
an allowance should be made for a maximum error in each
reading of not more than half of the verification scale inter-
val.
NOTE 2 When the display of the first element is discontinu-
ous, an allowance should be made for a maximum error in
each reading of not more than one digit of the verification
scale
External examination for all water meters
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
NOTE For combination meters with two indicating devices, the above requirements apply to both indicating
devices.
Marks and inscriptions
6.6.1 A place shall be provided on the meter for affixing the verifi-
cation mark, which shall be visible without dismantling the
meter
6.6.2 The water meter shall be clearly and indelibly marked with
the information listed in the following, either grouped or
distributed on the casing, the indicating device dial, an iden-
tification plate or on the meter cover if is not detachable
6.6.2 a) Unit of measurement: cubic metre
6.6.2 b) The accuracy class, where it differs from accuracy class 2
6.6.2 c) The numerical value of Q and the ratio Q /Q (may be
3 3 1
preceded by R). If the meter measures reverse flow and Q
and the ratio Q /Q are different in the two directions, both
3 1
values of Q and Q /Q shall be inscribed; the direction of
3 3 1
flow to which each pair of values refers shall be clear. If the
meter has different values of Q /Q in horizontal and verti-
3 1
cal positions, both values of Q /Q shall be inscribed, and the
3 1
orientation to which each value refers shall be clear
6.6.2 d) The type approval sign according to national regulations
6.6.2 e) The name or trademark of the manufacturer
6.6.2 f) The year of manufacture (or the last two digits of the year of
manufacture or the month and year of manufacture)
6.6.2 g) The serial number (as near as possible to the indicating
device)
6.6.2 h) The direction of flow (shown on both sides of the body; or on
one side only, provided the direction of flow arrow is easily
visible under all circumstances)
6.6.2 i) The maximum admissible pressure (MAP) if it exceeds
1 MPa (10 bar) or 0,6 MPa (6 bar) for nominal diameter
≥500 mm.
(The unit bar may be used where national regulations per-
mit)
6.6.2 j) The letter V or H, if the meter can only be operated in the
vertical or horizontal position
6.6.2 k) The temperature class where it differs from T30
6.6.2 l) The pressure loss class where it differs from Δp 63
6.6.2 m) The installation sensitivity class where it differs from U0/
D0
Additional markings for water meters with electronic devices
6.6.2 n) For an external power supply: the voltage and frequency
6.6.2 o) For a replaceable battery: the latest date by which the bat-
tery shall be replaced
6.6.2 p) For a non-replaceable battery: the latest date by which the
meter shall be replaced
6.6.2 q) Environmental classification
16 © ISO 2014 – All rights reserved

External examination for all water meters
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
6.6.2 r) Electromagnetic environmental class
Protection devices
6.8.1 Water meters shall include protection devices which can be
sealed so as to prevent, both before and after correct instal-
lation of the water meter, dismantling or modification of the
meter, its adjustment device or its correction device, without
damaging these devices. In the case of combination meters,
this requirement applies to both meters
Protection devices — Electronic sealing devices
6.8.2.1 When access to parameters that influence the determination
of the results of measurements is not protected by mechani-
cal sealing devices, the protection shall fulfil the following
provisions.
a) Access shall only be allowed to authorized people,
e.g. by means of a code (password) or of a special device
(hard key, etc.). The code shall be capable of being changed.
b) It shall be possible for at least the last intervention
to be memorized. The record shall include the date and a
characteristic element identifying the authorized person
making the intervention [see a)]. If it is possible to memorize
more than one intervention and if a previous intervention
requires deletion to permit a new record, the oldest record
shall be deleted
6.8.2.2 For meters with parts which can be disconnected one from
another by the user and which are interchangeable, the fol-
lowing provisions shall be fulfilled:
a) it shall not be possible to access parameters that
participate in the determination of results of measure-
ments through disconnected points unless the provisions of
ISO 4064-1:2014|OIML R 49-1:2013, 6.8.2.1 are fulfilled;
b) interposing any device which may influence the
accuracy shall be prevented by means of electronic and data
processing securities or, if this is not possible, by mechanical
means
6.8.2.3 For meters with parts which may be disconnected one from
the other by the user and which are not interchangeable, the
provisions in ISO 4064-1:2014|OIML R 49-1:2013, 6.8.2.2
apply.
Moreover, these meters shall be provided with devices
which do not allow them to operate if the various parts are
not connected according to the approved type.
NOTE Disconnections which are not allowed to the user
may be prevented, e.g. by means of a device that prevents
any measurement after disconnecting and reconnecting
Examination and testing of checking facilities
General requirements for examining checking facilities
External examination for all water meters
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
5.1.3 Water meters with electronic devices shall be provided with
the checking facilities specified in ISO 4064-1:2014|OIML
R 49-1:2013, Annex B, except in the case of non-resettable
measurements between two constant partners
5.1.3 All meters equipped with checking facilities shall prevent or
detect reverse flow, as laid down in ISO 4064-1:2014|OIML
R 49-1:2013, 4.2.7.
4.4.2 Checklist for water meter performance tests
4.4.2.1 Performance tests for all water meters
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
Static pressure test
4.2.10 The meter shall be capable of withstanding the following test
pressures without leakage or damage:
— 1,6 times the maximum admissible pressure for
15 min;
— 2 times the maximum admissible pressure for 1 min
Intrinsic errors (of indication)
18 © ISO 2014 – All rights reserved

ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
7.2.3 The errors (of indication) of the water meter (in the measure-
ment of the actual volume), shall be determined at least at the
following flow rate ranges:
a) Q to 1,1Q ;
1 1
b) Q to 1,1 Q ;
2 2
c) 0,33(Q + Q ) to 0,37(Q + Q );
2 3 2 3
d) 0,67(Q + Q ) to 0,74(Q + Q );
2 3 2 3
e) 0,9Q to Q ;
3 3
f) 0,95Q to Q ;
4 4
and for combination meters:;
g) 0,85Q to 0,95Q ;
x1 x1
h) 1,05Q to 1,15Q
x2 x2.
The water meter should be tested without its temporary sup-
plementary devices attached (if any).
During a test all other influence factors shall be held at refer-
ence conditions.
Other flow rates may be tested depending on the shape of the
error curve.
1) The relative errors (of indication) observed for each
of the flow rates shall not exceed the maximum permissible
errors (MPEs) given in ISO 4064-1:2014|OIML R 49-1:2013, 4.2.2
or 4.2.3. If the error observed on one or more meters is greater
than the MPE at one flow rate only, then if only two results have
been taken at that flow rate, the test at that flow rate shall be
repeated. The test shall be declared satisfactory if two out of the
three results at that flow rate lie within the MPE and the arith-
metic mean of the results for the three tests at that flow rate lies
within the MPE.
2) If all the relative errors (of indication) of the water
meter have the same sign, at least one of the errors shall not
exceed one-half of the MPE. In all cases, this requirement shall
be applied equitably with respect to the water supplier and the
consumer (see also ISO 4064-1:2014|OIML R 49-1:2013, 4.3.3 3)
and 8)
7.2.4 The meter shall be repeatable: the standard deviation of three
measurements at the same flow rate shall not exceed one-third
of the MPEs given in ISO 4064-1:2014|OIML R 49-1:2013, 4.2.2
or 4.2.3. Tests shall be carried out at nominal flow rates of Q ,
Q , and Q
2 3
Water temperature test
4.2.8 The requirements relating to the MPEs shall be met for all water
temperature variations within the rated operating conditions of
the meter
Water pressure test
4.2.8 The requirements relating to the MPEs shall be met for all water
pressure variations within the rated operating conditions of the
meter
Reverse flow test
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
4.2.7 A water meter designed to measure reverse flow shall either:
a) subtract the reverse flow volume from the indicated
volume; or
b) record the reverse flow volume separately.
The MPEs of ISO 4064-1:2014|OIML R 49-1:2013, 4.2.2 or 4.2.3
shall be met for both forward and reverse flow
4.2.7 A water meter not designed to measure reverse flow shall
either:
a) prevent it; or
b) be capable of withstanding an accidental reverse flow
at a flow rate up to Q without any deterioration or change in its
metrological properties for forward flow
Meter characteristics at zero flow rate
4.2.9 The water meter totalization shall not change when the flow
rate is zero
Pressure loss test
6.5 The pressure loss of the water meter, including its filter where
the latter forms an integral part of the water meter, shall not be
greater than 0,063 MPa (0,63 bar) between Q and Q
1 3
Flow disturbance test
6.3.4 If the accuracy of water meters is affected by disturbances
in the upstream or downstream pipeline, the meter shall be
provided with sufficient straight pipe lengths with or without
a flow straightener (as specified by the manufacturer) so that
the indications of the installed water meter do not exceed MPEs
according to the accuracy class of the meter.
Forward flow tests
Reverse flow tests (where applicable)
Overload temperature test
7.2.5 Water meters with MAT ≥50 °C shall be capable of withstanding
a water temperature of MAT + 10 °C for 1 h
Durability tests
7.2.6 The water meter shall undergo a durability test according to
the permanent flow rate Q and the overload flow rate Q of the
3 4
meter, simulating service conditions
7.2.6 Meters with Q ≤ 16 m /h:
a) 100 000 flow cycles between 0 and Q ;
b) 100 h at Q
7.2.6 Meters with Q > 16 m /h:
a) 800 h at Q ;
b) 200 h at Q ;
and for combination meters:
c) 50 000 flow cycles between Q ≥ 2Q and 0
x2
20 © ISO 2014 – All rights reserved

ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
Accuracy class 1 meters
7.2.6.2
The variation in the error curve shall not exceed 2 % for flow
rates in the lower zone (Q ≤ Q < Q ) and 1 % for flow rates in
1 2
the upper zone (Q ≤ Q ≤ Q ).
2 4
For the purpose of these requirements, the arithmetic mean
value of the errors (of indication) E for each flow rate shall
apply.
For flow rates in the lower flow rate zone (Q ≤ Q < Q ), the error
1 2
(of indication) curve shall not exceed a maximum error limit
of ±4 % for all temperature classes. For flow rates in the upper
flow rate zone (Q ≤ Q ≤ Q ), the error (of indication) curve shall
2 4
not exceed a maximum error limit of ±1,5 % for meters of tem-
perature class T30 and ±2,5 % for all other temperature classes
Accuracy class 2 meters
7.2.6.3
The variation in the error curve shall not exceed 3 % for flow
rates in the lower zone (Q ≤ Q < Q ) and 1,5 % for flow rates in
1 2
the upper zone (Q ≤ Q < Q ).
2 4
For the purpose of these requirements, the arithmetic mean
value of the errors (of indication) E for each flow rate shall
apply.
For flow rates in the lower flow rate zone (Q ≤ Q < Q ), the error
1 2
(of indication) curve shall not exceed a maximum error limit
of ±6 % for all temperature classes. For flow rates in the upper
flow rate zone (Q ≤ Q < Q ) the error (of indication) curve shall
2 4
not exceed a maximum error limit of ±2,5 % for meters of tem-
perature class T30 and ±3,5 % for all other temperature classes
7.2.7 It shall be demonstrated that cartridge meters and exchange-
able metrological modules for water meters with exchange-
able metrological modules are independent of the connection
interfaces they are made for, as far as their metrological per-
formance is concerned. The cartridge meters and exchangeable
metrological modules shall be tested in accordance with the test
specified in ISO 4064-2:2014|OIML R 49-2:2013, 7.4.6
7.2.8 All water meters where the mechanical components may be
influenced by a static magnetic field and all meters with elec-
tronic components shall be tested by applying a specified field.
The test shall be carried out at Q and show that the indications
of the installed water meter do not exceed MPEs of the upper
zone according to the accuracy class of the meter:
Forward flow tests
Reverse flow tests (where applicable)
Application of the field in different planes
4.4.2.2 Performance tests for electronic water meters and electronic devices fitted to mechani-
cal meters (first version)
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
Dry heat
A.5 To verify compliance with the provisions in 4.2 under condi-
tions of high temperature
(see ISO 4064-2:2014|OIML R 49-2:2013, 8.2)
Cold
ISO 4064-1:2014| Requirement + − Remarks
OIML R 49-1:2013,
subclause
A.5 To verify compliance with the provisions in 4.2 under condi-
...


NORME ISO
INTERNATIONALE 4064-3
Quatrième édition
2014-06-01
Compteurs d’eau potable froide et
d’eau chaude —
Partie 3:
Format du rapport d’essais
Water meters for cold potable water and hot water —
Part 3: Test report format
Numéro de référence
©
ISO 2014
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2014
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée
sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur
l’internet ou sur un Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à
l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse
ii © ISO 2014 – Tous droits réservés

Sommaire Page
Avant-propos .iv
Introduction .v
1 Domaine d’application . 1
2 Références normatives . 1
3 Termes, définitions, symboles et termes abrégés . 1
4 Rapport d’examen de type. 2
4.1 Généralités . 2
4.2 Informations concernant le type . 2
4.3 Informations générales concernant le matériel d’essai .11
4.4 Liste de contrôle des examens et des essais de performance des compteurs d’eau .12
4.5 Essais d’évaluation de type (applicables à tous les compteurs d’eau) .24
4.6 Essais d’évaluation de type (pour les compteurs d’eau électroniques et les compteurs
d’eau mécaniques équipés de composants électroniques) .53
5 Rapport de vérification primitive .92
5.1 Généralités .92
5.2 Informations concernant l’EST vérifié .92
5.3 Rapport des essais de vérification primitive (ISO 4064-2:2014|OIML R 49-2:2013,
Article 10).93
Annexe A (normative) Liste des documents relatifs au type (ISO 4064‑1:2014|OIML R 49‑1:2013,
7.2.10) .96
Annexe B (normative) Liste des matériels d’essais utilisés pour les examens et les essais .97
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui concerne
la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2, www.iso.
org/directives.
L’attention est appelée sur le fait que certains des éléments du présent document peuvent faire l’objet de
droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable
de ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant les
références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de l’élaboration
du document sont indiqués dans l’Introduction et/ou sur la liste ISO des déclarations de brevets reçues,
www.iso.org/patents.
Les éventuelles appellations commerciales utilisées dans le présent document sont données pour
information à l’intention des utilisateurs et ne constituent pas une approbation ou une recommandation.
Le comité chargé de l’élaboration du présent document est l’ISO/TC 30, Mesure de débit des fluides dans
les conduites fermées, sous-comité SC 7, Méthodes volumétriques, y compris les compteurs d’eau et du sous-
comité technique de l’OIML TC 8/SC 5 Compteurs d’eau.
Cette quatrième édition de l’ISO 4064-3 annule et remplace la troisième édition (ISO 4064-3:2005), qui
a fait l’objet d’une révision technique.
L’ISO 4064 comprend les parties suivantes, présentées sous le titre général Compteurs d’eau potable
froide et chaude:
— Partie 1: Exigences métrologiques et techniques
— Partie 2: Méthodes d’essai
— Partie 3: Format du rapport d’essais
— Partie 4: Exigences non métrologiques non couvertes par l’ISO 4064‑1
— Partie 5: Exigences d’installation
Cette édition de l’ISO 4064-3 est identique à l’édition correspondante de l’OIML R 49-3, qui a été publiée
en même temps. L’OIML R 49-3 a été approuvée pour la publication finale par le Comité International de
Métrologie Légale à sa 48ème réunion à Ho Chi Minh City, Vietnam en Octobre 2013. Elle sera soumise à
la Conférence Internationale sur la Métrologie Légale en 2016 pour une sanction formelle.
iv © ISO 2014 – Tous droits réservés

Introduction
La mise en œuvre de ce format de rapport d’essais est informative pour ce qui concerne la mise en œuvre
de l’ISO 4064-1|OIML R 49-1 et ISO 4064-2|OIML R 49-2 dans les réglementations nationales; cependant,
cette mise en œuvre est exigée dans le cadre du système de certification OIML des instruments de
mesure [ISO 4064-2|OIML R 49-2:2013, 11.1].
L’Article 4 présente le format requis d’un rapport d’examen de type d’un compteur d’eau complet ou de
combinaison.
Le rapport d’examen de type d’un calculateur séparable (incluant un dispositif indicateur) ou d’un
transducteur de mesure (incluant un capteur de débit ou de volume) nécessite un format similaire.
Cependant, il peut être nécessaire d’appliquer certaines modifications aux tableaux en raison du grand
nombre de possibilités de variations dans la conception de ces unités séparables.
Certains exemples de tableaux de présentation des résultats d’essai d’unités séparables sont montrés
dans l’Article 5 en vue de vérifications primitives. Ces tableaux peuvent également être adaptés à des
rapports d’examen de type.
NORME INTERNATIONALE ISO 4064-3:2014(F)
Compteurs d’eau potable froide et d’eau chaude —
Partie 3:
Format du rapport d’essais
1 Domaine d’application
La présente partie de l’ISO 4064|OIML R 49 spécifie le format du rapport d’essai à utiliser dans le contexte
de la mise en œuvre de l’ISO 4064-1|OIML R 49-1 et de l’ISO 4064-2|OIML R 49-2 pour les compteurs
d’eau potable froide et d’eau chaude.
2 Références normatives
Les documents suivants, en tout ou partie, sont référencés de manière normative dans le présent
document et sont indispensables pour son application. Pour les références datées, seule l’édition citée
s’applique. Pour les références non datées, la dernière édition du document de référence s’applique (y
compris les éventuels amendements).
ISO 4064-1:2014|OIML R 49-1:2013, Compteurs d’eau potable froide et d’eau chaude — Partie 1: Exigences
métrologiques et techniques
ISO 4064-2:2014|OIML R 49-2:2013, Compteurs d’eau potable froide et d’eau chaude — Partie 2: Méthodes
d’essai
3 Termes, définitions, symboles et termes abrégés
Pour les besoins du présent document, les termes et définitions donnés dans l’ISO 4064-1|OIML R 49-1
s’appliquent.
Les symboles utilisés dans les tableaux sont les suivants:
+ accepté
− refusé
n/a sans objet
EUT équipement soumis à essai
H horizontal
MAP pression maximale admissible
MAT température maximale admissible
MPE erreur maximale tolérée
V vertical
4 Rapport d’examen de type
4.1 Généralités
Pour chaque examen et chaque essai, la liste des points à vérifier (checklist) doit être remplie
conformément à cet exemple:
+ −
× Accepté
× Refusé
n/a n/a Sans objet
4.2 Informations concernant le type
4.2.1 Généralités
Demande No.: _________________________________
Demandeur: _________________________________
Représentant autorisé: _________________________________
Adresse: _________________________________
_________________________________
_________________________________
_________________________________
Laboratoire d’essai: _________________________________
Représentant autorisé: _________________________________
Adresse: _________________________________
_________________________________
_________________________________
_________________________________
4.2.2 Modèle soumis
Nouveau modèle: _________________________________
Variante de modèle(s) approuvé(s):
Numéro d’approbation: _________________________________
Variante du modèle approuvé: _________________________________
Voir Tableau 1.
2 © ISO 2014 – Tous droits réservés

Tableau 1 — Modèle présenté
a a
Présenté aux essais d’approbation Oui Non Remarques
Compteur d’eau mécanique (complet)
Compteur d’eau mécanique (de combinaison)
Compteur d’eau électronique (complet)
Compteur d’eau électronique (de combinaison)
Famille de compteurs d’eau
Calculateur séparable (incluant le dispositif indicateur)
Transducteur de mesure séparable (incluant le capteur de
débit ou de volume)
Dispositif(s) électronique(s) supplémentaire(s) d’essai
(fixé(s) de manière permanente au compteur)
Dispositif(s) électronique(s) supplémentaire(s) de trans-
mission de données (fixé(s) de manière permanente au
compteur)
Dispositif(s) électronique(s) supplémentaire(s) d’essai (fixé
de manière temporaire au compteur)
Dispositif(s) électronique(s) supplémentaire(s) de transmis-
sion de données (fixé(s) de manière temporaire au comp-
teur)
Dispositifs auxiliaires
a
Cocher la case appropriée.
4.2.3 Compteur d’eau mécanique (complet ou de combinaison)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Détails du type:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
Compteurs combinés:
Q _________ m /h
x1
Q _________ m /h
x2
Principe de mesurage: __________________________
Classe d’exactitude: _____________
Classe de température: _____________
Classe environnementale: _____________
Environnement électromagnétique: _____________
Température maximale admissible: _________ °C
Pression maximale admissible: ___________ MPa (_____ bar)
Limitation de l’orientation: _________________________
Exigences d’essai de l’EST (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Catégorie: _________________________
Cas: _________________________
Détails de l’installation:
Type de raccordement (bride, vis _________
filetée, collecteur concentrique):
Longueur droite minimale de la __________ mm
conduite d’entrée:
Longueur droite minimale de la __________ mm
conduite de sortie:
Conditionneur de flux (détails si _________________________
nécessaire):
Montage: _________________________
Orientation: _________________________
Autres informations utiles: _________________________
NOTE En cas de soumission d’une famille de compteurs, les informations énumérées dans le présent
paragraphe doivent être fournies pour chaque calibre de compteur d’eau.
4.2.4 Compteur d’eau électronique (complet ou de combinaison)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Détails du type:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
Compteurs combinés:
Q _________ m /h
x1
Q _________ m /h
x2
4 © ISO 2014 – Tous droits réservés

Principe de mesurage: __________________________
Classe d’exactitude: _____________
Classe de température: _____________
Classe environnementale: _____________
Environnement électromagnétique: _____________
Température maximale admissible: _________ °C
Pression maximale admissible: ___________ MPa (_____ bar)
Limitation de l’orientation: _________________________
Exigences d’essai de l’EST (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Catégorie: _________________________
Cas: _________________________
Détails de l’installation (mécanique):
Type de raccordement (bride, vis filetée, collecteur _________
concentrique):
Longueur droite minimale de la conduite d’entrée: __________ mm
Longueur droite minimale de la conduite de sortie: __________ mm
Conditionneur de flux (détails si nécessaire): _________________________
Montage: _________________________
Orientation: _________________________
Autres informations utiles: _________________________
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
Alimentation électrique:
Type d’alimentation (batterie, réseau CC, réseau CA): _________________________
U : ______________________ V
max
U : ______________________ V
min
Fréquence: _______________________ Hz
NOTE En cas de soumission d’une famille de compteurs, les informations énumérées dans le présent
paragraphe doivent être fournies pour chaque calibre de compteur d’eau.
4.2.5 Calculateur séparable (incluant le dispositif indicateur)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Détails du type:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
Compteurs combinés:
Q _________ m /h
x1
Q _________ m /h
x2
Principe de mesurage: __________________________
Classe d’exactitude: _____________
Classe de température: _____________
Classe environnementale: _____________
Environnement électromagnétique: _____________
Température maximale admissible: _________°C
Pression maximale admissible: ___________MPa (_____ bar)
Limitation de l’orientation: _________________________
Exigences d’essai de l’EST (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Catégorie: _________________________
Cas: _________________________
Erreur relative maximale spécifiée par le fabricant:
Zone de débit inférieure, Q ≤ Q < Q : ________________________%
1 2
Zone de débit supérieure, Q ≤ Q ≤ Q : ________________________%
2 4
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
6 © ISO 2014 – Tous droits réservés

Alimentation électrique:
Type d’alimentation (batterie, réseau CC, _________________________
réseau CA):
U : ________________________ V
max
U : ________________________ V
min
Fréquence: _______________________ Hz
Numéro(s) d’approbation du transducteur de
mesure compatible (incluant le capteur de
débit ou de volume): ____________________
4.2.6 Transducteur de mesure séparable (incluant le capteur de débit ou de volume)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Détails du type:
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q _________ m /h
Q /Q _________
3 1
Compteurs combinés:
Q _________ m /h
x1
Q _________ m /h
x2
Principe de mesurage: __________________________
Classe d’exactitude: _____________
Classe de température: _____________
Classe environnementale: _____________
Environnement électromagnétique: _____________
Température maximale admissible: _________°C
Pression maximale admissible: ___________MPa (_____ bar)
Limitation de l’orientation: _________________________
Exigences d’essai de l’EST (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Catégorie: _________________________
Cas: _________________________
Erreur relative maximale spécifiée par le fabricant:
Zone de débit inférieure, Q ≤ Q < Q : ________________________%
1 2
Zone de débit supérieure, Q ≤ Q ≤ Q : ________________________%
2 4
Détails de l’installation (mécanique):
Type de raccordement (bride, vis filetée, collecteur _________
concentrique):
Longueur droite minimale de la conduite d’entrée: __________mm
Longueur droite minimale de la conduite de sortie: __________mm
Conditionneur de flux (détails si nécessaire): _________________________
Montage: _________________________
Orientation: _________________________
Autres informations utiles: _________________________
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
Alimentation électrique:
Type d’alimentation (batterie, réseau CC, _________________________
réseau CA):
U : ________________________ V
max
U : ________________________ V
min
Fréquence: _______________________ Hz
Numéro(s) d’approbation du (des)
calculateur(s) compatible(s) (incluant le
dispositif indicateur): _________________________
4.2.7 Dispositif(s) électronique(s) supplémentaire(s) servant aux essais (fixé(s) de manière
permanente au compteur)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Alimentation électrique:
Type d’alimentation (batterie, _________________________
réseau CC, réseau CA):
U : ________________________ V
max
8 © ISO 2014 – Tous droits réservés

U : ________________________ V
min
Fréquence: _______________________ Hz
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
4.2.8 Dispositif(s) électronique(s) supplémentaire(s) servant à la transmission de données
(fixé(s) de manière permanente au compteur)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Alimentation électrique:
Type d’alimentation (batterie, _________________________
réseau CC, réseau CA):
U : ________________________ V
max
U : ________________________ V
min
Fréquence: _______________________ Hz
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
4.2.9 Dispositif(s) électronique(s) supplémentaire(s) servant aux essais (fixé(s) de manière
temporaire au compteur)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Alimentation électrique:
Type d’alimentation (batterie, _________________________
réseau CC, réseau CA):
U : ________________________ V
max
U : ________________________ V
min
Fréquence: _______________________ Hz
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
4.2.10 Dispositif(s) électronique(s) supplémentaire(s) servant à la transmission de données
(fixé(s) de manière temporaire au compteur)
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Alimentation électrique:
Type d’alimentation (batterie, _________________________
réseau CC, réseau CA):
U : ________________________ V
max
U : ________________________ V
min
Fréquence: _______________________ Hz
Exigences d’essai de l’EST (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Catégorie: _________________________
Cas: _________________________
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
4.2.11 Dispositifs auxiliaires
Fabricant: __________________________________________________
Numéro du modèle: __________________________________________________
Alimentation électrique:
Type d’alimentation (batterie, _________________________
réseau CC, réseau CA):
U : ________________________ V
max
U : ________________________ V
min
Fréquence: _______________________ Hz
Numéro(s) d’approbation
de la compatibilité du (des)
calculateur(s) (incluant le disposi-
tif indicateur): _________________________
10 © ISO 2014 – Tous droits réservés

Exigences d’essai de l’EST (ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8):
Catégorie: _________________________
Cas: _________________________
Détails de l’installation (électrique):
Instructions de câblage: _________________________
Système de montage: _________________________
Limitations de l’orientation: _________________________
Numéro(s) d’approbation des comp-
teurs d’eau, du (des) calculateur(s)
(incluant le dispositif indicateur) et
du (des) transducteur(s) de mesure
(incluant le capteur de débit ou de
volume) compatibles: _________________________
4.2.12 Documents concernant le type
Une liste des documents telle que celle figurant en Annexe A doit accompagner la demande d’approbation
de type.
4.3 Informations générales concernant le matériel d’essai
Les détails de tous les éléments des matériels de mesure et des instruments d’essai utilisés pour les
examens de type et les vérifications primitives doivent être énumérés dans l’Annexe B, y compris:
Fabricant
Numéro du modèle
Numéro de série,
Date du dernier étalonnage,
Date du prochain étalonnage prévu, par exemple des instruments destinés à mesurer:
— les dimensions linéaires
— des capteurs de pression
— des transmetteurs de pression
— des manomètres
— des transducteurs de température
— des compteurs de référence
— des réservoirs
— des appareils de pesée
— des générateurs de signaux (d’impulsions, de courant ou de tension)
4.4 Liste de contrôle des examens et des essais de performance des compteurs d’eau
4.4.1 Liste de contrôle des examens pratiqués sur les compteurs d’eau
Examen externe applicable à tous les compteurs d’eau
Paragraphe de Exigence + - Remarques
l’ISO 4064-1|
OIML R 49-1:2013
Fonction du dispositif indicateur
6.7.1.1 Le dispositif indicateur doit permettre une lecture
facile, et donner une indication visuelle fiable et non
ambiguë du volume indiqué.
6.7.1.1 Le dispositif indicateur doit inclure des éléments
visuels pour les essais et l’étalonnage.
6.7.1.1 Le dispositif indicateur peut inclure des éléments
supplémentaires pour les essais et l’étalonnage par
d’autres méthodes, par exemple pour des essais et un
étalonnage automatiques.
Unité de mesure et emplacement de l’unité de mesure
6.7.1.2 Le volume d’eau indiqué doit être exprimé en mètres
cubes.
doit apparaître sur le cadran ou figurer
6.7.1.2 Le symbole m
immédiatement à côté de l’afficheur numérique.
Étendue d’indication
6.7.1.3 Pour q ≤ 6,3, étendue d’indication minimale =
9 999 m
6.7.1.3 Pour 6,3 < q ≤ 63, étendue d’indication mini-
male = 99 999 m
6.7.1.3 Pour 63 < q ≤ 630, étendue d’indication mini-
male = 999 999 m
6.7.1.3 Pour 630 < q ≤ 6 300, étendue d’indication mini-
male = 9 999 999 m
Code couleur pour le dispositif indicateur
6.7.1.4 Il convient d’utiliser la couleur noire pour indiquer le
mètre cube et ses multiples.
6.7.1.4 Il convient d’utiliser la couleur rouge pour indiquer des
sous-multiples du mètre cube.
6.7.1.4 Ces couleurs doivent être appliquées soit aux aiguilles,
aux repères, aux chiffres, aux volants, aux disques, aux
cadrans ou aux ouvertures.
6.7.1.4 Il est possible d’utiliser d’autres procédés d’indication
du mètre cube à condition qu’il n’y ait pas d’ambiguïté
dans la distinction entre l’indication primaire et les
affichages alternatifs, par exemple les sous-multiples
pour la vérification et les essais.
Types de dispositif indicateur: Type 1 — Dispositif analogique
6.7.2.1 Le volume indiqué doit être donné par le mouvement
continu:
a) soit d’une ou plusieurs aiguilles mobiles par rapport
à des échelles graduées,
b) soit d’une ou plusieurs échelles circulaires ou tam-
bours chacun défilant devant un index.
12 © ISO 2014 – Tous droits réservés

Examen externe applicable à tous les compteurs d’eau
Paragraphe de Exigence + - Remarques
l’ISO 4064-1|
OIML R 49-1:2013
6.7.2.1 La valeur exprimée en mètres cubes pour chaque
n
division doit être de la forme 10 , où n est un nombre
entier positif ou négatif ou zéro, constituant de ce fait
un système de décades consécutives.
6.7.2.1 L’échelle doit être soit graduée en valeurs exprimées en
mètres cubes, soit accompagnée d’un facteur multipli-
cateur
(×0,001; ×0,01; ×0,1; ×1; ×10; ×100; ×1 000; etc.).
6.7.2.1 Le mouvement de rotation des aiguilles ou échelles
circulaires doit aller dans le sens des aiguilles d’une
montre.
6.7.2.1 Le mouvement linéaire des aiguilles ou échelles doit
aller de gauche à droite.
6.7.2.1 Le mouvement des indicateurs à cylindre gradué doit
être dirigé vers le haut.
Types de dispositif indicateur: Type 2 — Dispositif numérique
6.7.2.2 Le volume indiqué est donné par une ligne de chiffres
apparaissant dans une ou plusieurs ouvertures.
6.7.2.2 L’incrémentation d’une unité doit se faire en même
temps que le chiffre de la décade adjacente immédiate-
ment inférieure passe de 9 à 0.
6.7.2.2 La hauteur réelle ou apparente des chiffres doit être
d’au moins 4 mm.
6.7.2.2 En ce qui concerne les dispositifs non électroniques,
le mouvement des indicateurs à cylindre gradué (tam-
bours) doit être dirigé vers le haut.
6.7.2.2 En ce qui concerne les dispositifs non électroniques, la
valeur de la décade la plus basse peut avoir un mouve-
ment continu si l’ouverture est suffisamment grande
pour permettre la lecture sans ambiguïté d’un chiffre.
6.7.2.2 En ce qui concerne les dispositifs électroniques avec
affichage temporaire, le volume doit pouvoir être affi-
ché à tout moment pendant au moins 10 s.
6.7.2.2 En ce qui concerne les dispositifs électroniques, le
compteur doit permettre un contrôle visuel de l’en-
semble du système d’affichage. Ce contrôle doit suivre
les étapes suivantes:
— pour le type à sept segments, affichage de tous les
éléments (par exemple, un essai des «huit»); et
— pour le type à sept segments, suppression de tous
les éléments (un essai des «blancs»).
— pour l’affichage graphique, un essai équivalent
permettant de démontrer que les défauts d’affichage ne
peuvent pas fausser l’interprétation des chiffres.
Chaque étape de la séquence doit durer au moins 1 s.
Types de dispositif indicateur: Type 3 — Combinaison de dispositifs analogiques et numériques
6.7.2.3 Le volume indiqué est donné par une combinaison des
dispositifs de type 1 et de type 2, les exigences respec-
tives de chacun devant s’appliquer.
Dispositifs de vérification – Exigences générales
6.7.3.1 Chaque dispositif indicateur doit fournir des éléments
visuels, non ambigus pour les essais de vérification et
l’étalonnage.
Examen externe applicable à tous les compteurs d’eau
Paragraphe de Exigence + - Remarques
l’ISO 4064-1|
OIML R 49-1:2013
6.7.3.1 La vérification visuelle de l’indicateur peut avoir un
mouvement soit continu, soit discontinu.
6.7.3.1 En plus du dispositif de vérification visuelle, un dis-
positif indicateur peut inclure des dispositions pour
contrôle rapide par l’inclusion d’éléments complémen-
taires (par exemple, roues dentées ou disques), four-
nissant des signaux par l’intermédiaire de détecteurs
externes.
Dispositifs de vérification – Éléments de vérification visuelle
6.7.3.2.1 La valeur de l’échelon de vérification, exprimée en
n n
mètres cubes, doit être de la forme: 1 × 10 ou 2 × 10
n
ou 5 × 10 , où n est zéro ou un nombre entier positif ou
négatif.
6.7.3.2.1 Le volume indiqué est donné par une ligne de chiffres
apparaissant dans une ou plusieurs ouvertures.
6.7.3.2.1 Pour les dispositifs indicateurs analogiques ou numé-
riques à mouvement continu du premier élément,
l’échelon de vérification peut être formé par la division
en 2, 5 ou 10 parties égales de l’intervalle entre deux
chiffres consécutifs du premier élément. Le numéro-
tage ne doit pas être appliqué à ces divisions.
6.7.3.2.1 Pour les dispositifs indicateurs numériques à mouve-
ment discontinu du premier élément, l’échelon de véri-
fication est l’intervalle entre deux chiffres consécutifs
ou deux variations incrémentielles du premier élément.
6.7.3.2.2 Sur des dispositifs indicateurs à mouvement continu
du premier élément, la longueur apparente d’une
division d’échelle ne doit pas être inférieure à 1 mm ou
supérieure à 5 mm.
6.7.3.2.2 L’échelle doit consister
a) soit en lignes d’épaisseur égale ne dépassant pas un
quart de la longueur d’une division d’échelle et diffé-
rant seulement par leur longueur;
b) soit en bandes contrastées de largeur constante
égale à la longueur d’une division d’échelle.
6.7.3.2.2 La largeur apparente de la pointe de l’aiguille ne doit
pas dépasser un quart de la longueur d’une division
d’échelle et, en aucun cas, ne doit être supérieure à
0,5 mm.
Résolution du dispositif indicateur
6.7.3.2.3 Les sous-divisions de l’échelle de vérification doivent
être suffisamment faibles pour garantir que la réso-
lution du dispositif indicateur ne dépasse pas 0,25 %
du volume réel pour les compteurs de classe d’exacti-
tude 1, et 0,5 % du volume réel pour les compteurs de
classe d’exactitude 2, pendant un essai d’une durée de
90 min au débit minimal Q .
NOTE 1 Lorsque l’affichage du premier élément est
continu, il convient d’opérer une déduction pour une
erreur maximale de chaque mesure ne dépassant pas
la moitié de l’échelon de vérification.
NOTE 2 Lorsque l’affichage du premier élément est
discontinu, il convient d’opérer une déduction pour une
erreur maximale de chaque mesure ne dépassant pas
un chiffre de l’échelle de vérification.
14 © ISO 2014 – Tous droits réservés

Examen externe applicable à tous les compteurs d’eau
Paragraphe de Exigence + - Remarques
l’ISO 4064-1|
OIML R 49-1:2013
NOTE Pour les compteurs combinés équipés de deux dispositifs indicateurs, les exigences ci-dessus s’appliquent aux
deux dispositifs indicateurs.
Marquages et inscriptions
6.6.1 Un emplacement doit être prévu sur les compteurs
d’eau pour apposer la marque de vérification, qui doit
être visible sans démonter le compteur d’eau.
6.6.2 Le compteur d’eau doit porter, de manière lisible et
indélébile, les informations énumérées ci-dessous, ces
informations pouvant être regroupées ou réparties sur
le boîtier, le cadran du dispositif indicateur, une plaque
signalétique ou sur le coffrage du compteur si celui-ci
n’est pas amovible:
6.6.2 (a) Unité de mesure: mètre cube
6.6.2 (b) La classe d’exactitude, si elle diffère de la classe d’exac-
titude 2.
6.6.2 (c) La valeur numérique de q et le rapport Q /Q (peut
3 3 1
être précédé de la lettre R). Si le compteur mesure le
flux inversé et si Q et le rapport Q /Q sont différents
3 3 1
dans les deux sens, les deux valeurs de Q et Q /Q
3 3 1
doivent être inscrites: le sens d’écoulement auquel
chaque paire de valeurs se réfère doit être clair. Si le
compteur a différentes valeurs de Q /Q en positions
3 1
horizontale et verticale, les deux valeurs de Q /Q
3 1
doivent être inscrites, et l’orientation à laquelle chaque
valeur se réfère doit être claire.
6.6.2 (d) La marque d’approbation de type conformément aux
réglementations nationales.
6.6.2 (e) Le nom ou la marque commerciale du fabricant.
6.6.2 (f) L’année de fabrication (ou les deux derniers chiffres de
l’année de fabrication).
6.6.2 (g) Le numéro de série (aussi près que possible du disposi-
tif indicateur).
6.6.2 (h) Le sens de l’écoulement (indiqué des deux côtés du
boîtier, ou sur un côté seulement, à condition que le
sens indiqué par l’aiguille de sens d’écoulement soit
facilement visible en toutes circonstances).
6.6.2 (i) La pression maximale admissible (PMA) si elle dépasse
1 MPa (10 bar) ou 0,6 MPa (6 bar) pour un diamètre
nominal ≥ 500 mm.
(L’unité bar peut être utilisée si les réglementations
nationales l’autorisent)
6.6.2 (j) La lettre V ou la lettre H, si le compteur ne peut
fonctionner que dans la position verticale ou dans la
position horizontale.
6.6.2 (k) La classe de température si elle diffère de T30.
6.6.2 (l) La classe de perte de pression si elle diffère de Δp 63
6.6.2 (m) La classe de sensibilité de l’installation, si elle diffère
de U0/D0.
Marques supplémentaires pour les comptes d’eau équipés de dispositifs électroniques
6.6.2 (n) pour une alimentation électrique externe: la tension et
la fréquence.
6.6.2 (o) Pour une batterie remplaçable: la date limite de rem-
placement de la batterie.
Examen externe applicable à tous les compteurs d’eau
Paragraphe de Exigence + - Remarques
l’ISO 4064-1|
OIML R 49-1:2013
6.6.2 (p) Pour une batterie non remplaçable: la date limite de
remplacement du compteur.
6.6.2 (q) Classification environnementale
6.6.2 (r) Classe d’environnement électromagnétique
Dispositifs de protection
6.8.1 Les compteurs d’eau doivent inclure des dispositifs de
protection pouvant être scellés de manière à empêcher,
à la fois avant et après l’installation correcte du comp-
teur d’eau, que le compteur ou son dispositif de réglage
ou de correction puissent être démontés ou modifiés
sans provoquer la détérioration de ces dispositifs. Pour
les compteurs combinés, cette exigence s’applique aux
deux compteurs.
Dispositifs de protection – Dispositifs de scellement électroniques
6.8.2.1 Si l’accès aux paramètres influençant la détermination
des résultats de mesure n’est pas protégé par des dis-
positifs de scellement mécaniques, la protection doit
remplir les dispositions suivantes:
a) L’accès doit être permis uniquement aux personnes
autorisées, par exemple au moyen d’un code (mot de
passe) ou d’un dispositif spécial (clé matérielle, etc.).
Le code doit pouvoir être changé.
b) Il doit être possible de conserver au moins la der-
nière intervention en mémoire. L’enregistrement doit
inclure la date et un élément caractéristique identifiant
la personne autorisée effectuant l’intervention [voir a)
ci-dessus]. La traçabilité de la dernière intervention
doit être assurée pendant au moins deux ans, si celle-
ci n’est pas effacée à l’occasion d’une intervention
ultérieure. S’il est possible de mémoriser plusieurs
interventions et si l’effacement d’une intervention
antérieure doit être effectué pour permettre un nouvel
enregistrement, l’enregistrement le plus ancien doit
être supprimé.
6.8.2.2 Pour les compteurs ayant des parties pouvant être
déconnectées l’une de l’autre par l’utilisateur et qui
sont interchangeables, les dispositions suivantes
doivent être satisfaites:
a) il ne doit pas être possible d’accéder aux para-
mètres participant à la détermination des résultats de
mesures par des points déconnectés sauf si les dispo-
sitions de l’ISO 4064-1:2014|OIML R 49-1:2013, 6.8.2.1,
sont satisfaites,
b) l’interposition de tout dispositif pouvant influencer
la précision doit être empêchée au moyen de sécuri-
tés électroniques et de sécurités de traitement des
données ou, si cela n’est pas possible, par des procédés
mécaniques.
16 © ISO 2014 – Tous droits réservés

Examen externe applicable à tous les compteurs d’eau
Paragraphe de Exigence + - Remarques
l’ISO 4064-1|
OIML R 49-1:2013
6.8.2.3 Pour les compteurs ayant des parties pouvant être
déconnectées l’une de l’autre par l’utilisateur et
qui ne sont pas interchangeables, les dispositions
de l’ISO 4064-1:2014|OIML R 49-1:2013, 6.8.2.2,
s’appliquent.
En outre, ces compteurs doivent être munis de disposi-
tifs les empêchant de fonctionner si les diverses parties
ne sont pas reliées selon la configuration donnée par le
fabricant.
NOTE Les déconnexions interdites à l’utilisateur
peuvent être empêchées, par exemple au moyen d’un
dispositif bloquant tout mesurage après déconnexion
et reconnexion.
Examen et essai des systèmes de contrôle
Exigences générales en matière d’examen des systèmes de contrôle
5.1.3 Les compteurs d’eau comportant des dispositifs élec-
troniques doivent être munis des systèmes de contrôle
spécifiés dans l’ISO 4064-1:2014|OIML R 49-1:2013,
Annexe B, excepté en cas de mesurages non repro-
grammables entre deux partenaires constants.
5.1.3 Tous les compteurs d’eau équipés de systèmes de
contrôle doivent empêcher ou détecter le flux inversé,
comme stipulé dans l’ISO 4064-1:2014|OIML R 49-1:20
13, 4.2.7.
4.4.2 Liste de contrôle des essais de performance des compteurs d’eau
4.4.2.1 Essais de performance applicables à tous les compteurs d’eau
Paragraphe de Exigence + - Remarques
l’ISO 4064‑1:2014|
OIML R 49-1:2013
Essai de tenue à la pression statique
4.2.10 Le compteur doit pouvoir supporter les pressions
d’essai suivante, sans fuites, ni dommages:
— 1,6 fois la pression maximale admissible pendant
15 min,
— 2 fois la pression maximale admissible pendant
1 min.
Erreurs intrinsèques (d’indication)
Paragraphe de Exigence + - Remarques
l’ISO 4064‑1:2014|
OIML R 49-1:2013
7.2.3 Les erreurs (d’indication) des compteurs d’eau (lors
du mesurage du volume réel) doivent au moins être
déterminées pour les débits suivants:
a) Entre Q et 1,1 Q
1 1
b) Entre Q et 1,1 Q
2 2
c) Entre 0,33 (Q + Q ) et 0,37 (Q + Q )
2 3 2 3
d) Entre 0,67 (Q + Q ) et 0,74 (Q + Q )
2 3 2 3
e) Entre 0,9 Q et Q
3 3
f) Entre 0,95 Q et Q
4 4
et, pour les compteurs combinés:
(g) Entre 0,85 Q et 0,95 Q
x1 x1
(h) Entre 1,05 Q et 1,15 Q
x2 x2
Il convient que le compteur soit soumis à essai sans
raccordement de ses dispositifs temporaires supplé-
mentaires (le cas échéant).
Lors d’un essai, tous les autres facteurs d’influence
doivent être maintenus dans les conditions de réfé-
rence.
D’autres débits peuvent être mesurés en fonction de
la forme de la courbe d’erreur.
1) Les erreurs (d’indication) relatives observées pour
chaque débit ne doivent pas dépasser les erreurs
maximales tolérées données dans l’ISO 4064-1:2014|
OIML R 49-1:2013, 4.2.2 ou 4.2.3. Si l’erreur observée
sur un ou plusieurs compteurs est supérieure à l’er-
reur maximale tolérée à un débit seulement alors que
deux résultats seulement ont été relevés à ce débit,
l’essai à ce débit doit être répété. L’essai doit être
déclaré satisfaisant si deux des trois résultats à ce
débit se situent dans les limites de l’erreur maximale
tolérée et si la moyenne arithmétique des résultats
pour les trois essais à ce débit se situe dans les limites
de l’erreur maximale tolérée.
2) Si toutes les erreurs (d’indication) relatives du
compteur d’eau sont de même signe, au moins une de
ces erreurs ne doit pas dépasser la moitié de l’erreur
maximale tolérée. Dans tous les cas, cette exigence
doit être appliquée de manière équitable au four-
nisseur d’eau et au consommateur (voir également
l’ISO 4064-1:2014|OIML R 49-1:2014, 4.3.3, points 3
et 8).
7.2.4 Le compteur doit être répétable: l’écart-type des trois
mesurages au même débit ne doit pas dépasser un
tiers des erreurs maximales tolérées données dans
l’ISO 4064-1:2014|OIML R 49-1:2014, 4.2.2 ou 4.2.3.
Les essais doivent être réalisés aux débits nominaux
de Q , Q et Q .
1 2 3
Essai de température de l’eau
4.2.8 Les exigences relatives aux erreurs maximales
tolérées doivent être respectées pour toutes les varia-
tions de température de l’eau dans les limites des
conditions de fonctionnement assignées du compteur.
Essai de pression de l’eau
18 © ISO 2014 – Tous droits réservés

Paragraphe de Exigence + - Remarques
l’ISO 4064‑1:2014|
OIML R 49-1:2013
4.2.8 Les exigences relatives aux erreurs maximales
tolérées doivent être respectées pour toutes les
variations de pression de l’eau dans les limites des
conditions de fonctionnement assignées du compteur.
Essai d’inversion du flux
4.2.7 Un compteur d’eau conçu pour mesurer des flux
inversés doit
a) soit soustraire le volume de flux inversé du volume
indiqué,
b) soit enregistrer le volume de flux inversé séparé-
ment.
Les erreurs maximales tolérées dans l’ISO 4064-1:
2014|OIML R 49-1:2014, 4.2.2 ou 4.2.3 doivent être
respectées dans les deux sens du flux.
4.2.7 Un compteur d’eau qui n’est pas conçu pour mesurer
des flux inversés doit
a) soit l’empêcher,
b) soit pouvoir supporter un flux accidentel inversé à
un débit allant jusqu’à Q sans aucune détérioration
ou changement de ses caractéristiques métrologiques
en flux normal.
Caractéristiques du compteur d’eau avec un débit nul
4.2.9 La valeur indiquée sur le totalisateur du compteur
d’eau ne doit pas changer lorsque le débit est nul.
Essai pour la déterminati
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

記事タイトル:ISO 4064-3:2014 - 冷水および温水用水メーター- 第3部:テスト報告書の形式 記事内容:ISO 4064-3:2014|OIML R 49-3:2013は、冷水および温水用水メーターに関連して使用されるテスト報告書の形式を指定しています。ISO 4064-1:2014|OIML R 49-1:2013およびISO 4064-2:2014|OIML R 49-2:2013とも連動しています。

article title: ISO 4064-3:2014 - 冷たい飲用水と温水用の水メーターに関する試験報告書の形式 (Water meters for cold potable water and hot water - Part 3: Test report format) article content: ISO 4064-3:2014|OIML R 49-3:2013は、冷たい飲用水と温水用の水メーターに関連する試験報告書の形式を定めています。この形式は、ISO 4064-1:2014|OIML R 49-1:2013およびISO 4064-2:2014|OIML R 49-2:2013と組み合わせて使用する必要があります。

article title: ISO 4064-3:2014 - 냉정수 및 온수용 수도계를 위한 테스트 보고서 형식 (Water meters for cold potable water and hot water - Part 3: Test report format) article content: ISO 4064-3:2014|OIML R 49-3:2013은 냉정수 및 온수용 수도계를 위한 테스트 보고서 형식을 제시한다. 이 형식은 ISO 4064-1:2014|OIML R 49-1:2013과 ISO 4064-2:2014|OIML R 49-2:2013와 함께 사용되어야 한다.

기사 제목: ISO 4064-3:2014 - 냉수 및 온수용 수도계 - 3부: 시험 보고서 형식 기사 내용: ISO 4064-3:2014|OIML R 49-3:2013은 냉수 및 온수용 수도계에 대한 시험 보고서 형식을 지정한다. 이는 ISO 4064-1:2014|OIML R 49-1:2013 및 ISO 4064-2:2014|OIML R 49-2:2013와 함께 사용되어야 할 목적으로 제정되었다. 이 표준에서 제시된 형식은 이러한 수도계에 대한 시험 결과를 보고하는 가이드 라인을 제공한다.

ISO 4064-3:2014 is a standard that specifies a test report format to be used with ISO 4064-1:2014 and ISO 4064-2:2014 for water meters for cold potable water and hot water. It is also aligned with OIML R 49-3:2013.

ISO 4064-3:2014 is a standard that specifies a test report format for water meters used for cold potable water and hot water. It is meant to be used alongside ISO 4064-1:2014 and ISO 4064-2:2014. The format outlined in this standard provides guidelines for reporting the results of tests conducted on these water meters.

記事のタイトル:ISO 4064-3:2014 - 冷水および温水用水メーター- 第3部:試験報告書フォーマット 記事の内容:ISO 4064-3:2014|OIML R 49-3:2013は、冷水および温水用水メーターに関する試験報告書のフォーマットを規定しています。ISO 4064-1:2014|OIML R 49-1:2013およびISO 4064-2:2014|OIML R 49-2:2013と併用することを目的としています。この標準で指定されたフォーマットは、これらの水メーターの試験結果を報告するためのガイドラインを提供します。

기사 제목: ISO 4064-3:2014 - 냉수 및 온수용 수도계 - 제3부: 테스트 보고서 양식 기사 내용: ISO 4064-3:2014|OIML R 49-3:2013은 냉수 및 온수용 수도계에 사용되는 ISO 4064-1:2014|OIML R 49-1:2013 및 ISO 4064-2:2014|OIML R 49-2:2013와 함께 사용되는 테스트 보고서 양식을 지정한다.

ISO 4064-3:2014, also known as OIML R 49-3:2013, establishes a standard test report format to be used with ISO 4064-1:2014, OIML R 49-1:2013, ISO 4064-2:2014, and OIML R 49-2:2013. This format is specifically designed for water meters used to measure both cold potable water and hot water.