ISO 8124-5:2026
(Main)Safety of toys — Part 5: Determination of total concentration of certain elements in toys
General Information
- Abstract
This document specifies methods of sampling and analysing the total concentration of the elements antimony, arsenic, barium, cadmium, chromium, lead, mercury and selenium from toy materials and from parts of toys. Digestion methods for the elements mentioned in 1.1 are specified for the following types of toy materials: coatings of paint, varnish, lacquer, printing ink, polymer and similar coatings; polymeric and similar materials, including laminates, whether textile-reinforced or not, but excluding other textiles; paper, paperboard and cardboard; natural or synthetic textiles; metallic materials, whether coated or not; other materials, whether mass-coloured or not (e.g. wood, fibreboard, hardboard, bone, and leather); materials intended to leave a trace (e.g. the graphite materials in pencils and liquid ink in pens); pliable modelling materials, including modelling clays and gels; paints to be used as such in the toy, including finger paints, varnishes, lacquers and similar materials in solid or liquid form; packaging materials that form part of the toy or have intended play value (e.g. boxes containing jigsaw puzzles or packaging on which the instructions are included in the case of games. Items such as blister packs containing simple instructions are not included in this category). Digestion methods for glass, ceramic and other siliceous materials or fluorinated polymers or fluorinated polymer coatings are not described. These types of materials are outside the scope of this document. Screening determination (without the need for sample preparation and digestion) of the elements mentioned in 1.1 is specified for plastics using energy dispersive X-ray fluorescence (EDXRF) spectrometry (see Annex B).
- Status
- Published
- Publication Date
- 18-Aug-2026
- Technical Committee
- ISO/TC 181 - Safety of toys
- Drafting Committee
- ISO/TC 181 - Safety of toys
- Current Stage
- 6060 - International Standard published
- Start Date
- 19-Aug-2026
- Due Date
- 03-Nov-2026
- Completion Date
- 19-Aug-2026
Overview
ISO 8124-5:2026 is a key international standard published by the International Organization for Standardization (ISO), forming part of the broader ISO 8124 series focusing on toy safety. This document establishes recognized methods for sampling and analyzing the total concentration of specific heavy metals and elements within toy materials and components. The elements covered by this standard include antimony, arsenic, barium, cadmium, chromium, lead, mercury, and selenium. The correct application of ISO 8124-5:2026 is essential for ensuring that toys meet international chemical safety requirements and for supporting regulatory compliance across global markets.
Key Topics
- Elements Analyzed: The standard specifies procedures for measuring total concentrations of the eight hazardous elements: antimony, arsenic, barium, cadmium, chromium, lead, mercury, and selenium.
- Sample Materials: Testing applies to a wide variety of toy materials, including:
- Coatings (paint, varnish, lacquer, printing ink, polymers)
- Polymers and laminates (except textiles)
- Paper, paperboard, and cardboard
- Natural and synthetic textiles
- Metallic materials, coated or uncoated
- Other materials such as wood, fiberboard, bone, leather
- Modeling clays and gels
- Paints in solid or liquid form, including finger paints
- Packaging materials with intended play value
- Analytical Techniques: The document outlines the necessary digestion methods (hot plate, hot block, microwave digestion systems) for different material types to break down the matrix and allow accurate quantification of elements.
- Screening Method for Plastics: For plastic materials, ISO 8124-5:2026 introduces energy dispersive X-ray fluorescence (EDXRF) spectrometry as a rapid screening tool to determine if further quantitative testing is required.
- Test Portion Preparation: Procedures for sampling, cutting, grinding, and homogenising toy material to ensure representative testing and avoid cross-contamination are thoroughly described.
Applications
ISO 8124-5:2026 is applied by toy manufacturers, testing laboratories, and regulatory authorities to ensure that toys do not contain hazardous levels of toxic elements. Key practical applications include:
- Regulatory Compliance: Aligns toys with chemical safety regulations enforced in various markets, supporting conformity to legal requirements like those found in EN 71-3 and ASTM F963.
- Safety Assurance: Verifies that hazardous element concentrations are within safe limits, reducing risks of toxicity or health hazards to children.
- Quality Control within Manufacturing: Enables quality assurance teams to monitor and test raw materials and final products to prevent excessive levels of restricted substances entering the supply chain.
- Import/Export Assessment: Facilitates the global trade of toys by establishing common testing frameworks recognized worldwide.
- Rapid Screening: The EDXRF method allows for efficient screening of plastics for fast decision-making, helping manufacturers prioritize which batches require full digestion and analysis.
- Risk Assessment: Helps identify whether migration testing (per ISO 8124-3) is needed based on total element concentrations.
Related Standards
When implementing ISO 8124-5:2026, it is beneficial to be aware of related standards that provide further guidance on toy safety:
- ISO 8124-1:2022 - Safety aspects related to mechanical and physical properties of toys
- ISO 8124-3:2020/Amd 1:2023 - Migration of certain elements (soluble limits in toy materials)
- EN 71-3 - European standard for migration of certain elements in toys
- ASTM F963 - Standard consumer safety specification for toy safety in the United States
- ISO 3696 - Water for analytical laboratory use - Specifications and test methods
Together, these international standards form a comprehensive approach to toy safety, ensuring safer products for children through systematic assessment and control of chemical hazards.
Keywords: ISO 8124-5:2026, toy safety standard, determination of elements in toys, toy material testing, heavy metals in toys, EDXRF screening, international toy safety, ISO toy standards, regulatory compliance for toys, children’s toy safety.
Relations
- Effective Date
- 12-Nov-2022
Frequently Asked Questions
ISO 8124-5:2026 is a standard published by the International Organization for Standardization (ISO). Its full title is "Safety of toys — Part 5: Determination of total concentration of certain elements in toys". This standard covers: This document specifies methods of sampling and analysing the total concentration of the elements antimony, arsenic, barium, cadmium, chromium, lead, mercury and selenium from toy materials and from parts of toys. Digestion methods for the elements mentioned in 1.1 are specified for the following types of toy materials: coatings of paint, varnish, lacquer, printing ink, polymer and similar coatings; polymeric and similar materials, including laminates, whether textile-reinforced or not, but excluding other textiles; paper, paperboard and cardboard; natural or synthetic textiles; metallic materials, whether coated or not; other materials, whether mass-coloured or not (e.g. wood, fibreboard, hardboard, bone, and leather); materials intended to leave a trace (e.g. the graphite materials in pencils and liquid ink in pens); pliable modelling materials, including modelling clays and gels; paints to be used as such in the toy, including finger paints, varnishes, lacquers and similar materials in solid or liquid form; packaging materials that form part of the toy or have intended play value (e.g. boxes containing jigsaw puzzles or packaging on which the instructions are included in the case of games. Items such as blister packs containing simple instructions are not included in this category). Digestion methods for glass, ceramic and other siliceous materials or fluorinated polymers or fluorinated polymer coatings are not described. These types of materials are outside the scope of this document. Screening determination (without the need for sample preparation and digestion) of the elements mentioned in 1.1 is specified for plastics using energy dispersive X-ray fluorescence (EDXRF) spectrometry (see Annex B).
This document specifies methods of sampling and analysing the total concentration of the elements antimony, arsenic, barium, cadmium, chromium, lead, mercury and selenium from toy materials and from parts of toys. Digestion methods for the elements mentioned in 1.1 are specified for the following types of toy materials: coatings of paint, varnish, lacquer, printing ink, polymer and similar coatings; polymeric and similar materials, including laminates, whether textile-reinforced or not, but excluding other textiles; paper, paperboard and cardboard; natural or synthetic textiles; metallic materials, whether coated or not; other materials, whether mass-coloured or not (e.g. wood, fibreboard, hardboard, bone, and leather); materials intended to leave a trace (e.g. the graphite materials in pencils and liquid ink in pens); pliable modelling materials, including modelling clays and gels; paints to be used as such in the toy, including finger paints, varnishes, lacquers and similar materials in solid or liquid form; packaging materials that form part of the toy or have intended play value (e.g. boxes containing jigsaw puzzles or packaging on which the instructions are included in the case of games. Items such as blister packs containing simple instructions are not included in this category). Digestion methods for glass, ceramic and other siliceous materials or fluorinated polymers or fluorinated polymer coatings are not described. These types of materials are outside the scope of this document. Screening determination (without the need for sample preparation and digestion) of the elements mentioned in 1.1 is specified for plastics using energy dispersive X-ray fluorescence (EDXRF) spectrometry (see Annex B).
ISO 8124-5:2026 is classified under the following ICS (International Classification for Standards) categories: 97.200.50 - Toys. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 8124-5:2026 has the following relationships with other standards: It is inter standard links to ISO 8124-5:2015. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO 8124-5:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
International
Standard
ISO 8124-5
Second edition
Safety of toys —
2026-08
Part 5:
Determination of total
concentration of certain elements in
toys
Sécurité des jouets —
Partie 5: Détermination de la concentration totale de certains
éléments dans les jouets
Reference number
© ISO 2026
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 Principle . 3
5 Reagents and apparatus . 3
5.1 Reagents .3
5.2 Apparatus .4
6 Selection and composition of test portions . 5
6.1 Selection of test portions.5
6.2 Composition of test portions .5
7 Preparation of test portions . 5
7.1 Overview .5
7.2 Coatings of paint, varnish, lacquer, printing ink, polymer and similar coatings .6
7.3 Polymeric and similar materials, including laminates, whether textile-reinforced or
not, but excluding other textiles .6
7.4 Paper, paperboard and cardboard .6
7.5 Natural or synthetic textiles .6
7.6 Other materials, whether mass-coloured or not .6
7.7 Materials intended to leave a trace .7
7.8 Pliable modelling materials, including modelling clays and gels .7
7.9 Paints, including finger paints, varnishes, lacquers, and similar materials, in solid or
liquid form .7
7.9.1 Materials in solid form .7
7.9.2 Materials in liquid form . . .7
7.10 Metallic materials, whether or not partly coated .7
8 Digestion of test portions . 7
8.1 Overview .7
8.2 Microwave digestion .8
8.2.1 Non-metallic materials .8
8.2.2 Metallic materials, whether or not partly coated .8
8.2.3 Microwave digestion conditions .8
8.2.4 Cooling and dilution .9
8.3 Hot plate and hot block digestion .9
8.3.1 Non-metallic materials .9
8.3.2 Metallic materials whether or not partly coated .10
9 Elemental analysis .10
9.1 Methods of analysis .10
9.2 Method detection limit .10
10 Expression of results .11
11 Test report .11
Annex A (informative) Background and rationale .12
Annex B (informative) Screening determination of certain elements of antimony, arsenic,
barium, cadmium, chromium, lead, mercury and selenium in plastic toys — Energy
dispersive X-ray fluorescence spectrometry .16
Annex C (informative) Precision of the method .21
Bibliography .22
iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 181, Safety of toys.
This second edition cancels and replaces the first edition (ISO 8124-5:2015), which has been technically
revised.
The main changes are as follows:
— The choice of digestion mixtures has been changed from being based on the final quantification
instruments (ICP-MS or ICP-OES) to being based on the type of test part or sample material. A new
digestion mixture (1:9 mixture of hydrochloric acid and nitric acid) has been introduced to replace the
reverse aqua regia mixture.
— An energy dispersive X-ray fluorescence (EDXRF) method for screening the contents of eight elements in
plastic materials in toys has been included as an informative annex.
— Certain elements of additional information contained in Annex A have been integrated into the main
body of the text (e.g. Scope, 7.4).
A list of all parts in the ISO 8124 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
Introduction
This document specifies a method for the determination of the total concentration of certain elements in
toy materials. It can be used to decide whether there is a need to undertake migration testing in accordance
with the method specified in ISO 8124-3 or other equivalent standards, e.g. EN 71-3 or ASTM F963.
NOTE 1 A material can be considered to conform to the requirements of ISO 8124-3 if the total concentration
results are below the soluble limits as prescribed in ISO 8124-3:2020/Amd 1:2023, Table 1. If the soluble limits in
ISO 8124-3:2020/Amd 1:2023, Table 1 are exceeded, migration testing in accordance with ISO 8124-3 is required for
the determination of conformity.
This document presents a screening determination method for eight elements in plastics using the
analytical technique of energy dispersive X-ray fluorescence (EDXRF) spectrometry as an option for quickly
determining whether the screened part or section of a product requires quantitative analysis using the
digestion method.
This document can be used when taking decisions on the conformity of material with regulatory
requirements that impose restrictions on the total concentration of certain elements.
NOTE 2 Where legal conformity requires migration testing, this document can only be used to non-quantitatively
confirm compliance with regulatory limits.
This document has been developed only for the eight elements listed in Table 1. The method specified in
this document can be used for other elements provided adequate analytical performance is demonstrated.
Manufacturers are encouraged to apply the test methods of this document and the limits from ISO 8124-3 to
raw materials used in the manufacture of toys to give increased certainty of conformity to the requirements
of ISO 8124-3.
For further information on the background and rationale behind this document, see Clause A.1.
v
International Standard ISO 8124-5:2026(en)
Safety of toys —
Part 5:
Determination of total concentration of certain elements in
toys
1 Scope
1.1 This document specifies methods of sampling and analysing the total concentration of the elements
antimony, arsenic, barium, cadmium, chromium, lead, mercury and selenium from toy materials and from
parts of toys.
1.2 Digestion methods for the elements mentioned in 1.1 are specified for the following types of toy
materials:
— coatings of paint, varnish, lacquer, printing ink, polymer and similar coatings;
— polymeric and similar materials, including laminates, whether textile-reinforced or not, but excluding
other textiles;
— paper, paperboard and cardboard;
— natural or synthetic textiles;
— metallic materials, whether coated or not;
— other materials, whether mass-coloured or not (e.g. wood, fibreboard, hardboard, bone, and leather);
— materials intended to leave a trace (e.g. the graphite materials in pencils and liquid ink in pens);
— pliable modelling materials, including modelling clays and gels;
— paints to be used as such in the toy, including finger paints, varnishes, lacquers and similar materials in
solid or liquid form;
— packaging materials that form part of the toy or have intended play value (e.g. boxes containing jigsaw
puzzles or packaging on which the instructions are included in the case of games. Items such as blister
packs containing simple instructions are not included in this category).
Digestion methods for glass, ceramic and other siliceous materials or fluorinated polymers or fluorinated
polymer coatings are not described. These types of materials are outside the scope of this document.
1.3 Screening determination (without the need for sample preparation and digestion) of the elements
mentioned in 1.1 is specified for plastics using energy dispersive X-ray fluorescence (EDXRF) spectrometry
(see Annex B).
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 8124-1:2022, Safety of toys — Part 1: Safety aspects related to mechanical and physical properties
ISO 3696, Water for analytical laboratory use — Specification and test methods
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
base material
material upon which coatings (3.2) can be formed or deposited
[SOURCE: ISO 8124-3:2020, 3.1]
3.2
coating
layers of material formed or deposited on the base material (3.1) of a toy, including paints, varnishes,
lacquers, inks, polymers, or other substances of a similar nature, whether they contain metallic particles
or not, no matter how they have been applied to the toy, and which can be removed by scraping (3.8) with a
sharp blade
Note 1 to entry: This definition includes metallic coatings deposited on a metal surface such as an electroplated
coating. However, electroplating will only require testing if it can be removed by scraping (3.8); otherwise, it may be
tested with the base material.
[SOURCE: ISO 8124-3:2020, 3.2, modified — Note 1 to entry has been added.]
3.3
complete digestion
complete breakdown of the original material leaving only insoluble residues
3.4
composite test portion
test portion (3.9) that is composed of more than one similar material type or colour of material
3.5
method detection limit
MDL
three times the standard deviation of the result obtained in the method blank by the laboratory carrying out
the analysis
[SOURCE: ISO 8124-3:2020, 3.3]
3.6
laboratory sample
toy either in the form in which it is marketed, or in the form in which it is intended to be marketed
Note 1 to entry: The definition does not preclude the taking of raw materials used to manufacture the toy, provided
they are representative of the final toy. See 1.1 and Clause B.1.
3.7
method blank
solution that has undergone the same digestion processes used for the digestion of test portions (3.9) and
consists of all reagents excluding the test portion
3.8
scraping
mechanical process for removal of coatings (3.2) down to the base material (3.1)
[SOURCE: ISO 8124-3:2020, 3.7]
3.9
test portion
single material taken from an accessible part of a laboratory sample (3.6)
Note 1 to entry: This definition precludes the compositing of dissimilar materials, e.g. compositing textiles and paint
coatings is not permitted.
3.10
screening determination
analytical procedure to determine the presence or absence of substances in the representative part or
section of a product, relative to the value or values chosen as the criterion for presence, absence or further
testing
4 Principle
The prepared test portion is digested in highly acidic conditions at high temperature using a hot plate
digestion, a hot block digestion technique, or a microwave digestion system. Hot acid digestion destroys the
material matrix allowing the elements of interest to be solubilized and quantified by a suitable analytical
instrument (see 9.2).
NOTE The EDXRF screening determination (3.10) method is capable of directly analysing materials without the
need for sample preparation and digestion. See Annex B.
5 Reagents and apparatus
5.1 Reagents
Only reagents of recognized analytical grade or equivalent shall be used. The concentration of the analyte or
interfering substances in the reagents and water shall be negligible compared to the lowest concentration to
be determined.
“Trace metal” grade or equivalent reagents shall be used for the calibration standards used for the final
instrumentation quantification stage, as follows.
5.1.1 Nitric acid, concentrated, 1,40 g/ml, a mass fraction of 65 %, “analytical” grade.
5.1.2 Nitric acid, 1:9 dilution: Add 100 ml concentrated nitric acid (5.1.1) to 500 ml water (5.1.4). Dilute to
1 000 ml with water (5.1.4).
5.1.3 Hydrochloric acid, concentrated, 1,19 g/ml, a mass fraction of 37 %, “analytical” grade.
5.1.4 Water, of at least grade 3 purity, in accordance with ISO 3696.
5.1.5 Hydrogen peroxide, a mass fraction of 30 %.
NOTE Hydrogen peroxide which is not stabilized can be stored according to the instruction of the manufacturer.
5.1.6 Methylene chloride, “analytical” grade.
5.1.7 Acetone/ethanol solution, 1:1 mixture of absolute ethanol and acetone (“analytical” grades).
5.1.8 Acid mixture A, 1:9 mixture of hydrochloric acid (5.1.3) and nitric acid (5.1.1)
5.1.9 Acid mixture B, i.e. aqua regia, 3:1 mixture of hydrochloric acid (5.1.3) and nitric acid (5.1.1)
NOTE The acid mixtures A and B are best prepared in situ in the presence of test sample. The order of mixing the
acids is not critical but can be the same for consistency.
5.2 Apparatus
All glassware shall be soaked in nitric acid (5.1.2) for at least 2 h and then rinsed in deionized water before
use.
5.2.1 Microwave digestion system
Microwave sample preparation system equipped with a sample holder and high-pressure microwave
digestion vessels (5.2.2, high pressure microwave digestion vessel).
NOTE 1 Some newer models of microwave digestion systems do not utilize high-pressure digestion vessels and
these systems are considered as a suitable alternative provided they give an equivalent performance.
NOTE 2 There are many safety and operational recommendations specific to the model and manufacturer of the
microwave equipment used in individual laboratories. The analyst is required to consult the specific equipment
manual, manufacturer and literature for proper and safe operation of the microwave equipment and vessels (see
Clause A.2).
5.2.2 High pressure microwave digestion vessel
Closed-top vessel specifically designed for microwave digestion, of suitable capacity. It is recommended to
use a vessel capable of withstanding a temperature of at least 225 °C and an internal pressure of at least
3 000 kPa. The liner of the vessel shall be PTFE (polytetrafluoroethylene)/TFM [tris-(α-trifluoromethyl-β,β-
difluorovinyl)-1,3,5-benzenetricarboxylate], or PTFE/PFA (perfluoroalkoxy ethylene) or another chemically
inert material. Vessels shall also be equipped with a safety relief valve or disc that will prevent vessel
rupture or ejection of the vessel cap. The inner liners shall be inspected regularly to check for any chemical
or physical degradation.
NOTE Internal pressures in excess of 3 000 kPa can occur with some samples, e.g. crayons. A suitable pressure-
rated vessel (e.g. 5 000 kPa) can be used in these cases.
5.2.3 Scalpel, or other suitable scraping or cutting tools.
5.2.4 Laboratory grinding mill.
5.2.5 Rotary grinder, preferably with carbide burr grinders.
1)
5.2.6 Centrifuge, capable of centrifuging at (5 000 ± 500) g , with compatible tubes.
5.2.7 Analytical balance, capable of measuring accurately to 0,000 1 g.
5.2.8 Polypropylene or PTFE microfilters, pore size 0,45 µm.
5.2.9 Volumetric flasks, 10 ml, 25 ml or 100 ml capacity with stopper.
5.2.10 Pipettes, 1 ml, 5 ml, 10 ml, 20 ml, etc.
5.2.11 Beakers, 25 ml, 50 ml, 100 ml, etc.
1) 1 g = 9,806 65 m/s .
5.2.12 Electric hot plate, suitable for operation at surface temperatures up to at least 140 °C.
NOTE Provided that the hot plate is capable of handling the extra heating required, use of a 12 mm to 25 mm thick
heat-resistant glass plate placed on the hot plate can help reduce the presence of hot spots common to electric hot
plates.
5.2.13 Filter paper and funnel.
5.2.14 Hot block digester, heated metal block with variable temperature settings up to at least 140 °C
(optionally can have programmable settings and temperature ramps) with compatible digestion vessels of
suitable capacity.
6 Selection and composition of test portions
6.1 Selection of test portions
NOTE See A.1.2 for a discussion on practical considerations when deciding whether to use composite test
portions.
Test portions shall be taken from accessible parts (see ISO 8124-1) of the laboratory sample in accordance
with Clause 7 (preparation of test portions). When appropriate, the laboratory sample shall be subjected to
relevant tests in accordance with ISO 8124-1, before the accessibility is considered. Identical materials in the
laboratory sample can be combined and treated as a single test portion, but the use of additional laboratory
samples is not permitted. If it is not possible to obtain at least 10 mg, no further testing shall be conducted
and this shall be reported under Clause 11 c) (test report).
The test portion mass should be in the region of 100 mg where sufficient material is available.
6.2 Composition of test portions
Up to three test portions can be combined to form a composite test portion provided that the required
method detection limit can still be achieved (see A.1.2) and the combined materials are similar in nature.
The compositing of dissimilar materials is not permitted, e.g. compositing textiles and paint coatings. When
calculating the concentration of a target element in a material, it is assumed that all of that element found
in the digested sample originated from any one of the composited materials. Using this assumption and
the masses of the individual materials, the total concentration of the target element is calculated for each
individual material in the composite test portion.
7 Preparation of test portions
7.1 Overview
Materials from the laboratory sample are selected for testing in accordance with Clause 6 and removed
using cutting tools such as scalpels, razor blades, scissors and grinding and milling tools as described in
the following subclauses. If a grinding apparatus [such as a mill (5.2.4) or rotary grinding tool (5.2.5) with
disposable grinding bits] is used, then any contaminated parts shall be thoroughly cleaned or disposed of
between uses to prevent cross-contamination. Ensure that the device itself cannot contaminate the material
being prepared.
In 7.2 to 7.10, collect sufficient material to obtain a test portion of between 10 mg and 100 mg. In cases
where less than 10 mg of material is available (see 6.1) no further testing is required and this is reported
under Clause 11 c) (test report).
Digest the prepared test portion according to the procedures described in 8.2 (microwave digestion) or 8.3
(hot plate and hot block digestion of test portion).
NOTE The EDXRF screening determination (3.10) method is capable of directly analysing materials without the
need for sample preparation and digestion. See Annex B.
7.2 Coatings of paint, varnish, lacquer, printing ink, polymer and similar coatings
Remove each different coating from the laboratory sample by scraping down to the base material, taking
care to avoid the inclusion of the base material. Where lithographic coatings (dot printing) are present,
it is impractical to separate the individual colours. These coatings shall be removed in such a way that a
representative test portion is obtained.
For some coatings deposited on a non-polymeric base material, it is permissible to add a few drops of
solvent, such as acetone/ethanol (5.1.7) mixture or methylene chloride (5.1.6), to soften the paint and aid
in its removal from the base material. In the first instance, acetone/ethanol (5.1.7) should be used. If this
treatment is not effective in aiding removal, methylene chloride can be used under a fume hood.
If a solvent treatment is used, ensure that all traces of solvent have been removed by evaporation prior to
microwave digestion (see 8.2). Divide removed coatings into small pieces having a maximum length in any
direction of 2 mm in order to facilitate efficient digestion.
7.3 Polymeric and similar materials, including laminates, whether textile-reinforced or not,
but excluding other textiles
Scrape off, cut or grind the clean, dry material into pieces having a maximum length in any dimension of
2 mm using a scalpel or other suitable scraping or cutting tool.
7.4 Paper, paperboard and cardboard
Cut the material into pieces with a maximum length in any dimension of 2 mm using a suitable cutting tool.
If the paper or paperboard to be tested is coated with paint, varnish, lacquer, printing ink, adhesive or
similar coating, test portions of the coating shall not be taken separately. In such cases, take test portions
from the material so that they also include representative parts of the coated area.
Material that is printed, where the ink has become part of the base material, is prepared as though it is
unprinted. For example, a paper can be printed with ink and coated with a thick varnish. The varnish can be
scraped off and would be regarded as a coating but the remaining printed paper would be regarded as paper.
7.5 Natural or synthetic textiles
Cut the material into pieces having a maximum length in any dimension of 2 mm using a suitable cutting
tool.
If the sample is not of a uniform material or colour, where possible, obtain a test portion from each different
material or colour present in a mass greater than 100 mg. Materials or colours present in amounts between
10 mg and 100 mg shall form part of the test portion obtained from the main material.
Test portions taken from patterned textiles shall be representative of the whole material.
7.6 Other materials, whether mass-coloured or not
Cut, scrape or grind the material into pieces having a maximum length in any dimension of 2 mm using a
suitable cutting tool.
7.7 Materials intended to leave a trace
Obtain test portions from each different material in the laboratory sample, in the form in which it appears
in the laboratory sample, ensuring that the material is cut into pieces having a maximum length in any
dimension of 2 mm.
7.8 Pliable modelling materials, including modelling clays and gels
Obtain test portions from each different material or colour of material in the form in which it appears in
the laboratory sample, i.e. without allowing the material to dry out. Cut the material into pieces having a
maximum length in any dimension of 2 mm.
7.9 Paints, including finger paints, varnishes, lacquers, and similar materials, in solid or
liquid form
7.9.1 Materials in solid form
Using a suitable tool, grind, crush or cut the sample into particles having a maximum length in any dimension
of 2 mm.
Finger paint supplied in the form of powder shall be diluted with water (5.1.4) in accordance with the
manufacturer’s instructions and then prepared according to 7.9.2.
7.9.2 Materials in liquid form
Ensure that all settled material has been incorporated into the sample by scraping and long-term mechanical
shaking. Immediately prior to sampling, ensure the liquid is homogenized by stirring or shaking for 5 min.
Obtain a test portion of between 10 mg and 100 mg. In cases where less than 10 mg of material is available
(see 6.1) no further testing is required and this is reported under Clause 11 c) (test report). Digest the
prepared test portion according to 8.2 or 8.3.
If the liquid is intended to solidify or dry during intended use, it shall be coated onto a clean glass plate and
dried to constant weight before taking test portions as described in 7.2.
Report under Clause 11 g) (test report) whether the material was tested in the dry or liquid state.
7.10 Metallic materials, whether or not partly coated
Where practical, obtain the test portion from an uncoated part of the metallic material. If the metal
part is partially coated, remove any coatings (including electroplated coatings) that can be scraped off.
Electroplating that cannot be removed by scraping shall be tested together with the metallic base material.
NOTE Metallic materials that are completely coated so that no metal is accessible, as defined in ISO 8124-1, are
not tested. However, this approach differs in legal texts in certain regions of the world. For example, the United States
Consumer Product Safety Improvement Act of 2008 requires that surface coatings do not render the base material
inaccessible. Furthermore, ISO 8124-3 only requires the testing of metallic components that are small parts, i.e. fit
wholly within the test cylinder shown in ISO 8124-1:2022, Figure 26.
Using suitable cutting and grinding tools, obtain a test portion of between 10 mg and 100 mg. In cases where
less than 10 mg of material is available (see 6.1) no further testing is required and this is reported under
Clause 11 c) (test report).
8 Digestion of test portions
8.1 Overview
WARNING 1 Hydrogen peroxide can be used in hot plate and block digestions, but if used in microwave digestions,
great care shall be taken in order to avoid potential build-up of pressure during the heating cycle which can cause
accidents and consequential loss of sample.
WARNING 2 Fumes from nitric acid and hydrochloric acid are toxic; perform the operations described in 8.2 and
8.3 in a fume hood, especially when using hot plate and hot block digestion.
The steps in 8.2 and 8.3 are carried out under a fume hood or in the microwave digestion system (5.2.1).
Subclauses 8.2 and 8.3 describe how to digest the test portion using either acid mixture A (5.1.8) or acid
mixture B (5.1.9) followed by dilution and quantification using the instrumental technique of inductively
coupled plasma mass spectroscopy (ICP-MS) or inductively coupled plasma optical emission spectroscopy
(ICP-OES). Other instrumental techniques can be used provided that they fulfil the criteria relating to
method detection limits set out in Table 1.
Method blanks shall be run with each batch of test portions digested according to 8.2 or 8.3.
If arsenic or mercury is being quantified, the digestion technique shall use a closed vessel on the hot block
digester (5.2.14) or in the microwave digestion system (5.2.1).
NOTE Certain volatile elements such as arsenic and mercury are prone to be lost during hot plate digestion but
laboratory testing has shown that mercury is not lost when using aqua regia acid mix.
Neither the microwave digestion system (5.2.1) nor the conditions to be used are specified in this document.
The test laboratory may therefore use any appropriate microwave system and digestion conditions. Examples
of microwave digestion temperature programmes are given in 8.2.3. However, the test laboratory should
determine the optimum digestion conditions based upon the equipment manufacturer’s recommendations
and the laboratory’s own experiences with digestion of various material types. It is essential to achieve
complete digestion. If the safety relief valve or system activates during the digestion, the analysis shall be
repeated using a new test portion and a suitably pressure rated vessel.
8.2 Microwave digestion
8.2.1 Non-metallic materials
NOTE 1 Non-metallic materials include materials mentioned in 1.2, except for metallic materials.
Weigh accurately (to the nearest 0,1 mg) between 10 mg and 100 mg of the test portion prepared in Clause 7
(preparation of test portions) into a microwave digestion vessel (5.2.2) then carefully add 6 ml of acid
mixture A (5.1.8).
To reduce the potential for matrix interference, it is recommended to use the same grade reagents as the
calibration standards. The volume of the acid mixture can be adjusted to meet the minimum requirements
of the apparatus, provided that the specified ratio is maintained.
Allow the initial reaction to subside before sealing the vessel in accordance with the manufacturer's
instructions.
8.2.2 Metallic materials, whether or not partly coated
Replace the acids in 8.2.1 with acid mixture B, i.e. aqua regia (5.1.9) and proceed as described in
subclause 8.2.1.
8.2.3 Microwave digestion conditions
The microwave digestion system (5.2.1) shall be loaded with the sealed digestion vessel (5.2.2) and digested
according to the conditions and temperatures recommended by the equipment manufacturers, as optimized
by the test laboratory.
Examples of microwave digestion temperature programmes are given as below:
— For non-metallic materials: the temperature of each sample digestion solution rises to 210 °C in
approximately 20 min and holds for 15 min at 210 °C.
— For metallic materials: the temperature of each sample digestion solution rises to 175 °C in approximately
10 min and holds for 5 min at 175 °C.
8.2.4 Cooling and dilution
Following completion of the digestion programme, allow the vessels to cool for at least 5 min before
transferring to a fume hood for further cooling until the temperature of the sample is less than 40 °C
(typically at least 1 h). Carefully open the vessel and check that the test portion has been completely digested.
If the digestion is incomplete (e.g. evidence of original sample or solid lumps of charred test portion), the test
should be repeated using a new test portion and alternative digestion conditions until complete digestion is
achieved. If complete digestion cannot be obtained despite these efforts, an alternative method shall be used
(see A.3.2, incomplete digestion). Such alternative methods are outside the scope of this document, and their
use shall be clearly indicated in the final test report [see Clause 11 d), test report].
When the digestion is complete, quantitatively transfer the digestate with washings to a beaker and
evaporate to about 1 ml on a hot plate. Allow to cool and then add about 4 ml to 5 ml water followed by 3
to 4 drops of hydrochloric acid (5.1.3). Filter (5.2.8 or 5.2.13) the digest solution into a 25 ml volumetric
flask (5.2.9) or where limited sample was available, into a 10 ml volumetric flask. Dilute to the mark with
water (5.1.4), seal the flask with a stopper and mix thoroughly. This diluted digest shall be subjected to
instrumental quantification as soon as practically possible.
Alternatively, the step of evaporating the digestion solution described above may be omitted, provided the
instrument can tolerate high acidity matrix samples. Laboratories should prepare the working calibration
standards using the same acid matrix in the digestates in that case.
8.3 Hot plate and hot block digestion
8.3.1 Non-metallic materials
NOTE Non-metallic materials include materials mentioned in 1.2, except for metallic materials.
Weigh accurately (to the nearest 0,1 mg) between 10 mg and 100 mg of the test portion prepared in Clause 7
into a clean 25 ml beaker (5.2.11) or hot block digestion vessel, then carefully add 6 ml of acid mixture
A (5.1.8). The volume of the acid mixture may be adjusted to meet the minimum requirements of the
apparatus, provided that the specified ratio is maintained. Allow any reaction that occurs to subside, add
1 ml of hydrogen peroxide (5.1.5) (optional) drop wise and cover with a watch glass if the reaction vessel is
a beaker. Heat in a hot block (5.2.14) or on a hot plate (5.2.12) (surface temperature approximately 140 °C)
until most of the acid has evaporated.
In the case of hot plate digestion, evaporate to a final volume of approximately 1 ml. The composition of
this 1 ml digestate is predominantly concentrated nitric acid with reaction products from the sample. To
eliminate the possibility of cross-contamination or sample loss, avoid boiling or evaporating to complete
dryness.
If brown fumes are still observed after 1 h or particles are observed in the solution, continue the heating.
Add a few millilitres of acid mixture A (5.1.8) as necessary to prevent the sample from becoming dry. When
the sample ceases to emit brown fumes, the digestion is complete. Remove the reaction vessel from the hot
plate or hot block and allow to cool to room temperature.
If a complete digestion cannot be achieved, an alternative method shall be used (see A.3.2, incomplete
digestion).
Rinse the beaker and bottom of the watch glass with about 4 ml to 5 ml water (5.1.4) and then add 3 to 4
drops of concentrated hydrochloric acid. Filter (5.2.8 or 5.2.13) if necessary and then transfer the reaction
mixture with washings into a 25 ml volumetric flask or where limited sample was available, into a 10 ml
volumetric flask, make up to volume using water, seal the flask with a stopper and mix thoroughly. This
diluted digest shall be subjected to instrumental quantification as soon as practically possible.
Where the digestate is made up to 25 ml volume, the diluted solution contains approximately a volume
fraction of 4 % nitric acid (1 ml) with minor (<0,5 %) hydrochloric acid. Calibration standards used for
instrumental analysis should be made with these levels of acids. Similar matching of acidity should be
carried out if the digestate is made up to 10 ml.
Prior to instrumental quantification, remove any particulate in the solution by filtration (5.2.8 or 5.2.13), by
centrifugation (5.2.6), or by allowing the solution to settle.
This diluted digest shall be subjected to instrumental quantification as soon as practically possible.
8.3.2 Metallic materials whether or not partly coated
Replace the acids in 8.3.1 with acid mixture B i.e. aqua regia (5.1.9) and proceed as described in 8.3.1. To
reduce the potential for matrix interference, it is recommended to use the same grade reagents as the
calibration standards.
9 Elemental analysis
9.1 Methods of analysis
The final solution is analysed for the presence of elements by inductively coupled plasma optical emission
spectrometry (ICP-OES), inductively coupled plasma mass spectrometry (ICP-MS) or other suitable
techniques.
The precision of this method is shown in Annex C.
NOTE The EDXRF screening determination (3.10) method is capable of directly analysing materials without the
need for sample preparation and digestion. See Annex B.
9.2 Method detection limit
For the quantitative analysis of the diluted digests, the choice of the final instrumental a
...