IEC/TR 62348:2012(E) which is a technical report, provides a tool to assist users of IEC 60601-1:2005 to assess the impact of the most significant changes in Amendment 1:2012. This technical report also provides a tool to assist users of IEC 60601-1 to trace requirements between the third edition and their source in the documents that form the basis of the third edition; principally the second edition as amended. This report is intended to be used by:
- those who must align standards based on the second edition of IEC 60601-1 with the third edition as amended;
- manufacturers of medical electrical equipment or medical electrical systems; and
- health care regulatory authorities, test houses and other organizations responsible for implementing standards for medical electrical equipment and medical electrical systems. This second edition cancels and replaces the first edition published in 2006. The second edition retains the mapping that traces the requirements of IEC 60601-1:2005 and its Amendment A1:2012 (Edition 3.1) from their source in the documents that relate to IEC 60601-1:1998 and its amendments (Edition 2.2). The second edition adds an assessment of the impact of the most significant changes in Amendment 1:2012.

Status
Published
Publication Date
11-Dec-2012
Drafting Committee
JMT 1 - TC 62/SC 62A/JMT 1
Current Stage
PPUB - Publication issued
Start Date
12-Dec-2012
Completion Date
15-Mar-2013

Buy Documents

Technical report

IEC TR 62348:2012 - Assessment of the impact of the most significant changes in Amendment 1 to IEC 60601-1:2005 and mapping of the clauses of IEC 60601-1:2005 to the previous edition Released:12/12/2012

ISBN:978-2-8322-0548-8
English language (101 pages)
sale 15% off
Preview
sale 15% off
Preview

Overview

IEC TR 62348:2012 is a Technical Report from the International Electrotechnical Commission that helps users transition from IEC 60601-1:2005 (Edition 3) to the amended edition (Amendment 1:2012, often referred to as Edition 3.1). The report provides an impact assessment of the most significant changes introduced by Amendment 1:2012 and a clause-by-clause mapping between IEC 60601-1:2005 and the previous (second) edition of the standard. Edition 2.0 (2012) cancels and replaces the 2006 first edition, retaining the original mapping while adding the new impact assessment.

Key topics and technical requirements

IEC TR 62348:2012 focuses on practical points that affect compliance, design, testing and documentation for medical electrical equipment (ME EQUIPMENT) and medical electrical systems (ME SYSTEMS):

  • Risk management and essential performance: Restructuring of risk-management requirements and clarified application of essential performance concepts.
  • Mapping of clauses: Traceability from Edition 3.1 back to Edition 2.x and source documents to support standards alignment.
  • Patient-contact and applied parts: Requirements for parts that contact the patient, including defibrillation-proof applied parts and energy reduction testing.
  • Electrical safety and leakage current: Changes affecting allowable leakage currents, functional earth conductors, measurement methods and dielectric/insulation testing.
  • Supply mains and power interruptions: Revised SUPPLY MAINS test characteristics and requirements for interruption/standby behaviour.
  • Programmable Electrical Medical Systems (PEMS): Validation, design, and network/IT integration considerations.
  • Marking, labelling and accompanying documents: New requirements for legibility, unique version identification, disclosure of essential performance, and instructions for use.
  • Environmental and mechanical aspects: Humidity preconditioning, mechanical strength, instability/overbalancing, cleaning and disinfection, spillage resistance.
  • Special topics: Transformers construction, Y capacitors use, fuses/thermal cut-outs marking, microwave/laser radiations, sterilization- related tests.

The report categorizes changes by impact level (significant, moderate, substantial) to guide prioritization.

Practical applications

IEC TR 62348:2012 is a practical tool for:

  • Performing a gap analysis and transition plan when moving from IEC 60601-1 second edition to Edition 3.1.
  • Updating technical files, design controls, test plans and regulatory submissions to reflect Amendment 1:2012.
  • Identifying which clauses require design changes, new testing, or revised documentation.
  • Preparing test houses and conformity assessment bodies to interpret and enforce the amended requirements.

Who should use this standard

  • Manufacturers of medical electrical equipment and systems planning or executing a standards transition.
  • Regulatory authorities, notified bodies and compliance test labs assessing conformity to IEC 60601-1:2005 + A1:2012.
  • Standards committees and national standards bodies aligning national standards or normative references.
  • Design engineers, quality and regulatory affairs professionals involved in medical device compliance and risk management.

Related standards

  • IEC 60601-1:2005 (Edition 3) and Amendment 1:2012 (Edition 3.1)
  • IEC 60601-1 second edition and its amendments
  • IEC 60601-1-1, IEC 60601-1-4 (mapping tables included in the report)

Keywords: IEC TR 62348:2012, IEC 60601-1:2005, Amendment 1:2012, medical electrical equipment, risk management, PEMS, leakage current, clause mapping, standards transition.

Relations

Effective Date
05-Sep-2023

Buy Documents

Technical report

IEC TR 62348:2012 - Assessment of the impact of the most significant changes in Amendment 1 to IEC 60601-1:2005 and mapping of the clauses of IEC 60601-1:2005 to the previous edition Released:12/12/2012

ISBN:978-2-8322-0548-8
English language (101 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

IEC TR 62348:2012 is a technical report published by the International Electrotechnical Commission (IEC). Its full title is "Assessment of the impact of the most significant changes in Amendment 1 to IEC 60601-1:2005 and mapping of the clauses of IEC 60601-1:2005 to the previous edition". This standard covers: IEC/TR 62348:2012(E) which is a technical report, provides a tool to assist users of IEC 60601-1:2005 to assess the impact of the most significant changes in Amendment 1:2012. This technical report also provides a tool to assist users of IEC 60601-1 to trace requirements between the third edition and their source in the documents that form the basis of the third edition; principally the second edition as amended. This report is intended to be used by: - those who must align standards based on the second edition of IEC 60601-1 with the third edition as amended; - manufacturers of medical electrical equipment or medical electrical systems; and - health care regulatory authorities, test houses and other organizations responsible for implementing standards for medical electrical equipment and medical electrical systems. This second edition cancels and replaces the first edition published in 2006. The second edition retains the mapping that traces the requirements of IEC 60601-1:2005 and its Amendment A1:2012 (Edition 3.1) from their source in the documents that relate to IEC 60601-1:1998 and its amendments (Edition 2.2). The second edition adds an assessment of the impact of the most significant changes in Amendment 1:2012.

IEC/TR 62348:2012(E) which is a technical report, provides a tool to assist users of IEC 60601-1:2005 to assess the impact of the most significant changes in Amendment 1:2012. This technical report also provides a tool to assist users of IEC 60601-1 to trace requirements between the third edition and their source in the documents that form the basis of the third edition; principally the second edition as amended. This report is intended to be used by: - those who must align standards based on the second edition of IEC 60601-1 with the third edition as amended; - manufacturers of medical electrical equipment or medical electrical systems; and - health care regulatory authorities, test houses and other organizations responsible for implementing standards for medical electrical equipment and medical electrical systems. This second edition cancels and replaces the first edition published in 2006. The second edition retains the mapping that traces the requirements of IEC 60601-1:2005 and its Amendment A1:2012 (Edition 3.1) from their source in the documents that relate to IEC 60601-1:1998 and its amendments (Edition 2.2). The second edition adds an assessment of the impact of the most significant changes in Amendment 1:2012.

IEC TR 62348:2012 is classified under the following ICS (International Classification for Standards) categories: 11.040.01 - Medical equipment in general. The ICS classification helps identify the subject area and facilitates finding related standards.

IEC TR 62348:2012 has the following relationships with other standards: It is inter standard links to IEC TR 62348:2006. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

IEC TR 62348:2012 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


IEC/TR 62348 ®
Edition 2.0 2012-12
TECHNICAL
REPORT
colour
inside
Assessment of the impact of the most significant changes in Amendment 1 to
IEC 60601-1:2005 and mapping of the clauses of IEC 60601-1:2005 to the
previous edition
IEC/TR 62348:2012(E)
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.

IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.

Useful links:
IEC publications search - www.iec.ch/searchpub Electropedia - www.electropedia.org
The advanced search enables you to find IEC publications The world's leading online dictionary of electronic and
by a variety of criteria (reference number, text, technical electrical terms containing more than 30 000 terms and
committee,…). definitions in English and French, with equivalent terms in
It also gives information on projects, replaced and additional languages. Also known as the International
withdrawn publications. Electrotechnical Vocabulary (IEV) on-line.

IEC Just Published - webstore.iec.ch/justpublished Customer Service Centre - webstore.iec.ch/csc
Stay up to date on all new IEC publications. Just Published If you wish to give us your feedback on this publication
details all new publications released. Available on-line and or need further assistance, please contact the
also once a month by email. Customer Service Centre: csc@iec.ch.

IEC/TR 62348 ®
Edition 2.0 2012-12
TECHNICAL
REPORT
colour
inside
Assessment of the impact of the most significant changes in Amendment 1 to

IEC 60601-1:2005 and mapping of the clauses of IEC 60601-1:2005 to the

previous edition
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
PRICE CODE
XD
ICS 11.040 ISBN 978-2-83220-548-8

– 2 – TR 62348 © IEC:2012(E)
CONTENTS
FOREWORD . 6
INTRODUCTION . 8
1 Scope . 9
2 Normative references . 9
3 Terms and definitions . 9
4 Assessment of the changes in Amendment 1:2012 . 9
5 Changes impacting many users of the standard . 13
5.1 Updated normative references (Clause 2) [Moderate impact] . 13
5.2 Definition and rationale for EXPECTED SERVICE LIFE (Definition 3.28 and
rationale for 4.4) [Moderate impact] . 13
5.3 Restructuring of RISK MANAGEMENT (Subclause 4.2) [Significant impact] . 13
5.4 Application of ESSENTIAL PERFORMANCE (Subclause 4.3) [Significant impact] . 13
5.5 Revision to the concept of equivalent safety (Subclause 4.5) [Significant
impact] . 14
5.6 Requirements for parts that contact the PATIENT (Subclause 4.6) [Significant
impact] . 14
5.7 Steady-state measurement of input of the ME EQUIPMENT or ME SYSTEM
(Subclause 4.11) [Moderate impact] . 14
5.8 Simultaneous fault testing (Subclause 5.1) [Moderate impact] . 14
5.9 SUPPLY MAINS test characteristics (Subclause 5.5 a)) [Significant impact] . 15
5.10 Humidity preconditioning (Subclause 5.7) [Significant impact] . 15
5.11 Actuating mechanisms (Subclause 5.9.2.3) [Moderate impact] . 15
5.12 Legibility of markings (Subclause 7.1.2) [Moderate impact] . 15
5.13 Identification of ME EQUIPMENT, parts and ACCESSORIES (Subclauses 7.2.1
and 7.2.4) [Significant impact] . 15
5.14 Marking of fuses, THERMAL CUT-OUTS and OVER-CURRENT RELEASES
(Subclause 7.3.4) [Moderate impact] . 15
5.15 Marking of stand-by control (Subclause 7.4.2) [Moderate impact] . 15
5.16 Use of safety signs (Subclause 7.5) [Moderate impact] . 16
5.17 General requirements for ACCOMPANYING DOCUMENTS (Subclause 7.9.1)
[Significant impact] . 16
5.18 Disposal of waste (Subclause 7.9.2.15) [Moderate impact] . 16
5.19 Unique version identification (Subclauses 7.9.2.19 and 7.9.3.1) [Moderate
impact] . 16
5.20 Disclosure of ESSENTIAL PERFORMANCE (Subclause 7.9.3.1) [Significant
impact] . 16
5.21 Change to the fundamental rule of protection against electric shock
(Subclause 8.1 b)) [Moderate impact] . 16
5.22 Use of Y capacitors as a MEANS OF PROTECTION (Subclauses 8.5.1.2 and
8.5.1.3) [Moderate impact] . 16
5.23 Requirements for impedance and current-carrying capabilities (Subclause
8.6.4) [Moderate impact] . 17
5.24 Allowable values of LEAKAGE CURRENTS that can flow in a FUNCTIONAL EARTH
CONDUCTOR (Subclause 8.7.3) [Substantial impact] . 17
5.25 Measurement of the EARTH LEAKAGE CURRENT and current in functional earth
connection (Subclause 8.7.4.5) [Moderate impact] . 17
5.26 Clarification of the insulation to be tested (Subclause 8.8.1) [Significant
impact] . 17

TR 62348 © IEC:2012(E) – 3 –
5.27 Spacing for one MEANS OF OPERATOR PROTECTION for a WORKING VOLTAGE of
25 V r.m.s. (Table 16) [Moderate impact] . 17
5.28 Measurement of CREEPAGE DISTANCES AND AIR CLEARANCES (Subclause 8.9.4)
[Moderate impact]. 17
5.29 Spillage on ME EQUIPMENT and ME SYSTEMS (Subclause 11.6.3) [Moderate
impact] . 17
5.30 Cleaning and disinfection of ME EQUIPMENT and ME SYSTEMS (Subclause
11.6.6) [Moderate impact] . 18
5.31 Interruption of the power supply / SUPPLY MAINS to ME EQUIPMENT or
ME SYSTEM (Subclauses 11.8 and 16.8) [Significant impact] . 18
5.32 Emissions, deformation of ENCLOSURE or exceeding maximum temperature
(Subclause 13.1.2) [Significant impact] . 18
5.33 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) (Subclause 14.1)
[Moderate impact]. 18
5.34 Design and implementation (Subclause 14.9) [Moderate impact] . 18
5.35 PEMS VALIDATION (Subclause 14.11) [Moderate impact] . 18
5.36 Mechanical strength (Subclause 15.3.1) [Moderate impact] . 18
5.37 Actuating parts of controls of ME EQUIPMENT (Subclause 15.4.6.1 b)
[Moderate impact]. 19
5.38 Limitation of movement (Subclause 15.4.6.2) [Moderate impact] . 19
5.39 Transformers (Subclause 15.5.1.1) [Moderate impact] . 19
5.40 Overload test (Subclause 15.5.1.3) [Moderate impact] . 19
5.41 Dielectric strength (Subclause 15.5.2) [Moderate impact] . 19
5.42 Construction of transformers used to provide separation as required by 8.5
(Subclause 15.5.3) [Significant impact] . 19
5.43 Resistors bridging a MEANS OF PROTECTION (Rationale for Subclause 4.8)
[Moderate impact]. 19
6 Changes impacting particular users of the standard . 20
6.1 ME EQUIPMENT intended to receive power from other equipment (Subclause
7.2.5) [Significant impact] . 20
6.2 ME EQUIPMENT or ACCESSORIES supplied sterile (Subclauses 7.2.17 and
7.9.2.18) [Moderate impact] . 20
6.3 ME EQUIPMENT supplied from an external pressure source (Subclause 7.2.18)
[Moderate impact]. 20
6.4 Mass of MOBILE ME EQUIPMENT (Subclause 7.2.21) [Significant impact] . 20
6.5 General requirements for the instructions for use when the PATIENT is an
OPERATOR (Subclause 7.9.2.1) [Moderate impact] . 20
6.6 ME EQUIPMENT emitting radiation (Subclause 7.9.2.17) [Moderate impact] . 20
6.7 ME EQUIPMENT with DEFIBRILLATION-PROOF APPLIED PARTS (Subclause 8.5.5.1
a)) [Significant impact] . 21
6.8 Energy reduction test for ME EQUIPMENT with DEFIBRILLATION-PROOF APPLIED
PARTS (Subclause 8.5.5.2) [Significant impact] . 21
6.9 Measurement of LEAKAGE CURRENT for sterile equipment (Subclause 8.7.1 b))
[Moderate impact]. 21
6.10 Isolation of PERMANENTLY INSTALLED ME EQUIPMENT from the SUPPLY MAINS
(Subclause 8.11.1) [Moderate impact] . 21
6.11 Protective measures (Subclause 9.2.2.4.4) [Moderate impact] . 21
6.12 Continuous activation (Subclause 9.2.2.5) [Moderate impact] . 22
6.13 Speed of movement(s) (Subclause 9.2.2.6) [Moderate impact] . 22
6.14 Unintended movement (Subclause 9.2.3.1) [Significant impact] . 22
6.15 Overtravel end stops (subclause 9.2.3.2) [Moderate impact] . 22
6.16 Overbalancing from horizontal and vertical forces (Subclause 9.4.2.3 a))
[Significant impact] . 22

– 4 – TR 62348 © IEC:2012(E)
6.17 Movement over a threshold (Subclause 9.4.2.4.3) [Significant impact] . 22
6.18 Instability of MOBILE ME EQUIPMENT in transport position (Subclause
9.4.3.1 c)) [Moderate impact] . 22
6.19 Instability of MOBILE ME EQUIPMENT on an incline surface and in other than
transport position (Subclause 9.4.3.2 a)) [Moderate impact] . 22
6.20 Instability of MOBILE ME EQUIPMENT from lateral forces in other than transport
position (Subclause 9.4.3.2 b)) [Significant impact] . 22
6.21 Acoustic energy and vibration (Subclause 9.6.1) [Moderate impact] . 22
6.22 Audible acoustic energy measurements (Subclause 9.6.2.1) [Moderate
impact] . 23
6.23 Static forces due to loading from persons (Subclause 9.8.3.2) [Moderate
impact] . 23
6.24 Dynamic forces due to loading from persons (Subclause 9.8.3.3) [Moderate
impact] . 23
6.25 ME EQUIPMENT not intended to produce diagnostic or therapeutic X-radiation
(Subclause 10.1.1) [Moderate impact] . 23
6.26 ME EQUIPMENT intended to produce diagnostic or therapeutic X-radiation
(Subclause 10.1.2) [Moderate impact] . 23
6.27 Microwave radiation (Subclause 10.3) [Significant impact] . 23
6.28 Lasers and light-emitting diodes (LEDs) (Subclause 10.4) [Moderate impact] . 23
6.29 APPLIED PARTS not intended to supply heat to a PATIENT (Subclause 11.1.2.2)
[Moderate impact]. 23
6.30 Overflow in ME EQUIPMENT (Subclause 11.6.2) [Moderate impact] . 24
6.31 Diagnostic X-ray equipment (Subclause 12.4.5.2) [Moderate impact] . 24
6.32 PEMS intended to be incorporated into an IT-NETWORK (Subclause 14.13)
[Moderate impact]. 24
6.33 Rough handling test (Subclause 15.3.5) [Moderate impact] . 24
6.34 Application of temperature and overload control devices (Subclause
15.4.2.1 d)) [Moderate impact] . 24
6.35 Excessive current and voltage protection (Subclause 15.4.3.5) [Moderate
impact] . 24
6.36 Entry of liquids for foot-operated control devices (Subclause 15.4.7.3)
[Moderate impact]. 24
6.37 ME SYSTEM power supply (Subclause 16.3) [Moderate impact] . 24
6.38 General measurement conditions for ME SYSTEMS (Subclause 16.6.4.1)
[Moderate impact]. 25
6.39 An MSO combined with a separating transformer (Subclause 16.9.2.1 d))
[Moderate impact]. 25
6.40 Impedance of protective earth connections in ME SYSTEMS (Subclause
16.9.2.2) [Moderate impact] . 25
7 Mapping . 25
7.1 Mapping to the second edition of IEC 60601-1. 25
7.2 Mapping to the second edition of IEC 60601-1-1 . 59
7.3 Mapping to the first edition of IEC 60601-1-4, as amended . 61
7.4 Mapping from IEC 60601-1:2005 + A1:2012 . 64
Bibliography . 101

Table 1 – Amendment 1 changes assessed as having the potential for a significant
impact . 10
Table 2 – Amendment 1 changes assessed as having the potential for a moderate
impact (1 of 2) . 11

TR 62348 © IEC:2012(E) – 5 –
Table 3 – Mapping between the elements of the second edition of IEC 60601-1 as
amended and IEC 60601-1:2005 (1 of 35) . 25
Table 4 – Mapping between the elements of the second edition of IEC 60601-1-1 and
IEC 60601-1:2005 (1 of 2) . 60
Table 5 – Mapping between the elements of the first edition of IEC 60601-1-4 as
amended and IEC 60601-1:2005 (1 of 3). 62
Table 6 – Mapping between the elements of IEC 60601-1:2005 + A1:2012 and other
standards (1 of 36) . 65

– 6 – TR 62348 © IEC:2012(E)
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
ASSESSMENT OF THE IMPACT OF THE MOST SIGNIFICANT CHANGES
IN AMENDMENT 1 TO IEC 60601-1:2005 AND MAPPING OF THE
CLAUSES OF IEC 60601-1:2005 TO THE PREVIOUS EDITION

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
The main task of IEC technical committees is to prepare International Standards. However, a
technical committee may propose the publication of a technical report when it has collected
data of a different kind from that which is normally published as an International Standard, for
example "state of the art".
IEC 62348, which is a technical report, has been prepared by subcommittee 62A; Common
aspects of electrical equipment used in medical practice, of IEC technical committee 62:
Electrical equipment in medical practice.
This second edition cancels and replaces the first edition published in 2006. The second
edition retains the mapping that traces the requirements of IEC 60601-1:2005 and its
Amendment A1:2012 (Edition 3.1) from their source in the documents that relate to
IEC 60601-1:1998 and its amendments (Edition 2.2). See Clause 7. The second edition adds
an assessment of the impact of the most significant changes in Amendment 1:2012 (Clauses
4, 5 and 6).
TR 62348 © IEC:2012(E) – 7 –
The text of this technical report is based on the following documents:
Enquiry draft Report on voting
62A/831/DTR 62A/841/RVC
Full information on the voting for the approval of this technical report can be found in the
report on voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
A bilingual version of this Technical Report may be issued at a later date.

IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
– 8 – TR 62348 © IEC:2012(E)
INTRODUCTION
The first edition of this technical report was created by the Secretariat of SC 62A to assist
users of IEC 60601-1 by providing a tool to trace requirements between IEC 60601-1:2005
and their source in the documents that form the basis of the third edition; principally the
second edition as amended.
At the Auckland meeting in 2008, IEC Technical Committee (TC) 62 approved a project to
develop the 1st amendment to IEC 60601-1:2005 based on the issues outstanding at the time.
The TC approved developing the 1st amendment with a view to addressing outstanding
issues, including but not limited to:
– those issues reported to the Secretariat of IEC Subcommittee (SC) 62A since the
publication of IEC 60601-1:2005;
– the way in which risk management has been introduced into IEC 60601-1:2005; and
– the way the concept of essential performance is used in IEC 60601-1:2005.
Since the Auckland meeting, the Secretariat of SC 62A has received 73 additional issues from
National Committees or other interested parties for a total of 182 identified issues with the
2005 edition. Amendment 1 to IEC 60601-1:2005 is intended to address those issues.
The amendment process has resulted in 496 separate changes. Each change was assessed
by the experts developing the amendment for its potential impact on users of the standard.
Most of the changes are editorial corrections or clarifications and were assessed as having
minimal or no impact on the application of the standard. Others were assessed as having
moderate or significant impact because they represent a technical change, or they impact a
wide range of users, or both.
The second edition of this technical report was prepared by the Secretariat of IEC/SC 62A to
summarize those changes that were assessed during the development process as having a
moderate to significant impact on users of IEC 60601-1.
The tables from the first edition of this technical report were retained in the second edition
because there are counties that have not fully transitioned to the third edition of IEC 60601-1.
Therefore, the original contents of IEC/TR 62348 remain useful in those countries.
Table 6 has been updated to include new subclauses added in Amendment 1:2012
(highlighted in blue).
TR 62348 © IEC:2012(E) – 9 –
ASSESSMENT OF THE IMPACT OF THE MOST SIGNIFICANT CHANGES
IN AMENDMENT 1 TO IEC 60601-1:2005 AND MAPPING OF THE
CLAUSES OF IEC 60601-1:2005 TO THE PREVIOUS EDITION

1 Scope
This technical report provides a tool to assist users of IEC 60601-1:2005 to assess the impact
of the most significant changes in Amendment 1:2012.
This technical report also provides a tool to assist users of IEC 60601-1 to trace requirements
between the third edition and their source in the documents that form the basis of the third
edition; principally the second edition as amended.
This report is intended to be used by:
– those who must align standards based on the second edition of IEC 60601-1 with the third
edition as amended;
– manufacturers of medical electrical equipment or medical electrical systems;
– health care regulatory authorities, test houses and other organizations responsible for
implementing standards for medical electrical equipment and medical electrical systems.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
Amendment 1:2012
3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 +
A1:2012 apply.
4 Assessment of the changes in Amendment 1:2012
Amendment 1 contains 496 separate changes. During the development of the amendment,
each change was assessed by the experts involved for its potential impact on users of the
standard. Most of the changes are editorial corrections or clarifications and were assessed
and having minimal or no impact on the application of the standard.
However, 83 of the changes were assessed as having a moderate to significant impact on the
users of the standard. This assessment is based on the likelihood that some alterations to the
design documentation, testing, the product itself or its accompanying documents will be
required because of the change to a requirement in the amendment.
The changes have been divided into two groups:
– those assessed as having a significant impact (Table 1), and

– 10 – TR 62348 © IEC:2012(E)
– those assessed as having a moderate impact (Table 2).
Because any assessment is somewhat subjective, users of the standard are encouraged to
review the contents of the amendment and determine its impact on the sections that are
relevant to their products.
Table 1 – Amendment 1 changes assessed as
having the potential for a significant impact
Clause of Clause Title Impact Impact Section
IEC 60601-1 many users particular of this
users document
ISK MANAGEMENT PROCESS for ME EQUIPMENT or
R
4.2 X 5.3
ME SYSTEMS
4.3 Essential performance X 5.4
Alternative RISK CONTROL measures or test
4.5 X 5.5
methods for ME EQUIPMENT or ME SYSTEMS
ME EQUIPMENT or ME SYSTEM parts that contact
4.6 X 5.6
the PATIENT
5.1 TYPE TESTS X 5.8
Supply voltages, type of current, nature of
5.5 a) X 5.9
supply, frequency
5.7 Humidity preconditioning treatment X 5.10
Minimum requirements for marking on
7.2.1 X 5.13
ME EQUIPMENT and on interchangeable parts
7.2.4 ACCESSORIES X 5.13
ME EQUIPMENT intended to receive power from
7.2.5 X 6.1
other equipment
7.2.21 Mass of MOBILE ME EQUIPMENT X 6.4
7.9.1 General X 5.17
7.9.3.1 General X 5.20
8.5.5.1 a) Defibrillation protection X 6.7
8.5.5.2 Energy reduction test X 6.8
8.7.3 Allowable values X 5.24
8.8.1 General X 5.26
9.2.3.1 Unintended movement X 6.14
9.4.2.3 a) Instability from horizontal and vertical forces X 6.16
9.4.2.4.3 Movement over a threshold X 6.17
9.4.3.2 b) Instability excluding transport position X 6.20
10.3 Microwave radiation X 6.27
Interruption of the power supply / SUPPLY MAINS
11.8 X 5.31
to ME EQUIPMENT
Emissions, deformation of ENCLOSURE or
13.1.2 X 5.32
exceeding maximum temperature
Construction of transformers used to provide
15.5.3 X 5.42
separation as required by 8.5
TR 62348 © IEC:2012(E) – 11 –
Table 2 – Amendment 1 changes assessed as
having the potential for a moderate impact (1 of 2)
Clause of Clause Title Impact Impact Section
IEC 60601-1 many particular of this
users users document
2 Normative references X 5.1
3.28 EXPECTED SERVICE LIFE X 5.2
4.11 Power input X 5.7
5.9.2.3 Actuating mechanisms X 5.11
7.1.2 Legibility of markings X 5.12
Fuses, THERMAL CUT-OUTS and OVER-CURRENT
7.3.4 X 5.14
RELEASES
7.2.17 Protective packaging X 6.2
7.2.18 External pressure source X 6.3
7.4.2 Control devices X 5.15
7.5 Safety signs X 5.16
7.9.2.1 General X 6.5
7.9.2.15 Environmental protection X 5.18
7.9.2.17 ME EQUIPMENT emitting radiation X 6.6
ME EQUIPMENT and ACCESSORIES supplied
7.9.2.18 X 6.2
sterile
7.9.2.19 Unique version identifier X 5.19
7.9.3.1 General X 5.19
Fundamental rule of protection against electric
8.1 b) X 5.21
shock
8.5.1.2 MEANS OF PATIENT PROTECTION (MOPP) X 5.22
8.5.1.3 MEANS OF OPERATOR PROTECTION (MOOP) X 5.22
8.6.4 Impedance and current-carrying capability X 5.23
8.7.1 b) General requirements X 6.9
Measurement of the EARTH LEAKAGE CURRENT
8.7.4.5 X 5.25
and current in functional earth connection
Minimum CREEPAGE DISTANCES providing MEANS
Table 16 X 5.27
OF OPERATOR PROTECTION
Measurement of CREEPAGE DISTANCES AND AIR
8.9.4 X 5.28
CLEARANCES
8.11.1 Isolation from the SUPPLY MAINS X 6.10
9.2.2.4.4 Other RISK CONTROL measures X 6.11
9.2.2.5 Continuous activation X 6.12
9.2.2.6 Speed of movement(s) X 6.13
9.2.3.2 Overtravel end stops X 6.15
9.4.3.1 c) Instability in transport position X 6.18
9.4.3.2 a) Instability excluding transport position X 6.19
9.6.1 General X 6.21
9.6.2.1 Audible acoustic energy X 6.22
9.8.3.2 Static forces due to loading from persons X 6.23
9.8.3.3 Dynamic forces due to loading from persons X 6.24
ME EQUIPMENT not intended to produce
10.1.1 X 6.25
diagnostic or therapeutic X-radiation

– 12 – TR 62348 © IEC:2012(E)
Table 2 (2 of 2)
Subclause of Clause Title Impact Impact Section
IEC 60601-1 many particular of this
users users document
ME EQUIPMENT intended to produce diagnostic
10.1.2 X 6.26
or therapeutic X-radiation
10.4 Lasers X 6.28
APPLIED PARTS not intended to supply heat to a
11.1.2.2 X 6.29
PATIENT
11.6.2 Overflow in ME EQUIPMENT X 6.30
11.6.3 Spillage on ME EQUIPMENT and ME SYSTEMS X 5.29
Cleaning and disinfection of ME EQUIPMENT and
11.6.6 X 5.30
ME SYSTEMS
12.4.5.2 Diagnostic X-ray equipment X 6.31
14.1 General X 5.33
14.9 Design and implementation X 5.34
14.11 PEMS VALIDATION X 5.35
PEMS intended to be incorporated into an IT-
14.13 X 6.32
NETWORK
15.3.1 General X 5.36
15.3.5 Rough handling test X 6.33
15.4.2.1 d) Application X 6.34
14.4.3.5 Excessive current and voltage protection X 6.35
15.4.6.1 b) Fixing, prevention of maladjustment X 5.37
15.4.6.2 Limitation of movement X 5.38
15.4.7.3 Entry of liquids X 6.36
15.5.1.1 Transformers X 5.39
15.5.1.3 Overload test X 5.40
15.5.2 Dielectric strength X 5.41
16.3 Power supply X 6.37
16.6.4.1 General conditions for ME SYSTEMS X 6.38
Interruption of the power supply to parts of an
16.8 X 5.31
ME SYSTEM
16.9.2.1 d) MULTIPLE SOCKET-OUTLET X 6.39
16.9.2.2 PROTECTIVE EARTH CONNECTIONS in ME SYSTEMS X 6.40
Annex A, 4.8 Components of ME EQUIPMENT X 5.43
The following clauses break down the changes into two groups based on the breadth of their
anticipated impact.
– The first group are those changes that are anticipated to impact many, if not most, of the
users of IEC 60601-1 (Clause 5). An example would be the change to the requirements for
humidity preconditions testing to align with the IEC Certified Testing Laboratories (CTL)
decision to harmonize basic environmental testing conditions for electrical products.
Because of the impact on existing test protocols, the impact was assessed as significant.
– The second group are those changes that should impact only some of the users of
IEC 60601-1 depending on the nature of the equipment to which they are applying the
standard (Clause 6). Examples are the requirements that are applied only to sterile
ME EQUIPMENT or ME EQUIPMENT parts or to MOBILE ME EQUIPMENT.

TR 62348 © IEC:2012(E) – 13 –
5 Changes impacting many users of the standard
5.1 Updated normative references (Clause 2) [Moderate impact]
rd
A number of the standards that are referenced in the 3 edition have been revised or have
been withdrawn and replaced by different documents (e.g., ISO 14971-1:2000 has been
replaced by ISO 14971:2007). These normative references have been updated to reflect
current editions.
5.2 Definition and rationale for EXPECTED SERVICE LIFE (Definition 3.28 and rationale for
4.4) [Moderate impact]
The application of the term EXPECTED SERVICE LIFE has lead to a misunderstanding in some
quarters. To address the misunderstanding, the definition and the rationale have been
expanded.
5.3 Restructuring of RISK MANAGEMENT (Subclause 4.2) [Significant impact]
Subclause 4.2 and its rationale have been significantly modified and expanded. The
RISK MANAGEMENT PROCESS to be employed in
subclause describes in greater detail the
complying with IEC 60601-1.
Subclause 4.2.1 introduces the concepts and the purposes of RISK MANAGEMENT within the
framework of a TYPE TEST or design assurance standard. The subclause includes a reminder
that, "verification of compliance with the RISK MANAGEMENT requirements of this standard can
be accomplished by examination of the RECORDS and other documentation required by this
standard and assessment of the processes cited in this standard and does not require
auditing of the RISK MANAGEMENT PROCESS."
Subclause 4.2.2 is the heart of the subclause and sets out the basic PROCESS requirements,
which remain in compliance with ISO 14971 except for the requirements related to product
and post-production monitoring and periodic reviews of the suitability of the RISK MANAGEMENT
PROCESS.
Subclause 4.2.3 details how the requirements of this standard are to be applied when
evaluating RISK. There are four scenarios described. They are:
a) where IEC 60601-1 or its collateral or particular standards specify requirements
addressing particular HAZARDS or HAZARDOUS SITUATIONS, together with specific
acceptance criteria;
b) where IEC 60601-1 or its collateral or particular standards specify requirements
HAZARDS or HAZARDOUS SITUATIONS but do not provide specific
addressing particular
acceptance criteria;
c) where IEC 60601-1 or its collateral or particular standards identify particular HAZARDS or
HAZARDOUS SITUATIONS that have to be investigated without providing specific technical
requirements; and
d) where HAZARDS or HAZARDOUS SITUATIONS are identified for the particular ME EQUIPMENT or
ME SYSTEM but are not specifically addressed in this IEC 60601-1 or its collateral or
particular standards.
5.4 Application of ESSENTIAL PERFORMANCE (Subclause 4.3) [Significant impact]
The definition of ESSENTIAL PERFORMANCE has been modified from:
performance necessary to achieve freedom from unacceptable risk
rd
in the 3 edition to:
– 14 – TR 62348 © IEC:2012(E)
performance of a clinical function, other than that related to BASIC SAFETY, where loss or
degradation beyond the limits specified by the MANUFACTURER results in an unacceptable
RISK
Subclause 4.3 has been revised and extended to discuss the process the MANUFACTURER can
use to determine the ESSENTIAL PERFORMANCE of ME EQUIPMENT or an ME SYSTEM. The
fundamental goal was not to change the intent of the concept but make it easier to understand
and apply. The definition and the process in 4.3 have been changed to focus on clinical
functions and to try an exclude "performance" related to BASIC SAFETY, e.g., the performance
of insulation. However, remember, regardless of whether we are speaking of BASIC SAFETY or
ESSENTIAL PERFORMANCE, both must be managed.
The rationale for Subclause 4.3 has also been updated.
5.5 Revision to the concept of equivalent safety (Subclause 4.5) [Significant impact]
Recognizing that demonstrating "equivalent safety" can be difficult because many of the
requirements in IEC 60601-1 do not provide sufficient detail regarding the level of RESIDUAL
RISK, this subclause has been modified to allow alternative RISK CONTROL measures or test
methods when it can be demonstrated that that the RESIDUAL RISK resulting from application of
the alternative RISK CONTROL measure or test method remains acceptable and is comparable
to the RESIDUAL RISK that results from applying the requirements of this standard. The
RESIDUAL RISK must satisfy the MANUFACTURER'S RISK acceptance criteria and must be
comparable to that remaining after applying the requirements of the standard. The
MANUFACTURER can demonstrate this by applying:
– scientific data,
– clinical opinion,
– comparative studies, or
– a combination of all three methods.
5.6 Requirements for parts that contact the PATIENT (Subclause 4.6) [Significant impact]
The requirement for such parts was moved from 8.3 d). The RISK MANAGEMENT PROCESS is still
required to assess the parts that fall outside of the definition of APPLIED PARTS that need to be
subject to the requirements for APPLIED PARTS. However, the revision to 4.6 clarifies that
applicable requirements of the collateral standards also apply to these parts.
5.7 Steady-state measurement of input of the ME EQUIPMENT or ME SYSTEM (Subclause
4.11) [Moderate impact]
The provision in the IEC 60601-1:2005 that allowed a supplier certification to be used in place
of the power input measurements required by 4.11 has been replaced by a note that supplier
information can be used to supplement those measurements.
5.8 Simultaneous fault testing (Subclause 5.1) [Moderate impact]
Subclause 5.1 was modified to clarify the requirement that the combination of simultaneous
independent faults that could result in a HAZARDOUS SITUATION be documented in the RISK
MANAGEMENT FILE. The original text of IEC 60601-1:2005 required that the results of the RISK
ANALYSIS be used to determine which combination(s) of simultaneous faults were to be tested.
Since the RISK ANALYSIS is documented in the RISK MANAGEMENT FILE, this is not a new
requirement. When testing is necessary to demonstrate that BASIC SAFETY and ESSENTIAL
PERFORMANCE are maintained under such simultaneous independent faults, the related testing
may be limited to worst case situations. Because of the potential for additional documentation,
this was judged to be of moderate impact.

TR 62348 © IEC:2012(E) – 15 –
5.9 SUPPLY MAINS test characteristics (Subclause 5.5 a)) [Significant impact]
The requirement in 5.5 a) was modified to include other characteristics of the SUPPLY MAINS
other than just supply voltage. As this could involve additional testing, the impact was
assessed as significant.
5.10 Humidity preconditioning (Subclause 5.7) [Significant impact]
The requirements for humidity preconditions testing have been aligned with the IEC Certified
Testing Laboratories (CTL) decision to harmonize basic environmental testing conditions for
electrical products. The minimum time for soaking ME EQUIPMENT enclosures with an IPX0
rating is set to 48 h and for higher IP rating to 168 h. Because of the impact on existing test
protocols, the impact was assessed as significant.
5.11 Actuating mechanisms (Subclause 5.9.2.3) [Moderate impact]
The reference to RISK MANAGEMENT was removed. Any conductive parts of actuating
mechanisms are not to be considered ACCESSIBLE PARTS if removal of the handles, knobs, etc.
requires the use of a TOOL.
5.12 Legibility of markings (Subclause 7.1.2) [Moderate impact]
The test procedure for legibility of markings has been modified so the test is run from the
intended OPERATOR position, or, if that is not specified, a distance of 1 m. Also, the description
of the visual acuity of the observer is extended to include a near-vision specification.
5.13 Identification of ME EQUIPMENT, parts and ACCESSORIES (Subclauses 7.2.1 and 7.2.4)
[Significant impact]
The requirements for Identification of ME EQUIPMENT and ACCESSORIES have been expanded to
include several new items. These include:
– contact information of the MANUFACTURER (for example, this could be a mailing address, a
web site, or a telephone number);
– a serial number or lot or batch identifier; and
– the date of manufacture or use by date, if applicable.
The serial number, lot or batch identifier, and the date of manufacture may be provided in a
human readable code or through automatic identification technology such as barcodes or
RFID.
...