IEC 80601-2-35:2009/COR1:2012
(Corrigendum)Corrigendum 1 - Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use
Corrigendum 1 - Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use
Appareils électromédicaux - Partie 2-35: Exigences particulières pour la sécurité de base et les performances essentielles des dispositifs de réchauffage utilisant des couvertures, des coussins ou des matelas chauffants et destinés au réchauffage des patients en usage médical
General Information
- Status
- Replaced
- Publication Date
- 04-Mar-2012
- Technical Committee
- SC 62D - Particular medical equipment, software, and systems
- Current Stage
- DELPUB - Deleted Publication
- Start Date
- 21-Sep-2020
- Completion Date
- 14-Feb-2026
Relations
- Effective Date
- 05-Sep-2023
- Effective Date
- 05-Sep-2023
- Effective Date
- 05-Sep-2023
Buy Documents
IEC 80601-2-35:2009/COR1:2012 - Corrigendum 1 - Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use Released:3/5/2012
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
IEC 80601-2-35:2009/COR1:2012 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Corrigendum 1 - Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use". This standard covers: Corrigendum 1 - Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use
Corrigendum 1 - Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use
IEC 80601-2-35:2009/COR1:2012 is classified under the following ICS (International Classification for Standards) categories: 11.040.01 - Medical equipment in general; 11.140 - Hospital equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 80601-2-35:2009/COR1:2012 has the following relationships with other standards: It is inter standard links to IEC 60601-2-35:2020, IEC 80601-2-35:2009, IEC 80601-2-35:2009/COR2:2015. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
IEC 80601-2-35:2009/COR1:2012 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
IEC 80601-2-35
(First edition – 2009)
Medical electrical equipment –
Part 2-35: Particular requirements for the basic safety
and essential performance of heating devices using blankets,
pads or mattresses and intended for heating in medical use
CO RRI G E NDUM 1
201.4.3.101 Additional requirements for ESSENTIAL PERFORMANCE
In Table 201.101, replace the reference to "201.12.3.103" by "201.12.3.102" as follows:
Table 201.101 – *Additional ESSENTIAL PERFORMANCE requirements
Requirement Subclause
ESSENTIAL PERFORMANCE requirement 1 201.12.4.104 or generation of a TECHNICAL
ALARM CONDITION in compliance with
201.12.3.102
201.15.4.2.1.101 *Temperature and overload control devices – Additional requirements
for application
In the fifth paragraph, replace the reference to "42 °C" by "41 °C".
201.15.4.101.1 Securing the fixation of heating elements and internal wiring
In the fifth paragraph, replace the reference to "201.15.4.1.102.2" by "201.15.4.101.2".