IEC 62321-13:2026
(Main)Determination of certain substances in electrotechnical products - Part 13: Bisphenol A in plastics by liquid chromatography-diode array detection (LC-DAD), liquid chromatography-mass spectrometry (LC-MS) and liquid chromatography-tandem mass spectrometry (LC-MS/MS)
- Abstract
IEC 62321-13:2026 specifies three techniques for the determination of free Bisphenol A (BPA) in plastics of electrotechnical products.
This document describes the use of liquid chromatography–diode array detector (LC-DAD), liquid chromatography mass spectrometry (LC-MS), liquid chromatography tandem mass spectrometry (LC-MS/MS) with these test methods detailed in Annex A and Annex B.
These test methods have been evaluated for use with PC, PC/ABS, PP matrices containing free BPA between 20 mg/kg to 500 mg/kg as shown in the Pre-IIS 13 results in Annex C and IIS 13 results in Annex D. The use of these methods for BPA concentration ranges of plastics, other than those specified in Annex C and Annex D, has not been evaluated.
WARNING – Persons using this document should be familiar with normal laboratory practice. This document does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to ensure compliance with any national regulatory conditions.
- Status
- Published
- Publication Date
- 22-Jun-2026
- Technical Committee
- TC 111 - Environmental standardization for electrical and electronic products and systems
- Drafting Committee
- WG 3 - Test methods of certain substances
- Current Stage
- PPUB - Publication issued - 60 Publication
- Start Date
- 23-Jun-2026
- Due Date
- 17-Jul-2026
- Completion Date
- 17-Jul-2026
Buy Documents
iec62321-13{ed1.0}en - Determination of certain substances in electrotechnical products - Part 13: Bisphenol A in plastics by liquid chromatography-diode array detection (LC-DAD), liquid chromatography-mass spectrometry (LC-MS) and liquid chromatography-tandem mass spectrometry (LC-MS/MS)
iec62321-13{ed1.0}fr - Détermination de certaines substances dans les produits électrotechniques - Partie 13: Bisphénol A dans les plastiques par chromatographie en phase liquide avec détection a barrettes de diodes (LC-DAD), par chromatographie en phase liquide avec spectrométrie de masse (LC-MS) et par chromatographie en phase liquide avec spectrométrie de masse en tandem (LC-MS/MS)
IEC 62321-13:2026 is an IEC horizontal environmental standard for determining free Bisphenol A (BPA) in plastics from electrotechnical products. It specifies three laboratory techniques - LC-DAD, LC-MS and LC-MS/MS - and is written for people who need a reproducible test method for plastic parts used in electrical and electronic equipment.
What does IEC 62321-13:2026 specify?
IEC 62321-13:2026 specifies the scope, test principle, reagents, apparatus, sampling, analytical procedure, calculation, precision, quality control and test report content for measuring free BPA in plastics. It is a horizontal publication intended mainly for committees and laboratories that need a common environmental test method, but it can also be applied directly to products.
The document is organized into 12 clauses:
- scope
- normative references
- terms, definitions and abbreviations
- principle
- reagents and materials
- apparatus
- sampling
- procedure
- calculation of BPA concentration
- precision data
- quality assurance and quality control
- test report
| Annex | What it covers |
|---|---|
| Annex A | Instrument conditions for LC-DAD, LC-MS and LC-MS/MS |
| Annex B | Example chromatograms and mass spectra under the suggested conditions |
| Annex C | Results of the Preliminary International Interlaboratory Study (Pre-IIS 13) |
| Annex D | Results of the International Interlaboratory Study (IIS 13) |
The scope is limited to free BPA in plastics of electrotechnical products. The evaluation data in the document cover PC, PC/ABS and PP matrices with free BPA in the 20 mg/kg to 500 mg/kg range, so readers should treat those materials and levels as the documented basis for use.
What are the key requirements of IEC 62321-13:2026?
IEC 62321-13:2026 builds the analysis around ultrasonic extraction followed by LC-DAD, LC-MS or LC-MS/MS. In practice, the standard is mainly about controlling contamination, keeping the extraction consistent, and proving that the instrument is calibrated and stable before reporting BPA.
Sampling and extraction
Clause 7 requires sample pieces of about 2 mm x 2 mm, homogenized according to IEC 62321-2, while avoiding contamination and heat. Clause 4 and Clause 7 both matter in practice because BPA can be introduced or formed by contamination or thermal stress, especially when preparing PC or epoxy-like samples.
Clause 8.2.2 sets a fixed extraction sequence: weigh about 1 g of sample, add internal standard, sonicate at 40 °C for 60 min, cool, add mixed solvent to precipitate the matrix, let it stand or centrifuge, then filter through a 0.45 µm PTFE membrane. This gives the lab a repeatable extract and reduces matrix load before LC analysis.
Instrument setup and calibration
Clause 8.3 and Annex A define method-specific LC conditions, including the column type, mobile phases, injection volume, temperature and detector settings. The practical effect is that the lab must configure the system for the chosen technique rather than use a generic LC method.
Clause 8.5 requires at least four calibration solutions and uses peak-area ratios against an internal standard for quantification. The calibration should use the same solvent as the sample when possible, and the linear fit must give a relative standard deviation (RSD) of 15% or less; if not, the calibration is repeated.
Quantification and calculation
Clause 9 states that only detected free BPA is counted, and if none is found the result is reported as “non detected.” Clause 9.2 gives the calculation route from response ratio to final BPA concentration, including the need to account for extraction volume, sample mass and any dilution. That matters because an instrument result alone is not yet the reportable mass fraction in the plastic.
Quality control and method performance
Clause 11 requires a reagent blank with every sequence, a spiked sample at the set frequency, and a continuing calibration check (CCC) after calibration and every ten samples. In practice, these checks show whether the method is clean, the matrix effect is acceptable, and the calibration still holds during the run.
Clause 11 also sets recovery limits for the spike and the CCC, and it requires reanalysis or instrument checks if the CCC or internal standard response drifts outside limits. This prevents a batch from being reported when the system has lost control.
Clause 11.3 requires an MDL study for a new instrument, a major method change or a long idle period, and the calculated MDL for BPA must meet the document’s limit. It also ties LOQ to the MDL, so the laboratory knows the lowest level it can credibly report.
Precision basis
Clause 10 gives repeatability and reproducibility data separately for LC-DAD, LC-MS and LC-MS/MS, based on Pre-IIS 13 and IIS 13 results in Annex C and Annex D. For users, these tables show the performance basis behind the method and help compare the three techniques when selecting a test route.
What terms does IEC 62321-13:2026 define?
- free Bisphénol A (BPA) - Bisphenol A that is not chemically bound in the material and can be extracted and measured.
- Continuing Calibration Check (CCC) - A mid-range calibration check standard run as a sample to confirm that the calibration remains valid during the sequence.
- Certified Reference Material (CRM) - A reference material with an assigned value used to check method performance and accuracy.
- Internal Standard (IS) - A compound added to correct injection and instrumental variation during LC analysis.
- Method Detection Limit (MDL) - The minimum concentration that can be measured and reported with 99% confidence in a given matrix.
- Quality Control (QC) - The blank, spike, calibration check and other control steps used to verify that the method is under control.
- Relative Standard Deviation (RSD) - A measure of calibration scatter used here to judge whether the calibration fit is acceptable.
Who uses IEC 62321-13:2026?
IEC 62321-13:2026 is used by analytical laboratories testing plastics from electrotechnical products for BPA. It is also relevant to quality managers, compliance teams, material suppliers and buyers who need a common method for checking whether a plastic part meets a BPA-related specification.
The standard is especially useful for laboratories working with PC, PC/ABS and PP parts, and for organizations that need routine screening, batch release testing, method verification or interlaboratory comparison. Because it is a horizontal IEC publication, it also supports committees writing product or environmental standards that need a BPA test method.
Which standards are used with IEC 62321-13:2026?
- IEC 62321-1:2013 - Provides the general introduction and the terms used across the IEC 62321 series, and Clause 12 of this part refers back to its report content.
- IEC 62321-2 - Provides dismantling, disassembly and mechanical sample preparation, which supports the sampling step in Clause 7.
- ISO 3696:1987 - Specifies laboratory water quality, which is used for the water required in the reagents and mobile phases.
What does the IEC 62321-13:2026 document contain?
IEC 62321-13:2026 contains practical LC test conditions for BPA in plastics, including sample preparation, extraction, calibration and reporting steps. Annex A gives the operating conditions for each instrument, while Annex B shows example chromatograms and mass spectra that help an analyst confirm peak shape and identification under the suggested conditions.
The document also includes calibration solution tables, formulas for calibration and BPA concentration calculation, and detailed quality-control instructions for blanks, spikes, CCC checks and MDL studies. Annex C and Annex D provide interlaboratory result summaries and precision statistics for LC-DAD, LC-MS and LC-MS/MS, which is the performance evidence behind the method.
Buy Documents
iec62321-13{ed1.0}en - Determination of certain substances in electrotechnical products - Part 13: Bisphenol A in plastics by liquid chromatography-diode array detection (LC-DAD), liquid chromatography-mass spectrometry (LC-MS) and liquid chromatography-tandem mass spectrometry (LC-MS/MS)
iec62321-13{ed1.0}fr - Détermination de certaines substances dans les produits électrotechniques - Partie 13: Bisphénol A dans les plastiques par chromatographie en phase liquide avec détection a barrettes de diodes (LC-DAD), par chromatographie en phase liquide avec spectrométrie de masse (LC-MS) et par chromatographie en phase liquide avec spectrométrie de masse en tandem (LC-MS/MS)
Get Certified
Connect with accredited certification bodies for this standard

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.
Sponsored listings
Frequently Asked Questions
IEC 62321-13:2026 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Determination of certain substances in electrotechnical products - Part 13: Bisphenol A in plastics by liquid chromatography-diode array detection (LC-DAD), liquid chromatography-mass spectrometry (LC-MS) and liquid chromatography-tandem mass spectrometry (LC-MS/MS)". This standard covers: IEC 62321-13:2026 specifies three techniques for the determination of free Bisphenol A (BPA) in plastics of electrotechnical products. This document describes the use of liquid chromatography–diode array detector (LC-DAD), liquid chromatography mass spectrometry (LC-MS), liquid chromatography tandem mass spectrometry (LC-MS/MS) with these test methods detailed in Annex A and Annex B. These test methods have been evaluated for use with PC, PC/ABS, PP matrices containing free BPA between 20 mg/kg to 500 mg/kg as shown in the Pre-IIS 13 results in Annex C and IIS 13 results in Annex D. The use of these methods for BPA concentration ranges of plastics, other than those specified in Annex C and Annex D, has not been evaluated. WARNING – Persons using this document should be familiar with normal laboratory practice. This document does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to ensure compliance with any national regulatory conditions.
IEC 62321-13:2026 specifies three techniques for the determination of free Bisphenol A (BPA) in plastics of electrotechnical products. This document describes the use of liquid chromatography–diode array detector (LC-DAD), liquid chromatography mass spectrometry (LC-MS), liquid chromatography tandem mass spectrometry (LC-MS/MS) with these test methods detailed in Annex A and Annex B. These test methods have been evaluated for use with PC, PC/ABS, PP matrices containing free BPA between 20 mg/kg to 500 mg/kg as shown in the Pre-IIS 13 results in Annex C and IIS 13 results in Annex D. The use of these methods for BPA concentration ranges of plastics, other than those specified in Annex C and Annex D, has not been evaluated. WARNING – Persons using this document should be familiar with normal laboratory practice. This document does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to ensure compliance with any national regulatory conditions.
IEC 62321-13:2026 is classified under the following ICS (International Classification for Standards) categories: 13.020.01 - Environment and environmental protection in general; 43.040.10 - Electrical and electronic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 62321-13:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms, definitions and abbreviated terms
- 3.1 Terms and definitions
- 3.1.1 free bisphenol A
- 3.2 Abbreviated terms
- 4 Principle
- 5 Reagents and materials
- 6 Apparatus
- 7 Sampling
- 8 Procedure
- 8.1 General instructions for the analysis
- 8.2 Sample preparation
- 8.3 Instrumental parameters
- 8.4 Calibrants
- 8.5 Calibration
- 9 Calculation of BPA concentration
- 9.1 General
- 9.2 Calculation
- 10 Precision data
- 10.1 LC-DAD method
- 10.2 LC-MS method
- 10.3 LC-MS/MS method
- 11 Quality assurance and quality control
- 11.1 Performance
- 11.2 Internal control samples and blanks
- 11.3 Method detection limit and reporting limit
- 12 Test report
- Annex A
- Annex B
- Annex C
- Annex D
- Bibliography
INTRODUCTION The widespread use of electrotechnical products has drawn increased attention to their impact on the environment. In many countries this has resulted in the adaptation of regulations affecting waste, substances and energy use of electrotechnical products. The use of certain substances (e.g. lead (Pb), cadmium (Cd), mercury (Hg) and polybrominated diphenyl ethers (PBDEs)) and phthalates in electronical products is a source of concern in current and proposed regional legislation.
The purpose of the IEC 62321 series is therefore to provide test methods that will allow the electrotechnical industry to determine the levels of certain substances of concern in electrotechnical products on a consistent global basis.
This first edition of IEC 62321-13 introduces a new subject covering Bisphenol A in the IEC 62321 series.
1 Scope
This document specifies three techniques for the determination of free Bisphenol A (BPA) in plastics of electrotechnical products.
This document describes the use of liquid chromatography–diode array detector (LC-DAD), liquid chromatography mass spectrometry (LC-MS), liquid chromatography tandem mass spectrometry (LC-MS/MS) with these test methods detailed in Annex A and Annex B. These test methods have been evaluated for use with PC, PC/ABS, PP matrices containing free BPA between 20 mg/kg to 500 mg/kg as shown in the Pre-IIS 13 results in Annex C and IIS 13 results in Annex D [1], [2]. The use of these methods for BPA concentration ranges of plastics, other than those specified in Annex C and Annex D, has not been evaluated. This document is a basic environment horizontal publication focusing on test methods and is primarily intended for use by committees in the preparation of publications within the area of environment in accordance with the principles laid down in IEC Guide 123. Wherever applicable, it is the responsibility of committees to make use of environment basic publications in the preparation of their environment group and product publications. Committees can apply this document directly to products when they do not develop a product publication in the area of environment.
WARNING – Persons using this document should be familiar with normal laboratory practice. This document does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to ensure compliance with any national regulatory conditions.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
IEC 62321-1:2013, Determination of certain substances in electrotechnical products - Part 1:
Introduction and overview IEC 62321-2, Determination of certain substances in electrotechnical products - Part 2:
Disassembly, disjointment and mechanical sample preparation ISO 3696:1987, Water for analytical laboratory use - Specification and test methods
3 Terms, definitions and abbreviated terms
3.1 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 62321-1 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- IEC Electropedia: available at https://www.electropedia.org/
- ISO Online browsing platform: available at https://www.iso.org/obp
3.1.1 free bisphenol A
chemically unbound bisphenol A (BPA), as present in material
3.2 Abbreviated terms
For the purpose of this document, the following abbreviated terms apply. ABS acrylonitrile-butadiene-styrene BPA bisphenol A BPC bisphenol C CCC continuing calibration check standard CRM certified reference material IS internal standard LC liquid chromatography LC-DAD liquid chromatography-diode array detector LC-MS liquid chromatography-mass spectrometry LC-MS/MS liquid chromatography-tandem mass spectrometry MDL method detection limit PC polycarbonate PC/ABS polycarbonate/Acrylonitrile-butadiene-styrene PP polypropylene PTFE polytetrafluoroethylene QC quality control RSD relative standard deviation RM raw material
4 Principle
Free BPA is determined using ultrasonic extraction followed by high-performance liquid chromatography separation and tandem mass spectrometry and liquid chromatography-diode array detection. When preparing a test specimen (see Clause 7) of polycarbonate (PC) or similar type of epoxy resins, thermal stress of the specimen should be avoided, since heat can affect the test results.
Free BPA is determined using ultrasonic extraction followed by high-performance liquid chromatography separation with either mass-spectrometry (LC-MS), or tandem mass spectrometry (LC-MS/MS), or diode array detector (LC-DAD).
5 Reagents and materials
The reagent chemicals used for the tests described in this document shall have a minimum purity as described from a) to i) in Clause 5. This is intended to ensure the reagents are free from contamination and prevent proliferation of blank values. There is a risk of elevated blank values due to the ubiquitous nature of BPA (CAS #80-05-7). Laboratories should assure to verify the blank value of the test method especially in case the calibration curve shows an offset. These are the reagents used in the tests described in this document:
a) methanol (purity LC grade or higher); b) tetrahydrofuran (purity LC grade or higher);
c) water (purified water, for quality of the mobile phase ingredients and solution or according to ISO 3696:1987, Grade 1, applies); d) acetonitrile (Purity LC grade or higher); e) mixed solvent solution (1:1 Methanol and water): add 500 ml methanol and 500 ml water to 1 l beaker and mix; f) 10 mM ammonium acetate buffer solution: add 0,771 g ammonium acetate in a 1 l beaker (see Clause 6, item h)), dissolve with a little purified water (see Clause 5, item c)), then add 900 ml purified water. Transfer to 1 l volumetric flask and add 10 ml acetonitrile and dilute with purified water to the marked line; g) calibrants, BPA (purity of greater than a mass fraction of 99 %); h) Internal standard (IS) used to correct for injection (e.g. BPA-d , BPA-d , BPC); 16 8 i) 0,04 % Phosphoric acid buffer solution: add about 0,471 ml 85 % Phosphoric acid in a 1 l volumetric flask (see Clause 6, item b)) and dilute with purified water to the marked line. To prevent exothermal reaction, fill volumetric flask to half of water and then add acid. Then fill to the mark with purified water.
6 Apparatus
The following items shall be used for the analysis:
a) analytical balance having an accuracy of 0,000 1 g; b) 10 ml, 100 ml, 1 l volumetric flasks (glass);
c) ultrasonic bath with a minimum power of 200 W and a bath area of 706 cm , corresponding
to 0,28 W/cm , without a basket and with an internal or external thermostat; d) Pasteur pipette (glass); e) 1,5 ml sample vials with 100 µl glass insert and a screw cap with polytetrafluoroethylene (PTFE) gasket or, depending on the analytical system, a comparable sample receptacle; f) 0,45 µm PTFE membrane filter; g) centrifuge with maximum rotation speed over 3 000 r/min; h) 1 l beaker (glass); i) minimum 40 ml vial (glass).
7 Sampling
The following procedure shall be used as described in IEC 62321-2. Cut samples approximately to a size of 2 mm × 2 mm and homogenise. Contact with possible sources of BPA-contamination shall be avoided during sampling.
NOTE Heat can cause the formation of additional free BPA. It is important to avoid using devices that generate heat (e.g. circular saw).
8 Procedure
8.1 General instructions for the analysis
The following general instructions shall be followed.
a) In order to reduce blank values, ensure the cleanliness of all glass and analytical equipment. b) If the amount of BPA in the sample is considerably above the calibration range, it is necessary to carry out the analysis using an adjusted sample size or by repeating the analysis using an extract that has been appropriately diluted prior to internal standard addition.
8.2 Sample preparation
8.2.1 Stock solution
The following stock solutions shall be prepared.
a) Standard solution for calibration: prepare a standard solution containing 1 000 µg/ml Bisphenol A in Tetrahydrofuran (see Clause 5, item b)).
b) Internal standard for LC-MS and LC-MS/MS: 4 µg/ml in Tetrahydrofuran (see Clause 5, item b)) (e.g. BPC, BPA-d , BPA-d ).
16 8 c) Internal standard for LC-DAD: 40 µg/ml in Tetrahydrofuran (see Clause 5, item b)) (e.g. BPC).
d) Dilution solution: 10 ml of internal standard (8.2.1, item b) or 8.2.1, item c)) in 30 ml of mixed solvent (see Clause 5, item e)).
8.2.2 Extraction
The following steps shall be followed for sample extraction.
a) Weigh approximately 1 g of the sample and transfer it into a tared vial. Record the mass to the nearest 0,000 1 g. Lower sample weights can be tested.
b) Add 10 ml of IS solution (8.2.1, item b) or 8.2.1, item c)) to the vial. c) Tightly cap the sample vial. Place it in an ultrasonic bath (see Clause 6, item c)) and sonicate at about 40 °C for (60 ± 1) min.
d) After the sample has extracted, allow the vial to cool to ambient temperature. e) Accurately add 30 ml of mixed solvent (see Clause 5, item e)) dropwise into the vial to precipitate the sample matrix.
f) Allow the resulting extracted solution to stand at room temperature for 30 min (when necessary use centrifuge (see Clause 6, item g)) for 10 min to precipitate the sample matrix down with 3 000 r/min).
g) Filter the sample solution through a 0,45 µm PTFE membrane filter (see Clause 6, item f)) and transfer into a vial for LC analysis.
8.3 Instrumental parameters
8.3.1 General
Different conditions can be necessary to optimize a specific LC system to achieve effective determination of free BPA and meet the QC and MDL requirements. The following parameters (8.3.2 and 8.3.3) are suitable and are provided as an example. NOTE Additional detailed LC-DAD instrument parameters are described in Table A.1. Additional detailed LC-MS and LC-MS/MS parameters are described in Table A.2 and Table A.3.
8.3.2 Stationary phase
The following steps shall be followed when selecting a suitable column for analysis. a) C Column, 150 mm × 2,1 mm ID, 5 μm or equivalent;
b) C Column, 250 mm × 4,6 mm ID, 5 μm or equivalent.
8.3.3 Liquid (mobile) phase
Water (see Clause 5, item c)) and acetonitrile (see Clause 5, item d)) are used as liquid phases.
8.4 Calibrants
Table 1 shows BPA reference materials suitable for this analysis. An example of chromatogram of bisphenol A standard solution in LC-DAD is given in Figure B.1. Chromatogram of LC-MS and mass analysis spectrum of bisphenol A standard is given in Figure B.2 and Figure B.3. Chromatogram of LC-MS/MS and mass analysis spectrum of bisphenol A standard is given in Figure B.4 and Figure B.5.
Table 1 – CAS number of reference material and internal standards Compound name CAS number BPA 80-05-7 BPA-d 96210-87-6
BPA-d 92739-58-7
BPC 79-97-0
8.5 Calibration
8.5.1 General
Wherever possible, the solvent used for the sample and standard solutions are to be the same to avoid any potential solvent effects. A calibration curve shall be developed for quantitative analysis. At least four calibration solutions shall be prepared. Quantification is made on the basis of the measurement of the peak areas. The linear regression fit of the calibration curve is required to have a relative standard deviation (RSD) of less than or equal to 15 % of the linear calibration function. If the RSD of 15 % is exceeded, the calibration shall be repeated.
8.5.2 Standard solutions
A standard stock solution of BPA (reference material, Table 1) is prepared by dissolving appropriate amounts of BPA (see 8.2.1, item a)) using the appropriate dilution solution (see 8.2.1, item d)) to achieve a final BPA standard stock solution concentration of 10 µg/ml. This document stock solution can be diluted using the appropriate dilution solution (8.2.1, item d)) to achieve final calibration solutions shown in Table 2 and Table 3. This procedure ensures a constant IS concentration of 1 µg/ml (LC-MS or LC-MS/MS) or 10 µg/ml (LC-DAD), respectively). Table 2 – Preparation of low concentrations of calibration standard solutions of BPA Standard solution No. Final concentration of BPA Final concentration of internal standard (µg/ml) (µg/ml) 1 0,01 1,0 2 0,02 1,0 3 0,05 1,0 4 0,1 1,0 5 0,2 1,0
Table 3 – Preparation of high concentrations of calibration standard solutions of BPA Standard solution No. Final concentration of BPA Final concentration of internal standard (µg/ml) (µg/ml) 1 0,1 10 2 0,2 10 3 0,5 10 4 1,0 10 5 2,0 10
...
Contents
- 1 Domaine d'application
- 2 Références normatives
- 3 Termes, définitions et abréviations
- 3.1 Termes et définitions
- 3.1.1 bisphénol A libre
- 3.2 Abréviations
- 4 Principe
- 5 Réactifs et matériaux
- 6 Appareils
- 7 Échantillonnage
- 8 Procédure
- 8.1 Instructions générales pour l'analyse
- 8.2 Préparation de l'échantillon
- 8.3 Paramètres de l'instrument
- 8.4 Étalons
- 8.5 Étalonnage
- 9 Calcul de la concentration de BPA
- 9.1 Généralités
- 9.2 Calcul
- 10 Données de précision
- 10.1 Méthode LC-DAD
- 10.2 Méthode LC-MS
- 10.3 Méthode LC-MS/MS
- 11 Assurance qualité et contrôle de la qualité
- 11.1 Performances
- 11.2 Échantillons et témoins de contrôle internes
- 11.3 Limite de détection de la méthode et limite de déclaration
- 12 Rapport d'essai
- Annexe A
- Annexe B
- Annexe C
- Annexe D
- Bibliographie
INTRODUCTION L'utilisation largement répandue des produits électrotechniques suscite une attention accrue concernant leur incidence sur l'environnement. Dans de nombreux pays, ceci a conduit à l'adaptation de réglementations relatives aux déchets, aux substances et à la consommation d'énergie des produits électrotechniques.
L'utilisation de certaines substances (comme le plomb (Pb), le cadmium (Cd), le mercure (Hg) et les diphényléthers polybromés (PBDE, Polybrominated Diphenyl Ethers) et des phtalates dans les produits électroniques est une source de préoccupation dans la législation régionale en vigueur et en cours d'élaboration.
L'objet de la série IEC 62321 est par conséquent de fournir, à une échelle mondiale et de manière cohérente, des méthodes d'essai qui permettent à l'industrie électrotechnique de déterminer les niveaux de certaines substances, sources de préoccupation, dans les produits électrotechniques.
Cette première édition de l'IEC 62321-13 introduit un nouveau sujet qui couvre le bisphénol A dans la série IEC 62321.
1 Domaine d'application
Le présent document spécifie trois techniques de détermination du bisphénol A (BPA) libre présent dans des plastiques des produits électrotechniques.
Le présent document décrit l'utilisation de la chromatographie en phase liquide avec détecteur à barrettes de diodes (LC-DAD), la chromatographie en phase liquide avec spectrométrie de masse (LC-MS) et la chromatographie en phase liquide avec spectrométrie de masse en tandem (LC-MS/MS) dans le cadre des méthodes d'essai décrites en détail à l'Annexe A et à l'Annexe B.
Ces méthodes d'essai ont été évaluées pour une utilisation avec des matrices PC, PC/ABS et PP contenant entre 20 mg/kg et 500 mg/kg de BPA libre, comme cela est indiqué dans les résultats de l'Annexe C relative au Pre-IIS 13 et de l'Annexe D relative à l'IIS 13 [1], [2]. L'utilisation de ces méthodes pour des plages de concentration de BPA dans les plastiques, autres que celles spécifiées à l'Annexe C et à l'Annexe D, n'a pas été évaluée. Le présent document est une publication horizontale environnementale fondamentale qui met l'accent sur les méthodes d'essai et est principalement destiné à être utilisé par les comités pour l'élaboration de publications relevant du domaine de l'environnement, conformément aux principes énoncés dans le Guide 123 de l'IEC. Le cas échéant, il incombe aux comités d'utiliser les publications environnementales fondamentales dans le cadre de l'élaboration de leurs publications environnementales et leurs publications de produits. Les comités peuvent appliquer le présent document directement aux produits lorsqu'ils n'élaborent pas de publication de produit dans le domaine de l'environnement.
AVERTISSEMENT – Il convient que les personnes qui utilisent le présent document aient une bonne connaissance des pratiques normales de laboratoire. Le présent document ne prétend pas traiter tous les problèmes de sécurité éventuels associés à son utilisation. Il incombe à l'utilisateur de mettre en place les pratiques adéquates en matière de sécurité et de santé, mais aussi d'assurer la conformité aux conditions réglementaires nationales.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu'ils constituent, pour tout ou partie de leur contenu, des exigences du présent document. Pour les références datées, seule l'édition citée s'applique. Pour les références non datées, la dernière édition du document de référence s'applique (y compris les éventuels amendements).
IEC 62321-1:2013, Détermination de certaines substances dans les produits électrotechniques - Partie 1: Introduction et présentation IEC 62321-2, Détermination de certaines substances dans les produits électrotechniques - Partie 2: Démontage, défabrication et préparation mécanique de l'échantillon ISO 3696:1987, Eau pour laboratoire à usage analytique - Spécification et méthodes d'essai
3 Termes, définitions et abréviations
3.1 Termes et définitions
Pour les besoins du présent document, les termes et les définitions de l'IEC 62321-1 ainsi que les suivants s'appliquent.
L'ISO et l'IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation, consultables aux adresses suivantes:
- IEC Electropedia: disponible à l'adresse https://www.electropedia.org/
- ISO Online browsing platform: disponible à l'adresse https://www.iso.org/obp
3.1.1 bisphénol A libre
bisphénol A (BPA) non lié chimiquement, qui est présent dans les matériaux
3.2 Abréviations
Pour les besoins du présent document, les abréviations suivantes s'appliquent. ABS acrylonitrile butadiène styrène BPA bisphénol A BPC bisphénol C CCC (Continuing Calibration Check étalon de vérification continue de standard) l'étalonnage CRM (Certified Reference Material) matériau de référence certifié IS (Internal Standard) étalon interne LC (Liquid Chromatography) chromatographie en phase liquide LC-DAD (Liquid Chromatography-Diode chromatographie en phase liquide avec Array Detector) détecteur à barrettes de diodes LC-MS (Liquid Chromatography-Mass chromatographie en phase liquide- Spectrometry) spectrométrie de masse LC-MS/MS (Liquid Chromatography-Tandem chromatographie en phase liquide avec Mass Spectrometry) spectrométrie de masse en tandem MDL (Method Detection Limit) limite de détection de la méthode PC polycarbonate PC/ABS polycarbonate/acrylonitrile butadiène styrène PP polypropylène PTFE polytétrafluoroéthylène QC (Quality Control) contrôle de la qualité RSD (Relative Standard Deviation) écart-type relatif RM (Raw Material) matière première
4 Principe
Le BPA libre est déterminé en appliquant une extraction par ultrasons, suivie d'une séparation par chromatographie en phase liquide à haute performance couplée à la spectrométrie de masse simple (LC-MS), ou à la spectrométrie de masse en tandem (LC-MS/MS), ou à un détecteur à barrettes de diodes (LC-DAD). Lors de la préparation d'une éprouvette (voir
Article 7) de polycarbonate (PC) ou similairement d'un type de résine époxy, il convient d'éviter l'application de contraintes thermiques sur l'éprouvette, car la chaleur peut influencer le résultat.
5 Réactifs et matériaux
Les réactifs chimiques utilisés pour les essais décrits dans le présent document doivent avoir une pureté minimale, comme cela est décrit dans les points a) à i) de l'Article 5. L'objectif est de s'assurer que les réactifs ne présentent aucune contamination et d'empêcher la prolifération de valeurs de blancs témoins élevées. Il existe un risque de valeurs témoins élevées en raison du caractère omniprésent du BPA (numéro CAS 80-05-7). Il convient que les laboratoires vérifient la valeur témoin de la méthode d'essai, en particulier si la courbe d'étalonnage présente un décalage.
Les réactifs suivants sont utilisés pour les essais décrits dans le présent document:
a) méthanol (d'une pureté de qualité LC ou supérieure); b) tétrahydrofurane (d'une pureté de qualité LC ou supérieure); c) eau (eau purifiée, pour la qualité des ingrédients et de la solution de la phase mobile, ou conformément à l'ISO 3696:1987, la Qualité 1 s'applique); d) acétonitrile (d'une pureté de qualité LC ou supérieure); e) solution de solvant mélangé (mélange méthanol/eau à 1:1): ajouter 500 ml de méthanol et 500 ml d'eau dans un bécher de 1 l, puis mélanger; f) solution tampon contenant 10 mM d'acétate d'ammonium: ajouter 0,771 g d'acétate d'ammonium dans un bécher de 1 l (voir Article 6, point h)), dissoudre avec un peu d'eau purifiée (voir Article 5, point c)), puis ajouter 900 ml d'eau purifiée. Transvaser dans un flacon volumétrique de 1 l et ajouter 10 ml d'acétonitrile, puis diluer avec de l'eau purifiée jusqu'à la graduation; g) étalons; BPA (d'une pureté supérieure à une fraction volumique de 99 %); h) étalon interne (IS) utilisé pour corriger les erreurs d'injection (par exemple, BPA-d ,
BPA-d , BPC);
i) solution tampon contenant 0,04 % d'acide phosphorique: ajouter environ 0,471 ml d'acide phosphorique à 85 % dans un flacon volumétrique de 1 l (voir Article 6, point b)) et diluer avec de l'eau purifiée jusqu'à la graduation.
Pour éviter une réaction exothermique, remplir le flacon volumétrique à moitié d'eau et ajouter l'acide. Compléter avec de l'eau purifiée jusqu'à la graduation.
6 Appareils
Les appareils suivants doivent être utilisés pour l'analyse:
a) balance d'analyse avec une exactitude de 0,000 1 g; b) flacons volumétriques de 10 ml, 100 ml et 1 l (en verre);
c) bain à ultrasons; avec une puissance minimale de 200 W et une surface de bain de 706 cm ,
soit une densité de puissance de 0,28 W/cm , sans panier et avec un thermostat interne ou externe; d) pipette Pasteur (en verre); e) fioles d'échantillon de 1,5 ml avec insert en verre de 100 µl et bouchon vissé avec joint en polytétrafluoroéthylène (PTFE) ou, en fonction du système analytique, un récipient à échantillon comparable; f) filtre à membrane en PTFE de 0,45 µm; g) centrifugeuse, avec une vitesse de rotation supérieure à 3 000 r/min;
h) bécher de 1 l (en verre); i) fiole de 40 ml au minimum (en verre).
7 Échantillonnage
La procédure suivante doit être utilisée, comme cela est décrit dans l'IEC 62321-2. Découper des échantillons d'environ 2 mm × 2 mm, puis homogénéiser. Tout contact avec des sources potentielles de contamination par le BPA doit être évité pendant l'échantillonnage. NOTE La chaleur peut provoquer la formation de BPA libre supplémentaire. Il est important d'éviter d'utiliser des dispositifs qui produisent de la chaleur (scie circulaire, par exemple).
8 Procédure
8.1 Instructions générales pour l'analyse
Les instructions générales suivantes doivent être suivies.
a) Pour réduire la valeur des blancs témoins, s'assurer de la propreté de tous les équipements en verre et matériels d'analyse.
b) Si la quantité de BPA dans l'échantillon se situe largement au-dessus de la plage d'étalonnage, il est nécessaire d'effectuer l'analyse en adaptant la taille de l'échantillon ou en répétant l'analyse à l'aide d'un extrait dilué de manière appropriée avant d'ajouter l'étalon interne.
8.2 Préparation de l'échantillon
8.2.1 Solution mère
Les solutions mères suivantes doivent être préparées.
a) Solution d'étalonnage: préparer une solution étalon contenant 1 000 μg/ml de bisphénol A dans du tétrahydrofurane (voir Article 5, point b)).
b) Étalon interne pour les méthodes LC-MS et LC-MS/MS: 4 µg/ml dans du tétrahydrofurane (voir Article 5, point b)) (BPC, BPA-d , BPA-d , par exemple). 16 8 c) Étalon interne pour la méthode LC-DAD: 40 µg/ml dans du tétrahydrofurane (voir Article 5, point b)) (BPC, par exemple).
d) Solution de dilution: 10 ml d'étalon interne (8.2.1, point b) ou 8.2.1, point c)) dans 30 ml de solvant mélangé (voir Article 5, point e)).
8.2.2 Extraction
Pour l'extraction de l'échantillon, les étapes suivantes doivent être suivies. a) Peser environ 1 g d'échantillon et le transvaser dans une fiole préalablement tarée. Enregistrer la masse à 0,000 1 g près. Les poids d'échantillon inférieurs peuvent être soumis à l'essai.
b) Ajouter 10 ml de solution IS (8.2.1, point b) ou 8.2.1, point c)) dans la fiole. c) Fermer la fiole d'échantillon en serrant bien le bouchon. Placer la fiole dans un bain à ultrasons réglé à 40 °C (voir Article 6, point c)), puis appliquer les ultrasons pendant (60 ± 1) min.
d) Après l'extraction de l'échantillon, laisser refroidir la fiole jusqu'à atteindre la température ambiante.
e) Ajouter exactement 30 ml de solvant mélangé (voir Article 5, point e)) en goutte à goutte dans la fiole pour précipiter la matrice d'échantillon.
f) Laisser reposer la solution extraite à température ambiante pendant 30 min. (Si nécessaire, utiliser une centrifugeuse (voir Article 6, point g)) pendant 10 min pour précipiter la matrice d'échantillon à une vitesse de rotation de 3 000 r/min.) g) Filtrer la solution d'échantillon à travers un filtre à membrane en PTFE de 0,45 μm (voir Article 6, point f)) et la transvaser dans une fiole pour l'analyse LC.
8.3 Paramètres de l'instrument
8.3.1 Généralités
Différentes conditions peuvent être nécessaires pour optimiser un système LC spécifique afin de permettre une détermination efficace du BPA libre et de respecter les exigences en matière de contrôle de la qualité et de limite de détection de la méthode. Les paramètres suivants (8.3.2 et 8.3.3) sont appropriés et sont fournis à titre d'exemple.
NOTE D'autres paramètres détaillés de la méthode LC-DAD sont décrits dans le Tableau A.1. D'autres paramètres détaillés des méthodes LC-MS et LC-MS/MS sont décrits dans le Tableau A.2 et le Tableau A.3.
8.3.2 Phase stationnaire
Les étapes suivantes doivent être exécutées pour choisir la colonne appropriée pour l'analyse. a) colonne C , ID 150 mm × 2,1 mm, 5 μm ou équivalente;
b) colonne C , ID 250 mm × 4,6 mm, 5 μm ou équivalente.
8.3.3 Phase liquide (mobile)
L'eau (voir Article 5, point c)) et l'acétonitrile (voir Article 5, point d)) sont utilisés comme phases liquides.
8.4 Étalons
Le Tableau 1 répertorie des matériaux de référence BPA adaptés pour cette analyse. La Figure B.1 fournit un exemple de chromatogramme de la solution étalon de bisphénol A (LC-DAD). La Figure B.2 et la Figure B.3 représentent le chromatogramme de l'analyse LC-MS et le spectre d'analyse de masse de la solution étalon de bisphénol A. La Figure B.4 et la Figure B.5 représentent le chromatogramme de l'analyse LC-MS/MS et le spectre d'analyse de masse de la solution étalon de bisphénol A.
Tableau 1 – Numéros CAS des matériaux de référence et étalons internes Nom du composé Numéro CAS BPA 80-05-7 BPA-d 96210-87-6
BPA-d 92739-58-7
BPC 79-97-0
8.5 Étalonnage
8.5.1 Généralités
Dans la mesure du possible, le solvant utilisé pour l'échantillon doit être le même que celui utilisé pour la solution étalon afin d'éviter tout effet potentiel du solvant. Une courbe d'étalonnage doit être réalisée pour une analyse quantitative. Au moins quatre solutions d'étalonnage doivent être préparées. La quantification est effectuée en se fondant sur la mesure des surfaces des pics. L'ajustement par régression linéaire de la courbe d'étalonnage est exigé pour obtenir un écart-type relatif (RSD) inférieur ou égal à 15 % de la fonction d'étalonnage linéaire. Si le RSD de 15 % est dépassé, l'étalonnage doit être répété.
8.5.2 Solutions étalons
Préparer une solution mère étalon de BPA (matériau de référence du Tableau 1) en dissolvant des quantités appropriées de BPA (voir 8.2.1, point a)) avec la solution de dilution appropriée (voir 8.2.1, point d)) afin d'obtenir une concentration finale de 10 μg/ml. Cette solution mère étalon peut être diluée avec la solution de dilution appropriée (8.2.1, point d)) afin d'obtenir les concentrations finales indiquées dans les Tableaux 2 et 3. Cette procédure garantit une concentration de l'étalon interne constante de 1 μg/ml (LC-MS ou LC-MS/MS) ou de 10 μg/ml (LC-DAD), respectivement.
...