IEC 62321-3-1:2013
(Main)Determination of certain substances in electrotechnical products - Part 3-1: Screening - Lead, mercury, cadmium, total chromium and total bromine by X-ray fluorescence spectrometry
Determination of certain substances in electrotechnical products - Part 3-1: Screening - Lead, mercury, cadmium, total chromium and total bromine by X-ray fluorescence spectrometry
IEC 62321-3-1:2013 describes the screening analysis of five substances, specifically lead (Pb), mercury (Hg), cadmium (Cd), total chromium (Cr) and total bromine (Br) in uniform materials found in electrotechnical products, using the analytical technique of X-ray fluorescence (XRF) spectrometry.
IEC 62321-3-1:2013 has the status of a horizontal standard in accordance with IEC Guide 108.
Détermination de certaines substances dans les produits électrotechniques - Partie 3-1: Méthodes d'essai - Plomb, du mercure, du cadmium, du chrome total et du brome total par laspectrométrie par fluorescence X
L'IEC 62321-3-1:2013 décrit des procédures d'analyse de détection de cinq substances, spécifiquement le plomb (Pb), le mercure (Hg), le cadmium (Cd), le chrome total (Cr) et le brome total (Br) dans des matériaux uniformes utilisés dans les produits électrotechniques en utilisant la technique d'analyse de spectrométrie par fluorescence X (XRF).
L'IEC 62321-3-1:2013 a le statut de norme horizontale conformément au Guide IEC 108.
General Information
Relations
Standards Content (Sample)
IEC 62321-3-1 ®
Edition 1.0 2013-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
HORIZONTAL STANDARD
NORME HORIZONTALE
Determination of certain substances in electrotechnical products –
Part 3-1: Screening – Lead, mercury, cadmium, total chromium and total bromine
by X-ray fluorescence spectrometry
Détermination de certaines substances dans les produits électrotechniques –
Partie 3-1: Méthodes d'essai – Plomb, du mercure, du cadmium, du chrome total
et du brome total par la spectrométrie par fluorescence X
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les
microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
Useful links:
IEC publications search - www.iec.ch/searchpub Electropedia - www.electropedia.org
The advanced search enables you to find IEC publications The world's leading online dictionary of electronic and
by a variety of criteria (reference number, text, technical electrical terms containing more than 30 000 terms and
committee,…). definitions in English and French, with equivalent terms in
It also gives information on projects, replaced and additional languages. Also known as the International
withdrawn publications. Electrotechnical Vocabulary (IEV) on-line.
IEC Just Published - webstore.iec.ch/justpublished Customer Service Centre - webstore.iec.ch/csc
Stay up to date on all new IEC publications. Just Published If you wish to give us your feedback on this publication
details all new publications released. Available on-line and or need further assistance, please contact the
also once a month by email. Customer Service Centre: csc@iec.ch.
A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.
Liens utiles:
Recherche de publications CEI - www.iec.ch/searchpub Electropedia - www.electropedia.org
La recherche avancée vous permet de trouver des Le premier dictionnaire en ligne au monde de termes
publications CEI en utilisant différents critères (numéro de électroniques et électriques. Il contient plus de 30 000
référence, texte, comité d’études,…). termes et définitions en anglais et en français, ainsi que
Elle donne aussi des informations sur les projets et les les termes équivalents dans les langues additionnelles.
publications remplacées ou retirées. Egalement appelé Vocabulaire Electrotechnique
International (VEI) en ligne.
Just Published CEI - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications de la CEI.
Just Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur
Disponible en ligne et aussi une fois par mois par email. cette publication ou si vous avez des questions
contactez-nous: csc@iec.ch.
IEC 62321-3-1 ®
Edition 1.0 2013-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
HORIZONTAL STANDARD
NORME HORIZONTALE
Determination of certain substances in electrotechnical products –
Part 3-1: Screening – Lead, mercury, cadmium, total chromium and total bromine
by X-ray fluorescence spectrometry
Détermination de certaines substances dans les produits électrotechniques –
Partie 3-1: Méthodes d'essai – Plomb, du mercure, du cadmium, du chrome total
et du brome total par la spectrométrie par fluorescence X
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 13.020; 43.040.10 ISBN 978-2-83220-839-7
– 2 – 62321-3-1 © IEC:2013
CONTENTS
FOREWORD . 5
INTRODUCTION . 7
1 Scope . 8
2 Normative references . 10
3 Terms, definitions and abbreviations . 10
4 Principle . 10
4.1 Overview . 10
4.2 Principle of test . 11
4.3 Explanatory comments . 11
5 Apparatus, equipment and materials . 12
5.1 XRF spectrometer . 12
5.2 Materials and tools . 12
6 Reagents . 12
7 Sampling . 12
7.1 General . 12
7.2 Non-destructive approach . 12
7.3 Destructive approach. 12
8 Test procedure . 13
8.1 General . 13
8.2 Preparation of the spectrometer . 13
8.3 Test portion . 14
8.4 Verification of spectrometer performance . 14
8.5 Tests . 15
8.6 Calibration . 15
9 Calculations . 16
10 Precision . 17
10.1 General . 17
10.2 Lead . 17
10.3 Mercury . 17
10.4 Cadmium . 17
10.5 Chromium . 18
10.6 Bromine. 18
10.7 Repeatability statement for five tested substances sorted by type of tested
material . 18
10.7.1 General . 18
10.7.2 Material: ABS (acrylonitrile butadiene styrene), as granules and
plates . 18
10.7.3 Material: PE (low density polyethtylene), as granules . 19
10.7.4 Material: PC/ABS (polycarbonate and ABS blend), as granules . 19
10.7.5 Material: HIPS (high impact polystyrene) . 19
10.7.6 Material: PVC (polyvinyl chloride), as granules . 19
10.7.7 Material: Polyolefin, as granules . 19
10.7.8 Material: Crystal glass . 20
10.7.9 Material: Glass . 20
10.7.10 Material: Lead-free solder, chips . 20
62321-3-1 © IEC:2013 – 3 –
10.7.11 Material: Si/Al Alloy, chips . 20
10.7.12 Material: Aluminum casting alloy, chips . 20
10.7.13 Material: PCB – Printed circuit board ground to less than 250 µm . 20
10.8 Reproducibility statement for five tested substances sorted by type of tested
material . 20
10.8.1 General . 20
10.8.2 Material: ABS (Acrylonitrile butadiene styrene), as granules and
plates . 21
10.8.3 Material: PE (low density polyethtylene), as granules . 21
10.8.4 Material: PC/ABS (Polycarbonate and ABS blend), as granules . 21
10.8.5 Material: HIPS (high impact polystyrene) . 21
10.8.6 Material: PVC (polyvinyl chloride), as granules . 22
10.8.7 Material: Polyolefin, as granules . 22
10.8.8 Material: Crystal glass . 22
10.8.9 Material: Glass . 22
10.8.10 Material: Lead-free solder, chips . 22
10.8.11 Material: Si/Al alloy, chips . 22
10.8.12 Material: Aluminum casting alloy, chips . 22
10.8.13 Material: PCB – Printed circuit board ground to less than 250 µm . 22
11 Quality control . 23
11.1 Accuracy of calibration . 23
11.2 Control samples . 23
12 Special cases . 23
13 Test report . 23
Annex A (informative) Practical aspects of screening by X-ray fluorescence
spectrometry (XRF) and interpretation of the results . 25
Annex B (informative) Practical examples of screening with XRF . 31
Bibliography . 40
Figure B.1 – AC power cord, X-ray spectra of sampled sections . 32
Figure B.2 – RS232 cable and its X-ray spectra . 33
Figure B.3 – Cell phone charger shown partially disassembled . 34
Figure B.4 – PWB and cable of cell phone charger . 35
Figure B.5 – Analysis of a single solder joint on a PWB . 36
Figure B.6 – Spectra and results obtained on printed circuit board with two collimators . 36
Figure B.7 – Examples of substance mapping on PWBs . 38
Figure B.8 – SEM-EDX image of Pb free solder with small intrusions of Pb (size = 30 µm) . 39
Table 1 – Tested concentration ranges for lead in materials . 8
Table 2 – Tested concentration ranges for mercury in materials . 9
Table 3 – Tested concentration ranges for cadmium in materials . 9
Table 4 – Tested concentration ranges for total chromium in materials . 9
Table 5 – Tested concentration ranges for total bromine in materials . 9
Table 6 – Recommended X-ray lines for individual analytes . 14
Table A.1 – Effect of matrix composition on limits of detection of some controlled
elements . 26
– 4 – 62321-3-1 © IEC:2013
Table A.2 – Screening limits in mg/kg for regulated elements in various matrices . 27
Table A.3 – Statistical data from IIS2 . 29
Table A.4 – Statistical data from IIS4 . 30
Table B.1 – Selection of samples for analysis of AC power cord . 32
Table B.2 – Selection of samples (testing locations) for analysis after visual inspection
– Cell phone charger. 34
Table B.3 – Results of XRF analysis at spots (1) and (2) as shown in Figure B.6 . 37
62321-3-1 © IEC:2013 – 5 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
DETERMINATION OF CERTAIN SUBSTANCES
IN ELECTROTECHNICAL PRODUCTS –
Part 3-1: Screening – Lead, mercury, cadmium, total chromium
and total bromine by X-ray fluorescence spectrometry
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 62321-3-1 has been prepared by IEC technical committee 111:
Environmental standardization for electrical and electronic products and systems.
It has the status of a horizontal standard in accordance with IEC Guide 108.
The first edition of IEC 62321:2008 was a 'stand alone' standard that included an introduction,
an overview of test methods, a mechanical sample preparation as well as various test method
clauses.
This first edition of IEC 62321-3-1 is a partial replacement of IEC 62321:2008, forming a
structural revision and generally replacing Clauses 6 and Annex D.
Future parts in the IEC 62321 series will gradually replace the corresponding clauses in
IEC 62321:2008. Until such time as all parts are published, however, IEC 62321:2008 remains
valid for those clauses not yet re-published as a separate part.
– 6 – 62321-3-1 © IEC:2013
The text of this standard is based on the following documents:
FDIS Report on voting
111/298/FDIS 111/308/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
A list of all parts in the IEC 62321 series can be found on the IEC website under the general
title: Determination of certain substances in electrotechnical products
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
62321-3-1 © IEC:2013 – 7 –
INTRODUCTION
The widespread use of electrotechnical products has drawn increased attention to their impact
on the environment. In many countries this has resulted in the adaptation of regulations
affecting wastes, substances and energy use of electrotechnical products.
The use of certain substances (e.g. lead (Pb), cadmium (Cd) and polybrominated diphenyl
ethers (PBDEs)) in electrotechnical products, is a source of concern in current and proposed
regional legislation.
The purpose of the IEC 62321 series is therefore to provide test methods that will allow the
electrotechnical industry to determine the levels of certain substances of concern in
electrotechnical products on a consistent global basis.
WARNING – Persons using this International Standard should be familiar with normal
laboratory practice. This standard does not purport to address all of the safety
problems, if any, associated with its use. It is the responsibility of the user to establish
appropriate safety and health practices and to ensure compliance with any national
regulatory conditions.
– 8 – 62321-3-1 © IEC:2013
DETERMINATION OF CERTAIN SUBSTANCES
IN ELECTROTECHNICAL PRODUCTS –
Part 3-1: Screening – Lead, mercury, cadmium, total chromium
and total bromine by X-ray fluorescence spectrometry
1 Scope
Part 3-1 of IEC 62321 describes the screening analysis of five substances, specifically lead
(Pb), mercury (Hg), cadmium (Cd), total chromium (Cr) and total bromine (Br) in uniform
materials found in electrotechnical products, using the analytical technique of X-ray
fluorescence (XRF) spectrometry.
It is applicable to polymers, metals and ceramic materials. The test method may be applied to
raw materials, individual materials taken from products and “homogenized” mixtures of more
than one material. Screening of a sample is performed using any type of XRF spectrometer,
provided it has the performance characteristics specified in this test method. Not all types of
XRF spectrometers are suitable for all sizes and shapes of sample. Care should be taken to
select the appropriate spectrometer design for the task concerned.
The performance of this test method has been tested for the following substances in various
media and within the concentration ranges as specified in Tables 1 to 5.
Table 1 – Tested concentration ranges for lead in materials
Substance/
Lead
element
Medium/material tested
Unit of
a b d
ABS PE Low- Al, Lead- Ground Crystal PVC Poly-
Parameter
measure
c
alloy Al-Si free PWB glass olefine
steel alloy solder
Concentration
15,7 14 190 22 000 390
or 380 to
e
mg/kg to to 30 to 174 to 240 000 to
concentration 640
954 108 930 23 000 665
range tested
a
Acrylonitrile butadiene styrene.
b
Polyethylene.
c
Printed wiring board.
d
Polyvinyl chloride.
e
This lead concentration was not detectable by instruments participating in tests.
62321-3-1 © IEC:2013 – 9 –
Table 2 – Tested concentration ranges for mercury in materials
Substance/element Mercury
Medium/material tested
Parameter Unit of measure
a b
ABS PE
Concentration or concentration range
mg/kg 100 to 942 4 to 25
tested
a
Acrylonitrile butadiene styrene.
b
Polyethylene.
Table 3 – Tested concentration ranges for cadmium in materials
Substance/element Cadmium
Medium/material tested
Parameter Unit of measure
a b
Lead-free solder ABS PE
Concentration or concentration
c
mg/kg 3 10 to 183 19,6 to 141
range tested
a
Acrylonitrile butadiene styrene.
b
Polyethylene.
c
This cadmium concentration was not detectable by instruments participating in tests.
Table 4 – Tested concentration ranges for total chromium in materials
Substance/element Chromium
Medium/material tested
Unit of
Low-
Parameter
Al, Al-Si
measure a b
ABS PE alloy Glass
alloy
steel
Concentration or
concentration range mg/kg 16 to 944 16 to 115 240 130 to 1 100 94
tested
a
Acrylonitrile butadiene styrene.
b
Polyethylene.
Table 5 – Tested concentration ranges for total bromine in materials
Substance/element Bromine
Medium/material tested
Unit of
Parameter
measure c a d b
HIPS , ABS PC/ABS PE
Concentration or concentration
mg/kg 25 to 118 400 800 to 2 400 96 to 808
range tested
a
Acrylonitrile butadiene styrene.
b
Polyethylene.
c
High impact polystyrene.
d
Polycarbonate and ABS blend.
These substances in similar media outside of the specified concentration ranges may be
analysed according to this test method; however, the performance has not been established
for this standard.
– 10 – 62321-3-1 © IEC:2013
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC 62321-1, Determination of certain substances in electrotechnical products – Part 1:
Introduction and overview
IEC 62321-2, Determination of certain substances in electrotechnical products – Part 2:
Disassembly, disjointment and mechanical sample preparation
IEC/ISO Guide 98-1, Uncertainty of measurement – Part 1: Introduction to the expression of
uncertainty in measurement
3 Terms, definitions and abbreviations
For the purposes of this document, the terms, definitions and abbreviations given in
IEC 62321-1 and IEC 62321-2 apply.
4 Principle
4.1 Overview
The concept of 'screening' has been developed to reduce the amount of testing. Executed as
a predecessor to any other test analysis, the main objective of screening is to quickly
determine whether the screened part or section of a product:
– contains a certain substance at a concentration significantly higher than its value or values
chosen as criterion, and therefore may be deemed unacceptable;
– contains a certain substance at a concentration significantly lower than its value or values
chosen as criterion, and therefore may be deemed acceptable;
– contains a certain substance at a concentration so close to the value or values chosen as
criterion that when all possible errors of measurement and safety factors are considered,
no conclusive decision can be made about the acceptable absence or presence of a
certain substance and, therefore, a follow-up action may be required, including further
analysis using verification testing procedures.
This test method is designed specifically to screen for lead, mercury, cadmium, chromium and
bromine (Pb, Hg, Cd, Cr, Br) in uniform materials, which occur in most electrotechnical
products. Under typical circumstances, XRF spectrometry provides information on the total
quantity of each element present in the sample, but does not identify compounds or valence
states of the elements. Therefore, special attention shall be paid when screening for
chromium and bromine, where the result will reflect only the total chromium and total bromine
present. The presence of Cr(VI) or the brominated flame retardants PBB or PBDE shall be
confirmed by a verification test procedure. When applying this method to electronics “as
received”, which, by the nature of their design, are not uniform, care shall be taken in
interpreting the results. Similarly, the analysis of Cr in conversion coatings may be difficult
due to the presence of Cr in substrate material and/or because of insufficient sensitivity for Cr
in typically very thin (several hundred nm) conversion coating layers.
Screening analysis can be carried out by one of two means:
___________
To be published.
62321-3-1 © IEC:2013 – 11 –
• non-destructively – by directly analysing the sample “as received”;
• destructively – by applying one or more sample preparation steps prior to analysis.
In the latter case, the user shall apply the procedure for sample preparation as described in
IEC 62321-2. This test method will guide the user in choosing the proper approach to sample
presentation.
4.2 Principle of test
The representative specimen of the object tested is placed in the measuring chamber or over
the measuring aperture of the X-ray fluorescence spectrometer. Alternatively, a measuring
window/aperture of a handheld, portable XRF analyser is placed flush against the surface of
the object tested. The analyser illuminates the specimen for a preselected measurement time
with a beam of X rays which in turn excite characteristic X rays of elements in the specimen.
The intensities of these characteristic X rays are measured and converted to mass fractions
or concentrations of the elements in the tested sample using a calibration implemented in the
analyser.
The fundamentals of XRF spectrometry, as well as practical aspects of sampling for XRF, are
covered in detail in [1, 2 and 3].
4.3 Explanatory comments
To achieve its purpose, this test method shall provide rapid, unambiguous identification of the
elements of interest. The test method shall provide at least a level of accuracy that is
sometimes described as semi-quantitative, i.e. the relative uncertainty of a result is typically
30 % or better at a defined level of confidence of 68 %. Some users may tolerate higher
relative uncertainty, depending on their needs. This level of performance allows the user to
sort materials for additional testing. The overall goal is to obtain information for risk
management purposes.
This test method is designed to allow XRF spectrometers of all designs, complexity and
capability to contribute screening analyses. However, the capabilities of different XRF
spectrometers cover such a wide range that some will be relatively inadequate in their
selectivity and sensitivity while others will be more than adequate. Some spectrometers will
allow easy measurement of a wide range of sample shapes and sizes, while others, especially
research-grade WDXRF units, will be very inflexible in terms of test portions.
Given the above level of required performance and the wide variety of XRF spectrometers
capable of contributing useful measurements, the requirements for the specification of
procedures are considerably lower than for a high-performance test method for quantitative
determinations with low estimates of uncertainty.
This test method is based on the concept of a performance based measurement system.
Apparatus, sample preparation and calibration are specified in this standard in relatively
general terms. It is the responsibility of the user to document all procedures developed in the
laboratory that uses the test method. The user shall establish a written procedure for all cases
denoted in this method by the term “work instructions”.
The user of this test method shall document all relevant spectrometer and method
performance parameters.
WARNING 1 Persons using the XRF test method shall be trained in the use of XRF
spectrometers and the related sampling requirements.
WARNING 2 Xrays are hazardous to humans. Care shall be taken to operate the equipment
in accordance with both the safety instructions provided by the manufacturer and the
applicable local health and occupational safety regulations.
– 12 – 62321-3-1 © IEC:2013
5 Apparatus, equipment and materials
5.1 XRF spectrometer
An XRF spectrometer consists of an X-ray excitation source, a means of reproducible sample
presentation, an X-ray detector, a data processor and a control system [4, 5 and 6]:
a) source of X-ray excitation – X-ray tube or radio-isotope sources are commonly used;
b) X-ray detector (detection subsystem) – Device used to convert the energy of an X-ray
photon to a corresponding electric pulse of amplitude proportional to the photon energy.
5.2 Materials and tools
All materials used in the preparation of samples for XRF measurements shall be shown to be
free of contamination, specifically by the analytes of this test method. This means that all
grinding materials, solvents, fluxes, etc. shall not contain detectable quantities of Pb, Hg, Cd,
Cr and/or Br.
Tools used in the handling of samples shall be chosen to minimize contamination by the
analytes of this test method as well as by any other elements. Any procedures used to clean
the tools shall not introduce contaminants.
6 Reagents
Reagents, if any, shall be of recognized analytical grade and shall not contain detectable
quantities of Pb, Hg, Cd, Cr and/or Br.
7 Sampling
7.1 General
It is the responsibility of the user of this test method to define the test sample using
documented work instructions. The user may choose to define the test sample in a number of
ways, either via a non-destructive approach in which the portion to be measured is defined by
the viewing area of the spectrometer, or by a destructive approach in which the portion to be
measured is removed from the larger body of material and either measured as is, or
destroyed and prepared using a defined procedure.
7.2 Non-destructive approach
The user of this test method shall:
a) establish the area viewed by the spectrometer and place the test sample within that area,
taking care to ascertain that no fluorescent X-rays will be detected from materials other
than the defined test portion. Usually, the area viewed by the spectrometer is a section of
a plane delineated by the shape and boundary of the measuring window of the instrument.
The area of the test sample viewed by the spectrometer shall be flat. Any deviation from
the flat area requirement shall be documented;
b) make sure that a repeatable measurement geometry with a repeatable distance between
the spectrometer and the test portion is established;
c) document the steps taken to disassemble a larger object to obtain a test portion.
7.3 Destructive approach
The following points shall be taken into account in the destructive approach:
62321-3-1 © IEC:2013 – 13 –
a) the user shall create and follow a documented work instruction for the means of
destruction applied to obtain the test portion, as this information is critical for correct
interpretation of the measurement results;
b) a procedure that results in a powder shall produce a material with a known or controlled
particle size. In cases where the particles have different chemical, phase or mineralogical
compositions, it is critical to reduce their size sufficiently to minimize differential
absorption effects;
c) in a procedure that results in a material being dissolved in a liquid matrix, the quantity and
physical characteristics of the material to be dissolved shall be controlled and documented.
The resulting solution shall be completely homogeneous. Instructions shall be provided to
deal with undissolved portions to ensure proper interpretation of the measured results.
Instructions shall be provided for presentation of the test portion of the solution to the X-
ray spectrometer in a repeatable manner, i.e. in a liquid cell of specified construction and
dimensions;
d) in a procedure that results in a sample material being fused or pressed in a solid matrix,
the quantity and physical characteristics of the sample material shall be controlled and
documented. The resulting solid (fused or pressed pellet) shall be completely uniform.
Instructions shall be provided to deal with unmixed portions to ensure proper
interpretation of the measured results.
8 Test procedure
8.1 General
The test procedure covers preparation of the X-ray spectrometer, preparation and mounting of
test portions and calibration. Certain instructions are presented in general terms due to the
wide range of XRF equipment and the even greater variety of laboratory and test samples to
which this test method will be applied. However, a cardinal rule that applies without exception
to all spectrometers and analytical methods shall be followed; that is that the calibration and
sample measurements be performed under the same conditions and using the same sample
preparation procedures.
In view of the wide range of XRF spectrometer designs and the concomitant range of
detection capabilities, it is important to understand the limitation of the chosen instrument.
Certain designs may be incapable of detecting or accurately determining the composition of a
very small area or very thin samples. As a consequence, it is imperative that users carefully
establish and clearly document the performance of the test method as implemented in their
laboratories. One goal is to prevent false negative test results.
8.2 Preparation of the spectrometer
Prepare the spectrometer as follows:
a) switch on the instrument and prepare it for operation according to the manufacturer’s
manual. Allow the instrument to stabilize as per guidelines established by the
manufacturer or laboratory work instructions;
b) set the measurement conditions to the optimum conditions previously established by the
manufacturer or the laboratory.
Many instruments available on the market are already optimized and preset for a particular
application, and therefore this step might not be necessary. Otherwise, the laboratory should
establish optimum operating conditions for each calibration. Choices should be made to
optimize sensitivity and minimize spectral interferences. Excitation conditions may vary by
material, analyte and X-ray line energy. A list of recommended analytical X-ray lines is given
in Table 6. Detection system settings should optimize the compromise between sensitivity and
energy resolution. Guidance can usually be found in the instrument manual and in literature
on X-ray spectrometry [1, 2 and 3].
– 14 – 62321-3-1 © IEC:2013
a
Table 6 – Recommended X-ray lines for individual analytes
Analyte Preferred line Secondary line
Lead (Pb) L –M (Lβ ) L –M (Lα )
2 4 1 3 4,5 1,2
Mercury (Hg) L –M (Lα )
3 4,5 1,2
b
Cadmium (Cd) K–L (Kα )
2,3 1,2
Chromium (Cr) K–L (Kα )
2,3 1,2
Bromine (Br) K–L (Kα ) K–M (Kβ )
2,3 1,2 2,3 1,3
a
Other X-ray line choices may provide adequate performance. However, when deciding on alternative
analytical lines one should be aware of possible spectral interferences from other elements present in the
sample (e.g. BrKα on PbLα or AsKα on PbLα lines; see Clause A.2 b) for more typical examples).
b
K–L (Kα ) means that there are actually two transitions to the K shell, i.e. one from the L shell which
2,3 1,2 2
generates Kα X-rays and another from the L shell that generates Kα X- rays. However, since both energies
2 3 1
are very close, energy dispersive spectrometers cannot distinguish them and so they are analysed as one
combined K α energy.
1,2
8.3 Test portion
The creation of a test portion is described in Clause 7.
In the case of destructive sample preparation, measure the mass and dimensions of the test
portion as required by the calibration method and the work instruction established by the
laboratory to ensure repeatable sampling. The location of the test portion shall also be documented
in relation to its origin on the electrotechnical product.
8.4 Verification of spectrometer performance
Spectrometer performance shall be verified as follows:
a) Users shall provide objective evidence of the performance of the method as implemented
in their laboratories. This is necessary to enable the users and their customers to
understand the limitations of the method and to make decisions using the results of
analyses. Critical aspects regarding the performance of the method are as follows:
• sensitivity for each analyte;
• spectral resolution;
• limit of detection;
• demonstration of measured area;
• repeatability of sample preparation and measurement;
• accuracy of calibration, which will be checked according to Clause 10.
Given the variety of spectrometers and the associated software operating systems, it is
acceptable for the users to obtain this information in their own laboratory using their own
procedures or as a service provided by the manufacturer. It is important to obtain
verification of spectrometer and method performance when the method is implemented.
Evidence of the maintenance of performance may be obtained through the use of control
charts or by repeating the measurements and calculations made at the time of
implementation;
b) Spectrometer sensitivity is used as a figure of merit to compare spectrometers and to
ensure that a meaningful calibration is possible.
c) Spectral resolution is important to ensure that the analyte and interfering spectral lines are
handled correctly in the collection of data and in the calibration. For the purposes of this
standard, the correction of line overlaps is considered as part of the spectrometer
calibration.
d) The limit of detection, LOD, shall be estimated for each set of operating conditions
employed in the test method using Equation (1) below:
62321-3-1 © IEC:2013 – 15 –
LOD = 3σ (1)
where
LOD is the limit of detection (LOD) for given analyte expressed in units of concentration;
σ is the standard deviation of the results of multiple determinations using a blank
material. Standard deviation is usually estimated using a small (but not less than
seven) number of determinations, in which case the symbol, s (the unbiased
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...