EN IEC 60601-2-35:2021
(Main)Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads and mattresses and intended for heating in medical use
Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads and mattresses and intended for heating in medical use
IEC 60601-2-35:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the scope of this document. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. If a clause or subclause is specifically intended to be applicable to a specifically defined type of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-35:2020 does not apply to: - HEATING DEVICES intended for physiotherapy; - INFANT RADIANT WARMERS; for information, see IEC 60601-2-21; - INFANT INCUBATORS; for information, see IEC 60601-2-19; - INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20; - cooling devices. IEC 60601-2-35:2020 cancels and replaces IEC 80601-2-35 published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-35:2020 includes the following significant technical change with respect to the previous edition: re-dating of normative references.
Medizinische elektrische Geräte - Teil 2-35: Besondere Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale von Decken, Matten und Matratzen zur Erwärmung von Patienten in der medizinischen Anwendung
Appareils électromédicaux - Partie 2-35: Exigences particulières pour la sécurité de base et les performances essentielles des dispositifs de réchauffage utilisant des couvertures, des coussins ou des matelas chauffants et destinés au réchauffage des patients en usage médical
L'IEC 60601-2-35:2020 s’applique à la SECURITE DE BASE et aux PERFORMANCES ESSENTIELLES des DISPOSITIFS DE RECHAUFFAGE utilisant des COUVERTURES, des COUSSINS ou des MATELAS en usage médical, également désignés sous le terme APPAREILS EM. Les DISPOSITIFS DE RECHAUFFAGE destinés à préchauffer un lit sont compris dans le domaine d'application du présent document. Si un article ou un paragraphe est spécifiquement destiné à être applicable uniquement aux APPAREILS EM ou uniquement aux SYSTEMES EM, le titre et le contenu de cet article ou de ce paragraphe l'indiquent. Si cela n'est pas le cas, l'article ou le paragraphe s'applique à la fois aux APPAREILS EM et aux SYSTEMES EM, selon le cas. Si un article ou un paragraphe est spécifiquement destiné à être applicable à un type spécifiquement défini d’APPAREIL EM, comme c’est le cas avec les DISPOSITIFS A AIR PULSE, alors le titre et le contenu de cet article ou de ce paragraphe l’indiquent. Si cela n'est pas le cas, l'article ou le paragraphe s'applique à la fois aux APPAREILS EM et aux SYSTEMES EM, selon le cas. Les DANGERS inhérents à la fonction physiologique prévue des APPAREILS EM ou des SYSTEMES EM dans le cadre du domaine d'application du présent document ne sont pas couverts par des exigences spécifiques contenues dans le présent document, à l'exception de 7.2.13 et de 8.4.1 de la norme générale. Le présent document ne s'applique pas aux: – DISPOSITIFS DE RECHAUFFAGE destinés à la physiothérapie; – INCUBATEURS RADIANTS POUR NOUVEAU-NES; voir l’IEC 60601-2-21 à titre informatif; – INCUBATEURS POUR NOUVEAU-NES; voir l’IEC 60601-2-19 à titre informatif; – incubateurs de transport POUR NOUVEAU-NES; voir l’IEC 60601-2-20 à titre informatif; – dispositifs de refroidissement. L'IEC 60601-2-35:2020 annule et remplace l’IEC 80601-2-35 parue en 2009 et son Amendement 1 (2016).
Medicinska električna oprema - 2-35. del: Posebne zahteve za osnovno varnost in bistvene lastnosti za odeje, blazine in posteljne vložke, namenjene za ogrevanje pri medicinski uporabi (IEC 60601-2-35:2020)
Ta del mednarodnega standarda 60601 se uporablja za OSNOVNO VARNOST in BISTVENE LASTNOSTI za ODEJE, BLAZINE in POSTELJNE VLOŽKE, namenjene ogrevanju v medicinski uporabi, imenovane tudi ELEKTROMEDICINSKA OPREMA. NAPRAVE ZA OGREVANJE, namenjene predhodnemu ogrevanju postelje, so vključene v področje uporabe tega dokumenta.
Če je točka ali podtočka namenjena samo obravnavi ELEKTROMEDICINSKE OPREME ali ELEKTROMEDICINSKIH SISTEMOV, bo to zapisano v naslovu in vsebini točke ali podtočke. Če ni zapisano, točka ali podtočka obravnava ELEKTROMEDICINSKO OPREMO in ELEKTROMEDICINSKE SISTEME.
Če je točka ali podtočka namenjena samo obravnavi posebej opredeljene vrste ELEKTROMEDICINSKE OPREME, kot v primeru NAPRAVE Z UMETNIM PRETOKOM ZRAKA, bo to zapisano v naslovu in vsebini dane točke ali podtočke. Če ni zapisano, točka ali podtočka obravnava ELEKTROMEDICINSKO OPREMO in ELEKTROMEDICINSKE SISTEME.
Posebne zahteve tega dokumenta ne vključujejo temeljnih TVEGANJ pri predvideni fiziološki funkciji ELEKTROMEDICINSKE OPREME in SISTEMOV, ki spadajo na področje uporabe tega dokumenta, razen v poglavjih 7.2.13 in 8.4.1 splošnega standarda.
OPOMBA: Glej tudi točko 4.2 splošnega standarda.
Ta dokument se ne uporablja za:
– NAPRAVE ZA OGREVANJE za uporabo v fizioterapiji;
– OTROŠKE SEVALNE OGREVALNIKE; za informacije glej standard IEC 60601-2-21 [1]2;
– OTROŠKE INKUBATORJE; za informacije glej standard IEC 60601-2-19 [2];
– PRENOSNE OTROŠKE INKUBATORJE, za informacije glej standard IEC 60601-2-20 [3];
– naprave za hlajenje.
General Information
- Status
- Published
- Publication Date
- 15-Jul-2021
- Technical Committee
- CLC/TC 62 - Electrical equipment in medical practice
- Drafting Committee
- IEC/SC 62D - IEC_SC_62D
- Current Stage
- 6060 - Document made available - Publishing
- Start Date
- 16-Jul-2021
- Completion Date
- 16-Jul-2021
Relations
- Effective Date
- 23-Jan-2023
- Effective Date
- 07-Jun-2022
- Effective Date
- 20-Jul-2021
- Effective Date
- 20-Jul-2021
- Effective Date
- 26-Jul-2021
Overview
EN IEC 60601-2-35:2021 (IEC 60601-2-35:2020) specifies particular requirements for the basic safety and essential performance of medical electrical heating devices that use blankets, pads or mattresses and are intended for patient warming in medical use. The standard covers heating devices (including those intended to prewarm a bed) as Medical Electrical (ME) Equipment and ME Systems, and defines when clauses apply specifically to ME EQUIPMENT, ME SYSTEMS or to particular device types such as forced air devices.
This edition cancels and replaces the earlier IEC 80601-2-35 (2009) series and constitutes a technical revision (notably re‑dating of normative references). It does not apply to devices for physiotherapy, infant radiant warmers, infant incubators, infant transport incubators, or cooling devices.
Key topics and technical requirements
- Scope & application: Requirements for design, construction, testing and documentation to ensure safe operation and essential performance of patient warming blankets, pads and mattresses.
- Safety domains addressed:
- Electrical safety and protection against electrical hazards.
- Mechanical safety (construction, robustness, impact and ruck-resistance tests).
- Thermal safety: limits and tests for contact surface temperature, heat transfer to/from the patient, and protection against excessive temperatures (including single-fault conditions).
- Electromagnetic compatibility (EMC) and immunity per collateral standards.
- Alarms, controls and accuracy: requirements for controls, indicators and protection against hazardous outputs.
- Programmable systems (PEMS): guidance when devices include programmable electronic functions.
- Device‑specific clauses: Special requirements for forced air devices and other defined types are identified explicitly in clause titles.
- Normative test annexes: Several normative annexes define test procedures for heat transfer, contact surface temperature, forced air device tests and conditions of adequate heat discharge.
Practical applications and users
- Medical device manufacturers designing patient warming blankets, heating pads, mattress systems or prewarming bed devices.
- Test laboratories and conformity assessors performing safety and performance testing for CE marking and regulatory compliance.
- Regulatory affairs and quality managers preparing technical documentation and risk analyses to meet IEC 60601 series requirements.
- Hospitals and procurement teams evaluating product safety claims and specifications for patient warming equipment.
- Design and safety engineers implementing thermal controls, alarms and EMC mitigation in ME equipment.
Related standards
- IEC 60601-1 (General requirements for basic safety and essential performance)
- IEC 60601-1-2 (EMC)
- IEC 60601-1-10 (Physiologic closed-loop controllers)
- IEC 60601-2-19 / -2-20 / -2-21 (infant incubators / transport incubators / radiant warmers) - excluded from scope
- EN IEC 60601-2-35:2021 (European adoption / harmonized text)
Keywords: IEC 60601-2-35, EN IEC 60601-2-35:2021, medical electrical equipment, patient warming, heating blankets, pads, mattresses, forced air device, contact surface temperature, medical device safety, EMC, CE conformity.
Frequently Asked Questions
EN IEC 60601-2-35:2021 is a standard published by CLC. Its full title is "Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads and mattresses and intended for heating in medical use". This standard covers: IEC 60601-2-35:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the scope of this document. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. If a clause or subclause is specifically intended to be applicable to a specifically defined type of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-35:2020 does not apply to: - HEATING DEVICES intended for physiotherapy; - INFANT RADIANT WARMERS; for information, see IEC 60601-2-21; - INFANT INCUBATORS; for information, see IEC 60601-2-19; - INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20; - cooling devices. IEC 60601-2-35:2020 cancels and replaces IEC 80601-2-35 published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-35:2020 includes the following significant technical change with respect to the previous edition: re-dating of normative references.
IEC 60601-2-35:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the scope of this document. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. If a clause or subclause is specifically intended to be applicable to a specifically defined type of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard. IEC 60601-2-35:2020 does not apply to: - HEATING DEVICES intended for physiotherapy; - INFANT RADIANT WARMERS; for information, see IEC 60601-2-21; - INFANT INCUBATORS; for information, see IEC 60601-2-19; - INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20; - cooling devices. IEC 60601-2-35:2020 cancels and replaces IEC 80601-2-35 published in 2009 and Amendment 1:2016. This edition constitutes a technical revision. IEC 60601-2-35:2020 includes the following significant technical change with respect to the previous edition: re-dating of normative references.
EN IEC 60601-2-35:2021 is classified under the following ICS (International Classification for Standards) categories: 11.040.01 - Medical equipment in general; 11.140 - Hospital equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
EN IEC 60601-2-35:2021 has the following relationships with other standards: It is inter standard links to EN 80601-2-35:2009/A11:2011, EN 80601-2-35:2009, EN 80601-2-35:2009/A1:2016, EN 80601-2-35:2009/AC:2015, EN IEC 60601-2-35:2021/A1:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN IEC 60601-2-35:2021 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC, 93/44/EEC; Standardization Mandates: M/295. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
You can purchase EN IEC 60601-2-35:2021 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of CLC standards.
Standards Content (Sample)
SLOVENSKI STANDARD
01-september-2021
Nadomešča:
SIST EN 80601-2-35:2010
SIST EN 80601-2-35:2010/A1:2017
SIST EN 80601-2-35:2010/A11:2012
SIST EN 80601-2-35:2010/AC:2015
Medicinska električna oprema - 2-35. del: Posebne zahteve za osnovno varnost in
bistvene lastnosti za odeje, blazine in posteljne vložke, namenjene za ogrevanje
pri medicinski uporabi (IEC 60601-2-35:2020)
Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and
essential performance of heating devices using blankets, pads and mattresses and
intended for heating in medical use (IEC 60601-2-35:2020)
Medizinische elektrische Geräte - Teil 2-35: Besondere Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale von Decken, Matten und Matratzen
zur Erwärmung von Patienten in der medizinischen Anwendung (IEC 60601-2-35:2020)
Appareils électromédicaux - Partie 2-35: Exigences particulières pour la sécurité de base
et les performances essentielles des dispositifs de réchauffage utilisant des couvertures,
des coussins ou des matelas chauffants et destinés au réchauffage des patients en
usage médical (IEC 60601-2-35:2020)
Ta slovenski standard je istoveten z: EN IEC 60601-2-35:2021
ICS:
11.140 Oprema bolnišnic Hospital equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN IEC 60601-2-35
NORME EUROPÉENNE
EUROPÄISCHE NORM
July 2021
ICS 11.140; 11.040.01 Supersedes EN 80601-2-35:2009 and all of its
amendments and corrigenda (if any)
English Version
Medical electrical equipment - Part 2-35: Particular requirements
for the basic safety and essential performance of heating
devices using blankets, pads and mattresses and intended for
heating in medical use
(IEC 60601-2-35:2020)
Appareils électromédicaux - Partie 2-35: Exigences Medizinische elektrische Geräte - Teil 2-35: Besondere
particulières pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles des dispositifs de réchauffage utilisant des wesentlichen Leistungsmerkmale von Decken, Matten und
couvertures, des coussins ou des matelas chauffants et Matratzen zur Erwärmung von Patienten in der
destinés au réchauffage des patients en usage médical medizinischen Anwendung
(IEC 60601-2-35:2020) (IEC 60601-2-35:2020)
This European Standard was approved by CENELEC on 2020-10-26. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 60601-2-35:2021 E
European foreword
The text of document 62D/1765/FDIS, future edition 2 of IEC 60601-2-35, prepared by SC 62D
"Electromedical equipment" of IEC/TC 62 "Electrical equipment in medical practice" was submitted to
the IEC-CENELEC parallel vote and approved by CENELEC as EN IEC 60601-2-35:2021.
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2022-01-16
level by publication of an identical national
standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2024-07-16
document have to be withdrawn
This document supersedes EN 80601-2-35:2009 and all of its amendments and corrigenda (if any).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
Endorsement notice
The text of the International Standard IEC 60601-2-35:2020 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards
indicated:
IEC 60601-2-19:2020 NOTE Harmonized as EN IEC 60601-2-19:2021 (not modified)
IEC 60601-2-20:2020 NOTE Harmonized as EN IEC 60601-2-20:2020 (not modified)
IEC 60601-2-21:2020 NOTE Harmonized as EN IEC 60601-2-21:2020 (not modified)
IEC 60335-2-53 NOTE Harmonized as EN 60335-2-53
IEC 60529:1989 NOTE Harmonized as EN 60529:1991 (not modified)
IEC 60529:1989/A1:1999 NOTE Harmonized as EN 60529:1991/A1:2000 (not modified)
IEC 60529:1989/A2:2013 NOTE Harmonized as EN 60529:1991/A2:2013 (not modified)
ISO 2439:2008 NOTE Harmonized as EN ISO 2439:2008 (not modified)
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1 Where an International Publication has been modified by common modifications, indicated by (mod),
the relevant EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is
available here: www.cenelec.eu.
The Annex ZA of EN 60601-1:2006/A1:2013 applies with the following additions:
Publication Year Title EN/HD Year
IEC 60601-1 2005 Medical electrical equipment - Part 1: EN 60601-1 2006
General requirements for basic safety and
essential performance
- - + corrigendum Mar. 2010
+ A1 2012 + A1 2013
- - + A12 2014
IEC 60601-1-10 2007 Medical electrical equipment - Part 1-10: EN 60601-1-10 2008
General requirements for basic safety and
essential performance - Collateral
Standard: Requirements for the
development of physiologic closed-loop
controllers
+ A1 2013 + A1 2015
The Annex ZA of EN 60601-1:2006/A1:2013 applies with the following replacements:
IEC 60384-14 2013 Fixed capacitors for use in electronic EN 60384-14 2013
equipment - Part 14: Sectional
specification - Fixed capacitors for
electromagnetic interference suppression
and connection to the supply mains
+ A1 2016 + A1 2016
IEC 60601-1-2 2014 Medical electrical equipment - Part 1-2: EN 60601-1-2 2015
General requirements for basic safety and
essential performance - Collateral
Standard: Electromagnetic disturbances -
Requirements and tests
IEC 60601-2-35 ®
Edition 2.0 2020-09
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-35: Particular requirements for the basic safety and essential performance
of heating devices using blankets, pads or mattresses and intended for heating
in medical use
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 11.040.01; 11.140 ISBN 978-2-8322-8723-1
– 2 – IEC 60601-2-35:2020 © IEC 2020
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terms and definitions. 10
201.4 General requirements . 13
201.5 General requirements for testing ME EQUIPMENT . 14
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 14
201.7 ME EQUIPMENT identification, marking and documents . 14
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 18
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 24
201.10 Protection against unwanted and excessive radiation HAZARDS . 26
201.11 Protection against excessive temperatures and other HAZARDS . 26
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 28
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 33
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 38
201.15 Construction of ME EQUIPMENT . 38
201.16 ME SYSTEMS . 42
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 42
202 ELECTROMAGNETIC DISTURBANCES – Requirements and tests . 43
208 General requirements, tests and guidance for ALARM SYSTEMS IN MEDICAL
ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS . 43
210 * Requirements for the development of PHYSIOLOGIC CLOSED-LOOP CONTROLLERS . 43
Annexes . 44
Annex D (informative) Symbols on marking . 44
Annex AA (informative) Particular guidance and rationale . 45
Annex BB (normative) Determination of the LAGGING MATERIAL . 57
Annex CC (normative) * Determination of heat transfer towards the PATIENT . 58
Annex DD (normative) * Determination of heat transfer away from the PATIENT . 60
Annex EE (normative) CONDITIONS OF ADEQUATE HEAT DISCHARGE . 61
Annex FF (normative) Test procedure for maximum CONTACT SURFACE TEMPERATURE
for FORCED AIR DEVICES . 62
Annex GG (normative) Test procedure for maximum CONTACT SURFACE TEMPERATURE
for FORCED AIR DEVICES under SINGLE FAULT CONDITION. 64
Annex HH (normative) Safety test procedure for average CONTACT SURFACE
TEMPERATURE for FORCED AIR DEVICES . 65
Bibliography . 67
Index of defined terms used in this document . 69
Figure 201.101 – Positioning of temperature sensors on the contact surface of the
heated area of a HEATING DEVICE (see 201.12.4.101 and 201.12.4.105) . 11
IEC 60601-2-35:2020 © IEC 2020 – 3 –
Figure 201.102 – Example of the positioning of temperature sensors on the contact
surface of the heated areas of a HEATING DEVICE having more than one separately
heated area . 12
Figure 201.103 – Apparatus for the spark ignition test . 22
Figure 201.104 – Ramp for the impact test on PADS . 24
Figure 201.105 – Partial covering conditions . 26
Figure 201.106 – Method of folding BLANKETS . 35
Figure 201.107 – Examples of folds . 37
Figure 201.108 – Positions of a BLANKET for the RUCK-RESISTANCE test. 42
Figure AA.1 – Illustration of the main requirements of this document . 45
Figure HH.1 – Sensor locations – Average CONTACT SURFACE TEMPERATURE . 66
Table 201.101 – * Additional ESSENTIAL PERFORMANCE requirements . 14
Table 201.102 – Temperature limits in dependency to time . 39
– 4 – IEC 60601-2-35:2020 © IEC 2020
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-35: Particular requirements for the basic safety
and essential performance of heating devices using blankets,
pads or mattresses and intended for heating in medical use
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC
Publication(s)"). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60601-2-35 has been prepared by IEC technical committee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This second edition cancels and replaces IEC 80601-2-35 published in 2009 and
Amendment 1:2016.
IEC 60601-2-35:2020 © IEC 2020 – 5 –
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/1765/FDIS 62D/1777/RVD
Full information on the voting for the approval of this International Standard can be found in
the report on voting indicated in the above table.
This document has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this document, the following print types are used:
– requirements and definitions: roman type;
– test specifications: italic type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this document, the term
– "clause" means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– "subclause" means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this document are preceded by the term "Clause" followed by
the clause number. References to subclauses within this particular standard are by number
only.
In this document, the conjunctive "or" is used as an "inclusive or" so a statement is true if any
combination of the conditions is true.
The verbal forms used in this document conform to usage described in Clause 7 of the
ISO/IEC Directives, Part 2. For the purposes of this document, the auxiliary verb:
– "shall" means that compliance with a requirement or a test is mandatory for compliance
with this document;
– "should" means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this document;
– "may" is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the 60601 International Standard, published under the general title
Medical electrical equipment, can be found on the IEC website.
– 6 – IEC 60601-2-35:2020 © IEC 2020
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under "http://webstore.iec.ch" in the data related to
the specific document. At this date, the document will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
NOTE The attention of users of this document is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC publication in
which to make products in accordance with the new requirements and to equip themselves for conducting new or
revised tests. It is the recommendation of the committees that the content of this publication be adopted for
implementation nationally not earlier than 3 years from the date of publication.
IMPORTANT – The "colour inside" logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct understanding
of its contents. Users should therefore print this publication using a colour printer.
IEC 60601-2-35:2020 © IEC 2020 – 7 –
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation for HEATING DEVICES using BLANKETS,
PADS or MATTRESSES and intended for heating in medical use.
While K (degree Kelvin) is the recognized unit and symbol for absolute temperature and
temperature difference, °C has been used throughout this document because all
measurements are commonly made using equipment marked with the Celsius temperature
scale.
– 8 – IEC 60601-2-35:2020 © IEC 2020
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-35: Particular requirements for the basic safety
and essential performance of heating devices using blankets,
pads or mattresses and intended for heating in medical use
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 * Scope
Replacement:
This part of 60601 International Standard applies to the BASIC SAFETY and ESSENTIAL
PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also
referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the
scope of this document.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
If a clause or subclause is specifically intended to be applicable to a specifically defined type
of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that
clause or subclause will say so. If that is not the case, the clause or subclause applies both to
ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this document are not covered by specific requirements in this document,
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This document does not apply to:
– HEATING DEVICES intended for physiotherapy;
– INFANT RADIANT WARMERS; for information, see IEC 60601-2-21 [1] ;
– INFANT INCUBATORS; for information, see IEC 60601-2-19 [2];
– INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20 [3];
– cooling devices.
_____________
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance.
Figures in square brackets refer to the Bibliography.
IEC 60601-2-35:2020 © IEC 2020 – 9 –
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements, which minimize HAZARDS to PATIENTS, and OPERATORS for
HEATING DEVICES using BLANKETS, PADS or MATTRESSES and intended for heating in medical
use and to specify tests for demonstrating compliance with these requirements.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2:2014, IEC 60601-1-8:2006 and IEC 60601-1-8:2006/AMD1:2012, and
IEC 60601-1-10:2007 and IEC 60601-1-10:2007/AMD1:2013 apply as modified in Articles 202,
208 and 210 respectively. IEC 60601-1-3 does not apply. All other published collateral
standards in the IEC 60601-1 series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012 are referred to in this
particular standard as the general standard. Collateral standards are referred to by their
document number.
The numbering of clauses and subclauses of this document corresponds to that of the general
standard with the prefix "201" (e.g. 201.1 in this document addresses the content of Clause 1
of the general standard) or applicable collateral standard with the prefix "20x" where x is the
final digit(s) of the collateral standard document number (e.g. 202.4 in this document
addresses the content of Clause 4 of the IEC 60601-1-2 collateral standard, 203.4 in this
document addresses the content of Clause 4 of the IEC 60601-1-3 collateral standard, etc.).
The changes to the text of the general standard and applicable collateral standards are
specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
– 10 – IEC 60601-2-35:2020 © IEC 2020
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.147, additional definitions in this document are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where "x" is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the Bibliography.
Clause 2 of the general standard applies, except as follows:
Addition:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for the
development of physiologic closed-loop controllers
IEC 60601-1-10:2007/AMD1:2013
Replacement:
IEC 60384-14:2013, Fixed capacitors for use in electronic equipment – Part 14: Sectional
specification – Fixed capacitors for electromagneticinterference suppression and connection
to the supply mains
IEC 60384-14:2013/AMD1:2016
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic disturbances –
Requirements and tests
201.3 Terms and definitions
For the purposes of this document, the terms and definitions specified in IEC 60601-1:2005
and IEC 60601-1:2005/AMD1:2012, and the following apply.
IEC 60601-2-35:2020 © IEC 2020 – 11 –
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
NOTE An index of defined terms used in this document is found beginning on page 69.
Addition:
201.3.201.1
BLANKET
APPLIED PART of HEATING DEVICE, which can be folded, for
use under or over a PATIENT
201.3.201.2
BLANKET
APPLIED PART of HEATING DEVICE intended to be used with a CONTROLLER
PATIENT
to transfer thermal energy to all or part of the body of a
201.3.202
CONDITIONS OF ADEQUATE HEAT DISCHARGE
conditions achieved when a HEATING DEVICE is supported and covered as specified in
Annex EE
201.3.203.1
CONTACT SURFACE TEMPERATURE
temperature T at the reference point of the heated APPLIED
R
PART
Note 1 to entry: See Figures 201.101 and 201.102.
Note 2 to entry: The CONTACT SURFACE TEMPERATURE for FORCED AIR DEVICES is measured by the test methods
described in Annexes FF, GG and HH.
Key
T CONTACT SURFACE TEMPERATURE reference point on the contact surface
R
Some HEATING DEVICES may have unheated areas shown in the following figures as:
Figure 201.101 – Positioning of temperature sensors on the contact surface
of the heated area of a HEATING DEVICE
(see 201.12.4.101 and 201.12.4.105)
– 12 – IEC 60601-2-35:2020 © IEC 2020
The temperature at the centre point of any one of the heated areas closest to the centre of the HEATING DEVICE
(in the example shown above 2, 3, 6, or 7) is treated as T .
R
Figure 201.102 – Example of the positioning of temperature sensors
on the contact surface of the heated areas of a HEATING DEVICE having
more than one separately heated area
201.3.203.2
CONTACT SURFACE TEMPERATURE
< FORCED AIR DEVICES> temperature resulting from the heat transferred to a target surface by
the APPLIED PART
201.3.204
CONTROLLER
part of a HEATING DEVICE intended to supply and control thermal energy to a BLANKET, PAD or
MATTRESS
Note 1 to entry: This includes the HOSE, if present.
201.3.205
FORCED AIR DEVICE
HEATING DEVICE that uses air as the heat transfer medium to warm a PATIENT and is comprised
CONTROLLER and a BLANKET
of a
201.3.206
FREE HOSING
hazardous practice or condition of using the CONTROLLER without a BLANKET
201.3.207
HEATING DEVICE
ME EQUIPMENT intended to supply heat to the whole or part of the body of a PATIENT by means
of heated BLANKETS, PADS, or MATTRESSES
201.3.208
HIGH HEAT TRANSFER
thermal characteristic of a HEATING DEVICE as determined according to Annex CC or Annex DD
201.3.209
HOSE
component of the CONTROLLER that is the conduit for the heat transfer medium to and/or from
the BLANKET, PAD or MATTRESS
IEC 60601-2-35:2020 © IEC 2020 – 13 –
201.3.210
INFANT
PATIENT up to the age of three months and with a weight less than 10 kg
201.3.211
LAGGING MATERIAL
polyurethane or polystyrene insulation material used in the test methods of this specification
to assist in the determination of temperature
Note 1 to entry: Specifications for LAGGING MATERIAL are given in Annexes BB and FF.
201.3.212
LOW HEAT TRANSFER
thermal characteristic of a HEATING DEVICE as determined according to Annex CC or Annex DD
201.3.213
MATTRESS
APPLIED PART of a HEATING DEVICE, which provides resilient support to the whole body of a
PATIENT
201.3.214
NOZZLE
end of the HOSE that connects to the BLANKET, PAD or MATTRESS
201.3.215
OVER-BLANKET
BLANKET designed to be used over a PATIENT
201.3.216
PAD
APPLIED PART of HEATING DEVICE, which can be bent but not folded
201.3.217
RUCK
unintended fold in a normally even surface
201.3.218
RUCK-RESISTANT BLANKET
BLANKET having a construction such that RUCKING of the flexible part is unlikely
201.3.219
UNDER-BLANKET
BLANKET designed to be used under a PATIENT
201.4 General requirements
Clause 4 of the general standard applies except as follows:
201.4.3 ESSENTIAL PERFORMANCE
Addition:
201.4.3.101 Additional requirements for ESSENTIAL PERFORMANCE
Additional ESSENTIAL PERFORMANCE requirements are found in the subclauses listed in
Table 201.101.
– 14 – IEC 60601-2-35:2020 © IEC 2020
Table 201.101 – * Additional ESSENTIAL PERFORMANCE requirements
Requirement Subclause
ESSENTIAL PERFORMANCE requirement 1 201.12.4.104 or generation of a TECHNICAL ALARM
CONDITION in compliance with 201.12.3.102
201.4.5 Alternative RISK CONTROL measures or test methods for me equipment or me
systems
Addition:
This particular standard specifies safety requirements for HEATING DEVICES using BLANKETS,
PADS or MATTRESSES, but alternate methods of compliance with a specific clause or subclause
by demonstrating equivalent safety will not be judged non-compliant if the MANUFACTURER has
demonstrated in his RISK MANAGEMENT FILE that the RISKS presented by the HAZARDS are of an
acceptable level when weighed against the benefits of treatment using the device.
Additional subclause:
201.4.101 Combination of equipment
For equipment which combines several heat sources, the safety requirements of other
relevant particular standards shall be considered. Further, the safety requirements of this
particular standard shall be fulfilled with the combination of the other equipment, which is
approved by the MANUFACTURER as stated in the instructions for use according to Clause 16 of
the general standard (ME SYSTEMS).
201.5 General requirements for testing ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies, except as follows:
201.6.2 Protection against electric shock
Replacement of the last paragraph with the following new paragraph:
HEATING DEVICES shall have TYPE BF APPLIED PARTS or TYPE CF APPLIED PARTS.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
IEC 60601-2-35:2020 © IEC 2020 – 15 –
201.7.2.1 Minimum requirements for marking on ME EQUIPMENT and on
interchangeable parts
Addition:
201.7.2.1.101 * Additional minimum requirements for marking on ME EQUIPMENT and
on interchangeable parts
201.7.2.1.101.1 HEATING DEVICES (other than for FORCED AIR DEVICES)
A HEATING DEVICE shall be marked as follows:
NORMAL USE, whether:
a) to indicate how it is intended to be positioned in
– over the PATIENT;
PATIENT;
– under the
– directly in contact with the PATIENT;
PATIENT by an intermediate layer, or layers (for example by a water-
– separated from the
bed, other type of MATTRESS, or bedclothes);
– it has to be used flat (that is without creases), or whether it can be wrapped around the
PATIENT;
b) to warn against possible HAZARDOUS SITUATIONS from penetration by sharp objects;
c) to warn against possible HAZARDOUS SITUATIONS from folding or methods of storage, other
MANUFACTURER;
than those specified by the
d) to warn against the possibility of a HAZARDOUS SITUATION if a partial covering is caused by
pillows or other items having good thermal insulation being laid over part of the HEATING
DEVICE;
e) to specify, in the case of liquid-filled MATTRESSES intended to be used above a PAD, the
minimum amount of liquid to which the MATTRESS should be filled, and a warning that if this
minimum amount is not present a PATIENT could suffer a burn;
f) to specify, in the case of a HEATING DEVICE supplied or controlled by an external unit, that
the HEATING DEVICE shall only be used with the external unit specified by the
MANUFACTURER of the HEATING DEVICE.
201.7.2.1.101.2 CONTROLLERS
A CONTROLLER for FORCED AIR DEVICES shall be marked as follows.
HOSE shall be marked within 15 cm of the NOZZLE to caution that the NOZZLE needs to
a) The
be connected to a BLANKET. The following statement and the safety sign ISO 7010-M002
(see IEC 60601-1:2005, Table D.2, safety sign 10) shall accompany the "no FREE HOSING"
safety sign shown in Annex D of this particular standard:
"CAUTION! HOSE NOZZLE SHALL be connected to a compatible forced air BLANKET or
thermal injury may occur."
b) A caution that allowing the HOSE to contact the PATIENT can lead to thermal injury, if
appropriate.
c) A warning against using the device distal to arterial cross clamping and that non-
observance can lead to thermal injury.
201.7.2.1.101.3 * Temperature sensors
A temperature sensor which is designed to be attached to or inserted into a PATIENT shall
have its intended use identified clearly and unambiguously on or adjacent to the sensor.
– 16 – IEC 60601-2-35:2020 © IEC 2020
201.7.2.1.101.4 APPLIED PARTS (for other than FORCED AIR DEVICES) with a large
unheated border
For an APPLIED PART having an unheated border wider than 30 mm around the heated area,
and where the requirements of 201.12.4.101 are not satisfied, the outer boundary of the
heated area shall be marked on both sides of the APPLIED PART.
201.7.4.2 Control devices
Addition:
201.7.4.2.101 Additional requirements for control devices
NOTE See also 201.12.
Where provided, on ME EQUIPMENT other than FORCED AIR DEVICES, a control for setting the
CONTACT SURFACE TEMPERATURE of a HEATING DEVICE shall indicate the temperature in intervals
not greater than 1 °C (see also 201.12.1.101).
For FORCED AIR DEVICES, each heated temperature control position shall be marked in
degrees C. Such marking shall be CLEARLY LEGIBLE.
201.7.9.2.2 Warning and safety notices
Addition:
201.7.9.2.2.101 Additional requirements for warning and safety notices
The instructions for use shall additionally contain the following:
a) a strong recommendation that the surface of the HEATING DEVICE should be checked for
freedom from mechanical damage prior to each application;
b) an indication for the use of parts of HEATING DEVICES which are intended to be used
together;
c) statements, details and warnings on the use of the HEATING DEVICE in combination with
other heat sources, if applicable;
d) a warning statement that the use of materials of good thermal conductivity, such as water,
gel and similar substances, with the HEATING DEVICE not switched on can decrease the
temperature of the body of a PATIENT;
e) a warning statement regarding the RISK of electrical shock, burns or electromagnetic
interference with use of high frequency (HF) surgical instruments or endocardial catheters
while a HEATING DEVICE is in use, if applicable;
f) a statement that the OPERATOR should monitor the temperature of the PATIENT at regular
intervals;
g) a statement that means for drainage of liquid from a liquid-filled HEATING DEVICE is required
(see also 201.11.6.5.101);
h) a warning that a means for retaining a PATIENT either on or under a HEATING DEVICE may
need to be used and that the means for retaining a PATIENT should not block the fluid
pathways of the HEATING DEVICE;
i) a warning that warming transdermal medications (patches) can increase drug delivery,
resulting in possible HARM to the PATIENT;
j) a statement that the HEATING DEVICE contains an ALARM SYSTEM with an interruption of
power supply/SUPPLY MAINS ALARM CONDITION;
k) for CONTROLLERS for FORCED AIR DEVICES, the following warning statements:
– a caution that the HOSE, if allowed to contact the PATIENT, can lead to thermal injury, if
appropriate;
IEC 60601-2-35:2020 © IEC 2020 – 17 –
– a description of system operating modes and PATIENT conditions where the HEATING
DEVICE can be safely used;
– a statement that the OPERATOR should monitor the temperature of the PATIENT at
regular intervals;
– * a statement that the HOSE NOZZLE needs to be connected to a BLANKET. The following
statement shall accompany the "no FREE HOSING" safety sign shown in Annex D:
"CAUTION! HOSE NOZZLE SHALL be connected to a compatible forced air BLANKET or
thermal injury may occur."
201.7.9.2.9 Operating instructions
Addition:
201.7.9.2.9.101 Additional requirements for start-up PROCEDURE
The instructions for use shall include a method for testing the function of the ALARM SYSTEM for
each of the ALARM CONDITIONS specified in this document, if not performed automatically
during start up.
201.7.9.2.9.102 Additional requirements for operating instructions
The following shall appear in the instructions for use:
a) the approximate time required for the CONTACT SURFACE TEMPERATURE to heat up from
23 °C ± 2 °C to 37 °C, when operated under CONDITIONS OF ADEQUATE HEAT DISCHARGE as
specified in Annex EE;
ALARM SYSTEM.
b) a description of how and when to verify the functionality of the
201.7.9.2.9.103 Additional requirements for operating instructions for BLANKETS
The instructions for use for BLANKETS shall contain the following:
a) a description of system operating modes and PATIENT conditions where the system can be
safely used;
b) a statement that the OPERATOR should monitor the temperature of the PATIENT at regular
intervals;
c) an indication of how the BLANKET is to be positioned during NORMAL USE, for example:
i) over the PATIENT,
ii) under the PATIENT,
iii) directly in contact with the PATIENT,
iv) separated from the PATIENT by an intermediate layer, or layers, of materi
...
SLOVENSKI STANDARD
SIST EN IEC 80601-2-35:2021
01-september-2021
Nadomešča:
SIST EN 80601-2-35:2010
SIST EN 80601-2-35:2010/A1:2017
SIST EN 80601-2-35:2010/A11:2012
SIST EN 80601-2-35:2010/AC:2015
Medicinska električna oprema - 2-35. del: Posebne zahteve za osnovno varnost in
bistvene lastnosti za odeje, blazine in posteljne vložke, namenjene za ogrevanje
pri medicinski uporabi (IEC 60601-2-35:2020)
Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and
essential performance of heating devices using blankets, pads and mattresses and
intended for heating in medical use (IEC 60601-2-35:2020)
Medizinische elektrische Geräte - Teil 2-35: Besondere Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale von Decken, Matten und Matratzen
zur Erwärmung von Patienten in der medizinischen Anwendung (IEC 60601-2-35:2020)
Appareils électromédicaux - Partie 2-35: Exigences particulières pour la sécurité de base
et les performances essentielles des dispositifs de réchauffage utilisant des couvertures,
des coussins ou des matelas chauffants et destinés au réchauffage des patients en
usage médical (IEC 60601-2-35:2020)
Ta slovenski standard je istoveten z: EN IEC 60601-2-35:2021
ICS:
11.140 Oprema bolnišnic Hospital equipment
SIST EN IEC 80601-2-35:2021 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
SIST EN IEC 80601-2-35:2021
SIST EN IEC 80601-2-35:2021
EUROPEAN STANDARD EN IEC 60601-2-35
NORME EUROPÉENNE
EUROPÄISCHE NORM
July 2021
ICS 11.140; 11.040.01 Supersedes EN 80601-2-35:2009 and all of its
amendments and corrigenda (if any)
English Version
Medical electrical equipment - Part 2-35: Particular requirements
for the basic safety and essential performance of heating
devices using blankets, pads and mattresses and intended for
heating in medical use
(IEC 60601-2-35:2020)
Appareils électromédicaux - Partie 2-35: Exigences Medizinische elektrische Geräte - Teil 2-35: Besondere
particulières pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles des dispositifs de réchauffage utilisant des wesentlichen Leistungsmerkmale von Decken, Matten und
couvertures, des coussins ou des matelas chauffants et Matratzen zur Erwärmung von Patienten in der
destinés au réchauffage des patients en usage médical medizinischen Anwendung
(IEC 60601-2-35:2020) (IEC 60601-2-35:2020)
This European Standard was approved by CENELEC on 2020-10-26. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 60601-2-35:2021 E
SIST EN IEC 80601-2-35:2021
European foreword
The text of document 62D/1765/FDIS, future edition 2 of IEC 60601-2-35, prepared by SC 62D
"Electromedical equipment" of IEC/TC 62 "Electrical equipment in medical practice" was submitted to
the IEC-CENELEC parallel vote and approved by CENELEC as EN IEC 60601-2-35:2021.
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2022-01-16
level by publication of an identical national
standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2024-07-16
document have to be withdrawn
This document supersedes EN 80601-2-35:2009 and all of its amendments and corrigenda (if any).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
Endorsement notice
The text of the International Standard IEC 60601-2-35:2020 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards
indicated:
IEC 60601-2-19:2020 NOTE Harmonized as EN IEC 60601-2-19:2021 (not modified)
IEC 60601-2-20:2020 NOTE Harmonized as EN IEC 60601-2-20:2020 (not modified)
IEC 60601-2-21:2020 NOTE Harmonized as EN IEC 60601-2-21:2020 (not modified)
IEC 60335-2-53 NOTE Harmonized as EN 60335-2-53
IEC 60529:1989 NOTE Harmonized as EN 60529:1991 (not modified)
IEC 60529:1989/A1:1999 NOTE Harmonized as EN 60529:1991/A1:2000 (not modified)
IEC 60529:1989/A2:2013 NOTE Harmonized as EN 60529:1991/A2:2013 (not modified)
ISO 2439:2008 NOTE Harmonized as EN ISO 2439:2008 (not modified)
SIST EN IEC 80601-2-35:2021
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1 Where an International Publication has been modified by common modifications, indicated by (mod),
the relevant EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is
available here: www.cenelec.eu.
The Annex ZA of EN 60601-1:2006/A1:2013 applies with the following additions:
Publication Year Title EN/HD Year
IEC 60601-1 2005 Medical electrical equipment - Part 1: EN 60601-1 2006
General requirements for basic safety and
essential performance
- - + corrigendum Mar. 2010
+ A1 2012 + A1 2013
- - + A12 2014
IEC 60601-1-10 2007 Medical electrical equipment - Part 1-10: EN 60601-1-10 2008
General requirements for basic safety and
essential performance - Collateral
Standard: Requirements for the
development of physiologic closed-loop
controllers
+ A1 2013 + A1 2015
The Annex ZA of EN 60601-1:2006/A1:2013 applies with the following replacements:
IEC 60384-14 2013 Fixed capacitors for use in electronic EN 60384-14 2013
equipment - Part 14: Sectional
specification - Fixed capacitors for
electromagnetic interference suppression
and connection to the supply mains
+ A1 2016 + A1 2016
IEC 60601-1-2 2014 Medical electrical equipment - Part 1-2: EN 60601-1-2 2015
General requirements for basic safety and
essential performance - Collateral
Standard: Electromagnetic disturbances -
Requirements and tests
SIST EN IEC 80601-2-35:2021
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35 ®
Edition 2.0 2020-09
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-35: Particular requirements for the basic safety and essential performance
of heating devices using blankets, pads or mattresses and intended for heating
in medical use
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 11.040.01; 11.140 ISBN 978-2-8322-8723-1
SIST EN IEC 80601-2-35:2021
– 2 – IEC 60601-2-35:2020 © IEC 2020
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terms and definitions. 10
201.4 General requirements . 13
201.5 General requirements for testing ME EQUIPMENT . 14
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 14
201.7 ME EQUIPMENT identification, marking and documents . 14
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 18
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 24
201.10 Protection against unwanted and excessive radiation HAZARDS . 26
201.11 Protection against excessive temperatures and other HAZARDS . 26
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 28
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 33
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 38
201.15 Construction of ME EQUIPMENT . 38
201.16 ME SYSTEMS . 42
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 42
202 ELECTROMAGNETIC DISTURBANCES – Requirements and tests . 43
208 General requirements, tests and guidance for ALARM SYSTEMS IN MEDICAL
ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS . 43
210 * Requirements for the development of PHYSIOLOGIC CLOSED-LOOP CONTROLLERS . 43
Annexes . 44
Annex D (informative) Symbols on marking . 44
Annex AA (informative) Particular guidance and rationale . 45
Annex BB (normative) Determination of the LAGGING MATERIAL . 57
Annex CC (normative) * Determination of heat transfer towards the PATIENT . 58
Annex DD (normative) * Determination of heat transfer away from the PATIENT . 60
Annex EE (normative) CONDITIONS OF ADEQUATE HEAT DISCHARGE . 61
Annex FF (normative) Test procedure for maximum CONTACT SURFACE TEMPERATURE
for FORCED AIR DEVICES . 62
Annex GG (normative) Test procedure for maximum CONTACT SURFACE TEMPERATURE
for FORCED AIR DEVICES under SINGLE FAULT CONDITION. 64
Annex HH (normative) Safety test procedure for average CONTACT SURFACE
TEMPERATURE for FORCED AIR DEVICES . 65
Bibliography . 67
Index of defined terms used in this document . 69
Figure 201.101 – Positioning of temperature sensors on the contact surface of the
heated area of a HEATING DEVICE (see 201.12.4.101 and 201.12.4.105) . 11
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 3 –
Figure 201.102 – Example of the positioning of temperature sensors on the contact
surface of the heated areas of a HEATING DEVICE having more than one separately
heated area . 12
Figure 201.103 – Apparatus for the spark ignition test . 22
Figure 201.104 – Ramp for the impact test on PADS . 24
Figure 201.105 – Partial covering conditions . 26
Figure 201.106 – Method of folding BLANKETS . 35
Figure 201.107 – Examples of folds . 37
Figure 201.108 – Positions of a BLANKET for the RUCK-RESISTANCE test. 42
Figure AA.1 – Illustration of the main requirements of this document . 45
Figure HH.1 – Sensor locations – Average CONTACT SURFACE TEMPERATURE . 66
Table 201.101 – * Additional ESSENTIAL PERFORMANCE requirements . 14
Table 201.102 – Temperature limits in dependency to time . 39
SIST EN IEC 80601-2-35:2021
– 4 – IEC 60601-2-35:2020 © IEC 2020
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-35: Particular requirements for the basic safety
and essential performance of heating devices using blankets,
pads or mattresses and intended for heating in medical use
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC
Publication(s)"). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60601-2-35 has been prepared by IEC technical committee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This second edition cancels and replaces IEC 80601-2-35 published in 2009 and
Amendment 1:2016.
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 5 –
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/1765/FDIS 62D/1777/RVD
Full information on the voting for the approval of this International Standard can be found in
the report on voting indicated in the above table.
This document has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this document, the following print types are used:
– requirements and definitions: roman type;
– test specifications: italic type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this document, the term
– "clause" means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– "subclause" means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this document are preceded by the term "Clause" followed by
the clause number. References to subclauses within this particular standard are by number
only.
In this document, the conjunctive "or" is used as an "inclusive or" so a statement is true if any
combination of the conditions is true.
The verbal forms used in this document conform to usage described in Clause 7 of the
ISO/IEC Directives, Part 2. For the purposes of this document, the auxiliary verb:
– "shall" means that compliance with a requirement or a test is mandatory for compliance
with this document;
– "should" means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this document;
– "may" is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the 60601 International Standard, published under the general title
Medical electrical equipment, can be found on the IEC website.
SIST EN IEC 80601-2-35:2021
– 6 – IEC 60601-2-35:2020 © IEC 2020
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under "http://webstore.iec.ch" in the data related to
the specific document. At this date, the document will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
NOTE The attention of users of this document is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC publication in
which to make products in accordance with the new requirements and to equip themselves for conducting new or
revised tests. It is the recommendation of the committees that the content of this publication be adopted for
implementation nationally not earlier than 3 years from the date of publication.
IMPORTANT – The "colour inside" logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct understanding
of its contents. Users should therefore print this publication using a colour printer.
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 7 –
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation for HEATING DEVICES using BLANKETS,
PADS or MATTRESSES and intended for heating in medical use.
While K (degree Kelvin) is the recognized unit and symbol for absolute temperature and
temperature difference, °C has been used throughout this document because all
measurements are commonly made using equipment marked with the Celsius temperature
scale.
SIST EN IEC 80601-2-35:2021
– 8 – IEC 60601-2-35:2020 © IEC 2020
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-35: Particular requirements for the basic safety
and essential performance of heating devices using blankets,
pads or mattresses and intended for heating in medical use
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 * Scope
Replacement:
This part of 60601 International Standard applies to the BASIC SAFETY and ESSENTIAL
PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also
referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the
scope of this document.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
If a clause or subclause is specifically intended to be applicable to a specifically defined type
of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that
clause or subclause will say so. If that is not the case, the clause or subclause applies both to
ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this document are not covered by specific requirements in this document,
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This document does not apply to:
– HEATING DEVICES intended for physiotherapy;
– INFANT RADIANT WARMERS; for information, see IEC 60601-2-21 [1] ;
– INFANT INCUBATORS; for information, see IEC 60601-2-19 [2];
– INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20 [3];
– cooling devices.
_____________
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance.
Figures in square brackets refer to the Bibliography.
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 9 –
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements, which minimize HAZARDS to PATIENTS, and OPERATORS for
HEATING DEVICES using BLANKETS, PADS or MATTRESSES and intended for heating in medical
use and to specify tests for demonstrating compliance with these requirements.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2:2014, IEC 60601-1-8:2006 and IEC 60601-1-8:2006/AMD1:2012, and
IEC 60601-1-10:2007 and IEC 60601-1-10:2007/AMD1:2013 apply as modified in Articles 202,
208 and 210 respectively. IEC 60601-1-3 does not apply. All other published collateral
standards in the IEC 60601-1 series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012 are referred to in this
particular standard as the general standard. Collateral standards are referred to by their
document number.
The numbering of clauses and subclauses of this document corresponds to that of the general
standard with the prefix "201" (e.g. 201.1 in this document addresses the content of Clause 1
of the general standard) or applicable collateral standard with the prefix "20x" where x is the
final digit(s) of the collateral standard document number (e.g. 202.4 in this document
addresses the content of Clause 4 of the IEC 60601-1-2 collateral standard, 203.4 in this
document addresses the content of Clause 4 of the IEC 60601-1-3 collateral standard, etc.).
The changes to the text of the general standard and applicable collateral standards are
specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
SIST EN IEC 80601-2-35:2021
– 10 – IEC 60601-2-35:2020 © IEC 2020
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.147, additional definitions in this document are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where "x" is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the Bibliography.
Clause 2 of the general standard applies, except as follows:
Addition:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for the
development of physiologic closed-loop controllers
IEC 60601-1-10:2007/AMD1:2013
Replacement:
IEC 60384-14:2013, Fixed capacitors for use in electronic equipment – Part 14: Sectional
specification – Fixed capacitors for electromagneticinterference suppression and connection
to the supply mains
IEC 60384-14:2013/AMD1:2016
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic disturbances –
Requirements and tests
201.3 Terms and definitions
For the purposes of this document, the terms and definitions specified in IEC 60601-1:2005
and IEC 60601-1:2005/AMD1:2012, and the following apply.
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 11 –
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
NOTE An index of defined terms used in this document is found beginning on page 69.
Addition:
201.3.201.1
BLANKET
APPLIED PART of HEATING DEVICE, which can be folded, for
use under or over a PATIENT
201.3.201.2
BLANKET
APPLIED PART of HEATING DEVICE intended to be used with a CONTROLLER
PATIENT
to transfer thermal energy to all or part of the body of a
201.3.202
CONDITIONS OF ADEQUATE HEAT DISCHARGE
conditions achieved when a HEATING DEVICE is supported and covered as specified in
Annex EE
201.3.203.1
CONTACT SURFACE TEMPERATURE
temperature T at the reference point of the heated APPLIED
R
PART
Note 1 to entry: See Figures 201.101 and 201.102.
Note 2 to entry: The CONTACT SURFACE TEMPERATURE for FORCED AIR DEVICES is measured by the test methods
described in Annexes FF, GG and HH.
Key
T CONTACT SURFACE TEMPERATURE reference point on the contact surface
R
Some HEATING DEVICES may have unheated areas shown in the following figures as:
Figure 201.101 – Positioning of temperature sensors on the contact surface
of the heated area of a HEATING DEVICE
(see 201.12.4.101 and 201.12.4.105)
SIST EN IEC 80601-2-35:2021
– 12 – IEC 60601-2-35:2020 © IEC 2020
The temperature at the centre point of any one of the heated areas closest to the centre of the HEATING DEVICE
(in the example shown above 2, 3, 6, or 7) is treated as T .
R
Figure 201.102 – Example of the positioning of temperature sensors
on the contact surface of the heated areas of a HEATING DEVICE having
more than one separately heated area
201.3.203.2
CONTACT SURFACE TEMPERATURE
< FORCED AIR DEVICES> temperature resulting from the heat transferred to a target surface by
the APPLIED PART
201.3.204
CONTROLLER
part of a HEATING DEVICE intended to supply and control thermal energy to a BLANKET, PAD or
MATTRESS
Note 1 to entry: This includes the HOSE, if present.
201.3.205
FORCED AIR DEVICE
HEATING DEVICE that uses air as the heat transfer medium to warm a PATIENT and is comprised
CONTROLLER and a BLANKET
of a
201.3.206
FREE HOSING
hazardous practice or condition of using the CONTROLLER without a BLANKET
201.3.207
HEATING DEVICE
ME EQUIPMENT intended to supply heat to the whole or part of the body of a PATIENT by means
of heated BLANKETS, PADS, or MATTRESSES
201.3.208
HIGH HEAT TRANSFER
thermal characteristic of a HEATING DEVICE as determined according to Annex CC or Annex DD
201.3.209
HOSE
component of the CONTROLLER that is the conduit for the heat transfer medium to and/or from
the BLANKET, PAD or MATTRESS
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 13 –
201.3.210
INFANT
PATIENT up to the age of three months and with a weight less than 10 kg
201.3.211
LAGGING MATERIAL
polyurethane or polystyrene insulation material used in the test methods of this specification
to assist in the determination of temperature
Note 1 to entry: Specifications for LAGGING MATERIAL are given in Annexes BB and FF.
201.3.212
LOW HEAT TRANSFER
thermal characteristic of a HEATING DEVICE as determined according to Annex CC or Annex DD
201.3.213
MATTRESS
APPLIED PART of a HEATING DEVICE, which provides resilient support to the whole body of a
PATIENT
201.3.214
NOZZLE
end of the HOSE that connects to the BLANKET, PAD or MATTRESS
201.3.215
OVER-BLANKET
BLANKET designed to be used over a PATIENT
201.3.216
PAD
APPLIED PART of HEATING DEVICE, which can be bent but not folded
201.3.217
RUCK
unintended fold in a normally even surface
201.3.218
RUCK-RESISTANT BLANKET
BLANKET having a construction such that RUCKING of the flexible part is unlikely
201.3.219
UNDER-BLANKET
BLANKET designed to be used under a PATIENT
201.4 General requirements
Clause 4 of the general standard applies except as follows:
201.4.3 ESSENTIAL PERFORMANCE
Addition:
201.4.3.101 Additional requirements for ESSENTIAL PERFORMANCE
Additional ESSENTIAL PERFORMANCE requirements are found in the subclauses listed in
Table 201.101.
SIST EN IEC 80601-2-35:2021
– 14 – IEC 60601-2-35:2020 © IEC 2020
Table 201.101 – * Additional ESSENTIAL PERFORMANCE requirements
Requirement Subclause
ESSENTIAL PERFORMANCE requirement 1 201.12.4.104 or generation of a TECHNICAL ALARM
CONDITION in compliance with 201.12.3.102
201.4.5 Alternative RISK CONTROL measures or test methods for me equipment or me
systems
Addition:
This particular standard specifies safety requirements for HEATING DEVICES using BLANKETS,
PADS or MATTRESSES, but alternate methods of compliance with a specific clause or subclause
by demonstrating equivalent safety will not be judged non-compliant if the MANUFACTURER has
demonstrated in his RISK MANAGEMENT FILE that the RISKS presented by the HAZARDS are of an
acceptable level when weighed against the benefits of treatment using the device.
Additional subclause:
201.4.101 Combination of equipment
For equipment which combines several heat sources, the safety requirements of other
relevant particular standards shall be considered. Further, the safety requirements of this
particular standard shall be fulfilled with the combination of the other equipment, which is
approved by the MANUFACTURER as stated in the instructions for use according to Clause 16 of
the general standard (ME SYSTEMS).
201.5 General requirements for testing ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies, except as follows:
201.6.2 Protection against electric shock
Replacement of the last paragraph with the following new paragraph:
HEATING DEVICES shall have TYPE BF APPLIED PARTS or TYPE CF APPLIED PARTS.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 15 –
201.7.2.1 Minimum requirements for marking on ME EQUIPMENT and on
interchangeable parts
Addition:
201.7.2.1.101 * Additional minimum requirements for marking on ME EQUIPMENT and
on interchangeable parts
201.7.2.1.101.1 HEATING DEVICES (other than for FORCED AIR DEVICES)
A HEATING DEVICE shall be marked as follows:
NORMAL USE, whether:
a) to indicate how it is intended to be positioned in
– over the PATIENT;
PATIENT;
– under the
– directly in contact with the PATIENT;
PATIENT by an intermediate layer, or layers (for example by a water-
– separated from the
bed, other type of MATTRESS, or bedclothes);
– it has to be used flat (that is without creases), or whether it can be wrapped around the
PATIENT;
b) to warn against possible HAZARDOUS SITUATIONS from penetration by sharp objects;
c) to warn against possible HAZARDOUS SITUATIONS from folding or methods of storage, other
MANUFACTURER;
than those specified by the
d) to warn against the possibility of a HAZARDOUS SITUATION if a partial covering is caused by
pillows or other items having good thermal insulation being laid over part of the HEATING
DEVICE;
e) to specify, in the case of liquid-filled MATTRESSES intended to be used above a PAD, the
minimum amount of liquid to which the MATTRESS should be filled, and a warning that if this
minimum amount is not present a PATIENT could suffer a burn;
f) to specify, in the case of a HEATING DEVICE supplied or controlled by an external unit, that
the HEATING DEVICE shall only be used with the external unit specified by the
MANUFACTURER of the HEATING DEVICE.
201.7.2.1.101.2 CONTROLLERS
A CONTROLLER for FORCED AIR DEVICES shall be marked as follows.
HOSE shall be marked within 15 cm of the NOZZLE to caution that the NOZZLE needs to
a) The
be connected to a BLANKET. The following statement and the safety sign ISO 7010-M002
(see IEC 60601-1:2005, Table D.2, safety sign 10) shall accompany the "no FREE HOSING"
safety sign shown in Annex D of this particular standard:
"CAUTION! HOSE NOZZLE SHALL be connected to a compatible forced air BLANKET or
thermal injury may occur."
b) A caution that allowing the HOSE to contact the PATIENT can lead to thermal injury, if
appropriate.
c) A warning against using the device distal to arterial cross clamping and that non-
observance can lead to thermal injury.
201.7.2.1.101.3 * Temperature sensors
A temperature sensor which is designed to be attached to or inserted into a PATIENT shall
have its intended use identified clearly and unambiguously on or adjacent to the sensor.
SIST EN IEC 80601-2-35:2021
– 16 – IEC 60601-2-35:2020 © IEC 2020
201.7.2.1.101.4 APPLIED PARTS (for other than FORCED AIR DEVICES) with a large
unheated border
For an APPLIED PART having an unheated border wider than 30 mm around the heated area,
and where the requirements of 201.12.4.101 are not satisfied, the outer boundary of the
heated area shall be marked on both sides of the APPLIED PART.
201.7.4.2 Control devices
Addition:
201.7.4.2.101 Additional requirements for control devices
NOTE See also 201.12.
Where provided, on ME EQUIPMENT other than FORCED AIR DEVICES, a control for setting the
CONTACT SURFACE TEMPERATURE of a HEATING DEVICE shall indicate the temperature in intervals
not greater than 1 °C (see also 201.12.1.101).
For FORCED AIR DEVICES, each heated temperature control position shall be marked in
degrees C. Such marking shall be CLEARLY LEGIBLE.
201.7.9.2.2 Warning and safety notices
Addition:
201.7.9.2.2.101 Additional requirements for warning and safety notices
The instructions for use shall additionally contain the following:
a) a strong recommendation that the surface of the HEATING DEVICE should be checked for
freedom from mechanical damage prior to each application;
b) an indication for the use of parts of HEATING DEVICES which are intended to be used
together;
c) statements, details and warnings on the use of the HEATING DEVICE in combination with
other heat sources, if applicable;
d) a warning statement that the use of materials of good thermal conductivity, such as water,
gel and similar substances, with the HEATING DEVICE not switched on can decrease the
temperature of the body of a PATIENT;
e) a warning statement regarding the RISK of electrical shock, burns or electromagnetic
interference with use of high frequency (HF) surgical instruments or endocardial catheters
while a HEATING DEVICE is in use, if applicable;
f) a statement that the OPERATOR should monitor the temperature of the PATIENT at regular
intervals;
g) a statement that means for drainage of liquid from a liquid-filled HEATING DEVICE is required
(see also 201.11.6.5.101);
h) a warning that a means for retaining a PATIENT either on or under a HEATING DEVICE may
need to be used and that the means for retaining a PATIENT should not block the fluid
pathways of the HEATING DEVICE;
i) a warning that warming transdermal medications (patches) can increase drug delivery,
resulting in possible HARM to the PATIENT;
j) a statement that the HEATING DEVICE contains an ALARM SYSTEM with an interruption of
power supply/SUPPLY MAINS ALARM CONDITION;
k) for CONTROLLERS for FORCED AIR DEVICES, the following warning statements:
– a caution that the HOSE, if allowed to contact the PATIENT, can lead to thermal injury, if
appropriate;
SIST EN IEC 80601-2-35:2021
IEC 60601-2-35:2020 © IEC 2020 – 17 –
– a description of system operating modes and PATIENT conditions where the HEATING
DEVICE can be safely used;
– a statement that the OPERATOR should monitor the temperature of the PATIENT at
regular intervals;
– * a statement that the HOSE NOZZLE needs to be connected to a BLANKET. The following
statement shall accompany the "no FREE HOSING" safety sign shown in Annex D:
"CAUTION! HOSE NOZZLE SHALL be connected to a compatible forced air BLANKET or
thermal injury may occur."
201.7.9.2.9 Operating instructions
Addition:
201.7.9.2.9.101 Additional requirements for start-up PROCEDURE
The instructions for use shall include a method for testing the function of the ALARM SYSTEM for
each of the ALARM CONDITIONS specified in this document, if not performed automatically
during start up.
201.7.9.2.9.102 Additional requirements for operating instructions
The following shall appear in the instructions for use:
a) the approximate time required for the CONTACT SURFACE TEMPERATURE to heat up from
23 °C ± 2 °C to 37 °C, when operated under CONDITIONS OF ADEQUATE HEAT DISCHARGE as
specified in Annex EE;
ALARM SYSTEM.
b) a description of how and when to verify the functionality of the
201.7.9.2.9.103 Additional requirements for operating instructions for BLANKETS
The instructions for use for BLANKETS shall contain the following:
a) a description of system operating modes and PATIENT conditions where the system can be
safely used;
b) a statement that the OPERATOR should monitor the temperature of the PATIENT at regular
intervals;
c) an indication of how the BLANKET is to be positioned during NORMAL USE, for example:
i) over the PATIENT,
ii) under the PATIENT,
iii) directly in contact with the PATIENT,
iv) separated from the PATIENT by an intermediate layer, or layers, of materi
...
The standard EN IEC 60601-2-35:2021 provides essential guidelines concerning the basic safety and essential performance of heating devices, specifically those using blankets, pads, and mattresses intended for medical applications. This document extends the scope of IEC 60601-2-35:2020 by ensuring that medical electrical equipment (ME EQUIPMENT) delivers safe and reliable thermal management for patients, especially in medical settings where maintaining body temperature is crucial. One of the primary strengths of this standard is its clear delineation of applicability, which ensures that users can easily determine whether the guidelines apply to different types of ME EQUIPMENT or ME SYSTEMS. As it outlines specific cases, such as forced air devices, the standard provides clarity on criteria for those operating within the medical domain, helping manufacturers and healthcare providers maintain compliance with best practices. Moreover, the standard emphasizes the importance of addressing hazards related to the physiological functions of ME EQUIPMENT, although it omits certain inherent risks unless explicitly stated in specific clauses. This focus on both safety and performance highlights the relevance of EN IEC 60601-2-35:2021 in promoting the safe use of heating devices in clinical environments. The inclusion of significant technical changes with respect to normative references from previous editions reaffirms the continual evolution of the standard, adapting to advancements in technology and changing medical practices. By superseding IEC 80601-2-35 published in 2009, this updated standard reinforces best safety practices and fosters confidence among healthcare practitioners who rely on proper temperature management for patient care. Overall, EN IEC 60601-2-35:2021 stands as a critical reference for manufacturers and healthcare providers, ensuring that heating devices utilized in medical settings adhere to the highest standards of safety and effectiveness. The standard’s relevance in enhancing patient care cannot be overstated, marking it as an essential tool in the ongoing effort to develop safer medical electrical equipment.
Die Norm EN IEC 60601-2-35:2021 befasst sich mit den besonderen Anforderungen an die grundlegende Sicherheit und die wesentliche Leistung von medizinischen Heizgeräten, die Decken, Matten und Matratzensysteme verwenden und für die Wärmeerzeugung im medizinischen Bereich bestimmt sind. Dieser Standard ist von großer Bedeutung, da er sicherstellt, dass solche Heizgeräte, die in der medizinischen Nutzung eingesetzt werden, strengen Sicherheitsanforderungen entsprechen und gleichzeitig eine effektive Leistung bieten. Der Umfang dieser Norm schließt Heizgeräte ein, die zur Vorwärmung von Betten dienen, was ihre Relevanz im klinischen Umfeld erhöht. Es werden spezifische Anforderungen definiert, die sowohl für medizinische Geräte (ME EQUIPMENT) als auch für medizinische Systeme (ME SYSTEMS) gelten, was eine umfassende Anwendung und Praktikabilität in verschiedenen medizinischen Einrichtungen gewährleistet. Besonders hervorzuheben ist, dass die Norm klare Differenzierungen zwischen allgemein anwendbaren und spezifisch definierten Anforderungen für verschiedene Gerätetypen bietet, was die Transparenz und Benutzerfreundlichkeit erhöht. Ein weiterer Pluspunkt der Norm ist der Umgang mit Gefahren, die mit der beabsichtigten physiologischen Funktion von medizinischen Geräten verbunden sind. Die Erwähnung spezifischer Paragraphen, die sich damit beschäftigen, stellt sicher, dass potenzielle Risiken angemessen adressiert und berücksichtigt werden. Zusätzlich bringt die aktuelle Auflage signifikante technische Änderungen im Vergleich zur vorherigen Version, insbesondere die Neudatierung normativer Verweise, was die Aktualität und Relevanz der Norm im Rahmen der technisch-medizinischen Entwicklungen unterstreicht. Insgesamt bietet die EN IEC 60601-2-35:2021 eine gründliche, strukturierte und praxisnahe Basis für die Sicherheit und Leistung von Heizgeräten im medizinischen Umfeld und fördert somit die Patienten- und Anwendersicherheit. Diese Norm ist für Hersteller, Betreiber und Anwender von medizinischem Equipment unabdingbar und trägt zur Qualitäts- und Sicherheitssteigerung im Gesundheitswesen bei.
標準EN IEC 60601-2-35:2021は、医療用の暖房デバイス、特にブランケット、パッド、マットレスを使用する機器の基本的な安全性および本質的な性能に関する特定の要求を定義しています。この文書は、医療用機器(ME EQUIPMENT)の安全性を確保するために不可欠な要素を網羅しており、医療分野での実用性が求められる暖房デバイスの規範を提供しています。 この標準の範囲は、入院患者向けにベッドを予熱する目的で使用される暖房デバイスを含む全体的な枠組みをカバーしています。具体的には、ME EQUIPMENTおよびME SYSTEMSの両方に適用される条項があり、その適用範囲が明確に示されています。特定のME EQUIPMENTにのみ適用される条項も設けられており、例えば強制空気デバイスに関する要求が含まれています。これにより、異なるデバイスやシステムへの適切なガイダンスが提供されています。 標準の強みは、医療使用における暖房デバイスの特定のハザードや要求に関する明確な指針を示すことで、医療従事者が安全に機器を使用できるようサポートしている点です。基本的な安全性と本質的な性能が強化されることで、患者の安全が第一に考慮されています。 さらに、IEC 60601-2-35:2020は、2009年に発行されたIEC 80601-2-35および2016年の改訂版を置き換え、最新の技術的見解と世の中のニーズに基づいて見直されています。この改訂により、標準に対する信頼性が向上し、医療業界における適用の幅が広がっています。 この標準は、医療用の暖房デバイスの設計と使用に関して規制を遵守し、製品の品質を保証する上で重要な役割を果たします。EN IEC 60601-2-35:2021は、現代の医療実践におけるニーズに対して、適切で関連性の高い基準として位置付けられていると言えるでしょう。
EN IEC 60601-2-35:2021 표준 문서는 의료 전기 장비의 기본 안전성과 필수 성능에 대한 특정 요구사항을 다루고 있습니다. 이 표준은 의료 용도로 사용되는 담요, 패드 및 매트리스를 사용하는 난방 장치의 안전성을 보장하는 데 중점을 두고 있습니다. 문서의 범위는 의료 장비(ME EQUIPMENT)에 대한 기본 안전과 필수 성능을 모두 포함하며, 특히 병상을 예열하기 위한 난방 장치도 대상에 포함됩니다. 이 표준의 강점은 모든 난방 장치에 적합한 요구 사항을 명확히 하여 사용자 안전을 강화한다는 점입니다. 특히, 각 조항이나 하위 조항이 특정 ME EQUIPMENT에만 적용되거나 ME SYSTEMS에 적용됨을 명시하고 있어 독자가 기준을 정확히 이해하고 적용할 수 있도록 돕습니다. 이를 통해 의료 현장에 안전하고 효과적인 난방 장치를 제공할 수 있습니다. 또한, IEC 60601-2-35:2020은 이전 판에서 삭제된 기술적 요소들을 보완하며, 해당 난방 장치의 특정 안전 요구 사항을 설정하여 의료 장비의 일관성과 신뢰성을 높이고 있습니다. 이 문서는 강제 공기 장치와 같은 특정 ME EQUIPMENT 유형에 대한 조항을 명확히 구분하고 있으면서도, ME EQUIPMENT와 ME SYSTEMS에 관련된 조항들은 모두 포괄적으로 적용됩니다. 이 표준은 최신 기술 기준을 반영하고 있으며, 이전 판인 IEC 80601-2-35(2009) 및 Amendment 1(2016)을 대체합니다. 이러한 기술 개정들이 포함되어있어 의료기기 제조사들에게 필수적인 지침이 될 것입니다. EN IEC 60601-2-35:2021 표준은 난방 장치의 안전성과 성능을 보장하기 위해 꼭 필요한 기준으로, 의료 현장에서의 적용이 필수적입니다.
La norme EN IEC 60601-2-35:2021 se consacre aux équipements électriques médicaux, spécifiquement aux exigences de base relatives à la sécurité et à la performance essentielle des dispositifs de chauffage utilisant des couvertures, des coussins et des matelas destinés à un usage médical. Le champ d'application de cette norme est pertinent et bien défini, englobant les dispositifs de chauffage qui visent à préchauffer un lit, ce qui est crucial dans les environnements de soins de santé. Une des forces majeures de cette norme est sa capacité à clarifier les exigences applicables tant pour les Équipements Médicaux (ME) que pour les Systèmes Médicaux (ME SYSTEMS). La distinction est faite lorsque nécessaire, garantissant ainsi que les utilisateurs peuvent naviguer efficacement dans les clauses correspondantes sans ambiguïté. De plus, la norme stipule clairement les types de dispositifs exclus du champ d’application, tels que ceux destinés à la physiothérapie ou les incubateurs pour nourrissons, permettant ainsi de concentrer les exigences sur les appareils de chauffage spécifiques. Une autre force réside dans la mise à jour technique par rapport à l'édition précédente de 2009, qui intègre des références normatives remaniées, assurant que la norme reste en phase avec les évolutions technologiques et les meilleures pratiques actuelles. Cela démontre une réactivité aux besoins changeants du secteur médical et renforce la fiabilité des dispositifs qui en découlent. Globalement, l'EN IEC 60601-2-35:2021 s'avère être un document fondamental pour garantir la sécurité de ces dispositifs spécifiques. Sa structure claire et ses instructions détaillées font d'elle une norme incontournable pour les fabricants et les professionnels de la santé qui cherchent à assurer une performance essentielle et une sécurité de base dans l'utilisation de ces équipements de chauffage critiques en milieu médical.














Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...