EN ISO 8638:2014
(Main)Cardiovascular implants and extracorporeal systems - Extracorporeal blood circuit for haemodialysers, haemodiafilters and haemofilters (ISO 8638:2010)
Cardiovascular implants and extracorporeal systems - Extracorporeal blood circuit for haemodialysers, haemodiafilters and haemofilters (ISO 8638:2010)
ISO 8638:2010 specifies requirements for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators (hereafter referred to as “the device”) and (integral and non-integral) transducer protectors which are intended for use in haemodialysis, haemodiafiltration and haemofiltration.
Kardiovaskuläre Implantate und extrakorporale Systeme - Extrakorporaler Blutkreislauf bei Hämodialysatoren, Hämodiafiltern und Hämofiltern (ISO 8638:2010)
Diese Internationale Norm legt Anforderungen an Hämodialysatoren, Hämodiafilter, Hämofilter und Hämokonzentratoren (nachstehend als 'das Gerät' bezeichnet) sowie (integrale und nicht integrale) Schutzvorrichtungen für Messwandler ('Transducer-Protector') für die Anwendung bei Hämodialyse, Hämodiafiltration und Hämofiltration fest.
Die vorliegende Internationale Norm ist nicht anwendbar auf:
- Hämodialysatoren, Hämodiafilter oder Hämofilter;
- Plasmafilter;
- Hämoperfusionsgeräte;
- Geräte für den vaskulären Zugang;
- Blutpumpen;
- Druckaufzeichnungsgeräte für extrakorporale Blutkreisläufe;
- Luftnachweisgeräte;
- Systeme für die Herstellung, Aufbewahrung oder Überwachung der Dialyseflüssigkeit;
- Systeme oder Ausrüstungen für die Durchführung von Hämodialyse, Hämodiafiltration, Hämofiltration oder Hämokonzent¬ration.
ANMERKUNG Anforderungen an Hämodialysatoren, Hämodiafilter, Hämofilter und Hämokonzentratoren sind in ISO 8637 festgelegt.
Implants cardiovasculaires et systèmes extracorporels - Circuit sanguin extracorporel pour les hémodialyseurs, les hémodiafiltres et les hémofiltres (ISO 8638:2010)
L'ISO 8638:2010 spécifie des exigences relatives aux hémodialyseurs, hémodiafiltres, hémofiltres et hémoconcentrateurs (appelés ci-après «dispositifs») et aux dispositifs de protection des capteurs (intégrés et non intégrés) utilisables dans le cadre d'opérations d'hémodialyse, d'hémodiafiltration et d'hémofiltration.
Vsadki (implantati) za srce in ožilje ter zunajtelesni pretočni sistemi - Zunajtelesni krvni obtok za hemodializatorje, hemodiafiltre in hemofiltre (ISO 8638:2010)
General Information
- Status
- Withdrawn
- Publication Date
- 14-Jan-2014
- Withdrawal Date
- 20-Jan-2026
- Technical Committee
- CEN/TC 205 - Non-active medical devices
- Drafting Committee
- CEN/TC 205 - Non-active medical devices
- Current Stage
- 9960 - Withdrawal effective - Withdrawal
- Start Date
- 22-Aug-2018
- Completion Date
- 21-Jan-2026
Relations
- Effective Date
- 08-Jun-2022
- Effective Date
- 02-Mar-2016
- Effective Date
- 28-Jan-2026
- Effective Date
- 28-Jan-2026
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
EN ISO 8638:2014 is a standard published by the European Committee for Standardization (CEN). Its full title is "Cardiovascular implants and extracorporeal systems - Extracorporeal blood circuit for haemodialysers, haemodiafilters and haemofilters (ISO 8638:2010)". This standard covers: ISO 8638:2010 specifies requirements for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators (hereafter referred to as “the device”) and (integral and non-integral) transducer protectors which are intended for use in haemodialysis, haemodiafiltration and haemofiltration.
ISO 8638:2010 specifies requirements for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators (hereafter referred to as “the device”) and (integral and non-integral) transducer protectors which are intended for use in haemodialysis, haemodiafiltration and haemofiltration.
EN ISO 8638:2014 is classified under the following ICS (International Classification for Standards) categories: 11.040.20 - Transfusion, infusion and injection equipment; 11.040.40 - Implants for surgery, prosthetics and orthotics. The ICS classification helps identify the subject area and facilitates finding related standards.
EN ISO 8638:2014 has the following relationships with other standards: It is inter standard links to EN 1283:1996, EN ISO 8637-2:2018, EN ISO 14577-2:2015, EN 16068:2011. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN ISO 8638:2014 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC; Standardization Mandates: M/295. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
EN ISO 8638:2014 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-april-2014
1DGRPHãþD
SIST EN 1283:2000
9VDGNLLPSODQWDWL]DVUFHLQRåLOMHWHU]XQDMWHOHVQLSUHWRþQLVLVWHPL=XQDMWHOHVQL
NUYQLREWRN]DKHPRGLDOL]DWRUMHKHPRGLDILOWUHLQKHPRILOWUH,62
Cardiovascular implants and extracorporeal systems - Extracorporeal blood circuit for
haemodialysers, haemodiafilters and haemofilters (ISO 8638:2010)
Kardiovaskuläre Implantate und extrakorporale Systeme - Extrakorporaler Blutkreislauf
bei Hämodialysatoren, Hämodiafiltern und Hämofiltern (ISO 8638:2010)
Implants cardiovasculaires et systèmes extracorporels - Circuit sanguin extracorporel
pour les hémodialyseurs, les hémodiafiltres et les hémofiltres (ISO 8638:2010)
Ta slovenski standard je istoveten z: EN ISO 8638:2014
ICS:
11.040.20 Transfuzijska, infuzijska in Transfusion, infusion and
injekcijska oprema injection equipment
11.040.40 Implantanti za kirurgijo, Implants for surgery,
protetiko in ortetiko prosthetics and orthotics
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN ISO 8638
NORME EUROPÉENNE
EUROPÄISCHE NORM
January 2014
ICS 11.040.40 Supersedes EN 1283:1996
English Version
Cardiovascular implants and extracorporeal systems -
Extracorporeal blood circuit for haemodialysers, haemodiafilters
and haemofilters (ISO 8638:2010)
Implants cardiovasculaires et systèmes extracorporels - Kardiovaskuläre Implantate und extrakorporale Systeme -
Circuit sanguin extracorporel pour les hémodialyseurs, les Extrakorporaler Blutkreislauf bei Hämodialysatoren,
hémodiafiltres et les hémofiltres (ISO 8638:2010) Hämodiafiltern und Hämofiltern (ISO 8638:2010)
This European Standard was approved by CEN on 1 December 2013.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same
status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United
Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2014 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 8638:2014 E
worldwide for CEN national Members.
Contents Page
Foreword .3
Annex ZA (informative) Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC on medical devices .4
Foreword
The text of ISO 8638:2010 has been prepared by Technical Committee ISO/TC 150 “Implants for surgery” of
the International Organisation for Standardization (ISO) and has been taken over as EN ISO 8638:2014 by
Technical Committee CEN/TC 205 “Non-active medical devices” the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an identical
text or by endorsement, at the latest by July 2014, and conflicting national standards shall be withdrawn at the
latest by July 2014.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN 1283:1986.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association, and supports essential requirements of EU Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this document.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech
Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece,
Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.
Endorsement notice
The text of ISO 8638:2010 has been approved by CEN as EN ISO 8638:2014 without any modification.
Annex ZA
(informative)
Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC on medical devices
This European Standard has been prepared under a mandate given to CEN by the European Commission
and the European Free Trade Association to provide a means of conforming to Essential Requirements of the
New Approach Directive 93/42/EEC on medical devices.
Once this standard is cited in the Official Journal of the European Union under that Directive and has been
implemented as a national standard in at least one Member State, compliance with the clauses of this
standard given in Table ZA.1 confers, within the limits of the scope of this standard, a presumption of
conformity with the corresponding Essential Requirements of that Directive and associated EFTA regulations.
Table ZA.1 — Correspondence between this European Standard and Directive 93/42/EEC on
medical devices
Clause(s)/sub-clause(s) of this Essential Requirements (ERs) Qualifying remarks/Notes
EN of Directive 93/42/EEC
4.1, 4.2, 4.3 7.2
4.1 7.3
4.1, 6.3(q) 7.5 Addressed only in general terms.
Although these devices can
incorporate materials containing
phthalates, there is no specific
requirement that the presence of
phthalates be indicated in the
labelling.
4.4.1, 4.4.9 7.6
4.2, 4.4.1, 4.4.6, 4.4.9, 6.2(e), 8.1
6.2(j), 6.4(f), 6.4(i), 6.4(n)
4.2, 5.3 8.3 Addressed only in general terms.
4.2, 5.3 8.4
4.4.2, 4.4.3, 4.4.4, 4.4.9.2 9.1 Connectors are specified to
match tubing connectors
specified in ISO 8637 for the
blood compartment.
4.4.6.1, 4.4.10, 4.6 9.2
6 13.1
6.1, 6.2, 6.3, 6.4 13.2 The NOTE at the end of each
clause allows the use of symbols
from Harmonized Standards.
6.2(a), 6.3(a), 6.3(b), 6.4(a) 13.3 (a)
6.2(b), 6.2(c), 6.3(c), 6.3(d), 13.3 (b)
6.4(b), 6.4(c)
6.2(e), 6.3(f), 6.4(d) 13.3 (c)
Clause(s)/sub-clause(s) of this Essential Requirements (ERs) Qualifying remarks/Notes
EN of Directive 93/42/EEC
6.2(d), 6.3(e) 13.3 (d)
6.2(f), 6.3(g) 13.3 (e)
6.2(g), 6.4(e) 13.3 (f)
6.3(h) 13.3 (i)
6.2(j), 6.4(g), 6.4(i), 6.4(l), 13.3 (j)
6.4(m), 6.4(o)
6.2(j), 6.4(f) 13.3 (k)
6.2(i) 13.3 (m)
6.4(a), 6.4(b), 6.4(c), 6.4(d), 13.6 (a) There is no requirement for the
6.4(e), 6.4(f), 6.4(g), 6.4(i), 6.4(l), information in 13.3(i) in the
instructions for use. Instead, that
6.4(m), 6.4(o)
information is required to be
given on the outer container in
which the device is sold.
6.4(r) 13.6 (c)
WARNING — Other requirements and other EU Directives may be applicable to the product(s) falling within
the scope of this standard.
INTERNATIONAL ISO
STANDARD 8638
Third edition
2010-07-01
Cardiovascular implants and
extracorporeal systems — Extracorporeal
blood circuit for haemodialysers,
haemodiafilters and haemofilters
Implants cardiovasculaires et systèmes extracorporels — Circuit
sanguin extracorporel pour les hémodialyseurs, les hémodiafiltres et les
hémofiltres
Reference number
ISO 8638:2010(E)
©
ISO 2010
ISO 8638:2010(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2010
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2010 – All rights reserved
ISO 8638:2010(E)
Contents Page
Foreword .iv
Introduction.v
1 Scope.1
2 Normative references.1
3 Terms and definitions .2
4 Requirements.2
4.1 Biological safety.2
4.2 Sterility.3
4.3 Non-pyrogenicity .3
4.4 Mechanical characteristics.3
4.5 Expiry date .5
4.6 Tubing compliance.5
5 Test methods .5
5.1 General .5
5.2 Biological safety.6
5.3 Sterility.6
5.4 Non-pyrogenicity .6
5.5 Mechanical characteristics.6
5.6 Expiry date .9
5.7 Tubing compliance.9
6 Labelling.10
6.1 Labelling on the device.10
6.2 Labelling on the unit container .10
6.3 Labelling on the outer container.10
6.4 Accompanying documentation.11
Bibliography.13
ISO 8638:2010(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 8638 was prepared by Technical Committee ISO/TC 150, Implants for surgery, Subcommittee SC 2,
Cardiovascular implants and extracorporeal systems.
This third edition cancels and replaces the second edition (ISO 8638:2004), which has been technically
revised.
iv © ISO 2010 – All rights reserved
ISO 8638:2010(E)
Introduction
This International Standard is concerned with the extracorporeal blood circuit manufactured for single use and
intended for use in conjunction with haemodialysers, haemodiafilters and haemofilters. The requirements
specified in this International Standard for the extracorporeal blood circuit will help to ensure safety and
satisfactory function.
It was not found practicable to specify materials of construction. This International Standard therefore requires
only that materials have been tested and that the methods and results are made available upon request.
The dimensions of the connectors intended for connecting the extracorporeal blood circuit to a haemodialyser,
haemodiafilter or haemofilter have been specified to ensure compatibility with these devices, as specified in
ISO 8637. The design and dimensions have been selected in order to minimize the risk of leakage of blood
and ingress of air. Connectors with either fixed or loose locking shells are permitted.
This International Standard reflects the consensus of physicians, manufacturers and other interested parties
for devices that are approved for clinical use. Conformance with this International Standard is voluntary and it
is not intended to supersede any national regulation.
INTERNATIONAL STANDARD ISO 8638:2010(E)
Cardiovascular implants and extracorporeal systems —
Extracorporeal blood circuit for haemodialysers,
haemodiafilters and haemofilters
1 Scope
This International Standard specifies requirements for haemodialysers, haemodiafilters, haemofilters and
haemoconcentrators (hereafter referred to as “the device”) and (integral and non-integral) transducer
protectors which are intended for use in haemodialysis, haemodiafiltration and haemofiltration.
This International Standard does not apply to:
⎯ haemodialysers, haemodiafilters or haemofilters;
⎯ plasmafilters;
⎯ haemoperfusion devices;
⎯ vascular access devices;
⎯ blood pumps;
⎯ pressure monitors for the extracorporeal blood circuit;
⎯ air detection devices;
⎯ systems to prepare, maintain or monitor dialysis fluid;
⎯ systems or equipment intended to perform haemodialysis, haemodiafiltration, haemofiltration or
haemoconcentration.
NOTE Requirements for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators are specified in
ISO 8637.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 594-2, Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment —
Part 2: Lock fittings
ISO 7864, Sterile hypodermic needles for single use
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
ISO 10993-4, Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood
ISO 8638:2010(E)
ISO 10993-7, Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals
ISO 10993-11, Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
air capture chamber
component intended to capture air and which can provide compliance to the blood circuit or allow pressure to
be monitored
NOTE Air capture chambers are also known as drip chambers, bubble traps or venous and arterial blood chambers.
3.2
extracorporeal blood circuit
blood tubing and integral accessory tubing, including fluid and infusion tubing, for attaching the extracorporeal
blood circuit to pressure monitors and integral components
EXAMPLES (Of integral components.) Air-capture chambers and transducer protectors.
3.3
fluid pathway
internal surfaces of the extracorporeal blood circuit
3.4
labelling
written, printed, graphic or electronic matter that:
⎯ is affixed to a medical device or any of its containers or wrappers
or
⎯ accompanies a medical device and which is related to identification, technical description and use of that
medical device, but excluding shipping documents
3.5
pump segment
portion of the extracorporeal blood circuit (3.2) that is acted upon by the blood pump
3.6
transducer protector
pressure-transmitting sterile barrier
component of the extracorporeal blood circuit (3.2) that is intended to provide an interconnection between the
extracorporeal blood circuit and the haemodialysis machine while allowing the pressure within the
extracorporeal blood circuit to be measured by the machine
4 Requirements
4.1 Biological safety
Parts of the device that are intended to come into direct or indirect contact with blood shall be evaluated for
freedom from biological hazards, in accordance with 5.2.
2 © ISO 2010 – All rights reserved
ISO 8638:2010(E)
NOTE Attention is drawn to the need to establish whether national regulations or national standards governing
toxicology and biocompatibility testing exist in the country in which the device is produced and, if applicable, in the
countries in which the device is to be marketed.
4.2 Sterility
All fluid contacting surfaces of the device, and the mating surfaces of all connectors integral to the device,
shall be sterile. Compliance shall be verified in accordance with 5.3.
4.3 Non-pyrogenicity
The blood pathway of the device shall be non-pyrogenic. Compliance shall be verified in accordance with 5.4.
4.4 Mechanical characteristics
4.4.1 Structural integrity
The device shall be capable of withstanding a positive pressure of 1,5 × the manufacturer's recommended
maximum pressure and a negative pressure not exceeding 70
...