ASTM E1413-19
(Practice)Standard Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Dynamic Headspace Concentration onto an Adsorbent Tube
Standard Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Dynamic Headspace Concentration onto an Adsorbent Tube
SIGNIFICANCE AND USE
5.1 This practice is useful for preparing extracts from fire debris for subsequent qualitative analysis by gas chromatography mass spectrometry, see Test Method E1618.
5.2 The sensitivity of this practice is such that a sample consisting of a laboratory tissue onto which as little as 0.1 µL of ignitable liquid has been deposited, in an otherwise empty sample container, will result in an extract that is sufficient for identification and classification using Test Method E1618 (1).
5.2.1 Recovery from fire debris samples will vary, depending on factors including debris temperature, adsorbent temperature, container size, adsorptive material, headspace volume, sampling time and flow rate, and adsorptive competition from the sample matrix (2).
5.3 The principal concepts of dynamic headspace concentration are similar to those of static headspace concentration (Practice E3189). The dynamic headspace concentration technique can be more sensitive than the static headspace concentration technique. However, sample containers subjected to dynamic headspace concentration could be unsuitable for re-sampling.
5.3.1 Dynamic headspace concentration alters the original composition of the test sample because a portion of the original headspace from the sample container is removed and exchanged with dry inert gas or air. A portion of the concentrated headspace sample should be preserved for potential future analysis, if possible and if required, in accordance with Practice E2451.
5.4 Common solid adsorbent/desorption procedure combinations in use are activated carbon/solvent elution, and Tenax4 TA/thermal desorption.
5.5 Solid adsorbent/desorption procedure combinations not specifically described in this standard can be used as long as the practice has been validated as outlined in Section 11.
SCOPE
1.1 This practice describes the procedure for separation of ignitable liquid residues from fire debris samples using dynamic headspace concentration onto an adsorbent tube, with subsequent solvent elution or thermal desorption.
1.2 Dynamic headspace concentration onto an adsorbent tube takes place from a closed, rigid sample container (typically a metal can), using a source of dry inert gas or a vacuum system.
1.3 Both positive and negative applied pressure systems for dynamic headspace concentration onto an adsorbent tube are illustrated and described.
1.4 This practice is suitable for preparing extracts from fire debris samples containing a range of volumes (µL to mL) of ignitable liquid residues, with sufficient recovery for subsequent qualitative analysis (1).2
1.5 Alternative headspace concentration methods are listed in Section 2 (see Practices E1388, E1412, E3189, and E2154).
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.7 This standard cannot replace knowledge, skills, or abilities acquired through education, training, and experience (Practice E2917) and is to be used in conjunction with professional judgment by individuals with such discipline-specific knowledge, skills, and abilities.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
- Status
- Published
- Publication Date
- 31-May-2019
- Technical Committee
- E30 - Forensic Sciences
- Drafting Committee
- E30.01 - Criminalistics
Relations
- Refers
ASTM E1388-24 - Standard Practice for Static Headspace Sampling of Vapors from Fire Debris Samples - Effective Date
- 01-Mar-2024
- Effective Date
- 01-Feb-2024
- Effective Date
- 01-Feb-2024
- Effective Date
- 15-Feb-2019
- Effective Date
- 15-Jan-2019
- Effective Date
- 01-Nov-2018
- Effective Date
- 01-Jun-2018
- Effective Date
- 15-May-2018
- Effective Date
- 01-Sep-2017
- Effective Date
- 15-Jan-2016
- Effective Date
- 01-Mar-2015
- Effective Date
- 01-Jul-2014
- Effective Date
- 15-Feb-2013
- Effective Date
- 15-Jun-2012
- Effective Date
- 01-Jun-2012
Overview
ASTM E1413-19, Standard Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Dynamic Headspace Concentration onto an Adsorbent Tube, offers a robust procedure for forensic analysis of fire debris. This practice uses dynamic headspace concentration to extract and concentrate ignitable liquid residues onto an adsorbent tube, enabling subsequent qualitative analysis, often by gas chromatography-mass spectrometry (GC-MS). Recognized internationally, the method ensures sensitivity and efficiency in detecting even trace amounts of ignitable liquids, essential for fire investigation and forensic science applications.
Key Topics
Dynamic Headspace Concentration
This extraction technique involves exchanging vapors from a closed sample container and adsorbing them onto a solid medium under positive or negative pressure. Compared to static headspace methods, the dynamic approach enhances sensitivity, especially beneficial for low-concentration samples.Solid Adsorbents and Desorption
Common adsorbents include activated carbon (typically for solvent elution) and Tenax TA (typically for thermal desorption). This practice allows both desorption techniques, ensuring versatility for various laboratory setups.Applicability and Sensitivity
Suitable for fire debris samples containing tiny volumes-from microliters to milliliters-of ignitable liquid residues. The method offers sufficient recovery for identification and classification by analytical techniques.Quality Control and Validation
The standard emphasizes validation of the method under laboratory-specific conditions using a range of ignitable liquids and substrate materials. Users must validate flow rates, heating times, and analyte recovery to ensure reliable results.Limitations and Considerations
- Potential destructiveness: Part of the sample may not be available for re-analysis.
- Dynamic headspace techniques alter the original sample, making preservation of a portion of the extract for possible future analysis necessary.
- Requires professional judgment and proper training for accurate use.
Safety, Health, and Environment
Users are responsible for establishing safety practices and adhering to regulatory requirements related to solvents, gases, and material handling.
Applications
ASTM E1413-19 is highly relevant for the following fields and activities:
Forensic Science Laboratories
Used to analyze fire scene debris for the presence of ignitable liquid residues, supporting arson investigations and legal proceedings.Fire Investigation
Enables accurate identification of fuels or accelerants used in suspected fire scenes, contributing vital information to insurance and criminal casework.Analytical Chemistry
Applied in research or quality control environments where trace detection of volatile or semi-volatile compounds from complex matrices is required.Training and Method Development
Serves as a reference for training forensic practitioners (in accordance with ASTM E2917), and for validation of new or alternative headspace extraction techniques.
Related Standards
To ensure comprehensive fire debris analysis and maintain laboratory best practices, users should also consider:
- ASTM E1386 - Practice for Separation by Solvent Extraction
- ASTM E1388 - Practice for Static Headspace Sampling
- ASTM E1412 - Practice for Passive Headspace Concentration
- ASTM E1459 - Guide for Physical Evidence Labeling and Documentation
- ASTM E1492 - Practice for Evidence Management in Forensic Labs
- ASTM E1618 - Test Method for GC-MS Analysis of Extracts
- ASTM E2154 - Practice for Solid Phase Microextraction (SPME)
- ASTM E2451 - Practice for Preserving Extracts
- ASTM E2917 - Practice for Forensic Science Practitioner Training
- ASTM E3189 - Practice for Static Headspace Concentration onto an Adsorbent Tube
- ASTM E1732 - Terminology for Forensic Science
Keywords: ASTM E1413-19, dynamic headspace concentration, ignitable liquid residue, fire debris, adsorbent tube, forensic science, GC-MS, fire investigation, arson analysis, sample extraction, activated carbon, Tenax TA, analytical chemistry, standard practice, forensic laboratory procedures.
Buy Documents
ASTM E1413-19 - Standard Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Dynamic Headspace Concentration onto an Adsorbent Tube
REDLINE ASTM E1413-19 - Standard Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Dynamic Headspace Concentration onto an Adsorbent Tube
Get Certified
Connect with accredited certification bodies for this standard

NSF International
Global independent organization facilitating standards development and certification.
CIS Institut d.o.o.
Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

Kiwa BDA Testing
Building and construction product certification.
Sponsored listings
Frequently Asked Questions
ASTM E1413-19 is a standard published by ASTM International. Its full title is "Standard Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Dynamic Headspace Concentration onto an Adsorbent Tube". This standard covers: SIGNIFICANCE AND USE 5.1 This practice is useful for preparing extracts from fire debris for subsequent qualitative analysis by gas chromatography mass spectrometry, see Test Method E1618. 5.2 The sensitivity of this practice is such that a sample consisting of a laboratory tissue onto which as little as 0.1 µL of ignitable liquid has been deposited, in an otherwise empty sample container, will result in an extract that is sufficient for identification and classification using Test Method E1618 (1). 5.2.1 Recovery from fire debris samples will vary, depending on factors including debris temperature, adsorbent temperature, container size, adsorptive material, headspace volume, sampling time and flow rate, and adsorptive competition from the sample matrix (2). 5.3 The principal concepts of dynamic headspace concentration are similar to those of static headspace concentration (Practice E3189). The dynamic headspace concentration technique can be more sensitive than the static headspace concentration technique. However, sample containers subjected to dynamic headspace concentration could be unsuitable for re-sampling. 5.3.1 Dynamic headspace concentration alters the original composition of the test sample because a portion of the original headspace from the sample container is removed and exchanged with dry inert gas or air. A portion of the concentrated headspace sample should be preserved for potential future analysis, if possible and if required, in accordance with Practice E2451. 5.4 Common solid adsorbent/desorption procedure combinations in use are activated carbon/solvent elution, and Tenax4 TA/thermal desorption. 5.5 Solid adsorbent/desorption procedure combinations not specifically described in this standard can be used as long as the practice has been validated as outlined in Section 11. SCOPE 1.1 This practice describes the procedure for separation of ignitable liquid residues from fire debris samples using dynamic headspace concentration onto an adsorbent tube, with subsequent solvent elution or thermal desorption. 1.2 Dynamic headspace concentration onto an adsorbent tube takes place from a closed, rigid sample container (typically a metal can), using a source of dry inert gas or a vacuum system. 1.3 Both positive and negative applied pressure systems for dynamic headspace concentration onto an adsorbent tube are illustrated and described. 1.4 This practice is suitable for preparing extracts from fire debris samples containing a range of volumes (µL to mL) of ignitable liquid residues, with sufficient recovery for subsequent qualitative analysis (1).2 1.5 Alternative headspace concentration methods are listed in Section 2 (see Practices E1388, E1412, E3189, and E2154). 1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.7 This standard cannot replace knowledge, skills, or abilities acquired through education, training, and experience (Practice E2917) and is to be used in conjunction with professional judgment by individuals with such discipline-specific knowledge, skills, and abilities. 1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
SIGNIFICANCE AND USE 5.1 This practice is useful for preparing extracts from fire debris for subsequent qualitative analysis by gas chromatography mass spectrometry, see Test Method E1618. 5.2 The sensitivity of this practice is such that a sample consisting of a laboratory tissue onto which as little as 0.1 µL of ignitable liquid has been deposited, in an otherwise empty sample container, will result in an extract that is sufficient for identification and classification using Test Method E1618 (1). 5.2.1 Recovery from fire debris samples will vary, depending on factors including debris temperature, adsorbent temperature, container size, adsorptive material, headspace volume, sampling time and flow rate, and adsorptive competition from the sample matrix (2). 5.3 The principal concepts of dynamic headspace concentration are similar to those of static headspace concentration (Practice E3189). The dynamic headspace concentration technique can be more sensitive than the static headspace concentration technique. However, sample containers subjected to dynamic headspace concentration could be unsuitable for re-sampling. 5.3.1 Dynamic headspace concentration alters the original composition of the test sample because a portion of the original headspace from the sample container is removed and exchanged with dry inert gas or air. A portion of the concentrated headspace sample should be preserved for potential future analysis, if possible and if required, in accordance with Practice E2451. 5.4 Common solid adsorbent/desorption procedure combinations in use are activated carbon/solvent elution, and Tenax4 TA/thermal desorption. 5.5 Solid adsorbent/desorption procedure combinations not specifically described in this standard can be used as long as the practice has been validated as outlined in Section 11. SCOPE 1.1 This practice describes the procedure for separation of ignitable liquid residues from fire debris samples using dynamic headspace concentration onto an adsorbent tube, with subsequent solvent elution or thermal desorption. 1.2 Dynamic headspace concentration onto an adsorbent tube takes place from a closed, rigid sample container (typically a metal can), using a source of dry inert gas or a vacuum system. 1.3 Both positive and negative applied pressure systems for dynamic headspace concentration onto an adsorbent tube are illustrated and described. 1.4 This practice is suitable for preparing extracts from fire debris samples containing a range of volumes (µL to mL) of ignitable liquid residues, with sufficient recovery for subsequent qualitative analysis (1).2 1.5 Alternative headspace concentration methods are listed in Section 2 (see Practices E1388, E1412, E3189, and E2154). 1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.7 This standard cannot replace knowledge, skills, or abilities acquired through education, training, and experience (Practice E2917) and is to be used in conjunction with professional judgment by individuals with such discipline-specific knowledge, skills, and abilities. 1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM E1413-19 is classified under the following ICS (International Classification for Standards) categories: 13.220.99 - Other standards related to protection against fire. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM E1413-19 has the following relationships with other standards: It is inter standard links to ASTM E1388-24, ASTM E2917-24, ASTM E1732-24, ASTM E2917-19, ASTM E1732-19, ASTM E1732-18b, ASTM E1732-18a, ASTM E1732-18, ASTM E1732-17, ASTM E1412-16, ASTM E2154-15, ASTM E1618-14, ASTM E1459-13, ASTM E1732-12, ASTM E1412-12. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM E1413-19 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1413 − 19 An American National Standard
Standard Practice for
Separation of Ignitable Liquid Residues from Fire Debris
Samples by Dynamic Headspace Concentration onto an
Adsorbent Tube
This standard is issued under the fixed designation E1413; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.9 This international standard was developed in accor-
dance with internationally recognized principles on standard-
1.1 This practice describes the procedure for separation of
ization established in the Decision on Principles for the
ignitable liquid residues from fire debris samples using dy-
Development of International Standards, Guides and Recom-
namic headspace concentration onto an adsorbent tube, with
mendations issued by the World Trade Organization Technical
subsequent solvent elution or thermal desorption.
Barriers to Trade (TBT) Committee.
1.2 Dynamic headspace concentration onto an adsorbent
tube takes place from a closed, rigid sample container (typi-
2. Referenced Documents
cally a metal can), using a source of dry inert gas or a vacuum
system.
2.1 ASTM Standards:
E1388 Practice for Static Headspace Sampling of Vapors
1.3 Both positive and negative applied pressure systems for
from Fire Debris Samples
dynamic headspace concentration onto an adsorbent tube are
illustrated and described. E1412 Practice for Separation of Ignitable Liquid Residues
from Fire Debris Samples by Passive Headspace Concen-
1.4 This practice is suitable for preparing extracts from fire
tration with Activated Charcoal
debris samples containing a range of volumes (µL to mL) of
E1459 Guide for Physical Evidence Labeling and Related
ignitable liquid residues, with sufficient recovery for subse-
Documentation
quent qualitative analysis (1).
E1492 Practice for Receiving, Documenting, Storing, and
1.5 Alternative headspace concentration methods are listed
Retrieving Evidence in a Forensic Science Laboratory
in Section 2 (see Practices E1388, E1412, E3189, and E2154).
E1618 TestMethodforIgnitableLiquidResiduesinExtracts
1.6 The values stated in SI units are to be regarded as
from Fire Debris Samples by Gas Chromatography-Mass
standard. No other units of measurement are included in this
Spectrometry
standard.
E1732 Terminology Relating to Forensic Science
1.7 This standard cannot replace knowledge, skills, or E2154 Practice for Separation and Concentration of Ignit-
abilities acquired through education, training, and experience
able Liquid Residues from Fire Debris Samples by Pas-
(Practice E2917) and is to be used in conjunction with
sive Headspace Concentration with Solid Phase Microex-
professional judgment by individuals with such discipline-
traction (SPME)
specific knowledge, skills, and abilities.
E2451 Practice for Preserving Ignitable Liquids and Ignit-
1.8 This standard does not purport to address all of the able Liquid Residue Extracts from Fire Debris Samples
safety concerns, if any, associated with its use. It is the E2917 Practice for Forensic Science Practitioner Training,
responsibility of the user of this standard to establish appro-
Continuing Education, and Professional Development
priate safety, health, and environmental practices and deter-
Programs
mine the applicability of regulatory limitations prior to use.
E3189 Practice for Separation of Ignitable Liquid Residues
from Fire Debris Samples by Static Headspace Concen-
tration onto an Adsorbent Tube
This practice is under the jurisdiction of ASTM Committee E30 on Forensic
Sciences and is the direct responsibility of Subcommittee E30.01 on Criminalistics.
Current edition approved June 1, 2019. Published June 2019. Originally
approved in 1991. Last previous edition approved in 2013 as E1413 – 13. DOI: For referenced ASTM standards, visit the ASTM website, www.astm.org, or
10.1520/E1413-19. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
The boldface numbers in parentheses refer to a list of references at the end of Standards volume information, refer to the standard’s Document Summary page on
this standard. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1413 − 19
3. Terminology 5.3 The principal concepts of dynamic headspace concen-
tration are similar to those of static headspace concentration
3.1 Definitions—For definitions of terms used in this
(Practice E3189). The dynamic headspace concentration tech-
practice, refer to Terminology E1732.
nique can be more sensitive than the static headspace concen-
3.2 Definitions of Terms Specific to This Standard:
tration technique. However, sample containers subjected to
3.2.1 dynamic headspace concentration, n—an extraction
dynamic headspace concentration could be unsuitable for
technique in which a portion of the headspace vapors is
re-sampling.
exchangedfromthesamplecontainerandconcentratedontoan
5.3.1 Dynamic headspace concentration alters the original
adsorbent medium through applied positive or negative pres-
compositionofthetestsamplebecauseaportionoftheoriginal
sure.
headspace from the sample container is removed and ex-
3.2.2 static headspace concentration, n—an extraction tech- changed with dry inert gas or air.Aportion of the concentrated
nique in which a portion of the headspace vapors is removed
headspace sample should be preserved for potential future
from the sample container and concentrated onto an adsorbent analysis, if possible and if required, in accordance with
medium.
Practice E2451.
5.4 Common solid adsorbent/desorption procedure combi-
4. Summary of Practice
nations in use are activated carbon/solvent elution, and Tenax
4.1 Headspace vapors from the closed sample container are
TA/thermal desorption.
collected and concentrated onto an adsorbent tube by means of
5.5 Solid adsorbent/desorption procedure combinations not
headspace exchange performed through the use of a system
specifically described in this standard can be used as long as
based on either positive or negative applied pressure. The
the practice has been validated as outlined in Section 11.
adsorbent tube is subsequently eluted with solvent or desorbed
thermally prior to instrumental analysis (typically by gas
6. Apparatus
chromatography-mass spectrometry (GC-MS)).
6.1 Positive Applied Pressure-Based Sampling System—A
4.2 The technique of dynamic headspace concentration onto
system capable of increasing the pressure inside the sample
an adsorbent tube is illustrated in Fig. 1.The sample container,
container by applying dry inert gas at the sample container
and the adsorbent tube when necessary, can be heated during
inlet, thereby pushing the headspace vapors toward and
sampling, as outlined in Section 9.
through the outlet.
6.1.1 Sample Pressurization Device—A system capable of
5. Significance and Use
delivering pressurized dry inert gas at up to approximately 276
5.1 This practice is useful for preparing extracts from fire kPa (40 psi) to the inlet.
debris for subsequent qualitative analysis by gas chromatogra-
6.1.1.1 A needle valve capable of fine control of the flow
phy mass spectrometry, see Test Method E1618. rate of the dry inert gas, at up to approximately 400 mL/min.
6.1.2 A filtering tube, consisting of an adsorbent tube as
5.2 The sensitivity of this practice is such that a sample
described in 7.5, attached to the sample container inlet, for the
consisting of a laboratory tissue onto which as little as 0.1 µL
purpose of filtering incoming dry inert gas.
of ignitable liquid has been deposited, in an otherwise empty
6.1.3 Aflow meter capable of measuring the flow of the dry
sample container, will result in an extract that is sufficient for
inert gas through the end of the adsorption tube at the rate of
identification and classification using Test Method E1618 (1).
approximately 40 to 400 mL/min.
5.2.1 Recovery from fire debris samples will vary, depend-
6.1.4 Design and use the sampling system in a manner that
ing on factors including debris temperature, adsorbent
prevents the loss of volatile compounds from the container and
temperature, container size, adsorptive material, headspace
volume, sampling time and flow rate, and adsorptive compe-
tition from the sample matrix (2). Tenax is a trademark of Teijin Carbon America, Inc., Rockwood, TN.
Left: Positive applied pressure apparatus. Right: Negative applied pressure apparatus.
FIG. 1 Illustration of Dynamic Headspace Concentration onto an Adsorbent Tube
E1413 − 19
contamination of the sample from the apparatus itself or the 7.3 Solid Adsorbent—Activated carbon, Tenax TA or
laboratory environment. equivalent.
6.2 Negative Applied Pressure-Based Sampling System—A
7.4 Sampling Tubes—Glass Pasteur pipettes or equivalent
system capable of creating partial vacuum inside the sample
glasstubesforsolventelution,andstainlesssteelorglasstubes
container by connecting a vacuum system to the outlet, thereby
for thermal desorption.
pulling the headspace vapors toward and through the outlet.
7.5 Adsorbent Tubes—Sampling tubes packed with a solid
6.2.1 Avacuum system capable of pulling between approxi-
adsorbent.
mately 40 and 400 mL/min at the outlet.
7.5.1 Pre-packed adsorbent tubes are commercially avail-
6.2.2 A filtering tube, consisting of an adsorbent tube as
able. Empty sampling tubes that can be packed by the
described in 7.5, attached to the sample container inlet, for the
purchaser are also available.
purpose of filtering incoming air.
7.5.1.1 Activated carbon tubes and equivalent for solvent
6.2.3 Aflow meter capable of measuring the flow of filtered
elution can be made by inserting a filter into the bottom of a
air and headspace vapors through the outlet at the rate of
samplingtube(approximately5-mmdiameter),thenadding2.5
approximately 40 to 400 mL/min.
cm to 5 cm of activated carbon or equivalent, which is held in
6.2.4 Design and use the sampling system in a manner that
place with a second filter.
prevents the loss of volatile compounds from the container and
7.5.1.2 Tenax TA tubes and equivalent for thermal desorp-
contamination of the sample from the apparatus itself or the
tioncanbemadebyinsertingafilterintooneendofasampling
laboratory environment.
tube, adding approximately 80 mg of pre-conditioned Tenax
6.3 Heating System—An oven large enough to accommo-
TAor equivalent, and then packing tightly with a second filter.
date the sample container, or a heating mantel designed to fit
the sample container, capable of maintaining the required NOTE 1—Tenax TA or equivalent is conditioned by heating, in accor-
dance with instrument manufacturer or supplier instructions.
temperature range uniformly throughout.
7.6 Screw cap or crimp-top glass vials with polytetrafluo-
6.4 Temperature Measuring Device—A thermometer or
roethylene (PTFE) lined seals.
thermocouple capable of measuring the temperature of the
heating system in the required range of operation, to within
7.7 Tape, rubber sleeve stopper, or equivalent.
approximately 5°C.
7.8 Elution Solvent—Suitable elution solvents include car-
6.5 Thermal Desorption System—A system capable of des-
bon disulfide, n-pentane, dichloromethane and diethyl ether.
orbing trapped volatile compounds from an adsorbent tube by
meansofelevatedtemperature,refocusingthemonacold-trap,
8. Sample Preparation
and subsequently introducing them to a capillary GC column
8.1 Observe the appropriate procedures for handling and
by flash heating. The desorption apparatus is directly coupled
documentation of all submitted samples (see Guide E1459 and
to a GC-MS.
Practice E1492).
6.6 Puncturing Device—A device, such as a nail, that is
capable of puncturing small holes in the lid of the sample 8.2 Prepare the fire debris sample container for sampling by
container.The size of the holes is such that the adsorbent tubes dynamic headspace concentration.
can be introduced.
8.2.1 Thesamplingsystemisdesignedsothattheheadspace
sample can be extracted from the container in which it was
7. Reagents and Materials
received at the laboratory.
8.2.1.1 Alternatively, the fire debris sample, or a portion of
7.1 Purity of Reagents—Use reagent grade or better chemi-
it, can be transferred to a clean, rigid sample container suitable
cals in all tests. Unless otherwise indicated, it is intended that
for dynamic headspace sampling, such as a metal can with a
all reagents conform to the specifications of the Committee on
friction-seal lid.
Analytical Reagents of the American Chemical Society where
such specifications are available. Other grades may be used, 8.2.1.2 Verify the cleanliness of the transfer container prior
to sample transfer. Cleanliness is determined by means of
provided it is first ascertained that the reagent is of sufficiently
high purity through evaluation of appropriate blank(s) to analysis of a dynamic headspace concentration sample ob-
tained from the empty transfer container using the same
permit its use without lessening the accuracy of the determi-
nation. conditions as will be used for the questioned sample.
8.2.1.3 Allow the transfer container with sample to equili-
7.2 Filter—Permeablematerialsuchassilanizedglasswool,
brate for at least one hour before sampling
wire mesh or cotton held in place with a frit.
8.2.2 Create two holes in the lid of the rigid sample
container using a puncturing device, such that the adsorbent
tubes can be introduced.
Reagent Chemicals, American Chemical Society Specifications, American
8.2.2.1 Seal the holes with tape, rubber sleeve stopper, or
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Analar Standards for Laboratory equivalent.
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, NOTE 2—Cans that are designed for fire debris samples, with a hole
MD. pre-fitted with a rubber sleeve stopper, are commercially available.
E1413 − 19
9. Adsorption Procedure seal the openings in the container lid with tape, rubber sleeve
stoppers, or equivalent.
9.1 A laboratory typically uses either a positive applied
9.2.6.1 If an adsorption tube is to be thermally desorbed,
pressure-based system or a negative applied pressure-based
recap it immediately.
system. Either type of system can be used with the adsorbent
9.2.6.2 If an adsorption tube is to be desorbed by solvent
tubes described in 7.5.
elution and it was heated, allow the adsorption tube to cool to
9.2 Positive Applied Pressure-Based System:
room temperature prior to performing the desorption proce-
9.2.1 Place the inlet inert gas filtering adsorption tube and
dure.
theoutletsampleadsorptiontubethroughtheholesinthelidof
9.3 Negative Applied Pressure-Based System:
the sample container by penetrating the tape, rubber sleeve
9.3.1 Place the inlet air filtering adsorption tube and the
stoppers, or equivalent covering the holes.
outletsampleadsorptiontubethroughtheholesinthelidofthe
9.2.1.1 Useanewinletinertgasfilteringadsorptiontubefor
sample container by penetrating the tape, rubber sleeve
each sample container.
stoppers, or equivalent covering the holes.
9.2.2 Connect the sample pressurization device to the inlet
9.3.1.1 Use a new inlet air filtering adsorption tube for each
inert gas filtering adsorption tube. Then connect the exhaust
sample container.
tube to the outlet sample adsorption tube. Direct the system
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1413 − 13 E1413 − 19
Standard Practice for
Separation of Ignitable Liquid Residues from Fire Debris
Samples by Dynamic Headspace Concentration onto an
Adsorbent Tube
This standard is issued under the fixed designation E1413; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice describes the procedure for separation of small quantities of ignitable liquid residues from fire debris samples
using the method of dynamic headspace concentration.dynamic headspace concentration onto an adsorbent tube, with subsequent
solvent elution or thermal desorption.
1.2 Dynamic headspace concentration uses adsorption and subsequent solvent elution or thermal desorption.onto an adsorbent
tube takes place from a closed, rigid sample container (typically a metal can), using a source of dry inert gas or a vacuum system.
1.3 Both positive and negative applied pressure systems for adsorption are described, as well as a thermal desorption
system.dynamic headspace concentration onto an adsorbent tube are illustrated and described.
1.4 While this This practice is suitable for successfully extracting ignitable liquid residues over the entire range of
concentration, the headspace concentration methods are best used when a high level of sensitivity is required due to a very low
concentration preparing extracts from fire debris samples containing a range of volumes (μL to mL) of ignitable liquid residues
in the sample.residues, with sufficient recovery for subsequent qualitative analysis (1).
1.5 Alternate separation and concentration proceduresAlternative headspace concentration methods are listed in Section 2. If
archival of the extract is of importance, then this practice’s thermal desorption procedure, SPME (Practice (see Practices E2154),
and headspace (Practice E1388), E1412sample, E3189separation techniques are not recommended unless a portion of the extract
can be split and retained. In order to have an archivable extract, then this practice’s sample collection on charcoal, solvent
extraction (Practice , and E1386E2154), or passive headspace concentration (Practice ).E1412) is recommended.
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.7 This practice offers a set of instructions for performing one or more specific operations. This standard cannot replace
knowledge, skill,skills, or abilityabilities acquired through appropriate education, training, and experience (Practice E2917) and
should is to be used in conjunction with sound professional judgment.professional judgment by individuals with such
discipline-specific knowledge, skills, and abilities.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.9 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
E1386 Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Solvent Extraction
E1388 Practice for Static Headspace Sampling of Vapors from Fire Debris Samples
This practice is under the jurisdiction of ASTM Committee E30 on Forensic Sciences and is the direct responsibility of Subcommittee E30.01 on Criminalistics.
Current edition approved Aug. 15, 2013June 1, 2019. Published September 2013June 2019. Originally approved in 1991. Last previous edition approved in 20072013 as
E1413 – 07.E1413 – 13. DOI: 10.1520/E1413-13.10.1520/E1413-19.
The boldface numbers in parentheses refer to a list of references at the end of this standard.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1413 − 19
E1412 Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Passive Headspace Concentration With
Activated Charcoal
E1459 Guide for Physical Evidence Labeling and Related Documentation
E1492 Practice for Receiving, Documenting, Storing, and Retrieving Evidence in a Forensic Science Laboratory
E1618 Test Method for Ignitable Liquid Residues in Extracts from Fire Debris Samples by Gas Chromatography-Mass
Spectrometry
E1732 Terminology Relating to Forensic Science
E2154 Practice for Separation and Concentration of Ignitable Liquid Residues from Fire Debris Samples by Passive Headspace
Concentration with Solid Phase Microextraction (SPME)
E2451 Practice for Preserving Ignitable Liquids and Ignitable Liquid Residue Extracts from Fire Debris Samples
E2917 Practice for Forensic Science Practitioner Training, Continuing Education, and Professional Development Programs
E3189 Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Static Headspace Concentration onto
an Adsorbent Tube
3. Terminology
3.1 Definitions—For definitions of terms used in this practice, refer to Terminology E1732.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 dynamic headspace concentration, n—an extraction technique in which a portion of the headspace vapors is exchanged
from the sample container and concentrated onto an adsorbent medium through applied positive or negative pressure.
3.2.2 static headspace concentration, n—an extraction technique in which a portion of the headspace vapors is removed from
the sample container and concentrated onto an adsorbent medium.
4. Summary of Practice
4.1 The sample, preferably in its original container, is heated, forcing volatile compounds to vaporize. The headspace in the
sample is then drawn or pushed through a tube containing an adsorption media (typically activated charcoal or Tenax) which
adsorbs the vaporized compounds. Headspace vapors from the closed sample container are collected and concentrated onto an
adsorbent tube by means of headspace exchange performed through the use of a system based on either positive or negative applied
pressure. The adsorbent tube is subsequently eluted with solvent or desorbed thermally prior to instrumental analysis (typically by
gas chromatography-mass spectrometry (GC-MS)).
4.2 Other solid adsorbents The technique of dynamic headspace concentration onto an adsorbent tube is illustrated in Fig. 1and
collection systems can be used as long as the. The sample container, and the adsorbent tube when necessary, can be heated during
sampling, as outlined in Section 9method has been validated with a wide variety of ignitable liquids with a range of both flash
points and polarity.
5. Significance and Use
5.1 This practice is useful for preparing extracts from fire debris for later subsequent qualitative analysis by gas
chromatography-mass spectrometry (GC-MS),chromatography mass spectrometry, see Test Method E1618.
5.2 This is a very sensitive separation procedure, capable of isolating quantities smaller than The sensitivity of this practice is
such that a sample consisting of a laboratory tissue onto which as little as 0.1 μL of ignitable liquid residue from a samplehas been
deposited, in an otherwise empty sample container, will result in an extract that is sufficient for identification and classification
using Test Method E1618 (1).
Left: Positive applied pressure apparatus. Right: Negative applied pressure apparatus.
FIG. 1 Illustration of Dynamic Headspace Concentration onto an Adsorbent Tube
E1413 − 19
5.2.1 Actual recovery Recovery from fire debris samples will vary, depending on several factors, factors including
adsorptiondebris temperature, container size, and adsorbent temperature, container size, adsorptive material, headspace volume,
sampling time and flow rate, and adsorptive competition from the sample matrix.matrix (2).
5.3 This is a potentially destructive technique. Portions of the sample subjected to this procedure may not be suitable for
re-sampling. Therefore, a portion of the sample extract shouldThe principal concepts of dynamic headspace concentration are
similar to those of static headspace concentration (Practice E3189be saved for potential future analysis. Consider using passive
headspace concentration as described in Practice). The dynamic headspace concentration technique can be more sensitive than the
static headspace concentration technique. However, sample containers E1412.subjected to dynamic headspace concentration could
be unsuitable for re-sampling.
5.3.1 Dynamic headspace concentration alters the original composition of the test sample because a portion of the original
headspace from the sample container is removed and exchanged with dry inert gas or air. A portion of the concentrated headspace
sample should be preserved for potential future analysis, if possible and if required, in accordance with Practice E2451.
5.4 Common solid adsorbent/desorption procedure combinations in use are activated carbon/solvent elution, and Tenax
TA/thermal desorption.
5.5 Solid adsorbent/desorption procedure combinations not specifically described in this standard can be used as long as the
practice has been validated as outlined in Section 11.
6. Apparatus
6.1 Positive Pressure Apparatus: Applied Pressure-Based Sampling System—A system capable of increasing the pressure inside
the sample container by applying dry inert gas at the sample container inlet, thereby pushing the headspace vapors toward and
through the outlet.
6.1.1 Sample Pressurization Device—A system capable of delivering pressurized dry nitrogen (or other inert gas)inert gas at up
to 40 psi (276 kPa) to an orifice to be inserted into the bottom of the sample container.approximately 276 kPa (40 psi) to the inlet.
6.1.1.1 A needle valve capable of fine control of the flow rate of the dry nitrogen, inert gas, at up to 1500 cc/min.approximately
400 mL/min.
5.1.1.2 A flow meter capable of measuring the flow of dry nitrogen through the end of the adsorption tube at the rate of 0 to
1500 cc/min.
5.1.1.3 Container Closure—A device suitable for sealing the container and directing the effluent nitrogen (or other inert gas) and
vapors to the adsorption tube.
6.1.2 A filtering tube, consisting of an adsorbent tube as described in 7.5, attached to the sample container inlet, for the purpose
of filtering incoming dry inert gas.
6.1.3 A flow meter capable of measuring the flow of the dry inert gas through the end of the adsorption tube at the rate of
approximately 40 to 400 mL/min.
6.1.4 Design and use the sampling system in a manner that prevents the loss of volatile compounds from the container and
contamination of the sample from the apparatus itself or the laboratory environment.
6.2 Negative Pressure Apparatus: Applied Pressure-Based Sampling System—
5.2.1 Inlet and Outlet System—A tube containing approximately 1 cm of charcoal is fitted into the lid of the original container
with a suitable penetrable seal. This serves as a filter for incoming room air. A vacuum is pulled on an adsorption tube also fitted
into the lid of the original container with a suitable penetrable seal.A system capable of creating partial vacuum inside the sample
container by connecting a vacuum system to the outlet, thereby pulling the headspace vapors toward and through the outlet.
6.2.1 A vacuum system capable of pulling between 200 and 1500 cc/min on the sample collection tube.approximately 40 and
400 mL/min at the outlet.
6.2.2 A filtering tube, consisting of an adsorbent tube as described in 7.5, attached to the sample container inlet, for the purpose
of filtering incoming air.
6.2.3 A flow meter capable of measuring the flow of air through the end of the charcoal tube filtered air and headspace vapors
through the outlet at the rate of 200 to 1500 cc/min.approximately 40 to 400 mL/min.
6.2.4 Design and use the sampling system in a manner that prevents the loss of volatile compounds from the container and
contamination of the sample from the apparatus itself or the laboratory environment.
6.3 Heating System—An oven large enough to accommodate the sample container, or a heating mantel designed to fit the sample
container, capable of maintaining the required temperature range uniformly throughout.
6.4 Temperature Measuring Device—A thermometer or thermocouple capable of measuring the temperature of the heating
system in the required range of operation, to within approximately 5°C.
6.5 Thermal Desorption Apparatus: System—
Tenax is a trademark of Teijin Carbon America, Inc., Rockwood, TN.
E1413 − 19
5.3.1 Sampling System—An air-tight syringe is connected to the back-end of a Tenax tube. The front end of the Tenax tube is
fitted into the heated container.A system capable of desorbing trapped volatile compounds from an adsorbent tube by means of
elevated temperature, refocusing them on a cold-trap, and subsequently introducing them to a capillary GC column by flash
heating. The desorption apparatus is directly coupled to a GC-MS.
5.3.2 In order to prevent leakage and contamination, the system should be designed to minimize loss of vapors from the
container. For example by means of a septum mounted on top of the lid of the container.
5.3.3 Thermal Desorption Device—A system capable of desorbing the volatiles by means of elevated temperature and trapping
the volatiles in a cold-trap. This apparatus is directly coupled to a GC-MS.
6.6 Adsorption Tubes: Puncturing Device—
5.4.1 Charcoal Tubes—Suitable charcoal filter and sample adsorption tubes may be made by inserting a small (approximately
1 cm) plug of glass wool or cotton in the bottom of a Pasteur pipette (approximately 5 mm diameter), then adding 2.5 to 5 cm
of activated charcoal, and finally, holding the charcoal in place with an additional plug of glass wool or cotton.
5.4.1.1 Alternatively, charcoal tubes are available from commercial sources.A device, such as a nail, that is capable of
puncturing small holes in the lid of the sample container. The size of the holes is such that the adsorbent tubes can be introduced.
5.4.2 Tenax Tubes—Suitable Tenax tubes are commercially available. Note that solvent extraction of these tubes results in
complications in the desorbing phase because certain solvents cause the dissolution of the Tenax. Tenax should be employed when
thermal desorption is to be performed.
5.5 Heating System—A heating mantel designed to fit the evidence container or an oven or a hot plate.
5.5.1 An oven may be set up with any number of stations to allow for multiple sample preparation.
5.6 Temperature Measuring Device—A thermometer or thermocouple capable of measuring temperatures in the range of 40 to
150°C.
7. Reagents and Materials
7.1 Purity of Reagents—Reagent Use reagent grade or better chemicals shall be used in all tests. Unless otherwise indicated,
it is intended that all reagents conform to the specifications of the Committee on Analytical Reagents of the American Chemical
Society where such specifications are available. Other grades may be used, provided it is first ascertained that the reagent is of
sufficiently high purity through evaluation of appropriate blank(s) to permit its use without lessening the accuracy of the
determination.
7.2 Solid Adsorbent: Filter—
6.2.1 Activated Charcoal (coconut).
6.2.1.1 Charcoal may be activated and cleaned by heating in a 400°C oven for approximately 4 hours and cooling in a
desiccator.
6.2.1.2 Test each new or reconditioned batch of charcoal for residual hydrocarbons by analyzing a concentrated extract of blank
adsorption media according to Test Method E1618.Permeable material such as silanized glass wool, wire mesh or cotton held in
place with a frit.
6.2.2 Tenax:
6.2.2.1 Tenax may be cleaned and conditioned according to the manufacturer’s instructions.
6.2.2.2 Test each new or reconditioned batch of Tenax for residual hydrocarbons by analyzing a concentrated extract of blank
adsorption media according to Test Method E1618.
7.3 Solid Adsorbent—Activated carbon, Tenax TA or equivalent.
7.4 Sampling Tubes—Glass Pasteur pipettes or equivalent glass tubes for solvent elution, and stainless steel or glass tubes for
thermal desorption.
7.5 Glass Wool, Adsorbent Tubes—or cotton, free of extractable hydrocarbons.Sampling tubes packed with a solid adsorbent.
7.5.1 Pre-packed adsorbent tubes are commercially available. Empty sampling tubes that can be packed by the purchaser are
also available.
7.5.1.1 Activated carbon tubes and equivalent for solvent elution can be made by inserting a filter into the bottom of a sampling
tube (approximately 5-mm diameter), then adding 2.5 cm to 5 cm of activated carbon or equivalent, which is held in place with
a second filter.
7.5.1.2 Tenax TA tubes and equivalent for thermal desorption can be made by inserting a filter into one end of a sampling tube,
adding approximately 80 mg of pre-conditioned Tenax TA or equivalent, and then packing tightly with a second filter.
NOTE 1—Tenax TA or equivalent is conditioned by heating, in accordance with instrument manufacturer or supplier instructions.
Reagent Chemicals, American Chemical Society Specifications, American Chemical Society, Washington, DC. For suggestions on the testing of reagents not listed by
the American Chemical Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia and National
Formulary, U.S. PharmaceuticalPharmacopeial Convention, Inc. (USPC), Rockville, MD.
E1413 − 19
7.6 Extraction solvent (for example,Screw cap carbonor disulfide, pentane, ethylcrimp-top glass vials ether).with polytetrafluo-
roethylene (PTFE) lined seals.
6.4.1 Read and follow the safety precautions described in the Safety Data Sheets (SDS) of the extraction solvent that is used.
6.4.2 Test each new lot number of the extraction solvent by analyzing a concentrated portion according to Test Method E1618.
7.7 Tape, rubber sleeve stopper, or equivalent.
7.8 Elution Solvent—Pasteur pipettes and glass vials freeSuitable elution solvents include carbon disulfide, ofn extractable
hydrocarbons.-pentane, dichloromethane and diethyl ether.
8. Sample Preparation
8.1 Observe the appropriate procedures for handling and documentation of all submitted samples (Guide (see Guide E1459 and
Practice E1492).
7.1.1 Examine the fire debris sample in order to determine that it is consistent with its description.
7.1.1.1 Resolve any discrepancies between the submitting agent’s description of the evidence and the analyst’s observation with
the submitting agent prior to the completion of the report.
7.2 This is a potentially destructive technique. Portions of the sample subjected to this procedure may not be suitable for
re-sampling. Therefore, a portion of the sample extract should be saved for potential future analysis. Consider using passive
headspace concentration as described in Practice E1412 which is essentially nondestructive.
8.2 The system should be designed so that the majority of the samples can be extracted from the container in which they are
delivered to the laboratory.Prepare the fire debris sample container for sampling by dynamic headspace concentration.
8.2.1 Alternatively, the sample or a portion of the sample can be placed in an appropriate, clean sampling container which is
designed to be flushed by positive or negative pressure.The sampling system is designed so that the headspace sample can be
extracted from the container in which it was received at the laboratory.
8.2.1.1 Alternatively, the fire debris sample, or a portion of it, can be transferred to a clean, rigid sample container suitable for
dynamic headspace sampling, such as a metal can with a friction-seal lid.
8.2.1.2 Verify the cleanliness of the transfer container prior to sample transfer. Cleanliness is determined by means of analysis
of a dynamic headspace concentration sample obtained from the empty transfer container using the same conditions as will be used
for the questioned sample.
8.2.1.3 Allow the transfer container with sample to equilibrate for at least one hour before sampling
8.2.2 Create two holes in the lid of the rigid sample container using a puncturing device, such that the adsorbent tubes can be
introduced.
8.2.2.1 Seal the holes with tape, rubber sleeve stopper, or equivalent.
NOTE 2—Cans that are designed for fire debris samples, with a hole pre-fitted with a rubber sleeve stopper, are commercially available.
9. Adsorption Procedure
9.1 A laboratory typically has one method of adsorption and elution that is employed. The most common combinations in use
are charcoal adsorption followed by solvent elution, and Tenax adsorption followed by thermal desorption. Both adsorbents can
be collected using uses either a positive applied pressure-based system or a negative pressure system. Other solid adsorbents and
collection systems applied pressure-based system. Either type of system can be used as long as the method has been validated with
a wide variety of ignitablewith the adsorbent tubes described in 7.5liquids with a range of both flash points and polarity.
8.1.1 The volume of air sampled is generally less when thermal desorption is employed as the entire collected sample is
commonly injected into the GC-MS system for analysis.
8.1.2 With thermal desorption there is typically no material that can be archived once GC-MS analysis has been performed. If
archival of a portion of the sample extract is needed, then a different adsorbent and elution combination should be chosen either
following this practice, or following one of the other sample collection methods for fire debris analysis (Practices E1386 or E1412).
Some instrument manufacturers now have the facility to collect the unused sample from the split; if your instrument has such
capability, then there will be material available for archival when thermal desorption is used.
9.2 Positive Pressure: Applied Pressure-Based System:
9.2.1 Place the inlet inert gas filtering adsorption tube and the outlet sample adsorption tube through the holes in the lid of the
sample container by penetrating the tape, rubber sleeve stoppers, or equivalent covering the holes.
9.2.1.1 Use a new inlet inert gas filtering adsorption tube for each sample container.
9.2.2 Place the sample container in the heat
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...