Standard Test Method for Measurement of Flow Properties and Evaluation of Wear, Contaminants, and Oxidative Properties of Lubricating Grease by Die Extrusion Method and Preparation

SIGNIFICANCE AND USE
5.1 Trending the wear, contamination, consistency, and oxidative properties of a lubricating grease is a crucial part of condition-monitoring programs. Changes in these properties or deviations from the new grease can be indicative of problems within the lubricated component, such as the mixing of incompatible thickener types, excessive wear or contaminant levels, or significant depletion of antioxidant levels. These test methods also makes it possible to develop trends that can be used to predict failures before they occur and allow for corrective action to be taken.
SCOPE
1.1 This test method covers the determination and evaluation of flow properties, wear levels, contaminants, and oxidative condition of new and in-service lubricating grease.  
1.2 This test method provides guidance on evaluating in-service grease samples, NLGI grades 00 to 3, for wear, consistency, contamination, and oxidation.  
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.  
1.3.1 Exception—The exception to this will be where units of references were developed by the developers of the test equipment and necessary to report the results of the test.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
30-Sep-2023

Relations

Effective Date
01-Oct-2023
Effective Date
01-Mar-2024
Effective Date
01-Feb-2024
Effective Date
01-Dec-2023
Effective Date
01-Jul-2023
Effective Date
01-Apr-2015

Overview

ASTM D7918-23 is the internationally recognized standard test method developed by ASTM International for the measurement of flow properties and the evaluation of wear, contaminants, and oxidative properties in lubricating grease using the die extrusion method. This standard is essential for condition monitoring programs in industries reliant on lubricating greases, both new and in-service, particularly for those using NLGI grades 00 to 3. By assessing changes in grease properties, users can detect issues such as contamination, excessive wear, incompatibility, or antioxidant depletion, helping to predict and prevent equipment failures.

Key Topics

  • Flow Properties: Evaluates the consistency and movement resistance of greases under stress, often referred to as penetration.
  • Wear Levels: Measures ferrous debris and other wear indicators using specialized analyzers and sensors.
  • Contaminants: Identifies the presence and concentration of particulate matter or other contaminants within the grease.
  • Oxidative Properties: Determines antioxidant content using linear sweep voltammetry and assesses oxidative degradation.
  • Sampling Methods: Covers both passive and active grease sampling devices, ensuring representative samples from gears, bearings, or drive shafts for accurate analysis.
  • Test Apparatus: Utilizes equipment such as die extrusion testers, load cells, temperature chambers, and spectrophotometers for thorough evaluation.

Applications

ASTM D7918-23 provides practical value in several industrial applications:

  • Predictive Maintenance: Trending grease properties supports proactive identification of component issues and enables timely corrective actions, reducing the risk of catastrophic equipment failures.
  • Quality Assurance: Ensures that in-service lubricating greases remain within specification, maintaining optimal machinery performance and extending equipment life.
  • Lubricant Development and Research: Facilitates comparison between new and used grease formulations, supporting R&D in lubrication technology.
  • Compliance and Reporting: Provides a consistent and traceable methodology for reporting and documenting the condition of lubricants in regulated industries.
  • Asset Management Programs: Supports condition-based maintenance strategies in sectors such as manufacturing, power generation, transportation, and heavy industry, where asset reliability is critical.
  • Environmental and Safety Monitoring: Helps maintain compliance with environmental and safety regulations by detecting potential lubricant failures that could lead to leaks, spills, or excessive wear.

Related Standards

Organizations implementing ASTM D7918-23 may also find these ASTM standards relevant to comprehensive grease condition monitoring:

  • ASTM D217: Test methods for cone penetration of lubricating grease
  • ASTM D6300: Practice for determining precision and bias of test methods for petroleum products
  • ASTM D6595: Determination of wear metals and contaminants in used lubricating oils
  • ASTM D7527: Measurement of antioxidant content in lubricating greases by linear sweep voltammetry
  • ASTM D7546: Determination of moisture in lubricating oils and additives by humidity sensor
  • ASTM D7718: Practice for obtaining in-service samples of lubricating grease
  • ASTM D7843: Measurement of lubricant generated insoluble color bodies in turbine oils using membrane patch colorimetry

Conclusion

ASTM D7918-23 establishes a robust, systematic approach for evaluating critical properties of lubricating grease through die extrusion. Its implementation enhances equipment reliability, supports maintenance planning, and improves the overall efficiency of industrial operations dependent on lubricated machinery. By providing a clear, standardized method for grease analysis, this test method is a cornerstone in the predictive maintenance and quality control of lubricating greases across multiple sectors.

Buy Documents

Standard

ASTM D7918-23 - Standard Test Method for Measurement of Flow Properties and Evaluation of Wear, Contaminants, and Oxidative Properties of Lubricating Grease by Die Extrusion Method and Preparation

English language (9 pages)
sale 15% off
sale 15% off
Standard

REDLINE ASTM D7918-23 - Standard Test Method for Measurement of Flow Properties and Evaluation of Wear, Contaminants, and Oxidative Properties of Lubricating Grease by Die Extrusion Method and Preparation

English language (9 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

ABS Quality Evaluations Inc.

American Bureau of Shipping quality certification.

ANAB United States Verified

Element Materials Technology

Materials testing and product certification.

UKAS United Kingdom Verified

ABS Group Brazil

ABS Group certification services in Brazil.

CGCRE Brazil Verified

Sponsored listings

Frequently Asked Questions

ASTM D7918-23 is a standard published by ASTM International. Its full title is "Standard Test Method for Measurement of Flow Properties and Evaluation of Wear, Contaminants, and Oxidative Properties of Lubricating Grease by Die Extrusion Method and Preparation". This standard covers: SIGNIFICANCE AND USE 5.1 Trending the wear, contamination, consistency, and oxidative properties of a lubricating grease is a crucial part of condition-monitoring programs. Changes in these properties or deviations from the new grease can be indicative of problems within the lubricated component, such as the mixing of incompatible thickener types, excessive wear or contaminant levels, or significant depletion of antioxidant levels. These test methods also makes it possible to develop trends that can be used to predict failures before they occur and allow for corrective action to be taken. SCOPE 1.1 This test method covers the determination and evaluation of flow properties, wear levels, contaminants, and oxidative condition of new and in-service lubricating grease. 1.2 This test method provides guidance on evaluating in-service grease samples, NLGI grades 00 to 3, for wear, consistency, contamination, and oxidation. 1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only. 1.3.1 Exception—The exception to this will be where units of references were developed by the developers of the test equipment and necessary to report the results of the test. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

SIGNIFICANCE AND USE 5.1 Trending the wear, contamination, consistency, and oxidative properties of a lubricating grease is a crucial part of condition-monitoring programs. Changes in these properties or deviations from the new grease can be indicative of problems within the lubricated component, such as the mixing of incompatible thickener types, excessive wear or contaminant levels, or significant depletion of antioxidant levels. These test methods also makes it possible to develop trends that can be used to predict failures before they occur and allow for corrective action to be taken. SCOPE 1.1 This test method covers the determination and evaluation of flow properties, wear levels, contaminants, and oxidative condition of new and in-service lubricating grease. 1.2 This test method provides guidance on evaluating in-service grease samples, NLGI grades 00 to 3, for wear, consistency, contamination, and oxidation. 1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only. 1.3.1 Exception—The exception to this will be where units of references were developed by the developers of the test equipment and necessary to report the results of the test. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM D7918-23 is classified under the following ICS (International Classification for Standards) categories: 75.100 - Lubricants, industrial oils and related products. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM D7918-23 has the following relationships with other standards: It is inter standard links to ASTM D7918-17a, ASTM D6300-24, ASTM D7546-24, ASTM D6300-23a, ASTM D6300-23, ASTM D7546-15. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM D7918-23 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D7918 − 23
Standard Test Method for
Measurement of Flow Properties and Evaluation of Wear,
Contaminants, and Oxidative Properties of Lubricating
Grease by Die Extrusion Method and Preparation
This standard is issued under the fixed designation D7918; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope* 2. Referenced Documents
2.1 ASTM Standards:
1.1 This test method covers the determination and evalua-
D217 Test Methods for Cone Penetration of Lubricating
tion of flow properties, wear levels, contaminants, and oxida-
Grease
tive condition of new and in-service lubricating grease.
D6300 Practice for Determination of Precision and Bias
1.2 This test method provides guidance on evaluating in-
Data for Use in Test Methods for Petroleum Products,
service grease samples, NLGI grades 00 to 3, for wear,
Liquid Fuels, and Lubricants
consistency, contamination, and oxidation.
D6595 Test Method for Determination of Wear Metals and
Contaminants in Used Lubricating Oils or Used Hydraulic
1.3 The values stated in SI units are to be regarded as the
Fluids by Rotating Disc Electrode Atomic Emission Spec-
standard. The values given in parentheses are for information
trometry
only.
D7527 Test Method for Measurement of Antioxidant Con-
1.3.1 Exception—The exception to this will be where units
tent in Lubricating Greases by Linear Sweep Voltammetry
of references were developed by the developers of the test
D7546 Test Method for Determination of Moisture in New
equipment and necessary to report the results of the test.
and In-Service Lubricating Oils and Additives by Relative
1.4 This standard does not purport to address all of the
Humidity Sensor
safety concerns, if any, associated with its use. It is the
D7718 Practice for Obtaining In-Service Samples of Lubri-
responsibility of the user of this standard to establish appro-
cating Grease
priate safety, health, and environmental practices and deter-
D7843 Test Method for Measurement of Lubricant Gener-
mine the applicability of regulatory limitations prior to use.
ated Insoluble Color Bodies in In-Service Turbine Oils
1.5 This international standard was developed in accor-
using Membrane Patch Colorimetry
dance with internationally recognized principles on standard-
E177 Practice for Use of the Terms Precision and Bias in
ization established in the Decision on Principles for the
ASTM Test Methods
Development of International Standards, Guides and Recom-
E691 Practice for Conducting an Interlaboratory Study to
mendations issued by the World Trade Organization Technical
Determine the Precision of a Test Method
Barriers to Trade (TBT) Committee.
3. Terminology
3.1 Definitions:
This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.96.07 on Integrated Testers, Instrumentation Techniques for
In-Service Lubricants. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Oct. 1, 2023. Published November 2023. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2015. Last previous edition approved in 2017 as D7918 – 17a. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/D7918-23. the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7918 − 23
3.1.1 active grease sampling device, n—device designed to pressed as a percent when compared to the average of three
take an active sample of a lubricating grease from a bearing, unique test conditions of the unused lubricating grease base-
gear, or drive shaft located in a grease lubricated component. line.
D7718
3.2.2 die extrusion tester, n—a device that measures the
3.1.1.1 Discussion—The Grease Thief (trademarked) Type
consistency of in-service lubricating greases and compares the
2 is a tool which meets this description and can be used to
measurement to the values obtained from a baseline sample of
obtain an active grease sample from a lubricated component. A
lubricating grease.
full description and dimensions of this device can be found in
3.2.3 ferrous debris level, n—the total amount of ferrous
Annex A2.
metal as measured by a Hall-effect sensor in the grease sample
3.1.2 active sampling, v—to use a sampling device to
independent of particle size.
actively gather an in-service lubricating grease sample from a
3.2.4 full attenuation, n—the loss of intensity in signal
grease-lubricated component. D7718
strength.
3.1.3 calibration, n—the determination of the values of the
3.2.5 grease colorimetry, n—the science of color measure-
significant parameters by comparison with values indicated by
ment by the evaluation of the CIE LAB values of grease in the
a set of reference standards. D6595
visible light region of the electromagnetic spectrum.
3.1.4 consistency, n—of lubricating grease, the degree of
3.2.6 grease ribbon, n—grease that is extruded onto a
resistance to movement under stress. D217
substrate.
3.1.4.1 Discussion—The term “consistency” is used some-
3.2.7 indexing die, n—a single-use orifice die to measure the
what synonymously with “penetration.” Generally, consistency
consistency as compared to the baseline.
refers to the worked penetration of a grease.
3.2.8 spring cell calibration cylinder, n—a cylinder that
3.1.5 linear sweep voltammetry, n—a test method designed
houses a spring and a push rod to compress the spring.
to monitor the anti-oxidant additive content in lubricating
3.2.9 substrate, n—the single-use strip onto which a ribbon
greases. D7527
of grease is extruded.
3.1.6 lubricating grease, n—a semi-fluid to solid product of
3.2.10 substrate segment, n—a peel-away section of the
dispersion of a thickener in a liquid lubricant. D217
substrate containing a portion of the extruded grease; the
3.1.6.1 Discussion—The dispersion of the thickener forms a
segments are pre-cut to hold 0.25 g of extruded grease ribbon.
two-phase system and immobilizes the liquid lubricant by
surface tension and other physical forces. Other ingredients are
3.2.11 speed 1, n—the slowest speed in which the grease is
commonly included to impart special properties. extruded from the indexing die.
3.1.7 in-service lubricating grease, n—lubricating grease 3.2.12 speed 2, n—the fastest speed in which the grease is
that has been applied as a lubricant to a gear, bearing, or drive extruded from the indexing die.
screw for any period of time. D7718
3.2.13 speed 3, n—the intermediate speed at which the
3.1.8 passive grease sampling device, n—a device designed grease is extruded from the indexing die.
to gather a sample from the equipment by being attached to the
4. Summary of Test Method
grease reservoir at the purge point. D7718
4.1 Testing of a grease sample includes steps that charac-
3.1.8.1 Discussion—The Grease Thief (trademarked) Type
terize the ferrous wear within the sample, the consistency of
1 is a tool which meets this description and can be used to
the sample, and properties related to its chemistry.
obtain a representative grease sample from a purge path. A full
description and dimensions of this device can be found in
4.2 A grease sample of known volume is measured to
Annex A1.
determine the density of ferrous material in the sample.
3.1.9 passive sampling, v—to use a passive grease sampling
4.3 A grease sample held with a defined geometry sample
device to collect a purged sample of in-service lubricating
holder is placed into a temperature-controlled instrument and
grease from a purge path. D7718
extruded onto a substrate as a thin ribbon or strip of grease. The
extrusion process provides a measurement of the consistency
3.2 Definitions of Terms Specific to This Standard:
of the grease.
3.2.1 die extrusion index, n—an average of three unique test
conditions of the in-service lubricating grease sample ex-
4.4 The test method requires the grease to be tested at three
different rates to reflect the non-Newtonian nature of greases.
Testing at several different rates creates a series of step changes
that are then compared to an unused baseline grease that has
The Grease Thief Type I and Type II is described in US Patent No. 7984661.
Interested parties are invited to submit information regarding the identification of an also been tested under the same conditions.
alternative(s) to this patented item to the ASTM International Headquarters. Your
4.5 While the flow properties are being measured, the grease
comments will receive careful consideration at a meeting of the responsible
is simultaneously deposited onto a thin-film substrate contain-
technical committee, which you may attend. The sole source of the Grease Thief
Type I and Type II are known to the committee is York Laboratories, LLC located
ing substrate segments. Each substrate segment contains ap-
at 410 Kings Mill Rd., York, PA 17401. If you are aware of alternative suppliers,
proximately 0.25 g of grease. The individual substrate seg-
please provide this information to ASTM International Headquarters. Your com-
ments are used for further testing of wear, contamination, and
ments will receive careful consideration at a meeting of the responsible technical
committee, which you may attend. oxidative properties.
D7918 − 23
4.6 The substrate is removed from the instrument and 7. Apparatus
processed further to obtain information related to the chemistry
7.1 Passive Grease Sampling Device—The device that holds
and content of the grease sample including linear sweep
the grease sample. The device is typically capable of holding
voltammetry and grease colorimetry.
approximately 1.5 g to 2 g of grease when full. A full
description and dimensions of this device can be found in
5. Significance and Use
Annex A1.
5.1 Trending the wear, contamination, consistency, and
7.2 Active Grease Sampling Device—The device that holds
oxidative properties of a lubricating grease is a crucial part of
an active grease sample from a grease-lubricated component. A
condition-monitoring programs. Changes in these properties or
full description and dimensions of this device can be found in
deviations from the new grease can be indicative of problems
Annex A2.
within the lubricated component, such as the mixing of
7.3 Die Extrusion Tester —Designed for the passive grease
incompatible thickener types, excessive wear or contaminant
sampling device, the die extrusion tester is a temperature-
levels, or significant depletion of antioxidant levels. These test
controlled chamber consisting of a table that is controlled by a
methods also makes it possible to develop trends that can be
linear actuator that holds the substrate, a bracket that holds the
used to predict failures before they occur and allow for
passive grease sampling device, and a second linear actuator to
corrective action to be taken.
push the handle of the passive grease sampling device. Both
6. Interferences the passive grease sampling device and the second linear
actuator are mounted perpendicular to the substrate table. A
6.1 Particulate Matter—When the in-service grease sample
load cell is mounted between the second actuator and the
is heavily contaminated with particulate matter, it may be
handle of the passive grease sampling device to measure the
necessary to perform a dilution of the sample in order to
force during the extrusion. The die extrusion tester is shown in
properly extrude the sample onto the substrate. Larger pieces of
Fig. 1.
particulate matter may also cause spikes in the load profile
graph. The data from these spikes should be discounted from 7.4 Constant Temperature Chamber—The test specimens in
the passive grease sampling device should be heated in the
the average force calculations because they are not indicative
of changes in the flow properties of the sample. temperature-controlled chamber of the die extrusion tester
prior to performing any test. The passive grease sampling
6.1.1 Heavily contaminated grease samples are considered
any grease that contains solid particulate of size or quantity that device shall have the indexing die firmly engaged on the
threaded open end of the device.
results in clogging of the die, which inhibits the completion of
the test.
7.5 Ferrous Debris Analyzer—An analyzer that allows the
passive grease sampling device to be inserted completely in the
NOTE 1—The specific steps to perform a dilution of grease samples is
not addressed in this test method.
coil geometry, designed to measure the total amount of ferrous
debris in the grease sample. By use of a Hall-effect sensor, the
6.2 Sample Size—The passive sampling device shall be at
total amount of ferrous debris is designed to measure quanti-
least 70 % filled to perform the testing indicated within this test
tatively in parts per million (ppm). Upon receipt of the sample,
method. If the sampling device is insufficiently filled, the
it is the first test performed on the grease.
instrument may not record a force reading over the entirety of
the program.
6.3 Color—The grease colorimetry test may become insig-
Wurzbach, R. and Williams, L., “Die Extrusion Method For Comparing
nificant if the color of the grease causes full attenuation and
Changes in Grease Consistency and Flow Characteristics,” Viscosity and Rheology
minimal transmittance of the signal making it incomparable to
of In-Service Fluids as They Pertain to Condition Monitoring, ASTM STP 1564,
the new grease. 2013, ASTM International.
FIG. 1 Die Extrusion Tester
D7918 − 23
7.6 Linear Sweep Voltammetry—Linear sweep voltammetry 10.2 Passive Grease Sampling Tool—Upon completion of
measures the amount of hindered phenols, amines, and zinc the ferrous debris analyzer test, the protective cap shall be
dithio dialkyl phosphate (ZDDP) in a lubricating grease. Using removed and an indexing die shall be threaded onto the open
0.25 g of grease from the substrate, one substrate segment is end of the passive grease sampling device.
NOTE 2—If the in-service grease sample is received in an active grease
inserted into a vial and the antioxidant concentration remaining
sampling device, the stinger probe must be removed and replaced with the
in the grease is measured and compared to an unused sample of
handle of a passive grease sampling device prior to use.
the same grease.
NOTE 3—If the in-service grease sample is received in any sampling
container other than the passive or active grease sampling device, it must
7.7 Grease Colorimetry—The grease colorimetry test pro-
be transferred into a passive grease sampling device by hand, with a clean,
vides a spectrum in the 400 nm to 700 nm region of the visible
single-use glove and/or a clean, single-use 10 mL slip tip syringe.
light spectrum. An optical spectroscopy cell is used and the
substrate segment is placed into the same holder. The delta-E
11. Preparation of Apparatus
values within the CIE LAB scale are recorded and a spectrum
11.1 Preparation of the Die Extrusion Tester:
can be generated in a method similar to Test Method D7843.
11.1.1 Heating the Chamber—Prior to the performance of
7.8 Direct Imaging System—A particle sizing and identifi-
any tests, a thermocouple is used to regulate the temperature of
cation component consisting of a CCD (charge-coupled de-
the heating tray to achieve a temperature sufficient to pre-heat
vice) video chip, pulsed diode backlight, magnifying lens, fine
samples to meet the test temperature requirements for the trial
focus dial, software, and readout system.
specified in 12.3.3. The samples shall be on the heating tray for
7.9 Relative Humidity Sensor Analyzer—The test specimens
at least 20 min to come to thermal equilibrium, which is
in the passive grease sampling device are extruded on the
verified with an embedded thermocouple.
substrate of the die extrusion tester, to create an approximate
11.1.2 Validation Check—A check shall be performed with a
0.25 g thin-film deposition on a plastic substrate. One plastic
spring cell calibration cylinder to confirm calibration prior to
substrate is weighed and placed in a clean, dry glass vial with
the analysis of routine samples. The procedure and accuracy
a septum and sealing cap. The vial is placed into the heating
guidelines for the validation check are described in Annex A3.
chamber of the relative humidity sensor analyzer, and a dry
This procedure shall be performed at the beginning of each
carrier gas is introduced to transfer the moisture driven from
shift or if the instrument has not been used for an extended
the sample by the heating, and passed over the humidity sensor
period of time.
to calculate moisture content based on calculated total flow and
11.2 Preparation of the Optical Spectroscopy Cell—Check
original sample mass.
the mirror, optical path, and the tray for cleanliness.
11.3 This test method requires no additional preparation
8. Reagents and Materials
steps for the other instruments.
8.1 Spring Cell Calibration Cylinder—This device is used
to perform a calibration check on the die extrusion tester prior
12. Testing Procedure
to use. See Annex A3 for the full calibration procedure.
12.1 Procedure for Evaluation of Ferrous Debris Level:
8.2 Indexing Die—Threads onto the end of the passive
12.1.1 Factory Calibration—The parameters for the analyti-
grease sampling device prior to performing the die extrusion
cal instrumentation are pre-set by the manufacturer of the
test.
equipment. Daily validation checks must be performed and
8.3 Substrate—The three-layered plastic substrate shall be
analyses of test specimens must be within the linear range of
long enough to hold the grease ribbon once it is extruded from
response.
the indexing die. The substrate’s three layers consist of a rigid
12.1.2 Routine Standardization—A minimum two-point
bottom layer of high-density polyethylene, a double-sided
routine standardization shall be performed if the instrument
releasable adhesive, and a top layer of low-density polyethyl-
fails the validation check.
ene. The top layer of the substrate is divided into peel-away
12.1.3 Set the instrument to grease mode and drop the
segments that, when fully covered with the grease ribbon, each
sample into the coil.
contains approximately 0.25 g of grease.
12.1.4 Record the results in the nearest whole number in
ppm ferrous content.
9. Sampling
12.2 Analysis of Grease Samples by Die Extrusion:
9.1 Good grease-sampling procedures are critical to good
12.2.1 Factory Calibration—The parameters for the analyti-
analyses, and samples should be taken in accordance with
cal instrumentation are pre-set by the manufacturer of the
Practice D7718.
equipment. A calibration curve must be determined prior to use
to calculate a K constant and an offset for the reference spring
10. Preparation of Test Specimen
cell. Daily validation checks must be performed and results
10.1 Ferrous Debris Analyzer—Prior to analysis, the pas- must be within the linear range of response.
sive grease sampling tool shall be removed from the storage 12.2.2 Routine Standardization—A minimum two-point
container. The passive grease sampling tool shall be placed into routine standardization shall be performed if the instrument
the ferrous debris analyzer to determine the total amount of fails the validation check. The procedure for the routine
ferrous wear present (in ppm). standardization is described in Annex A4.
D7918 − 23
12.3 Steps for Standard and In-service Sample Preparation 12.4.3.1 Place the blank substrate segment into the drawer
Procedure: of the optical spectroscopy cell to collect the background
spectrum.
12.3.1 Load a substrate, with the segments facing up.
12.4.3.2 Attach the surface head to the cell so that it is
12.3.2 Upon preparation of the sample, move the heated
perpendicular to the substrate (or the separator is parallel to the
passive grease sampling device from the heating tray and
side of the cell containing the drawer).
immediately place into the sample-holding bracket with the
12.4.3.3 Turn the i-Lab on using the power button in the
opening of the indexing die facing the same direction that the
center of the unit.
table moves. A new, clean indexing die and substrate shall be
12.4.3.4 Using the arrow keys, highlight GreaseBk and
used for each analysis.
press the power button to select.
12.3.3 The die extrusion tester shall be programmed to only
12.4.3.5 Remove the protective cap from the i-Lab and
perform the test if the sample is in the appropriate temperature
place it onto surface head, which should already be attached to
range. Therefore, if the sample loses too much heat in the
the cell.
loading process, it will need to be reheated. The appropriate
12.4.3.6 Press the Power button again to obtain the back-
temperature is verified using a non-contact thermometer that
ground spectrum.
has been modeled and calibrated by the manufacturers of the
12.4.3.7 Remove the unit from the cell.
instrument to ensure internal grease temperature of 30 °C 6
12.4.4 Procedure for Sample Data Collection:
2 °C.
12.4.4.1 Using the arrow keys, select the GreaseCIE method
12.3.4 The die extrusion tester shall be programmed to run
and press power.
the following steps:
12.4.4.2 Invert the unit back onto the surface head and press
12.3.4.1 The prime must be set so that it allows sufficient
power again to obtain the spectru
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D7918 − 17a D7918 − 23
Standard Test Method for
Measurement of Flow Properties and Evaluation of Wear,
Contaminants, and Oxidative Properties of Lubricating
Grease by Die Extrusion Method and Preparation
This standard is issued under the fixed designation D7918; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
1.1 This test method covers the determination and evaluation of flow properties, wear levels, contaminants, and oxidative
condition of new and in-service lubricating grease.
1.2 This test method provides guidance on evaluating in-service grease samples, NLGI grades 00 to 3, for wear, consistency,
contamination, and oxidation.
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.3.1 Exception—The exception to this will be where units of references were developed by the developers of the test equipment
and necessary to report the results of the test.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
D217 Test Methods for Cone Penetration of Lubricating Grease
D6300 Practice for Determination of Precision and Bias Data for Use in Test Methods for Petroleum Products, Liquid Fuels, and
Lubricants
D6595 Test Method for Determination of Wear Metals and Contaminants in Used Lubricating Oils or Used Hydraulic Fluids by
Rotating Disc Electrode Atomic Emission Spectrometry
D7527 Test Method for Measurement of Antioxidant Content in Lubricating Greases by Linear Sweep Voltammetry
D7546 Test Method for Determination of Moisture in New and In-Service Lubricating Oils and Additives by Relative Humidity
Sensor
This test method is under the jurisdiction of ASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcommittee
D02.96.07 on Integrated Testers, Instrumentation Techniques for In-Service Lubricants.
Current edition approved Nov. 15, 2017Oct. 1, 2023. Published December 2017November 2023. Originally approved in 2015. Last previous edition approved in 2017 as
D7918 – 17.D7918 – 17a. DOI: 10.1520/D7918-17A.10.1520/D7918-23.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7918 − 23
D7718 Practice for Obtaining In-Service Samples of Lubricating Grease
D7843 Test Method for Measurement of Lubricant Generated Insoluble Color Bodies in In-Service Turbine Oils using
Membrane Patch Colorimetry
E177 Practice for Use of the Terms Precision and Bias in ASTM Test Methods
E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
3. Terminology
3.1 Definitions:
3.1.1 active grease sampling device, n—device designed to take an active sample of a lubricating grease from a bearing, gear, or
drive shaft located in a grease lubricated component. D7718
3.1.1.1 Discussion—
The Grease Thief (trademarked) Type 2 is a tool which meets this description and can be used to obtain an active grease sample
from a lubricated component. A full description and dimensions of this device can be found in Annex A2.
3.1.2 active sampling, v—to use a sampling device to actively gather an in-service lubricating grease sample from a
grease-lubricated component. D7718
3.1.3 calibration, n—the determination of the values of the significant parameters by comparison with values indicated by a set
of reference standards. D6595
3.1.4 consistency, n—of lubricating grease, the degree of resistance to movement under stress. D217
3.1.4.1 Discussion—
The term “consistency” is used somewhat synonymously with “penetration.” Generally, consistency refers to the worked
penetration of a grease.
3.1.5 linear sweep voltammetry, n—a test method designed to monitor the anti-oxidant additive content in lubricating greases.
D7527
3.1.6 lubricating grease, n—a semi-fluid to solid product of dispersion of a thickener in a liquid lubricant. D217
3.1.6.1 Discussion—
The dispersion of the thickener forms a two-phase system and immobilizes the liquid lubricant by surface tension and other
physical forces. Other ingredients are commonly included to impart special properties.
3.1.7 in-service lubricating grease, n—lubricating grease that has been applied as a lubricant to a gear, bearing, or drive screw for
any period of time. D7718
3.1.8 passive grease sampling device, n—a device designed to gather a sample from the equipment by being attached to the grease
reservoir at the purge point. D7718
3.1.8.1 Discussion—
The Grease Thief (trademarked) Type 1 is a tool which meets this description and can be used to obtain a representative grease
sample from a purge path. A full description and dimensions of this device can be found in Annex A1.
3.1.9 passive sampling, v—to use a passive grease sampling device to collect a purged sample of in-service lubricating grease from
a purge path. D7718
3.2 Definitions of Terms Specific to This Standard:
3.2.1 die extrusion index, n—an average of three unique test conditions of the in-service lubricating grease sample expressed as
a percent when compared to the average of three unique test conditions of the unused lubricating grease baseline.
The Grease Thief Type I and Type II is described in US Patent No. 7984661. Interested parties are invited to submit information regarding the identification of an
alternative(s) to this patented item to the ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the responsible technical
committee, which you may attend. The sole source of the Grease Thief Type I and Type II are known to the committee is York Laboratories, LLC located at 410 Kings Mill
Rd., York, PA 17401. If you are aware of alternative suppliers, please provide this information to ASTM International Headquarters. Your comments will receive careful
consideration at a meeting of the responsible technical committee, which you may attend.
D7918 − 23
3.2.2 die extrusion tester, n—a device that measures the consistency of in-service lubricating greases and compares the
measurement to the values obtained from a baseline sample of lubricating grease.
3.2.3 ferrous debris level, n—the total amount of ferrous metal as measured by a Hall-effect sensor in the grease sample
independent of particle size.
3.2.4 full attenuation, n—the loss of intensity in signal strength.
3.2.5 grease colorimetry, n—the science of color measurement by the evaluation of the CIE LAB values of grease in the visible
light region of the electromagnetic spectrum.
3.2.6 grease ribbon, n—grease that is extruded onto a substrate.
3.2.7 indexing die, n—a single-use orifice die to measure the consistency as compared to the baseline.
3.2.8 spring cell calibration cylinder, n—a cylinder that houses a spring and a push rod to compress the spring.
3.2.9 substrate, n—the single-use strip onto which a ribbon of grease is extruded.
3.2.10 substrate segment, n—a peel-away section of the substrate containing a portion of the extruded grease; the segments are
pre-cut to hold 0.25 g of extruded grease ribbon.
3.2.11 speed 1, n—the slowest speed in which the grease is extruded from the indexing die.
3.2.12 speed 2, n—the fastest speed in which the grease is extruded from the indexing die.
3.2.13 speed 3, n—the intermediate speed at which the grease is extruded from the indexing die.
4. Summary of Test Method
4.1 Testing of a grease sample includes steps that characterize the ferrous wear within the sample, the consistency of the sample,
and properties related to its chemistry.
4.2 A grease sample of known volume is measured to determine the density of ferrous material in the sample.
4.3 A grease sample held with a defined geometry sample holder is placed into a temperature-controlled instrument and extruded
onto a substrate as a thin ribbon or strip of grease. The extrusion process provides a measurement of the consistency of the grease.
4.4 The test method requires the grease to be tested at three different rates to reflect the non-Newtonian nature of greases. Testing
at several different rates creates a series of step changes that are then compared to an unused baseline grease that has also been
tested under the same conditions.
4.5 While the flow properties are being measured, the grease is simultaneously deposited onto a thin-film substrate containing
substrate segments. Each substrate segment contains approximately 0.25 g of grease. The individual substrate segments are used
for further testing of wear, contamination, and oxidative properties.
4.6 The substrate is removed from the instrument and processed further to obtain information related to the chemistry and content
of the grease sample including linear sweep voltammetry and grease colorimetry.
5. Significance and Use
5.1 Trending the wear, contamination, consistency, and oxidative properties of a lubricating grease is a crucial part of
condition-monitoring programs. Changes in these properties or deviations from the new grease can be indicative of problems
D7918 − 23
within the lubricated component, such as the mixing of incompatible thickener types, excessive wear or contaminant levels, or
significant depletion of antioxidant levels. These test methods also makes it possible to develop trends that can be used to predict
failures before they occur and allow for corrective action to be taken.
6. Interferences
6.1 Particulate Matter—When the in-service grease sample is heavily contaminated with particulate matter, it may be necessary
to perform a dilution of the sample in order to properly extrude the sample onto the substrate. Larger pieces of particulate matter
may also cause spikes in the load profile graph. The data from these spikes should be discounted from the average force
calculations because they are not indicative of changes in the flow properties of the sample.
6.1.1 Heavily contaminated grease samples are considered any grease that contains solid particulate of size or quantity that results
in clogging of the die, which inhibits the completion of the test.
NOTE 1—The specific steps to perform a dilution of grease samples is not addressed in this test method.
6.2 Sample Size—The passive sampling device shall be at least 70 % filled to perform the testing indicated within this test method.
If the sampling device is insufficiently filled, the instrument may not record a force reading over the entirety of the program.
6.3 Color—The grease colorimetry test may become insignificant if the color of the grease causes full attenuation and minimal
transmittance of the signal making it incomparable to the new grease.
7. Apparatus
7.1 Passive Grease Sampling Device—The device that holds the grease sample. The device is typically capable of holding
approximately 1.5 g to 2 g of grease when full. A full description and dimensions of this device can be found in Annex A1.
7.2 Active Grease Sampling Device—The device that holds an active grease sample from a grease-lubricated component. A full
description and dimensions of this device can be found in Annex A2.
7.3 Die Extrusion Tester —Designed for the passive grease sampling device, the die extrusion tester is a temperature-controlled
chamber consisting of a table that is controlled by a linear actuator that holds the substrate, a bracket that holds the passive grease
sampling device, and a second linear actuator to push the handle of the passive grease sampling device. Both the passive grease
sampling device and the second linear actuator are mounted perpendicular to the substrate table. A load cell is mounted between
the second actuator and the handle of the passive grease sampling device to measure the force during the extrusion. The die
extrusion tester is shown in Fig. 1.
FIG. 1 Die Extrusion Tester
Wurzbach, R. and Williams, L., “Die Extrusion Method For Comparing Changes in Grease Consistency and Flow Characteristics,” Viscosity and Rheology of In-Service
Fluids as They Pertain to Condition Monitoring, ASTM STP 1564, 2013, ASTM International.
D7918 − 23
7.4 Constant Temperature Chamber—The test specimens in the passive grease sampling device should be heated in the
temperature-controlled chamber of the die extrusion tester prior to performing any test. The passive grease sampling device shall
have the indexing die firmly engaged on the threaded open end of the device.
7.5 Ferrous Debris Analyzer—An analyzer that allows the passive grease sampling device to be inserted completely in the coil
geometry, designed to measure the total amount of ferrous debris in the grease sample. By use of a Hall-effect sensor, the total
amount of ferrous debris is designed to measure quantitatively in parts per million (ppm). Upon receipt of the sample, it is the first
test performed on the grease.
7.6 Linear Sweep Voltammetry—Linear sweep voltammetry measures the amount of hindered phenols, amines, and zinc dithio
dialkyl phosphate (ZDDP) in a lubricating grease. Using 0.25 g of grease from the substrate, one substrate segment is inserted into
a vial and the antioxidant concentration remaining in the grease is measured and compared to an unused sample of the same grease.
7.7 Grease Colorimetry—The grease colorimetry test provides a spectrum in the 400 nm to 700 nm region of the visible light
spectrum. An optical spectroscopy cell is used and the substrate segment is placed into the same holder. The delta-E values within
the CIE LAB scale are recorded and a spectrum can be generated in a method similar to Test Method D7843.
7.8 Direct Imaging System—A particle sizing and identification component consisting of a CCD (charge-coupled device) video
chip, pulsed diode backlight, magnifying lens, fine focus dial, software, and readout system.
7.9 Relative Humidity Sensor Analyzer—The test specimens in the passive grease sampling device are extruded on the substrate
of the die extrusion tester, to create an approximate 0.25 g thin-film deposition on a plastic substrate. One plastic substrate is
weighed and placed in a clean, dry glass vial with a septum and sealing cap. The vial is placed into the heating chamber of the
relative humidity sensor analyzer, and a dry carrier gas is introduced to transfer the moisture driven from the sample by the heating,
and passed over the humidity sensor to calculate moisture content based on calculated total flow and original sample mass.
8. Reagents and Materials
8.1 Spring Cell Calibration Cylinder—This device is used to perform a calibration check on the die extrusion tester prior to use.
See Annex A3 for the full calibration procedure.
8.2 Indexing Die—Threads onto the end of the passive grease sampling device prior to performing the die extrusion test.
8.3 Substrate—The three-layered plastic substrate shall be long enough to hold the grease ribbon once it is extruded from the
indexing die. The substrate’s three layers consist of a rigid bottom layer of high-density polyethylene, a double-sided releasable
adhesive, and a top layer of low-density polyethylene. The top layer of the substrate is divided into peel-away segments that, when
fully covered with the grease ribbon, each contains approximately 0.25 g of grease.
9. Sampling
9.1 Good grease-sampling procedures are critical to good analyses, and samples should be taken in accordance with Practice
D7718.
10. Preparation of Test Specimen
10.1 Ferrous Debris Analyzer—Prior to analysis, the passive grease sampling tool shall be removed from the storage container.
The passive grease sampling tool shall be placed into the ferrous debris analyzer to determine the total amount of ferrous wear
present (in ppm).
10.2 Passive Grease Sampling Tool—Upon completion of the ferrous debris analyzer test, the protective cap shall be removed and
an indexing die shall be threaded onto the open end of the passive grease sampling device.
NOTE 2—If the in-service grease sample is received in an active grease sampling device, the stinger probe must be removed and replaced with the handle
of a passive grease sampling device prior to use.
D7918 − 23
NOTE 3—If the in-service grease sample is received in any sampling container other than the passive or active grease sampling device, it must be
transferred into a passive grease sampling device by hand, with a clean, single-use glove and/or a clean, single-use 10 mL slip tip syringe.
11. Preparation of Apparatus
11.1 Preparation of the Die Extrusion Tester:
11.1.1 Heating the Chamber—Prior to the performance of any tests, a thermocouple is used to regulate the temperature of the
heating tray to achieve a temperature sufficient to pre-heat samples to meet the test temperature requirements for the trial specified
in 12.3.3. The samples shall be on the heating tray for at least 20 min to come to thermal equilibrium, which is verified with an
embedded thermocouple.
11.1.2 Validation Check—A check shall be performed with a spring cell calibration cylinder to confirm calibration prior to the
analysis of routine samples. The procedure and accuracy guidelines for the validation check are described in Annex A3. This
procedure shall be performed at the beginning of each shift or if the instrument has not been used for an extended period of time.
11.2 Preparation of the Optical Spectroscopy Cell—Check the mirror, optical path, and the tray for cleanliness.
11.3 This test method requires no additional preparation steps for the other instruments.
12. Testing Procedure
12.1 Procedure for Evaluation of Ferrous Debris Level:
12.1.1 Factory Calibration—The parameters for the analytical instrumentation are pre-set by the manufacturer of the equipment.
Daily validation checks must be performed and analyses of test specimens must be within the linear range of response.
12.1.2 Routine Standardization—A minimum two-point routine standardization shall be performed if the instrument fails the
validation check.
12.1.3 Set the instrument to grease mode and drop the sample into the coil.
12.1.4 Record the results in the nearest whole number in ppm ferrous content.
12.2 Analysis of Grease Samples by Die Extrusion:
12.2.1 Factory Calibration—The parameters for the analytical instrumentation are pre-set by the manufacturer of the equipment.
A calibration curve must be determined prior to use to calculate a K constant and an offset for the reference spring cell. Daily
validation checks must be performed and results must be within the linear range of response.
12.2.2 Routine Standardization—A minimum two-point routine standardization shall be performed if the instrument fails the
validation check. The procedure for the routine standardization is described in Annex A4.
12.3 Steps for Standard and In-service Sample Preparation Procedure:
12.3.1 Load a substrate, with the segments facing up.
12.3.2 Upon preparation of the sample, move the heated passive grease sampling device from the heating tray and immediately
place into the sample-holding bracket with the opening of the indexing die facing the same direction that the table moves. A new,
clean indexing die and substrate shall be used for each analysis.
12.3.3 The die extrusion tester shall be programmed to only perform the test if the sample is in the appropriate temperature range.
Therefore, if the sample loses too much heat in the loading process, it will need to be reheated. The appropriate temperature is
verified using a non-contact thermometer that has been modeled and calibrated by the manufacturers of the instrument to ensure
internal grease temperature of 30 °C 6 2 °C.
12.3.4 The die extrusion tester shall be programmed to run the following steps:
D7918 − 23
12.3.4.1 The prime must be set so that it allows sufficient time for the die to be filled. Grease shall be exiting the die before
continuing to Rate 1.
12.3.4.2 Rate 1—0.15 mm/s 6 0.01 mm/s for 19 s 6 1 s.
12.3.4.3 Rate 2—0.73 mm/s 6 0.04 mm/s for 13 s 6 1 s.
12.3.4.4 Rate 3—0.44 mm/s 6 0.02 mm/s for 22 s 6 1 s.
12.3.5 Start the program, which will extrude the lubricating grease onto the substrate and collect the flow response data. This
methodology is good for grades 1, 2, and 3 greases. Consult the vendor manual for how to run grades 00 and 0.
NOTE 4—Die extrusion testing only uses the speeds that show a force response in the final calculation of the die extrusion index. The passive grease
sampling device must be at least 70 % full to run the die extrusion test.
12.3.6 Verify during the prime and prior to Rate 1 step that the load data shows an increase from the start load. Inspect the piston
for movement. Its movement should correspond to an increase in the load cell reading. If no change of the load cell reading is
observed during the prime while the piston is moving, immediately terminate the test by stopping the instrument. If the load cell
displays a zero reading during extrusion, it could be an indication of misalignment or instrument malfunction. Return to 12.3.1 and
repeat the test until a satisfactory validation step is achieved.
12.3.7 Once the test is complete, the data shall be transferred to a computer to be analyzed. For calculations see Section 13.
12.3.8 Remove the empty passive grease sampling device from the die extrusion tester and dispose in the trash.
12.3.9 Remove the substrate containing the grease ribbon for additional testing.
12.4 Procedure for Grease Colorimetry:
12.4.1 General Test Procedure—The test procedure will consist of an unused grease reference reading and the sample (in-service
grease) reading.
12.4.2 Prepare the grease according to 12.3.
12.4.3 Calibration of the Spectrophotometer:
12.4.3.1 Place the blank substrate segment into the drawer of the optical spectroscopy cell to collect the background spectrum.
12.4.3.2 Attach the surface head to the cell so that it is perpendicular to the substrate (or the separator is parallel to the side of
the cell containing the drawer).
12.4.3.3 Turn the i-Lab on using the power button in the center of the unit.
12.4.3.4 Using the arrow keys, highlight GreaseBk and press the power button to select.
12.4.3.5 Remove the protective cap from the i-Lab and place it onto surface head, which should already be attached to the cell.
12.4.3.6 Press the Power button again to obtain the background spectrum.
12.4.3.7 Remove the unit from the cell.
12.4.4 Procedure for Sample Data Collection:
12.4.4.1 Using the arrow keys, select the GreaseCIE method and press power.
12.4.4.2 Invert the unit back onto the surface head and press power again to obtain the spectrum of the sample.
D7918 − 23
12.4.4.3 Repeat test as needed for multiple samples.
12.4.5 Procedure for Linear Sweep Voltammetry:
12.4.5.1 Prepare the grease sample according to 12.3.
12.4.5.2 Add one substrate segment with 0.25 g of grease to the vial and follow the procedure outlined in section 8.5.1 in Test
Method D7527 to perform a linear sweep voltammetry test.
NOTE 5—Step 8.5.1.3 in Test Method D7527 shall be replaced with a single grease strip from the substrate; instead of smearing the grease evenly on the
inside of the 7 mL vial, the user shall take one full strip of the grease substrate and curl it to fit inside the 7 mL vial. User may proceed following the
remainder of the steps in section 8.5.1 in Test Method D7527.
12.5 Analysis of Particle Concentration by Direct Imaging:
12.5.1 For each sample, the grease is extruded in a thin film and collected on a translucent polyethylene substrate as described
in 12.3.
12.5.2 The linear actuator that controls the moving table the plastic substrate sits upon is used to parade the grease ribbon
transversely across the fixed view of the high-resolution camera.
12.5.3 The grease ribbon width shall be equal to or greater than the frame width of the camera, and the length large enough to
capture a sufficient number of unique non-overlapping frames during video analysis.
12.5.4 Prior to analysis, the backlighting is adjusted to provide a consistent average background intensity between samples even
as grease color and opacity may vary.
12.5.5 The focus is also adjusted as necessary for each sample, with a fine vertical position dial.
12.5.6 The camera is used to capture a video recording of the grease ribbon as it passes.
12.5.7 The travel speed of the table is set to maximize the number of frames to be collected during video recording, while
minimizing overall test duration.
12.5.8 The video recording is then processed with particle sizing and counting software to determine the average particle area per
frame in the sample.
12.5.9 Out-of-focus frames, or frames that are otherwise compromised by grease distribution inconsistency, are excluded from the
count.
12.5.10 The results in μm are correlated to m
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...