Standard Practice for Preparing an Equilibrium Liquid/Vapor Sample of Live Crude Oil, Condensates, or Liquid Petroleum Products Using a Manual Piston Cylinder for Subsequent Liquid Analysis or Gas Analysis

SIGNIFICANCE AND USE
5.1 This practice allows for compositional analysis of the gases in equilibrium with crude oil, condensate, and liquid petroleum products at a 4:1 vapor/liquid ratio at ambient temperature for analysis using typical instrumentation (RGA) already available in typical refinery laboratories. These highly volatile components can result in vapor pressure conditions above atmospheric pressure, so this mechanically simple system is easily adaptable to day-to-day application at low cost/effort using existing analytical equipment.  
5.2 This practice allows for compositional analysis and day-to-day tracking or trending of the light hydrocarbons in crude oil for the purpose of identifying unusual blending of NGL, LPG, butane etc. into individual crude oil batch receipts.  
5.3 This practice allows identification of gases: including: CO, CO2, H2, H2S, N2, O2, CH4, C2H6, C3H8, etc. that can contribute to vapor pressure by Test Method D6377, but are not identified using Test Method D8003 (see Note 1). These components can originate from production or can be the result of the use of pad gas and may not be native to the original product. Significant difference in Test Method D6377 vapor pressure measurements at low V/L (for example, 0.1:1) versus high V/L (for example, 4:1) indicate the contribution of high vapor pressure gases such as those in 5.2.
Note 1: Test Method D8003 does identify: CH4, C2H6, and C3H8. Test Method D8003 does not identify: CO, CO2, H2, H2S, N2, and O2.  
5.4 Nitrogen and combustion gases (mostly nitrogen and CO2 with minor concentrations of air) at positive pressures up to 2500 mm water column (nominal 4 psig) is required by International Marine Organization (IMO) Marine Pollution (MARPOL) and Safety of Life at Sea (SOLAS) regulations for the marine transport of crude oil. Analysis of the equilibrium vapor may be required to determine the contribution of inert gases to the total vapor pressure of the crude oil on receipt at the discharge port or refine...
SCOPE
1.1 This practice covers the preparation of an equilibrium gas sample of live crude oil, condensate, or liquid petroleum products, using a Practice D8009 manual piston cylinder (MPC) as a vapor tight expansion chamber to generate an equilibrium vapor/liquid pair at a known temperature and vapor/liquid ratio (V/L). Inert gas such as helium or argon is injected to the equilibrium vapor space of the MPC to provide an equilibrium vapor sample sufficiently above atmospheric pressure for subsequent analysis using a standard refinery gas analyzer (RGA) such as described in Test Method D7833. Other gas analysis methods may be used provided they meet the minimum performance criteria stated in 7.4.1.  
1.2 This practice is suitable for UN Class 3 Liquid samples having vapor pressures between 0 kPa and 300 kPa at 50.0 °C, and 0.1:1 to 4:1 vapor/liquid ratio, spanning the nominal range near bubble point (Test Method D6377 VPCr,0.1) to Test Methods D323 (RVP), D4953, and D5191 (V/L=4). The temperature may vary over a wide range, provided that the cylinder is maintained at isothermal and isobaric conditions to prevent condensation of equilibrium vapor upon cooling either in the cylinder or in the injection system of the Refinery Gas Analyzer (RGA, Test Method D7833). The method is best suited for preparation of an equilibrium gas/liquid pair near ambient conditions, typical of routine daily operations in a typical refinery quality assurance or marine terminal laboratory, to routinely monitor the light ends content of crude oil receipts.  
1.3 This practice is suitable to prepare an equilibrium liquid/vapor sample pair in a sealed sampling system (no light ends loss from either phase). The equilibrium gas phase is suitable for subsequent gas analysis of both hydrocarbon and fixed/inert gases in the sample, including: hydrogen, oxygen, nitrogen, carbon dioxide, carbon monoxide, hydrogen sulfide, C1 to C7 hydrocarbons at levels consist...

General Information

Status
Published
Publication Date
14-Dec-2018
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D8236-18 - Standard Practice for Preparing an Equilibrium Liquid/Vapor Sample of Live Crude Oil, Condensates, or Liquid Petroleum Products Using a Manual Piston Cylinder for Subsequent Liquid Analysis or Gas Analysis
English language
10 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D8236 − 18
Standard Practice for
Preparing an Equilibrium Liquid/Vapor Sample of Live Crude
Oil, Condensates, or Liquid Petroleum Products Using a
Manual Piston Cylinder for Subsequent Liquid Analysis or
1
Gas Analysis
This standard is issued under the fixed designation D8236; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope nitrogen, carbon dioxide, carbon monoxide, hydrogen sulfide,
C1 to C7 hydrocarbons at levels consistent with the Test
1.1 This practice covers the preparation of an equilibrium
Method D7833 method used. The equilibrium liquid phase can
gas sample of live crude oil, condensate, or liquid petroleum
be subsequently analyzed by Test Method D8003 to obtain
products, using a Practice D8009 manual piston cylinder
paired analytical results on both the equilibrium liquid and
(MPC) as a vapor tight expansion chamber to generate an
vapor pair with a sealed sample system.
equilibrium vapor/liquid pair at a known temperature and
vapor/liquid ratio (V/L). Inert gas such as helium or argon is
1.4 Addition of the diluent gas provides a positive pressure
injected to the equilibrium vapor space of the MPC to provide
sample to allow the use of a typical RGA-type gas injection
an equilibrium vapor sample sufficiently above atmospheric
system that operates only slightly above barometric pressure.
pressure for subsequent analysis using a standard refinery gas
The preferred diluent gas shall be the same as the carrier gas
analyzer (RGA) such as described in Test Method D7833.
used in the RGA(typically helium or argon). Choice of diluent
Other gas analysis methods may be used provided they meet
or carrier gas may affect the ability to detect some inert gases
the minimum performance criteria stated in 7.4.1.
(especially O or H ) in some RGAconfigurations conforming
2 2
1.2 This practice is suitable for UN Class 3 Liquid samples
to Test Method D7833.
having vapor pressures between 0 kPa and 300 kPa at 50.0 °C,
1.5 The VLE gas generation and subsequent RGAoutput is
and 0.1:1 to 4:1 vapor/liquid ratio, spanning the nominal range
usedasascreeningmethodtoidentifygascomponentsthatcan
near bubble point (Test Method D6377 VPCr,0.1) to Test
be present in the crude oil affecting the total vapor pressure.
Methods D323 (RVP), D4953, and D5191 (V/L=4). The
The RGA output only represents the equilibrium vapor com-
temperature may vary over a wide range, provided that the
ponents present and relative to one another. Due to dilution of
cylinder is maintained at isothermal and isobaric conditions to
the VLE gas with inert gas, the RGA output does not purport
prevent condensation of equilibrium vapor upon cooling either
to accurately provide the actual vapor composition at VLE
in the cylinder or in the injection system of the Refinery Gas
Analyzer (RGA, Test Method D7833). The method is best conditions and is definitely not representative of the composi-
tion of the whole sample.
suited for preparation of an equilibrium gas/liquid pair near
ambient conditions, typical of routine daily operations in a
1.6 The values stated in SI units are to be regarded as
typical refinery quality assurance or marine terminal
standard. The values given in parentheses after SI units are
laboratory, to routinely monitor the light ends content of crude
provided for information only and are not considered standard.
oil receipts.
1.7 This standard does not purport to address all of the
1.3 This practice is suitable to prepare an equilibrium
safety concerns, if any, associated with its use. It is the
liquid/vapor sample pair in a sealed sampling system (no light
responsibility of the user of this standard to establish appro-
ends loss from either phase). The equilibrium gas phase is
priate safety, health, and environmental practices and deter-
suitable for subsequent gas analysis of both hydrocarbon and
mine the applicability of regulatory limitations prior to use.
fixed/inert gases in the sample, including: hydrogen, oxygen,
1.8 This international standard was developed in accor-
dance with internationally recognized principles on standard-
1
This practice is under the jurisdiction ofASTM Committee D02 on Petroleum
ization established in the Decision on Principles for the
Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-
Development of International Standards, Guides and Recom-
mittee D02.08 on Volatility.
mendations issued by the World Trade
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.