This document specifies a test method for determining short-term deformation of a press-fit acetabular component for total hip joint replacement under specific laboratory conditions. This document also defines the conditions of testing so that the important parameters that affect the components are taken into account and it describes how the specimen is set up for testing. The described method is intended to be used to measure the amount of deformation under load and plastic deformation after unloading of various designs and materials used for acetabular components in total hip joint replacement. These measurements are then used in an evaluation of risks associated with acetabular cup deformation for the acetabular component under evaluation to determine if its performance can be adversely affected. In the evaluation of risks associated with acetabular component deformation, it can be useful to take into consideration various design, material and surgical implantation factors (e.g. those identified in REF Section_sec_8 \r \h Clause 8 08D0C9EA79F9BACE118C8200AA004BA90B02000000080000000E000000530065006300740069006F006E005F007300650063005F0038000000 ), and, if necessary, a comparison of results to a reference implant tested under the same conditions. Depending on this evaluation, either additional testing or clinical data, or both, which is outside the scope of this document, can be necessary. The loading of the acetabular components in vivo differs, in general, from the loading defined in this test method. The results obtained here cannot be used to directly predict in vivo performance. This document does not cover methods that examine the test specimens.

  • Standard
    10 pages
    English language
    sale 15% off

This document specifies surface finish requirements for the articulating surfaces of total and partial knee joint prostheses classified in ISO 7207-1. This document is intended to provide guidance for periodic validation of production processes.

  • Standard
    3 pages
    English language
    sale 15% off
  • Standard
    3 pages
    French language
    sale 15% off
  • Standard
    1 page
    English language
    sale 15% off
  • Standard
    1 page
    French language
    sale 15% off
  • Standard
    1 page
    French language
    sale 15% off

This document specifies methods of determining the compressive (fracture) or the tension (disassembly) loads required, under specific laboratory conditions, to cause failure of a modular head system. This document applies to components made of metallic and non-metallic materials, such as femoral heads of partial or total hip-joint replacements of modular construction (i.e. a head/neck conical taper connection). This document excludes methods of examining and reporting the test specimens.

  • Standard
    11 pages
    English language
    sale 15% off
  • Standard
    11 pages
    French language
    sale 15% off

ISO 7206-13:2016 describes a method of determining the torque required, under specified laboratory conditions, to loosen the fixation of the head of hip joint prostheses in which the head is not intended to rotate relative to the neck. It applies to the femoral component of total or partial hip joint replacements in which the head and neck/stem (in the following referred to as cone) are secured together by a locking conical taper or any other means and in which the head and cone are separate components, and which are made of metallic or non-metallic materials. It does not cover methods of examining the test specimens; these should be agreed between the test laboratory and the party submitting the specimen for test.

  • Standard
    6 pages
    English language
    sale 15% off

This document specifies a test method for determining the endurance properties, under specified laboratory conditions, of tibial trays used in knee-joint prostheses to support and secure the plastic articulating surface. It applies to tibial trays which cover both the medial and lateral plateaux of the tibia. The test method does not apply to tibial components manufactured solely from plastic materials. This document does not cover methods of examining and reporting the final condition of the test specimen; these can be the subject of agreement between the test laboratory and the parties submitting the specimen for test. NOTE Correlation of test results with in vivo performance has not been established.

  • Standard
    9 pages
    English language
    sale 15% off
  • Standard
    10 pages
    French language
    sale 15% off

ISO 7206-12:2016 specifies a test method for determining short-term deformation of a press-fit acetabular component for total hip joint replacement under specific laboratory conditions. It also defines the conditions of testing so that the important parameters that affect the components are taken into account and it describes how the specimen is set up for testing. Furthermore, this part of ISO 7206 specifies the test parameters of press-fit acetabular components tested in accordance with this part of ISO 7206. The described method is intended to be used to evaluate the comparison of various designs and materials used for acetabular components in total hip joint replacement when tested under similar conditions. The loading of the acetabular components in vivo will, in general, differ from the loading defined in this test method. The results obtained here cannot be used to directly predict in vivo performance. ISO 7206-12:2016 does not cover methods of examining the test specimen.

  • Standard
    8 pages
    English language
    sale 15% off
  • Standard
    8 pages
    French language
    sale 15% off
  • Standard
    3 pages
    English language
    sale 15% off
  • Standard
    3 pages
    French language
    sale 15% off
  • Standard
    3 pages
    French language
    sale 15% off

This standard defines a test method and laboratory conditions so that the important parameters that affect the components are taken into account and describes how the specimen is set up for testing. The test method has been developed for prostheses that have a plane of symmetry and may not be suitable for other designs. The standard does not cover methods of examining and reporting of the test specimen. The informative Annex A includes a guidance on composition of high-alumina cement embedding medium.

  • Standard
    7 pages
    English language
    sale 15% off
  • Standard
    7 pages
    French language
    sale 15% off
  • Standard
    7 pages
    French language
    sale 15% off

Frequently Asked Questions

ISO/TC 150/SC 4/WG 1 is a Subcommittee within the International Organization for Standardization (ISO). It is named "Mechanical testing". This committee has published 13 standards.

ISO/TC 150/SC 4/WG 1 develops ISO standards. Currently, there are 13 published standards from this subcommittee.

The International Organization for Standardization (ISO) is an independent, non-governmental international organization that develops and publishes international standards. Founded in 1947 and headquartered in Geneva, Switzerland, ISO brings together experts from 170+ member countries to share knowledge and develop voluntary, consensus-based standards that support innovation and provide solutions to global challenges.

A Subcommittee (SC) in ISO operates under a Technical Committee and focuses on a specific subset of the TC's scope. Subcommittees develop standards and technical specifications in their specialized area, reporting to their parent Technical Committee. They may also have working groups for detailed technical work.